HMS1000E
HMS1000E
1GHz Spectrum Analyzer
HMS1000E
Comfortable automatic
measurement functions
with up to 8 markers
R Frequency Range 100 kHz…1 GHz
R Amplitude Measurement Range -104…+20 dBm
R Sweep Time 20 ms…1000s
R R
esolution Bandwidth 10 kHz…1 MHz in 1–3 Steps, 200 kHz (-3 dB)
1 GHz EMI Near Field Probe
Set HZ530
R Spectral Purity <-100 dBc/Hz (@100 kHz)
R Video Bandwidth 1 kHz…1 MHz in 1–3 Steps
R Integrated AM and FM Demodulator (Phone and int. Speaker)
R Detectors: Auto-, Min-, Max-Peak, Sample, RMS
R 8 Marker with Delta Marker, miscellaneous Peak Functions
Ethernet/USB-interface
HO730 (Option)
R Crisp 16.5 cm (6.5") TFT VGA Display
R 3
x USB for Mass-Storage, Printer and Remote Control,
optional IEEE-488 (GPIB) or Ethernet/USB Interface
1 GHz Spectrum Analyzer HMS1000E
All data valid at 23 °C after 30 minute warm-up.
Frequency
100 kHz…1 GHz
Frequency range:
±2 ppm (0…30 °C)
Temperature stability:
±1 ppm/year
Aging:
0 Hz (zero span) and 1 MHz…1 GHz
Span setting range:
Spectral purity, SSB phase noise:
100 kHz from carrier <-100 dBc/Hz
(500M Hz, +20…30 °C)
1 MHz from carrier <-120 dBc/Hz
(500MHz, +20…30 °C)
Sweep time:
20 ms…100 s
Span = 0 Hz
20 ms…1000 s, min. 20 ms/600 MHz
Span > 0 Hz
Resolution bandwidths (-3 dB): 10 kHz…1 MHz in 1–3 steps, 200 kHz
Tolerance:
±5 % typ.
≤300kHz
±10 % typ.
1MHz
1 kHz…1 MHz in 1–3 steps
Video bandwidths:
Amplitude
Display range:
Average noise level displayed up to
+20 dBm
Amplitude measurement
Typ. -104…+20 dBm
range:
Max. permissible DC 80 V
at HF input:
20 dBm, 30 dBm for max. 3 Min.
Max. power at HF input:
Intermodulation free range:
TOI products, 2x -20 dBm 66 dB typ.
(-10 dBm ref. level)
(typ. +13 dBm third-order intercept)
(at distance between
signals ≤2 MHz)
60 dB typ. (+10 dBm TOI)
(at distance between
signals >2 MHz)
66 dB typ. (typ. +13 dBm TOI)
DANL (Displayed average noise level):
(RBW 10 kHz, VBW 1 kHz,
ref. level ≤-30 dBm
10 MHz…1 GHz)
-95 dBm, typ. -104 dBm
Inherent spurious:
(ref. level ≤-20 dBm,
f >30 MHz, RBW ≤100 kHz) <-80 dBm
Input related spurious:
(Mixer level ≤-40 dBm,
carrier offset >1 MHz)
-70 dBc typ.
2nd harmonic receive frequency:
(mixer level -40 dBm)
-60 dBc typ.
Level display:
-80…+20 dBm in 1 dB steps
Reference level
100 dB, 50 dB, 20 dB, 10 dB
Display range
Logarithmic display
dBm, dBµV, dBmV
scaling
1 curve and 1 memory curve
Measured curves:
A-B (curve-stored curve), B-A
Trace mathematics:
Auto-, Min-, Max-Peak, Sample, RMS,
Detectors:
Average
<1.5 dB, typ. 0.5 dB
Failure of level display:
(ref. level -50 dBm, 20…30 °C)
Marker/Deltamarker
Number of marker:
Marker functions:
Marker displays:
Inputs/Outputs
HF Input
Input Impedance:
VSWR (10 MHz…1 GHz):
Trigger input:
Trigger voltage
Ext. reference input/output:
Reference frequency
Essential level (50 Ω)
Supply output for field
probes:
Audio output (Phone):
Demodulation
Miscellaneous
Display:
Save/Recall memory:
Trigger:
Interfaces:
Power supply:
Power consumption:
Protection class:
Operating temperature:
Storage temperature:
Rel. humidity:
Dimensions (W x H x D):
Weight:
3.5 mm DIN jack
AM and FM (internal speaker)
16.5 cm (6.5“) TFT Color VGA Display
10 complete device settings
Free run, Single Trigger, external Trigger
Dual-Interface USB/RS-232 (HO720),
USB-Stick (frontside),
USB-Printer (rear side)
105…253 V, 50…60 Hz, CAT II
Max. 40 W at 230 V, 50 Hz
Safety class I (EN61010-1)
+5…+40 °C
-20…+70 °C
5…80 % (non condensing)
285 x 175 x 220 mm
3.6 kg
Accessories supplied: Line cord, Operating manual, HZ21 Adapter plug,
N plug to BNC socket, CD, Software
Recommended accessories:
HO730
Dual-Interface Ethernet/USB
HO740
Interface IEEE-488 (GPIB), galvanically isolated
HZ13
Interface cable (USB) 1.8 m
HZ14
Interface cable (serial) 1:1
HZ20
Adapter, BNC to 4 mm banana
HZ33
Test cable 50 Ω, BNC/BNC, 0.5 m
HZ34
Test cable 50 Ω, BNC/BNC, 1 m
HZ46
4RU 19‘‘ Rackmount Kit
HZ72
GPIB-Cable 2 m
HZ99
Carrying Case for protection and transport
HZ520
Plug-in Antenna with BNC connection
HZ525
50 Ω-Termination, N plug
HZ530
Near-Field Probe Set 1 GHz for EMV diagnostics
HZ540/550
Near-Field Probe Set 3 GHz for EMV diagnostics
HZ540L/550L Near-Field Probe Set 3 GHz for EMV diagnostics
HZ560
Transient limiter
HZ575
75/50 Ω Converter
HZO30
active probe 1 GHz (0.9 pF, 1 MΩ, including many accessories)
8
Peak, next peak, minimum,
center = marker, frequency,
reference level = marker level,
all marker on peak
Normal (level & log.), delta marker,
noise marker
N socket
50 Ω
<1.5 typ.
BNC female
TTL
BNC females
10 MHz
10 dBm
6 Vdc, max. 100 mA (2.5 mm DIN jack)
HMS1000E/E/240112 · C&E · Subject to change without notice · © HAMEG Instruments GmbH® · DQS-certified in accordance with DIN EN ISO 9001:2008, Reg.-No.: 071040 QM08
HAMEG Instruments GmbH · Industriestr. 6 · D-63533 Mainhausen · Tel +49 (0) 6182 8000 · Fax +49 (0) 6182 800100 · www.hameg.com · [email protected]
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF