9010 30P

PMM 9010/30P
Full CISPR 16-1-1 Compliant 9 kHz-30 MHz
Field Upgradable up to 18 GHz
(CISPR 16-1-1 Compliant)
All new CISPR detectors included
Save
Time
& Money!
2
Conducted emissions can be the most critical part of the
interference to measure, particularly for industrial equipments and household appliances with switching power supplies, motors, contactors, speed regulators etc.
General purpose and pre-test instruments may be quite ineffective for correct evaluation of such complex disturbance signals; meantime, building a fullcompliant test setup is accessible to all (see example at page 4).
PMM 9010/30P features full-compliance to CISPR-16-1-1 and
MIL-STD 461 for conducted measurements from 10 Hz up to
30 MHz. The required accessories are summarized at page 10.
PMM 9010/30P: one instr ument, many applications
REVOLUTIONARY CONCEPT
PMM 9010/30P is the first EMI
Receiver that fits in any budget and grows with your needs:
• On-field expandable to Radiated
Full Compliance (see page 6)
• Safe and easy Firmware and
Software upgrading from Web
• Digital Technology for best
reliability
• Reduced turnaround time and cost
for periodic re-calibration
• Full choice of Ancillary Equipments
• PEMS - PMM Emission Suite included
• Small size, lightweight, AC/DC and
battery for portable operation
10 Hz—30 MHz RF input
30 MHz - 3 GHz RF input
Radiated emissions may require specific setups to be
performed in full compliance mode, e.g. in an anechoic chamber; viceversa pre-test measurements can be carried out with less expensive methods (see example at page 5).
A specifically designed pre-test EMI Receiver like the PMM
9010/30P used in debugging can significantly improve the results and speed up the development time of new products.
PMM 9010/30P’s High Frequency section allows for Radiated
Emissions pre-testing in the frequency range 30 MHz-3 GHz, featuring all the CISPR and MIL-STD RBW filters and Detectors, including those of recent introduction in standards (C-AVG,
RMS AVG, APD).
The new model PMM 9010/30P is based on the same structure of the PMM 9010: the first compact, worldwide successful Fully Digital
EMI Receiver.
3
PMM 9010/30P
Conducted
Conducted emission in Full Compliance mode
Reference Standard Generic Standards Product Standards
4
Receiver:
CISPR 16-1-1
IEC61000-6-3
IEC61000-6-4
MIL-STD 461
EN 55011-CISPR 11
EN
EN
EN
55014-
55015-CISPR 15
55022-CISPR
14
15
55022-CISPR 22
AMN (Artificial Mains Network)
Probes
16-1-2
Interference
Voltage
Interference
Voltage
0,15 30 MHz
Remark:
All above Standards appears as undated for the PMM 9010/30P can updated future changes.
Typical set-up configuration for EMI conducted measurements; construction details can be found in CISPR-16-1-2.
Its realization can be afforded by most of the laboratories and industries with limited costs.
40 cm bonded to plane
Power cord
40 cm
80 cm
Coax clable
TIPS:
• Care perfect groundings: loose or longer connections are likely to cause severe measuring errors
• Replace any cable looking crushed, overbent or scratched
• Make sure of all connections being tightened and efficient
• Make sure that LISNs are compliant to CISPR-16-1-2. Most of the old PMM
LISN you may own can be upgraded: contact Factory for details
• LISNs ground currents may be hazardous and/or cause ground protection systems to switch off. An isolation transformer of rated current adequate to the equipment under test between mains and LISN input is recommended or mandatory
• Keep connections between Receiver and LISN as short as possible;
9010/30P can be easily placed close by it and operated by battery
• EUT line voltage transients may cause RF noise exceeding the receiver max input level: make sure there are no heavy machinery, contactors, power inverters, inductive or capacitive loads etc. on the same power line or nearby
• An external attenuator at RF input may be useful in case of doubt about the max noise level
• EMI mains filters can be useful to avoid noise pick-up by LISN & cables.
Excessive ambient noise may require moving into a shielded room
• Sweep mode + Smart Detector function can dramatically reduce the test time
• Analyzer and Manual modes allow for debugging critical frequencies or spectrum portions
• The Audio Output can be helpful in debugging to discriminate the noise source (other machinery connected to the same line, broadcast, etc.) receiver
> 80 cm
Radiated emission measurements of luminaries by means of the Loop Antenna according to CISPR15/
EN 55015.
Radiated
Radiated Emissions in Pre-Test mode (VHQP-C)
Conducted
Conducted emission in Full Compliance mode
Radiated Emissions in Pre-Test mode (VHQP-C)
Reference Standard Generic Standards Product Standards
Receiver:
CISPR 16-1-1
IEC61000-6-3
IEC61000-6-4
EN…
MIL-STD 461
CISPR 14
55015-CISPR 15
EN 55022-CISPR 22
EN 55025-CISPR 25
Clamp:
CISPR 16-1-3
Test
CISPR
Site:
16-1-4
Radiated power
emissions
GHz
Radiated
and
30
Remark:
All
above Standards appears as undated for the PMM 9010/30P can be updated to future changes.
power
emissions
MHz GHz
Open Air Test Site (OATS) for radiated emissions
5 m absorbing clamp mains insulated platform
> 80 cm
USB
PW
9010/30P
DUT test receiver coax able
Radiated power measurement set-up with absorbing clamp
TIPS:
• Optimal conditions may allow for debugging even in unshielded environment
• Reduce ambient noise by proper antenna orientation
• Care perfect grounding
• Keep cable connection between receiver and antenna as short as possible: PMM
9010/30P can be easily placed near the antenna
• To save time start tests in Analyzer mode with selected Limit & max hold on, then use
Sweep mode with Smart Detector to focus on critical frequencies only
• Measurements should be taken at min. two heights on four sides, possibly at 3 m distance
• Biconic and Log-periodic antennas are most suitable; see PMM AS-02 Antenna Set
• Consider G-TEM cells for both emission (debugging) and immunity (full compliance)
• For better repeatability reduce variables (e.g. same equipments positioning, cables layout, etc.)
• Demodulated AF output can be helpful to discriminate the noise source (e.g. other machinery, broadcast, etc.) in debugging
• Ambient reflections can increase or decrease results unpredictably
• Recognize other possible RF sources hidden in the environment
• Noise cancellation methods (e.g. via pre-scan and post-processing) may be uneffective if not leading to wrong measurements
• Near field probes may be useful for bench debugging of subassemblies
• In case of excessive ambient noise move into a shielded/anechoic room
G-TEM cell for emission and immunity
Compact Shielded Room
5
PMM 9010/30P
PMM 9010/30P helps saving your time & money, today and tomorrow
•
•
• Expandable any time, in any
• Fits in all requirements,
Brings
More
requirements,
value-added
any
from
innovation, quality,
place, by
reliability
yourself
debugging to final, full compliant tests
than any other solution
30 - 3000 MHz
Extension Receiver
Module PMM 9030
30 - 6000 MHz
Extension Receiver
Module PMM 9060
4-Channel Click
Analyzer Module
6 - 18 GHz
Extension Receiver
Module PMM 9180
1-Channel Click
Analyzer Optional
Function (Internal)
6
PMM’s unique Extension Receiver Modules are real EMI Receivers fully compliant to CISPR 16-1-1 that boost the performance of PMM 9010/30P main unit to the edge of the most sophisticated radiated emission measurements up to 18 GHz. Higher frequency bands are planned.
• Fiber Optic Digital Link for interference-free measurements
• Direct connection to the antenna
• Better overall uncertainty without cable loss
• Automatic Discontinuous Disturbance (Click) Analyzer full compliant to CISPR 14-1, including all latest
Exceptions requirements (C-AVG detector)
For further details please consult the brochures PMM 9010/9030/9060 and 9180 downloadable from our web page: www.narda-sts.it
Technical specifications
Operating modes
Frequency range
Resolution
Accuracy
RF input
VSWR
Attenuator
Pulse limiter
Preamplifier
Max input level without damage
Sinewave AC voltage
Pulse spectral density
Preselector
IF RBW
Normal
CISPR-16-1-1
MIL-STD-461 (option)
Noise level
Spurious response
Mode: Sweep RBW: Auto
Preamp.: ON RF Att.: 0 dB
Detectors
(simultaneous on PMM
Emission Suite)
10
10
Hz
Hz
0,1 Hz
20 dB
–
to
<1 ppm
30
30
MHz
MHz
Input
50 Ω, BNC female
<1,2 (10 dB RF att.)
0 to 35 dB, 5 dB step
Built-in, selectable
1 x LP; 6 x BP filters
(3 dB BW) 3, 10, 30, 100, 300 kHz
200 Hz; 9 kHz
10, 100 Hz; 1, 10 kHz
(Preamplifier ON)
9 – 150 kHz
RBW 200 Hz, QP <-8 dBµV
RBW 200 Hz, Avg <-15 dBµV
0,15 – 30 MHz
RBW 9 kHz, QP <-4 dBµV
RBW 9 kHz, Avg <-10 dBµV
<0 dBµV; < 10 dBµV over 150 kHz
<2 ppm
50 Ω, N-F
<1,2 ; <2 over 1 GHz
0 to 50 dB, 2 dB step
n.a.
n.a.
n.a.
MHz
Scanning receiver
Manual tuning
Analyzer
137 dbµV – 1 W
97 dbµV/MHz
–
receiver
3 GHz Input
(6 dB BW) 3, 10, 30, 100, 300 kHz, 1 MHz
9 kHz, 120 kHz, 1 MHz
100 kHz, 1 MHz
30 MHz – 3 GHz
RBW 120 kHz, QP <8 dBµV
RBW 120 kHz, Avg <4 dBµV
< 15 dBµV
Peak, Quasi-Peak
C-Average, Average
RMS-Average (*) , RMS
APD
Smart Detector function
Hold time
Stand-alone display &
measure functions
Measuring units
Stand-alone
PMM Emission Suite
Displayed dynamic
Marker; marker peak; marker to center; highest peaks;
Move peak to Analyzer & Manual modes
Store & Load:
- up to 11 traces (sweep mode)
- two panels
- 4 conversion factors
Built-in limits: CISPR 11, 14, 22 (can be changed by the PMM Emission Suite)
Battery charge and voltage
Display style, contrast, backlight
Click functions (option required)
1 ms to 30 s
Measurement accuracy
Autocalibration
(1)
Demodulation
I/O Interface
RF output, rear panel
Frequency range
Level
Level accuracy (10 Hz to 30 MHz)
Operating temperature
S/N > 20 dB
10 Hz to 9 kHz ±1,0 dB
9 kHz to 30 MHz ±1,0 dB
Internal reference source
AM with variable volume dBm, dBµV dBµA, dBpW, dBµV/m, dBµA/m
80, 100, 120 dB selectable
S/N > 20 dB
30 MHz to 1 GHz ±1,0 dB
1 to 3 GHz ±1,5 dB
n.a.
USB; RS-232
High Speed Optical
User Port (drives PMM LISNs/accessories)
Bluetooth (optional)
50 Ω, BNC fem.
CW: 10 Hz to 50 MHz; Tracking mode: 10 Hz – 30 MHz
60 to 90 dBµV (0.1 dB step)
± 0,5 dB
0° to 40°C
Power supply
Battery operation time
Dimensions
Weight (including battery option)
AC universal adapter/charger
External 10 - 15 Vdc, 2.5A
Li-Ion rechargeable plug-in battery (Option)
3 h (2)
235x105x335 mm
4,95 kg
(1) RF front-end only. All RBW filters and detectors are digital and do not require any re-calibration, any time.
(2) Minimum value; may be higher in relation with the selected operating mode.
(*) rms-average detector manufactured under license of Rohde & Schwarz GmbH & Co. KG.
7
PMM 9010/30P
8
Powerful SCAN TABLE that meets the most demanding applications for flexibility
PEMS
PMM Emission Suite
The PMM Emission Suite comes with PMM 9010/30P EMI receiver to expand its applications:
• Full control of all auto and manual Receiver functions
• Real-time display on PC
• One-click operating mode change: Scan/Sweep, Analyzer,
Manual
• Editing of Standard Limits
• Creation of Correction Factors Tables for ancillary
equipments (antennas, cables, etc.)
• Retrieve, save, recall and compare measurements
• Simultaneous Marker on all Detectors and Zoom
• “n” Highest Peaks Finder and Scan Table generation
• Measured LISN lines scroll by mouse wheel
• GTEM correlation to OATS
• Warning messages for incorrect settings
The pictures show examples of the most commonly used functions, some of them peculiar of PMM’s receivers.
Spectrum Analyzer
Function
The very high scan speed (<100 ms for fullspan 10 Hz-30 MHz @ IF resolution 300 kHz), is very useful for any kind of debugging with the outstanding performances of an EMIspecific instrument.
Scalar Network Analyzer
Function
A powerful function useful for designers and test engineers provides an easy-to-use scalar network analyzer for characterizing components, filters and much more.
Correlates radiated emission tests to those performed in an OATS.
Fast and precise, this feature is already built in the PMM software.
IEC/EN 55015
(Lighting equipment)
Specific function featuring simultaneous
Avg - Q-Peak Smart Detector for complete measurement in one single scan.
Automatic measurement and calculation of
Insertion Loss.
Automatize and speeds up the tests with
Large Loop Antenna.
IEC 62493
(Human exposure to EMF)
The preset scan table - according to IEC
62493 - allows for immediate measurement with the Van Der Hoofden test-head; just one-click for automatic calculation of factor F.
PEAKS FINDER with generation of Frequency tables
COMPARE function of stored measurements
Automatic
Pass/Fail
IEC 62493 Test
ANALYZER mode for fast and accurate debugging
MANUAL mode: up to four simultaneous detectors
Just one click to move across functions
9
10
PMM 9010/30P
Ancillary Equipment & Accessories
Conducted measurements
AMN-Artificial
Mains
Networks
Also known as LISN (Line Impedance Stabilization
these equipments are used for RF
measurements in the frequency range from 9 kHz to 30 MHz on AC single and three-phase lines, from DC to 60 Hz. The V-network design is fully compliant with latest CISPR publ. 16 and FCC part 15 regulation.
General features:
• automatic line switching control from PMM EMI
receivers
• artificial hand circuit
• 150 kHz high-pass filter control
Most common models:
Model L1-150
Single-phase
250 VAC/600 VDC,
150 A
Model L2-L216-B
Single-phase 230 V/16 A
Models L3-64, L3-100
3-ph. 4-wires
400 V/64 A, 100 A
Model L3-32
3-phase 4-wires
400 V/32 A
Other models up to 500 A on request
Model L1-150M 0,1 - 200 MHz 250
VAC/600 VDC, 150 A
Multi-standard: ISO 7637-2
ISO 11452-4
CISPR 16-1-2 • CISPR 25
Passive Probes mod. SHC
When LISN are not applicable (higher line current) the conducted interferences are measured by means of a high impedance voltage probe.
• compliant with CISPR 16-1-2
• internal resistance >1500 Ω
• 1 kV AC max. voltage
Radiated measurements
Measuring Antennas
Measuring antennas of excellent electrical performances and mechanical stability are essential in radiated measurements for repeatable and reliable results.
General features:
• Robust construction in passivated Aluminium
• Weather-resistant paint
• Lightweight
• Hi-quality RF connectors
• Easy to assembly and store
The following models are perfect for operation in conjunction with the EMI Analyzer PMM 9010/30P:
Model BC-01
Biconical Antenna
30-200 MHz
Model LP-02
Log Periodic Antenna
200-3000 MHz
PMM AS-02 is a complete, convenient Antenna System inclusive of the models BC-01,
LP- 02 and RF cable stored in a practical, robust soft carrying case.
Model TR-01
Antenna Tripod Extensible 60 - 180 cm
Mounting adapter for quick horizontal to vertical polarization change
Other EMC valuable solutions by PMM
RF Generators up to 3 GHz
RF Power Meters 9 kHz - 3 GHz, USB and Fiber Optic
Systems for conducted and radiated immunity
VDH-01 Van der Hoofden test-head for lighting equipment
Magnetic Immunity
Test System for
IEC/EN 61000-4-8
Isotropic RF Field Probes up to 40 GHz.
Accredited calibration
Accessories for
IEC/EN 55015 - IEC 62493
Services for your PMM EMI Receiver
• Free updating of PC Software and instrument Firmware from our web page.
• Failsafe, User-upgradeable Firmware whenever required by future standards, measuring features and test solutions.
An exclusive “parking memory”makes upgrading the PMM Receivers Firmware totally failsafe.
• The fastest and most convenient recalibration and service ever seen before, thanks to the modular construction
and pre-calibrated subassemblies; provided by our Qualified Sales Partners worldwide network.
Other Services
• Calibration of antennas in OATS in the 20 MHz - 18 GHz frequency range.
• Accredited calibrations for RF power, Frequency, Electric and Magnetic fields strength, Magnetic induction
• Notified Body NB 2050 as per 2004/108/EC directive.
11
Ordering information
9010-30P EMI receiver 10 Hz - 3 GHz, two RF inputs:
- 9 kHz - 30 MHz, CISPR 16-1-1 full-compliance
- 30 MHz - 3 GHz, Pre-compliance with CISPR 16-1-1 RBW and Detectors
Including:
- internal tracking generator 10 Hz- 30 MHz
- AC adapter (mod. 9010/AC)
- PC software PMM Emission Suite
- Standard Calibration Certificate
- RS232/USB adapter (for FW upgrades)
- N-BNC adapter
- Control cables (USB, RS-232), BNC-BNC cable
Optional accessories and functions
9010/MIL
9010/CLICK
9010/CLICK4E
9010/RAV
9010/BTA
BP01
9010/AC
9010/CC
9010-RMA
9010/UKAS
9010/UKAS-Click
MIL-STD-461 RBW Filters
1-channel Click Analyzer function, CISPR 14-1 full-compliance, including:
- Switching Operation Box, control cables - 2x20 dB attenuator
NOTE: field-installable function (advice S/N for upgrading confirmation)
External box to connect to a receiver equipped with 9010/click option.
Allows four-channel simultaneous click measurements according to CISPR-14-1. AC power only
CISPR 16-1-1 RMS-Average Detector
RS-232 to BlueTooth adapter for 9010
Li-Ion Battery Pack
Additional AC adapter/charger for BP01 (one already supplied with 9010/30P)
Rigid carrying case
19” Rack Mount Adapter for one unit mod. 9010
UKAS CISPR-16-1-1 accredited calibration certificate for Bands A, B (9 kHz - 30 MHz)
UKAS accredited calibration certificate for 9010 + 9010/Click to CISPR-16-1-1 & CISPR-14-1
Bands C – D – E upgrades to full CISPR-16-1-1
9030
9060
Extension unit 30 MHz - 3 GHz (UKAS accredited calibration on option)
Extension unit 30 MHz - 6 GHz (UKAS accredited calibration on option)
Ancillary equipments
LISN - controlled by the PMM 9010/30P receiver to automatically select the lines to measure
• L1-150M: Single line LISN,150A, multi-standard CISPR 16-1-2, CISPR 25, ISO 11452-4, ISO 7637-2
• L1-150 Single line LISN, 150A
• L2-16B Two lines, Single phase, 16A LISN
• L3-64 Four lines, 3-phase, 64A LISN
• L3-100 Four lines, 3-phase, 100A LISN
CISPR 16-1-2 Voltage Probes
• L3-32 Four lines, 3-phase, 32A LISN • L3-500 Four lines, 3-phase, 350A LISN
• SHC-1/1000 35 dB CISPR, 1500
Ω, 1 kV
• SHC-2 30 dB CISPR, 1500
Ω, 300 V
Antennas
• RA-01-HV: Rod Antenna (10 kHz – 30 MHz) suitable for EMI measurements of Overhead Power Lines
• RA-01 Rod Antenna 9 kHz-30 MHz
• BC-01 Biconical Antenna 30-200 MHz
• TR-01 Tripod for PMM Antennas
Components for: IEC/EN 55015 (CISPR 15) IEC 62493
• LP-02 Log Periodic Antenna 200 MHz - 3 GHz
• LP-03 Log Periodic Antenna 800 MHz - 6 GHz
• F-300M-16 CDN 150 kHz - 300 MHz; 250 VAC - 16A, 50/60 Hz
for power circuitry testing with phase, neutral and PE
• RF-300 3-axis Loop Antenna System for CISPR 15 EN 55015
• RF-300C Calibration kit for RF-300
• SBRF4: 4-ways RF Switching Box for automatic operation of RF-300
• TRF-1 Balance/unbalance transformer
• DL-xx Dummy Lamps according to the Standard
• VDH-01 Van der Hoofden test-head with protection network a brand of
® Names and Logo are registered trademarks of Narda Safety Test Solutions GmbH and L3 Communications Holdings, Inc. – Trade names are trademarks of the owners.
® Names and Logo are registered trademarks of Narda Safety Test Solutions GmbH and L3 Communications Holdings, Inc. – Trade names are trademarks of the owners.
Narda Safety Test Solutions srl
Via Leonardo da Vinci, 21/23
20090 Segrate (MI) ITALY
Phone: +39 02 26 998 71
Fax: +39 02 26 998 700
E-Mail: [email protected]
www.narda-sts.it
* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project
Related manuals
advertisement