Assembly instructions thermoMETER CS


Add to my manuals
2 Pages

advertisement

Assembly instructions thermoMETER CS | Manualzz

 

 

Assembly Instructions thermoMETER CS

 

 

 

 

 

 

The sensors of the CS series are non-contact infrared temperature sensors.

They calculate the surface temperature based on the emitted infrared energy of objects.

You can download a detailed instruction manual as PDF file from our web site under http://www.micro-epsilon.de/download/manuals/man--thermoMETER-CS--com.pdf

Scope of Supply

ƒ CS incl. connection cable, two mounting nuts and assembly instructions

Cleaning

Blow off loose particles using clean compressed air. The lens surface can be cleaned with a soft, humid tissue moistened with water or a water based glass cleaner.

Factory Default Settings

The unit has the following presetting at time of delivery:

CS

Transmission:

Average

CSTK

Ambient temperature source: internal (head) internal (head)

Status-LED function:

Input green):

Self diagnostic

inactive

Self diagnostic

inactive

Output (OUT/ yellow):

Temperature range:

Output voltage:

Thermocouple

Vcc

Signal processing:

Calibration: mV output

0 ... 400 °C

0 ... 4.0 V

Hold mode: off

Gain 1,000/ Offset 0,0 inactive

-40 … 1030 °C according to t/c type K

Hold mode: off

Gain 1,000/ Offset 0,0

 

 

 

 

 

 

 

1) only at model CSTK / The t/c wires are indicated with an additional cable marker to avoid wrong connections due to the identical cable colours of other wires (white, green).

Technical Data

Ambient temperature

Storage temperature

Relative humidity

Dimensions

Cable length

Output impedances

-20 ... 80 °C

-40 ... 85 °C

10 ... 95 %, non condensing

M12x1, 85 mm long

1 m (standard), 3 m, 8 m, 15 m min. 10 kΩ load impedance

 

 

 

 

Power supply

Temperature range

Spectral range

 

Accuracy 1) 2)

5 ... 30 VDC

-40 ... 1030 °C (scalable via software)

8 ... 14 µm

±1,5 °C or ±1,5 % of reading (whichever is greater)

1)

2)

at ambient temperature 23 ±5 °C and object temperatures > 0 °C

Accuracy for thermocouple output: ±2,5 °C or ±1 %

LED Functions

The green LED can be programmed for the following functions. For the programming the USB adapter cable incl. software (option) is necessary. The factory default setting for the LED is self diagnostic.

LED Alarm LED lights up if the object temperature exceeds or deceeds an alarm threshold

Automatic aiming support Sighting feature for an accurate aiming of the CS to hot or cold objects

Self diagnostic LED is indicating different states of the sensor

Temperature Code indication Indication of the object temperature via the LED

Off

Mechanical Installation

The CS is equipped with a metric M12x1 thread and can be installed either directly via the sensor thread or with the help of the both hex nuts (standard) to the mounting bracket available.

Optical Chart CS (15:1) -2-

 

 

Assembly Instructions thermoMETER CS

Electrical Installation

Please use a stabilized power supply unit with an output voltage in the range of 5 – 30 VDC which can supply 100 mA. The residual ripple should be max. 200 mV.

Analog device (mV output on OUT pin)

Analog device (thermocouple type K output on OUT t/c K pins / only at model CSTK)

Software (Optional)

Installation

Insert the installation CD into the according drive on your computer. If the autorun option is activated the installation wizard will start automatically.

Otherwise please start setup.exe from the CD-ROM. Follow the instructions of the wizard until the installation is finished.

The installation wizard will place a launch icon on the desktop and in the start menu.

If you want to uninstall the software from your system please use the uninstall icon in the start menu.

Main Features:

ƒ Graphic display for temperature trends and automatic data logging for analysis and documentation

ƒ Complete sensor setup and remote controlling

ƒ Adjustment of signal processing functions

ƒ Programming of outputs and functional inputs

CE-Conformity

The product complies with the following standards:

EMC: EN 61326-1: 2006 (Basic requirements)

EN

EN 2001

This product is in conformity with Directive 2011/65/EU of the European Parliament and of the Council of 8

June 2011 on the restriction of the use of certain hazardous substances in electrical and electronic equipment.

Digital Communication

For a digital communication the optional USB programming kit is required.

The sensor is offering two ways of digital communication:

ƒ bidirectional

ƒ unidirectional communication (burst mode – the sensor is sending data only)

Open Collector Output

The open collector output is an additional alarm output on the CS and can control an external relay e.g. In addition the analog output can be used simultaneously.

-3-

Warranty

All components of the device have been checked and tested for perfect function in the factory. In the unlikely event that errors should occur despite our thorough quality control, this should be reported immediately to

MICRO-EPSILON. The warranty period lasts 12 months following the day of shipment. Defective parts, except wear parts, will be repaired or replaced free of charge within this period if you return the device free of cost to MICRO-EPSILON. This warranty does not apply to damage resulting from abuse of the equipment and devices, from forceful handling or installation of the devices or from repair or modifications performed by third parties. No other claims, except as warranted, are accepted. The terms of the purchasing contract apply in full. MICRO-EPSILON will specifically not be responsible for eventual consequential damages. MICRO-

EPSILON always strives to supply the customers with the finest and most advanced equipment.

Development and refinement is therefore performed continuously and the right to design changes without prior notice is accordingly reserved. For translations in other languages, the data and statements in the

German language operation manual are to be taken as authoritative.

MICRO-EPSILON Messtechnik GmbH & Co. KG

Königbacher Straße 15

94496 Ortenburg

Tel. 0 85 42/1 68-0

Fax 0 85 42/1 68 90 [email protected] / www.micro-epsilon.com

X9771203-A011123HDR

*X9771203-A01*

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

advertisement