HJ Density Measurement kit

HJ Density Measurement kit
HJ Density Measurement Option
Operation Manual
Introduction
Thank you for your purchasing the Density Measurement option. This Density Measurement option
Operation Manual is intended to describe only density measurements. Please use this operation
manual together with the operation manual of the main unit.
Check for the following items before use
Density measurement kit
Water tank platform
Measurement pan
Pan cradle
Pan in
空中皿
the air
Pan in
液中皿
the liquid
Water tank
Tweezers
Thermometer
Height adjuster
Contents
Page
1 Assembly Procedure for the Density Measurement Kit
1
2 Procedure for Use
3
3 Solid Density Measurement Mode
4
4 Function 1
6
5 Coefficient Setting
7
6 Display
8
7 Gravimeter measurement data output
8
1
1
Assembly Procedure for the Density Measurement Kit
1 Remove the measurement pan and
Pan base.
2
Attach the pan cradle
as shown to the right.
2
3 Attach the water tank platform
as shown to the right.
4 Put the water tank in
position as shown to the right.
5 Set the measurement pan in
position shown to the right.
3
2
Procedure for Use
2.1 Measuring solid density
1
Set-up Function 1. (See operation manual of the balance「4
Function 1」)
・Set the system in the solid density measurement mode [ ].
・Select the media to be used. If water is to be used, set [] to [].
If any other medium to be used, set to [].
・For enabling data output, set Output Data [] and Automatic Output
[].
2
Put water or other medium into the accessory tank as appropriate.
3
If water is used, measure the water temperature using accessory thermometer.
4
Set applicable coefficients ( See ”5 Coefficient Setting” ).
If the water is used
5 Set the density measurement kit and the water tank in the place. Turn off the balance
power and then turn it on again.
6
Measure the density of a solid object according to ”3
Solid Density measurement
mode”.
2.2 Precautions relating to Use
1
If water is used as medium, check the water temperature as frequently as appropriate.
2
Be careful not to spill water or other liquid on the balance. If liquids is
spilled on the balance, wipe off immediately with soft cloth.
3
If water in used as medium, add one or two drops of surface-active agent (liquid
detergent for kitchen use). It will effectively prevent air bubbles from adhering
to any measurement object.
œ Hints œ
Measurements with smaller size samples may result in some variations in the measured
specific gravity values. In particular, any sample whose specific gravity is larger tends to
show more variation.
Use as large a sample as possible for obtaining more stable measurements.
4
3 Solid Density Measurement Mode
If you want to interrupt an operation, press the Print key. The display will show
[] for a few seconds and then go back into the Weight mode.
3.1 Measurement for samples that sink down in the medium
1 Set the density measurement kit and press the ON/OFF
key to turn on the balance.
2 Push Zero/Tare key for zero adjustment or for
2
Zero/Tare
tare range.
d

 

 
g
3 Put a sample to weigh on the pan in the air.
4 When the safety mark() is displayed, press
3
d
g
the Set key.
4
Set
d
5

Push Zero/Tare key for zero adjustment or for
tare range.

5
Zero/Tare
d
6 Put the sample on the pan submerged in the medium.

Make sure the sample is free from air bubbles.
7 Push Set key when () mark is displayed.
g


6

Then measure the weight in the medium.
7

g
 
d
  g
Set
8 When the measurement is finished, the density is
displayed. The symbol [] is lit in the top of the
d
8
 
right of the display while the density is displayed.
Set
9
Press Set key to return to the weight display.
9
d
 
g
5
3.2 Measurement for samples that can float in the medium
1 Attach the height adjuster on top of the pan cradle
1
as shown to the right.
2 Set the density measurement kit and press the ON/OFF key
to turn on the balance.
3
Push Zero/Tare key for zero adjustment or for
tare range.
3
Zero/Tare

d
  g

d
  g
4 Put a sample to weight on the pan in the air.
5 When the safety mark() is displayed, press
4
Set key.
5
Set
6 When the measurement is done, the weight measured
In the air will be displayed for a few seconds.
6
d

7
Push Zero/Tare key for zero adjustment for
tare range.

7
8 Put the sample under the pan submerged in the medium.
g

Zero/Tare


d
 g
Make sure that the measurement pan is not floating,
and that the sample is free from air bubbles.
8

9 When the safety mark() is displayed, press Set key.
Measurement will be made in the water.
10 After the measurement is done, it displays the density.
The symbol  is lit on the display when the density
d
9

g
 
Set
10
d
 
is displayed.
Set
d
  g
6
4
Function 1
The following paragraphs describe Function 1 items to be set for the density measurement mode.
Also refer to “Section 6.6”of the operation manual of the balance.
4.1 Item details
Media
The medium to be used in sold density measurements is set for this item.
If it is specified to use water, it is necessary to set the water
temperature upon the measurement. Otherwise, it is necessary to set
the density of the alternative medium. Refer to “5.2 Coefficient Setting”.
Output data
Output items of measurement results are to be set. This item is
functional only for solid density measurement mode. When it is set to ,
only the display value of the sample is output. When set to ,
mass of the sample, water temperature (or the density of the medium if
any other medium is used) are printed in addition to the density value.
Refer to “7
Auto output
Print Details” for details of output data.
The timing to output measurement result is set. When “Auto Output”
is disabled, pressing the Print key upon weight display (after density
measurement is done) will print the results. The most recent measurement
before the output timing will be printed. Don’t press the Print key
after power on and before any density measurement is made. You cannot
expect normal values to be printed if no measurement has been made.
When “Auto Output” is enabled, the results will be printed upon returning
to the weight display after finishing the density measurement.
Regardless of the Auto Output setting, the items set in the above
“Output Data” setting will be printed.
7
5 Coefficient Setting
5.1
Coefficient setting items
The following coefficients must be set before density measurement. The setting values are
different depending on the mode and medium selected. Note that the values once set are held
in the balance even after power is turned off.
Water as media
Media other than Water
5.2
Setting value:Temperature of the water used
Displayed Unit:t
Setting Range:0.0∼99.9℃
Setting value:Density of the liquid used
Displayed Unit:(upper), (lower)
Setting Range:0.0001∼9.9999
Coefficient setting
1 Keep pressing the Zero/Tare key until the applicable
setting value (*1) is displayed
*1 Refer to “5.1
1
Coefficient setting items”
2 Press Zero/Tare key to initiate the numerical setting mode.
d
 g
Zero/Tare
dt
2

Zero/Tare
3 Every time you press the Zero/Tare key, the number of the
3
flashing digit(the digit for value entry) will increment.
Select an appropriate number by pressing the key as many
dt

Zero/Tare
times as required.
4
4 Press the Function key to move the flashing digit (the digit
dt

Function
for value entry).
5
5 Press the Set key liquid density, the item will be shifted
from “Water temperature (media density)” to “Mass of
dt

Set
weight”. Repeat the above steps 2 to 5.
d
 g
8
6
Display
6.1 Minimum indication
Minimum indication values of density is as follows; 0.0001
6.2
7
Unit notation
d
To be indicated in density measurement mode
(Upper)
To be indicated when density is displayed
(Lower)
To be indicated while density measurement is in progress
Gravimeter measurement data output
(1) Output before gravimeter measurement
With the gravimeter function, the operation is with the [  ] (output once when Print
key is pressed after it is stabilized) mode, regardless of the setting by [].
(2) Output while specific gravity is displayed
[ ] is used for selecting the output contents.
The output method can also be selected by [ ].
(3) Output (print) format
When "specific gravity, weight, and actual water temperature or density of media" are all selected
in [ ], …. If "specific gravity only" is selected, only the second line value will be output.
In addition, if any statistical calculation is made for the printer, the values in the second line will
be printed being prefixed by a serial number.
Language selection between English and Japanese for printing is made by the GLP function
[ ] of Function 1.
1)
1
When water is selected
1
2
3
4
5
6
7
D
E
N
S
I
T
Y
S
A M P
2
3
4
5
T
6
L
×
E M P E
×
8
×
E
W
× × ×
R A T
× × ×
9
10
11
12
13
14
15
S O L I D
× . × × ×
E I G H T
× . × × ×
g
U R E
N O W
× × × . ×
C
7
2)
1
When other than water is selected
1
2
3
4
5
6
7
D
E
N
S
I
T
Y
S
A M P
D
E
2
3
4
5
6
N
L
×
S I
×
8
×
E
W
× × ×
T Y
× × ×
9
10
11
12
13
S O L I D
× . × × ×
E I G H T
× . × × ×
M E D . L
. × × × ×
14
15
I
g
Q
7
9