Model DS100 Scale Indicator  Instruction Manual
Model DS100
Scale Indicator
Instruction Manual
Doran Scales, Inc.
1315 Paramount Pkwy
Batavia, IL 60510
1-800-262-6844
www.doranscales.com
MAN0244 Revision 0.2
i
Section 1. Unpacking and Installation ....................................................1
Unpacking ...............................................................................................................................................................1
Installation ...............................................................................................................................................................1
Electrical Connections.............................................................................................................................................1
Care & Cleaning ......................................................................................................................................................1
Section 2. Scale Operation......................................................................2
Display Functions....................................................................................................................................................2
Power.......................................................................................................................................................................2
Units Select..............................................................................................................................................................2
Print .........................................................................................................................................................................2
Motion .....................................................................................................................................................................2
Zero Weighing.........................................................................................................................................................3
Net / Gross Mode.....................................................................................................................................................3
Net / Gross Weighing ..............................................................................................................................................3
Section 3. Setup and Calibration Guide .................................................4
Fig. 2 DS100 Rear Panel .........................................................................................................................................4
Power connections...................................................................................................................................................4
RS-232 Connections................................................................................................................................................4
Load Cell Connections ............................................................................................................................................4
Entering and Exiting Setup Mode - Pushbutton ......................................................................................................5
Entering and Exiting Setup Mode – Front Panel .....................................................................................................5
Setup Mode Navigation ...........................................................................................................................................5
Capacity and Resolution Configuration...................................................................................................................5
Calibration ...............................................................................................................................................................6
Calibration Troubleshooting....................................................................................................................................7
Section 4. Parameter Setup.....................................................................8
Entering and Exiting Setup Mode - Pushbutton ......................................................................................................8
Entering and Exiting Setup Mode – Front Panel .....................................................................................................8
Changing a Parameter..............................................................................................................................................8
Legal for Trade Restrictions ....................................................................................................................................8
Capacity Setup Menu...............................................................................................................................................9
Resolution Setup Menu ...........................................................................................................................................9
Calibration Menu.....................................................................................................................................................9
Reset All Parameters to Default Settings.................................................................................................................9
Operation Mode.......................................................................................................................................................9
Auto Zero Tracking ...............................................................................................................................................10
Motion Aperture ....................................................................................................................................................10
Start Up Zero .........................................................................................................................................................10
Data Output ...........................................................................................................................................................11
Data Format ...........................................................................................................................................................11
Baud Rate ..............................................................................................................................................................12
Units Enabled ........................................................................................................................................................12
Startup Units..........................................................................................................................................................12
Front Panel Calibration..........................................................................................................................................13
Raw Counts ...........................................................................................................................................................13
ii
Section 5. Data Communications .........................................................14
Introduction to data communications ....................................................................................................................14
Remote Scale Commands ......................................................................................................................................17
UPS Application Note: UPS On-Line WorldShip Software Rev 3.1 and Higher .................................................18
Hardware Setup: ....................................................................................................................................................18
Software Setup: .....................................................................................................................................................18
Scale Configuration:..............................................................................................................................................19
Troubleshooting:....................................................................................................................................................19
Section 6. Specifications........................................................................20
Section 7. Troubleshooting...................................................................21
General Problem Resolution..................................................................................................................................21
Error Messages ......................................................................................................................................................22
iii
Section 1. Unpacking and Installation
Unpacking
Before unpacking your Doran scale, please read the instructions in this
section. Your new scale is a durable industrial product, but it is also a
sensitive weighing instrument. Normal care should be taken when
handling and using this product. Improper handling or abuse can
damage the scale and result in costly repairs that may not be covered by
the warranty. If you notice any shipping damage, notify the shipper
immediately. Please observe the following precautions to insure years
of trouble free service from your new scale.
! DO NOT drop the scale.
! DO NOT drop objects on the scale.
Carefully remove the scale from the shipping carton. Be sure to retain
all shipping materials in case the scale must be shipped elsewhere.
Installation
Place the scale on a stable flat surface. Verify that the bubble level
located under the platter shows that the scale is level. Adjust the four
feet to obtain a level condition (bubble in center.)
Electrical Connections
The DS100 uses a wall mounted transformer to provide power to the
scale. The transformer requires 115 VAC, 50/60 Hz power. Be sure the
AC power is not excessively noisy – this can occur if large inductive
loads, such as solenoids or motors, are on the same power line.
Subjecting electronics to problematic AC power lines may result in
damage not covered by the warranty. Also be sure that the power outlet
and transformer are not exposed to water while the scale is plugged in.
Care & Cleaning
With reasonable care, this product will last for many years. Here are
some tips to care for your DS100 Scale Indicator.
• Hand clean with a damp cloth using mild detergent.
• Do not use strong solvents or abrasive cleaners as this can damage
the touch panel or other plastic parts.
• Do not drop or overload the scale.
• Do not use sharp objects to press any of the buttons.
• Do not immerse.
1
Section 2. Scale Operation
Fig. 1 DS100 Front Panel
Display Functions
The Model DS100 controls consist of GROSS/NET, TARE, ZERO (ON/OFF),
UNITS and PRINT buttons located under to the LED display. The display is used
to provide weight indications and operator messages describing scale operation.
Power
The DS100 is powered by an AC wall transformer. Turn the scale off by pressing
and holding the ZERO (ON/OFF) button for three seconds. Turn the scale on by
pressing the ZERO (ON/OFF) button.
Units Select
Press the UNITS button to change weight display units from lb, lb:oz, oz, kg, g.
The units annunciator to the right of the LED display will indicate the current
weight.
Print
To transmit the selected print string through the standard RS-232 port, press
PRINT. This will cause the selected print string to be printed. The scale will not
transmit while the scale is in motion. If the scale is in motion when the PRINT
button is pressed, the DS100 will transmit the selected print string once the scale
becomes stable.
Motion
Motion is indicated by the MOT annunciator to the left of the LED display.
2
Zero Weighing
1. Empty the scale base or place an empty container on the base and press
ZERO to zero the scale. If the scale is in motion when the ZERO button is
pressed, the DS100 will zero the scale once the scale becomes stable.
2. Place an item on the scale and wait for the MOTION indicator to go out
3. Read the weight on the display
4. Press ZERO again to weigh additional items
Net / Gross Mode
The NET annunciator to the left of the LED display will indicate the NET or
GROSS weigh status of the indicator. The indicator enters the NET mode when
the TARE button is pressed with an item on the base. Alternatively, to enter the
NET mode, press the GROSS/NET button. The NET mode is not available when
a TARE weight is not active.
To clear an active TARE weight:
1. If the scale is in the NET mode, enter the GROSS mode by pressing ZERO.
2. When the scale display reads zero in the GROSS mode, press TARE.
Net / Gross Weighing
1. Empty the scale base or place an empty container on the base and press
TARE
2. The indicator will read zero and place the scale in NET mode
3. Place an item on the scale platter and wait for the MOTION indicator to go out
4. Read the weight on the display
5. Press TARE again to weigh additional items
6. Press ZERO, then TARE to clear the tare weight
3
Section 3. Setup and Calibration Guide
Fig. 2 DS100 Rear Panel
Power connections
The DS100 is powered from a wall mounted transformer. The transformer has a
power cord which plugs into the power jack located on the back of the scale. A
strain relief is provided to the left of the power jack. The power cord can be
looped around the strain relief to prevent accidental unplugging.
RS-232 Connections
The DS100 has a standard RS-232 output. To use this feature, the optional RS232 cable (P/N DSOPT001 or DSOPT002) is required. To access the RS-232
terminals at the rear of the scale, remove the rear panel from the indicator as
shown in Figure 2.
Insert the cable conductors by placing a small flat blade screw driver into the
terminal slot and then push the handle of the screw driver upwards towards the
top of the indicator. This forces the terminal below to open. Insert the conductor
into the terminal and remove the screwdriver.
Load Cell Connections
The DS100 accepts 4 wire and 6 wire load cells. For 4-wire operation, jumpers
JP1 and JP2, located to the right of the load cell connections, must be inserted.
For 6-wire operation, remove JP1 and JP2. To protect against ESD, use the
ground wire provided with your Doran base load cell cable and connect to the
Shield (SHD) terminal. If a ground wire is not available, ground the base to the
meter with the SHD connection. Verify proper grounding by measuring the
resistance from the SHD terminal on the DS100 to the base chassis.
Insert the cable conductors by placing a small flat blade screw driver into the
terminal slot and then push the handle of the screw driver upwards towards the
4
top of the indicator. This forces the terminal below to open. Insert the conductor
into the terminal and remove the screwdriver.
Doran Standard Load Cell Color Code
+ Signal
Red
- Signal
White
+ Excitation
Green
- Excitation
Black
+ Sense
Blue
- Sense
Brown or Yellow
Entering and Exiting Setup Mode - Pushbutton
The setup mode can be entered by pressing the setup pushbutton located behind
the rear panel as shown in Figure 2. This pushbutton is provided so a lead and
wire seal can be utilized to metrologically seal the scale. To exit the setup mode,
press the setup pushbutton or the PRINT button.
Entering and Exiting Setup Mode – Front Panel
To enter setup mode from the front panel, turn off the indicator. Then press and
hold PRINT and ZERO(ON/OFF). When the indicator turns on, all annunciators
will be activated. Release both PRINT and ZERO(ON/OFF). To exit the setup
mode, press PRINT to immediately exit or press and release UNITS until the
indicator automatically exits.
Setup Mode Navigation
To navigate through the parameter menus:
• Press ZERO to change the parameter selection
• Press UNITS to go to the next parameter
• Press TARE to back up to the previous parameter
For changing the capacity and scale calibration, refer to the sections below for
detailed instructions.
Capacity and Resolution Configuration
After entering the setup mode, the display will briefly show the Calibration and
Parameter audit counters. The display will then show C00100, in this case, the
capacity is set to 100. The capacity unit can be either lb or kg. The units
annunciator to the right of the weight display will indicate the active unit.
1. To change the capacity, press ZERO
2. The active digit will flash
3. Press ZERO to change the active digit
4. Once the active digit is properly set, press UNITS to advance to the next
digit to the right or press TARE to move one digit to the left
5. After setting the right most digit, the units annunciator will flash
6. To change the capacity unit, press ZERO to change between lb and kg
5
7. Press UNITS to advance to the resolution parameter once the proper unit
is selected
8. The display will read the current “count by” setting for example: d 0.01
9. Only resolutions that are available for the defined capacity can be
selected. The DS100 is limited from 100 to 12500 displayed divisions.
10. Press ZERO to select the desired resolution
11. Press UNITS to advance to calibration
NOTE: If CAP E appears on the display during step 7, the capacity selected is
out of range. Ensure the capacity is within 1 to 99,999 lb or 1 to 45,000
kg. Verify the desired unit annunciator is lit. Press ZERO to clear this
error.
Calibration
Once the capacity and resolution are set, the scale can be calibrated. The
DS100 utilizes a two point zero and span calibration. Note: the minimum
calibration weight is 5% of scale capacity. For the most accurate results, it is
recommended to calibrate as close to full capacity as possible.
1. After the capacity is set the scale reads CAL 0
2. To perform a zero calibration, remove all items from the base and ensure
the base is free from any disturbance (air currents or vibration)
3. Press ZERO. The indicator will count down from 7 to 0. If there is any
disturbance, the countdown will reset back to 7 to ensure an accurate
calibration.
4. Once zero calibration is successfully completed, the display will show the
calibration weight value. For example 00100, represents 100 lb or kg.
The units annunciator to the right of the LED display will indicate the active
unit. This unit must be the same as the capacity and resolution unit as
defined above.
5. To change the span calibration weight, press ZERO.
6. The active digit will flash
7. Press ZERO to change the active digit
8. Once the active digit is properly set, press UNITS to advance to the next
digit to the right or press TARE to move one digit to the left
9. After setting the right most digit, place the calibration weight on the base
10. Press UNITS to perform the span calibration
11. After span calibration, the scale returns to the weigh mode to confirm
calibration
NOTE: If rG Err appears on the display during step 3, the calibration zero is out
of range. Press ZERO to clear this error. Refer to the Calibration
Troubleshooting below for additional information.
NOTE: If SPn E appears on the display during step 4, the entered calibration
weight value is outside the required 5% to 100% of capacity. Press
ZERO to clear this error.
6
NOTE: If SPnL or SPnH appears on the display, the calibration span is out of
range. Verify that the calibration weight is correct and repeat the
calibration. Refer to the Calibration Troubleshooting below for additional
information.
Calibration Troubleshooting
The following table shows the acceptable Raw Counts for no load and full load.
Raw Counts can be viewed through the last setup parameter. Refer to Section 4,
Parameter Setup, for more information regarding parameter viewing.
If the scale’s raw counts are not within the ranges specified in the following table,
contact a Doran Certified Scale Distributor, or Doran’s Technical Assistance
department at 800-262-6844.
Calibration Requirements in Raw
Counts
Zero Calibration Point
Required Calibration Span
Minimum
-8,000
10,000
Maximum
55,000
220,000
500
11,000
(100% of Capacity Scale Cal Point – Zero
Cal Point)
Required Calibration Span
(5% of Capacity Cal Point – Zero Cal Point)
7
Section 4. Parameter Setup
The DS100 has adjustable setup parameters, which control the scale’s operation.
Entering and Exiting Setup Mode - Pushbutton
The setup mode can be entered by pressing the setup pushbutton located behind
the rear panel as shown in Figure 2. This pushbutton is provided so a lead and
wire seal can be utilized to metrologically seal the scale. To exit the setup mode,
press the setup pushbutton or the PRINT button.
Entering and Exiting Setup Mode – Front Panel
To enter setup mode from the front panel, turn off the indicator. Then press and
hold PRINT and ZERO(ON/OFF). When the indicator turns on, all annunciators
will be activated. Release both PRINT and ZERO(ON/OFF). To exit the setup
mode, press PRINT to immediately exit or press and release UNITS until the
indicator automatically exits.
Parameters are accessed by pressing the UNITS button. Pressing this button
will cause the scale to step to the next parameter on the list.
NOTE: No setup information is saved until the DS100 exits the setup mode. A
power failure while in the setup mode will cause changes to be lost.
If it is necessary to physically seal the scale, run a lead and wire seal through
the three cross drilled bolts at the rear of the indicator.
Changing a Parameter
Once the desired parameter has been found, it may adjusted. Press and release
ZERO to step through the individual options. See Section 3 for setting scale
capacity and resolution. When you have stepped through all of the parameters,
the scale will return to the weigh mode.
Legal for Trade Restrictions
The Legal for Trade mode disables the lb:oz mode. Setup mode front panel
access can be disabled for applications that require a physical lead and wire
seal.
8
Capacity Setup Menu
C
Select Scale Capacity (lb or kg)
Capacity setting:
1−99999(lb) 1 to 99,999 in lb
1 to 45,000 in kg
1−45000(kg) See Capacity and Resolution Configuration
in Section 3 for configuration instructions.
Resolution Setup Menu
Select Scale Capacity (lb or kg)
d
0.0001 to 50 resolution setting in same unit
0.0001− as capacity unit above. The resolution
selections are limited by the capacity
50
setting from 200 to 12,500 divisions.
Calibration Menu
CAL
0
Zero Calibration Point
Press ZERO to calibrate zero point
000000
Span Calibration Point
Span calibration weight setting:
1 to 99,999 in lb
1−99999(lb) 1 to 45,000 in kg.
See Calibration in Section 3 for calibration
1−45000(kg) instructions.
NOTE: This parameter is only visible after
a zero calibration is performed.
Reset All Parameters to Default Settings
Default parameters
rst
No. Parameters are not defaulted.
n
Yes. Set all parameters to the factory
y
default values when UNITS is pressed.
Operation Mode
oP
St
44
Sets the scale for Legal for Trade mode
Standard mode. NOT legal for trade.
Legal for trade setting. This setting
automatically disables certain parameter
menus below as indicated by an (*).
9
Auto Zero Tracking
2t *
of
0.5
3.0
1.0
# of displayed divisions that are
automatically zeroed from displayed
zero, while the scale is stable.
Zero tracking is disabled
1/2 division (Legal for Trade Setting)
3 divisions
1 division
Motion Aperture
nnA *
1
3
5
# of displayed divisions that must
change before motion is detected.
1 division (Legal for Trade Setting)
3 divisions
5 divisions
Start Up Zero
Suo *
no
FS
20
Controls the start-up zero point
Scale will NOT zero on power up.
Calibrated zero is loaded on power-up.
Scale will zero up to 100% of scale capacity
on power-up.
Scale will zero up to 20% of scale capacity
on power-up.
(Legal for Trade Setting)
NOTE: The Legal for Trade requirements disables certain parameter menus listed
above. These items have been indicated by an asterisk ( * ).
10
Data Output
d.o.
C.P.
A.1
A.2
t.d.
Controls when information is
transmitted from the scale though the
serial port
Continuous data transmission. Transmits
data each time the display is updated.
Readings that occur when the scale is in
motion are identified by the abbreviation
“MOT.” following the weight data.
Auto Print 1 transmits the first scale reading
after the scale leaves motion. The reading
must be stable and must be a valid reading
before it can be sent.
Auto Print 2 transmits the first scale reading
following the scale leaving motion. In Auto
Print 2, no further readings will be sent until
the scale returns to displayed zero. The
reading must be stable and must be a valid
reading before it can be sent.
Transmit on Demand transmits when the
PRINT button is pressed and when the “W”
data request command is received through
the serial port. The scale must be stable
and the weight must be valid (no error
codes displayed) before data is printed.
Data Format
For.
FO
2P
F1
SP
gn
UPS
Format of transmitted data from the
serial port
Basic data format.
Basic dual print format. Includes metric
weight.
Legacy data format.
Output NOT Legal for Trade.
Basic format for an Eltron SSP printer.
Call Doran for details.
Gross, Net and Tare format
UPS print string compatible with UPS
Online World Ship program
11
Baud Rate
b.r.
↑
12
24
48
96
192
Serial port transmission rate
1200 Baud (bits per second)
2400 Baud (bits per second)
4800 Baud (bits per second)
9600 Baud (bits per second)
19,200 Baud (bits per second)
lb Units Enabled
on
of
Enables/Disables pounds
lb is enabled.
lb is disabled.
lb:oz Units Enabled
*
†
on
of
Enables/Disables pounds:ounces
lb:oz is enabled.
lb:oz is disabled.
oz Units Enabled
†
on
of
Enables/Disables ounces
oz is enabled.
oz is disabled.
kg Units Enabled
on
of
Enables/Disables kilograms
kg is enabled.
kg is disabled.
g Units Enabled
†
on
of
Enables/Disables grams
g is enabled.
g is disabled.
Startup Units
Units
lb
lb:oz † *
oz †
kg
g†
Determines scale startup units
Note: Unit annunciator to the right of the
display indicates parameter setting
Scale starts up in lb.
Scale starts up in lb:oz.
Scale starts up in oz.
Scale starts up in kg.
Scale starts up in g.
12
NOTE: Depending on capacity and resolution selection, some units are not available
due to display constraints. These limitations are indicated by (†).
Front Panel Calibration
Enables/Disables Front Panel
FCL
Calibration
Note: This menu is only accessible by
pressing the calibration pushbutton at the
rear of the DS100.
Front panel calibration is enabled.
on
Front panel calibration is disabled. Access
of
to the setup mode is allowed through the
calibration pushbutton located behind the
rear panel.
Raw Counts
######
Raw counts from A/D converter
View these numbers if calibration is
unsuccessful. Refer to Section 3 for
interpretation of these numbers. Press
UNITS to exit setup mode.
NOTE: Depending on capacity and resolution selection, some units are not available
due to display constraints. These limitations are indicated by (†).
13
Section 5. Data Communications
Introduction to data communications
Basic understanding of serial data communications is needed when setting up the
DS100 to communicate with a printer or PC.
When setting up a serial communications system, there are two concerns which
affect the configuration of that system. These are:
•
•
Baud Rate
Data Bits and Parity
The baud rate determines how fast the data is sent from the scale. The sending
and receiving units must be set to the same baud settings. Typical values are
1200, 2400, 4800 and 9600.
The DS100 is factory set for eight bits, no parity and one stop bit also known as
8n1. The receiving units must be set to 8n1 for proper communications.
“F0” Example (Negative weight, in motion)
☻- 2.452 lb MOT.♪◙
“F0” Print String Definition for Each Weight Unit:
Pounds
STX | POL | WEIGHT| SP | lb | SP | ST | CR | LF
Ounces
STX | POL | WEIGHT| SP | oz | SP | ST | CR | LF
Kiligrams
STX | POL | WEIGHT| SP | kg | SP | ST | CR | LF
Grams
STX | POL | WEIGHT| SP | g | SP | SP ST | CR | LF
Pounds-ounces
STX | POL | WEIGHTLB| SP | lb | POL | WEIGHTOZ | SP | oz | SP | ST | CR | LF
STX (☻) = ASCII 02
POL = minus sign for negative weight or a space for a positive weight
WEIGHT = 6 character field plus decimal if needed
WEIGHTLB = pound portion of lb-oz weight
WEIGHTOZ = ounce portion of lb-oz weight. (WEIGHTLB and WEIGHTOZ total 5 characters
plus decimal)
SP = ASCII space
ST = MOT. if in motion or a space if stable
CR (♪) = Carriage return
LF (◙) = Linefeed
| = Separator, not printed
14
“F1” Example (Negative weight, in motion)
☻- 2.452 LBM♪◙
“F1” Print String Definition for Each Weight Unit:
Pounds
STX | POL | WEIGHT| SP | LB | ST | CR | LF
Ounces
STX | POL | WEIGHT| SP | OZ | ST | CR | LF
Kiligrams
STX | POL | WEIGHT| SP | KG | ST | CR | LF
Grams
STX | POL | WEIGHT| SP | G | SP | ST | CR | LF
Pounds-ounces
STX | POL | WEIGHTLB| SP | LB | POL | WEIGHTOZ | SP | OZ | SP | ST | CR | LF
STX (☻) = ASCII 02
POL = minus sign for negative weight or a space for a positive weight
WEIGHT = 6 character field plus decimal if needed
WEIGHTLB = pound portion of lb-oz weight
WEIGHTOZ = ounce portion of lb-oz weight. (WEIGHTLB and WEIGHTOZ total 5 characters
plus decimal)
SP = ASCII space
ST = MOT. if in motion or a space if stable
CR (♪) = Carriage return
LF (◙) = Linefeed
| = Separator, not printed
“2P” Example (Negative weight, in motion)
☻- 2.452 lb MOT.♪◙
☻(- 1112 kg MOT.)♪◙
The dual print mode provides the DS100 with the ability to print the current
scale reading followed by the equivalent value in kilograms.
The weight is first printed using the “F0” format. Then the weight is
recalculated in kilograms and is sent as a second line of text. The kilogram
data follows the “F0” data format except where parentheses are placed after
the STX character and before the carriage return & line feed.
15
“SP” Example (Negative weight, in motion)
FR"L1"◙
?◙
- 1.052◙
lb◙
GS◙
MOT. ◙
- 0.478◙
kg◙
P1,1◙
“SP” Print String Definition for Each Weight Unit:
Pounds
FR”L1” | LF | ? | LF | POL | WEIGHT | LF | lb | LF | GS | LF | ST | LF | POL | WEIGHT2 | LF | kg | LF | P1,1 | LF
Ounces
FR”L1” | LF | ? | LF | POL | WEIGHT | LF | oz | LF | GS | LF | ST | LF | POL | WEIGHT2 | LF | kg | LF | P1,1 | LF
Kilograms
FR”L1” | LF | ? | LF | POL | WEIGHT | LF | kg | LF | GS | LF | ST | LF | POL | WEIGHT2 | LF | kg | LF | P1,1 | LF
Grams
FR”L1” | LF | ? | LF | POL | WEIGHT | LF | g | SP | LF | GS | LF | ST | LF | POL | WEIGHT2 | LF | kg | LF | P1,1 | LF
Pounds - ounces
FR”L1” | LF | ? | LF | POL | WEIGHTLB | SP | lb | POL | WEIGHTOZ | LF | oz | LF | GS | LF | ST | LF | POL | WEIGHT2 | LF |
kg | LF | P1,1 | LF
POL = minus sign for negative weight or a space for a positive weight
WEIGHT = 6 character field plus decimal if needed
WEIGHT2 = Kilogram weight. 6 character field plus decimal if needed
WEIGHTLB = pound portion of lb-oz weight
WEIGHTOZ = ounce portion of lb-oz weight. (WEIGHTLB and WEIGHTOZ total 5 characters
plus decimal)
SP = ASCII space
ST = MOT. if in motion or four (4) spaces if stable
CR (♪) = Carriage return
LF (◙) = Linefeed
| = Separator, not printed
16
“Gn” Example (Negative weight, in motion)
25.15 lb GR MOT.♪◙
20.05 lb NT MOT.♪◙
5.10 lb TR MOT.♪◙
☻
☻
☻
“Gn” Print String Definition for Each Weight Unit:
Pounds
STX | POL | WEIGHTGROSS | SP | lb | SP | GR | SP | ST | CR | LF
STX | POL | WEIGHTNET | SP | lb | SP | NT | SP | ST | CR | LF
STX | POL | WEIGHTTARE | SP | lb | SP | TR | SP | ST | CR | LF
Ounces
STX | POL | WEIGHTGROSS | SP | oz | SP | GR | SP | ST | CR | LF
STX | POL | WEIGHTNET | SP | oz | SP | NT | SP | ST | CR | LF
STX | POL | WEIGHTTARE | SP | oz | SP | TR | SP | ST | CR | LF
Kiligrams
STX | POL | WEIGHTGROSS | SP | kg | SP | GR | SP | ST | CR | LF
STX | POL | WEIGHTNET | SP | kg | SP | NT | SP | ST | CR | LF
STX | POL | WEIGHTTARE | SP | kg | SP | TR | SP | ST | CR | LF
Grams
STX | POL | WEIGHTGROSS | SP | g | SP | GR | SP | ST | CR | LF
STX | POL | WEIGHTNET | SP | g | SP | NT | SP | ST | CR | LF
STX | POL | WEIGHTTARE | SP | g | SP | TR | SP | ST | CR | LF
Pounds-ounces
STX | POL | WEIGHTLBGROSS | SP | lb | POL | WEIGHTOZGROSS | SP | oz | SP | GR | SP | ST | CR | LF
STX | POL | WEIGHTLBNET | SP | lb | POL | WEIGHTOZNET | SP | oz | SP | NT | SP | ST | CR | LF
STX | POL | WEIGHTLBTARE | SP | lb | POL | WEIGHTOZTARE | SP | oz | SP | TR | SP | ST | CR | LF
STX (☻) = ASCII 02
POL = minus sign for negative weight or a space for a positive weight
WEIGHT[GROSS/NET/TARE] = 6 character field plus decimal if needed
WEIGHTLB[GROSS/NET/TARE] = pound portion of lb-oz weight
WEIGHTOZ[GROSS/NET/TARE] = ounce portion of lb-oz weight. (WEIGHTLB and
WEIGHTOZ total 5 characters plus decimal)
SP = ASCII space
ST = MOT. if in motion or a space if stable
CR (♪) = Carriage return
LF (◙) = Linefeed
| = Separator, not printed
Remote Scale Commands
The scale will respond to the following single letter ASCII commands.
“W”
“U”
“Z”
Initiates transmission of current weight data (if in motion, scale will
wait until stable, then print).
Changes the displayed weight units.
Zeroes the scale (if in motion, scale will wait until stable, then zero).
17
UPS Application Note: UPS On-Line WorldShip Software Rev 3.1 and Higher
Description:
The following instructions will allow a Scale to be connected to the UPS On-Line
shipping system. The UPS On-Line software operates by continuously polling the
scale for weight several times a second. The UPS program then displays the
current weight in a window marked “Scale Weight:”. This package weight, in
pounds is then used to generate a shipping label if the “Electronic Scale Activated”
check box is selected.
Hardware Setup:
Before starting the UPS On-Line software, connect the Scale's serial port to the
computer’s serial port (com1 or com2) by way of the optional serial cable (P/N
DSOPT001 or DSOPT002). Apply power to the scale and press the ZERO push
button.
WARNING: The scale must be on and connected at all times while the UPS OnLine program is running. If the scale is disconnected, turned off, or placed in the
calibration mode, an error message will be displayed on the computer screen. To
correct this error, you must perform a scale setup again or restart the UPS
software – see Software Setup for instructions.
Software Setup:
If the UPS On-Line program has been already installed, follow the configuration
steps below only. If the UPS On-Line program is not installed, power up the scale
and connect the scale's serial cable to an available PC serial port. Next, install the
UPS On-Line program as per instruction on the UPS installation CD. The
installation program will prompt you to select if a scale is connected, select yes.
To configure the UPS On-Line WorldShip program:
1. Click on the “Tools” menu located at top of screen, then select “System
Preferences”.
2. Click on the “Hardware” tab and set the “Scale Type:” drop down menu to
“Fairbanks 70-2453-4”.
3. Click on the “Scale Port:” drop down menu, choose which serial port the
scale is connected to, com1, com2 or other com port. This is the port that
the DB9 cable is connected to at the rear of your PC from the scale.
4. Finally, click on the “Test Scale” button. Place weight on the scale. The
computer should then display the current weight on the scale.
5. Click the “OK” button, and return back to the service menu.
6. Check the “Electronic Scale Activation” box to accept weight readings from
the scale. The current weight will be displayed under the “Package:” section,
in the “Scale Weight” window. NOTE: UPS software does not show zero or
negative weights on the main display page.
18
Scale Configuration:
In order to be compatible with the UPS On-Line program. The Scale's Parameters
must be configured to the following settings:
d.o.
For.
br.
Units
=
=
=
=
t.o.d.
UPS
96
lb
Data Output Mode is set to Transmit On Demand
Data Output Format is set to UPS protocol
Baud Rate is set to 9600 baud
Start up units must be set to pounds
Troubleshooting:
If any problems occur with your UPS On-Line Software, a trouble-shooting guide
can be found on the United Parcel Service web site under Shipping Systems
Scales at address:
http://www.ups.com/using/custserv/techfaq/scales.html
Or alternatively, contact Doran Scale’s Technical Support Department at 800-2626844 for assistance over the phone.
19
Section 6. Specifications
Resolution
200 to 12500 divisions
Indicator Load Cell
Input Range
Excitation
0.35 mV/V to 3.0 mV/V
Power Supply
Wall Transformer output: (scale input)
12VDC, 300mA
Neg. (-) center
Display
0.56” high red LED
Displayed Units
lb, oz, kg, g and lb-oz
Indicator Capacities
1 to 99,999 lb
1 to 45,000 kg
Bi-directional RS-232
5V
Printer Interface
Calibration
Zero and Span
(Minimum Span = 5% of Capacity)
Controls
GROSS/NET, TARE, ZERO(ON/OFF),
UNITS AND PRINT buttons
Construction
Painted Mild Steel
Options
RS232 cable (Female) P/N DSOPT001
RS232 cable (Male) P/N DSOPT002
230 VAC Transformer P/N DSOPT003
20
Section 7. Troubleshooting
General Problem Resolution
Problem
Weight reading will not repeat or
scale does not return to zero when
weight is removed.
What to Do or Check
Make sure that the scale platter is not
rubbing or touching the scale cover. Verify
that there is nothing caught in the platform,
under or around the load cell.
Scale overloads early
Verify scale calibration is correct. If
problem persists, recalibrate the scale.
Scale will not come to zero when the
ZERO button is pressed.
Make sure that the scale is becoming
stable (Motion annunciator is off). After
pressing the ZERO button, the scale
should zero as soon as it becomes stable.
If problem persists, there may be a
problem with the touch panel or
motherboard.
Weight readings don’t seem to be
correct.
Verify the scale calibration with an
accurate test weight. If the readings are
not correct, recalibrate.
Scale drifts off zero.
Check for air currents and/or vibration
around the scale. If that is the cause, it
may be necessary to set the AZ T and nnA
parameters to wider settings to
compensate (see the parameter section.)
Verify that no mechanical restrictions exist,
i.e. platter rubbing, something caught
under or around the load cell.
Scale shuts itself off or will not turn
on.
Press the ZERO button to turn on the
indicator.
The transformer may be bad or the power
connector at the rear of the scale may
have an intermittent connection. Check
the power connector at the rear of the
scale.
21
Error Messages
Error Message
Er EP
What to Do or Check
The setup parameters loaded in nonvolatile memory
have become corrupted.
Verify scale parameters and calibrate.
The A/D communication is not detected.
Er Ad
If problem persists, recalibrate.
If problem still persists, the motherboard will need to
be replaced.
Verify load cell wiring connections. Ensure the load
cell is wired properly. Ensure that all load cell
conductors are inserted into the terminal properly
and that the conductor jacket in not interfering.
AD of
Verify that JU1 and JU2 have the jumper inserted
for a four wire load cell or removed for a six wire
load cell.
rg Err
Ldg 0
ov−Ld
If problem still persists, the motherboard will need to
be replaced.
The calibration zero is out of range.
Error is displayed after a ZERO calibration attempt.
Press zero to clear this error.
Refer to the analog setup section for additional
information.
Motherboard or load cell may need to be replaced.
The scale is attempting to zero on power-up.
This message will remain until the scale is stable.
Air currents or vibration may be the cause. If
problem persists, the pcb or load cell may be
damaged.
NOTE: This message will not appear if parameter
Suo = no.
The scale is in overload. The load on the scale
platform exceeds the scale capacity by more than
105%.
Remove excess weight from scale platform.
If problem persists, recalibrate.
If problem still persists, the motherboard or load cell
may need to be replaced.
22
Error Message
gS−oL
SPnL
SPnH
SPn E
ERR 0
CAP E
What to Do or Check
The scale is in gross overload. The load exceeds
the scale rating and might result in damage to the
scale.
Remove excess weight immediately.
If problem persists, recalibrate.
If problem still persists, the motherboard or load cell
will need to be replaced.
Raw counts for the span calibration is too low.
Refer to the Calibration Troubleshooting section for
raw count ranges.
Raw counts for the span calibration is too high.
Refer to the Calibration Troubleshooting section for
raw count ranges.
The span calibration weight must be between 5%
and 100% of full capacity.
Load on the scale exceeds 20%. Remove excess
weight. This error only occurs when the Start Up
Zero Suo parameter is set to 20. Change this
parameter for FS to allow for automatic start up
zeroing up to 100% of capacity.
The capacity has a zero value or the value exceeds
45,000 kg. Adjust capacity and ensure the capacity
is defined in the desired unit.
23
Limited One Year Warranty
Doran Scales, Inc. warrants its products to be free from defects in material and
workmanship for a period of one (1) year from date of shipment. Any product found to
be defective within this time period may be returned to Doran’s factory, freight prepaid,
with prior return authorization and proof of purchase showing date of original sale, for
repair or
replacement at no charge.
Doran’s liability under this warranty is limited to the repair or
replacement of the defective product and in no event shall Doran Scales, Inc. be liable
for consequential or indirect damages to equipment or personnel. Nor shall Doran
Scales. Inc. be liable for damages to equipment or for personal injury caused by
misuse, overload, accidental damage, alteration, improper installation, or unauthorized
opening of the equipment. Under no circumstances will Doran Scales, Inc. be
responsible for any indirect or consequential damages due to errors in weighing or
failure of a Doran Scales, Inc. product to perform properly.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR
IMPLIED. THIS WARRANTY CONSTITUTES DORAN’S EXCLUSIVE WARRANTY.
THERE ARE NO OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING ANY
WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
Doran Scales, Inc.
1315 Paramount Pkwy
Batavia, IL 60510
1-800-262-6844
www.doranscales.com
24