Below you will find brief information for RTD transmitter MXR. This document provides information on installation, connection and basic maintenance procedures. It also includes information on the product's conformity with EC directives and power input ratings. The device also features burnout protection, reset all settings function, and calibration procedures. It's suitable for indoor environments and has an operating temperature range of -5 to +55°C.

PDF Download
Document
RTD transmitter MXR Instruction Manual | Manualzz

Advertisement

Advertisement

Advertisement

Key features

  • Installation instructions
  • Connection and wiring diagrams
  • Burnout protection
  • Reset all settings function
  • Calibration procedures
  • Temperature range (-5 to +55°C)
  • Conformity with EC directives
  • Power input ratings
  • Indoor environment use
  • Moving average function

Frequently asked questions

The operating temperature range is -5 to +55°C.

The power input rating is 100-240V AC, 85-264V, 47-66Hz.

Warm up the unit for at least 10 minutes. Apply 0%, 25%, 50%, 75% and 100% input signal. Check that the output signal remains within the accuracy described in the datasheet. When the output is out of tolerance, recalibrate the unit using ITEM 07 and 08.

The burnout function protects the device from wire breakdown or invalid overrange input. If the RTD wires are broken, the output goes over 100% (below 0% with downscale) due to the burnout function. In such a case, check the leadwires.