Cross Technologies 2016-75-213 is a downconverter that can translate signals from 719 MHz to 201.143 or 213.143 ± 3 MHz with low group delay and flat frequency response. It has a range of gain from 0 to +40 dB and image rejection of less than 50 dB. The device is rack-mountable and has an alarm function. It can be controlled remotely using a serial interface and has several options like external 10 MHz reference input and output, high stability internal reference, and RS485 remote interface. Possible use cases include signal processing, frequency conversion, and telecommunications.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Manual
Cross Technologies 2016-75-213 Downconverter, 350-750 MHz 213 ±3 MHz in 1 MHs steps Owner's Manual | Manualzz

Instruction Manual

Model 2016-75-213

Downconverter

April 2009 Rev A

REMOTE POWER MUTE ALARM

719 to 201 MHZ

G=+10 REF=INT

MENU

EXECUTE

MODEL 2016

DOWNCONVERTER

C ROSS T ECHNOLOGIES INC.

Data, drawings, and other material contained herein are proprietary to Cross Technologies, Inc., but may be reproduced or duplicated without the prior permission of Cross Technologies, Inc. for purposes of operating the equipment.

When ordering parts from Cross Technologies, Inc., be sure to include the equipment model number, equipment serial number, and a description of the part.

C ROSS T ECHNOLOGIES, INC.

6170 Shiloh Road

Alpharetta, Georgia 30005

(770) 886-8005

FAX (770) 886-7964

Toll Free 888-900-5588

WEB www.crosstechnologies.com

E-MAIL [email protected]

INSTRUCTION MANUAL

MODEL 2016-75-213 Downconverter

TABLE OF CONTENTS PAGE

Warranty

1.0 General

2

3

1.1 Equipment Description

1.2 Technical Characteristics

1.3 Monitor & Control Interface

1.4 Environmental Use Information

3

4

5

7

2.0 Installation

2.1 Mechanical

2.2 Rear I/O’s

2.3 Front Panel Controls, Indicators

2.4 Operation

2.5 Menu Settings

8

8

9

10

11

12

WARRANTY - The following warranty applies to all Cross Technologies, Inc. products.

All Cross Technologies, Inc. products are warranted against defective materials and workmanship for a period of one year after shipment to customer. Cross Technologies,

Inc.’s obligation under this warranty is limited to repairing or, at Cross Technologies, Inc.’s option, replacing parts, subassemblies, or entire assemblies. Cross Technologies, Inc. shall not be liable for any special, indirect, or consequential damages. This warranty does not cover parts or equipment which have been subject to misuse, negligence, or accident by the customer during use. All shipping costs for warranty repairs will be prepaid by the customer. There are not other warranties, express or implied, except as stated herein.

C ROSS T ECHNOLOGIES, INC.

2016-75-213, Rev. A Manual

6170 Shiloh Road

Alpharetta, Georgia 30005

(770) 886-8005

FAX (770) 886-7964

Toll Free 888-900-5588

WEB www.crosstechnologies.com

E-MAIL [email protected]

Page 2 4/16/09

MODEL 2016-75-213 Downconverter

1.0 General

1.1 Equipment Description

The 2016-75-213 Downconverter converts 719 MHz to 201.143 or 213.143 ± 3 MHz with low group delay and flat frequency response. Synthesized local oscillators (LO) provide output frequency selection. Multi-function push button switches select the output frequency, gain, and other parameters. Front panel LEDs provide indication of DC power (green), remote operation (yellow), and PLL alarm (red). Variable attenuators for the

RF input provide a gain range of 0 to +40 dB as adjusted by the front panel multi-function pushbutton switches.

Remote operation allows selection of output frequency and gain. Parameter selection and frequency translation and gain settings appear on the LCD display. Connectors are Type F female for the RF and BNC female for IF and optional 10 MHz input and output signals (option -E). A high stability (±0.01ppm) (option -H) is also available. The unit is powered by a 100-240 ± 10% VAC power supply, and housed in a 1 3/4” X 19“ X 16” rack mount chassis.

AC

REMOTE POWER MUTE ALARM

719 to 201 MHZ

G=+10 REF=INT

MENU

EXECUTE

FRONT PANEL

MODEL 2016

DOWNCONVERTER

C ROSS T ECHNOLOGIES INC.

GND

10 MHZ REF

OUTPUT

MONITOR

AND

CONTROL

5 4

9

3

8 7

2

6

1

J18 J10

REAR PANEL

10 MHZ

EXT REF

INPUT

IF IN

J3 J4

FIGURE 1.1 Model 2016-75-213 Front and Rear Panels

RF OUT

VDC on RF

DS7 J5

719

MHz

IN

VAR

ATT

LO1

IF

BP

1751

LO2

350 to 750 MHz

BP

VAR

ATT

EXT

10 M*

10MHz*

+24VDC*

201.143 or

213.143

MHz

OUT

CONTROLLER

INT/EXT

10 M

201 to 719 MHZ

G=+10 REF=INT

Block Diagram

FIGURE 1.2 Model 2016-75-213 Downconverter Block Diagram

2016-75-213, Rev. A Manual Page 3 4/16/09

1.2 Technical Characteristics

TABLE 1.0 2016-75-213 Downconverter Specifications*

Input Characteristics

Impedance/Return Loss

Frequency

Input Level

75 Ω /12 dB (see TABLE 2.2 for connector options)

719 MHz ± 3 MHZ

-60 to -20 dBm

Output Characteristics

Impedance/Return Loss

Frequency

75 Ω/12 dB (see Table 2.2 for connector options)

201.143 or 213.143 ± 3 MHz

Level Range -30 to -20 dBm

Output 1 dB Compression -15 dBm

Channel Characteristics

Gain range (adjustable)

Image Rejection

Frquency Reponse

Spurious Response

Group Delay, max.

Frequency Sense

0.0 to +40 dB

<50 db, min.

201.143 or 213.143 ± 3 MHz out; ± 0.5 dB

<-45 dBC, in band

0.01 ns/MHz2 parabolic; 0.03 ns/MHz linear; 1 ns ripple

Non-inverting

Synthesizer Characteristics

Frequency Accuracy

Frequency Step

10 MHz In/Out Level

Phase Noise @ Freq

± 1.0 ppm max over temp (± 0.01 ppm option -H) internal reference

1.0 MHz minimum

3 dBm, ± 3 dB, 75 (option -E)

10Hz 1kHz 10kHz 100kHz 1MHz

dBC/Hz -70 -70 -80 -95 -110

Controls, Indicators

Frequency/Gain Selection

Power

Alarm

Remote direct readout LCD; manual or remote

Green LED

Red LED

Yellow LED; RS232C (RS485, option -Q), 9600 baud

Other

RF, IF Connector

IF Connector

Type F (female) (see TABLE 2.2 for other options)

BNC (female) (see TABLE 2.2 for other options)

Ext 10 MHz Conn. (In/Out) BNC (female) (option -E)

Alarm/Remote Connector DB9 - NO or NC contact closure on Alarm

Size

Power

19 inch, 1RU standard chassis 1.75”high X 16.0” deep

100-240 ±10% VAC, 47-63 Hz, 45 watts max

Options Available

E - External 10 MHz ref input & output

H - High Stability Internal Ref (±0.01 ppm)

Q - RS485 Remote Interface

X - 125 kHz Frequency Steps

Connectors/Impedance

B - 75Ω BNC (RF), 75Ω BNC (IF)

C - 50Ω BNC (RF), 75Ω BNC (IF)

D - 50Ω BNC (RF), 50Ω BNC (IF)

N - 50Ω N-type (RF), 75Ω BNC (IF)

M - 50Ω N-type (RF), 50Ω BNC (IF)

S - 50Ω SMA (RF), 50Ω BNC (IF)

*+10˚C to +40˚C; Specifications subject to change without notice.

2016-75-213, Rev. A Manual Page 4 4/16/09

1.3 Monitor and Control Interface

A) Remote serial interface

Protocol: RS-232C, 9600 baud rate, no parity, 8 data bits, 1 start bit, and 1 stop bit.

(RS-232C, RS-422, or RS-485 - option -Q)

7

8

5

6

9

3

4

1

2

M&C Cable Diagram - Cross Technologies Frequency Converters

Female DB-9

PC Com Port

Male DB-9

2015/16/17 M&CPort

7

8

5

6

9

3

4

1

2

Connector: Rear panel, DB-9 male

6

7

4

5

Pin

1

2

3

8

9

Function

Rx-

Rx+ (RS-232C)

Tx+ (RS-232C)

Tx-

GND

Alarm Relay: Common

Alarm Relay: Normally Open

Not Used

Alarm Relay: Normally Closed

2016-75-213, Rev. A Manual Page 5 4/16/09

B) Status Requests Table 1.1 lists the status requests for the 2016-75-40 and briefly describes them

* PLEASE NOTE: The two character {aa}(00-31) prefix, in the table below,

should be used ONLY when RS-485, (OPTION-Q), is selected.

TABLE 1.1 2016-75-213

Command

Command Status

Reference Status

Syntax *

{aaS1}

{aaS2}

Description

Returns {aaS1bccAM} where:

• b = 0 if input frequency = 201 MHz

• b = 1 if input frequency = 213 MHz

• cc = Tx gain

• A = 0 - summary alarm

Returns {aaS2E} where:

• E = Ext 10MHz Status (1 = on, 0 = off)

C) Commands Table 1.2 lists the commands for the 2016-75-213 and briefly describes them. After a command is sent the 2016-75-213 sends a return “>” indicating the command has been received and executed.

General Command Format - The general command format is {aaCND...}, where:

{ = start byte aa = address (RS-485 only - option -Q)

C = 1 character, either C (command) or S (status)

N = 1 character command or status request

D = 1 character or more of data (depends on command)

} = stop byte

* PLEASE NOTE: The two character {aa}(00-31) prefix, in the table below,

should be used ONLY when RS-485, (OPTION-Q), is selected.

Talbe 1.2 2016-75-213

Command

Set Input Frequency

Set Transmit Gain

Set External Reference

Enable Remote

Disable Remote

COMMANDS

Syntax*

{aaC1x}

{aaC3xx}

{aaCEx}

#

{aaCR0}

Description where:

• x = 1 characters

• x = 0 for 201 MHz

• x = 1 for 213 MHz

• Range: 0 to 40 dB where x =

• 0 to disable External 10MHz ref signal

• 1 to enable External 10MHz ref signal

Just # sign

{CR and zero}

2016-75-213, Rev. A Manual Page 6 4/16/09

1.4 Environmental Use Information

A. Rack-Mounting - To mount this equipment in a rack, please refer to the installation instructions located in the user manual furnished by the manufacturer of your eqipment rack.

B. Mechanical loading - Mounting of equipment in a rack should be such that a hazardous condition does not exist due to uneven weight distribution.

C. Elevated operating ambient temperature - If installed in a closed or multi-unit rack assembly, the operating ambient temperature of the rack may be greater than room ambient temperature.

Therefore, consideration should be given to Tmra.

D. Reduced air flow - Installation of the equipment in a rack should be such that the amount of air flow required for safe operation of the equipment is not compromised. Additional space between unit may be required.

E. Circuit Overloading - Consideration should be given to the connection of the equipment to the supply circuit and the effect that overloading of circuits could have on over current protection and supply wiring.

Appropriate consideration of equipment name plate rating should be used, when addressing this concern.

F. Reliable Earthing - Reliable earthing of rack-mounted equipment should be maintained. Particular attention should be given to supply connections other than direct connection to the Branch

(use of power strips).

G. Top Cover - There are no servicable parts inside the product so, the Top Cover should not be removed.

If the Top Cover is removed the ground strap and associated screw MUST BE RE-INSTALLED prior to Top Cover screw replacement. FAILURE TO DO this may cause INGRESS and/or EGRESS emission problems.

2016-75-213, Rev. A Manual Page 7 4/16/09

2.0 Installation

2.1 Mechanical - The 2016-75-213 consists of one RF/Controller PCB housed in a 1 RU (1 3/4 inch high) by

16 inch deep chassis. A switching, ± 12, +24, +5 VDC power supply provides power for the assemblies.

The 2016-75-213 can be secured to a rack using the 4 holes on the front panel.

Figure 2.0 shows how the 2016-75-213 is assembled.

POWER

SUPPLY

FIGURE 2.0 2016-75-213 Mechanical Assembly

2016-75-213, Rev. A Manual Page 8 4/16/09

2.2 Rear Panel Input/Output Signals - Figure 2.1 shows the input and output connectors on the rear panel.

J10 - MONITOR AND CONTROL

DB9 female connector. see Table 2.1.

J3 - 10 MHz EXT REF INPUT (Option -E)

10 MHz external reference input, 3 ± 3 dBm, 75 ohms, BNC female connector.

AC

GND

10 MHZ REF

OUTPUT

J18

MONITOR

AND

CONTROL

5

9

4

8

3

7

2

6

1

J10

10 MHZ

EXT REF

INPUT

J3

IF IN

J4

RF OUT

VDC on RF

DS7 J5

AC1 - POWER IN

AC input for switching power supply.

100-240 ±10% VAC,

47-60 Hz.

J18 - 10 MHz REF OUTPUT (option -E)

10 MHz reference output. 75Ω BNC female connector.

J4 - IF IN

70 MHz,

-10 to -40 dBm input see Table 2.2.

DS7 - SSPB ALARM LED (option -V)

Lights yellow when +24 VDC SSPB voltage is present on RF OUT, J5, center pin.

J5 - RF OUT

250-750 MHz,

0 to -20 dBm output see Table 2.2.

FIGURE 2.1 2016-75-213 Rear Panel I/Oʼs

6

7

4

5

Pin

1

2

3

8

9

Function

Rx-

Rx+ (RS-232C)

Tx+ (RS-232C)

Tx-

GND

Alarm Relay: Common

Alarm Relay: Normally Open

Not Used

Alarm Relay: Normally Closed

*Remote Serial Interface

Interface: DB-9 Male

Protocol: RS-232C (RS-232C/422/485 option -Q), 9600 baud rate, no parity,

8 data bits, 1 start bit, 1 stop bit.

Option

STD

-B

-C

-D

-N

-M

-S

IF

BNC, 75Ω

BNC, 75Ω

BNC, 75Ω

BNC, 50Ω

BNC, 75Ω

BNC, 50Ω

BNC, 50Ω

RF

Type F, 75Ω

BNC, 75Ω

BNC, 50Ω

BNC, 50Ω

Type N, 50Ω

Type N, 50Ω

SMA, 50Ω

2016-75-213, Rev. A Manual Page 9 4/16/09

2.3 Front Panel Controls and Indicators -The following are the front panel controls and indicators.

DS1 - REMOTE LED

Yellow LED indicates remote operation.

LCD DISPLAY

Display shows frequency in MHz and Gain in dB

S1 - MENU/EXECUTE BUTTON

Press this to get into Program mode and to execute any changes.

719 to 213 MHz

G=+10 REF=INT

MENU

EXECUTE

ALARM REMOTE POWER

DS2 - UP ALARM LED

Red LED indicates upconverter alarm.

DS6 - POWER LED

Green LED indicates presence of DC power.

S2 - VERT. TOGGLE

Vertical toggle switch that controls values in the Menu items when in program mode. Does not function in the normal display mode

FIGURE 2.2 2016-75-213 Front Panel Controls and Indicators

2016-75-213, Rev. A Manual Page 10 4/16/09

2.4 Installation / Operation

2.4.1 Installing and Operating the 2016-75-213, Upconverter Section

1. Connect a -10 dBm to -40 dBm signal to IF In, J4 (Figure 2.1)

2. Connect the RF OUT, J5, to the external equipment

3. Connect 100-240 ± 10% VAC, 47 - 63 Hz to AC1 on the back panel.

4. Set the desired output frequency (See Section 2.5 Menu Settings).

5. Set the input level (See Section 2.5 Menu Settings).

6. Set the gain for -10 to +30 dB. Make sure the output stays within -20 to 0 dBm with the gain selected and the input level provided. The firmware will prevent setting gain and input level outside this range.

(See Section 2.5 Menu Settings).

7. Be sure DS6 (green, DC Power) is on and DS2 (red, Alarm) is off (Figure 2.2).

8. Option -V only - To insert SSPB +24 VDC on the RF center pin install 2.5A fast blo fuse in F2 and check that DS7 lights yellow (Figure 2.1)

9. AC Fuse - The fuse is a 5 mm X 20 mm, 2 amp slow blow (Type T) and is inserted in the far slot in the drawer below the AC input as shown in Figure 2.3. There is a spare fuse in the near slot. If a fuse continues to open, the power supply is most likely defective.

FUSE DRAWER

SPARE FUSE DRAWER

AC Fuse - 2 amp slow blow (Type T 2A GDC),

5 mm X 20 mm

INPUT

~

100-240± 10%VAC

47-63 Hz

~

2A MAX

FUSE

TYPE T 2A GDC

250 VOLT

FOR 100 - 240 V~

FIGURE 2.3 Fuse Location and Spare Fuse

4/16/09 2016-75-213, Rev. A Manual Page 11

2.5 Menu Settings

2.5.1 Functions - This section describes operation of the front panel controls. There are three operator switches, the LCD display and alarm indicator LEDs. All functions for the equipment are controlled by these components. The functions are (see Figure 2.4):

Power Up

Normal Display

Menu 1 Frequency in MHz

Menu 2 Gain

Menu 3 Select External 10 MHz

Menu 4 Set Remote

Menu 5 Remote Mode (Option Q)

Menu 6 RS-485 Address (Option Q)

Menu 7 Save Changes

Alarm indications appear on the LEDs (see figure 2.2).

All program changes must start with the operation of the Menu/Execute switch and must also end with the operation of the Menu/Execute switch verified by the “Save Settings?” Menu. If this sequence is not followed, none of the changes will take effect. If programming is initiated and no operator action takes place for approximately 12 seconds (before the final press of the Menu/Execute switch) the display will revert to its previous status and you will need to start over.

2016-75-213, Rev. A Manual Page 12 4/16/09

2.5.2. Power On Settings

NOTE: THE LAST STATUS OF A UNIT IS RETAINED EVEN WHEN POWER IS REMOVED.

WHEN POWER IS RESTORED, THE UNIT WILL RETURN TO IT'S PREVIOUS SETTINGS.

When power is first applied, the LCD display goes through three steps.

1. The LCD goes black to show all segments are functioning.

2. The software version will be displayed.

R E V 1 . 0 0

3. The present frequency and gain of the upconverter is shown.

719 to 201

G = +10 REF=INT

The unit is now operational and ready for any changes the operator may desire.

2.5.3 Control Switches

1. Menu/Execute - Any change to the programming of the unit must be initiated by pressing the Menu/Execute switch and completed by pressing the Menu/Execute switch.

2. Horizontal Switch - This switch is mounted so its movement is horizontal and moves the cursor left or right.

3. Vertical Switch - This switch is mounted so its movement is vertical and has two functions: a. During frequency , gain changes, the vertical movement will raise or lower the number in the direction of the arrows.

b. For other functions such Mute on/off, the vertical switch will alternately turn the function on or off regardless of the direction operated.

2016-75-213, Rev. A Manual Page 13 4/16/09

2.5.4 Frequency Changes

At any time during the modification process, if you have made a mistake and do not wish to save the changes you have made, do not press the Menu/Execute switch; simply do nothing for approximately 12 seconds, and the system will return to the normal operating mode or scroll to “R” and push the Menu/Execute switch and select “NO” in the “SAVE SETTINGS?” window.

NOTE: CHANGES DO NOT TAKE PLACE ON FREQUENCY UNTIL YOU GO TO THE SAVE

MENU AND INDICATE YOU WANT TO SAVE THE CHANGES.

When the display indicates the value desired you can push the

Menu/Execute switch to the next item:

719 to 201

G = +10 REF=INT

OR you can scroll to “R”, push the

Menu/Execute switch to get to:

S A V E S E T T I N G S ? Y N

Selecting Y will save the new settings.

Selecting N will revert to the previous settings.

Pushing the Menu/Execute switch then takes you to the :

719 to 201

G = +10 REF=INT

Figure 2.4 gives the menu items and how to make changes

2016-75-213, Rev. A Manual Page 14 4/16/09

2.5.5 Gain Changes

When you get to this menu note that the gain changes will be made as you make them but if you do not wish to save the changes you have made, scroll to “R” and push the menu/Execute switch and select “NO” in the

“SAVE SETTINGS?” window or do not press the Menu/Execute switch; simply do nothing for approximately 12 seconds, and the system will return to the normal operating mode.

NOTE: CHANGES TAKE PLACE ON LEVEL AND GAIN IMMEDIATELY BUT DO NOT GET

SAVED UNTIL YOU GO TO THE SAVE MENU AND INDICATE YOU WANT TO SAVE THE

CHANGES.

Press the Up/Down switch to change the level in 1 dB steps and then push the Menu/Execute switch to get to the Gain setting:

G A I N = + 2 0 R

Pressing the Up/Down switch to change the gain in 1 or 10 dB steps and then push the Menu/Execute switch to get to the Gain setting:

G A I N = + 2 0 R

By using the horizontal rocker switch the cursor can be moved left or right .

Pressing the Up/Down switch down will toggle the display digit selected until you have the desired gain.

NOTE: THE GAIN WILL BE CHANGED AS YOU ADJUST THE NUMBERS. HOWEVER, THE

VALUE WILL NOT BE STORED UNTIL YOU INDICATE YES IN THE SAVE SETTINGS

WINDOW.

When the display indicates the value desired you can push the

Menu/Execute switch to the next item OR you can scroll to “R”, push the Menu/Execute switch to get to:

S A V E S E T T I N G S ? Y N

Selecting Y will save the new settings.

Selecting N will revert to the previous settings.

Pushing the Menu/Execute switch then takes you to the :

719 to 201

G = +10 REF=INT

Figure 2.4 gives the menu items and how to make changes

2016-75-213, Rev. A Manual Page 15 4/16/09

2.5.5 Alarm Indications

An alarm condition for will occur if any local oscillator phase lock loop (PLL) comes out of lock.

The Mute LED will light if you select Mute and the Remote LED will light when you select the Remote mode.

Power Up

Normal Display

Menu 1 Frequency in MHZ

Menu 2 Gain (-10.0 to +20.0)

Menu 3 Select External

10MHz Ref (option -E

Menu 4 Set Unit to

Remote Operation

Menu 5 Save Settings

ON POWER UP

REV 1.00

NORMAL DISPLAY

719 to 213 MHz

G = +10 REF=INT

PUSHING MENU/EXECUTE SEQUENCE

719 to 213 MHz R

G = +10 R

EXT REF OFF

REMOTE OFF

SAVE SETTINGS?

R

Y N

R SCROLL <>

SCROLL

SCROLL <>

SCROLL

SCROLL <>

SCROLL

PUSH BUTTON

PUSH BUTTON

PUSH BUTTON

FIGURE 2.4 Menu Display and Sequences

2016-75-213, Rev. A Manual Page 16 4/16/09