CAS CI-201A manual
Advertisement
Advertisement
.
2
Cautions for Your Safety
Please comply with 'Cautions for Your Safety', which will lead you to use the product safely and properly to prevent any dangerous situations.
■
Cautions are divided into 'Warning' and 'Alert', which mean as follows.
■
Keep this manual in a place where product users can find out, after finish reading it.
'Warning' means a great possibility led to the death or heavy injury when instructions are violated.
'Alert' means a great possibility led to the injury or material damage when instructions are violated.
3
Warning
Never disassemble, repair or retrofit the product.
It might exclude the product from the quality assurance and cause the damage to devices, electric shock or fire.
Ensure the power plug to be fully inserted to prevent shaking.
Any instable connection might cause electric sparks to set fire.
Ensure the grounding of the product.
Poor grounding might cause failure or electric shock upon electric leak.
Do not damage, process, excessively jerk, bend or twist the power cord.
It might damage the power cord to cause fire or electric shock.
Keep any combustible spray or fire source away.
It might cause fire.
Do not spray water to the outside of the product or use it in any humid place.
It might deteriorate the insulation of electric parts that can cause the electric shock, fire risk or weighing errors.
Do not place the product to the direct sunlight or near any hot object like a heater.
It might cause fire.
4
Attention
Check the weighing error anytime for the accurate weighing.
Any use out of the allowed tolerance for the careless use or other causes might not ensure the accurate weighing.
Customer Service:080-022-0022
Avoid any sudden shock to the product.
It might damage the product to fail the accurate weighing.
Find a proper place to attach the rubber pad at the bottom of the indicator, which was shipped together.
Do not use the product at a place with sudden temperature changes or severe vibrations.
It might cause the weighing error or failure.
Do not install the produce at a place with the excessive electromagnetic wave.
It might cause the wrong weighing.
5
Contents
1. Features ...............................................................................
9
1-1. Features........................................................................................9
1-2. Major Functions..........................................................................9
1-3. Analog and A/D Conversion ....................................................10
1-4. Digital and Display ...................................................................10
1-5. General Specifications.............................................................. 11
1-6. Communication and Option ................................................... 11
2. Specifications in Appearance
...................................................12
2-1. External Dimension ..................................................................12
2-2. Front Panel Descriptions ..........................................................14
2-3. Rear Panel Descriptions............................................................20
3. How to Install
...................................................................................22
3-1. How to Connect Load Cell.......................................................22
4. Weight Setup Mode
......................................................................23
4-1. Weight Setup Menu...................................................................23
4-2. How to Seal the Indicator .........................................................30
5. Set Mode
.............................................................................................31
5-1. How to Enter the Set Mode......................................................31
5-2. Descriptions on key operations in the Set Mode ....................31
5-3. Set Menu Descriptions (F00 ~ F82) ........................................32
5-3-1. General Function...................................................................................
35
5-3-2. RS-232 (Serial Communication) Function........................................
41
5-3-3. Print Function ........................................................................................
46
5-3-4. Checker Function..................................................................................
51
5-3-5. Other Functions.....................................................................................
52
6. Test Mode
...........................................................................................53
6-1. How to Enter the Test Mode ....................................................53
6-2. Test Menu ..................................................................................53
6
7. System Mode
....................................................................................58
7-1. How to Enter the System Mode ...............................................58
7-2. PCS Mode..................................................................................60
7-3. Percent Mode.............................................................................62
8. General Function Descriptions
...............................................64
8-1. Item Number Input Method......................................................64
8-2. Key Tare Input Method.............................................................64
8-3. How to Check Subtotal, Total and Weighing Count...............65
8-4. How to Enter High Limit..........................................................66
8-5. How to Enter Low Limit...........................................................66
9. Weighing Mode
...............................................................................67
9-1. Zeroing Function – LED...........................................................67
9-2. Tare Function – LED.................................................................67
9-3. Hold Function – LED................................................................68
9-4. Zeroing Function – LCD ..........................................................69
9-5. Tare Function – LCD ................................................................69
9-6. Hold Function – LCD ...............................................................70
10. Charge and Use Time
................................................................71
10-1. How to Use and Charge the Chargeable Batter.....................71
10-2. Use Time of the Battery..........................................................71
11. RS-232C Interface in Detail
...................................................72
11-1. RS-232C Port Connection ......................................................72
11-2. How to Connect Serial Communication Devices .................73
11-2-1. How to Connect an Auxiliary Display.............................................
73
11-2-2. How to Connect a Label Printer .......................................................
73
11-3. RS-232 Communication Protocol..........................................73
11-3-1. 22 Bytes for CAS................................................................................
73
11-3-2. 10 Bytes for CAS................................................................................
74
11-3-3. A18 Bytes for AND............................................................................
74
12. Error Message
.............................................................................76
12-1. Error Message from the Weight Setup Mode.............................76
12-2. Error Message from the Weighing Mode...................................77
7
Preface
Thank you very much for purchasing CAS International Indicator.
This produce is characterized by the excellent performance and luxurious features through strict examinations, as well as elaboration for each part under our strict quality control.
CA Indicator (CI series) is a product with rich functions and various external interfaces, which is designed to comply well with special requirements in a variety of industrial fields under strong and beautiful designs in appearance.
In addition, it is designed for the user-friendly programs for the easier use of indicator by any user with the built-in message display functions to help users understand the product.
Please use the product right and sufficiently utilize functions of CI-200 series as you read this manual thoroughly before using CI-200 series.
8
1. Features
1-1. Features
Suitable for the platform and bench type scale and weighing system
Easy operations
Simple and prompt full digital calibration (automatic weight setup at once)
Weight backup functions [restoring weight at the power supply On/Off]
Multiple weights setup functions [5 point input weight setup]
Command mode functions [PC control functions - data request and setup]
6 line [basic] / 4 line load cell Input
Front panel key lock
User message output functions
High & low limit, zero, OK signal output functions (only for LCD, SC)
System functions [count, percent, summation] (only for LCD)
Tare input functions using key
Gravity calibration functions
1-2. Major Functions
Various printer connection supports [roll DEP & label DLP printer]
Free to set the maximum weight and a division value as a user desires
Independent zeroing functions
Built-in hardware test functions
9
1-3. Analog and A/D Conversion
Applied voltage for load cell
Zeroing range
Input sensitivity
Non-straightness
A/D internal resolution
A/D external resolution
A/D conversion speed
Weight setup
DC 5V (350Ω maximum 8 possible connections)
0 ~ 2mV/V
0.5 uV / D (OIML, )Ntep, KS
0.5 uV / D (Non OIML, )Ntep, KS
0.01% Full Scale
1 / 520,000
1 / 10,000 (NTEP, )OIML, KS
1 / 20,000 (Non NTEP, )OIML, KS
(Possible with the use of sufficient output at 2mV/V L/C)
Maximum 80 rounds/second
Full Digital Calibration : SPACTM
(Automatic weight setup at once)
1-4. Digital and Display
* Communication (RS 232/422) ensures the free setup of independent use.
Weight display
CI-200A,
CI-200S, CI-200SC
CI-201A
LED (6 digit)
LCD (6 digit + Sign)
Character size
Sign below zero point
CI-200A
CI-201A
"-" minus sign
25 mm (Height)
24 mm (height)
Sign for status
ZERO, TARE, NET, STABLE, HOLD,
UNIT(kg)
10
1-5. General Specifications
AC Adapter
AC 100~240 V (DC 12V, 1.25A)
Operating temperature -10℃ ~ 40℃
Product size
Product weight
CI-200A
CI-201A
CI-200S
CI-200SC
CI-200A
CI-201A
CI-200S
CI-200SC
139mm(H) x 206mm(L) x 91.05mm(W)
169.5mm(H) x 250mm(L) x 83mm(W)
About 1.3kg
About 1.5kg
1-6. Communication and Option
Basic
Optional
COM1 (RS-232 Printer & PC Interface )
COM2 (RS-232 Printer & Auxiliary Display)
RS-485 Multi Drop Interface
* COM2 can be selectively used for a printer (RS-232).
11
2. Specifications in Appearance
2-1. External Dimension (CI-200A, CI-201A)
12
(CI-200S, CI-200SC)
13
2-2. Front Panel Descriptions
CI-200A
CI-201A
14
-
Tare
Communicat ion
-
-
-
(1) Main Display (Weight Display)
A. Displaying the value of gross or net weight.
B. Displaying error messages for any abnormal motion or weigh setup error/
C. Displaying the status value for the Set Mode and weight setup mode.
(2) Status Display (Lamp)
LED Lamp LCD Status Display Descriptions
Stable
Net weight
Zero point
Hold
-
-
The weighed weight is stable.
The current display of weight is a net weight.
The current weight is 0 kg.
The current status is under hold.
Displayed when the battery should be charged (chargeable battery).
◁
High limit The weight is heavier than the upper limit.
Low limit
Normal
The display of lower limit is lit if the value set at F50 is smaller than the lower limit, or greater than the lower limit or smaller than the upper limit.
The weight is greater than the lower limit and smaller than the upper limit.
The current status is at the tare status.
◁
Summatio n lamp
Quantity lamp
Percent lamp
The current status is at the communication status.
The current weight is the value of summation.
The current mode is at the count mode.
The current mode is at the percent mode.
15
CI-200S
CI-200SC
16
(1) Main Display (Weight Display)
A. Displaying the value of gross or net weight.
B. Displaying error messages for any abnormal motion or weigh setup error/
C. Displaying the status value for the Set Mode and weight setup mode.
(2) Status Display (Lamp)
Stable
Tare
The weighed weight is stable.
The current status is at the tare status.
Net weight
Hold
Zero point
The currently displayed weight is a net weight.
The current status is under hold.
The current weight is 0 kg.
Communication
Kg
The current status is at the communication status.
The current weight unit is set to kg. lb
Lack
The current weight unit is set to lb.
The weight is less than the setup value. (SC Only)
Fixed Quantity The weight is within the setup range. (SC Only)
The weight is greater than the setup value. (SC Only) Excess
17
(3) Keyboard
Function Key
* Some functions can be defined to the needs.
(The function set at F17 in the Set Mode will be operated.)
* Some functions can be defined to the needs.
(The function set at F18 in the Set Mode will be operated.)
Number Key
* It enters 1 in the input mode.
* It sets the weight display near zero point to 0.
( A range of 2%, 5%, 10%, 20% and 100% can be selected.)
* Long press to enter the test mode.
* It enters 2 in the input mode.
* Each press after setting up the tare displays the gross weight and the net weight in turn.
(The displayed weight is the net weight when the net weight lamp is on, but the displayed weight is the gross weight when the net weight lamp is off.)
* Long press to enter the setup mode.
* It enters 3 in the input mode.
* Use it to weigh with the tare.
* The current weight is memorized as the tare by pressing the key.
* Press the key when the load tray is empty to release the tare.
* Long press to enter the system selection mode.
(CI-201A Only)
* It enters 4 in the input mode.
* Use it to check the subtotal (partial summation).
* Long press to enter the system weight setup mode.
(CI-201A Only)
* It enters 5 in the input mode.
* Use it to check the grand total (entire summation).
* It enters 6 in the input mode.
* Use it to check the weighing count.
* It enters 7 in the input mode.
* Use it for the manual print. (manual print key)
(Print format can be changed in the Set Mode.)
* It enters 8 in the input mode.
* Use it to fix the shaking weight.
18
* It enters 9 in the input mode.
* Use it to check the remaining capacity of battery.
* Use it to correct any wrong input while entering data.
* Use it to enter a decimal point (.) in the weight setup mode and weighing mode.
* It enters 0 in the input mode.
* Use it to register an item number. (0 ~ 19)
Double Key
+
+
+
* Use it to save the current status and exit from the weight setup mode, Set Mode and test mode.
* Use it to check the current weight value in PCS and percent mode.
(CI-201A Only)
* Use it to print the subtotal.
* Use it to print the grand total.
* Use it for the tare key.
* If the tare is known, enter it using the numeric keys.
(If the remaining value occurs when the input value is divided into the minimum unit, the value is rounded and entered.)
The key tare function cannot be used during the PCS and percent functioning.
19
2-3. Rear Panel Descriptions
CI-200A, CI-201A
SEALING
(CAL S/W)
POWER
Use it to set the weight (calibration).
* Use it for the power supply.
LOAD CELL A port to connect load cell.
RS-232C
F/G
Serial Com 1 and Com 2 port (connect PC or printer)
It is a terminal for grounding to improve electric noises, which is connected to the grounding line upon any abnormalities in the product.
(If the grounding terminal of the product is not connected, it might cause failures.)
20
CI-200S, CI-200SC
SEALING, F/G RS-232C LOADCELL POWER
SEALING
(CAL S/W)
POWER
Use it to set the weight (calibration).
* Use it for the power supply.
LOAD CELL A port to connect load cell.
RS-232C
F/G
Serial Com 1 and Com 2 port (connect PC or printer)
It is a terminal for grounding to improve electric noises, which is connected to the grounding line upon any abnormalities in the product.
(If the grounding terminal of the product is not connected, it might cause failures.)
21
3. How to Install
3-1. How to Connect Load Cell
Connect the load cell connector to the load cell port on the back of the indicator.
* How to connect the load cell to the connector.
Pin
Number
Pin Function
1 EXC+
6 SEN+
2 EXC-
7 SEN-
3 SIG+
4 SIG-
5 SHIELD
Note 1. When 4 line load cell is used, connect EXC+ and SEN+ to '+' power supply terminal in the load cell input, and connect EXC- and SEN- to '-' power supply terminal in the load cell input.
* Relationship between the load cell output and input sensitivity.
The input sensitivity of this product is maximum 0.2uV/digit or more.
The following equation should be satisfied upon the system design.
Applied voltage of load cell x Output voltage of load cell x Value of a division
0.2uV ≤ ------------------------------------------------------------------------------------------------
Rated capacity of load cell x Number of load cell
Example 1) Number of load cell: 4 ea
Rated capacity of load cell: 500 Kg
Rated output of load cell: 2mV/V
Value of a division: 0.10 Kg
Applied voltage of load cell: 10V (= 10,000 mV)
According to the equation (10000 mV* 2mV * 0.1Kg)/(500Kg * 4) = 1 ≥ 0.2uV
As the calculated value is greater than 0.2uV, this weight system design has no problem.
22
4. Weight Setup (Calibration) Mode
What is the weight setup?
It refers to the calibration to set the displayed value to the actual weight in displaying weights.
How to Access to the Weight Setup Mode
Turn on the power supply to access to the weight setup mode while pressing Cal S/W after removing the sealing. Press the setup key long in the weight setup mode to return to the weighing mode.
4-1. Weight Setup Menu (CAL1 – CAL9)
CAL 1: Maximum capacity
CAL 2: Minimum division and decimal position setting
CAL 3: Weight calibration
3-1. Setting the range of multiple calibration
3-2. Zero calibration
3-3. Setting weight
3-4. Span calibration
CAL 7: Gravity adjustment
CAL 8: Zero adjustment
CAL 9: Factor calibration
CAL 10: Setting dual range
23
CAL 1
(CAL 1 automatically starts.)
Function: Setting Maximum Value
Range of set value: 1 ~ 99,999
Used key
:Save and next
Menu navigation
~ : Set value change
:End
Display
C= 10000
C= 10
Descriptions
Max. value = 10000kg
Max. value = 10kg
Note 1. It means the maximum weight value to be weighed by the scale.
CAL 2
Function: Minimum division and decimal position setting
Range of set value: 0.001 ~ 9999
Used key Display Descriptions d= 0.001 Minimum division 0.001 kg
:Save and next
Menu navigation
~ : Set value change
:Set decimal point and end d= 0.01 d= 0.1 d= 1
Minimum division 0.01 kg
Minimum division 0.1 kg
Minimum division 1 kg d= 10 Minimum division 10 kg
Note 1. To end CAL2, press key when a decimal point is set.
Note 2. The minimum division means the value of a division.
Note 3. Set the external resolution within 1/30,000 as the value by dividing the maximum weight by the minimum division.
If the external resolution is 1/30,000 or more, Err 21 is shown.
Note 4. The position of a decimal point is decided by the position of a decimal point for the minimum division set in CAL2.
Note 5. If the minimum division is set to any value out of 1, 2 and 5 unit, "ERR DIV" is shown.
24
3 CAL
CAL 3-1
Function: Setting Multi Calibration Step
Range of set value: 1 ~ 5
Used key Display Descriptions
:Save and next
Menu navigation
~ : Set value
:End change
STEP- 1
STEP- 3
STEP- 5
Setting multi calibration for step 1
(CAL3-3 and CAL 3-4 are carried out once)
Setting multi calibration for step 3
(CAL3-3 and CAL 3-4 are carried out three times.)
Setting multi calibration for step 5
(CAL3-3 and CAL 3-4 are carried out five times.)
* If the actual curve of load cell is a straight line, set the range of weight setup to 1.
* A function to use, when the output of load cell is corrected by setting multiple points in some sections because the actual curve of load cell is not a straight line.
mV/V
Real curve
Multi-point
Multi-point (full span)
Multi-point
Multi-point
Ideal
Multi-point (zero span) weight
Zero and Span points to interpolate weight from Load Cell
25
CAL 3-2
Function: Zero Calibration
Used key
:Zeroing
:End
Display
UnLoAd
1234
Descriptions
Empty the load tray and press the setup key.
The current weight value is displayed.
Confirm 'Stable' and press the setup key.
- - - Zeroing in progress...
Note 1. If zeroing finished with no error, it moves to Setting Weight (CAL 3-3) although no key is pressed.
Note 2. When zero point is too low, an error message "ERR27" is displayed.
Note 3. When zero point is too high, an error message "ERR26" is displayed.
CAL 3-3
Function: Setting Weight
Range of set value: 1 ~ 99,999
Used key
:Save and next
Menu navigation
~ : Set value change
:End
Display
LOAD 1
Descriptions
It means the weight setting mode.
(Number = multi calibration number)
W=100.00 100.00 (unit: Kg or Ton)
W= 0.10 0.10 (unit: Kg or Ton)
Note 1. Set the weight within a range of 10% ~ 100%.
Although 100% of the maximum weight is given as the initial value, enter the desired weight again if the
weight is different from the initial value.
(If the weight exceeds the maximum weight, "ERR 23" is displayed.)
If the weight is 10% or less, "Err 20" is displayed and if the calibration is set to 10% or less, the accuracy is lowered.
26
CAL 3-4
Function: Weight Calibration)
Used key Display Descriptions
:Span adjustment
:End
LoAd
12345
- - -
Load the weight set in CAL 4-3 and press the setup key.
The current weight value is displayed.
Confirm 'Stable' and press the setup key.
Span adjustment in progress...
Note 1. CAL 3-3 and CAL 3-4 are repeated as many as STEP is set in CAL 3-1.
At this time, the weight value should be set to a value greater than the previous one.
Note 2. Move to CAL-1 if the span adjustment is over with no error.
Note 3. When zero point is low, an error message "ERR24" is displayed.
Note 4. When zero point is high, an error message "ERR25" is displayed.
Note 5. After finishing calibration, press the setup key for 2 seconds or more to convert to the weighing mode.
CAL 7
Function: Gravity Adjustment
Used key
:Save and next
Menu navigation
~ : Set value change
:End
Display
G-CAL
Gr-CAL
9.XXXX
Gr-SET
9.XXXX
Descriptions
It means you accessed to the menu for the gravity adjustment.
Set the gravity for the production place.
Set the gravity for the place to use the product.
Note 1. If the gravity of the indicator production place is different from that of the place to use, the gravity adjustment can be done using this function.
27
CAL 8
Function: Zero adjustment - calibration when any zeroing error occurs.
Used key Display Descriptions
2-CAL
Empty the load tray and press the setup key.
:Zeroing
:End
1234
The current weight value is displayed.
Confirm 'Stable' and press the setup key.
- - - Zero adjustment in progress...
Note 1. Use this function when zeroing is not passed for any shock to the load cell.
The range of zero adjustment is 0 ~ 2mV/V.
Note 2. Move to CAL-1 if the zero adjustment is over with no error.
Note 3. When zero point is too low, an error message "Err27" is displayed.
Note 4. When zero point is too high, an error message "Err 26" is displayed.
CAL 9
Function: Factor Calibration
Used key
:Save and next
Menu navigation
~ : Set value change
:End
Display
NOT USE
FACtor
12345
Descriptions
This function cannot be used because of multi calibration.
It means you entered the factor correction mode.
The current factor is displayed.
Note 1. As this is a menu to set the weight setup with no weight, general users have no need to use it.
Note 2. This can be used only when the range of multi calibration in CAL 4-1 is set to 1.
"NOT USE" is displayed when the range of CAL 4-1 is set to 2 or larger.
Note 3. Enter a password to enter the factor correction mode.
28
CAL 10
CAL 10-1
Function: Setting Dual Range
Range of set value: 0 ~ 1
Used key
:Save and next
Menu navigation
~ : Set value change
:End
Display
DUAL- 0
DUAL- 1
Descriptions
Dual range function is not used.
Dual range function is used.
Note 1. If the resolution capability is 1/10,000 or higher, "OVER" message is displayed and return to the
CAL menu mode.
CAL 10-2
Function: Setting the applied section for the Dual Range
Range of set value: 0 ~ 99999
Used key Display
:Save and next
Menu navigation
Descriptions
M 1000 Dual range is applied to less than 1000kg.
~ : Set value change
:End
M 5000 Dual range is applied to less than 5,000kg.
M 10000 Dual range is applied to less than 10,000kg.
Note 1. If the input value is greater than the maximum value, "ERR SET" message is displayed and
returned to the CAL menu mode.
29
4-2. How to Seal the Indicator (Sealing)
After the calibration mode is carried out, proceed to the following step.
1. Tighten the CAL switch bolt.
2. Connect the sealing wire as shown in the picture.
3. Press the sealing wax as shown in the picture.
Sealing Point
30
5. Set Mode
5-1. How to Enter the Set Mode
Turn on the power while pressing
Or, Hold
key at the indicator front to start the Set Mode.
key for about 3 seconds to move from other mode to the convesion mode.
After finishing the setup in the Set Mode, press key for a long time
5-2. Descriptions on key operations in the Set Mode
~ : Use them to change the setup value.
: Save changes in the setup value and move to the higher menu
: Cancel the set value and move to the higher menu
31
5-3. Set Menu Descriptions (F00 ~ F99)
General Function
F03 (00) Auto Power Off
F04 (10) A/D Converting Speed
F05 (10) Digital Filter
F06 (00) Vibration Filter
F07 (02) Motion Detection Condition
F12
F13
F16
F17
F18
(00) Set Auto Hold Range
(10) Set Zero Range
Availability
(00) Set the Front Key Input to be Allowed
(00) Set "F1" Key
(00) Set "F2" Key
F21 (10) Set Initial Zero Range
F23
F24
(09)
(00)
Set Excessive Weight Check
Set Backlight Operational Condition (LCD)
F25 (03) Set LED Brightness or Backlight Brightness
* Note. Number in () is the default at the factory shipment.
32
RS-232 Serial Communication Function
F28 (04) COM1 Baud Rate
F30
F31
F32
(00) COM1 Output Format
(00) COM1 - Output Mode
(04) COM2 Baud Rate
F35 (00) COM2 - Output Mode
Print Function
F40
F41
(02) Set Printer in Use
(00) Set Print Format
F43
F44
(01) Print Line Feed
- User Print Message Input
F47 (01) Data Initialization after Summation Print
33
Checker Function
Set Mode Initialization
F99 - Set the Set value of Set Mode to the Factory Default
* Note. Number in () is the default at the factory shipment.
34
5-3-1. General Function
F01
Function Date Change
Numeric key Display Meaning
: assigning data 02.01.10 January 10, 2002
F02
Function Time Change
Numeric key Display Meaning
: assigning data 11.30.10 11 o’clock 30 minutes and 10 seconds AM
F03
Function Auto Power OFF
Display Meaning
Setting range
(00 ~ 30)
F03. 00
F03. 10
F03. 30
Not used.
Automatic power off after 10 minutes in the waiting mode.
Automatic power off after 30 minutes in the waiting mode.
Note 1. The power is automatically off if the defined time continues at the zero point after the automatic power off is set.
F04
Function Setting A/D Converting Speed
Display Meaning
Setting range
(00 ~ 99)
F04. 10
F04. 20
F04. 80
10 rounds/second
20 rounds/second
80 rounds/second
F05
Function Setting digital filter
Display Meaning
Setting range
(00 ~ 50)
F05. 10
F05. 30
Display of average for No. 10
Display of average for No. 30
F05. 50 Display of average for No. 50
35
F06
Function Setting vibration filter
Display Meaning
Setting range
(00 ~ 99)
F06. 00
F06. 10
F06. 99
Vibration filter OFF
Compensation for the vibration value of 5 divisions (0.5d * 10)
Compensation for the vibration value of 49.5 divisions (0.5d * 99)
Note 1. Apply this function to a place with heavy vibrations.
(The display response speed becomes slower when the vibration filter is applied.)
Note 2. This function should be adjusted appropriately to the site while the speed of weight variations in
F04 is being lowered little by little.
F07
F08
Function Setting Motion Detection Condition
Display Meaning
Setting range
(1 ~ 99)
F07. 1
F07. 2
F07. 10
The 'Stable' lamp is lit if the weight changes within 0.5 division.
The 'Stable' lamp is lit if the weight changes within 1 division.
The 'Stable' lamp is lit if the weight changes within 5 division.
Function Setting Automatic Zero Tracking Compensation
Display Meaning
Setting range
(0 ~ 9)
F08. 0
F08. 1
Automatic zero function is not used.
If it changes slowly to 0.5 divisions or less, it is compensated.
F08. 2
F06. 9
If it changes slowly to 1.0 divisions or less, it is compensated.
If it changes slowly to 4.5 divisions or less, it is compensated.
Note 1. This function compensates zero automatically if the weight at the zero point does not exceed the division in a certain range within a specific time.
Ex) If F08 is set to "4" when the maximum displayed division is 120.0kg and the value of a division is set to 0.05kg;
2.0digit
(0.1Kg)
0.75 Sec
If a greater change than 100g occurs within 0.75 seconds,, the changed
36
F09
Function Weight Backup Function
Setting range
(0, 1)
Display Meaning
F09. 0
F09. 1
Weight backup is not used.
Weight backup is used.
Note 1. As the backup state memorizes the initial status at zero for the weighing machine even during the blackout or when the power is turned off, the weight value is displayed if there is any weighing object in the weighing machine when the power is turned on.
If the weighing tray is empty, press the "ZERO" key to memorize the zero again.
F10
Function Set Hold Type
Display Meaning
Setting range
(0 ~ 3)
F10. 0
F10. 1
F10. 2
F10. 3
Ordinary hold: calculating the average of weights for shaking objects
Peak hold: calculating the maximum value for shaking objects
Sampling hold: calculating the sampling value for shaking objects
Automatic hold: automatically calculating the average weight of shaking objects
Note 1. If any load more than 'Over' is applied or at the zero, the hold is automatically released.
Note 2. Use automatic hold function, when you weight an animal or moving.
F12
Function Auto Hold Range
Display Meaning
Setting range
(0~99)
F12. 09
F12. 99
Auto hold range is 9 division
Auto hold range is 99 division
F13
Function Set Zero Range
Display Meaning
Setting range
(0~99)
F13. 2
F13. 10
The 'Zero' Key is operated within 2% of the maximum weight.
The 'Zero' Key is operated within 10% of the maximum weight.
F13. 99 The 'Zero' Key is operated within 99% of the maximum weight.
Note. Be aware that the load cell can be damaged if you set the value to F13=10% or more.
F14
Function ZERO and TARE Keys Availability
Setting range
(0, 1)
Display Meaning
F14. 0 Always operated.
F14. 1 Operated when the weight is 'Stable'.
37
F16
Function Set the front key input to be allowed.
Setting range
(0 ~ 1)
Display Meaning
F16. 0
F16. 1
The front keys are unlocked.
The front keys are locked.
Note 1. If it is set to 1, some function keys among the front keys cannot be used.
(Print, Hold, Tare, Step, Subtotal, Grand total, Weighing count, Item number, Setup, etc)
F17
Function Set the use of function key 1
Setting range Display Meaning
(0~15) F17. XX Set function key 1 to the key in the code table.
Note 1. Set the desired functions using <Table 1. Function Key Code>.
(LCD product = "11" and LED product = "0" as the default at the product shipment)
F18
Function Set the use of function key 2
Setting range Display Meaning
(0~15) F18. XX Set function key 2 to the key in the code table.
Note 1. Set the desired functions using <Table 1. Function Key Code>.
(LCD product = "12" and LED product = "0" as the default at the product shipment)
Table 1> Function Key Code Table
Function Name Key Code Function Name Key Code
Empty 00 Hold 08
Zero Point
Gross Weight * Net Weight
01
02
Battery
Item Number
09
10
Tare
Subtotal
Grand Total
Weighing Count
03
04
05
06
High Limit (LCD, SC Only)
Low Limit(LCD, SC Only)
Tare Lease
Piece Weight Value
(LCD Only)
11
12
13
15
Print 07
F19
Function Set the use of unit
Display Meaning
Setting range
(0, 1)
F19. 0
F19. 1
The unit is set the ‘kg’
The unit is set the ‘lb’
38
F21
Function Set the initial zero range
Display Meaning
Setting range
(02~20)
F21. 02
F21. 10
F21. 20
Set the initial zero up to 2% of the maximum weight
Set the initial zero up to 10% of the maximum weight
Set the initial zero up to 20% of the maximum weight
Note 1. Please consult with an engineer because setting 10 or larger value might affect the load cell greatly.
F23
Function Setting the range of check for the excessive weight (weighing unit)
Display Meaning
Setting range
(00~99)
F23 09
F23. 99
Excessive weight from the maximum weight + 9 divisions
Excessive weight from the maximum weight + 99 divisions
F24(CI-201A)
Function Backlight Operation
Display Meaning
Setting range
(0~5)
F24 0
F24 1
F24 2
F24 3
Backlight off
Backlight on when any key is operated.
Backlight on when the weight changes.
Backlight on when it is 'Stable' after the weight changes.
F24 4
F24 5
Backlight on when a key operates or the weight changes.
Backlight on all the time
Note. Although it is set to 5, press the power key shortly to turn off the backlight.
39
F25
Function Set Backlight and LED Brightness
Display Meaning
F25 1
F25 2
Set 10% of brightness
Set 30% of brightness
Setting range
(1~7)
F25 3
F25 4
F25 5
F25 6
F25 7
Set 50% of brightness
Set 60% of brightness
Set 70% of brightness
Set 90% of brightness
Set 100% of brightness
Note 1. Any value out of the setting range, the brightness will be set to '3'.
40
5-3-2. RS-232 (Serial Communication) Function
F26
Function Set Device ID
Setting range
(00 ~ 99)
Display Meaning
F26. 00
F26. 99
Device ID 00
Device ID 99
Note 1. This function enables to use the unique indicator ID in the command mode.
F27
Function Set Parity Bit – RS232C & PRT
Display Meaning
Setting range
(0 ~ 2)
F27. 0
F27. 1
F27. 2
Data bit 8, stop bit 1, parity bit: none
Data bit 7, stop bit 1, parity bit: even number
Data bit 7, stop bit 1, parity bit: odd number
Note 1. F26 and F27 apply commonly to 2 serial communications (RS23C and PRT).
41
Serial Communication COM1 Function
F28
Function Set COM1 Baud Rate
Display Meaning
F28. 0
F28. 1
600 bps
1200 bps
Setting range
(0 ~ 8)
F28. 2
F28. 3
F28. 4
F28. 5
F28. 6
F28. 7
F28. 8
2400 bps
4800 bps
9600 bps
19200 bps
38400 bps
57600 bps
115200 bps
F29
Function Set COM1 - Usage
Setting range
(0 ~ 1)
Display Meaning
F29 0
F29 1
Connect to a printer
Connect to a computer or auxiliary display
* If F29: 0 and F33 : 0, "ERR-Set" is displayed with no print.
F30
Function
Setting range
(0 ~ 2)
Set COM1 - Output Format
Display Meaning
F30 0
F30 1
22 bytes for CA
10 bytes for CA
F30 2 18 bytes for AND
F31
Function Set COM1 - Output Mode
Display Meaning
Setting range
(0 ~ 4)
F31 0
F31 1
F31 2
F31 3
F31 4
No data out
Transmission for both the stable and instable time (stream mode)
One time transmission after the weight is stabilized.
Transmission only if data is requested.
* Data request signal: device ID (F26) _ 1 byte communication
(Data on request: 1= 0x01, 10 = 0x0A)
Response to the data request - Command Mode
Set the value of F31 to '1' or more if the print mode is used.
42
Note 1. Command Mode Table
Data Request Signal of CI-200 Descriptions on CI-200
0 1 2 3 4 5
D
D
D
D
D
D
D
D
D
D
D
D dd dd dd dd dd dd dd dd
K Z CR
K T CR
K G CR
6
LF
LF
LF
7
K N CR LF
H
K B CR
K C CR
LF
LF
K W CR LF
8 9 10 11
Zero Point Key
Zero Point Key
Gross Weight Key
Net Weight Key
Print Key
Total Print Key
Weight Data Request Signal Received Data Return dd I 0 0 0 0 0 dd dd dd
H
H I
H L
0 0 0
0
0
0
0
0
0
0
0
CR
0 CR
0 CR
LF
LF
LF
Request Signal
Key Tare Value
High Limit(LCD Only)
Low Limit (LCD Only)
Output Signal
Received Data Return
Received Data Return
Received Data Return
Received Data Return
Received Data Return
Received Data Return
Received Data Return
Received Data Return
Received Data Return
Received Data Return
Received Data Return
Note 1. (D : 0x44, dd:00~99, K:0x4B , Z:0x5A , CR : 0×0D, LF: 0×0A) dd = Device Number (2byte), CR = 0×0D, LF: 0×0A
Ex) If a device number is 10, dd becomes 0x31 and 0x30.
Ex) If you want to operate the zero point key when a device number is 11, the indicator operates zeroing if the hex code of “ 44 31 31 4B 5A 0D 0A” is sent.
43
Note 1. NT-200 Command Mode Table
Command (ASCII Code) Description Status
HI
KT
High Limit
Key Tare Value
Read / Write
Limit Read / Write
Read / Write
CO Code Write
ZE Operation with ZERO Key Read
TR Operation Read
GN
ID
Operation with Gross/Net Key
Device Number (ID) Change
Read
Read
HD
PR
TP
Operation with HOLD Key
Operation with PRINT Key
Operation with Total Print Key
Read
Read
Read
Read
Device ID Command CR LF
Note 1. Device ID is hex value and Command is ASCII value.
[Ex] If Device ID is 13, a user wants to know the current weight value -> 0x0d
0x57 0x54 0x0d 0x0a
Write
DATA (Not include decimal point) Device ID Command
Format for Device ID Change
CR LF
Device ID Command DATA CR LF
Note 2. When you change code and device number, the data value is HEX 1byte.
44
Serial Communication COM2 Function
F32
Function Set COM2 Baud Rate
Display Meaning
F32 0
F32 1
600 bps
1200 bps
Setting range
(0 ~ 8)
F32 2
F32 3
F32 4
F32 5
2400 bps
4800 bps
9600 bps
19200 bps
F32 6
F32 7
F32 8
38400 bps
57600 bps
115200 bps
F33
Function Set COM2 - Usage
Setting range
(0 ~ 1)
Display Meaning
F33 0 Connect to a printer
F33 1 Connect to a computer or auxiliary display
* If F29: 0 and F33 : 0, "ERR-Set" is displayed with no print.
* COM1 and COM2 cannot be used together as the printer function.
F34
Function Set COM2 - Output Format
Display Meaning
Setting range
(0 ~ 2)
F34 0
F34 1
F34 2
22 bytes for CA
10 bytes for CA
18 bytes for AND
F35
Function Set COM2 - Output Mode
Display Meaning
Setting range
(0 ~ 2)
F35 0
F35 1
F35 2
No data out
Transmission for both the stable and instable time (stream mode)
One time transmission after the weight is stabilized.
Set the value of F35 to '1' or more if the print mode is used.
45
5-3-3. Print Function
F40
F41
Function Set a printer to use
Display Meaning
Setting range
(0 ~ 2)
F40 0
F40 1
Not used.
DLP (Label Printer)
F40 2
DEP (Roll Printer)
F42
Function Set print format
Display Meaning
Setting range
(0 ~ 2)
F41 0
F41 1
Set print format 0
Set print format 1
F41 2 Set print format 2
Function Set automatic print
Display Meaning
Setting range
(0, 1)
F42 0
F42 1
Manual print
Automatic print
Note 1. If the automatic print is set, print can be done with no press of print key when the weight is stable.
F43
Function Set Line Feed
Display Meaning
Setting range
(0 ~ 9)
F43 1
F43 9
1 Line feed
9 Line feed
46
【 Print Format 0 】 【 Print Format 1 】
Date, Time, Weighing No. (Item No.), Net Weight Date, Time, Weighing No. (Item No.), Net
Weight
2002. 1. 1 12:30
0001 ID_01: 50.0 kg
0002 ID_01: 100.0 kg
0003 ID_01: 200.5 kg
2002. 1. 1 12:30
0001 ID_01: 50.0 kg
2002. 1. 1 12:40
0002 ID_01: 50.0 kg
2002. 1. 1 12:50
0003 ID_01: 50.0 kg
【 Print Format 2 】
Date, Time, Weighing No. (Item No.), Net Weight
2002. 1. 112:30
No.0001 ID_01
Gross : 1000.0 kg
Tare : 0.0 kg
Net : 1000.0 kg
2002. 1. 112:40
No.0002 ID_01
Gross : 2000.0 kg
Tare : 500.0 kg
Net : 1500.0 kg
Note 1. If the power is turned off and then on, the number and total are initialized to 0001.
Note 2. The output of item number (ID_XX) depends on the setting in “F48”.
Note 3. The possible number for print is a range of 1~9999.
47
【 Total Print Format 】
Total Format
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
ID_01 TOTAL
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
2004.06.24 14:32:54
COUNT 22
WEIGHT 4500.05kg
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
GRAND TOTAL
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
2004.06.24 14:32:58
COUNT 123
WEIGHT 12500.10kg
Note 1. When a label printer (DLP-50) is used, the subtotal and grand total functions are not supported and Err-12 is displayed.
Note 2. After summation, data are maintained or initialized depending on the set value in F47.
48
CAS DLP Protocol
Variable Descriptions
V02
V03
V04
V05
Net Weight (8 bytes)
Barcode (Net Weight) (8 bytes)
Count in the Count Mode (8 bytes)
Percent in the Percent Mode (8 bytes)
The weight, count and percent cannot be printed at the same time.
Values that can be accurately printed are those for [weight, count and percent].
User’s Output Message Protocol
Command (ASCII code) Descriptions Status
The maximum length is 40 bytes. 0xFF should be put in the last byte.
20 bytes are printed in a line and the message starts from the upper left corner.
F44
Function Enter the user output message
Set Range
(32 ~ 255)
Display Meaning
12 - 065 Designate a character “A” equivalent to ASCII code 65 in the 12th data
00 - 032
18 - 255
To print out the added contents, designate ASCII code 32 to 0th data.
The end has to be meant by designating ASCII code 255 next to the last data.
49
~ : set number, : coordinate increase, : end entry
(If a coordinate increase is done when the input range exceed a range of 32 ~ 255, it will be cleared with
“255”)
Note 1. This function adds something to write down on the print format.
(Ex: company name, Phone number)
Note 2. Coordinates that can be designated have a range from 0 to 71, of which 0th data designates whether or not to print the added contents (032: printed, others: not printed). Accordingly, the actually printed contents will include contents from 1st data to the part right before the coordinate where data 255 is assigned.
Note 3. If you want to add the company name “CAS” to the existing print format, you might assign as follows;
P00-032 (ASCII code 32: data starts),
P01-067 (ASCII code 67: character C)
P02-065 (ASCII code 65: character A)
P03-083 (ASCII code 83: character S)
P04-255 (ASCII code 255: data ends)
F45
Function Set print output
Setting range
(0, 1)
Display Meaning
F45 0 Print on both the stable and instable time
F45 1 Print when the weight is stable.
F47
Function Initialize data after the summation is printed.
Setting range
(0, 1)
Display Meaning
F45 0 Maintain the status
F45 1
Initialize data after the summation is printed.
F48
Function Setting print item number
Setting range
(0, 1)
Display Meaning
F45 0 Not printing item number on print output
F45 1 Printing item number on print output
50
5-3-4. Checker Function
F50
Function Select the weighing mode (LCD, SC Only)
Display Meaning
Setting range
(0 ~ 2)
F50 0
F50 1
F50 2
Not used.
Use as the checker mode
Use as the limit mode
[CHECKER MODE]
Comm
Signal
Weight
(Low Limit) (High Limit)
0 kg 50 kg 100 kg
OUT
PUT
LOW
1
0
HIGH
1
0
OK
1
0
Note 1. All the outputs are generated regardless of the stable status.
[LIMIT MODE]
Comm
Signal
Weight
(Low Limit) (High Limit)
0 kg 50 kg 100 kg
LOW
OUT
PUT
1
0
HIGH
1
0
OK
1
0
Note 1. OK signal is displayed only for the stable status.
51
F51
Function Set Buzzer On/Off on the Checker Function (LCD, SC Only)
Setting range
(0, 1)
Display Meaning
F51 0
F51 1
General functions are operated as the buzzer.
Buzzer ON when the checker function is OK.
5-3-5. Other Functions
F90
Function Password Change
Display Meaning
Setting range
(0, 1)
F98. 0
F98. 1
Password not changed.
Password Changed
Password
Change
_ _ _ _
Good
_ _ _ _
PASS
_ _ _ _
Change
Enter the current password using numeric keys.
Enter a new password.
Enter the new password again.
F99
Function Set default
Setting range
(0, 1)
Display Meaning
0
1
No initialization functions for indicator.
Carry out the initialization functions for indicator.
Note 1. To set values to the same as the factory default for the indicator, press the setup key after setting F99 to 1.
52
6. Test Mode
6-1. How to Enter the Test Mode
Test mode starts when the power is turned on while pressing of the indicator.
Press the number for the test menu as you wish.
key in the front
To enter the weighing mode during test, press
6-2. Test Menu (TEST 1 - TEST10)
key for a long time.
Test 1: Key test
Test 2: Display test
Test 3: Load cell test and A/D conversion test
Test 4: RS-232 serial communication test (COM1, COM2)
Test 5: Printer test
Test 8: EEPROM test
Test 9: Battery test
Test 10: Clock (RTC) test
53
Test 1
Function: Key test
Used key
: Higher Menu
Other keys: Test
<Key List>
Display
1 1
Descriptions
When you press any key to test, the number and code for the key are displayed on the screen.
Key Number Code Key Number Code
1 1
Key
6 6
Number Code
0 0
2 2 7 7 70 30
3 3 8 8 28 28
9 9 29 29
11 27
4 4
5 5
Test 2
Function: Display Screen Test
Used key
: Higher Menu
Other keys: Test
Display Descriptions
An LCD lamp is on.
An LED lamp is on.
54
Test 3
Function: Load cell test and A/D conversion test
Used key
: Higher Menu
Display
XXXXXX
X.XX
Descriptions
The internal value for the current weight value is displayed.
The output value of the current load cell is displayed in mv/V.
Note 1. If key is pressed, the internal value of the current weight and the output of load cell (mv/V) are displayed repeatedly.
Note 2. Check this number to see if it moves well, while loading or unloading a weight to the
load tray. If the number is fixed or “0” is displayed, check the connection of load cell once again.
Test 4
Function: Serial Communication Test
Used key Display Descriptions
: Higher Menu
Other keys: Test
Tx -- Rx
------
05 --13
Status to wait for transmission or reception
Transmission: 5, Reception: 13
Note 1. Run this test while the communication program in the computer (ex: Hyper Terminal) is executing after connecting a serial port in the computer to the serial port on the back.
Note 2. Send ‘1’ from the computer keyboard, check whether or not ‘1’ is received properly on the indicator’s screen, and then check whether or not ‘1’ is received properly on the computer after pressing ‘1’ from the indicator’s keyboard.
55
TEST 5
Function: Printer Test
Used key Display Descriptions
: Higher Menu
Other keys: Test
No abnormality in the printer.
Check the connection of the printer connector..
Note 1. Designate a printer used in the Set Mode (F30) in advance.
Note 2. If the printer connection and the designation are done correctly, the following details
will be shown in the printer.
CAS Corporation
Come And Succeed
TEL 1577-5578
TEST OK
TEST 8
Function: EEPROM Test
Used key
: Higher Menu
Display
ROM OK
Descriptions
Displaying the status of EEPROM operation
56
Test 9
Function: Battery test
Used key
: Higher Menu
Test 10
Function: RTC Test
Display Descriptions b 6.15
Displaying the current voltage of battery (6.15V)
Used key
: Higher Menu
Display Descriptions
SEC XX
XX : Displaying the progress of seconds (SEC)
Note 1. If key is pressed, the current second changes to '0'.
57
7. System Mode
7-1. How to Enter the System Mode
Step
1
2
Display Screen and Key Input
Load
Tray
Descriptio ns
Press mode.
key for about 3 seconds in the weighing
Screen display: “1. PCS” characters are blinking after
“SYSTEM” is displayed.
Empty
key is pressed, “1. PCS” characters are blinking.
3
4
If
If
If
If
key is pressed, “2. PER” characters are blinking.
key is pressed, “3. WGT” characters are blinking.
key is pressed, the selected mode is set.
Select the mode to which you want to move.
58
Weighing Mode (I)
Initial Screen Descriptions
Weighing Mode
Counting Mode (II)
Initial Screen
Percent Mode (III)
Initial Screen
Counting Mode
Descriptions
Percent Mode
Descriptions
59
7-2. PCS MODE
7-2-1. PCS MODE Sample Input Method
Step
1
2
Display Screen and Key Input
Press key for about 3 seconds in the PCS Mode.
Screen display: “1.SAMPL” characters are blinking.
3
If
If key is pressed , “1. SAMPL” characters are blinking.
key is pressed , “2. WEIGH” characters are blinking.
Load
Tray
Descriptions
Empty
Select the input method as desired.
4
Press + keys.
5
Screen display: A/D value is displayed after “SAMPLE” ->
“LoAd” is shown.
(Wait until the weight is stable.)
Sample
Put samples on the load tray
6
7
Press key
Screen display: “SUCCES” -> “NUMBER” is displayed.
Sample
Sample
Save sample weight
8
9
Enter the number of samples using ~ keys, and then press key.
(Ex) If 10kg (sample) and 10 pieces, then the unit weight becomes 1kg.
Sample
Screen display: It moves to PCS Mode after displaying “End”. Sample
Note 1. The current weight is displayed when key is pressed during operating “1. PCS
MODE”.
Note 2. If the value of 1 PCS is smaller than 0.7 divisions of maximum resolution capacity when the number of samples is entered, Err-21 is displayed.
60
7-2-2. PCS Mode Direct Input Method
Step
1
2
Display Screen and Key Input
Press key for about 3 seconds in the PCS Mode.
Screen display: “1.SAMPL” characters are blinking.
Load
Tray
Descriptions
Empty
3
If
If key is pressed , “1. SAMPL” characters are blinking.
key is pressed , “2. WEIGH” characters are blinking.
Select the input method as desired.
4
Press + keys.
5
Screen display: After “WEIGHT” is displayed, “0.000 KG” is displayed.
Weight input mode
6
Enter the weight of PCS using ~ keys, and Save sample weight
7 then press key.
Screen display: It moves to PCS Mode after displaying “End”.
Note 1. If key is pressed during operations in the PCS MODE, it shows the current weight for 3 seconds and then returns to the PCS MODE.
Note 2. If the value of Piece Weight to a function key (F17 or F18), you may confirm the unit weight of 1 PCS.
61
7-3. PERCENT MODE
7-3-1. Percent Mode Sample Input Method
Step
1
2
Display Screen and Key Input
Press key for about 3 seconds in the Percent Mode.
Screen display: “1.SAMPL” characters are blinking.
Load
Tray
Descriptions
Empty
3
4
5
6
7
If
If key is pressed , “1. SAMPL” characters are blinking.
key is pressed , “2. WEIGH” characters are blinking.
Press + keys.
Screen display: A/D value is displayed after “SAMPLE” ->
“LoAd” is shown.
(Wait until the weight is stable.)
Sample
Select the input method as desired.
Put samples on the load tray
Press key
Screen display: “SUCCES” -> “NUMBER” is displayed.
Sample
Sample
Save sample weight
8
9
Enter the number of samples using ~ keys, and then press key.
(Ex) If 10kg (sample) and 10 pieces, then the unit weight becomes 1kg.
Screen display: It moves to Percent Mode after displaying
“End”.
Sample
Sample
Note 1. The current weight is displayed when key is pressed during operating in the
Percent Mode.
Note 2. If the value of 1 PCS is smaller than 0.7 divisions of maximum resolution capacity when the number of samples is entered, Err-21 is displayed.
62
7-3-2. Percent Mode Direct Input Method
Step
1
2
Display Screen and Key Input
Press key for about 3 seconds in the PCS Mode.
Screen display: “1.SAMPL” characters are blinking.
3
If
If key is pressed , “1. SAMPL” characters are blinking.
key is pressed , “2. WEIGH” characters are blinking.
Load
Tray
Descriptions
Empty
Select the input method as desired.
4
Press + keys.
5
Screen display: After “WEIGHT” is displayed, “0.000 KG” is displayed.
Weight input mode
6
7
Enter the weight of 100% using ~ keys, and then press key.
Screen display: It moves to Percent Mode after displaying
“End”.
Save sample weight
Note 1. If key is pressed during operations in the Percent MODE, it shows the current weight for 3 seconds and then returns to the PCS Mode.
Note 2. If the value of Piece Weight to a function key (F17 or F18), you may confirm the unit weight of 1 PCS.
63
8. General Function Descriptions
8-1. Item Number (Unique Number of Weighing Item: ID) Input
Method
Step Display Screen and Key Input
Load
Tray
Descriptions
1
Press key
Screen display: “ID = XX”
“Meaning the value of the current item number”
2
Enter a desired ID using number keys Input ID(=10)
3
Press key to save and exit
Note 1. Product ID has a range of 0 ~ 19.
8-2. Key Tare Input Method
Step Display Screen and Key Input
1
Press + keys
2
Screen display: “t = 0.000
Item
An item number is registered. The weight is displayed.
Load
Tray
Empty
Descriptions
Empty
“Meaning the value of the current item number”
3
Enter a desired ID using number keys
7
Press key to save and exit
Note 1. If the remainder occurs when the input value is divided by the minimum unit, it is rounded and entered.
64
8-3. How to Check Subtotal, Total and Weighing Count
Key Descriptions
The current subtotal (partial summation) is displayed.
The current total (entire summation) is displayed.
+
The current subtotal (partial summation) is printed.
Subtotal is erased after it is printed.
+
The current total (entire summation) is printed.
Total is erased after it is printed.
The current weighing count is displayed.
Note 1. While printing subtotal and total, an error (Err 12) is displayed with no connection to printer, and total and weighing count are erased.
1% unit of weight can be confirmed.
65
8-4. How to Enter High Limit
Step Display Screen and Key Input
Load
Tray
Descriptions
1 Press key .
Screen display: “H 0.000”
It means the current high limit.
Enter a desired value
2
Change the high limit using ~ keys .
3
Press key to save and exit.
Item
The weight is displayed after the high limit is saved.
Note 1. If the remainder occurs when the input value is divided into the minimum unit, the
value is rounded and entered.
8-5. How to Enter Low Limit
Step Display Screen and Key Input
Load
Tray
Descriptions
1 Press key .
Screen display: “L 0.000”
It means the current low limit.
Enter a desired value
2
Change the low limit. using ~ keys .
3
Press key to save and exit.
Item
The weight is displayed after the low limit is saved.
Note 1. If the remainder occurs when the input value is divided into the minimum unit, the value is rounded and entered.
Note 2. If the key code value of F17 and 18 was changed from the initial value, the key code
should be set again.
* F1 key's basic value is set to the high limit.
* F2 key's basic value is set to the low limit.
* If the weight is greater than the high limit, the "HI" lamp appears on the screen.
If the weight is smaller than the low limit, the "LO" lamp appears on the screen.
If the weight is smaller than the low limit, the "LO" lamp appears on the screen.
66
9. Weighing Mode
9-1. Zeroing Function (used when the zero point changes) - LED
■ Range of zero point: within a range set in F13
Zero chanced. Press Zero Key to set the zero lamp on and 0.
9-2. Tare Function (used for weighing with a container) - LED
■ Maximum tare set range: maximum weight
* Caution: the weight including the tare cannot exceed the maximum weight.
Put a container on the load tray.
(Container weight: 10kg)
Press the tare key.
(Tare is saved.)
Put an object on the load tray.
(Net weight: 20kg)
■ If you want to know the total weight;
Press the 'total * net weight' key (the value of object's weight + tare is displayed.)
■ If you want to know the net weight;
Press the 'total * net weight' key (the value of object's weight is displayed.)
Remove the container and object from the load tray to display the saved tare.
■ If the tare is removed;
Remove the container and object from the load tray, and press the tare key (picture on the right) if the saved tare is only displayed (picture on the left).
67
9-3. Hold Function (used for weighing moving objects) - LED
■ Ordinary Hold Function (hold function is performed when the hold key is pressed.)
Put an object on the load tray. Press the hold key. Hold message is displayed for a second.
The hold weight is calculated for about 3~5 seconds.
The hold weight is displayed.
It returns to the normal status and the hold lamp is turned off if you empty the load tray or press the hold key to release the held weight value.
■ Automatic hold function (the hold function is performed by automatically calculating
the maximum weight of moving objects.)
Empty the load tray. The weight on the load tray is displayed.
The maximum value is displayed with 'HOLD'.
■ It returns to the normal status and the hold lamp is turned off if you empty the load tray
or press the hold key to release the held weight value.
※ Note. The hold function carries out operations according to the set value of F10.
68
9-4. Zeroing Function (used when the zero point changes) - LCD
■ Range of zero point: within a range set in F13
Zero chanced. Press Zero Key to set the zero lamp on and 0.
9-5. Tare Function (used for weighing with a container) - LCD
■ Maximum tare set range: maximum weight
* Caution: the weight including the tare cannot exceed the maximum weight.
Put a container on the load tray.
(Container weight: 10kg)
Press the tare key.
(Tare is saved.)
Put an object on the load tray.
(Net weight: 20kg)
■ If you want to know the total weight;
Press the 'total * net weight' key (the value of object's weight + tare is displayed.)
■ If you want to know the net weight;
Press the 'total * net weight' key (the value of object's weight is displayed.)
Remove the container and object from the load tray to display the saved tare.
■ If the tare is removed;
Remove the container and object from the load tray, and press the tare key (picture on the right) if the saved tare is only displayed (picture on the left).
69
9-6. Hold Function (used for weighing moving objects) - LCD
■ Ordinary Hold Function (hold function is performed when the hold key is pressed.)
Put an object on the load tray. Press the hold key. Hold message is displayed for a second.
The hold weight is calculated for about 3~5 seconds.
The hold weight is displayed.
It returns to the normal status and the hold lamp is turned off if you empty the load tray or press the hold key to release the held weight value.
■ Automatic hold function (the hold function is performed by automatically calculating
the maximum weight of moving objects.)
Empty the load tray. The weight on the load tray is displayed.
The maximum value is displayed with 'HOLD'.
■ It returns to the normal status and the hold lamp is turned off if you empty the load tray
or press the hold key to release the held weight value.
※ Note. The hold function carries out operations according to the set value of F10.
70
10. Charge and Use Time
■ Charge the battery sufficiently when you use the product after storing it for a long time.
■ During the use of device, sign is shown (LCD) or ‘LOW BAT’ sign (LED) on the upper right corner, and then the power is turned off after a specific time.
When the power supply of battery reaches 5.6V, the battery alert lamp is turned on. When it reaches 5.2V, the power is automatically turned off.
■ When the battery alert lamp is turned on, charge the battery.
10-1. How to Use and Charge the Chargeable Battery
■ When an adapter is connected, a red light in the power supply lamp and another red light in the charge lamp are turned on.
When the charging is completed, a green light in the charge lamp is turned on.
■ The charging takes about 12 hours.
■ The complete charge mark is turned on if an adaptor is connected with no battery.
10-2. Use Time of the Battery
CI-200A
CI-200S
Condition
-
Use Time
About 30 hours
Backlight OFF About 180 hours
C-I201A(LCD)
Backlight ON About 33 hours
CI-200SC - About 26 hours
※ Note. The time stated above is subject to change depending on the period of battery use and the number of batteries.
To use the battery for a longer time, adjust the automatic power switch function in F03 and the brightness of display in F25.
71
11. RS-232C Interface in Detail
11
-
1. RS-232C Port Connection
(1) COM1 - RXD: Pin No. 2, TXD: Pin No. 3, GND: Pin No. 7
RXD 2 O
TXD 3 O
GND 7 O
----------------
O 2 Transmit Data
----------------
O 3 Receive Data
----------------
O 7 Signal Ground
┌─
O 8 Carrier Detect
├─
O 20 Data Terminal Ready
└─
O 6 Data Set Ready
┌─
O 4 Request to Send
└─
O 5 Clear to Send
9 pin port (male) 25 pin port (female)
RS-232C port of CI-200 Serial port of the computer
(2) COM2 - RXD: Pin No. 2, TXD: Pin No. 3 , GND: Pin No. 7 (Option)
RXD 2 O
TXD 3 O
GND 7 O
----------------
O 2 Transmit Data
----------------
O 3 Receive Data
----------------
O 5 Signal Ground
┌─
├─
└─
┌─
O 1 Carrier Detect
O 4 Data Terminal Ready
O 6 Data Set Ready
O 7 Request to Send
└─ O 8 Clear to Send
9 pin port (male) 9 pin port (female)
RS-232C port of CI-200 Serial port of the computer
72
11
-
2. How to Connect Serial Communication Devices
11-2-1. How to Connect an Auxiliary Display
TXD 3 O ----------------
O 3 Receive Data
GND 7 O ----------------
O 7 Signal Ground
9 pin port (male) 9 pin port (male)
RS-232C port of CI-200 Serial port of an auxiliary display
11-2-2. How to Connect a Label Printer (DLP)
RXD 2 O ---------------
---------------
O 3 Transmit Data
TXD 3 O
---------------
O 2 Receive Data
GND 7 O
O 5 Signal Ground
9 pin port (male) 9 pin port (male)
RS-232C port of CI-200 Serial port of DLP printer
Note. Refer to page 38 (Set Mode) for RS-232C communication and setting method.
11
-
3. RS-232 Communication Protocol
11-3-1. 22 Bytes for CAS
(1) Data bit: 8, Stop bit: 1, Parity bit: none
(2) Code: ASCII
(3) Set the time to send data to the computer in the Set Mode.
▣ Send all the time: if F30 and F35 are set to 1.
▣ Send when the weight is stable: if F30 and F35 are set to 2.
▣ Send upon data request: if F30 and F35 are set to 3.
▣ Only if the computer send 1 byte of the indicator’s device ID to the indicator, the indicator makes the defined output format.
73
(4) Transmission Data Format (22 bytes)
, , , Data (8 bytes) CR LF
└─┘ └─┘ └┐ └────┐ ┏──┘ └─┘
US (Unstable) GS (Gross weight) Device ID Lamp Status byte
Empty Unit (kg/t)
ST (Stable) NT (Net weight)
OL (Overload)
▣ Device ID: Send ing1 byte of device ID to selectively receive the information from the indicator to the receiver.
(Device ID is set in F26 .)
▣ Data (8 bytes): When the weight date including a decimal, for example, 13.5 kg, 8 bytes of ASCII code corresponding to 0’,‘0’,‘0’, ‘0’,‘1’,‘3’,‘.’and‘5’ are sent.
▣ Lamp Status Byte
Bt7
1
Bt6
Stable
Bt5
0
Bt4
Hold
Bt3
Printer
Bt2
Gross
Weight
Bt1
Tare
11-3-2. 10 Bytes for CAS
(1) Data bit: 8, Stop bit: 1, Parity bit: none
(2) Code: ASCII
(3) Transmission data format: (10 bytes)
Data (8 bytes) CR LF
11-3-3. 18 Bytes for AND
(1) Data bit: 7, Stop bit: 1, Parity bit: odd number / even number
(2) Code: ASCII
(3) Transmission data format (18 bytes)
, , Data (8 Byte)
└──┘ └──┘ └──┘
US (Unstable) GS (Gross Weight) Unit (kg/t)
ST (Stable) NT (Net Weight)
OL (Overload)
Bt0
Zero Point
CR LF
74
RS-422 & 485 Serial Communications (COM2)
RS-422 & 485 transmit signals with the voltage difference, which are more stable for electric noises than other communication methods.
In addition, the AC Power Cable or other electric wires should be placed separately, and the shield cable (0.5Φ or more) dedicated to communications should be applied.
The recommended use distance is within 1.2km.
▶ Setting output method
The same as RC232C before
▶ Signal Format and Data Format
The same as RC232C before
- 422 Connection Diagram -
- 485 Connection Diagram -
Note. RS-422 & 485 communication is optional specifications, which are supported through COM2.
Refer to page 42 (Set Mode) for the setup.
75
12. Error Message
12-1. Error Message from the Weight Setup Mode
Error Cause
Err 20
The resolution was set in excess of the tolerance 1/10,000.
Solution
Lower the resolution.
As the resolution = maximum tolerance / value of one division, adjust the resolution to 1/10,000 or less by correcting either the maximum allowable weight in CAL 1 or the value of one division in CAL3 in the weight setup mode.
Err 21
The resolution was set in excess of the tolerance 1/30,000.
Lower the resolution.
As the resolution = maximum tolerance / value of one division, adjust the resolution to 1/30,000 or less by correcting either the maximum allowable weight in CAL 1 or the value of one division in CAL3 in the weight setup mode.
Err 22
Err 23
The weight for the span adjustment was set to less than 10% of the maximum capacity.
The weight for the span adjustment was set to more than 100% of the maximum capacity.
Set the weight to 10% or more of the maximum capacity (set in
CAL 1) from CAL 4 in the weight setup mode.
Set the weight within the maximum capacity (set in CAL 1) from
CAL 4 in the weight setup mode.
Err 24
Err 25
Too low span.
Too high span.
Set the weight again by lowering the resolution as the setting of the current resolution is not possible because of either abnormality or lower output in the load cell.
Two low weight for PCS and percent sample.
There is either any abnormality or too high output in the load cell.
Execute steps from the zeroing step in CAL 4 in the weight set up again.
Two high weight for PCS and percent sample.
Err 26
Too high zero point.
Check whether or not the load tray is empty.
Retry the weight setup after check at the test mode 3.
Err 27
Too low zero point.
Err 28
Weight is shaking.
Set the weight setting again after confirming what force is given to the load tray of the scale in the test mode 3.
Check the connection of the load cell connector.
76
12-2. Error Message from the Weighing Mode
Error Cause Solution
Err 01
The initialization of the scale cannot be done because of the shaking weight.
Turn on the power after placing the scale at a flat place with no vibration.
Err 02
Either the connection of load cell is wrong or there is abnormality in the A/D conversion section.
Check the connection between the load tray and the body.
Err 05
Either you are pressing a key for a long time or there is abnormality in the key section.
Make an inquiry to A/S.
Err 08
The zero key, tare key and start key were disabled at the instable weight.
Set the zero key, tare key and start key to the proper user conditions at F14 in the Set Mode.
Err 09
The current weight is out of the range of zero point.
Set the range of operations for the zero key to within
2% or 10% at F13 in the Set Mode.
Err 10
The tare to set is out of the maximum weight of the scale.
Set the tare to less than the maximum weight.
Err 12
The type of the configured printer is one that cannot support the total print.
DLP printers cannot make the total print.
Set "F40" to '2' when DEP printers are used.
Err 13
The set value of zero point on the weight setting is out of range.
Check the status of the load tray and set the weight again.
Err 15
The range exceeded during setting the item code in the command mode.
Check the range of item code.
Err 82
There is abnormalities in the A/D set section.
. Make an inquiry to A/S.
Over
The current weight on the load tray is too heavy and out of the allowable tolerance.
Avoid any weight in excess of the maximum allowable limit on the scale.
If the load cell is damaged, it should be replaced.
77
Descriptions on Abbreviation on the Display
Abbreviation Descriptions Abbreviation
"UnLoad"
Descriptions
Unload the load tray
"PASS" Enter Password "LoAd" Load a weight
"Discord" Re-enter Execution
""CAL
"SET"
Weight Set Mode
Set Mode
"SyS"
"PCS"
System Mode
PCS Mode
"TEST" Test "Per” Mode
"OUEr"
Exceeding Maximum
Load
Appendix 1. ASCII Code Table
Character Code Character Code Character Code Character Code Character Code Character Code
Space 32 0 48 @ 64 P 80 ` 96 p 112
! 33 1 49 A 65 Q 81 a 97 q 113
“ 34 2 50 B 66 R 82 b 98 r 114
# 35 3 51 C 67 S 83 c 99 s 115
$ 36 4 52 D 68 T 84 d 100 t 116
% 37 5 53 E 69 U 85 e 101 u 117
& 38 6 54 F 70 V 86 f 102 v 118
‘ 39 7 55 G 71 W 87 g 103 w 119
( 40 8 56 H 72 X 88 h 104 x 120
) 41 9 57 I 73 Y 89 i 105 y 121
* 42 : 58 J 74 Z 90 j 106 z 122
+ 43 ; 59 K 75 [ 91 k 107 { 123
- 45 = 61 M 77 ] 93 m 109 } 125
. 46 > 62 N 78 ^ 94 n 110 ~ 126
78
MEMO
79
MEMO
80
MEMO
81
82
83