Datasheet HMP110 (english)

Datasheet HMP110 (english)
www.vaisala.com
Vaisala HUMICAP® Humidity and Temperature
Probe HMP110
The HMP110 with excellent stability and high chemical tolerance.
Features/Benefits
HMP110
Several Outputs
▪
Miniature-size humidity
transmitter
▪
Low power consumption
and fast start-up for battery
powered applications
The temperature measurement
is a standard feature, dew point
measurement is optional. Three
standard voltage outputs are
available.
▪
Measurement range:
0 ... 100 %RH; -40 ... +80°C
▪
Cable detachable with
standard M8 quick connector
▪
Reliable: Latest generation
HUMICAP® 180R sensor for
best stability and high chemical
tolerance. IP65 metal housing.
The HMP110 is a trouble-free and
cost-effective humidity transmitter
with high accuracy and good
stability. It is suitable for volume
applications or integration into other
manufacturers’ equipment. The
HMP110 is also suitable for glove
boxes, greenhouses, fermentation
and stability chambers, data loggers,
and incubators.
▪
▪
Optional RS485 digital output
Traceable: Comes with
calibration certificate. ±1.7%RH
measurement accuracy
(0 ... 90%RH)
▪
HMP110R replacement probe
service available for easy
maintenance
▪
Optional dew point calculation
Easy Installation
The probe cable has a screw-on
quick connector for easy installation.
Different cable lengths and
accessories are available.
Low Current Consumption
HMP110 is suitable for batterypowered applications because of its
very low current consumption. It also
has a fast start-up time.
Robust Design
The stainless steel body of the
HMP110 is classified as IP65. Thus,
it survives rough conditions.
The HMP110 has high chemical
tolerance because of the HUMICAP®
180R sensor.
Easy Maintenance
Maintaining measurement
traceability is easy using the
HMP110R replacement probe. We
send you a replacement probe, you
detach the old probe and send it back
to us. In this way the measurement
is available at all times without
interruptions.
Technical Data
Performance
TEMPERature
Measurement range
-40 ... +80 °C
Accuracy over temperature range
0 ... +40 °C,
±0.2 °C
-40 ... 0 °C, +40 ... +80 °C
±0.4 °C
Temperature sensor
Pt1000 RTD Class F0.1 IEC 60751
DEW POINT
Measurement range
-40 ... +80 °C
Accuracy (incl. non-linearity, hysteresis and repeatability)
temperature range
0 ... +40 °C
when dew point depression < 15 °C
±1 °C
when dew point depression 15 ... 25 °C
±2 °C
temperature range
-40 ... 0 °C, +40 ... +80 °C
when dew point depression < 15 °C - dew point
±2 °C
depression = ambient temperature - dew point
Inputs and Outputs
Operating voltage
(Use lowest available operating
voltage to minimize heating)
Current consumption
Start-up time
probes with analog output
5 ... 28 VDC / 8 ... 28 VDC with
5 V output
8 ... 28 VDC with loop power
converter
1 mA average, max. peak 5 mA
4 s at operating voltage
13.5 ... 16.5 VDC
2 s at other valid operating voltages
1s
probes with digital output
Outputs
2 channels
0 ... 1 VDC / 0 ... 2.5 VDC / 0 ... 5 VDC / 1 ... 5 VDC
1-channel loop-power converter (separate module,
compatible with humidity accuracy only)
4 ... 20 mA
digital output (optional)
RS485 2-wire half duplex
External loads
0 ... 1 V
0 ... 2.5 V /0 ... 5 V
Operating Environment
Operating temperature range
-40 ... +80 °C
Electromagnetic compatibility
EN 61326-1: Electrical equipment
for measurement, control and laboratory use – EMC
requirements – for use in industrial locations.
Mechanics
Materials
body
grid filter
cable
Housing classification
Body thread
Cable connector
Weight
probe
probe with 0.3 m cable
stainless steel (AISI 316)
chrome coated ABS plastic
polyurethane or FEP
IP65
MI2x1 / 10 mm
4-pin M8 (IEC 60947-5-2)
17 g
28 g
Options and Accessories
Sensor protection
plastic grid
membrane filter
stainless steel sintered filter
4 ... 20 mA loop power converter
Mounting bracket for converter
Plastic M12 installation nuts, pair
USB cable for PC connection
Probe mounting clamp set, 10 pcs
Probe mounting flange
Connection cables
standard 0.3 m
standard 3 m
80 °C 1.5 m
80 °C 3 m
180 °C 3 m FEP
DRW010522
DRW010525
HM46670SP
UI-CONVERTER-1CB
225979
18350SP
219690
226067
226061
HMP50Z032SP
HMP50Z300SP
225777SP
225229SP
226902SP
Dimensions
Dimensions in mm (inches)
Please contact us at
www.vaisala.com/requestinfo
www.vaisala.com
RL min 10 kΩ
RL min 50 kΩ
(2.8)
71
(0.47)
12
Measurement range
0 ... 100 %RH
Accuracy (incl. non-linearity, hysteresis and repeatability)
temperature range
0 ... +40 °C
0 ... 90 %RH
±1.7 %RH
90 ... 100 %RH
±2.5 %RH
temperature range
-40 ... 0 °C, +40 ... +80 °C
0 ... 90 %RH
±3.0 %RH
90 ... 100 %RH
±4.0 %RH
Factory calibration uncertainty (+20 °C)
± 1.5 %RH
Humidity sensor
Vaisala HUMICAP® 180R
Stability
±2 %RH over 2 years
M12 x1
Relative humidity
Ref. B210852EN-C ©Vaisala 2012
Scan the code for
more information
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertisement