HM8115_2
HM8115-2
8kW Power Meter
HM8115-2
HZ815 Power Adapter
R Wide Measurement Range 1 mW…8 kW
R Voltage Range 100 mV…500 V, Current Range 1 mA…16 A
R Frequency Range DC…1 kHz
R Simultaneous Voltage, Current and Power Display
R Display of apparent, effective and reactive Power
RMS Value
R Power Factor Display
R Autoranging, simple Operation
R M
onitor Output (BNC) representing the instantaneous
Active Power
R Suitable for Measurements on Frequency Converters
R Software for Remote Control and Data Acquisition included
Active Power
R G
alvanically isolated USB/RS-232 Interface,
optional IEEE-488 (GPIB)
8 kW Power Meter HM8115-2
All data valid at 23 °C after 30 minutes warm-up.
Voltage
Ranges:
Resolution:
Accuracy:
Input impedance:
Crest factor:
Input protection:
Current
Ranges:
Resolution:
Accuracy:
Crest factor:
Input protection:
True RMS voltage measurement (AC + DC)
50 V
150 V
500 V
0.1 V
1 V
1 V
20 Hz…1 kHz:
±(0.4 % + 5 digit)
DC:
±(0.6 % + 5 digit)
1 MΩ II 100 pF
max. 3.5 at full scale
max. 500 Vp
True RMS current measurement (AC + DC)
160 mA
1.6 A
16 A
1 mA
1 mA
10 mA
20 Hz…1 kHz:
±(0.4 % + 5 digit)
DC:
±(0.6 % + 5 digit)
max. 4 at full scale
fuse, FF 16A 6.3 x 32mm (superfast)
Active power measurement
Ranges:
8 W
24 W
80 W 240 W 800 W 2400 W 8000 W
Resolution:
1 mW 10 mW 10 mW 100 mW 100 mW
1 W
1 W
20 Hz…1 kHz:
±(0.8 % + 10 digit)
Accuracy:
DC:
±(0.8 % + 10 digit)
4-digit, 7-segment LED
Display:
Reactive power measurement
Ranges:
8 var
24 var
80 var 240/800 var 2400/8000 var
Resolution: 10 mvar 100 mvar 100 mvar
1 var
1 var
20…400 Hz: ±(2.5 % + 10 digit + 0.02x P)
Accuracy:
P = active power
4-digit, 7-segment LED
Display:
Apparent power measurement
Ranges:
8 VA
24 VA
80 VA 240/800 VA 2400/8000 VA
Resolution: 1 mVA 10 mVA
10 mVA
100 mVA
1 VA
20 Hz…1 kHz:
±(0.8 % + 5 digit)
Accuracy:
4-digit, 7-segment LED
Display:
Functions and displays
Measurement functions:
Range selection:
Overrange alarm:
Display resolution
Voltage:
Current:
Power:
Power factor:
Interface
Interface:
Connection RS-232:
voltage, current, power, power factor
automatic/manual
visual and acoustic
3-digit, 7-segment LED
4-digit, 7-segment LED
4-digit, 7-segment LED
3-digit, 7-segment LED
Protocol:
Data rate:
Functions:
USB/RS-232 (HO820), IEEE-488 (option)
D-sub connector (galvanic isolation to test
circuit and monitor output)
Xon/Xoff
9600 Baud
control / data fetch
Miscellaneous
Safety Class:
Power supply:
Power consumption:
Operating temperature:
Storage temperature:
Rel. humidity:
Dimensions (W x H x D):
Weight:
Safety Class I (EN 61010)
115…230 V ±10 %, 50/60Hz, CAT II
approx. 15 W at 50 Hz
+5…+40 °C
-20…+70 °C
5…80 % (non condensing)
285 x 75 x 365 mm
approx. 4 kg
Accessories supplied: Line cord, Operating manual, software
Recommended accessories:
HO880 IEEE-488 (GPIB) Interface (galvanically isolated)
HZ10S 5 x silicone test lead black
HZ10R 5 x silicone test lead red
HZ10B 5 x silicone test lead blue
HZ13
Interface cable (USB) 1.8 m
HZ14
Interface cable (serial) 1:1
HZ33
Test cable 50 Ω, BNC/BNC, 0.5 m
HZ34
Test cable 50 Ω, BNC/BNC, 1 m
HZ42
19” Rackmount kit 2RU
HZ72
GPIB-Cable 2 m
HZ815 Socket adapter
Power factor measurement
0.00…+1.00
Display:
50…60 Hz: ±(2 % + 3 digit) (sine wave)
Accuracy:
voltage and current >1/10 of full scale
Monitor output (analog)
BNC connector (galvanic isolation
Connection:
to test circuit and RS-232 interface)
protective earth
Reference potential:
1 Vac at full scale (2400/8000 digit)
Level:
typ. 5 %
Accuracy:
approx. 10 kΩ
Output impedance:
DC…1 kHz
Bandwidth:
±30 V
Protected up to:
HM8115-2E/121010 · C&E · Subject to change without notice · © HAMEG Instruments GmbH® · DQS-certified in accordance with DIN EN ISO 9001:2008, Reg.-No.: 071040 QM08
HAMEG Instruments GmbH · Industriestr. 6 · D-63533 Mainhausen · Tel +49 (0) 6182 8000 · Fax +49 (0) 6182 800100 · www.hameg.com · [email protected]
Was this manual useful for you? yes no