Extech Instruments EX330 is a versatile multimeter that offers a range of functions for electrical measurements, including AC/DC voltage and current, resistance, capacitance, frequency, and temperature. It features a non-contact AC voltage detector for added safety, and a continuity beeper for quickly identifying open circuits. The EX330 also has a large backlit display for easy reading, and a data hold function for capturing and storing measurements.
advertisement
User's Guide
Mini Multimeter with Non-Contact Voltage Detector (NCV)
Model EX330
Introduction
Congratulations on your purchase of the Extech EX330 Meter. The EX330 offers AC/DC
Voltage, AC/DC Current, Resistance, Diode, Continuity, non-contact Voltage Detection,
Capacitance, Frequency, Duty Cycle, and Temperature (Type K) functions. Proper use and care of this meter will provide many years of reliable service.
Safety
WARNING
CAUTION
MAX
600V
This symbol adjacent to another symbol, terminal or operating device indicates that the operator must refer to an explanation in the Operating
Instructions to avoid personal injury or damage to the meter.
This WARNING symbol indicates a potentially hazardous situation, which if not avoided, could result in death or serious injury.
This CAUTION symbol indicates a potentially hazardous situation, which if not avoided, may result damage to the product.
This symbol advises the user that the terminal(s) so marked must not be connected to a circuit point at which the voltage with respect to earth ground exceeds 600V.
This symbol adjacent to one or more terminals identifies them as being associated with ranges that may, in normal use, be subjected to particularly hazardous voltages. For maximum safety, the meter and its test leads should not be handled when these terminals are energized.
This symbol indicates that a device is protected throughout by double insulation or reinforced insulation.
SAFETY INSTRUCTIONS
This meter has been designed for safe use, but must be operated with caution. The rules listed below must be carefully followed for safe operation.
1. NEVER apply voltage or current to the meter that exceeds the specified maximum:
Input Protection Limits
V DC or V AC mA AC/DC
A AC/DC
Frequency, Resistance, Capacitance, Duty
Cycle, Diode Test, Continuity
600V AC and DC
500mA DC/AC
10A DC/AC (for 30 seconds max. every 15 minutes
250V DC/AC
2. USE EXTREME CAUTION when working with high voltages.
3. DO NOT measure voltage if the voltage on the "COM" input jack exceeds 600V above earth ground.
4. NEVER connect the meter leads across a voltage source while the function switch is in the current, resistance, or diode mode. Doing so can damage the meter.
5. ALWAYS discharge filter capacitors in power supplies and disconnect the power when making resistance or diode tests.
6. ALWAYS turn off power and disconnect test leads before opening the covers to replace the fuse or battery.
7. NEVER operate the meter unless the back cover and the battery and fuse covers are in place and fastened securely.
8. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
2
Model EX330 - Version 4.0 – 05/06
Controls and Jacks
1. AC Voltage Detector Sensor
2. AC Voltage Detector indicator light
1
2
3
6. Non-contact AC Voltage Detector test button
7. Rotary function dial
8. 10 ampere test lead jack
9. COM test lead jack
10. Test lead jack for voltage, milli-amp, micro-amp, resistance, capacitance, frequency, and temperature functions
11. RANGE button
12. HOLD button
13. Protective rubber holster (must be removed to access the rear battery compartment)
4
5
6
7
8
Display Symbols and Annunciators
•))) Continuity
Diode test
Battery status n nano
(10
-3
) (volts, amps) A Amps
F Farads (capacitance)
Ω
Ohms
Hz Hertz V Volts
% Percent (duty ratio) REL Relative
DC
ºF
Direct current
Degrees Fahrenheit
HOLD Display hold
ºC Degrees Centigrade
9
10
13
12
11
3
Model EX330 - Version 4.0 – 05/06
Operating Instructions
WARNING
: Risk of electrocution. High-voltage circuits, both AC and DC, are very dangerous and should be measured with great care.
1. ALWAYS turn the function switch to the OFF position when the meter is not in use.
2. Press the HOLD button to freeze a displayed reading
NOTE
: On some low AC and DC voltage ranges, with the test leads not connected to a device, the display may show a random, changing reading. This is normal and is caused by the high-input sensitivity. The reading will stabilize and give a proper measurement when connected to a circuit.
NON-CONTACT AC VOLTAGE DETECTOR
The EX330 can detect the presence of AC voltage (from 100 to 600VAC) simply by being held very near to a voltage source.
WARNING
: Test the AC voltage detector on a known live circuit before each use.
WARNING
: Before using the meter in the AC Voltage Detector mode, verify that the battery is fresh by confirming characters appear on the LCD when the function dial is turned to any position. Do not attempt to use the meter as an AC Voltage Detector if the battery is weak or bad.
NON-CONTACT VOLTAGE (NCV)
The NCV function works on any rotary switch position.
1. Test the detector on a known live circuit before use.
2. Press and hold the NCV button for the duration of the test. The meter will beep once when the button is pushed.
3. Hold the top of the meter very close to the voltage source as shown.
4. If voltage is present, the rim of the LCD display will flash a bright orange and an audible warning will sound.
4
Model EX330 - Version 4.0 – 05/06
AC VOLTAGE MEASUREMENTS
WARNING:
Risk of Electrocution. The probe tips may not be long enough to contact the live parts inside some 240V outlets for appliances because the contacts are recessed deep in the outlets. As a result, the reading may show 0 volts when the outlet actually has voltage on it. Make sure the probe tips are touching the metal contacts inside the outlet before assuming that no voltage is present.
CAUTION:
Do not measure AC voltages if a motor on the circuit is being switched ON or OFF. Large voltage surges may occur that can damage the meter.
1. Set the function switch to the VAC position.
2. Insert the black test lead banana plug into the negative COM jack.
Insert red test lead banana plug into the positive V jack.
3. Touch the black test probe tip to the neutral side of the circuit.
4. Touch the red test probe tip to the “hot” side of circuit.
5. Read the voltage in the display.
6. If the measured AC voltage exceeds the highest range of the meter (refer to the specification table) an audible tone will sound.
DC VOLTAGE MEASUREMENTS
CAUTION:
Do not measure DC voltages if a motor on the circuit is being switched ON or
OFF. Large voltage surges may occur that can damage the meter.
1. Set the function switch to the VDC position.
2. Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive V jack.
3. Touch the black test probe tip to the negative side of the circuit.
Touch the red test probe tip to the positive side of the circuit.
4. Read the voltage in the display.
5
Model EX330 - Version 4.0 – 05/06
AC / DC CURRENT MEASUREMENTS
CAUTION:
Do not make current measurements at 10 Amps for longer than 30 seconds.
Exceeding 30 seconds may cause damage to the meter and/or the test leads.
1. Insert the black test lead banana plug into the negative COM jack.
2. For current measurements up to 4000µA, set the function switch to the µA position and insert the red test lead banana plug into the mA/µA jack
3. For current measurements up to 400mA, set the function switch to the mA position and insert the red test lead banana plug into the mA/µA jack.
4. For current measurements up to 10A, set the function switch to the 10A range and insert the red test lead banana plug into the 10A jack.
5. Use the MODE button to select AC or DC current. The display will reflect the selection.
6. Remove power from the circuit under test, then open up the circuit at the point where you wish to measure current.
7. Touch the black test probe tip to the negative side of the circuit. Touch the red test probe tip to the positive side of the circuit.
8. Apply power to the circuit.
9. Read the current in the display.
6
Model EX330 - Version 4.0 – 05/06
RESISTANCE MEASUREMENTS
WARNING:
To avoid electric shock, disconnect power to the unit under test and discharge all capacitors before taking any resistance measurements. Remove the batteries and unplug the line cords.
1. Set the function switch to the
Ω position.
2. Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive
Ω jack.
3. Touch the test probe tips across the circuit or component under test. It is best to disconnect one side of the circuit under test so the rest of the circuit will not interfere with the resistance reading.
4. Read the resistance in the display.
CONTINUITY CHECK
WARNING:
To avoid electric shock, never measure continuity on circuits or wires that have voltage on them.
1. Set the function switch to the position.
2. Insert the black lead banana plug into the negative COM jack.
3. Insert the red test lead banana plug into the positive
Ω jack.
4. Use the MODE button to view the
icon on the display.
5. Touch the test probe tips to the circuit or wire you wish to check.
6. If the resistance is less than approximately
100
Ω, the audible signal will sound. If the circuit is ‘open’ (bad), the display will indicate “OL”.
7
Model EX330 - Version 4.0 – 05/06
DIODE TEST
1. Set the function switch to the position.
2. Insert the black test lead banana plug into the negative COM jack and the red test lead banana plug into the positive jack.
3. Use the MODE button to view the icon on the display.
4. Touch the test probes to the diode under test.
Forward voltage will typically indicate 0.400 to
0.700V. Reverse voltage will indicate “OL”.
Shorted devices will indicate near 0V and an open device will indicate “OL” in both polarities.
8
Model EX330 - Version 4.0 – 05/06
CAPACITANCE MEASUREMENTS
WARNING:
To avoid electric shock, disconnect power to the unit under test and discharge all capacitors before taking any capacitance measurements.
Remove the batteries and unplug the line cords.
1. Set the rotary function switch to the CAP position.
2. Insert the black test lead banana plug into the negative COM jack.
3. Insert the red test lead banana plug into the positive CAP jack.
4. Touch the test leads to the capacitor to be tested.
5. The test may take up to 3 minutes or more for large capacitors to charge. Wait until the readings settle before ending the test.
6. Read the capacitance value in the display
FREQUENCY MEASUREMENTS
1. Use the MODE button to view the Hz unit of measure on the LCD display.
2. Insert the black test lead banana plug into the negative COM jack and the red test lead banana plug into the positive Hz jack.
3. Touch the test probe tips to the circuit under test.
4. Read the frequency on the display.
% DUTY CYCLE
1. Set the rotary function switch to the Hz/% position.
2. Use the MODE button to view the % unit of measure on the LCD display.
3. Insert the black lead banana plug into the negative COM jack and the red test lead banana plug into the positive Hz jack.
4. Touch the test probe tips to the circuit under test.
5. Read the % duty cycle on the display.
CONTACT TEMPERATURE MEASUREMENTS
1. Set the function switch to the ºF or ºC position.
2. Insert the Temperature Probe into the input jacks, making sure to observe polarity.
3. The probe can be pressed against a device under test to read its temperature or the probe can be held in air to read the ambient temperature. Allow 30 seconds for the display to stabilize.
4. Read the temperature in the display.
9
Model EX330 - Version 4.0 – 05/06
AUTO-MANUAL RANGE SELECTION
When the meter is first turned on, it automatically goes into the Auto Range mode. This automatically selects the best range for the measurements being made and is generally the best mode for most measurements. For measurement situations requiring that a range be manually selected, perform the following:
1. Press the RANGE key. The AUTO display indicator will turn off.
2. Press RANGE to step through the available ranges until the desired range is selected.
3. To exit the Manual Ranging mode and return to Auto Range, press and hold the
RANGE key for 2 seconds.
Note:
Manual range does not apply to Capacitance, Frequency or Temperature modes.
RELATIVE MODE
The relative measurement feature allows you to make measurements relative to a stored reference value. A reference voltage, current, etc. can be stored so that subsequent measurements can be made in comparison to that value. The displayed value is the difference between the reference value and the measured value.
1. Perform the measurement as described in the operating instructions.
2. Press the REL button to store the reading (the REL indicator will appear on the display).
3. The display will now indicate the difference between stored value and subsequent measurements.
4. Press the REL button to exit the relative mode.
Note:
The Relative mode is not available when measuring Frequency or Duty Cycle.
Maintenance
WARNING
: To avoid electric shock, disconnect the test leads from any source of voltage before removing the back cover or the battery or fuse covers.
WARNING:
To avoid electric shock, do not operate your meter until the battery and fuse covers are in place and fastened securely.
This MultiMeter is designed to provide years of dependable service, if the following care instructions are performed:
1.
KEEP THE METER DRY
. If it gets wet, dry it immediately.
2.
USE AND STORE THE METER IN NORMAL TEMPERATURES.
Temperature extremes can shorten the life of the electronic parts and distort or melt plastic parts.
3.
HANDLE THE METER GENTLY AND CAREFULLY.
Dropping it can damage the electronic parts or the case.
4.
KEEP THE METER CLEAN.
Wipe the case occasionally with a damp cloth. DO
NOT use chemicals, cleaning solvents, or detergents.
5.
USE ONLY FRESH BATTERIES OF THE RECOMMENDED SIZE AND TYPE.
Remove old or weak batteries so they do not leak and damage the unit.
6.
IF THE METER IS TO BE STORED FOR A LONG PERIOD OF TIME
, the batteries should be removed to prevent damage to the unit.
10
Model EX330 - Version 4.0 – 05/06
BATTERY INSTALLATION and LOW BATTERY INDICATION
WARNING
: To avoid electric shock, disconnect the test leads from any source of voltage before removing the battery cover.
LOW BATTERY INDICATION
The icon will appear in the lower left-hand corner of the display when the battery voltage becomes low. Replace the batteries when this appears.
BATTERY REPLACEMENT
1. Disconnect the test leads from the meter.
2. Remove the protective rubber holster as shown in the diagram.
3. Remove the Phillips head screw located on the lower back of the instrument.
4. Flip up the fuse/battery compartment cover to access the batteries.
5. Gently remove the batteries and install two new 1.5V ‘AAA’ batteries observing polarity.
6. Secure the fuse/battery compartment cover.
7. Place the protective rubber holster on the meter.
WARNING:
To avoid electric shock, do not operate the meter until the batteries and the fuses are in place and fastened securely.
Holster
2. Meter
3. Battery
4. Fuses
REPLACING THE FUSES
WARNING
: To avoid electric shock, disconnect the test leads from any source of voltage before removing the fuse cover.
1. Disconnect the test leads from the meter.
2. Remove the protective rubber holster as shown in the diagram.
3. Remove the Phillips head screw located on the lower back of the instrument.
4. Flip up the fuse/battery compartment cover to access the fuses.
5. Gently remove the fuse(s) and install new fuse(s) into the holder(s).
6. Always use fuses of the proper size and value (500mA/250V fast blow for the mA /
µA ranges, 10A/250V fast blow for the A range).
7. Secure the fuse/battery compartment cover.
8. Place the protective rubber holster on the meter.
11
Model EX330 - Version 4.0 – 05/06
UL LISTED
The UL mark does not indicate that this product has been evaluated for the accuracy of its readings.
AUTO POWER OFF
The meter will automatically turn off after 15 minutes of inactivity. This will conserve battery energy. To turn the meter on after an Auto Power OFF, simply turn the rotary switch to OFF and then back to the desired setting.
12
Model EX330 - Version 4.0 – 05/06
Specifications
Non-contact AC
Voltage detector
DC Voltage
(V DC)
100 to
600VAC
Resolution & accuracy do not apply since the meter does not display the voltage in this mode. The lamp at the top of the meter’s display flashes when voltage is sensed and an audible warning will sound.
400mV 0.1mV ±(0.5% reading + 2 digits)
4V 0.001V
40V 0.01V
±(1.0% reading + 2 digits)
AC Voltage
(V AC)
(50 / 60Hz)
400V 0.1V
600V 1V
400mV 0.1mV
±(1.5% reading + 2 digits)
±(1.0% reading + 30 digits)
4V 0.001mV
40V 0.01V
400V 0.1V
±(1.5% reading + 3 digits)
DC Current
(A DC)
AC Current
(A AC)
(50 / 60Hz)
Resistance
600V 1V
400
µA 0.1µA
4000
µA 1µA
40mA 0.01mA
±(2.0% reading + 4 digits
±(1.0% reading + 3 digits)
±(1.5% reading + 3 digits)
400mA 0.1mA
10A 0.01A ±(2.5% reading + 5 digits)
400
µA 0.1µA ±(1.5% reading + 5 digits)
4000
µA 1µA
40mA 0.01mA
400mA 0.1mA
±(1.8% reading + 5 digits)
10A 0.01A ±(3.0% reading + 7 digits)
400
Ω 0.1Ω ±(1.2% reading + 4 digits)
4k
Ω
1
Ω
40k
Ω 0.01kΩ
400k
Ω 0.1kΩ
4M
Ω 0.001MΩ
40M
Ω 0.01MΩ
±(1.2% reading + 2 digits)
±(2.0% reading + 3 digits)
Capacitance
40nF 0.01nF
400nF 0.1nF
4
µF 0.001µF
40
µF 0.01µF
200
µF 0.1µF
±(2.5% reading + 4 digits)
±(3.5% reading + 4 digits)
±(3.5% reading + 10 digits)
13
Model EX330 - Version 4.0 – 05/06
Frequency
Duty Cycle
Temp (type-K)
5.000Hz 0.001Hz
50.00Hz 0.01Hz
500.0Hz 0.1Hz
5.000kHz 0.001kHz
50.00kHz 0.01kHz
500.0kHz 0.1kHz
5.000MHz 0.001MHz
10.00MHz 0.01MHz
±(0.1% reading + 1 digits)
Sensitivity: 0.8V rms min. @ 20% to 80% duty cycle and <100kHz; 5Vrms min @ 20% to
80% duty cycle and > 100kHz.
0.1 to 99.9% 0.1% ±(1.2% reading + 2 digits)
For duty cycle the pulse width range is 100µs - 100ms (Frequency: 5Hz to 150kHz)
-4 to 1382
°F 1°F
-20 to 750
°C 1°C
±(3.0% reading + 8 digits)
(probe accuracy not included)
NOTES:
Accuracy specifications consist of two elements:
• (% reading) – This is the accuracy of the measurement circuit.
• (+ digits) – This is the accuracy of the analog to digital converter.
Accuracy is stated at 65 o
F to 83 o
F (18 o
C to 28 o
C) and less than 75% RH.
14
Model EX330 - Version 4.0 – 05/06
General Specifications
Diode Test
Continuity Check
Temperature Sensor
Test current: 0.3mA max., Open circuit voltage: 1.5V DC typ.
Audible signal will sound if the resistance is less than 100
Ω
Requires type K thermocouple
Input Impedance
10MΩ (VDC & VAC)
AC Bandwidth
Display
Overrange indication
Auto Power Off
Polarity
50 / 60Hz
4000 count (0 to 3999 digits) backlit LCD
For all functions “OL” is displayed (Note: For ACV measurements only, an audible beep will also sound)
After 15 minutes (approximately) of inactivity
No indication for positive; Minus (-) sign for negative
Measurement Rate
Low Battery Indication
Battery
Fuses
2 times per second, nominal
“ ”
is displayed if battery voltage is critically low
Two (2) 1.5V ‘AAA’ batteries mA, µA ranges: 500mA/250V fast blow;
’A’ range: 10A/250V fast blow
Operating Temperature
32ºF to 122ºF (0ºC to 50ºC)
Storage Temperature
-4 o
F to 140 o
F (-20 o
C to 60 o
C)
Operating Humidity
<70% RH
Storage Humidity
Operating Altitude
Weight
Size
Approvals
Safety
UL LISTED
<80% RH
7000ft. (2000meters) maximum.
9.17 oz (260g) (includes holster).
5.8” x 2.9” x 1.6” (147 x 76 x 42mm) (includes holster)
UL, CE
This meter is intended for indoor use and protected, against the users, by double insulation per EN61010-1 and IEC61010-
1 2nd Edition (2001) to CAT II 1000V & CAT III 600V; Pollution
Degree 2. The meter also meets UL 61010-1, Second Edition
(2004), CAN/CSA C22.2 No. 61010-1, Second Edition (2004), and UL 61010B-2-031, First Edition (2003)
The UL mark does not indicate that this product has been evaluated for the accuracy of its readings.
PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY
OVERVOLTAGE CATEGORY I
Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in which measures are taken to limit the transient overvoltages to an appropriate low level.
Note – Examples include protected electronic circuits.
OVERVOLTAGE CATEGORY II
Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from the fixed installation.
Note – Examples include household, office, and laboratory appliances.
OVERVOLTAGE CATEGORY III
Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations.
Note – Examples include switches in the fixed installation and some equipment for industrial use with permanent connection to the fixed installation.
OVERVOLTAGE CATEGORY IV
Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the installation.
Note – Examples include electricity meters and primary over-current protection equipment
15
Model EX330 - Version 4.0 – 05/06
Warranty
EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date of shipment (a six month limited warranty applies on sensors and cables). If it should become necessary to return the instrument for service during or beyond the warranty period, contact the Customer Service Department at (781) 890-7440 ext. 210 for authorization or visit our website at www.extech.com. A Return Authorization (RA) number must be issued before any product is returned to Extech. The sender is responsible for shipping charges, freight, insurance and proper packaging to prevent damage in transit. This warranty does not apply to defects resulting from action of the user such as misuse, improper wiring, operation outside of specification, improper maintenance or repair, or unauthorized modification. Extech specifically disclaims any implied warranties or merchantability or fitness for a specific purpose and will not be liable for any direct, indirect, incidental or consequential damages.
Extech's total liability is limited to repair or replacement of the product. The warranty set forth above is inclusive and no other warranty, whether written or oral, is expressed or implied.
Calibration and Repair Services
Extech offers complete repair and calibration services for most of the products we sell. For periodic calibration, NIST certification or repair for most Extech products, call customer service for details. Extech recommends that calibration be performed on an annual basis to ensure calibration integrity.
Support line (781) 890-7440
Technical support: Extension 200; E-mail: [email protected]
Repair & Returns: Extension 210; E-mail: [email protected]
Product specifications subject to change without notice
For the latest version of this User’s Guide, Software updates, and other up-to-the-minute product information, visit our website: www.extech.com
Extech Instruments Corporation, 285 Bear Hill Rd., Waltham, MA 02451
Copyright
© 2005 Extech Instruments Corporation.
All rights reserved including the right of reproduction in whole or in part in any form.
16
Model EX330 - Version 4.0 – 05/06
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement