Omega PRTXD Series Owner Manual
Advertisement
Advertisement
MADE IN
PRTXD Series RTD Temperature Transmitters
M-4115/0507
INSTRUCTION
SHEET
Shop online at: omega.com e-mail: [email protected]
For latest product manuals: omegamanual.info
PRTXD Series RTD Transmitters
MODEL
PROBE
LENGTH
PROBE
TYPE
PRTXD-*-4
PRTXD-*-4-SL
PRTXD-*-12
PRTXD-*-12-SL
4"
4"
12"
12"
Fixed
Spring-Loaded †
Fixed
Spring-Loaded †
† Thermowell required for spring-loaded versions
* Insert temperature code
CODE RANGE
100C
150C
0 to 100°C
0 to 150°C
100F
200F
300F
0 to 100°F
0 to 200°F
0 to 300°F
Accuracy
Display: ±0.1% of span, plus sensor error, ±1 digit
Analog output: ±0.1% of span, plus sensor error
Sensor: ±0.3°C at 0°C, ±1.1°C at 150°C
Sensor
100
Ω RTD element, 0.00385 alpha coefficient
IEC-751 Class B: ±0.3°C at 0°C, ±1.1°C at 150°C
10 second time constant
Display (update rate, type, size)
3 readings per second nominal display update rate
3 digit LCD, 1 /
2
" digit height
Controls
Non-interactive zero and span, ±10% range
Test calibration level; 0-100% range
Loop Supply Voltage
Any DC supply/loop resistance that maintains 8 to 32
VDC at gauge terminals. Minimum 8 V required at gauge terminals at all times for proper operation.
Reverse polarity protected
Loop Output Characteristics
4-20 mA output
See chart below right for loop voltage characteristics
Test Function
Front panel TEST button when held, sets loop current and display to test calibration level, independent of temperature input, to allow testing of system operation.
Wiring Connections (see drawing)
3 ft long, 2-conductor 22 AWG cable
Weight
Transmitter: Approx. 12 ounces
Shipping weight: Approx. 1 pound
Material and Color
NEMA 4X ABS/Polycarbonate housing, light gray
Gasketed rear cover
Connection and Probe Material
1 /
2
" NPT male hex nipple, 316 stainless steel
316 stainless steel probe sheath
Storage Temperature
–40 to 203°F (–40 to 95°C)
Operating Temperature
–4 to 185°F (–20 to 85°C) at housing
Installation and Precautions
A ther mowell is required with flowing mater ial, pressurized applications, and all applications using spring-loaded versions of the PRTXD.
Install or remove using wrench on probe hex fitting only.
Do not attempt to tighten by turning housing or any other part of the transmitter. Due to the hardness of 316 stainless steel, it is recommended that a thread sealant be used to ensure leak-free operation.
Electrical Connections
The 2-wire connection allows the PRTXD to be used as a temperature display powered by a low-voltage DC source, or as a loop-powered 4-20 mA transmitter.
Connect the loop (+) supply to the RED lead and the loop (–) supply to the BLACK lead. Reversing the connections will not harm the transmitter but it will not operate with incorrect polarity.
Select a loop power supply voltage and total loop resistance so that when the loop current is 20 mA, the transmitter will have at least 8 VDC at its terminals. For correct operation and to avoid erratic or erroneous readings, the terminal voltage must not fall below 8
VDC. Too large a loop resistance will cause the output to
“limit” or saturate before reaching its full 20 mA output.
The minimum loop supply voltage may be calculated from the formula:
Vmin = 8V + (20mA x Total loop resistance)
If the terminal voltage falls below about 7.8 VDC erratic operation may occur. This is an indication that the loop supply/resistance may not allow adequate power for reliable operation. This should never occur in normal use. If it does, examine the loop supply and resistance.
Operation
The PRTXD series is designed for continuous operation.
Warm-up time is negligible. The display will show the temperature, and the loop current will be proportional to the indicated temperature.
4 mA = Zero or low end
20 mA = Span, full-scale or high end
3.5"
3.0"
2.0"
PRTXD
TEST Button
When the front-panel TEST button is held depressed, the display and loop current are switched, independent of the temperature, to a test level determined by the setting of the Test adjustment.
Remove the nylon screw to access the Test potentiometer. To set the test output level, press and hold the front-panel TEST button and adjust the Test level to set the display and loop current. When the button is released, normal operation is resumed.
Calibration
The PRTXD is factory calibrated and there is generally no need to alter calibration settings. Zero and span calibration should only be attempted if the user has access to temperature calibration equipment that has at least four times the transmitter accuracy.
Remove nylon cover screws to access the front calibration potentiometers. These potentiometers calibrate the display. Zero calibration must be done before span calibration. Record readings at three to five points over the transmitter range and adjust span control to minimize error and meet specifications.
Internal zero and span controls adjust the agreement between the displayed value and the 4-20 mA loop current. These are set at the factory and should not normally be adjusted. If adjustment is necessary, accurate current measurement equipment are required to successfully complete this calibration.
3 ft long,
2-conductor
22 AWG cable
(+)RED
Loop Power
Supply
24 VDC typical
(+) (–)
(–)BLACK
Remote Device
(–) (+)
°F
Span
TEST
Test Zero
Omega Engineering Made in USA
.75"
Voltage Compliance for
4-20 mA Current Loop
1400
1.37" 0.75" 1.25"
1200
Typical thermowell
1/2" NPT
316SS hex nipple
1000
800
Spring-loaded probes
0.25" nom.
compression
(0.50" max.)
Stem length
“L”
Continuous weld on non-spring loaded models
0.25" dia.
600
400
200
0
8 12 16 20 24 28
Loop Supply Voltage (DC)
32
OMEGAnet
®
Online Service Internet e-mail www.omega.com [email protected]
Servicing North America:
USA:
ISO 9001 Certified
One Omega Drive, Box 4047
Stamford CT 06907-0047
Tel: (203) 359-1660 e-mail: [email protected]
FAX: (203) 359-7700
Canada:
976 Bergar
Laval (Quebec) H7L 5A1, Canada
Tel: (514) 856-6928 e-mail: [email protected]
FAX: (514) 856-6886
For immediate technical or application assistance:
USA and Canada:
Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA ®
Customer Service: 1-800-622-2378 / 1-800-622-BEST ®
Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN ®
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
Mexico:
En Espan˜ol: (001) 203-359-7803
FAX: (001) 203-359-7807 e-mail:[email protected]
Servicing Europe:
Benelux:
Postbus 8034, 1180 LA Amstelveen, The Netherlands
Tel: +31 (0)20 3472121 FAX: +31 (0)20 6434643
Toll Free in Benelux: 0800 0993344 e-mail: [email protected]
Czech Republic:
Frystatska 184, 733 01 Karviná, Czech Republic
Tel: +420 (0)59 6311899 FAX: +420 (0)59 6311114
Toll Free: 0800-1-66342 e-mail: [email protected]
France:
11, rue Jacques Cartier, 78280 Guyancourt, France
Tel: +33 (0)1 61 37 2900 FAX: +33 (0)1 30 57 5427
Toll Free in France: 0800 466 342 e-mail: [email protected]
Germany/Austria:
Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
Tel: +49 (0)7056 9398-0 FAX: +49 (0)7056 9398-29
Toll Free in Germany: 0800 639 7678 e-mail: [email protected]
United Kingdom:
One Omega Drive, River Bend Technology Centre
ISO 9002 Certified Northbank, Irlam, Manchester
M44 5BD United Kingdom
Tel: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622
Toll Free in United Kingdom: 0800-488-488 e-mail: [email protected]
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the
European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, human applications.
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase.
OMEGA’s WARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and free of defects.
OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS OR IMPLIED, EXCEPT THAT OF
TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE
ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of
OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic
WARRANTY / DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS / INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO
OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER
TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
PATENT NOTICE: U. S. Pat. No. 6,074,089; 5,465,838 / Canada 2,228,333; 2,116,055 / UK GB 2,321,712 / Holland 1008153 / Israel 123052 / France
2 762 908 / EPO 0614194. Other patents pending.
FOR WARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA:
1. Purchase Order number under which the product was PURCHASED,
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems relative to the product.
FOR NON-WARRANTY REPAIRS, consult OMEGA for current repair charges.
Have the following information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the COST of the repair,
2. Model and serial number of the product, and
3. Repair instructions and/or specific problems relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
© Copyright 2004 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
M-4115/0507

Public link updated
The public link to your chat has been updated.
Advertisement