Tens 210T Specifications

Tens 210T Specifications
TENS 210 / 210T
Specifications
Channels:
Output Current:
Output Voltage:
Timer*:
Battery:
Battery Life:
Weight:
Dimensions:
Patient Compliance Meter:
Wave Form:
Pulse Rate:
Pulse Width:
Two, independently controlled
1 – 80 mA into a 500 Ω load
Constant Current
0 – 40 Volts, 500 Ω
30 or 60 minutes, or continuous
DC 9 V battery, alkaline or rechargeable nickel-cadmium
Approximately 70 hours at nominal settings
4 oz. (battery included)
3.7 in, (H) x 2.6 in. (W) x 0.9 in. (T)
Shows treatment times in hours
Symmetric, biphasic rectangular pulse
2 – 150 Hz
30 – 260 s
Function Mode Table:
Mode
Intensity (Vp)
Pulse width (Vp)
Pulse rate (Vp-p)
Mode type
B
C
0 ~ 80 mA
0 ~ 80 mA
30 ~ 260 uS
30 ~ 260 uS
100 Hz
2 ~150 Hz
Burst
Continuous
0 ~ 80 mA
30 ~ 260 uS
2 ~150 Hz
Modulation
M **
** Modulates down from preset width setting by 60% then back to original setting.
* TENS 210T has a timer.
1333 South Claudina Street • Anaheim, California 92805 USA • (800) 854–9305 • (714) 533–2221 • FAX (714) 635–7539