Rigol DS1104Z 100 MHz Digital Oscilloscope User manual

Add to my manuals
262 Pages

advertisement

Rigol DS1104Z 100 MHz Digital Oscilloscope User manual | Manualzz

RIGOL

User’s Guide

MSO1000Z/DS1000Z Series

Digital Oscilloscope

Jul. 2014

RIGOL Technologies, Inc.

RIGOL

Guaranty and Declaration

Copyright

© 2014 RIGOL Technologies, Inc. All Rights Reserved.

Trademark Information

RIGOL is a registered trademark of RIGOL Technologies, Inc.

Publication Number

UGA19104-1110

Notices

RIGOL products are covered by P.R.C. and foreign patents, issued and pending.

RIGOL reserves the right to modify or change parts of or all the specifications and pricing policies at company’s sole decision.

 Information in this publication replaces all previously corresponding material.

 Information in this publication is subject to change without notice.

RIGOL shall not be liable for either incidental or consequential losses in connection with the furnishing, use or performance of this manual as well as any information contained.

 Any part of this document is forbidden to be copied, photocopied or rearranged without prior written approval of RIGOL.

Product Certification

RIGOL guarantees this product conforms to the national and industrial standards in

China as well as the ISO9001:2008 standard and the ISO14001:2004 standard.

Other international standard conformance certification is in progress.

Contact Us

If you have any problem or requirement when using our products or this manual, please contact RIGOL.

E-mail: [email protected]

Website: www.rigol.com

MSO1000Z/DS1000Z User’s Guide I

RIGOL

Safety Requirement

General Safety Summary

Please review the following safety precautions carefully before putting the instrument into operation so as to avoid any personal injury or damage to the instrument and any product connected to it. To prevent potential hazards, please use the instrument only specified by this manual.

Use Proper Power Cord.

Only the power cord designed for the instrument and authorized for use within the local country could be used.

Ground the Instrument.

The instrument is grounded through the Protective Earth lead of the power cord. To avoid electric shock, it is essential to connect the earth terminal of the power cord to the Protective Earth terminal before connecting any inputs or outputs.

Connect the Probe Correctly.

If a probe is used, do not connect the ground lead to high voltage since it has the isobaric electric potential as ground.

Observe All Terminal Ratings.

To avoid fire or shock hazard, observe all ratings and markers on the instrument and check your manual for more information about ratings before connecting the instrument.

Use Proper Overvoltage Protection.

Make sure that no overvoltage (such as that produced by a thunderstorm) can reach the product, or else the operator might be exposed to the danger of electrical shock.

Do Not Operate without Covers.

Do not operate the instrument with covers or panels removed.

Do Not Insert Anything into the Holes of Fan.

Do not insert anything into the holes of the fan to avoid damaging the instrument.

Use Proper Fuse.

Please use the specified fuses.

Avoid Circuit or Wire Exposure.

Do not touch exposed junctions and components when the unit is powered.

II MSO1000Z/DS1000Z User’s Guide

RIGOL

Do Not Operate with Suspected Failures.

If you suspect damage occurs to the instrument, have it inspected by qualified service personnel before further operations. Any maintenance, adjustment or replacement especially to circuits or accessories must be performed by RIGOL authorized personnel.

Keep Well Ventilation.

Inadequate ventilation may cause an increase of temperature which would damage the device. So please keep the instrument well ventilated and inspect the intake and fan regularly.

Do Not Operate in Wet Conditions.

In order to avoid short circuiting to the interior of the device or electric shock, please do not operate the instrument in a humid environment.

Do Not Operate in an Explosive Atmosphere.

In order to avoid damage to the device or personal injury, it is important to operate the device away from an explosive atmosphere.

Keep Product Surfaces Clean and Dry.

To avoid the influence of dust and/or moisture in the air, please keep the surface of the device clean and dry.

Electrostatic Prevention.

Operate the instrument in an electrostatic discharge protective area to avoid damage induced by static discharges. Always ground both the internal and external conductors of the cable to release static before connecting it to the instrument.

Proper Use of Battery.

If a battery is supplied, it must not be exposed to high temperature or in contact with fire. Keep it out of the reach of children. Improper change of battery (note: lithium battery) may cause explosion. Use RIGOL specified battery only.

Handling Safety.

Please handle with care during transportation to avoid damage to buttons, knob interfaces and other parts on the panels.

MSO1000Z/DS1000Z User’s Guide III

RIGOL

Safety Terms and Symbols

Terms Used in this Manual. These terms may appear in this manual:

WARNING

Warning statements indicate the conditions or practices that could result in injury or loss of life.

CAUTION

Caution statements indicate the conditions or practices that could result in damage to this product or other property.

Terms Used on the Product. These terms may appear on the Product:

DANGER indicates an injury or hazard may immediately happen.

WARNING

CAUTION

indicates an injury or hazard may be accessible potentially. indicates potential damage to the instrument or other property might occur.

Symbols Used on the Product. These symbols may appear on the product:

Hazardous

Voltage

Safety

Warning

Protective

Earth

Terminal

Chassis

Ground

Test

Ground

IV MSO1000Z/DS1000Z User’s Guide

RIGOL

Allgemeine Sicherheits Informationen

Überprüfen Sie diefolgenden Sicherheitshinweise sorgfältigumPersonenschädenoderSchäden am Gerätundan damit verbundenen weiteren Gerätenzu vermeiden. Zur Vermeidung vonGefahren, nutzen Sie bitte das

Gerät nur so, wiein diesem Handbuchangegeben.

Um Feuer oder Verletzungen zu vermeiden, verwenden Sie ein ordnungsgemäßes Netzkabel.

Verwenden Sie für dieses Gerät nur das für ihr Land zugelassene und genehmigte

Netzkabel.

Erden des Gerätes.

Das Gerät ist durch den Schutzleiter im Netzkabel geerdet. Um Gefahren durch elektrischen Schlag zu vermeiden, ist es unerlässlich, die Erdung durchzuführen. Erst dann dürfen weitere Ein- oder Ausgänge verbunden werden.

Anschluss einesTastkopfes.

Die Erdungsklemmen der Sonden sindauf dem gleichen Spannungspegel des

Instruments geerdet. SchließenSie die Erdungsklemmen an keine hohe Spannung an.

Beachten Sie alle Anschlüsse.

Zur Vermeidung von Feuer oder Stromschlag, beachten Sie alle Bemerkungen und

Markierungen auf dem Instrument. Befolgen Sie die Bedienungsanleitung für weitere

Informationen, bevor Sie weitere Anschlüsse an das Instrument legen.

Verwenden Sie einen geeigneten Überspannungsschutz.

Stellen Sie sicher, daß keinerlei Überspannung (wie z.B. durch Gewitter verursacht) das Gerät erreichen kann. Andernfallsbestehtfür den Anwender die

GefahreinesStromschlages.

Nicht ohne Abdeckung einschalten.

Betreiben Sie das Gerät nicht mit entfernten Gehäuse-Abdeckungen.

Betreiben Sie das Gerät nicht geöffnet.

Der Betrieb mit offenen oder entfernten Gehäuseteilen ist nicht zulässig. Nichts in entsprechende Öffnungen stecken (Lüfter z.B.)

Passende Sicherung verwenden.

Setzen Sie nur die spezifikationsgemäßen Sicherungen ein.

Vermeiden Sie ungeschützte Verbindungen.

Berühren Sie keine unisolierten Verbindungen oder Baugruppen, während das Gerät in Betrieb ist.

MSO1000Z/DS1000Z User’s Guide V

RIGOL

Betreiben Sie das Gerät nicht im Fehlerfall.

Wenn Sie am Gerät einen Defekt vermuten, sorgen Sie dafür, bevor Sie das Gerät wieder betreiben, dass eine Untersuchung durch qualifiziertes Kundendienstpersonal durchgeführt wird.Jedwede Wartung, Einstellarbeiten oder Austausch von Teilen am

Gerät, sowie am Zubehör dürfen nur von RIGOL autorisiertem Personal durchgeführt werden.

Belüftung sicherstellen.

Unzureichende Belüftung kann zu Temperaturanstiegen und somit zu thermischen

Schäden am Gerät führen. Stellen Sie deswegen die Belüftung sicher und kontrollieren regelmäßig Lüfter und Belüftungsöffnungen.

Nicht in feuchter Umgebung betreiben.

Zur Vermeidung von Kurzschluß im Geräteinneren und Stromschlag betreiben Sie das

Gerät bitte niemals in feuchter Umgebung.

Nicht in explosiver Atmosphäre betreiben.

Zur Vermeidung von Personen- und Sachschäden ist es unumgänglich, das Gerät ausschließlich fernab jedweder explosiven Atmosphäre zu betreiben.

Geräteoberflächen sauber und trocken halten.

Um den Einfluß von Staub und Feuchtigkeit aus der Luft auszuschließen, halten Sie bitte die Geräteoberflächen sauber und trocken.

Schutz gegen elektrostatische Entladung (ESD).

Sorgen Sie für eine elektrostatisch geschützte Umgebung, um somit Schäden und

Funktionsstörungen durch ESD zu vermeiden. Erden Sie vor dem Anschluß immer

Innen- und Außenleiter der Verbindungsleitung, um statische Aufladung zu entladen.

Die richtige Verwendung desAkku.

Wenneine Batterieverwendet wird, vermeiden Sie hohe Temperaturen bzw. Feuer ausgesetzt werden.Bewahren Sie es außerhalbder Reichweitevon Kindern auf.UnsachgemäßeÄnderung derBatterie(Anmerkung:Lithium-Batterie)kann zu einer

Explosion führen. VerwendenSie nur von RIGOLangegebenenAkkus.

Sicherer Transport.

Transportieren Sie das Gerät sorgfältig (Verpackung!), um Schäden an

Bedienelementen, Anschlüssen und anderen Teilen zu vermeiden.

VI MSO1000Z/DS1000Z User’s Guide

RIGOL

Sicherheits Begriffe und Symbole

Begriffe in diesem Guide. Diese Begriffe können in diesem Handbuch auftauchen:

WARNING

Die Kennzeichnung WARNING beschreibt Gefahrenquellen die leibliche

Schäden oder den Tod von Personen zur Folge haben können.

CAUTION

Die Kennzeichnung Caution (Vorsicht) beschreibt Gefahrenquellen die

Schäden am Gerät hervorrufen können.

Begriffe auf dem Produkt. Diese Bedingungen können auf dem Produkt erscheinen:

DANGER weist auf eine Verletzung oder Gefährdung hin, die sofort geschehen kann.

WARNING weist auf eine Verletzung oder Gefährdung hin, die möglicherweise nicht sofort geschehen.

CAUTION bedeutet, dass eine mögliche Beschädigung des Instruments oder anderer Gegenstände auftreten kann.

Symbole auf dem Produkt. Diese Symbole können auf dem Produkt erscheinen:

GefährlicheS

pannung

Sicherheits-

Hinweis

Schutz-erde Gehäusemasse Erde

MSO1000Z/DS1000Z User’s Guide VII

RIGOL

Measurement Category

Measurement Category

MSO1000Z/DS1000Z series digital oscilloscopes can make measurements in

Measurement Category I.

WARNING

This oscilloscope can only be used for measurements within its specified measurement categories.

Measurement Category Definitions

Measurement category I is for measurements performed on circuits not directly connected to MAINS. Examples are measurements on circuits not derived from

MAINS, and specially protected (internal) MAINS derived circuits. In the latter case, transient stresses are variable. For that reason, the transient withstand capability of the equipment is made known to the user.

Measurement category II is for measurements performed on circuits directly connected to low voltage installations. Examples are measurements on household appliances, portable tools and similar equipment.

Measurement category III is for measurements performed in building installations.

Examples are measurements on distribution boards, circuit-breakers, wiring

(including cables, bus-bars, junction boxes, switches and socket-outlets) in fixed installations, equipment for industrial use and some other equipment. For example, stationary motors with permanent connection to a fixed installation.

Measurement category IV is for measurements performed at the source of a low-voltage installation. Examples are electricity meters and measurements on primary overcurrent protection devices and ripple control units.

VIII MSO1000Z/DS1000Z User’s Guide

RIGOL

Ventilation Requirement

This oscilloscope uses fan to force cooling. Please make sure that the air intake and exhaust areas are free from obstructions and have free air. When using the oscilloscope in a bench-top or rack setting, provide at least 10 cm clearance beside, above and behind the instrument for adequate ventilation.

WARNING

Inadequate ventilation may cause a temperature increase which can damage the instrument. So please keep the instrument well ventilated during operation and inspect the intake and fan regularly.

MSO1000Z/DS1000Z User’s Guide IX

RIGOL

Working Environment

Temperature

Operating: 0℃ to +50℃

Non-operating: -40℃ to +70℃

Humidity

0℃ to +30℃: ≤95% relative humidity

+30℃ to +40℃: ≤75% relative humidity

+40℃ to +50℃: ≤45% relative humidity

WARNING

To avoid short circuits inside the instrument or electric shocks, please do not operate in humid environment.

Altitude

Operating: less than 3 km

Non-operating: less than 15 km

Installation (overvoltage) Category

This product is powered by mains conforming to installation (overvoltage) category

II.

WARNING

Make sure that no overvoltage (such as that produced by a thunderstorm) can reach the product, or else the operator might be exposed to the danger of electric shock.

Installation (overvoltage) Category Definitions

Installation (overvoltage) category I refers to signal level which is applicable to equipment measurement terminals connected to the source circuit. In these terminals, precautions are done to limit the transient voltage to the corresponding low level.

Installation (overvoltage) category II refers to the local power distribution level which is applicable to equipment connected to the AC line (AC power).

Pollution Degree

Degree 2

Pollution Degree Definitions

Pollution degree 1: No pollution or only dry, non-conductive pollution occurs. The pollution has no influence. For example, a clean room or air-conditioned office environment.

X MSO1000Z/DS1000Z User’s Guide

RIGOL

Pollution degree 2: Normally only dry, non-conductive pollution occurs. Occasionally a temporary conductivity caused by condensation may occur. For example, general indoor environment.

Pollution degree 3: Conductive pollution occurs, or dry, non-conductive pollution occurs which becomes conductive due to condensation which is expected. For example, sheltered outdoor environment.

Pollution degree 4: Pollution that generates persistent conductivity through conductive dust, rain, or snow. For example, outdoor locations.

Safety Class

Class 1 – Grounded Product

MSO1000Z/DS1000Z User’s Guide XI

RIGOL

General Care and Cleaning

General Care:

Do not store or leave the instrument at places where the instrument will be exposed to direct sunlight for long periods of time.

Cleaning:

Clean the instrument regularly according to its operating conditions. To clean the exterior surface:

1. Disconnect the instrument from all power sources.

2. Clean the loose dust on the outside of the instrument with a lint- free cloth (with mild detergent or water). When cleaning the LCD, take care to avoid scarifying it.

CAUTION

To avoid damage to the instrument, do not expose it to corrosive liquids.

WARNING

To avoid injury resulting from short circuit, make sure the instrument is completely dry before reconnecting it to a power source.

Environmental Considerations

The following symbol indicates that this product complies with the WEEE Directive

2002/96/EC.

Product End-of-Life Handling

The equipment may contain substances that could be harmful to the environment or human health. In order to avoid release of such substances into the environment and harm to human health, we encourage you to recycle this product in an appropriate system that will ensure that most of the materials are reused or recycled appropriately. Please contact your local authorities for disposal or recycling information.

XII MSO1000Z/DS1000Z User’s Guide

RIGOL

MSO1000Z/DS1000Z Series Overview

MSO1000Z is a mixed signal digital oscilloscope aimed at the embedded design and test fields. It allows users to measure analog and digital signals at the same time.

MSO1000Z/DS1000Z series is a multifunctional and high-performance digital oscilloscope designed on the basis of the UltraVision technique developed by RIGOL.

Featuring extremely high memory depth, wide dynamic range, clear display, excellent waveform capture rate and comprehensive triggering functions, it is a useful commissioning instrument for various fields such as communication, aerospace, defense, embedded systems, computers, research and education.

MSO1000Z/DS1000Z is the one with the most comprehensive functions and the most outstanding specifications among the 100 MHz bandwidth digital oscilloscopes.

Main features:

 1 GSa/s real-time sample rate of the analog channels; up to 24 Mpts memory depth (option) and 12 Mpts standard memory depth

 1 GSa/s real-time sample rate of the digital channels

 100 MHz, 70 MHz and 50 MHz analog channel bandwidth

 4+16-channel MSO1000Z series mixed signal digital oscilloscope

 Built-in dual-channel 25 MHz signal source (only applicable to MSO1000Z-S and

DS1000Z-S models)

 30 000 wfms/s (dots display) waveform capture rate

 Real-time hardware waveform recording and playback functions; up to 60000 frames of waveform can be recorded

 Multi-degree gray scale display

 Low base noise, 1 mV/div to 10 V/div ultra-wide vertical dynamic range

 7.0 inch WVGA (800*480) TFT LCD, with ultra-wide screen, vivid picture, low power consumption and long service life

 Adjustable waveform brightness

 Auto setting of waveform display (AUTO)

 Up to 15 kinds of trigger functions, including various protocol triggers

 Standard parallel decoding, support multiple serial decoding options

 Auto measurement of 33 waveform parameters (with statistics)

 Fine delayed sweep function

 Built-in FFT function

 Multiple waveform math operation functions

 Pass/fail test function

 Standard interfaces: USB Device, USB HOST, LAN, Aux, GPIB (extended via the

USB HOST interface)

 Conform to LXI CORE 2011 DEVICE class instrument standards; enable quick, economic and efficient creation and reconfiguration of test system

 Support remote command control

 Built-in help to facilitate information acquisition

 Support multiple languages and Chinese/English input

 Novel and delicate industrial design and easy operation

MSO1000Z/DS1000Z User’s Guide XIII

RIGOL

Document Overview

Main Topics of this Manual:

Chapter 1 Quick Start

Introduce the preparations before using the oscilloscope and provide a basic introduction of the instrument.

Chapter 2 To Set the Vertical System

Introduce the vertical system functions of the oscilloscope.

Chapter 3 To Set the Horizontal System

Introduce the horizontal system functions of the oscilloscope.

Chapter 4 To Set the Sample System

Introduce the sample system functions of the oscilloscope.

Chapter 5 To Trigger the Oscilloscope

Introduce the trigger mode, trigger coupling, trigger holdoff, external trigger and various trigger types of the oscilloscope.

Chapter 6 MATH and Measurement

Introduce how to make math operation, cursor and auto measurements.

Chapter 7 Digital Channel

Introduce how to use the digital channels of the mixed signal digital oscilloscope.

Chapter 8 Protocol Decoding

Introduce how to decode the input signal using those common protocols.

Chapter 9 Reference Waveform

Introduce how to compare the input waveform with the reference waveform.

Chapter 10 Pass/Fail Test

Introduce how to monitor the input signal using the Pass/Fail test.

Chapter 11 Waveform Record

Introduce how to analyze the input signal using waveform record.

Chapter 12 Display Control

Introduce how to control the display of the oscilloscope.

Chapter 13 Signal Source

Introduce how to use the built-in signal source.

XIV MSO1000Z/DS1000Z User’s Guide

RIGOL

Chapter 14 Store and Recall

Introduce how to store and recall the measurement result and the setting of the oscilloscope.

Chapter 15 Accessibility Setting

Introduce how to set the remote interfaces and system-related functions.

Chapter 16 Remote Control

Introduce how to control the oscilloscope remotely.

Chapter 17 Troubleshooting

Introduce how to deal with the common failures of the oscilloscope.

Chapter 18 Specifications

Provide the specifications and general specifications of the oscilloscope.

Chapter 19 Appendix

Provide common information such as the options and accessories.

Format Conventions in this Manual:

1. Key

The front panel keys are denoted by the format of “Button Name (Bold) + Text

Box”, for example, Utility denotes the “Utility” key.

2. Menu

The menu items are denoted by the format of “Menu Word (Bold) + Character

Shading”, for example, System denotes the “System” menu item under Utility.

3. Operation Step

The next step of operation is denoted by an arrow “”, for example, Utility

System denotes pressing Utility at the front panel and then pressing System.

4. Knob

Label

HORIZONTAL

SCALE

HORIZONTAL

POSITION

TRIGGER

LEVEL

Knob Label

Horizontal Scale Knob VERTICAL

SCALE

Horizontal Position

Knob

VERTICAL

POSITION

Trigger Level Knob

--

Knob

Vertical Scale Knob

Vertical Position

Knob

--

MSO1000Z/DS1000Z User’s Guide XV

RIGOL

Content Conventions in this Manual:

MSO1000Z/DS1000Z series includes the following models. This manual takes

MSO1104Z-S for example to illustrate the functions and performances of

MSO1000Z/DS1000Z series.

Model

MSO1104Z-S

MSO1074Z-S

MSO1104Z

MSO1074Z

DS1104Z-S

DS1074Z-S

DS1104Z

DS1074Z

DS1054Z

Analog

Bandwidth

100 MHz

70 MHz

100 MHz

70 MHz

100 MHz

70 MHz

100 MHz

70 MHz

50 MHz

4

4

4

4

4

4

4

4

Number of

Analog

Channels

4

Number of Source

Channels

2

2

--

--

--

2

2

--

--

--

--

--

--

16

16

16

--

Number of

Digital

Channels

16

Manuals of this Product:

The manuals of this product include the quick guide, user’s guide, programming guide and data sheet. The latest versions of the manuals can be downloaded from www.rigol.com

.

XVI MSO1000Z/DS1000Z User’s Guide

Contents RIGOL

Contents

Guaranty and Declaration ......................................................................... I

Safety Requirement ................................................................................ II

General Safety Summary ........................................................................... II

Safety Terms and Symbols ....................................................................... IV

Allgemeine Sicherheits Informationen ......................................................... V

Sicherheits Begriffe und Symbole ............................................................. VII

Measurement Category .......................................................................... VIII

Ventilation Requirement ........................................................................... IX

Working Environment ............................................................................... X

General Care and Cleaning ...................................................................... XII

Environmental Considerations .................................................................. XII

MSO1000Z/DS1000Z Series Overview ................................................ XIII

Document Overview ............................................................................. XIV

Chapter 1 Quick Start ......................................................................... 1-1

General Inspection ................................................................................ 1-2

Appearance and Dimensions ................................................................... 1-3

To Prepare the Oscilloscope for Use ......................................................... 1-4

To Adjust the Supporting Legs .......................................................... 1-4

To Connect to Power Supply............................................................. 1-4

Power-on Inspection ....................................................................... 1-5

To Connect the Probe ...................................................................... 1-5

Function Inspection ......................................................................... 1-7

Probe Compensation ....................................................................... 1-8

Front Panel Overview ............................................................................. 1-9

Rear Panel Overview ............................................................................. 1-10

Front Panel Function Overview............................................................... 1-12

VERTICAL ..................................................................................... 1-12

Logic Analyzer ............................................................................... 1-13

Signal Source ................................................................................ 1-13

HORIZONTAL ................................................................................ 1-14

TRIGGER ...................................................................................... 1-15

CLEAR .......................................................................................... 1-15

AUTO............................................................................................ 1-15

RUN/STOP .................................................................................... 1-16

SINGLE ......................................................................................... 1-16

Multifunction Knob ......................................................................... 1-16

Function Menus ............................................................................. 1-17

Print ............................................................................................. 1-18

User Interface ...................................................................................... 1-19

Parameter Setting Method ..................................................................... 1-23

To Use the Security Lock ....................................................................... 1-24

MSO1000Z/DS1000Z User’s Guide XVII

RIGOL Contents

To Use the Built-in Help System ............................................................. 1-25

Chapter 2 To Set the Vertical System ................................................. 2-1

To Enable the Analog Channel ................................................................. 2-2

Channel Coupling ................................................................................... 2-2

Bandwidth Limit ..................................................................................... 2-3

Probe Ratio ........................................................................................... 2-3

Waveform Invert .................................................................................... 2-4

Vertical Scale ......................................................................................... 2-4

Amplitude Unit ....................................................................................... 2-5

Channel Label ........................................................................................ 2-5

Delay Calibration of the Analog Channel ................................................... 2-7

Chapter 3 To Set the Horizontal System ............................................ 3-1

Delayed Sweep ...................................................................................... 3-2

Time Base Mode .................................................................................... 3-3

YT Mode ......................................................................................... 3-3

XY Mode ......................................................................................... 3-3

Roll Mode ....................................................................................... 3-5

Chapter 4 To Set the Sample System ................................................. 4-1

Acquisition Mode .................................................................................... 4-2

Normal ........................................................................................... 4-2

Peak Detect .................................................................................... 4-2

Average .......................................................................................... 4-2

High Resolution ............................................................................... 4-3

Sin(x)/x ................................................................................................. 4-4

Sample Rate .......................................................................................... 4-4

Memory Depth ....................................................................................... 4-6

Antialiasing ............................................................................................ 4-7

Chapter 5 To Trigger the Oscilloscope ................................................ 5-1

Trigger Source ....................................................................................... 5-2

Trigger Mode ......................................................................................... 5-3

Trigger Coupling..................................................................................... 5-4

Trigger Holdoff ....................................................................................... 5-5

Noise Rejection ...................................................................................... 5-5

Trigger Type .......................................................................................... 5-6

Edge Trigger ................................................................................... 5-7

Pulse Trigger ................................................................................... 5-8

Slope Trigger ................................................................................. 5-10

Video Trigger................................................................................. 5-13

Pattern Trigger .............................................................................. 5-15

Duration Trigger ............................................................................ 5-17

TimeOut Trigger (Option) ............................................................... 5-19

Runt Trigger (Option) ..................................................................... 5-21

Window Trigger (Option) ................................................................ 5-23

Delay Trigger (Option).................................................................... 5-25

XVIII MSO1000Z/DS1000Z User’s Guide

Contents RIGOL

Setup/Hold Trigger (Option) ............................................................ 5-27

Nth Edge Trigger (Option) ............................................................... 5-29

RS232 Trigger (Option) ................................................................... 5-31

I2C Trigger (Option) ....................................................................... 5-33

SPI Trigger (Option) ....................................................................... 5-35

Trigger Output Connector ...................................................................... 5-37

Chapter 6 MATH and Measurement .................................................... 6-1

Math Operation ..................................................................................... 6-2

Addition ......................................................................................... 6-2

Subtraction .................................................................................... 6-3

Multiplication .................................................................................. 6-3

Division .......................................................................................... 6-4

FFT ............................................................................................... 6-6

“AND” Operation ............................................................................. 6-9

“OR” Operation .............................................................................. 6-10

“XOR” Operation ............................................................................ 6-11

“NOT” Operation ............................................................................ 6-12

Intg .............................................................................................. 6-13

Diff ............................................................................................... 6-14

Sqrt .............................................................................................. 6-15

Lg (Use 10 as the Base) ................................................................. 6-16

Ln ................................................................................................ 6-16

Exp .............................................................................................. 6-17

Abs .............................................................................................. 6-18

Fx Operation ................................................................................. 6-19

Math Operation Label ..................................................................... 6-19

Auto Measurement ............................................................................... 6-20

Quick Measurement after AUTO ...................................................... 6-20

One-key Measurement of 33 Parameters .......................................... 6-21

Frequency Counter Measurement .................................................... 6-26

Measurement Setting ..................................................................... 6-26

To Clear the Measurement .............................................................. 6-27

All Measurement ............................................................................ 6-28

Statistic Function ........................................................................... 6-29

Measurement History ..................................................................... 6-29

Measurement Result Display Type .................................................... 6-30

Cursor Measurement............................................................................. 6-30

Manual Mode ................................................................................. 6-31

Track Mode ................................................................................... 6-35

Auto Mode ..................................................................................... 6-1

XY Mode ........................................................................................ 6-2

Chapter 7 Digital Channel ................................................................... 7-1

To Select the Digital Channel .................................................................. 7-2

To Turn on/off the Digital Channel ........................................................... 7-2

Group Set ............................................................................................. 7-3

MSO1000Z/DS1000Z User’s Guide XIX

RIGOL Contents

To Set the Waveform Display Size ............................................................ 7-4

Reorder Setting ...................................................................................... 7-4

Auto View.............................................................................................. 7-4

To Set the Threshold .............................................................................. 7-4

To Set the Label ..................................................................................... 7-5

Probe Calibration .................................................................................... 7-5

Digital Channel Delay Calibration ............................................................. 7-6

Chapter 8 Protocol Decoding ............................................................. 8-1

Parallel Decoding ................................................................................... 8-2

RS232 Decoding (Option) ....................................................................... 8-6

I2C Decoding (Option) .......................................................................... 8-11

SPI Decoding (Option) .......................................................................... 8-15

Chapter 9 Reference Waveform ......................................................... 9-1

To Enable REF Function .......................................................................... 9-2

To Select REF Source .............................................................................. 9-2

To Adjust REF Waveform Display ............................................................. 9-2

To Save to Internal Memory .................................................................... 9-2

To Set the Color ..................................................................................... 9-3

To Reset the Reference Waveform ........................................................... 9-3

To Export to Internal or External Memory ................................................. 9-3

To Import from Internal or External Memory ............................................. 9-3

Chapter 10 Pass/Fail Test ............................................................ 10-1

To Enable Pass/Fail Test ........................................................................ 10-2

To Select Source .................................................................................. 10-2

Mask Range ......................................................................................... 10-2

Test and Output ................................................................................... 10-3

To Save the Test Mask .......................................................................... 10-3

To Load the Test Mask .......................................................................... 10-4

Chapter 11 Waveform Record ...................................................... 11-1

Common Settings ................................................................................. 11-2

Playback Option ................................................................................... 11-3

Record Option ...................................................................................... 11-4

Chapter 12 Display Control .......................................................... 12-1

To Select the Display Type .................................................................... 12-2

To Set the Persistence Time .................................................................. 12-2

To Set the Waveform Intensity .............................................................. 12-3

To Set the Screen Grid .......................................................................... 12-3

To Set the Grid Brightness..................................................................... 12-3

Chapter 13 Signal Source ............................................................ 13-1

To Output Basic Waveform .................................................................... 13-2

To Output Sine .............................................................................. 13-2

To Output Square .......................................................................... 13-3

To Output Ramp ............................................................................ 13-3

XX MSO1000Z/DS1000Z User’s Guide

Contents RIGOL

To Output Pulse ............................................................................. 13-5

To Output DC ................................................................................ 13-5

To Output Noise ............................................................................. 13-6

To Output Built-in Waveform .................................................................. 13-6

To Output Arbitrary Waveform ............................................................. 13-10

To Select Waveform ..................................................................... 13-11

To Create Waveform ..................................................................... 13-12

To Edit Waveform ........................................................................ 13-14

Modulation ........................................................................................ 13-15

AM ............................................................................................. 13-16

FM ............................................................................................. 13-17

Chapter 14 Store and Recall ......................................................... 14-1

Storage System .................................................................................... 14-2

Storage Type ....................................................................................... 14-2

Internal Storage and Recall ................................................................... 14-4

External Storage and Recall ................................................................... 14-5

Disk Management ................................................................................. 14-6

To Select File Type ......................................................................... 14-6

To Create a New File or Folder ........................................................ 14-7

To Delete a File or Folder .............................................................. 14-10

To Rename a File or Folder ........................................................... 14-10

To Clear the Local Memory ............................................................ 14-10

Factory .............................................................................................. 14-11

Chapter 15 Accessibility Setting ................................................... 15-1

Remote Interface Configuration ............................................................. 15-2

LAN Configuration .......................................................................... 15-3

USB Device ................................................................................... 15-6

GPIB Address ................................................................................ 15-6

System-related ..................................................................................... 15-7

Sound ........................................................................................... 15-7

Language ...................................................................................... 15-7

System Information ........................................................................ 15-7

Vertical Reference .......................................................................... 15-7

Power-off Recall ............................................................................. 15-7

Self-calibration ............................................................................... 15-8

Print Setting .................................................................................. 15-8

Aux Output ................................................................................. 15-10

Option Management ..................................................................... 15-11

Auto Options ............................................................................... 15-12

Key Lock ..................................................................................... 15-12

Chapter 16 Remote Control .......................................................... 16-1

Remote Control via USB ........................................................................ 16-2

Remote Control via LAN ........................................................................ 16-6

Remote Control via GPIB ....................................................................... 16-8

MSO1000Z/DS1000Z User’s Guide XXI

RIGOL Contents

Chapter 17

Chapter 18

Troubleshooting ........................................................ 17-1

Specifications ............................................................ 18-1

Chapter 19 Appendix ................................................................... 19-1

Appendix A: Accessories and Options ..................................................... 19-1

Appendix B: Warranty ........................................................................... 19-2

Index ........................................................................................................ 1

XXII MSO1000Z/DS1000Z User’s Guide

Chapter 1 Quick Start

RIGOL

Chapter 1 Quick Start

This chapter introduces the precautions when using the oscilloscope for the first time, the front/rear panels of the oscilloscope, the user interface and the using method of the built-in help system.

The contents of this chapter:

 General Inspection

 Appearance and Dimensions

 To Prepare the Oscilloscope for Use

 Front Panel Overview

 Rear Panel Overview

 Front Panel Function Overview

 User Interface

 Parameter Setting Method

 To Use the Security Lock

 To Use the Built-in Help System

MSO1000Z/DS1000Z User’s Guide 1-1

RIGOL

Chapter 1 Quick Start

General Inspection

1. Inspect the shipping container for damage.

Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has passed both electrical and mechanical tests.

The consigner or carrier shall be liable for the damage to instrument resulting from shipment. RIGOL would not be responsible for free maintenance/rework or replacement of the unit.

2. Inspect the instrument.

In case of any damage, or defect, or failure, notify your RIGOL sales representative.

3. Check the Accessories

Please check the accessories according to the packing lists. If the accessories are incomplete or damaged, please contact your RIGOL sales representative.

1-2 MSO1000Z/DS1000Z User’s Guide

Chapter 1 Quick Start

Appearance and Dimensions

RIGOL

Figure 1-1 Front View Unit: mm

Figure 1-2 Top View Unit: mm

MSO1000Z/DS1000Z User’s Guide 1-3

RIGOL

Chapter 1 Quick Start

To Prepare the Oscilloscope for Use

To Adjust the Supporting Legs

Adjust the supporting legs properly to use them as stands to tilt the oscilloscope upwards for stable placement of the oscilloscope as well as better operation and observation.

Figure 1-3 To Adjust the Supporting Legs

To Connect to Power Supply

The power requirements of the oscilloscope are 100-240 V, 45-440 Hz. Please use the power cord supplied with the accessories to connect the oscilloscope to the AC power source as shown in the figure below.

1-4

Figure 1-4 To Connect to Power Sypply

Power Socket

MSO1000Z/DS1000Z User’s Guide

Chapter 1 Quick Start

RIGOL

Power-on Inspection

When the oscilloscope is energized, pressing the power key at the lower-left corner of the front panel can start the oscilloscope. During the start-up process, the oscilloscope performs a series of self-tests. After the self-test, the welcome screen is displayed. The instrument is installed with the trial versions of the options before leaving factory and the remaining trial time is about 2000 minutes. The “Current

Options” dialog box will be displayed if your instrument currently installs the trial versions of options. From this dialog box you can view the names, detail, versions and the remaining trial time of the options currently installed.

To Connect the Probe

RIGOL provides passive probe for the DS1000Z series as well as passive probe and logic probe for MSO1000Z series. For detailed technical information of the probes, please refer to the corresponding Probe User’s Guide. The following are the probes recommended for this oscilloscope.

Model

RP2200

RPL1116

Description Applicable Model

150 MHz, passive probe, standard MSO1000Z/DS1000Z

Logic probe, standard MSO1000Z

Connect the passive probe:

1. Connect the BNC terminal of the probe to an analog channel input terminal of the oscilloscope at the front panel.

2. Connect the ground alligator clip or spring of the probe to the circuit ground terminal and connect the probe tip to the circuit point to be tested.

Figure 1-5 To Connect the Passive Probe

MSO1000Z/DS1000Z User’s Guide 1-5

RIGOL

Chapter 1 Quick Start

Connect the logic probe:

1. Connect the single-wire terminal of the logic probe to the digital channel interface at the front panel of MSO1000Z in the correct direction.

2. Connect the signal under test to the other terminal of the logic probe. MSO1000Z is provided with the RPL1116 logic probe which provides two connecting methods with the signal under test to fulfill the requirements of different application environment. For the details, please refer to the

RPL1116 Logic Probe

User’s Guide

.

Digital Channel Interface

Figure 1-6 To Connect the Logic Probe

1-6 MSO1000Z/DS1000Z User’s Guide

Chapter 1 Quick Start

RIGOL

Function Inspection

1. Press Storage Default to restore the oscilloscope to its default configuratiuon.

2. Connect the earth alligator clip of the probe to the “Ground Terminal” as shown in the figure below.

3. Use the probe to connect the input terminal of CH1 of the oscilloscope and the

“Compensation Signal Output Terminal” of the probe.

Compensation Signal Output Terminal

Ground Terminal

Figure 1-7 To Use the Compensation Signal

4. Press AUTO.

5. Observe the waveform on the display. In normal condition, the square waveform as shown in the figure below should be displayed.

Figure 1-8 Square Waveform

6. Use the same method to test the other channels. If the square waveforms

actually shown do not match that in the figure above, please perform “Probe

Compensation”.

WARNING

To avoid electric shock when using the probe, please make sure that the insulated wire of the probe is in good condition and do not touch the metallic part of the probe when the probe is connected to high voltage source.

Tip

The signal output from the probe compensation connector can only be used for probe compensation adjustment and can not be used for calibration.

MSO1000Z/DS1000Z User’s Guide 1-7

RIGOL

Chapter 1 Quick Start

Probe Compensation

When the probes are used for the first time, you should compensate the probes to make them match the input channels of the oscilloscope. Non-compensated or poorly compensated probes may cause measurement inaccuracy or error. The probe compensation procedures are as follows.

1. Perform steps 1, 2, 3 and 4 of “Function Inspection”.

2. Check the displayed waveforms and compare them with the following figures.

Over compensated Perfectly compensated Under compensated

Figure 1-9 Probe Compensation

3. Use a nonmetallic driver to adjust the low-frequency compensation adjustment hole on the probe until the displayed waveform is as the “Perfectly compensated” in the figure above.

1-8 MSO1000Z/DS1000Z User’s Guide

Chapter 1 Quick Start

RIGOL

Front Panel Overview

1 2 3 4 5 6 7 8 9 10

11 12 13 14 15 16 17 18 19 20

Figure 1-10 Front Panel Overview

Table 1-1 Front Panel Descriptions

No. Description No. Description

1 Measurement Menu Softkeys 11 Power Key

2 LCD

3 Function Menu Softkeys

12

13

USB HOST Interface

Digital Channel Input Interface [1]

Note

Note

4 Multifunction Knob

5 Common Operation Keys

6 CLEAR

7 AUTO

8 RUN/STOP

9 SINGLE

10 Help/Print

14

15

Analog Channel Input Interface

Logic Analyzer Control Key

16 Signal Source [2]

17 VERTICAL Control

18 HORIZONTAL Control

19 TRIGGER Control

[1]

20 Probe Compensation Signal

Output Terminal/Ground Terminal

[1] : Only applicable to MSO1000Z and MSO1000Z-S models mixed signal digital

[2] oscilloscopes.

: Only applicable to MSO1000Z-S and DS1000Z-S models digital oscilloscopes.

MSO1000Z/DS1000Z User’s Guide 1-9

RIGOL

Chapter 1 Quick Start

Rear Panel Overview

1 2 3

4 5 6 7 8

Figure 1-11 Rear Panel Overview

1 Handle

Pull up the handle vertically for easy carry of the instrument. When you do not need the handle, press it down.

2 LAN

Connect the instrument to network via this interface for remote control. This oscilloscope conforms to the LXI CORE 2011 DEVICE class instrument standards and can quickly build test system with other instruments.

3 USB Device

You can connect the oscilloscope to a PictBridge printer or PC via this interface.

When a PC is connected, users can send SCPI commands using the PC software or control the oscilloscope via user-defined programming. When a printer is connected, users can print the waveform displayed on the screen using the printer.

4 Trigger Out and Pass/Fail

Trigger Out:

The oscilloscope can output a signal that can reflect the current capture rate of the oscilloscope at each trigger via this interface. Connect the signal to a waveform display device and measure the frequency of the signal. The measurement result is the same with the current capture rate.

1-10 MSO1000Z/DS1000Z User’s Guide

Chapter 1 Quick Start

RIGOL

Pass/Fail:

The instrument can output a positive pulse via this connector when a failed waveform is detected during the pass/fail test. The instrument continuously outputs a low level via this connector when no failed waveform is detected.

5 Source Output

The output terminals of the built-in dual-channel source of the oscilloscope.

When Source1 or Source2 is enabled, the signal currently set can be output through the [Source1] or [Source2] connector at the rear panel.

6 Lock Hole

You can lock the instrument to a fixed location using the security lock (please buy it yourself) via the lock hole.

7 Fuse

If a new fuse is required, please use the specified fuse (250V, T2A). The replacing method is as follows. a) Turn off the instrument, switch off the power supply and remove the power cord. b) Insert a small straight screwdriver into the slot at the power socket and pry out the fuse seat gently. c) Take out the fuse and replace it with a specified fuse. Then, install the fuse seat.

8 AC Power Socket

AC power input terminal. The power requirements of this oscilloscope are

100-240 V, 45-440 Hz. Use the power cord provided with the accessories to connect the instrument to AC power. Then, you can press the power key at the front panel to start the instrument.

MSO1000Z/DS1000Z User’s Guide 1-11

RIGOL

Chapter 1 Quick Start

Front Panel Function Overview

VERTICAL

CH1, CH2, CH3, CH4: analog channel setting keys. The 4 channels are marked by different colors which are also used to mark both the corresponding waveforms on the screen and the channel input connectors. Press any key to open the corresponding channel menu and press again to turn off the channel.

MATH: press MATHMath to open the math operation menu under which A+B, A-B,

A×B, A/B, FFT, A&&B, A||B, A^B, !A, Intg, Diff,

Sqrt, Lg, Ln, Exp and Abs operations are provided. You can also press MATH to open the decoding menu and set the decoding options.

REF: press this key to enable the reference waveform function to compare the waveform actually measured with the reference waveform.

VERTICAL POSITION: modify the vertical position of the current channel waveform. Turn clockwise to increase the position and turn counterclockwise to decrease. During the modification, the waveform would move up and down and the position message (e.g. ) at the lower-left corner of the screen would change accordingly. Press down this knob to quickly reset the vertical position to zero.

VERTICAL SCALE: modify the vertical scale of the current channel. Turn clockwise to decrease the scale and turn counterclockwise to increase. During the modification, the display amplitude of the waveform would enlarge or reduce. The scale information (e.g. ) at the lower side of the screen would change accordingly. Press down this knob to quickly switch the vertical scale adjustment mode between “Coarse” and “Fine”.

Tip

How to set the vertical scale and vertical position of each channel?

The 4 channels of MSO1000Z/DS1000Z use the same VERTICAL POSITION and VERTICAL SCALE knobs. If you want to set the vertical scale and vertical position of a channel, please press CH1, CH2, CH3 or CH4 at first to select the desired channel. Then turn the VERTICAL POSITION and VERTICAL

SCALE knobs to set the values.

1-12 MSO1000Z/DS1000Z User’s Guide

Chapter 1 Quick Start

RIGOL

Logic Analyzer

Press this key to open the logic analyzer control menu. You can turn on or off any channel or channel group, modify the display size of the digital channel, modify the logic threshold of the digital channel as well as group the 16 digital channels. You can also set a label for each digital channel.

Note:

― This function is only applicable to the MSO1000Z and MSO1000Z-S series oscilloscopes.

― Press LAD7-D0; when “On” is selected, CH4 function is automatically disabled; when “Off” is selected, CH4 function recovers automatically. Press LA

D15-D8; when “On” is selected, CH3 function is automatically disabled; when “Off” is selected, CH3 function recovers automatically.

Signal Source

Press this key to enter the source setting interface. You can enable or disable the output of the [Source1] or [Source2] connector at the rear panel, set the output signal waveform and parameters, turn on or off the state display of the current signal.

Note: This function is only available for MSO1000Z-S and DS1000Z-S models oscilloscopes.

MSO1000Z/DS1000Z User’s Guide 1-13

RIGOL

Chapter 1 Quick Start

HORIZONTAL

HORIZONTAL POSITION: modify the horizontal position. The trigger point would move left or right relative to the center of the screen when you turn the knob. During the modification, waveforms of all the channels would move left or right and the horizontal position message (e.g.

) at the upper-right corner of the screen would change accordingly. Press down this knob to quickly reset the horizontal position (or the delayed sweep position).

MENU: press this key to open the horizontal control menu where you can turn on or off the delayed sweep function and switch between different time base modes.

HORIZONTAL SCALE: modify the horizontal time base.

Turn clockwise to reduce the time base and turn counterclockwise to increase the time base. During the modification, waveforms of all the channels will be displayed in expanded or compressed mode and the time base message

(e.g. ) at the upper side of the screen would change accordingly. Press down this knob to quickly switch to the delayed sweep state.

1-14 MSO1000Z/DS1000Z User’s Guide

Chapter 1 Quick Start

RIGOL

TRIGGER

MODE: press this key to switch the trigger mode to Auto,

Normal or Single and the corresponding state backlight of the current trigger mode would be illuminated.

TRIGGER LEVEL: modify the trigger level. Turn clockwise to increase the level and turn counterclockwise to reduce the level. During the modification, the trigger level line would move up and down and the value in the trigger level message box

(e.g. ) at the lower-left corner of the screen would change accordingly. Press down the knob to quickly reset the trigger level to zero point.

MENU: press this key to open the trigger operation menu. This oscilloscope provides various trigger types. For more details,

please refer to “ To Trigger the Oscilloscope”.

FORCE: press this key to generate a trigger signal forcefully.

CLEAR

Press this key to clear all the waveforms on the screen. If the oscilloscope is in the “RUN” state, new waveforms will still be displayed.

AUTO

Press this key to enable the waveform auto setting function. The oscilloscope will automatically adjust the vertical scale, horizontal time base and trigger mode according to the input signal to realize optimum waveform display.

Note: Waveform auto setting function requires that the frequency of sine is no lower than 41Hz; the duty cycle should be greater than 1% and the amplitude must be at least 20mVpp for square. Otherwise, the Waveform auto setting function may be invalid and the quick parameter measurement function displayed in the menu will also be unavailable.

MSO1000Z/DS1000Z User’s Guide 1-15

RIGOL

Chapter 1 Quick Start

RUN/STOP

Press this key to “RUN” or “STOP” waveform sampling. In the

“RUN” state, the key is illuminated in yellow. In the “STOP” state, the key is illuminated in red.

SINGLE

Press this key to set the trigger mode to “Single”. In single trigger mode, press FORCE to generate a trigger signal immediately.

Multifunction Knob

Adjust waveform brightness:

In non-menu-operation mode, turn this knob to adjust the brightness of waveform display. The adjustable range is from

0% to 100%. Turn clockwise to increase the brightness and counterclockwise to reduce. Press down this knob to reset the brightness to 60%.

You can also press DisplayWaveIntensity and use the knob to adjust the waveform brightness.

Multifunction:

In menu operation, the backlight of the knob goes on. Press any menu softkey and turn the knob to select the submenus under this menu and then press down the knob to select the current submenu. It can also be used to modify parameters (please

refer to the introduction in “Parameter Setting Method”) and input filename.

1-16 MSO1000Z/DS1000Z User’s Guide

Chapter 1 Quick Start

Function Menus

RIGOL

Measure: press this key to open the measurement setting menu. You can set the measurement source and turn on or off the frequency counter, all measure, statistic function etc.

Press MENU at the left of the screen to open the measurement menus of 33 waveform parameters. Then, press down the corresponding menu softkey to quickly realize one-key measurement and the measurement result will be displayed at the bottom of the screen.

Acquire: press this key to enter the sample setting menu to set the acquisition mode, Sin(x)/x and memory depth of the oscilloscope.

Storage: press this key to enter the file store and recall interface. The storable file types include picture, traces, waves, setups, CSV and parameters. Internal and external storage as well as disk management are also supported.

Cursor: press this key to enter the cursor measurement menu. The oscilloscope provides four cursor modes: manual, track, auto and XY. Note that XY cursor mode is only available when the horizontal time base is set to XY.

Display: press this key to enter the display setting menu to set the display type, persistence time, wave intensity, grid type and grid brightness.

Utility: press this key to enter the system function setting menu to set the system-related functions or parameters, such as the I/O, sound and language.

Besides, some advanced functions (such as the pass/fail test, waveform record, etc.) are also supported.

MSO1000Z/DS1000Z User’s Guide 1-17

RIGOL

Print

Chapter 1 Quick Start

Press this key to print the screen or save the screen to a USB storage device.

― If a PictBridge printer is connected currently and the printer is in idle state, pressing this key can print the screen.

― If no printer is connected but a USB storage device is inserted, pressing this key can save the screen to the USB storage device in “.png” format. When the storage type is picture, you can store the screen image to the USB storage device in the specified picture format. For more details, please refer to the

introduction in “Storage Type”.

― If both a printer and a USB storage device are connected, the printer enjoys higher priority when pressing this key.

1-18 MSO1000Z/DS1000Z User’s Guide

Chapter 1 Quick Start

RIGOL

User Interface

MSO1000Z/DS1000Z provides 7.0 inch, WVGA (800*480) 160,000 color TFT LCD.

1 2 3 4 5 6 7 8 9 10 11

12 13 14 15 16 17 18 19 20 21 22

Figure 1-12 User Interface

1. Auto Measurement Items

Provide 16 horizontal (HORIZONTAL) and 17 vertical (VERTICAL) measurement parameters. Press the softkey at the left of the screen to activate the corresponding measurement item. Press MENU continuously to switch between the horizontal and vertical parameters.

2. Digital Channel Label/Waveform

The logic high level of the digital waveform is displayed in blue and the logic low level in green. Its edge is displayed in white. The waveform of the digital channel currently selected and the channel label are displayed in red. The digital channels can be divided into 4 channel groups by the grouping setting function of the logic analyzer function menu. The channel labels of the same channel group are displayed in the same color; different channel groups are marked with different colors.

Note: This function is only applicable to MSO1000Z-S and MSO1000Z models oscilloscopes.

MSO1000Z/DS1000Z User’s Guide 1-19

RIGOL

Chapter 1 Quick Start

3. Status

Available states include RUN, STOP, T’D (triggered), WAIT and AUTO.

4. Horizontal Time Base

 Represent the time per grid on the horizontal axis on the screen.

 Use HORIZONTAL SCALE to modify this parameter. The range available is from 5 ns to 50 s.

5. Sample Rate/Memory Depth

 Display the current sample rate and memory depth of the oscilloscope.

 The sample rate and memory depth will change in accordance with the horizontal time base.

6. Waveform Memory

Provide the schematic diagram of the memory position of the waveform currently on the screen.

Waveform on the screen

7. Trigger Position

Display the trigger position of the waveform in the waveform memory and on the screen.

8. Horizontal Position

Use HORIZONTAL POSITION to modify this parameter. Press down the knob to automatically set the parameter to zero.

9. Trigger Type

Display the currently selected trigger type and trigger condition setting.

Different labels are displayed when different trigger types are selected.

For example, represents triggering on the rising edge in “Edge” trigger.

10. Trigger Source

Display the trigger source currently selected (CH1-CH4, AC or D0-D15).

Different labels are displayed when different trigger sources are selected and the color of the trigger parameter area will change accordingly.

For example, denotes that CH1 is selected as the trigger source.

11. Trigger Level

 When an analog channel is selected as the trigger source, you need to set a proper tigger level.

 The trigger level label is displayed at the right of the screen and the trigger level value is displayed at the upper-right corner of the screen.

 When using TRIGGER LEVEL to modify the trigger level, the trigger

1-20 MSO1000Z/DS1000Z User’s Guide

Chapter 1 Quick Start

RIGOL level value will change with the up and down of .

Note: In slope trigger, runt trigger and window trigger, two trigger level labels

( and ) are displayed.

12. CH1 Vertical Scale

 Display the voltage value per grid of CH1 waveform vertically.

 Press CH1 to select CH1, and use VERTICAL SCALE to modify this parameter.

 The following labels will be displayed according to the current channel setting: channel coupling (e.g. ) and bandwidth limit (e.g. ).

13. Analog Channel Label/Waveform

Different channels are marked with different colors and the colors of the channel label and waveform are the same.

14. CH2 Vertical Scale

 Display the voltage value per grid of CH2 waveform vertically.

 Press CH2 to select CH2, and use VERTICAL SCALE to modify this parameter.

 The following labels will be displayed according to the current channel setting: channel coupling (e.g. ) and bandwidth limit (e.g. ).

15. CH3 Vertical Scale

 Display the voltage value per grid of CH3 waveform vertically.

 Press CH3 to select CH3, and use VERTICAL SCALE to modify this parameter.

 The following labels will be displayed according to the current channel setting: channel coupling (e.g. ) and bandwidth limit (e.g. ).

16. CH4 Vertical Scale

 Display the voltage value per grid of CH4 waveform vertically.

 Press CH4 to select CH4, and use VERTICAL SCALE to modify this parameter.

 The following labels will be displayed according to the current channel setting: channel coupling (e.g. ) and bandwidth limit (e.g. ).

17. Message Box

Display the prompt messages.

18. Digital Channel Status Area

Display the current status of the 16 digital channels. The digital channels currently turned on are displayed in green and the digital channel currently selected is displayed in red. The digital channels turned off are displayed in grey.

Note: This function is only applicable to MSO1000Z-S and MSO1000Z models oscilloscopes.

MSO1000Z/DS1000Z User’s Guide 1-21

RIGOL

Chapter 1 Quick Start

19. Source1 Waveform

 Display the type of waveform currently set for Source1.

 When the modulation of source1 is enabled, will be displayed below the

Source1 waveform.

 When the impedance of source1 is set to 50 Ω, will be displayed below the Source1 waveform.

 This function is only applicable to MSO1000Z-S and DS1000Z-S models oscilloscopes.

20. Source2 Waveform

 Display the type of waveform currently set for Source2.

 When the modulation of source2 is enabled, will be displayed below the

Source2 waveform.

 When the impedance of source2 is set to 50 Ω, will be displayed below the Source2 waveform.

 This function is only applicable to MSO1000Z-S and DS1000Z-S models oscilloscopes.

21. Notification Area

Display the sound icon and USB storage device icon.

 Sound Icon: Press UtilitySound to enable or disable the sound. When the sound is enabled, will be displayed; when the sound is disabled,

will be displayed.

 USB Storage Device Icon: when a USB storage device is detected, be displayed.

22. Operation Menu

will

Press any softkey to activate the corresponding menu. The following symbols might be displayed in the menu:

Denote that the multifunction knob can be used to modify the parameters. The backlight of turns on in the parameter modification status.

Denote that you can use to select the desired items and the item currently selected is displayed in blue. Press down to enter the menu bar corresponding to the selected item. The backlight of is constant on after menus with this symbol are selected.

Denote that you can press to open the pop-up numeric keyboard and input the desired paramrter values directly. The backlight of is constant on after menus with this symbol are selected.

Denote that the current menu has several options.

Denote that the current menu has a lower level menu.

Press this key to return to the previous menu.

The number of dots indicates the number of pages of the current menu.

1-22 MSO1000Z/DS1000Z User’s Guide

Chapter 1 Quick Start

RIGOL

Parameter Setting Method

MSO1000Z/DS1000Z supports the following two parameter setting methods.

Method one:

For the parameters with displayed on the menu, you can turn the multifunction knob directly to set the desired values.

Method two:

For the parameters with displayed on the menu, press down the multifunction knob and the numeric keyboard as shown in the figure below is displayed. Turn the knob to select the desired value and press down the knob to input the value.

After entering all the values, turn the knob to select the desired unit and press down the knob to complete the parameter setting.

Figure 1-13 Numeric Keyboard

MSO1000Z/DS1000Z User’s Guide 1-23

RIGOL

Chapter 1 Quick Start

To Use the Security Lock

If needed, you can use the security lock (please buy it yourself) to lock the oscilloscope to a fixed location. The method is as follows, align the lock with the lock hole and plug it into the lock hole vertically, turn the key clockwise to lock the oscilloscope and then pull the key out.

Security Lock Hole

Figure 1-14 To Use the Security Lock

Note: Please do not insert other articles into the security lock hole to avoid damaging the instrument.

1-24 MSO1000Z/DS1000Z User’s Guide

Chapter 1 Quick Start

RIGOL

To Use the Built-in Help System

The help system of this oscilloscope provides instructions for all the function keys

(including the menu keys) at the front panel. Press Help to open the help interface and press again to close the interface. The help interface mainly consists of two parts.

The left are “Help Options”. The right is the “Help Display Area”.

Help Options Help Display Area

Figure 1-15 Help Interface

You can press the button (except the power key , the multifunction knob and the menu page up/down key / ) directly at the front panel to get the corresponding help information in the “Help Display Area”.

MSO1000Z/DS1000Z User’s Guide 1-25

Chapter 2 To Set the Vertical System

Chapter 2 To Set the Vertical System

The contents of this chapter:

 To Enable the Analog Channel

 Channel Coupling

 Bandwidth Limit

 Probe Ratio

 Waveform Invert

 Vertical Scale

 Amplitude Unit

 Channel Label

 Delay Calibration of the Analog Channel

RIGOL

MSO1000Z/DS1000Z User’s Guide 2-1

RIGOL

Chapter 2 To Set the Vertical System

To Enable the Analog Channel

MSO1000Z/DS1000Z provides 4 analog input channels (CH1-CH4). As the setting methods of the vertical systems of the four channels are the same, this chapter takes

CH1 as an example to illustrate the setting method of the vertical system.

Connect a signal to the channel connector of CH1 and then press CH1 in the vertical control area (VERTICAL) at the front panel to enable CH1. At this point, the channel setting menu is displayed at the right side of the screen and the channel status label at the bottom of the screen (as shown in the figure below) is highlighted. The information displayed in the channel status label is related to the current channel setting.

After the channel is turned on, modify the parameters such as the vertical scale, horizontal time base, trigger mode and trigger level according to the input signal for easy observation and measurement of the waveform.

Channel Coupling

The undesired signals can be filtered out by setting the coupling mode. For example, the signal under test is a square waveform with DC offset.

 When the coupling mode is “DC”: the DC and AC components of the signal under test can both pass the channel.

 When the coupling mode is “AC”: the DC components of the signal under test are blocked.

 When the coupling mode is “GND”: the DC and AC components of the signal under test are both blocked.

Press CH1Coupling and use to select the desired coupling mode (the default is DC). The current coupling mode is displayed in the channel status label at the bottom of the screen as shown in the figure below. You can also press Coupling continuously to switch the coupling mode.

DC AC GND

2-2 MSO1000Z/DS1000Z User’s Guide

Chapter 2 To Set the Vertical System

RIGOL

Bandwidth Limit

MSO1000Z/DS1000Z supports the bandwidth limit function which can reduce the display noise. For example, the signal under test is a pulse with high frequency oscillation.

 When the bandwidth limit is disabled, the high frequency components of the signal under test can pass the channel.

 When limiting the bandwidth to 20 MHz, the high frequency components of the signal under test that exceed 20 MHz are attenuated.

Press CH1, then press BW Limit continuously to switch the bandwidth limit state

(the default is OFF). When the bandwidth limit is enabled, the character “B” will be displayed in the channel status label at the bottom of the screen.

Note: Bandwidth limit can reduce the noise as well as attenuate or eliminate the high frequency components of the signal.

Probe Ratio

MSO1000Z/DS1000Z allows user to set the probe attenuation ratio manually. Press

CH1Probe and use to select the desired probe ratio. The probe ratio values available are as shown in the table below.

Table 2-1 Probe Ratio

Menu

0.01X

0.02X

0.05X

0.1X

0.2X

0.5X

1X

2X

5X

10X (the default value)

20X

50X

100X

200X

500X

1000X

Attenuation Coefficient (the measured

signal: the displayed waveform)

1:100

1:50

1:20

1:10

1:5

1:2

1:1

2:1

5:1

10:1

20:1

50:1

100:1

200:1

500:1

1000:1

MSO1000Z/DS1000Z User’s Guide 2-3

RIGOL

Chapter 2 To Set the Vertical System

Waveform Invert

Press CH1Invert to enable or disable waveform invert. When waveform invert is enabled, the waveform display rotates 180 degree relative to the ground potential.

When waveform invert is disabled, the waveform display is normal (as shown in the figure below).

(a) Turn Off “Invert” (b) Turn On “Invert”

Figure 2-1 Waveform Invert

Vertical Scale

Vertical scale refers to the voltage value per grid in the vertical direction on the screen and is usually expressed as V/div.

Press CH1 and rotate VERTICAL SCALE to adjust the vertical scale (clockwise to reduce the scale and counterclockwise to increase). The size of the displayed waveform will be changed accordingly. The scale information (as shown in the figure below) in the channel label at the bottom of the screen will change accordingly during the adjustment. The adjustable range of the vertical scale is related to the probe ratio currently set. By default, the probe ratio is 10X and the adjustable range of the vertical scale is from 10 mV/div to 100 V/div.

The vertical scale can be adjusted in “Coarse” or “Fine” mode. Press CH1

Volts/Div to switch the adjustment mode.

 Coarse adjustment (take counterclockwise as an example): set the vertical scale in 1-2-5 step namely 10 mV/div, 20 mV/div, 50 mV/div, 100 mV/div…100 V/div.

 Fine adjustment: further adjust the vertical scale within a relatively smaller range to improve vertical resolution. If the amplitude of the input waveform is a little bit greater than the full scale under the current scale and the amplitude would be a little bit lower if the next scale is used, fine adjustment can be used to improve the amplitude of waveform display to view signal details.

2-4 MSO1000Z/DS1000Z User’s Guide

Chapter 2 To Set the Vertical System

RIGOL

Note: You can also press VERTICAL SCALE to quickly switch between “Coarse” and “Fine” adjustments.

When changing the vertical scale of the analog channel by ratating VERTICAL

SCALE, you can select to expand or compress the waveform around the “Center” or

“Ground”. For more details, please refer to the introduction in “Vertical

Reference”.

Amplitude Unit

Select the amplitude display unit for the current channel. The available units are W, A,

V and U. When the unit is changed, the unit displayed in the channel label will change accordingly.

Press CH1Unit to select the desired unit and the default is V.

Channel Label

The instrument uses the number of the channel to mark the corresponding channel

“ by default. For ease of use, you can also set a label for each channel, for example,

”. Press CH1Label to enter the label setting menu. You can use the built-in label or manually input a label. Manual input does not support the Chinese input, and the length of the label cannot exceed 4 characters.

Press Display to turn on or off the channel label display. The default is CH1 when the channel label display is enabled.

Press Template to select the preset labels such as CH1, ACK, ADDR, BIT, CLK, CS,

DATA, IN, MISO, MOSI, OUT, RX and TX, ect.

Press Label Edit and the label editing interface is automatically displayed as shown in the figure below. You can input the label manually.

Name Input Area Keyboard Upper/Lower Case Switch

Figure 2-2 Label Editing Interface

MSO1000Z/DS1000Z User’s Guide 2-5

RIGOL

Chapter 2 To Set the Vertical System

For example, set the label to “

Select “A a

”. Press Keyboard to select the “Keyboard” area.

” using and press down to switch it to “ a A”. Select “C” using and press down to input the character. Use the same method to input “hn1”.

After finishing the input, press OK to finish the editing. If the Display is enabled, the label will be displayed at the left of CH1 waveform.

To modify or delete the input character, press Name to select the “Name Input Area” and use to select the character to be modified or deleted. Enter the desired character or press Delete to delete the character selected.

2-6 MSO1000Z/DS1000Z User’s Guide

Chapter 2 To Set the Vertical System

RIGOL

Delay Calibration of the Analog Channel

When using an oscilloscope for actual measurement, the transmission delay of the probe cable may bring relatively greater error (zero offset). MSO1000Z/DS1000Z supports user to set a delay time to calibrate the zero offset of the corresponding channel. Zero offset is defined as the offset of the crossing point of the waveform and trigger level line relative to the trigger position, as shown in the figure below.

Figure 2-1 Zero Offset

Press CH1Delay-Cal and use to set the desired delay time. The range available is from -100 ns to 100 ns. Pressing down the multifunction knob will reset the delay time to 0.00 s.

Note: This parameter is related to the instrument model and the current horizontal time base setting. The larger the horizontal time base is, the larger the setting step will be. Take MSO1104Z-S as an example. The step values under different horizontal time base are as shown in the table below.

Table 2-2 Relation between delay calibration time step and horizontal time base

Horizontal Time Base Delay Calibration Time Step

5ns

10ns

20ns

50ns

100ns

200ns

500ns

1μs to 10μs

100ps

200ps

400ps

1ns

2ns

4ns

10ns

20ns

Note: When the horizontal time base is equal to or greater than 10μs, the delay calibration time cannot be adjusted.

MSO1000Z/DS1000Z User’s Guide 2-7

Chapter 3 To Set the Horizontal System

RIGOL

Chapter 3 To Set the Horizontal System

The contents of this chapter:

 Delayed Sweep

 Time Base Mode

MSO1000Z/DS1000Z User’s Guide 3-1

RIGOL

Chapter 3 To Set the Horizontal System

Delayed Sweep

Delayed sweep can be used to enlarge a length of waveform horizontally to view waveform details.

Press MENU in the horizontal control area (HORIZONTAL) at the front panel and press Delayed to enable or disable delayed sweep.

Note: To enable delayed sweep, the current time base mode must be “YT”.

In delayed sweep mode, the screen is divided into two display areas as shown in the figure below.

The waveform before enlargement

Main time base

Delayed

Sweep

Time Base

放大后的波形

The waveform after enlargement

Figure 3-1 Delayed Sweep Mode

The waveform before enlargement:

The waveform in the area that is not covered by subtransparent blue in the upper part of the screen is the waveform before enlargement. You can turn HORIZONTAL

POSITION to move the area left and right or turn HORIZONTAL SCALE to enlarge or reduce this area.

The waveform after enlargement:

The waveform in the lower part of the screen is the horizontally expanded waveform.

Compared to the main time base, the delayed time base has increased the waveform resolution (as shown in the figure above).

Note: The delayed time base should be less than or equal to the main time base.

3-2 MSO1000Z/DS1000Z User’s Guide

Chapter 3 To Set the Horizontal System

RIGOL

Tip

You can also press down HORIZONTAL SCALE (delayed sweep shortcut key) to directly switch to the delayed sweep mode.

Time Base Mode

Press MENU in the horizontal control area (HORIZONTAL) at the front panel and then press Time Base to select the time base mode of the oscilloscope. The default is YT.

YT Mode

In this mode, the Y axis represents voltage and the X axis represents time.

Note:

Only when this mode is enabled, can “Delayed Sweep” be turned on. In this

mode, when the horizontal time base is greater than or equal to 200 ms, the instrument enters slow sweep mode. For the details, please refer to the introduction

of slow sweep in “Roll Mode”.

XY Mode

In this mode, the oscilloscope changes the two channels from voltage-time display mode to voltage-voltage display mode. The phase deviation between two signals with the same frequency can be easily measured via Lissajous method. The figure below shows the measurement schematic diagram of the phase deviation.

The signal must be centered horizontally

D A B

C

Figure 3-2 Measurement Schematic Diagram of Phase Deviation

MSO1000Z/DS1000Z User’s Guide 3-3

RIGOL

Chapter 3 To Set the Horizontal System

According to sinθ=A/B or C/D (wherein, θ is the phase deviation angle between the two channels and the definitions of A, B, C and D are as shown in the figure above), the phase deviation angle is obtained, that is:

θ=±arcsin (A/B) or ±arcsin( C/D)

If the principal axis of the ellipse is within quadrant I and III, the phase deviation angle obtained should be within quadrant I and IV, namely within (0 to π/2) or (3π/2 to 2π). If the principal axis of the ellipse is within quadrant II and IV, the phase deviation angle obtained should be within quadrant II and III, namely within (π/2 to

π) or (π to 3π/2).

The XY function can be used to measure the phase deviation occurred when the signal under test passes through a circuit network. Connect the oscilloscope to the circuit to monitor the input and output signals of the circuit.

Application example: measure the phase deviation of the input signals of two channels.

Method 1: Use Lissajous method

1. Connect a sine signal to CH1 and then connect a sine signal with the same frequency and amplitude but a 90° phase deviation to CH2.

2. Press AUTO and then adjust the vertical positions of CH1 and CH2 to 0 V.

3. Set the time base mode to XY, press X-Y and select “CH1-CH2”. Rotate

Horizontal SCALE to adjust the sample rate properly to get better

Lissajous figure for better observation and measurement.

4. Rotate VERTICAL SCALE of CH1 and CH2 to make the signals easy to observe. At this point, the circle as shown in the figure below should be displayed.

5. Observe the measurement result shown in the figure above. According to the

measurement schematic diagram of the phase deviation (as shown in Figure

3-2), A/B (C/D) = 1. Thus, the phase deviation angle θ=±arcsin1=90°.

3-4 MSO1000Z/DS1000Z User’s Guide

Chapter 3 To Set the Horizontal System

RIGOL

Note:

 The maximum sample rate of XY mode is 500 MSa/s. Generally, longer sample waveform can ensure better display effect of Lissajous figure. But due to the limitation of the memory depth, you have to reduce the waveform sample rate

to acquire longer waveform (refer to the introduction in “Memory Depth”).

Therefore, during the measurement, reducing the sample rate properly can acquire better display effect of Lissajous figure.

 When XY mode is enabled, “Delayed Sweep” will be disabled automatically.

 Press X-Y to select “CH1-CH2, CH1-CH3, CH1-CH4, CH2-CH3, CH2-CH4,

CH3-CH4”. After you choose any of the options, the instrument automatically turns on the two corresponding channels and turns off the other two channels.

The X-axis tracks the voltage of the first channel in each option; the Y-axis tracks the voltage of the second channel in each option.

 The following functions are not available in XY mode:

Delayed Sweep”, “Vectors”, “Protocol Decoding”, “Acquisition Mode”,

Pass/Fail Test”, “Waveform Record” “Digital Channel” and “To Set the

Persistence Time”.

Method 2: Use the shortcut measurement function

Please refer to “Phase 12 ” and “Phase 12 ” measurement functions of “Delay

and Phase” on page 6-23.

Roll Mode

In this mode, the waveform scrolls from right to left to update the display.

The horizontal position and trigger control of the waveform are not available. The range of horizontal scale adjustment is from 200 ms to 50.0 s.

Note:

When Roll mode is enabled, the waveform “horizontal position”, “Delayed

Sweep”, “Protocol Decoding”, “Pass/Fail Test”, “Waveform Record”, “To Set

the Persistence Time” and “To Trigger the Oscilloscope” are not available.

Slow Sweep

Slow sweep is similar to Roll mode. In YT mode, when the horizontal time base is set to 200 ms/div or slower, the instrument enters “slow sweep” mode in which the instrument first acquires the data at the left of the trigger point and then waits for a trigger event. After the trigger occurs, the instrument continues to finish the waveform at the right of the trigger point. When slow sweep mode is used to

observe low frequency signal, DC “Channel Coupling” mode is recommended.

MSO1000Z/DS1000Z User’s Guide 3-5

Chapter 4 To Set the Sample System

Chapter 4 To Set the Sample System

The contents of this chapter:

 Acquisition Mode

 Sin(x)/x

 Sample Rate

 Memory Depth

 Antialiasing

RIGOL

MSO1000Z/DS1000Z User’s Guide 4-1

RIGOL

Chapter 4 To Set the Sample System

Acquisition Mode

The acquisition mode is used to control how to generate waveform points from sample points.

Press AcquireMode at the front panel and use to select the desired acquisition mode (the default is normal), then press down the knob to select this mode. You can also press Mode continuously to switch the acquisition mode.

Normal

In this mode, the oscilloscope samples the signal at equal time interval to rebuild the waveform. For most of the waveforms, the best display effect can be obtained using this mode.

Peak Detect

In this mode, the oscilloscope acquires the maximum and minimum values of the signal within the sample interval to get the envelope of the signal or the narrow pulse of the signal that might be lost. In this mode, signal confusion can be prevented but the noise displayed would be larger.

In this mode, the oscilloscope can display all the pulses with pulse widths at least as wide as the sample period.

Average

In this mode, the oscilloscope averages the waveforms from multiple samples to reduce the random noise of the input signal and improve the vertical resolution.

Greater number of averages can lower the noise and increase the vertical resolution; while at the same time, it will slow the response of the displayed waveform to the waveform changes.

The available range of the number of averages is from 2 to 1024 and the default is 2.

When “Average” mode is selected, press Averages and use to set the desired number of averages as the power function of 2.

4-2 MSO1000Z/DS1000Z User’s Guide

Chapter 4 To Set the Sample System

RIGOL

Figure 4-1 The Waveform before Average

Figure 4-2 The Waveform after 256 Averages

High Resolution

This mode uses a kind of ultra-sample technique to average the neighboring points of the sample waveform to reduce the random noise on the input signal and generate much smoother waveforms on the screen. This is generally used when the sample rate of the digital converter is higher than the storage rate of the acquisition memory.

Note: “Average” and “High Res” modes use different averaging methods. The former uses “Multi-sample Average” and the latter uses “Single Sample Average”.

MSO1000Z/DS1000Z User’s Guide 4-3

RIGOL

Chapter 4 To Set the Sample System

Sin(x)/x

Press Sin(x)/x to enable or disable the dynamic sine interpolation function which can acquire better restoration of the original waveform.

Note: If the number of channels currently turned on is less than three, Sin(x)/x is grayed out and disabled.

Sample Rate

The maximum sample rate of MSO1000Z/DS1000Z is 1 GSa/s.

Note: The sample rate is displayed in the status bar at the upper side of the screen and in the Sa Rate menu and can be changed by adjusting the horizontal time base through HORIZONTAL

SCALE or modifying the “Memory Depth”.

The influence on the waveform when the sample rate is too low:

1. Waveform Distortion: when the sample rate is too low, some waveform details are lost and the waveform displayed is rather different from the actual signal.

2. Waveform Confusion: when the sample rate is lower than twice the actual signal frequency (Nyquist Frequency), the frequency of the waveform rebuilt from the sample data is lower than the actual signal frequency.

4-4 MSO1000Z/DS1000Z User’s Guide

Chapter 4 To Set the Sample System

RIGOL

3. Waveform Leakage: when the sample rate is too low, the waveform rebuilt from the sample data does not reflect all the actual signal information.

Pulse disappeared

MSO1000Z/DS1000Z User’s Guide 4-5

RIGOL

Chapter 4 To Set the Sample System

Memory Depth

Memory depth refers to the number of waveform points that the oscilloscope can store in a single trigger sample and it reflects the storage ability of the sample memory. MSO1000Z/DS1000Z provides up to 24 Mpts memory depth (option) and 12

Mpts standard memory depth.

Trigger Point

T

Pre-sample Delayed Sample

Memory Depth

Figure 4-1 Memory Depth

The relation of memory depth, sample rate and horizontal time base scale fulfills the equation below:

MDepth

=

SRate TScale HDivs

(4-1)

MDepth : Memory depth. The unit is pts.

SRate : Sample rate. The unit is Sa/s.

TScale

: Horizontal time base scale. The unit is s/div.

HDivs : Number of grids horizontally. The unit is div. For MSO1000Z/DS1000Z, this value is 12.

Therefore, under the same horizontal time base scale, higher memory depth can ensure higher sample rate.

Press AcquireMem Depth, use to switch to the desired memory depth (the default is auto) and then press down the knob to select the option. You can also press Mem Depth continuously to switch the memory depth.

 For analog channels:

― When a single channel is enabled, the memory depths available include

Auto, 12kPoints, 120kPoints, 1.2MPoints, 12MPoints and 24MPoints

(option).

― When dual channels are enabled, the memory depths available include Auto,

6kPoints, 60kPoints, 600kPoints, 6MPoints and 12MPoints (option).

― When three or four channels are enabled, the memory depths available include Auto, 3kPoints, 30kPoints, 300kPoints, 3MPoints and 6MPoints

4-6 MSO1000Z/DS1000Z User’s Guide

Chapter 4 To Set the Sample System

RIGOL

(option).

 For digital channels:

― When 8 channels are enabled, the memory depths available include Auto,

6kPoints, 60kPoints, 600kPoints, 6MPoints, 12MPoints and 24MPoints

(option).

― When 16 channels are enabled, the memory depths available include Auto,

3kPoints, 30kPoints, 300kPoints, 3MPoints, 6MPoints and 12MPoints

(option).

Note: In “Auto” mode, the oscilloscope selects the memory depth automatically according to the current sample rate.

Antialiasing

At slower sweep speed, the sample rate is reduced and a dedicated display algorithm can be used to minimize the possibility of aliasing.

Press AcquireAnti-aliasing to enable or disable the antialiasing function. By default, antialiasing is disabled and waveform aliasing is more possible.

MSO1000Z/DS1000Z User’s Guide 4-7

Chapter 5 To Trigger the Oscilloscope

RIGOL

Chapter 5 To Trigger the Oscilloscope

As for trigger, you set certain trigger condition according to the requirement and when a waveform in the waveform stream meets this condition, the oscilloscope captures this waveform as well as the neighbouring part and displays them on the screen. For digital oscilloscope, it samples waveform continuously no matter whether it is stably triggered, but only stable trigger can ensure stable display. The trigger module ensures that every time base sweep or acquisition starts from the user-defined trigger condition, namely every sweep is synchronous to the acquisition and the waveforms acquired overlap to display stable waveform.

Trigger setting should be based on the features of the input signal, thus you need to have some knowledge of the signal under test to quickly capture the desired waveform. This oscilloscope provides abundant advanced trigger functions which can help you to focus on the desired waveform details.

The contents of this chapter:

 Trigger Source

 Trigger Mode

 Trigger Coupling

 Trigger Holdoff

 Noise Rejection

 Trigger Type

 Trigger Output Connector

MSO1000Z/DS1000Z User’s Guide 5-1

RIGOL

Chapter 5 To Trigger the Oscilloscope

Trigger Source

Press MENUSource in the trigger control area (TRIGGER) at the front panel to select the desired trigger source. Analog channels CH1-CH4, digital channels D0-D15 or the AC Line can all be used as trigger source.

Analog channel input:

Signals input from analog channels CH1-CH4 can all be used as trigger source. No matter whether the channel selected is enabled, the channel can work normally.

Note: When any channel among D7-D0 is enabled, CH4 cannot be used as trigger source; when any channel among D15-D8 is enabled, CH3 cannot be used as trigger source.

Digital channel input:

Only the digital channels turned on can be used as trigger source. Please refer to the

introduction of “To Turn on/off the Digital Channel” to turn on the desired

digital channel.

AC line:

The trigger signal is obtained from the AC power input of the oscilloscope. AC trigger is usually used to measure signals relevant to the AC power frequency. For example, stably trigger the waveform output from the transformer of a transformer substation; it is mainly used in related measurements of the power industry.

5-2 MSO1000Z/DS1000Z User’s Guide

Chapter 5 To Trigger the Oscilloscope

RIGOL

Trigger Mode

Trigger mode affects the way in which the oscilloscope searches for a trigger. The following is the schematic diagram of the acquisition memory. As shown in the figure below, the position of the trigger event is determined by the reference time point and the delay setting.

Note: The acquisition memory of the oscilloscope is a cyclic buffer and the new data would overwrite the old data until the acquisition finishes.

Trigger Event

Pre-trigger Buffer Delayed Trigger Buffer

Acquisition Memory

Figure 5-1 Schematic Diagram of the Acquisition Memory

Pre-trigger/Delayed trigger:

Acquire data before/after the trigger event. The trigger position is usually at the horizontal center of the screen. In full-screen display, six-grid pre-trigger and delayed trigger information are displayed respectively. You can adjust the horizontal position of the waveform through HORIZONTAL POSITION to view more pre-trigger information or delayed trigger information, through which the signal information before/after the trigger (such as capture the glitch generated by the circuit and analyze the pre-trigger data to find out the reasons for glitch) can be obtained.

Press MODE in the trigger control area (TRIGGER) at the front panel or press

MENU Sweep to select the desired trigger mode. The corresponding status light of the mode currently selected turns on.

Auto:

After this mode is selected, the oscilloscope starts searching for trigger signals that meet the specified condition. If trigger signals that meets the specified condition are found, “T’D” is displayed at the upper-left corner of the user interface and stable waveform is displayed. Otherwise, “AUTO” is displayed and unstable waveform is displayed.

Note: When the horizontal time base setting is 50 ms/div or greater, you can also observe the input waveform without a trigger signal in this mode.

Normal:

After this mode is selected, the oscilloscope enters the wait-for-trigger state and starts searching for trigger signals that meet the specified condition. If trigger signals that meets the specified condition are found, “T’D” is displayed in the running status

MSO1000Z/DS1000Z User’s Guide 5-3

RIGOL

Chapter 5 To Trigger the Oscilloscope

bar and stable waveform is displayed. Otherwise, “WAIT” is displayed and the waveform of the last trigger (there is a last trigger) or no waveform (there is not a last trigger) is displayed.

Single:

When this mode is selected, the backlight of SINGLE turns on. The oscilloscope enters the wait-for-trigger state and starts searching for trigger signals that meet the specified condition. If trigger signals that meets the specified condition are found,

“T’D” is displayed in the running status bar and stable waveform is displayed. Then, the oscilloscope stops sweeping, the backlight of RUN/STOP turns on and “STOP” is displayed in the running status bar. Otherwise, “WAIT” is displayed in the running status bar and no waveform is displayed.

Note: In this mode, pressing FORCE can generate a trigger signal forcefully.

Trigger Coupling

Trigger coupling decides which kind of components will be transmitted to the trigger

module. Please distinguish it from “Channel Coupling”.

 DC: allow DC and AC components into the trigger path.

 AC: block all the DC components and attenuate signals lower than 75 Hz.

 LF Reject: block the DC components and reject the low frequency components

(lower than 75 kHz).

 HF Reject: reject the high frequency components (higher than 75 kHz).

Press MENUSettingCoupling in the trigger control area (TRIGGER) at the front panel to select the desired coupling type (the default is DC).

Note: Trigger coupling is only valid in edge trigger.

5-4 MSO1000Z/DS1000Z User’s Guide

Chapter 5 To Trigger the Oscilloscope

RIGOL

Trigger Holdoff

Trigger holdoff can be used to stably trigger complex waveforms (such as modulated waveform). Holdoff time is the amount of time that the oscilloscope waits for re-arming the trigger module after generating a correct trigger. The oscilloscope will not trigger even if the trigger condition is met during the holdoff time and will only re-arm the trigger module after the holdoff time expires.

Trigger Position

Holdoff Time

Figure 5-2 Schematic Diagram of Trigger Holdoff

Press MENUSettingHoldoff in the trigger control area (TRIGGER) at the front panel and use the multifunction knob to modify the holdoff time (the default is 16 ns) until the waveform triggers stably. The adjustable range of holdoff time is from 16 ns to 10 s.

Note: Trigger holdoff is not available for video trigger, timeout trigger, setup/hold trigger, Nth edge trigger, RS232 trigger, I2C trigger and SPI trigger.

Noise Rejection

Noise rejection can reject the high frequency noise in the signal and reduce the possibility of miss-trigger of the oscilloscope.

Press MENUSettingNoiseReject in the trigger control area (TRIGGER) at the front panel to enable or disable noise rejection.

Note: This function is not available when the current trigger source is a digital channel.

MSO1000Z/DS1000Z User’s Guide 5-5

RIGOL

Chapter 5 To Trigger the Oscilloscope

Trigger Type

MSO1000Z/DS1000Z provides various trigger functions, including various serial bus triggers.

 Edge Trigger

 Pulse Trigger

 Slope Trigger

 Video Trigger

 Pattern Trigger

 Duration Trigger

 TimeOut Trigger (Option)

 Runt Trigger (Option)

 Window Trigger (Option)

 Delay Trigger (Option)

 Setup/Hold Trigger (Option)

 Nth Edge Trigger (Option)

 RS232 Trigger (Option)

 I2C Trigger (Option)

 SPI Trigger (Option)

5-6 MSO1000Z/DS1000Z User’s Guide

Chapter 5 To Trigger the Oscilloscope

RIGOL

Edge Trigger

Trigger on the trigger threshold of the specified edge of the input signal.

Trigger Type:

Press Type, rotate to select “Edge” and press down . At this point, the trigger setting information is displayed at the upper right corner of the screen.

For example, . The trigger type is edge trigger; the trigger source is CH1; the trigger level is 0.00 V.

Source Selection:

Press Source to open the signal source list and select CH1-CH4, AC or D0-D15. For

the details, please refer to the introduction in “ Trigger Source”. The current trigger

source is displayed at the upper right corner of the screen.

Note: Select channel with signal input as trigger source to obtain stable trigger.

Edge Type:

Press Slope to select the kind of edge of the input signal on which the oscilloscope triggers. The current edge type is displayed at the upper right corner of the screen.

 : trigger on the rising edge of the input signal when the voltage level meets the preset trigger level.

: trigger on the falling edge of the input signal when the voltage level meets the preset trigger level.

: trigger on the rising or falling edge of the input signal when the voltage level meets the preset trigger level.

Trigger Mode:

Press Sweep to open the trigger mode list and select auto, normal or single. For the

details, please refer to “Trigger Mode”. The corresponding status light of the

current trigger mode turns on.

Trigger Setting:

Press Setting to set the trigger parameters (trigger coupling, trigger holdoff and noise rejection) under this trigger type.

Trigger Level:

When the trigger source is an analog channel, trigger occurs only when the signal reaches the preset trigger level. You can modify the level using TRIGGER

LEVEL. At this point, an orange trigger level line and the trigger mark “ ” appear on the screen and move up and down with the rotation of the knob. The trigger level value (such as ) at the lower left corner of the screen also changes accordingly. When stopping turning the knob, the trigger level line disappears in about 2 s.

MSO1000Z/DS1000Z User’s Guide 5-7

RIGOL

Chapter 5 To Trigger the Oscilloscope

Pulse Trigger

Trigger on the positive or negative pulse with a specified width. In this mode, the oscilloscope will trigger when the pulse width of the input signal satisfies the specified pulse width condition.

Trigger Type:

Press Type, rotate to select “Pulse” and press down . At this point, the trigger setting information is displayed at the upper right corner of the screen.

For example, . The trigger type is pulse trigger; the trigger source is CH1; the trigger level is 168 mV.

Source Selection:

Press Source to open the signal source list and select CH1-CH4 or D0-D15. For the

details, please refer to the introduction in “ Trigger Source”. The current trigger

source is displayed at the upper right corner of the screen.

Note: Select channel with signal input as trigger source to obtain stable trigger.

Pulse Condition:

In this oscilloscope, positive pulse width is defined as the time difference between the two crossing points of the trigger level and positive pulse; negative pulse width is defined as the time difference between the two crossing points of the trigger level and negative pulse, as shown in the figure below.

A B

Trigger level

Positive pulse width

Negitive pulse width

A B

Trigger level

Figure 5-3 Positive Pulse Width/Negative Pulse Width

Press When to select the desired pulse width condition.

 : trigger when the positive pulse width of the input signal is greater than the specified pulse width.

: trigger when the positive pulse width of the input signal is lower than the specified pulse width.

: trigger when the positive pulse width of the input signal is greater than the specified lower limit of pulse width and lower than the specified upper limit

5-8 MSO1000Z/DS1000Z User’s Guide

Chapter 5 To Trigger the Oscilloscope

RIGOL

 of pulse width.

: trigger when the negative pulse width of the input signal is greater than the specified pulse width.

: trigger when the negative pulse width of the input signal is lower than the specified pulse width.

: trigger when the negative pulse width of the input signal is greater than the specified lower limit of pulse width and lower than the specified upper limit of pulse width.

Pulse Width Setting:

 When the Pulse Condition is set to , , or , press

Setting and use to input the desired value. The range available is from 8 ns to 10 s.

 When the Pulse Condition is set to or , press Upper Limit and

Lower Limit and use to input the desired values respectively. The range of the upper limit is from 16 ns to 10 s. The range of the lower limit is from 8 ns to

9.99 s.

Note : The lower limit of the pulse width must be lower than the upper limit.

Trigger Mode:

Press Sweep to open the trigger mode list and select auto, normal or single. For the

details, please refer to “Trigger Mode”. The corresponding status light of the

current trigger mode turns on.

Trigger Setting:

Press Setting to set the trigger parameters (trigger holdoff and noise rejection) under this trigger type.

Trigger Level:

When the trigger source is an analog channel, you can use TRIGGER LEVEL to

modify the level. For the details, please refer to the introduction of “ Trigger Level”.

MSO1000Z/DS1000Z User’s Guide 5-9

RIGOL

Chapter 5 To Trigger the Oscilloscope

Slope Trigger

In slope trigger, the oscilloscope triggers on the positive or negative slope of the specified time. This trigger mode is applicable to ramp and triangle waveforms.

Trigger Type:

Press Type, rotate to select “Slope” and press down . At this point, the trigger setting information is displayed at the upper right corner of the screen.

For example, . The trigger type is slope trigger; the trigger source is CH1; the difference between the upper limit of trigger level and the lower limit of trigger level is 400 mV.

Source Selection:

Press Source to switch the signal source and select CH1-CH4. For the details, please

refer to the introduction in “ Trigger Source”.

Note: Select channel with signal input as trigger source to obtain stable trigger.

Slope Condition:

In this oscilloscope, positive slope time is defined as the time difference between the two crossing points of trigger level line A and B with the rising edge; negative slope time is defined as the time difference between the two crossing points of trigger level line A and B with the falling edge as shown in the figure below.

B

A

The upper limit of trigger level (Up Level)

The lower limit of trigger level (Low Level)

Positive

Slope Time

Negative

Slope Time

Figure 5-4 Positive Slope Time/Negative Slope Time

Press When to select the desired slope condition.

: trigger when the positive slope time of the input signal is greater than the specified time.

: trigger when the positive slope time of the input signal is lower than the specified time.

: trigger when the positive slope time of the input signal is greater than the specified lower limit of time and lower than the specified upper limit of time.

: trigger when the negative slope time of the input signal is greater than the specified time.

: trigger when the negative slope time of the input signal is lower than the specified time.

5-10 MSO1000Z/DS1000Z User’s Guide

Chapter 5 To Trigger the Oscilloscope

RIGOL

 : trigger when the negative slope time of the input signal is greater than the specified lower limit of time and lower than the specified upper limit of time.

Time Setting:

 When the Slope Condition is set to , , or , press Time and use to input the desired value. The range available is from 8 ns to 10 s.

 When the Slope Condition is set to or , press Upper Limit and Lower Limit and use to input the desired values respectively. The range of time upper limit is from 16 ns to 10 s. The range of the time lower limit is from 8 ns to 9.99 s.

Note: The time lower limit must be lower than the upper limit.

Vertical Window and Trigger Level:

After the trigger condition setting is completed, adjust the trigger level using

TRIGGER LEVEL to correctly trigger the signal and obtain stable waveform. The adjustment mode of the trigger level is different when different vertical window is selected in slope trigger. Press Vertical and use to select the desired vertical window or press down Vertical continuously to switch the vertical window. You can choose to only adjust the upper limit, the lower limit or both of them.

: only adjust the upper limit of the trigger level. During the adjustment,

“UP Level” and “Slew Rate” change accordingly but “Low Level” remains unchanged.

: only adjust the lower limit of the trigger level. During the adjustment,

“Low Level” and “Slew Rate” change accordingly but “UP Level” remains unchanged.

 : adjust the upper and lower limits of the trigger level at the same time.

During the adjustment, “UP Level” and “Low Level” change accordingly but

“Slew Rate” remains unchanged.

When the Slope Condition is set to , , screen, as shown in figure 5-5 (a). The formula of slew rate is:

, or , the current trigger level and slew rate will be displayed at the lower left corner of the

SlewRate =

UpLevel - LowLevel

Time

(5-1)

When the Slope Condition is set to or , the current trigger level and slew rate range will be displayed at the lower left corner of the screen, as shown in figure 5-5 (b). The formula of slew rate range is:

SlewRate =

UpLevel - LowLevel UpLevel - LowLevel

~

UpperLimit LowerLimit

(5-2)

MSO1000Z/DS1000Z User’s Guide 5-11

RIGOL

Chapter 5 To Trigger the Oscilloscope

(a) (b)

Figure 5-5 Trigger Level Information Display

Note: Under the “Slope” trigger menu, you can also press down the trigger level knob continuously to switch the vertical window.

During the adjustment of trigger level, two orange trigger level lines and two trigger marks ( and ) appear on the screen and move up and down with the rotation of the knob. At the same time, the current trigger level information is displayed at the lower right corner of the screen according to different slope condition settings. After stopping turning the knob, the trigger level lines and trigger level information disappear in about 2 s.

Trigger Mode:

Press Sweep to open the trigger mode list and select auto, normal or single. For the

details, please refer to “Trigger Mode”. The corresponding status light of the

current trigger mode turns on.

Trigger Setting:

Press Setting to set the trigger parameters (trigger holdoff and noise rejection) under this trigger type.

5-12 MSO1000Z/DS1000Z User’s Guide

Chapter 5 To Trigger the Oscilloscope

RIGOL

Video Trigger

The waveform of video signal includes image signal and timing signal and different signals adopt different standards and formats. MSO1000Z/DS1000Z provides basic measurement function for video signals. The oscilloscope triggers on the standard video signal field or line of NTSC (National Television Standards Committee), PAL

(Phase Alternating Line) or SECAM (Sequential Couleur A Memoire).

Trigger Type:

Press Type, rotate to select “Video” and press down . At this point, the trigger setting information is displayed at the upper right corner of the screen.

For example, . The trigger type is video trigger; the trigger source is CH1; the trigger level is 0.00 V.

Source Selection:

Press Source to switch the signal source and select CH1-CH4. For the details, please

refer to the introduction in “ Trigger Source”.

Note: Select channel with signal input as trigger source to obtain stable trigger.

Video Polarity:

Press Polarity to select the desired video polarity. The polarities available are positive polarity ( ) and negative polarity ( ).

Sync:

Press Sync to select the desired sync type.

 All Line: trigger on all the horizontal sync pulses.

 Line: for NTSC and PAL/SECAM video standards, trigger on the specified line in the odd or even field.

Note: When this sync trigger mode is selected, you can modify the line number using in Line menu with a step of 1.

The range of the line number is from 1 to 525 (NTSC), 1 to 625 (PAL/SECAM), 1 to 525 (480P) or 1 to 625 (576P).

 Odd: trigger on the rising edge of the first sync signal in the odd field.

 Even: trigger on the rising edge of the first sync signal in the even field.

Video Standard:

Press Standard to select the desired video standard.

 NTSC: the field frequency is 60 fields per second and the frame frequency is 30 frames per second. The TV sweep line is 525 with the even field goes first and the odd field follows behind.

 PAL/SECAM: the frame frequency is 25 frames per second. The TV sweep line is

625.

 480P: the frame frequency is 60 frames per second; the TV sweep line is 525; line-by-line sweep; the line frequency is 31.5 kHz.

 576P: the frame frequency is 60 frames per second; the TV sweep line is 625;

MSO1000Z/DS1000Z User’s Guide 5-13

RIGOL

Chapter 5 To Trigger the Oscilloscope

line-by-line sweep.

Trigger Mode:

Press Sweep to open the trigger mode list and select auto, normal or single. For the

details, please refer to “ Trigger Mode”. The corresponding status light of the

current trigger mode turns on.

Trigger Setting:

Press Setting to set the trigger parameter (noise rejection) under this trigger type.

Trigger Level:

Use TRIGGER LEVEL to modify the level. For the details, please refer to the

introduction of “ Trigger Level”.

Tips

 For a better observation of the waveform details in the vedio signal, you can set a larger memory depth first.

 In the trigger debugging process of vedio signals, the frequencies in different parts of the signal can be reflected by different brightnesses as RIGOL digital oscilloscopes provides the multi-level gray scale display function. Experienced users can quickly judge the signal quality and discover abnormalities.

5-14 MSO1000Z/DS1000Z User’s Guide

Chapter 5 To Trigger the Oscilloscope

RIGOL

Pattern Trigger

Identify a trigger condition by looking for a specified pattern. This pattern is a logical

“AND” combination of channels. Each channel can have a value of high (H), low (L), don’t care (X), rising or falling edge. When a rising or falling edge is specified for one channel, the oscilloscope will trigger at the edge of the specified channel if the input signals for other channels are consistent with the pattern currently set. If no edge is specified, the oscilloscope will trigger on the last edge that makes the pattern true. If all channels in the pattern are set to “Don’t Care”, the oscilloscope will not trigger.

(CH1-CH4 or D0-D15)

H

L

(CH1-CH4 or D0-D15)

Figure 5-6 Pattern Trigger

Trigger Type:

Press Type, rotate to select “Pattern” and press down . At this point, the trigger setting information is displayed at the upper right corner of the screen.

For example, . The trigger type is pattern trigger; the current trigger source is CH1; the trigger level is 0.00 V.

Source Selection:

Press Source to open the signal source list and select CH1-CH4 or D0-D15. The current trigger source is displayed at the upper right corner of the screen.

Pattern Setting:

Press Code to set the pattern of the current signal source. At this point, the corresponding pattern is displayed at the bottom of the screen. The patterns of channels CH1-CH4 and D0-D15 are presented from left to right as shown in the figure below.

 : set the pattern of the channel selected to “H”, namely the voltage level is higher than the trigger level of the channel.

 : set the pattern of the channel selected to “L”, namely the voltage level is lower than the trigger level of the channel.

 : set the pattern of the channel selected to “Don’t Care”, namely this channel is not used as a part of the pattern. When all channels in the pattern are set to

“Don’t Care”, the oscilloscope will not trigger.

or : set the pattern to the rising or falling edge of the channel selected.

MSO1000Z/DS1000Z User’s Guide 5-15

RIGOL

Chapter 5 To Trigger the Oscilloscope

Note:

 When digital channels D7-D0 are enabled, CH4 is automatically disabled; its the corresponding pattern cannot be set and is replaced by X. When D15-D8 are enabled, CH3 is automatically disabled; its corresponding pattern cannot be set and is replaced by X.

 Only one rising or falling edge can be specified in the pattern. If one edge item is currently defined and then another edge item is defined in another channel in the pattern, the former edge item defined will be replaced by X.

All Bits:

Press AllBits to set the patterns of all trigger sources to the pattern settings selected currently.

Trigger Mode:

Press Sweep to open the trigger mode list and select auto, normal or single. For the

details, please refer to “Trigger Mode”. The corresponding status light of the

current trigger mode turns on.

Trigger Setting:

Press Setting to set the trigger parameters (trigger holdoff and noise rejection) under this trigger type.

Trigger Level:

For the analog channels, the trigger level of each channel needs to be set independently. For example, set the trigger level of CH1. Press Source to select CH1, and then use TRIGGER LEVEL to modify the level. For the details, please refer

to the introduction of “Trigger Level”.

5-16 MSO1000Z/DS1000Z User’s Guide

Chapter 5 To Trigger the Oscilloscope

RIGOL

Duration Trigger

In duration trigger, the instrument identifies a trigger condition by looking for the duration of a specified pattern. This pattern is a logical “AND” combination of the channels. Each channel can have a value of high (H), low (L) or don’t care (X). The instrument triggers when the duration (△T) of this pattern meets the preset time, as shown in the figure below.

Figure 5-7 Duration Triger

Trigger Type:

Press Type, rotate to select “Duration” and press down . At this point, the trigger setting information is displayed at the upper right corner of the screen.

For example, . The trigger type is duration trigger; the current trigger source is CH1; the trigger level is 0.00 V.

Source Selection:

Press Source to open the signal source list and select CH1-CH4 or D0-D15. The current trigger source is displayed at the upper right corner of the screen.

Pattern Setting:

Press Code to set the pattern of the current channel. At this point, the pattern setting area (as shown in the figure below) is displayed at the bottom of the screen.

 : set the pattern of the channel selected to “H”, namely the voltage level is higher than the trigger level of the channel.

 : set the pattern of the channel selected to “L”, namely the voltage level is lower than the trigger level of the channel.

MSO1000Z/DS1000Z User’s Guide 5-17

RIGOL

Chapter 5 To Trigger the Oscilloscope

 : set the pattern of the channel selected to “Don’t Care”, namely this channel is not used as a part of the pattern. When all channels in the pattern are set to

“Don’t Care”, the oscilloscope will not trigger.

Note: When digital channels D7-D0 are enabled, CH4 is automatically disabled; its corresponding pattern cannot be set and is replaced by X. When D15-D8 are enabled,

CH3 is automatically disabled; its corresponding pattern cannot be set and is replaced by X.

All Bits:

Press AllBits to set the patterns of all trigger sources to the pattern settings selected currently.

Trigger Condition:

Press When to select the desired trigger condition.

>: trigger when the duration of the pattern is greater than the preset time.

Press Time to set the duration of the pattern and the range is from 8 ns to 10 s.

<: trigger when the duration of the pattern is lower than the preset time. Press

Time to set the duration of the pattern and the range is from 8 ns to 10 s.

<>: trigger when the duration of the pattern is lower than the upper limit of the preset time and greater than the lower limit of the preset time. Press Upper

Limit to set the upper limit of the duration of the pattern and the range is from

16 ns to 10 s. Press Lower Limit to set the lower limit of the duration of the pattern and the range is from 8 ns to 9.99 s.

Note: The time lower limit must be lower than the time upper limit.

Trigger Mode:

Press Sweep to open the trigger mode list and select auto, normal or single. For the

details, please refer to “ Trigger Mode”. The corresponding status light of the

current trigger mode turns on.

Trigger Setting:

Press Setting to set the trigger parameters (trigger holdoff and noise rejection) under this trigger type.

Trigger Level:

Press Source to select CH1 to CH4 respectively; use TRIGGER LEVEL to modify the trigger level for each channel. For the details, please refer to the introduction of

Trigger Level”.

5-18 MSO1000Z/DS1000Z User’s Guide

Chapter 5 To Trigger the Oscilloscope

RIGOL

TimeOut Trigger (Option)

In timeout trigger, the instrument triggers when the time interval (△T) from when the rising edge (or falling edge) of the input signal passes through the trigger level to when the neighbouring falling edge (or rising edge) passes through the trigger level is greater than the timeout time set, as shown in the figure below.

T Time Out< T

Time Out

Figure 5-8 TimeOut Trigger

Trigger Type:

Press Type, rotate to select “TimeOut” and press down . At this point, the trigger setting information is displayed at the upper right corner of the screen.

For example, . The trigger type is timeout trigger; the trigger source is CH1; the trigger level is 0.00 V.

Source Selection:

Press Source to open the signal source list and select CH1-CH4 or D0-D15. For the

details, please refer to the introduction in “ Trigger Source”. The current trigger

source is displayed at the upper right corner of the screen.

Note: Select channel with signal input as trigger source to obtain stable trigger.

Edge Type:

Press Slope to select the type of the first edge of the input signal that passes through the trigger level.

: start timing when the rising edge of the input signal passes through the trigger level.

: start timing when the falling edge of the input signal passes through the trigger level.

: start timing when any edge of the input signal passes through the trigger level. Note: When the trigger source is a digital channel, this option will be hidden automatically.

MSO1000Z/DS1000Z User’s Guide 5-19

RIGOL

Chapter 5 To Trigger the Oscilloscope

Timeout Time:

Timeout time represents the minimum time that the clock signal must be in the idle state before the oscilloscope starts searching for the data meets the trigger condition.

Press TimeOut to set the timeout time of timeout trigger and the range is from 16 ns to 10 s.

Trigger Mode:

Press Sweep to open the trigger mode list and select auto, normal or single. For the

details, please refer to “Trigger Mode”. The corresponding status light of the

current trigger mode turns on.

Trigger Setting:

Press Setting to set the trigger parameter (noise rejection) under this trigger type.

Trigger Level:

When the trigger source is an analog channel, you can use TRIGGER LEVEL to

modify the level. For the details, please refer to the introduction of “Trigger Level”.

5-20 MSO1000Z/DS1000Z User’s Guide

Chapter 5 To Trigger the Oscilloscope

RIGOL

Runt Trigger (Option)

This trigger mode is used to trigger pulses that pass through one trigger level but fail to pass through the other trigger level as shown in the figure below.

Figure 5-9 Runt Trigger

Trigger Type:

Press Type, rotate to select “Runt” and press down . At this point, the trigger setting information is displayed at the upper right corner of the screen.

For example, . The trigger type is runt trigger; the trigger source is CH1; the difference between the upper limit of trigger level and the lower limit of trigger level is 10.00 V.

Source Selection:

Press Source to switch the signal source and select CH1-CH4. For the details, please

refer to the introduction in “Trigger Source”. The current trigger source is displayed

at the upper right corner of the screen.

Pulse Polarity:

Press Polarity to select the pulse polarity of runt trigger.

 : positive polarity. The instrument triggers on the positive runt pulse.

: negative polarity. The instrument triggers on the negative runt pulse.

Qualifier:

Press Qualifier to set the trigger conditions of runt trigger.

None: do not set the trigger condition of runt trigger.

>: trigger when the runt pulse width is greater than the lower limit of pulse width. Press Lower Limit to set the minimum pulse width of runt trigger. The range available is from 8.00 ns to 10 s.

<: trigger when the runt pulse width is lower than the upper limit of pulse width.

Press Upper Limit to set the maximum pulse width of runt trigger. The range available is from 16.0 ns to 10.0 s.

<>: trigger when the runt pulse width is greater than the lower limit and lower than the upper limit of pulse width. Press Upper Limit to set the maximum

MSO1000Z/DS1000Z User’s Guide 5-21

RIGOL

Chapter 5 To Trigger the Oscilloscope

pulse width of runt trigger and the range is from 16.0 ns to 10.0 s; press Lower

Limit to set the minimum pulse width of runt trigger and the range is from 8.00 ns to 10 s.

Note: The lower limit of the pulse width must be lower than the upper limit.

Vertical Window and Trigger Level:

After the trigger condition setting is completed, you need to adjust the trigger level to correctly trigger the signal and obtain stable waveform. The adjustment mode of the trigger level is different when different vertical window is selected in the runt trigger. Press Window and use to select the desired vertical window or press down Window continuously to switch the vertical window. You can choose to only adjust the upper limit, the lower limit or both of them. For more details, please refer to the following introduction.

Note: Under the “Runt” trigger menu, you can also press down TRIGGER

LEVEL continuously to switch the vertical window.

After the vertical window type is selected, you can use Trigger LEVEL to adjust the trigger level. During the adjustment, two orange trigger level lines and trigger marks ( and ) appear on the screen and move up and down with the rotation of the knob. At the same time, the current trigger level values are displayed at the lower left corner of the screen (as shown in the figure below). The trigger level lines and trigger marks disappear after you stop rotating the knob for about 2 s.

The adjustment mode of the trigger level differs when different vertical window is selected.

 : only adjust the upper limit of the trigger level. During the adjustment, the “Up Level” changes accordingly and “Low Level” remains unchanged.

 : only adjust the lower limit of the trigger level. During the adjustment, the “Low Level” changes accordingly and the “Up Level” remains unchanged.

 : adjust the upper and lower limits of the trigger level at the same time.

During the adjustment, the “Up Level” and “Low Level” change accordingly.

Trigger Mode:

Press Sweep to open the trigger mode list and select auto, normal or single. For the

details, please refer to “Trigger Mode”. The corresponding status light of the

current trigger mode turns on.

Trigger Setting:

Press Setting to set the trigger parameters (trigger holdoff and noise rejection) under this trigger type.

5-22 MSO1000Z/DS1000Z User’s Guide

Chapter 5 To Trigger the Oscilloscope

RIGOL

Window Trigger (Option)

Window trigger provides a high trigger level and a low trigger level. The instrument triggers when the input signal passes through the high trigger level or the low trigger level.

Trigger Type:

Press Type, rotate to select “Window” and press down . At this point, the trigger setting information is displayed at the upper right corner of the screen.

For example, . The trigger type is window trigger; the trigger source is CH1; the difference between the upper limit of trigger level and the lower limit of trigger level is 2.00 V.

Source Selection:

Press Source to switch the trigger source and select CH1-CH4. For the details,

please refer to the introduction in “ Trigger Source”. The current trigger source is

displayed at the upper right corner of the screen.

Note: Select channel with signal input as trigger source to obtain stable trigger.

Window Type:

Press WndType to select the kind of edge of the input signal on which the oscilloscope triggers.

: trigger on the rising edge of the input signal when the voltage level is greater than the preset high trigger level.

: trigger on the falling edge of the input signal when the voltage level is lower than the preset low trigger level.

 : trigger on any edge of the input signal when the voltage level meets the preset trigger level.

Trigger Position:

After selecting the window type, press Position to further specify the time point of trigger by selecting the trigger position.

Enter: trigger when the trigger signal enters the specified trigger level range.

Exit: trigger when the input signal exits the specified trigger level range.

Time: trigger when the accumulated hold time after the trigger signal enters the specified trigger level range is equal to the window time. When this type is selected, press Time to set the window time. The range is from 8 ns to 10 s. For

the setting method, please refer to the introduction in “Parameter Setting

Method”.

Vertical Window and Trigger Level:

Press Window to select the desired vertical window type and use Trigger

LEVEL to adjust the trigger level. For detailed operation, please refer to the

introduction in “Vertical Window and Trigger Level”.

MSO1000Z/DS1000Z User’s Guide 5-23

RIGOL

Chapter 5 To Trigger the Oscilloscope

Trigger Mode:

Press Sweep to open the trigger mode list and select auto, normal or single. For the

details, please refer to “Trigger Mode”. The corresponding status light of the

current trigger mode turns on.

Trigger Setting:

Press Setting to set the trigger parameters (trigger holdoff and noise rejection) under this trigger type.

5-24 MSO1000Z/DS1000Z User’s Guide

Chapter 5 To Trigger the Oscilloscope

RIGOL

Delay Trigger (Option)

In delay trigger, you need to set signal source A and signal source B. The oscilloscope triggers when the time difference (△T) between the specified edges of source A

(Edge A) and source B (Edge B) meets the preset time limit, as shown in the figure below.

Note: Edge A and Edge B must be neighbouring edges.

Edge A=

Edge B=

Source A

Source B

T

Figure 5-10 Delay Trigger

Trigger Type:

Press Type to open the trigger type list. Rotate to select “Delay” and press down

. At this point, the trigger setting information is displayed at the upper right corner of the screen.

For example, . The trigger type is delay trigger; the current trigger source is CH1; the trigger level is 0.00 V.

Source A:

Press SourceA to select CH1-CH4 as the trigger source of signal source A.

Edge A:

Press EdgeA to select the trigger edge type of signal source A in delay trigger. It can be set to the rising edge or falling edge.

Source B:

Press SourceB to select CH1-CH4 as the trigger source of signal source B.

Edge B:

Press EdgeB to select the trigger edge type of signal source B in delay trigger. It can be set to the rising edge or falling edge.

Delay Type:

Press DelayType to set the time limit condition of delay trigger.

MSO1000Z/DS1000Z User’s Guide 5-25

RIGOL

Chapter 5 To Trigger the Oscilloscope

>: trigger when the time difference ( △T) between the specified edges of source

A and source B is greater than the preset time lower limit. Press Lower Limit to set the delay time lower limit in delay trigger and the range is from 8.00 ns to 10

s. For the setting method, please refer to the introduction in “Parameter

Setting Method”.

<: trigger when the time difference ( △T) between the specified edges of source

A and source B is lower than the preset time upper limit. Press Upper Limit to set the delay time upper limit in delay trigger and the range is from 16.0 ns to

10.0 s. For the setting method, please refer to the introduction in “Parameter

Setting Method”.

<>: trigger when the time difference ( △T) between the specified edges of source A and source B is greater than the lower limit of the preset time and lower than the upper limit of the preset time. Press Upper Limit to set the upper limit of the delay time in delay trigger and the range is from 16.0 ns to

10.0 s. Press Lower Limit to set the lower limit of the delay time in delay trigger and the range is from 8.00 ns to 9.99 s. For the setting method, please

refer to the introduction in “Parameter Setting Method”.

Note: The time lower limit must be lower than the time upper limit.

><: trigger when the time difference ( △T) between the specified edges of source A and source B is lower than the lower limit of the preset time or greater than the upper limit of the preset time. Press Upper Limit to set the upper limit of the delay time in delay trigger and the range is from 16.0 ns to 10.0 s. Press

Lower Limit to set the lower limit of the delay time in delay trigger and the range is from 8.00 ns to 9.99 s. For the setting method, please refer to the

introduction in “Parameter Setting Method”.

Note: The time lower limit must be lower than the time upper limit.

Trigger Mode:

Press Sweep to open the trigger mode list and select auto, normal or single. For the

details, please refer to “ Trigger Mode”. The corresponding status light of the

current trigger mode turns on.

Trigger Setting:

Press Setting to set the trigger parameters (trigger holdoff and noise rejection) under this trigger type.

Trigger Level:

The trigger level of each source needs to be set independently. For example, set the trigger level of source A. Press SourceA to select the desired source, and then use

TRIGGER LEVEL to modify the level. For the details, please refer to the

introduction of “Trigger Level”.

5-26 MSO1000Z/DS1000Z User’s Guide

Chapter 5 To Trigger the Oscilloscope

RIGOL

Setup/Hold Trigger (Option)

Setup time (△T1) indicates the time that the data stays stable and constant before the rising edge of the clock signal appears. The hold time (△T2) indicates the time that the data stays stable and constant after the clock signal appears, as shown in the figure below. In setup/hold trigger, the instrument triggers when the setup time or the hold time of the data signal detected by the trigger is less than the preset setup time or hold time.

T1

T2

T1 is the setup time

T2 is the hold time

Data Source

Clock Source

The edge type is The data type is H

Figure 5-11 Setup/Hold Trigger

Trigger Type:

Press Type to open the trigger type list. Rotate to select “StpHold” and press down . At this point, the trigger setting information is displayed at the upper right corner of the screen.

For example, . The trigger type is setup/hold trigger; the current trigger source is CH1; the trigger level is 0.00 V.

Source Selection:

Press DataSource and ClkSource to set the signal sources of the data line and clock line respectively. They can be set to CH1-CH4. The current trigger source is displayed at the upper right corner of the screen.

Edge Type:

Press Slope to select the desired clock edge type and it can be set to the rising edge or falling edge.

Data Type:

Press DataType to set the effective pattern of the data signal to H (high level) or L

(low level).

Setup Type:

It is used to select the different trigger types of setup/hold trigger. Press SetupType to select the desired setup type.

MSO1000Z/DS1000Z User’s Guide 5-27

RIGOL

Chapter 5 To Trigger the Oscilloscope

Setup: the oscilloscope triggers when the setup time is less than the specified setup time. After selecting this type, press Setup to set the setup time and the range is from 8 ns to 1 s.

Hold: the oscilloscope triggers when the hold time is less than the specified hold time. After selecting this type, press Hold to set the hold time and the range is from 8 ns to 1 s.

StpHold: the oscilloscope triggers when the setup time and hold time are less than the specified times. After selecting this type, press Setup and Hold to set the setup time and hold time respectively and the range is from 8 ns to 1 s.

Trigger Mode:

Press Sweep to open the trigger mode list and select auto, normal or single. For the

details, please refer to “ Trigger Mode”. The corresponding status light of the

current trigger mode turns on.

Trigger Setting:

Press Setting to set the trigger parameter (noise rejection) under this trigger type.

Trigger Level:

Press DataSource and use TRIGGER LEVEL to modify the trigger level of the data source channel. Press ClkSource and use TRIGGER LEVEL to modify the trigger level of the clock source channel. For the details, please refer to the

introduction of “Trigger Level”.

5-28 MSO1000Z/DS1000Z User’s Guide

Chapter 5 To Trigger the Oscilloscope

RIGOL

Nth Edge Trigger (Option)

Trigger on the nth edge that appears after the specified idle time. For example, in the waveform shown in the figure below, the instrument should trigger on the second rising edge after the specified idle time (the time between two neighbouring rising edges) and the idle time should be set to P<Idle Time<M. Wherein, M is the time between the first rising edge and its previous rising edge and P is the maximum time between the rising edges participate in the counting.

Edge Type=

Edge Num=2

M P

P<Idle Time<M

Figure 5-12 Nth Edge Trigger

Trigger Type:

Press Type to open the trigger type list. Rotate to select “Nth” and press down

. At this point, the trigger setting information is displayed at the upper right corner of the screen.

For example, . The trigger type is Nth edge trigger; the trigger source is CH1; the trigger level is 0.00 V.

Source Selection:

Press Source to open the signal source list and select CH1-CH4 or D0-D15. For the

details, please refer to the introduction in “ Trigger Source”. The current trigger

source is displayed at the upper right corner of the screen.

Note: Select channel with signal input as trigger source to obtain stable trigger.

Edge Type:

Press Slope to select the kind of edge of the input signal on which the oscilloscope triggers.

 : trigger on the rising edge of the input signal when the voltage level meets the specified trigger level.

 : trigger on the falling edge of the input signal when the voltage level meets the specified trigger level.

MSO1000Z/DS1000Z User’s Guide 5-29

RIGOL

Chapter 5 To Trigger the Oscilloscope

Idle Time:

Press Idle to set the idle time before the edge counting in Nth edge trigger. The range available is from 16 ns to 10 s.

Edge Number:

Press Edge to set the value of “N” in Nth edge trigger and the range available is from

1 to 65535.

Trigger Mode:

Press Sweep to open the trigger mode list and select auto, normal or single. For the

details, please refer to “Trigger Mode”. The corresponding status light of the

current trigger mode turns on.

Trigger Setting:

Press Setting to set the trigger parameter (noise rejection) under this trigger type.

Trigger Level:

When the trigger source is an analog channel, you can use TRIGGER LEVEL to

modify the level. For the details, please refer to the introduction of “Trigger Level”.

5-30 MSO1000Z/DS1000Z User’s Guide

Chapter 5 To Trigger the Oscilloscope

RIGOL

RS232 Trigger (Option)

RS232 bus is a serial communication mode used in the data transimmision between

PCs or between PC and terminal. In RS232 serial protocol, a character is transimitted as a frame of data which consists of 1 bit start bit, 5~8 bits data bits, 1 bit check bit and 1~2 bits stop bit(s). Its format is as shown in the figure below.

MSO1000Z/DS1000Z oscilloscope triggers when the start frame, error frame, check error or the specified data of the RS232 signal is detected.

Figure 5-13 Schematic Diagram of RS232 Protocol

Trigger Type:

Press Type to open the trigger type list. Rotate to select ”RS232” and press down . At this point, the trigger setting information is displayed at the upper right corner of the screen.

For example, . The trigger type is RS232 trigger; the current trigger source is CH1; the trigger level is 0.00 V.

Source Selection:

Press Source to open the signal source list and select CH1-CH4 or D0-D15. For the

details, please refer to the introduction in “ Trigger Source”. The current trigger

source is displayed at the upper right corner of the screen.

Polarity

Press Polarity to select the polarity of data transmission. It can be set to “ or “ ” and the default is .

Trigger Condition:

Press When to select the desired trigger condition.

 Start: trigger on the start frame position.

 Error: trigger when error frame is detected. After this trigger condition is selected:

--press Stop Bit to select “1 bit” or “2 bits”;

--press EvenOdd to select “None”, “Odd” or “Even”.

The oscilloscope will determine error frame according to the preset parameters.

 ChkError: trigger when check error is detected. When this trigger condition is selected, press EvenOdd to select “Odd” or “Even”. The oscilloscope will determine check error according to the preset parameters.

 Data: trigger on the last bit of the preset data bits. When this trigger condition is selected:

MSO1000Z/DS1000Z User’s Guide 5-31

RIGOL

Chapter 5 To Trigger the Oscilloscope

--press Data Bits to select “5 bits”, “6 bits”, “7 bits” or “8 bits”;

--press Data and use to set the data value of RS232 trigger. According to the setting in Data Bits and the ranges can be from 0 to 31, from 0 to 63, from

0 to 127 or from 0 to 255 respectively.

Baud Rate:

Set the baud rate of data transmission (equal to specifying a clock frequency). Press

BaudRate to set the desired baud rate to 2400 bps, 4800 bps, 9600 bps (default),

19200 bps, 38400 bps, 57600 bps, 115200 bps, 230400 bps, 460800 bps, 921600 bps, 1 Mbps and user-defined. When “User” is selected, press Setup to set a more specific value from 110 to 20000000 with an adjustment step of 1 bps. For the

setting method, please refer to the introduction in “Parameter Setting Method”.

Trigger Mode:

Press Sweep to open the trigger mode list and select auto, normal or single. For the

details, please refer to “Trigger Mode”. The corresponding status light of the

current trigger mode turns on.

Trigger Setting:

Press Setting to set the trigger parameter (noise rejection) under this trigger type.

Trigger Level:

When the trigger source is an analog channel, you can use TRIGGER LEVEL to

modify the level. For the details, please refer to the introduction of “Trigger Level”.

5-32 MSO1000Z/DS1000Z User’s Guide

Chapter 5 To Trigger the Oscilloscope

RIGOL

I2C Trigger (Option)

I2C is a 2-wire serial bus used to connect the microcontroller and its peripheral device and is a bus standard widely used in the microelectronic communication control field.

The I2C serial bus consists of SCL and SDA. Its transmission rate is determined by

SCL and its transmission data is determined by SDA, as shown in the figure below.

MSO1000Z/DS1000Z triggers on the start condition, restart, stop, missing acknowledgement, specific device address or data value. Besides, it can also trigger on the specific device address and data values at the same time.

Figure 5-14 Schematic Diagram of I2C Protocol

Trigger Type:

Press Type to open the trigger type list. Rotate to select “I2C” and press down

. At this point, the trigger setting information is displayed at the upper right corner of the screen.

For example, . The trigger type is I2C trigger; the current trigger source is CH1; the trigger level is 0.00 V.

Source Selection:

Press SCL and SDA to specify the signal sources of SCL and SDA respectively. They can be set to CH1-CH4 or D0-D15. The current trigger source is displayed at the upper right corner of the screen.

Trigger Condition:

Press When to select the desired trigger condition.

 Start: trigger when SDA data transitions from high level to low level while SCL is high level.

 Restart: trigger when another start condition occurs before a stop condition.

 Stop: trigger when SDA data transitions from low level to high level while SCL is high level.

 MissedACK: trigger when the SDA data is high level during any acknowledgement of SCL clock position.

 Address: the trigger searches for the specified address value. When this event

MSO1000Z/DS1000Z User’s Guide 5-33

RIGOL

Chapter 5 To Trigger the Oscilloscope

occurs, the oscilloscope will trigger on the read/write bit. After this trigger condition is selected:

--press AddrBits to select “7 bits”, “8 bits” or “10 bits”;

--press Address to set the address value of I2C trigger. According to the setting in AddrBits, the range can be from 0 to 0x7F, from 0 to 0xFF or from 0 to

0x3FF;

--press Direction to select “Read”, “Write” or “R/W”. Note: This setting is not available when AddrBits is set to “8 bits”.

 Data: the trigger searches for the specified data value on the data line (SDA).

When this event occurs, the oscilloscope will trigger on the clock line (SCL) transition edge of the last bit of data. After this trigger condition is selected:

--press Bit X to select the desired data bit and the range is from 0 to (Byte

Length×8-1);

--press Data to set the data pattern of the current data bit to X, H or L;

--press Bytes to set the length of the data and the range is from 1 to 5;

--press AllBits to set the data pattern of all the data bits to the data pattern specified in Data.

 A&D: the oscilloscope searches for the specified address and data at the same time and triggers when the specified address and data are found. After this trigger condition is selected:

--press AddrBits to select “7 bits”, “8 bits” or “10 bits”;

--press Address and use to set the address value. According to the setting in AddrBits, the range can be from 0 to 0x7F, from 0 to 0xFF or from 0 to

0x3FF;

--press Bit X to select the desired data bit and the range is from 0 to (Byte

Length×8-1);

--press Data to set the data pattern of the current data bit to X, H or L;

--press Bytes to set the length of the data and the range is from 1 to 5;

--press AllBits to set the data pattern of all the data bits to the data pattern specified in Data;

--press Direction to select “Read”, “Write” or “R/W”.

Note: This setting is not available when AddrBits is set to “8 bits”.

Trigger Mode:

Press Sweep to open the trigger mode list and select auto, normal or single. For the

details, please refer to “ Trigger Mode”. The corresponding status light of the

current trigger mode turns on.

Trigger Setting:

Press Setting to set the trigger parameter (noise rejection) under this trigger type.

Trigger Level:

When SCL is an analog channel, press SCL and use TRIGGER LEVEL to modify the trigger level of the SCL channel. When SDA is an analog channel, press SDA and use TRIGGER LEVEL to modify the trigger level of the SDA channel. For the

details, please refer to the introduction of “Trigger Level”.

5-34 MSO1000Z/DS1000Z User’s Guide

Chapter 5 To Trigger the Oscilloscope

RIGOL

SPI Trigger (Option)

In SPI trigger, after the CS or timeout condition is satisfied, the oscilloscope triggers when the specified data is found. When using SPI trigger, you need to specify the

SCL clock sources and SDA data sources. Below is the sequential chart of SPI bus.

CS

SCL

SDA

D7 D6 D5 D4 D3 D2 D1 D0

Figure 5-15 Sequential Chart of SPI Bus

Trigger Type:

Press Type to open the trigger type list. Rotate to select “SPI”and press down

. At this point, the trigger setting information is displayed at the upper right corner of the screen.

For example, . The trigger type is SPI trigger; the current trigger source is CH1; the trigger level is 0.00 V.

Source Selection:

Press SCL and SDA to specify the data sources of SCL and SDA respectively. They can be set to CH1-CH4 or D0-D15. The current trigger source is displayed at the upper right corner of the screen.

Data Line Setting:

Press Data to set the data bits and data of SPI trigger.

 Press DataBits to set the number of bits of the serial data character string. It can be set to any integer between 4 and 32.

 Press CurrentBit to set the number of the data bit and the range is from 0 to

(value specified in DataBits – 1).

 Press Data to set the value of the current bit to H, L or X.

 Press AllBits to set all the data bits to the value specified in Data.

Trigger Condition:

Press When to select the desired trigger condition.

CS: if the CS signal is valid, the oscilloscope will trigger when the data (SDA) satisfying the trigger conditions is found.

― After selecting this condition, you can press CS to select the chip selection signal line. The available channels are CH1-CH4 or D0-D15 (only the channels currently enabled can be selected).

― After selecting this condition, you can press Mode to set the CS mode to

MSO1000Z/DS1000Z User’s Guide 5-35

RIGOL

Chapter 5 To Trigger the Oscilloscope

“ ” (high level is valid) or “ ” (low level is valid).

Timeout: the clock (SCL) signal need to maintain a certain idle time before the oscilloscope searches for a trigger. The oscilloscope will trigger on when the data

(SDA) satisfying the trigger conditions is found. After selecting this condition, you can press Timeout to set the minimum idle time and the range is from 100 ns to 1 s.

Clock Edge:

Press ClockEdge to select the desired clock edge.

 : sample the SDA data on the rising edge of the clock.

 : sample the SDA data on the falling edge of the clock.

Trigger Mode:

Press Sweep to open the trigger mode list and select auto, normal or single. For the

details, please refer to “ Trigger Mode”. The corresponding status light of the

current trigger mode turns on.

Trigger Setting:

Press Setting to set the trigger parameter (noise rejection) under this trigger type.

Trigger Level:

When SCL is an analog channel, press SCL and use TRIGGER LEVEL to modify the trigger level of the SCL channel. When SDA is an analog channel, press SDA and use TRIGGER LEVEL to modify the trigger level of the SDA channel. For the

details, please refer to the introduction of “ Trigger Level”.

5-36 MSO1000Z/DS1000Z User’s Guide

Chapter 5 To Trigger the Oscilloscope

RIGOL

Trigger Output Connector

The trigger output connector ([Trigger Out]) at the rear panel can output trigger signals determined by the current setting.

Trigger Output Connector

A signal which reflects the current oscilloscope capture rate can be output from

[Trigger Out] connector each time a trigger is generated by the oscilloscope. If this signal is connected to a waveform display device to measure the frequency, the measurement result is equal to the current capture rate.

Note: Press UtilityAux Out to select “Pass/Fail” and press UtilityPass/Fail

Aux Out to select “ON”. In the pass/fail test, when a failure is detected, the oscilloscope outputs a positive pulse from the [Trigger Out] connector; when no failure is detected, the oscilloscope outputs a low level continuously from this connector.

MSO1000Z/DS1000Z User’s Guide 5-37

Chapter 6 MATH and Measurement

RIGOL

Chapter 6 MATH and Measurement

MSO1000Z/DS1000Z can make math operation, cursor measurement and auto measurement after data is sampled and displayed.

The contents of this chapter:

 Math Operation

 Auto Measurement

 Cursor Measurement

MSO1000Z/DS1000Z User’s Guide 6-1

RIGOL

Chapter 6 MATH and Measurement

Math Operation

MSO1000Z/DS1000Z can realize various math operations, including:

 Algebra operations: A+B, A-B, A×B and A/B

 Spectrum operation: FFT

 Logic operations: A&&B, A||B, A^B and !A

 Function operations: Intg, Diff, Sqrt, Lg, Ln, Exp and Abs

 Fx operations: combination of two operations. For the details, please refer to the

introduction in “Fx Operation”.

The results of math operation also allow further measurement.

Press MATHMathOperator in the vertical control area (VERTICAL) at the front panel to select the desired operation function. Press Operation to enable the operation. The result of math operation is displayed on the waveform marked with

“M” on the screen.

Addition

Add the waveform voltage values of signal source A and B point by point and display the results.

Press MATHMathOperator to select “A+B”:

 Press Operation to enable or disable the addition operation function.

 Press SourceA and SourceB to select the desired channels (CH1, CH2, CH3,

CH4 or fx (please refer to the introduction in “Fx Operation”)).

 Press Offset and use to adjust the vertical position of the operation results.

 Press Scale and use to adjust the vertical scale of the operation results.

 Press Scale Reset to adjust the vertical scale of the operation results to the optimal value according to the current configuration.

 Press Options to set the start and end points of the operation results, enable or disable waveform invert, etc.

― Press Start and use to set the start point of the operation results.

― Press End and use to set the end point of the operation results.

― Press Invert to enable or disable the inverted display function of the waveform.

― Press Auto Scale to enable or disable the auto scale function. When auto scale is enabled, the instrument will adjust the vertical scale of the operation results to the optimal value according to the current configuration.

― Press fx Operator, fx A and fx B to set the operator and signal sources of the inner layer operation of Fx operation (please refer to the introduction in

Fx Operation”).

Note: Sens. and Smooth are grayed out and disabled. Sens. is only available when a digital channel is selected as the source. Smooth is only available for differential operation.

6-2 MSO1000Z/DS1000Z User’s Guide

Chapter 6 MATH and Measurement

RIGOL

 You can also use HORIZONTAL POSITION and HORIZONTAL

SCALE to adjust the horizontal position and scale of the operation results.

Subtraction

Subtract the waveform voltage values of signal source B from that of source A point by point and display the results.

Press MATHMathOperator to select “A-B”:

 Press Operation to enable or disable the subtraction operation function.

 Press SourceA and SourceB to select the desired channels (CH1, CH2, CH3,

CH4 or fx (please refer to the introduction in “Fx Operation”)).

 Press Offset and use to adjust the vertical position of the operation results.

 Press Scale and use to adjust the vertical scale of the operation results.

 Press Scale Reset to adjust the vertical scale of the operation results to the optimal value according to the current configuration.

 Press Options to set the start and end points of the operation results, enable or disable waveform invert, etc.

― Press Start and use to set the start point of the operation results.

― Press End and use to set the end point of the operation results.

― Press Invert to enable or disable the inverted display function of the waveform.

― Press Auto Scale to enable or disable the auto scale function. When auto scale is enabled, the instrument will adjust the vertical scale of the operation results to the optimal value according to the current configuration.

― Press fx Operator, fx A and fx B to set the operator and signal sources of the inner layer operation of Fx operation (please refer to the introduction in

Fx Operation”).

Note: Sens. and Smooth are grayed out and disabled. Sens. is only available when a digital channel is selected as the source. Smooth is only available for differential operation.

 You can also use HORIZONTAL POSITION and HORIZONTAL

SCALE to adjust the horizontal position and scale of the operation results.

Multiplication

Multiply the waveform voltage values of signal source A and B point by point and display the results.

Press MATHMathOperator to select “A×B”:

 Press Operation to enable or disable the multiplication operation function.

 Press SourceA and SourceB to select the desired channels (CH1, CH2, CH3,

MSO1000Z/DS1000Z User’s Guide 6-3

RIGOL

Chapter 6 MATH and Measurement

CH4 or fx (please refer to the introduction in “Fx Operation”)).

 Press Offset and use to adjust the vertical position of the operation results.

 Press Scale and use to adjust the vertical scale of the operation results.

 Press Scale Reset to adjust the vertical scale of the operation results to the optimal value according to the current configuration.

 Press Options to set the start and end points of the operation results, enable or disable waveform invert, etc.

― Press Start and use to set the start point of the operation results.

― Press End and use to set the end point of the operation results.

― Press Invert to enable or disable the inverted display function of the waveform.

― Press Auto Scale to enable or disable the auto scale function. When auto scale is enabled, the instrument will adjust the vertical scale of the operation results to the optimal value according to the current configuration.

― Press fx Operator, fx A and fx B to set the operator and signal sources of the inner layer operation of Fx operation (please refer to the introduction in

Fx Operation”).

Note: Sens. and Smooth are grayed out and disabled. Sens. is only available when a digital channel is selected as the source. Smooth is only available for differential operation.

 You can also use HORIZONTAL POSITION and HORIZONTAL

SCALE to adjust the horizontal position and scale of the operation results.

Division

Divide the waveform voltage values of signal source A by that of source B point by point and display the results. It can be used to analyze the multiple relationships of waveforms of two channels.

Note: When the voltage of signal source B is 0V, the division result is invalid and

“NAN” is displayed at the bottom of the screen.

Press MATHMathOperator to select “A/B”:

 Press Operation to enable or disable the division operation function.

 Press SourceA and SourceB to select the desired channels (CH1, CH2, CH3,

CH4 or fx (please refer to the introduction in “Fx Operation”)).

 Press Offset and use to adjust the vertical position of the operation results.

 Press Scale and use to adjust the vertical scale of the operation results.

 Press Scale Reset to adjust the vertical scale of the operation results to the optimal value according to the current configuration.

 Press Options to set the start and end points of the operation results, enable or disable waveform invert, etc.

― Press Start and use to set the start point of the operation results.

― Press End and use to set the end point of the operation results.

6-4 MSO1000Z/DS1000Z User’s Guide

Chapter 6 MATH and Measurement

RIGOL

― Press Invert to enable or disable the inverted display function of the waveform.

― Press Auto Scale to enable or disable the auto scale function. When auto scale is enabled, the instrument will adjust the vertical scale of the operation results to the optimal value according to the current configuration.

― Press fx Operator, fx A and fx B to set the operator and signal sources of the inner layer operation of Fx operation (please refer to the introduction in

Fx Operation”).

Note: Sens. and Smooth are grayed out and disabled. Sens. is only available when a digital channel is selected as the source. Smooth is only available for differential operation.

 You can also use HORIZONTAL POSITION and HORIZONTAL

SCALE to adjust the horizontal position and scale of the operation results.

MSO1000Z/DS1000Z User’s Guide 6-5

RIGOL

Chapter 6 MATH and Measurement

FFT

FFT (Fast Fourier Transform) is used to transform time domain signals to frequency domain components (frequency spectrum). MSO1000Z/DS1000Z oscilloscope provides FFT operation function which enables users to observe the time domain waveform and spectrum of the signal at the same time. FFT operation can facilitate the following works:

 Measure harmonic components and distortion in the system

 Display the characteristics of the noise in DC power

 Analyze vibration

Press MATHMathOperator to select “FFT”. You can set the parameters of

FFT operation.

FFT Vertical

Scale

FFT Horizontal

Scale

FFT Sample

Rate

Time Domain

Waveform

Frequency Domain

Waveform

Figure 6-1 FFT Operation

1. Operation

Press Operation to enable or disable the FFT operation function. When FFT operation is enabled, the time domain waveform and frequency domain

waveform are displayed separately on the screen by default as shown in Figure

6-1. The FFT sample rate is equal to 100 divided by the current horizontal time

base in figure above.

2. Source Selection

Press Source to select the desired channel (CH1, CH2, CH3 or CH4).

3. Center Frequency

Press Center and use to adjust the frequency of the frequency domain

6-6 MSO1000Z/DS1000Z User’s Guide

Chapter 6 MATH and Measurement

RIGOL waveform corresponding to the horizontal center of the screen.

4. Horizontal Scale

Press Hz/Div and use to adjust the horizontal scale of the frequency domain waveform.

5. Vertical Position

Press Offset and use to adjust the vertical position of the operation results.

6. Vertical Scale

Press Scale and use to adjust the vertical scale of the operation results.

7. Select Window Function

Spectral leakage can be considerably decreased when a window function is used.

MSO1000Z/DS1000Z provides six kinds of FFT window functions which have different characteristics and are applicable to measure different waveforms. You need to select the window function according to the waveform to be measured and its characteristics. Press Window to select the desired window function and the default is “Rectangle”.

Table 6-1 Window Function

Window Characteristics

Waveforms Suitable for

Measurement

Transient or short pulse, the signal levels before and after the

Rectangle The best frequency resolution; the poorest amplitude resolution; similar to the situation when no window is multiplied.

Blackman The best amplitude resolution; the poorest frequency resolution multiplication are basically the same;

Sine waveforms with the same amplitude and rather similar frequencies;

Wide band random noise with waveform spectrum changing relatively slower.

Single frequency signal, search for higher order harmonics.

Hanning window.

Hamming A little bit better frequency resolution than Hanning

Flattop

Better frequency resolution and poorer amplitude resolution compared to Rectangle

Sine, periodic and narrow band random noise.

Transient or short pulse, the signal levels before and after the multiplication are rather different.

Accurate measurement Signal without accurate reference and

MSO1000Z/DS1000Z User’s Guide 6-7

RIGOL

Chapter 6 MATH and Measurement

of signal. require to accurately measurement.

Triangle Better frequency resolution.

8. Set the Display Mode

Narrow band signal with stronger interference noise.

Press View to select “Half” (default) or “Full” display mode.

 Half: the source channel and the FFT operation results are displayed separately. The time domain and frequency domain signals are displayed clearly.

 Full: the source channel and the FFT operation results are displayed in the same window to view the frequency spectrum more clearly and to perform more precise measurement.

9. Set the Vertical Unit

The vertical unit can be selected as dB/dBm or Vrms. Press Unit to select the desired unit and the default is dB/dBm. dB/dBm and Vrms use logarithmic mode and linear mode to display vertical amplitude respectively. If you need to display the FFT frequency spectrum in a relatively larger dynamic range, dB/dBm is recommended.

10. Press Scale Reset to adjust the vertical scale of the operation results to the optimal value according to the current configuration.

Tips

 You can adjust the center frequency and horizontal scale at the same time using HORIZONTAL SCALE.

 Signals with DC components or deviation would cause error or deviation of the

FFT waveform components. To reduce the DC components, set the “Channel

Coupling” to “AC”.

 To reduce the random noise and aliasing frequency components of repetitive

or single pulse, set the “Acquisition Mode” of the oscilloscope to “Average”.

6-8 MSO1000Z/DS1000Z User’s Guide

Chapter 6 MATH and Measurement

RIGOL

“AND” Operation

0

0

1

1

Perform logic “AND” operation on the waveform voltage values of the specified sources point by point and display the results. In operation, when the voltage value of the source channel is greater than the threshold of the corresponding channel, it is regarded as logic “1”; otherwise logic “0”.

The results of logic AND operation of two binary bits are as follows.

Table 6-2 Logic “AND” Operation

A B

0

1

0

1

A&&B

0

0

0

1

Press MATHMathOperator to select “A&&B”:

 Press Operation to enable or disable the “AND” operation function.

 Press SourceA and SourceB to select the desired channels (CH1-CH4 or

D0-D15).

― If source A selects any channel in CH1-CH4, press Thre.A and use set the threshold of source A in logic operation. If source A selects any channel in D0-D15, Thre.A will be hidden automatically.

― If source B selects any channel in CH1-CH4, press Thre.B and use set the threshold of source B in logic operation. If source B selects any channel in D0-D15, Thre.B will be hidden automatically.

 Press Offset and use to adjust the vertical position of the operation results.

 Press Scale and use to adjust the vertical scale of the operation results.

 Press Scale Reset to adjust the vertical scale of the operation results to the optimal value according to the current configuration.

 Press Options to set the start and end points of the operation results, enable or disable waveform invert, etc.

― Press Start and use to set the start point of the operation results.

― Press End and use to set the end point of the operation results.

― Press Invert to enable or disable the inverted display function of the waveform.

― Press Sens. and use to set the sensitivity of the digital signal converted from the analog signal on the source. The setting range is from 0.00 Div to

0.96 Div.

― Press Auto Scale to enable or disable the auto scale function. When auto scale is enabled, the instrument will adjust the vertical scale of the operation results to the optimal value according to the current configuration.

― Press fx Operator, fx A and fx B to set the operator and signal sources of the inner layer operation of Fx operation (please refer to the introduction in

MSO1000Z/DS1000Z User’s Guide 6-9

RIGOL

Chapter 6 MATH and Measurement

Fx Operation”).

Note: Smooth is grayed out and disabled. It is only available for differential operation.

 You can also use HORIZONTAL POSITION and HORIZONTAL

SCALE to adjust the horizontal position and scale of the operation results.

“OR” Operation

0

0

1

1

Perform logic “OR” operation on the waveform voltage values of the specified sources point by point and display the results. In operation, when the voltage value of the source channel is greater than the threshold of the corresponding channel, it is regarded as logic “1”; otherwise logic “0”.

The results of logic OR operation of two binary bits are as follows.

Table 6-3 Logic “OR” Operation

A B

0

1

0

1

A||B

0

1

1

1

Press MATHMathOperator to select “A||B”:

 Press Operation to enable or disable the “OR” operation function.

 Press SourceA and SourceB to select the desired channels (CH1-CH4 or

D0-D15).

― If source A selects any channel in CH1-CH4, press Thre.A and use set the threshold of source A in logic operation. If source A selects any channel in D0-D15, Thre.A will be hidden automatically.

― If source B selects any channel in CH1-CH4, press Thre.B and use set the threshold of source B in logic operation. If source B selects any channel in D0-D15, Thre.B will be hidden automatically.

 Press Offset and use to adjust the vertical position of the operation results.

 Press Scale and use to adjust the vertical scale of the operation results.

 Press Scale Reset to adjust the vertical scale of the operation results to the optimal value according to the current configuration.

 Press Options to set the start and end points of the operation results, enable or disable waveform invert, etc.

― Press Start and use to set the start point of the operation results.

― Press End and use to set the end point of the operation results.

― Press Invert to enable or disable the inverted display function of the waveform.

― Press Sens. and use to set the sensitivity of the digital signal converted from the analog signal on the source. The setting range is from 0.00 Div to

6-10 MSO1000Z/DS1000Z User’s Guide

Chapter 6 MATH and Measurement

RIGOL

0.96 Div.

― Press Auto Scale to enable or disable the auto scale function. When auto scale is enabled, the instrument will adjust the vertical scale of the operation results to the optimal value according to the current configuration.

― Press fx Operator, fx A and fx B to set the operator and signal sources of the inner layer operation of Fx operation (please refer to the introduction in

Fx Operation”).

Note: Smooth is grayed out and disabled. It is only available for differential operation.

 You can also use HORIZONTAL POSITION and HORIZONTAL

SCALE to adjust the horizontal position and scale of the operation results.

“XOR” Operation

0

0

1

1

Perform logic “XOR” operation on the waveform voltage values of the specified sources point by point and display the results. In operation, when the voltage value of the source channel is greater than the threshold of the corresponding channel, it is regarded as logic “1”; otherwise logic “0”.

The results of logic XOR operation of two binary bits are as follows.

Table 6-4 Logic “XOR” Operation

A B

0

1

0

1

A^B

0

1

1

0

Press MATHMathOperator to select “A^B”:

 Press Operation to enable or disable the “XOR” operation function.

 Press SourceA and SourceB to select the desired channels (CH1-CH4 or

D0-D15).

― If source A selects any channel in CH1-CH4, press Thre.A and use set the threshold of source A in logic operation. If source A selects any channel in D0-D15, Thre.A will be hidden automatically.

― If source B selects any channel in CH1-CH4, press Thre.B and use set the threshold of source B in logic operation. If source B selects any channel in D0-D15, Thre.B will be hidden automatically.

 Press Offset and use to adjust the vertical position of the operation results.

 Press Scale and use to adjust the vertical scale of the operation results.

 Press Scale Reset to adjust the vertical scale of the operation results to the optimal value according to the current configuration.

 Press Options to set the start and end points of the operation results, enable or

MSO1000Z/DS1000Z User’s Guide 6-11

RIGOL

Chapter 6 MATH and Measurement

disable waveform invert, etc.

― Press Start and use to set the start point of the operation results.

― Press End and use to set the end point of the operation results.

― Press Invert to enable or disable the inverted display function of the waveform.

― Press Sens. and use to set the sensitivity of the digital signal converted from the analog signal on the source. The setting range is from 0.00 Div to

0.96 Div.

― Press Auto Scale to enable or disable the auto scale function. When auto scale is enabled, the instrument will adjust the vertical scale of the operation results to the optimal value according to the current configuration.

― Press fx Operator, fx A and fx B to set the operator and signal sources of the inner layer operation of Fx operation (please refer to the introduction in

Fx Operation”).

Note: Smooth is grayed out and disabled. It is only available for differential operation.

 Use HORIZONTAL POSITION and HORIZONTAL SCALE can also be used to adjust the horizontal position and scale of the operation results.

“NOT” Operation

Perform logic “NOT” operation on the waveform voltage values of the specified sources point by point and display the results. In operation, when the voltage value of the source channel is greater than the threshold of the corresponding channel, it is regarded as logic “1”; otherwise logic “0”.

The results of logic NOT operation of two binary bits are as follows.

Table 6-5 Logic “NOT” Operation

A

0

1

!A

1

0

Press MATHMathOperator to select “!A”:

 Press Operation to enable or disable the “NOT” operation function.

 Press SourceA to select the desired channels (CH1-CH4 or D0-D15).

If source A selects any channel in CH1-CH4, press Thre.A and use set the threshold of source A in logic operation. If source A selects any channel in

D0-D15, Thre.A will be hidden automatically.

 Press Offset and use to adjust the vertical position of the operation results.

 Press Scale and use to adjust the vertical scale of the operation results.

 Press Scale Reset to adjust the vertical scale of the operation results to the optimal value according to the current configuration.

6-12 MSO1000Z/DS1000Z User’s Guide

Chapter 6 MATH and Measurement

RIGOL

 Press Options to set the start and end points of the operation results, enable or disable waveform invert, etc.

― Press Start and use to set the start point of the operation results.

― Press End and use to set the end point of the operation results.

― Press Invert to enable or disable the inverted display function of the waveform.

― Press Sens. and use to set the sensitivity of the digital signal converted from the analog signal on the source. The setting range is from 0.00 Div to

0.96 Div.

― Press Auto Scale to enable or disable the auto scale function. When auto scale is enabled, the instrument will adjust the vertical scale of the operation results to the optimal value according to the current configuration.

― Press fx Operator, fx A and fx B to set the operator and signal sources of the inner layer operation of Fx operation (please refer to the introduction in

Fx Operation”).

Note: Smooth is grayed out and disabled. It is only available for differential operation.

 You can also use HORIZONTAL POSITION and HORIZONTAL

SCALE to adjust the horizontal position and scale of the operation results.

Intg

Calculate the integral of the selected source. You can use integral to measure the area under a waveform or the pulse energy.

Press MATHMathOperator to select “Intg”:

 Press Operation to enable or disable the “Intg” operation function.

 Press Source to select the desired channel (CH1, CH2, CH3, CH4 or fx (please

refer to the introduction in “Fx Operation”)).

 Press Offset and use to adjust the vertical position of the operation results.

 Press Scale and use to adjust the vertical scale of the operation results.

 Press Scale Reset to adjust the vertical scale of the operation results to the optimal value according to the current configuration.

 Press Options to set the start and end points of the operation results, enable or disable waveform invert, etc.

― Press Start and use to set the start point of the operation results.

― Press End and use to set the end point of the operation results.

― Press Invert to enable or disable the inverted display function of the waveform.

― Press Auto Scale to enable or disable the auto scale function. When auto scale is enabled, the instrument will adjust the vertical scale of the operation results to the optimal value according to the current configuration.

MSO1000Z/DS1000Z User’s Guide 6-13

RIGOL

Chapter 6 MATH and Measurement

― Press fx Operator, fx A and fx B to set the operator and signal sources of the inner layer operation of Fx operation (please refer to the introduction in

Fx Operation”).

Note: Sens. and Smooth are grayed out and disabled. Sens. is only available when a digital channel is selected as the source. Smooth is only available for differential operation.

 You can also use HORIZONTAL POSITION and HORIZONTAL

SCALE to adjust the horizontal position and scale of the operation results.

Diff

Calculate the discrete time differentiate of the selected source. For example, you can use differentiate to calculate the instantaneous slope of a waveform.

Press MATHMathOperator to select “Diff”:

 Press Operation to enable or disable the “Diff” operation function.

 Press Source to select the desired channel (CH1, CH2, CH3, CH4 or fx (please

refer to the introduction in “Fx Operation”)).

 Press Offset and use to adjust the vertical position of the operation results.

 Press Scale and use to adjust the vertical scale of the operation results.

 Press Scale Reset to adjust the vertical scale of the operation results to the optimal value according to the current configuration.

 Press Options to set the start and end points of the operation results, enable or disable waveform invert, etc.

― Press Start and use to set the start point of the operation results.

― Press End and use to set the end point of the operation results.

― Press Invert to enable or disable the inverted display function of the waveform.

― Press Smooth to set the smoothing window width of differential operation.

The setting range is from 3 to 201. The smoothing window is rectangular which can increase the smoothing for differential operation.

― Press Auto Scale to enable or disable the auto scale function. When auto scale is enabled, the instrument will adjust the vertical scale of the operation results to the optimal value according to the current configuration.

― Press fx Operator, fx A and fx B to set the operator and signal sources of the inner layer operation of Fx operation (please refer to the introduction in

Fx Operation”).

Note: Sens. is grayed out and disabled. It is only available when a digital channel is selected as source.

 You can also use HORIZONTAL POSITION and HORIZONTAL

SCALE to adjust the horizontal position and scale of the operation results.

6-14 MSO1000Z/DS1000Z User’s Guide

Chapter 6 MATH and Measurement

RIGOL

Tip

Because the differential operation is very sensitive to noise, you can set the

Acquisition Mode” to “Average”.

Sqrt

Calculate the square root of the selected source point by point and display the results.

When the operation is invalid, “NAN” is displayed at the bottom of the screen.

Press MATHMathOperator to select “Sqrt”:

 Press Operation to enable or disable the “Sqrt” operation function.

 Press Source to select the desired channel (CH1, CH2, CH3, CH4 or fx (please

refer to the introduction in “Fx Operation”)).

 Press Offset and use to adjust the vertical position of the operation results.

 Press Scale and use to adjust the vertical scale of the operation results.

 Press Scale Reset to adjust the vertical scale of the operation results to the optimal value according to the current configuration.

 Press Options to set the start and end points of the operation results, enable or disable waveform invert, etc.

― Press Start and use to set the start point of the operation results.

― Press End and use to set the end point of the operation results.

― Press Invert to enable or disable the inverted display function of the waveform.

― Press Auto Scale to enable or disable the auto scale function. When auto scale is enabled, the instrument will adjust the vertical scale of the operation results to the optimal value according to the current configuration.

― Press fx Operator, fx A and fx B to set the operator and signal sources of the inner layer operation of Fx operation (please refer to the introduction in

Fx Operation”).

Note: Sens. and Smooth are grayed out and disabled. Sens. is only available when a digital channel is selected as the source. Smooth is only available for differential operation.

 You can also use HORIZONTAL POSITION and HORIZONTAL

SCALE to adjust the horizontal position and scale of the operation results.

MSO1000Z/DS1000Z User’s Guide 6-15

RIGOL

Chapter 6 MATH and Measurement

Lg (Use 10 as the Base)

Calculate the logarithm of the selected source (use10 as the base) point by point and display the results. When the operation is invalid, “NAN” is displayed at the bottom of the screen.

Press MATHMathOperator to select “Lg”:

 Press Operation to enable or disable the “Lg” operation function.

 Press Source to select the desired channel (CH1, CH2, CH3, CH4 or fx (please

refer to the introduction in “Fx Operation”)).

 Press Offset and use to adjust the vertical position of the operation results.

 Press Scale and use to adjust the vertical scale of the operation results.

 Press Scale Reset to adjust the vertical scale of the operation results to the optimal value according to the current configuration.

 Press Options to set the start and end points of the operation results, enable or disable the inverted waveform, etc.

― Press Start and use to set the start point of the operation results.

― Press End and use to set the end point of the operation results.

― Press Invert to enable or disable the inverted display function of the waveform.

― Press Auto Scale to enable or disable the auto scale function. When auto scale is enabled, the instrument will adjust the vertical scale of the operation results to the optimal value according to the current configuration.

― Press fx Operator, fx A and fx B to set the operator and signal sources of the inner layer operation of Fx operation (please refer to the introduction in

Fx Operation”).

Note: Sens. and Smooth are grayed out and disabled. Sens. is only available when a digital channel is selected as the source. Smooth is only available for differential operation.

 You can also use HORIZONTAL POSITION and HORIZONTAL

SCALE to adjust the horizontal position and scale of the operation results.

Ln

Calculate the natural logarithm of the selected source point by point and display the results. When the operation is invalid, “NAN” is displayed at the bottom of the screen.

Press MATHMathOperator to select “Ln”:

 Press Operation to enable or disable the “Ln” operation function.

 Press Source to select the desired channel (CH1, CH2, CH3, CH4 or fx (please

refer to the introduction in “Fx Operation”)).

 Press Offset and use to adjust the vertical position of the operation results.

 Press Scale and use to adjust the vertical scale of the operation results.

6-16 MSO1000Z/DS1000Z User’s Guide

Chapter 6 MATH and Measurement

RIGOL

 Press Scale Reset to adjust the vertical scale of the operation results to the optimal value according to the current configuration.

 Press Options to set the start and end points of the operation results, enable or disable waveform invert, etc.

― Press Start and use to set the start point of the operation results.

― Press End and use to set the end point of the operation results.

― Press Invert to enable or disable the inverted display function of the waveform.

― Press Auto Scale to enable or disable the auto scale function. When auto scale is enabled, the instrument will adjust the vertical scale of the operation results to the optimal value according to the current configuration.

― Press fx Operator, fx A and fx B to set the operator and signal sources of the inner layer operation of Fx operation (please refer to the introduction in

Fx Operation”).

Note: Sens. and Smooth are grayed out and disabled. Sens. is only available when a digital channel is selected as the source. Smooth is only available for differential operation.

 You can also use HORIZONTAL POSITION and HORIZONTAL

SCALE to adjust the horizontal position and scale of the operation results.

Exp

Calculate the exponent of the selected source point by point and display the results.

Press MATHMathOperator to select “Exp”:

 Press Operation to enable or disable the “Exp” operation function.

 Press Source to select the desired channel (CH1, CH2, CH3, CH4 or fx (please

refer to the introduction in “Fx Operation”)).

 Press Offset and use to adjust the vertical position of the operation results.

 Press Scale and use to adjust the vertical scale of the operation results.

 Press Scale Reset to adjust the vertical scale of the operation results to the optimal value according to the current configuration.

 Press Options to set the start and end points of the operation results, enable or disable waveform invert, etc.

― Press Start and use to set the start point of the operation results.

― Press End and use to set the end point of the operation results.

― Press Invert to enable or disable the inverted display function of the waveform.

― Press Auto Scale to enable or disable the auto scale function. When auto scale is enabled, the instrument will adjust the vertical scale of the operation results to the optimal value according to the current configuration.

― Press fx Operator, fx A and fx B to set the operator and signal sources of the inner layer operation of Fx operation (please refer to the introduction in

MSO1000Z/DS1000Z User’s Guide 6-17

RIGOL

Chapter 6 MATH and Measurement

Fx Operation”).

Note: Sens. and Smooth are grayed out and disabled. Sens. is only available when a digital channel is selected as the source. Smooth is only available for differential operation.

 You can also use HORIZONTAL POSITION and HORIZONTAL

SCALE to adjust the horizontal position and scale of the operation results.

Abs

Calculate the absolute value of the selected source and display the results.

Press MATHMathOperator to select “Abs”:

 Press Operation to enable or disable the “Abs” operation function.

 Press Source to select the desired channel (CH1, CH2, CH3, CH4 or fx (please

refer to the introduction in “Fx Operation”)).

 Press Offset and use to adjust the vertical position of the operation results.

 Press Scale and use to adjust the vertical scale of the operation results.

 Press Scale Reset to adjust the vertical scale of the operation results to the optimal value according to the current configuration.

 Press Options to set the start and end points of the operation results, enable or disable waveform invert, etc.

― Press Start and use to set the start point of the operation results.

― Press End and use to set the end point of the operation results.

― Press Invert to enable or disable the inverted display function of the waveform.

― Press Auto Scale to enable or disable the auto scale function. When auto scale is enabled, the instrument will adjust the vertical scale of the operation results to the optimal value according to the current configuration.

― Press fx Operator, fx A and fx B to set the operator and signal sources of the inner layer operation of Fx operation (please refer to the introduction in

Fx Operation”).

Note: Sens. and Smooth are grayed out and disabled. Sens. is only available when a digital channel is selected as the source. Smooth is only available for differential operation.

 You can also use HORIZONTAL POSITION and HORIZONTAL

SCALE to adjust the horizontal position and scale of the operation results.

6-18 MSO1000Z/DS1000Z User’s Guide

Chapter 6 MATH and Measurement

RIGOL

Fx Operation

MSO1000Z/DS1000Z supports the fx operation function which can achieve more complex operations.

The procedures are as follows:

1. Split the complex operation

Users can split a more complex operation into inner and outer layers (the inner layer can only be one of the algebra operations; the outer layer can be one of the algebra or function operations) according to their needs.

2. Set the operator and signal sources of the inner layer

Press MATHMathOptionsfx Operator select “A+B”, “A-B”, “A×B” or

“A/B”. Press fx A or fx B to select the source A and source B of the inner layer operation respectively.

3. Set the operator and signal source of the outer layer

Press MATHMathOperator to select the desired operator. The outer layer supports multiple operations, including A+B, A-B, A×B, A/B, Intg, Diff, Sqrt,

Lg, Ln, Exp and Abs. Then, you can set the result (“fx”) of the inner layer operation as the signal source of the outer layer operation.

For example, Intg(CH1*CH2). The procedure is as follows:

Press MATHMathOptionsfx Operator to select “A×B”. Press fx A to select “CH1” and press fx B to select “CH2”. At this point, the inner layer operation setting is completed. Press MATHMathOperator to select “Intg”. You can set the parameters of the function operation (Intg). Press Operation to select “ON” and press Source to select “fx”.

Math Operation Label

Press MATHMathLabelDisplay to enable or disable the MATH label.

When “ON” is selected, you can add the MATH label via two modes.

To Use Preset Label

Press Preset to select ADD, SUB, MUL, DIV, FFT, AND, OR, XOR, NOT, Intg, Diff,

Sqrt, Lg, Ln, Exp or Abs.

To Edit Label Manually

Press Label Edit to open the label input interface. You can input the label

manually. For the details, please refer to the introduction in “Channel Label”.

MSO1000Z/DS1000Z User’s Guide 6-19

RIGOL

Chapter 6 MATH and Measurement

Auto Measurement

MSO1000Z/DS1000Z provides auto measurements of 33 waveform parameters and the statistics and analysis of the measurement results. What’s more, you can also use the frequency counter to realize more precise frequency measurement.

Quick Measurement after AUTO

When the oscilloscope is correctly connected and has detected valid input signal, press AUTO to enable the waveform auto setting function and open the following function menu:

Single-period: set the oscilloscope to display single-period signal automatically. Measure the “Period” and “Frequency” of the current signal within a single period and display the measurement results at the bottom of the screen.

Multi-period: set the oscilloscope to display multi-period signal automatically. Measure the “Period” and “Frequency” of the current signal within multiple periods and display the measurement results at the bottom of the screen.

Rise Time: measure the “Rise Time” of the current signal and display the measurement result at the bottom of the screen.

Fall Time: measure the “Fall Time” of the current signal and display the measurement result at the bottom of the screen.

Back: return to the menu displayed at the last user setting.

Undo: cancel the automatic settings and restore the parameters set by users lastly.

Figure 6-2 Function Menu after AUTO

Note: Waveform auto setting function requires that the frequency of sine is no lower than 41Hz; the duty cycle should be greater than 1% and the amplitude must be at least 20mVpp for square. Otherwise, the Waveform auto setting function may be invalid and the quick parameter measurement function displayed in the menu will also be unavailable.

6-20 MSO1000Z/DS1000Z User’s Guide

Chapter 6 MATH and Measurement

RIGOL

One-key Measurement of 33 Parameters

Press MENU at the left of the screen to turn on the measurement menu of the 33 parameters and then press the corresponding menu softkey to quickly realize

“One-key” measurement. The measurement result can be displayed at the bottom of the screen in two font sizes (press MeasureFont Size to select “Normal” or

“Large”.

The time and voltage parameters icons in the measurement items and the measurement results on the screen are always marked in the same color with the channel (MeasureSource) currently used.

The parameter icons and the measurement results of delay and phase are always marked in white. The colors of the numbers (1 and 2) in the icons and results are related to the measurement source currently selected. When the measurement source is an analog channel, the color of 1 or 2 is the same with that of the channel selected. When the measurement source is a digital channel, 1 or 2 is marked in green.

Note: If the measurement result is displayed as “*****”, it means that there is no signal input in the current source or the measurement result is not within the valid range (too large or too small).

MSO1000Z/DS1000Z User’s Guide 6-21

RIGOL

Chapter 6 MATH and Measurement

Time Parameters

RiseTime

+Width

FallTime

-Width

Threshold

Upper Limit

Threshold

Middle Value

Threshold

Lower Limit

Period

Figure 6-3 Time Parameters

1. Period: defined as the time between the middle threshold points of two consecutive, like-polarity edges.

2. Frequency: defined as the reciprocal of period.

3. Rise Time: the time for the signal amplitude to rise from the threshold lower limit to the threshold upper limit.

4. Fall Time: the time for the signal amplitude to fall from the threshold upper limit to the threshold lower limit.

5. + Width: the time difference between the threshold middle value of a rising edge and the threshold middle value of the next falling edge of the pulse.

6. - Width: the time difference between the threshold middle value of a falling edge and the threshold middle value of the next rising edge of the pulse.

7. + Duty: the ratio of the positive pulse width to the period.

8. - Duty: the ratio of the negative pulse width to the period.

9. tVmax: the time corresponding to the waveform maximum value (Vmax).

10. tVmin: the time corresponding to the waveform minimum value (Vmin).

Note: The default values of the threshold upper limit, threshold middle value and threshold lower limit in the figure above are 90%, 50% and 10% respectively. You can set them via MeasureSettingType  “Threshold” and for the setting

method, please refer to the introduction in “Threshold Measurement Setting”.

6-22 MSO1000Z/DS1000Z User’s Guide

Chapter 6 MATH and Measurement

Delay and Phase

RIGOL

Period

Source 1

Delay

Source 2

Figure 6-4 Delay and Phase

Source 1 and source 2, the same as source A and source B in the measurement setting menu can be any channel of CH1-CH4 or D0-D15. For the setting method,

please refer to the introduction in “Measurement Setting”.

1. Delay 12 : the time difference between the rising edges of source 1 and source 2. Negative delay indicates that the selected rising edge of source 1 occurred after that of source 2.

2. Delay 12 : the time difference between the falling edges of source 1 and source 2. Negative delay indicates that the selected falling edge of source 1 occurred after that of source 2.

3. Phase 12 : phase difference calculated according to “Delay 1→2 ” and the period of source 1, expressed in degree. The calculation formula is as shown in

equation (6-1).

4. Phase 12 : phase difference calculated according to “Delay 1→2 ” and the period of source 1, expressed in degree. The calculation formula is as shown in

equation (6-1).

The phase calculation formula:

Wherein,

Phase

Phase =

Delay

Period1

× 360

°

represents “Phase 12 ” or “Phase 12

Delay represents “Delay 12 ” or “Delay 12

Period 1 represents the period of source 1

(6-1)

MSO1000Z/DS1000Z User’s Guide 6-23

RIGOL

Chapter 6 MATH and Measurement

Voltage Parameters

Overshoot

Vmax

Vpp

Vtop

Vamp

Vupper

Vmid

Vmin

Vbase

Preshoot

Figure 6-5 Voltage Parameters

Vlower

1. Vmax: the voltage value from the highest point of the waveform to the GND.

2. Vmin: the voltage value from the lowest point of the waveform to the GND.

3. Vpp: the voltage value from the highest point to the lowest point of the waveform.

4. Vtop: the voltage value from the flat top of the waveform to the GND.

5. Vbase: the voltage value from the flat base of the waveform to the GND.

6. Vamp: the voltage value from the top of the waveform to the base of the waveform.

7. Vupper: the actual voltage value corresponding to the threshold maximum value.

8. Vmid: the actual voltage value corresponding to the threshold middle value.

9. Vlower: the actual voltage value corresponding to the threshold minimum value.

10. Vavg: the arithmetic average value on the whole waveform or on the gating area. The calculation formula is as follows:

Wherein, i

Average =

n x i

(6-2) x is the i th point being measured, n is the number of points being measured.

11. Vrms: the root mean square value on the whole waveform or the gating area.

The calculation formula is as follows.

12. Per.Vrms: the root mean square value within a period. The calculation formula is as follows.

RMS = n

i=1 x i

2 n

(6-1)

6-24 MSO1000Z/DS1000Z User’s Guide

Chapter 6 MATH and Measurement

RIGOL

Wherein, x i is the i th point being measured, n is the number of points being measured.

13. Overshoot: the ratio of the difference of the maximum value and top value of the waveform to the amplitude value.

14. Preshoot: the ratio of the difference of the minimum value and base value of the waveform to the amplitude value.

15. Variance: the average of the sum of the squares for the difference between the amplitude value of each waveform point and the waveform average value on the whole waveform or on the gating area. The variance reflects the fluctuation degree of the waveform. The calculation formula is as follows:

Variance = i n

=1

2 ( )

(6-3) n i th poin, Average is the waveform Wherein, average value, n is the number of points being measured.

Other Parameters

1. +Rate: divide the difference of the upper value and lower value on the rising edge by the corresponding time.

2. -Rate: divide the difference of the lower value and upper value on the falling edge by the corresponding time.

3. Area: the area of the whole waveform within the screen and the unit is

Vs. The area meadured above the zero reference (namely the vertical offset) is positive and the area measured below the zero reference is negative. The area measured is the algebraic sum of the area of the whole waveform within the screen.

4. Period Area: the area of the first period of waveform on the screen and the unit is Vs. The area above the zero reference (namely the vertical offset) is positive and the area below the zero reference is negative. The area measured is the algeraic sum of the area of the whole period waveform.

Note: When the waveform on the screen is less than a period, the period area measured is 0.

MSO1000Z/DS1000Z User’s Guide 6-25

RIGOL

Chapter 6 MATH and Measurement

Frequency Counter Measurement

The hardware frequency counter supplied with MSO1000Z/DS1000Z oscilloscope can make more precise measurement of the input signal frequency.

Press MeasureCounter to select CH1-CH4 or D0-D15 as the measurement source. The measurement result is displayed at the upper right corner of the screen and you can identify the current measurement source according to the color of the icon and channel number. The following figure is the result of the frequency measurement of the input signal of CH1.

Select “OFF” to disable the frequency counter measurement function.

Measurement Setting

Source Selection

Press MeasureSource to select the desired channel for measurement

(CH1-CH4, MATH or D0-D15). The color of the parameter icons under MENU at the left of the screen will change with the source selected.

Delay Measurement Setting

Specify the source1 and source2 in the measurement items “Delay 12 ” and

Delay 12 ”.

Press MeasureSettingType  “Delay” and then press SourceA and

SourceB to set the two channel sources (CH1-CH4 or D0-D15) of delay measurement respectively.

Phase Measurement Setting

Specify the source1 and source2 in the measurement items “Phase 12 ” and “Phase 12 ”.

Press MeasureSettingType  “Phase” and then press SourceA and

SourceB to set the two channel sources (CH1-CH4 or D0-D15) of phase measurement respectively.

Threshold Measurement Setting

Specify the upper limit, lower limit and middle value of the auto measurement of the analog channel. Measurements of all the time, delay and phase parameters will be influenced by this setting.

Press MeasureSettingType  “Threshold” and then:

6-26 MSO1000Z/DS1000Z User’s Guide

Chapter 6 MATH and Measurement

RIGOL

 Press Max and use to set the upper limit of the measurement.

Reducing the upper limit to the current middle value will automatically reduce the middle value and lower limit to keep them lower than the upper limit. The default is 90% and the range available is from 7% to 95%.

 Press Mid and use to set the middle value of measurement. The middle value is limited by the settings of the upper limit and lower limit. The default is 50% and the range available is from 6% to 94%.

 Press Min and use to set the lower limit of the measurement.

Increasing the lower limit to the current middle value will automatically increase the middle value and upper limit to keep them higher than the lower limit. The default is 10% and the range available is from 5% to 93%.

To Clear the Measurement

If you have currently enabled one or more items in the 33 measurement parameters, you can “Delete” or “Recover” the last five parameters. The measurement results of measurement items 1 to 5 are displayed from left to right at the bottom of the screen.

Note: The last five parameters are determined according to the order in which you turned them on and they will not change as you delete one or more measurement items.

Press MeasureClearItem n (n=1~5) to “Delete” or “Recover” the specified measurement item.

 When the state of Item n (n=1~5) is “Delete”, you can press Item n (n=1~5) to delete the corresponding item. At this point, the Nth measurement result at the bottom of the screen is grayed out. If a new measurement item is opened, all items will move one-item left.

 When the state of Item n (n=1~5) is “Recover”, you can press Item n (n=1~5) to recover the corresponding item. At this point, the Nth measurement result

(from the left) at the bottom of the screen is illuminated automatically.

Press MeasureClearAll Items to “Delete” or “Recover” the last five parameters at the same time. When all measurement items are deleted, all measurement results will be grayed out transiently and then be hidden. At this point, you can press All Items to recover the last five items deleted at the same time.

Note: You can press and hold Measure to clear all the measurement items currently turned on. At this point, you can recover the last five parameters turned on by pressing MeasureClearItem n (n=1~5) or MeasureClearAll

Items.

MSO1000Z/DS1000Z User’s Guide 6-27

RIGOL

Chapter 6 MATH and Measurement

All Measurement

All measurement could measure all the time and voltage parameters (each measurement source has 28 items) of the current measurement source and display the results on the screen. Press MeasureMeasure All to enable or disable the all measurement function. Press All Measure Source and use to select the channel(s) to be measured (CH1-CH4 and MATH).

 When all measurement is enabled, “One-key” measurement is also valid.

 “To Clear the Measurement” will not clear the results of all measurement.

All Measurement One-key Measurement

Figure 6-6 All Measurement

6-28 MSO1000Z/DS1000Z User’s Guide

Chapter 6 MATH and Measurement

RIGOL

Statistic Function

Make statistic and display the current, average, minimum (or standard deviation) and maximum (or count) values of at most 5 measurement items that are turned on last.

Press MeasureStatistic to turn the statistic function on or off. When the statistic function is enabled, press Stat.Sel. to select “Extremum” or “Difference” measurement. When “Extremum” is selected, minimum and maximum values are displayed. When “Difference” is selected, standard deviation and count values are displayed. Press MeasureReset Stat. to clear the history data and make statistic again.

Extremum Measurement

Measurement History

Difference Measurement

The measurement history function can display the history measurement data of the measurement items currently turned on in graph or table form.

Press MeasureHistory Disp Type to select “Graph” or “Table”. Press

MeasureHistory DispHistory to turn on or off the measurement history function. In “Graph” display mode, the oscilloscope draws the measurement value curve in real time with the horizontal axis denoting time and the vertical axis denoting the measurement value. In “Table” display mode, the oscilloscope displays the 8 latest measurement values in real time. As shown in the figures below.

MSO1000Z/DS1000Z User’s Guide

Graph

6-29

RIGOL

Chapter 6 MATH and Measurement

Table

Measurement Result Display Type

The measurement results can be displayed at the bottom of the screen in two font sizes. Press MeasureFont Size to select “Normal” or “Large”. The display effects are as shown in the figures below.

Normal (Statistic Function Disabled)

Normal (Statistic Function Enabled)

Large (Statistic Function Disabled)

Large (Statistic Function Enabled)

Cursor Measurement

Cursor measurement can measure the X axis values (usually Time) and Y axis values

(usually Voltage) of the selected waveform. Before making cursor measurement,

connect the signal to the oscilloscope and acquire stable display. All the “Auto

Measurement” parameters can be measured through cursor measurement.

Press CursorMode at the front panel and use to select the desired cursor mode (the default is “OFF”) and then press down the knob. The modes available are

“Manual”, “Track” and “Auto”. When the Time Base Mode is set to XY, you can

select the XY cursor measurement mode.

6-30 MSO1000Z/DS1000Z User’s Guide

Chapter 6 MATH and Measurement

RIGOL

Manual Mode

In this mode, a pair of cursors will appear. You can adjust the cursors manually to measure the X (or Y), X increment (or Y increment) between cursors and the reciprocal of

X increment on the waveform of the selected source (CH1-CH4, LA or MATH). When the measurement source is set to LA, the logic level value of the digital channel currently turned on will be displayed in hexadecimal forms (high level is 1 and low level is 0).

Press CursorMode  “Manual” to turn the manual cursor function on. To easily read data, the display form of measurement results on the upper left corner of the screen changes with the measurement source and unit selected.

When the measurement source is set to CH1-CH4 or MATH:

 When the measurement source is set to FFT (press MATHMath

Operator  “FFT” and press Operation to select “ON”; press Cursor

Source  “MATH”), if the horizontal unit selects “s”, the measurement form is as shown in figure (a); if other units are selected, the measurement forms are as shown in figure (b).

 When the measurement source is non-FFT, if the horizontal unit selects “s”, the measurement form is as shown in figure (c); if the horizontal unit selects “Hz”, the measurement form is as shown in figure (d); if other units are selected, the measurement forms are as shown in figure (e).

Figure (a) Figure (b)

Figure (c) Figure (d) Figure (e)

 AX: the X value at cursor A. X value takes the trigger position as reference. Its unit is “s” or “Hz” (when measuring FFT waveform).

 AY: the Y value at cursor A. Y value takes the channel GND of CH1 as reference.

Its unit is the same as that of the current signal source.

 BX: the X value at cursor B. X value takes the trigger position as reference. Its unit is “s” or “Hz” (when measuring FFT waveform).

 BY: the Y value at cursor B. Y value takes the channel GND of CH1 as reference.

MSO1000Z/DS1000Z User’s Guide 6-31

RIGOL

Chapter 6 MATH and Measurement

Its unit is the same as that of the current signal source.

 BX-AX: the horizontal difference between cursors A and B.

 BY-AY: the vertical difference between cursors A and B.

 |dX|: when the horizontal unit selects “s”, |dx| represents the time difference between cursors A and B; when the horizontal unit selects “Hz”, |dx| represents the frequency difference between cursors A and B;

 1/|dX|: the frequency difference between cursors A and B.

When the measurement source is set to LA, the measurement results will be displayed at the upper-left corner of the screen in the following form.

 AX: the X value at cursor A. X value takes the trigger position as reference.

 D15-D0: display the logic level values at cursor A (D15-D0 from left to right) in hexadecimal forms. If the digital channel is currently turned off, it will be denoted by “*”.

 BX: the X value at cursor B. X value takes the trigger position as reference.

 D15-D0: display the logic level values at cursor B (D15-D0 from left to right) in hexadecimal forms. If the digital channel is currently turned off, it will be denoted by “*”.

If needed, please refer to the following steps to modify the parameters of manual cursor measurement.

1. Select Cursor Type

Press Select to select “ ” or “ ”.

: the X cursors are a pair of vertical solid (cursor A)/dotted (cursor B) lines and are usually used to measure time parameters.

: the Y cursors are a pair of horizontal solid (cursor A)/dotted

(cursor B) lines and are usually used to measure voltage parameters.

Note: When the current measurement source is LA, the cursor type is

“ ” and cannot be set.

2. Select Measurement Source

Press Source to select the waveform of the analog channels (CH1-CH4), math operation results (MATH) or LA waveform for measurement.

Note: Only channel currently enabled can be selected.

3. Adjust Cursor Position

 Adjust cursorA: press CursorA and use to adjust the position of cursorA. The measurement result will change accordingly during the

6-32 MSO1000Z/DS1000Z User’s Guide

Chapter 6 MATH and Measurement

RIGOL adjustment and the adjustable range is limited within the screen.

 Adjust cursorB: press CursorB and use to adjust the position of cursorB. The measurement result will change accordingly during the adjustment and the adjustable range is limited within the screen.

 Adjust cursorA and cursorB at the same time: press CursorAB and use to adjust the positions of cursorA and cursorB at the same time. The measurement result will change accordingly during the adjustment and the adjustable range is limited within the screen.

Note: You can also press down knob continuously to switch the current cursor.

4. Select X (Y) Axis Unit

Press Units to set the horizontal unit and vertical unit of the cursor measurement.

Press Hori.Unit to select “s”, “Hz”, “Degree” or “Percent”.

s: when this unit is selected, in the measurement results, AX, BX and BX-AX are in “s” and 1/|dX| is in “Hz”.

Hz: when this unit is selected, in the measurement results, AX, BX and |dX| are in “Hz” and 1/|dX| is in “s”.

Degree: when this unit is selected, in the measurement results, AX, BX and

BX-AX are in “°”. At this point, AX, BX and BX-AX will change to “0°”, “360°” and “360°” respectively when you press Set Range, no matter where cursors A and B are currently located. At the same time, the two cursor lines

(unmovable) appear on the screen as the reference positions.

Percent: when this unit is selected, in the measurement results, AX, BX and BX-AX are in “%”. At this point, AX, BX and BX-AX will change to “0%”,

“100%” and “100%” respectively when you press Set Range, no matter where cursors A and B are currently located. At the same time, the two cursor lines (unmovable) appear on the screen as the reference positions.

Press Vert. Unit to select “Source” or “Percent”.

 Source: when this unit is selected, in the measurement results, the units of

AY, BY and BY-AY will be automatically set to the unit of the current source.

 Percent: when this unit is selected, in the measurement results, AY, BY and

BY-AY are in “%”. At this point, AY, BY and BY-AY will change to “0%”,

“100%” and “100%” respectively when you press Set Range, no matter where cursors A and B are currently located. At the same time, the two cursor lines (unmovable) appear on the screen as the reference positions.

5. Select Screen Region

 When delayed sweep is enabled (pressing the HORIZONTAL SCALE can enable Zoom), the screen is divided into two regions, Main and Zoom.

Press CursorRegion to select “Main” or “Zoom” and measure the corresponding value. When “Main” is selected, the cursor values are displayed in the zoom region. When “Zoom” is selected, the cursors values are displayed in the main region.

MSO1000Z/DS1000Z User’s Guide 6-33

RIGOL

Chapter 6 MATH and Measurement

Note:

― When the cursor measurement source is LA and the main time base is measured (namely the “Main” region is selected), the measurement results are as shown in the figure below. There are no measurement values in digital channels and the measurement values are replaced by

“*”.

― When the current measurement source is MATH, you can only measure the zoom region using the cursor. At this point, “Zoom” is automatically selected and Region is grayed out.

 When the XY time base mode is selected, the measurement values of the manual cursor are displayed on the lower-half screen.

6. Measurement Example

Use manual cursor measurement to measure the period (BX-AX=1 ms) of a square waveform and the result is equal to the period from auto measurement, as shown in the figure below.

Cursor

Measurement

Auto

Measurement

Figure 6-7 Manual Cursor Measurement Example

6-34 MSO1000Z/DS1000Z User’s Guide

Chapter 6 MATH and Measurement

RIGOL

Track Mode

In this mode, one or two pairs of cursors will appear. You can adjust the two pairs of cursors (cursor A and cursor B) to measure the X and Y values on two different sources respectively. The points being measured on cursors A and B are marked by

and respectively. When the cursors are moved horizontally, the marks will position on the waveform automatically. When the waveform is expanded or compressed horizontally, the marks will track the points being marked at the last adjustment of the cursors.

Press CursorMode  “Track” to turn on the cursor track function and the measurement results will be displayed at the upper left corner of the screen in the following mode.

 AX: the X value at cursor A. The X value takes the trigger position as reference and “s” or “Hz” (when measuring FFT waveform) as its unit.

 AY: the Y value at cursor A. The Y value takes the channel GND as reference and use the same unit as the current source.

 BX: the X value at cursor B. The X value takes the trigger position as reference and “s” or “Hz” (when measuring FFT waveform) as its unit.

 BY: the Y value at cursor B. The Y value takes the channel GND as reference and use the same unit as the current source.

 BX-AX: the horizontal difference between cursors A and B.

 BY-AY: the vertical difference between cursors A and B.

 1/|dX|: the reciprocal of the horizontal difference between cursors A and B.

If needed, please refer to the following steps to modify the track cursor measurement parameters.

1. Select Measurement Source

Press CursorA to select the waveform of analog channels (CH1-CH4) or math operation results (MATH) as the measurement source of cursor A (only channels enabled are available). You can also select “None”, namely do not use cursor A.

Press CursorB to select the waveform of analog channels (CH1-CH4) or math operation results (MATH) as the measurement source of cursor B (only channels enabled are available). You can also select “None”, namely do not use cursor B.

2. Adjust Cursor Position

 Adjust cursor A: press CursorA and use to adjust the position of cursor

MSO1000Z/DS1000Z User’s Guide 6-35

RIGOL

Chapter 6 MATH and Measurement

A. During the adjustment, the measurement result will change accordingly.

The adjustable range is limited within the screen.

 Adjust cursor B: press CursorB and use to adjust the position of cursor

B. During the adjustment, the measurement result will change accordingly.

The adjustable range is limited within the screen.

 Adjust cursor A and B at the same time: press CursorAB and use to adjust the position of cursor A and B at the same time. During the adjustment, the measurement results will change accordingly. The adjustable range is limited within the screen.

Note:

― Under the current menu, you can rotate to switch the current cursor.

― In track mode, the cursor will track the marked point (namely jumps up and down with the transient change of the waveform). Thus, the Y value might change even though you do not adjust the cursor.

6-36 MSO1000Z/DS1000Z User’s Guide

Chapter 6 MATH and Measurement

RIGOL

Auto Mode

In this mode, one or more cursors will appear. You can use auto cursor measurement to measure any of the 33 waveform parameters. Before using this mode, you need to at least enable one auto measurement parameter and the number of cursors will change with the measurement parameter enabled.

Press CursorMode  “Auto” and the number of cursors displayed on the screen is determined by the measurement parameter enabled (different measurement parameter needs different number of cursors).

Note: No cursor will be displayed if no auto cursor measurement parameter is enabled or the measurement source has no input. When the waveform is expanded or compressed horizontally, the cursor will move accordingly.

If multiple measurement parameters are turned on later, you can use Auto Item to switch among at most five measurement parameters that are turned on last. If

“NONE” is selected, cursor measurement will not be used.

The following figure shows the auto measurement example of sine signal frequency in CH2.

Figure 6-8 Auto Cursor Measurement

MSO1000Z/DS1000Z User’s Guide 6-1

RIGOL

Chapter 6 MATH and Measurement

XY Mode

XY mode is only available when the horizontal time base mode is “XY”. In this mode, two pairs of cursors will appear. You can adjust the cursor positions and the instrument will automatically calculate the chord length and curvature of the ellipses corresponding to the crossing points of the two pairs of cursor lines.

Press CursorMode  “XY” to enable the XY mode cursor measurement function.

The measurement results are displayed at the upper left corner of the screen in the following mode.

 AX: the X value at cursor AX.

 AY: the Y value at cursor AY.

 BX: the X value at cursor BX.

 BY: the Y value at cursor BY.

 BX-AX: the horizontal difference between cursors BX and

AX.

 BY-AY: the vertical difference between cursors BY and AY.

 dX*dY: the product of the horizontal difference between cursors BX and AX times the vertical difference between cursors BY and AY.

 dX/dY: the quotient of the horizontal difference between cursors BX and AX divided by the vertical difference between cursors BY and AY.

 absAA: the chord length of the crossing point of cursors

AX and AY relative to the center point. The center point is defined as the crossing point of the horizontal position

(X-axis) and the vertical position (Y-axis) of two channel labels on the screen as

shown in Figure 6-9.

 absAB: the chord length of the crossing point of cursors AX and BY relative to the center point.

 absBA: the chord length of the crossing point of cursors BX and AY relative to the center point.

 absBB: the chord length of the crossing point of cursors BX and BY relative to the center point.

 argAA: the angle between X-axis and absAA. The range is from -180° to +180°.

 argAB: the angle between X-axis and absAB. The range is from -180° to +180°.

 argBA: the angle between X-axis and absBA. The range is from -180° to +180°.

 argBB: the angle between X-axis and absBB. The range is from -180° to +180°.

6-2 MSO1000Z/DS1000Z User’s Guide

Chapter 6 MATH and Measurement

RIGOL

AY

Center

Point absAA argAA

X-axis

BY

BX AX

Y-axis

Figure 6-9 XY Cursor Measurement Definition

1. Select Values

Press Values and use to turn the value(s) to be measured on or off. The measurement items opened are displayed at the upper left corner of the screen.

 Open the BX-AX and BY-AY measurement items.

 Open the dX*dY measurement item.

 Open the dX/dY measurement item.

 Open the dY/dX measurement item.

 Open the absAA and argAA measurement items at the same time.

 Open the absAB and argAB measurement items at the same time.

 Open the absBA and argBA measurement items at the same time.

 Open the absBB and argBB measurement items at the same time.

2. Adjust Cursor Position

 Adjust cursor AX: press AX and use to adjust the position of cursor AX.

 Adjust cursor BX: press BX and use to adjust the position of cursor BX.

 Adjust cursor AY: press AY and use to adjust the position of cursor AY.

 Adjust cursor BY: press BY and use to adjust the position of cursor BY.

 Adjust cursors AX and BX at the same time: press ABX and use to adjust the position of cursors AX and BX at the same time.

 Adjust cursors AY and BY at the same time: press ABY and use to adjust the position of cursors AY and BY at the same time. During the adjustment, the measurement results will change accordingly. The adjustable range is limited within the screen.

Note: You can also press down knob continuously to switch the current cursor.

MSO1000Z/DS1000Z User’s Guide 6-3

RIGOL

Chapter 6 MATH and Measurement

3. Display Lissajous Legend

This oscilloscope provides the lissajous schematic diagram under different frequencies and phases.

Phase difference between two signals

Frequency ratio between two signals

Figure 6-10 Lissajous Schematic Diagram

6-4 MSO1000Z/DS1000Z User’s Guide

Chapter 7 Digital Channel

RIGOL

Chapter 7 Digital Channel

MSO1000Z series mixed signal digital oscilloscope provides 4 analog channels as well as 16 digital channels. For the digital channels, the oscilloscope compares the voltages acquired in each sample with the preset logic threshold. If the voltage of the sample point is greater than the logic threshold, it will be stored as logic 1; otherwise, it will be stored as logic 0. The oscilloscope displays the logic 1 and logic 0 in figure form for users to easily detect and analyze the errors in circuit design (hardware design and software design). This chapter introduces how to use the digital channels of MSO1000Z mixed signal digital oscilloscope.

Before using the digital channels, connect the oscilloscope and the device under test using RPL1116 logic probe provided in the accessories. For the using method of the logic probe, refer to the

RPL1116 Logic Probe User’s Guide

The contents of this chapter:

.

 To Select the Digital Channel

 To Turn on/off the Digital Channel

 Group Set

 To Set the Waveform Display Size

 Reorder Setting

 Auto View

 To Set the Threshold

 To Set the Label

 Probe Calibration

 Digital Channel Delay Calibration

MSO1000Z/DS1000Z User’s Guide 7-1

RIGOL

Chapter 7 Digital Channel

To Select the Digital Channel

Press LA to open the logic analysis control menu. You can select any of the digital channels or user-defined digital channel groups. Press Current to open the channel/group dropdown menu and rotate the multifunction knob to select the desire channel or channel group. You can also press Current or VERTICAL

POSITION continuously to switch among the options in order.

 D0-D15: select any channel of D0-D15. The channel label and waveform corresponding to the channel selected are displayed in red.

 Group1-Group4: select any of the user-defined groups (Group1-Group4). The channel labels and waveforms of all the channels of the channel group selected are displayed in red.

 None: do not select any channel/group.

Note: Only digital channels or user-defined groups that are turned on can be selected.

For how to turn on the digital channel/group, please refer to the introduction in “To

Turn on/off the Digital Channel”.

For how to define channel group, please refer to the introduction in “Group Set”.

To Turn on/off the Digital Channel

MSO1000Z provides two methods for turning on or off the digital channels.

Method One:

Press D7-D0 to turn on or off channels D7-D0 at the same time.

Press D15-D8 to turn on or off channels D15-D8 at the same time.

Method Two:

Press ON/OFF to enter the digital channel/group “ON/OFF” setting menu.

 Press SelectCH to turn on channels D0-D15 selection list. Rotate to select any channel and press to turn on or off the channel. You can also press

SelectCH continuously to turn on or off the selected channel. Channels turned on are marked with and channels that are not turned on are marked with .

You can turn on or off any channel via the channel selection menu.

 Press Group, rotate to select any channel group and press down to turn on or off all the channels in the group at the same time. You can also press

Group continuously to turn on or off the selected group. The channel groups turned on are marked with and the channel groups that are not turned on are marked with .

7-2 MSO1000Z/DS1000Z User’s Guide

Chapter 7 Digital Channel

RIGOL

Note:

― When you have turned off channels D7-D0 at the same time using D7-D0, you can still turn on or off any of the channels via the channel selection menu. When you have turned on channels D7-D0 at the same time using

D7-D0, the channels in the list switch to ON status automatically. The situation of channels D15-D8 is similar.

― Only user-defined digital channel groups can be selected.

For how to define channel groups, please refer to the introduction in “Group

Set”.

Group Set

Press Group Set to enter the user-defined group setting menu. You can perform group and ungroup operations on the 16 digital channels.

Group:

The group operations of Group1-Group4 are the same and Group1 is taken as an example for illustration.

Press Group1 to turn on channels D0-D15 selection list (the status label is located at the left side of each channel), rotate to select the channel to be added into Group1 and press down or Group1 to add the channel selected to Group1. Channels added into Group1 are marked with and channels that are not added into Group1 is marked with .

Use the same method to group the other channels. Each channel can only be added into a single group and channels that are already added into other group cannot be selected (its status label is gray). You can group the 16 digital channels (D0-D15) via the group setting.

Ungroup:

Press UnGroup to open the channel/group list and rotate to select the desired channel/group. If any of channels D0-D15 is selected, pressing down will cancel the group operation of the channel; if any of Group1-Group4 is selected, pressing down will cancel the group operations of all the channels in the group.

Note:

― You can only perform the ungroup operation on the grouped digital channels or digital channel groups.

― If none group is set, UnGroup will be grayed out and disabled.

MSO1000Z/DS1000Z User’s Guide 7-3

RIGOL

Chapter 7 Digital Channel

To Set the Waveform Display Size

Press WaveSize to open the waveform size selection list. You can use or press

WaveSize to set the waveform display size of the channel currently turned on to S

(small) and L (large).

Note:

― L (large) can only be used when the number of channels currently turned on is no more than 8.

― You can also directly rotate VERTICAL SCALE to set the waveform display size. Turn clockwise to set it to L; turn counterclockwise to set it to S.

Reorder Setting

Press ReOrder to select the waveform order mode on the screen of the channels currently selected. You can select “D0-D15” or “D15-D0” and the default is “D0-D15”.

 D0-D15: the waveforms on the screen are D0-D15 from top to bottom.

 D15-D0: the waveforms on the screen are D15-D0 from top to bottom.

Auto View

Press Auto View and the analog channel waveform and digital channel waveform will be displayed respectively on the upper and lower half screen respectively. By default, the digital channel waveforms are displayed on the upper half screen and the analog channel waveforms are displayed on the lower half screen.

Note: When the number of channels currently turned on is more than 8 or the waveform display size is L, Auto View is grayed out and disabled.

To Set the Threshold

Press Threshold to enter the threshold setting menu. When the voltage of the input signal is greater than the threshold currently set, it is treated as logic 1; otherwise, it is treated as logic 0.

The threshold levels of channels D7-D0 and channels D15-D8 can be set separately according to the need. You can select a preset value or user-defined threshold. The range of the user-defined threshold is from -15.0V to +15.0V.

7-4 MSO1000Z/DS1000Z User’s Guide

Chapter 7 Digital Channel

RIGOL

Press Low Type to open the threshold selection list to select the threshold of D7-D0.

The selectable preset values include TTL (1.40 V), CMOS (2.50 V), ECL (-1.30 V),

PECL (3.70 V), LVDS (1.20 V) and 0V. You can also press D7-D0 and use to set the desired threshold.

Press High Type to open the threshold selection list to select the threshold of

D15-D8. The selectable preset values include TTL (1.40 V), CMOS (2.50 V), ECL

(-1.30 V), PECL (3.70 V), LVDS (1.20 V) and 0V. You can also press D15-D8 and use

to set the desired threshold.

To Set the Label

You can set another label for each digital channel using the following two methods to easily distinguish the digital buses.

Use preset labels

Press Select CH to select the digital channel (any channel of D0-D15) that needs to set a label. Press Preset to select ACK, AD0, ADDR, BIT, CAS, CLK, CS,

DATA, HALT, IN, INT, LOAD, MISO, MOSI, NMI, OUT, PAS, PIN, RDY, RST, RX, TX or WR.

Input a label manually

Press Label Edit to enter the label input interface to input a label manually. For

the input method, refer to the introduction in “ Channel Label”.

Press Clear to delete the label of the current channel.

Probe Calibration

MSO1000Z oscilloscope can calibrate the digital probe to reduce the measurement error of the digital signal. Connect RPL1116 digital probe to the oscilloscope and make sure that no signal is connected to the digital probe. Press Probe-CalStart to start calibrating the zero point of the digital probe. Press Exit to exit the calibration.

Note: Before calibrating the probe, please make sure that there is no signal input into the connected digital probe.

MSO1000Z/DS1000Z User’s Guide 7-5

RIGOL

Chapter 7 Digital Channel

Digital Channel Delay Calibration

When using the oscilloscope for actual measurement, the transmission delay of the probe cable may cause relatively larger error (zero offset). Zero offset is defined as the offset of the crossing point of the waveform and threshold level line relative to the trigger position. For MSO1000Z, users can set a delay time to calibrate the zero offset of the corresponding channel.

Press Delay-Cal and rotate to set the delay calibration time. The range is from

-100 ns to 100 ns.

Note: The delay calibration setting is related to the instrument model and the current horizontal time base. The larger the horizontal time base is, the larger the setting step will be. Take MSO1104Z as an example; the steps under different horizontal time bases are as shown in the table below.

Table 7-1 Relationship between the delay calibration step and horizontal time base

Horizontal Time Base

5ns

10ns

20ns

50ns

100ns

200ns

500ns

Delay Calibration Time Step

100ps

200ps

400ps

1ns

2ns

4ns

10ns

1μs to 10μs 20ns

Note: When the horizontal time base is equal to or greater than 10μs, the delay calibration time cannot be adjusted.

7-6 MSO1000Z/DS1000Z User’s Guide

Chapter 8 Protocol Decoding

RIGOL

Chapter 8 Protocol Decoding

Protocol analysis can be used by users to discover errors, debug hardware and accelerate development easily, so as to guarantee quick and high-quality accomplishment of projects. Protocol decoding is the basis of protocol analysis. Only protocol analyses with correct protocol decoding are acceptable and only correct protocol decoding can provide more error information. MSO1000Z/DS1000Z provides two bus decoding modules (Decode 1 and Decode 2) to make common protocol decoding (including Parallel (standard), RS232 (option), I2C (option) and SPI

(option)) for the input signals of the analog channels (CH1-CH4) and digital channels

(D0-D15). As the decoding functions and setting methods of Decode 1 and Decode 2 are the same, this chapter only takes Decode1 for illustration.

To get the decoding option information, please refer to the introduction in

Appendix A: Accessories and Options”. When you have ordered the decoding

option, please refer to the introduction in “Option Management” to activate

corresponding option.

The contents of this chapter:

 Parallel Decoding

 RS232 Decoding (Option)

 I2C Decoding (Option)

 SPI Decoding (Option)

MSO1000Z/DS1000Z User’s Guide 8-1

RIGOL

Chapter 8 Protocol Decoding

Parallel Decoding

Parallel bus consists of clock line and data line. As shown in the figure below, CLK is the clock line, while Bit0 and Bit1 are the 0 bit and 1st bit on the data line respectively.

The oscilloscope will sample the channel data on the rising edge, falling edge or the rising/falling edge of the clock and judge each data point (logic “1” or logic “0”) according to the preset threshold level.

Figure 8-1 Parallel Decoding

Press MATHDecode1Decoder to select “Parallel” and open the parallel decoding function menu.

1. Press Decode to turn on or off the decoding function.

2. Clock Line Setting (CLK)

 Press CLK to select any channel (CH1-CH4 or D0-D15) as the clock channel.

If “OFF” is selected, no clock channel is set.

 Press Edge to set the oscilloscope to sample the channel data on the rising edge ( ), falling edge ( ) or rising/falling edge ( ) of the clock. If no clock channel is selected, the instrument will sample when the channel data jumps during the decoding.

3. Digital Bus

Press BUS to select the digital bus for parallel decoding. This setting will automatically modify the settings of Width, Bit X and CH, as shown in the table below.

BUS

D0-D7

Width BitX CH Note

D7-D0 8

D15-D8 8

D15-D0 16

8

D8-D15 8

D0-D15 16

0

0

0

0

0

0

D0

D8

D0

D7

D15

D15

Bit0 to Bit7 are set to D0 to D7 respectively.

Bit0 to Bit7 are set to D8 to D15 respectively.

Bit0 to Bit15 are set to D0 to D15 respectively.

Bit0 to Bit7 are set to D7 to D0 respectively.

Bit0 to Bit7 are set to D15 to D8 respectively.

Bit0 to Bit15 are set to D15 to D0 respectively.

8-2 MSO1000Z/DS1000Z User’s Guide

Chapter 8 Protocol Decoding

RIGOL

In additional, you can also modify the settings of Width, Bit X and CH manually.

Note: This function is only available for MSO1000Z-S and MSO1000Z model oscilloscope.

4. Data Line Setting

Set the bus width

Press Width to set the data width of the parallel bus namely the number of bits per frame. The default is 8 and the range is from 1 to 16.

Specify data channel for each bit.

Press Bit X to select the bit that needs to specify a channel. Press CH to specify a channel source from CH1-CH4 or D0-D15.

5. Analog Channel Threshold Setting

To judge logic “1” and logic “0” of the buses, you need to set a threshold for each analog channel (CH1-CH4). When the signal amplitude is greater than the preset value, it is considered as “1”; otherwise “0”.

Threshold

Level

1

0 0

Press MATHDecode OptionsAuto Thre. to turn on or off the auto threshold function. When auto threshold is turned on, the middle value of the channel waveform is defined as the digital threshold level. When auto threshold is turned off, you can press Thre.Set to enter the threshold setting menu. Press

CH1, CH2, CH3 and CH4 respectively and use to set the threshold of each channel. Press 50% and set the current threshold to 50% of the current waveform trace manually.

6. Display-related Setting

 Press Format to set the display format of the bus to HEX, DEC, BIN, ASC or

LINE.

Note: In LINE format, the actual values of the bus are displayed in binary form and the order is consistent with the bus transmission order. This format

 Press Pos and use to adjust the vertical display position of the bus. is only valid for the serial bus which includes LSB and MSB endian. If MSB endian is selected, LINE format is the same as binary.

MSO1000Z/DS1000Z User’s Guide 8-3

RIGOL

Chapter 8 Protocol Decoding

Tip

Press MATHDecode OptionsASC List, characters, numbers and ASCII table of the common control characters will be displayed on the screen. You can instantly query the corresponding ASCII value.

7. Noise Rejection

Press NRJ to enable or disable the noise rejection function. Noise rejection can remove the data without enough duration on the bus to eliminate the burst glitch of the actual circuit. When the noise rejection is enabled, press NRJ.Time to set the desired rejection time and the range is from 0.00 s to 100 ms.

8. Clock Compensation

Press CLK Tune to set the compensation time. You can fine-tune the phase difference between the clock line and data line. The setting range is from -100 ms to 100 ms. A negative value indicates the clock moves forward and a positive value indicates the clock moves backward.

9. Curve

Press Plot to enable or disable the curve function. When “ON” is selected, the change trend of the bus data will be displayed in vector mode.

10. Polarity

Press Polarity to select the data polarity.

11. Decoding Configuration

Press Configure to enter the decoding configuration submenu.

 Press Label to enable or disable the label display function. When “ON” is selected, the bus label will be displayed at the upper left of the bus (the bus display is turned on).

 Press Line to enable or disable the bus display function. When “ON” is

selected, the bus will be displayed on the screen. You can use “Pos” to

adjust the vertical display position of the bus.

 Press Format to enable or disable the format display function. When “ON” is selected, the current bus display format will be displayed on the right side

of the label display (the bus display is turned on). You can use “Format” to

set the display format of the bus.

 Press Width to enable or disable the width display function. When “ON” is selected, the current bus width will be displayed on the right side of the

format display (the bus display is turned on). You can use “Width” to set

the bus width.

Dig.Sa menu displays the current digital sample rate which relates to the data source selected currently. When the data source is “Trace”, the digital sample rate relates to the horizontal time base.

Note: By default, MSO1000Z/DS1000Z uses “Trace” as the data source.

8-4 MSO1000Z/DS1000Z User’s Guide

Chapter 8 Protocol Decoding

RIGOL

12. Event Table

The event table displays the decoded data and the corresponding line number and time in table format. It can be used to observe relatively longer decoded data. Press Evt.TableEventTable to select “ON” ( Note: This operation is only available when MATHDecode1 / Decode2Decode is set to “ON”)

to enter the event table interface as shown in Figure 8-2.

Format: set “Data” display format in the event table to HEX, DEC or ASC.

Focus: press this key and turn to browse the “Data” item.

View: select the display form of the event table. If “Packets” is selected, the time and data are displayed in the event table; if “Details” is selected, the detailed data of the specified row will be displayed in the event table; if

“Payload” is selected, all data of the specified column will be displayed in the event table. When different view is selected, the export format of the data table will change accordingly.

Data: select the data column to be viewed in the “Datails” or “Payload” view. It is valid for the decoder of multiple data output.

Order: select the display type of decoding result in the event table to “Ascend” or “Descend”.

Export: if an USB storage device is currently connected to the instrument, press this softkey to export the data table to the external USB storage device in CSV format (“Packets” is selected) or HEX format (“Payload” or “Details” is selected).

Figure 8-2 Parallel Decoding Event Table

MSO1000Z/DS1000Z User’s Guide 8-5

RIGOL

Chapter 8 Protocol Decoding

RS232 Decoding (Option)

RS232 serial bus consists of the transmitting data line (TX) and the receiving data line (RX).

Rx Tx

DeviceA

Tx

DeviceB

Rx

Figure 8-3 RS232 Serial Bus Schematic Diagram

The industry standard of RS232 uses “Negative Logic”, namely high level is logic “0” and low level is logic “1”.

0 0 0 0

1 1 1 1 1 1

Figure 8-4 Negative Logic Schematic Diagram

By default, RS232 uses LSB (Least Significant Bit) transmission sequence, namely the lowest bit of the data is transmitted first. While for MSB (Most Significant Bit), the highest bit of the data is transmitted first. t t

Endian (LSB) Endian (MSB))

In RS232, baud rate is used to represent the transmitting rate (namely bits per second) of the data. The commonly used baud rates include 2400 bps, 4800 bps,

9600 bps, 19200 bps, 38400 bps, 57600 bps and 115200 bps.

In RS232, you need to set the start bit, data bits, check bit (optional) and stop bit of each frame of data.

Data Bit

8-6 MSO1000Z/DS1000Z User’s Guide

Chapter 8 Protocol Decoding

RIGOL

Start Bit: represent when the data starts outputting.

Data Bits: represent the number of data bits actually contained in each frame of data.

Check Bit: used to check the correctness of the data transmission.

 Odd Checkout: the total number of “1” in the data bit and check bit is an odd.

For example, when 0x55 (01010101) is sent, a 1 needs to be filled in the check bit to make the number of 1 be an odd.

 Even Checkout: the total number of “1” in the data bit and check bit is an even.

For example, when 0x55 (01010101) is sent, a 0 should be filled in the check bit.

 None: there would not be check bit during the transmission.

Stop Bit: represent when the data stops outputting.

Press MATHDecode1Decoder to select “RS232” to open the RS232 decoding function menu.

1. Press Decode to turn on or off the decoding function.

2. TX and RX Channel Setting

Press Tx to select any channel (CH1-CH4 or D0-D15) as the transmitting channel and when “OFF” is selected, no transmitting channel is set. Use the same method the set the Rx channel.

3. Baud Rate Setting

Press Baud and use to set the desired baud rate. The range is from 110 to

20000000. You can also press Preset to select 2400, 4800, 9600, 19200, 38400,

57600, 115200, 230400, 460800, 9216000, 1M, 2M, 5M, 10M and 20M. The default is 9600 and the unit is bps.

4. Copy Trigger

Press CopyTrig to copy the protocol trigger settings from the trigger system and set the corresponding decoding parameters automatically. During RS232 decoding, this function is valid only when the trigger type is RS232 trigger. You can copy the settings of data channel, width, baud rate, stop bit, check mode and polarity.

5. Polarity Setting

Press Polarity to select “

.

(positive)” or “ (negative)” and the default is

6. Order Setting

Press Order to select “LSB” or “MSB” and the default is “LSB”.

7. Data Line Setting

As mentioned before, in RS232, you need to set the start bit, data bits, check bit

(optional) and stop bit of each frame of data. “Start Bit” is specified by the

“Polarity Setting”. The setting methods of other parameters are as follows.

MSO1000Z/DS1000Z User’s Guide 8-7

RIGOL

Chapter 8 Protocol Decoding

 Press Data to set the data width of each frame. It can be set to 5, 6, 7 or 8 and the default is 8.

 Press Stop to set the stop bit after each frame of data. It can be set to 1 bit,

1.5 bits or 2 bits.

 Press Parity to set the even-odd check mode of the data transmission. It can be set to None, Odd or Even.

8. Analog Channel Threshold Setting

For more details, please refer to the introduction in “Analog Channel

Threshold Setting” of parallel decoding.

9. Display-related Setting

For more details, please refer to the introduction in “Display-related Setting

of parallel decoding.

10. Event Table

Press Evt.TableEventTable to select “ON” ( Note: This operation is only available when MATHDecode1 / Decode2Decode is set to “ON”) to enter the event table interface.

Format: set “Data” display format in the event table to HEX, DEC or ASC.

Focus: press this key and turn to browse the “Data” item.

View: select the display form of the event table. If “Packets” is selected, the time and data are displayed in the event table; if “Details” is selected, the detailed data of the specified row will be displayed in the event table; if

“Payload” is selected, all data of the specified column will be displayed in the event table. When different view is selected, the export format of the data table will change accordingly.

Data: select the data column to be viewed in the “Datails” or “Payload” view.

During RS232 decoding, if TX and RX channels are set at the same time, you can specify to display TX or RX data.

Order: select the display type of decoding result in the event table to “Ascend” or “Descend”.

Export: if a USB storage device is currently connected to the instrument, press this softkey to export the data table to the external USB storage device in CSV format (“Packets” is selected) or HEX format (“Payload” or “Details” is selected).

11. Decoding Configuration

Press Configure to enter the decoding configuration submenu.

 Press Label to enable or disable the label display function. When “ON” is selected, the bus label will be displayed at the upper left of the bus (the bus display is turned on).

 Press Line to enable or disable the bus display function. When “ON” is

selected, the bus will be displayed on the screen. You can use “Pos” to

adjust the vertical display position of the bus.

8-8 MSO1000Z/DS1000Z User’s Guide

Chapter 8 Protocol Decoding

RIGOL

 Press Format to enable or disable the format display function. When “ON” is selected, the current bus display format will be displayed on the right side

of the label display (the bus display is turned on). You can use “Format” to

set the display format of the bus.

 Press Endian to enable or disable the endian display function. When “ON” is selected, the current bus endian will be displayed on the right side of the

format display (the bus display is turned on). You can use “Order” to set the

bus endian.

 Press Width to enable or disable the width display function. When “ON” is selected, the width per frame data will be displayed on the right side of the

endian display (the bus display is turned on). You can use “Width” to set

the width per frame data.

Did.Sa menu displays the current digital sample rate which relates to the data source selected currently. When the data source is “Trace”, the digital sample rate relates to the horizontal time base.

Note: By default, MSO1000Z/DS1000Z uses “Trace” as the data source.

12. The Error Expression during Decoding

End Frame Error: errors generated when the end frame condition is not met.

When the stop bit is set to 1, error mark “?” will be displayed if the stop bit is less than 1.

The stop bit is less than 1

Check Error: when check bit error is detected during the decoding, error mark

“?” will be displayed. For example, when the transmitting terminal is set to none check and the decoder is set to odd check, the following check error occurs.

MSO1000Z/DS1000Z User’s Guide 8-9

RIGOL

(01111101, LSB)

Chapter 8 Protocol Decoding

The check bit detected is 0

Wherein, there are even number of 1 in the 5-bit data 10111 and the check bit should be 1; but the check bit detected on the TX is 0, thus check error occurs.

Note: Two error marks will be displayed when end frame error and check error are detected at the same time.

Frame Check Error Frame End Error

8-10 MSO1000Z/DS1000Z User’s Guide

Chapter 8 Protocol Decoding

I2C Decoding (Option)

I2C serial bus consists of the clock line (SCLK) and the data line (SDA).

Vcc

SCLK

Device

A2

A1

A0

SDA

Host

RIGOL

SCLK

SDA

Figure 8-5 I2C Serial Bus

Press MATHDecode1Decoder to select “I2C” and open the I2C decoding function menu.

1. Press Decode to turn on or off the decoding function.

2. CLK Setting

Press CLK to select any channel (CH1-CH4 or D0-D15) as the clock channel.

3. SDA Setting

Press DATA to select any channel (CH1-CH4 or D0-D15) as the data channel.

Note: Press Exchange to switch the current signal sources of the clock channel and data channel.

4. Copy Trigger

Press CopyTrig to copy the protocol trigger settings from the trigger system and set the corresponding decoding parameters automatically. During I2C decoding, this function is valid only when the trigger type is I2C trigger. You can copy the settings of clock and data channels.

5. Analog Channel Threshold Setting

For more details, please refer to the introduction in “Analog Channel

Threshold Setting” of parallel decoding.

6. Display-related Setting

For more details, please refer to the introduction in “Display-related Setting

of parallel decoding.

MSO1000Z/DS1000Z User’s Guide 8-11

RIGOL

Chapter 8 Protocol Decoding

7. Event Table

The event table displays the decoded data, the corresponding line number and time in table format. Press Evt. TableEventTable to select “ON” ( Note:

This operation is only available when MATHDecode1 / Decode2

Decode is set to “ON”) to enter the event table interface.

Format: set “Data” display format in the event table to HEX, DEC or ASC.

Focus: press this key and turn to browse the “Data” item.

View: select the display form of the event table. If “Packets” is selected, the time and data are displayed in the event table; if “Details” is selected, the detailed data of the specified row will be displayed in the event table; if

“Payload” is selected, all data of the specified column will be displayed in the event table. When different view is selected, the export format of the data table will change accordingly.

Data: select the data column to be viewed in the “Datails” or “Payload” view. It is valid for the decoder of multiple data output.

Order: select the display type of decoding result in the event table to “Ascend” or “Descend”.

Export: if a USB storage device is currently connected to the instrument, press this softkey to export the data table to the external USB storage device in CSV format (“Packets” is selected) or HEX format (“Payload” or “Details” is selected).

8. Decoding Configuration

Press Configure to enter the decoding configuration submenu.

 Press Label to enable or disable the label display function. When “ON” is selected, the bus label will be displayed at the upper left of the bus (the bus display is turned on).

 Press Lineto enable or disable the bus display function. When “ON” is

selected, the bus will be displayed on the screen. You can use “Pos” to

adjust the vertical display position of the bus.

 Press Format to enable or disable the format display function. When “ON” is selected, the current bus display format will be displayed on the right side

of the label display (the bus display is turned on). You can use “Format” to

set the display format of the bus.

 Press Endian to enable or disable the endian display function. When “ON” is selected, the current bus endian will be displayed on the right side of the format display (the bus display is turned on). The default is MSB.

 Press Width to enable or disable the width display function. When “ON” is selected, the width per frame data will be displayed on the right side of the endian display (the bus display is turned on). The default is 8.

Dig.Sa menu displays the current digital sample rate which relates to the data source selected currently. When the data source is “Trace”, the digital sample rate relates to the horizontal time base.

Note: By default, MSO1000Z/DS1000Z uses “Trace” as the data source.

8-12 MSO1000Z/DS1000Z User’s Guide

Chapter 8 Protocol Decoding

RIGOL

9. Address Information during Decoding

In I2C bus, the front part of each frame of data contains the address information and blue patches are used to represent the address ID. In the ID, “Write” is used to represent writing address and “Read” is used to represent reading address.

Press ADDR to select “Normal” or “R/W”. When “R/W” is selected, “AddrBits

will include the “R/W” bit as a part of the address value.

Address ID

Figure 8-6 I2C Address Information during Decoding

When the ACK (ACKnowledge Character) is not met, error marks “?” as shown in the figure below will be displayed.

MSO1000Z/DS1000Z User’s Guide 8-13

RIGOL

Chapter 8 Protocol Decoding

ACK=1

Figure 8-7 I2C Error Expressions during Decoding

8-14 MSO1000Z/DS1000Z User’s Guide

Chapter 8 Protocol Decoding

RIGOL

SPI Decoding (Option)

SPI bus is based on the master-slave configuration and usually consists of chip select line (CS), clock line (SCLK) and data line (SDA). Wherein, the data line includes MISO and MOSI.

SCLK

MOSI

Master Slave

MISO SCLK

CS

SDA

(MISO/MOSI)

Figure 8-1 SPI Serial Bus

SCLK: SDA is sampled on the rising or falling edge of clock.

SDA: represents the data channel.

Press MATHDecode1Decoder to select “SPI” and open SPI decoding function menu.

1. Press Decode to turn on or off the decoding function.

2. CLK Setting

Press CLK to select any channel (CH1-CH4 or D0-D15) as the clock channel.

3. MISO and MOSI Settings

Press MISO to select any channel (CH1-CH4 or D0-D15) as the MISO data channel. If “OFF” is selected, this data line is not set. You can use the same method to set the MOSI data line.

Note: Press Exchange to switch the current signal sources of the clock channel and data channel.

4. Copy Trigger

Press CopyTrig to copy the protocol trigger settings from the trigger system and set the corresponding decoding parameters automatically. During SPI decoding, this function is valid only when the trigger type is SPI trigger. You can copy the settings of clock, data channel, databits, edge, type (CS/Timeout), CS channel, CS polarity and timeout time as well as set the data polarity to positive.

MSO1000Z/DS1000Z User’s Guide 8-15

RIGOL

Chapter 8 Protocol Decoding

5. Mode

 Timeout: you can perform frame synchronization according to the timeout and the timeout must be greater than half of the clock cycle.

 CS: it contains a chip select line (CS). You can perform frame synchronization according to CS. Press Mode to select “CS” and enter the

CS mode. At this point, you can select CS signal line and CS polarity. The CS signal line can be CH1-CH4 or D0-D15. The CS polarity can be “ ”

(high level is valid, namely the instrument starts transmitting data when the

CS is high level) or “ ” (low level is valid, namely the instrument starts transmitting data when the CS is low level).

6. Timeout

When there are only two lines, the decoder can find the right start frame according to the timeout. The timeout must be greater than the maximum clock pulse width and less than the idle time between frames. For the setting method,

please refer to the introduction in “Parameter Setting Method”.

7. Edge Setting

Press Edge to set the instrument to sample MISO and MOSI on the rising edge

( ) or falling edge ( ) of CLK.

8. Polarity Setting

Press Polarity to set the polarity of the SDA data line to (high level is 1) or

(low level is 1).

9. Width Setting

Press Width to set the number of digits per frame of data and the setting range is from 4 to 32.

10. Order

Press Order to select “LSB” or “MSB” and the default is MSB.

11. Display-related Setting

For more details, please refer to the introduction in “Display-related Setting

of parallel decoding.

12. Event Table

The event table displays the decoded data, the corresponding line number and time in table format. Press Evt.TableEventTable to select “ON” ( Note:

This operation is only available when MATHDecode1 / Decode2

Decode is set to “ON”) to enter the event table interface.

Format: set “Data” display format in the event table to HEX, DEC or ASC.

Focus: press this key and turn to browse the “Data” item.

View: select the display form of the event table. If “Packets” is selected, the

8-16 MSO1000Z/DS1000Z User’s Guide

Chapter 8 Protocol Decoding

RIGOL time and data are displayed in the event table; if “Details” is selected, the detailed data of the specified row will be displayed in the event table; if

“Payload” is selected, all data of the specified column will be displayed in the event table. When different view is selected, the export format of the data table will change accordingly.

Data: select the data column to be viewed in the “Datails” or “Payload” view.

During SPI decoding, if the MISO and MOSI data lines are set at the same time, you can specify to display the data of MISO or MOSI.

Order: select the display type of decoding result in the event table to “Ascend” or “Descend”.

Export: if a USB storage device is currently connected to the instrument, press this softkey to export the data table to the external USB storage device in CSV format (“Packets” is selected) or HEX format (“Payload” or “Details” is selected).

13. Decoding Configuration

Press Configure to enter the decoding configuration submenu.

 Press Label to enable or disable the label display function. When “ON” is selected, the bus label will be displayed at the upper left of the bus (the bus display is turned on).

 Press Line to enable or disable the bus display function. When “ON” is

selected, the bus will be displayed on the screen. You can use “Pos” to

adjust the vertical display position of the bus.

 Press Format to enable or disable the format display function. When “ON” is selected, the current bus display format will be displayed on the right side

of the label display (the bus display is turned on). You can use “Format” to

set the display format of the bus.

 Press Endian to enable or disable the endian display function. When “ON” is selected, the current bus endian will be displayed on the right side of the

format display (the bus display is turned on). You can use “Order” to set the

bus endian.

 Press Width to enable or disable the width display function. When “ON” is selected, the number of digits per frame data will be displayed on the right side of the endian display (the bus display is turned on). You can use

Width” to set the number of digits per frame data.

Dig.Sa menu displays the current digital sample rate which relates to the data source selected currently. When the data source is “Trace”, the digital sample rate relates to the horizontal time base.

Note: By default, MSO1000Z/DS1000Z uses “Trace” as the data source.

MSO1000Z/DS1000Z User’s Guide 8-17

Chapter 9 Reference Waveform

Chapter 9 Reference Waveform

In the actual test process, you can compare the waveform with the reference waveform to determine the cause of failure.

The contents of this chapter:

 To Enable REF Function

 To Select REF Source

 To Adjust REF Waveform Display

 To Save to Internal Memory

 To Set the Color

 To Reset the Reference Waveform

 To Export to Internal or External Memory

 To Import from Internal or External Memory

RIGOL

MSO1000Z/DS1000Z User’s Guide 9-1

RIGOL

Chapter 9 Reference Waveform

To Enable REF Function

Press REF in the vertical control area (VERTICAL) at the front panel to enable the

REF function.

MSO1000Z/DS1000Z provides 10 reference waveform channels. Press Channel and use to set the desired reference channel to on or off and the channel icon (for example, ) of the channel enabled will be displayed at the left side of the screen grid.

When the REF function is enabled, you can select different color for each reference waveform, set the source of each reference channel, adjust the vertical scale and position of the reference waveform and save the reference waveform to internal or external memory as well as recall it when needed. For the details, please refer to the introductions below.

To Select REF Source

Press Current and use to select any of the reference channels (Ref1 to Ref10) enabled and then press Source to specify a reference source (CH1-CH4, MATH or

D0-D15) for this channel.

To Adjust REF Waveform Display

You can adjust the reference waveform specified in Current.

Press REF to enable the REF function. Press Offset and use to adjust the vertical position of the reference waveform and press Scale and use to adjust the vertical scale of the reference waveform. At this point, you can also directly rotate VERTICAL POSITION and VERTICAL SCALE to adjust the vertical position and scale respectively.

To Save to Internal Memory

Press Save to save the waveform (screen region) in the specified source to the internal memory as reference waveform.

Note: This operation only saves the reference waveform in the volatile memory and the waveform will be cleared at power-off.

9-2 MSO1000Z/DS1000Z User’s Guide

Chapter 9 Reference Waveform

RIGOL

To Set the Color

MSO1000Z/DS1000Z series oscilloscope provides five colors (light gray, green, light blue, magenta and orange) to mark the reference waveforms of different channels in order to distinguish them.

Press Current and use to select any of the reference channels (Ref1-Ref10) enabled. Then, press Color to specify a different color for the reference waveform of that channel. The corresponding icon at the left of the channel currently selected will be filled with the specified color, for example, .

To Reset the Reference Waveform

Press Reset and the reference waveform returns to the position where the source channel waveform is located when the Save operation was executed.

To Export to Internal or External Memory

Users can also save the reference waveform to the internal Flash memory or external

USB storage device. The file format of the reference waveform is “*.ref”.

Press Export to enter the file store interface. Please refer to the relative descriptions

in “Store and Recall” to save the reference waveform to the internal or external

memory.

To Import from Internal or External Memory

Users can also import the reference waveform stored in the internal Flash memory or external USB storage device to the internal memory.

Press Import to enter the file recall interface. Please refer to the relative

descriptions in “Store and Recall” to import the reference waveform to the internal

memory of the instrument.

MSO1000Z/DS1000Z User’s Guide 9-3

Chapter 10 Pass/Fail Test

RIGOL

Chapter 10 Pass/Fail Test

During the product design and manufacture process, you usually need to monitor the change of the signal or judge whether the product is up to standard. The pass/fail test function of MSO1000Z/DS1000Z oscilloscope can fulfill this task perfectly.

The contents of this chapter:

 To Enable Pass/Fail Test

 To Select Source

 Mask Range

 Test and Output

 To Save the Test Mask

 To Load the Test Mask

MSO1000Z/DS1000Z User’s Guide 10-1

RIGOL

Chapter 10 Pass/Fail Test

To Enable Pass/Fail Test

Press UtilityPass/FailEnable to select “ON”. The blue shadow area denotes the failure area and if the waveform measured at certain time point during the measurement process passes through the failure area, it will be treated as failed frame. Press Operate to select “ ” to start testing and select “■” to stop testing.

Note: When the time base is set to X-Y, ROLL or the horizontal time base is set to

200 ms/div or slower under YT mode and the instrument enter the “slow sweep” mode, the pass/fail test function is not available.

When the test function is enabled, you can set the output form of the test results. For

the details, please refer to the introduction in “Test and Output”. You can also

select the signal source, set the test mask range as well as save and recall the test mask. For the details, refer to the introductions below.

To Select Source

Before selecting the source, you need to connect the signal under test to the analog channel input terminal of the oscilloscope. Press Source to select the channel

(CH1-CH4) to be tested.

Note: Only channels enabled can be selected.

Mask Range

Users can define their desired test mask range. Press RangeMaskRange to set the mask region to “Screen” or “Cursor”. The default is “Screen”.

When “Screen” is selected, the entire waveform display area is defined as the mask region. When “Cursor” is selected, two gray cursor lines appear on the screen. At this point, press CursorA and CursorB; then, use to adjust the positions of the two cursor lines respectively. Or, press CursorAB and use to adjust the positions of cursor A and cursor B at the same time. Thereby the mask region is determined.

When the mask region is determined, press X Mask and Y Mask respectively, rotate

to adjust the pass/fail mask range. Press Create to apply the mask currently created. The horizontal and vertical adjustment ranges are 0.02 div to 4.0 div and

0.04 div to 5.12 div respectively.

10-2 MSO1000Z/DS1000Z User’s Guide

Chapter 10 Pass/Fail Test

RIGOL

Test and Output

You can set the output form of the test results by the following method.

Press Stat.Disp to select “ON” or “OFF”. When “ON” is selected, the test results will be displayed at the upper right corner of the screen as shown in the figure below.

Press Stat.Reset to clear the current data and perform statistics on the test results again.

Press StopOnFail to select “ON” or “OFF”.

 ON: when failed waveforms are detected, the oscilloscope will stop the test and enter the “STOP” state. At this point, if Stat.Disp is turned on, the last statistic results will be displayed on the screen; if Aux Out is turned on, a pulse is output from the [Trigger Out] connector at the rear panel.

 OFF: the oscilloscope will continue with the test even though failed waveforms are detected. The test results on the screen will update continuously. If Aux

Out is turned on, a pulse will be output from the [Trigger Out] connector at the rear panel each time a failed waveform is detected.

Press Output to select “ ” or “ ”.

 : when failed waveforms are detected, the instrument displays the statistic results and outputs a pulse.

 : when failed waveforms are detected, the instrument displays the statistic results and outputs a pulse and the beeper sounds (not related to the on/off state of the sound).

Press AuxOut to quickly turn “ON” or “OFF” the output of test results from the

[Trigger Out] connector at the rear panel. When failed waveforms are detected, a positive pulse is output from the [Trigger Out] connector at the rear panel; when failed waveforms are not detected, the instrument will output low level continuously.

To Save the Test Mask

Users can save the current test mask to the internal Flash memory or external USB storage device. The file format of the test mask file is “*.pf”.

Press Save to enter the file store interface. Please refer to the relative descriptions in

Store and Recall” to save the test mask file to the internal or external memory.

MSO1000Z/DS1000Z User’s Guide 10-3

RIGOL

Chapter 10 Pass/Fail Test

To Load the Test Mask

Users can also load the test mask files (*.pf) stored in the internal Flash memory or external USB storage device to the internal memory.

Press Load to enter the file recall interface. Please refer to the relative descriptions

in “Store and Recall” to load the test masks to the internal memory of the

instrument.

10-4 MSO1000Z/DS1000Z User’s Guide

Chapter 11 Waveform Record

RIGOL

Chapter 11 Waveform Record

Waveform record can record the waveforms of the analog input channels (CH1-CH4) and the digital channels (D0-D15). Waveform playback can provide better waveform analysis effect.

Note: The horizontal time base must be set to Y-T mode during waveform record.

The contents of this chapter:

 Common Settings

 Playback Option

 Record Option

MSO1000Z/DS1000Z User’s Guide 11-1

RIGOL

Chapter 11 Waveform Record

Common Settings

Press UtilityRecordRecord to turn on or off the waveform record function.

1. Record Waveform

Before recording the waveform, you can refer to the introduction in “Record

Option” to set the waveform recording parameters. Press Record to start

recording the waveform. During the recording, the current recording information as shown in the figure below is displayed at the upper right corner of the screen.

“●” in the menu changes to “■” automatically. At this point, press Record again to stop recording.

Tip

Under the waveform recording menu, pressing RUN/STOP can also start and stop the recording.

2. Play

Press Play to start playing the waveform recorded. For more details, please

refer to the introduction in “ Playback Option”.

3. Stop

During the playback, you can press Stop to stop the playback.

4. Current Frame

Press Current and use to set the current frame. The default is the maximum number of frames recorded currently. During the setting, the screen will display the corresponding waveform of the current frame synchronously, namely manual playback.

5. Step Direction

Press Step Dir to select the single frame playback direction of the recorded waveform.

 : press SINGLE to play the single frame forward (namely the single frame playback sequence is the same as the waveform record sequence) and the number of the frame is incremental.

 : press SINGLE to play the single frame backward (namely the single frame playback sequence is opposite to the waveform record sequence) and the number of the frame is decremental.

6. The Last/First Frame

You can view the last or first frame of waveform. Pressing the last frame of the recorded waveform; pressing frame of the recorded waveform.

can jump to

can jump to the first

11-2 MSO1000Z/DS1000Z User’s Guide

Chapter 11 Waveform Record

RIGOL

Playback Option

Waveform playback can play back the waveforms currently recorded. During the playback, the information as shown in the figure below is displayed at the upper right corner of the screen. The data at the left side of the figure indicates the specific frame currently displayed on the screen. During the playback, this value would change continuously. The data at the right side of the figure indicates the maximum number of frames recorded currently.

Before playing back the waveform, you can press Play Opt to set the playback parameters.

1. Play Mode

Press Mode to set the playback mode to cycle or single.

 : cycle playback. Play from the start frame to the end frame and then repeat until you stop it manually.

 : single playback. Play from the start frame to the end frame and then stop.

2. Play Direction

Press Dir to set the play direction to positive or negative direction.

 : positive direction. Play from the start frame to the end frame.

: negative direction. Play from the end frame to the start frame. 

3. Interval

Press Interval to set the time interval of playback. The range availbale is from

100 ns to 10 s and the default is 100 ns. For the setting method, please refer to

the introduction in “Parameter Setting Method”.

4. Start Frame

Press Start to set the start frame of playback. The default is 1 and the maximum is the maximum number of frames recorded. For the setting method,

please refer to the introduction in “Parameter Setting Method”.

5. End Frame

Press End to set the end frame of the playback. The default is the total number of frames of the waveform recorded. For the setting method, please refer to the

introduction in “Parameter Setting Method”.

MSO1000Z/DS1000Z User’s Guide 11-3

RIGOL

Chapter 11 Waveform Record

Record Option

Before recording the waveform, you can press Record Opt to set the following parameters.

1. Interval

Press Interval to set the time interval between the frames in waveform recording and the range available is from 100 ns to 10 s. For the setting method,

please refer to the introduction in “Parameter Setting Method”.

2. Length

Press Length to set the number of waveform frames to be recorded currently.The range availbale is from 1 to the maximum number of frames can be recorded currently. For the setting method, please refer to the introduction in

Parameter Setting Method”. Press Set Max to set the number of waveform

frames to be recorded to the maximum number of frames can be recorded currently.

3. Max Length

The menu shows the maximum number of frames that can be recorded currently.

As the capacity of the waveform memory is fixed, the more the number of points each frame of waveform has, the less the number of waveform frames can be recorded. Thus, the maximum end frame of waveform recording is decided by the memory depth currently selected. The less the memory depth, the more the number of waveform frames can be recorded.

The current memory depth refers to the number of waveform points per frame.

Memory depth=Sample rate×Horizontal time base×Number of grids horizontally on the screen; wherein, for MSO1000Z/DS1000Z, “Number of grids horizontally on the screen” is 12. Therefore, the maximum end frame of waveform recording is also related to the “Sample rate” and “Horizontal time base”.

4. Beep

: the beeper does not sound at the end of recording.

: the beeper sounds at the end of recording.

11-4 MSO1000Z/DS1000Z User’s Guide

Chapter 12 Display Control

RIGOL

Chapter 12 Display Control

You can set the type, persistence time and brightness of waveform display as well as the grid type and grid brightness of the screen display.

The contents of this chapter:

 To Select the Display Type

 To Set the Persistence Time

 To Set the Waveform Intensity

 To Set the Screen Grid

 To Set the Grid Brightness

MSO1000Z/DS1000Z User’s Guide 12-1

RIGOL

Chapter 12 Display Control

To Select the Display Type

Press DisplayType to set the waveform display mode to “Vectors” or “Dots”.

 Vectors: the sample points are connected by lines and displayed, as shown in the figure below. Normally, this mode can provide the most vivid waveform to view the steep edge of the waveform (such as square waveform).

 Dots: display the sample points directly as shown in the figure below. You can directly view each sample point and use the cursor to measure the X and Y values of the sample point.

Vectors Display Dots Display

To Set the Persistence Time

Press DisplayPersis.Time to set the persistence time of the oscilloscope to Min, specific values (100 ms, 200 ms, 500 ms, 1 s, 5 s and 10 s) or Infinite.

In the following part, the frequency sweep signal of a sine waveform is used to demonstrate the waveform effects in different persistence times.

1. Min

Enable to view waveform changing in high refresh rate.

2. Specific Values

Enable to observe glitch that changes relatively slowly or glitch with low occurrence probability. The persistence time can be set to 100 ms, 200 ms, 500 ms, 1 s, 5 s or 10 s.

12-2 MSO1000Z/DS1000Z User’s Guide

Chapter 12 Display Control

RIGOL

3. Infinite

In this mode, the oscilloscope displays the waveform newly acquired without clearing the waveforms acquired formerly. The waveforms acquired formerly will be displayed in relatively low-brightness color and the waveform newly acquired will be displayed in normal brightness and color. Infinite persistence can be used to measure noise and jitter and to capture incidental events.

To Set the Waveform Intensity

Press DisplayIntensity or turn in non-menu operation to adjust the waveform brightness of the channel. The default is 60% and the range available is from 0% to 100%.

To Set the Screen Grid

Press DisplayGrid to set the screen grid type.

: turn the background grid and coordinate on.

: turn the background grid off.

 : turn the background grid and coordinate off.

To Set the Grid Brightness

Press DisplayBrightness to set the brightness of the screen grid. Turn to adjust the grid brightness. The default is 50% and the range available is from 0% to

100%.

MSO1000Z/DS1000Z User’s Guide 12-3

Chapter 13 Signal Source

RIGOL

Chapter 13 Signal Source

MSO1000Z/DS1000Z combines signal generator and oscilloscope in one instrument by providing a built-in dual-channel, 25MHz signal generator, thus providing convenience for engineers who need to use the signal generator and oscilloscope at the same time. This chapter introduces how to use the built-in signal generator. As the functions and setting methods of source 1 and source 2 are the same, this chapter takes Source 1 for illustration.

Press Source to enter the signal generator control menu.

 Press Output to enable or disable the signal output. When it is enabled, the

[Source1] connector at the rear panel outputs a signal according to the current setting.

 Press Src1Conf to set the signal type and the signal parameters.

 Press StatusDisp to view the parameter settings of SOURCE1 and SOURCE2, such as the frequency, amplitude, offset, phase, modulation type and modulation frequency, etc.

 Press RecoveryCalls to set the output signal type and parameters to be recalled when the oscilloscope is powered on again after power-off. You can select “Last” or “Default”.

 Pressing Factory can restore the default settings of the signal source separately.

Note: The function of SourceOutput is the same with that of Source

Src1ConfOutput.

The contents of this chapter:

 To Output Basic Waveform

 To Output Built-in Waveform

 To Output Arbitrary Waveform

 Modulation

MSO1000Z/DS1000Z User’s Guide 13-1

RIGOL

Chapter 13 Signal Source

To Output Basic Waveform

MSO1000Z/DS1000Z built-in signal source can output a variety of basic waveforms, including sine, square, ramp, pulse, DC and noise.

To Output Sine

Press Src1Conf to open the waveform setting menu. Press Wave and select “Sine”.

At this point, you can set the sine parameters.

1. Source 1 output

Press Output to turn on or off the signal output. This function is the same with that of SourceOutput. When “ON” is selected, the signal set currently is output from the [Source1] connector at the rear panel.

2. Set the frequency

Press Frequency to set the frequency of the signal. For the setting method,

please refer to the introduction in “Parameter Setting Method”. For different

waveforms, the frequency ranges are different.

Sine: 100 mHz to 25 MHz

Square: 100 mHz to 15 MHz

Ramp: 100 mHz to 100 kHz

Pulse: 100 mHz to 1 MHz

DC and noise: there is no frequency parameter.

3. Set the amplitude

Press Amplitude to set the amplitude of the signal. For the setting method,

please refer to the introduction in “Parameter Setting Method”. When the

impedance is set to HighZ, the range is from 20 mVpp to 5 Vpp; when the impedance is set to 50 Ω, the range is from 10 mVpp to 2.5 Vpp.

4. Set the DC offset voltage

Press Offset to set the DC offset of the signal. For the setting method, please

refer to the introduction in “Parameter Setting Method”. When the

impedance is HighZ, the range is from (-2.5 V+the current amplitude/2) to (2.5

V-the current amplitude/2); when the impedance is set to 50 Ω, the range is from

(-1.25 V+the current amplitude/2) to (1.25 V-the current amplitude/2).

5. Set the start phase

Press StartPhase to set the start phase of the signal. For the setting method,

please refer to the introduction in “Parameter Setting Method”. The range is

from 0°to 360°.

13-2 MSO1000Z/DS1000Z User’s Guide

Chapter 13 Signal Source

RIGOL

6. Align phase

Pressing AlignPhase will re-configure the two channels to output according to the preset frequency and phase. For two signals of which the frequencies are the same or are multiples, this operation can align their phases. Acquire the waveforms of the two channels using the oscilloscope and display the waveforms stably. After switching the channel status, the phase deviation between the two waveforms is changed. At this point, press AlignPhase and the phase deviation shown on the oscilloscope will restore to the current phase deviation between the two waveforms automatically.

7. Modulation

Press Src1Mod to enter the modulation setting interface. For detailed

introduction of the modulation function, please refer to “ Modulation”.

Note: When “Pulse”, “DC” or “Noise” is selected by pressing Wave, there is not this setting.

8. Set the impedance

Press Impedance to set the output impedance of the signal generator. It can be set to “HighZ” or “50Ω”.

To Output Square

Press Src1Conf to open the waveform setting menu. Press Wave and select

“Square”. At this point, you can set the square parameters. For the detailed setting

method, refer to the introduction in “To Output Sine”. The duty cycle of square is

fixed at 50%.

To Output Ramp

Press Src1Conf to open the waveform setting menu. Press Wave and select “Ramp”.

At this point, you can set the ramp parameters. For the detailed setting method, refer

to the introduction in “To Output Sine”. This section only introduces “Symmetry”.

Symmetry is defined as the percentage that the rising period of the ramp takes up in the whole period.

MSO1000Z/DS1000Z User’s Guide 13-3

RIGOL t Symmetry=t/T*100%

Chapter 13 Signal Source

T

Figure 13-1 Symmetry Definition

Press Symmetry to set the symmetry of the ramp waveform. For the setting method,

please refer to the introduction in “Parameter Setting Method”. The range is from

0% to 100%.

13-4 MSO1000Z/DS1000Z User’s Guide

Chapter 13 Signal Source

RIGOL

To Output Pulse

Press Src1Conf to open the waveform setting menu. Press Wave and select “Pulse”.

At this point, you can set the pulse parameters. For the detailed setting method, refer

to the introduction in “To Output Sine”. This section only introduces “DutyCycle”.

Duty cycle is defined as the percentage that the high level takes up in the whole pulse period. t

Duty Cycle=t/T*100%

T

Figure 13-2 Duty Cycle Definition

Press DutyCycle to set the duty cycle of the pulse. For the setting method, please

refer to the introduction in “Parameter Setting Method”. The range is from 10%

to 90%.

To Output DC

Press Src1Conf to open the waveform setting menu. Press Wave and select “DC”.

At this point, you can output a DC signal with the specified offset and impendance.

1. Source 1 output

Press Output to turn on or off the signal output.

2. Set the offset

Press Offset to set the offset of the DC signal. For the setting method, please

refer to the introduction in “Parameter Setting Method”. When the

impedance is HighZ, the range is from -2.5 V to +2.5 V; when the impedance is set to 50 Ω, the range is from -1.25 V to +1.25 V.

3. Set the impedance

Press Impedance to set the output impedance of the signal generator. It can be set to “HighZ” or “50Ω”

MSO1000Z/DS1000Z User’s Guide 13-5

RIGOL

Chapter 13 Signal Source

To Output Noise

Press Src1Conf to open the waveform setting menu. Press Wave and select “Noise”.

At this point, you can output a noise signal with the specified amplitude, offset and impendance.

1. Source 1 output

Press Output to turn on or off the signal output.

2. Set the amplitude

Press Amplitude to set the amplitude of the signal. For the setting method,

please refer to the introduction in “Parameter Setting Method”. When the

impedance is set to HighZ, the range is from 20 mVpp to 5 Vpp; when the impedance is set to 50 Ω, the range is from 10 mVpp to 2.5 Vpp.

3. Set the offset

Press Offset to set the DC offset of the noise signal. For the setting method,

please refer to the introduction in “ Parameter Setting Method”. When the

impedance is HighZ, the range is from (-2.5 V+the current amplitude/2) to (2.5

V-the current amplitude/2); when the impedance is set to 50 Ω, the range is from

(-1.25 V+the current amplitude/2) to (1.25 V-the current amplitude/2).

To Output Built-in Waveform

MSO1000Z/DS1000Z series oscilloscope provides 7 kinds of built-in waveforms, including Sinc, ExpRise, ExpFall, ECG, Gauss, Lorentz and Haversine. Press

Src1Conf to open the waveform setting menu. Press Wave and select “Built-in”. At this point, you can set the output signal parameters according to the type of the built-in waveform selected.

1. Source 1 output

Press Output to turn on or off the signal output.

2. Select built-in waveform

Press Built-In to select any of the 7 built-in waveforms (as shown in the figures below).

13-6 MSO1000Z/DS1000Z User’s Guide

Chapter 13 Signal Source

RIGOL

Figure 13-3 Built-in Waveform-Sinc

Figure 13-4 Built-in Waveform-ExpRise

Figure 13-5 Built-in Waveform-ExpFall

MSO1000Z/DS1000Z User’s Guide 13-7

RIGOL

Chapter 13 Signal Source

Figure 13-6 Built-in Waveform-ECG

Figure 13-7 Built-in Waveform-Gauss

13-8

Figure 13-8 Built-in Waveform-Lorentz

MSO1000Z/DS1000Z User’s Guide

Chapter 13 Signal Source

RIGOL

Figure 13-9 Built-in Waveform-Haversine

3. Set the frequency

Press Frequency to set the frequency of the signal. For the setting method,

please refer to the introduction in “ Parameter Setting Method”. The range is

from 100 mHz to 1 MHz.

4. Set the amplitude

Press Amplitude to set the amplitude of the signal. For the setting method,

please refer to the introduction in “Parameter Setting Method”. When the

impedance is set to HighZ, the range is from 20 mVpp to 5 Vpp; when the impedance is set to 50 Ω, the range is from 10 mVpp to 2.5 Vpp.

5. Set the DC offset voltage

Press Offset to set the DC offset of the signal. For the setting method, please

refer to the introduction in “ Parameter Setting Method”. When the

impedance is HighZ, the range is from (-2.5 V+the current amplitude/2) to (2.5

V-the current amplitude/2); when the impedance is set to 50 Ω, the range is from

(-1.25 V+the current amplitude/2) to (1.25 V-the current amplitude/2).

6. Set the start phase

Press StartPhase to set the start phase of the signal. For the setting method,

please refer to the introduction in “Parameter Setting Method”. The range is

from 0°to 360°.

7. Align Phase

Pressing AlignPhase will re-configure the two channels to output according to the preset frequency and phase. For two signals of which the frequencies are the same or are multiples, this operation can align their phases. Acquire the waveforms of the two channels using the oscilloscope and display the waveforms stably. After switching the channel status, the phase deviation between the two waveforms is changed. At this point, press AlignPhase and the phase deviation shown on the oscilloscope will restore to the current phase deviation between the two waveforms automatically.

MSO1000Z/DS1000Z User’s Guide 13-9

RIGOL

Chapter 13 Signal Source

8. Modulation

Press Src1Mod to enter the modulation setting interface. For detailed

introduction of the modulation function, please refer to “Modulation”.

Note: When “Pulse”, “DC” or “Noise” is selected, the modulation menu will be hidden automatically.

9. Set the impedance

Press Impedance to set the output impedance of the signal generator. It can be set to “HighZ” or “50Ω”.

To Output Arbitrary Waveform

MSO1000Z/DS1000Z allows users to define their own arbitrary waveform and save it in the internal or external memory. At most 10 arbitrary waveforms can be stored in the internal memory. The user-defined waveform can contain 1 to 16384 data points

(namely 1 pts to 16 kpts).

1. Source 1 output

Press Output to turn on or off the signal output.

2. Set the frequency

Press Frequency to set the frequency of the signal. For the setting method,

please refer to the introduction in “Parameter Setting Method”. The available

range is from 100 mHz to 10 MHz.

3. Set the amplitude

Press Amplitude to set the amplitude of the signal. For the setting method,

please refer to the introduction in “Parameter Setting Method”. When the

impedance is set to HighZ, the range is from 20 mVpp to 5 Vpp; when the impedance is set to 50 Ω, the range is from 10 mVpp to 2.5 Vpp.

4. Set the DC offset voltage

Press Offset to set the DC offset of the signal. For the setting method, please

refer to the introduction in “Parameter Setting Method”. When the

impedance is HighZ, the range is from (-2.5 V+the current amplitude/2) to (2.5

V-the current amplitude/2); when the impedance is set to 50 Ω, the range is from

(-1.25 V+the current amplitude/2) to (1.25 V-the current amplitude/2).

5. Set the start phase

Press StartPhase to set the start phase of the signal. For the setting method,

please refer to the introduction in “Parameter Setting Method”. The range is

from 0°to 360°.

13-10 MSO1000Z/DS1000Z User’s Guide

Chapter 13 Signal Source

RIGOL

6. Align Phase

The align phase operation is for the two channels. Pressing AlignPhase will re-configure the two channels to output according to the preset frequency and phase. For two signals of which the frequencies are the same or are multiples, this operation can align their phases. Acquire the waveforms of the two channels using the oscilloscope and display the waveforms stably. After switching the channel status, the phase deviation between the two waveforms is changed. At this point, press AlignPhase and the phase deviation shown on the oscilloscope will restore to the current phase deviation between the two waveforms automatically.

7. Select waveform

Select the arbitrary waveform in the internal or external memory. For the detailed

operation, please refer to the introduction in “To Select Waveform”.

8. Create waveform

Users define their own arbitrary waveform. For the detailed operation, please

refer to the introduction in “To Create Waveform”.

9. Edit waveform

Edit the arbitrary waveform stored. For the detailed operation, please refer to the

introduction in “To Edit Waveform”.

10. Modulation

Press Src1Mod to enter the modulation setting interface. For detailed

introduction of the modulation function, please refer to “ Modulation”.

11. Set the impedance

Press Impedance to set the output impedance of the signal generator. It can be set to “HighZ” or “50Ω”.

The following section introduces how “To Select Waveform”, “To Create

Waveform” and “To Edit Waveform”.

To Select Waveform

Users can load the waveform stored or recall the channel signal waveform according to their needs.

Load

Select arbitrary waveforms in the internal or external memory (D disk) for output.

Press SelectLoad, use to select the desired waveform. For the detailed

operation, please refer to the introduction in “Store and Recall”.

MSO1000Z/DS1000Z User’s Guide 13-11

RIGOL

Chapter 13 Signal Source

Channel Signal

Select the signals of analog channel (CH1-CH4) turned on currently for output.

Press SelectCHSignal to set the desired channel signal.

― Press Source to select the waveform of any of the channels turned on

(CH1-CH4).

― Press WaveRange to set the waveform in the “Cursor” or “Screen” region for output. When “Cursor” is selected, Press CursorA or CursorB and use

to adjust the positions of the two orange cursor lines respectively; or press CursorAB and rotat to adjust the positions of cursors A and B at the same time.

― Press Probe to set the probe ratio when the analog channel signal is recalled. You can select “0.1X”, “0.2X”, “0.5X”, “1X”, “2X” or “5X”. For example, if the selected analog channel amplitude is 10 VPP and the probe ratio is “0.5X”, the amplitude of the arbitray waveform recalled is 5 VPP.

― Pressing Load can recall the channel signal selected.

You can also edit the waveform currently selected; For the details, refer to the

introduction in “To Edit Waveform”.

To Create Waveform

Users can create arbitrary waveforms according to their needs. Press Create to enter the waveform creation menu.

1. Set the Initial Points

Initial points refers to the number of editable points. By default, the number of editable points of arbitrary waveform created can be prset to 2 points. Point 1 is fixed at 0 s and point 2 is fixed at the middle of the period. Press InitPoint to set the number of editable points of the new waveform. For the setting method,

please refer to the introduction in “Parameter Setting Method”. The arbitrary

waveform can contain up to 16384 (16 kpts) points.

2. Interpolation

Press Interp to turn on or off the interpolation mode between the points of the arbitrary waveform.

ON: the waveform editor connects two points using a straight line.

13-12 MSO1000Z/DS1000Z User’s Guide

Chapter 13 Signal Source

RIGOL

OFF: the waveform editor will keep a constant voltage level between two points and create a ladder-like waveform.

3. Zoom

Press Zoom to turn on or off the zoom function.

ON: the waveform editing window only displays the current point.

OFF: the waveform editing window displays all the initial points.

4. Current Point

Press CurPoint to set the point to be edited. The range is from 1 to the initial

points. For the setting method, please refer to the introduction in “Parameter

Setting Method”.

5. Voltage

Press Voltage to set the voltage of the current point. The range is from -2.5 V to

+2.5 V and the default is 0.000 mV. For the setting method, please refer to the

introduction in “Parameter Setting Method”.

6. Time

Press Time to set the duration of the current point. This setting is limited by the times of the previous point and following point. The time of point 1 is fixed at 0 s.

7. Insert

Press Insert to insert a new waveform point between the current point and the next point. At this time, the initial points increases by one. You can press Insert continuously to increase the editable point successively.

MSO1000Z/DS1000Z User’s Guide 13-13

RIGOL

Chapter 13 Signal Source

8. Delete

Press Delete to delete the current point from the waveform and connect the remaining points using the current interpolation mode.

Note: Point 1 cannot be deleted.

9. APPly

Press Apply to finish editing the current waveform and output the edited waveform.

10. Save

Press Save to enter the file store interface. Please refer to the introduction in

Store and Recall” to save the current waveform file in “.arb” format in the

internal memory or external memory (you can overwrite the original file or save the current waveform again). You can select the arbitrary waveforms in the internal or external memory for output. For the details, refer to the introduction

in “To Select Waveform”.

To Edit Waveform

Users can edit the waveform stored. Press Edit to enter the waveform editing menu.

1. Interp

Press Interp to turn on or off the interpolation mode between the points of the waveform.

ON: the waveform editor connects two points using a straight line.

OFF: the waveform editor will keep a constant voltage level between the two points and create a ladder-like waveform when the interpolation is disabled.

2. Zoom

Press Zoom to turn on or off the zoom function.

ON: the waveform editing window only displays the current point.

OFF: the waveform editing window displays all the initial points.

3. Current Point

Press CurPoint to set the point to be edited. The range is from 1 to the initial

points. For the setting method, please refer to the introduction in “Parameter

Setting Method”.

4. Voltage

Press Voltage to set the voltage of the current point. The range is from -2.5 V to

+2.5 V and the default is 0.000 mV. For the setting method, please refer to the

introduction in “Parameter Setting Method”.

13-14 MSO1000Z/DS1000Z User’s Guide

Chapter 13 Signal Source

RIGOL

5. Time

Press Time to set the duration of the current point. This setting is limited by the times of the previous point and following point. The time of point 1 is fixed at 0 s.

For the setting method, please refer to the introduction in “Parameter Setting

Method”.

6. Insert

Press Insert to insert a new waveform point in the middle position between the current point and the next point.

7. Delete

Press Delete to delete the current point from the waveform and connect the remaining points using the current interpolation mode.

Note: Point 1 cannot be deleted.

8. Apply

Press Apply to finish editing the current waveform.

9. Save

Press Save to enter the file store interface. Please refer to the introduction in

Store and Recall” to save the current waveform file in “.arb” format in the

internal memory or external memory (you can overwrite the original file or save the current waveform again).You can select the arbitrary waveforms in the internal or external memory for output. For the details, refer to the introduction

in “To Select Waveform”.

Modulation

The built-in signal generator of MSO1000Z/DS1000Z series oscilloscope supports amplitude modulation (AM) and frequency modulation (FM). The modulated waveform consists of the carrier waveform and modulating waveform. The carrier waveform is the waveform output from the signal generator and the modulating signal can be the built-in sine, square, ramp waveform or noise signal.

Press Src1Mod to open the source 1 modulation setting menu. Press Modulation to turn on or off the modulation function. Press Mod.Type to set the modulation type of the current signal to “AM” or “FM”. Then, you can set the modulation parameters according to the selected modulation type.

MSO1000Z/DS1000Z User’s Guide 13-15

RIGOL

Chapter 13 Signal Source

AM

AM (Amplitude Modulation), namely the amplitude of the carrier waveform changes with the amplitude of the modulating waveform, as shown in the figure below.

Modulating Waveform

Carrier Waveform

Modulated Waveform

Figure 13-10 Amplitude Modulation

1. Select the Carrier Waveform

Press Src1Conf to enter the waveform setting interface. Press Wave to select the desired carrier.

Note: When “Pulse”, “DC” or “Noise” is selected, the modulation menu will be hidden automatically.

2. Set the Carrier Parameters

After selecting the desired carrier, you can refer to the introduction in

Parameter Setting Method” to set the carrier parameters (frequency,

amplitude, etc).

3. Select the Modulating Waveform

Press Src1Mod to open the source 1 modulation setting menu. Press Shape to select the desired modulating waveform, including sine, square, triangle waveform and noise.

4. Set the Modulation Frequency

Press Frequency to set the frequency of the modulating waveform. The range is from 1 Hz to 50 kHz. For the setting method, please refer to the introduction in

Parameter Setting Method”.

13-16 MSO1000Z/DS1000Z User’s Guide

Chapter 13 Signal Source

RIGOL

5. Set the Modulation Depth

The modulation depth refers to the strength of the AM and is expressed in percentage. Press Depth to set the modulation depth of the modulating waveform. The range is from 0% to 120%. When it is set to 0%, the output amplitude is half of the carrier amplitude. When it is set to 100%, the output amplitude equals the carrier amplitude. When it is set to a value greater than

100%, envelop distortion will occur which must be avoided in actual circuit; at this point, the output of the instrument will not exceed 5 Vpp (the load is 50 Ω).

FM

FM (Frequency Modulation), namely the frequency of the carrier waveform changes with that of the modulating waveform, as shown in the figure below.

Modulating Waveform

Carrier Waveform

Modulated Waveform

Figure 13-11 Frequency Modulation

1. Select the Carrier Waveform

Press Src1Conf to enter the waveform setting interface. Press Wave to select the desired carrier.

Note: When “Pulse”, “DC” or “Noise” is selected, the modulation menu will be hidden automatically.

2. Set the Carrier Parameters

After selecting the desired carrier, you can refer to the introduction in

Parameter Setting Method” to set the carrier parameters (frequency,

amplitude, etc).

3. Select the Modulating Waveform

Press Src1Mod to open the source 1 modulation setting menu. Press Shape to select the desired modulating waveform, including sine, square, triangle waveform and noise.

MSO1000Z/DS1000Z User’s Guide 13-17

RIGOL

Chapter 13 Signal Source

4. Set the Modulation Frequency

Press Frequency to set the frequency of the modulating waveform. The range is from 1 Hz to 50 kHz.

5. Set the Modulation Deviation

Press Deviation to set the deviation of the frequency of the modulating waveform relative to the carrier frequency. The range is from 0 Hz to the current carrier frequency and the sum of the frequency deviation and carrier frequency cannot be greater than the current carrier frequency upper limit. For the setting

method, please refer to the introduction in “Parameter Setting Method”.

13-18 MSO1000Z/DS1000Z User’s Guide

Chapter 14 Store and Recall

RIGOL

Chapter 14 Store and Recall

Users can save the current settings, waveforms, screen image and parameters of the oscilloscope in internal memory or external USB mass storage device (such as USB storage device) in various formats and recall the stored traces, settings or waveforms when needed.

The contents of this chapter:

 Storage System

 Storage Type

 Internal Storage and Recall

 External Storage and Recall

 Disk Management

 Factory

MSO1000Z/DS1000Z User’s Guide 14-1

RIGOL

Chapter 14 Store and Recall

Storage System

Press Storage to enter the store and recall setting interface.

The internal memory (Local Disk) of this oscilloscope is 90.5 MByte. It provides a USB

Host interface at the front panel to connect USB storage device for external storage.

The USB storage device connected is marked as “Disk D”.

Storage Type

Press StorageStorage to select the desired storage type. The default is

“Picture”. The storage and recall descriptions of each type are as follows.

1. Picture

Save the screen image to external memory in “.png”, “.bmp8”, “.bmp24”, “.jpeg” or “.tiff” format. You can specify the file name and saving directory and save the corresponding parameter file (*.txt) under the same directory using the same

file name. This parameter file is the same as the “Parameters” function. The

recall of image and parameter files is not supported.

After selecting this type:

Press PicType to select the desired storage format.

Press Param to enable or disable the parameter save function.

Press Invert to enable or disable the invert function.

Press Color to select the desired storage color. It can be set to “Gray” or “Color”.

Tip

After a USB storage device is connected, press at the front panel to quickly save the current screen image under the root directory of the USB storage device.

2. Traces

Save the waveform data to external memory in “*.trc” format. The data of all the channels turned on can be saved in the same file. At recall, the data will be displayed on the screen directly.

3. Waves

Save the waveform data to external memory in “*.wfm” format. The stored files contain the waveform data of the four analog channels and the main setting information of the oscilloscope and all the data can be recalled.

14-2 MSO1000Z/DS1000Z User’s Guide

Chapter 14 Store and Recall

RIGOL

4. Setups

Save the settings of the oscilloscope to internal or external memory in “*.stp” format. The stored settings can be recalled.

5. CSV

Save the waveform data displayed on the screen or of the specified channels in external memory in a single “*.csv” file. You can specify the file name and the saving directory and save the corresponding parameter file (*.txt) under the same directory using the same file name. This parameter file is the same as the

Parameters” function. The recall of CSV and parameter files is not supported.

After selecting this type:

Press DataSrc to select “Screen” or “Memory”. After selecting “Memory”, press

Channel to select the desired channel ( Note: Only channels currently enabled can be selected).

Press Param to enable or disable the parameter save function.

Press Sequence to choose whether to add the sequence number for the CSV file. You can select “ON” or “OFF” and the default is “ON” (namely add the sequence number for the CSV file).

6. Parameters

Save the parameters of the screen waveform to external memory in “*.txt” format. The parameter storage contain the system information (such as the software and hardware version numbers, time, date, etc) and the current related setting information of the instrument (such as vertical, horizontal, trigger and channel switching state, etc).

MSO1000Z/DS1000Z User’s Guide 14-3

RIGOL

Chapter 14 Store and Recall

Internal Storage and Recall

“Setups” and “Param” (internal recall is not supported) in Storage support internal storage. In the following part, the storage and recall method and procedures of internal storage are introduced.

1. Save the specified oscilloscope setting in internal memory.

1) Connect a signal to the oscilloscope and obtain stable display.

2) Press StorageStorage to select “Setups” or “Param”and press Save and use to select “Local Disk” (blue shading) and press down to open the local disk.

3) Press New File and use the pop-up keyboard to create a new file name.

For the details, please refer to the introduction in “To Create a New File

or Folder”. If this type of file is saved in the internal memory, use

to select this file and then Save and Delete are illuminated. At this point, pressing down Save will execute the storage operation and the original file will be overwritten; pressing Delete will delete the original file. Use to select and then press down to return to the previous directory.

2. Load the specified type of file in internal memory.

1) Press StorageStorage to select “Setups” and then press Load and use

to select “Local Disk” and then press down to open the local disk.

2) If this type file is saved in the internal memory, use to select the desired file and press Load to load the file selected.

14-4 MSO1000Z/DS1000Z User’s Guide

Chapter 14 Store and Recall

RIGOL

External Storage and Recall

Before using external storage and recall, make sure that a USB storage device is connected correctly. External storage supports all the types of files in Storage but in recall, “Picture”, “CSV” and “Param” are not supported. In the following part, we will take “Traces” as an example to introduce the storage and recall method and procedures.

1. Save the specified type of file in the external USB storage device

1) Connect a signal to the oscilloscope and obtain stable display.

2) Press StorageStorage to select “Traces”. Press Save and use to select “Disk D” and then press down to open the USB storage device.

3) Use to select the desired storage position. The file can be stored under the root directory or in a certain folder under the root directory of the USB storage device.

Note: Pressing NewFolder can create a new folder. For the detailed,

please refer to the introduction in “ To Create a New File or Folder”.

4) After the storage position is selected, press New File to create a new filename using the pop-up keyboard. For the details, please refer to the

descriptions in “To Create a New File or Folder”. If this type file is saved

in the USB storage device, use to select this file and then Save and

Delete are illuminated. At this point, pressing down Save will execute the storage operation and the original file will be overwritten; pressing Delete will delete the original file. Use to select

to return to the previous directory.

5) Press OK to execute the saving operation.

and then press down

2. Load the specified type of file in the external USB storage device

1) Press StorageStorage to select “Traces” and then press Load and use

to select “Disk D”. Press down to open the USB storage device.

2) If this type file is saved in the USB storage device, use to select the desired file and press Load to load the file selected.

MSO1000Z/DS1000Z User’s Guide 14-5

RIGOL

Chapter 14 Store and Recall

Disk Management

Press StorageDiskManage to turn on the disk management interface as shown

in Figure 14-1 and use

to select the desired disk. The disk currently selected is displayed in blue shading. Press down to open the disk selected.

Execute the following operations through the disk management menu:

 To Select File Type

 To Create a New File or Folder

 To Delete a File or Folder

 To Rename a File or Folder

 To Clear the Local Memory

Figure 14-1 Disk Management Interface

To Select File Type

Except the file types in Storage, the oscilloscope can also display, save or read some files for advanced applications such as mask file of the Pass/Fail test (*.pf), waveform record file (*.rec), upgrade file (.gel) and reference waveform file (*.ref), etc.

Press StorageDiskManageFile Type to select the desired file type. The default is “*.*”. Under the current directory, only files of which the suffix of the file name matches with the file type selected will be displayed in the current disk.

14-6 MSO1000Z/DS1000Z User’s Guide

Chapter 14 Store and Recall

RIGOL

To Create a New File or Folder

This operation is only valid in external storage. Before using external disk, make sure that a USB storage device is connected correctly.

First, press StorageDiskManage and use to select and open the external disk (“Disk D”). Then, select the desired file type and select the desired directory under which to create a new file or folder. The default is the root directory of the USB storage device.

Press New File or New Folder to turn on the interface as shown in the figure below.

Figure 14-2 To Create a New File or Folder

This oscilloscope supports Chinese/English input method. The file name or folder name can contain letters, numbers, underscores, spaces and Chinese characters and the length of the characters is limited to 64 bytes. The following part introduces how to input a file name or folder name using Chinese/English input method.

Operation Tip

During the name input, use the menu softkeys to select different operation areas, then turn to select the desired content and press down to input the content selected.

MSO1000Z/DS1000Z User’s Guide 14-7

RIGOL

Chapter 14 Store and Recall

English Input Method

For example, create a file or folder with the name “Filename”.

1. Press Keyboard.

1) Use to select English input method “En” and uppercase input state “A a

”.

2) Use to input the letter “F”. If the input is wrong, press Delete to delete the character input.

3) Use to select lowercase input state “a

A

”.

4) Use to input the remaining letters “ilename”.

Name Input Area Keyboard Upper-lower Case Switch Input Method Switch

2. During the name input, you can press Name to select the “Name Input Area” and use to move the cursor, then press Delete to delete the characters at the left of the cursor one by one.

3. After finishing the input, press OK and the oscilloscope will create a folder or a specified type of file with this name under the current directory.

14-8 MSO1000Z/DS1000Z User’s Guide

Chapter 14 Store and Recall

RIGOL

Chinese Input Method

For example, create a file or folder with the name “文件名”.

1. Press Keyboard.

1) Use to select Chinese input method “中”.

Note: Chinese is added in the menu items at the right of the screen.

2) Use to input the pinyin “wen”. If the input is wrong, press Delete to delete the pinyin input. After “wen” is input, a series of Chinese characters appear in the “Chinese Chraracter Selecting Area”.

3) Press Chinese and use to select and input “文”.

4) Use the same method to input “件” and “名”.

Pinyin Input Area Chinese Character Selecting Area Input Method Switch

Name Input Area Keyboard Upper-lower Case Switch

2. During the name input, you can press Name to select the “Name Input Area” and then press Delete to delete the Chinese characters at the left of the cursor one by one.

3. After finishing the input, press OK and the oscilloscope will create a folder or a specified type of file with this name under the current directory.

MSO1000Z/DS1000Z User’s Guide 14-9

RIGOL

Chapter 14 Store and Recall

To Delete a File or Folder

Folder operation is valid only in external storage. Before using the external disk, make sure that a USB storage device is connected correctly.

1. Delete a file in internal memory.

1) Press StorageDiskManage and use to select and open the local disk (“local Disk”).

2) Press File Type to select the desired type of file to be deleted.

3) Use to select the desired file to be deleted.

4) Press DeleteOK to delete the file selected.

2. Delete a file or folder in external memory.

Press StorageDiskManage and use to select and open the external disk (“Disk D”). Use to select the file (or folder) to be deleted and then press

DeleteOK to delete the selected file (or folder).

To Rename a File or Folder

Rename operation is valid only in external storage. Before using the external disk, make sure that a USB storage device is connected correctly.

Press StorageDiskManage and use to select and open the external disk

(“Disk D”). Use to select the desired file or folder to rename and then press

Rename to turn on the rename interface. For specific operations, please refer to the

descriptions in “To Create a New File or Folder”.

To Clear the Local Memory

Press StorageDiskManage and select “Local Disk”. Press FlashEraseOK to delete all the files and setting stored in the local memory. At the same time, the instrument will be restored to the default setting.

14-10 MSO1000Z/DS1000Z User’s Guide

Chapter 14 Store and Recall

RIGOL

Factory

Press Storage Default to return the oscilloscope to its factory state (refer to the table below).

Table 14-1 Factory

Parameter Factory

Horizontal Setting (HORIZONTAL)

Vertical Setting (VERTICAL)

Acquisition Setting (Acquire)

Trigger Setting (TRIGGER)

Display Setting (Display)

Signal Source (Source)

Cursor Setting (Cursor)

Storage Setting (Storage)

Utility Function Setting (Utility)

Math Operation Setting (MATHMath)

LA

Protocol Decoding (MATH Decode 1/Decode 2)

Reference Waveform Setting (REF)

Horizontal Setting (HORIZONTAL)

Horizontal Time Base 1 μs

Horizontal Offset

Delayed Sweep

Time Base Type

0 s

OFF

YT

Vertical Setting (VERTICAL)

Vertical Scale

Vertical Offset

CH1 Switch

CH2 Switch

CH3 Switch

CH4 Switch

Channel Coupling

Bandwidth Limit

Probe Ratio

1 V

0.00 V

ON

OFF

OFF

OFF

DC

OFF

Invert

Amplitude Scale

Channel Unit

Display Label

10X

OFF

Coarse

[V]

Template Label

Delay Calibration

OFF

CH1

0.00 s

Acquisition Setting (Acquire)

Acquisition Mode Normal

MSO1000Z/DS1000Z User’s Guide 14-11

RIGOL

Sin(x)/x

Memory Depth

Anti-Aliasing

ON

Auto

OFF

Trigger Setting (TRIGGER)

Trigger Type Edge

CH1 Source

Slope

Trigger Mode

Trigger Coupling

Trigger Holdoff

Noise Reject

Rising Edge

Auto

DC

16 ns

OFF

Display Setting (Display)

Display Type

Persistence Time

Vectors

Min

Waveform Intensity 60%

Screen Grid

50% Brightness

Signal Source (Source) [1]

Source 1 Switch

Source 2 Switch

OFF

OFF

Status Display OFF

Source 1 / Source 2 Setup

Waveform

Channel Switch

Frequency

Amplitude

Offset

Start Phase

Modulation

Impedance

Sine

OFF

1.00 kHz

5.000 V

0.000 uV

0.0°

OFF

HighZ

Cursor Setting (Cursor)

Mode

Manual

OFF

Select

Source

CursorA

CursorB

Vertical Unit

Horizontal Unit

Track

CH1 s

-4.000 μs

4.000 μs

Source

14-12

Chapter 14 Store and Recall

MSO1000Z/DS1000Z User’s Guide

Chapter 14 Store and Recall

A+B

Operation

Source A

Source B

Offset

Scale

A-B

Operation

Source A

Source B

Offset

Scale

CursorA Source

CursorB Source

CursorA

CursorB

XY

AX

BX

AY

BY

CH1

CH1

-4.000 μs

4.000 μs

2.000 V

-2.000 V

2.000 V

-2.000 V

Storage Setting (Storage)

Storage Type Picture

Utility Function Setting (Utility)

OFF Sound

Pass/Fail Test

Enable Test

Source

Operate

Mask Range

X Mask

Y Mask

Statistics Display

Stop On Fail

Output

OFF

CH1

OFF

Screen Region

0.24 div

0.48 div

OFF

OFF

Aux Out

System Setting

Vertical Reference

Power Set

OFF

Ground

Last

Math Operation Setting (MATHMath)

OFF

CH1

CH1

0.00 V

1.00 V

OFF

CH1

CH1

0.00 V

1.00 V

MSO1000Z/DS1000Z User’s Guide

RIGOL

14-13

RIGOL

A×B

Operation

Source A

Source B

Offset

Scale

A/B

Operation

Source A

Source B

Offset

Scale

FFT

Operation

Source

Center Frequency

Horizontal Scale

Offset

Scale

Window Function

View

Unit

A&&B

Operation

Source A

Source B

Offset

Scale

Threshold A

Threshold B

A||B

Operation

Source A

Source B

Offset

Scale

Threshold A

Threshold B

A^B

Operation

Source A

Source B

Offset

Scale

14-14

OFF

CH1

CH1

0.00 U

2.00 U

OFF

CH1

CH1

0.00 U

50.0 mU

OFF

CH1

5.00 MHz

5.00 MHz

0.00 dBV

10.0 dBV

Rectangle

Half dB/dBm

OFF

D0

D1

0.00 V

1.00 V

0.00 V

0.00 V

OFF

D0

D1

0.00 V

1.00 V

0.00 V

0.00 V

OFF

D0

D1

0.00 V

1.00 V

Chapter 14 Store and Recall

MSO1000Z/DS1000Z User’s Guide

Chapter 14 Store and Recall

Offset

Scale

Sqrt

Operation

Source

Offset

Scale

Lg

Operation

Source

Offset

Scale

Ln

Threshold A

Threshold B

!A

Operation

Source A

Offset

Scale

Threshold A

Intg

Operation

Source

Offset

Scale

Diff

Operation

Source

Operation

Source

Offset

Scale

Exp

Operation

Source

Offset

Scale

Abs

Operation

Source

Offset

Scale

0.00 V

0.00 V

OFF

D0

0.00 V

1.00 V

0.00 V

OFF

CH1

0.00 U

10.0 μU

OFF

CH1

0.00 V/s

50.0 MV/s

OFF

CH1

0.00 U

2.00 U

OFF

CH1

0.00 U

500 mU

OFF

CH1

0.00 U

1.00 U

OFF

CH1

0.00 U

5.00 U

OFF

CH1

0.00 V

50.0 V

MSO1000Z/DS1000Z User’s Guide

RIGOL

14-15

RIGOL

LA [2]

Current Channel

D7-D0

D15-D8

D0

OFF

OFF

Wave Size S

ReOrder

Delay Calibration

Switch Channel

Channel Selection

Threshold Settings

Low Type

D7-D0

High Type

D0-D15

0.00 s

D7-D0

TTL

1.40 V

TTL

1.40 V D15-D8

Label

Select Channel

Preset Label

D0

ACK

Protocol Decoding (MATH Decode 1/Decode 2)

Parallel

OFF

ASC

Decoder

Decode

Format

Parallel

CLK

Edge

Bus Width

Bit X

Channel

RS232

TX

Polarity

Baud

RX

Order

Data Bits

Stop Bit

Parity

I2C

CLK

DATA

Address Mode

SPI

CLK

MISO

CH1

Rising Edge

8

0

D0

CH1

9600

OFF

LSB

8

1

None

CH1

CH2

Normal

CH1

OFF

Chapter 14 Store and Recall

14-16 MSO1000Z/DS1000Z User’s Guide

Chapter 14 Store and Recall

MOSI

Mode

Timeout

Edge

Polarity

Width

Order

Decode Configuration

Label

8

MSB

Line

Format

Endian

Width

Data Source

CH2

Timeout

1.00 us

Rising Edge

ON

ON

ON

OFF

OFF

Reference Waveform Setting (REF)

Channel Setting

Current Channel

Source

Offset

Scale

Color

Note

Note

Traces

Ref1

Ref1

CH1

0.00 nV

100 mV

Light Gray

[1] : Only applicable to MSO1000Z-S and DS1000Z-S models digital oscilloscopes.

[2] : Only applicable to MSO1000Z and MSO1000Z-S models digital oscilloscopes.

RIGOL

MSO1000Z/DS1000Z User’s Guide 14-17

Chapter 15 Accessibility Setting

Chapter 15 Accessibility Setting

The contents of this chapter:

 Remote Interface Configuration

 System-related

RIGOL

MSO1000Z/DS1000Z User’s Guide 15-1

RIGOL

Chapter 15 Accessibility Setting

Remote Interface Configuration

MSO1000Z/DS1000Z provides three remote control interfaces: USB Device (TMC and

SICD static image capture devices share this port), LAN and GPIB (expanded from the USB HOST interface by the USB-GPIB interface converter). The priority order of the three ports is USB Device, LAN and GPIB. You can press UtilityIO Setting

RemoteIO to select the desired communication interface according to the particular condition.

 When only the USB-Device interface is connected and press UtilityIO

SettingUSB Device ”Computer”, the system will select the USB-Device interface automatically as the only valid port no matter whether the LAN and

GPIB interfaces are connected or not. At this point, the RemoteIO menu is disabled as shown in figure (a).

 When only the GPIB or LAN interface is connected, the system will automatically detect the corresponding port of which the display status is “ON”. At this point, pressing the correnponding key can turn on or off the remote connection. The

RemoteIO menu is as shown in figure (b) and figure (c).

 When GPIB and LAN are connected at the same time, the system will set LAN

(the display status is “ON”) as the current control port by default. At this point,

LAN and GPIB are available. You can select the communication interface according to the needs. The RemoteIO menu is as shown in figure (d).

Figure (a) Figure (b)

15-2

Figure (c) Figure (d)

MSO1000Z/DS1000Z User’s Guide

Chapter 15 Accessibility Setting

RIGOL

LAN Configuration

Press UtilityIO SettingLAN Conf. to turn on the LAN setting interface. You can view the network connection status and configure the network parameters.

Network Status

Current IP

Config Type

MAC Address

VISA Address

IP Config Type

Status

Figure 15-1 LAN Setting Interface

Network Status

Connect the oscilloscope to your local area network using a network cable. The network interface of the oscilloscope is at the rear panel. The oscilloscope will give different prompts according to the current network connection status.

 Net Config Success!

 Acquire IP...

 IP Conflict!

 Unconnected!

 DHCP Fail!

 Read Status Fail!

IP Configuration Type

The configuration type of the IP address can be DHCP, auto IP or static IP. In different IP configuration type, the configuration mode of the network parameters

(such as the IP address) is different.

MSO1000Z/DS1000Z User’s Guide 15-3

RIGOL

Chapter 15 Accessibility Setting

DHCP

Press Configure and use to select “DHCP”. Then press down to select this type. When DHPC is valid, the DHCP server in the current network will assign the network parameters (such as the IP address) for the oscilloscope.

Auto IP

Press Configure and use to select “Auto IP”. Then press down to select this type. When the auto IP type is valid, disable DHCP manually, Gate and DNS are added to the right side of the screen and users can define the gateway and DNS server address of the oscilloscope. In auto IP mode, the oscilloscope will get the IP address ranging from 169.254.0.1 to 169.254.255.254 and the subnet mask

255.255.0.0 automatically according to the current network configuration.

Static IP

Press Configure and use to select “Static IP”. Then press down to select this type. When this type is valid, disable DHCP and auto IP manually, IP Address,

Mask, Gate and DNS are added to the right of the screen. At this point, users can define their own network parameters (such as the IP address) of the oscilloscope.

1. Set the IP Address

The format of IP address is nnn.nnn.nnn.nnn. The range of the first nnn is from

0 to 255 (except 127); wherein, the valid range is from 0 to 223. The ranges of the other three nnn are from 0 to 255. You are recommended to ask your network administrator for an IP address available.

Press IP Address and use to input the desired IP address. This setting will

be saved in the non-volatile memory and if “Power-off Recall” is set to “Last”,

DHCP and Auto IP will be “Off” and the oscilloscope will load the preset IP address automatically at the next power-on.

2. Set the Subnet Mask

The format of the subnet mask is nnn.nnn.nnn.nnn; wherein, the range of the nnn is from 0 to 255. You are recommended to ask your network administrator for a subnet mask available.

Press Mask and use to input the desired subnet mask. This setting will be

saved in the non-volatile memory and if “Power-off Recall” is set to “Last”,

15-4 MSO1000Z/DS1000Z User’s Guide

Chapter 15 Accessibility Setting

RIGOL

DHCP and Auto IP will be “Off” and the oscilloscope will load the preset subnet mask automatically at the next power-on.

Set the Gate

You can set this paramter in Auto IP and Static IP mode.

The format of the gate is nnn.nnn.nnn.nnn; wherein, the range of the first nnn is from 0 to 223 (except 127) and the ranges of the other three nnn are from 0 to 255.

You are recommended to ask your network administrator for a gate address available.

Press Gate and use to input the desired gate address. This setting will be saved

in the non-volatile memory and if “Power-off Recall” is set to “Last”, DHCP and

Auto IP will be “Off” and the oscilloscope will load the preset gate address automatically at the next power-on.

Set the Domain Name Server

You can set this paramter in Auto IP and Static IP mode.

The address format of the domain name server is nnn.nnn.nnn.nnn; wherein, the range of the first nnn is from 0 to 223 (except 127) and the ranges of the other three nnn are from 0 to 255. You are recommended to ask your network administrator for an address available. press DNS and use to input the desired address. Generally, users do not need to set the DNS, therefore this parameter setting can be ignored.

Tips

 When the three IP configuration types are all turned on, the priority of the parameter configuration from high to low is “DHCP”, “Auto IP” and “Static IP”.

 The three IP configuration types cannot be all turned off at the same time.

Apply the Network Parameter Setting

Press Apply to validate the current network parameter setting.

Initialize the Network Parameters

Press Initialize to return the network parameters to the default state.

MSO1000Z/DS1000Z User’s Guide 15-5

RIGOL

Chapter 15 Accessibility Setting

MAC Address

For each oscilloscope, the MAC address is unique. When attributing IP address for the oscilloscope, the MAC address is usually used to identify the instrument.

VISA Address

Display the VISA address currently used by the oscilloscope.

USB Device

Press UtilityIO SettingUSB Device to select the device type (“Computer” or “PicBridge”) connected to the USB device interface and the default is “Computer”.

When “Computer” is selected, the oscilloscope can communicate with the PC. When

“PictBridge” is selected, you can print the content displayed on the screen via a

PicBridge printer.

GPIB Address

When using the GPIB mode to control the oscilloscope, you need to extend a GPIB interface for the oscilloscope using the USB-GPIB interface converter (need to be ordered separately).

To set the GPIB address of this interface, press UtilityIO SettingGPIB and use to input the desired address. The default is 1 and the range is from 1 to 30.

15-6 MSO1000Z/DS1000Z User’s Guide

Chapter 15 Accessibility Setting

RIGOL

System-related

Sound

When the sound is enabled, you can hear the sound of the beeper when you press a function key or a menu softkey or when the prompt message pops up.

Press UtilitySound to select (off) or (on). The default is off.

Language

This oscilloscope supports multiple languages. Press UtilityLanguage and use

to select the desired language. Then press down to select the language. This oscilloscope will display the corresponding menus, help and prompt message using this language.

System Information

Press UtilitySystemSystem Info to view the system information of your oscilloscope. The system information includes the manufacturer, model, serial number as well as software and hardware version numbers.

Vertical Reference

When rotating VERTICAL SCALE to change the vertical scale of the analog channel, you can expand and compress the signal vertically around the screen center or signal ground level position.

Press UtilitySystemVerticalRef. to select “Center” or “Ground”. The default is “Ground”.

 Center: when the vertical scale is changed, the waveform will be expanded or compressed around the screen center.

 Ground: when the vertical scale is changed, the waveform will be expanded or compressed around the signal ground level position.

Power-off Recall

You can set the system configuration to be recalled when the oscilloscope is powered on again after power-off. Press UtilitySystemPower Set to select “Last” or

“Default”.

 Last: return to the setting of the system at last power-off.

 Default: return to the factory setting of the system.

MSO1000Z/DS1000Z User’s Guide 15-7

RIGOL

Chapter 15 Accessibility Setting

Self-calibration

The self-calibration program can quickly make the oscilloscope reach the best working state to get the most precise measurement values. You can perform self-calibration at any time especially when the change of the environment temperature is up to or more than 5℃. Make sure that the oscilloscope has been warmed up or operated for more than 30 minutes before the self-calibration.

Disconnect all the input channel connections, and then press UtilitySelf-Cal and the self-calibration interface as shown in the figure below is displayed.

Figure 15-2 Self-calibration Interface

Press Start and the oscilloscope will start to execute the self-calibration program.

Press Exit to give up the self-calibration operation at any time and return to the previous menu.

Note: Most of the keys are disabled during the self-calibration.

Print Setting

PictBridge is a new print standard. If your oscilloscope and the printer both comply with the PictBridge standard, you can connect the oscilloscope to the printer using a USB cable to print the screen image directly. Devices complies with the PictBridge standard always have the icon as shown in the figure on the right side.

15-8 MSO1000Z/DS1000Z User’s Guide

Chapter 15 Accessibility Setting

RIGOL

MSO1000Z/DS1000Z digital oscilloscope supports the PictBridge printer function.

Printers that support PictBridge can be connect directly via the USB Device connector at the rear panel of the oscilloscope. You can set the printer parameters using the oscilloscope to output the corresponding picture.

After connecting the printer, you need to first set the USB device type of the oscilloscope. Press UtilityIO Setting USB Device to select “PictBridge”.

Then, press UtilityPrint Set to set the print parameters.

1. Print

After setting the print parameters, press this key or press panel to execute the print operation.

at the front

2. Continue

When the print is paused, press this key to resume the unfinished print.

Note: This key is only available after the print is paused.

3. Abort

Pressing this key can stop the current print via the oscilloscope.

Note: This key is only available during the print process.

4. Status

Press this key and pop-up message box will be displayed on the screen showing the current print status.

5. Print Range

Press this key to set the print range to “Screen” or “Wave” and the default is

“Screen”.

 Screen: print the whole screen image.

 Wave: only print the waveform area.

6. Palette

Press this key to set the print color to “Gray Scale” or “Color” and the default is

“Color”.

7. Paper Size

Press this key to open the size list and rotate the multifunction knob to select the desired paper size. You can also press this key continuously to switch among the options. The sizes include “Default”, “A2”, “A3”, “A4”, “A5”, “A6” or

“B5”.

Note: The paper sizes available are related to the property of the printer connected. Sizes not supported by the printer cannot be selected.

8. File Type

Press this key to open the file type list and rotate the multifunction knob to

MSO1000Z/DS1000Z User’s Guide 15-9

RIGOL

Chapter 15 Accessibility Setting

select the desired file type. You can also press this key continuously to switch among the options. The file types include “Default”, “Jpeg” or “Bmp”.

Note: The file types available are related to the property of the printer connected. File types not supported by the printer cannot be selected.

9. Print Quality

Press this key to open the print quality list and rotate the multifunction knob to select the desired print quality. You can also press this key continuously to switch among the options. The print qualities include “Default”, “Normal”,

“Draft” or “Fine”.

Note: The print qualities available are related to the property of the printer connected. Print qualities not supported by the printer cannot be selected.

10. Copies

Press this key to set the number of copies to be printed. Rotate the multifunction knob to select the desired number of copies. The range is from 1 to 999.

11. Invert

Press this key to set the print color invert to “ON” or “OFF” and the default is

“OFF”.

Aux Output

Users can set the type of the signal output from the [Trigger Out] connector at the rear panel.

Press UtilityAux Out to select the desired output type.

1. TrigOut

After this type is selected, the oscilloscope outputs a signal that can reflect the current capture rate of the oscilloscope at each trigger. Connect this signal to a waveform display device, measure the signal frequency and the measurement result equals the current acquisition rate.

2. PassFail

After this type is selected, the oscilloscope will output a positive pulse from this connector when failed waveforms are detected in the pass/fail test. The instrument will output low level continuously if no failure is detected.

15-10 MSO1000Z/DS1000Z User’s Guide

Chapter 15 Accessibility Setting

RIGOL

Option Management

This oscilloscope provides multiple options to fulfill your measurement requirements.

Please contact your RIGOL sales representative or RIGOL technical support to order the corresponding options. You can view the options currently installed on the oscilloscope or activate the newly bought option serial number through this menu.

Press UtilityOptionsInstalled to view the options currently installed on the oscilloscope and their information. Press Setup to enter the serial number activation operation menu.

 Editor: press this softkey to select “ON” to turn on the serial number input interface as shown in the figure below. Use to select the character on the virtual keyboard and press down the knob to input the character.

Serial Number Input Area Character Selecting Area

Figure 15-3 Option Serial Number Input Interface

 Backspace: press this softkey to delete the characters in the “Serial Number

Input Area” from the right to the left.

 Clear: press this softkey to clear all the characters in the “Serial Number Input

Area”.

 Apply: press this softkey and the oscilloscope will activate the corresponding option using the serial number currently input.

MSO1000Z/DS1000Z User’s Guide 15-11

RIGOL

Chapter 15 Accessibility Setting

Auto Options

As mentioned before, pressing AUTO at the front panel can enable the waveform automatic setting function. The oscilloscope will automatically adjust the vertical scale, horizontal time base and trigger mode according to the input signal to realize optimum waveform display. The oscilloscope allows users to set the related parameters of the waveform auto setting function.

Press UtilityAuto Options to enter the auto options setting menu. You can set the following parameters.

 Press Lock to lock AUTO. At this point, this key is disabled.

Note: You can only unlock the key through the remote command

(:SYSTem:AUToscale 1). For the remote command, please refer to

MSO1000Z/DS1000Z Programming Guide

.

 Pres Pk.Pk to enable or disable the peak-peak priority function. When “ON” is selected, the peak-peak value is displayed on the screen with the best scale. For signals with offset, this function is more practical for observing signal variations.

 Press CH to select the channel to use the AUTO operation. You can select

“OPENed” (the channel turned on currently) or “ALL”. The default is “ALL”.

Note: If no channel is turned on currently, the instrument will execute the AUTO operation for all channels.

 Press Menu Hold to enable or disable the menu hold function. When “ON” is

selected, the screen does not display the menus as shown in Figure 6-2 (namely

the instrument holds the current menus) after the AUTO operation is successful.

 Press Overlay to enable or disable the overlay display function. When “ON” is selected and signals are input through multiple channels, the signals of multiple channels are overlayed on the screen. Each channel can occupy 8 grids of vertical range; amplitude resolution is small and the trigger is more stable.

When “OFF” is selected, the signals of multiple channels are displayed on the screen respectively. Each channel can only copy about 2 grids of vertical range; amplitude resolution is large and the trigger is instable.

 Press Coupling to enable or disable the coupling hold function. When “ON” is selected, the channel coupling settings of the signal detected will be held.If the channel is set to DC coupling, the instrument keeps the DC coupling after the signal is detected; if the channel is set to AC coupling, the instrument keeps the

AC coupling after the signal is detected; if the channel is set to GND, DC coupling is used by default. When “OFF” is selected, DC coupling is used by default after the signal is detected.

Key Lock

Press UtilityKeyLockLock to lock all softkeys and buttons of the oscilloscope except the Unlock softkey. Press Unlock to unlock these keys.

15-12 MSO1000Z/DS1000Z User’s Guide

Chapter 16 Remote Control

RIGOL

Chapter 16 Remote Control

MSO1000Z/DS1000Z oscilloscope can be controlled remotely mainly through the following two methods.

User-defined programming

Users can program and control the oscilloscope by using the SCPI (Standard

Commands for Programmable Instruments) commands. For more information about the commands and programming, refer to the

Programming Guide

.

Use PC software provided by RIGOL or other manufacturers

Users can use the PC software Ultra Sigma of RIGOL, Measurement &

Automation Explorer of NI (National Instruments Corporation) or Agilent IO

Libraries Suite of Agilent (Agilent Technologies, Inc.) to send commands to control the oscilloscope remotely.

This oscilloscope can communicate with PC through USB, LAN or GPIB (with the

USB-GPIB interface converter provided by RIGOL) instrument bus. This chapter gives a detailed introduction of how to use Ultra Sigma to control the oscilloscope remotely through various interfaces. For the Ultra Sigma software, please contact

RIGOL salesmen or technical support.

The contents of this chapter:

 Remote Control via USB

 Remote Control via LAN

 Remote Control via GPIB

MSO1000Z/DS1000Z User’s Guide 16-1

RIGOL

Chapter 16 Remote Control

Remote Control via USB

1. Connect the device

Connect the USB Device interface of the oscilloscope at the rear panel with the

USB HOST interface of your PC using a USB cable

2. Install the USB driver

This oscilloscope is a USB-TMC device. Assuming that your PC has already been installed with Ultra Sigma, after you connect the oscilloscope to the PC and turn both on for the first time (the oscilloscope is automatically configured to USB interface), the Found New Hardware Wizard as shown in the figure below is displayed on the PC. Please install the “USB Test and Measurement Device (IVI)” driver following the directions in the wizard. The steps are as follows.

2

1

16-2 MSO1000Z/DS1000Z User’s Guide

Chapter 16Remote Control

3

4

5

6

RIGOL

MSO1000Z/DS1000Z User’s Guide 16-3

RIGOL

Chapter 16 Remote Control

7

3. Search for device resource

Start up Ultra Sigma and the software will automatically search for the oscilloscope resources currently connected to the PC. You can also click

to search for the resources.

16-4 MSO1000Z/DS1000Z User’s Guide

Chapter 16Remote Control

RIGOL

4. View the device resource

The resources found will appear under the “RIGOL Online Resource” directory and the model number and USB interface information of the instrument will also be displayed as shown in the figure below.

5. Communication test

Right click the resource name “MSO1104Z

(USB0::0x1AB1::0x04CE::DS1A0000000001::INSTR)” to select “SCPI Panel

Control” to turn on the remote command control panel (as shown in the figure below) through which you can send commands and read data.

MSO1000Z/DS1000Z User’s Guide 16-5

RIGOL

Chapter 16 Remote Control

Remote Control via LAN

1. Connect the device

Connect the oscilloscope to your LAN using a network cable.

2. Configure network parameters

Configure the network parameters of the oscilloscope according to the

description in “LAN Configuration”.

3. Search for device resource

Start the Ultra Sigma, click local network.

and the window as shown below is displayed. Click “Search” to search for the instrument resource connected to the

The name of the instrument resource found will be displayed in the window at the right side, as shown in the figure below. Click “OK” to add the instrument resource.

4. View device resource

The resources found will appear under the “RIGOL Online Resource” directory as shown in the figure below.

16-6 MSO1000Z/DS1000Z User’s Guide

Chapter 16Remote Control

RIGOL

5. Communication test

Right click the resource name “MSO1104Z (TCPIP::172.16.3.94::INSTR)” to select “SCPI Panel Control” to turn on the remote command control panel (as shown in the figure below) through which you can send commands and read data.

6. Load LXI webpage

As this oscilloscope conforms to LXI CORE 2011 DEVICE standards, you can load

LXI webpage through Ultra Sigma (right-click the resource name and select

LXI-Web; or directly input the IP address in the browser). Various important information about the oscilloscope (including the model number, manufacturer, serial number, description, MAC address and IP address) will be displayed on the webpage.

MSO1000Z/DS1000Z User’s Guide 16-7

RIGOL

Chapter 16 Remote Control

Remote Control via GPIB

1. Connect the device

Use the USB to GPIB interface converter to extend a GPIB interface for the oscilloscope. Then connect the oscilloscope to your PC using a GPIB cable

2. Install the driver of GPIB card

Install the driver of the GPIB card which has been connected to the PC correctly.

3. Set the GPIB address

Set the GPIB address of the oscilloscope according to the description in “GPIB

Address”.

4. Search for device resource

Start up the Ultra Sigma and click to open the panel as shown in the figure below. Click “Search” and the software will search for the GPIB instrument resources connected to the PC.

The device resources found are displayed at the right side of the panel, as shown in the figure below. Click “OK” to add the resource.

16-8

If resources cannot be found automatically:

 Select the GPIB card address of the PC from the comboBox of “GPIB0::” and

MSO1000Z/DS1000Z User’s Guide

Chapter 16Remote Control

RIGOL select the GPIB address set in the oscilloscope from the comboBox of

“0::INSTR”.

 Click “Test” to check whether the GPIB communication works normally; if not, please follow the corresponding prompt messages to solve the problem.

5. View device resource

Click “OK” to return back to the main interface of Ultra Sigma. The GPIB instrument resources found will appear under the “RIGOL Online Resource” directory.

6. Communication Test

Right-click the resource name “MSO1104Z (GPIB0::1::INSTR)” to select “SCPI

Panel Control” to turn on the remote command control panel through which you can send commands and read data as shown in the figure below.

MSO1000Z/DS1000Z User’s Guide 16-9

Chapter 17 Troubleshooting

RIGOL

Chapter 17 Troubleshooting

The commonly encountered failures and their solutions are listed below. When you encounter those problems, please solve them following the corresponding steps. If the problem remains still, please contact RIGOL and provide your device information (Utility SystemSystem Info).

1. The screen is still dark (no display) after pressing the power key:

(1) Check whether the power switch is really on.

(2) Check whether the power is correctly connected.

(3) Check whether the fuse is burned out. If the fuse needs to be changed, please use the specified fuse.

(4) Restart the instrument after finishing the above inspections.

(5) If it still does not work correctly, please contact RIGOL.

2. The signal is sampled but no waveform of the signal is displayed:

(1) Check whether the probe is correctly connected to the oscilloscope and item to be tested.

(2) Check whether there are signals generated from the item to be tested (you can connect the probe compensation signal to the problematic channel to determine which has problem, the channel or the item to be tested).

(3) Resample the signal.

3. The tested voltage amplitude is greater or lower than the actual value

(this problem usually occurs when probe is used):

Check whether the probe ratio of the channel complies with the attenuation ratio of the probe.

4. There is waveform display but not stable:

(1) Check the trigger source: press MENUSource in the trigger control area (TRIGGER) at the front panel to check whether the trigger source selected complies with the signal channel actually used.

(2) Check the trigger type: general signals should use “Edge” trigger and video signal should use “Video” trigger. Only when the proper trigger type is used, can the waveform be displayed stably.

(3) Check the trigger level: adjust the trigger level to the middle of the signal.

(4) Change the trigger holdoff setting.

5. No display after pressing RUN/STOP:

Check whether the trigger mode in the trigger control area (TRIGGER) at the front panel is “Normal” or “Single” and whether the trigger level exceeds the waveform range. If yes, set the trigger level to the middle or set MODE to

“Auto”.

Note: Using AUTO could automatically finish the above setting.

MSO1000Z/DS1000Z User’s Guide 17-1

RIGOL

Chapter 17 Troubleshooting

6. The display of waveform is ladder-like:

(1) The horizontal time base might be too low. Increase the horizontal time base to increase the horizontal resolution and improve the display.

(2) If the display type is “Vectors”, the lines between the sample points may cause ladder-like display. Press DisplayType to set the display type to

“Dots” to solve the problem.

7. Fail to connect PC or PictBridge printer through USB:

(1) Press UtilityIO Setting USB Device to check whether the current setting matches the device currently connected.

(2) Check whether the USB cable is correctly connected to the oscilloscope and

PC.

(3) Check whether the USB cable is in good condition. If needed, restart the oscilloscope.

8. The USB storage device cannot be recognized:

(1) Check whether the USB storage device can work normally.

(2) Make sure that the USB storage device being used is flash storage type.

This oscilloscope does not support hardware storage type.

(3) Make sure whether the capacity of the USB storage device is too large. It is recommended that the capacity of the USB storage device being used with this oscilloscope is no larger than 8 GBytes.

(4) Restart the instrument and then insert the USB storage device to check it.

(5) If the USB storage device still cannot be used normally, please contact

RIGOL.

17-2 MSO1000Z/DS1000Z User’s Guide

Chapter 18 Specifications

RIGOL

Chapter 18 Specifications

All the specifications are guaranteed except the parameters marked with “Typical” and the oscilloscope needs to operate for more than 30 minutes under the specified operation temperature.

Sample

Sample Mode Real-time Sample

Real Time

Sample Rate

Analog channe: 1 GSa/s (single-channel), 500 Msa/s

(dual-channel), 250 MSa/s (3/4-channel)

Digital channel: 1 GSa/s (8-channel), 500 MSa/s (16-channel)

Peak Detect

Averaging

Analog channe: 4 ns

Digital channel: 4 ns

After all the channels finish N samples at the same time, N can be 2, 4, 8, 16, 32, 64, 128, 256, 512 or 1024

High Resolution 12 bit (max)

Interpolation Sin(x)/x (optional)

Min Detect

Pulse Width

Digital channel: 10 ns

Memory Depth

Analog channel: standard 12M pts (single-channel), 6M pts

(dual-channel), 3M pts (3/4-channel); optional 24 Mpts

(single-channel), 12 Mpts(dual-channel), 6 Mpts (3/4-channel)

Digital channel: standard 12 Mpts (8-channel)/6

Mpts(16-channel); optional 24 Mpts(8-channel)/12

Mpts(16-channel)

Input

Number of

Channels

MSO1XX4Z/1XX4Z-S: 4 analog channels+16 digital channels

DS1XX4Z/1XX4Z-S: 4 analog channels

Input Coupling DC, AC or GND

Input

Impedance

Analog channe: (1 MΩ±1%) || (15 pF±3 pF)

Digital channel: (100 kΩ±1%) || (8 pF±3 pF)

Probe

Attenuation

Coefficient

Maximum Input

Voltage (1MΩ)

Analog channe: 0.01X to 1000X, in 1-2-5 step

Analog Channel:

CAT I 300 Vrms, CAT II 100 Vrms, Transient Overvoltage 1000

Vpk

With RP2200 10:1 probe: CAT II 300 Vrms

Digital channel: CAT I 40Vrms, Transient Overvoltage 800 Vpk

MSO1000Z/DS1000Z User’s Guide 18-1

RIGOL

Horizontal

Timebase Scale 5 ns/div to 50 s/div

Max Record

Length

Timebase

Accuracy [1]

Clock Drift

Max Delay

Range

24 Mpts (optional)

≤±25 ppm

≤±5 ppm/year

Negative delay: ≥1/2 screen width

Positive delay: 1 s to 5000 s

Timebase Mode YT, XY, Roll

Number of X-Y 1

Waveform

Capture Rate [2]

30,000 wfms/s (dots display)

Vertical

Bandwidth

(-3dB)

Single-shot

Bandwidth

Vertical

Resolution

Vertical Scale

Offset Range

(Probe ratio is

1X)

Bandwidth

Limit [1]

Low Frequency

Response

(AC Coupling,

-3dB)

Calculated Rise

Time [1]

DC Gain

Accuracy [3]

DC Offset

Accuracy

Channel to

Channel

Isolation

MSO/DS 1104Z/1104Z-S: DC to 100 MHz

MSO/DS 1074Z/1074Z-S: DC to 70 MHz

DS1054Z: DC to 50 MHz

MSO/DS 1104Z/1104Z-S: DC to 100 MHz

MSO/DS 1074Z/1074Z-S: DC to 70 MHz

DS1054Z: DC to 50 MHz

Analog channe: 8 bit

Digital channel: 1 bit

1 mV/div to 10 V/div

1 mV/div to 499 mV/div: ±2 V

500 mV/div to 10 V/div: ±100 V

20 MHz

≤5 Hz (on BNC)

MSO/DS 1104Z/1104Z-S: 3.5 ns

MSO/DS 1074Z/1074Z-S: 5 ns

DS1054Z: 7 ns

<10 mV: ±4% full scale

≥10 mV: ±3% full scale

±0.1 div ±2 mV ±1% offset value

DC to maximum bandwidth: >40 dB

Chapter 18 Specifications

18-2 MSO1000Z/DS1000Z User’s Guide

Chapter 18 Specifications

RIGOL

Vertical (Digital Channel)

Threshold

Threshold

Selection

TTL (1.4 V)

5.0 V CMOS (+2.5 V), 3.3 V CMOS (+1.65 V)

2.5 V CMOS (+1.25 V), 1.8 V CMOS (+0.9 V)

ECL (-1.3 V)

PECL (+3.7 V)

LVDS (+1.2 V)

0 V

User

±15.0V, 10 mV step Threshold

Range

Threshold

Accuracy

±(100 mV+3% threshold setting)

Dynamic Range ±10.0 V + threshold

Minimum

Voltage Swing

Vertical

Resolution

Adjustable threshold of 8 channels per group

500 mVpp

1 bit

Trigger

Trigger Level

Range

Trigger Mode

± 5 div from center of the screen

Auto, Normal, Single

Holdoff Range 16 ns to 10 s

75 kHz High Frequency

Rejection [1]

Low Frequency

Rejection [1]

Trigger

Sensitivity [1]

Edge Trigger

75 kHz

1.0 div (below 5 mV or noise rejection is enabled)

0.3 div (above 5 mV and noise rejection is disabled)

Edge Type

Pulse Trigger

Rising, Falling, Rising/Falling

Pulse Condition Positive Pulse Width (greater than, lower than, within specific interval)

Pulse Width

Negative Pulse Width (greater than, lower than, within specific interval)

8 ns to 10 s

Runt Trigger (Option)

Pulse Width

Condition

Pulse Polarity

None, >, <, <>

Positive, Negative

MSO1000Z/DS1000Z User’s Guide 18-3

RIGOL

Chapter 18 Specifications

Pulse Width

Range

8 ns to 10 s

Window Trigger (Option)

Window Type Rising, Falling, Rising/Falling

Trigger Position Enter, Exit, Time

Window Time 8 ns to 10 s

Nth Edge Trigger (Option)

Edge Type Rising, Falling

Idle Time 16 ns to 10 s

Edge Number 1 to 65535

Slope Trigger

Slope Condition Positive Slope (greater than, lower than, within specific interval)

Negative Slope (greater than, lower than, within specific

Time Setting interval)

8 ns to 10 s

Video Trigger

Signal Standard NTSC, PAL/SECAM, 480P, 576P

Pattern Trigger

Pattern Setting H, L, X, Rising, Falling

Delay Trigger (Option)

Edge Type Rising, Falling

Delay Type >, <, <>, ><

Delay Time 8 ns to 10 s

TimeOut Trigger (Option)

Edge Type Rising, Falling, Rising/Falling

Timeout time 16 ns to 10 s

Duration Trigger (Option)

Pattern

Trigger

Condition

H, L, X

>, <, <>

Duration Time 8 ns to 10 s

Setup/Hold Trigger (Option)

Edge Type

Data Type

Setup Time

Hold Time

Rising, Falling

H, L, X

8 ns to 1 s

8 ns to 1 s

RS232/UART Trigger (Option)

Polarity

Trigger

Condition

Baud Rate

Normal, Invert

Start, Error, Check Error, Data

2400 bps, 4800 bps, 9600 bps, 19200 bps, 38400 bps, 57600 bps, 115200 bps, 230400bps, 460800bps, 921600bps, 1Mbps and User

18-4 MSO1000Z/DS1000Z User’s Guide

Chapter 18 Specifications

RIGOL

Data Bits 5 bit, 6 bit, 7 bit, 8 bit

I2C Trigger (Option)

Trigger

Condition

Address Bits

Start, Restart, Stop, Missing ACK, Address, Data, A&D

7 bits, 8 bits, 10 bits

Address Range 0x0 to 0x7F, 0x0 to 0xFF, 0x0 to 1023

Byte Length 1 to 5

SPI Trigger (Option)

Trigger

Condition

Timeout, CS

Timeout Value 16 ns to 10 s

Data Bits 4 bit to 32 bit

Data Line

Setting

H, L, X

Measure

Cursor

Auto

Measurement

Manual Mode

Track Mode

Voltage Deviation between Cursors (△V)

Time Deviation between Cursors (△T)

Reciprocal of △T (Hz) (1/△T)

Voltage and Time Values of the Waveform

Point

Auto Mode Allow to display cursors during auto measurement

Analog channel:

Period, Frequency, Rise Time, Fall Time, Positive Pulse Width,

Negative Pulse Width, Positive Duty Cycle, Negative Duty Cycle, tVmax, tVmin, Positive Rate, Negative Rate, Delay 12 , Delay

12 , Phase 12 , Phase 12 , Maximum, Minimum,

Peak-Peak Value, Top Value, Bottom Value, Amplitude, Upper

Value, Middle Value, Lower Value, Average, Mean Square Root,

Overshoot, Pre-shoot, Area, Period Area, Variance

Digital channel:

Frequency, Period, Positive Pulse Width, Negative Pulse Width,

Positive Duty Cycle, Negative Duty Cycle, Delay 12 , Delay

12 , Phase 12 , Phase 12

Number of

Measurements

Measurement

Range

Measurement

Statistic

Frequency

Counter

Display 5 measurements at the same time.

Screen or cursor

Average, Max, Min, Standard Deviation, Number of

Measurements

Hardware 6 bit frequency counter

(channels are selectable)

MSO1000Z/DS1000Z User’s Guide 18-5

RIGOL

Chapter 18 Specifications

Math Operation

Waveform

Operation

A+B, A-B, A×B, A/B, FFT, A&&B, A||B, A^B, !A, intg, Diff, Sqrt,

Lg, Ln, Exp, Abs

FFT Window

Function

FFT Display

Rectangle, Hanning, Blackman, Hamming, Flat Top, Triangle

Half, Full

FFT Vertical

Scale

Number of

Buses for dB/dBm, Vrms

2

Decoding

Decoding Type Parallel (standard), RS232/UART (option), I2C (option), SPI

(option)

Display

Screen Type

Display

Resolution

Display Color

Persistence

Time

Display Type

7.0 inch (203 mm) TFT LCD display

800 Horizontal ×RGB×480 Vertical Pixel

160,000 Color (TBD)

Min, 100 ms, 200 ms, 500 ms, 1 s, 5 s, 10 s, Infinite

Dots, Vectors

I/O

Standard Ports USB HOST, USB Device, LAN, Aux Output (TrigOut/PassFail),

GPIB (use USB HOST interface to expand)

Singnal Source (MSO1000Z-S/DS1000Z-S)

Channels

Sample Rate

Vertical

Resolution

2

200 MSa/s

14 bits

Max. Frequency 25 MHz

Standard Sine, Square, Pulse, Ramp, Noise, DC

Waveform

Built-in

Waveform

Sine

Sinc, Exponential Rise, Exponential Fall, ECG, Gauss, Lorentz,

Haversine

Frequency Range 0.1 Hz to 25 MHz

Flatness

Harmonic

Distortion

Stray

±0.5 dB (relative to 1 kHz)

-40 dBc

-40 dBc

18-6 MSO1000Z/DS1000Z User’s Guide

Chapter 18 Specifications

Square/Pulse

Ramp

Noise [1]

Built-in

Waveform

Arbitrary

Waveform

Frequency

Amplitude

DC Offset

RIGOL

(Non-harmonic)

Total Harmonic

Distortion

S/N Ratio

1%

40 dB

Frequency Range 0.1 Hz to 15 MHz

Rise/Fall Time

Overshoot

Duty Cycle

Duty Cycle

<15 ns

<5%

10% to 90%

1% or 10 ns (the larger of the two)

Resolution

Min. Pulse Width 20 ns

10 ns or 5 bits (the larger of the two) Pulse Width

Resolution

Jitter 500 ps

Frequency Range 0.1 Hz to 100 kHz

Linearity

Symmetry

1%

0 to 100%

Bandwidth 25 MHz

Frequency Range 0.1 Hz to 1 MHz

Frequency Range 0.1 Hz to 10 MHz

Waveform Length 2 to 16k points

Accuracy

Resolution

Output Range

100 ppm (lower than 10 kHz)

50 ppm (higher than 10 kHz)

0.1 Hz or 4 bits, the larger of the two

20 mVpp to 5 Vpp, HighZ

Resolution

Accuracy

Range

Resolution

Accuracy

10 mVpp to 2.5 Vpp, 50 Ω

100 μV or 3 bits, the larger of the two

2% (1 kHz)

±2.5 V, HighZ

±1.25 V, 50 Ω

100 μV or 3 bits, the larger of the two

2% (1 kHz)

MSO1000Z/DS1000Z User’s Guide 18-7

RIGOL

Chapter 18 Specifications

General Specifications

Probe Compensation Output

Output Voltage

Frequency [1]

Power

Power Voltage

Power

Fuse

Environment

Temperature

[1] About 3 V, peak-peak

1 kHz

100 V-240 V, 45 Hz-440 Hz

Maximum 50 W

2 A, T degree, 250 V

Range

Cooling Method Fan cooling

Humidity Range

Operating: 0 ℃ to +50 ℃

Non-operating: -40 ℃ to +70 ℃

0 ℃ to +30 ℃: ≤95% Relative Humidity

+30 ℃ to +40 ℃: ≤75% Relative Humidity

+40 ℃ to +50 ℃: ≤45% Relative Humidity

Altitude

Size

Operating: under 3,000 meters

Non-operating: under 15,000 meters

Physical Characteristics

[4]

Weight [4]

Width×Height×Depth = 313.1 mm× 160.8 mm×122.4 mm

Package 3.2 kg±0.2 kg

Excluded

Package

Included

Calibration Interval

3.8 kg±0.5 kg

The recommended calibration interval is one year.

Regulatory Information

Electromagnetic

Compatibility

Safety

2004/108/EC

Execution standard EN 61326-1:2006 EN 61326-2-1:2006

UL 61010-1:2004; CAN/CSA-C22.2 NO. 61010-1-2004;

EN 61010-1:2001; IEC 61010-1:2001

Note

Note

[1]

[2]

: Typical.

Note

Note

[3]

[4]

: Maximum value. 50ns, single-channel mode, dots display, auto memory depth.

: Supporting legs and handle folded, knob height included.

: Standard configuration.

18-8 MSO1000Z/DS1000Z User’s Guide

Chapter 19 Appendix

RIGOL

Chapter 19 Appendix

Appendix A: Accessories and Options

Model

Description

DS1104Z (100MHz, 4-analog channel)

DS1104Z-S (100MHz, 4-analog channel + 2-channel

25MHz signal source)

DS1074Z (70MHz, 4-analog channel)

DS1074Z-S (70MHz, 4-analog channel + 2-channel 25MHz signal source)

MSO1104Z (100 MHz, 4-analog channel+16-digital channel)

MSO1104Z-S (100 MHz, 4-analog channel+16-digital channel +2-channel 25MHz signal source)

MSO1074Z (70 MHz, 4-analog channel+16-digital channel)

MSO1074Z-S (70 MHz, 4-analog channel+16-digital channel +2-channel 25MHz signal source)

DS1054Z (50 MHz, 4-analog channel)

Power Cord conforming to the standard of the country

Order Number

DS1104Z

DS1104Z-S

DS1074Z

DS1074Z-S

MSO1104Z

MSO1104Z-S

MSO1074Z

MSO1074Z-S

Standard

Accessories

USB Cable

4 Passive Probes (150 MHz)

1 Logic Analyzer Pobe (only MSO1000Z/MSO1000Z-S)

Quick Guide

Resource CD (include User’s Guide and application software)

Rack Mount Kit

DS1054Z

-

CB-USBA-USBB-

FF-150

RP2200

RPL1116

-

-

Optional

Accessories

Memory

Depth

Options

Waveform

Record

Options

Advanced

Trigger

Options

24Mpts (single channel)/ 12Mpts (dual-channel)/ 6 Mpts

(four-channel)

This option provides the waveform record and playback function.

RS232/UART trigger, I2C trigger, SPI trigger, Runt trigger,

Window trigger, Nth edge trigger, delay trigger, timeout trigger, Setup/Hold trigger

RM-DS1000Z

MEM-DS1000Z

REC-DS1000Z

AT-DS1000Z

Serial

Protocol

Analysis

Options

RS232/UART decoding kit

I2C decoding kit

SPI decoding kit

SA-DS1000Z

Note: All the options or accessories can be ordered from you local RIGOL Office.

MSO1000Z/DS1000Z User’s Guide 19-1

RIGOL

Chapter 19 Appendix

Appendix B: Warranty

RIGOL warrants that its products mainframe and accessories will be free from defects in materials and workmanship within the warranty period.

If a product is proven to be defective within the respective period, RIGOL guarantees the free replacement or repair of products which are approved defective.

To get repair service, please contact with your nearest RIGOL sales and service office.

RIGOL does not provide any other warranty items except the one being provided by this summary and the warranty statement. The warranty items include but not being subjected to the hint guarantee items related to tradable characteristic and any particular purpose. RIGOL will not take any responsibility in cases regarding to indirect, particular and ensuing damage.

19-2 MSO1000Z/DS1000Z User’s Guide

Index

Index

- Duty ...................................... 6-22

- Width .................................... 6-22

+ Duty ..................................... 6-22

+ Width ................................... 6-22

+Rate ...................................... 6-25

Acquisition Mode ........................ 4-2

AM ........................................ 13-16 amplitude................................. 13-2

Antialiasing ................................ 4-7

Area ........................................ 6-25

Auto IP .................................... 15-4

Aux Output ............................ 15-10

Average ..................................... 4-2

Bandwidth Limit .......................... 2-3

Blackman ................................... 6-7

Built-in Waveform ..................... 13-6

Channel Coupling ........................ 2-2

Channel Label ............................. 2-5

CSV ......................................... 14-3

DC .......................................... 13-5

DC offset voltage ...................... 13-2

Delay ....................................... 6-23

Delay Calibration ......................... 2-7

Delay Trigger ............................ 5-25

Delayed Sweep ........................... 3-2 digital channel ............................ 7-1

Domain Name Server ................ 15-5

Dots ........................................ 12-2

Duration Trigger ....................... 5-17

Duty cycle ................................ 13-5

ECG ......................................... 13-8

Edge Trigger ............................... 5-7 edge type .......................... 5-7, 5-29

ExpFall ..................................... 13-7

ExpRise.................................... 13-7

Factory .................................. 14-11

Fall Time .................................. 6-22

Flattop ....................................... 6-7

FM ........................................ 13-17

Frequency ........................ 6-22, 13-2

Frequency Counter .................... 6-26

Gate ........................................ 15-5

Gauss ...................................... 13-8

MSO1000Z/DS1000Z User’s Guide

RIGOL

Hamming .................................. 6-7

Hanning .................................... 6-7

Haversine ................................. 13-9

High Resolution .......................... 4-3

I2C Decoding ............................ 8-11

I2C Trigger ............................... 5-33

Idle Time .................................. 5-30 impedance ................................ 13-3 interpolation ........................... 13-14

IP Address ................................ 15-4

LAN ......................................... 15-3 logic analyzer ............................ 1-13

Lorentz ..................................... 13-8

MAC ......................................... 15-6

Memory Depth ........................... 4-6

Modulating Waveform .............. 13-16

Modulation ............................. 13-15

Modulation Depth .................... 13-17

Modulation Deviation ............... 13-18

Modulation Frequency .............. 13-16

Network Status ......................... 15-3

Noise ....................................... 13-6

Noise Rejection .......................... 5-5

Normal ...................................... 4-2

Nth Edge Trigger ....................... 5-29

Overshoot................................. 6-25

Parallel Decoding ........................ 8-2

Parameters ............................... 14-3

Pattern Setting ................. 5-15, 5-17

Peak Detect ............................... 4-2

Period ...................................... 6-22

Period Area ............................... 6-25

Persistence Time ....................... 12-2

Phase ....................................... 6-23

Picture ..................................... 14-2

Preshoot ................................... 6-25

Probe Ratio ................................ 2-3

Pulse ........................................ 13-5

Pulse Condition .......................... 5-8

Pulse Polarity ............................ 5-21

Pulse Trigger .............................. 5-8

Pulse Width Setting .................... 5-9

Qualifier ................................... 5-21

1

RIGOL

Ramp ...................................... 13-3

-Rate ....................................... 6-25

Rectangle .................................. 6-7

Rise Time ................................ 6-22

Roll ........................................... 3-5

RS232 Decoding ......................... 8-6

RS232 Trigger .......................... 5-31

Runt Trigger ............................. 5-21

Sample Rate .............................. 4-4

Setup/Hold Trigger ................... 5-27

Setups ..................................... 14-3

Sin(x)/x ..................................... 4-4

Sinc......................................... 13-7

Sine ........................................ 13-2

Slope Condition ........................ 5-10

Slope Trigger ............................ 5-10

SPI Decoding ........................... 8-15

SPI Trigger ............................... 5-35

Square .................................... 13-3 start phase .............................. 13-2

Static IP .................................. 15-4

Subnet Mask ............................ 15-4

Symmetry ................................ 13-3

Sync ........................................ 5-13

Threshold ........................ 6-22, 6-26

Time Setting ............................ 5-11

To Create Waveform ................ 13-12

To Edit Waveform .................... 13-14

To Select Waveform ................. 13-11

Traces ..................................... 14-2

Triangle ..................................... 6-8

Trigger Coupling ......................... 5-4

Trigger Holdoff ........................... 5-5

Trigger Level .............................. 5-7

Index

Trigger Mode .............................. 5-3

Trigger Position ......................... 5-23

Trigger Source ............................ 5-2 tVmax ...................................... 6-22 tVmin ....................................... 6-22

USB ......................................... 15-6

Vamp ....................................... 6-24

Vavg ........................................ 6-24

Vbase ...................................... 6-24

Vectors .................................... 12-2

Vertical Scale .............................. 2-4

Vertical Window ............... 5-11, 5-22

Video Polarity ........................... 5-13

Video Standard ......................... 5-13

Video Trigger ............................ 5-13

VISA ........................................ 15-6

Vlower ..................................... 6-24

Vmax ....................................... 6-24

Vmid ........................................ 6-24

Vmin ........................................ 6-24

Vpp ......................................... 6-24

Vrms ........................................ 6-24

Vtop ........................................ 6-24

Vupper ..................................... 6-24

Waveform Confusion ................... 4-4

Waveform Distortion .................... 4-4

Waveform Leakage ...................... 4-5

Waves ...................................... 14-2

Window Function ........................ 6-7

Window trigger ......................... 5-23

Window Type ............................ 5-23

XY Mode .................................... 3-3

YT Mode .................................... 3-3

2 MSO1000Z/DS1000Z User’s Guide