Millipore Milli-Q Plus Technical information
Add to my manuals40 Pages
Millipore Milli-Q Plus is a state-of-the-art water purification system designed to provide ultrapure water for laboratory applications requiring the absence of interfering inorganic, organic, particulate, or microbiological material. With its patented purification pack concept, the system consistently produces Type 1 standards for Reagent Grade Water. It features a user-friendly control panel with an alphanumeric display for easy monitoring of system status. The system also includes a final filter unit to remove microorganisms and particles larger than 0.22 microns.
advertisement
iilli-Q* Plus
Water Purification System
Installation and Operation
M1LUPORE
MILLI-Q PLUS
A. Introduction
1. System Description
2. External Features
3. Schematic
4. Dimensions
5. Function of Components/Controls
B. Installation
1. Unpacking
2. Battery Insertion
3. Water Connection
4. Electrical Connection
5. Bench/Wall Mounting
6. Pack Installation
7. Initial Startup
C. Operation
1. Switch On - Auto Test
2. Operate
3. Standby
4. Auto Recirculation (Intermittent)
5. Temperature
6. Battery
7. Remote Standby
8. Pack Replacement
9. Final Filter - (Millipak) - Replacement
0. Maintenance
1. Access to PC Board
2. Battery Replacement
3. Fuse Replacement
4. Set Point Adjustment
5. Pack Replacement Frequency Setting
6. Changing the Frequency of Automatic Recirculation
1. Troubleshooting
E. Technical Information
1. Components
2. Specifications
3. Ordering Information
4. Warranty
Milli-Q Plus
A. INTRODUCTION
1. System Description
The Milli-Q Plus system is a state of the art water purification system, designed to provide the final polish to water which has been pretreated by primary purification methods such as
Reverse Osmosis (RO), Distillation, or Deionization (DI).
The system provides ultrapure water for the laboratory for use in critical applications requiring the absence of interfering inorganic, organic, particulate, or microbiological material.
As a result of attention to design detail, plus strict selection and testing of materials of construction, resins, and other purification media, the system consistently produces water which exceeds Type 1 standards for Reagent Grade Water.
The Milli-Q Plus system utilizes a patented purification pack concept. The design of this pack ensures minimal dead water volume and as a consequence extremely low TOC values in the purified water. The pack is constructed of pure polypropylene and is heat welded to minimize extractables associated with gluing.
The pack concept makes maintenance of the system extremely easy.
The system indicates when the pack requires to be changed and this can be done quickly and easily without the use of any tools.
The control box of the system utilizes a membrane keypad combined with an alphanumeric display. The alphanumeric display allows the user to monitor the status of the Milli-Q Plus system at all times.
A. Introduction
2. External Features
5 3 1
11 8 9 10
6 7 13 15 14 12
Front View (left)
1. Alphanumeric display
2. Control keypad
3. Pack lock/unlock arm
4. Purification pack
5. Manual delivery valve (ultrapure water from point of use)
6. Pure water point of delivery - Millipak Final Filter Unit
7. Front cabinet
Rear View (right)
8. On/off switch
9. Fuse holder
10. Power cable
11. Remote control cable
12. Cabinet locking nut
13. Serial number/catalogue number
14. Plug for entry of wall mounting rod option
15. Feed water inlet
A. Introduction
3. Schematic
Flow Schematic
1 - Inlet solenoid valve
2 - Pump
3 - Purification pack
4 - Resistivity sensor
5 - 3-way valve (manual)
6 - Check valve
7 - Millipak Filter Unit
A. Introduction
4. Dimensions
Height 495 mm
Depth 433 mm
Width 297 mm
Weight (without purification pack) 9.3 kg
(with purification pack installed and 15.4 kg full of water)
A. Introduction
5. Function of Components/Controls
On/Off Power Switch
This switch, located on the rear side of the unit, controls the power to the control box. The power inlet socket is located below the ON/OFF switch. This socket has a built in fuse holder which contains the main fuse plus a spare.
Pump
This magnetic drive gear pump provides a draw-off flow rate of
1.5 1/min. A built-in pressure relief valve regulates discharge pressure to 2.8 bars (40 psi).
Sensor
The sensor, located at the outlet of the purification pack, is electrically connected to the PC board. The sensor continuously measures the resistivity of the product water as it exits the pack.
Meg-0-Meter Display
Resistivity of the water is displayed on the alphanumeric display as product , Megohm x cm.
The unit is internally temperature-compensated to normalize readings to 25° C.
Three Way Point-of-Use Valve
This PVDF valve, located behind the front panel of the unit, is manually operated by the knob on the front panel of the unit.
This valve has two positions: RECIRCULATION - where the water is directed through the system's recirculation loop; or
PRODUCTION - where the water is directed through the final filter to the point-of-use.
CONTROL KEYBOARD
Operate/Standby, Operate Mode
This button, when first depressed, will switch the system to the operate, mode. During operate mode the inlet solenoid valve is open and the pump is running. If the manual point-of-use valve is in the RECIRCULATION position, water will recirculate around the system's internal recirculation loop. If the point-of-use valve is open in the PRODUCTION position, water will be delivered through the point-of-use. During operate, the display will show the resistivity of the purified water as product , Megohm x cm.
If this resistivity is less than the set point of 14 Megohm x cm then the display of product , _ Megohm x cm will flash on and off. When the resistivity builds to greater than the set point, the display of product , _ Megohm x cm becomes a steady display.
The set point is adjustable (see Section D., Maintenance, 4., Set
Point Adjustment, in this manual).
Standby Mode
During normal operate mode, if the button is depressed, this will switch the system to the standby mode.
During standby the inlet solenoid valve is closed and the pump is not running. Therefore, the system is not producing water.
During standby the system will switch on automatically every
55 minutes and recirculate for 5 minutes. This inhibits bacterial growth and degradation of pure water
Auto-Test
Each time the unit goes into operate mode, it carries out an automatic test on the accuracy of the resistivity measuring system.
The test result is displayed for 4 seconds before the display changes to show product water resistivity. The display should show TEST:
15 Megohm x cm.
Temperature
During the standby mode, if the temperature button is depressed, the display will show the calibration temperature of 25° C.
(If 25° C is not displayed then there is a calibration problem.)
During operate mode, if the button is pressed, the unit will display the current temperature of the water in the system in ° C.
Power Light
This light, when illuminated, indicates that the unit is switched on.
Exchange Pack Light
When this lamp is flashing, it indicates that the pack has been installed longer than 4 months and should be exchanged.
(The exchange interval is set at 4 months, but can be changed to 6 months (see Section D., Maintenance, 5., Pack Replacement
Frequency Setting, in this manual).
When this lamp is steadily illuminated, it indicates the absence of the pack; or if the pack is there, then it has not been pushed fully into place.
ALPHANUMERIC DISPLAY
Keyboard
Press OPERATE/STANDBY
Mode
Operate
Display
The first display is
Test , _ Megohm x cm
(for 4 sec), then product , _ Megohm x cm
Press OPERATE/STANDBY Standby
Automatic recirculation 5 minutes Standby every 55 minutes
Press TEMPERATURE during Operate
Standby
Recirculation
Press TEMPERATURE during Standby
Temperature: ° C
(temperature of water in system)
Temperature: 25° C
(calibration temperature)
Fault Displays
Fault in resistivity sensing system Test incorrect
Fault in temperature sensing system Temperature:
Temperature:
< 0° C
> +° C
(See Section D., Maintenance, in this manual.)
PURIFICATION PACK
The purification pack utilizes a chain of purification media to polish the feed water to high quality:
QPAK1 (to be used if feed water has been pretreated by reverse osmosis or distillation)
I
Activated
Carbon
Organex
Mixed
Bed
QPAK2 (to te used if feed water has been pretreated by deionization)
I
I
Activated
Carbon
Prefilter
Mixed
Bed
Final Filter Unit
The pre-sterilized, vented microporous membrane filter unit is used as a final filter at the point-of-use. It removes micro-organisms and particulate matter larger than 0.22 um.
B. Installation
1. Unpacking
Unpack the system carefully from its container.
Verify Contents: 1. Milli-Q Plus system
2. Operation and maintenance manual
3. Power cable
4. Cover for point-of-use
5. 3 meters of feed water tubing - white, 8 mm O.D.
6. Roll Teflon tape
7. 1/4" male - 8 mm diameter tubing adapter feed water
8. 1/2" female to 1/4" female reducing bushing with stainless steel filter screen
9. 1/4" male - 1/4" hose barb
10. 2 meters tubing, 12 mm O.D.
Purification Pack
The system requires a purification pack which is not included with the unit. The purification pack must be ordered separately,
The choice of purification pack depends on feed water quality.
Feed Water
Reverse Osmosis or distilled - use CPMQ 004 Rl
Deionized - use CPMQ 004 D2
Pack Contents 1. Purification pack
2. Millipak final filter unit
B. Installation
2. Battery Activation
Before making water, or electrical connections (initial startup) the battery, to power the system memory, must be activated.
Open rear cabinet cover by unlocking cabinet locking nut
Pull out PC board slightly
Hold battery with protective cloth (to avoid touching it directly) and pull out small plastic sheet located underneath the battery
Push PC board back, replace Rear cover and tighten lock nut.
•Ml
Note: Switch on power immediately after battery activation.
Switch system to ON for a few seconds. This will initialise the microprocessor and avoid discharging the battery.
3. Water Connection
In order to protect the Milli-Q Plus Unit from any gross particulate matter in the feed water line, a 1/2" female to 1/4" female reducing bushing, which incorporates a stainless steel filter screen, is supplied with the system.
This fitting should be used to connect the system to the feed water line. The fitting should be periodically dismantled and the filter screen should be rinsed in running water to remove entrapped particulate material.
For details, please refer to exploded view shown below.
, 1/2" female Stainless steel screen 1/4" female
This side to be connected to feed water line
Rubber gasket
This side to be connected to 1/4" male quick connect tube fitting (to water purification system)
The system has a single "quick connect" connection located at the rear of the system.
Simply push-in tube to attach
Tube is secured in position
Push-in collet to release tube
Isolating Valve
A valve should be installed on the inlet water line to facilitate maintenance of the Milli-Q.
Tank feed
0,1m min.
Distribution Loop Feed
Pressure
Regulator
Valve
HX—I
Rec ire. Loop
Feed Water Pressure
The pump requires a positive feed of water. 10 cm head is sufficient. The maximum pressure is 1 bar (15 psi).
If the feed pressure is higher than 1 bar then a pressure regulator should be fitted. Pressure regulator -
Cat. No. ZFMQ 000 PR (full details in Accessories Section).
CAUTION: Although the pump is self-priming, it should never be started without an adequate water supply.
Feed Water Temperature
Feed water temperature must be between 5" C and 40° C.
B. Installation
4. Electrical Connection
ON/OFF Switch
Fuse holder
Power Inlet
Main Connection: 110-120 V/60 Hz (0.8 A) (System Cat. No. ZD52 115 84)
Grounded
220-240 V/50 Hz (0.4 A) (System Cat. No. ZFMQ 050 01)
Grounded
Power: 85 VA maximum
Fuse: 0,5 A (220 - 240 V)
1A (110 - 120 V)
Remote Standby: Normally closed switch plugged in the socket,
5 V/10 mA.
B. Installation
5. Bench/Wall Mounting
The system is compact, and has four rubber feet for easy installation on the laboratory bench.
Alternatively, the system can be mounted on the wall using the wall mounting bracket accessory,
Cat. No. ZFMQ 00W MB
Full mounting instructions included with the mounting bracket.
B. Installation
6. Pack Installation
Remove plugs from pack.
Remove plugs from pack receptor inside Milli-Q.
Unlock pack retaining arm on Milli-Q by pushing down on tab (marked press) and swinging arm to the left.
Moisten 0-rings with water.
Maintain position of Milli-Q with left hand. Push pack into place with right hand.
Swing locking arm back into place.
By pushing gently on the underside of the arm it will snap back into original position.
B. Installation
7. Initial Start Up
Checklist: - battery in place and activated
- water connected to rear of unit
- power cable plugged into rear socket
- pack inserted/handle locked
- isolating valve on feed water line open
- switch system ON/OFF switch to ON position to initiate Microprocessor
Control and Display Layout
- "battery" is displayed 2 seconds for 3 seconds to reset battery display
Air Purge
In order to purge all of the air from a newly installed pack it is necessary to run some water to drain.
New Pack in Place - Checklist OK
Connect adapter and tubing from point-of-use to drain.
Open point-of-use valve. Turn to
PRODUCTION.
Press OPERATE/STANDBY to switch the system to operate mode.
Allow water to collect in a vessel with the tubing submerged in the vessel.
Allow system to run for 5 minutes to vessel.
Close point-of-use valve. Turn to RECIRCULATION.
Allow system to recirculate for 5 minutes.
Open point-of-use valve. Turn to PRODUCTION.
Allow system to run to vessel for 2 minutes.
Close point of use valve. Turn to RECIRCULATION.
Press OPERATE/STANDBY to switch to standby mode.
Remove adapter and tubing from point-of-use.
Put 1-2 turns of Teflon tape on threaded end of Millipak final filter unit.
Screw Millipak firmly into point-of-use adapter until the vent cap is oriented to the system.
Fit Millipak inside point-of-use cover.
Open vent cap on Millipak.
Turn point-of-use valve to
PRODUCTION.
Press OPERATE/STANDBY to put the system into operate mode.
Threaded adapter
Vent cap
Allow water to purge air from
Millipak.
Close Millipak vent by replacing cap.
Close point-of-use valve. Turn to RECIRCULATION.
Press OPERATE/STANDBY to go into standby mode.
Press on tabs at side of point-of-use cover.
Position point-of-use cover and allow tabs to pop into place.
The system is now ready for use and is sitting in the standby mode. Please proceed to the next section of the manual which describes the normal functions and operation of the system.
C. Operation
1. Auto Test
When the operate button is pressed, after a standby mode, the unit performs an automatic test of the accuracy of the resistivity sensing system.
P0WSR
I •
mum®.—" r
OPERATE '
STANDBY
Display will show test resistivity for 4 seconds. This is preset at 15 Megohm x cm
"t 1 Megohm x cm
After this 4 second period the display will switch to show the actual resistivity of the water in the system at that moment.
2. Operate
After the 4 second display of the auto test resistivity the unit will display the actual resistivity in the system.
If this value is below the set point (14 Megohm x cm) resistivity, the value will flash on and off until the resistivity builds to the pre-adjusted set point resistivity.
Resistivity will steadily build to a level above the set point.
Resistivity is now shown as a steady display.
3. Standby
During periods of no demand, the user can press the OPERATE/
STANDBY button to convert the system to STANDBY mode (when running in operate mode). During STANDBY the display will show
STANDBY.
4. Auto Recirculation (Intermittent)
During STANDBY the system will automatically recirculate water for a period of 5 minutes every one hour (or 3 hours if selected)
During this period the display will show RECIRCULATION.
Intermittent recirculation discourages bacterial proliferation, and maintains water quality levels between uses.
5. Temperature
The temperature button can be depressed during operate mode.
This will display the current temperature of the water in the system in ° C.
The display will last 4 seconds when TEMPERATURE button is depressed. During standby mode, the unit will display the auto test temperature which is 25° C. Display: TEMPERATURE 25° C.
The temperature display range is 0° C to 40° C. Outside of these limits the display will show TEMPERATURE < 0° C, or TEMPERATURE
> +° C
This may be a true reading indicating a problem with feed water temperature.
Check feed water temperature. If feed water temperature is satisfactory, refer to troubleshooting.
6. Battery
The display may show BATTERY followed by a normal function display - RECIRCULATION, etc. This sequence will be repeated
(battery 2 seconds, function 5 seconds). This indicates that the battery is exhausted and should be replaced. (As in Section B.,
Installation, 2., Battery.)
The battery is used to maintain some data in the event of a power failure. During a power cut, the battery stores the data telling it which function it was performing at the time of the power cut.
When power is restored, the unit will restart in this function.
The battery also stores elapsed time from pack installation.
7. Remote Standby
When the system is connected to a remote controlling unit, via the inlet jack plug at the rear of the unit, the display will show REMOTE STANDBY. (Remote control, e.g., clinical analyzer.)
The Milli-Q Plus should initially be switched to operate mode by the user. In this mode the remote unit is able to switch the
Milli-Q Plus from OPERATE to
STANDBY and vice versa when the remote unit requires water.
During REMOTE STANDBY, the system will continue to perform automatic intermittent recirculation as in normal standby. During this period, the unit will show RECIRCULATION, as normal.
3ack plug for remote control
When the contact from the remote unit is closed, the Milli-Q Plus will run through its normal sequence of AUTO TEST, ACTUAL
RESISTIVITY flashing till set point is reached, etc.
8. Pack Replacement
The orange EXCHANGE PACK light will flash to indicate that it is time to replace the purification pack with a new one.
If the EXCHANGE PACK light is steadily illuminated, it indicates that the pack sensing microswitch is open. This indicates the absence of the pack, or if the pack is installed then it has not been pushed fully into place.
Switch the Milli-Q Plus to STANDBY mode
Unlock pack retaining arm.
Maintain Milli-Q Plus in position with left hand. Pull out pack smoothly with right hand.
Dispose of old pack.
Remove caps on new pack.
Moisten 0-rings.
Maintain Milli-Q Plus in position with left hand. Slide pack into position with right hand.
Relock pack retaining arm.
Press OPERATE/STANDBY.
Refer to Section C , Operation, for sequence of steps from this point, Normal Operation.
Pack Replacement Frequency
See Section D., Maintenance, 5., Pack Replacement Frequency
Setting.
9. Final Filter (Millipak) Replacement
(Refer also to Section B., Installation, 7., Initial Start-Up)
The final filter unit should be replaced when the flow rate decreases to an unacceptable level (0.5 1/min).
Press OPERATE/STANDBY to switch the system to STANDBY mode.
Press on tabs at side of point-of-use cover. Remove point-of-use cover and lay aside.
Unscrew Millipak from final 1/4" female fitting and dispose of
Millipak.
Wrap Teflon tape once around the threads on the 1/4" NPTM end fitting on Millipak.
Screw Millipak into 1/4" female fitting.
Press OPERATE/STANDBY to switch the system to operate mode.
Loosen vent cap on top of Millipak final filter unit (to vent air from filter).
Open point-of-use valve to PRODUCTION position.
Allow water to purge air from vent.
Close the vent.
Close point-of-use valve switch to RECIRCULATION.
Press OPERATE/STANDBY to switch the system to STANDBY.
Replace Millipak inside point-of-use cover and replace point-of-use cover in position by pressing on tabs on side of cover and allowing them to pop into place.
D. Maintenance
1. Access to PC Board
(Refer to Section B., Installation)
Remove rear cabinet by unlocking the cabinet locking nut at rear.
Disconnect 5 traces.
Pull out PC board.
2. Battery Replacement
Battery lifetime: 5-8 years when system is powered for 24 hrs/day;
2-3 months when system power is off.
System Memory: when battery is exhausted, the system will store data (FUNCTION position and elapsed time from pack insertion) for only 20 seconds after power is cut.
The display BATTERY (flashing) indicates an exhausted battery, and that power has been off for longer than 30 seconds. This means stored data has been lost (elapsed time, etc.).
Replace exhausted battery with a fully charged one.
Battery replacement procedure: see Section B., Installation, 2.,
Battery Activation.
3. Fuse Replacement
If power lamp is off, check fuses as follows:
Main ON/OFF switch, rear of unit, and socket on ring main.
Check main fuse. Remove main cable. Remove fuse — insert holder.
Replace main fuse by spare found inside fuse insert holder.
Check PC board fuse. Replace if necessary.
PC board fuse: 220-240V version has 0.1 amp.
115V version has 0.25 amp.
a b c d e m #
** # *--—" a -
> t
* !
*>?l
4. Set Point Adjustment
The resistivity set point is preadjusted to 14 Megohm x cm.
This can be adjusted from 1 Megohm x cm to 18 Megohm x cm, but should be done by a Millipore service engineer.
Call Millipore for assistance.
5. Pack Replacement Frequency Setting
The EXCHANGE PACK light will begin to flash after the pack has been in place for longer than a preset time.
The pack should be changed for optimum performance in terms of
T.O.C. and microorganisms, even though all ion exchange Resin may not be 100% exhausted.
There are two preset time intervals available: 4 months, or 6 months.
Jumper (Refer to illustration in section 3., Fuse Replacement)
Jumper (position b) in place gives 4 months preset time interval.
Jumper removed gives 6 months preset time interval.
6. Changing the Frequency of Automatic Recirculation
In order to maintain water quality, the system will automatically recirculate water for 5 minutes every 55 minutes when it is in
STANDBY mode. If you wish, you can change the setting so that water will recirculate for 5 minutes every 3 hours.
To change setting:
- Open rear cabinet and pull out PC board as described in section B.2.
- Remove jumper A (see illustration in section D.3.)
Note: If this jumper is connected to the PC board, water will recirculate for 5 minutes every 55 minutes. If jumper is removed, water will recirculate for 5 minutes every three hours.
- Replace PC board and close rear cabinet.
7. Troubleshooting
Fault: Initial AUTO TEST:
Resistivity is not 15 Megohm x cm ± 1 Megohm x cm
When the OPERATE/STANDBY button is first pressed, the unit carries out an internal check on the accuracy of its resistivity monitoring system.
The display should show 15 Megohm x cm (+ 1 Megohm x cm) for 4 seconds).
If the value by test is less than 14 Megohm x cm or more than
16 Megohm x cm, then the display will show TEST INCORRECT for 2 seconds.
Then actual resistivity in the system for 4 seconds.
This sequence of TEST INCORRECT/ACTUAL RESISTIVITY is repeated.
This fault may be due to a problem of calibration of temperature or resistivity.
To Check Test Temperature Calibration
. Put system in STANDBY mode.
. Press TEMPERATURE button - display test temperature.
. The test temperature is 25° C.
. If the display shows anything other than 25° C then you will need to recalibrate the temperature setting.
To Recalibrate Temperature
Remove rear cabinet.
Remove Dumper C (this locks PC board to TEST mode).
Press temperature button and maintain.
Display shows the TEST calibration TEMPERATURE
(should be 25° C ) .
If not 25° C, adjust temperature (TEMP) potentiometer as shown until 25° C is displayed.
Release TEMPERATURE button on keypad.
Displays TEST RESISTIVITY
(should be 15 Megohm x cm).
If display does not show TEST: 15 Megohm x cm after Temperature
Recalibration:
Adjust RESISTIVITY (F/RES) potentiometer until display shows
15 Megohm x cm.
Replace test Jumper C in position.
Replace rear cover and lock in place.
Fault: Display shows TEMPERATURE < 0° C or TEMPERATURE
> + ° C for 2 seconds, followed by product XX,X Megohm x cm for 2 seconds.
Check the TEST TEMPERATURE as previously described for Test
Temperature Calibration.
Readjust to 25° C if incorrect.
Check if the temperature of your feed water is elevated as this could be a true reading of feed temperature > 40° C.
If the temperature of the feed water source is satisfactory then call Millipore. This indicates a problem with the thermistor.
E. Technical Information
1. Components
Components
Cabinets front-rear
Internal skeleton
Hardware (screws, nuts, washers)
3-way valves
0-rings
Microsensor tips
Check valve
Valve operating handle
Pack locking handle
Sanitization port access lid
Point-of-use cover
Tubing
Pump Body
Pump materials (contacting fluid)
Inlet solenoid valve
Point-of-use fittings
Point-of-use tubing
Purification pack
Materials of Construction
Polypropylene
Noryl (polyamide)
Stainless steel
PVDF
Nitrile - EPDM
316 S.S. electropolished
Stainless steel and copolymer acetyl resin
Polyamide
Polyamide
Stainless steel
Ryton, Teflon,
316 S.S.
PVDF
PTFE
Polypropylene
2. Specifications
Dimensions
Height 495 mm
Width 297 mm
Depth 433 mm
Connections
Feed connector Polyamide tubing 8 mm O.D.
quick connect fitting
Electrical Requirements
220-240V/50Hz, 85 VA (0,4A) Grounded
Fuse size: 0.5 amp/250V "Slow Blow" or
110-220V/60Hz, 85 VA (0,8A) Grounded
Fuse size: 1 amp/250V "Slow Blow"
Feed Water Requirements
Feed water should be preteated by:
1) Reverse Osmosis
2) Distillation
3) Deionization
System Performance
Water production rate up to 1.5 1/min. Product quality exceeds all standards for type 1 reagent grade water.
Note: Product water quality will degrade if stored.
Pump and Motor
Type Positive displacement magnetic gear pump
Electrical 220-240V/50Hz 0.25 amps
115V/60Hz 0.43 amps
Maximum discharge pressure 60 psi (4.0 bar)
Bypass adjustment 40 psi (2.8 bar)
3. Ordering Information
Description Catalogue Number
Milli-Q Plus system 220-240V/50 Hz ZFMQ 050 01
Milli-Q Plus system 110-120V/60 Hz ZD52 115 84
Expendables
Purification pack 1 CPMQ 004 Rl
Feed water RO or distilled, 1/pk
Purification pack 1 with final filter CPMQ K05 Rl
Feed water RO or distilled, 1/pk
Purification pack 2 CPMQ 004 D2
Feed water deionized, 1/pk
Purification pack 2 with final filter CPMQ K05 D2
Feed water deionized, 1/pk
Millipak GV final filter unit MPGL 04S K2
0.22 urn, sterile, 1/4" NPTM inlet with filling bell, 2/pk
Accessories
60 L reservoir complete ZFRE 000 60 with float switch, bacterial filter on air vent, sanitary drain line and all connections to RO and Milli-Q systems
Pressure regulator ZFMQ 000 PR polypropylene, 1/4" NPTF, with pressure gauge
(Two 1/4" NPT x 8 mm tube connectors, are included to connect this regulator to the feed line)
Pressure switch FTPF 006 08
FTPF 006 09
System wall mounting bracket ZFMQ 00W MB
4. Warranty
Millipore Corporation (the "Company") warrants its Water purification products against defects in materials and workmanship when used in accordance with the Company's instructions for use, for a period of one year from date of purchase of the products. The company's sole obligation shall be to repair or replace, at its option, any product or part thereof that proves defective in material or workmanship within the warranty period provided the customer notifies Millipore promptly of any such defect. The Company reserves the right to charge the user for the labor costs involved in repairing or replacing any defective part after
90 days from installation. The Company further reserves the right not to honor this warranty if it believes said warranty is being abused. MILLIPORE MAKES NO OTHER WARRANTY, EXPRESS
OR IMPLIED. THERE IS NO WARRANTY OF FITNESS FOR A PARTICULAR
PURPOSE. The company shall not be liable for consequential damages resulting from economic loss or property damage sustained by any customer from the use of its product.
While Millipore personnel are available to advise customers concerning general applications of the Company's products, oral representation are not warranties with respect to particular applications and should not be relied upon if inconsistent with the terms stated herein. Any variation of these terms must be in writing and must be signed by an officer of Millipore.
These warranty conditions do not affect the statutory rights of the purchaser.
Catalogue No. EU303/U. Printed in France, 1/90. Copyright C 1990,
Millipore Corporation, U.S.A. Millipore is a registered trademark of Millipore Corporation.
Milli-Q, QPAK, Millipak and Organex are trademarks of Millipore
Corporation.
Milli-Q* Plus
Reagent Grade
Water Purification System
Catalogue Numbers
ZFMQ 05001 (230 V/50 Hz)
ZD5211584(115V/60Hz)
Components and Spare Parts
MILLIPORE
— LEFT SIDE VIEW
1 ) ficfl (S) ( 5
ITEM DESCRIPTION
1 IWAKI PUMP
2 FUSE 5x20 mm
3 CONTROL P.C. BOARD
4 SOLENOID VALVE COIL
5 RUBBER PAD
6 POWER SUPPLY SOCKET
7 MALE CONNECTOR 1/4" M-06T
ADAPTOR 1/4"F-1/4"F
9 ELBOW 0 6 MALE - 06T
10 POLYAMIDE TUBING 0 6 / 4
SOLENOID VALVE MOUNTING KIT
12 SOLENOID VALVE BODY SST
13 MALE CO N N ECTO R 1 /4"M - 08T
14 PVDF ELBOW 1/4"F-06T
•15 POINT-OF-USE COVER
16 PTFE T U B I N G 0 6 / 4
17 PVDF 3 - W A Y VALVE 1/4"
18 PVDF CONNECTOR 1/4"M-06T
19 KEYBOARD STICKER
LIST OF COMPONENTS
CATALOGUE NO.
MILLI-Q PLUS
220/240 V - 50 Hz
CAT. N O . 2FMQ 050 01
MILLI-Q PLUS
115 V/60 Hz
CAT. N O . ZD52 115 84
ITEM
FTPF01782
FTPF 00306
ZF20 000 50
FTPF 02467
FTPF 02560
FTPF 01562
FTPF 02710
FTPF 02716
FTPF 02473
FTPF 00756
ZF20 000 51
FTPF 02468
FTPF 02713
FTPF 02720
FTPF 02569
•FTPF 02466
FTPF 02630
FTPF 02723
FTPF 02557
FTPF 02724
FTPF 01776
FTPF 02722
FTPF 02611
DESCRIPTION
20
21
22
23
24
25
26
27
28
29
30
31
3-WAY VALVE KNOB
TOP DOOR
KNURLED KNOB
SOLENOID VALVE CONNECTOR
CHECK VALVE
KEYBOARD P.CIRCUIT
TEE3x06T
LITHIUM BATTERY 3V
RESISTIVITY CELL 0.02 WITH PACK ADAPTOR
MICROSWITCH WITH CABLE
LCD DISPLAY 16 DIGITS WITH CABLE
JACK FEMALE CONNECTOR WITH CABLE
NOT SHOWN:
POLYAMIDE TUBING 8 / 6
CORD LINE 220V
CORD LINE 115V
CATALOGUE NO.
FTPF 02558
FTPF 02556
FTPF 02561
FTPF 00396
FTPF 02718
FTPF 02605
FTPF 02715
FTPF 02480
ZF20 000 10
ZF20 000 20
ZF20 000 30
ZF20 000 21
FTPF 02721
FTPF 01866
FTPF 02471
r
1
MILLIPORE
FRONT VIEW
MILLIPORE
For more information please call the nearest
Millipore subsidiary
North America
Call toll-free 800-225-1380
In Western States 800-632-2708
In Canada 800-268-4881
In Puerto Rico 809-747-8444
In Toronto 416-678-2161
In Massachusetts 617-275-9200
Africa
St Quentin Yvelines -
France 111 30.45.70.00
Australia
Toll-free 10081 222 - 111
Sydney 1021 427 - 0511
Austria, USSR,
Eastern Europe,
Middle East
Vienna 102221 82 89 26
Belgium - Luxemburg
Brussels (021 242.17.40
Brazil
Sao Paulo [0111 548 - 7011
Denmark
Tastrup 42 - 52 88 11
Finland
Espoo 90 - 801 90 77
France
St Quentin Yvelines
(II 30.45.70.00
India
108121394657
Italy
Vimodrone 1021 25078 -
Roma (061 57 33 600
Japan
Tokyo 1031 474 - 9111
Mexico
Mexico 19051 576 - 9688
Norway
Oslo 02 - 67 82 53
Singapore
Singapore 2532733
Spain
Madrid 1911 729 03 00
Barcelona (93) 325 96 16
Sweden
Vastra Frolunda 031-289860
Sundbyberg 08 - 98 89 60
Switzerland
Kloten 1011 814 13 63/65
The Netherlands
Etten-Leur 1016081 22000
UK - Ireland
Watford 10923) 816 375
West Germany
Eschborn 106196] 494 - 0
In all other Countries
Bedford - USA (6171275 - 9200
Cat. No. EU 288/U. Printed in France. 11/89. Copyright © 1989, Millipore Corporation, USA. Millipore is a registered trademark of Millipore Corporation.
Milli-Q is a trademark of Millipore Corporation.
PLEASE NOTE
PRODUCT IMPROVEMENT
Milli-Q Plus Milli-Q UF Plus Alpha-Q
Milli-RO 6 Milli-RO 6 Plus Milli-RO 10 Milli-RO 10 Plus
All units are now shipped with a rechargeable battery on the p.c.
Board in place of the disposable battery on the initial series of units.
Please ignore the sections on BATTERY ACTIVATION and BATTERY
REPLACEMENT in this Operation and Maintenance manual. These sections are no longer valid with the rechargeable battery.
New start up procedure
. Switch on power to system at mains connection.
. Flick ON/OFF switch at rear of unit to ON.
. Display may show BATTERY if the battery is not fully charged.
In this case switch unit OFF for three seconds and then switch back ON.
Continue start up as normal from this point referring to
B. INSTALLATION 3. WATER CONNECTION.
advertisement