datasheet for U2542A by Agilent
Agilent U2500A Series
USB Modular Simultaneous Sampling
Multifunction DAQ Devices
Data Sheet
Introduction
Features
• Simultaneous sampling with
up to 2 MSa/s sampling rate
for each channel
The Agilent U2500A Series USB modular simultaneous sampling multifunction
data acquisition (DAQ) devices are high-performance modules that consist of
three models – the U2531A, U2541A and U2542A. The U2500A Series has up to
four channels with resolutions of 14-bit and 16-bit. The U2531A can sample up
to 2 MSa/s for each channel with a resolution of 14 bits, while the U2541A and
U2542A can sample up to 250 kSa/s and 500 kSa/s for each channel
respectively with a resolution of 16 bits.
• Multifunction DAQ solution AI, AO, DIO, counter
• Dedicated ADC per channel
Various features of the U2500A Series
• 14-bit or 16-bit resolution
• Quick and easy USB setup
• 24-bit programmable digital
input/output
• High sampling rate of up to 2 MSa/s for each channel
• Functions as a standalone or
modular unit
• Flexible standalone or modular capability that enables lower startup cost
• Dedicated analog-to-digital (ADC) that allows simultaneous sampling of data
• Supports SCPI and IVI-COM
• SCPI and IVI-COM supported with a wide range of ADE compatibility that
minimizes work time and increases software choices
• Compatible with a wide range
of ADEs
• Easy-to-use application software and command logger function for easy SCPI
command conversion into snippets of VEE, VB, C++, and C# code
• Easy-to-use bundled software
• Command logger function
• USB 2.0 and USBTMC-USB488
standards
High sampling rate of up to 2 MSa/s
The U2500A Series provides a high analog input sampling rate coverage of up to
2 MSa/s for each channel. The high sampling rate coverage offered is ideal for
transient signal applications such as sonar analysis.
Simultaneous sampling of data
The U2500A Series has dedicated ADCs that enable simultaneous signals
acquisition, which makes the U2500A Series suitable for your phase-sensitive
applications.
2
Product outlook and dimensions
Front view
Product characteristics and General Specifications
REMOTE INTERFACE
• Hi-Speed USB 2.0
• USBTMC-USB488[1]
Rear view
POWER REQUIREMENT
• +12 VDC (TYPICAL)
• 2 A (MAX) input rated current
• Installation Category II
44.00 mm
POWER CONSUMPTION
+12 VDC, 480 mA maximum
OPERATING ENVIRONMENT
• Operating temperature from 0 °C to +55 °C
• Relative humidity at 15% to 85% RH (non-condensing)
• Altitude up to 2000 meters
• Pollution Degree 2
• For indoor use only
Top view
STORAGE COMPLIANCE
–20 °C to 70 °C
SAFETY COMPLIANCE
Certified with:
• IEC 61010-1:2001/EN 61010-1:2001 (2nd Edition)
• USA: ANSI/UL 61010-1:2004
• Canada: CSA C22.2 No.61010-1:2004
182.40 mm
EMC COMPLIANCE
• IEC 61326-1:2002/EN 61326-1:1997+A2:2001+A3:2003
• CISPR 11: 1990/EN 55011:1990-Group 1 Class A
• Canada: ICES-001:2004
• Australia/New Zealand: AS/NZS CISPR 11:2004
120.00 mm
Standard shipped accessories
•
•
•
•
•
•
•
•
•
AC/DC Power adapter
Power cord
USB extension cable
L-Mount kit (used with modular
product chassis)
Agilent USB Modular Products Quick
Start Guide
Agilent Measurement Manager for
U2500A Series Quick Start Guide
Agilent USB Modular Products
Reference CD-ROM
Agilent Automation-Ready CD-ROM
(contains the Agilent IO Libraries Suite)
Certificate of Calibration
SHOCK AND VIBRATION
Tested to IEC/EN 60068-2
IO CONNECTOR
68-pin female VHDCI Type
DIMENSION (W × D × H)
Module dimension:
• 120.00 mm × 182.40 mm × 44.00 mm (with plastic casing)
• 105.00 mm × 174.54 mm × 25.00 mm (without plastic casing)
Terminal block dimension:
• 103.00 mm × 85.20 mm × 42.96 mm
WEIGHT
• 565 g (with plastic casing)
• 400 g (without plastic casing)
WARRANTY
Three years for U2500A series DAQ devices
Three months for standard shipped accessories
Optional accessories
• U2901A Terminal block and SCSI-II 68pin connector with 1-meter cable
• U2902A Terminal block and SCSI-II 68pin connector with 2-meter cable
1. Compatible with Microsoft Windows operating systems only.
3
Product specifications
Model number
U2531A
U2541A
U2542A
Analog input
Resolution
14 bits
Number of channels
16 bits
4 differential input channels (software selectable/channel)
Maximum sampling rate
Programmable bipolar input range
2 MSa/s
250 kSa/s
[1]
500 kSa/s
± 10 V, ± 5 V, ± 2.5 V, ± 1.25 V
Programmable unipolar input range
0 to 10 V, 0 to 5 V, 0 to 2.5 V, 0 to 1.25 V
Input coupling
DC
Input impedance
1 GΩ/100 pF
Operational common mode voltage range
Overvoltage range
± 8.0 V maximum
Power-on: Continuous ± 30 V, Power-off: Continuous ± 15 V
Trigger sources
External analog/digital trigger, SSI/star trigger [2]
Trigger modes
Pre-trigger, delay-trigger, post-trigger, and middle-trigger
FIFO buffer size
Up to 8 MSa
Analog output
Resolution
12 Bits
Number of channels
2
Maximum update rate
1 MSa/s
0 to 10 V, ±10 V, 0 to AO_EXT_REF, ±AO_EXT_REF [3]
Output ranges
Output coupling
DC
Output impedance
0.1 Ω Typical
Stability
Any passive load up to 1500 pF
Power-on state
0 V steady state
Trigger sources
External analog/digital trigger, SSI/star trigger [2]
Trigger modes
Delay trigger, post trigger
FIFO buffer size
1 Channel used: Maximum 8 MSa
4 Channels used: Maximum 2 MSa/ch
Glitch energy
5 ns-V (Typical), 80 ns-V (Maximum)
Driving capability
5 mA
Function generation mode
Sine, square, triangle, sawtooth, and noise waveforms
Digital input/output
Number of bits
24-bit programmable input/output
Compatibility
TTL
Input voltage
VIL = 0.7 V maximum; IIL = 10 µA maximum
VIH = 2.0 V minimum; IIH = 10 µA maximum
Input voltage range
–0.5 V to +5.5 V
VOL = 0.45 V maximum; IOL = 8 mA maximum
VOH = 2.4 V minimum; IOH = 400 µA maximum
Output voltage
General purpose digital timer/counter
Maximum count
(231 – 1) bits
Number of channels
2 Independent up/down counter
Compatibility
TTL
Clock source
Internal or external
Base clock available
48 MHz
Maximum clock source frequency
Input frequency range
12 MHz
[4]
0.1 Hz to 6 MHz at 50% duty cycle
Pulse width measurement range
0.167 µs to 178.956 s ± 0.0833 µs
4
Product specifications (continued)
Model number
U2531A
U2541A
U2542A
Analog input
Trigger source
All analog input channels, External analog trigger (EXTA_TRIG)
± Full scale for internal
± 10 V for external
Trigger level
Trigger conditions
Above high, below low, and window (software selectable)
Trigger level resolution
8 bits
Bandwidth
400 kHz
Input impedance for EXTA_TRIG
20 kΩ
Coupling
DC
Overvoltage protection
Continuous for ± 35 V maximum
Digital trigger
Compatibility
TTL/CMOS
Response
Rising or falling edge
Pulse width
Calibration
20 ns minimum
[5]
On board reference voltage
5V
Temperature drift
± 2 ppm/°C
Stability
± 6 ppm/1000 hours
Power consumption
Input voltage (DC)
Input current
+12 VDC
480 mA maximum
390 mA maximum
Physical attributes
Dimensions (W × D × H)
120.00 mm × 182.40 mm × 44 mm (with plastic casing)
105.00 mm × 174.54 mm × 25.00 mm (without plastic casing)
IO connector
68-pin female VHDCI type
565 g with plastic casing
400 g without plastic casing
Weight
Environmental condition
Operating temperature
0 to 55 °C
Storage temperature
–20 °C to 70 °C
Relative humidity
15% to 85% RH (non-condensing)
General
Remote interface
Hi-Speed USB 2.0
Device class
USBTMC-USB488
Programmable interface
SCPI and IVI-COM
1. Maximum input voltage for analog input is ± 10 V.
2. System Synchronous Interface (SSI) and star trigger commands are applicable when modular devices are used in modular product chassis (U2781A).
3. Maximum external reference voltage for analog output (AO_EXT_REF) is ± 10 V.
4. Measurement frequency’s resolution:
= 12 MHz/n, n = 2, 3, 4, 5, ..., 120 M
= 6 MHz, 4 MHz, 3 MHz, 2.4 MHz, 2.0 MHz, ..., 0.1 Hz (up to six decimal points)
5. Recommended for 20 minutes warm-up time.
5
Electrical Specifications and Characteristics
Analog input characteristics [1]
Model number
U2531A
0 °C to 18 °C
28 °C to 55 °C
23 °C ± 5 °C
Offset error [2]
Gain error
[2]
± 2 mV
± 6 mV
–3 dB Small signal bandwidth
U2541A
23 °C ± 5 °C
± 2 mV
± 1 mV
± 6 mV
± 2 mV
1.2 MHz
1% THD Large signal bandwidth
U2542A
0 °C to 18 °C
28 °C to 55 °C
0 °C to 18 °C
28 °C to 55 °C
23 °C ± 5 °C
± 1 mV
± 1 mV
± 2.5 mV
± 2 mV
600 KHz
± 1 mV
± 2.5 mV
1.0 MHz
400 KHz
400 KHz
400 KHz
2.0 mVrms
0.5 mVrms
0.5 mVrms
CMRR (DC to 60 Hz)
64 dB
80 dB
80 dB
Spurious-Free Dynamic Range (SFDR)
76 dB
88 dB
86 dB
Signal-to-Noise and Distortion Ratio
(SINAD)
70 dB
82 dB
80 dB
Total Harmonic Distortion (THD)
–72 dB
–86 dB
–84 dB
Signal-to-Noise Ratio (SNR)
72 dB
84 dB
82 dB
11.3-bit
13.3-bit
13.0-bit
66 dB
84 dB
80 dB
U2531A
U2541A
U2542A
System noise [3]
Effective Number of Bits (ENOB)
Channels crosstalk [4]
Analog output characteristics [1]
Model number
23 °C ± 5 °C
28 °C to 55 °C
23 °C ± 5 °C
28 °C to 55 °C
23 °C ± 5 °C
28 °C to 55 °C
Offset error
± 1 mV
± 3 mV
± 1 mV
± 3 mV
± 1 mV
± 3 mV
Gain error
± 3 mV
± 4 mV
± 2 mV
± 4 mV
± 2 mV
± 4 mV
Slew rate
15 V/µs
15 V/µs
15 V/µs
Rise time
1.1 µs
1.2 µs
1.1 µs
1.2 µs
1.1 µs
1.2 µs
Fall time
1.1 µs
1.2 µs
1.1 µs
1.2 µs
1.1 µs
1.2 µs
Settling time(s) to 1% output error
2 µs
2 µs
1. Specifications are based on 20 minutes warm-up, self-calibration temperature at 23 °C, and bipolar input range of ± 10 V.
2. The measurements are calculated with 100 points averaging of data.
3. The noise rms value is the standard deviation of 20000 points.
4. The crosstalk measurements are tested up to input frequency of Fin = MaxSamplingRate/2.
6
2 µs
Test Condition
Dynamic range test for U2500A Series DAQ devices
Dynamic range test
SFDR, THD, SINAD, SNR, ENOB
Model
Test conditions (DUT setting at ± 10 V bipolar)
U2531A
•
•
•
•
Sampling rate:
Fundamental frequency:
Number of points:
Fundamental input voltage:
2 MSa/s
19.927 kHz
65536
FSR –1 dB FS
U2541A
•
•
•
•
Sampling rate:
Fundamental frequency:
Number of points:
Fundamental input voltage:
250 kSa/s
2.4109 kHz
8192
FSR – 1 dBFS
U2542A
•
•
•
•
Sampling rate:
Fundamental frequency:
Number of points:
Fundamental input voltage:
500 kSa/s
4.974 kHz
16384
FSR – 1 dBFS
Bandwidth test for U2500A Series DAQ devices
Bandwidth test
–3 dB Small signal bandwidth:
1% THD Large signal bandwidth:
Model
Test conditions (DUT setting at ± 10 V bipolar)
2 MSa/s
U2531A
Sampling rate:
Input voltage
• –3 dB Small signal bandwidth:
• 1% THD Large signal bandwidth:
250 kSa/s
U2541A
Sampling rate:
Input voltage
• –3 dB Small signal bandwidth:
• 1% THD Large signal bandwidth:
500 kSa/s
U2542A
Sampling rate:
Input voltage
• –3 dB Small signal bandwidth:
• 1% THD Large signal bandwidth:
7
10% FSR
FSR – 1 dBFS
10% FSR
FSR – 1 dBFS
10% FSR
FSR – 1 dBFS
Typical Performance
8
DC Characteristics
Accuracy specifications[1]
Model
U2541A, U2542A
Analog input
Unipolar range (V)
Offset error (mV) [2]
Gain error (mV)
10
1.0
1.0
0.02% + 1.0 mV
Accuracy (% of reading + offset error)
5
1.0
1.0
0.04% + 1.0 mV
2.5
1.0
1.0
0.08% + 1.0 mV
1.25
1.0
1.0
0.16% + 1.0 mV
10
1.0
2.0
0.02% + 1.0 mV
Bipolar range (V)
5
1.0
1.0
0.02% + 1.0 mV
2.5
1.0
1.0
0.04% + 1.0 mV
1.25
1.0
1.0
0.08% + 1.0 mV
Unipolar range (V)
Offset error (mV) [2]
Gain error (mV)
Model
U2531A
Accuracy (% of reading + offset error)[3]
10
2.0
3.0
0.06% + 2.0 mV
5
1.5
1.5
0.06% + 1.5 mV
2.5
1.0
1.0
0.08% + 1.0 mV
1.25
1.0
1.0
0.16% + 1.0 mV
10
2.0
6.0
0.06% + 2.0 mV
Bipolar range (V)
5
1.5
3.0
0.06% + 1.5 mV
2.5
1.0
2.0
0.08% + 1.0 mV
1.25
1.0
1.0
0.08% + 1.0 mV
Model
U2541A, U2542A
Analog output
Unipolar range (V)
Offset error (mV) [2]
Gain error (mV)
Accuracy (% of reading + offset error) [4]
10
1.0
2.0
0.02% + 1.0 mV
1.0
2.0
Bipolar range (V)
10
Model
0.02% + 1.0 mV
U2531A
Unipolar range (V)
Offset error (mV) [2]
Gain error (mV)
Accuracy (% of reading + offset error) [4]
10
1.0
3.0
0.03% + 1.0 mV
1.0
3.0
0.03% + 1.0 mV
Bipolar range (V)
10
1. Specifications are based on 20 minutes warm-up, and self-calibration temperature at 23 °C.
2. Offset error is measured at 0 V.
3. Accuracy = ± [% of Gain error/(Measured value – Midscale) + Offset error]
4. Accuracy = ± (% of Gain error/Output value + Offset error)
9
Agilent Measurement Manager
The Agilent Measurement Manager
(AMM) is an application data viewer
software that comes with the
standard purchase of the U2500A
Series USB modular data acquisition.
This software is designed to help you
perform quick device configuration,
data logging and data acquisition
using the products.
Supported features found in the
U2500A Series USB modular
simultaneous sampling multifunction
DAQ device:
•
•
•
•
•
•
•
•
•
Averaging
Command logger
Self-test
Self-calibration
Option to save the current
instrument configuration to a file
Data logging and export feature to
CSV, HTML and text only format
files that can be printed
Data viewer to load and review
previously logged data
Trigger settings between modules
in the instrument chases with Star
trigger and Master/Slave trigger
Synchronization display and
data logging for modules in the
instrument chassis
Agilent measurement manager prerequisites
Prior to installing the Agilent Measurement Manager software, ensure that
your PC meets the following minimum system requirements for installation and
operations.
Hardware requirements
Processor
1.6 GHz Pentium IV or higher
Operating system
One of the following Microsoft Windows versions:
• Windows XP Professional or Home edition
(Service Pack 1 or later)
• Windows Vista 32-bit (Business, Ultimate, Enterprise,
Home Basic and Home Premium edition)
• Windows 7 32-bit (Home Basic, Home Premium,
Professional, Enterprise and Ultimate edition)
• Windows 7 (64-bit) support for 32-bit application running on
a WOW64 (Windows-on-Windows 64 bit) emulator
Hard disk space
1 GB
RAM
512 MB or higher recommended
Video
Super VGA (800 x 600), 256 colors or more
Software requirements
Agilent IO Libraries Suite 15.1 and above [1]
Agilent T&M Toolkit Runtime version 2.1 [2]
Agilent T&M Toolkit Redistributable Package 2.1 patch [2]
Microsoft .NET Framework version 2.0 [2]
1. Available on the Agilent Automation-Ready CD-ROM
2. Bundled with Agilent Measurement Manager software application installer
10
Other products in the Agilent USB Modular Data Acquisition (DAQ) Family
U2300A Series USB Modular Multifunction DAQ
Features:
•
•
•
•
High analog input sampling rate coverage of up to 3 MSa/s for a single channel
High analog input up to 64 channels
High speed USB 2.0
Multifunction capabilities — analog input (AI), analog output (AO), digital
input output (DIO), and counter
For more information: http://www.agilent.com/find/U2300A
U2600A Series USB Modular Isolated Digital I/O
Features:
• 64 opto-isolated lines that can meet demand up to 24 V
• High speed USB 2.0
• Isolation voltage of 1250 Vrms for protection from transient voltage spikes
For more information: http://www.agilent.com/find/U2600A
U2781A USB modular product chassis
Features:
•
•
•
•
•
Expansion of channels for each modular product
Multiple instrument synchronization
Internal and external 10 MHz reference clock
High-speed USB 2.0
SSI/Star trigger bus synchronization between external trigger source and
modules
For more information: http://www.agilent.com/find/U2781A
11
Optional Accessories
U2901A/U2902A -Terminal
block and SCSI- II 68-pin
connector with
1-meter/2-meter cable
Terminal block overview
Front view
The U2901A/U2902A is a terminal
block and SCSI-II 68-pin connector
with 1 meter cable or 2 meter cable
that can be used conjunction with the
U2300A Series and U2500A Series.
85.20 mm
103.00 mm
Side view
42.96 mm
28.40 mm
12
Ordering Information
Model
Description
U2541A
250 kSa/s USB modular simultaneous sampling multifunction DAQ
U2542A
500 kSa/s USB modular simultaneous sampling multifunction DAQ
U2531A
2 MSa/s USB modular simultaneous sampling multifunction DAQ
Optional accessories
Model
Description
U2901A
Terminal block and SCSI-II 68-pin connector with 1-meter cable
U2902A
Terminal block and SCSI-II 68-pin connector with 2-meter cable
13
www.agilent.com
www.agilent.com/find/U2500A
For more information on Agilent
Technologies’ products, applications or
services, please contact your local Agilent
office. The complete list is available at:
Agilent Email Updates
www.agilent.com/find/emailupdates
Get the latest information on the
products and applications you select.
www.lxistandard.org
LAN eXtensions for Instruments puts
the power of Ethernet and the Web
inside your test systems. Agilent
is a founding member of the LXI
consortium.
Agilent Channel Partners
www.agilent.com/find/channelpartners
Get the best of both worlds: Agilent’s
measurement expertise and product
breadth, combined with channel
partner convenience.
Agilent Advantage Services is committed
to your success throughout your equipment’s lifetime. To keep you competitive,
we continually invest in tools and
processes that speed up calibration and
repair and reduce your cost of ownership.
You can also use Infoline Web Services
to manage equipment and services more
effectively. By sharing our measurement
and service expertise, we help you create
the products that change our world.
www.agilent.com/find/advantageservices
www.agilent.com/find/contactus
Americas
Canada
Brazil
Mexico
United States
(877) 894 4414
(11) 4197 3600
01800 5064 800
(800) 829 4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Other AP Countries
1 800 629 485
800 810 0189
800 938 693
1 800 112 929
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
(65) 375 8100
www.agilent.com/quality
Europe & Middle East
Belgium
32 (0) 2 404 93 40
Denmark
45 45 80 12 15
Finland
358 (0) 10 855 2100
France
0825 010 700*
*0.125 €/minute
Germany
Ireland
Israel
Italy
Netherlands
Spain
Sweden
United Kingdom
49 (0) 7031 464 6333
1890 924 204
972-3-9288-504/544
39 02 92 60 8484
31 (0) 20 547 2111
34 (91) 631 3300
0200-88 22 55
44 (0) 118 927 6201
For other unlisted countries:
www.agilent.com/find/contactus
Revised: January 6, 2012
Product specifications and descriptions
in this document subject to change
without notice.
© Agilent Technologies, Inc. 2012
Published in USA, June 29, 2012
5991-0651EN