SIGNAL SOURCE 7016
SIGNAL SOURCE 7016
SIGNAL GENERATOR
Constant voltage,
constant current,
pulse generation and measurement
The SIGNAL SOURCE 7016 incorporates high-performance DMM functions into a handy signal
generator to provide a convenient, multi-function calibrator at a low price. In addition to
generating constant voltage and constant current, it generates pulse signals such as the voltage
pulses used for calibrating flow meters and similar devices. Along with its many signal generator
functions, the substantial stand-alone DMM measurement functions of Model 7016 include AC
and DC voltage and current, resistance, frequency and temperature, and continuity checking.
Moreover, with the optional Communication Package 3856-02, measured values can be sent to a
PC, which can also control functions such as source voltage settings and measurement range
selection. The 7016 gives you multiple functions, high precision and high performance at low cost
in an instrument that is suitable for use in the laboratory as well as in the field.
1
As a DMM
As a signal generator
● DCCV [±1.5000 V/±15.000 V range]
● DCCA [±25.000mA range]
● PULSE [0.5 Hz to 4800 Hz, 5 V/12 V/±5 V/±12 V]
● Other Standard Features
Bipolar sink/source
generation, pulse
generation with
variable duty ratio,
pulse width, and
amplitude, memory
generation, scan
generation, and
ramp generation
function
Pulse signal generation for
calibration of flow meters
■ Convenient pulse source for calibrating flow
meters, as well generation of constant current
and constant voltage
Meters that use pulse output as sensor signals, such as flow
meters, can easily be calibrated in the field by using the
7016 as a pulse generator to supply reference signal input.
Its ability to generate constant voltage and constant current
in the range from 1-5 V and 4-20 mA makes it ideally suited to a variety of maintenance needs, such as calibration of
equipment instrumentation in the 1-5 V/4-20 mA range.
Pulse input
■ Bipolar output expands
test utility
Ability to function both as source and
sink makes the 7016 well suited for
signal loop testing in instrumentation
systems or testing charge/discharge of
secondary batteries.
■ Up to 16 steps of memory
scan output
Memory scan output allows the 7016
to quickly accommodate calibration
requirements that involve repetitive
checks.
● DC/ACV [50 mV to 250 V range]
● DC/ACA [50 mA to 500 mA range]
● OHM [500 Ω to 50 MΩ range]
● FREQ [measurement range 1 Hz to 200 kHz]
● Continuity check
● Diode check
● Temperature
[-40˚C to 1372˚C]
[-40˚F to 2502˚F]
● AC+DC RMS
measurement of
voltage and current,
1 ms peak hold
A high-performance,
function
multi-function DMM
■ High resolution, high accuracy and
advanced measurement functions
The 7016 achieves unparalleled performance for a handy
DMM, providing DC voltage measurement accuracy of ±0.03%
rdg. ±5 dgt. (excluding 50 mV range), with display switchable to
51000 count. Also, in addition to the basic measurements of DC
voltage, DC current, AC voltage, resistance, diode, and grounding, this multi-function instrument also supports frequency and
temperature measurement.
■ AC+DC measurement function provides RMS
measurement of full- and half-wave rectified
waveforms
+
0
_
The 7016 can measure RMS values
of full-wave and half-wave rectified
waveforms used in household electrical equipment.
■ 1ms Peak Hold Function maintains
maximum/minimum peak values
The waveform peak values can be acquired and the crest
factors calculated from measurement
of the instantaneous peak value and
calculated true RMS value.
Peak value
Crest factor = Peak value/effective value
■ Temperature measurement function
Measuring temperature by connecting the 7016 to the
optional 9180-9183 or 9472-9476 temperature probes.
■ Signal generator and measurement
functions can be used simultaneously
to measure input/output insulation
■ 3-way power
supply for use in
any location
When measuring insulation between inputs and outputs, the 7016 can be simultaneously used as both a
signal generator and DMM without compromising the
functionality of either.
A dual display makes it
easy to simultaneously
check inspection results
for both inputs and outputs.
Power the 7016 with a
choice of 3 different power
supplies for easy use
regardless of your location: AA alkaline batteries, Ni-MH battery pack,
and AC adapter.
■ A wide variety of accessories
A wide variety of accessories, such as an AC adapter,
Ni-MH battery and three types of test leads are provided with the 7016 as standard features. The 7016 is
also equipped with a carrying case for transporting the
unit together with all of its accessories.
2
■ Full data transfer compatibility included as
a standard feature
Communication Package Import Screen
(Dedicated cable and communication software sold separately)
● Functions
The optional COMMUNICATION PACKAGE 3856-02 consists
of a dedicated USB cable and software for transferring measurement data and control signals to and from a PC. Imported data
can be stored in text format on the PC, enabling efficient data
management using commercial spreadsheet software.
• Change settings,
record measurements
• Data points: up to 65,525
• File operations: open/save
• Printing: data printing
• Recording: text format
• Interface: USB
● Operating environment
COMMUNICATION PACKAGE 3856-02 (USB)
■ Generation Range and Accuracy
Accuracy is guaranteed at 23±5˚C and at 80% rh or less after 5 minutes warm-up.
In other conditions, add ±(50 ppm setting + 0.5dgt.)/˚C.
● DC constant voltage generation (CV)
Range
Resolution
Accuracy
0.1 mV
±1.5000 V
±(0.03% setting +3 dgt.)
1 mV
±15.000 V
Remarks
sink/source output
max. output: ±25 mA
■ Load regulation: 0.012 mV/mA
■ Maximum input voltage: ±30 Vdc
● DC constant current generation (CC)
Range
Resolution
Accuracy
±25.000 mA
1 µA
±(0.03% setting +5 dgt.)
* Windows 98/ME/2000/XP are registered trademarks of Microsoft Corp., USA
● Pulse generation (PULSE)
Parameter
Output range
Resolution
0.5, 1, 2, 5, 10, 15, 20,
25, 30, 40, 50, 60, 75,
80, 100, 120, 150, 200,
Frequency 240, 300, 400, 480, 600,
800, 1200, 1600, 2400,
4800 Hz
Duty ratio
Remarks
sink/source output
max. output: ±12 V
■ General Specifications
: DC constant voltage, DC constant current, pulse generation
: AC voltage, DC voltage, AC+DC voltage, AC current,
DC current, AC+DC current, resistance, diode,
continuity, temperature, frequency, duty ratio, pulse
width measurement
: Bipolar sink/source output
Output method
AC measurement method : True RMS
Additional functions : Settable duty ratio, pulse width and amplitude pulse
generation, memory generation (16 memory data
settings per range), scan generation (single/continuous); ramp generation, AC+DC RMS voltage/current
measurement, 1-ms peak hold (for voltage/current
measurement), recording, data hold/refresh hold,
relative display, 4-20 mA current-loop percentage
display, 0-20 mA percentage display, RS-232C data
communications, power-on option.
: Full auto or manual
Range selection
: LCD with backlight
Display device
: Two 5-digit numeric digital displays (for generation
Display contents
and measurement functions, one large main display
and one small sub display)
Max. measurement count : 51,000 counts
: Settable 0 to 99 minutes in 1-minute intervals
Auto power off
Battery charge state : Warning on LCD when battery voltage falls below 9V
: 3/s (except AC+DC and frequency measurement)
Sampling rate
1/s for AC+DC and frequency measurement
0.25 - 4/s for duty ratio and pulse width measurement
Noise susceptibility : NMRR DCV; - 60dB or more(50/60 Hz)
CMRR DCV; - 90dB or more(50/60 Hz)
Pulse width output
voltage
f
Accuracy
–
±(0.005% setting + 0.01Hz)
0.390625%
±(0.01% setting + 0.2%) *1
1
256 × f
±(0.01% setting + 0.3 ms)
5 V, 12 V
–
±(2% setting + 0.2 V)
±5 V, ±12 V
–
±(2% setting + 0.4 V)
0.39 % to 99.60 %
1
■ Load regulation: 1µA/V
■ Maximum input voltage: ±30 Vdc
Generator functions
Measurement
functions
PC requirements: At least 200-MHz Pentium running Windows 98SE/ME/2000/XP* /
RAM: At least 128 MB / Display: At least 800 × 600 SVGA /
Hard disk: At least 40 MB free space
× (0.39 % to 99.60 %)
■ Duty ratio and pulse width should be set so that positive-polarity pulse width is at least 50 µs.
■ Duty ratio, pulse width and output voltage accuracy specifications are based on at least 50 µs
positive-polarity pulse width.
■ Maximum applied voltage: ±30 VDC
*1. At 1 kHz or more, add ±0.1%/ kHz of setting.
Withstand voltage
: [Case]-[Combined power supply terminals]
[Case]-[Combined output terminals]
[Combined output terminals]-[Combined power supply terminals]
510-Vrms 50/60-Hz sine wave for one minute
[Case]-[Combined input terminals]
[Combined power supply terminals]-[Combined input terminals]
[Combined output terminals]-[Combined input terminals]
2.3-kVrms 50/60-Hz sine wave for one minute
Operating temp. & humidity : 0 to 40˚C, 0 to 80% rh (non-condensating)
Storage temp. & humidity : -20 to 60˚C, 0 to 80% rh (non-condensating, w/o batteries)
: Indoors, below 2,000m altitude
Operating location
: 1.5V AA-size alkaline batteries (LR6) × 8
Power supply
1.2V Ni-MH batteries × 8 (supplied)
AC Adapter (Model SA-141A0F-11 supplied, for 100
to 250 VAC, 47 to 63 Hz)
Maximum rated power : 5 VA
Continuous operation : 20 h or more (measurement only),
4 h or more ( generation and measurement)
(with supplied Ni-MH batteries, new and after full charge)
: 8 h or more
Charging time
Dimensions and mass : Approx. 90W × 192H × 54D mm, 735 g (instrument only)
Conforming standards : Safety; EN61010
Measurement category II (anticipated transient
overvoltage 2.5 kV), pollution level 2
EMC; EN61326-1, EN61326-2-2, EN61000-3-2, EN61000-3-3
Related Product
The DC SIGNAL SOURCE SS7012: a DC signal
generator for calibrating thermocouples.
In addition to measuring and generating ±25 V, ±25
mA, the SS7012 is capable of generating eight types of
thermoelectromotive force by temperature settings.
3
■ Measurement Range and Accuracy
is guaranteed at 23±5˚C and at 80% rh or less after 5 minutes warm-up. [ Accuracy
]
In other conditions, add ±(measurement accuracy 0.15)/ºC
● DC Voltage (DCV) / AC Voltage (ACV) / AC+DC Voltage (ACDCV) / 1-ms peak-hold Voltage (V)
Range Measurement
Resolution DC V Accuracy
range
50 mV 0 to ±51.000 mV
500 mV 0 to ±510.00 mV
0 to ±5.1000 V
5 V
0 to ±51.000 V
50 V
0 to ±250.00 V
250 V
1 µV
10 µV
100 µV
1 mV
10 mV
AC V Accuracy*2
ACDC V Accuracy*2
±(0.05% rdg.+ 5 dgt.)*1
45 Hz to 5 kHz 5 kHz to 20 kHz 45 Hz to 5 kHz 5 kHz to 20 kHz
±(0.7% rdg.+ 40 dgt.) ±(1.5% rdg.+ 40 dgt.) ±(0.8% rdg.+ 70 dgt.) ±(1.6% rdg.+ 70 dgt.)
±(0.03% rdg.+ 5 dgt.)
±(0.7% rdg.+ 20 dgt.) ±(1.5% rdg.+ 20 dgt.) ±(0.8% rdg.+ 25 dgt.) ±(1.6% rdg.+ 25 dgt.)
Input
V Accuracy*3 impedance
1000 MΩ
±(2% rdg.+ 400 dgt.)
10 MΩ
(100 pF or less)
*1. When measuring REL after shorting input terminals before measurement. Accuracy when not measuring REL is ±(0.05% rdg +50 dgt.). *2. Specified for 5% or more from the low end of range. Crest factor: 3 or less.
*3. Specified for signals with at least 1ms pulse width. ■ Overvoltage protection: 250 Vrms AC. Upper limit frequency product: 106 VHz.
● DC Current (DCA) / AC Current (ACA) / AC+DC Current (ACDCA) / 1-ms peak-hold Current (A)
Range Measurement
Resolution
range
50 mA 0 to ±51.000 mA
500 mA 0 to ±510.00 mA
1 µA
10 µA
AC A Accuracy*2
ACDC A Accuracy*2
±(0.6% rdg.+ 20 dgt.)
±(0.7% rdg.+ 40 dgt.)
DC A Accuracy*1
±(0.03% rdg.+ 5 dgt.)
45 Hz to 2 kHz
Maximum voltage
Shunt resistance
A Accuracy*3
45 Hz to 2 kHz
0.06V/1 Ω
0.6V/1 Ω
±(2% rdg.+ 400 dgt.)
*1. When measuring REL after opening input terminals before measurement, or for “0 mA” input. Accuracy when not measuring REL is ±(0.03% rdg + 25 dgt.). *2. Specified for 5% or more from the low end of range. Crest factor: 3 or less.
*3. Specified for signals with at least 1ms pulse width. ■ Overcurrent protection: fast-blow fuse (630 mA /250 V)
● Resistance (OHM)
Measurement
Range Measurement
Resolution
Accuracy
range
current
1
500 Ω* 0 to 510.00 Ω 0.01 Ω ±(0.15% rdg.+ 8 dgt.)
0.45 mA
5 kΩ*1 0 to 5.1000 kΩ 0.1 Ω
45 µA
50 kΩ 0 to 51.000 kΩ 1 Ω
±(0.15% rdg.+ 5 dgt.)
4.5 µA
500 kΩ 0 to 510.00 kΩ 10 Ω
450 nA
5 MΩ 0 to 5.1000 MΩ 0.1 kΩ
±(1% rdg.+ 8 dgt.)
45 nA
50 MΩ 0 to 51.000 MΩ 1 kΩ
Open terminal
voltage
< +4.8 VDC
● Diode test (DIODE), Continuity test (CONT)
Range Resolution
DIODE
CONT
Measurement Open terminal
current
voltage
Accuracy
0.1 mV
–
±(0.05% rdg.+ 5 dgt.)
Beeps below about 50 mV
Sleeve
● Temperature(TEMP)
(K type30 thermocouple)
500
55
Rounded sheath tip
Range
˚C
˚F
Measurement range
– 40 ˚C to 1372 ˚C
– 40 ˚F to 2502 ˚F
Ø 3.2
Resolution
6
< +4.8 VDC
0.45 mA
■ Overvoltage protection: 250 Vrms AC.
Sheath Ø 8
Ø 3.2
Approx.
0.1 ˚C
0.1 Ø˚F
13
Accuracy
±(0.3% rdg.+ 3 ˚C )
±(0.3% rdg.+ 6 ˚F )
*1. When measuring REL.
■ Beeps when measured value is 1000 dgt or less (can be set on or off)
■ Accuracy of 50-MΩ range is specified for humidity up to 60% rh. ■ Overvoltage protection: 250 Vrms AC
Grip
■ The above accuracy doesRounded
not include
accuracy of thermocouple.
■ Overvoltage protection: 250 Vrms AC.
sheath tip
● Frequency (FREQ)
● Probes
Range
Measurement range Resolution
0.001 Hz
100 Hz 0.500 Hz to 99.999 Hz
0.50 Hz to 999.99 Hz
0.01 Hz
1 kHz
0.1 Hz
10 kHz 0.5 Hz to 9.9999 kHz
1 Hz to 99.999 kHz
1 Hz
100 kHz
10 Hz
200 kHz 10 Hz to 199.99 kHz
Accuracy
±(0.02% rdg. + 3 dgt.)
(Voltage measurement sensitivity)
Minimum input level
Trigger level
Input
(rms sine wave)
(DC coupling)
range
1 Hz to 100 kHz >100 kHz
15 mV
25 mV
35 mV
200 mV
0.3 V
0.5 V
3 V
5 V
30 V
–
< 20 kHz 20 kHz to 200 kHz
20 mV
30 mV
50 mV
80 mV
0.5 V
0.8 V
5 V
8 V
60 V
–
Minimum
Input input level
range (rms sine wave)
30Hz to 20kHz
50 mA 2.5 mA
500 mA 25 mA
* SIGNAL
Measurement range: 0.1 to 1999 ms
Accuracy: < ±(0.2% rdg. + 3 dgt.)
When pulse width >10µs, specification.
Measurement range depends on signal frequency.
SOURCE 7016
[Accessories: Carrying case × 1, AC adapter × 1, Ni-MH battery × 8,
Test lead L9207-10 × 1 set (for measurement),
Test lead × 1 set (for generation), Test lead × 1(yellow)]
● OPTION
COMMUNICATION PACKAGE 3856-02 (USB)
* SHEATH TYPE TEMPERATURE PROBE 9180
* SURFACE TYPE TEMPERATURE PROBE 9181
Sheath
150
100
6
SHEATH TYPE TEMPERATURE SURFACE TYPE
SHEATH TYPE
PROBE 9180, 9183
TEMPERATURE PROBE 9181 TEMPERATURE PROBE 9182
Rounded sheath tip
Ø 13
Sheath
150
Grip
6
Sensor
Ø 15
100
Ø 13
8
Sheath Ø 8
Ø 3.2
Grip
Rounded sheath tip
100
500
6
Sleeve
55
30
9180, 9183
Type of element wire
(Current
measurement
sensitivity)
Sleeve
55
30
6
Model name
(Duty ratio, Pulse width) [Accuracy is specified for 5V pulse in 5V DC range]
DUTY RATIO
PULSE WIDTH
Measurement range:
0.1 to 99.9% (DC coupling)
5 to 95% (AC coupling)
Accuracy: < ±(0.3%/kHz + 0.3% f.s.)
500
Ø 3.2
■ Minimum input frequency: 0.5 Hz (set by power-on option). ■ Overvoltage protection: 250 Vrms AC.
50 mV
500 mV
5 V
50 V
250 V
Sheath Ø 8
Ø 3.2
Rounded sheath tip
Tolerance
Ø 15
Sheath/ Ødimensions
Ø 2.3 (9472)
4.8 (9473)
Rounded sheath tip
Compensation lead
9181
9182
K (CA) Chromel/Alumel
Ø
3.2
Ø 13
Sheath
JIS C 1602, class 2*(class 1* for 9183)
Sensor Ø 13
Grip
3.2×150
15×8 mm Rounded sheath tip 1503.2×500 mm
Grip mm
Compensation
conductor
100
general
Gripfor 100
8
100
150 (9472) / 300 (9473)
6 -4˚F to 302˚F)
2m (-20 to 150˚C,
use 1m (-20 to 90˚C, -4˚F to 194˚F)
Ø 2.3 (9472) / Ø 4.8 (9473)
1000
Rounded
sheath (752˚F)
tip
750˚C (1382˚F)
400˚C
150 (9472) / 300 (9473)
150˚C (302˚F)
Max. use temperature
Grip heat resistance
Compensation conductor
Grip
100
750˚C (1382˚F)
1000(194˚F)
90˚C
Ø 2.3TYPE
(9474) / TEMPERATURE
Ø 4.8 (9475)
13
Sensor
Ø 15 PROBE
SHEATH
PROBEconductor
9472, 9473 SHEATH TYPE TEMPERATURE
9474,Ø 9475
Compensation
Rounded sheath tip
Ø 2.3 (9472) / Ø 4.8 (9473) Grip
Rounded sheath tip
100
100
150 (9472) / 300 (9473)
Compensation conductor
Grip
100
1000
Ø 2.3 (9474) / Ø 4.8 (9475)
Rounded sheath tip
1000
100
Grip
Compensation conductor
Grip
100
1000
100
8
SURFACE TYPE TEMPERATURE PROBE 9476
Ø 20
25
Compensation conductor
Ø 2.3 (9474) / Ø 4.8 (9475)
Compensation conductor
Grip
Rounded sheath tip
1000
100 Grip
90
Model name
100
Type of element wire
Tolerance
Ø 20
Sheath dimensions
100
9472
1000
Compensation conductor
Ø 20
25
Grip
1000
100
90
9272-9275
are water
resistant
9473
9474
9475
K (CA) Chromel/Alumel (JIS C 1602: 1995)
9476
Class 1 (the greater
of ±1.5˚C
(±2.7˚F) or ±0.4% of the measurement temperature)
Compensation
conductor
Class 2*
20 mm
Grip
2.3 × 100 mm
4.8 × 30 mm 2.3 × 100 mm 4.8 × 100 mm
1000
100
90
for general
use 1 m (-20 to 90˚C, -4˚F to 194˚F)
Max. use temperature 300˚C (572˚F) 800˚C (1472˚F) 300˚C (572˚F)
500˚C (932˚F)
Grip heat resistance
80˚C (176˚F)
*Class 2 is the greater of ±2.5˚C (±4.5˚F) or ±0.75% of the measurement temperature)
25
Compensation lead
* SHEATH TYPE TEMPERATURE PROBE 9182
SHEATH TYPE TEMPERATURE PROBE 9183 (class 1)
SHEATH TYPE TEMPERATURE PROBE 9472 (class 1)
SHEATH TYPE TEMPERATURE PROBE 9473 (class 1)
SHEATH TYPE TEMPERATURE PROBE 9474 (class 1)
SHEATH TYPE TEMPERATURE PROBE 9475 (class 1)
SURFACE TYPE TEMPERATURE PROBE 9476
TEST LEAD L9207-10 (Lead length: Approx.1 m / standard accessories)
* Non-CE mark products
Note: Company names and Product names appearing in this catalog are trademarks or registered trademarks of various companies.
HIOKI (Shanghai) Sales & Trading Co., Ltd. :
Headquarters :
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
http://www.hioki.co.jp / E-mail: os-com@hioki.co.jp
HIOKI USA CORPORATION :
6 Corporate Drive, Cranbury, NJ 08512 USA
TEL +1-609-409-9109 / FAX +1-609-409-9108
http://www.hiokiusa.com / E-mail: hioki@hiokiusa.com
1608-1610,Shanghai Times Square Office, 93 Huai Hai Zhong Road
Shanghai, P.R.China POSTCODE: 200021
TEL +86-21-63910090/63910092 FAX +86-21-63910360
http://www.hioki.cn / E-mail: info@hioki.com.cn
Beijing Office : TEL +86-10-84418761 / 84418762
Guangzhou Office : TEL +86-20-38392673 / 38392676
DISTRIBUTED BY
HIOKI INDIA PRIVATE LIMITED :
Khandela House, 24 Gulmohar Colony Indore 452 018 (M.P.), India
TEL +91-731-4223901, 4223902 FAX +91-731-4223903
http://www.hioki.in / E-mail: info@hioki.in
HIOKI SINGAPORE PTE. LTD. :
33 Ubi Avenue 3, #03-02 Vertex, Singapore 408868
TEL +65-6634-7677 FAX +65-6634-7477
E-mail: info@hioki.com.sg
All information correct as of Jul. 20, 2011. All specifications are subject to change without notice.
7016E2-17M
Printed in Japan