Manning EC-P2 Gas Sensor Instruction and Maintenance Manual
Below you will find brief information for Gas Sensor EC-P2. The EC-P2 is a portable, battery operated instrument for the measurement of gas concentrations in ambient air. This sensor can be easily converted to detect another gas by simply plugging in a different SMART-CELL. The EC-P2 can log periodic gas concentration readings to be downloaded to a file which can be read by Microsoft Excel or other popular spreadsheet applications. The instrument will store instantaneous gas values every one, five, 10, or 15 minutes. The EC-P2 also has advanced capabilities such as data logging to an Excel spreadsheet, 0 to 1.0 Volt DC analog output capability, and a sampling mode for taking a single reading in a controlled, repeatable manner.
Advertisement
Advertisement
Instruction and
Maintenance Manual
Rev G
EC-P2
Gas Sensor
11511 West 83 rd
Terrace email [email protected]
Lenexa, Kansas 66214 www.gasalarm.com
1.800.444.9935 or 913.894.1185
Manning EC-P2 Sensor fax 913.894.1296
1
Contents
Section Title
1 System
2 Operation
3 Maintenance
4 Programming and Advanced Features
5 Appendices
6 Limited
Introduction
Serial number:
Page
3
5
7
9
13
15
This manual has been prepared to help in the use of the Manning Systems EC-P2 portable gas detector.
This manual must be carefully followed by all individuals who have or will have the responsibility for using or servicing the sensor.
Warranties made by Manning Systems, Inc. with respect to this equipment will be voided if the equipment is not used and serviced in accordance with the instructions in this manual. If in doubt about a procedure, please contact
Manning Systems, Inc. before proceeding.
Manning EC-P2 Sensor
18272 ECP2 04/2005 REVG Copyright © 2005 Manning Systems, Inc. All Rights Reserved.
2
1
System Description
The EC-P2 is a portable, battery operated instrument for the measurement of gas concentrations in ambient air.
The EC-P2 can be converted from one target gas to another by simply plugging in a different SMART-CELL for the desired target gas (takes about 30 seconds). This allows one instrument to detect a number of different gases.
All calibration and alarm setpoint data is stored on the individual SMART-CELL so reconfiguration is not necessary when changing target gases. Extra SMART-
CELLs should be kept on the optional sensor keeper when not in use to keep them “warmed up” and ready for immediate use.
The EC-P2 utilizes two different batteries to insure that you’ll never be without power when you need it. The primary power source is a “C” size alkaline battery housed in the handle. It will run the unit approximately
48 hours. When this battery is removed or dead, the unit will switch to its standby power source, an internal rechargeable NiCad battery, which will run the unit approximately six hours. When not in use, the unit should be plugged into the charger to keep this standby battery charged. If desired, the unit can be run continuously when plugged into the charger.
The EC-P2 also has advanced capabilities such as data logging to an Excel spreadsheet, 0 to 1.0 Volt DC analog output capability, and a sampling mode for taking a single reading in a controlled, repeatable manner.
Figure 1: Basic Parts of the EC-P2 Sensor
Outlet Tube
Sampling Wand
Assembly
Data Logger Port
Standby Battery
Recharge Receptacle
Primary Battery
Manning EC-P2 Sensor
18272 ECP2 04/2005 REVG Copyright © 2005 Manning Systems, Inc. All Rights Reserved.
3
1
System Description
continued
System Specifications
Primary Power:
Alkaline C cell battery, 48 hours operation.
Note:
An audible alarm single beep every
60 seconds indicates that either the Alkaline C cell battery is low or the internal rechargeable NiCad battery is low.
Standby Power:
Internal rechargeable NiCad battery, 6 hours operation.
Note:
An audible alarm single beep every 60 seconds indicates that either the Alkaline C cell battery is low or the internal rechargeable NiCad battery is low.
Charger:
120 VAC plug-in charger for internal standby battery
Display:
Backlit LCD
Pump:
Miniature internal diaphragm type, flowrate 0.5 SCFH
Alarms:
Two concentration alarms, zero drift caution, Low flow and Low battery alarms. Alarms displayed on LCD and indicated by audible beeper.
Standard Accessories:
Storage case, 10" extension wand, battery charger, filter material, datalogging software, RS-232 output cable (DB9 connector)
Optional Accessories:
Analog output cable. Sensor
Keeper for storage of additional SMART-CELLs
Sensor Type:
Target gas selective electrochemical
SMART-CELL
Accuracy:
Sensor dependant but generally ± 5% of reading
Repeatability:
± 1% full scale
Operating Ambient Temperature Range:
0 o
to +120 o
F
Storage Temperature:
Recommend storing on the charger in an air conditioned office
Weight:
3 lbs. (shipping weight), 1.8 lbs. (detector only)
Dimensions:
Approx 3.5" x 5" x 8.5"
Outputs:
RS-232 output of stored gas values. Analog output 0 to 1.0 VDC (requires optional analog output cable)
Memory:
12,000 data points, storage interval between points programmable at 1, 5, 10, or 15 minutes. Typical capacity: 8 days at 1 minute storage interval
Manning EC-P2 Sensor
18272 ECP2 04/2005 REVG Copyright © 2005 Manning Systems, Inc. All Rights Reserved.
4
2
Operation
A
Operating the Sensor
The EC-P2 is shipped ready for use. To power up the instrument, push the PWR button. The power up sequence displays first the self check, the sensor full scale value, the downscale caution level, warning concentration setpoint, alarm concentration setpoint, output ranges, software version, then the normal operating screen.
The unit will then display the instantaneous gas concentration and is ready for use.
The normal operating screen displays the information as shown in Figure 2.
The EC-P2 is designed to avoid accidental shut-off of the power.
To turn the unit off, the PWR button must be held down for approximately three seconds.
Release the button after the “Power Down” message appears.
B
Changing
Technology
Target using SMART-CELL
The EC-P2 can easily be changed to sense another target gas by installing the SMART-CELL for the desired target gas. Each SMART-CELL contains all the calibration and alarm setpoint information on its internal memory chip, so recalibration or reprogramming is not necessary when changing the EC-P2 from one gas to another. This procedure can be performed with the unit powered on or off (see Figure 3).
1)
Unscrew the two stainless steel thumbscrews and remove the black inlet / outlet manifold lid from the back of the unit.
2)
To remove the existing SMART-CELL, grip the sides of the cell and pull it straight out.
3)
Install the desired SMART-CELL by gently inserting it into the socket. Notice that the SMART-CELL is keyed so it cannot be inserted incorrectly.
4)
Replace the manifold and hand tighten the thumbscrews.
5)
Verify proper flow through the sampling wand using the flowmeter provided.
Extra SMART-CELLS should be kept on the optional sensor keeper when not in use to keep them “warmed up” and ready for immediate use. The sensor keeper is powered by an Alkaline “C” size battery which should be replaced every 6 months.
For a variety of target gases, see page 13 for a list of available SMART-CELLs. Contact Manning Systems if the desired target gas is not listed, as new cells may be available.
Figure 2: Normal Operating Screen of the EC-P2 Sensor
BATTERY IN USE CURRENT GAS CONCENTRATION
P = Primary (“C” cell)
S = Standby (Internal NiCad)
(Blinking indicates sensor over range)
Flashing indicates low battery
TARGET GAS
SMART-CELL dependent
P
A
OFF
0
PPM
NH3
MENU SMPL
Battery Light
ADVANCED FEATURES ONLY
Do not push for normal operation
ALARM INDICATION CODE
Blank = No alarm condition present
C = Caution—Cell has excessive zero downdrift
W = Warning—Gas concentration has exceeded warning level
A = Alarm—Gas concentration has exceeded alarm level
P = Pump trouble
Lg = Data logging in progress
Manning EC-P2 Sensor
18272 ECP2 04/2005 REVG Copyright © 2005 Manning Systems, Inc. All Rights Reserved.
5
2
Operation
continued
C
Test
Prior to use, the unit should first be tested for proper response. With the detector operating, the unit should be exposed to a sample of the target gas. The display should show an increasing concentration. If not, do not use it for field measurements.
D
Sample Inlet Port
The EC-P2 is provided with a quick disconnect flexible extension wand. It will work with or without the wand.
When using the extension wand, the internal walls must be kept dry. Water on the walls of the tubing can potentially absorb the target gas. To dry, allow the unit to pump dry, ambient air for 15 minutes.
E
Response/Recovery Time
Under normal conditions, the instrument will reach 90% of final value within two minutes. This is dependent on concentration and temperature.
Recovery time for the sensor depends on duration and concentration of exposure. Short exposures of concentrations at the lower end of the target gas range result in rapid recovery. Long exposures to levels above the middle of the target gas range or short exposures to levels exceeding the target gas range can extend recovery times to hours. Repeated exposures above the target gas range will reduce cell life and should be avoided.
F
Interference Gases
The EC-P2 is generally quite specific to the target gas, but depending on the target gas, other gases may cause a reading. See Appendix B for a listing of potential interference gases for the various SMART-CELLs.
G
Alarm Functions
The EC-P2 provides both visible and audible gas concentration alarms and system alarms. An alarm condition will be displayed on the normal operating screen as described on page 5. In addition to the alarm indication code in the left side of the screen, an alarm message will flash intermittently on the screen, the audible beeper will sound and the ACK button will appear. Pushing the
ACK button will silence the beeper and stop the display flashing. The Alarm Indication Code will remain on the left side of the screen until the alarm condition clears.
Alarm setpoints are stored on the individual SMART-
CELL and can be changed by the user as described in
Section 4.
Note:
An audible alarm single beep every 60 seconds indicates that either the Alkaline C cell battery is low or the internal rechargeable NiCad battery is low.
H
Resolution
The EC-P2 displays gas concentration with a resolution that depends upon the range of the unit. Full scale ranges of 0-4.00 or below will provide resolution of 0.01. Ranges from 0-5.0 up to 0-49.9 will provide resolution of 0.1.
Full scale ranges above 50 will have a resolution of 1.
Figure 3: Manifold Assembly of the EC-P2 Sensor
Sampling wand
Outlet tube barb fitting
INLET
TEST UNIT PRIOR TO USE
OUTLET
Thumbscrews
(loosen to remove manifold lid)
Sensor
(SMART-CELL)
Filter Material
Inlet fitting
CAUTION
DO NOT BLOCK INLET
OR OUTLIET PORTS
Manifold lid
Sensor alignment groove
Sensor alignment pin
Manning EC-P2 Sensor
18272 ECP2 04/2005 REVG Copyright © 2005 Manning Systems, Inc. All Rights Reserved.
6
3
Maintenance
A
Maintenance
It is essential that the test and calibration schedule be adhered to. Manning Systems recommends the following maintenance schedule:
• Response test prior to each use. Expose sensor to a sample of the target gas to verify proper sensor response and alarm functions. should performed with certified calibration gas every six months, and all tests must be logged.
• Check filter material every six months. Replace if dirty or moist.
• If the optional sensor keeper is used for extra
SMART-CELLs, change the sensor keeper battery once every 6 months. Use only alkaline “C” size batteries.
B
Life
The electrochemical SMART CELL is extremely reliable, but the following can cause the cell chemicals to become depleted:
• A period of time
• Exposure to high temperatures
• Exposure to high concentrations of the target gas
• Exposure to high moisture
When the cell becomes depleted, the unit will give no indication of failure other than that the sensor will not respond. For this reason it is absolutely essential that these units be exercised with a gas sample on a regular basis.
Typical sensor life under normal use is two years or more. When the SMART-CELL will no longer calibrate properly, simply purchase a new SMART-CELL and replace.
C
Calibration
Calibration is recommended every six months in normal use. Each SMART-CELL stores its calibration and alarm setpoint information on its own memory chip. When the
SMART-CELL is plugged into the EC-P2, this information is used by the EC-P2. The SMART-CELL can be field calibrated by the user or returned to Manning Systems for factory calibration.
The calibration procedure requires the use of a bypass tee to allow the gas to be drawn into the flow cell without creating variable pressure or dilution problems.
Calibration gas cannot be fed into the flow cell under pressure.
When applying calibration gas use the bypass tee as shown in figure 4 on page 8. The flowrate of calibration gas should be high enough that at least 0.5 SCFH is flowing out of the calibration tee. This can be measured with the supplied flowmeter and insures that air is not being drawn into the unit and diluting the calibration gas.
The unit may be zeroed without spanning it; however, never span the unit without first zeroing it.
The following procedure will calibrate the SMART-CELL.
To Enter ZERO / SPAN Mode
1)
Push power button to turn unit on. Wait for power up sequence to finish (30 seconds).
2) Push and hold the unnamed button (second from left) first, then Push and hold the Menu button. Hold both buttons until the screen goes blank (2 seconds). Release both buttons and the unit is now in Programming Mode.
3) Push the SENS button and the unit is now in the zero/span mode.
Manning EC-P2 Sensor
18272 ECP2 04/2005 REVG Copyright © 2005 Manning Systems, Inc. All Rights Reserved.
7
3
Maintenance
continued
Zero Calibration
4)
Be sure the unit is in clean air or apply zero air for two minutes.
5)
Push ZERO button.
6)
Push SAVE button to save the new zero calibration.
Span Calibration
7) Push SPAN button.
8) Apply span gas for two minutes.
9) Push INC button (increase) or DEC button (decrease) until the display matches the span gas concentration.
10) Push SAVE button to store new span calibration.
To Exit ZERO / SPAN Mode
11) Push DONE button twice slowly to get back into normal operation mode.
D
Troubleshooting
The detector does not start:
Check to see that a fresh battery is installed. Both the primary and standby batteries must be discharged for this problem to occur.
Plug the charger into the EC-P2 to charge the standby battery. Wait 10 minutes and retry.
The detector does not respond to the target gas:
Verify the proper flow rate using the supplied flowmeter.
Perform calibration. If sensor won’t calibrate, the
SMART-CELL may be depleted. Contact Manning
Systems for a replacement SMART-CELL.
Pump alarm:
This will occur if the pump motor stops or if an internal pressure sensor detects blockage of the inlet. Check that the internal filter is not clogged or wet and that the inlet tube is clear.
External flowmeter indicates no flow:
Verify that pump is running. Check that the manifold screws are tight and the extension wand connection is tight.
Unit does not respond when keys are pressed:
Press and hold all four buttons for five seconds. This will reset the EC-P2. Turn the unit back on by pressing the PWR button.
High reading won’t clear:
When the sensor is exposed to very high gas concentrations, it takes a relatively long time for the sensor to clear. Leave the unit running in a clean environment for a few hours to clear the sensor. If it doesn’t clear after six hours, the sensor may have been damaged. Contact Manning Systems for a replacement SMART-CELL.
Unit is in alarm with gas reading of 0 ppm:
Check the alarm setpoints and alarm functions as described on page 9. Verify that alarm is not programmed as a downscale alarm, and that the setpoint is above zero.
Figure 4: Calibration of the EC-P2 Sensor using a Bypass Tee
Vent through flowmeter
Bypass tee
.4
.2
LPM AIR
1.0
.8
.6
Calibration gas flowmeter
Calibration gas bottle
(use only Certified
Calibration Gas)
TEST UNIT PRIOR TO USE
OUTLET
INLET
CAUTION
DO NOT BLOCK INLET
OR OUTLIET PORTS
Manning EC-P2 Sensor
18272 ECP2 04/2005 REVG Copyright © 2005 Manning Systems, Inc. All Rights Reserved.
8
4
Programming and Advanced Features
A
Alarm Setpoints and
Functions
The EC-P2 has three adjustable alarm setpoints. These setpoints are stored on each SMART-CELL. The Caution setpoint is to warn of excessive negative zero drift. The
WARNING and ALARM setpoints are to warn of potentially dangerous target gas concentrations. The following procedure will change the WARNING setpoint.
The function of each alarm can be selected to alarm on increasing gas concentration (UP arrow), alarm on decreasing gas concentration (DOWN arrow), or be disabled (off). The following procedure will set the
Warning function to alarm on decreasing concentration
(Useful for Oxygen Depletion warning).
To Enter CAUT / WARN / ALRM Mode
1) Push power button to turn unit on. Wait for power up sequence to finish (30 seconds).
To Enter CAUT / WARN / ALRM Mode
1) Push power button to turn unit on. Wait for power up sequence to finish (30 seconds).
2) Push and hold the unnamed button first, then Push and hold the Menu button. Hold both buttons until the screen goes blank (two seconds). Release both buttons and the unit is now in Programming Mode.
2) Push and hold the unnamed button first, then Push and hold the Menu button. Hold both buttons until the screen goes blank (two seconds). Release both buttons and the unit is now in Programming Mode.
3) Push the ALRMS button and the unit is now in the
CAUT / WARN / ALRM mode.
Warning Function
3) Push the ALRM button and the unit is now in the
CAUT / WARN / ALRM mode.
4) Push WARN button.
Warning Setpoint
5) Push FUNC (function) button.
4) Push WARN button.
5)
Push S.P. button (setpoint).
6) Push INC button (increase) or DEC button (decrease) until the display matches the desired warning setpoint.
Notice the minus sign above the arrow if the setpoint is below zero.
7) Push SAVE button to store new warning setpoint.
8)
Push DONE button to return to Alarms mode.
To Exit CAUT / WARN / ALRM Mode
6) Push SELECT button while viewing the arrow in the lower left hand corner of the screen to toggle through the following three possibilities:
• UP arrow – Warn on increasing concentration
(gas goes higher than setpoint).
• DOWN arrow – Warn on decreasing concentration
(gas goes lower than setpoint).
• Blank – warning function turned OFF.
7) Push SAVE button when the DOWN arrow is displayed to store new warning function.
8) Push DONE button to return to Alarms mode.
9) Push DONE button twice slowly to get back into normal operation mode.
The Caution and Alarm setpoints are programmed in the same manner, except for step four above in which you would push either the CAUT or ALRM button.
To Exit CAUT / WARN / ALRM Mode
9)
Push DONE button two times slowly to get back into normal operation mode.
The Caution and Alarm functions are programmed in the same manner, except for step four above in which you would push either the CAUT or ALRM button.
Manning EC-P2 Sensor
18272 ECP2 04/2005 REVG Copyright © 2005 Manning Systems, Inc. All Rights Reserved.
9
4
Programming and Advanced Features
continued
B
Display Variables
C
Mode
The EC-P2 has two adjustable display variables. These variables control signal conditioning of the LCD display.
Variables are stored on each SMART-CELL. The AVG
(Average) variable determines the number of one second samples to average to calculate the displayed value. The lower the averaging value, the faster the display will change and the more “wiggle” will be displayed. The BLANK (zero blanking) variable controls the lowest reading that will be displayed. This is to eliminate the misleading display of signals within the zero noise of the sensor. The following procedure will change the AVG variable.
To Enter AVG / BLANK Mode
1) Push power button to turn unit on. Wait for power up sequence to finish (30 seconds).
2) Push and hold the unnamed button first, then Push and hold the Menu button. Hold both buttons until the screen goes blank (two seconds). Release both buttons and the unit is now in Programming Mode.
3) Push the MORE button.
4) Push the DISP (display) button and the unit is now in the
AVG / BLANK mode.
The Sample Mode feature allows the user to take a measurement using a predefined repeatable method to eliminate user “interpretation” of a gas concentration that is changing. This is similar to a digital scale, for example, so the exact same reading will be taken regardless of who is taking it.
The measuring sequence used in Sample Mode is as follows:
Sample
for a user programmed sampling time to allow unit to stabilize to the existing gas concentration.
•
Measure
the gas concentration and average it for a user programmed measuring time.
•
Clear
the gas concentration to below a user programmed recovery value before allowing another cycle to start. (Unit should be moved to fresh air during the clear time.)
AVG Variable
5) Push AVG button.
6) Push INC button (increase) or DEC button (decrease) until the display matches the desired value.
7) Push SAVE button to store new AVG variable.
To Exit AVG / BLANK Mode
8) Push DONE button twice slowly to get back into normal operation mode.
The Blanking variable is programmed in the same manner, except for step five above in which you would push the BLANK button.
Taking a sample:
To take a sample, locate the detector at the desired sample location, and push the SMPL (sample) button on the normal operation screen. The unit will countdown the Sample time, followed by the Measure time, at the end of which the average value will be displayed.
Note
the displayed sample value. Pushing the CLEAR button will return the unit to the normal operation screen once the gas concentration is below the predefined recovery value. The unit is now ready to take another sample if desired.
Manning EC-P2 Sensor
18272 ECP2 04/2005 REVG Copyright © 2005 Manning Systems, Inc. All Rights Reserved.
10
4
Programming and Advanced Features
continued
Programming sample mode variables:
The sample mode variables are
Sample, Measure
and
Clear
as described above. The following procedure will change the Sample variable.
To Enter SAMP / MEAS / CLEAR Mode
1) Push power button to turn unit on. Wait for power up sequence to finish (30 seconds).
2) Push and hold the unnamed button (second from left) first, then Push and hold the Menu button. Hold both buttons until the screen goes blank (two seconds). Release both buttons and the unit is now in Programming Mode.
3) Push the MORE button.
4) Push the SMPL (sample) button and the unit is now in the
SAMP / MEAS / CLEAR mode.
EC-P2 software installation:
The EC-P2 is supplied with a CD containing the “Manning
Systems EC-P2” program. It will run on Windows 95,
Windows 98, or Windows 2000. To install the program, place the disk in your CD drive. The CD should automatically begin the installation process. Follow the prompts as they appear on your screen. Accepting the default directory locations is recommended for most installations. If the CD does not automatically start, click on “My Computer” and then your CD drive. Double-click on the file named “Setup.exe.”
Prior to your first datalogging session it is recommended that you connect the EC-P2 to your computer with the data logging program running. The computer can then set the time in the EC-P2 to match its own clock.
SAMP Variable
5) Push SAMP (sample) button.
6) Push INC button (increase) or DEC button (decrease) until the display matches the desired value. The display is in minutes and seconds.
7) Push SAVE button to store new SAMP variable.
To Exit SAMP / MEAS / CLEAR Mode
8) Push DONE button twice slowly to get back into normal operation mode.
The Measure and Clear variables are programmed in the same manner, except for step five above in which you would push the MEAS or CLEAR button.
D
Logging
The EC-P2 can log periodic gas concentration readings to be downloaded to a file which can be read by
Microsoft Excel or other popular spreadsheet applications. The EC-P2 will store instantaneous gas values every one, five, 10, or 15 minutes. Multiple sessions can be logged and then downloaded as separate files.
Data logging with the EC-P2:
To Start the Data Logging Session
1) Push power button to turn unit on. Wait for power up sequence to finish (30 seconds).
2) Push the MENU button.
3) Push the LOG button and the unit is now in the
CLEAR / INT / START mode.
4) If the CLEAR button is available, that means there is data currently stored in the unit. If desired, you can erase the data already in the unit by pushing the CLEAR button.
(If you don’t push CLEAR, each data logging session can be downloaded as a separate file).
5) Push INT button (interval) to cycle through the choices for how often to log a data point. Stop when the desired interval is displayed. Notice how much time can be covered with the remaining free memory.
6) Push and hold the START button until the screen goes blank (two seconds) to begin a new data logging session.
The unit will now display the gas concentration. The Lg symbol in the left side of the display indicates that the unit is logging data. If you watch this symbol carefully, you will see it change to an “s” each time a data point is stored.
Manning EC-P2 Sensor
18272 ECP2 04/2005 REVG Copyright © 2005 Manning Systems, Inc. All Rights Reserved.
11
4
Programming and Advanced Features
continued
To Stop the Data Logging Session
E
Output
7) Push the MENU button.
8)
Push the LOG button.
9) Push and hold the STOP button until the screen changes to CLEAR / INT / START / MODE (two seconds).
10) Push the DONE button to get back into normal operation mode.
The EC-P2 has a 0 to 1.0 volt DC analog output capability when used with the optional analog output cable. With the analog output cable plugged into the communication port of the EC-P2, the user can measure the voltage between the two wires and correlate it to gas concentration. The voltage output will be linear between zero and the full scale range of the SMART-
CELL as programmed below. Each SMART-CELL has a wide range of choices of full scale values.
The full scale range as programmed below only affects the analog output and the data logging full-scale output.
The display will always cover the entire operating range of the SMART-CELL. Changing the range can affect the display resolution for some SMART-CELLs.
Working with the “EC-P2” program:
The “EC-P2” program downloads the logging files from the EC-P2. It does not display the data. The files it downloads can be opened by or imported into Excel,
Lotus 123 or other programs for graphing and analysis.
Turn on the EC-P2 and connect it to your PC using the
RS-232 cable supplied. From Windows START menu, select “PROGRAMS,” then “Manning Systems EC-P2,” and then “EC-P2”. As soon as you start the “EC-P2” program, your computer will try to establish a connection to the EC-P2. If the EC-P2 is currently logging data, the “EC-P2” program will automatically stop the logging function.
The program contains two tabs, one marked “Download” and one marked “Configure”. Review the settings on the
“Configure” tab, making changes as necessary.
From the “Download” tab, review the Link Status block in the lower right hand corner. It should indicate
“Link OK”. Link Trouble can be caused by selection of the wrong Com port in the Configure Tab, improperly connecting the RS-232 cable, or not turning on the
EC-P2.
When you are ready to transfer data from the EC-P2 to your PC, click on “get data” and the files will be downloaded to your computer. Note the file names in the destination window so you can locate the files.
These files can now be opened with Excel (use .csv filename), Lotus 123, or other user supplied programs.
Changing the full scale range on EC-P2:
To Enter RANGE Mode
1) Push power button to turn unit on. Wait for power up sequence to finish (30 seconds).
2) Push and hold the unnamed button (second from left) first, then Push and hold the Menu button. Hold both buttons until the screen goes blank (two seconds). Release both buttons and the unit is now in Programming Mode.
3) Push the MORE button.
4) Push the DISP (display) button.
5) Push the MORE button and the unit is now in the RANGE mode.
Adjusting the RANGE
6) Push RANGE button.
7) Push INC button (increase) or DEC button (decrease) until the display matches the desired full scale range.
8) Push SAVE button to store new full scale range.
To Exit RANGE Mode
9) Push DONE button twice slowly to get back into normal operation mode.
Manning EC-P2 Sensor
18272 ECP2 04/2005 REVG Copyright © 2005 Manning Systems, Inc. All Rights Reserved.
12
5
Appendices
APPENDIX A: SMART-CELLs Available
EC-P2-SMART-CELL-NH
3
-0/500ppm
EC-P2-SMART-CELL-NH
3
-0/2000ppm
Bromine 0-5 EC-P2-SMART-CELL-Br
2
-0/5ppm
Carbon Monoxide 0-1000 ppm EC-P2-SMART-CELL-CO-0/1000ppm
Chlorine 0-5 EC-P2-SMART-CELL-Cl
2
-0/5ppm
EC-P2-SMART-CELL-Cl
2
-0/200ppm
Chlorine Dioxide 0-5 ppm EC-P2-SMART-CELL-ClO
2
-0/5ppm
EC-P2-SMART-CELL-Fl
2
-0/5ppm
Hydrogen Chloride
Hydrogen Peroxide
Hydrogen Sulfide
Nitric Oxide
Nitrogen Dioxide
Sulfur Dioxide
0-200ppm
0-200 ppm
0-200 ppm
0-500 ppm
0-200 ppm
0-500 ppm
EC-P2-SMART-CELL-H
2
-0/2000ppm
EC-P2-SMART-CELL-HCl-0/200ppm
EC-P2-SMART-CELL-H
2
O
2
-0/200ppm
EC-P2-SMART-CELL-H
2
S-0/200ppm
EC-P2-SMART-CELL-NO-0/500ppm
EC-P2-SMART-CELL-NO
2
-0/200ppm
EC-P2-SMART-CELL-O
2
-0/25%
EC-P2-SMART-CELL-O
3
-0/5ppm
EC-P2-SMART-CELL-O
3
-0/100ppm
EC-P2-SMART-CELL-SO
2
-0/500ppm
Manning EC-P2 Sensor
18272 ECP2 04/2005 REVG Copyright © 2005 Manning Systems, Inc. All Rights Reserved.
13
5
Appendices
continued
APPENDIX B: SMART-CELL Cross-Sensitivity Table (Interference Gases)
R
F
E
R
E
I
N
T
E
G
A
S
N
C
E
E
S
SENSOR AND TARGET GAS
Br
2
Cl
2
ClO
2
O
3
NH
3
CO H
2
O
2
NO H
2
S NO
2
SO
2
HCHO OH Fl
2
HCl H
2
O
2
Bromine ----- 1.0 1.0 0.7 -0.1 None None None None 0.5 -0.1 None None 1.0 None None
Chlorine 1.0 ----- 1.0 0.7 -0.1 None None None None 0.5 -0.1 None None 1.0 None None
Chlorine dioxide 1.0 1.0 ----- 0.7 -0.1 None None None None None 0.5 -0.1 None None 1.0 None None
Ozone 1.5 1.5 1.5 ----- -0.1 None None None None None 0.65 -0.1 None None 1.5 None None
Carbon monoxide
None None ----- None None None None None None None None None None None None
None None None None 0.05 ----- 0.1 None None 0.002 0.001 None 0.5 0.5 None 0.005 0.005
Oxygen
None ----- None 0.001 0.001 0.001 0.05 0.05 None 0.01 0.01
None None None None None None None ----- None None None None None None None None None
None None None None None 0.1 None None
Formaldehyde
0.1 ----- -0.08
0.1 -1 ----- 0.4 0.4 -0.01
None None None None 0.03 0.1 0.1 None 0.05 0.01 None 0.02 ----- 1.2 None None 0.1
Flourine 1.0 1.0 1.0 0.7 -0.1 None None -0.1 None None ----- None None
Hydrochloric acid None None None None None None None None None None -0.3 None 0.2 0.2 None ----- 0.1
Hydrogen peroxide None None None None 0.03 0.1 0.1 None 0.05 0.01 None 0.02 None None None None -----
Hydrogen -0.08 -0.08 -0.08 -0.05 None 0.1 None None None None -0.07 0.15 0.1 0.1 -0.08 0.01 0.1
Hydroflouric acid
Silica hydride
Carbon dioxide
None None None None None None None None None None None None None None None None None
None None None None 1 0.5
None None None None 0.5 None None
None None None None None None None None None None None None None None None None None
Methane None None None None None None None None None None None None None None None None None
Carbosulfonic -0.8
None 0.1 None None None None None 1 1 None None None
None None None None 0.01 None None None 0.001 0.005 0.001 0.05 2 2 None 0.01 0.02
Notes: 1) None means there is no interference.
2) A number indicates that when exposed to the interference gas, the sensor will react the specified amount for a given concentration.
For example, on the first line of the table above, if the NO
2
sensor is exposed to 1 ppm Bromine (Br
2
), the sensor will read 0.5 ppm.
This interference is in addition to the target gas present. For example, if an NH
3
sensor is exposed to 1 ppm NH
3
and 1 ppm Br
2
at
the same time, the reading will be 0.9 ppm (1ppm NH
3
+ -0.1 ppm Br
2
= 0.9 ppm reading).
Manning EC-P2 Sensor
18272 ECP2 04/2005 REVG Copyright © 2005 Manning Systems, Inc. All Rights Reserved.
14
6
Limited Warranty
Warranty
Manning Systems, Inc. (“Manning”) warrants to the original purchaser and/or ultimate customer
(“Purchaser”) of Manning’s Products (“Product”) that if any part thereof proves to be defective in material or workmanship within eighteen (18) months of the date of shipment by Manning or twelve (12) months from the date of first use by the purchaser, whichever comes first, such defective part will be repaired or replaced, free of charge, at Manning’s discretion if shipped prepaid to Manning at 11511 W. 83rd Terrace, Lenexa,
Kansas 66214, in a package equal to or in the original container. The Product will be returned freight prepaid and repaired or replaced if it is determined by Manning that the part failed due to defective materials or workmanship. The repair or replacement of any such defective part shall be
Manning’s sole and exclusive responsibility and liability under this limited warranty.
2. Exclusions
A. If gas sensors are part of the Product, the gas sensor is covered by a twelve (12) month limited warranty of the manufacturer.
B. If gas sensors are covered by this limited warranty, the gas sensor is subject to inspection by Manning for extended exposure to excessive gas concentrations if a claim by the Purchaser is made under this limited warranty. Should such inspection indicate that the gas sensor has been expended rather than failed prematurely, this limited warranty shall not apply to the Product.
C. This limited warranty does not cover consumable items, such as batteries, or items subject to wear or periodic replacement, including lamps, fuses, valves, vanes, sensor elements, cartridges, or filter elements.
3. Warranty Limitation and Exclusion
Manning will have no further obligation under this limited warranty. All warranty obligations of
Manning are extinguishable if the Product has been subject to abuse, misuse, negligence, or accident or if the Purchaser fails to perform any of the duties set forth in this limited warranty or if the Product has not been operated in accordance with instructions, or if the Product serial number has been removed or altered.
4. Disclaimer of Unstated Warranties
THE WARRANTY PRINTED ABOVE IS THE ONLY
WARRANTY APPLICABLE TO THIS PURCHASE.
ALL OTHER WARRANTIES, EXPRESS OR
IMPLIED, INCLUDING, BUT NOT LIMITED TO,
THE IMPLIED WARRANTIES OF MERCHANT-
ABILITY OR FITNESS FOR A PARTICULAR
PURPOSE ARE HEREBY DISCLAIMED.
5. Limitation of Liability
IT IS UNDERSTOOD AND AGREED THAT
MANNING’S LIABILITY, WHETHER IN
CONTRACT, IN TORT, UNDER ANY WARRANTY,
IN NEGLIGENCE OR OTHERWISE SHALL NOT
EXCEED THE AMOUNT OF THE PURCHASE
PRICE PAID BY THE PURCHASER FOR THE
PRODUCT AND UNDER NO CIRCUMSTANCES
SHALL MANNING BE LIABLE FOR SPECIAL,
INDIRECT, OR CONSEQUENTIAL DAMAGES.
THE PRICE STATED FOR THE PRODUCT IS A
CONSIDERATION LIMITING MANNING’S
LIABILITY. NO ACTION, REGARDLESS OF
FORM, ARISING OUT OF THE TRANSACTIONS
UNDER THIS WARRANTY MAY BE BROUGHT BY
THE PURCHASER MORE THAN ONE YEAR
AFTER THE CAUSE OF ACTION HAS
OCCURRED.
Manning EC-P2 Sensor
18272 ECP2 04/2005 REVG Copyright © 2005 Manning Systems, Inc. All Rights Reserved.
15
Advertisement
Key features
- Portable, battery operated gas detector
- SMART-CELL technology for easy gas type conversion
- Data logging to an Excel spreadsheet
- 0 to 1.0 Volt DC analog output capability
- Sampling mode for controlled measurements