Burster 4462 Datasheet

Burster 4462 Datasheet
High Precision Calibration Source
Code:
4462 EN
For voltage, current and thermocouples
Delivery:
upon request
DIGISTANT®
Warranty:
24 months
4462 EN
Model 4462
n
High precision current and
n
voltage source ± 52 mA, ± 30 V
Option: ± 22 mA, ± 60 V
Precision simulation for all conventional
n
n
thermocouples types (optional)
Error limit 0.003 % Rdg.
RS232/USB and optional IEEE488 interface
n
Application
The precision calibration unit combines high accuracy, low
drift, low noise and superior long-term stability with multiple
functionality and simple operation.
Ramps, ∆+/∆-, and multiple setpoint storage make the operation of the device easier for the user.
For that reason the application possibilities are many:
► Calibration of current and DC-voltage meters
► Precise testing of thermocouple temperature measuring instruments
► Calibration of controllers, sensor, detection devices and other devices used in process control
► Open-loop process control with the aid of integrated ramp function
The DIGISTANT® model 4462 can be used both as a standalone table-top device, as well as in automatic, computer-assisted manufacturing and testing systems.
Technical changes reserved Latest updates of data sheet always under www.burster.com
Current "SINK"
Description
It is possible to set current of ± 200 nA ... ± 52 mA, voltage of
± 1 µV ... ± 30 V and, optionally, temperature setpoint value of
14 thermocouples types.
The output value is fed back via the sense line to eliminate
voltage drops across the measuring leads.
The device has an adjustable current/voltage limitation. An
external voltage divider of 1 up to 1:1000 can be considered
internally.
With the thermovoltage sourcing option you can enter °C,
°F and K, the temperature scales ITS 90 or IPTS 68 and the
comparison point mode constant/external. Furthermore, when
sourcing thermocouples a calibrated external comparison
point can be used, whereby the data for calibration in the
device can be taken into consideration.
Indication of the source value is carried out in large 12 mm
figures on an illuminated graphics-LCD.
The device can be operated both via the keyboard as well as
the interface.
burster praezisionsmesstechnik gmbh & co kg . Tel. +49-7224-6450 . Fax 64588
Talstr. 1-5 . D-76593 Gernsbach . www.burster.com . [email protected]
4462 EN - 2
Source main menu
External sense
Divider factor of the external divider.
The voltage value is fed with the value
which should appear on the output.
Manual range selection
����� ��� ���������
������� �������������
Current limitation
�����������
A new delta value
can be entered
��
����
���� ����
Currently set
delta value
Source value
Source value
will be negative
0 are spent.
The last source value
remains stored
Source value will be
positive
Operating Examples
Ramp 1 Configuration menu
������������������
����������������
�����������������
�����������������
�����������������
���������������������
Ramp function:
► Ramp 1
with constant delta values and delta time
► Ramp 2
with variable delta values and internal time.
The ramp function allows single or repeated outputs in sawtooth or triangular form. The
number of steps can set from 0 to 99 (0 is continuos). The START, END and DELTA values
can be entered in µV, mV, V, mA and temperature values. DELTA time is displayed as
shown in the menu.
����
Current/voltage, limit setting
����������
�������������
���������������
Current/voltage limit:
If a voltage or temperature value is given, the current limit is automatically active. If current
is sourced the voltage limit is active. The voltage limit ranges from 1 V to 32 V and the
current limit ranges from 1 mA to 55 mA.
������������������
���� ����
TC/Temperature menu
��������������������
����������������
�����������������
���� ����
�������������
��������
��������������
��������������
�����������������
������������������
���
Technical changes reserved Latest updates of data sheet always under www.burster.com
��
���� ����
Optionally the thermocouples types R,
S, B, J ,T, E, K, U, L, N, M, C, D and
G2 can be simulated. For the "manual"
reference junction at 0 °C the accuracy
depends on the thermocouples model
starting at 0.1 K.
The connection occurs direct at the
standard terminals or "externally" via
an external, attachable reference junction model 4485-V001, at which the
temperature is detected with a Pt100
sensor (see application 1)
burster praezisionsmesstechnik gmbh & co kg . Tel. +49-7224-6450 . Fax 64588
Talstr. 1-5 . D-76593 Gernsbach . www.burster.com . [email protected]
724-004462EN-5072-111521
�����������������
������������
Pt100 scale
(measurement with external RJ)
4462 EN - 3
Technical Data
Voltage source
3 V 10 µV 0.003%(to ± 450 mV) +20 µV 8 ppm/K +1 µV/K
(> ± 450 mV) +110 µV
300mV 1 µV 0.003% (to ± 45 mV) +3 µV 8 ppm/K+0,35µV/K
(>±45 mV) +11 µ
Option: 60 V (Range 30 V will be dropped)
Range Reso-
Error limits
TC with resp,
±
lution
at 23°C ± of reading
to 23 °C
60 V 0.2mV 0.003%
(to ± 9 V) +500 µV 8 ppm/K+10 µV/K
(>± 9 V)
+2.2 mV
Output current: max. 52 mA at 30 V, source resistance < 10 mΩ
(max. 22 mA at 60 V, model -VXX1)
Current source
Range Reso-
±
lution
Error limits
at 23°C ± of reading
TC with resp.
to 23 °C
52 mA 200 nA 0.007% (to ± 7.5 mA) +0.6 µA10 ppm/K+10 nA/K
(22 mA)
(>± 7.5 mA) +3 µA
Burden voltage: max. 30 V at 52 mA, source resistance > 500 MΩ
Confidence coefficient for the specified errors: 95 % (K=2).
(Burden voltage: max. 60 V at 22 mA, model -VXX1)
R
S
B
J
T
E
K
U
L
N
M
C
D
G2
Range
Error (K)*
- 50.0 °C ... 1768 °C
0.4 (+ 250 ... 1768 °C)
- 50.0 °C ... 1768 °C
0.4 (+ 350 ... 1768 °C)
0.0 °C ... 1820 °C
0.5 (+ 800 ... 1820 °C)
- 210 °C ... 1200 °C
0.2 (- 210 ... 900 °C)
- 270 °C ... 400 °C
0.2 (-170 ... 400 °C)
- 270 °C ... 1000 °C
0.2 (- 220 ... 1000 °C)
- 270 °C ... 1372 °C
0.1
(- 50 ... 800 °C)
- 200 C ... 600 °C
0.3 (- 100 ... 600 °C)
- 200 °C ... 900 °C
0.2 (- 100 ... 750 °C)
- 270 °C ... 1300 °C
0.2 (- 120 ... 1200 °C)
- 50 °C ... 1410 °C
0.1
(- 50 ... 900 °C)
0,0 °C ... 2315 °C
0.2 (+ 100 ... 900 °C)
0,0 °C ... 2315 °C
0.2 ( 300 ... 1100 °C)
0,0 °C ... 2315 °C
0.3 ( 300 ... 2100 °C)
*The errors are defined at "manual" reference junction 0 °C.
Reference junction
EXTERNAL: The temperature are measured with an external Pt100 sensor.
MANUAL: The reference junction temperature is entered manually.
Temperature recording in an external reference junction
or temperature measurement with Pt100
Range
- 200 ... 850 °C
Outputs and terminals on the rear side
Standard RS232C interface:
Optional IEEE488 interface:
Instruction language:
9 pin subminiatur D-socket
Baud rate 300 - 38 400
Protocoll ANSI X 3.28 1976 Subcategory 2.1, A3
24 pin, open collector outputs
(E1) SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP0, DC1, DT1, C0
SCPI, version 1997.0
Order Information
Option: Thermocouples simulation
Model
Relative humidity:
80 % up to 31 °C above linear: decreasing
that temperature to 50 % at Tmax, no condensation
Storage temperature:
- 10 °C ... 60 °C
Device construction:
metal housing in protection class Ι
in accordance with DIN EN 61010 part 1
Power supply:
230 V ± 10 %, 45 Hz ... 65 Hz,
can be changed on device to 115 V
Power requirement:
approx. 30 VA
Dimensions: (L x W x H)
237 x 285 x 151 [mm]
(with handles u = 325 mm)
Weight:
ca. 6 kg
Output:
potential-free
Resolution Current (mA)
Tolerance
0.01 °C approx. 0.6 mA 0.00006*x °C + 0,045 °C
General Technical Data
Long-term stability:U-Drift< 20 ppm / year + 2 µV / year(300 mV)
U-Drift< 20 ppm / year + 6 µV / year
(3 V)
U-Drift< 20 ppm / year +10 µV / year(30/60 V)
Ι-Drift < 70 ppm / year + 0.5µA / year
Warm-up time:
30 minutes, until specified error limit
External divider:
1 to 1010
An external voltage divider can be connected. In this case the divider
function is activated and the division factor of the external divider is
entered. Then the source value is entered as the value that should be
present at the divider’s output.
Display:
graphics LCD display with LED illumination
Visual field:
56.3 mm x 38 mm, resolution 128 x 64 dots
Sockets:
+ output, - output, + sensor, - sensor,
, gold-plated
4 mm terminals and 6 pin LEMO socket 1B for the optional
Pt100 connection.
Potential to ground:
≤ 50 V between analog ground and ground
Temperature range:
5 °C ... 23 °C ... 40 °C
Technical changes reserved Latest updates of data sheet always under www.burster.com
DIGISTANT®
Model 4462 - V
Standard with RS232
1 0 0
Additional with IEEE488
2 0 0
Option thermocouple simulation 1 0 RJ temperature recording
(temperature measurement with Pt100)
subsequent installation is possible
Option 60 V/ 22 mA 1
A test certificate with traceability verification is part of the delivery.
Accessories (Option)
4 measuring leads with low thermal voltage Cu/Te
safety connectors, length 1 m
Model 9900-K342
RS232 data cable
for PC connection
Model 9900-K333
interface set consisting of USB/RS232 converter Model 9900-K351
External reference junction
for DIGISTANT® model 4462
for an accurate simulation of thermocouples,
Connection: Miniature thermo plug connection
(For description refer to page 4)
Model 4485-V001
Assembly set suitable for 19" rack mounting Model 2329-Z004
DKD Certificates
DKD Calibration (basic system)
Each range (voltage, current) is calibrated at ±12,5 %, 25 %, 50 % and
90 % of full scale.
Model 44DKD-4462-V100
DKD Calibration (extended system)
Each range (voltage, current) is calibrated at ±12,5 %, 25 %, 50 % and
90 % of full scale. With 2 points for 10 thermocouples, temperature of
the reference junction 0 °C and two points for Pt100.
Model 44DKD-4462-V110
Calibration Certificate for the external reference junction
At 3 points (0 °C, + 23 °C and 40 °C). If the built-in Pt100 of the reference junction is calibrated (NAMAS, DKD or others) and you enter the
probe calibration into the DIGISTANT® model 4462-VX1X the accuracy
of the temperature measurement is ≤ 0.1 K (in the temperature range
+ 15 °C to + 35 °C).
Model 44DKD-4485
burster praezisionsmesstechnik gmbh & co kg . Tel. +49-7224-6450 . Fax 64588
Talstr. 1-5 . D-76593 Gernsbach . www.burster.com . [email protected]
4462 EN
Range Reso-
Error limits
TC with resp.
±
lution
at 23°C ± of reading
to 23 °C
30 V 0.1 mV 0.003% (to ± 4.5 V) +200 µV 8 ppm/K+10 µV/K
(> ± 4.5 V) +1.1mV
4462 EN - 4
External reference junction model 4485-V001 for thermocouples
►
►
►
►
For an accurate simulation of thermocouples
A built-in Pt100 for cold junction compensation
Thermally stable and isolated construction
Plug type: Miniature TC connector
Technical Data
►
►
►
►
►
Limits:
Long term stability:
Insulation resistance between pin
in disconnected status:
Operating temperature range:
Storage temperature range:
Note: ± 0.3 K
typical 0.05 K/year
>= 20 MΩ
0 °C ... 23 °C ... 40 °C
- 10 °C ... 60 °C
Thermo cable and connector cause an additional error.
We recommend to use the class 1.
Application Example
1. Calibration of a PC card with a thermocouple measurement input
Instead of the thermocouple the calibration source DIGISTANT® model 4462 is connected. Using an external DKD-calibrated reference junction the PC card is retraceable calibrated with the optimum accuracy. Up to 14 thermocouples can be selected.
External reference junction
����� � �������
Test object:
Measurement card
with TC input
Plug type:
Miniature TC connector
2. Calibration of measuring system in the medicine engineering
In the sweep function you set different current and voltage
values with individual steps. The output happens once ore
repetitioned in triangular or sawtooth wave.
�
�
Technical changes reserved Latest updates of data sheet always under www.burster.com
burster praezisionsmesstechnik gmbh & co kg . Tel. +49-7224-6450 . Fax 64588
Talstr. 1-5 . D-76593 Gernsbach . www.burster.com . [email protected]
724-004462EN-5072-111521
Synthesis processes
in the production of
medicine require a
careful check.
A highly secured production process is
life saving.