Campbell CS455 Product manual

Add to my manuals
34 Pages

advertisement

Campbell CS455 Product manual | Manualzz

CS450/CS455 Pressure

Transducer

Revision: 1/12

C o p y r i g h t © 2 0 0 9 - 2 0 1 2

C a m p b e l l S c i e n t i f i c , I n c .

Warranty

“PRODUCTS MANUFACTURED BY CAMPBELL SCIENTIFIC, INC. are warranted by Campbell Scientific, Inc. (“Campbell”) to be free from defects in materials and workmanship under normal use and service for twelve (12) months from date of shipment unless otherwise specified in the corresponding

Campbell pricelist or product manual. Products not manufactured, but that are re-sold by Campbell, are warranted only to the limits extended by the original manufacturer. Batteries, fine-wire thermocouples, desiccant, and other consumables have no warranty. Campbell's obligation under this warranty is limited to repairing or replacing (at Campbell's option) defective products, which shall be the sole and exclusive remedy under this warranty. The customer shall assume all costs of removing, reinstalling, and shipping defective products to Campbell. Campbell will return such products by surface carrier prepaid within the continental United States of America. To all other locations, Campbell will return such products best way CIP (Port of Entry)

INCOTERM® 2010, prepaid. This warranty shall not apply to any products which have been subjected to modification, misuse, neglect, improper service, accidents of nature, or shipping damage. This warranty is in lieu of all other warranties, expressed or implied. The warranty for installation services performed by Campbell such as programming to customer specifications, electrical connections to products manufactured by Campbell, and product specific training, is part of Campbell’s product warranty. CAMPBELL

EXPRESSLY DISCLAIMS AND EXCLUDES ANY IMPLIED

WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A

PARTICULAR PURPOSE. Campbell is not liable for any special, indirect, incidental, and/or consequential damages.”

Assistance

Products may not be returned without prior authorization. The following contact information is for US and international customers residing in countries served by Campbell Scientific, Inc. directly. Affiliate companies handle repairs for customers within their territories. Please visit www.campbellsci.com

to determine which Campbell Scientific company serves your country.

To obtain a Returned Materials Authorization (RMA), contact CAMPBELL

SCIENTIFIC, INC., phone (435) 227-2342. After an applications engineer determines the nature of the problem, an RMA number will be issued. Please write this number clearly on the outside of the shipping container. Campbell

Scientific's shipping address is:

CAMPBELL SCIENTIFIC, INC.

RMA#_____

815 West 1800 North

Logan, Utah 84321-1784

For all returns, the customer must fill out a "Statement of Product Cleanliness and Decontamination" form and comply with the requirements specified in it.

The form is available from our web site at www.campbellsci.com/repair . A completed form must be either emailed to [email protected]

or faxed to

(435) 227-9579. Campbell Scientific is unable to process any returns until we receive this form. If the form is not received within three days of product receipt or is incomplete, the product will be returned to the customer at the customer's expense. Campbell Scientific reserves the right to refuse service on products that were exposed to contaminants that may cause health or safety concerns for our employees.

CS450/CS455 Table of Contents

PDF viewers: These page numbers refer to the printed version of this document. Use the

PDF reader bookmarks tab for links to specific sections.

1. Introduction..................................................................1

1.1 Initial Inspection and Handling Guidelines ..............................................1

2. Specifications ..............................................................2

3. Configuration ...............................................................3

3.1 Connection................................................................................................4

3.1.1 Using the A200 ...............................................................................4

3.1.1.1 Driver Installation .................................................................4

3.1.1.2 Wiring ...................................................................................4

3.1.1.3 Powering the Sensor..............................................................4

3.1.1.4 Determining which COM Port the A200 has been

Assigned.............................................................................4

3.1.2 Device Configuration Utility (version 1.12 or higher) ...................6

3.1.3 SDI-12 Transparent Mode ..............................................................8

3.1.4 RS-232 Connection via PC Terminal Software ..............................9

3.2 SDI-12 Commands .................................................................................11

3.2.1 SDI-12 Addresses .........................................................................11

3.2.2 Extended SDI-12 Commands........................................................12

3.2.2.1 Units ....................................................................................12

3.2.2.2 Sample Number...................................................................13

4. Installation..................................................................13

4.1 Vent Tubes..............................................................................................13

4.2 Dislodging Bubbles ................................................................................13

4.3 Water Depth............................................................................................13

4.4 Well Installations....................................................................................14

4.5 Other Installations...................................................................................14

4.6 Calculate Offset ......................................................................................14

5. Wiring..........................................................................15

5.1 SDI-12 ....................................................................................................15

5.2 RS-232....................................................................................................16

6. Programming .............................................................16

6.1 Using SCWIN.........................................................................................16

6.2 Using CRBasic or Edlog.........................................................................16

6.2.1 CRBasic Programming .................................................................16

6.2.2 Edlog Programming ......................................................................18 i

CS450/CS455 Table of Contents

7. Maintenance ...............................................................19

7.1 Every Visit ............................................................................................. 20

7.2 Every Two to Three Years or on a Rotating Schedule........................... 20

8. Troubleshooting.........................................................20

Appendix

A. Calibration Certificate ............................................. A-1

Figures

3-1. A200 Sensor-to-PC Interface.................................................................. 5

3-2. Device Configuration Utility .................................................................. 6

3-3. Connect screen........................................................................................ 7

3-4. Configure units ....................................................................................... 7

3-5. Terminal Emulator.................................................................................. 8

3-6. HyperTerminal...................................................................................... 10

A-1. Calibration Certificate........................................................................ A-2

Tables

3-1. Factory Settings ...................................................................................... 3

3-2. A200 Wiring........................................................................................... 5

3-3. RS-232 Settings ...................................................................................... 9

3-4. RS-232 Terminal Commands ............................................................... 10

3-5. SDI-12 Commands ............................................................................... 11

5-1. SDI-12 Wiring ...................................................................................... 15

5-2. RS-232 Wiring...................................................................................... 16 ii

CS450/CS455 Pressure Transducer

The CS450/CS455 Submersible Pressure Transducer probe is designed for general pressure measurements. It utilizes the SDI-12 or RS-232 communications protocols to communicate with a SDI-12 or RS-232 recorder simplifying installation and programming.

1. Introduction

The CS450/CS455 Submersible Pressure Transducer is designed to provide a reliable, accurate pressure/level measurement that is fully temperature compensated.

Combining the sensor, analog measurement and signal processing within a single housing results in the integration of state-of-the-art sensor and measurement technology. The 24 Bit A/D has simultaneous 50/60 Hz rejection and automatic calibration for each measurement. A number of additional advanced measurement techniques are employed to harness the best possible performance available from today’s state-of-the-art pressure sensor technology. The

Transducer reverts to a low power sleep state between measurements. A series of measurements are performed yielding a temperature and pressure value. This measurement cycle takes less than 1.5 second. The sensor can also be configured to output pressure only in less than 1 sec. The measurement cycle is activated by commands via SDI-12 or RS-232 Terminal commands.

The design utilizes a piezoresistive sensor housed in a 316L Stainless Steel

(CS450) or Titanium (CS455) package to enhance reliability. The rugged construction makes the CS450/CS455 suitable for water level measurement in irrigation applications, water wells, lakes, streams and tanks. The Titanium package of the CS455 makes it ideal for salt water or other harsh environments.

The cable incorporates a vent tube to compensate for atmospheric pressure fluctuations and the jacket is made of rugged Hytrel, designed to remain flexible and tough, even under harsh environmental conditions.

The CS450/CS455 has two communication options: SDI-12 or RS-232. The

CS450/CS455 is shipped from the factory with both communications options enabled; there is no configuration required. As an SDI-12 sensor, the

CS450/CS455 is shipped with an address of 0.

Two values are output by the CS450/CS455, pressure or level and temperature.

The CS450/CS455 is shipped from the factory to output pressure in psig and temperature in degrees Celsius.

1.1 Initial Inspection and Handling Guidelines

Upon receipt of the CS450/CS455, inspect the packaging for any signs of shipping damage, and, if found, report the damage to the carrier in accordance with policy. The contents of the package should also be inspected and a claim filed if any shipping related damage is discovered.

Care should be taken when opening the package not to damage or cut the cable jacket. If there is any question about damage having been caused to the cable jacket, a thorough inspection is prudent.

1

CS450/CS455 Pressure Transducer

CAUTION

The model number and pressure range is etched on the housing. Check this information against the shipping documentation to ensure that the expected model number and range were received.

Gauge pressure (Vented) devices must always have a desiccant tube attached.

New desiccant is blue in color. As the desiccant material absorbs water vapor, it begins to turn pink, and eventually white when fully expired. Desiccant tubes should be inspected regularly (i.e. every two months) and historical information will help to baseline desiccant replacement requirements for a given application/climate.

The desiccant tube is shipped with a black cap to cover the vent hole. This cap MUST be removed prior to installation.

Remember that although the CS450/CS455 is designed to be a rugged and reliable device for field use, it is also a highly precise scientific instrument and should be handled as such. There are no user serviceable parts and any attempt to disassemble the device will void the warranty.

2. Specifications

Power Requirements:

Power Consumption:

Measurement Time:

6 to 18 volts D.C.

Quiescent current < 80 µA

Measurement/Communication Current: 8 mA for 1 second measurement

Maximum Peak Current: 40 mA

Outputs:

Measurement Ranges:

Pressure (PSIG)

0 to 2.9

0 to 7.25

0 to 14.5

0 to 29

0 to 72.5

0 to 145

Accuracy:

Resolution:

Overpressure:

Less than 1.5 seconds

SDI-12 (version 1.3) 1200 Baud

RS-232 9600 Baud

Pressure (kPa)

0 to 20

0 to 50

0 to 100

0 to 200

0 to 500

0 to 1000

±0.1% FS TEB* or

±0.05% FS TEB**

0.0035% FS

2x pressure range

Operating Temperature: -10 to 80°C ft of fresh water

0 to 6.7

0 to 16.7

0 to 33.4

0 to 67

0 to 167

0 to 334.5

2

CS450/CS455 Pressure Transducer

Compensated Temperature: 0 to 60°C

Temperature Accuracy: ±0.2°C

Maximum Cable Length: SDI-12 - one sensor connected to a single port can communicate up to approximately 1500 ft. (457 m)

10 sensors connected to a single port, each sensor can have up to 200 ft. (60 m)

RS-232 - 50 ft. (15 m)

Cable Type: 5 Conductor, 26 AWG

Body Material:

Element Material:

CS450 – 316L Stainless Steel

CS455 – Titanium

CS450 – 316L Stainless Steel

CS455 – Hastelloy

Top Cone Material: Delrin

Dimensions: Length 8.4 in (213.36 mm)

Diameter 0.84 in (21.34 mm)

Distance from pressure sensor interface

(black line etched on housing) to:

End of NPT fitting: 1.13 in. (2.86 cm)

End of standard nose cone: 1.58 in (4 cm)

End of weighted nose cone: 4.56 in (11.59 cm)

Weight: CS450: 0.40 lbs (0.18 kg)

CS455: 0.25 lbs (0.11 kg)

Cable: 0.283 lbs/ft (0.421 kg/m)

* Total ERROR Band (TEB) includes the combined errors due to nonlinearity, hysteresis, nonrepeatability, and thermal effects over the compensated temperature range, per ISA S51.1.

** 0.05% FS accuracy not available in the 0 – 2.9 psig range.

3. Configuration

The CS450/CS455 is configured at the factory with default settings:

TABLE 3-1. Factory Settings

SDI-12 Address

Temperature Units

0

RS-232 Baud Rate 9600

Pressure/Level Units PSIG

Celsius

3

CS450/CS455 Pressure Transducer

3.1 Connection

Communicating with the CS450/CS455 requires the sensor to be either connected to a PC or to an SDI-12 Recorder. The sensor typically connects to a

PC via the A200 sensor to PC interface. Many SDI-12 Recorders allow communication to the sensor via a terminal screen. Configurable settings can be changed via SDI-12 commands or by using Campbell Scientific's software

Device Configuration Utility.

3.1.1 Using the A200

The A200 or another device is required to connect the CS450/CS455 to a PC.

This allows sensor settings to be changed via our Device Configuration Utility.

3.1.1.1 Driver Installation

If the A200 has not been previously plugged into your PC and your PC operating system is not Windows 7, the A200 driver needs to be loaded onto your PC.

NOTE Drivers should be loaded before plugging the A200 into the PC.

The A200 drivers can be downloaded, at no charge, from: www.campbellsci.com/downloads .

3.1.1.2 Wiring

One end of the A200 has a terminal block while the other end has a type B female USB port. The terminal block provides 12V, G, TX, and RX terminals for connecting the sensor (see Figure 3-1 and Table 3-2).

A data cable, CSI part number 17648, ships with the A200. This cable has a

USB type-A male connector that attaches to a PC’s USB port, and a type B male connector that attaches to the A200’s USB port.

3.1.1.3 Powering the Sensor

The A200 provides power to the sensor when it is connected to a PC’s USB port. An internal DC/DC converter boosts the 5 Vdc supply from the USB connection to a 12 Vdc output that is required to power the sensor.

3.1.1.4 Determining which COM Port the A200 has been Assigned

When the A200 driver is loaded, the A200 is assigned a COM port number.

This COM port number is needed when using the Device Configuration Utility or a PC terminal software such as HyperTerminal.

Often, the assigned COM port will be the next port number that is free.

However, if other devices have been installed in the past (some of which may no longer be plugged in), the A200 may be assigned a higher COM port number.

4

CS450/CS455 Pressure Transducer

To check which COM port has been assigned to the A200, you can monitor the appearance of a new COM port in the list of COM ports offered in your software package (e.g., LoggerNet) before and after the installation, or look in the Windows Device Manager list under the ports section (access via the control panel).

FIGURE 3-1. A200 Sensor-to-PC Interface

TABLE 3-2. A200 Wiring

Color

Sensor

Cable Label

Red 12V

Black G

White C

Blue G

Yellow G

A200

Terminal

+12Vdc

G

Tx

Rx

G

5

CS450/CS455 Pressure Transducer

3.1.2 Device Configuration Utility (version 1.12 or higher)

The Device Configuration Utility (DevConfig) allows you to change the settings of the CS450/CS455. DevConfig is shipped on the CSI ResourceDVD included with the CS450/CS455.

To use DevConfig, the transducer needs to be connected to the PC via the

A200 (see Section 3.1.1). After installing DevConfig and connecting the transducer to the PC, select CS450 from the Device Type list (see Figure 3-2).

FIGURE 3-2. Device Configuration Utility

In the PC Serial Port box, select the COM port that was assigned to the A200

(see Section 3.1.1.4). Click on the Connect button to enable communication with the sensor. Once successfully connected, the DevConfig screen should look like Figure 3-3.

6

CS450/CS455 Pressure Transducer

FIGURE 3-3. Connect screen

There are three settings that can be changed: SDI-12 address, Pressure/Level

Units and Temperature Units. Double-click on the window of the units to be changed. This will open a pick menu box. Select the desired units and Apply the changes (see Figure 3-4). kPa

FIGURE 3-4. Configure units

7

CS450/CS455 Pressure Transducer

3.1.3 SDI-12 Transparent Mode

Transparent Mode allows direct communication with the CS450/CS455. This may require waiting for programmed datalogger commands to finish before sending responses. While in the transparent mode, datalogger programs may not execute. Datalogger security may need to be unlocked before transparent mode can be activated.

Transparent mode is entered while the PC is in telecommunications with the datalogger through a terminal emulator program. It is most easily accessed through Campbell Scientific datalogger support software, but is also accessible with terminal emulator programs such as Windows HyperTerminal.

To enter the SDI-12 transparent mode, enter Terminal Emulator from

LoggerNet, PC400 or PC200W datalogger support software. A terminal emulator screen is displayed. Click the “Open Terminal” button as shown in

Figure 3-5.

For CR800 series, CR1000, CR3000 loggers, press <Enter> until the datalogger responds with the prompt (“CR1000>” for the CR1000). Type

“SDI12” at the prompt (without quotes) and press <Enter>. In response, the query “Enter Cx Port 1,3,5 or 7” will appear. Enter the control port integer to which the CR450 is connected. An “Entering SDI12 Terminal” response indicates that SDI-12 Transparent Mode is active.

8

FIGURE 3-5. Terminal Emulator

For CR10X and CR510 standard mixed-array loggers: The datalogger telecommunications command to enter SDI-12 transparent mode is

“nX<Enter>” where n is the control port being used for SDI-12. In this example the selected control port is C1, so the command would be

“1X<Enter>”. In response, the datalogger opens the link to control port 1 and

CS450/CS455 Pressure Transducer responds with a prompt. CR10X and CR510 dataloggers reply with “entering

SDI-12”.

CR10X-PB and CR510-PB dataloggers require “*#n” being entered at the prompt, where n is the control port being used.

Now check for response from the sensor with address zero by typing the SDI-

12 Identify command “0I!<Enter>” (that’s a zero, not the letter O). The sensor should respond with an identification string similar to

“013CSI450.Std.01_xxxxxxxx”, where xxxxxxxx represents the eight digit serial number. Note that the SDI-12 standard allows for multiple probes to be connected to one datalogger control port. For example, if you have another

SDI-12 probe on the C1 that has address 7, you could issue the identify command “7I!<Enter>”.

Only one sensor of the same address can be connected when using the change address command.

3.1.4 RS-232 Connection via PC Terminal Software

PC terminal software can be used to communicate with the CS450/CS455 via the RS-232 communication mode. For this example HyperTerminal will be used (see Figure 3-6).

The CS450/CS455 is connected to the PC via the A200 (see Section 3.1.1).

Upon setup, HyperTerminal will request you enter the Communication connection; defaults to a phone connection. Change the communication to appropriate “Com” in the “Connect Using” box (see Section 3.1.1.4 to determine the COM port that was assigned to the A200). HyperTerminal will then prompt for the proper “Port Settings”. These should be:

TABLE 3-3. RS-232 Settings

Bits per Second

Data bits

9600

8

Parity None

Stop bits

Flow control

1

None

From the HyperTerminal screen, you will now be able to communicate with the CS450/CS455. At the prompt, push the ENTER key several times. This will wake-up the RS-232 mode of the sensor. By default the CS450/CS455 is in the SDI-12 mode for communication. Once in the RS-232 mode, if there is no communication for 20 seconds, the sensor will return to the SDI-12 mode.

9

CS450/CS455 Pressure Transducer

FIGURE 3-6. HyperTerminal

TABLE 3-4. RS-232 Terminal Commands

Terminal Commands Values Returned

1 Serial Number, Pressure/Level, Temperature

(in configured units)

2

3

5

H or h

Serial Number, Pressure (kPa), Temperature

(°C)

Serial Number, ΔR(ohms), Rb(ohms),

Temperature (°C ), Element Serial Number,

Product Name

Copyright information, OS Version and Date,

Serial Number, Element Serial Number,

Product Name, User Defined Name (Station

Name), SDI-12 Address

Help menu

10

CS450/CS455 Pressure Transducer

3.2 SDI-12 Commands

The CS450/CS455 uses an SDI-12 compatible hardware interface and supports a subset of the SDI-12 commands. The most commonly used command is the aM! command, issued by the datalogger. Here, ‘a’ represents the sensor address. The communication sequence begins with the datalogger waking the sensor and issuing the aM! command. The transducer responds to the logger indicating that two measurements will be ready within two (2) seconds.

Subsequent communications handle data reporting from the sensor to the datalogger.

The SDI-12 protocol has the ability to support various measurement commands. The CS450/CS455 supports the commands that are listed in the following table.

TABLE 3-5. SDI-12 Commands

SDI-12 Command Command Function Values Returned settings

Temperature aM1! aM2! aM3! aM4!

PSIG, °C

PSIG, °F kPa, °C kPa, °F

Pressure, Temperature

Pressure, Temperature

Pressure, Temperature

Pressure, Temperature aM6! Ohms, ohms, °C

(provides data in less than

0.8 seconds)

ΔR, Rb, Temperature

Pressure/Level

(provides average of data based on user selected samples)

Pressure/Level,

Temperature

As the measurement data is transferred between the probe and datalogger digitally, there are no offset errors incurred with increasing cable length as seen with analog sensors. However, with increasing cable length there is still a point when the digital communications will break down, resulting in either no response or excessive SDI-12 retries and incorrect data due to noise problems.

(Using SDI-12 commands like aMC!, which adds a CRC check, can significantly improve incorrect data issues.)

3.2.1 SDI-12 Addresses

Multiple sensors can be connected to a single digital I/O channel (control port) on an SDI-12 compatible datalogger; each sensor must have a unique SDI-12 address. Possible addresses are 0 to 9, A to Z, and a to z.

11

CS450/CS455 Pressure Transducer

The CS450/CS455 is shipped from the factory with the address set to 0. The address on the CS450/CS455 can be changed by sending an SDI-12 change address command ‘aAb!’, where a is the original address and b is the new address. The change address command can be issued from most SDI-12 recorders.

To change the address of a sensor that has a default address of 0 to the address of 1 the following command can be sent:

“0A1!”

When it is necessary to measure more than one CS450/CS455, it is easiest to use a different control port for each CS450/CS455 instead of changing the address. If additional control ports are not available, then the address will need to be changed.

3.2.2 Extended SDI-12 Commands

Extended SDI-12 commands can be used to configure the CS450/CS455 data output and sample number. These commands are used to select the temperature units: Celsius or Fahrenheit and pressure/level units: psig, kPa, bar, feet, meter, inches, millimeter, and the integration time for each measurement. If level units are selected, they will represent level of fresh water. The Multiplier

(slope) and offset should be used to correct for relative density of water.

Sample number represents the number of values used to provide the output value received by the datalogger. This output value is an average of the samples.

NOTE Extended commands were introduced in Version 4. Versions previous to this did not have the extended commands.

3.2.2.1 Units

The extended SDI-12 command used to configure output units is

"aXCONFIG1=tt,pp,mmm.mm,000.00!" where a = the SDI-12 address of the sensor, tt = temperature units, pp = measurement unit, mmm.mm = multiplier

(slope), and 000.00 = offset.

Where valid entries for tt (temperature) is:

0 = Celsius

1 = Fahrenheit and valid entries for pp (pressure/level) are:

0 = psig

1 = kPa

2 = Bar

3 = Feet

4 = Meter

5 = Inch

6 = Millimeter

12

CS450/CS455 Pressure Transducer

Only SDI-12 instruction aM!, aM7!, and aM8! output the results obtained when using the multiplier and offset. The offset and multiplier is only applied to the pressure/level value, not to the temperature.

3.2.2.2 Sample Number

The extended SDI-12 command used to configure sample number is

"aXCONFIG2=nnn!", where nnn is the number of samples that will be measured to obtain the final output value, which is an average of the samples taken. This value only applies to the aM8! command. The integration time is a result of the number of samples selected. This value can be derived by adding

2 seconds to the number of samples. For example, if nnn = 50, then 50 samples would be averaged. The integration time for this process is 50 plus 2, or 52 seconds.

4. Installation

The CS450/CS455 is designed for water level measurements. Typical applications include agricultural water level/flow, water wells, lakes, streams and tanks. If the device is to be installed in a liquid other than water or in contaminated water, check the compatibility of the wetted material. The CS455 should be used in harsh water applications, including salt water.

4.1 Vent Tubes

A vent tube incorporated in the cable vents the sensor diaphragm to the atmosphere. This eliminates the need to compensate for changes in barometric pressure.

To prevent water vapor from entering the inner cavity of the sensor, the vent tube opening should terminate inside a desiccated enclosure or a desiccant tube. Moisture inside the vent tube can cause erroneous data and damage to the sensor.

4.2 Dislodging Bubbles

While submersing the sensor, air bubbles may become trapped between the pressure plate and the water surface, causing small offset errors until the bubbles dissolve. Dislodge these bubbles by gently shaking the CS450/CS455 while it is under water.

CAUTION Do not hit the sensor against the well casing or other solid surface while dislodging the bubbles, this could damage the sensor.

4.3 Water Depth

For water level measurements, the CS450/CS455 must be installed below the water at a fixed depth. This depth should be chosen so that the water pressure will never exceed the sensor’s pressure range. Otherwise, the output reading will not be correct, and the sensor can be damaged if pressure is excessive (2X full scale).

13

CS450/CS455 Pressure Transducer

Pressure can be converted to feet of fresh water using the following simple equation:

1 psi = 2.31 feet of water

For example, the maximum depth with a pressure range of 0 to 7.25 psig is

16.748 feet of water.

4.4 Well Installations

Lower the transducer to an appropriate depth.

CAUTION Do not drop the instrument or allow it to “free fall” down a well as this may damage the sensor.

With long drops it may be necessary to use the weighted nose cone (CSI

#25414).

Fasten the cable to the well head using cable ties or a weather proof strainrelief system (CSI #7421). The cable ties should wrap around the cable jacket; never suspend the CS450/CS455 from the connections at the top end of the cable. When securing the cable, care should be taken to avoid cable damage.

Sharp bends or excessive pinching of the cable can cause damage and may pinch off the vent tube causing measurement errors.

Several readings should be taken to insure proper operations after installation.

4.5 Other Installations

The CS450/CS455 can be installed in any position; however, when it leaves the factory it is tested in the vertical position. There will be an offset error if not installed vertically. Strapping the transducer body with tie wraps or tape will not hurt it.

If the CS450/CS455 is being installed in a fluid environment other than water, check the compatibility of the fluid with the wetted parts of the probe and cable.

Because the CS450/CS455 is vented, a desiccant tube must always be attached to the CS450/CS455. When you receive the CS450/CS455, ensure that the desiccant is blue. Replace the desiccant if it is pink or white.

4.6 Calculate Offset

The pressure created is directly proportional to the water column above the sensor. An offset is used to correct the final measurement to any error due to sensor zero offset or installation.

14

CS450/CS455 Pressure Transducer

For example, if the correct elevation of the water, as measured by a staff gauge or other measurement device, is 2015.50 feet, and the CS450 provides a reading of 5.76 psig, then:

5.76 psig * 2.31 ft/psig = 13.3056 ft.

So, the offset is calculated:

2015.50 ft – 13.3056 ft = 2002.1944 ft

This offset can be accounted for in the program instruction of the SDI-12 recorder.

5. Wiring

5.1 SDI-12

It is recommended to power down your system before wiring the

CS450/CS455. The shield wire plays an important role in noise emissions and susceptibility as well as transient protection.

TABLE 5-1. SDI-12 Wiring

Color

CS450/CS455

Function

CR800

CR5000

CR3000

CR1000

Red +12VDC 12V

CR200

Series

CR510

CR23X

CR10X

Battery+ 12V

G G

White SDI-12 Signal *Control Port C1/SDI-12 Control Port

Clear Shield GND

*dedicated SDI-12 port on CR5000

G G

15

CS450/CS455 Pressure Transducer

5.2 RS-232

TABLE 5-2. RS-232 Wiring

Color

CS450/CS455

Function Connection RS-232 9-pin

Red +12vdc

Black

White

Power Ground Power Ground

RS-232 Tx (Output) Transmit Pin 2 Rx (Input)

Blue RS-232 Rx (Input)

Yellow Digital Ground

Receive

Ground

Pin 3 Tx (Output)

Pin 5 GND

Ground

6. Programming

6.1 Using SCWIN

SCWIN is the easiest and typically the preferred method for programming the datalogger. SCWIN generates a wiring diagram that shows how to connect the

CS450/SC455 to your Campbell Scientific datalogger.

NOTE The sections that immediately follow are for CRBasic and Edlog.

Keyboard/Display users. SCWIN users can jump ahead to the Maintenance section.

6.2 Using CRBasic or Edlog

6.2.1 CRBasic Programming

Dataloggers that use CRBasic include our CR200-series, CR800, CR850,

CR1000, CR3000, and CR5000. These dataloggers use the SDI12Recorder to read the CS450/CS455. A multiplier of 1.0 and an offset of 0.0 yield water level in psig and temperature in degrees C.

The SDI12Recorder instruction has the following form:

SDI12Recorder(Destination, Output String, Multiplier, Offset)

16

Sample Program for CR200 Series Datalogger

‘CR200 Series

‘Declare the variable for the water level measurement

Public CS450(2)

‘Rename the variable names

Alias CS450(1)=Level

Alias CS450(2)=Temp_C

‘Define a data table for 60 minute maximum and minimums

DataTable(Hourly,True,-1)

DataInterval(0,60,Min)

Maximum(1,Level,0,0)

Minimum(1,Level,0,0)

Maximum(1,Temp_C,0,0)

Minimum(1,Temp_C,0,0)

EndTable

‘Read sensor every 60 seconds

BeginProg

Scan(60,sec)

‘Code for SDI-12 measurements:

SDI12Recorder(CS450,0M!,1,0)

‘Call the data table:

CallTable(Hourly)

NextScan

EndProg

Sample Program for CR1000 Datalogger

‘CR1000 Series Datalogger

‘Declare the variable for the water level measurement

Public CS450(2)

‘Rename the variable names

Alias CS450(1)=Level

Alias CS450(2)=Temp_C

‘Define a data table for 60 minute maximum and minimums

DataTable(Hourly,True,-1)

DataInterval(0,60,Min,10)

Maximum(1,Level,FP2,0,0)

Minimum(1,Level,FP2,0,0)

Maximum(1,Temp_C,FP2,0,0)

Minimum(1,Temp_C,FP2,0,0)

EndTable

CS450/CS455 Pressure Transducer

17

CS450/CS455 Pressure Transducer

‘Read sensor every 60 seconds

BeginProg

Scan(60,sec,1,0)

‘Code for SDI-12 measurements:

SDI12Recorder(CS450,1,”0”,”M!”,1,0)

‘Call the data table:

CallTable(Hourly)

NextScan

EndProg

6.2.2 Edlog Programming

Our CR500, CR510, CR10(X), and CR23X dataloggers are programmed with

Edlog.

These dataloggers use Instruction 105 to read the CS450/CS455. Your datalogger manual has a detailed explanation of Instruction 105.

Please note that Edlog only allocates one input location for Instruction 105.

Two input locations are required—one for the pressure measurement and one for the temperature measurement. The additional input location needs to be inserted manually using the Input Location Editor. To get into the Input

Location Editor, select Edit/Input Labels or press the F5 key. Once in the

Input Location Editor, do the following steps:

1. Choose Edit/Insert Block.

2. After the Insert Block dialog box appears, type in a base name for the input locations. Each input location will have the base name with an underscore and a consecutive number.

3. In the Start Address field, type in the number of the first input location.

4. In the Number of InLocs field, type in 2 and select OK.

Below is a portion of a CR10X program that measures the CS450/CS455.

NOTE The instructions below do not store data in final storage.

Instruction 92, Instruction 77 and processing instructions such as

Instruction 70 are required to store the data permanently.

18

CS450/CS455 Pressure Transducer

;{CR10X}

;

*Table 1 Program

01: 60 Execution Interval (seconds)

1: SDI-12 Recorder (P105)

1: 0

2: 0

3: 1

4: 1

5: 1.0

6: 0.0

*Table 2 Program

Port ;this is where the white wire is connected

Loc[Data_1

Mult

Offset

]

02: 0.000 Execution (seconds)

*Table 3 Subroutines

End Program

After Instruction 105 is executed, the input location called “Data_1” will hold the measured pressure, reported in PSI, and the input location called Data_2 will hold the measured temperature, in degrees Celsius. The results can be further processed by the logger. Note that Port 1 specifies that the SDI-12 data line is to be connected to the Port C1.

7. Maintenance

Campbell Scientific recommends that the CS450/CS455 be factory recalibrated and check every 24 months. Before a CS450/CS455 sensor is sent to Campbell

Scientific, the customer must get an RMA (returned material authorization) and fill out the Declaration of Hazardous Material and Decontamination form.

The CS450/CS455 has no user-serviceable parts. Cable can be damaged by abrasion, rodents, sharp objects, twisting, crimping or crushing and pulling.

Take care during installation and use to avoid cable damage. If a section of cable is damaged, it is recommended that you send your sensor back to replace the bale harness assembly.

Periodic evaluation of the desiccant is vital for keeping the vent tube dry. The

CS450/CS455 ships with the desiccant tube attached. To assess the effectiveness of the desiccant, use on of the following:

• The desiccant in the tube changes color from blue to pink when the drying power is lost.

• An enclosure humidity indicator such as CSI PN6571 humidity indicator card.

19

CS450/CS455 Pressure Transducer

7.1 Every Visit

• Collect data

• Visually inspect wiring and physical conditions

• Check indicating desiccant or enclosure humidity indicator; service if necessary

• Check battery condition (inspect physical appearance and use a keyboard display, PDA, or laptop to view the battery voltage)

• Check all sensor readings: adjust transducer offsets if necessary

• Check recent data

7.2 Every Two to Three Years or on a Rotating Schedule

• Send the CS450/CS455 in for inspection and have them served and or/replaced as needed.

8. Troubleshooting

The most common causes for erroneous pressure transducer data include:

• poor sensor connections to the datalogger

• damaged cables

• damaged transducers

• moisture in the vent tube

Problem:

Unit will not respond when attempting serial communications.

Suggestion:

Check the power (Red is +V and Black is Ground) and signal (White is SDI-12

Data) lines to ensure proper connection to the logger. Check the logger program to ensure that the same port the SDI-12 data line is connected to is specified in the measurement instruction.

Problem:

Transducer appears to be operating properly but data shows a periodic or cyclic fluctuation not attributable to water level changes.

20

CS450/CS455 Pressure Transducer

Suggestion:

A kinked or plugged vent tube will not effectively vent a gauge pressure

(Vented) type of device. Normal changes in barometric pressure will appear as water level fluctuations and these types of errors are typically on the order of 1 foot of water level. If the desiccant chamber has not been properly maintained, water may have condensed in the vent tube and the device should be returned to the factory for service.

21

CS450/CS455 Pressure Transducer

22

Appendix A. Calibration Certificate

Each CS450/CS455 has been calibrated to meet printed accuracy specification at multiple temperature and pressure ranges. If additional verification is required, a Calibration Certificate can be purchased for each CS450/CS455

Submersible Pressure Transducer.

This document included both the Certificate of Calibration and Instrument Data

Report. A sample for illustration is provided (Figure A-1).

The Instrument Data Report provides a list of the pressure and temperature at which the sensor was tested.

Pressure [kPa] is the pressure applied (listed in kilopascals) to the sensor.

Temperature [°C] is the temperature inside the test chamber at the time of testing. Pressure After [kPa] represents the resulting measurement output by the CS450/CS455 at the give pressure and temperature. Finally, Deviation

After [%F.S.], provides the difference between the actual pressure applied to the sensor and the pressure measurement output by the sensor. This value is listed as a percentage of the Full Scale range of the sensor.

When a CS450/CS455 is returned to CSI for calibration, the sensor will be returned with an Instrument Data Report. This report will include values in the

Pressure Before [kPa] column. These values represent the measured pressure the sensor returns at the specified pressure and temperature, BEFORE calibration.

A-1

Appendix A. Calibration Certificate

A-2

Appendix A. Calibration Certificate bars = kPa / 100.0 psi = kPa * 0.145

FIGURE A-1. Calibration certificate

A-3

Appendix A. Calibration Certificate

A-4

Campbell Scientific Companies

Campbell Scientific, Inc. (CSI)

815 West 1800 North

Logan, Utah 84321

UNITED STATES www.campbellsci.com • [email protected]

Campbell Scientific Africa Pty. Ltd. (CSAf)

PO Box 2450

Somerset West 7129

SOUTH AFRICA www.csafrica.co.za • [email protected]

Campbell Scientific Australia Pty. Ltd. (CSA)

PO Box 444

Thuringowa Central

QLD 4812 AUSTRALIA www.campbellsci.com.au • [email protected]

Campbell Scientific do Brazil Ltda. (CSB)

Rua Luisa Crapsi Orsi, 15 Butantã

CEP: 005543-000 São Paulo SP BRAZIL www.campbellsci.com.br • [email protected]

Campbell Scientific Canada Corp. (CSC)

11564 - 149th Street NW

Edmonton, Alberta T5M 1W7

CANADA www.campbellsci.ca • [email protected]

Campbell Scientific Centro Caribe S.A. (CSCC)

300 N Cementerio, Edificio Breller

Santo Domingo, Heredia 40305

COSTA RICA www.campbellsci.cc • [email protected]

Campbell Scientific Ltd. (CSL)

Campbell Park

80 Hathern Road

Shepshed, Loughborough LE12 9GX

UNITED KINGDOM www.campbellsci.co.uk • [email protected]

Campbell Scientific Ltd. (France)

3 Avenue de la Division Leclerc

92160 ANTONY

FRANCE www.campbellsci.fr • [email protected]

Campbell Scientific Spain, S. L.

Avda. Pompeu Fabra 7-9, local 1

08024 Barcelona

SPAIN www.campbellsci.es • [email protected]