1000 Series Arm
MODEL NUMBER:
Document Number:
1000 Series CMT Arm
Better Solutions Are Within Reach®
Installation Instruction for 1000 Series CMT Arm
PD196-201R4
Page 1 of 7
THIS INSTALLATION SHOULD BE MADE BY A QUALIFIED SERVICE PERSON AND SHOULD CONFORM TO ALL
LOCAL CODES. READ ANDFOLLOW THE SAFETY INSTRUCTIONS BEFORE ATTEMPTING THIS INSTALLATION.
NOTE to CATV system installer: This reminder is provided to call the CATV system installer's attention to Article 820-40
of the NEC that provides guidelines for proper grounding and, in particular, specifies that the cable ground shall be
connected to the grounding system of the building, as close to the point of cable entry as practical.
WARNING: OXYGEN ENVIRONMENT DO
NOT use in an oxygen tent or oxygen
chamber. Such use may cause a fire hazard.
WARNING: To avoid the hazards of fire or
electric shock, DO NOT expose this TV to rain
or moisture.
DANGER: ARM RECOIL HAZARD The
safety brake pin must remain in the
SAFETY BRAKE PIN HOLE whenever the
television set is removed from the arm or
when the arm is removed from the wall
bracket to prevent the arm from springing
open.
Apparatus shall not be exposed to dripping or splashing and no objects filled with liquids, such as vases, shall be placed
on the apparatus. Unplug this unit from the wall outlet before cleaning. Do not use liquid cleaners or aerosol cleaners. Use
a damp cloth for cleaning. This unit REQUIRES periodic maintenance. Refer to the maintenance section for instructions.
1. Read these instructions.
2. Keep these instructions.
3. Heed all warnings.
4. Follow all instructions.
5. Do not use this apparatus near water.
6. Clean only with dry cloth.
7. Do no block any ventilation openings. Install in
accordance with the manufacturer's instructions.
8. Do not install near any heat source such as radiators,
heat registers, stove, or other apparatus (including
amplifiers) that produce heat.
9. Use only with the cart, stand, tripod, bracket or table
specified by the manufacturer, or sold
with the apparatus. When a cart is used,
use caution when moving the
cart/apparatus combination to avoid
injury from tip-over.
10. Do not defeat the safety purpose of the polarized or
grounding-type plug. A polarized plug has two blades
with one wider than the other. A grounding type plug has
two blades and a third grounding prong. The wide blade
or the third prongs are provided for your safety. If the
provided plug does not fit into your outlet, consult an
electrician for replacement of the obsolete outlet.
11. Protect the power cord from being walked on or
pinched particularly at plugs, convenience receptacles,
and the point where they exit from the apparatus.
12. Only use attachments/accessories specified by the
manufacturer.
13. Unplug this apparatus during lightning storms or
when unused for long periods of time.
14. Refer all servicing to qualified service personnel.
Servicing is required when the apparatus has been
damaged in any way, such as power-supply cord or plug
is damaged, liquid has been spilled or objects have
fallen into the apparatus, the apparatus has been
exposed to rain or moisture, does not operate normally,
or has been dropped.
Contact PDi for complete list of approved TV model numbers for which this arm can be used.
PDi Communication Systems, Inc. 40 Greenwood Lane Springboro, Ohio 45066 USA PH 1-800-628-9870 FX 937-743-5664
MODEL NUMBER:
1000 Series CMT Arm
Better Solutions Are Within Reach®
Installation Instruction for 1000 Series CMT Arm
Document Number:
PD196-201R4
Page 2 of 7
Scope:
How to install a PDI-CMT-SUPPORT.
How to install a 1000 series PDi Ceiling Mount ARM to a PDI-CMT support.
Tool List:
1) Safety glasses/goggles
2) P1 OR P2 Phillips screw driver
3) 5/32 Hex key or hex bit and driver
4) 3/16 Hex key or hex bit and driver
5) SUITABLE tools needed for mounting the PDI-CMT-SUPPORT to the building structure.
6) (2) 9/16" wrenches
7) Drill with a 3/8" drill bit
Safety steps:
1) Always wear safety glasses/goggles to prevent injury from debris.
2) Follow the installation/removal instructions for your existing products.
Procedure Overview:
A complete mounting system for attachment to a solid roofing structure above a suspended tile ceiling. Outer tube mounts
to a solid surface, inner tube allows adjustment of height to clear the tile ceiling, and then a special arm section attaches
to the bottom of the tube. The various CEILING MOUNT ARM model can support up to 21 pounds.
PDi Communication Systems, Inc. 40 Greenwood Lane Springboro, Ohio 45066 USA PH 1-800-628-9870 FX 937-743-5664
MODEL NUMBER:
1000 Series CMT Arm
Better Solutions Are Within Reach®
Installation Instruction for 1000 Series CMT Arm
Document Number:
PD196-201R4
Page 3 of 7
Step 1:
LOCATE AND MOUNT OUTER TUBE
1. Remove the ceiling tile and locate a mounting position above the patient's bed at a minimum distance of 52 inches from
walls or other obstacles.
2. Mount the outer tube PDI-592 securely to the ceiling using the twelve mounting holes provided in the plate.
PDi Communication Systems, Inc. 40 Greenwood Lane Springboro, Ohio 45066 USA PH 1-800-628-9870 FX 937-743-5664
MODEL NUMBER:
1000 Series CMT Arm
Better Solutions Are Within Reach®
Installation Instruction for 1000 Series CMT Arm
Document Number:
PD196-201R4
Page 4 of 7
Adjust Height and Drill Inner Tube
1. Slide the inner tube PDI-591 up to the desired height (see note “A”). Push the plastic sleeve PDI-590 until the shoulder
is seated in the outer tube. At the cross hole near the bottom of the outer tube, drill through the plastic sleeve and the
inner tube from each side. Use a 3/8” diameter drill bit. Slide the 3/8” x 2-1/2” bolt through and tighten nut. Cut hole for
tube in ceiling
tile and replace tile.
2. Drop a fish wire through the cable exit hole located near the base plate of the outer tube. (not shown). The wire should
exit the bottom of the tube. Remove the four ¼-20 socket head bolts from the swivel collar PDI-589. Attach the fish wire to
the cable pigtail(s) that exit the arm swivel cap elbow. Being sure to first thread the cable(s) through the plastic shim 1.
washer PDI-586.
Attach the Arm Section
3. Pull the fish wire through the tube until the cables exit the base plate hole. Hold the plastic shim washer PDI-586
against the bottom of the inner tube PDI-591. Be sure it does not overlap the swivel collar. Hold the arm assembly against
the bottom of the inner tube. The supplied thread locking compound LOCTITE® #242 (Blue) must be used on the ¼-20
socket head bolts. Slide the ¼-20 socket head bolts through the swivel cap PDI-688 and thread back into the swivel collar
PDI-589. Tighten the ¼-20 socket head bolts to a torque of 10 ft-lbs.
PDi Communication Systems, Inc. 40 Greenwood Lane Springboro, Ohio 45066 USA PH 1-800-628-9870 FX 937-743-5664
MODEL NUMBER:
1000 Series CMT Arm
Better Solutions Are Within Reach®
Installation Instruction for 1000 Series CMT Arm
Document Number:
PD196-201R4
Page 5 of 7
Step 3:
Attach the TV to the nose of the arm as pictured below, using a PD209-002 socket head cap screw and a PDISLW25 lock
washer. Make the coaxial cable connection per the installation documents supplied with the LCD TV.
After verifying that the LCD TV is correctly connected and functioning properly, begin installing the arm cover kit.
PDi Communication Systems, Inc. 40 Greenwood Lane Springboro, Ohio 45066 USA PH 1-800-628-9870 FX 937-743-5664
MODEL NUMBER:
1000 Series CMT Arm
Better Solutions Are Within Reach®
Installation Instruction for 1000 Series CMT Arm
Document Number:
PD196-201R4
Page 6 of 7
NOTE:
1) Attach arm to wall bracket and mount TV to arm before beginning cover installation.
2) Arm Safety Brake Pin must be removed from arm nose to install the new Nose Arm Cover.
3) DO NOT DISCARD SAFETY BRAKE PIN. Store the safety pin under one of the extra clamps in the nose of the
arm for future service of the arm and/or TV.
STEP 4.
Snap the (2) PD216-001 End Caps into the ends of the arm section as pictured.
STEP 5.
Securely snap the Elbow Covers around the elbow joint. Hook over the top edges of the elbow plates then snap the lower
hooks along the bottom edges.
When the halves are properly installed there should only be a thin seam visible where the halves meet.
STEP 6.
Attach the left and right Nose covers to the nose using (4) PD209-004 screws as pictured.
Tighten the (4) PD209-004 screws to secure the nose covers.
Now install the PDIPPTL61950 Plastite screw into the nose cvores to secure the "bell" around the top of the LCD TV
housing. When the halves are properly installed there should only be a thin seam visible where the halves meet.
STEP 7.
Rotate the TV and move the arm thru its range of travel to check for any inference issues.
PARTS LIST for KIT:
(2) PD216-001 End Caps
(2) PD216-003 Elbow Covers
(1) PD216-004L Left Nose Cover
(1) PD216-004R Right Nose Cover
(1) PDIPPTL61950 Plastite Screw
(4) PD209-004 Painted Cover Screws
The TV is not shown for installation clarity.
PDi Communication Systems, Inc. 40 Greenwood Lane Springboro, Ohio 45066 USA PH 1-800-628-9870 FX 937-743-5664
MODEL NUMBER:
1000 Series CMT Arm
Better Solutions Are Within Reach®
Installation Instruction for 1000 Series CMT Arm
Document Number:
PD196-201R4
Page 7 of 7
ARM MAINTENANCE
Very little maintenance should be required for your PDI Arm to give years of trouble free service if the following lubrication
and inspection procedures are followed.
1. INSPECTION - Occasionally check the elbow swivel joint for any evidence of bending or distortion due to a patient
using the ARM for assistance in getting in or out of bed or any other misuse. If the ARM is not mounted with thesupport
tube assembly absolutely vertical, it will have a tendency to drift and will not properly counter balance the television. Be
certain that the support tube assembly that the arm is mounted to is vertical.
2. SPRING ADJUSTMENT - The gas spring in this arm support a specific weight range for the model construction. These
springs are not "adjustable. The pivot points for some arm models can be moved to alter the weight capacity. Please
contact the manufacturer for details. Otherwise the gas springs will have to be changed to alter the weight capacity of the
arm.
3. LUBRICATION - At least once a year a drop of light machine oil should be put on each pivot point in the nose, elbow,
and base. DO NOT PUT OIL ON THE FRICTION DAMPER PADS!
4. REPLACEMENT OF COAX CABLE - See cable replacement procedure in the full maintenance manual for your arm.
5. TORQUE ADJUSTMENT OF PIVOT SCREWS - All screws in the ARM are tightened securely, and Loctite, at the
factory and do not require specific checks with a torque wrench. IF THE THREADS ARE STRIPPED IN THE NOSE,
ELBOW, OR BASE, THAT PART NEEDS REPLACEMENT - DISCONTINUE USE IMMEDIATELY.