Pulsed I-V Testing for Components and Semiconductor Devices

Pulsed I-V Testing for Components and Semiconductor Devices
www.keithley.com
appl icat i ons
gui de
Pulsed I-V Testing for Components and
Semiconductor Devices
a
g r e a t e r
m e a s u r e
o f
c o n f i d e n c e
Pulsed I-V Testing for Components and Semiconductor Devices
Pulsed I-V testing is ideal for preventing device self-heating or minimizing charge trapping effects
when characterizing devices. By using narrow pulses and/or low duty cycle pulses rather than DC
signals, important parameters are extracted while maintaining the DUT performance. Transient I-V
measurements allow scientists and engineers to capture ultra-high-speed current or voltage waveforms
in the time domain in order to study dynamic properties. This pulsed I-V testing applications e-guide
features a concentration of application notes on pulsed I-V testing methods and techniques using
Keithley’s Model 4200-SCS Parameter Analyzer.
Contents
Making Proper Electrical Connections to Ensure
Semiconductor Device Measurement Integrity . . . . . 3
An Ultra-Fast Single Pulse (UFSP) Technique for
Channel Effective Mobility Measurement . . . . . . . . . . 7
Performing Charge Pumping
Measurements with the Model 4200-SCS
Parameter Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Electrical Characterization of Carbon Nanotube
Transistors (CNT FETs) with the Model 4200-SCS
Parameter Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Pulsed Characterization of Charge-trapping
Behavior in High-k Gate Dielectrics . . . . . . . . . . . . . . 27
Pulse I-V Characterization of
Non-Volatile Memory Technologies . . . . . . . . . . . . . . 31
Electrical Characterization of Photovoltaic
Materials and Solar Cells with the
Model 4200-SCS Parameter Analyzer . . . . . . . . . . . . 57
www.keithley.com1
www.keithley.com
Making Proper Electrical Connections
to Ensure Semiconductor Device
Measurement Integrity
Introduction
VDS–IDS curves
The most straightforward way to assess a MOSFET’s overall
performance is to take V DS –IDS curves for a set of gate voltages
because these curves define the operating regions of the device.
Pulsed I-V characterization, wherein voltages and currents are
applied for a very short time and at a limited duty cycle, is a
common way to measure these curves. Pulsed I-V measurements
can reduce test times and allow characterizing a device without
exceeding its safe operating area or causing device self-heating
and the associated parameter shifts.
Two pulsed I-V channels are typically used to measure these
curves on a MOSFET, with one connected to the gate and the
other to the drain. The ground of each channel is connected to
the MOSFET source pin.
To construct the transistor curves, the gate channel first
applies voltage to the gate, then the drain channel sweeps V DS
through a range of values, measuring the resulting current at
each point. Next, the gate channel applies a different voltage to
the gate and the process repeats, constructing the next transistor
curve in the set.
Modern pulse instruments can produce very short voltage
pulses (100ns or less) with rise times as fast as 20ns, so wide
bandwidth is critical to obtaining good measurements. For
optimal performance, both pulse generator channels should be
connected with coaxial cables matched to the generator’s 50W
output impedance. However, it is impossible to maintain 50W
impedance all the way to a device on a wafer in practice; at some
point, the ground must be connected to one pin and the signal
to another, breaking 50W characteristic impedance (Figure 1).
Minimizing the length of these non-coaxial ground and signal
connections is crucial.
Figures 2a and 2b illustrate how inadequate grounding
and coaxial cabling can produce erroneous data when
ground return pa
th
Long
Prober
Bulkhead
To 50Ω Pulse
Instrument
Prober
Bulkhead
50 Ω Co
ax c
ab
le
ground wi
re
ort
Sh
Probe
Holder
50Ω coax
Wafer
Not 50Ω
50
Poor-quality electrical connections to the device under test
(DUT) can compromise the measurement integrity of even the
most powerful and sophisticated semiconductor test system.
For high speed pulse measurements, interconnect quality
typically determines maximum bandwidth; for low current
measurements, it often affects measurement speed and accuracy.
This application note discusses the problems created by
poor connections. Although it specifically addresses MOSFET
measurements, the techniques and results discussed also apply
to many other devices.
Ω
ble
ax ca
Co
To 50Ω Pulse
Instrument
Probe
Holder
50Ω coax
Figure 1. Prober hookups usually do not maintain 50W all the way to the
DUT. If the short ground wire is not used, the much longer ground return
path will limit bandwidth.
characterizing a MOSFET using pulses with 20ns rise times,
20ns fall times, and a width of 200ns. Figure 2a shows data
taken with 100W impedance triaxial cables and 1–2 ft. ground
connections. Figure 2b shows data taken with 50W impedance
coaxial cables and ground connections just a few inches long.
The difference is striking: the data from an improperly cabled
system are compressed (Figure 2a), measuring about half the
level expected, while the data from the correctly cabled system
(Figure 2b) match conventionally measured data.
Maximum gm and V TH
Transconductance (gm) is a critical parameter widely used to
determine the threshold voltage (V TH) of MOSFETs [1]. Pulsed I-V
testing is ideal for this application because these parameters can
be determined without violating safe operating area (SOA) limits
or damaging the device.
For a small signal analysis about a given gate and drain bias
point, gm is defined as:
∂iDS
∂iDS
gm =_____
gm = _____
∂VGS
∂VGS(1)
When measuring gm, the same connection scheme used to
obtain the data in Figures 2A and 2B is used, but this time a
DC voltage is applied to the drain while the gate voltage (VGS)
is swept over the voltage range that transitions the device from
off to on. Because the connections and pulsing speeds are the
same, the earlier bandwidth discussion applies to gm and V TH
testing as well.
Figures 3a and 3b show the results of another example of
gm and V TH measurement, with the data shown in each taken
with the same pulse rise and fall times, pulse widths, connections, and grounding. As Eq. 1 states, calculating gm requires
www.keithley.com3
80.0E–03
70.0E–03
60.0E–03
50.0E–03
IDS (A)
IDS at 3V VGS = 34mA
40.0E–03
30.0E–03
20.0E–03
10.0E–03
00.0E–03
0.0
0.5
1.0
1.5
2.0
VDS (V)
Figure 2A. MOSFET curves with poor cabling and grounding. IDS at a VGS of 3V is measured as 34mA.
IDS at 3V VGS = 72mA
80.0E–03
70.0E–03
50.0E–03
40.0E–03
30.0E–03
20.0E–03
10.0E–03
00.0E–03
0.0
0.5
1.0
1.5
2.0
VDS (V)
Figure 2B. MOSFET curves with optimal cabling and grounding. IDS at a VGS of 3V is measured as 72mA.
computing the first derivative of the drain
current with respect to the gate voltage.
The blue curve is the drain current and
the red curve is its calculated derivative.
Just as with the V DS –IDS curves, the
difference is significant. The results produced using the improperly cabled setup
would mislead the user to think the DUT’s
gm is half its actual value and its V TH
roughly 200mV higher than it actually is.
Capacitance-Voltage Testing
Capacitance-voltage (C-V) measurements are often used to characterize a
MOSFET’s gate oxide thickness, oxide
defect density, doping profiles, etc. In this
measurement, as the gate voltage varies,
the capacitance of the gate to the drain
and source changes. The Model 4210-CVU
www.keithley.com
Although using cables longer than
those provided with the instrument will
cause large changes in data at high frequencies, a feature known as cable length
compensation can mitigate this problem.
Using cables of different characteristic
impedances, such as when coaxial cables
are improperly adapted to triaxial cables,
will also produce increased measurement
error at high frequencies.
Off-State Leakage
Characterizing off-state leakage is critical
to understanding quiescent power dissipation and transistor quality for small-scale
parameter transistors. A source measure
unit (SMU) instrument capable of measuring picoamp-level (1E–12A) currents is
essential.
60.0E–03
IDS (A)
significantly. For optimal C-V measurements, DUTs must be connected to the
instrumentation using coaxial cables of
the proper length and impedance level;
for example, the Model 4210-CVU uses
1.5-meter, 100W cables.
option allows Keithley’s Model 4200-SCS
Parameter Analyzer to make 1kHz–10MHz
C-V measurements. It has four terminals
wired to two device connections: CVU
high and CVU low. In a standard C-V
measurement, the source, bulk, and drain
of a MOSFET are connected together and
tied to CVU low; the gate is connected to
CVU high. The Model 4210-CVU applies a
DC bias and a small AC voltage to bias the
transistor and simultaneously measure its
capacitance, returning capacitance values
for a wide range of bias voltages.
C-V measurements can be made at
many frequencies (sometimes greater
than 1MHz), depending on the parameters to be extracted. At higher frequencies, transmission line effects and cable
length can impact measurement integrity
As with any low current measurement,
properly guarded triaxial cables are essential. Using coaxial cables will produce
erroneous results due to high cable capacitance (50W coaxial cables typically have
30pF of capacitance per foot). This capacitance must be charged with high currents
before small currents flowing in the DUT
can be measured, extending settling times
substantially.
Triaxial cables alleviate this problem
by providing a third conductor between
the center pin and shell, called the
guard, driven at the same potential as the
center pin. The guard conductor drastically reduces the effective capacitance
and reduces settling times. Never use
poor-quality triaxial cables, which can
introduce problematic dielectric absorption, triboelectric effects, or high noise.
Unguarded cable segments should be
minimized. For best performance, triaxial
cables should be run as close to the DUT
as possible.
Match the Cable to
the Measurement
As these examples illustrate, matching
cabling and grounding to the measurement type enhances measurement integrity. However, changing cables for each
measurement type is so time-­consuming
many users simply tolerate the sub-­
optimal results. Moreover, whenever
cables are rearranged, users run the risk
of reconnecting them improperly, thereby
causing errors and demanding extra
­troubleshooting time. Worse still, these
errors may go unnoticed for a long time.
One alternative is to use a remote
switch capable of handling I-V, C-V and
pulsed I-V signals, such as Keithley’s
Model 4225-RPM Remote Amplifier/
Switch. When combined with a multi-­
measurement performance cable kit,
such as Keithley’s 4210-MMPC kit, it
can often eliminate the need to recable
between tests. This kit delivers the correct impedance and cable type, allowing
low c­ urrent, high speed pulse, and C-V
measurements with no manipulator
reconnection, eliminating the need to
reprobe wafers.
Given the importance of characterizing semiconductor devices quickly and
accurately, the value that the latest cable
solutions can provide to device researchers is obvious.
References
1. X. Zhou, K. Y. Lim, and D. Lim. A Simple and
Unambiguous Definition of Threshold Voltage
and Its Implications in Deep-Submicron MOS
Device Modeling. IEEE Transactions on Electron
Devices, Vol 46, No. 4, p. 807. April 1999.
0.06
0.09
0.08
0.05
0.07
0.04
0.06
Max gm = 0.45 S
IDS (A)
0.05
0.03
0.04
0.02
gm (S)
0.03
0.02
0.01
0.01
0.00
VTH = 2.88V
1.0
1.5
2.0
2.5
3.0
3.5
4.0
0.00
VGS (V)
Figure 3a. gm test with poor cabling and grounding.
0.06
0.09
Max gm = 0.81 S
0.08
0.05
0.07
0.04
IDS (A)
0.06
0.05
0.03
0.04
0.02
gm (S)
0.03
0.02
0.01
0.01
0.00
VTH = 2.70V
1.0
1.5
2.0
2.5
3.0
3.5
4.0
0.00
VGS (V)
Figure 3b. gm test with optimal cabling and grounding. From this measurement, g m is 0.81 S and
V TH is 2.70V.
www.keithley.com5
www.keithley.com
An Ultra-Fast Single Pulse (UFSP)
Technique for Channel Effective
Mobility Measurement
Introduction
(a)
The channel effective mobility (µeff ) influences the MOSFET
performance through the carrier velocity and the driving current.
It is one of the key parameters for complementary metal-oxidesemiconductor (CMOS) technologies. It is widely used for
benchmarking different processes in technology development
and material selection [1, 2]. It is also a fundamental parameter
for device modelling [3]. With device scaling down to Nanosize regime and the introduction of new dielectric materials,
conventional measurement technique for mobility evaluation
encountered a number of problems described in the following
section, leading to significant measurement errors. As a result, a
new mobility extraction technique is needed.
This application note describes a novel Ultra-Fast Single
Pulse technique (UFSP) [4, 5] for accurate mobility evaluation,
including the technique principle, how to connect the device,
and how to use the provided software in the Model 4200-SCS
Parameter Analyzer.
Figure 1. Configuration for (a) conduction current measurement and (b) gateto-channel capacitance, Cgc , measurement.
Qi measurement. This difference in Vd used in two measurements
can lead to significant errors in evaluating mobility for thin
oxides, especially in the low electric field region. One example
is given in Figure 2, where a higher |Vd| results in a substantial
reduction of mobility near its peak. This is because |Vg–Vd|
reduces for high |Vd|, so that the real charge carrier density for
the Ich is smaller than the Qi measured at Vd = 0.
120
Conventional Mobility Measurement
and Challenges
Vg
(1)
where Vd is a small bias applied on the drain terminal of the
device, Qi is the mobile channel charge density (C/cm2), and Ich is
the conduction current flowing in the channel.
Traditionally, Ich is measured at the drain terminal of
the device with the configuration shown in Figure 1(a). Qi
is extracted from integrating the measured gate-to-channel
capacitance, Cgc, with respect to Vg, i.e.,
by using the connection configuration shown in Figure 1(b).
The principle of conventional mobility measurement is
deceptively simple. However, many challenges and pitfalls are
associated with this testing. Several sources of error are often
ignored in the past.
Vd-dependence: The conventional technique applies a non-zero
Vd (usually 50mV–100mV) for Ich measurement but a zero Vd for
μeff (cm2/V-s)
100
We use a p-channel device of gate length L and width W as an
example. When the channel charge is fairly uniform from source
to drain in the linear region, the channel effective mobility (µeff )
can be written as
L Ich
µeff = ___·——–—
WQi · Vd
(b)
Id ~ Vg config
Vd
A
80
Vg
60
40
~
Ccg ~ Vg config
A
Vd = -25 mV
Vd = -50 mV
Vd = -100 mV
20
0
0
2
4
6
8
10
12
Ns (x1012 cm-2)
Figure 2. Effective channel mobility measured by conventional technique.
Ich was measured under various non-zero drain biases, V DS, but Qi was
measured under Vd = 0. The extracted mobility clearly reduces for higher |Vd|.
Insets illustrate the carrier distribution in the channel.
Charge trapping: The conventional technique used slow
measurement with typical measurement time in seconds. The
fast charge trapping becomes significant for both thin SiON and
high-k dielectric. For slow measurements, trapping can respond
during the measurement and give rising to hysteresis and stretchout of the Cgc–Vg curve and a reduction of Ich. This results in an
underestimation of mobility.
www.keithley.com7
Leaky dielectric: As gate oxide downscales, high gate leakage
current becomes a main challenge for mobility extraction. It
affects both Ich and Qi measurements and in turn the mobility.
To minimize its impact on Cgc measurement, frequency up
to gigahertz has been used, which requires devices with RF
structure. The RF structure requires more processing and die
space and is not always available.
Cable switching: The conventional technique involves cable
changing between Ich and Qi measurements. This slows down
the measurement and can potentially cause breakdown of the
device under test.
(a)
(b)
The Ultra-Fast Single Pulse
Technique (UFSP Technique)
To overcome the challenges mentioned above, a novel technique
called the Ultra-Fast Single Pulse technique (UFSP) is developed
and described below.
A p-channel device is used here for illustrating the working
principle of the UFSP technique as shown in Figure 3. The
considerations for n-channel devices are similar. To perform
the UFSP measurement, a single pulse with edge time of several
micro-seconds is applied on the gate terminal of the device. The
gate voltage sweeps toward negative during the falling edge
of the pulse and turns the device on. The transient currents
are recorded at both the source and the drain terminal of the
device. The device is then switched off during the subsequent
rising edge where gate voltage are swept toward positive. The
corresponding transient currents are also to be recorded.
Channel effective mobility can be extracted from these four
transient currents measured within several micro-seconds.
Figure 4. Schematic diagram of current flow during the transient
measurement.
direction as Ich at the source, but Idis_d is in opposite direction to
Ich at the drain. When device is switched on-to-off, Idis_s and Idis_d
change direction, but Ich does not. Ig_s and Ig_d are independent
of Vg sweep direction and always flow from the source and drain
towards gate under negative Vg. Based on the above analysis,
channel current, Ich, gate current, Ig, and displacement current,
Idis can be separated by using Equations (2)–(4). Cgc can be
calculated using (5).
(2)
(3)
0V
A
A
p+
p+
n-sub
Figure 3. Illustration of the working principle of UFSP technique.
To facilitate the analysis, we define currents measured at
drain and source terminal during switching on and off as Idon,
Ison, Idoff, and Isoff. The current flow in the channel during the
transient measurement is shown in Figure 4 (a) and (b). Three
types of current are present: channel conduction current, Ich,
displacement current between gate and source/drain, Idis_s and
Idis_d, and the leakage current between gate and source/drain,
Ig_s and Ig_d. When device is switched off-to-on, the direction of
Idis_s and Idis_d is toward the channel centre: Idis_s has the same
www.keithley.com
Vd =
-100mV
Vg
(4)
(5)
To calibrate the UFSP technique, a p-channel MOSFET
with thick oxide is used which has negligible IG current. The
measurement time (=edge time) is set at 3µs. the measured four
currents are shown in Figure 5. The extracted Ich, Ig and Cgc
by using Equations (2) to (5) are shown in Figure 6(a). Once
Cgc and Ich are evaluated accurately, Qi can be obtained by
integrating Cgc against Vg and channel effective mobility, μeff, is
calculated through Equation (1) as shown in Figure 6(b).
Since the UFSP measured Ich and Cgc under the same Vd, µeff
should be independent of Vd. The µeff evaluated under three
different Vd biases is compared in Figure 7. Good agreements
are obtained confirming the errors induced by Vd for the
conventional techniques has been removed.
The UFSP also works well on leaky gate dielectric of standard
structure. When it was applied on one ‘leaky’ n-channel MOSFET
with an EOT of 1.28nm, the four currents measured from the
μeff (cm2/V-s)
100
80
60
40
Vd = -25 mV
20
Vd = -50 mV
Vd = -100 mV
0
0
2
4
6
Ns
Figure 5. Four currents measured from source and drain corresponding to the
off-to-on and on-to-off Vg sweep. Schematic Vg waveform is shown in inset.
8
(x1012
10
12
cm-2)
Figure 7. The effective channel mobility, µeff, extracted under three different
Vd by using UFSP technique.
(a)
50.0
(a)
nMOSFET. EOT = 1.28nm
W/L = 10µm/10µm
Vd = +50mV
45.0
40.0
Current (μA)
35.0
30.0
Idon
25.0
Ison
Idoff
20.0
15.0
Is
10.0
Isoff
Ison
off
Idoff
Idon
5.0
0.0
0
1
2
3
Vg (V)
(b)
Ich or Ig (μA)
40
(b)
120
20
80
300
60
3
2.5
2
Cgc
1.5
Ich
10
100
0
1
Ig
1
0.5
0
2
Vg (V)
(c)
250
40
μeff (cm2/V-s)
μeff (cm2/V-s)
30
0
3.5
nMOSFET. EOT = 1.28nm
Vd = +50mV
Cgc (µF/cm2)
50
20
0
0
2
4
6
8
10
12
200
150
100
nMOSFET
EOT = 1.28nm
Vd = 50mV
50
0
Ns (x1012 cm-2)
0
10
20
Ns (x1012 cm-2)
30
Figure 6.
Figure 8.
(a). Ich, Ig , and Cgc extracted simultaneously from the currents in Figure 5 by
using Equations (2)–(5).
(a) Four currents measured from the source and drain corresponding to the
off-to-on and on-to-off Vg sweeps by UFSP technique on an nMOSFET with
EOT of 1.28nm.
(b) Channel effective mobility extracted from Ich and Cgc from (a).
(b) Ich (‘n’), Ig (‘o’) and Cgc (‘x’) are extracted from the currents in (a) with
Equations (2)-(5). The blue line is the leakage current obtained by DC
measurement.
(c) Channel effective mobility, µeff, is calculated by using the extracted Ich
and Cgc with Eqn (1).
www.keithley.com9
source and drain terminals corresponding to the off-to-on and
on-to-off VG sweep are shown in Figure 8(a). By using Equations
(2)–(5), Ich (‘n’), Ig (‘o’) and Cgc (‘x’) are extracted and plotted
in Figure 8(b). Ig from DC measurement is also plotted for
comparison in Figure 8(b). Good agreement is obtained. Figure
8(c) shows that electron mobility can be reliably measured for
this leaky device where Ig is as high as 45A/cm2. Since the UFSP
can tolerate high gate leakage, it does not need using the special
RF structure for mobility evaluation.
To demonstrate the applicability of UFSP to devices with
significant charge trapping, one pMOSFET with HfO2/SiO2 stack
was used. Large amount of traps locate close to the Si/SiO2
interface in this dielectric stack and they can exchange charges
with substrate rapidly. The conventional technique takes seconds,
making them indistinguishable from channel mobile charges.
As a result, inversion charges will be overestimated and in turn
the channel effective mobility will be underestimated. The UFSP
technique only takes microseconds, minimizing charge trapping
effect. Figure 9 compares the mobility extracted by these two
techniques. It clearly shows that after suppressing the trapping,
the mobility extracted from the UFSP is considerably higher than
that by the conventional technique.
100
7% increases
@ low E
μeff (cm2/V-s)
80
60
20% increases
@ high E
40
0
1
4
7
10 13 16
Ns (x1012 cm-2)
19
22
25
Figure 9. A comparison of mobility extracted by UFSP and conventional
technique for a device with HfO2/SiON dielectric of considerable
fast trapping.
Required Hardware for UFSP Measurement
Selecting appropriate measurement equipment is critical to the
success implementation of ultra-fast single pulse method. The
following hardwares are required: Two Keithley Ultra-Fast I-V
Modules (4225-PMU);
• Two Keithley Ultra-Fast I-V Modules (4225-PMU);
• Four Remote Amplifier/Switch (4225-RPM);
• 4 sets of high Performance Triaxial Cable Kit (4210-MMPC-C).
A photo of the cabling configuration for the test is shown
in Figure 10. 4225-PMU is the latest addition to the growing
range of instrumentation options for the Model 4200-SCS
www.keithley.com
Parameter Analyzer. The module integrates ultra-fast voltage
waveform generation and signal observation capabilities into the
Model 4200-SCS’s already powerful test environment to deliver
unprecedented I-V testing performance. It makes ultra-fast I-V
sourcing and measurement as easy as making DC measurements
with a traditional high resolution source measure unit (SMU)
instrument. Each plug-in Model 4225-PMU module provides two
channels of integrated sourcing and measurement. Each channel
of the Model 4225-PMU combines high speed voltage outputs
(with pulse widths ranging from 60 nanoseconds to DC) with
simultaneous current and voltage measurements. 4225-RPM
Remote Amplifier/Switch further expands the Model 4225-PMU’s
capabilities by providing ultra-low current measurement (below
100 nA) and reducing cable capacitance effects.
Connections to the Device
UFSP technique
Conventional techique
20
Figure 10. Photo of the UFSP technique setup using Keithley instruments.
The connection for the UFSP measurement is shown in Figure
11. Each terminal of the device is connected to one 4225RPM using two 11-inch blue cables (provided in the cable set
4210-MMPC-C). Then each 4225-RPM is connected to one
channel of PMU using two triaxial cables. All the measurements
are controlled by the Keithley KTEI software.
Using the KTEI Software to Perform
UFSP measurement
Performing UFSP for channel effective mobility
measurement using Keithley 4200-SCS system is quite
simple. One example project can be downloaded from
http://www.keithley.com/data?asset=57747. As shown in Figure
12, each terminal of the device is connected to one channel of
PMU. Users can modify the parameters for each PMU channel
in the definition tab. Table 1 lists one set of user-defined
parameters for a p-channel MOSFET.
In the timing tab, users can input the desired measurement
speed which is the edge time of the pulse. The recommended
values are listed in Table 2.
Figure 11. Experiment connection for the Ultra-fast Single Pulse (UFSP) technique. Two Keithley dual-channel 4225-PMUs are used for performing transient
measurements. Four Keithley 4225-RPMs are used to reduce cable capacitance effect and achieve accurate measurement below 100nA.
Figure 12. Example project in KTEI software for UFSP measurement. Each of the four terminals of the device is connected to one channel of PMU respectively.
www.keithley.com11
Table 1. Recommended setting in the definition tab for each channel of PMU.
PMU Setting for Gate Terminal
Parameters
Value
Description
Forcing Function
Pulse Train
To generate a single pulse or a pulse train with same shape
Pulse Train Settings
Voltage Amplitude
–2V
To define the Vg sweep range
Voltage Base
0V
Vrange
10V
Maximum possible voltage applied on the gate
Measurement Range
Irange
10µA
Measurement range for current
Sample I waveform
untick
Do not record current at the gate
Sample V waveform
tick
Record applied voltage at the gate
Measurement Setting
Timestamp
tick
Record total time for the measurement
PMU Setting for Drain Terminal
Value
Description
Forcing Function
Pulse Train To generate a single pulse or a pulse train with the same shape
Pulse Train Settings
DC voltage
To apply a constant Vd bias used for mobility measurement
Voltage base (V)
–0.1
Vrange
10V
Maximum possible voltage applied on the gate
Irange
10µA
Measurement range for current
Sample I waveform
tick
Record current at the drain
Sample V waveform
untick
Do not record applied voltage at the drain
Timestamp
untick
Do not record total time for the measurement
Parameters
Pulse Train Settings
Measurement Range
Measurement Setting
PMU Setting for Source Terminal
Value
Description
Forcing Function
Pulse Train To generate a single pulse or a pulse train with the same shape
Pulse Train Settings
DC voltage
To apply a zero Vs bias used for mobility measurement
Voltage base (V)
0
Vrange
10V
Maximum possible voltage applied on the gate
Irange
10µA
Measurement range for current
Sample I waveform
tick
Record current at the source
Sample V waveform
untick
Do not record applied voltage at the source
Timestamp
untick
Do not record total time for the measurement
Parameters
Measurement Range
Measurement Setting
PMU Setting for Bulk Terminal
Value
Description
Forcing Function
Pulse Train To generate a single pulse or a pulse train with the same shape
Pulse Train Settings
DC voltage
To apply a zero Vbulk bias used for mobility measurement
Voltage base (V)
0
Sample I waveform
untick
Do not record current at the bulk
Sample V waveform
untick
Do not record applied voltage at the bulk
Timestamp
untick
Do not record total time for the measurement
Parameters
Measurement Setting
www.keithley.com
Table 2. Recommended setting in the timing tab.
Parameters
Test Mode
Value
Waveform capture
Measurement Mode
Discrete Pulses
Sweep parameter
Period (s)
Width (s)
Rise Time (s)
Fall Time (s)
Pulse Delay (s)
None
5.00E-05
6.00E-06
3.00E-06
3.00E-06
2.00E-06
Description
Discrete Pulse and Average pulses, then you need to
input number of Pulses, 10 is enough.
No sweeping required
Period of the pulse
Pulse width
Pulse rise time
Pulse fall time, set to be the same as rise time
Pulse delay time, keep the same as rise time
Once the test is executed, transient currents during switching
on and off at source and drain terminals will be recorded and
stored in the sheet tab and can be saved as an .xls file. These
currents can also be plotted on the graph tab. From these
currents, the channel effective mobility can be extracted based
on Equations (2) to (5).
Conclusion
Channel carrier mobility is a key parameter for material selection
and process development. The conventional technique suffers
from several shortcomings: slow speed and vulnerability to
fast trapping, Vd-dependence, cable-changing, sensitivity to
gate leakage, and complex procedure. An ultra-fast single
pulse technique (UFSP) has been proposed and developed to
overcome these shortcomings. ICH and Qi can be simultaneously
measured within several micro-seconds without cable switching.
UFSP measurement can be easily performed using the Keithley
4200-SCS Parameter Analyzer with four 4225-RPMs. It provides a
complete solution for robust and accurate mobility evaluation in
a convenient way and serves as a tool for process development,
material selection, and device modelling for CMOS technologies.
References
1. P. R. Chidambaram, C. Bowen, S. Chakravarthi, C. Machala, and R. Wise,
“Fundamentals of silicon material properties for successful exploitation of
strain engineering in modern CMOS manufacturing,” IEEE Trans. Electron
Dev., vol. 53, no. 5, pp. 944-964, 2006.
2. R. Chau, S. Datta, M. Doczy, B. Doyle, J. Kavalieros, and M. Metz, “Highkappa/metal-gate stack and its MOSFET characteristics,” IEEE Electron Dev.
Lett., vol. 25, no. 6, pp. 408-410, Jun, 2004.
3. K. Chain, J. Huang, J. Duster, K. K. Ping, and C. Hu, “A MOSFET electron
mobility model of wide temperature range (77 - 400 K) for IC simulation,”
Semiconductor Science and Technology, vol. 12, no. 4, pp. 355, 1997.
4. Z. Ji, J. F. Zhang and W. Zhang, “A New Mobility Extraction Technique Based
on Simultaneous Ultrafast Id-Vg and Ccg-Vg Measurements in MOSFETs,”
IEEE Trans. Electron Dev., vol. 59, no. 7, pp. 1906, 2012.
5. Z. Ji, J. Gillbert, J. F. Zhang and W. Zhang, “A new Ultra-Fast Single Pulse
technique (UFSP) for channel effective mobility evaluation in MOSFETs,”
IEEE Int. Conf. Microelectronic Test Structures, pp. 64, 2013.
Acknowledgements
Author: Dr. Zhigang Ji, School of Engineering, Liverpool John Moores University
www.keithley.com13
www.keithley.com
Performing Charge Pumping
Measurements with the
Model 4200-SCS Parameter Analyzer
Introduction
HI
Charge pumping (CP) is a well-known measurement technique
for analyzing the semiconductor–dielectric interface of MOS
structures. Important information about the quality and
degradation of a device can be extracted from charge pumping
current (ICP) measurement results, including the interface trap
density and the mean capture cross section. Pulsing a gate
voltage and measuring a DC substrate current simultaneously is
the basis for the various charge pumping methods, so a pulse
generator and sensitive DC ammeter are required to make
these measurements.
The Model 4200-SCS Parameter Analyzer offers a complete
solution for charge pumping measurements because it contains
the necessary hardware to make the sensitive measurements,
as well as software to automate the measurements and analyze
the results. This system is provided with predefined tests for
making most of the common charge pumping tests, such as a
pulsed base voltage sweep or a pulsed voltage amplitude sweep.
This application note explains how to make charge pumping
measurements using the Model 4200-SCS with the optional
Model 4225-PMU Ultra Fast I-V Module (PMU) or Model 4220PGU Pulse Generator Unit (PGU).
Charge Pumping Overview
Figure 1 is a charge pumping measurement circuit diagram.
Basically, the gate of the MOSFET is connected to a pulse
generator, which repeatedly switches the transistor from
accumulation to inversion. While the gate is pulsed, a
recombination process of majority/minority carriers occurs on
the rising and falling edges of the pulses. This causes a current
to flow in the opposite direction of the normal drain-to-source
current. This induced current is known as the charge pumping
current (ICP) and can be measured by connecting a sensitive
ammeter to the substrate, or bulk terminal, of the MOSFET.
Although several charge pumping methods have been
developed, the basic charge pumping technique involves
measuring the substrate current while applying voltage pulses
of fixed amplitude, rise time, and frequency to the gate of the
transistor. The source and drain are either tied to ground or
slightly reverse-biased. The voltage pulse can be applied with a
fixed amplitude while sweeping the base voltage or with a fixed
base voltage while sweeping the amplitude of the pulse.
In the fixed-amplitude/voltage-base sweep, the amplitude
and period (width) of the pulse are kept constant while the base
voltage is swept from inversion to accumulation. This waveform
Gate
Source
Drain
n+
n+
p
Bulk
HI
ICP
A
Figure 1. Basic charge pumping measurement circuit
and the corresponding curve of the charge pumping current
shown as a function of the base voltage are both illustrated in
Figure 2. From the data, it’s possible to extract the interface trap
density (Nit) using this equation:
I____
Nit = CP
qfA
where:
Nit= interface trap charge density (cm–2)
ICP= charge pumping current (A)
f = test frequency (Hz)
q = electron charge, 1.6022 × 10 –19 C
A = channel area (cm2)
The interface trap density as a function of band bending can
also be extracted from the following equation:
I
CP
Dit = _______
qfADE
where:
Dit = interface trap charge density (cm–2eV–1)
ICP = charge pumping current (A)
f
= test frequency (Hz)
q = electron charge, 1.6022 × 10 –19 C
A = channel area (cm2)
ΔE = the difference between the inversion Fermi level and
the accumulation Fermi level [1]
www.keithley.com15
Hardware Configuration
VG
Vthreshold
VFlatBand
The pulser (4225-PMU or 4220-PGU) is connected to the gate
of the MOSFET in order to apply pulses of sufficient amplitude to
drive the device between inversion and accumulation. Depending
on the charge pumping method, the PGU or PMU can sweep the
pulse amplitude, sweep the base voltage, vary the rise/fall time,
and vary the test frequency. The test frequency is usually in the
kilohertz to megahertz range.
VBASE
Depletion
ICP
Accum.
Inversion
VBASE
Figure 2. Pulse waveform for fixed-amplitude/voltage-base sweep and
corresponding charge pumping current curve
The fixed-base/variable-amplitude sweep method is another
common technique for determining the charge pumping
current. With this method, the base voltage is kept constant in
accumulation and the variable voltage amplitude is pulsed into
inversion. As shown in Figure 3, as the voltage amplitude (VAMP)
of the pulses increases, the charge pumping current saturates
and stays saturated.
VG
Figure 4 is the basic circuit diagram for making charge pumping
measurements using the Model 4200-SCS. For this application,
the Model 4200-SCS is configured with either a Model 4220-PGU
Pulse Generator Unit (PGU) or a Model 4225-PMU Ultra Fast I-V
Module (PMU), one or two Model 4200-SMU Source Measure
Unit (SMU) Instruments, and one Model 4200-PA Preamp.
VAMP
Vthreshold
VFlatBand
SMU1 is connected to the Bulk terminal and measures the
resulting substrate current. This charge pumping current (ICP)
is often in the nanoamp or picoamp range. For measuring
currents of less than one nanoamp, the optional Model 4200-PA
should be used.
The source and drain terminals of the MOSFET are tied
together and connected to SMU2, which applies a slight reversebias (Vr). If Vr = 0, then the source/drain terminals can be
connected to the ground unit (GNDU) instead of to SMU2. To
prevent oscillations and minimize noise, it is very important to
connect the LO (common) terminals of all the SMU instruments
and the pulser (4225-PMU or 4220-PGU) as close as possible to
the device. The LO terminal of the SMU instrument is the outside
shell of the triax connector. The LO terminal of the PMU and
PGU is the outside shield of the SMA cable.
To minimize noise in low current measurements due to
electrostatic interference, make sure the device is shielded by
placing it in a metal enclosure with the shield connected to the
LO terminal of the SMU instrument. Further information on
making low current measurements with the Model 4200-SCS is
VBASE
Drain
Gate
Bulk
ICP
Source
ICP
Channel 1
VAMP
50Ω
Force HI
Force HI
A
A
4200-SMU
4200-SMU
SMU2 Voltage Source
supplies reverse bias to
Source/Drain, Vr or
SMU2 can be set
to Common
SMU1 with Preamp
measures ICP
Figure 3. Pulse waveform for fixed-base/variable-amplitude sweep with
corresponding charge pumping current curve
Other charge pumping techniques are used in addition to
the fixed-amplitude/variable-base sweep and the fixed-base/
variable-amplitude sweep. In some cases, the voltage waveform
can have various shapes, the rise and fall times can be varied,
or the charge pumping current can be measured as a function
of frequency.
www.keithley.com
4220-PGU
or
4225-PMU
PGU or PMU applies
pulses to Gate, VG
Figure 4. Model 4200-SCS configuration for charge pumping measurements
available in Keithley Application Note Number 2959, “Optimizing
Low Current Measurements with the Model 4200-SCS
Semiconductor Characterization System.”
project is opened, a list of the tests is displayed in the Project
Navigator (Figure 5). Table 1 lists the user test modules (UTMs)
included in the project and a brief description of each test.
Using the KITE Software to Automate
the Charge Pumping Measurements
The user selects the desired test and then inputs the
appropriate values for the test parameters on the Definition Tab.
The parameters vary depending on the particular test, but they
usually include the magnitude of the pulse, sweep values, rise/
fall time, test frequency, duty cycle, etc. Specific information on
these tests, including the test parameters, is available in Section
16 of the Model 4200-SCS Complete Reference Manual.
The Model 4200-SCS comes with a project that contains a
library of tests used in many of the common charge pumping
measurement techniques. This chargepumping project is located
in the projects folder: C:\S4200\kiuser\Projects\_Pulse. When this
After the hardware and the software have been configured,
the measurement can be executed in the project by clicking the
on-screen Run button. The measurements are displayed in the
Graph tab and listed in the Sheet tab. The Sheet tab is used to
record and manipulate the data. The measurements (ICP, QCP,
pulsed voltage, etc.) can be saved in a worksheet as an .xls, .txt,
or .csv file.
The graph from executing the BaseSweep user test module is
shown in Figure 6. This user test module measures the charge
pumping current as a function of the base voltage.
Charge Pumping Current (ICP) vs. Base Voltage (V)
15.0E–9
14.0E–9
13.0E–9
12.0E–9
11.0E–9
10.0E–9
Figure 5. The chargepumping project
Table 1. Charge pumping user test modules
User Test Modules
BaseSweep
Description
The base voltage of the waveform is swept while the
amplitude of the pulse is kept constant. The resulting
charge pumping current is measured and graphed
as a function of the base voltage. The source/drain
terminals are tied to ground.
BaseSweep_2SMU
Same as BaseSweep test, except it adds a second SMU
instrument to apply a DC voltage bias to the source/
drain terminals.
AmplitudeSweep
The amplitude of the pulse is swept while the base
voltage is kept constant. The charge pumping current
is measured and graphed as a function of the pulse
amplitude voltage. The source/drain terminals are
tied to ground.
AmplitudeSweep_2SMU Same as AmplitudeSweep test, except it adds a
second SMU instrument to apply a DC voltage bias to
the source/drain terminals.
RiseTimeLin
Performs a linear sweep of the rising transition
time of the pulse. ICP is measured and graphed as a
function of the rise time. The source/drain terminals
are tied to ground.
FallTimeLin
Performs a linear sweep of the falling transition
time of the pulse. ICP is measured and graphed as a
function of the fall time. The source/drain terminals
are tied to ground.
FreqLin
With the amplitude, offset voltage, rise/fall times
constant, the ICP is measured as a function of a
linear sweep of the test frequency. The source/drain
terminals are tied to ground.
FreqLog
With the amplitude, offset voltage, and rise/fall
times constant, the ICP is measured and graphed as
a function of a log sweep of the test frequency. The
source/drain terminals are tied to ground.
9.0E–9
ICP
8.0E–9
7.0E–9
6.0E–9
5.0E–9
4.0E–9
3.0E–9
2.0E–9
1.0E+0
0.0E–9
–3.0E+0
–2.5E+0
–2.0E+0
–1.5E+0
–1.0E+0
–500.0E–3
Base V
Figure 6. Graphical results of BaseSweep user test module
The input parameter values of a user test module can be
updated and the measurements can be repeated. One way to do
this is by using the on-screen Append Run button, which can be
used to show multiple test results on one graph. Figure 7 shows
the test results of increasing the test frequency from 1MHz to
6MHz. The data is appended to the graph and a new worksheet
is added in the Sheet tab for each appended test run.
www.keithley.com17
on the Definition tab of the test setup window. Enter the formula
for Dit as shown in Figure 9. The resulting Dit value can also be
plotted in the graph.
Charge Pumping Current (ICP) vs. Base Voltage (V)
and Test Frequency (1MHz to 6MHz)
70.0E–9
60.0E–9
50.0E–9
ICP
40.0E–9
30.0E–9
20.0E–9
10.0E–9
0.0E+0
–3.0E+0
–2.5E+0
–2.0E+0
–1.5E+0
–1.0E+0
–500.0E–3
Base V
Figure 7. Charge pumping current measurement results at multiple
test frequencies
The AmplitudeSweep is another common charge pumping
test. It measures the charge pumping current as the amplitude
of the pulse is swept. The base voltage is kept constant. The
resulting charge pumping measurements are shown in Figure 8.
Figure 9. Entering formulas in the Formulator function
Charge Pumping (ICP) vs. Pulse Amplitude (V)
Conclusion
26.0E–9
24.0E–9
22.0E–9
20.0E–9
18.0E–9
16.0E–9
ICP
14.0E–9
12.0E–9
10.0E–9
8.0E–9
6.0E–9
4.0E–9
2.0E–9
0.0E+0
–2.0E–9
1.0E+0
2.0E+0
3.0E+0
4.0E+0
5.0E+0
6.0E+0
7.0E+0
8.0E+0
9.0E+0
Pulse Amplitude
Figure 8. Charge pumping current as a function of pulse amplitude
Simple analyses, such as extracting the interface trap density,
can be performed on the data using the built-in Formulator
function. To activate this function, click the Formulator button
www.keithley.com
The Model 4200-SCS is the ideal tool for characterizing interface
properties of gate dielectrics. With the built-in pulse generator,
Model 4225-PMU or Model 4220-PGU, and the KTE Interface
software, the user need not do any programming, which
simplifies measurement and analysis. When equipped with the
Model 4225-PMU, the Model 4200-SCS is a powerful tool for
performing many tests commonly required in DC and ultra-fast
I‑V electrical characterization of devices, including the charge
pumping application detailed here. The 4225-PMU is not simply a
pulse generator; it can also measure current and voltage and be
used for transient I-V (waveform capture) applications.
References
1. G. Groeseneken, H.E. Maes, N. Beltran, and R.T. De Keersmaecker, “A
Reliable Approach to Charge-Pumping Measurements in MOS Transistors,”
IEEE Trans. Electron. Dev., Vol. ED-31, pp. 42-53, 1984.
Electrical Characterization of Carbon
Nanotube Transistors (CNT FETs) with
the Model 4200-SCS Parameter Analyzer
Introduction
Carbon nanotubes (CNTs) have been the subject of a lot of
scientific research in recent years, due not only to their small
size but to their remarkable electronic and mechanical properties
and many potential applications. The problems associated with
attempting to scale down traditional semiconductor devices
have led researchers to look into CNT-based devices, such
as carbon nanotube field effect transistors (CNT FETs), as
alternatives. Because they are not subject to the same scaling
problems as traditional semiconductor devices, CNT FETs
are being studied for a wide variety of applications, including
logic devices, memory devices, sensors, etc. The research on
these devices typically involves determining various electrical
parameters, which may include current-voltage (I-V), pulsed
I-V, and capacitance (C) measurements. Characterizing the
electrical properties of delicate nanoelectronic devices requires
instruments and measurement techniques optimized for low
power levels and high measurement sensitivity.
The Model 4200-SCS Parameter Analyzer offers a variety
of advantages for electrical characterization of CNT FETs. This
configurable test system can simplify these sensitive electrical
measurements because it combines multiple measurement
instruments into one integrated system that includes hardware,
interactive software, graphics, and
analysis capabilities. The system comes
with pre-configured tests for performing
electrical measurements that have been
optimized to ensure accurate results on
CNT FETs. This application note explains
how to optimize DC, pulsed I-V, and C-V
measurements on a CNT FET using the
Model 4200-SCS Parameter Analyzer. It
includes detailed information on proper
cabling and connections, guarding,
shielding, noise reduction techniques,
and other important measurement
considerations when testing carbon
nanotube transistors.
The Carbon Nanotube Transistor
A single semiconducting CNT can be
used as the conducting channel between
the source and drain of a FET. Figure 1
illustrates a back-gated Schottky barrier
CNT FET. Two metal contacts are located
across both ends of the CNT to form the
Figure 2.
CNT
Source
Drain
SiO2
Si Gate
Figure 1. Back-gated carbon nanotube transistor
Source and Drain terminals of the FET. The CNT is placed atop
an oxide that sits above a doped silicon substrate, which forms
the Gate terminal. Connections are made to the three DUT
terminals to perform the electrical measurements.
Making Electrical Measurements with the Model 4200-SCS
The Model 4200-SCS is supplied with a test project for making
some of the most commonly used CNT FET measurements.
This project (CNTFET) includes tests for I-V, pulsed I-V, and
C-V measurements. The I-V tests are performed using two of
the Model 4200-SMU Source Measure Unit (SMU) Instruments,
both with the Model 4200-PA Preamp option. The pulsed and
transient I-V measurements are made using the Model 4225PMU Ultra Fast I-V Module with two Model 4225-RPM Remote/
Preamplifier Switch options. Finally, the C-V measurements are
CNTFET project for the Model 4200-SCS
www.keithley.com19
performed using the Model 4210-CVU C-V
Measurement module.
CNT
The CNTFET project is included with
all Model 4200-SCS systems running
KTEI Version 8.1 or later. Figure 2 shows
the CNTFET project running in the
Keithley Interactive Test Environment
(KITE) software.
Current-Voltage Measurements
The I-V characteristics of a CNT transistor
can be used to extract many of the
device’s parameters, study the effects
of fabrication technique and process
variations, determine the quality of the
contacts, etc. Figure 3 illustrates a DC
I-V test configuration that incorporates
two Model 4200-SMU Instruments. These
SMU instruments are capable of sourcing
and measuring both current and voltage;
they have picoamp sensitivity and can
be current-limited to prevent damage
to the device. In this diagram, SMU1 is
connected to the Gate of the CNT FET
and SMU2 is connected to the Drain.
The Source terminal is connected to the
Ground Unit (GNDU) or to a third SMU
instrument if it is necessary to source
and measure from all three terminals
of the FET.
Source
Drain
SiO2
Si Gate
Ground Unit
or SMU3
(if required)
A
SMU1
Steps V
A
SMU2
Sweeps V
Measures I
Figure 3. Circuit diagram for measuring the DC I-V characteristics of a CNT FET
In this example, the Model 4200-SCS’s
KITE software is set up to measure a DC
drain family of curves (Vds–Id). As SMU1
steps the gate voltage (Vg), SMU2 sweeps
the drain voltage (Vd) and measures the
resulting drain current (Id). Figure 4
shows the resulting FET characteristics
generated using the CNTFET project.
Figure 4. DC I-V drain family of curves measured by the Model 4200-SMU Source Measure Unit
Without changing connections to the
device, Model 4200-SCS’s interactive KITE
amount of current that can flow through the device. This can
software simplifies performing other
be done in the software by setting the Current Compliance of
common I-V tests such as the drain current (Id) vs. gate voltage
each SMU to a safe level, such as 20µA. This is a programmed
(Vg) curves. For this test, the gate voltage is swept and the
limit to ensure the current doesn’t exceed the user-defined
resulting drain current is measured at a constant drain voltage.
compliance.
The results of an Id–Vg curve at a constant drain voltage are
shown in Figure 5. The drain voltage can also be stepped as the
• Provide Sufficient Settling Time: Because CNT FET
gate voltage is swept.
measurements often involve measuring low current (<1µA),
Optimizing DC measurements
The following techniques will improve the quality of DC
measurements made on CNT FETs with the Model 4200-SCS:
• Limit Current: To prevent damage to the device while
performing I-V characterization, the user should limit the
www.keithley.com
it is important to allow sufficient settling time to ensure the
measurements are stabilized after a current or voltage has
been applied. Some of the factors that affect the settling time
of the measurement circuit include the cables, test fixtures,
switches, probers, the DUT resistance, and the current range
of the measurement instrument. To ensure settled readings,
• Keep Probes Up: Make sure the
probes are in the up position (not
contacted to the device) when
connecting and disconnecting
instruments from the terminals of the
device. The process of moving cables
has the potential to inject charge
into the device and cause damage.
This is due to both triboelectric and
piezoelectric effects.
Pulsed I-V Measurements
Figure 5. Drain current vs. gate voltage of CNT FET
additional delay time can be added to
the voltage or current step time prior
to the measurement. This delay time
can be easily adjusted in the Timing
Menu in the KITE software.
• Use Proper Speed Modes: The
Timing Menu also offers Speed Modes,
including Delay and Filter Factor
settings, which affect the settling
time of the reading, as well as the
integration time of the measurement.
Increasing the Delay Factor, Filter
Factor, and the A/D Aperture Time can
decrease noisy measurements.
• Minimize Noisy Measurements:
Noise may be generated from a
variety of sources, including particle
collisions, defects, AC pick-up, and
electrostatic interference. Noisy
measurements result when a noise
signal is superimposed on the DC
signal being measured. This can
result in inaccurate or fluctuating
measurements.
The most common form of external
noise “pick-up” is 60Hz (or 50Hz) line
cycle pick-up. This can be a common
occurrence near fluorescent lights.
Millivolts of noise are not uncommon.
Keithley uses a technique called LineCycle Integration to minimize the
effects of 60Hz (or 50Hz) line pick-up.
Line-cycle noise will “average out”
when the integration time is equal
to an integral number of power line
cycles. The number of power line
cycles can be adjusted in the KITE
software in the Timing Menu.
Electrostatic interference is another
cause of noisy measurements when
measuring low currents. This coupling
occurs when an electrically charged
object approaches the circuit under
test. In high impedance circuits, this
charge doesn’t decay rapidly and can
result in unstable measurements. The
erroneous readings may be due to
either DC or AC electrostatic fields,
so electrostatic shielding will help
minimize the effects of these fields.
The electrostatic shield can be just a
simple metal box that encloses the test
circuit. Probe stations often include
an electrostatic/EMI shield or optional
dark box. The shield should be
connected to the measurement circuit
LO, which is the Force LO terminal of
the SMU. The Force LO terminal is the
outside shield of the triax cable of the
SMU or is located on the GNDU. All
cables need to be of a low-noise design
and shielded. Each Model 4200-SMU
comes with two low-noise triax cables.
In addition to making traditional DC
I-V measurements, it may be desirable
to perform ultra-fast pulsed I-V
measurements for various reasons. First,
it may be important to observe the
high speed response of the CNT device.
In some cases, nanostructures can be
destroyed by the heat generated when
making traditional DC measurements.
Pulsed I-V measurements can reduce the
total energy dissipated in a device, and
therefore reduce the potential for damage.
Finally, pulsed electrical testing can
prevent current drifting in measurements
that can occur during DC measurements.
The pulsed I-V measurements on
the CNT FET can be easily made using
the Model 4225-PMU Ultra Fast I-V
Module. The Model 4225-PMU provides
two channels of high speed, multi-level
voltage pulse output while simultaneously
measuring current and voltage. This
module replaces traditional pulse/measure
test configurations, which consisted of
a pulse generator, digital oscilloscope,
interconnect hardware, and software.
The Model 4225-PMU has two modes
of ultra-fast I-V source with measure:
pulsed I-V and transient I-V. These two
modes are illustrated in Figure 6.
Pulsed I-V refers to any test with a
pulsed source and a corresponding high
speed, time-based measurement that
provides DC-like results. The current
and/or voltage measurement is an
average of readings taken in a predefined
measurement window on the pulse. This
average of readings is called the “spot
mean.” The user defines the parameters of
the pulse, including the pulse width, duty
cycle, rise/fall times, amplitude, etc.
www.keithley.com21
Transient I-V, or waveform capture, is a time-based current
and/or voltage measurement that is typically the capture of
a pulsed waveform. A transient test is typically a single pulse
waveform that is used to study time-varying parameters, such as
the drain current degradation versus time due to charge trapping
or self-heating. Transient I-V measurements can be used to test
a dynamic test circuit or as a diagnostic tool for choosing the
appropriate pulse settings in the pulsed I-V mode.
Given that the Model 4225-PMU has two channels, only one
module is needed to test a three-terminal CNT FET. A typical
test configuration for connecting the PMU module to a CNT
FET is shown in Figure 7. In this diagram, Ch 1 of the PMU is
connected to the Gate terminal and Ch 2 is connected to the
Drain terminal. The Source terminal is connected to the PMU
Common, which is the outside shield of the PMU coax connector.
To connect this Common terminal to the probe tip, use a BNC
or triax shorting plug that will connect the outside of the coax
to the manipulator probe. To generate a Vds–Id curve, Ch 1 steps
the gate voltage and Ch 2 sweeps the drain voltage and measures
the resulting drain current.
Voltage
Samples
Time
Measure Window
Pulsed I-V
Transient I-V
Pulse/Measure with DC-like results
Train, Sweep, Step modes
General device characterization
Time-based I and V measurements
Waveform capture
Dynamic device testing
Figure 6. Two modes of ultra fast I-V source with measure: Pulsed I-V and
Transient I-V
Figure 8 illustrates a pulsed I-V drain family of curves taken
with the Model 4225-PMU. For this measurement, a pulse width
of 500µs was used to generate the curves. However, each PMU
channel has the ability to output voltage pulses as short as 70ns
with a rise time as short as 20ns. The minimum duration of the
pulse width will depend on several factors, including the test
circuit RC time constant and the magnitude of the test current.
Each dot on the curves represents a “spot mean” measurement
on the pulsed waveform.
The Model 4225-PMU has five ranges full scale from 800mA
down to 100µA. To measure lower currents, using the Model
4225-RPM optional Remote Amplifier/Switch is recommended
because it adds six measurement ranges, down to 100nA full
scale. The pulsed I-V curves shown in Figure 8 were taken on the
100µA range. The threshold current was set to 20µA so that the
test will stop if the threshold current level is reached.
CNT
Source
Drain
SiO2
Si Gate
PMU1 CH 2
Sweeps Drain V
Measures Drain I
PMU1 CH 1
Steps Gate V
Common
(outside shield
of PMU
coaxial cable)
A
A
V
50Ω
CH1
V
50Ω
CH2
For some applications, it may be necessary to study the
transient response of a CNT FET. If this is the case, the waveform
capture mode (transient I-V) can be used to capture the current
and voltage time-based response to the device. Figure 9 shows
the transient response of the CNT FET. The blue curve is the
pulsed drain voltage and the red curve is the resulting current
response as a function of time.
Figure 7. Circuit diagram for measuring the pulsed I-V characteristics of
a CNT FET
The blue voltage output curve looks close to the defined rise
and fall times of 10µs with a pulse width of 50µs. Note that the
pulse width is measured at one-half of the input amplitude of 1V.
Therefore, the pulse width is measured at 500mV. The sample
period in this example is 25ns (40MHz rate). With proper cabling
and connections, the voltage shape should be output as defined
by the user with minimal deviation.
Optimizing Pulsed I-V Measurements
To improve the quality of pulsed I-V measurements made with
the Model 4200-SCS, follow these guidelines:
The red curve shows the drain current and is plotted on the
right Y-axis. The drain current is measured at constant drain and
gate voltages. The peaks in the curve are caused by charging
www.keithley.com
4225-PMU
and discharging of the cabling, as well as the current flow
through the device. Note that these peaks occur during the pulse
transitions. Reducing the pulse amplitude or increasing the pulse
transition time reduces the dV/dt, which reduces the peak height.
• Use the Right Cables and Connections: Using proper
cabling and connections is important for ultra-fast I-V
applications in order to achieve the highest frequency output
and to avoid signal distortions and capacitive charging effects.
–– Use cabling and connections
optimized for high frequency (at
least 150MHz).
–– Use a signal path that matches
the impedance of the instrument
(50 ohms).
–– Tie the low side of the DUT to the
shield of the PMU coax cable.
–– Connect the shields from each PMU
channel together as close as possible
to the DUT.
–– Minimize the loop area once the
center conductor and shield are
separate in the test circuit.
–– Minimize the cable length.
• Make the Right Chuck Connections:
CNT FETs and other nanotransistors
may be either back-gated or topgated. For back-gated devices, one
of the PMU channels needs to be
connected to the chuck of the prober.
When making PMU connections
to the chuck, the user will give up
some functionality of the PMU: fast
transitions, high frequency, low
current, etc. This is because the output
of the high frequency PMU channel is
connected to the chuck capacitance
and the chuck cabling, which slows
down the source response and
couples noise into the measurement.
If possible, it is better to use a third
manipulator and probe directly to the
chuck. For high speed sourcing and
measuring, it is best to use all top-side
connections and avoid connecting the
PMU to the chuck.
• Verify Pulse Width: Ensure the pulse
width is long enough to ensure a
settled reading. Verify the resulting
current measurement is settled by
outputting a single pulse using the
Waveform Capture mode. Both the
current and voltage can be plotted as a
function of time in the Graph tab.
Figure 8. Pulsed I-V drain family of curves of CNT FET
• Minimize Noise: To minimize noisy
results, multiple waveforms can
be averaged or a moving average
function can be created in the
built-in Formulator to smooth out the
measurements further.
Capacitance-Voltage
Measurements
In addition to performing DC and
pulsed I-V measurements on CNT FETs,
measuring the capacitance of the FET can
also provide information about the device,
including the mobility, timing effects, and
gate dielectrics. Figure 10 outlines the
connections of the Model 4210-CVU to
the CNT FET. In this configuration, the
gate-to-drain capacitance is measured as a
function of the gate voltage.
Figure 9. Waveform of single drain voltage pulse and resulting drain current of CNT FET
The HCUR/HPOT terminals that
connect the high of the voltage source
to the gate should be connected to the
chuck. The LCUR/LPOT terminals that
measure the capacitance should be
connected to the drain terminal of the
www.keithley.com23
4210-CVU
AC
Source
HCUR
CNT
Source
HPOT
Drain
SiO2
Si Gate
AC
Ammeter
LPOT
ABB
Feedback
AC
Voltmeter
LCUR
Figure 10. Connections of the Model 4210-CVU to a CNT FET
DUT. For best results, the measurement
terminals should never be connected to
the chuck. For top-gated CNT FETs, both
the measure and voltage source can be
output to the gate of the FET from the
same terminals (either HCUR/HPOT or
LPOT/LCUR) of the CVU. The HI and LO
terminals of the CVU are interchangeable
in the Force Measure Window of the
CVU in the KITE software. The results of
generating a C-V sweep between the gate
and drain of the CNT FET are shown in
Figure 11.
Optimizing Capacitance
Measurements
To improve the quality of capacitance
measurements made with the Model 4200SCS, follow these guidelines:
• Perform Open Compensation (for
Measurements <10pF): The open
correction feature compensates for
capacitance offsets in the cabling
and connections. Performing the
correction is a two-part process. The
corrections are performed, and then
they are enabled within a test module.
To perform the corrections, open
the Tools Menu and select CVU
Connection Compensation. For an
Open correction, click on Measure
Open. Probes must be up or the DUT
removed from the test fixture. Enable
the correction by clicking on the
Compensation button in the Forcing
Functions/Measure Options window.
www.keithley.com
Figure 11. C-V sweep of gate-to-drain capacitance
• Use Proper
Shield
Connections: Connect the shields
of the coax cables together as close
as possible to the DUT. This reduces
the loop area of the shields, which
minimizes the inductance. This also
helps to maintain the transmission
line effects. If the shields are not
connected together, offsets may occur.
The higher the frequency, the more
important this becomes.
• Choose Appropriate Hold and
Sweep Delay Times: The condition
of a device when all internal
capacitances are fully charged after
an applied voltage step is referred
to as “equilibrium.” If capacitance
measurements are made before the
device is in equilibrium, inaccurate
results may occur.
To choose the delay times for a C-V
sweep, step an applied voltage using
the Sampling Mode, and plot the
capacitance as a function of time. Observe the settling time
from the graph. Use this time for the Hold Time for the initial
applied voltage or for the Sweep Delay Time applied at each
step in the sweep. The Sweep Delay Time may not need to be
as long as the first step. The user will need to experiment to
verify the appropriate time.
• Choose Appropriate Speed Mode in Timing Menu: The
Speed mode function enables the user to adjust the time
for settling and integration of the measurement. For small
capacitances (pico-Farads or less) use the Quiet or Custom
Speed modes for best results.
• Use Guarding: When making very small capacitance
measurements, guarding will help prevent stray capacitance
from unused terminals of the device from affecting
measurement accuracy. For example, if measuring the
capacitance between only the gate and drain terminals, the
source terminal of the FET can be connected to the guard.
The guard terminal of the Model 4210-CVU is the outside
shield of the coax cable.
Conclusion
When using the appropriate instrumentation and measurement
techniques, optimal electrical characterization of CNT FETs can
be achieved. The Model 4200-SCS is an ideal tool for performing
electrical characterization of CNT FETs and other nanostructures
because of its integrated hardware, software, and analysis tools.
The Model 4200-SMU Source Measure Unit Instrument can
be used to determine Vds–Id, VGS –Id, resistance, and other I-V
measurements on the CNT FET. The Model 4225-PMU Ultra-Fast
I-V Module can be used to make pulsed I-V measurements or
observe the transient response of a pulsed waveform applied to
the DUT. The Model 4210-CVU Capacitance Meter can be used
to generate C-V, C-f, or C-t curves. Using the CNTFET project
that comes with the Model 4200-SCS can further simplify
measurement setup and execution.
Acknowledgement
The author appreciates the assistance provided by Sandia
National Labs, Livermore, California, who supplied the CNT FETs
used in the device testing process during the development of this
application note.
www.keithley.com25
www.keithley.com
Pulsed Characterization of Charge-trapping
Behavior in High-k Gate Dielectrics
Introduction
Development of High-k Gates for
Advances CMOS Devices
High dielectric constant (high-k) materials, such as hafnium
oxide (HfO2), zirconium oxide (ZrO2), alumina (Al2O3), and their
silicates, have drawn a great deal of attention in recent years for
potential use as gate dielectrics in advanced CMOS processes [1].
With high dielectric constants, gate dieletrics can be made thicker
than SiO2 while achieving the same capacitance. The result is
leakage current that can be lower by as much as several orders
of magnitude. However, there are still technical challenges to
overcome, such as Vt instability [2-4], carrier channel mobility
degradation [5-9], and long-term device reliability [10-13].
One of the important issues preventing implementation of
high-k gates is the trapping of charges in the pre-existing traps
inside these dielectrics [14-15]. When the transistor is turned
on, some of the channel carriers will be accumulated in the
gate dielectric due to the vertical electrical field, resulting in a
shift of threshold voltage and a reduction in drain current. Fully
understanding charge-trapping and these related mechanisms
is the key to understanding channel mobility degradation and
device reliability problems. However, traditional DC testing
techniques may not accurately characterize these mechanisms.
Limitation of DC Characterization Techniques
As charges are trapped in the gate dielectric, the threshold
voltage of the transistor increases due to the built-in voltage
in the gate capacitor; therefore, the drain current decreases. It
appears that charge trapping and de-trapping times strongly
depend on the composition of the gate stacks, i.e., physical
thickness of the interfacial SiO2 layer and high-k film, as well as
process techniques [16-18]. The time scale varies from several
microseconds to tens of milliseconds [19]. The de-trapping of the
charges is also strongly gate voltage and polarity dependent.
The wide dynamic range of charge trapping, and the voltage
dependent trapping and de-trapping, make it very difficult
to use one type of characterization technique (especially a
DC technique) to get a complete picture of what is going on
1.20E–009
15ms per point
1s per point
1.00E–009
C (F)
This application note discusses charge trapping and the
limitation of DC characterization techniques in quantifying
trapped charge. It also describes an ultra-fast pulse I-V technique
for characterizing the intrinsic (“trap free”) performance of
high-k gate transistors that exhibit the fast transient charging
effect (FTCE).
1.40E–009
8.00E–010
6.00E–010
4.00E–010
2.00E–010
0.00E–000
–2.0
–1.5
–1.0
–0.5
0.0
0.5
Vg (V)
Figure 1. Time-dependent hysteresis of C-V: Slower measurements produce
less hysteresis, indica­ting an equilibrium condition has been reached.
Moreover, hysteresis cannot quantify fast transient trapping, since a
significant portion of the fast transient trapping could be lost in the DC
­measurement.
inside the stacked gate dielectric. For example, commonly
used methods employ a double sweep in either DC Vgs-Id or
high frequency C-V measurements. These techniques involve
ramping gate voltage back and forth while drain current or gate
capacitance is measured. If hysteresis is seen on the resulting I-V
or C-V curves, it is a clear indication of charge trapping inside the
gate stacks.
The issue with this technique is that the hysteresis is strongly
dependent on measurement time. The hysteresis measured in the
DC I-V test could be different from that in a C-V test, because the
time taken for each measurement may be dramatically different.
This is illustrated by dual sweep C-V measurements taken at
different speeds (Figure 1). Test speed is strongly dependent on
instrumentation and not easily controlled. Even if it were, there
is no model to quantify how much charge is really trapped in the
gate during the test; i.e., hysteresis cannot quantify the amount
of charge trapped, since a significant portion of the fast transient
trapping could be lost in the DC measurement.
Another method involves DC stress voltage to inject charges
intentionally into the gate, then the use of C-V or I-V method
to measure the flat band voltage or threshold voltage shift [20].
The amount of charge trapped can be calculated based on
this formula:
Qtrap = Cgate · ∆Vfb , or Qtrap = Cgate · ∆V1.
The issue with this technique is the transition period between
DC stress and I-V or C-V measurement, when typically there is
www.keithley.com27
PGU
Trigger
DC Bias/Control
Scope
Ch. 1
VDD
4200-SCS
Ch. 2
VD
100mV
Scope
GPIB
Trigger
RL
Pulse
Generator Pick-Off Bias
Tee
Bias
Tee
VG
Tee
Figure 2. Two different pulse I-V test setups to study transient charge trapping: (a) pulse generator unit (PGU) and scope; (b) integrated test system with
hardware and software designed specifically for fast transient trapping measurements.
no applied voltage, or voltage is very low compared to the stress
condition. When stress voltage is off, the charges trapped in the
gate can de-trap in as little as tens of microseconds. So, only a
fraction of the total trapped charges are measured due to the
relaxation effect, resulting in an overly “optimistic” view of the
film’s quality.
Ultra-short Pulse Characterization Techniques
Better methods have been developed over the past few years
[3, 4, 19, 21] for capturing the fast transient behavior of charge
trapping. Figure 2 shows two different test configurations for
a Single Pulse Charge Trapping (SPCT) measurement. In both
setups a pulse is applied to the gate of the transistor while its
drain is biased at a certain voltage. The change in drain current,
resulting from the gate pulse, appears on the digital oscilloscope.
The difference between these two configurations is that the
one in Figure 2b has much higher bandwidth than the one in
Figure 2a; therefore, it can capture much faster pulse responses
(down to tens of nanoseconds). At such high speed, the bulk
traps in a high-k layer are unlikely to respond; therefore, an
“intrinsic” transistor response with negligible charge-trapping
effect can be measured.
Gate Voltage
Pulse
Charge trapping during
transistor “on”
Up Trace
Trapping
Down Trace
De-trapping
0V
De-trap before ramp
Clean up interface charges
Figure 3. Trapping and de-trapping in single gate voltage pulse.
Figure 4a shows an example of a pulse I-V measurement
with different pulse speeds using setups from Figures 2a and
2b respectively. A DC I-V curve is overlaid for comparison. There
is hysteresis with the slower pulse due to charge trapping. By
using the faster pulse, hysteresis is eliminated because there’s
insufficient response time for charges to be trapped.
The key to using SPCT is to look at charge trapping
and de-trapping within a single, well-configured gate pulse
(Figure 3). The pulse usually starts in a position that discharges
the gate capacitor before the voltage ramp begins. This is to clean
up any residual charges trapped in the gate. Then, during the
rise time of the voltage ramp, the corresponding drain current
response is captured, allowing a Vgs-Id curve to be formed.
Slower pulse measurements should only be used for high-k
devices with relatively low charge-trapping effects. Although
the hysteresis of a slower pulse I-V curve can quantify charge
trapping better than a DC method, results must be interpreted
with caution, since de-trapping effects are influenced by the
specific structure of the gate stack and the pulse fall time. Also,
when two traces in the tr and tf portions are not parallel, it is
difficult to define the point at which hysteresis is measured.
If the pulse rise time is fast enough that there is no charge
trapping, then the Vgs-Id curve represents the transistor’s intrinsic
behavior, free from error due to spurious charge-trapping effects
[16]. During the plateau of the pulse, the transistor is turned
on, and some of the channel carriers might be trapped in the
gate, which changes the threshold voltage and causes the drain
current to drop. During the fall time of the pulse, another Vgs-Id
curve is formed, but with the charge-trapping effect that should
be measured.
An alternative approach is to plot pulse Id versus time
(Figure 4b). An evaluation of the pulse width portion of Id
(where Id degrades) can be used to quantify trapped charge [15,
16]. Still, to ensure negligible charge trapping during the entire
pulse, one can use a fast pulse with short rise and fall times. This
can be achieved easily with the ultra-short-pulse test system in
Figure 2b. It produces Id-Vg values by using a single pulse per
point of less than 100ns.
www.keithley.com
Drain Current (µA)
800
“Single Pulse” Id-Vg
Ramped Pulse Id-Vg
DC Id-Vg
Because of complications associated with trying to
characterize charge-trapping effects for transistors with different
functionalities, including high frequency operation, the best
solution for modeling engineers is to use instrumentation and
a test configuration that avoids artifacts associated with DC or
slower pulse measurements. The ultra-short pulse measurement
technique provides a better depiction of real device performance
[22], so the resulting models help optimize designs for actual
operating conditions.
Id degradation
PW = Pulse Width
600
400
tr
tf
tr
Vg = –1 to 2.2V
PW = 60µs
tr , tf = 5µs
tf
200
0
0.0
0.5
1.0
1.5
2.0
Gate Voltage (V)
20
40
60
80
100
Time (µs)
Figure 4. Single Pulse Charge Trapping measurements: (a) the ultra-short
ramped pulse Id-Vg and slow “single pulse” pulse with hysteresis (DC result
is shown as a reference), and (b) corresponding slow pulse versus time
illustrating an alternative approach to determine the degradation of Id.
Process engineers also need these pulse measurement
techniques to characterize and track improvements in their
continuing efforts to improve film quality and remove chargetrapping degradation. Finally, an understanding of the intrinsic
behavior of high-k gate dielectric devices is key to understanding
the physics of charge trapping and de-trapping. This physicalbased understanding leads to the correct extrapolation of
device lifetime.
Benefits of Short Pulse Characterization
Sources of Error
Since there is much less charge-trapping effect with very short
pulse widths, the drain current measured is higher than under
DC conditions (red curve, Figure 4a). This results in a higher
predicted channel carrier mobility when pulse I-V data are used
to generate a model, which is more representative of transistors
that are switching very fast (i.e., those that will not experience
full charge-trapping effects). Another advantage of the ultrashort pulse system is that the pulse I-V measurements, with
pulse widths on the order of nanoseconds, can be performed
very easily and results can be compared to a DC measurement
without resetting the system hardware or moving the wafer to
another station. Such a comparison is shown in Figure 5.
Although ultra-short pulse measurements are a powerful
technique, recognize that speed characteristics are in the radio
frequency (RF) domain. Therefore, it is easy to introduce
measurement errors if the test system is not optimized for high
bandwidth. There are three main sources of errors: signal losses
due to cables and connectors, losses due to device parasitics, and
contact resistance.
Drive current measured using the pulse I-V technique
can be significantly higher than DC due to the lack of
charge-trapping effects. This illustrates the close-to-intrinsic
performance of transistors with high-k gate dielectrics, and
ultimately demonstrates the advantage of the ultra-short pulse
I-V technique.
Signal losses due to cables and connectors, as well as contact
resistance effects, can be calibrated out using a transistor with
a SiO2 gate, where there is no charge-trapping effect. The pulse
I-V curve should overlay the DC I-V curve precisely once the
calibration is complete.
However, device parasitics, usually parasitic capacitance
between pad contacts, cannot be easily removed. This will
reduce the pulse current from the drain terminal. This problem
can be designed out by using RF-compatible transistor structures
(Ground-Signal-Ground), and by increasing pad pitch so that
400
400
nFET W/L = 10/0.5µm
Vd = 100mV
300
200
DC Ramp
tr = tf = 2ns,
35ns pulse
100
0
0.0
0.5
1.0
1.5
Gate Voltage (V)
2.0
Drain Current (µA)
Drain Current (µA)
500
300
Vg = 1.8V
200
DC Ramp
tr = tf = 2ns,
35ns pulse
100
0
0.0
0.5
1.0
1.5
Drain Voltage (V)
2.0
Figure 5. Overlay of ramped pulse I-V and DC I-V measurements on NMOSFET with high-k (HfO2) gate dielectric: (a) DC and pulsed Id-Vg measurement at
linear region, (b) DC and pulsed Id-Vd measurement. Short pulse parameters allowed close-to-intrinsic drive current to be measured and compared to the
DC measurement.
www.keithley.com29
capacitance between the pads is small enough not to affect pulse
current measurements.
Conclusion
DC characterization techniques are inadequate for capturing
the dynamic nature of charge trapping in high-k gate stack
structures. Pulse I-V is a powerful technique that allows
characterizing transistors with high-k gate stacks in a trap-free
environment that permits assessing their intrinsic qualities.
When current is measured in a shorter time than that associated
with the charge-trapping time scale, improved intrinsic transistor
characteristics can be achieved, including a drive current that
is significantly higher than that measured with traditional DC
I-V techniques.
References
1. E. Gusev, E. Cartier, D. Buchanan, M. Gribelyuk, M. Copel, H. OkornSchmidt, and C. D’Emic, in Proc. of Conference on Insulating Films on
Semiconductors, 2001.
2. A. Kerber, E. Cartier, L. Pantisano, R. Degraeve, T. Kauerauf, Y. Kim, A. Hou,
G. Groeseneken, H. E. Maes, and U. Schwalke, “Origin of the threshold
voltage instability in SiO2/HfO2 dual layer gate dielectrics,” IEEE Electron
Device Lett., vol. 24, pp. 87, 2003.
3. A. Kerber, E. Cartier, L. Pantisano, M. Rosmeulen, R. Degraeve, T. Kauerauf,
G. Groeseneken, H. E. Maes, and U. Schwalke, “Characterization of the VT
instability in SiO2/HfO2 gate dielectrics,” in Proc. 41st Annual IEEE Intl.
Reliability Physics Symp., p. 41-45, 2003.
4. C. D. Young, R. Choi, J. H. Sim, B. H. Lee, P. Zeitzoff, Y. Zhao, k. Matthews,
G. A. Brown, and G. Bersuker, “Interfacial Layer Dependence of HfSixOy
Gate Stacks on Vt Instability and Charge Trapping Using Ultra-short Pulse I-V
Characterization,” in Proc. 43rd Annual IEEE Intl. Reliability Physics Symp.,
p. 75-79, 2005.
5. M. Fischetti, E. Cartier and D. Neumayer, J. Appl. Phys., 90, p. 4587, 2001.
6. A. Kerber, et al., VLSI, p. 159 (2003).
7. W. Zhu, J.-P. Han, and T. P. Ma, “Mobility Measurements and Degradation
Mechanisms of MOSFETs Made With Ultrathin High-k Dielectrics,” IEEE
Transactions on Electron Devices, vol. 51, p. 98-105, 2004.
8. C.D. Young, IEEE Electron Device Letters, vol. 26, no. 8, p. 586, 2005.
9. D. V. Singh, P. Solomon, E. P. Gusev, G. Singco, and Z. Ren, “Ultra-fast
Measurements of the Inversion Charge in MOSFETs and Impact on Measured
www.keithley.com
Mobility in High-k MOSFETs,” in IEEE Intl. Electron Devices Meeting Tech.
Digest, p. 863-866, 2004.
10.E.Gusev, et al., Appl. Phys. Lett., vol. 83, p.5223, 2003.
11.B.H. Lee, R. Choi, J.H. Sim, S.A. Krishnan, J.J. Peterson, G.A. Brown, and G.
Bersuker, “Validity of Constant Voltage Stress Based Reliability Assessment of
High-k Devices,” IEEE Transactions on Device and Materials Reliability, vol.
5, no. 1, p. 20-25, 2005.
12.[S. Zafar, A. Kumar, E. Gusev, and E. Cartier, “Threshold Voltage Instabilities
in High-k Gate Dielectric Stacks,” IEEE Transactions on Device and Materials
Reliability, vol. 5, no. 1, p. 45-64, 2005.
13.G. Ribes, J. Mitard, M. Denais, S. Bruyere, F. Monsieur, C. Parthasarathy, E.
Vincent, G. Ghibaudo, “Review on High-k Dielectrics Reliability Issues,” IEEE
Transactions on Device and Materials Reliability, vol. 5, no. 1, p. 5-19, 2005.
14.G. Bersuker, et al., Materials Today, 7, p.26., 2004.
15.G. Bersuker, J. H. Sim, C. D. Young, R. Choi, B. H. Lee, P. Lysaght, G. A.
Brown, P. Zeitzoff, M. Gardner, R. W. Murto, and H. R. Huff, “Effects of
Structural Properties of Hf-Based Gate Stack on Transistor Performance,”
presented at 2004 Spring Meeting of the Material Research Society, p.
31-35, 2004.
16.C. D. Young, R. Choi, J. H. Sim, B. H. Lee, P. Zeitzoff, Y. Zhao, k. Matthews,
G. A. Brown, and G. Bersuker, “Interfacial Layer Dependence of HfSixOy
Gate Stacks on Vt Instability and Charge Trapping Using Ultra-short Pulse I-V
Characterization,” presented at the 43rd Annual IEEE Intl. Reliability Physics
Symp. Proc, p. 75-79, 2005.
17.R. Choi, S.J. Rhee, J.C. Lee, B.H. Lee, and G. Bersuker, “Charge Trapping and
Detrapping Characteristics in Hafnium Silicate Gate Stack Under Static and
Dynamic Stress,” IEEE Electron Device Lett., vol. 26, no. 3, p. 197-199, 2005.
18.J.H. Sim, R. Choi, B.H. Lee, C.D. Young, P. Zeitzoff, D.-L. Kwong, and G.
Bersuker, “Trapping/De-trapping Gate Bias Dependence of Hf-Silicate
Dielectrics with Poly and TiN Gate Electrode,” Jpn. J. of Appl. Phys., vol. 44,
no. 4B, p. 2420-2423, 2005.
19.C. Leroux, J. Mitard, G. Ghibaudo, X. Garros, G. Reimbold, B. Guillaumor, F.
Martin, in IEEE Intl Electron Devices Meeting Tech Digest. p.737-740, 2004.
20.S. Zafar, S. Callegari, V. Narayanan,, and S. Guha, “Impact of Moisture on
Charge Trapping and Flatband Voltage in Al2O3 Gate Dielectric Films,”
Applied Physics Letters, vol. 81, p. 2608, 2002.
21.C.D. Young, Y. Zhao, M. Pendley, B.H. Lee, k. Matthews, J.H. Sim, R. Choi,
G.A. Brown, R.W. Murto and G. Bersuker, “Ultra-Short Pulse Current-Voltage
Characterization of the Intrinsic Characteristics of High-k Devices”, Jpn. J. of
Appl. Phys., vol. 44, no. 4B, 2437, 2005.
22.C. Y. Kang, R. Choi, J. H. Sim, C. Young, B. H. Lee, G. Bersuker, Jack C.
Lee, “Charge trapping effects in HfSiON dielectrics on the ring oscillator
circuit and the single stage inverter operation,” in IEEE Intl Electron Devices
Meeting Tech Digest, p. 485-488, 2004.
Pulse I-V Characterization of
Non-Volatile Memory Technologies
Introduction
for testing floating gate flash, phase-change cell, and
ferro-electric cell devices.
Until recently, floating gate (FG) NAND flash memory
technology has successfully met the growing requirement for
non-volatile memory (NVM) for digital cameras, MP3 players
and smartphones. However, there is increasing concern in the
consumer electronics industry that floating gate NVM may not
be able to continue providing higher storage capacities at the
ever-lower cost-per-bit requirements that drive the NVM market
[1]. The potential for the floating gate approach to “hit the wall”
means that research into alternative technologies has become
increasingly critical.
Brief History of NVM
Scientists around the world are investigating NVM alternatives
that can replace FG NAND technology, including phase-change
memory (PCM/PRAM), charge trap flash (CTF/SONOS), resistive
memory (ReRAM), ferro-electric memory (FeRAM), and
magnetoresistive memory (MRAM) (Figure 1). These device
technologies have been studied for years, and each is currently
available in the market in some form. Other NVM technologies,
including spin-transfer torque (STT) MRAM, floating body
(FBRAM), and various types of carbon-nanotube-based memory
(CNT RAM), are being actively researched to determine their
suitability for memory product applications.
This wide-ranging research into an expanding number
of materials and technologies requires an electrical test
system with wide dynamic range and flexible parameter
control. This application note provides a brief history of NVM,
an overview of the test parameters required for electrical
characterization of NVM materials and devices, and an
explanation of the capabilities of the Model 4225-PMU UltraFast I-V Module with the Model 4225-RPM Remote Amplifier/
Switch, two instrument options designed for use with
Keithley’s Model 4200-SCS Parameter Analyzer. The Model
4225-PMU/4225-RPM combination has integrated simultaneous
measurement of current and voltage on each channel, making
investigating transient pulse responses much simpler than with
previous hardware. With the system’s multi-pulse waveform
generation capability, the Model 4200-SCS with the Model
4225-PMU may be used to characterize the memory device’s
switching mechanism in both the transient and I-V domains.
After a discussion of emerging test requirements, this note
provides an overview of the NVM project, tests, and parameters
Floating Gate
• Increasingly low cost and high density based on
predictable scalability
Crystalline GST
S
D
Active Area
I
B
Floating Gate Flash
Insulator
Control Gate
Metal
Ferro-electric
Metal
D
Resistor
(heater)
S
The ideal memory should have characteristics of both
dynamic and non-volatile memory:
Top Electrode
Control Gate
Tunnel Oxide
In addition to the traditional uses of NVM in portable
consumer electronic devices, the success of FG NVM has created
new product categories, such the ubiquitous “thumb” or USB
drive and, more recently, the high performance Solid State Disk
(SSD) products now being used as replacements for traditional
computer hard drives in high performance applications. These
products, as well as the possibility of a “universal” memory that
would replace both existing flash and dynamic memory (DRAM),
have justified ongoing research at universities and semiconductor
organizations and companies [2].
Bottom Electrode
Phase-Change Memory
B
Ferro-electric Memory
Figure 1. Various non-volatile memory devices.
www.keithley.com31
• Fast read/write (similar to or faster than existing
DRAM speeds)
• High endurance (to address DRAM or SSD applications)
• Long retention
• Low power and voltage requirements
• Compatible with existing logic circuits and
semiconductor processes
As the applications for FG NVM devices have increased, so
too has the pressure on the FG technology. This has opened
up the market to multiple technologies that may replace the
FG approach. The future may have multiple NVM technologies
addressing the different requirements for each product type or
category. In fact, the 2010 International Technology Roadmap
for Semiconductors (ITRS) just recommended two additional
NVM technologies that should receive accelerated research and
development leading to commercial NVM products: Spin Transfer
Torque MRAM (STT-MRAM) and Redox RRAM [1].
NVM Test Requirement
Evolution and Overview
With floating gate flash memory, electrical characterization was
traditionally performed using DC instruments, such as source
measure unit (SMU) instruments, after pulse generators had
programmed and/or erased the memory cell. This requires some
type of switch to apply the DC or pulse signal alternately to the
test device. Occasionally, oscilloscopes were used to verify pulse
fidelity (pulse width, overshoot, pulse voltage level, rise time,
fall time) at the device under test (DUT). Measuring the pulse is
important because the flash memory state is quite sensitive to
the pulse voltage level. However, the use of oscilloscopes was
relatively rare, even in research, because the required setup for
oscilloscope measurements differed from that for the pulsesource/DC-measure approach. Even when scopes were used for
flash characterization, the complexity of measuring the transient
current meant that voltage was the only measurement taken
while pulsing.
Recently, standard instrumentation has improved, and now
it is possible to measure the current and voltage simultaneously
with a single instrument while applying pulses to a memory
device or material. Although this capability was possible
earlier, it required a rack of instruments and involved making
various tradeoffs related to cost, performance, and complexity.
In addition, these custom systems were typically created and
maintained by an in-house test instrumentation expert who
had the wide-ranging skills, experience, and significant time
necessary to integrate the various instruments into a system that
provided pulse source and measure. These earlier systems were
functional, but they were typically one-off creations with limited
test envelopes and cumbersome test controls and required
time-consuming data extraction. The measurement approach
typically used a load or sense resistor with an oscilloscope or
digitizer to measure the current. This is a proven technique,
www.keithley.com
but the effect of the load resistor on the voltage delivered to the
device has significant downsides for many pulse measurements.
Also, correlation across multiple systems and obtaining traceable
system-level calibration was effectively impossible.
The new instrumentation provides researchers with
additional data to gain a better understanding of NVM material
and device behavior in less time. Applying pulses while
simultaneously measuring the voltage and current with highspeed sampling provides better insight into the electrical and
physical mechanisms that provide the memory behavior. Adding
this transient characterization capability to DC characterization
provides fundamental data on intrinsic material properties and
device response.
Many materials and technologies are currently under
investigation for NVM, and each has unique aspects for the
physical memory behavior. However, the overall electrical
characterization of these various approaches share important
test parameters and methods. This commonality means that a
single test instrument is useful for characterizing a wide range of
memory technologies and device types.
Electrical characterization is crucial to a better understanding
of the physical aspects of the underlying technology. Regardless
of the particular memory technology under investigation,
pulsing is required to exercise the switching behavior. Pulsing
with simultaneous measurement provides the data necessary to
understand the dynamic behavior of the switching mechanism.
Different materials use different terms. For example, the terms
program/erase, set/reset, and write/erase are used to indicate
the fundamental storage of a 1 or 0 bit. These write/erase
procedures are done in a pulse mode to provide the overall
speed required for typical memory operations and simulate the
final product environment. The next section addresses important
test parameters that are common to a wide range of non-volatile
memory technologies.
Common NVM Test Parameters
Pulse amplitude is the required pulse height used to program
and erase the memory cell. Floating gate memory can require
15–20V, or even greater, during the write pulse. Most NVM
candidates require 3–5V. The goal for replacement NVM
technologies is lower pulse amplitudes, but early research prior
to any dimensional scaling or material optimization can require
6–8V. Many technologies require bipolar pulses at these voltage
levels, but some recent pulse I-V solutions do not provide these
higher voltages when used in a bipolar sourcing mode.
Pulse amplitude fidelity is one of the most important
parameters in NVM testing because memory state switching
behaviors are non-linear (for example, Fowler-Nordheim current,
phase transformation, filament creation/annihilation); therefore,
these devices are sensitive to the amplitude of the voltage pulse.
Pulse amplitude fidelity parameters are specified in terms of
pulse level accuracy, ringing, overshoot, and undershoot.
Minimizing the ringing, overshoot, and undershoot is critical to
the design of the pulse instrument. Modern pulse I-V systems
can provide overshoot and ringing specifications of 3% or less.
However, it’s important to be aware that pulse shape fidelity
at the test device is significantly affected by the connection
setup, cables, pulse parameters, device impedance, timing, and
impedance mismatches.
In addition to accurate pulse levels, newer technologies
require complicated and easily adjustable waveforms, not just
a single standard square pulse. For example, testing of ReRAM
devices often requires pulse sweep up/sweep down profiles while
simultaneously measuring the current. FeRam testing requires
the PUND (Positive, Up, Negative, Down) four-pulse sequence.
PRAM (phase-change memory) testing requires the ability to do
sweeps of virtually any pulse parameter, such as sweeping the
fall time of one of the four pulses in the RESET-measure-SETmeasure multi-pulse waveform. All of these memory technologies
require the ability to output multi-pulse waveforms consisting
of arbitrary segments, together with multiple measurements
within each waveform. Endurance testing requires the ability to
output complex arbitrary waveforms quickly without the need
for additional setup time or overhead. This requirement further
separates the capabilities of newer pulse instrumentation from
those of the traditional two-level pulse generator.
Pulse timing parameters, such as rise time, fall time, and
pulse width in test equipment will continue to be very important,
especially with the general trend toward faster pulsing, with
pulse widths trending from 100ns down to <10ns. For some
technologies, such as PRAM, fall time is a critical parameter to
define how the RESET operation takes place. This is important
because traditional pulse generators typically have limited rise/
fall timing ranges and do not permit, for example, a 20ns rise
time combined with a 2ms fall time. In general, shorter pulse
widths and faster transition times are preferred, but there are
practical limitations due to typical interconnect impedances,
measurement constraints, and instrumentation tradeoffs.
The need for dynamic, simultaneous ultra-fast current and
voltage measurement is driven by emerging NVM technologies
such as the phase-change and ferro-electric approaches discussed
previously. The Model 4225-PMU/4225-RPM combination
provides simultaneous voltage and current measurement,
which is important when the dynamic resistance of the material
represents the electrical manifestation of the physical mechanism
of the bit storage.
The scaling trend leads to smaller and smaller devices.
This trend requires measuring smaller currents while pulsing,
which calls for some type of a pre-amplifier for the current
measurement. To minimize parasitic effects of the cable
capacitance and have better control of the amount of energy to
the test device, a remote pulse amplifier, connected within
15–25cm (6–10 inches) of the DUT, presents a significant benefit.
This is especially important for phase change memory (PCM) and
ReRAM characterization.
Current compliance or current control is important for
testing some NVM technologies, such as ReRAM and PRAM.
Usually, this is done using DC instruments and sometimes
implemented in custom pulse setups. It is not clear if the current
compliance in DC instrumentation is providing sufficiently fast
control of the current to meet the typical requirements. For
pulsing current control, it is desirable to install the current
control device as close as possible to the test device to avoid
the possibility of current discharging from the interconnect
capacitance into the test device.
To simplify and speed up testing, switching between pulse
and DC instruments is necessary. To float connections in flash
program/erase cycles for endurance testing, switching has to
be fast (10–100µs) because the switching must occur between
the program and erase pulses to allow for a very large number
of stress waveforms. This type of switch should be controlled
directly by the pulse generator and located within the pulse
instrument for fast control. Typically, this switching is performed
by a solid state relay (SSR) for each pulse channel.
Channel synchronization is necessary for NVM testing
that requires multiple pulse source and measure channels. Two
channels are sufficient for PRAM and ReRAM characterization,
to force and measure on both sides of two-terminal devices.
For NVM that utilizes transistors as access devices, three or
four channels of pulse I-V may be required. In flash memory,
two or four channels are needed. Traditional pulse instruments
are difficult to synchronize because there are a variety of
trigger synchronization methods, each of which has different
complexity/trigger performance tradeoffs associated with it.
Modern pulse I-V instruments offer internal trigger routing
and automatic synchronization in addition to the integrated
measurement capability.
As mentioned previously, different NVM technologies have
slightly different measurement needs. Table 1 summarizes the
important test parameters for a few memory technologies.
Capabilities of Model 4225-PMU and
Model 4225-RPM for NVM Characterization
The Model 4225-PMU Ultra-Fast I-V Module (Figure 2) is a singleslot instrument card for the Model 4200-SCS. It has two channels
of voltage pulse sourcing with integrated, simultaneous real-time
current and voltage measure for each channel. There are two
types of measurements: sample and spot mean. The sample
type is used to capture the time-based current and voltage
waveforms critical for understanding transient or dynamic
behaviors. The spot mean type provides DC-like current and
voltage measurements for I-V characterization. The real-time
sampling capability is critical to capture the transient behavior of
NVM materials in a single waveform, because applying repetitive
waveforms will cause the memory switching behavior or even
damage to the material itself.
www.keithley.com33
Table 1. Summary of important test parameters for NVM technologies.
NVM
Type
Pulse
Level
Pulse Transition, Pulse Fidelity
Minimizing pulse overshoot is critical
Floating
because of the non-linear nature
15–20V+
Gate Flash
of the tunneling current used to
program or erase the cell.
Asymmetric rise and fall required:
Fast fall time (≤20ns) for RESET pulse,
PRAM
~±8V
(PCRAM)
slow fall time for crystallization (tens
to hundreds of nanoseconds).
ReRAM
~±6V
Accurate pulse levels and transition
control allow for the study of ion
transport or filament formation.
FeRAM
~±5V
Pulse transition control allows for
polarization change characterization.
Transient Measurement
Emerging need to increase
understanding of charge transport.
Multi-Channel
Multi-Level Pulses
Synch
Required for bipolar pulse waveform for
write and erase. Also require high-speed
Emerging need
control for solid state relay (high impedance)
on source and drain during FN erase.
Necessary when
transitioning to
1T1R 1 device
Critical to understand material and Unipolar, multi-level pulsing for resetcell behavior.
measure-set-measure waveforms.
Emerging requirement to help
understand material and cell
behavior, assist in search for lowercurrent variants.
High speed charge measurements
are critical to characterizing
capacitive memory effect.
Required for both unipolar and bipolar
technologies.
Necessary
for 1T1R 1
device structure
Bipolar pulsing for write and erase
(PUND method).
Useful for 1T1C 2
device structure
1. 1T1R = memory cell consisting of one transistor and one resistor, with the transistor providing the control and access to the cell.
2. 1T1C = memory cell consisting of one transistor and one capacitor, with the transistor providing the control and access to the cell.
4225-PMU Source/Measure Block Diagram
10V
40V
Output
Relay
SSR
50Ω
I
Ch1
V
Controller
RPM1
Source
Pulse
Seg Arb
RPM2
V
10V
40V
I
50Ω
Choose standard (2-level)
or multi-level pulsing
Figure 2. Model 4225-PMU Ultra-Fast I-V Module and two Model 4225-RPM
Remote Amplifier/Switch Modules.
The Model 4225-RPM Remote Amplifier/Switch is an optional
addition to the Model 4225-PMU. This small box is located
near the DUT and provides the lower-current measurement
ranges necessary for characterization of many NVM materials
and technologies. In addition, the Model 4225-RPM provides
switching for the Model 4200-SCS’s source measurement
units (SMUs) and CVU signals to allow for high resolution DC
measurements and C-V measurements. The Model 4225-RPM is
a single-channel device, so two Model 4225-RPM modules are
required to match the two channels from the Model 4225-PMU.
The Model 4225-RPM module is designed to be located close to
the test device (≤30cm or one foot), minimizing cabling effects to
provide improved pulse shape and high speed measurements.
The PMU/RPM combination provides the test capabilities
to characterize the existing and emerging NVM technologies
described previously. Figure 3 is a block diagram of the Model
4225-PMU. Note that both channels have both current and
voltage measurement (two A/D converters per channel). Each
channel can independently source ±10V or ±40V (into high
impedance). Figure 4 is a block diagram of the Model 4225RPM: on the left side, note the inputs from the Model 4200-SCS
www.keithley.com
Ch2
SSR
Output
Relay
Integrated V & I Simultaneous Sampling on each
channel (2 A/D converters per channel)
Figure 3. Model 4225-PMU Block Diagram. It is possible to use the Model
4225-PMU with two Model 4225-RPMs by using the RPM1 and RPM2
connections. The SSR shows the solid-state relay, which is useful for high
impedance mode when performing program or erase on flash memory
devices via Fowler-Nordheim tunneling.
Sense
Triax
Force
Pot
SMA
Cur
4225-RPM Block Diagram
Inputs
Outputs
SMU
Force
Triax
CVU
Pulse
RPM
Sense
Current Measure Ranges
10 mA
1 mA
100 µA
10 µA
1 µA
100 nA
Figure 4. Model 4225-RPM Block Diagram. The Model 4225-RPM is both a
current pre-amplifier for high speed current measure (in Pulse mode) and
a switch for use in choosing between Pulse I-V, SMU, or C‑V measurement
modes for other instruments in the Model 4200-SCS chassis.
Pulse Level
Each Model 4225-PMU channel has two source ranges. The 10V
source range can output from –10V to +10V (20V amplitude)
into high impedance and covers most of the modern NVM
candidates. For testing of existing floating gate flash or early
non-optimized emerging materials that require higher voltages,
the Model 4225-PMU also has a 40V range, outputting from –40V
to +40V (80V amplitude) into high impedance. Because both
source ranges are bipolar, both unipolar and bipolar memory
technologies can be characterized.
Transient Measurement and Multi-Channel Synch
The Model 4225-PMU has two channels, each of which has
two A/D converters that sample the voltage and current
simultaneously. With the two channels, both the gate and drain
of a device may be synchronously pulsed while sampling the
voltage and current on both channels to capture the dynamic
response of the DUT. If more than two channels of pulse I-V
are required, multiple Model 4225-PMU cards may be installed
in a single chassis and all channels will be automatically
synchronized to within ±2ns. This simultaneous, synchronous
measurement is critical to understanding the switching behaviors
of NVM materials and devices.
The Model 4225-RPM adds measure ranges from 10mA down
to 100nA to permit transient analysis of the small currents that
are characteristic of the switching states. In addition, the RPM
has switching capability to permit routing SMU instrument or C-V
signals to the test device.
Pulse Transition and Multi-Level Waveforms
The Model 4225-PMU has two independent channels, with
timing parameters adjustable from 20ns to 40s. The shortest
pulse possible is 40ns (FWHM, Full Width at Half Maximum
of pulse amplitude), but wider pulses are required to measure
smaller currents.
Multi-level or multi-pulse waveforms can be created by
linking together linear segments (voltage vs. time) using the
Segment ARB® feature. Each channel has a maximum of 2048
segments that may be used in a single sequence or used across
multiple sequences. A sequence is a set of segments that allows
for looping, typically to provide stresses from a looped sequence
interspersed with a measure sequence.
Measurements are controlled on a segment-by-segment basis,
so only the required data is collected, maximizing the use of
the available sample storage memory, similar to the segmented
memory feature available in high-end oscilloscopes. In addition,
two measurement types are available: sample and mean. The
sample type is used to capture time-based signals, useful for
capturing transient behaviors and validating proper connections
by evaluating the pulse shape behavior. The mean measurement
type is used when measuring I-V characteristics. Both types
can be applied to the entire segment, a partial segment, or all
segments in the waveform on a channel-by-channel basis.
The ability to output dozens or hundreds of unique pulses
within a single waveform decreases test time over that of
traditional pulse instruments, which could only pulse between
two pulse levels. Traditional floating gate flash used two pulse
channels to create a two-pulse waveform of three voltage
levels for a single test device terminal (Figure 5). In addition,
it required an external switch between the pulse generators
and the DUT to route each pulse alternately to the terminal.
This external switch added complexity and cost, and most
importantly, increased test time. Pulsing the gain and drain
simultaneously required four pulse channels and two sets of
switches. Figure 6 shows a more complicated waveform used for
testing phase-change memory. The red boxes on the waveform
represent measurements. Note that the entire waveform,
consisting of four pulses of varying widths and heights, is output
by a single Model 4225-PMU channel, using 16 segments out of a
maximum of 2048.
V
0
Program
Erase
Switch Pulse
Channels
chassis; on the right side, note the output to the DUT. The top
half (blue, red, and heavy green lines) is the switching part of the
Model 4225-RPM. Note that the SMU and CVU pathways support
four-wire connections throughout. The thinner green lines near
the bottom represent the various pulse current measure ranges.
Time
Figure 5. Flash memory Program and Erase waveform: two pulses, three
levels. Traditional pulse generators can only output two level pulses, so
flash testing required two pulse channels for each test pin and a time-costly
switch to apply the second (negative) pulse to the test device. Note that this
setup is only for one device terminal. For a typical pulsing on the gate and
drain, this setup would have to be duplicated.
V
3
10
2
1
11
9
4
5
6
7
8
12
13 14 15 16
Time
Figure 6. Multi-level pulse waveform consisting of 16 linear voltage segments
(gray numbers) and four measurements (red boxes). New pulse source
hardware allows creating multi-segment voltage waveforms to provide
several pulses within a single waveform, as well as adding integrated V and I
sampling (not shown).
www.keithley.com35
Making Connections to the Device
Figure 7 shows the general method for connecting to a twoterminal device using two Model 4225-RPM Remote Amplifier/
Switches. The Model 4225-RPM is an optional item for the dualchannel Model 4225-PMU Ultra-Fast I-V Module and is necessary
for NVM characterization. Specific interconnect diagrams are
shown for each of the three devices in the NVM_Examples
project discussed later in this application note. Note that the
ground connection for NVM devices, which is usually necessary
on memory cells that use a transistor as the selection device,
should be made to the local shield and the shields for each
channel should be connected together (Figure 8). These
4225-RPM #1
Inputs
Optional:
CVU
Outputs
SMU 1
Force
Sense
to PMU
NOTE: Connect outer
shields together
Inputs
Outputs
4225-RPM #2
SMU 2
Optional:
CVU
DUT
Force
shield connections are important to ensure the relatively high
bandwidth required for the fast transitions and narrow pulses
required for today’s emerging NVM technologies.
It is possible to connect a two-terminal device using a single
channel, with the low side of the device connected to the shield,
or ground return, of the channel. This is the traditional way
to test a simple two-terminal device and is reasonable for DC
characterization. However, because of transient effects during
the pulse, using two channels on a two-terminal device provides
better results. See the “Optimizing Measurements” section for
additional information.
Using the KTEI Software to Control
the Models Model 4225-PMU and
Model 4225-RPM for NVM Testing
A set of example tests for NVM characterization is included with
KTEI 8.2 or higher, contained in the NVM_Examples project (in
the C:\sModel 4200\kiuser\Projects\_Memory folder). This project
(Figure 9), which includes tests and data for flash, PRAM, and
FeRAM devices, demonstrates the capabilities of the Model 4200SCS and particularly the Model 4225-PMU with the Model 4225RPM. The Model 4225-PMU/4225-RPM combination provides the
fundamental pulse and transient I-V test capabilities to investigate
and characterize a wide range of NVM materials and devices. A
brief explanation of the user test modules (UTMs) follows.
Sense
to PMU
Figure 7. Connection to two-terminal test device.
Inputs
Optional:
CVU
Outputs
4225-RPM #1
SMU 1
Force
Sense
to PMU
G
NOTE: Connect outer
shields together
B
D
DUT
Inputs
Outputs
4225-RPM #2
SMU 2
Optional:
CVU
S
Force
Sense
to PMU
Figure 8. Connection to four-terminal test device with two Model 4225-RPMs.
www.keithley.com
Figure 9. The devices and tests included in the NVM_Examples project for the
Keithley Model 4200-SCS.
Table 2. User modules in the nvm user library.
User Module
Used in
Description
Outputs one or two V-shaped waveforms while sampling the voltage and current. Also calculates the accumulated charge.
doubleSweep.c
N/A
Requires one Model 4225-PMU with two Model 4225-RPMs.
PRAM, FeRAM Similar to doublesweep but returns the waveform data split into segments used by the PRAM I-V sweep. Also calculates the
doubleSweepSeg.c
accumulated charge, used in FeRAM hysteresis test. Requires one Model 4225-PMU with two Model 4225-RPMs.
Performs the pulse stress and DC measure cycling. Applies the maximum number of requested waveforms while measuring
flashEndurance.c
Flash
every log10 waveform count. The stress waveform is the typical program and erase and the measurement is performed by the
SMU instruments. Requires two SMU instruments and one Model 4225-PMU with two Model 4225-RPMs.
Applies the program and erase pulse waveform while sampling the voltage and current. Can choose to output both program
flashProgramErase.c
Flash
and erase or just one pulse. Requires one Model 4225-PMU with two Model 4225-RPMs.
This utility module does not test but saves intra-test status information to a file (C:\nvmlog.txt). This is useful for debugging test
nvmDebug.c
issues, especially when modifying existing test modules or writing new ones.
Performs the stress and measure cycling for phase-change memory. The test uses the RESET-SET waveform for stresses and
pramEndurance.c
PRAM
every log10 count uses a RESET-measure-SET-measure to obtain the resistance measurements post-RESET and post-SET.
Requires one Model 4225-PMU with two Model 4225-RPMs.
This outputs the RESET-measure-SET-measure waveform while sampling the voltage and current. Requires one Model 4225pramSweep.c
PRAM
PMU with two Model 4225-RPMs.
This is the underlying test routine used by all of the tests. Requires one Model 4225-PMU with two Model 4225-RPMs. For
Flash,
pulse_test.c
PRAM, FeRAM additional information, see the read_me.rtf file in C:\s4200\kiuser\usrlibs\nvm.
Performs the stress and measure cycling for FeRAM. The stress waveform is output a maximum number of times and is
pundEndurance.c
FeRAM
interrupted on a log10 basis to obtain the PUND and Psw and Qsw extractions.The stress waveform is the PUND waveform and
the measure is the two V-pulses controlled by doubleSweepSeg. Requires one Model 4225-PMU with two Model 4225-RPMs.
This test outputs the four-pulse PUND waveform while sampling voltage and current. The routine extracts the P, U, N, and D
pundTest.c
FeRAM
values from the waveform. Requires a Model 4225-PMU with two Model 4225-RPMs.
reramSweep.c
ReRAM
This test outputs 2 pulses, one pulse to be used for “SET” and another for “RESET”. Pulses can be generated by the SMU
instrument or by the PMU. SMU instrument sweeps are slow (ms range and higher) and PMU sweeps are fast (ns to ms range).
Both voltage sweeps (PMU and SMU instrument) can be used with current limit. Requires one PMU with 2 RPMs and 2 SMU
instruments.
reramEndurance.c
ReRAM
Performs the stress and measure cycling for ReRAM memory. The test uses the RESET-SET waveform for stresses and every
log10 count uses a RESET-measure-SET-measure waveform to obtain the resistance measurements for the post-RESET and
post-SET states. Requires one PMU with 2 RPMs and two SMU instruments.
util.c
vt_ext.c
Flash
This utility module does not test but has support routines for the other tests, such as hardware initialization, calculation of
total points, and calculation of maximum allowable sample rate.
Performs a VG-ID sweep and extracts the voltage threshold (V T ). Requires two SMU instruments.
The tests in the NVM_Examples project are UTMs, which
are customized for each memory type but share an underlying
approach to minimize the effort required to accommodate
additional NVM materials or device types. All of the UTMs used
in this project are contained in the nvm user library (Table 2).
These modules are used in NVM_Examples project, but they may
also be added to another project. If additional capabilities or test
types are required, the source code for the modules is included
with the Model 4200-SCS and may be modified by using the
Keithley User Library Tool (KULT) with the optional compiler
(order number: 4200-Compiler).
Flash Testing
Flash cells are the dominant type of NVM and because they are
implemented on the foundation of MOSFET transistors, they have
standard source, gate (actually, control gate or CG), drain, and
bulk/substrate connections (Figure 12). Fowler-Nordheim current
tunneling through gate oxide and hot carrier injection represent
the two standard methods for storing and removing charge from
the floating gate (Figure 10). These methods are degradation
mechanisms of standard (non-NVM) MOSFET transistors, which
also explains the limited endurance of flash memory. Because
the majority of the NVM market is Flash, this is an active area for
research and development [3, 4].
Initial flash memory characterization usually consists of
determining the appropriate values for the voltage pulse height
and the pulse width to provide the target threshold values for
the program and erase states. Instead of the traditional onebit (1/0) cell, most modern flash memory utilizes two or three
bits per cell, which maps to four or eight unique V T levels. The
increasing number of V T levels requires more precise pulse level
performance and higher pulse voltages.
There are six UTMs for the flash device in the NVM_Examples
project (Figure 11), which is located in the C:\sModel 4200\
kiuser\Projects\_Memory folder. These tests support a standalone NAND or NOR cell. There are three tests that measure the
Program
Tunnel Oxide
Erase
Tunnel Oxide
CG
FG
- - -
S
D
-
B
-
0V
CG
FG
- - -
S
D
B
0V
Figure 10. Flash memory structure, showing program and erase conditions
for Fowler-Nordheim tunneling.
www.keithley.com37
Sense
Triax cables
CVU Pot
CVU Cur
4200-SMU 2
SENSE
FORCE
Force
Sense
RPM cable
G
NOTE: Connect outer
shields together
Sense
Force
4225-PMU
• Two or more SMU instruments, either
medium-power Model 4200-SMUs or
high-power Model 4210-SMUs
CVU Pot
CVU Cur
RPM1
RPM2
S
B
D
DUT
4225-RPM #2
Triax cables
Outputs
• Model 4200-SCS with KTEI
8.2 or higher
• One Model 4225-PMU with two
Model 4225-RPMs
4225-RPM #1
Force
Inputs
SENSE
FORCE
Outputs
4200-SMU 1
Inputs
V T of the flash memory transistor, two
that apply program or erase pulses, and
one endurance test. The connection to a
stand-alone flash memory cell is shown in
Figure 12. In these tests, only two pulse
I-V channels are used, so both the source
and bulk are connected to the Model
4225-RPM shields. These tests require the
following hardware:
Force
Sense
RPM cable
Figure 12. Connection to four-terminal floating gate flash device.
Program Voltage and Current Waveform
14.0E+0
60.0E–6
12.0E+0
50.0E–6
10.0E+0
40.0E–6
8.0E+0
30.0E–6
The three Vt tests are the same and
use two SMU instruments to measure the
V T of the flash transistor. The Program
and Erase tests apply pulse waveforms
to program and erase a stand-alone flash
memory cell. Because the Model 4225PMU has integrated high speed sampling,
the Program and Erase tests also capture
the voltage and current waveforms
(Figure 13). Both of these tests use
Fowler-Nordheim tunneling to perform
the charge transfer, so the drain voltage =
0V and the graphs show the gate voltage
and current only. However, both the
gate and drain current and voltages are
measured in each test, so they can be
displayed if desired.
Measuring the current while pulsing
was not previously practical. Because the
tunneling current is non-linear with the
applied voltage, the measured current
provides additional information on
how close the voltage is to providing a
sufficient electrical field for programming
or erasing. The transient current provides
www.keithley.com
Data:gateV
Data:gateI
4.0E+0
2.0E+0
20.0E–6
10.0E–6
0.0E–6
0.0E+0
–2.0E+0
–10.0E–6
–4.0E+0
–20.0E–6
–6.0E+0
–30.0E–6
–8.0E+0
–40.0E–6
–10.0E+0
–50.0E–6
–12.0E+0
–60.0E–6
–14.0E+0
0.0E+3
Gate Current (A)
Figure 11. Tests for the flash device in the NVM_
Examples project.
Gate Voltage (V)
6.0E+0
1.0E+3
2.0E+3
3.0E+3
4.0E+3
5.0E+3
6.0E+3
7.0E+3
–70.0E–6
8.0E+3
Time (s)
Figure 13. Pulse current and voltage waveforms for a Program pulse on a NAND cell using a Model 4225PMU with Model 4225-RPMs. Note the capacitive charging and discharging current during the pulse
transitions (see “Optimizing Measurements” for an explanation). This graph is from the Flash Program
test (user module flashProgramErase).
information on the dynamic current
flow and overall charge transfer. The
use of the transient current and voltage
information can be correlated to the
DC-based V T results to provide additional
understanding of the program and erase
processes, which may be unique for
different structures, dimensions, and
materials. The results of the program and
erase pulses are in the Vt_Programmed
and Vt_Erased tests (Figure 14).
The last test is flashEndurance, where
the program+erase waveform is applied
in increasing log intervals and then the
program and erase V T are measured and
plotted (Figure 15).
Flash Test Modules
There are three modules for testing
of either NAND or NOR floating gate
devices: flashProgramErase, vt_ext, and
flashEndurance. The flashProgramErase
module applies a pulse waveform of
either the program pulse, the erase
pulse, or both. The vt_ext module
performs a VG-ID sweep, using the
the pulses or the SMU instrument voltages
alternately to the device, so no recabling
is required.
NAND Flash Vg-Id Curves
60.0E–6
The flashProgramErase module uses
the Model 4225-PMU and the Model 4225RPMs to output single-pulse or dual-pulse
waveforms using the Segment ARB feature
of the Model 4225-PMU. The module
supports the two channels of the Model
4225-PMU, so unique pulse voltages can
be output for the gate and drain. This
module also measures the waveforms for
diagnostic purposes, such as verifying
pulse shape performance and proper
voltage levels.
50.0E–6
Program Vg-Id
Erase Vg-Id
Drain Current (A)
40.0E–6
30.0E–6
20.0E–6
10.0E–6
0.0E–6
0.0E+0
200.0E–3
400.0E–3
600.0E–3
800.0E–3
1.0E+0
1.2E+0
1.4E+0
1.6E+0
1.8E+0
2.0E+0
Gate Voltage (V)
Figure 14. V T sweeps for a programmed and erased cell taken with SMUs, from the Vt_Erased UTM (user
module vt_ext). In this graph, the difference in the V T after program and after erase is about 180mV.
NAND Endurance
2.500E+0
2.400E+0
2.300E+0
Voltage Threshold (V)
2.200E+0
Program Vt
Erase Vt
2.100E+0
2.000E+0
1.900E+0
1.700E+0
1.600E+0
1.500E+0
1.400E+0
10.0E+0
100.0E+0
1.0E+3
Iterations
Figure 15. NAND flash endurance results from the Flash UTM Endurance test (user module
flashEndurance).
SMU instruments, and extracts the
threshold voltage. The flashEndurance
module applies an increasing number
of program+erase waveforms while
periodically measuring (on a log interval
of the number of waveforms applied) the
program and erase voltage thresholds and
plotting them.
Table 3 lists the input parameters for the
flashProgramErase module. This module
has the settings for both the gate and
drain program and erase pulses. It uses
both channels of the Model 4225-PMU,
each connected to a Model 4225-RPM
(Figure 12). This module defines and
outputs Program and Erase waveforms
as shown in Figure 16. To measure
the voltage threshold after sending
the program and/or erase pulses, use
the vt_ext module, described in the
next section.
If only one pulse is desired, set the
other pulse levels to 0. For example, if
the erase pulse is not desired, set drainE
= gateE = 0. This will make the erase
portion of the waveform stay at 0V.
1.800E+0
1.0E+0
Setting Up the Parameters in
the flashProgramErase Module
For initial characterization and
determination of appropriate pulse
parameters for program and erase, the
flashProgramErase and vt_ext modules
allow for adjustments of the pulse
parameters, then the measurement of
the voltage threshold using the SMU
instruments. Note that the Model 4225RPM switching capability is used to apply
For Fowler-Nordheim tunneling,
usually the drain voltage is 0V, and just
the gate pulse (gateP), is set to either
pushing charge into the floating gate
(gateP = +V) or clearing the charge from
the floating gate (gateE = –V). For the
hot carrier injection method, the drain
voltage is positive, creating the field
in the channel necessary to create the
hot carriers.
Setting Up the Parameters
in the vt_ext Module
Table 4 lists the input parameters for the
vt_ext module. This module performs
a VGS-ID sweep and returns transistor
www.keithley.com39
threshold voltage using the maximum gm method, where the
threshold voltage is defined as:
to the same value (vgs_ pts = ids_ pts = gm_ pts). Typically, a
30-point sweep is sufficient to obtain a reliable V T.
Note that this module assumes that DrainSMU is connected
to RPM2 and GateSMU is connected to RPM1. As shown in Figure
12, the test device Source and Bulk connections are to the RPM
shield, so the appropriate setting in vt_ext are SourceSMU =
BulkSMU = “”.
V T = VG MAX – ID MAX /gm MAX – 1/2 * V DS
VG MAX is the maximum gate voltage
ID MAX is the maximum drain current
gm MAX is the transconductance (gm) at the maximum gm.
The number of points in the V T sweep is set by vgs_ pts and
ids_ pts, gm_ pts. Note that all three parameters must be set
Setting Up the Parameters in the
flashEndurance Module
Table 3. Parameters in the flashProgramErase module.
Table 5 lists the input parameters for the flashEndurance
module. This module has the settings for both the gate and
drain program and erase pulses and the maximum number
of stress loops. It uses both channels of the Model 4225-PMU,
each connected to a Model 4225-RPM (Figure 12). This module
outputs a number (max_loops) of Program and Erase waveforms
(Figure 16) to the test device. The module will use log10 stress
counts from the max_loops and the desired number of iterations
in iteration_size to determine how many program+erase pulse
waveforms to apply for each stress interval. After each stress
interval, the vt_ext is run, once after the program pulse is
applied and once after erase. Note that iteration_size, vtE_size,
and vtP_size must have the same value (iteration_size = vtE_size
= vtP_size).
Parameter
gateP
drainP
widthP
gateE
drainE
widthE
Range
–40V to +40V
–40V to +40V
20ns to 1s
–40V to +40V
–40V to +40V
20ns to 1s
riseT
20ns to 33ms
delayT
20ns to 1s
loops
1 to 1012
V
Description
Gate Program Pulse Voltage
Drain Program Voltage
Program Pulse top width
Gate Erase Pulse Voltage
Drain Erase Pulse Voltage
Erase Pulse top width
Pulse transition time, both rise and fall for
Program, Erase
Pulse delay time, before Program pulse and
between the Program and Erase pulses
Number of times to output the
ProgramErase waveform
widthP
delayT
riseT
0
Table 5. Parameters in the flashEndurance module.
Gate
Program
riseT
GateP
Parameter
delayT
delayT
Time
Gate
Erase
GateE
0
Drain
Erase
riseT
Drain
Program
riseT
DrainP
widthE
DrainE
drainP
–40V to +40V Drain Program Voltage
20ns to 1s
–40V to +40V Gate Erase Pulse Voltage
drainE
–40V to +40V Drain Erase Pulse Voltage
widthE
20ns to 1s
delayT
Pulse transition time, both rise and fall for
Program+Erase pulses
20ns to 1s
Pulse delay time, before Program pulse and
between the Program and Erase pulses
Table 4. Parameters in the vt_ext module.
vlow
–40V to +40V
vhigh
vds
–40V to +40V
–40V to +40V
vbs
–40V to +40V
vgs_ pts
ids_ pts
gm_ pts
10 to 100
10 to 100
10 to 100
vt
www.keithley.com
Description
SMU instrument for DUT Drain
SMU instrument for DUT Gate
SMU instrument for DUT Source, if used
SMU instrument for DUT Bulk, if used
Starting sweep voltage for the Gate SMU
instrument
Final sweep voltage for the Gate SMU instrument
Drain bias voltage for the Drain SMU instrument
Pulse delay time, before Program pulse and
between the Program and Erase pulses
Number of points in the Vgs-Id sweep
Number of points in the Vgs-Id sweep
Number of points in the Gm array
Resulting voltage threshold from the
max Gm calculation
max_loops
Erase Pulse top width
20ns to 33ms
Figure 16. Waveform parameters for the flashProgramErase module.
Range
SMU1 to SMUn
SMU1 to SMUn
SMU1 to SMUn
SMU1 to SMUn
Program Pulse top width
gateE
riseT
Parameter
DrainSMU
GateSMU
SourceSMU
BulkSMU
Description
–40V to +40V Gate Program Pulse Voltage
widthP
V
Range
gateP
1 to 1012
Number of times to output the
Program+Erase waveform to stress the
test device
vds
–40V to +40V
V T sweep drain bias voltage for
the Drain SMU
vgsstart
–40V to +40V
V T sweep starting sweep voltage for
the Gate SMU
vgsstop
–40V to +40V
V T sweep final sweep voltage for
the Gate SMU
vtP_size
2 to 100
Number of points in the VGS -ID sweep after
the Program pulse
vtE_size
2 to 100
Number of points in the VGS -ID sweep after
the Erase pulse
Iteration_size
2 to 100
Number of times to measure the V T during
the max_loops stress
PRAM Material Testing
Top Electrode
Active area
I
Resistor (heater)
Crystalline GST
Bottom Electrode
Figure 17. Tests for the PRAM two-terminal device in the
NVM_Examples project.
Figure 18. GST [Germanium (Ge), Antimony (Sb), and Tellurium (Te)]
structure.
Phase-change memory (PRAM, PCRAM, or PCM) cells are made
of a chalcogenide alloy (i.e., with at least one element from the
VI group of the periodic table of the elements, plus one element
each from the V and IV groups). These same types of materials
are also widely used in the active layers of rewritable optical
media such as CDs and DVDs. PRAM is one type of resistive
memory. Other types of resistive memory include OxRRAM and
TMO-RRAM (both are types of Redox RAM) and cation-based,
conductive-bridge memory (CBRAM). Figure 17 shows the tests
included for phase-change memory testing in the NVM_Examples
project (in the C:\sModel 4200\kiuser\Projects\_Memory folder).
at the top of the pulse, when the cell cools from melting point
to crystallization. After reaching crystallization temperature, the
crystalline order is frozen. For example, the pulse fall time may
be 20ns, but it might take 5ns to go from TMELT to TCRYSTALLIZE.
In addition, some PRAM structures have a “Select Diode” or
“Protection Diode” integrated in series with the PRAM cell. In
addition to some drawbacks for the I-V curve, this diode will
make the effective fall time shorter. The current flowing through
a diode is exponentially dependent on the voltage, so a small
decrease in the voltage results in a dramatic reduction in the
current flowing through the diode, especially at small currents
(~<50µA). Therefore, when testing devices with the series
selection diode, the pulse fall time is not as critical for small
current test devices and permits testing with standard pulse I-V
instrumentation, such as the models 4225-PMU and 4225-RPM.
Through the application of heat in the form of an electrical
pulse (or a laser pulse in CDs/DVDs), PCM cells can be switched
rapidly from an ordered crystalline phase (which has low
resistance) to a disordered, amorphous phase (which has
much higher resistance). The switch from the crystalline to the
amorphous phase and back is triggered by melting and quick
cooling (or a slightly slower process known as re-crystallization).
GST [germanium (Ge), antimony (Sb), and tellurium (Te)], with a
melting temperature from 500° to 600°C, has emerged as one of
the most promising materials for PCM devices (Figure 18).
These devices can store binary data because of the differing
levels of resistivity of the crystalline and amorphous phases of
these alloys. The high resistance amorphous state represents
a binary 0; the low resistance crystalline state represents a 1.
Multiple resistive levels will permit multi-bit PCM, which has
been demonstrated, allowing PCM to scale and provide lower
cost-per-bit. [5, 6]. These states are stable over time, which is
important for any commercial application [7].
In the amorphous phase, the GST material has short-range
atomic order and low free electron density, which means higher
resistivity. This is sometimes referred to as the RESET phase
because it is usually formed after a RESET operation, in which
the temperature of the cell is raised slightly above the melting
point, then the material is suddenly quenched to cool it. The
cooling rate is critical to the formation of the amorphous state.
If the rate is too slow, then the material will be less amorphous.
For so-called “slow materials,” the cooling rate is about 30
nanoseconds; for ’”fast materials,” it is in the range of single
nanoseconds or faster [8]. The fall time of the pulse can be
slower than the required speed; what is important is the fall rate
Much like other types of NVM technologies, a PCRAM cell
must be formed before it displays the consistent switching
necessary to be a memory element. One way to explain the
forming process is that it creates the active area of the PCRAM
cell. The active area is the portion of the chalcogenide material
that transitions between the amorphous and crystalline states.
Figure 18 shows a half-circle shape that represents the active
area and is the result of the forming process. The goal of the
forming process is reproducible cycling between the SET and
RESET states. One challenge to testing a new or unknown cell
is to determine the appropriate pulse parameters (amplitude,
rise/fall, width) for the RESET and SET pulses. This is usually an
iterative process, first starting with a reasonable RESET pulse,
then optimizing the SET pulse. The PRAM_Reset and PRAM_Set
tests are useful for this initial determination of the pulse
parameters. Just as with flash memory, it is possible to over-stress
the cell and permanently damage it. Because the RESET voltage
is the largest, the search for the appropriate RESET voltage must
be done with care.
There are five UTMs for testing phase-change memory devices
in the NVM_Examples project (Figure 17). The first test applies
a waveform with a RESET pulse and measure R pulse to the
PRAM device (Figure 20). The second test is similar to the first
test but applies a waveform consisting of a SET pulse followed
by a measure pulse. There is an I-V sweep test that shows the
switching effect (Figure 22). The next test shows the typical R-I
www.keithley.com41
curve, which is the resistance of the cell vs. the current (R-I) of
the SET pulse (Figure 24). The last test is an endurance test,
showing the change in the SET resistance vs. the number of
RESET-SET waveforms (Figure 26).
RESET Voltage and Current Waveform
8.0E+0
7.0E+0
Outputs
Inputs
3.0E+0
Data:VForce
Data:Imeas
2.0E–3
2.0E+0
1.0E–3
Data Variables 1
Data:resetR = 1.00000e+6
1.0E+0
Measure
0.0E+0
2.0E–6
4.0E–6
6.0E–6
8.0E–6
10.0E–6
12.0E–6
0.0E+0
14.0E–6
16.0E–6
18.0E–6
Time (s)
Figure 20. PRAM RESET waveform generated and measured by the Model
4225-PMU with Model 4225-RPMs (user module pramSweep).
4225-RPM #1
SMU Force
3.0E–3
Reset Pulse
The connection diagram is shown in Figure 19. To
obtain the best possible measurements, the connection
uses the source high, measure low method described in the
“Optimizing Measurements” section. No SMUs are used in these
tests, although adding SMU tests is possible.
4.0E+0
Current (A)
• One Model 4225-PMU with two Model 4225-RPMs
Voltage (V)
5.0E+0
• Model 4200-SCS with KTEI 8.2 or higher
CVU Pot
CVU Cur
4.0E–3
6.0E+0
These tests require the following hardware:
SMU Sense
5.0E–3
Force
Sense
RPM cable
NOTE: Connect outer
shields together
SMU Sense
4225-RPM #2
RPM2
CVU Pot
CVU Cur
Outputs
RPM1
Inputs
SMU Force
4225-PMU
R GST
Force
Sense
RPM cable
Figure 19. Connection diagram for PRAM tests. Note that the PRAM tests
included with NVM_Examples do not use the SMUs or CVU.
The Reset test applies a two-pulse RESET and measure
waveform to the test device and captures the voltage and current
waveforms (Figure 20). This transient response data is useful to
determine the resistance of the device and the degree of reset.
This test and the Set test are used to determine the proper
voltage and pulse width for reset and set pulses as well as a way
to send a reset-only or set-only waveform to the test device. Use
these tests to optimize the RESET and SET pulse parameters
before performing an endurance test.
Figure 21. I-V waveform from the PRAM IV_CURVE test generated and
measured by the Model 4225-PMU with Model 4225-RPMs (user module
doubleSweepSeg).
The I-V sweep test uses a single inverted-V-shaped pulse
to capture the PRAM switching characteristics. Because this
test uses a pulse with V and I sampling, there are two ways to
view the data. Figure 21 shows the waveform data, with I and
V plotted vs. time. This is not the typical way to display the
characteristics but shows how the test is performed. The second
graph (Figure 22) shows the typical PRAM I-V curve (current
vs. voltage).
The R-I curve is a typical phase-change measurement.
The SET pulse voltage is increased while the RESET and SET
resistance values are being measured. Figure 23 shows the fourpulse waveform for one point in the R-I curve. For the curve,
spot means are taken (green boxes) to capture the measurements
for the R-I curve.(Figure 24). Note that the RESET resistance is
traditionally displayed with the SET resistance, even though the
RESET pulse is not changing throughout the sweep, causing the
www.keithley.com
Figure 22. I-V curve from the PRAM IV_CURVE test generated and measured
by the Model 4225-PMU with Model 4225-RPMs. This test uses the
same data shown in Figure 21, but the curve plots I vs. V (user module
doubleSweepSeg).
RESET R to be a straight line. Plotting the RESET R does indicate
if the RESET pulse is adequately resetting the material after each
SET pulse. Figure 25 shows how the waveform measurements
map to the R-I results. Some test systems may use SMU
instruments for the resistance measurements, but this requires
additional switching and much longer test times. The PMU+RPM
combination measures the voltage and current simultaneously at
multiple points in the test, providing flexibility while ensuring
that the proper measurements are made.
PRAM I-V Curve
SET Resistance (ohms)
1.0E+6
Data:setR
Data:resetR
PRAM Endurance
1.0E+6
SET Resistance (ohms)
Figure 23. R-I RESET-Measure-SET-Measure waveform generated and
measured by the Model 4225-PMU with Model 4225-RPMs (user module
pramSweep). This graph shows the voltage (blue, left y-axis) and current
(red, right y‑axis) waveforms. There are four pulses within this waveform.
The first pulse is the RESET pulse, which resets the PRAM material, putting
it into a high resistance, amorphous state. The second pulse measures the
resistance of the RESET state. The third pulse is the SET pulse, which will
put the material into a low resistance, crystalline state. The fourth and last
pulse measures the resistance of the material in the SET state. Figure 24 is
an R-I curve that results from sweeping the SET pulse height and plotting the
results in the R-I curve.
The PRAM endurance test applies the RESET+SET waveform
(Figure 23) and measures the SET resistance every log n
iterations. Figure 26 shows an example of an endurance curve.
The routine also measures the RESET resistance, so it can also be
graphed if desired.
100.0E+3
10.0E+3
1.0E+3
1.0E+0
10.0E+0
100.0E+0
1.0E+3
10.0E+3
100.0E+3
Iterations
Figure 26. PRAM endurance curve generated and measured by the Model
4225-PMU with Model 4225-RPMs (user module pramEndurance).
10.0E+3
PRAM Test Modules
1.0E+3
0.0E+0 200.0E–6 400.0E–6 600.0E–6 800.0E–6
1.0E–3
1.2E–3
1.4E–3
1.6E–3
1.8E–3
2.0E–3
2.2E–3
SET Current (A)
Figure 24. R-I curve shows the resistance variation of the SET state resistance
generated and measured by the Model 4225-PMU with Model 4225-RPMs
(user module pramSweep). This test uses the data captured from the tops of
the pulses shown in Figure 23.
On subsequent waveforms,
sweep SET pulse height to
create R
RESET to amorphous
M
V
M
M
SET to
crystalline
M
Time
R
R/I “Bathtub” Curve
Reset R
Set R
ISET
Figure 25. Diagram explains how the various measurements are used to
define the R-I curve. The SET curve is swept while the RESET resistance
(red M), SET current (green M), and SET resistance (blue M) are measured.
There are three modules for testing of phase-change materials
or devices: doubleSweep, pramSweep, and pramEndurance,
using one Model 4225-PMU and two Model 4225-RPMs. The
doubleSweep module applies a single V-shaped pulse while
sampling the voltage and current.
Setting Up the Parameters in the doubleSweep
and doubleSweepSeg Modules
Table 6 lists the input parameters for the doubleSweep module,
which performs a transient I-V sweep by plotting the I-V samples
from one or two V-shaped pulses (Figure 27) using one Model
4225-PMU and two Model 4225-RPMs. This module is useful for
a variety of NVM technologies, whether unipolar (V1 and V2 +V)
or bipolar (V1 = +V, V2 = –V). For PRAM, only the first pulse is
used, so V2=0.
Two waveforms are returned from this test—the voltage
waveform on RPM1 and the current waveform from RPM2. (See
connection diagram, Figure 19.) Note that the V1 and V2 can
be of either polarity. The base voltage for the pulses is 0V. The
returned data is the voltage from RPM1 (V FORCE) and the current
from RPM2 (IMEAS). The data can be plotted vs. time (Figure 21,
www.keithley.com43
showing V1 only) or as I-V (Figure 22). The difference between
doubleSweep and doubleSweepSeg is that the doubleSweepSeg
module returns each up and down segment as a separate array,
which allows plotting each portion of the curve in a separate
color to help distinguish the contributions of each to hysteresis
or other relatively complicated I‑V results. An example of this
routine is shown in Figure 22, with the upsweep in green and
the downsweep in blue.
For FeRAM, both pulses are used (Figure 31). This module
is also useful for material characterization for ReRAM, CBRAM,
and other conductive bridge or ion transport technologies,
many based on transition metal oxide (TMO) materials. This test
returns data that can be used to create “butterfly” curves, which
is the standard technique for TMO ReRAM characterization.
Setting up the Parameters in
the pramSweep Module
Range
20ns to 33ms
–10V to +10V
–10V to +10V
Description
Transition time for the pulses
Voltage amplitude for first pulse
Voltage amplitude for second pulse
Current measure range for RPM1 (RPM1
100nA to 10mA
forces voltage)
Current measure range for RPM2 (RPM2
100nA to 10mA
measures current)
Delay
setWidth
setFallTime
setStartV
setStopV
Irange1
Irange2
iteration
setR_size,
resetR_size,
setV_size,
setI_size
pts
Range
20ns to 10ms
–10V to +10V
20ns to 1s
–10V to +10V
20ns to 1s
Description
Rise and fall time for the RESET pulse
Voltage for the RESET pulse
Pulse top width of the RESET pulse
Voltage for the measure V pulse
Pulse top width for the measure V pulse
Delay time between the pulses and rise/fall
20ns to 1s
time for measure pulses
20ns to 1s
Pulse top width for the SET pulse
20ns to 10ms
Fall time for the SET pulse
Start voltage for the SET pulse
–10V to +10V
amplitude sweep
Stop voltage for the SET pulse
–10V to +10V
amplitude sweep
Current measure range for RPM1 (RPM1
100 nA to 10 mA
forces voltage)
Current measure range for RPM2 (RPM2
100 nA to 10 mA
measures current)
Which waveform to capture from
1 to 10,000
the SET sweep
10 to 10,000
Number of steps in the SET sweep and the size
of the measurement arrays. All sizes must be
set to the same value.
10 to 10,000
Number of waveform points returned
½ * riseTime
Figure 27. The multi-pulse waveform of doubleSweep.
Figure 28. PRAM waveform definition.
SET R
measV
Time
Delay
Delay
Delay
Delay
setFallTime
Set
measWidth
setV
setWidth
measWidth
Delay
Delay
riseTime
Delay
Delay
RESET R
Reset
V2
resetWidth
Time
V2
Measurement
resetV
Delay
riseTime
riseTime
½ * riseTime
V1
riseTime
riseTime
V1
www.keithley.com
Parameter
riseTime
resetV
resetWidth
measV
measWidth
V
V
0
Table 7. Parameters in the pramSweep module.
riseTime
Irange2
Setting Up the Parameters in the
pramEndurance Module
SET V, I
Table 6. Parameters in the doubleSweep and doubleSweepSeg modules.
Irange1
To enhance the visibility of the changes in the resistances,
values >1MW are set equal to 1MW. This prevents the occasional
current measurements at the noise floor of the range from
causing a very large R value (5MW to hundreds of mega-ohms).
Note that the size parameters (setR_size, resetR_size, setV_size,
setI_size) for the pulse top measurements must all be the same
value (setR_size = resetR_size = setV_size = setI_size). Similarly,
for the sample waveform arrays, the sizes must be set to the same
value (VForce_size = IMeas_size = Time_size).
Table 8 lists the input parameters for the pramEndurance
module. This module stresses the PRAM cell by applying the
Table 7 lists the input parameters for the pramSweep module.
This module performs a SET voltage amplitude sweep from
setStartV to setStopV, with the number of steps set by the
parameter iteration. The waveform diagram is shown in
Figure 28. At each step in the sweep, several measurements
are extracted (green boxes in Figure 28) from the voltage
waveform on RPM1 and the current waveform from RPM2:
RESET resistance (measurement of cell resistance after the RESET
pulse), SET resistance (measurement of cell resistance after the
SET pulse), and Set V and Set I (current and voltage at the top
of the SET pulse). Measurements are extracted from 30% to 90%
of the pulse top (green boxes on Figure 28), which avoids the
current settling issue and provides a relatively wide window to
reduce the noise in the measurements. In addition to the scalar
measurements for each sweep (Reset R, Set V, Set I, Set R), this
Parameter
riseTime
V1
V2
routine also returns one current and voltage waveform from one
step of the sweep. The choice of which waveform is captured and
returned is set by the value in iteration.
Table 8. Parameters in the pramEndurance module.
Parameter
riseTime
resetV
resetWidth
measV
measWidth
delayT
setWidth
setFallTime
setV
Irange1
Irange2
max_loops
Range
20ns to 10ms
–10V to +10V
20ns to 1s
–10V to +10V
20ns to 1s
Description
Rise and fall time for the RESET pulse
Voltage height of the RESET pulse
Pulse top width of the RESET pulse
Voltage for the measure V pulse
Pulse top width for the measure V pulse
Delay time between the pulses and rise/fall
20ns to 1s
time for measure pulses
20ns to 1s
Pulse top width for the SET pulse
20ns to 10ms
Fall time for the SET pulse
–10V to +10V
Voltage height for the SET
Current measure range for RPM1 (RPM1
100nA to 10mA
forces voltage)
Current measure range for RPM2 (RPM2
100nA to 10mA
measures current)
Maximum
number of Reset+Set waveforms
1 to 1012
applied to the test device
setR_size,
resetR_size,
2 to 10,000
setI_size,
iteration_size
Size of arrays for returned measurements.
All sizes must be set to the same value.
RESET+SET waveform (Figure 28) max_loops times and
measuring the RESET and SET resistances. The number of stress/
measure cycles is determined by the value of iteration_size,
which occurs in a log10 spacing within max_loops.
FeRAM Material Testing
The NVM_Examples project has three tests for two-terminal
ferro-electric materials used in FeRAM (or FRAM) and
stand-alone (1C) devices (Figure 29) in the NVM_Examples
project (in the C:\sModel 4200\kiuser\Projects\_Memory
folder). The connection diagram is shown in Figure 30.
To obtain the best possible measurements, the connection
uses the source high, measure low method described in the
“Optimizing Measurements” section.
These tests require the following hardware:
• Model 4200-SCS with KTEI 8.2 or higher
• One Model 4225-PMU with two Model 4225-RPMs
FeRAM memory effect relies on charge storage in a capacitor
but uses a ferro-electric layer instead of the dielectric layer of a
typical capacitor. The memory mechanism for FeRAM is based
on polarization shift in ferro-electric materials [9, 10, 11, 12].
Ferro-electric materials have strong non-linear dependency
between applied electrical field (E) and polarization (P). When
the electric field reaches a critical level, the ions inside the
crystalline structure move from one stable location to another
one. This shift is accompanied by a shift of the ferro-electric
domain walls. Electrically, it is represented by the hysteresis
chart (Figure 32), showing the dependency between electrical
field and polarization. The switch between one state and another
is characterized by the area of hysteresis, which represents the
amount of charge moved during re-polarization.
The challenge of characterizing the ferro-electric capacitor is
that the fundamental behavior is switching of the polarization
Figure 29. Tests for the FeRAM device in the NVM_Examples project.
state of the ferro-electric material, which requires measuring
the polarization charge on the capacitor as it changes. Typically,
a load capacitor, pulse generator, and oscilloscope are used,
usually in a Sawyer-Tower circuit. In this approach, measuring
the transient voltage, using an oscilloscope or sampler, across
the load capacitor is a proxy for the charge flowing into the FE
material. However, this method has several drawbacks.
The load capacitance must be relatively large compared to
the FE capacitance to ensure that the voltage drop across the
load capacitor is not significant; otherwise, some somewhatunsatisfactory assumptions have to be applied to obtain
the voltage across the FE element. However, this large load
capacitance means that the sense voltage is fairly small.
This small voltage is difficult to measure accurately with an
oscilloscope or digitizer. The PMU+RPM solution does not
require the load capacitor method because it measures the
current and voltage directly and simultaneously so that the
total charge can be accurately determined. The charge is
calculated from the high speed current measurements, which are
sampled consistently over time, resulting in high speed charge
measurements. Note that the value reported is charge, not charge
per unit area. However, it’s easy to calculate the charge per unit
area from the provided charge values by using the Formulator in
the UTM definition tab.
Much like PCRAM and other NVM materials, FeRAM cells
require a forming step or process before the cell exhibits
reproducible switching behavior. The PUND or FERAM
endurance tests can be used to apply the forming pulses. The
PUND test is appropriate when a small number of pulses are
required for forming, whereas the endurance test is better when
a larger number of pulses are required. The hysteresis curve can
indicate if the test device was sufficiently formed (see the gap at
0V in Figure 32).
The hysteresis test applies positive and negative V-shaped
pulses to measure the current. Because the current is sampled
continuously, calculating the total charge is straightforward.
Figure 31 shows the test signal applied to the FE capacitor, as
measured by the PMU+RPM. Figure 32 shows the hysteresis
curve, which was extracted from the data in Figure 31. Proper
test parameters on a good device should show a complete loop,
with the beginning and end at 0V.
The characteristic shape of the hysteresis curve is not only a
measure of the inherent ferro-electric material performance but
may also show degradation due to the semiconductor processing
that occurs after the FE capacitor fabrication.
www.keithley.com45
SMU Sense
4225-RPM #1
Inputs
CVU Pot
CVU Cur
Outputs
SMU Force
Force
Sense
RPM cable
NOTE: Connect outer
shields together
4225-RPM #2
SMU Sense
RPM2
CVU Pot
CVU Cur
Outputs
RPM1
Inputs
SMU Force
4225-PMU
C FE
Force
Sense
RPM cable
Figure 30. Connection for FeRAM tests. Note that the FeRAM tests included
with NVM_Examples do not use the SMUs or CVU.
Figure 32. Hysteresis curve generated and measured by the Model 4225-PMU
with Model 4225-RPMs (user module doubleSweepSeg). This graph shows
the variation in the polarity charge as the voltage across the material varies.
Ec is the coercive field and Pr is the remnant polarization, which are key
parameters for FeRAM performance. Proper test parameters on a good, nonleaky device should show a complete loop, with the beginning and end at 0V.
Figure 31. Hysteresis waveform generated and measured by the Model 4225PMU with Model 4225-RPMs (user module doubleSweepSeg). This graph
shows the voltage waveform (blue) applied to the FE capacitor. The red curve
shows the current flow. Figure 32 shows the typical hysteresis charge vs.
voltage curve.
The PUND test characterizes the polarity change in the
ferro-electric material. It is called PUND because four pulses are
applied sequentially: Positive, Up, Negative, Down (Figure 33).
Note the change in the shape of the current between the two up
pulses (red P and red U) and similarly for the two down pulses
(red N and red D). The first pulse requires additional charge/
current to change the polarization (red P or red N), compared to
the second pulse, which has the capacitive charging only (red U
or red D). The difference between them (P-U or N-D) represents
the polarization charge or memory effect. PSW is the polarization
change during the up pulses (PSW = red P – red U). QSW is an
average of the two charges from the two polarities (two up and
two down pulses, QSW = ((red P – red U) + (red N – red D)) / 2).
In addition to fundamental device and material characterization,
the PUND test is used to determine the proper voltages and
timing for the endurance test.
The endurance test shows the reduction in the polarization
charge as the number of pulses applied increases. For some
devices, the polarization charge may increase slightly during the
initial pulse stress, before declining with an even larger number
www.keithley.com
Figure 33. PUND waveforms generated and measured by the Model 4225PMU with Model 4225-RPMs (user module pundTest). This graph shows the
applied voltage pulses and the current response. The values for P, U, N, and
D are extracted from the current waveform.
of stresses [13]. Figure 34 shows example endurance results
for QSW and PSW. The degradation in QSW begins at 11 million
cycles. The onset of degradation is a strong function of the pulse
amplitude, so different materials and PUND voltages can provide
significantly different degradation curves. Figure 35 shows the
change in the charges for P, U, N, and D, which determines the
data shown in Figure 34.
FeRAM Test Modules
There are three modules for testing ferro-electric materials or
devices: doubleSweep, pundTest, and pundEndurance.
RPM2), and time arrays. In addition, the charge for each pulse
transition is returned. Figure 36 shows the pulse parameter
definitions.
FeRAM Endurance: PSW, QSW
460.0E–12
PSW and QSW (charge)
450.0E–12
Table 9. Input parameters in the pundTest module.
440.0E–12
430.0E–12
Parameter
420.0E–12
Vp
410.0E–12
400.0E–12
390.0E–12
Data: PSW
Data: QSW
380.0E–12
370.0E–12
360.0E–12
1.0E+0
10.0E+0
100.0E+0
1.0E+3
10.0E+3
100.0E+3
1.0E+6
10.0E+6
100.0E+6
Iterations
Figure 34. FeRAM endurance, showing the degradation in P SW and QSW (user
module pundEndurance). Some FeRAM endurance curves only show QSW.
This data was taken by the Model 4225-PMU with Model 4225-RPMs.
Figure 35. FeRAM endurance, showing the degradation in P, U, N, and D
values (user module pundEndurance). This data is used to create the data
in Figure 34. This data was taken by the Model 4225-PMU with Model
4225-RPMs.
Setting Up the Parameters in the
doubleSweep Module
This module is the same module used in the PRAM test IV_Curve.
However, the PRAM test used just one of the V-shaped pulses,
while the FE capacitor hysteresis test uses two pulses (Figure
27 shows the definition, Figures 31 and 32 show test results
on FeRAM).
Table 6 lists the input parameters for the doubleSweep module.
Setting Up the Parameters in
the pundTest Module
This module applies the four pulses that make up the PUND test,
with the parameters briefly explained in Tables 9 and 10. This
test uses both Model 4225-RPMs connected to the first Model
4225-PMU in the Model 4200-SCS chassis (Figure 30). RPM1
outputs the voltage pulses and measures the applied voltage.
RPM2 measures the current flowing through the test device. The
returned values include the voltage (from RPM1), current (from
tp
td
trf
Irange1
Irange2
Range
Description
Voltage level for the four PUND pulses. The first
–10V to +10V and second pulses are +Vp and the third and
fourth pulses are at –Vp (Figure 36)
20ns to 1s
Width of each pulse top
20ns to 1s
Delay time between each pulse
20ns to 10ms Pulse transition time for all pulses
Current measure range for RPM1
100nA to 10mA
(RPM1 forces voltage)
Current measure range for RPM2
100nA to 10mA
(RPM2 measures current)
Table 10. Output parameters in the pundTest module.
Parameter
Type
V
Double Array
I
Double Array
t
P
Pa
U
Ua
N
Na
D
Da
Psw
Qsw
pts
Double Array
Double
Double
Double
Double
Double
Double
Double
Double
Double
Double
Double
Description
Array of voltage measurements for the
PUND waveform
Array of current measurements for the
PUND waveform
Array of timestamps for the PUND waveform
Total charge on rise of pulse 1
Total charge on fall of pulse 1
Total charge on rise of pulse 2
Total charge on fall of pulse 2
Total charge on rise of pulse 3
Total charge on fall of pulse 3
Total charge on rise of pulse 4
Total charge on fall of pulse 4
Polarization charge, defined as P – U
Average polarization charge: ((p-u)+(n-d))/2
Total number of points recorded
Setting Up the Parameters in the
pundEndurance Module
This module is a stress/measure test that applies the four pulses
that make up the PUND test max_loops times to the test device,
with the parameters briefly explained in Tables 11 and 12.
While applying the stress PUND waveform max_loops times,
the number of measurement intervals is set by fatigue_count.
The number of stress waveforms between each measurement is
calculated on a log10 basis. The returned parameters are similar
to pundTest and the pulse parameter definitions are shown in
Figure 36.
ReRAM and CBRAM Testing
As mentioned previously, ReRAM and CBRAM are types of Redox
memory [14]. Both ReRAM and CBRAM are typically tested in the
DC realm using SMU instruments, but SMU instruments may not
be the best instrument for testing some memory types.
In traditional ReRAM test setups, to create, or form, the low
resistance state initially, SMU instrument current compliance
is used to limit the maximum current flowing through the test
device during the forming or reset operation. The desire is to
limit the amount of current to reduce the stress on the cell and
www.keithley.com47
tp
+VP
P
td
trf
U
trf td
Time
N
-VP
D
Figure 36. Pulse waveform parameters for pundTest.
Table 11. Input parameters in the pundEndurance module.
Parameter
Range
Description
Vp
–10V to +10V
Voltage level for the four PUND pulses.
The first and second pulses are +Vp and
the third and fourth pulses are at – Vp
Vfat
–10V to +10V
Voltage pulse level for the two pulses of
the fatigue waveform. First pulse is +(Vp),
last pulse is – (Vp)
Tp
20ns to 1s
Width of each pulse top
td
20ns to 1s
Delay time between each pulse.
Trf
20ns to 10ms
Pulse transition time for all pulses
Irange1
100nA to 10mA
Current measure range for RPM1
(RPM1 forces voltage)
Irange2
100nA to 10mA
Current measure range for RPM2
(RPM2 measures current)
pts_ per_waveform
10 to 10,000
max_loops
1 to 1012
Total number of stress waveforms applied
to the test device
fatigue_count
2 to 100
Number of measurement intervals during
max_loops.
Number of measure points per PUND
measure waveform
Table 12. Output parameters in the pundEndurance module.
Parameter
Iteration
Type
Description
Integer: Number of measurement intervals, must be same or
1 to 100 greater than fatigue_count
P
Double Total charge on rise of the P pulse
Pa
Double Total charge on fall of the P pulse
U
Double Total charge on rise of the U pulse
Ua
Double Total charge on fall of the U pulse
N
Double Total charge on rise of the N pulse
Na
Double Total charge on fall of the N pulse
D
Double Total charge on rise of the D pulse
Da
Double Total charge on fall of the D pulse
Psw
Double Polarization charge, defined as P – U
Qsw
Double Average polarization charge: ((p-u)+(n-d))/2
also improve the quality of the switching process. However, the
compliance circuit in an SMU instrument is not instantaneous,
and takes microseconds to milliseconds to engage fully. Before
the circuit is active, the amount of current flowing is not fully
known or controlled. The actual transient response of the
current compliance and the detailed interplay between the
changing test device impedance and the reaction of the SMU
instrument is complicated and not well understood.
Pulse I-V characterization improves the situation by providing
strict timing control of the voltage signal applied to the test
device. Tests in the ReRAM section of the NVM_Examples project
can be used for characterization of ReRAM and CBRAM.
In the project, ReRAM device results are shown. ReRAM
memory devices are two terminal devices, with a “low” and
“high” side. A voltage pulse is applied across the oxide, and
according to the mainstream literature, this creates conductive
filaments. The process for creation of filaments is called
“forming” and is considered the most important aspect that
determines the ReRAM switching behavior. Forming is usually
performed with DC voltage sweep with current limit enabled.
The following process takes place, as generally accepted. As
voltage ramps up, the electrical field grows. When it becomes
sufficient, a conductive filament begins to grow through the
material from one electrode to the opposite electrode. As
soon as it happens, the current limit engages and arrests the
further growth of the conductive filament section. Resistance
of the device drops from high resistance to low, for example,
from several mega-ohms to the kilo-ohm range. This plausible
explanation can bring into question that the speed of currentlimit engagement may be important for ReRAM forming and
therefore worthy of additional characterization. In the industry
it still not common knowledge that the speed of a DC (SMU
instrument) current limit is relatively slow, in the ~100µs range.
An ideal solution to this problem is to design structures with
current-limiting control transistors directly on-wafer. This will
ensure minimum parasitic capacitance and minimum response
time. If, however, a current limit is not available on the wafer, it
is suggested to use both SMU instrument current limiting and
Model 4225-RPM current limiting to better understand forming
mechanisms.
After forming, the applied RESET pulse changes resistance
of the structure from low resistance to high. It is thought to
be caused by the destruction of the conductive filament in the
vicinity of the one of the electrodes. ReRAM structures can be
symmetrical or polar. Polar devices require certain polarity of
the forming pulse, and have non-symmetric material layouts.
For polar devices, the Forming (and Set Pulse) have an opposite
polarity from the Reset pulse.
Figure 37 shows the tests used for ReRAM testing. The first
one is a standard characterization test, where a ReRAM device is
tested and verified that it is valid and can be used.
www.keithley.com
The remaining tests include a Forming and a Set/Reset
sequence, that is usually called the “Butterfly Curve.” The last
test is an Endurance test. Figure 38 shows the connection
diagram for a two terminal ReRAM device, showing both the
pulse and SMU instrument connections. Note that if there is
a chuck or backside connection, please connect RPM #1 to
this terminal.
In Figure 39, the blue curve, using the left y-axis, shows
transient voltage applied to the ReRAM device. The red curve,
using the right y-axis, is the current response. Note that the
x-axis is time, in hundreds of microseconds. The voltage
amplitude is ±2.3V. The same data is also plotted as I vs V as
shown on the right. Exponential dependency of the current on
the bias demonstrates that the device is not yet formed and can
be used for characterization.
This data was collected with a PMU and two RPMs, using
test routine, reramSweep; it can also be configured to use
SMU instruments to collect similar data. There are several
main differences between tests with PMU with RPMs and
SMU instruments. First, the SMU instrument sweep takes
significant time, up to several seconds, while the PMU+RPM
sweep can be as fast as hundreds of nanoseconds. Second, the
current limit capability for an RPM is much faster than an SMU
instrument. An SMU instrument current compliance “engage
time” is approximately 100–500µs, depending on the particular
configuration and SMU instrument model, while an RPM limit
can be as fast as several hundred nanoseconds, effectively
~3 orders of magnitude faster. While a control transistor,
located on-wafer, would be faster and better in terms of energy
control, the RPM current limit does provide research benefits
4210-CVU HCURR
HPOT
LPOT
LCURR
RPM cable
Cur
Force
Sense
R
Sense
Triax cables
Force
Pot
RPM1
Cur
RPM2
4225-RPM #2
RPM cable
Figure 38. Connection to the 2 terminal ReRAM device.
Outputs
SENSE
FORCE
4225-RPM #1
Outputs
Force
Pot
4200-SMU 2
4225-PMU
Sense
Inputs
Triax cables
Inputs
SENSE
FORCE
Force
Sense
A detailed explanation of
Capacitive Charging Effects during Pulse Transitions is
given below.
ReRAM forming results are shown in the two graphs of
Figure 40. The graph to the left is voltage and current transients
(V and I vs. time), and the graph to the right is the same data
plotted as current vs voltage. Current exponentially grows as
voltage reaches a certain value. When the filament is established,
the device switches into low resistance state and the current limit
prevents further growth of the filament. Forming was performed
using a PMU+RPM with current limit enabled. A standard
test configuration, used by most of the researchers in ReRAM
field is to use DC instruments, SMU instruments, with current
compliance. Forming performed with a PMU+RPM, unlike the
the SMU instrument approach, is able to measure current and
power transients during forming, while forming done with
an SMU instrument completely hides details of the process.
According to the literature, the forming process is controlled
mostly by the value of the current limit and the speed of its
activation. The right graphs shows the same data, but with the
current plotted vs the voltage. The added arrows show the time
progression of the curve, to allow for comparison to the timebased graph on the left.
The same test routine that was used for Characterization and
Forming was applied to an already formed device to collect the
“Butterfly” curve, as shown in Figure 41. The chart to the left, as
before, shows the current & voltage transient (I and V vs. time).
The chart to the right is current vs. voltage in the shape of butterfly
wings. The same routine, reramSweep [use fixed-width font], with
different test conditions, can obtain both DC (SMU instrument)
data and Pulse (PMU) data. This allows convenient switching
between DC and PULSE modes of ReRAM characterization.
Figure 37: Screenshot of the tests for the ReRAM device in the NVM_
Examples project.
4200-SMU 1
compared to the SMU instrument testing. Using the PMU+RPM
provides information that may help during the complicated
and potentially costly transition to adding an on-wafer current
limit capability. Another benefit to the pulse approach is using a
second PMU+RPM channel to avoid capacitive charging effects by
measuring the current through the test device on the low or side
opposite the pulsing (Figure 50). The faster current limit is also
useful during constant voltage stressing (CVS). Traditionally, CVS
was performed using SMU instruments, but it can also be done
using the PMU+RPM approach.
The left wing of the butterfly curve is a RESET transition,
when resistance switches from low resistive state to high resistive
state; note that the current limit is not enabled. Voltage polarity
is opposite to the polarity used for forming and SET. During
RESET process, it is commonly assumed, that conductive filament
loses continuity from one electrode to another one, but this
affects a small section of the filament. The SET process is very
similar to the Forming process, in that it uses pulses of the same
polarity to re-establish continuity of the conductive filament
from one electrode to another one. Since the SET bias does not
have to grow the whole filament, just reestablish the connection;
www.keithley.com49
Figure 39. Characterization test of ReRAM structure.
Figure 40. ReRAM forming.
www.keithley.com
Figure 41. ReRAM “Butterfly” curve.
it requires less bias than the Forming process does. The SET
procedure, similar to Forming, requires the use of the current
limit capability but at a lower current level.
Selection of the test parameters for ReRAM testing, including
Forming, SET, and RESET, follows a logical progression for initial
forming and characterization.
1.First, prior to any forming, the test device is verified to be
good and that it demonstrates a non-linear I-V dependency
(Figure 39). During this initial pre-screening, care should
be taken to use a voltage lower than the forming, SET, and
RESET voltages.
2.Forming is a critical process (Figure 40). The value of the
maximum bias should be selected to trigger filament growth.
At the same time the current limit should be small enough,
thereby preventing the filament from becoming too stable.
If a conductive filament is too stable, than no amount of
the Reset bias will disconnect the filament and no sufficient
electrical field in the filament can be achieved. The value of
the voltage bias for the RESET pulse should be just enough to
break the filament.
3.Current limit for the SET bias, similar to Forming, should be
just large enough to reconnect the filament, but not too large to
prevent a “disconnect” in the next SET/RESET cycle. Selection
of the Forming/SET and RESET parameters are somewhat tricky,
require sufficient investigation, and in a critical way control the
endurance of the ReRAM device. The endurance is defined as
maximum number of SET/RESET cycles, where the LOW and
HIGH resistance states are sufficiently different.
Figure 42 gives an example of the Endurance test of a ReRAM
structure. The endurance test is similar to endurance tests for
PRAM, Flash, and FeRAM. The control parameter, which in case
of resistive memory, is resistance, and has two states, low and
high. Low resistance corresponds to the state when conductive
filament is formed and connected opposite electrodes across
the oxide. High resistance is the state where a small section
of the filament is converted back to high resistive material.
Selection of the Forming, SET, and RESET processes defines
how many cycles it will take for the low resistance state to
become indistinguishable from the higher resistance state, which
is one way to determine the maximum number of cycles for
device endurance.
Setting up the Parameters in
the reramSweep Module
The reramSweep module is used to perform a double sweep with
a flat section at the peak of each sweep. To test a ReRAM device,
the user chooses appropriate values for the two peaks, either
positive or negative, and then sets the timing (Figure 43, Table
12). Either SMU instruments or a PMU with 2 RPMs can be used,
depending on the setting of useSmu parameter. The Low side
(Figure 38)of the device needs to be connected to PMU channel
1 using a RPM. “Low” side is defined as the side connected to the
bulk, substrate, or chuck if there is connection between DUT and
the substrate. SMU1 is also connected to channel 1, RPM1. SMU2
is connected to the channel 2 RPM. The voltage bias is applied
to channel 1, channel 2 is kept at virtual ground potential and
current is measured on channel 2.
www.keithley.com51
Table 13. List of Input Parameters in the reramSweep module (see Figure 43)
Parameter
Figure 42: ReRAM endurance test.
Setting up the Parameters in the
reramEndurance Module
The reramEndurance routine is used to perform a series of
double sweeps (SET+RESET) using the same parameters used
for the single sweeps as described in the reramSweep routine.
To test a ReRAM device, choose appropriate values for the two
peaks, either positive or negative, and then set the timing you
would like to implement. Choose the number of times you
would like to run the double sweep (max _ loops) and how
often you want to take measurements (fatigue _ count). The
routine will use a log10 approach to spacing the total number
of measurements (fatigue _ count) across the total pulses
(max _ loops). For example, if max _ loops = 10000 and
fatigue _ count = 4, then there will be RSET and R RESET
measurements after 10, 1000, 1000, and 10000 SET+RESET
pulses applied to the test device.
RPM current limit calibration
The pulse current limiting capability is provided by the Model
4225-RPM. The short distance of the RPM to the DUT minimizes
parasitic capacitance and time it takes for the current limit to
activate and become effective. As mentioned above, this time is
delayTime
riseTime
delayTime
riseTime
resTestV
SMU measures
RSET
-V
widthTime
resetV
SET
resTestV
delayTime
RESET
setV
SMU measures
RSET
widthTime
riseTime
delayTime
0
riseTime
V
Figure 43. Pulse waveform with measure for reramSweep and
reramEndurance.
www.keithley.com
Range
Description
The time it takes for voltage to ramp to the SET
or RESET voltage. This value is slew rate-limited.
riseTime
2e-8s to 33 ms
To reach the slowest rise time, a higher voltage
must be used.
The time to wait at the top of the pulse at
widthTime
2e-8s to 1s
full voltage
delayTime
20 ns to 1 s The time between the SET and RESET pulses
On which SMU instrument channel to enforce
compliance
1 or 2
current compliance
The peak voltage of the reset pulse. For ReRAM
resetV
–20V to 20V
devices, this value should be negative.
The peak voltage of the set pulse. For ReRAM
setV
–20V to 20V
devices, this value should be positive.
100nA to 10mA
The range at which to measure current. The
for RPM
SMU instrument has more ranges and also
Irange
10 pA to 1A
accept 0 for autorange.
for SMU
Please note that this variable is used both for
PMU and SMU instrument control during RESET.
If it is set to 0, then no current limit is applied.
–100mA to
If useSmu = 1, and sweep is performed with
resetIcomp
+100mA
the SMU instrument, and current limit is 0,
autorange is used. With current limit not zero
and useSmu = 1, the SMU instrument is set in
the fixed current range.
Please note that this variable is used both for
PMU and SMU instrument control during SET.
If it is set to 0, then no current limit is applied.
If useSmu = 1, and sweep is performed with
–100mA to
setIcomp
+100mA
the SMU instrument, and current limit is 0,
autorange is used. With current limit not zero
and useSmu = 1, the SMU instrument is set in
the fixed current range.
SMU instrument voltage used to measure device
resTestV
–10V to +10V
resistance.
Whether or not to take resistance measurements.
1 means yes, take resistance measurements,
takeRmeas
0 or 1
while 0 means no, do not take resistance
measurements.
Whether or not to use an SMU instrument to
take DC measurements instead of using the
useSmu
0 or 1
PMU+RPM for pulsing measurements. A 1 means
yes, use the SMU instrument and a 0 means no,
don’t use the SMU instrument, use the PMU.
numIter
1 to 100
This parameter should be set to 1.
These three values should be the same
Vforce_size
Imeas_size
10 to 10000 and represent the number of items in the
Time_size
output arrays.
Table 14. List of Output Parameters in the reramSweep module.
Parameter
Range
Description
Vforce
Array of forced voltages
Imeas
Array of measured currents
Time
Array of measured times
resetResistance
Resistance of DUT after the reset pulse
setResistance
Resistance of DUT after the set pulse
initResistance
Resistance of DUT before any pulse
Time
~3 orders of magnitude smaller than the current compliance
activation time for a typical DC SMU instrument. The RPM
current limit capability requires initial and periodic calibration.
Before using the current limit feature of the Model 4225-RPM,
calibration has to be performed manually by calling routine
Table 15. List of In put Parameters in the reramEndurance module (see
Figure 43 and Figure 44).
V
delayTime
widthTime
delayTime
riseTime
SET
widthTime
resetV
riseTime
setV
riseTime
RESET
delayTime
riseTime
widthTime
riseTime
delayTime
0
Parameter
Time
-V
delayTime
useSmu
compliance
resetV
setV
Figure 44. Pulse Waveform parameters for the stress reramEndurance.
Do_RPM_Ilimit_Cal (from user library RMP_ILimit_Control). This
test is included in the NVM_Examples project in the Initialization
Steps section at the top of the project tree.
Irange
resetIcomp
Note that the Model 4225-RPM-LR does not have the current
limit feature. Attempting to perform a current limit calibration
on or set the current limit of a Model 4225-RPM-LR module will
cause an error.
setIcomp
resTestV
Figure 45. RPM Current limit UTMs in the NVM_Examples project tree.
max_loops
fatigue_count
Range
Description
The time it takes for voltage to ramp to the
4e-8 to 1e-2s
final value
The time to wait at the top of the pulse at
4e-8 to 1e-2s
full voltage
The time between the two pulses and between
4e-8 to 1e-2s
the sweeps
0 or 1
Should stay zero. Used for debugging purposes
On which SMU instrument channel to enforce
CH 1 or 2
current compliance
The peak voltage of the reset pulse. For ReRAM
–20 to 20V
devices, this value should be negative
The peak voltage of the set pulse. For ReRAM
–20 to 20V
devices, this value should be positive
0 to 0.2
The range at which to measure current
Please note that this variable is used both
for PMU and SMU instrument control during
RESET. If it is set to 0, then no current limit is
–100mA to applied. If useSmu = 1, and sweep is performed
+100mA
with the SMU instrument, and current limit is
0, autorange is used. With current limit not zero
and useSmu = 1, the SMU instrument is set in
the fixed current range.
Please note that this variable is used both for
PMU and SMU instrument control during SET.
If it is set to 0, then no current limit is applied.
–100mA to If useSmu = 1, and sweep is performed with
+100mA
the SMU instrument, and current limit is 0,
autorange is used. With current limit not zero
and useSmu = 1, the SMU instrument is set in
the fixed current range.
The voltage at which to measure the resistance
of the DUT. This should be much less than
0.1 to 2V
the set and reset voltages as to not set or
reset the DUT
Number of pulses to output and stress the test
device. Set the fatigue_count to determine the
1 to 1012
number of measure intervals within the max_
loops number of pulse waveforms.
Number of times to measure during the max_
loops tests, using a log interval. For example, if
max_loops = 1000 and fatigue_count = 4, there
2 to 100
will be measurements of the SET and RESET
resistance after 1, 10, 100, 1000 SET+RESET
pulse waveforms are applied to the test device.
Table 16. List of Output Parameters in the reramSweep module.
Parameter
Figure 46. Settings for current limit calibration of the PMU channel 1 RPM.
reset
set
init
Range
Resistance
Resistance
Resistance
Description
Resistance of DUT after the reset pulse
Resistance of DUT after the set pulse
Resistance of DUT before any pulse
SMU1 for parameter smuname. The quality of the calibration is
determined by correlation factors called PosRsq and NegRsq.
Figure 47. Example good results for RPM Current limit calibration, as shown
in the Sheet.
Here, two calls are shown to calibrate both RPMs (Figure
45). To execute a calibratation procedure for channel 1 of
PMU1, set up input conditions to the routine as in Figure 46,
and connect SMU2 to the force output connector of the RPM1.
To calibrate the channel 2 RPM, use the RPM2 test (Figure 45)
and connect SMU1 to the force output connector RPM2 and use
Calibration is considered successful if both these numbers are
close to 1.000, with at least the first 4 or 5 digits = 9 (Figure 47,
showing 999.995E-3 or 999.999E-3). The calibration will drift due
to temperature, so for better performance, daily calibration is
recommended. The current limit range is from ~10µA to 10mA.
www.keithley.com53
Status and Error Codes for the nvm User Library
When debugging an error, start with the error code below (Table
13) and also look at the debug log file (C:\nvmlog.txt).
Optimizing Measurements
Because the Models 4225-PMU and 4225-RPM provide waveforms
for each of the pulse waveform tests, determining the pulse
fidelity is greatly simplified compared to source-only pulse
systems. During initial setup, use the pulse waveforms to
confirm proper voltage levels and timing response in addition to
the extracted values (V T, SET R, Hysteresis, PUND).
When measuring, there is always a tradeoff between the size
of the measure window vs. the noise floor. A longer measure
time will provide a less noisy result. The models 4225-PMU and
4225-RPM also follow this fundamental fact of measurement.
However, because timing is a primary parameter, the tradeoff is
more visible and therefore more controllable.
In addition to noise, note that lower current measure ranges
have slower response than higher ranges. This is true for SMU
current ranges, as well as Model 4225-PMU and Model 4225-RPM
ranges. For example, the Model 4225-RPM current measurement
settles much faster on the 10mA range (100ns best case) than the
100μA range (750ns best case). This means that sometimes using
a higher current measure range will permit shorter pulse timing
parameters because of the faster settling time but at the expense
of a bit more noise.
Table 13. Error codes and descriptions for the nvm user library.
Code Description
1 Test ran successfully
–10 Could not initialize NVM structure
–20
Return arrays are not the same size, or iter_size is smaller than
fatigue_count
–40
Current range on one of the channels is too large, should be 10mA
(10e-2) or less
–50 Error in pg2_init
–60 Error in pulse_load
–70 Error in pulse_ranges
–80 Error in pulse_burst_count
–90 Error in pulse_output
–100 Error in pulse_standby
–110 Error in pulse_sample_rate
–120 Error in Set_RPM_ICompliance
–130 Error in seq_arb_sequence
–140 Error in pulse_exec
–150 Error in pulse_ fetch
–160 No points returned
–170 Error in seg_arb_waveform
–210 Error completing the test
–220 Error during measurement
www.keithley.com
Capacitive Charging Effects
During Pulse Transitions
Figure 48 shows a simplified block diagram of a configuration
used to create the waveforms shown in Figure 49. During pulse
transitions, current flows to charge and discharge capacitance
in the test system. This current is not flowing through the DUT
but does flow out from the instrument and into the cabling, as
shown in Figure 48. This charging current is measured by the
PMU+RPM. Figure 49 shows the capacitive charging effect using
the pulse waveforms. The equation for this effect is:
I = C dV / dt
where:
I is the capacitive charging current.
C is the capacitance of the system (that is, the capacitance “seen”
by the measurement circuit).
dV is the change in voltage.
dt is the change in time, in this case, the pulse transition or
pulse rise and fall time.
This effect is fundamental and not implementation-specific,
meaning that all instruments that measure high speed current
will measure this effect. Note that measurements during the
pulse top are not affected, which means that pulse I-V results are
not affected by capacitive charging/discharging. This capacitive
charging effect has not been previously seen because nearly all
waveform measurements were voltage only.
Based on the equation, the effect is caused by the dV/dt and
the C. Generally, the capacitance of the cable is larger than
the DUT or instrument capacitance. The capacitance can be
minimized by using shorter cable, but there will always be some
capacitance in the instrument and in the interconnect. The dV/dt
can be reduced by either reducing the voltage or increasing the
rise and fall times. Reducing the voltage is usually not an option,
as it is the desired signal level for characterization. Reducing the
time may be possible in certain situations, but the time is a key
test parameter for transient characterization.
The dV/dt provides the hint to a solution to the capacitive
charging effect. Measuring the current on the other side of the
DUT, away from the applied pulse, is the key to avoiding the
effect. Figure 50 shows the connection setup. Note that the side
of the device connected to Channel 2 has a dV/dt that is essentially 0 for most cases. The results of the two-channel setup are
shown in Figure 51. Note that there is no charging or discharging effect seen during the pulse transitions. This approach is
sometimes called pulsing on the high side (red curve in Figure
51) and measuring the current on the low side. Note that high
and low are referring to the relative voltages across the device.
Test Equipment
Minimum configuration:
• Model 4200-SCS with KTEI 9.0 or higher.
Pulse I-V Instrument
A
Pulse I-V Channel 1
Coax Cable
A
V
V
DUT
Pulse I-V Channel 2
A
V
Figure 48. Block diagram of pulse I-V instrument showing the capacitive
charging current during pulse transitions.
dV/dt ≈ 0
Figure 50. Block diagram of two pulse I-V channels connected to a twoterminal device.
2.0E+0
1.8E+0
0.0E+0
Channel 1 Voltage
1.6E+0
–200.0E–6
–400.0E–6
1.2E+0
1.0E+0
–600.0E–6
800.0E–3
–800.0E–6
600.0E–3
–1.0E–3
400.0E–3
–1.2E–3
Channel 2 Current
200.0E–3
–1.4E–3
0.0E–3
–1.6E–3
• Two SMU instruments, either medium-power (Model 4200SMU) or high-power (Model 4210-SMU). Note that the SMU
instruments are used in the Flash and ReRAM tests only.
• One Model 4225-PMU with two Model 4225-RPMs
• Optional: Compiler to modify the nvm user modules (Order
Number: 4200-Compiler)
References
1. J. Hutchby and M. Garner, “Assessment of the Potential & Maturity
of Selected Emerging Research Memory Technologies Workshop &
ERD/ERM Working Group Meeting (April 6-7, 2010),” International
Technology Roadmap for Semiconductors, July 23, 2010. [Online].
Available: http://www.itrs.net/links/2010itrs/2010Update/ToPost/ERD_
ERM_2010FINALReportMemoryAssessment_ITRS.pdf. [Accessed: September
26, 2011].
2. E. Harari, “The Non-Volatile Memory Industry – A Personal Journey,”
presented at the 3rd IEEE International Memory Workshop (IMW),
Monterey, CA, 2011, pp. 1-4.
3. H. Yoo, E. Choi, H. Joo, G. Cho, S. Park, S. Aritome, S. Lee, S. Hong, “New
Read Scheme of Variable Vpass-Read for Dual Control Gate with Surrounding
Floating Gate (DC-SF) NAND Flash Cell,” presented at the 3rd IEEE
International Memory Workshop (IMW), Monterey, CA, 2011, pp. 53-56.
4. M. Seo, B. Lee, S. Park, T. Endoh, “A Novel 3-D Vertical FG NAND Flash
Memory Cell Arrays Using the Separated Sidewall Control Gate (S-SCG) for
Highly Reliable MLC Operation,” presented at the 3rd IEEE International
Memory Workshop (IMW), Monterey, CA, 2011, pp. 61-64.
5. “IBM scientists demonstrate computer memory breakthrough,” June 30,
2011. [Online]. Available: http://www.zurich.ibm.com/news/11/pcm.html.
[Accessed: September 26, 2011].
6. T. Nirschl, J.B. Phipp, T.D. Happ, G.W. Burr, B. Rajendran, M.-H. Lee, A.
Schrott, M. Yang, M. Breitwisch, C.-F. Chen, E. Joseph, M. Lamorey, R. Cheek,
S.-H. Chen, S. Zaidi, S. Raoux, Y.C. Chen, Y. Zhu, R. Bergmann, H.-L. Lung,
1.6E–6
1.5E–6
1.4E–6
1.3E–6
1.2E–6
1.1E–6
1.0E–6
900.0E–9
700.0E–9
800.0E–9
500.0E–9
600.0E–9
300.0E–9
400.0E–9
100.0E–9
200.0E–9
–200.0E–3
Figure 49. Capacitive charging effect. The blue curve is the voltage pulse
applied to a short length of coax cable. The red curve is the resulting current
flow, measured from the same side as the applied voltage pulse (Figure 48,
or Channel 1 in Figure 50). The charging effects are the humps in the red
curve during the voltge pulse transition.
Current
Voltage
1.4E+0
Time
Figure 51. Pulse I-V waveforms for the configuration shown in Figure 50. No
capacitive charging current is seen during the pulse transitions, only the
current flowing through the resistor DUT.
C. Lam, “Write Strategies for 2 and 4-bit Multi-Level Phase-Change Memory,”
IEDM 2007. pp. 461-464.
7. A. Pirovano, A. L. Lacaita, F. Pellizzer, S. A. Kostylev, A. Benvenuti, and R.
Bez, “Low-Field Amorphous State Resistance and Threshold Voltage Drift in
Chalcogenide Materials,” IEEE Trans. on Electron. Devices, vol. 51, May 2004,
pp. 714-719.
8. D.S. Suh, K.H.P. Kim, J.S. Noh, W.C. Shin, Y.S. Kang, C. Kim, Y. Khang, and
I.K. Yoo, “Critical Quenching Speed Determining Phase of Ge2Sb2Te5 in
Phase-Change Memory,” IEDM 2006, pp. 1-4.
9. G. R. Fox, R. Bailey, W. B. Kraus, F. Chu, S. Sun, and T. Davenport, “The
Current Status of FeRAM,” Topics in Applied Physics, vol. 93, 2004,
pp. 139-148.
10.L. Hai, M. Takahashi, S. Sakai, “Downsizing of Ferroelectric-Gate FieldEffect-Transistors for Ferroelectric-NAND Flash Memory Cells,” presented at
the 3rd IEEE International Memory Workshop (IMW), Monterey, CA, 2011,
pp. 175-178.
11.AIST press release, “Development of the 1T FeRAM: Towards the Realization
of the Ultra-Gbit Next-Generation Semiconductor Memory,” Oct 24, 2002.
12.J. S. Cross, S.-H. Kim, S. Wada, and A. Chatterjee, “Characterization of Bi
and Fe co-doped PZT capacitors for FeRAM,” Sci. Technol. Adv. Mater. 11 (4)
(August 2010), 044402.
13.F. Chu and T. Davenport, “The Endurance Performance of 0.5 m FRAM
Products,” [Online]. Available: http://www.ramtron.com/files/tech_papers/FRAM_Endurance.pdf. (Accessed: September 26, 2011).
14.R. Waser, R. Dittmann, G. Staikov, and K. Szot, “Redox-Based Resistive
Switching Memories – Nanoionic Mechanisms, Prospects, and Challenges,”
Adv. Mater. 2009, 21, 2632–2663.
15.“Hot forming to improve memory window and uniformity of low-power
HfOx-based RRAMs”, B. Butcher, , G. Bersuker, K. G. Young-Fisher, D.C.
Gilmer, A. Kalantarian, Y. Nishi, R. Geer, P.D. Kirsch, R. Jammy, Proceedings
of 2012 International Memory Workshop, pages 49-52.
www.keithley.com55
www.keithley.com
Electrical Characterization of Photovoltaic
Materials and Solar Cells with the
Model 4200-SCS Parameter Analyzer
I‑V, C‑V, C-f, DLCP, Pulsed I-V, Resistivity, and
Hall Voltage Measurements
Introduction
The increasing demand for clean energy and the largely
untapped potential of the sun as an energy source is making
solar energy conversion technology increasingly important.
As a result, the demand for solar cells, which convert sunlight
directly into electricity, is growing. Solar or photovoltaic (PV)
cells are made up of semiconductor materials that absorb
photons from sunlight and then release electrons, causing an
electric current to flow when the cell is connected to a load.
A variety of measurements are used to characterize a solar
cell’s performance, including its output and its efficiency. This
electrical characterization is performed as part of research and
development of photovoltaic cells and materials, as well as
during the manufacturing process.
Some of the electrical tests commonly performed on solar
cells involve measuring current and capacitance as a function
of an applied DC voltage. Capacitance measurements are
sometimes made as a function of frequency or AC voltage.
Some tests require pulsed current-voltage measurements.
These measurements are usually performed at different light
intensities and under different temperature conditions. A variety
of important device parameters can be extracted from the DC
and pulsed current-voltage (I‑V) and capacitance‑voltage (C‑V)
measurements, including output current, conversion efficiency,
maximum power output, doping density, resistivity, etc. Electrical
characterization is important in determining how to make the
cells as efficient as possible with minimal losses.
Instrumentation such as the Model 4200-SCS Parameter
Analyzer can simplify testing and analysis when making these
critical electrical measurements. The Model 4200-SCS is an
integrated system that includes instruments for making DC and
ultra fast I‑V and C‑V measurements, as well as control software,
graphics, and mathematical analysis capability. The Model 4200SCS is well-suited for performing a wide range of measurements,
including DC and pulsed current-voltage (I‑V), capacitancevoltage (C‑V), capacitance-frequency (C-f), drive level capacitance
profiling (DLCP), four-probe resistivity (ρ, σ), and Hall voltage
(V H) measurements. This application note describes how to use
the Model 4200-SCS to make these electrical measurements
on PV cells.
Making Electrical Measurements
with the Model 4200-SCS
To simplify testing photovoltaic materials and cells, the Model
4200-SCS is supported with a test project for making many of
the mostly commonly used measurements easily. These tests,
which include I‑V, capacitance, and resistivity measurements, also
include formulas for extracting common parameters such as the
maximum power, short circuit current, defect density, etc. The
SolarCell project (Figure 1) is included with all Model 4200-SCS
systems running KTEI Version 8.0 or later. It provides thirteen
tests (Table 1) in the form of ITMs (Interactive Test Modules)
and UTMs (User Test Modules) for electrical characterization.
Table 1. Test modules in the SolarCell project
Subsite Level
IV_sweep
Test Module
fwd-ivsweep
Description
Performs I‑V sweep and calculates Isc,
Voc, Pmax, Imax, Vmax, FF
rev-ivsweep
Performs reversed bias I‑V sweep
CV_sweep
cvsweep
Generates C‑V sweep
C-2vsV
Generates C‑V sweep and calculates 1/C2
cfsweep
Sweeps the frequency and measures
capacitance
DLCP
Measures capacitance as AC voltage is
swept. DC voltage is applied so as to
keep the total applied voltage constant.
The defect density is calculated.
Pulse-IV
pulse-iv-sweep Performs pulse I-V sweep using one
channel of PMU
4PtProbe_resistivity HiR
Uses 3 or 4 SMUs to source current and
measure voltage difference for high
resistance semiconductor materials.
Calculates sheet resistivity.
LoR
Uses 1 or 2 SMUs to source current and
measure voltage using remote sense.
Calculates sheet resistivity. Uses current
reversal method to compensate for
thermoelectric voltage offsets.
vdp_resistivity
I1_V23
First of 4 ITMs that are used to measure
the van der Pauw resistivity. This ITM
sources current between terminals 1 and
4 and measures the voltage difference
between terminals 2 and 3.
I2_V34
Sources current between terminals 2 and
1 and measures the voltage difference
between terminals 3 and 4.
I3_V41
Sources current between terminals 3 and
2 and measures the voltage difference
between terminals 4 and 1.
I4_V12
Sources current between terminals 4 and
1 and measures the voltage difference
between terminals 1 and 2.
www.keithley.com57
DC I-V
C-V
Pulsed I-V
Resistivity
Figure 1. Screenshot of SolarCell project for the Model 4200-SCS
DC Current-Voltage (I‑V) Measurements
As described previously, many solar cell parameters can be
derived from current-voltage (I‑V) measurements of the cell.
These I‑V characteristics can be measured using the Model
4200-SCS’s source measure units (SMU) instruments, which can
source and measure both current and voltage. Because these
SMU instruments have four-quadrant source capability, they
can sink the cell current as a function of the applied voltage.
Two types of SMU instruments are available for the Model
4200-SCS: the Model 4200-SMU, which can source/sink up to
100mA, and the Model 4210-SMU, which can source/sink up
to 1A. If the output current of the cell exceeds these current
levels, it may be necessary to reduce it, possibly by reducing the
area of the cell itself. However, if this is not possible, Keithley’s
Series 2400 or 2600B SourceMeter® SMU Instruments, which
are capable of sourcing/sinking higher currents, offer possible
alternative solutions.
Photon hυ
IL
rsh
Figure 2. Idealized equivalent circuit of a photovoltaic cell
www.keithley.com
If a load resistor (R L) is connected to an illuminated solar
cell, then the total current becomes:
I = IS(eqV/kT – 1) – IL
where: IS = current due to diode saturation
IL = current due to optical generation
rs
PV Cell
Parameters Derived from I‑V Measurements
A solar cell may be represented by the equivalent circuit
model shown in Figure 2, which consists of a light-induced
current source (IL), a diode that generates a saturation current
[IS(eqV/kT –1)], series resistance (rs), and shunt resistance (rsh).
The series resistance is due to the resistance of the metal
contacts, ohmic losses in the front surface of the cell, impurity
concentrations, and junction depth. The series resistance is an
important parameter because it reduces both the cell’s shortcircuit current and its maximum power output. Ideally, the series
resistance should be 0W (rs = 0). The shunt resistance represents
the loss due to surface leakage along the edge of the cell or to
crystal defects. Ideally, the shunt resistance should be infinite
(rsh = ∞).
Load RL
Several parameters are used to characterize the efficiency of
the solar cell, including the maximum power point (Pmax), the
energy conversion efficiency (η), and the fill factor (FF). These
points are illustrated in Figure 3, which shows a typical forward
bias I‑V curve of an illuminated PV cell. The maximum power
point (Pmax) is the product of the maximum cell current (Imax)
and the voltage (Vmax) where the power output of the cell is
greatest. This point is located at the “knee” of the curve.
I sc
Cell Current (mA)
Imax
SMU1
A
Pmax
Force HI
Sense HI
200
V-Source
150
V
Solar Cell
100
Maximum Power Area
Pmax = ImaxVmax
Sense LO
50
Force LO
0
0.0
0.2
0.4
0.6
Cell Voltage (V)
0.8
Vmax Voc
Figure 3. Typical forward bias I‑V characteristics of a PV cell
The fill factor (FF) is a measure of how far the I‑V
characteristics of an actual PV cell differ from those of an ideal
cell. The fill factor is defined as:
I
maxVmax
FF =_____________
IscVoc
where:
Imax = the current at the maximum power output (A)
Vmax= the voltage at the maximum power output (V)
Isc = the short-circuit current (A)
Voc = the open-circuit voltage (V)
As defined, the fill factor is the ratio of the maximum power
(Pmax = ImaxVmax) to the product of the short circuit current (Isc)
and the open circuit voltage (Voc). The ideal solar cell has a fill
factor equal to one (1) but losses from series and shunt resistance
decrease the efficiency.
Another important parameter is the conversion efficiency (η),
which is defined as the ratio of the maximum power output to
the power input to the cell:
Pmax
η = _______
Pin
where:
Pmax= the maximum power output (W)
Pin = the power input to the cell defined as the total radiant
energy incident on the surface of the cell (W)
Making Connections to the Solar Cell
for I‑V Measurements
Figure 4 illustrates a solar cell connected to the Model 4200-SCS
for I‑V measurements. One side of the solar cell is connected
to the Force and Sense terminals of SMU1; the other side is
connected to the Force and Sense terminals of either SMU2 or
the ground unit (GNDU) as shown.
SMU2 or GNDU
Figure 4. Connection of Model 4200-SCS to a solar cell for I‑V measurements
Using a four-wire connection eliminates the lead resistance
that would otherwise affect this measurement’s accuracy.
With the four-wire method, a voltage is sourced across the
solar cell using one pair of test leads (between Force HI and
Force LO), and the voltage drop across the cell is measured
across a second set of leads (across Sense HI and Sense LO).
The sense leads ensure that the voltage developed across the
cell is the programmed output value and compensate for the
lead resistance.
Forward-Biased I‑V Measurements
Forward-biased I‑V measurements of the solar cell are made
under controlled illumination. The SMU instrument is set up to
output a voltage sweep and measure the resulting current. This
forward bias sweep can be performed using the “fwd-ivsweep”
ITM, which allows adjusting the sweep voltage to the desired
values. As previously illustrated in Figure 3, the voltage source
is swept from V1 = 0 to V2 = VOC. When the voltage source is 0
(V1 = 0), the current is equal to the source-circuit current (I1 =
ISC). When the voltage source is an open circuit (V2 = VOC), then
the current is equal to zero (I2 = 0). The parameters, VOC and
ISC, can easily be derived from the sweep data using the Model
4200-SCS’s built-in mathematical analysis tool, the Formulator.
For convenience, the SolarCell project has the commonly derived
parameters already calculated, so the values automatically
appear in the Sheet tab every time the test is executed. Figure 5
shows some of the derived parameters in the Sheet tab. These
parameters include the short-circuit current (ISC), the open
circuit voltage (VOC), the maximum power point (Pmax), the
maximum cell current (Imax), the maximum cell voltage (Vmax),
and the fill factor (FF).
Th user can easily add other formulas depending on the
required parameters that need to be determined.
Using the Formulator, the conversion efficiency (η) can also
be calculated if the user knows the power input to the cell and
inputs the formula. The current density (J) can also be derived
by using the Formulator and inputting the area of the cell.
www.keithley.com59
Figure 5. Results of calculated parameters shown in Sheet tab
Figure 6 shows an actual I‑V sweep of an illuminated
silicon PV cell generated with the Model 4200-SCS using the
“fwd-ivsweep” ITM. Because the system’s SMU instruments can
sink current, the curve passes through the fourth quadrant and
allows power to be extracted from the device (I–, V+). If the
current output spans several decades as a function of the applied
voltage, it may be desirable to generate a semilog plot of I vs.
V. The Graph tab options support an easy transition between
displaying data graphically on either a linear or a log scale.
go to the Graph Settings tab, select Axis Properties, select the
Y1 Axis tab, and click on the Invert checkbox. The inverse of the
graph will appear as shown in Figure 7.
Figure 7. Inversion of the forward-biased I‑V curve about the voltage axis
Figure 6. I‑V sweep of silicon PV cell generated with the 4200-SMU
If desired, the graph settings functions make it easy to create
an inverted version of the graph about the voltage axis. Simply
www.keithley.com
The series resistance (rs) can be determined from the forward
I‑V sweep at two or more light intensities. First, make I‑V curves
at two different intensities (the magnitudes of the intensities
are not important). Measure the slope of this curve from the
far forward characteristics where the curve becomes linear. The
inverse of this slope yields the series resistance:
∆V
rs = ____
∆I
By using additional light intensities, this technique can be
extended using multiple points located near the knee of the
curves. As illustrated in Figure 8, a line is generated from which
the series resistance can be calculated from the slope.
Current (mA)
When considered as ammeters, one important feature of the
Model 4200-SCS’s SMU instruments is their very low voltage
burden. The voltage burden is the voltage drop across the
ammeter during the measurement. Most conventional digital
multimeters (DMMs) will have a voltage burden of at least 200mV
at full scale. Given that only millivolts may be sourced to the
sample in solar cell testing, this can cause large errors. The
Model 4200-SCS’s SMU instruments don’t produce more than a
few hundred microvolts of voltage burden, or voltage drop, in
the measurement circuit.
rS =
VReverse Bias
Linear region used
∆VReverse Bias
rsh ≈
∆VReverse Bias
∆IReverse Bias
to estimate rsh
∆IReverse Bias
log IReverse Bias
Figure 9. Typical reverse-biased characteristics of a PV cell
Figure 10 shows an actual curve of a reverse-biased solar cell,
generated using the ITM “rev-ivsweep”. In this semi-log graph,
the absolute value of the current is plotted as a function of the
reverse-biased voltage that is on an inverted x-axis.
∆V
∆I
Voltage (V)
Figure 8. Slope method used to calculate the series resistance
Reverse-Biased I‑V Measurements
The leakage current and shunt resistance (rsh) can be derived
from the reverse-biased I‑V data. Typically, the test is performed
in the dark. The voltage is sourced from 0V to a voltage level
where the device begins to break down. The resulting current
is measured and plotted as a function of the voltage. Depending
on the size of the cell, the leakage current can be as small as
picoamps. The Model 4200-SCS has a preamp option that allows
making accurate measurements well below a picoamp. When
making very sensitive low current measurements (nanoamps
or less), use low noise cables and place the device in a shielded
enclosure to shield it electrostatically. This conductive shield is
connected to the Force LO terminal of the Model 4200-SCS. The
Force LO terminal connection can be made from the outside
shell of the triax connectors, the black binding post on the
ground unit (GNDU), or from the Force LO triax connector
on the GNDU.
One method for determining the shunt resistance of the PV
cell is from the slope of the reverse-biased I‑V curve, as shown
in Figure 9. From the linear region of this curve, the shunt
resistance can be calculated as:
∆VReverse Bias
rsh = _______________
∆ IReverse Bias
Figure 10. Reverse-biased I‑V measurement of silicon solar cell using the
Model 4200‑SMU
Capacitance Measurements
Capacitance-voltage measurements are useful in deriving
particular parameters about PV devices. Depending on the type
of solar cell, capacitance-voltage (C‑V) measurements can be
used to derive parameters such as the doping concentration
and the built-in voltage of the junction. A capacitance-frequency
(C-f) sweep can be used to provide information on the existence
of traps in the depletion region. The Model 4210-CVU, the
Model 4200-SCS’s optional capacitance meter, can measure
the capacitance as a function of an applied DC voltage (C‑V), a
function of frequency (C-f), a function of time (C-t), or a function
of the AC voltage. The Model 4210-CVU can also measure
conductance and impedance.
To make capacitance measurements, a solar cell is connected
to the Model 4210-CVU as shown in Figure 11. Like I‑V
measurements made with the SMU instrument, the capacitance
measurements also involve a four-wire connection to compensate
for lead resistance. The HPOT/HCUR terminals are connected to
www.keithley.com61
the anode and the LPOT/LCUR terminals are connected to the
cathode. This connects the high DC voltage source terminal of
the Model 4210-CVU to the anode.
4210-CVU
AC
Source
HCUR
HPOT
AC
Voltmeter
Solar
Cell
LPOT
ABB
Feedback
AC
Ammeter
LCUR
Figure 12. C‑V sweep of a silicon solar cell
Figure 11. Connecting the solar cell to the Model 4210-CVU capacitance meter
Figure 11 shows the shields of the four coax cables coming
from the four terminals of the capacitance meter. The shields
from the coax cables must be connected together as close as
possible to the solar cell to obtain the highest accuracy because
this reduces the effects of the inductance in the measure circuit.
This is especially important for capacitance measurements made
at higher test frequencies.
Performing an Open and Short Connection Compensation
will reduce the effects of cable capacitance on measurement
accuracy. This simple procedure is described in Section 15 of the
Model 4200-SCS Reference Manual.
Given that the capacitance of the cell is directly related to
the area of the device, it may be necessary to reduce the area of
the cell itself, if possible, to avoid capacitances that may be too
high to measure. Also, setting the Model 4210-CVU to measure
capacitance at a lower test frequency and/or lower AC drive
voltage will allow measuring higher capacitances.
C‑V Sweep
C‑V measurements can be made either forward-biased or reversebiased. However, when the cell is forward-biased, the applied
DC voltage must be limited; otherwise, the conductance may get
too high for the capacitance meter to measure. The maximum
DC current cannot be greater than 10mA; otherwise, the
instrument’s DC voltage source will go into compliance and the
DC voltage output will not be at the desired level.
Figure 12 illustrates a C‑V curve of a silicon solar cell
generated by the Model 4210-CVU using the “cvsweep”
ITM. This test was performed in the dark while the cell was
reversed-biased.
Rather than plotting dC/dV, it is sometimes desirable to view
the data as 1/C2 vs. voltage because some parameters are related
to the 1/C2 data. For example, the doping density (N) can be
www.keithley.com
Figure 13. 1/C 2 vs. voltage of a silicon solar cell
derived from the slope of this curve because N is related to the
capacitance by:
2
N(a) = ______________________
2
qES A [d(1/C2)/dV]
where:
N(a)= the doping density (1/cm3)
q = the electron charge (1.60219 × 10 –19C)
Es = semiconductor permittivity (1.034 × 10 –12F/cm for silicon)
A = area (cm2)
C = measured capacitance (F)
V = applied DC voltage (V)
The built-in voltage of the cell junction can be derived from
the intersection of the 1/C2 curve and the horizontal axis. This
plot should be a fairly straight line. An actual curve taken with
the Model 4210-CVU, generated using the “C-2vsV” ITM, is shown
in Figure 13. The Formulator function is used to derive both
the doping density (N) and the built-in voltage on the x-axis
(x-intercept). The doping density is calculated as a function of
voltage in the Formulator and appears in the Sheet tab in the
ITM. The user must input the area of the cell in the Constants
area of the Formulator. The built-in voltage source value is
derived both in the Formulator and by using a Linear Line Fit
option in the Graph settings. Notice the value of the x-intercept
appears in the lower left corner of the graph.
C-f Sweep
The Model 4210-CVU option can also measure capacitance,
conductance, or impedance as a function of the test frequency.
The range of frequency is from 1kHz to 10MHz. The curve
in Figure 14 was generated by using the “cfsweep” ITM. Both
the range of sweep frequency and the bias voltage can be
adjusted. The desired parameters, such as the trap densities,
can be extracted from the capacitance vs. frequency data. The
measurements can be repeated at various temperatures.
Thus, in DLCP, the position (xe) can be varied by adjusting the
DC voltage bias to the sample. This also allows determining the
defect density as a function of the distance, or special profiling.
The test frequency and temperature of the measurement can also
be varied to show a profile that is energy dependent.
Once the measurements are taken, a quadratic fit of the C‑V
data is related to the impurity density at a given depletion depth
as follows for a p-type semiconductor:
E V + Ee
ρ
C 03
NDL ≡
= e = p + ∫ 0 g ( E , xe )dE
2
EF
2 qε A C 1
q
where:
NDL = defect density (cm–3)
C1, C0 = coefficients of quadratic fit of C‑V data
q
= electron charge (1.60 × 10 –19C)
ε
= permittivity (F/cm)
A
= area of solar cell (cm2)
ρe
= charge density (C/cm3)
p
= hole density (cm–3)
xe = distance from interface where EF – Ev = Ee
The coefficients C0 and C1 are determined via a full leastsquares best fit of the data to a quadratic equation:
dQ/dV = C2 (dV)2 + C1*(dV) + C0
However, only the C0 and C1 coefficients are used in the analysis.
Figure 14. C-f Sweep of Solar Cell
Drive Level Capacitance Profiling (DLCP)
Drive Level Capacitance Profiling (DLCP) is a technique for
determining the defect density (NDL) as a function of depth of a
photovoltaic cell1. During the DLCP measurement, the applied
AC voltage (peak-to-peak) is swept and the DC voltage is varied
while the capacitance is measured. This is in contrast to the
conventional C‑V profiling technique, in which the AC rms
voltage is fixed and the DC voltage is swept.
In DLCP, the DC voltage is automatically adjusted to keep the
total applied voltage (AC + DC) constant while the AC voltage is
swept. By maintaining a constant total bias, the exposed charge
density (ρe) inside the material stays constant up to a fixed
location (xe), which is defined as the distance from the interface
where EF – Ev = Ee. This is also in contrast to conventional C‑V
profiling, the analysis of which assumes that the only charge
density changes occur at the end of the depletion region.1
1J. T. Heath, J. D. Cohen, W. N. Shafarman, “Bulk and metastable defects in CuIn1–xGa xSe2
thin films using drive-level capacitance profiling,” Journal of Applied Physics, vol. 95, no.
3, p. 1000, 2004
The “DLCP” UTM allows making C‑V measurements for drive
level capacitance profiling. During these measurements, the
total applied voltage remains constant as the DC voltage bias is
automatically adjusted as the AC voltage drive level amplitude
varies. The AC amplitude of the 4210-CVU can vary from 10mVrms
to 100mVrms (14.14mV to 141.4mVp-p). The range of frequency can
also be set from 1kHz to 10MHz. The capacitance is measured as
the AC voltage is sweeping.
Table 2 lists the input parameters used in the UTM, the
allowed range of input values, and descriptions. The user
inputs the total applied voltage (VmaxTotal), the AC start, stop,
and step voltages (VacppStart, VacppStop, and VacppStep), the
time between voltage steps (SweepDelay), the test frequency
(Frequency), the measurement speed (Speed), the measurement
range (CVRange), and offset compensation (OpenComp,
ShortComp, LoadComp, and LoadVal).
Table 2. Adjustable parameters for the DLCP UTM
Parameter
VmaxTotal
VacppStart
VacppStop
VacppStep
SweepDelay
Frequency
Speed
CVRange
OpenComp
Range
–10 to 10 volts
.01414 to .1414
.02828 to .1414
.0007070 to .1414
0 to 100
1E+3 to 10E+6
0, 1, 2
0, 1E–6,
30E–6, 1E–3
1, 0
Description
Applied DC Volts and ½ AC Volts p-p
Start Vac p-p
Stop Vac p-p
Step Vac p-p
Sweep delay time in seconds
Test Frequency in Hertz
0=Fast, 1=Normal, 2=Quiet
0=autorange, 1µA, 30µA, 1mA
Enables/disables open compensation for CVU
www.keithley.com63
ShortComp
LoadComp
LoadVal
1, 0
1, 0
1 to 1E+9
Enables/disables short compensation for CVU
Enables/disables load compensation for CVU
Load value
Once the test is executed, the capacitance, AC voltage, DC
voltage, time stamp, frequency, and the defect density (NDL)
are determined and their values are listed in the Sheet tab. The
defect density is calculated in the Formulator using a quadratic
line fit of the C‑V data. The coefficients (C0 and C1) of the
quadratic equation are also listed in the Sheet tab. The user
inputs the area and permittivity of the solar cell to be tested into
the Constants/Values/Units area of the Formulator.
To make pulsed I-V measurements on a solar cell, the Model
4225-PMU is connected to the cell as shown in Figure 16. Each
PMU has two channels so the solar cell can be connected using
either one or two channels. In the one-channel case shown, one
end of the cell is connected to the HI terminal of PMU CH1 and
the other side of the cell is connected to the shield of the coax
cable, which is the LO terminal of the PMU.
Solar Cell
Figure 15 shows the measurement results in the graph of
capacitance vs. AC voltage p-p. Notice the coefficients of the
derived quadratic line fit and the defect density are displayed on
the graph.
Connection to
common terminal
SMA Coax
Cable
The capacitance measurements can be repeated at various
applied total voltages in order to vary the position of xe. The
energy (Ee) can be varied by changing the test frequency (1kHz
to 10MHz) or the temperature. To change the temperature of the
measurement, the user can add a User Test Module (UTM) to
control a temperature controller via the Model 4200-SCS’s GPIB
interface. The Model 4200-SCS is provided with user libraries for
operating the Temptronics and Triotek temperature controllers.
A
A
V
50Ω
CH1
V
50Ω
4225-PMU
CH2
Figure 16. Connecting the solar cell to the Model 4225-PMU Ultra-Fast
I-V Module
Unlike the DC I-V and C-V measurements, the 4225-PMU uses
a two-wire technique. The Short Compensation feature can be
used to “zero out” the voltage drops due to the cables so that a
4-wire measurement technique isn’t necessary.
Because solar cells are fairly capacitive, it is important to
ensure the pulse width is long enough for the pulsed I-V sweep.
The waveform capture mode should be used to verify the pulse
Figure 15. Capacitance vs. AC voltage p-p of a solar cell
Pulsed I-V Measurements
Pulsed I-V measurements can be useful for studying parameters
of solar cells. In particular, pulsed I-V measurements have been
used to determine the conversion efficiency, minimum carrier
lifetime, and the effects of cell capacitance. The Model 4225-PMU,
the Model 4200-SCS’s optional Ultra-Fast I-V Module, can output
pulsed voltage and measure current, and can capture ultrahigh-speed current or voltage waveforms in the time domain. In
addition to sourcing a pulsed voltage, the PMU can sink current
so it can measure a solar cell’s current output.
Figure 17. Pulsed I-V measurement on solar cell using Model 4225-PMU
www.keithley.com
width prior to generating the pulsed I-V sweep. The waveform
capture mode enables a time-based current and/or voltage
measurement that is typically the capture of a pulsed waveform.
This can be used to perform a dynamic test on the cell or used
as a diagnostic tool for choosing the appropriate pulse settings
in the pulsed I-V mode. Given that larger solar cells have larger
capacitances, it may be necessary to reduce the area of the cell
itself to avoid a long settling time in the measurement.
The results of generating a pulsed I-V measurement sweep on
a silicon solar cell are shown in Figure 17. Note that the current
is in the fourth quadrant of the curve. This indicates that the
PMU is sinking current; in other words, the current is flowing
out of the solar cell and into the PMU.
Resistivity and Hall Voltage Measurements
Determining the resistivity of a solar cell material is a common
electrical measurement given that the magnitude of the resistivity
directly affects the cell’s performance. Resistivity measurements
of semiconductor materials are usually performed using a
four-terminal technique. Using four probes eliminates errors
due to the probe resistance, spreading resistance under each
probe, and the contact resistance between each metal contact
and the semiconductor material. Two common techniques for
determining the resistivity of a solar cell material are the fourpoint collinear probe method and the van der Pauw method.
The SolarCell project contains several ITMs for making both
types of measurements. More detailed information about making
resistivity measurements on semiconductor materials using
the Model 4200-SCS can be found in Keithley Application Note
#2475, “Four-Probe Resistivity and Hall Voltage Measurements
with the Model 4200-SCS.”
Four-Point Collinear Probe Measurement Method
The four-point collinear probe technique involves bringing four
equally spaced probes in contact with a material of unknown
resistance. The probe array is placed in the center of the material
as shown in Figure 18. The two outer probes are used to source
current and the two inner probes are used to measure the
resulting voltage difference across the surface of the material.
Current
Source
Voltmeter
Figure 18. Four-point collinear probe resistivity configuration
From the sourced current and the measured voltage, the
surface or sheet resistivity is calculated by:
pV
σ = ____×___
ln2 I
where:
σ= surface resistivity (W/ ■)
V = the measured voltage (V)
I = the source current (A)
Note that the units for sheet resistivity are expressed as ohms
per square (W/ ■) in order to distinguish this number from the
measured resistance (V/I), which is simply expressed in ohms.
Correction factors to the resistivity calculation may be required
for extremely thin or thick samples or if the diameter of the
sample is small relative to the probe spacing.
If the thickness of the sample is known, the volume resistivity
can be calculated as follows:
pV
ρ = ____×___× t × k
ln2 I
where:
ρ=volume resistivity (W-cm)
t = the sample thickness (cm)
k = a correction factor* based on the ratio of the probe spacing
to wafer diameter and on the ratio of wafer thickness to
probe spacing
*The correction factors can be found in a standard four-point
probe resistivity test procedure such as Semi MF84: Standard
Test Method for Measuring Resistivity of Silicon Wafers With
an In-Line Four-Point Probe. This standard was originally
published by ASTM International as ASTM F 84.
Using the Four-Point Probe ITMs, HiR and LoR
The “HiR” and “LoR” ITMs are both used for making four-point
collinear probe measurements. The “HiR” ITM can be used
for materials over a wide resistance range, ~1mW to 1TW.
The Model 4200-PA preamps are required for making high
resistance measurements (>1MW). The “LoR” ITM is intended for
measurements of lower resistance materials (~1mW–1kW).
A screenshot of the “HiR” ITM for measuring four-probe
resistivity is shown in Figure 19.
The “HiR” ITM uses either three or four SMU instruments to
make the resistivity measurements. One SMU instrument (SMU1)
and the ground unit (GNDU) are used to source current between
the outer two probes. Two other SMU instruments (SMU2 and
SMU3) are used to measure the voltage drop between the two
inner probes. The Force HI terminal of each SMU instrument
is connected to each of the four probes. The SMU instrument
designation for this configuration is shown in Figure 20.
In the Formulator, the voltage difference between SMU2 and
SMU3 is calculated and the resistance and sheet resistivity are
derived from the voltage difference. The results appear in the
Sheet tab of the ITM.
www.keithley.com65
Figure 19. “HiR” test module for measuring resistivity
When making high resistance measurements, potential
sources of error need to be considered in order to make optimal
measurements. Use a probe head that has a level of insulation
resistance between the probes that is sufficiently higher than
the resistance of the material to be measured. This will help
prevent errors due to leakage current through the probe head.
Ensure that the measurement circuit is electrostatically shielded
by enclosing the circuit in a metal shield. The shield is connected
to the LO terminal of the 4200. The LO terminal is located on
the GNDU or on the outside shell of the triax connectors. Use
triax cables to produce a guarded measurement circuit. This will
prevent errors due to leakage current and significantly reduce
the test time. Finally, the Model 4200-PA preamp option is
required to source very small currents (nanoamp and picoamp
The “LoR” ITM is only used for lower resistance materials
and requires only one or two SMU instruments. In this case, the
Force and Sense terminals of the SMU instruments are connected
to the four-point probe as shown in Figure 20.
SMU1
SMU2 or GNDU
Sense HI
Sense HI
Force HI
Force HI
1
Using the Formulator,
calculate the voltage
difference between
SMU2 and SMU3.
SMU1: Set to Current
Bias (VMU) – Set
current level to have
about a 10mV drop
between SMU2 and
SMU3.
SMU2: Set to Current
Bias (VMU) – Use as
high impedance voltmeter, and set
current to 0A on 1nA
range.
SMU3: Set to Current
Bias (VMU) – Use as
high impedance voltmeter, and set
current to 0A on 1nA
range.
GNDU: Common
connection for all
SMUs. Or, this can be
SMU4 set to
Common.
Force HI
Force HI
Force HI
Force HI
Figure 20. SMU instrument designation for four-point collinear probe
measurements
www.keithley.com
range) and to provide high input impedance (>1E16 ohms) to
avoid loading errors when measuring the voltage difference.
2
3
4
Figure 21. Connecting two SMU instruments for four-point probe
measurements
In the configuration shown in Figure 21, the Force HI
terminal SMU1 sources the current through Probe 1. The voltage
difference between Probes 2 and 3 is measured through the
Sense terminals of the two SMU instruments.
V5
1
2
1
2
1
2
1
2
4
3
4
3
4
3
V3
4
3
V7
V1
V6
1
2
1
2
1
2
1
2
4
3
4
3
4
3
V4
4
3
V8
V2
Figure 22. Van der Pauw resistivity measurement conventions
To compensate for thermoelectric offset voltages, two voltage
measurements are made with currents of opposite polarity.
The two measurements are combined and averaged to cancel
the thermoelectric EMFs. The “LoR” ITM performs this offset
correction automatically by sourcing the two current values in
the List Sweep and then mathematically correcting for the offsets
in the Formulator. The corrected resistance and sheet resistivity
are displayed in the Sheet tab.
V
6 – V5
QB = __________
V
8 – V7
Also, Q and f are related as follows:
(
Q – 1 _______
f
e0.693/f
________
=
arc cosh _______
Q + 1 0.693
2
A plot of this function is shown in Figure 23. The value of “f”
can be found from this plot once Q has been calculated.
Measuring Resistivity with the van der Pauw Method
The van der Pauw (vdp) technique for measuring resistivity
uses four isolated contacts on the boundary of a flat,
arbitrarily shaped sample. The resistivity is derived from eight
measurements made around the sample as shown in Figure 22.
Once all the voltage measurements have been taken, two
values of resistivity, ρA and ρB, are derived as follows:
p(V2 + V4 – V1 – V3)
ρA = ____fAts________________________
ln2
4I
p(V6 + V8 – V5 – V7)
ρB =____fBts________________________
ln2
4I
where:
ρA and ρB are volume resistivities in ohm-cm;
ts is the sample thickness in cm;
V1–V8 represent the voltages measured by the voltmeter;
I is the current through the sample in amperes;
f A and f B are geometrical factors based on sample symmetry, and
are related to the two resistance ratios QA and QB as shown in
the following equations (f A = f B = 1 for perfect symmetry).
QA and QB are calculated using the measured voltages
as follows:
V
2 – V1
QA = __________
V4 – V3
)
1.0
0.9
0.8
f
0.7
0.6
0.5
0.4
1
10
100
Q
Figure 23. Plot of f vs. Q
Once ρA and ρB are known, the average resistivity (ρAVG) can
be determined as follows:
ρA + ρB
ρAVG =__________
2
Using the vdp_resistivity subsite and vdp method ITMs
To automate the vdp resistivity measurements, the SolarCell
project has a vdp-resistivity subsite with four ITMs: “I1_V23”,
www.keithley.com67
Figure 24. Screenshot of van der Pauw test
“I2_V34”, “I3_V41”, and “I4_V12.” A screenshot of the test is shown
in Figure 24.
Each terminal of the sample is connected to the Force HI
terminal of an SMU instrument, so a Model 4200-SCS with four SMU
instruments is required. The four SMU instruments are configured
differently in each of the four ITMs – one SMU instrument supplies
the test current, two are configured as voltmeters, and one is set to
common. This measurement setup is repeated around the sample,
with each of the four SMU instruments serving a different function
ITM NAME:
I2_V34
Common
Current Bias (+)
SMU1
SMU2
ITM NAME:
I4_V12
Common
Current Bias (–)
SMU1
SMU2
Voltmeter
Voltmeter
SMU1
SMU2
V12
Voltmeter
SMU1
SMU2
V12
1
2
1
2
1
2
1
2
4
3
4
3
4
3
4
3
SMU4
SMU3
V34
Voltmeter
Voltmeter
SMU4
SMU3
V34
Voltmeter
Voltmeter
SMU4
SMU3
Current Bias (+)
Common
ITM NAME:
I3_V41
Voltmeter
Common
SMU2
SMU1
1
Voltmeter
Common
SMU2
SMU1
1
Current Bias (+)
Voltmeter
SMU1
2
SMU2
1
3
SMU4
SMU3
SMU4
Voltmeter
3
4
SMU3
Current Bias (–)
Figure 25. SMU instrument configurations for van der Pauw measurements
www.keithley.com
Common
Current Bias (–)
Voltmeter
SMU1
SMU2
1
2
4
3
V23
4
Current Bias (+)
SMU3
Current Bias (–)
2
V41
4
SMU4
ITM NAME:
I1_V23
2
V41
Voltmeter
Voltmeter
SMU4
Common
V23
3
SMU3
Voltmeter
SMU4
Common
SMU3
Voltmeter
in each of the four ITMs. A diagram of the function of each SMU
instrument in each ITM is shown in Figure 25.
Adjusting the Test Parameters
Before executing the test, some of the test parameters must be
adjusted based on the sample to be tested. In particular, it’s
necessary to specify the source current, the settling time, and the
thickness of the material.
Input Source Current: Before running the project, input the
current source values based on the expected sample resistance.
Adjust the current so that the voltage difference will not exceed
approximately 25mV to keep the sample in thermal equilibrium.
In each of the four ITMs, enter both polarities of the test current.
The same magnitude must be used for each ITM.
Input the Settling Time: For high resistance samples, it will be
necessary to determine the settling time of the measurements.
This can be accomplished by creating an ITM that sources
current into two terminals of the samples and measures the
voltage drop on the adjacent two terminals. The settling time
can be determined by taking multiple voltage readings and then
graphing the voltage difference as a function of time.
This settling time test can be generated by copying and
then modifying one of the existing vdp ITMs. Switch the source
function from the sweep mode to the sampling mode. Then, in
the Timing menu, take a few hundred or so readings with a delay
time of one second. Make sure that the “Timestamp Enabled”
box is checked. After the readings are done, plot the voltage
difference vs. time on the graph. The settling time is determined
by observing the graph and finding the time when the reading is
within the desired percentage of the final value.
Input the Thickness of the Sample: Enter the thickness of the
sample into the Calc sheet at the subsite level. Select the subsite
vdp_resistivity. Go to the Subsite Data vdp-device tab. It contains
the output values of the voltage differences and test current.
From the Calc tab, the thickness can be adjusted. The default
thickness is 1cm.
Input Correction Factor: The resistivity formula found in the
Calc sheet at the subsite level also allows inputting a correction
factor, if necessary. The resistivity is multiplied by this number,
which may be based on the geometry or uniformity of the
sample. By default, the correction factor is 1.
Running the Project
The van der Pauw resistivity measurements must be run at the
subsite level. Make sure that all four checkboxes in front of the
vdp ITMs (“I1_V23,” “I2_V34,” “I3_V41,” and “I4_V12”) are checked
and then click on the subsite vdp_resistivity. Execute the project
by using the subsite Run button (circular arrow). Each time the
test is run, the subsite data is updated. The voltage differences
from each of the four ITMS will appear in the Subsite Data
Figure 26. Executing the vdp_resistivity test
www.keithley.com69
where:
R HC and R HD are Hall coefficients in cm3/C;
t is the sample thickness in cm;
V represents the voltages measured in V;
I is the current through the sample in A;
B is the magnetic flux in Vs/cm2
B+
i
1
2
4
Once R HC and R HD have been calculated, the average Hall
coefficient (R HAVG) can be determined as follows:
3
t
______________
R + RHD
RHAVG = HC
2
From the resistivity (ρAVG) and the Hall coefficient (R H), the
Hall mobility (µH) can be calculated:
V2–4+
|RH|
µH = ______
ρAVG
Figure 27. Hall voltage measurement
vdp-device Sheet tab. The resistivity will appear in the Subsite
Data Calc sheet as shown in Figure 26.
Hall Voltage Measurements
Hall effect measurements are important to semiconductor
material characterization because the conductivity type, carrier
density, and Hall mobility can be derived from the Hall voltage.
With an applied magnetic field, the Hall voltage can be measured
using the configuration shown in Figure 27.
With a positive magnetic field (B–), apply a current between
Terminals 1 and 3 of the sample, and measure the voltage drop
(V2–4+) between Terminals 2 and 4. Reverse the current and
measure the voltage drop (V4–2+). Next, apply current between
Terminals 2 and 4, and measure the voltage drop (V1–3+)
between Terminals 1 and 3. Reverse the current and measure the
voltage drop (V3–1+) again.
Reverse the magnetic field (B–) and repeat the procedure,
measuring the four voltages: (V2–4–), (V4–2–), (V1–3–), and (V3–1–).
Table 3 summarizes the Hall voltage measurements.
Table 3. Summary of Hall Voltage Measurements
Voltage
Designation
V2–4+
V4–2+
V1–3+
V3–1+
V2–4–
V4–2–
V1–3–
V3–1–
Magnetic
Flux
B+
B+
B+
B+
B–
B–
B–
B–
Current Forced
Between Terminals
1–3
3–1
2–4
4–2
1–3
3–1
2–4
4–2
Voltage Measured
Between Terminals
2–4
4–2
1–3
3–1
2–4
4–2
1–3
3–1
From the eight Hall voltage measurements, the average Hall
coefficient can be calculated as follows:
t(V
4–2+ – V2–4+ + V2–4– – V4–2–)
RHC = _______________________________________
4BI
t(V3–1+ – V1–3+ + V1–3– – V3–1–)
RHD = _______________________________________
4BI
www.keithley.com
Using the Model 4200-SCS to Measure the Hall Voltage
The SolarCell project does not include a specific test to measure
the Hall voltage; however, four ITMs can be added to the subsite
for determining the Hall coefficient and mobility. Given that the
configuration for the Hall measurements is very similar to the
van der Pauw resistivity measurements, the vdp ITMs can be
copied and modified for making the Hall voltage measurements.
The modifications involve changing the functions of the SMU
instruments. Figure 28 illustrates how to configure the four SMU
instruments in the ITMs to measure the Hall voltage. Use the
Output Value checkboxes on the Definition tab to return the Hall
voltages to the subsite-level Calc sheet.
User Test Modules (UTMs) must be added to control the
magnet. For a GPIB-controlled electromagnet, users can write a
program using KULT (the Keithley User Library Tool) to control
the magnitude and polarity of the electromagnet. The code
can be opened up in a UTM within the project. Information on
writing code using KULT is provided in Section 8 of the Model
4200-SCS Reference Manual.
If a permanent magnet is used, UTMs can be employed to
create a Project Prompt that will stop the test sequence in the
project tree and instruct the user to change the polarity of the
magnetic field applied to the sample. A Project Prompt is a dialog
window that pauses the project test sequence and prompts
the user to perform some action. See Section A of the Model
4200-SCS Reference Manual for a description of how to use
Project Prompts.
Finally, the Hall coefficient and mobility can be derived in the
subsite-level Calc sheet. These math functions can be added to
the other equations for determining resistivity.
Conclusion
Measuring the electrical characteristics of a solar cell is critical
for determining the device’s output performance and efficiency.
The Model 4200-SCS simplifies cell testing by automating the
I‑V, C‑V, pulsed I-V, and resistivity measurements and provides
B+
Current Bias (+)
Voltmeter
SMU1
Current Bias (–)
Voltmeter
SMU1
SMU2
SMU2
1
2
1
2
4
3
4
3
SMU4
Voltmeter
SMU3
SMU4
Common
Voltmeter
Voltage Measured: V2–4+
graphics and analysis capability. For measurements of currents
greater than 1A, Keithley offers Series 2400 and Series 2600B
SourceMeter Instruments that can be used for solar cell testing.
Information on these models and further information on making
solar cell measurements can be found on Keithley’s website:
www.keithley.com.
SMU3
V34
Common
Voltage Measured: V4–2+
B+
Voltmeter
Current Bias (+)
Voltmeter
SMU2
SMU1
1
2
4
3
Current Bias (–)
SMU2
SMU1
V41
1
2
4
3
V41
SMU4
SMU3
Common
SMU4
Voltmeter
SMU3
Common
Voltage Measured: V1–3+
Voltmeter
Voltage Measured: V3–1+
B–
Current Bias (+)
SMU1
V12
Voltmeter
SMU2
Current Bias (–)
Voltmeter
SMU2
SMU1
V12
1
2
1
2
4
3
4
3
SMU4
SMU3
Voltmeter
SMU4
Common
SMU3
Voltmeter
Voltage Measured: V2–4–
Common
Voltage Measured: V4–2–
B–
Voltmeter
SMU1
SMU4
Common
Current Bias (+)
SMU2
Voltmeter
SMU1
Current Bias (–)
SMU2
1
2
1
2
4
3
4
3
SMU3
Voltmeter
Voltage Measured: V1–3+
SMU4
Common
SMU3
Voltmeter
Voltage Measured: V3–1–
Figure 28. SMU configurations for Hall voltage measurements
www.keithley.com71
www.keithley.com
Specifications are subject to change without notice. All Keithley trademarks and trade names are the property of Keithley Instruments, Inc.
All other trademarks and trade names are the property of their respective companies.
A Greater Measure of Confidence
KEITHLEY INSTRUMENTS, INC. ■ 28775 AURORA RD. ■ CLEVELAND, OH 44139-1891 ■ 440-248-0400 ■ Fax: 440-248-6168 ■ 1-888-KEITHLEY ■ www.keithley.com
BENELUX
+31-40-267-5506
www.keithley.nl
FRANCE
+33-01-69-86-83-60
www.keithley.fr
ITALY
+39-049-762-3950
www.keithley.it
MALAYSIA
60-4-643-9679
www.keithley.com
SINGAPORE
01-800-8255-2835
www.keithley.com.sg
BRAZIL
55-11-4058-0229
www.keithley.com
GERMANY
+49-89-84-93-07-40
www.keithley.de
JAPAN
81-120-441-046
www.keithley.jp
MEXICO
52-55-5424-7907
www.keithley.com
TAIWAN
886-3-572-9077
www.keithley.com.tw
CHINA
86-10-8447-5556
www.keithley.com.cn
INDIA
080-30792600
www.keithley.in
KOREA
82-2-6917-5000
www.keithley.co.kr
RUSSIA
+7-495-664-7564
www.keithley.ru
UNITED KINGDOM
+44-1344-39-2450
www.keithley.co.uk
For further information on how to purchase or to locate a sales partner please visit www.keithley.com/company/buy
© Copyright 2014 Keithley Instruments, Inc.