MDS-400A Series Medical AC-DC Open Frame Power Supply 24V/400W 3

MDS-400A Series Medical AC-DC Open Frame Power Supply 24V/400W 3
Medical AC-DC Open Frame Power Supply
24V/400W 3” x 5” High Density MDS-400A Series
MDS-400A Series
Highlights & Features
 3" x 5" x 1.4" Package
 Power Good, Remote sense, Remote ON/OFF,
5V Standby Power and 12V Fan Power
 200W convection, 400W force air
 IT & Medical Safety Approvals
Safety Standards
CB Certified for worldwide use
Model Number:
MDS-400A
Unit Weight:
405g
Dimensions (W x L x H): 76.2 x 127 x 35.6 mm
3 x 5 x 1.4 inch
General Description
The MDS-400A series of internal open frame power supplies come with universal AC input range from 90Vac to 264Vac.
Other features include low leakage, Type BF Patient Access Leakage Currents, and electric shock protection compliance
with 2 x MOPPrequirements. The MDS series is certified for EMC standards according to EN 55011for industrial, scientific
and medical (ISM) radio-frequency equipment; and, EN 55022 for Industrial Technology Equipment (ITE) radio-frequency
equipment. In addition, only recognized Japanese capacitors are used to ensure long product life.
The MDS-400A series comes with both medical and ITE safety approvals, including UL/CE/CCC (5000meters), and CB
certification. Designs are compliant with RoHS Directive 2011/65/EU for environmental protection.
Model Information
Medical AC-DC Open Frame
ModelNumber
MainOutput Main Output Standby
Voltage
Current
Output
Voltage
Standby
Output
Current
Fan Output
Voltage
Fan Output
Current
MDS-400APB24AA
24Vdc
0-2A
12Vdc
0.05-0.6A
0-16.66A
5Vdc
1)
Total Max Output Power
400W
2)
1) Fan output will be presenting when 24V main output is available
2) With 12CFM force air
Model Numbering
MDS
400
A
P
B
24
Delta Medical
power Supply
Max wattage in
the product
Series.
Maybe lower at
some voltage.
060→60W
150 →150W
1K2 → 1,200W
Family Code
Product Type
Input Type Code
Output Voltage
A~ Z
P: Open Frame
B: 3pin Class I
24 for24V
1
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (Jun 2015, Rev. 04)
A
Revision Code
A
Revision code
Medical AC-DC Open Frame Power Supply
24V/400W 3” x 5” High Density MDS-400A Series
Specifications
Input Ratings / Characteristics
Nominal Input Voltage
100-240Vac
Input Voltage Range
90-264Vac
Nominal Input Frequency
50-60Hz
Input Frequency Range
47-63Hz
InputCurrent (max)
5.5A
Input Surge Voltage (max)
300Vac for 100ms
Full load Efficiency (typ.)
92% @ 115Vac/60Hz
93% @ 230Vac/50Hz, Reference Fig.1
Standby Power (max)
0.5W (only standby working with Inhibit signal high)
@ 115Vac/60Hz, 230Vac/50Hz
InrushCurrent (max)
40A@230Vac, cold start
Input-PE (protective earth) leakage current (max)
0.1mA @ NC, 0.3mA @ SFC1)
Output-PE (protective earth) leakage current for Type BF application (max)
0.1mA @ NC, 0.5mA @ SFC 1)
Power Factor (min)
0.95 @ 115V/50Hz, 230V/50Hz, full load
1) NC: normal condition, SFC: single fault condition
Leakage Current
Input-PE Leakage Current
100Vac/60Hz (Typ.) 264Vac/60Hz (Typ.)
Delta Limit
IEC60601-1 Limit
Normal Condition 17.5uA
43.5uA
100uA max
5000uA max
Single Fault Condition 32.9uA
90.7uA
300uA max
10000uA max
86.7uA
100uA max
100uA max
128.6uA
500uA max
500uA max
Output-PE Leakage Current forType BF application
Normal Condition 28.5uA
Single Fault Condition 42.9uA
100.000
90.000
Efficiency (%)
80.000
115Vac
70.000
230Vac
60.000
50.000
25%
50%
Output Load
Fig.1 Efficiency versus output load
2
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (Jun 2015, Rev. 04)
75%
100%
Medical AC-DC Open Frame Power Supply
24V/400W 3” x 5” High Density MDS-400A Series
Output Ratings / Characteristics
Nominal Output Voltage (Vrated)
24V
OutputVoltageTolerance
±3%
OutputPower
400W max with 12CFM air cooling
LineRegulation (max)
±0.5%
LoadRegulation (max)
±1%
Ripple& Noise(typ.)
1%pk-pkVrated@ Full load, Reference Fig. 2
Start-upTime(max)
2000ms@115Vac
Hold-up Time (typ.)
16ms@115Vac
DynamicResponse(Overshoot&UndershootO/PVoltage)
±5% @50-100% load
Capacitive load (max)
1500uF for 400W
Rise time (max)
100ms
Remote Sense
Compensate up to 500mV lead drop with remote sense
Short and reverse connection protected. PSU can work
normally with remote sense pins left open.
(a) 115V (measured value=106mV)
(b) 230V(measured value=104mV)
Fig.2Ripple & Noise example, 20MHz BW
Ripple & Noise measurement circuit
PSU
Load
5cm
0.47uF Ceramic
10uF Electrolytic
Test Fixture
3
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (Jun 2015, Rev. 04)
Scope
Medical AC-DC Open Frame Power Supply
24V/400W 3” x 5” High Density MDS-400A Series
Mechanical
Case Cover
TBD
Dimensions (W x L x H typ.)
76.2 x 127 x 35.6 mm 3 x 5 x 1.4inch
Unit Weight
405g
Indicator
NA
Cooling System
NA
Environment
Surrounding Air Temperature
Operating
Storage
Absolute Max
-10°C to+70°C, supported power linearly de-rate from
50°C to 50% rated p to 70°C for forced air. Convection
power de-rate from 200W @ 40°C to 100W @ 60°C
Note: see power de-rating curve
-40°C to+85°C
Operating Humidity
5-95% RH (Non-Condensing)
Operating Altitude
5,000 meters
Shock Test (Non-Operating)
50G, 11ms, 3 shocks for each direction
Vibration (Operating)
5-500Hz, 2Grms, 15 minute for each three axis
Power De-rating curve
4
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (Jun 2015, Rev. 04)
Medical AC-DC Open Frame Power Supply
24V/400W 3” x 5” High Density MDS-400A Series
Protections
Overvoltage (max)
135% of rated voltage, Latch Mode
Overload / Overcurrent (max)
Main output 160% of ratedcurrent
Standby 3A max
Hiccup Mode(Non-Latching, Auto-Recovery)
Over Temperature
Latch Mode
Short Circuit
Hiccup Mode,
(Non-Latching, Auto-Recovery)
Reliability
MTBF(Minimum) at 115Vac, 400W, 35°C,
Convection A
12CFM Air Flow
Operating life at 115Vac, 400W, ambient 25 °C,12CFM Air Flow
800 kHrs based on Telecordia SR-332
26,280Hrs
Safety Standards / Directives
nd
rd
Medical Safety
IEC60601-12 and 3 edition CB report
TUV EN60601-1:2006
UL60601-1+CAN/CSA 60601-1: (Ed.3.2005)
ITE Safety
IEC60950-1 CB report
TUV60950-1
UL60950-1+CAN/CSA60950-1
GB4943.1-2011, GB9254-2008, GB17625.1-2003
CE
MDD Directive 93/42/EEC
Environmental
RoHS Directive 2011/65/EU Compliant
Galvanic Isolation
1)
5
Input to/Output (2XMOPP) 4000 Vac
1)
Input to/Ground(1XMOPP) 1500Vac
Output to/Ground(1XMOPP) 1500Vac ( Type BF application rated)
PSU can support PoE applications with Primary to FG 2500Vac test.
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (Jun 2015, Rev. 04)
Medical AC-DC Open Frame Power Supply
24V/400W 3” x 5” High Density MDS-400A Series
EMC
EMC / Emissions
Harmonic Current Emissions
ClassI model: EN55011, EN55022, FCC Title 47: Class B
IEC61000-3-2
Meet Class D limit
Immunity to
Voltage Flicker
IEC61000-3-3
Electrostatic Discharge
IEC61000-4-2
Level 3 Criteria A
Air Discharge: 8kV
Contact Discharge: 6kV
Radiated Field
IEC61000-4-3
Level 2 Criteria A
80MHz-2.5GHz, 3V/M with 1kHz tone / 80% modulation
Electrical Fast Transient / Burst
IEC61000-4-4
Level 3 Criteria A
2kV
Surge
IEC61000-4-5
Level 3 Criteria A
2)
Common Mode : 2kV
3)
Differential Mode : 1kV
Conducted
IEC61000-4-6
Level 2 Criteria A
150kHz-80MHz, 3Vrms
Power Frequency Magnetic Fields
IEC61000-4-8
Criteria A
Magnetic field strength 3A/Meter
Voltage Dips
Voltage Dips and interruption
IEC61000-4-11
IEC60601-1-2
1)
1)
1)
1)
1)
30% 10ms Criteria A; 60% 100ms and 100% 5000ms
Criteria B
1)
100% 10ms, Criteria A ;
2)
60% 100ms, Criteria B ;
2)
30% 500ms, Criteria B ;
2)
100% 5000ms, Criteria B ;
1) Criteria A: Normal performance within the specification limits
2) Criteria B: Output out of regulation, or shuts down during test. Automatically restored to normal operation after test.
3) Asymmetrical: Common mode (Line to earth)
3) Symmetrical: Differential mode (Line to line)
6
1)
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (Jun 2015, Rev. 04)
Medical AC-DC Open Frame Power Supply
24V/400W 3” x 5” High Density MDS-400A Series
Dimensions
W x L x H: 76.2 x 127.0 x 35.6 mm
Suggested air flow
direction, draw air
from input side to
output side
Mechanical drawing (3Pin input type)
Notes
―Dimensions are in mm(inches)
Input Connector CN1
Pin 1
Control and STANDBY connector CN101
Neutral
Pin 2
Line
CN1 mates with Molex housing 26033031 and
Molex series 6838 crimp terminals.
Output Connector
Pin 1
Remote sense +
Pin 2
Remote sense –
Pin 3
Power Good +
Pin 4
Power Good -(DC RTN)
Pin 5
Remote ON_OFF/INHIBIT +
Pin 6
Remote ON_OFF/INHIBIT -(DC RTN)
Pin 7
5V Standby
Pin 8
DC RTN
CN102
Vo
Pin 9
5V Standby
CN103
DC RTN
Pin 10
DC RTN
Fan Connector CN202
Pin1
CN101 mates with Molex housing 1041421000 and Molex series 104539 crimp
terminals.
12V Fan
Pin2
DC RTN
CN202 mates with JST housing PHR-2 and JST SPH-002T-P0.5S terminals.
Two mounting points in mechanical drawing need to be connected to system earth case together, Protective
bonding conductor from the end product protective earth terminal (if any) can be tied to CN2 for open frame
model.
7
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (Jun 2015, Rev. 04)
Medical AC-DC Open Frame Power Supply
24V/400W 3” x 5” High Density MDS-400A Series
Functions
Start-up Time
The time required for the output voltage to reach 90% of its set value, after the input voltage is applied.
Rise Time
The time required for the output voltage to change from 10% to 90% of its set value.
Hold-up Time
Hold up time is the time when the AC input collapses and output voltage retains regulation for a certain period of time. The time
required for the output to reach 90% of its set value, after the input voltage is removed.
■Graph illustrating the Start-up Time, Rise Time, and Hold-up Time
8
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (Jun 2015, Rev. 04)
Medical AC-DC Open Frame Power Supply
24V/400W 3” x 5” High Density MDS-400A Series
Dynamic Response
Short Circuit Protection
The power supply output voltage will remain within ±5% of its
steady state value, when subjected to a dynamic load 50 to 100%
of its rated current.
■ 50 to 100% Load
The power supply’s output OLP/OCP function also provides
protection against short circuits. When a short circuit is applied,
the output current will operate in “Hiccup mode”, as shown in the
illustration in the OLP/OCP section on this page. The power
supply will return to normal operation after the short circuit is
removed.
Inrush Current
Inrush current is the peak, instantaneous, input current measured
and, occurs when the input voltage is first applied. For AC input
voltages, the maximum peak value of inrush current will occur
during the first half cycle of the applied AC voltage. This peak
value decreases exponentially during subsequent cycles of AC
voltage.
Overload & Overcurrent Protections
The power supply’s Overload (OLP) and Overcurrent (OCP)
Protections will be activated before output current under 160% of
IO (Max load). Upon such occurrence, VO will start to drop. Once
the power supply has reached its maximum power limit, the
protection will be activated and the power supply will go into
“Hiccup mode” (Auto-Recovery). The power supply will recover
once the fault condition causing the OLP and OCP is removed
and IO is back within the specified limit.
Additionally, if the Iout is <160% but >100% for a prolong period
of time (depending on the load), the Over Temperature
Protection (OTP) will be activated due to high temperature on
critical components. The power supply will then go into
latchmode.
Over Temperature Protection
Overvoltage Protection
The power supply’s overvoltage circuit will be activated when its
internal feedback circuit fails. The output voltage shall not exceed
its specifications defined on Page 4 under “Protections”. Power
supply will latch off, and require removal/re-application of input
AC voltage in order to restart.
9
As mentioned above, the power supply also has Over
Temperature Protection (OTP). This is activated when the
overload condition persists for an extended duration and the
output current is below the overload trigger point but >100% load.
In the event of a higher operating temperature condition at 100%
load, the power supply will run into OTP when the surrounding air
temperature is higher than the operating temperature. When
activated, the output voltage will go into latch mode until the input
voltage is removed; then, reapplied, and the surrounding air
temperature drops to its normal operating temperature.
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (Jun 2015, Rev. 04)
Medical AC-DC Open Frame Power Supply
24V/400W 3” x 5” High Density MDS-400A Series
Power Good
Remote On_Off/Inhibit
Power Good+ pin is an open collector transistor (40V/600mA
rating). A resistor (suggested value 10Kohm, 1/8W) can be added
between 5V STANDBY pin (or, other available pull-up voltage
that is no greater than 30V) and the Power Good+ pin (refer to
figure below). Value of pull-up resistor may have to be adjusted,
depending on voltage used, and other end-use conditions of the
Power Good+ pin connection to the product. When AC input is
on, Power Good+ pin will be high. When AC input is off, Power
Good+ pin will be low. There will be a minimum of 3 milliseconds
between the time the power good goes to low level, and the time
when the output reaches 90% of its rated value.
Remote ON_OFF/INHIBIT can be used to enable or disable only
the main output. When the main output is disabled, the +5V
Standby output will continue to operate. This signal can be pulled
down to a low level of 0.3 volts, or shorted to DC-Return, in order
for the main output to be enabled; and, floated (no connection to
the signal), or pulled up to a value greater than or equal to 3 volts,
in order to disable the main output.
10
Remote Sense
Remote sense feature can be used to compensate for the extra
voltage drop on output wires that are connected from the main
output terminals, to the load. With wires connected from the
remote sense pins, at the same locations as the wires from the
main output, the remote sense function can compensate up to
500mV voltage drop. The power supply will not be damaged if the
remote sense pins are shorted, or if a reverse/inverted polarity
connection is made to the load.
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (Jun 2015, Rev. 04)
Medical AC-DC Open Frame Power Supply
24V/400W 3” x 5” High Density MDS-400A Series
Certificate
All Delta Medical Power products conform to the European directive 2011/65/EU.
RoHS is the abbreviation for “Restriction of the use of certain hazardous
substances
Delta has been certified as meeting the requirement of ISO 13485: 2003 and EN
ISO 13485:2012 for the design and manufacture of switching power supply and
adaptor for medical device.
In addition to a UL Total Certification Program (TCP) approved client laboratory for
IEC60950 and IEC60065. Delta also has participated UL Client Test Data Program
(CDTP) for IEC 60601
11
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (Jun 2015, Rev. 04)