MANUAL PCE-SST 60-150
[email protected]
1 Introduction ........................................................................................................................... 3
2 keypad functions ................................................................................................................... 3
3 Display status ....................................................................................................................... 3
4 Technical Description ........................................................................................................... 4
4.1 Load cells to indicator D-B 9 Pin plug .......................................................................................... 4
4.2 Load cells to indicator 5 Pin plug .................................................................................................. 4
4.3 RS-232C D-B 9 .............................................................................................................................. 4
4.4 Continuous ASCII RS-232 data output format .............................................................................. 4
4.4.1 Automatic Output........................................................................................................................ 4
4.4.1.1 Address: Adr = 00 .................................................................................................................... 4
4.4.1.2 Address: Adr = 99 .................................................................................................................... 4
4.4.1.3 Manual and Automatic Printing Output ................................................................................... 4
Address: Adr = 01—98 ........................................................................................................................ 4
5 Power ................................................................................................................................... 5
6 Configuration ........................................................................................................................ 5
7 Calibration ............................................................................................................................ 6
8 Zero ...................................................................................................................................... 6
9 Tare ...................................................................................................................................... 6
9.1 Digital tare ..................................................................................................................................... 6
9.2 Acquire tare .................................................................................................................................... 6
9.3 Remove tare ................................................................................................................................... 6
10 Manual weight accumulation .......................................................................................................... 7
11 High resolution display mode ......................................................................................................... 7
12 Automatic Memory Accumulation ................................................................................................. 7
13 Clear memory ..................................................................................................................... 8
14 How to set datum ................................................................................................................ 8
15 Battery capacity .................................................................................................................. 9
16 Unit of weighing:( kg and lb ) .............................................................................................. 9
17 Connection to Mini-printers ................................................................................................. 9
18 Precaution .......................................................................................................................... 9
19 Services ............................................................................................................................ 10
[email protected]
1 Introduction
·Simple operation
·Keyboard calibration and configuration
·Weighing up to 1:15,000 divisions
·Excitation voltage: DC 5V; Up to 4,350 Ω load cells can be used
·Manual or automatic memory accumulation facility
·High Resolution (×10) Display
·Selectable digital filtering
·Selectable automatic power off
·6-bit Display: 0.8 inches LED
·Selectable display resolution:
1/2/5/0.1/0.2/0.5/0.01/0.02/0.05/0.001/0.002/0.005/10/20/50/100/200/500/0.10/0.20/0.50/0.010/0.020/0.050
·RS–232C The interface , optional RS–485 interface 1200 2400 4800 9600 Baud rate continuous ASCII data
output
·External Power supply: 230V(115V) AC. Internal rechargeable 6V DC Battery
·Operating temperature: 0 to 40 .Storage temperature: -25º to 55º
·Relative humidity: ≤ 85% non–condensing
·Dimensions: 256mm wide × 165mm high ×130mm deep
2 keypad functions
【STOP】 Whenever the indicator is dead locked or keeps beeping accompanied by abnormal
【①】 On/Off function
【∴】 Manual accumulation function
【﹡】 Function selection during normal operation and configuration
【→】 Move the flashing digit to the right during configuration or setting Preset tare
【↑】 The display is temporarily set to high resolution increment the flashing digit during configuration or
setting preset tare;
【→0←】Zero the display, set the zero point or enter a tare value
3 Display status
〖AC〗
Mains power is applied to the indicator
〖→T←〗
A weight has been tare display is showing the net weight
〖
Battery capacity less than 30%
〗
〖→0←〗
The scale is Zero
〖-AUTO-〗 The automatic accumulation function is active
〖◣◢〗
The Weight is stable
〖- 0〗 The unit of weighing is lb
〖
0.〗 ×10 The display is temporarily set to high resolution
[email protected]
4 Technical Description
4.1 Load cells to indicator D-B 9 Pin plug
4.2 Load cells to indicator 5 Pin plug
I ndi cat or
+E 1
+S 2
-S 3
-E 4
GND 5
Load cel l
+Exci t at i on
+Si gnal
- Si gnal
- Exci t at i on
4.3 RS-232C D-B 9
Pin 3 Output (TXD)
Pin 5 Signal Ground(GND)
4.4 Continuous ASCII RS-232 data output format
4.4.1 Automatic Output
4.4.1.1 Address: Adr = 00
The ASCII data format is “=,X1,X2, X3,X4,X5,X6”
<stx> =,X1 , X2, X3 , X4, X5, X6 <cr>
X1, X2, X3, X4, X5, X6 is weight data.
If the weight is 〖 100.00〗kg, the continuous output is = 00.001 = 00.001=
4.4.1.2 Address: Adr = 99
The ASCII data format is “=,X6,X5,X4,X3,X2,X1”
<stx> =,X6 , X5, X4 , X3, X2, X1 <cr>
X6, X5, X4, X3, X2, X1 is weight data..
If the weight is 〖 100.00〗kg, the continuous output is = 100.00 = 100.00=
4.4.1.3 Manual and Automatic Printing Output
Address: Adr = 01—98
GND 5
[email protected]
5 Power
In power off states, press 【①】key turn on the indicator . The indicator will check the LED and display battery
capacity〖bPt 82〗for 1.5 second.
In power on states, press 【①】 key for two seconds will display battery capacity 〖bPt 82〗 for 1.5 second
and then powers off the indicator.
6 Configuration
Connect load cells to the indicator and set following configuration parameters
Step
1
Operation
Press
【→0←】
【①】
Displaying
Self test from
〖UEr
2.6〗〖0〗to〖9
〗〖
2
3
In power off states, press 【→0←】key and
【 】 turn on power, segment check, and display
〖UE--2.6〗edition No 1.5 second.
0〗
Enter of the scale
Enter the configuration setting modal
Press【﹡】
〖CAL SP〗
Press【﹡】
〖-SEt-〗
Press【→】
〖d 1〗
The number of scale divisions selected
Press【↑】
〖d 2〗
0.001-0.002-0.005-10-20-50-100-200-500-0.10-0.200.50-0.010-0.020-0.050-1-2-5-0.1-0.2-0.5-0.01-0.020.05 For example:d = 0.1
Press【↑】
4
Contents
〖d 0.1〗
Press【﹡】
〖 6000〗
Press【→】
〖000000〗
Sets scale F·S
Press【→】
Press【→】
5
Press【↑】Pr
〖001000〗
ess【↑】
〖002000〗
Press【↑】
〖003000〗
Press【﹡】
Sets digit Filter parameters :00-99
The display will updata faster and filter faster as
〖FLt 00〗
The filter parameter is changed from 99-00
For example: FLt = 00
Sets Automatic Power Off function
〖AUtP00〗
AUtP=00 Not automatic power off
〖AUtP00〗
AUtP=01 Automatic power off
digit
express the choice of zero trace range (1--9):
〖AUtP00〗
:0.4 d 2:0.8 d 3:1.2 d 4:1.6 d 5:2d
9:3.6 d
e choice of zero set
o set at start operation
decimal digit 1 zero set at start operation 20%FS
For example:AUtP=10
(AUtP=10 when leaving the factory)
〖AUtP10〗
continuous output :Adr=00
〖Adr 00〗
Press【﹡】
〖b 2400〗
Baud rate range:1200→2400→4800→9600
Press【↑】
〖b 4800〗
For example:b=4800
Press【﹡】
〖 1000〗
No :1 setpoint output Weight<1000 display LO
Press【﹡】
Press【→】
6
Press【﹡】
Press【→】P
ress【→】
Press【↑】
7
8
9
For example: F·S=3000
〖FLt 10〗
[email protected]
10
11
Press【﹡】
Press【﹡】
〖 2000〗
〖
0〗
No :2 setpoint output
1000< Weight<2000 display OK
Weight>2000 display HI
Press to confirm configurations and go to calibration menu
★ Note: one time accumulation is allowed for weighing once. Following accumulation is allowed for weighing
only when displayed value is below 20 d.
☆ Optional LO—OK—HI and two setpoint output PCB
☆ Press【→】key for two seconds, LCD display for LED are turned on(A1GB-L)
7 Calibration
Calibration should be done after setting the parameters
Step
Operation
Displaying
Comments
1
Enters
calibration
Press【﹡】
〖CAL SP〗
Zero the scale
Press【→】
〖CAL 00〗
2
Starts zero calibration and wait for calibration to
Press【﹡】
〖------〗
complete
〖 3000〗
3
Load the standard
Starts calibration and wait for calibration to
〖------〗
weight for F·S on the
complete
〖 3000〗
platform press【﹡】
Pressed【﹡】key three times during calibration or configuration, the indicator will display. To view the A/D counts,
press the 【→】key when 〖-A-d-〗is displayed press the【﹡】key to return to weighing mode.
8 Zero
When the weight is stable, Press 【→0←】key for two seconds to set the zero point and zero the display. the
【→0←】status LED is turned on.
9 Tare
9.1 Digital tare
press【→】key , set tare with【→】and【↑】key , then press【→0←】key , the input data is tare , Tare status
LED is turned on.
9.2 Acquire tare
When Tare status LED is off and the weight is stable, press the 【→0←】key to acquire tare and switch to net
mode. The Tare status LED is turned on.
9.3 Remove tare
When Tare status LED is on, press【→0←】key will switch to gross mode and removed tare. The Tare status
LED is off.
[email protected]
10 Manual weight accumulation
When weight is stable, press【∴】key to accumulation the current weight to the total weight, The total
number of accumulation will be displayed for 〖n 12〗1.5 seconds.
11 High resolution display mode
In this display mode , press 【↑】, switch to high resolution display mode. (10 times normal display) The last
decimal point is light on. Press【↑】key return to normal weight display mode.
12 Automatic Memory Accumulation
Selection of manual/automatic accumulation function (Selection of manual /automatic print function). Selection
of animal scale, peak value retain and counting function.
Step
1
2
Operation
Displaying
Contents
Press【﹡】
〖n 12〗
to display times of accumulation
Press【﹡】
〖AUt 0〗
the selection of manual/automatic accumulation
AUt = 0, manual accumulation
AUt = 1, automatic accumulation and print when weight is added,
〖AUTO〗indicator is lit.
AUt=2,automatic memorize displayed value when weight is added.
Accumulate and print final stable values after load down to below 20d.
〖AUTO〗indicator is lit.
AUt=3,dynamic weighing method. At weighing >20d: the buzzer
sounds “du”and lock is displayed for 6 seconds. When a new
data is fixed, lock is displayed again for 6 seconds. Then lock
is released for weighing <20d; automatic accumulation and
print. Suggest FLt>30, 〖AUTO〗indicator is lit.
AUt=4, peak value fixed weighing method. At weighing>20d, the
buzzer sounds“du”and lock is displayed.
When weighing <20d, fixed data displays with flash,
automatic accumulation and print. Lock can be released by pressing
any key. 〖AUTO〗indicator is lit.
AUt=5, dynamic weighing method. Manual accumulation and print.
AUt=6, peak value fixed weighing method. Manual accumulation and
print.
AUt=7, counting function. note
3
Press【↑】
〖AUt 0〗
digit displays with flash
4
Press【↑】
〖AUt 0〗
Press【↑】
〖AUt 1〗
Sets F·S
Move binking digit to the right bit
Press【↑】
〖AUt
e.g. AUt=3 expresses dynamic weighing method
2〗〖AUt
3〗
5
Press【﹡】
〖
0〗
return to normal weighing status
[email protected]
note: (1) sampling: When net weight on scale is zero (tare can be removed by pressing tare key if net
weight is not zero), the sample , which must be <200 pieces, i.e. between 1 to 199, is put on the scale. Press
【→】and 【↑】, input quantity of the sample (e.g.30), 〖Cnt030〗 is displayed. Press 【﹡】, confirm the
completion of sampling. Weighing status is redisplayed. Sampling is memorized even with power off.
Step
1
2
operation
display
description
place sample
place selected sample, weight: 27, quantity:30
press
〖Cnt000〗〖Cnt000〗 ready to input sample’s quantity
decimal digit display with flash
【→】press
【→】
3
press【↑】
〖Cnt010〗
press【↑】
〖Cnt020〗
press【↑】
〖Cnt030〗
4
press 【﹡】
〖
5
press【↑】
〖C 30〗
27〗
display sample’s weight:27, 【*】is a confirmation key,
sample collection completed
display sample’s quantity, 【↑】is change-over key
between weight and quantity display
(2) Counting operation: place the object on scale, weight is displayed, press【↑】,〖C 255〗is displayed, and the
display changes over to the quantity of the object. When the display is stable, press 【∴】, accumulate the weight
and quantity. Accumulation can be done only at counting status.
Step
operation
display
description
1
place object
object
weight:230
〖 230〗
2
press【↑】
〖C 255〗
object quantity:255
3
press 【∴】
〖n
display after 1.5 seconds
at counting status
4〗
〖C 255〗
(3) Accumulate inquiries and delete: both at weighing stage and counting status
Step
1
operation
display
press 【﹡】
〖C 1203〗
description
display the total quantity of the object:1203
2
press【↑】
〖H
display accumulated weight 4 digits higher
3
press【↑】
〖L 1085〗
display accumulated weight 4 digits lower=1085
4
press【↑】
〖C 1203〗
back to counting status
5
press【→0←】 〖C
0〗
0〗
delete accumulated quantity
13 Clear memory
Press【→0←】key If the memory has previously been cleared the display will show 〖n 0〗followed by zero
weight when the【﹡】key is pressed.
14 How to set datum
Press【→】key move and blinks the current digit to.
Press【↑】key increment the current digit to the next available value
[email protected]
15 Battery capacity
When the indictor is being power on/off. battery capacity will be displayed 〖PBt 85〗for1.5 seconds
·When the battery is capacity less than 20%,the display will start to blink, power off the indictor to avoid over
discharging battery or connect the external main. Power supply
·When the battery capacity ≤ 10%,the indictor automatic turn off to avoid battery over discharge
·In auto power off mode, If weight is stable and no key operation for more than >3 mints, The indictor displays
〖- 〗to conserve battery
·When auto power off setting is action, If weight is stable and no key operation > 30 mints automatic turn off
power
·A full charged battery life is approximately 30 hours
16 Unit of weighing:( kg and lb )
kg or lb is selectable. The unit of weighing is kg normally. You can change it to lb by pressing and holding
【﹡】key 2 seconds at least.
17 Connection to Mini-printers
Connect serial port printer Type Up-16TS as follows:
Indicator
Serial Port Printer
9 pin hole
25 pin needle
3
TXD
2
5
GND
7
Note: Before connection to printer, communication address is set as Adr=01; baud rate is set as b=2400.
Print operation with serial port printer is as follows:
17.1.Print: at weighing status, weighing data >20d and display is stable, press 【∴】, weighing sheet is printed
out. The second printing can be operated only when the weighing data is back to <20d.
17.2 Accumulated print: at weighing status, press 【﹡】, then press 【∴】, accumulated printing can be
operated.
17.3 Set to be automatic accumulation status, i.e. automatic print. At weighing status, weighing data>20d and
display is stable, weighing sheet is printed out. The second printing can be operated only when the weighing data
is back to <20d and more weight is loaded.
Attached with print sample
Normal print
accumulation print
NO:1 (serial No.)
NO:
9 (number of accumulation)
Gross: 3940kg (gross weight)
W: 8225kg (accumulated weight)
Tare: 2000kg (tare weight)
Net: 1940kg (net weight)
18 Precaution
·Indicator should be far away from heat resource while using.
·Do not place the indicator in the dusty surroundings or the site vibrant.
·Cannot use full capacity. Over load stops hitting platform support is not permitted.
·To ensure to keep out of chemical erosion, Operating temperature range will be
–10-50 ,relative humidity is no less than 85%,without any corrupt gas in air.
·Never pour the water into the indicator.
·Housing, head pallet, wire connector should be sealed entirely. Users do not open sealed device or connect with
wire without any expert advice. In case any malfunction of indicator occurs, please sent the indicator for
maintenance.
·The indicator will charge the internal battery at all times when it is connected to the main power.
[email protected]
19 Services
Offer a full range of technical services such as on site and workshop repair, Preventative maintenance and
calibration facilities.
(Battery is not in the services range )
In this direction will find a vision of the measurement technique:
http://www.industrial-needs.com/measuring-instruments.htm
NOTE: "This instrument doesn’t have ATEX protection, so it should not be used in potentially explosive
atmospheres (powder, flammable gases)."