AG Neovo | GM678 | EMC COMPLIANCE TEST REPORT

EMC COMPLIANCE TEST REPORT
Report Number: 030355-E
April 29, 2003
EMC COMPLIANCE TEST REPORT
for
LCD Monitor
Trade Name
Model Number
Serial Number
Report Number
Date
Regulations
:
:
:
:
:
:
COMPAL; AG neovo
GM678; F-417
N/A
030355-E
April 29, 2003
See below
Standards
Results (Pass/Fail)
EN 55022: 1998
PASS
EN 61000-3-2: 1995 + A1: 1998 + A2: 1998
PASS
EN 61000-3-3: 1995
PASS
EN 55024: 1998
PASS
- IEC 61000-4-2: 2001
PASS
- IEC 61000-4-3: 1995
PASS
- IEC 61000-4-4: 1995
PASS
- IEC 61000-4-5: 1995
PASS
- IEC 61000-4-6: 1996
PASS
- IEC 61000-4-8: 1993
PASS
- IEC 61000-4-11: 1994
PASS
Prepared for:
Compal Electronics Inc.
No. 581, Jui Kuang Rd., Neihu,
Taipei, (114) Taiwan, R.O.C.
Prepared by:
C&C LABORATORY, CO., LTD.
No. 81-1, 210 Lane, Pa-de 2nd Road, Lu-Chu Hsiang,
Taoyuan, Taiwan, R.O.C.
TEL: (03) 324-0332
FAX: (03) 324-5235
This report shall not be reproduced, except in full, without the written approval of
C&C LABORATORY, CO., LTD.
The client to claim product certification, approval or endorsement by NVLAP, NIST or any agency
of the federal government must not use this report.
Page 1 of 58
Rev. 00
Report Number: 030355-E
April 29, 2003
EC-Declaration of Conformity
For the following equipment:
LCD Monitor
( Product Name )
GM678; F-417 / COMPAL; AG neovo
( Model Designation / Trade name )
1.) Compal Electronics Inc.
2.) Compal Electronics (China) Co., Ltd.
( Manufacturer Name )
1.)
2.)
No. 8, Nan-Tung Rd., Pin-Cheng City, Tao-Yuan Hsien, Taiwan, R.O.C.
No. 988, Tung Fen East Rd., Economic & Technical Development Zone Kunshan, Jiangsun, P.R. China
( Manufacturer Address )
is herewith confirmed to comply with the requirements set out in the Council Directive on the
Approximation of the Laws of the Member States relating to Electromagnetic Compatibility
Directive (89/336/EEC, Amended by 92/31/EEC, 93/68/EEC & 98/13/EC), For the evaluation
regarding the Electromagnetic Compatibility (89/336/EEC, Amended by 92/31/EEC, 93/68/EEC &
98/13/EC), the following standards are applied:
V
V
V
V
EN 55022: 1998
EN 61000-3-2: 1995 + A1: 1998 + A2: 1998
EN 61000-3-3: 1995
EN 55024: 1998
IEC 61000-4-2: 2001; IEC 61000-4-3: 1995; IEC 61000-4-4: 1995;
IEC 61000-4-5: 1995; IEC 61000-4-6: 1996; IEC 61000-4-8: 1993; IEC 61000-4-11: 1994
The following manufacturer / importer or authorized representative established within the EUT is
responsible for this declaration:
( Company Name )
( Company Address )
Person responsible for making this declaration:
( Name, Surname )
( Position / Title )
( Place )
( Date )
( Legal Signature )
Page 2
Rev. 00
Report Number: 030355-E
April 29, 2003
TABLE OF CONTENTS
1
VERIFICATION OF COMPLIANCE .................................................................................................................................4
2
GENERAL INFORMATION .................................................................................................................................................5
3
SYSTEM DESCRIPTION .......................................................................................................................................................6
4
PRODUCT INFORMATION .................................................................................................................................................7
5
SUPPORT EQUIPMENT ........................................................................................................................................................8
6
TEST FACILITY .......................................................................................................................................................................9
7
TEST EQUIPMENT LIST (EMISSION) ..........................................................................................................................10
8
TEST EQUIPMENT LIST (IMMUMITY) .......................................................................................................................11
9
SECTION 1 EN 55022 (LINE CONDUCTED & RADIATED EMISSION) .........................................................12
9.1
MEASUREMENT PROCEDURE (PRELIMINARY LINE CONDUCTED EMISSION TEST) ............12
9.2
MEASUREMENT PROCEDURE (PRELIMINARY RADIATED EMISSION TEST) .............................14
10 BLOCK DIAGRAM OF TEST SETUP .............................................................................................................................16
11 SUMMARY DATA...............................................................................................................................................................17
12
SECTION 2 EN 61000-3-2 & EN 61000-3-3 (POWER HARMONICS & VOLTAGE FLUCTUATION / FLICKER) .............20
13 SECTION 3 IEC 61000-4-2 (ELECTROSTATIC DISCHARGE) ...........................................................................28
14 SECTION 4 IEC 61000-4-3 (RADIATED ELECTROMAGNETIC FIELD) .......................................................31
15 SECTION 5 IEC 61000-4-4 (FAST TRANSIENTS/BURST) ....................................................................................34
16 SECTION 6 IEC 61000-4-5 (SURGE IMMUNITY) ....................................................................................................36
17 SECTION 7 IEC 61000-4-6 (CONDUCTED DISTRBANCE/INDUCED RADIO-FREQUENCY FIELD) .38
18 SECTION 8 IEC 61000-4-8 (POWER FREQUENCY MAGNETIC FIELD IMMUNITY TEST) .................40
19
SECTION 9 IEC 61000-4-11 (VOLTAGE DIPS, SHORT INTERRUPTIONS AND VOLTAGE VARIATIONS) ........42
20 APPENDIX 1 PHOTOGRAPHS OF TEST SETUP....................................................................................................44
20.1 LINE CONDUCTED EMISSION TEST (EN 55022)...........................................................................................45
20.2 RADIATED EMISSION TEST (EN 55022)............................................................................................................46
20.3 POWER HARMONIC & VOLTAGE FLUCTUATION / FLICKER TEST (EN 61000-3-2, EN 61000-3-3)............47
20.4 ELECTROSTATIC DISCHARGE TEST (IEC 61000-4-2) ................................................................................48
20.5 RADIATED ELECTROMAGNETIC FIELD (IEC 61000-4-3) ........................................................................49
20.6 FAST TRANSIENTS/BURST TEST (IEC 61000-4-4) .........................................................................................50
20.7 SURGE IMMUNITY TEST (IEC 61000-4-5) ........................................................................................................51
20.8 CONDUCTED DISTURBANCE, INDUCED BY RADIO-FREQUENCY FIELDS TEST (IEC 61000-4-6) ...........52
20.9 POWER FREQUENCY MAGNETIC FIELD IMMUNITY TEST (IEC 61000-4-8) ..................................53
20.10 VOLTAGE DIPS / INTERRUPTION TEST (IEC 61000-4-11) .........................................................................54
21 APPENDIX 2 PHOTOGRAPHS OF EUT .....................................................................................................................55
Page 3
Rev. 00
Report Number: 030355-E
April 29, 2003
1 VERIFICATION OF COMPLIANCE
Equipment Under Test:
Trade Name:
Model Number:
Serial Number:
Applicant:
LCD Monitor
COMPAL; AG neovo
GM678; F-417
N/A
Compal Electronics Inc.
No. 581, Jui Kuang Rd., Neihu, Taipei, (114) Taiwan, R.O.C.
Manufacturer:
1.) Compal Electronics Inc.
No. 8, Nan-Tung Rd., Pin-Cheng City, Tao-Yuan Hsien, Taiwan, R.O.C.
2.) Compal Electronics (China) Co., Ltd.
No. 988, Tung Fen East Rd., Economic & Technical Development Zone
Kunshan, Jiangsun, P.R. China
EMC Directive 89/336/EEC for CE Marking
EN 55022: 1998
EN 61000-3-2: 1995 + A1: 1998 + A2: 1998
EN 61000-3-3: 1995
EN 55024: 1998 (IEC 61000-4-2: 2001, IEC 61000-4-3: 1995,
IEC 61000-4-4: 1995, IEC 61000-4-5: 1995,
IEC 61000-4-6: 1996, IEC 61000-4-8: 1993,
IEC 61000-4-11: 1994)
030355-E
File Number:
April 24 ~ 26, 2003
Date of Test:
None
Deviation:
Condition of Test Sample: Normal
Pass
Final Result:
See below
Worst Data:
Test Item
Freq. (MHz)
Measured Data
Margin (MµC)
Remark
Radiated Emission
179.76
27.7 (dB/m)
-2.3 dB (± 3.3498 dB)
Conducted Emission
0.155
42.3 (dB)
-23.4dB (± 2.8104 dB)
The negative sign in Margin cell means under the specific limit.
This test result traceable to national or international standards.
The above equipment was tested by C&C Laboratory Co., Ltd. for compliance with the requirements
set forth in EMC Directive 89/336/EEC and the Technical Standards mentioned above. This said
equipment in the configuration described in this report shows the maximum emission levels
emanating from equipment and the level of the immunity endurance of the equipment are within the
compliance requirements.
Type of Test:
Technical Standards:
The test results of this report relate only to the tested sample identified in this report.
Approved by:
Reviewed by:
Jonson Lee / EMC Director
Susan Su / Section Manager
Page 4
Rev. 00
Report Number: 030355-E
April 29, 2003
2 GENERAL INFORMATION
Applicant:
Compal Electronics Inc.
No. 581, Jui Kuang Rd., Neihu, Taipei, (114) Taiwan, R.O.C.
Contact Person:
Manufacturer:
File Number:
Wilson Pan
1). Compal Electronics Inc.
No. 8, Nan-Tung Rd., Pin-Cheng City, Tao-Yuan Hsien, Taiwan, R.O.C.
2). Compal Electronics (China) Co., Ltd.
No. 988, Tung Fen East Rd., Economic & Technical Development Zone
Kunshan, Jiangsun, P.R. China
030355-E
Date of Test:
April 24 ~ 26, 2003
Equipment Under Test: LCD Monitor
Model Number:
GM678; F-417
Serial Number:
N/A
Type of Test:
Technical Standards:
EMC Directive 89/336/EEC for CE Marking
EN 55022: 1998
EN 61000-3-2: 1995 + A1: 1998 + A2: 1998
EN 61000-3-3: 1995
EN 55024: 1998 (IEC 61000-4-2: 2001, IEC 61000-4-3: 1995,
IEC 61000-4-4: 1995, IEC 61000-4-5: 1995,
IEC 61000-4-6: 1996, IEC 61000-4-8: 1993,
IEC 61000-4-11: 1994)
150kHz to 30MHz for Line Conducted Test
30MHz to 1000MHz for Radiated Emission Test
Frequency Range
(EN 55022):
Test Site:
C&C LABORATORY CO., LTD.
No. 81-1, 210 Lane, Pa-de 2nd Road, Lu-Chu Hsiang,
Taoyuan, Taiwan, R.O.C.
Page 5
Rev. 00
Report Number: 030355-E
April 29, 2003
3 SYSTEM DESCRIPTION
EUT Test Program:
1. EMI test program was loaded and executed in Windows 98 mode.
2. Data was sent to EUT filling the screen with upper case of “H” patterns.
3. Test program sequentially exercised printer and modem, then sent “H” patterns to them
individually.
4. Repeat 2 to 3. Test program is self-repeating throughout the test.
Page 6
Rev. 00
Report Number: 030355-E
April 29, 2003
4 PRODUCT INFORMATION
Housing Type:
Plastic
EUT Power Rating:
100~240VAC, 50 / 60Hz
AC power during Test:
230VAC/ 50Hz
AC Power Cord Type:
Unshielded, 1.8m (Detachable)
OSC/Clock Frequencies:
14.318MHz
LCD Panel Manufacturer:
Hyundai
Model:
HT17E12-200
Power Board Manufacturer:
COMPAL
Model:
VP-719
Main Board Manufacturer:
COMPAL
Model:
VL-720
Key Board Manufacturer:
COMPAL
Model:
VK-718
VGA Cable Type:
Shielded, 1.8m (Non-detachable) with two cores
I/O Port of EUT
I/O Port Type
1). Video Out Port (VGA)
Q’TY
Tested with
1
1
Note: The differences between of two model numbers (list on this report) are identical, just for
marketing purpose only.
Page 7
Rev. 00
Report Number: 030355-E
April 29, 2003
5 SUPPORT EQUIPMENT
No. Equipment
Model
#
Serial
#
FCC
ID
Trade
Name
Data
Cable
Power
Cord
Compaq
N/A
Unshielded, 1.8m
1.
PC
EVO D300
6K1BKF83F18F
FCC DoC
2.
Modem
2400
94-364-176277
DK467GSM24
3.
Printer
EPSON
STYLUS
C20SX
DW4E130540
FCC DoC
EPSON
Shielded, 1.8m
Unshielded, 1.8m
4.
PS/2
Keyboard
SK-2800C
B1C790BCPJCN6L
GYUR79SK
Compaq
Shielded, 1.8m
N/A
M-CAA43
LZA11750827
FCC DoC
Logitech
Shielded, 1.8m
N/A
5. PS/2 Mouse
Computer
Shielded, 1.8m
Peripherals
Unshielded, 1.8m
Note: All the above equipment/cables were placed in worse case positions to maximize emission
signals during emission test.
Grounding: Grounding was in accordance with the manufacturer’s requirements and conditions for
the intended use.
Page 8
Rev. 00
Report Number: 030355-E
April 29, 2003
6 TEST FACILITY
Location:
No. 81-1, 210 Lane, Pa-de 2nd Road, Lu-Chu Hsiang, Taoyuan,
Taiwan, R. O. C.
Description:
There are four 3/10m open area test sites and three line conducted
labs for final test. The Open Area Test Sites and the Line
Conducted labs are constructed and calibrated to meet the FCC
requirements in documents ANSI C63.4: 2001 and CISPR 16
requirements.
Site Filing:
A site description is on file with the Federal Communications
Commission, 7435 Oakland Mills Road, Columbia, MD 21046.
Registration also was made with Voluntary Control Council for
Interference (VCCI).
Site Accreditation:
Accredited by NEMKO (Authorization #: ELA 124) for EMC &
A2LA (Certificate #: 824.01) for Emission
Accredited by NVLAP (Certificate #: 200600-0)
Also accredited by BSMI for the product category of Information
Technology Equipment.
Instrument Tolerance:
All measuring equipment is in accord with ANSI C63.4 and CISPR
22 requirements that meet industry regulatory agency and
accreditation agency requirement.
Ground Plane:
Two conductive reference ground planes were used during the Line Conducted
Emission, one in vertical and the other in horizontal. The dimensions of these ground planes are
as below. The vertical ground plane was placed distancing 40 cm to the rear of the wooden test
table on where the EUT and the support equipment were placed during test. The horizontal
ground plane projected 50 cm beyond the footprint of the EUT system and distanced 80 cm to the
wooden test table. For Radiated Emission Test, one horizontal conductive ground plane extended
at least 1m beyond the periphery of the EUT and the largest measuring antenna, and covered the
entire area between the EUT and the antenna. It has no holes or gaps having longitudinal
dimensions larger than one-tenth of a wavelength at the highest frequency of measurement up to
1GHz.
Site # 3 & # 4 Line Conducted Test Site: At Shielding Room
Page 9
Rev. 00
Report Number: 030355-E
April 29, 2003
7 TEST EQUIPMENT LIST (EMISSION)
Instrumentation: The following list contains equipment used at C & C Laboratory, Co., Ltd. for
testing. The equipment conforms to the CISPR 16-1 / ANSI C63.2-1988 Specifications for
Electromagnetic Interference and Field Strength Instrumentation from 9kHz to 1.0 / 2.0 GHz.
Equipment used during the tests:
Open Area Test Site:
#1
Open Area Test Site # 1
EQUIPMENT
TYPE
Spectrum Analyzer
S.P.A Display
Q.P Adaptor
RF Pre-selector
Spectrum Analyzer
Pre-Amplifier
Bilog Antenna
Turn Table
Antenna Tower
Controller
RF Switch
Site NSA
MFR
HP
HP
HP
HP
Anritsu
HP
CHASE
EMCO
EMCO
EMCO
ANRITSU
C&C
Conducted Emission Test Site:
MODEL
NUMBER
8568B
85662A
85650A
85685A
MS2601A
8447D
CBL6112A
2081-1.21
2075-2
2090
MP59B
N/A
SERIAL
NUMBER
3001A05004
3014A18846
2811A01399
2947A01064
MT09950
2944A08432
2309
N/A
9707-2604
N/A
M54367
N/A
LAST
CAL.
07/03/2002
07/03/2002
07/03/2002
07/03/2002
N/A
N/A
02/28/2003
N.C.R
N.C.R
N.C.R
N.C.R
08/31/2002
CAL.
DUE
07/02/2003
07/02/2003
07/02/2003
07/02/2003
N/A
N/A
02/27/2004
N.C.R
N.C.R
N.C.R
N.C.R
08/30/2003
#3
Conducted Emission Test Site # 3
EQUIPMENT
TYPE
EMI Test Receiver
LISN
LISN
MFR
R&S
R&S
EMCO
MODEL
NUMBER
ESHS30
ENV 4200
3825/2
SERIAL
NUMBER
828144/003
830326/016
9003/1382
LAST
CAL.
CAL.
DUE
08/08/2002 08/07/2003
03/05/2003 03/04/2004
02/26/2003 02/25/2004
Power Harmonic & Voltage Fluctuation/Flicker Measurement:
Power Harmonic & Voltage Fluctuation/Flicker Measurement (61000-3-2&-3-3)
EQUIPMENT
TYPE
Harmonic & Flicker
Tester
MFR
MODEL
NUMBER
SERIAL
NUMBER
HAEFELY
TRENCH
PHF555
080 419-25
LAST
CAL.
CAL DUE.
10/14/2002 10/13/2003
The calibrations of the measuring instruments, including any accessories that may effect such
calibration, are checked frequently to assure their accuracy. Adjustments are made and correction
factors applied in accordance with instructions contained in the manual for the measuring
instrument.
Page 10
Rev. 00
Report Number: 030355-E
April 29, 2003
8 TEST EQUIPMENT LIST (IMMUMITY)
ESD test (61000-4-2)
EQUIPMENT
TYPE
MFR
MODEL
NUMBER
SERIAL
NUMBER
LAST
CAL.
CAL DUE.
ESD Generator
EM TEST
P30C-RFCI
0603-011
02/27/2003
02/26/2004
Radiated Electromagnetic Field immunity Measurement (61000-4-3)
EQUIPMENT
TYPE
MFR
MODEL
NUMBER
SERIAL
NUMBER
LAST
CAL.
CAL DUE.
S.G.
R&S
SMY02
100094
08/08/2002
08/07/2003
Power Amplifier
ar
150W1000
300300
N/A
N/A
Power Antenna
EMCO
93141
9712-1083
N/A
N/A
EM PROBE
GW
EMR-30
L-0013
05/23/2002
05/22/2003
Fast Transients/Burst test (61000-4-4)
EQUIPMENT
TYPE
Fast Transients/Burst
Generator
MFR
MODEL
NUMBER
SERIAL
NUMBER
LAST
CAL.
CAL DUE.
HAEFELY
TRENCH
PEFT- JUNIOR
583 333-117
08/22/2002
08/21/2003
Surge Immunity test (61000-4-5)
EQUIPMENT
TYPE
MFR
MODEL
NUMBER
SERIAL
NUMBER
LAST
CAL.
CAL DUE.
Surge Tester
HAEFELY
TRENCH
PSUGER 4010
583 334-71
09/03/2002
09/02/2003
CS test (61000-4-6)
EQUIPMENT
TYPE
MFR
MODEL
NUMBER
SERIAL
NUMBER
LAST
CAL.
CAL DUE.
S.G.
R&S
SMY02
100094
08/08/2002
08/07/2003
Power Amplifier
ar
500A100A
300299
N/A
N/A
CDN
Lüthi
801-M3
1879
02/26/2003
02/25/2004
CDN
MEB
M2
A3002010
04/24/2002
04/23/2003
Power Frequency Magnetic Field Immunity test (61000-4-8)
EQUIPMENT
TYPE
TRIAX ELF Magnetic
Field Meter
Magnetic Field Tester
MFR
MODEL
NUMBER
SERIAL
NUMBER
LAST
CAL.
CAL DUE.
F.W.BELL
4090
9711
10/21/2002
10/20/2003
HAEFELY
TRENCH
MAG 100.1
080 938-01
N/A
N/A
Voltage Dips/Short Interruption and Voltage Variation Immunity test (61000-4-11)
EQUIPMENT
TYPE
Dips/Interruption and
Variations Simulator
MFR
MODEL
NUMBER
SERIAL
NUMBER
LAST
CAL.
CAL DUE.
HAEFELY
TRENCH
PLINE 1610
080 344-05
03/28/2003
03/27/2004
Page 11
Rev. 00
Report Number: 030355-E
April 29, 2003
9
9.1
SECTION 1 EN 55022 (LINE CONDUCTED & RADIATED EMISSION)
MEASUREMENT PROCEDURE (PRELIMINARY LINE CONDUCTED EMISSION TEST)
1) The equipment was set up as per the test configuration to simulate typical actual usage per the
user’s manual. When the EUT is a tabletop system, a wooden table with a height of 0.8 meters
is used and is placed on the ground plane as per EN 55022 (see Test Facility for the dimensions
of the ground plane used).
When the EUT is a floor-standing equipment, it is placed on the
ground plane which has a 3-12 mm non-conductive covering to insulate the EUT from the
ground plane.
2) Support equipment, if needed, was placed as per EN 55022.
3) All I/O cables were positioned to simulate typical actual usage as per EN 55022.
4) The EUT received AC power through a Line Impedance Stabilization Network (LISN) which
supplied power source and was grounded to the ground plane.
5) All support equipment received power from a second LISN supplying power of 110VAC/60Hz,
if any.
6) The EUT test program was started. Emissions were measured on each current carrying line of
the EUT using a spectrum analyzer / Receiver connected to the LISN powering the EUT. The
LISN has two monitoring points: Line 1 (Hot Side) and Line 2 (Neutral Side). Two scans
were taken: one with Line 1 connected to Analyzer / Receiver and Line 2 connected to a 50
ohm load; the second scan had Line 1 connected to a 50 ohm load and Line 2 connected to the
Analyzer / Receiver.
7) Analyzer / Receiver scanned from 150kHz to 30MHz for emissions in each of the test modes.
8) During the above scans, the emissions were maximized by cable manipulation.
9) The following test mode(s) were scanned during the preliminary test:
Mode(s):
1. 1280 × 1024 Resolution 75Hz
2. 1024 × 768 Resolution 75Hz
3. 800 × 600 Resolution 75Hz
10) After the preliminary scan, we found the following test mode producing the highest emission
level.
Mode:
1.
Then, the EUT configuration and cable configuration of the above highest emission level were
recorded for reference of final testing.
Page 12
Rev. 00
Report Number: 030355-E
April 29, 2003
MEASUREMENT PROCEDURE (FINAL LINE CONDUCTED EMISSION TEST)
1) EUT and support equipment was set up on the test bench as per step 10 of the preliminary test.
2) A scan was taken on both power lines, Line 1 and Line 2, recording at least the six highest
emissions. Emission frequency and amplitude were recorded into a computer in which
correction factors were used to calculate the emission level and compare reading to the
applicable limit. If EUT emission level was less –2dB to the A.V. limit in Q.P. mode, then the
emission signal was re-checked using an A.V. detector.
3) The test data of the worst case condition(s) was reported on the Summary Data page.
Data Sample:
Q.P.
Raw
dBuV
43.95
Freq.
MHz
x.xx
Average
Raw
dBuV
---
Freq.
Raw dBuV
Q.P.
Limit
dBuV
56
Average
Limit
dBuV
46
Q.P.
Margin
dB
-12.05
Average
Margin
dB
---
Note
L1
= Emission frequency in MHz
= Uncorrected Analyzer/Receiver reading +
Insertion loss of LISN, if it > 0.5 dB
= Limit stated in standard
= Reading in reference to limit
= Current carrying line of reading
= The emission level complied with the Average
limits, with at least 2dB margin limits, so no
further recheck.
Limit dBuV
Margin dB
Note
“---“
Calculation example:
Margin (dB) = RAW (dBuV) – Limit (dBuV)
LINE CONDUCTED EMISSION LIMIT (EN 55022)
Frequency
Maximum RF Line Voltage
Q.P.
AVERAGE
150kHz-500kHz
66-56dBuV
56-46dBuV
500kHz-5MHz
56dBuV
46dBuV
5MHz-30MHz
60dBuV
50dBuV
Note: The lower limit shall apply at the transition frequency.
Page 13
Rev. 00
Report Number: 030355-E
April 29, 2003
9.2 MEASUREMENT PROCEDURE (PRELIMINARY RADIATED EMISSION TEST)
1) The equipment was set up as per the test configuration to simulate typical actual usage per the
user’s manual. When the EUT is a tabletop system, a wooden turntable with a height of 0.8
meters is used which is placed on the ground plane as per EN 55022 (see Test Facility for the
dimensions of the ground plane used).
When the EUT is a floor-standing equipment, it is
placed on the ground plane which has a 3-12 mm non-conductive covering to insulate the EUT
from the ground plane.
2) Support equipment, if needed, was placed as per EN 55022.
3) All I/O cables were positioned to simulate typical actual usage as per EN 55022.
4) The EUT received AC power source from the outlet socket under the turntable. All support
equipment received 110VAC/60Hz power from another socket under the turntable, if any.
5) The antenna was placed at 10 meter away from the EUT as stated in EN 55022. The antenna
connected to the Analyzer via a cable and at times a pre-amplifier would be used.
6) The Analyzer / Receiver quickly scanned from 30MHz to 1000MHz. The EUT test program
was started. Emissions were scanned and measured rotating the EUT to 360 degrees and
positioning the antenna 1 to 4 meters above the ground plane, in both the vertical and the
horizontal polarization, to maximize the emission reading level.
7) The following test mode(s) were scanned during the preliminary test:
Mode(s):
1. 1280 × 1024 Resolution 75Hz
2. 1024 × 768 Resolution 75Hz
3. 800 × 600 Resolution 75Hz
8)
After the preliminary scan, we found the following test mode producing the highest emission
level.
Mode:
1.
Then, the EUT and cable configuration, antenna position, polarization and turntable position of the
above highest emission level were recorded for final testing.
Page 14
Rev. 00
Report Number: 030355-E
April 29, 2003
MEASUREMENT PROCEDURE (FINAL RADIATED EMISSION TEST)
1) EUT and support equipment were set up on the turntable as per step 8 of the preliminary test.
2) The Analyzer / Receiver scanned from 30MHz to 1000MHz. Emissions were scanned and
measured rotating the EUT to 360 degrees, varying cable placement and positioning the antenna
1 to 4 meters above the ground plane, in both the vertical and the horizontal polarization, to
maximize the emission reading level.
3) Recorded at least the six highest emissions. Emission frequency, amplitude, antenna position,
polarization and turntable position were recorded into a computer in which correction factors
were used to calculate the emission level and compare reading to the applicable limit and only
Q.P. reading is presented.
4) The test data of the worst case condition(s) was reported on the Summary Data page.
Data Sample:
=======================================================================
Freq.
Raw
Corr.
Emiss.
Limits
Margin
Data
Factor
Level
(MHz)
(dBuV)
(dB)
(
dBuV/m )
(dB)
=======================================================================
xx.xx
14.0
11.2
26.2
30
-3.8
======================================================================
Freq.
Raw Data (dBuV)
Corr. Factor (dB)
Emiss. Level (dBuV/m)
Limit (dBuV/m)
Margin (dB)
P
Q
A
= Emission frequency in MHz
= Uncorrected Analyzer / Receiver reading
= Antenna factor + Cable loss – Amplifier gain
= Raw reading converted to dBuV/m and CF added
= Limit stated in standard
= Reading in reference to limit
= Peak Reading
= Quasi-peak Reading
= Average Reading
Calculation example:
Margin (dB) = Emiss. Level (dBuV/m) – Limits (dBuV/m)
Emission Level (dBuV/m)=Raw Data (dBuV) + Corr Factor (dB)
RADIATED EMISSION LIMIT
Frequency
(MHz)
Distance
(m)
30-230
10
Maximum Field Strength Limit
(dBu V/m/ Q.P.)
30
230-1000
10
37
Note: The lower limit shall apply at the transition frequency.
Page 15
Rev. 00
Report Number: 030355-E
April 29, 2003
10 BLOCK DIAGRAM OF TEST SETUP
System Diagram of Connections between EUT and Simulators
EUT: LCD Monitor
Trade Name: COMPAL
Model Number: GM678
Power Cord: Unshielded, 1.8m
VGA Cable
LCD Monitor
1. PC
(EUT)
2. Modem
4. PS/2 Keyboard
3. Printer
5. PS/2 Mouse
Page 16
Rev. 00
Report Number: 030355-E
April 29, 2003
11 SUMMARY DATA
(LINE CONDUCTED TEST)
Model Number: GM678
Location: Site # 3
Tested by: Hank Huang
Test Mode: Mode 1
Test Results: Passed
Temperature: 20oC
Humidity: 65%RH
(The chart below shows the highest readings taken from the final data)
FREQ
Q.P.
AVG
Q.P.
AVG
Q.P.
AVG
NOTE
MHz
RAW
RAW
Limit
Limit
M argin
M argin
0.155
2.301
2.433
17.448
18.369
19.350
dBuV
41.00
32.40
30.20
31.60
34.00
31.40
dBuV
-------------
dBuV
65.70
56.00
56.00
60.00
60.00
60.00
dBuV
55.70
46.00
46.00
50.00
50.00
50.00
dB
-24.70
-23.60
-25.80
-28.40
-26.00
-28.60
dB
-------------
L1
L1
L1
L1
L1
L1
0.155
2.172
2.435
6.307
9.150
25.843
42.30
32.10
31.00
26.50
26.20
25.70
-------------
65.70
56.00
56.00
60.00
60.00
60.00
55.70
46.00
46.00
50.00
50.00
50.00
-23.40
-23.90
-25.00
-33.50
-33.80
-34.30
-------------
L2
L2
L2
L2
L2
L2
L1 = Line One (Hot side) / L2 = Line Two (Neutral side)
**NOTE: “---” denotes the emission level was or more than 2dB below the Average limit,
so no re-check anymore.
Page 17
Rev. 00
Report Number: 030355-E
April 29, 2003
SUMMARY DATA
(RADIATED EMISSION TEST)
Model Number: GM678
Location: Site # 1
Tested by: Hank Huang
Polar: Vertical--10m
Test Mode: Mode 1
Test Results: Passed
Detector Function: Quasi-Peak
Temperature: 21°C
Humidity: 65%RH
(The chart below shows the highest readings taken from the final data)
==============================================================
Freq.
Raw
Corr.
Emiss.
Limits
Margin
Data
Factor
Level
(MHz)
(dBuV)
(dB)
(
dBuV/m
)
(dB)
==============================================================
78.75
15.5
6.8
22.3
30.0
-7.7
--------------------------------------------------------------------------------------------------------157.29
16.3
10.2
26.5
30.0
-3.5
--------------------------------------------------------------------------------------------------------168.08
15.1
10.5
25.6
30.0
-4.4
--------------------------------------------------------------------------------------------------------179.76
16.2
11.5
27.7
30.0
-2.3
--------------------------------------------------------------------------------------------------------200.80
12.2
10.8
23.0
30.0
-7.0
--------------------------------------------------------------------------------------------------------315.15
18.1
16.2
34.3
37.0
-2.7
--------------------------------------------------------------------------------------------------------571.17
7.5
22.3
29.8
37.0
-7.2
--------------------------------------------------------------------------------------------------------614.42
10.6
22.5
33.1
37.0
-3.9
---------------------------------------------------------------------------------------------------------
Page 18
Rev. 00
Report Number: 030355-E
April 29, 2003
SUMMARY DATA
(RADIATED EMISSION TEST)
Model Number: GM678
Location: Site # 1
Tested by: Hank Huang
Polar: Horizontal--10m
Test Mode: Mode 1
Test Results: Passed
Detector Function: Quasi-Peak
Temperature: 21°C
Humidity: 65%RH
(The chart below shows the highest readings taken from the final data)
==============================================================
Freq.
Raw
Corr.
Emiss.
Limits
Margin
Data
Factor
Level
(MHz)
(dBuV)
(dB)
(
dBuV/m
)
(dB)
==============================================================
157.85
12.1
10.1
22.2
30.0
-7.8
-------------------------------------------------------------------------------------------------------189.54
9.6
11.2
20.8
30.0
-9.2
-------------------------------------------------------------------------------------------------------215.30
13.4
10.5
23.9
30.0
-6.1
-------------------------------------------------------------------------------------------------------271.80
11.8
15.9
27.7
37.0
-9.3
-------------------------------------------------------------------------------------------------------501.60
8.4
21.2
29.6
37.0
-7.4
-------------------------------------------------------------------------------------------------------577.20
6.2
22.3
28.5
37.0
-8.5
-------------------------------------------------------------------------------------------------------610.80
10.0
22.4
32.4
37.0
-4.6
-------------------------------------------------------------------------------------------------------865.60
5.1
28.3
33.4
37.0
-3.6
--------------------------------------------------------------------------------------------------------
Page 19
Rev. 00
Report Number: 030355-E
April 29, 2003
12 SECTION 2 EN 61000-3-2 & EN 61000-3-3 (POWER HARMONICS &
VOLTAGE FLUCTUATION / FLICKER)
POWER HARMONICS MEASUREMENT
Port
Basic Standard
: AC mains
: EN 61000-3-2 (1995 + A1: 1998 + A2: 1998)
Limits
Tester
Temperature
Humidity
:
:
:
:
V CLASS A ;
CLASS D
Lung Tsai
25oC
51%
VOLTAGE FLUCTUATION/FLICKER MEASUREMENT
Port
Basic Standard
Limits
: AC mains
: EN 61000-3-3 (1995)
: § 5 of EN 61000-3-3
Tester
Temperature
Humidity
: Lung Tsai
: 25oC
: 51%
Block Diagram of Test Setup:
Harmonics & Flicker
Analyzer
+
Power Source
( PF 555 )
Power cord
EUT
Support Units
0.8m
Result:
Please see the attached test data.
Page 20
Rev. 00
Report Number: 030355-E
April 29, 2003
---------------------------------------------------------------------EN 61000-3-2 TEST REPORT 2003/4/26 11:19 AM
---------------------------------------------------------------------Unit:
LCD Monitor
Model No.:
GM678
Remarks:
Temp: 25oC
Operator:
Lung Tsai
Humid: 51%
=====================================================================
TEST SETUP
----------------Test Freq.:
Waveform :
Classification :
50.00 Hz.
SINE
CLASS A
Prog. Zo Enabled:
YES
Test Voltage:
Test Time:
Test Type:
230.0 vac
2.5 min.
STEADY-STATE
Prog. Zo:
0.000
Motor Driven with Phase Angle Control:
Impedance selected:
Synthetic R+L Enabled:
Resistance: 0.380 Ohms
NO
DIRECT
NO
Inductance: 460.000 uH
MAX WATTS: 37.1W
Page 21
Rev. 00
Report Number: 030355-E
April 29, 2003
TEST DATA
----------------
Result:
PASS
Harmonic Current Results
------------------------------------------Hn
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
AMPS LO Limit
0.000
0.162
0.001
0.134
0.001
0.121
0.001
0.109
0.001
0.094
0.001
0.078
0.001
0.061
0.001
0.044
0.000
0.029
0.000
0.017
0.000
0.000
NaN
1.080
2.300
0.430
1.140
0.300
0.770
0.230
0.400
0.184
0.330
0.153
0.210
0.131
0.150
0.115
0.132
0.102
0.118
0.092
HI Limit
0.000
NaN
1.080
2.300
0.430
1.140
0.300
0.770
0.230
0.400
0.184
0.330
0.153
0.210
0.131
0.150
0.115
0.132
0.102
0.118
0.092
Result
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
Page 22
Rev. 00
Report Number: 030355-E
April 29, 2003
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
0.009
0.000
0.008
0.000
0.011
0.000
0.012
0.000
0.011
0.000
0.010
0.000
0.007
0.000
0.004
0.000
0.002
0.000
0.003
0.000
0.107
0.084
0.098
0.077
0.090
0.071
0.083
0.066
0.078
0.061
0.073
0.058
0.068
0.054
0.064
0.051
0.061
0.048
0.058
0.046
0.107
0.084
0.098
0.077
0.090
0.071
0.083
0.066
0.078
0.061
0.073
0.058
0.068
0.054
0.064
0.051
0.061
0.048
0.058
0.046
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
END OF REPORT
Page 23
Rev. 00
Report Number: 030355-E
April 29, 2003
---------------------------------------------------------------------EN 61000-3-3 TEST REPORT 2003/4/26 11:34 AM
---------------------------------------------------------------------Unit:
LCD Monitor
Model No.:
GM678 (Continue)
Remarks:
Temp: 25oC
Operator:
Lung Tsai
Humid: 51%
=====================================================================
TEST SETUP
----------------Test Freq.:
Waveform :
Test Time:
Prog. Zo Enabled:
50.00 Hz.
SINE
10.0 min.
YES
Test Voltage:
230.0 vac
Tshort:
10.0 min.
Prog. Zo:
Voltage Change less than once per Hour:
Impedance selected:
Synthetic R+L Enabled:
Resistance: 0.380 Ohms
0.000
NO
DIRECT
NO
Inductance: 460.000 uH
Page 24
Rev. 00
Report Number: 030355-E
April 29, 2003
TEST DATA
----------------
Result:
PASS
EUT Data
Pst max
Plt max
dc %
dmax %
d(t) sec.
Source Pst max
% THD
0.001
0.001
0.00
0.00
0.00
Power Source Data
0.020
0.03
Limit
Result
Test Enabled
1.00
0.65
3.00
4.00
0.20
PASS
PASS
PASS
PASS
PASS
true
true
true
true
true
0.400
3.00
PASS
PASS
true
true
END OF REPORT
Page 25
Rev. 00
Report Number: 030355-E
April 29, 2003
---------------------------------------------------------------------EN 61000-3-3 TEST REPORT 2003/4/26 11: 47 PM
---------------------------------------------------------------------Unit:
LCD Monitor
Model No.:
GM678 (Manual Switch)
Remarks:
Temp: 25oC
Operator:
Lung Tsai
Humid: 51%
=====================================================================
TEST SETUP
----------------Test Freq.:
Waveform :
Test Time:
Prog. Zo Enabled:
50.00 Hz.
SINE
10.0 min.
YES
Test Voltage:
230.0 vac
Tshort:
10.0 min.
Prog. Zo:
0.000
Voltage Change less than once per Hour:
Impedance selected:
Synthetic R+L Enabled:
Resistance: 0.380 Ohms
NO
DIRECT
NO
Inductance: 460.000 uH
Page 26
Rev. 00
Report Number: 030355-E
April 29, 2003
TEST DATA
----------------
Result:
PASS
EUT Data
Pst max
Plt max
dc %
dmax %
d(t) sec.
Source Pst max
% THD
0.001
0.001
0.00
0.00
0.00
Power Source Data
0.020
0.03
Limit
Result
Test Enabled
1.00
0.65
3.00
4.00
0.20
PASS
PASS
PASS
PASS
PASS
true
true
true
true
true
0.400
3.00
PASS
PASS
true
true
END OF REPORT
Page 27
Rev. 00
Report Number: 030355-E
April 29, 2003
13 SECTION 3 IEC 61000-4-2 (ELECTROSTATIC DISCHARGE)
ELECTROSTATIC DISCHARGE (ESD) IMMUNITY TEST
Port
Basic Standard
Test Level
: Enclosure
: IEC 61000-4-2
: ± 8 kV (Air Discharge)
± 4 kV (Contact Discharge)
± 4 kV (Indirect Discharge)
Performance Criteria
Tester
Temperature
Humidity
Pressure
:
:
:
:
:
B (Standard Require)
Lung Tsai
23oC
46%
1018mbar
Block Diagram of Test Setup:
( The 470 k ohm resistors are installed per standard requirement )
VCP
EUT & its
direct
Support Units
Indirect
Support units
HCP
>1m
Wooden Table
0.8m
Ground Reference Plane
Page 28
Rev. 00
Report Number: 030355-E
April 29, 2003
Test Procedure:
1. The EUT was located 0.1 m minimum from all side of the HCP.
2. The indirect support units were located 1 m minimum away from the EUT, but direct support unit
was/were located at same location as EUT on the HCP and keep at a distance of 10 cm with EUT.
3. A scroll ‘H’ test program was loaded and executed in Windows 98 mode.
4. The Host PC sent above message to EUT and related peripherals through the test.
5. Active the communication function if the EUT with such port(s).
6. As per the requirement of EN 55024; applying direct contact discharge at the sides other than
front of EUT at minimum 50 discharges (25 positive and 25 negative) if applicable, can’t be
applied direct contact discharge side of EUT then the indirect discharge shall be applied. One of
the test points shall be subjected to at least 50 indirect discharge (contact) to the front edge of
horizontal coupling plane.
7. Other parts of EUT where it is not possible to perform contact discharge then selecting
appropriate points of EUT for air discharge, a minimum of 10 single air discharges shall be applied.
8. The application of ESD to the contact of open connectors is not required.
9. The EUT direct connection units also need to be applied ESD at the port of EUT cable connected.
10. Putting a mark on EUT to show tested points. The following test condition was followed during the tests.
Note: As per IEC 61000-4-2:2001, with two 470k bleed resistors cable is connected between the
EUT and HCP during the test applicable for power ungrounded or battery operating unit only.
The electrostatic discharges were applied as follows:
Amount of Discharges
Voltage
Coupling
Result (Pass/Fail)
Mini 10 /Point
± 8 kV
Air Discharge
Pass
Mini 25 /Point
± 4 kV
Contact Discharge
Pass
Mini 25 /Point
± 4 kV
Indirect Discharge HCP (Front)
Pass
Mini 25 /Point
± 4 kV
Indirect Discharge VCP (Right)
Pass
Mini 25 /Point
± 4 kV
Indirect Discharge VCP (Left)
Pass
Mini 25 /Point
± 4 kV
Indirect Discharge VCP (Back)
N/A
**The tested points to EUT, please refer to attached page.
(Blue arrow mark for Contact Discharge and red arrow mark for Air Discharge)
Performance & Result:
V Criteria A: The apparatus continues to operate as intended. No degradation of performance
V
V
or loss of function is allowed below a performance level specified by the
manufacturer, when the apparatus is used as intended. In some cases the
performance level may be replaced by a permissible loss of performance.
Criteria B: The apparatus continues to operate as intended after the test. No degradation of
performance or loss of function is allowed below a performance level specified
by the manufacturer, when the apparatus is used as intended. In some cases the
performance level may be replaced by a permissible loss of performance.
During the test, degradation of performance is however allowed.
Criteria C: Temporary loss of function is allowed, provided the functions self recoverable or
can be restored by the operation of controls.
V
PASS
FAILED
Observation: No any function degraded during the tests.
Page 29
Rev. 00
Report Number: 030355-E
April 29, 2003
The Tested Points of EUT
Photo 1 of 2
Photo 2 of 2
Page 30
Rev. 00
Report Number: 030355-E
April 29, 2003
14 SECTION 4 IEC 61000-4-3 (RADIATED ELECTROMAGNETIC FIELD)
RADIATED ELECTROMAGNETIC FIELD IMMUNITY TEST
Port
Basic Standard
Requirements
Performance Criteria
Tester
Temperature
Humidity
Pressure
:
:
:
:
:
:
:
:
Enclosure
IEC 61000-4-3
3 V/m / with 80% AM. 1kHz Modulation.
A (Standard Require)
Lung Tsai
23oC
46%
1018mbar
Block Diagram of Test Setup:
3 meter
7x3x3 Chamber
EUT &
Support Units
1.5 meter
0.8m
Power Amp
Signal
EUT Monitoring by
Generator
using a camera
PC Controller to control
S.G. & PA as well as
forward power
Control Room
Page 31
Rev. 00
Report Number: 030355-E
April 29, 2003
Test Procedure:
1. The EUT was located at the edge of supporting table keep 3 meter away from transmitting
antenna, it just the calibrated square area of field uniformity. The support units were located
outside of the uniformity area, but the cable(s) connected with EUT were exposed to the
calibrated field as per IEC 61000-4-3.
2. A scroll ‘H’ messages were displayed on part of screen of EUT and an enlarged ‘H’ characters
were displayed on the other part of screen of EUT.
3. Adjusting the monitoring camera to monitor the ‘H’ message as clear as possible.
4. Setting the testing parameters of RS test software per IEC 61000-4-3.
5. Performing the pre-test at each side of with double specified level (6V/m) at 4% steps.
6. From the result of pre-test in step 5, choice the worst side of EUT for final test from 80 MHz to
1000 MHz at 1% steps.
7. Recording the test result in following table.
8. It is not necessary to perform test as per annex A of EN 55024 if the EUT doesn’t belong to ITE
product.
IEC 61000-4-3 Preliminary test conditions:
Test level
: 6V/m
Steps
: 4 % of fundamental
Dwell Time
: 3 sec
Range (MHz)
80-1000
80-1000
80-1000
80-1000
80-1000
80-1000
80-1000
80-1000
Field
6V/m
6V/m
6V/m
6V/m
6V/m
6V/m
6V/m
6V/m
Modulation
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Polarity
H
V
H
V
H
V
H
V
Position (°) Result (Pass/Fail)
Front
Pass
Front
Pass
Right
Pass
Right
Pass
Back
Pass
Back
Pass
Left
Pass
Left
Pass
Polarity
H
V
Position (°) Result (Pass/Fail)
Back
Pass
Back
Pass
IEC 61000-4-3 Final test conditions:
Test level
: 3V/m
Steps
: 1 % of fundamental
Dwell Time
: 3 sec
Range (MHz)
80-1000
80-1000
Field
3V/m
3V/m
Modulation
Yes
Yes
Page 32
Rev. 00
Report Number: 030355-E
April 29, 2003
Performance & Result:
V Criteria A: The apparatus continues to operate as intended. No degradation of performance
V
V
or loss of function is allowed below a performance level specified by the
manufacturer, when the apparatus is used as intended. In some cases the
performance level may be replaced by a permissible loss of performance.
Criteria B: The apparatus continues to operate as intended after the test. No degradation of
performance or loss of function is allowed below a performance level specified
by the manufacturer, when the apparatus is used as intended. In some cases the
performance level may be replaced by a permissible loss of performance.
During the test, degradation of performance is however allowed.
Criteria C: Temporary loss of function is allowed, provided the functions self recoverable or
can be restored by the operation of controls.
V
PASS
FAILED
Observation: No any function degraded during the tests.
Page 33
Rev. 00
Report Number: 030355-E
April 29, 2003
15 SECTION 5 IEC 61000-4-4 (FAST TRANSIENTS/BURST)
FAST TRANSIENTS/BURST IMMUNITY TEST
Port
Basic Standard
: On Power Supply Lines
: IEC 61000-4-4
Requirements
Performance Criteria
Tester
Temperature
Humidity
Pressure
:
:
:
:
:
:
± 1 kV for Power Supply Line
B (Standard Require)
Lung Tsai
23oC
46%
1018mbar
Block Diagram of Test Setup:
EUT
AC Line
EFT/Burst/Surge
Generator
0.8m
Support Units
Non-Conductive Table
Controller Computer
Page 34
Rev. 00
Report Number: 030355-E
April 29, 2003
Test Procedure:
1. The EUT and support units were located on a wooden table 0.8 m away from ground reference
plane.
2. A 1.0 meter long power cord was attached to EUT during the test.
3. The length of communication cable between communication port and clamp was keeping within
1 meter.
4. A test program was loaded and executed in Windows 98 mode.
5. The data was sent to EUT filling the screens with upper case of “H” patterns.
6. The test program exercised related support units sequentially.
7. Repeating step 3 to 6 through the test and increase test voltage to the EUT ports from minimum
to standard request or client request.
8. Recording the test result as shown in following table.
Test conditions:
Impulse Frequency
Tr/Th
Burst Duration
Burst Period
Inject Line
L
N
PE
L+N
L + PE
N + PE
L1 + N + PE
: 5kHz
: 5/50ns
: 15ms
: 3Hz
Voltage kV
±1
±1
±1
±1
±1
±1
±1
Inject Method
Direct
Direct
Direct
Direct
Direct
Direct
Direct
Result (Pass/Fail)
Pass
Pass
Pass
Pass
Pass
Pass
Pass
Performance & Result:
V Criteria A: The apparatus continues to operate as intended. No degradation of performance
V
V
or loss of function is allowed below a performance level specified by the
manufacturer, when the apparatus is used as intended. In some cases the
performance level may be replaced by a permissible loss of performance.
Criteria B: The apparatus continues to operate as intended after the test. No degradation of
performance or loss of function is allowed below a performance level specified
by the manufacturer, when the apparatus is used as intended. In some cases the
performance level may be replaced by a permissible loss of performance.
During the test, degradation of performance is however allowed.
Criteria C: Temporary loss of function is allowed, provided the functions self recoverable or
can be restored by the operation of controls.
V
PASS
FAILED
Observation: No any function degraded during the tests.
Page 35
Rev. 00
Report Number: 030355-E
April 29, 2003
16 SECTION 6 IEC 61000-4-5 (SURGE IMMUNITY)
SURGE IMMUNITY TEST
Port
Basic Standard
: Power Cord
: IEC 61000-4-5
: ± 1 kV (Line to Line)
± 2 kV (Line to Ground)
Performance Criteria : B (Standard Require)
Tester
: Lung Tsai
Temperature
: 23oC
Humidity
: 46%
Pressure
: 1018mbar
Requirements
Block Diagram of Test Setup:
To AC
Source
Surge Immunity
Test
EUT
&
Support Units
0.8m
Controller Computer
Page 36
Rev. 00
Report Number: 030355-E
April 29, 2003
Test Procedure:
1. The EUT and support units were located on a wooden table 0.8 m away from ground floor.
2. A test program was loaded and executed in Windows 98 mode.
3. The data was sent to EUT filling the screens with upper case of “H” patterns.
4. The test program exercised related support units sequentially.
5. Repeating step 3 to 4 through the test and increase test voltage to the EUT ports from minimum
to standard request or client request.
6. Recording the test result as shown in following table.
Test conditions:
Voltage Waveform
Current Waveform
Polarity
Phase angle
Number of Test
Coupling Line
L1-L2
L1-PE
L2-PE
L1-L2
L1-PE
L2-PE
: 1.2/50 us
: 8/20 us
: Positive/Negative
: 0o, 90o, 270o
:5
Voltage (kV)
1
2
2
1
2
2
Polarity
Positive
Positive
Positive
Negative
Negative
Negative
Coupling Method Result (Pass/Fail)
Capacitive
Pass
Capacitive
Pass
Capacitive
Pass
Capacitive
Pass
Capacitive
Pass
Capacitive
Pass
Performance & Result:
V Criteria A: The apparatus continues to operate as intended. No degradation of performance
V
V
or loss of function is allowed below a performance level specified by the
manufacturer, when the apparatus is used as intended. In some cases the
performance level may be replaced by a permissible loss of performance.
Criteria B: The apparatus continues to operate as intended after the test. No degradation of
performance or loss of function is allowed below a performance level specified
by the manufacturer, when the apparatus is used as intended. In some cases the
performance level may be replaced by a permissible loss of performance.
During the test, degradation of performance is however allowed.
Criteria C: Temporary loss of function is allowed, provided the functions self recoverable or
can be restored by the operation of controls.
V
PASS
FAILED
Observation: No any function degraded during the tests.
Page 37
Rev. 00
Report Number: 030355-E
April 29, 2003
17 SECTION 7
IEC 61000-4-6 (CONDUCTED DISTRBANCE/INDUCED
RADIO-FREQUENCY FIELD)
CONDUCTED DISTRBANCE/INDUCED RADIO-FREQUENCY FIELD IMMUNITY TEST
Port
Basic Standard
Requirements
Injection Method
Performance Criteria
Tester
Temperature
Humidity
Pressure
:
:
:
:
:
:
:
:
:
AC Port
IEC 61000-4-6
3 V / with 80% AM. 1kHz Modulation.
CDN-M3
A (Standard Require)
Lung Tsai
23oC
46%
1018mbar
Block Diagram of Test Setup:
0.1m< L <0.3m
10 cm isolation
supporter
EUT and
Support units
CDN
Power
Amplifier
PC
Controller
Ground Reference Plane
Page 38
Rev. 00
Report Number: 030355-E
April 29, 2003
Test Procedure:
1. The EUT and support units were located at a ground reference plane with the interposition of a
0.1 m thickness insulating support and the CDN was located on GRP directly.
2. A ‘H’ messages were displayed on screen of EUT.
3. Adjusting the monitoring camera to monitor the ‘H’ message as clear as possible.
4. Setting the testing parameters of CS test software per IEC 61000-4-6.
5. Recording the test result in following table.
Test conditions:
Frequency Range : 0.15MHz-80MHz
Frequency Step
: 1% of fundamental
Dwell Time
: 3 sec
Range (MHz)
Field
Modulation
Result (Pass/Fail)
0.15-80
3V
Yes
Pass
Performance & Result:
V Criteria A: The apparatus continues to operate as intended. No degradation of performance
V
V
or loss of function is allowed below a performance level specified by the
manufacturer, when the apparatus is used as intended. In some cases the
performance level may be replaced by a permissible loss of performance.
Criteria B: The apparatus continues to operate as intended after the test. No degradation of
performance or loss of function is allowed below a performance level specified
by the manufacturer, when the apparatus is used as intended. In some cases the
performance level may be replaced by a permissible loss of performance.
During the test, degradation of performance is however allowed.
Criteria C: Temporary loss of function is allowed, provided the functions self recoverable
or can be restored by the operation of controls.
V
PASS
FAILED
Observation: No any function degraded during the tests.
Page 39
Rev. 00
Report Number: 030355-E
April 29, 2003
18 SECTION 8 IEC 61000-4-8 (POWER FREQUENCY MAGNETIC FIELD
IMMUNITY TEST)
POWER FREQUENCY MAGNETIC FIELD IMMUNITY TEST
Port
Basic Standard
Requirements
Performance Criteria
Tester
Temperature
Humidity
Pressure
:
:
:
:
:
:
:
:
Enclosure
IEC 61000-4-8
1 A/m
A (Standard Require)
Lung Tsai
23oC
46%
1018mbar
Block Diagram of Test Setup:
Induction
Coil
1/2
Dimension of
EUT
EUT
10mm thick
Insulation Support
Signal
Generator
To
Earth Ground
To
Earth Ground
Page 40
Rev. 00
Report Number: 030355-E
April 29, 2003
Test Procedure:
1.
2.
3.
4.
5.
6.
7.
8.
9.
The EUT and support units were located on Ground Reference Plane with the interposition of a
0.1 m thickness insulation support.
Putting the induction coil on horizontal direction.( X direction )
A test program was loaded and executed in Windows 98 mode.
The data was sent to the screen of EUT and filling the screen with upper case of “H” patterns.
The test program exercised related support units sequentially.
Repeating step 3 to 5 through the test.
Recording the test result as shown in following table.
Rotating the induction coil by 90o ( Y direction ) then repeat step 3 to 7.
Rotating the induction coil by 90o again ( Z direction ) then repeat step 3 to 7.
*. Test conditions:
Field Strength:
1A/m
Power Freq.:
50Hz
Orientation:
X, Y, Z
Orientation
Field
Result (Pass/Fail)
X
1A/m
Pass
Y
1A/m
Pass
Z
1A/m
Pass
Remark
Performance & Result:
V
Criteria A:
The apparatus continues to operate as intended. No degradation of performance
or loss of function is allowed below a performance level specified by the
manufacturer, when the apparatus is used as intended. In some cases the
performance level may be replaced by a permissible loss of performance.
V
Criteria B:
The apparatus continues to operate as intended after the test. No degradation of
performance or loss of function is allowed below a performance level specified
by the manufacturer, when the apparatus is used as intended. In some cases the
performance level may be replaced by a permissible loss of performance.
During the test, degradation of performance is however allowed.
V
Criteria C:
Temporary loss of function is allowed, provided the functions self recoverable or
can be restored by the operation of controls.
V
PASS
V
FAILED
Observation: No any function degraded during the test.
Page 41
Rev. 00
Report Number: 030355-E
April 29, 2003
19
SECTION 9
IEC 61000-4-11 (VOLTAGE DIPS, SHORT INTERRUPTIONS AND
VOLTAGE VARIATIONS)
VOLTAGE DIPS / SHORT INTERRUPTIONS
Port
Basic Standard
Requirement
: AC mains
: IEC 61000-4-11 (1994)
: PHASE ANGLE 0, 45, 90, 135, 180, 225, 270, 315 degrees
Voltage
Dips
Test Level
% UT
<5
70
Reduction
(%)
>95
30
Duration
( periods )
0.5
25
Performance
Criteria
B
C
Voltage
Interceptions
Test Level
% UT
<5
Reduction
(%)
>95
Duration
( periods )
250
Performance
Criteria
C
Test Interval
Tester
Temperature
Humidity
Pressure
:
:
:
:
:
Min. 10 sec.
Lung Tsai
23oC
46%
1018mbar
Block Diagram of Test Setup:
To AC
Source
EUT
&
Support Units
Dips/Interruption and
Variations Simulator
0.8m
Controller Computer
Page 42
Rev. 00
Report Number: 030355-E
April 29, 2003
Test Procedure:
1. The EUT and support units were located on a wooden table, 0.8 m away from ground floor.
2. A test program was loaded and executed in Windows 98 mode.
3. The data was sent to EUT filling the screens with upper case of “H” patterns.
4. The test program exercised related support units sequentially.
5. Setting the parameter of tests and then Perform the test software of test simulator.
6. Conditions changes to occur at 0 degree crossover point of the voltage waveform.
7. Repeating step 3 to 4 through the test.
8. Recording the test result in test record form.
Test conditions:
The duration with a sequence of three dips/interruptions with interval of 10 s minimum
( Between each test event )
Voltage Dips:
Test Level
% UT
0
70
Reduction
(%)
100
30
Duration
( periods )
0.5
25
Observation
Normal
Normal
Voltage Interruptions:
Test Level
Reduction
% UT
(%)
0
100
Duration
Observation
( periods )
250
EUT shut down, but can
be auto recovered as the
events disappear.
Normal: No any functions degrade during and after the test.
Meet Performance
Criteria
A
A
Meet Performance
Criteria
B
Performance & Result:
Criteria A:
The apparatus continues to operate as intended. No degradation of performance
or loss of function is allowed below a performance level specified by the
manufacturer, when the apparatus is used as intended. In some cases the
performance level may be replaced by a permissible loss of performance.
Criteria B:
The apparatus continues to operate as intended after the test. No degradation of
performance or loss of function is allowed below a performance level specified
by the manufacturer, when the apparatus is used as intended. In some cases the
performance level may be replaced by a permissible loss of performance.
During the test, degradation of performance is however allowed.
Criteria C:
Temporary loss of function is allowed, provided the functions self recoverable or
can be restored by the operation of controls.
V
PASS
FAILED
Page 43
Rev. 00
Report Number: 030355-E
April 29, 2003
20
APPENDIX 1 PHOTOGRAPHS OF TEST SETUP
Page 44
Rev. 00
Report Number: 030355-E
April 29, 2003
20.1
LINE CONDUCTED EMISSION TEST (EN 55022)
Front View
Back View
Page 45
Rev. 00
Report Number: 030355-E
April 29, 2003
20.2
RADIATED EMISSION TEST (EN 55022)
Front View
Back View
Page 46
Rev. 00
Report Number: 030355-E
April 29, 2003
20.3 POWER HARMONIC & VOLTAGE FLUCTUATION / FLICKER TEST
(EN 61000-3-2, EN 61000-3-3)
Page 47
Rev. 00
Report Number: 030355-E
April 29, 2003
20.4
ELECTROSTATIC DISCHARGE TEST (IEC 61000-4-2)
Page 48
Rev. 00
Report Number: 030355-E
April 29, 2003
20.5
RADIATED ELECTROMAGNETIC FIELD (IEC 61000-4-3)
Page 49
Rev. 00
Report Number: 030355-E
April 29, 2003
20.6
FAST TRANSIENTS/BURST TEST (IEC 61000-4-4)
Page 50
Rev. 00
Report Number: 030355-E
April 29, 2003
20.7
SURGE IMMUNITY TEST (IEC 61000-4-5)
Page 51
Rev. 00
Report Number: 030355-E
April 29, 2003
20.8 CONDUCTED DISTURBANCE, INDUCED BY RADIO-FREQUENCY FIELDS
TEST (IEC 61000-4-6)
Page 52
Rev. 00
Report Number: 030355-E
April 29, 2003
20.9
POWER FREQUENCY MAGNETIC FIELD IMMUNITY TEST
(IEC 61000-4-8)
Page 53
Rev. 00
Report Number: 030355-E
April 29, 2003
20.10 VOLTAGE DIPS / INTERRUPTION TEST (IEC 61000-4-11)
Page 54
Rev. 00
Report Number: 030355-E
April 29, 2003
21
APPENDIX 2 PHOTOGRAPHS OF EUT
Page 55
Rev. 00
Report Number: 030355-E
April 29, 2003
Front View of EUT
Back View of EUT
Page 56
Rev. 00
Report Number: 030355-E
April 29, 2003
Left View of EUT
Right View of EUT
Page 57
Rev. 00
Report Number: 030355-E
April 29, 2003
I/O Port of EUT
Page 58
Rev. 00
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising