Below you will find brief information for Digital Pressure Gage DPG-000. This manual provides instructions for installing and operating the Dwyer DPG-000 Digital Pressure Gage. It has a 4 digit digital display which reduces the potential for errors in readings by eliminating parallax error commonly produced with analog gages. The DPG-000 is packaged in a durable extruded aluminum case designed to meet NEMA 4/4X. The unit is battery powered and has an auto-shut off to conserve battery life. Readings may be converted to various engineering units. You can find complete instructions for operating the gage in the flow chart provided on the back side of the manual.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Dwyer DPG-000 Digital Pressure Gage User Manual | Manualzz
A34B Bulletin
1/26/05
9:52 AM
Page 1
Bulletin A-34B
Series DPG-000 Digital Pressure Gage
Specifications - Installation and Operating Instructions
3
[76.20]
1-5/8
[40.39]
3-5/8
[91.39]
1/4 NPT
The Series DPG-000 Digital Pressure Gage has a precise +/- 0.50% full scale accuracy. The 4 digit digital display
will reduce the potential for errors in readings by eliminating
parallax error commonly produced with analog gages.
The DPG-000 is packaged in a durable extruded aluminum
case designed to meet NEMA 4/4X. The unit is battery
powered and has an auto-shut off to conserve battery life.
A four button key pad allows easy access to features without the need to work through complex menus or difficult
key combinations. These features include backlight, peak
and valley, auto zero and conversion of the pressure units.
Readings may be converted to various engineering units.
See reverse page for available units.
FEATURES
For complete instructions on the four button operation see
the flow chart provided on back.
On/Backlight: Push to turn on; Press again for backlight.
Press and hold for 5 seconds to turn unit off.
Zero: Zeros the display - Push & hold for 2 seconds to zero
the display. If after pressing zero the display does not rezero, the unit is no longer meeting specified accuracy and
should be sent back to the factory for calibration. Do not
use the zero button when pressure is applied.
Units: Scroll (see reverse).
P/V: Peak recorded reading retrieved by initial push of
button; push again retrieves the valley or lowest value
recorded.
INSTALLATION
When installing gage always use 1˝ hex at the base of the
housing to tighten the gage to a mating fitting. Do not apply
wrench to housing.
MAINTENANCE
Battery Removal: Remove the three screws on backplate.
Lift backplate off by hand. To reassemble, ensure O-Ring
seats into housing evenly and press until seated completely. Replace screws.
SPECIFICATIONS
Service: Liquids and combustible compatible gases.
Wetted Materials: Type 316L SS.
Housing Materials: Black Polycarbonate front & back cover,
anodized aluminum extruded housing with recessed grooves,
Polycarbonate Overlay, Buna-N O-Rings, 316L SS Sensor
Construction.
Accuracy: 0.50% F.S. +/- 1 least significant digit 32 to 130°F (0
to 55°C). (Includes linearity, hysteresis, repeatability) ±2% F.S. @
10°F. ±5% F.S. @ 0°F.
Pressure Limit: 2x pressure range for models ≤1000 psi; 5000
psi for 3000 psi range; 7500 psi for 5000 psi range.
Enclosure Rating: Designed to meet NEMA 4/4X (IP65).
Temperature Limits: 0 to 130°F (-18 to 55°C).
Thermal Effect: Between 70 to 130°F is 0.016%/F.
Between 32 to 70°F is 0.026%/F. Between 10 to 32°F is
0.09%/F. Between 0 to 10°F is 0.50%/F.
Size: 3.00˝ OD x 1.90 Deep (max).
Process Connection: 1/4˝ male NPT.
Weight: 8.84 oz (275 g).
Display: 4 digit (.425 H x .234 W digits).
Power Requirements: Two AAA batteries.
Battery Life: 2000 hours typical; Low battery indicator.
Auto Shut-Off:
Gage: 60 minute auto shut off. Auto shut-off may
be disengaged.
Backlight: 2 minute auto shut-off.
Agency Approvals:
CE EMC Directive:
EN 61000-4-2
EN 61000-4-3 & ENV50204
EN 61000-6-2
EN 61000-6-4
EN 55011.
WARNING: DO NOT pressurize DPG over maximum
allowable pressure limits. Extreme damage can occur if limits are exceeded (see Pressure Limit under Specifications).
An overflow “OFL” designator will flash to acknowledge that
the gages pressure range has been exceeded.
Replace batteries per polarity indicators.
A “LOW BAT” descriptor indication will appear on the display when batteries need to be replaced.
DWYER INSTRUMENTS, INC.
P.O. Box 373 • Michigan City, IN 46361-0373, U.S.A.
Phone: 219/879-8000
Fax: 219/872-9057
www.dwyer-inst.com
e-mail: [email protected]
A34B Bulletin
1/26/05
9:52 AM
Page 2
Bulletin A-34B
Page 2
DPG BUTTON OPERATION
Press to turn unit on.
Press again to turn backlight on (backlight: 2 minute auto shut-off).
Press again to turn backlight off.
Push and hold for 5 seconds to turn unit off.
Press and hold to zero display if a value is read when no pressure is applied.
If after pressing zero the display does not re-zero, the unit is no longer meeting specified
accuracy and should be sent back to the factory for calibration.
DO NOT use the zero button when pressure is applied.
Press to scroll through engineering units.
(psi, in. Hg, oz in.2, in. w.c., ft. w.c., mm Hg, mm w.c., kg/cm2, kPa, mbar, bar, cm w.c.)
Selection of Units on Compound Ranges
Engineering units for vacuum readings may be set independently from engineering units
for positive pressure readings. When selecting engineering units for vacuum readings a
vacuum (negative pressure) must be applied to the instrument. The engineering units for
positive pressure readings must be selected when a positive pressure is applied to the
instrument. Once selected, the DPG will recall and display the engineering units chosen.
[Some units may not be present depending on range]
Press to read peak reading.
Press again to read valley reading.
Press and hold to reset peak and valley.
Press
at any time to return to home display.
Auto Shut-off Disengage
Press and hold
, then press and hold
. A decimal will blink indicating that auto shutoff is disengaged and the unit is in “always-on” mode.
Repeat the process to return unit to auto shut-off mode. Unit will shut-off after 60 minutes.
©Copyright 2004 Dwyer Instruments, Inc.
Printed in U.S.A. 7/04
DWYER INSTRUMENTS, INC.
P.O. Box 373 • Michigan City, IN 46361-0373, U.S.A.
Phone: 219/879-8000
Fax: 219/872-9057
FR# 18-443202-50
www.dwyer-inst.com
e-mail: [email protected]
PAGE 1 OF 4
PAG

advertisement

Key Features

  • Precise +/- 0.50% Full Scale Accuracy
  • 4 Digit Digital Display
  • Durable Extruded Aluminum Case
  • NEMA 4/4X Enclosure Rating
  • Battery Powered
  • Auto-Shut Off
  • Peak and Valley Reading
  • Engineering Unit Conversion
  • Backlight

Frequently Answers and Questions

How do I zero the display?
Press and hold the 'Zero' button for 2 seconds to zero the display.
What happens if the display does not re-zero after pressing the 'Zero' button?
If the display does not re-zero after pressing the 'Zero' button, the unit is no longer meeting specified accuracy and should be sent back to the factory for calibration.
What are the pressure limits for this gage?
The pressure limit is 2x pressure range for models ≤1000 psi; 5000 psi for 3000 psi range; 7500 psi for 5000 psi range.
How do I remove the batteries?
To remove the batteries, remove the three screws on the backplate, lift the backplate off by hand, and replace the batteries per polarity indicators.