Sampling Accessories

Sampling Accessories
Sampling Accessories
Sampling Accessories
Our modular components can be
easily configured for absorbance,
transmission, reflectance, fluorescence, emission or scattering experiments. These fiber optic sampling
accessories create the optical interface of our modular spectrometer
systems.
And, with so many sampling
accessories to choose from, Ocean
Optics can readily meet the demands
of your unique application needs.
Tip
Collimating lenses are among our
most versatile sampling accessories.
Lenses convert divergent beams of
radiation into a parallel beam and are
optimized for UV-NIR wavelengths.
With collimating lenses, you can
adjust for variables such as FOV,
collection efficiency, spatial
resolution and lens transmission and
acceptance angles.
Sampling Accessories
Sampling Accessories
A Sea of Tools for Sample Handling
Your optimal choice of sampling accessory depends on the
application. Are you measuring absorbance, reflectance or
emission? Is your sample a solid or solution? Does it have a smooth,
even surface or an irregular surface? Do you need to collect
radiation from a particular field of view?
Those are just a few of the many questions that enter into your
sampling accessory decision. And, much like your choice of
spectrometer and optical bench accessories, the key question is a
simple one: Why are you making the measurements?
Example Sampling Accessory Options: Absorbance
With an extensive line of sampling accessories – and the flexibility
to switch among them with ease – you have many options for
tackling your experiment. Consider absorbance: accessories run the
gamut from standard labware such as cuvettes and cuvette holders
to more sophisticated tools such as flow injection analysis devices.
Here are some criteria to consider for your absorbance setup:
Where will you make the measurement?
If you can isolate the sample in a container such as cuvette or test
tube, your options extend to cuvette holders and similar devices.
If you need to make the measurement in situ – in a process stream,
for example -- you’ll need a transmission dip probe (see Fibers and
Probes section) or flow cell that can be embedded into the
sample stream.
What is the optical density of your sample?
The OD of your sample directly affects the pathlength of the device
you’ll need for your measurement. Cuvette holders, for example, are
available with pathlengths of 1 cm and 10 cm. The more optically
dense the sample, the shorter the pathlength that’s required. On
the other hand, low OD solutions require longer pathlengths. Use
Beer’s Law to help you determine the best pathlength.
What’s Your Field of View?
Four of our devices are used to control field of view (FOV) and aperture:
1. Optical Fiber (25° FOV, aperture = fiber diameter)
2. Collimating Lens (0° - 45° FOV, aperture = 3 mm)
3. Cosine Corrector (180° FOV, aperture = 3.9 mm)
4. Integrating Sphere (360° FOV, aperture = 25 mm)
Optical Fiber: 25°
Adjustable Collimating Lens: ~0-45°
What is the sample volume?
The amount of sample you measure also affects your choice of
sampling accessory. We offer microvolume-level cuvettes and
pipetters for applications with sample volumes as low as 2 µl.
Cosine Corrector: ~180°
Integrating Sphere: 360°
Does your sample require continuous flow?
We offer sampling accessories such as flow cells that can be
coupled with pumps, tubing and fittings to stream sample fluids for
absorbance and other measurements.
That’s just absorbance. We also offer nearly 500 sampling accessories for reflectance, emission, fluorescence and other measurements. Our Applications Scientist can help you choose the best
option for your application.
106
www.oceanoptics.com Tel: +1 727-733-2447
Technical Tip
The divergence (α) of a beam focused using
a single lens is: tan(α) = d/f where f is the focal
length of the lens and d is the aperture of the
spectrometer.
Choosing Your Accessory by Measurement Type
We proudly offer a full range of Sampling Accessories that help you obtain accurate, reliable measurements in virtually any environment.
Absorbance and Transmission
1-cm Cuvette Holder
See Page 110
10-cm Cuvette Holder
See Page 110
Longpass Flow Cells
See Page 116
Sampling Accessories
Sampling Accessories
Cuvettes
See Page 113
Fluorescence
4-way Cuvette Holder
See Page 110
Flow Cells
See Page 114
Linear Variable Filters
See Page 127
Emission
Cosine Correctors
See Page 118
LED Power Supply
See Page 118
Integrating Spheres
See Page 119
Collimating Lenses
See Page 108
Reflectance
Diffuse Reflectance Standards
See Page 125
Integrating Spheres
See Page 120
Multifunction Reflection Stage
See Page 123
www.oceanoptics.com Tel: +1 727-733-2447
107
Sampling Accessories
Sampling Accessories
Collimating Lenses
74-Series Lens Fixtures
Our 74-Series Collimating Lens Fixtures are the common fiber optic-coupled lens fixtures used throughout our extensive line of sampling
accessories. They feature an inner barrel threaded for SMA 905 Connectors.
74-UV Collimating Lens
(200-2000 nm) The 74-UV has an f/2 fused silica lens for 200-2000 nm. When focused for collimation, beam divergence is
2° or less, depending on fiber diameter. The 74-UV can be adjusted for UV-VIS or VIS-NIR setups.
74-90-UV Collimating Lens
(200-2000 nm) The 74-90-UV features a mirror located under its cap that reflects light from the collimating lens to 90°.
The mirror is coated with a UV-enhanced aluminum substrate that is >80% reflective from 200-2000 nm.
74-VIS Collimating Lens
(350-2000 nm) The 74-VIS has a BK7 lens suitable for the VIS-NIR. These single-lens systems have the disadvantage of
chromatic aberration, due to dispersion or variation in refractive index with wavelength.
74-ACR Collimating Lens
(350-2000 nm) The 74-ACR has two optical elements cemented together to form an achromatic doublet, optimized to
correct for spherical and chromatic aberrations.
74-DA Collimating Lens
(200-2000 nm) The 74-DA Direct-attach Collimating Lens screws on to a spectrometer’s SMA 905 Connector for increased
light throughput. The lens collects collimated light in a straight path of open air and focuses it onto a spectrometer’s slit.
COL-UV-30 Collimating Lens
(200-2000 nm) This 74-series Collimating Lens is our largest-diameter lens – 30 mm – and couples to SMA 905 connectors
for simple use.
84-Series Lens Fixtures
(200-2000 nm) The 84-Series Collimating Lens is designed for coupling larger free-space beams to optical fibers. The fiber
is coupled to the assembly with an inner 17.85 mm threaded barrel. The barrel positions the fiber ~100 mm from the lens
surface and is turned to achieve a fine focus.
Item
108
Diameter
Focal Length
Material
Range
Operating Temp
Connector
74-UV
5 mm
10 mm
f/2 fused silica Dynasil
200-2000 nm
150 °C
SMA 905, 6.35-mm ferrule, 3/8-24
external thread
74-90-UV
5 mm
10 mm
f/2 fused silica Dynasil
200-2000 nm
120 °C
SMA 905, 6.35-mm ferrule, 3/8-24
external thread
74-VIS
5 mm
10 mm
f/2 BK7 glass
350-2000 nm
150 °C
SMA 905, 6.35-mm ferrule, 3/8-24
external thread
74-ACR
5 mm
10 mm
f/2 BaF10 and FD10 fused silica
350-2000 nm
150 °C
SMA 905, 6.35-mm ferrule, 3/8-24
external thread
74-DA
5 mm
10 mm
f/2 fused silica Dynasil
200-2000 nm
150 °C
SMA 905, 1/4-36 internal thread, 3/8-24
external thread
COL-UV-30
30 mm
30 mm
f/2 fused silica Suprasil
200-2000 nm
200 °C
SMA 905, 6.35-mm ferrule, 1/4-36
external thread
84-UV-25
25.4 mm
100 mm
f/2 fused silica Dynasil
200-2000 nm
70 °C
SMA 905, 6.35-mm ferrule, 1/4-36
external thread
www.oceanoptics.com Tel: +1 727-733-2447
Collimating Lenses
ACH-CUV-VAR
Adjustable Collimating Lens and Cuvette Holder
The ACH-CUV-VAR Adjustable Collimating Lens and Cuvette Holder is
two products in one: a fixture for positioning collimating lenses at
various heights or for holding extra-large or especially thick samples, and
a holder for accepting cuvettes for transmission measurements.
Sampling Accessories
Sampling Accessories
As a lens holder, the ACH-CUV-VAR has an anodized aluminum base and
adjustable mount bars with 3/8-24 threaded holes for collimating lenses
(two 74-UV Collimating Lenses are included). The bars can be adjusted
to accept samples up to ~150 mm thick and the base is marked in 1-mm
intervals as a pathlength guide. The ACH-CUV-VAR has a cuvette holder
component that fits securely between the two mount bars and can
accept cuvettes up to ~150-mm wide.
OPM-SMA
Optical Post Mount
The OPM-SMA is a fixture for mounting 74-Series Collimating Lenses and
SMA 905-terminated optical fibers. The OPM-SMA consists of a 1.5-inch
OD disk with 3/8-24 threads for use with lenses and an adapter for use
with SMA 905-terminated optical fibers. A special adapter that holds an
SMA 905-terminated optical fiber flush against one surface of the
OPM-SMA is included. The OPM-SMA also includes 8-32 (Imperial) and
M6 (metric) threads for attachment to an optical post.
OPM-M
Optical Post Mount Assemblies
We offer four stainless steel optical posts (in 25.4-mm, 50.8-mm,
76.2-mm and 101.6-mm heights) to attach to the OPM-SMA. These posts
have a 12.7 mm OD and screw into optical breadboards via a 1/4”-20
(M6) tapped hole in the bottom of each post. The posts also have a
removable 8-32 (M4) threaded stud.
Item Codes: OPM-1, OPM-2, OPM-3, OPM-4
Specification
OPM-SMA
OPM-M
Dimensions (in mm):
38.1 OD x 10.2 width
30 mm OD x 6.5 width
Weight:
130 g (including post)
220 g (including post)
Collimating lens included:
No
No
Threads:
3/8-24 (lens holder)
8-32 (bore for mounting)
3/8-24 (lens holder)
M4 (bore for mounting)
Material:
Black anodized Al
(mount) and stainless
steel (post)
Black anodized Al
(mount) and stainless
steel (post)
www.oceanoptics.com Tel: +1 727-733-2447
109
Sampling Accessories
Sampling Accessories
Cuvette Holders
CUV-UV
Holder for 1-cm Cuvettes
The CUV-UV Cuvette Holder for 1-cm pathlength cuvettes couples via SMA 905-terminated optical
fibers to Ocean Optics high-sensitivity miniature fiber optic spectrometers and light sources to
create small-footprint spectrophotometric systems for absolute absorbance measurements of
aqueous solutions. This compact cuvette holder is optimized for UV-VIS-NIR.
(~200-2000 nm) applications.
CUV-UV-10
Holder for 10-cm Cuvettes
The CUV-UV-10 Cuvette Holder for 10-cm pathlength cuvettes couples via SMA 905-terminated
optical fibers to Ocean Optics high-sensitivity miniature fiber optic spectrometers and light sources
to create small-footprint spectrophotometric systems for absorbance/transmission measurements of
aqueous solutions and gases. This compact cuvette holder is optimized for UV-VIS-NIR.
(~200-2000 nm) applications.
CUV-ALL
4-way Holder for 1-cm Cuvettes
The CUV-ALL-UV 4-way Cuvette Holder for 1-cm pathlength cuvettes is equipped with fiber optic
couplings at each of four quartz f/2 collimating lenses, which couple to optical fibers to either read
or illuminate the sample. When used with Ocean Optics modular spectrometers and light sources,
CUV-ALL-UV Cuvette Holders can measure absorbance, fluorescence, scattering or any combination
of these optical phenomena.
CUV-FL-DA
Direct-attach Cuvette Holder
The CUV-FL-DA Cuvette Holder attaches directly to Ocean Optics light sources and couples via
SMA 905-terminated optical fibers to our spectrometers, creating an incredibly small-footprint
spectrophotometric system for fluorescence and relative absorbance experiments. The CUV-FL-DA,
optimized for UV-VIS-NIR (200-2000 nm) applications, holds 1-cm square cuvettes.
74-MSP
Mirror Screw Plugs
Designed for use with our fluorescence cuvette holders, our Mirror Screw Plugs are inserted into a
collimating lens port on the cuvette holder to redirect energy back to the sample or back into a
collimating lens. This increases signal collection for fluorescence measurements.
110
Specifications
CUV-UV
CUV-UV-10
CUV-ALL-UV
CUV-FL-DA
Dimensions:
58 mm x 140 mm x 38 mm
97 mm x 147 mm x 40 mm
147 mm x 147 mm x 40 mm
57 mm x 61 mm x 29 mm
Weight:
230 g
1.4 kg
540 g
80 g
Pathlength:
1 cm
10 cm
1 cm
1 cm
Z dimension:
15 mm
15 mm
15 mm
15 mm
Filter slot:
Up to 6 mm, screw clamp
Up to 6 mm, wheel clamp
Up to 6 mm, screw clamp
Up to 6 mm, screw clamp
Water input fittings:
3.175 mm (1/8”) NPT
3.175 mm (1/8”) NPT
3.175 mm (1/8”) NPT
NA
Collimating lens:
2 each 74-UV
2 each 74-UV
4 each 74-UV
2 each 74-UV
Fiber termination:
SMA 905
SMA 905
SMA 905
SMA 905
www.oceanoptics.com Tel: +1 727-733-2447
Temperature-Regulated Cuvette Holders
CUV-QPOD-2e
Temperature-Controlled Cuvette Holder
The qpod™ is a temperature-regulated sample compartment for fiber optic spectroscopy that controls the temperature of standard 1-cm square
cuvettes to +/-0.05 ºC. The unit includes a Peltier controller, magnetic stirrer and fused silica focusing lenses, and has SMA 905 connectors for easy
coupling to Ocean Optics spectrometers and accessories. The qpod is built for cuvettes with a Z-dimension of 8.5 mm.
Sampling Accessories
Sampling Accessories
When combined with Ocean Optics spectrometers and accessories, the qpod is
especially useful for absorbance and fluorescence measurements that require
stringent control of the sample’s temperature. Each unit is calibrated against a
NIST-traceable thermometer; performance data is provided.
Here are some typical applications:
- DNA melting and annealing
- Protein thermodynamics
- Fluorophore characterization
- Enzyme kinetics
- On-line thermocycling of biological
particles
Features
- R
apid and precise temperature control
over a wide range of temperatures
from -30 ºC to +105 ºC +/- 0.05 ºC
(controllable to even lower temperatures under special conditions)
- Calibrated against a NIST-traceable
thermometer with performance data
provided
- D
esigned for standard 1 x 1 cm square
cuvettes or standard microcuvettes
- All optical components have focusing
and position adjustments to maximize
light throughput
- Light-tight cover with access cap
providing a means of holding a
thermistor probe in the cuvette
Available Items
Item
The qpod was conceived by our channel partner
Quantum Northwest as an improvement on its
CUV-TLC-Series Temperature-regulated Cuvette Holders.
Description
CUV-QPOD-ABSKIT
Temperature Controlled Sample Compartment - Absorbance
CUV-QPOD-2E-ABSKIT
Temperature Controlled Sample Compartment -- Absorbance;
USB/Bluetooth interface
CUV-QPOD-FLKIT
Temperature Controlled Sample Compartment - Fluorescence
CUV-QPOD-2E-FLKIT
Temperature Controlled Sample Compartment -- Fluorescence;
USB/Bluetooth interface
CUV-QPOD-MPKIT
Temperature Controlled Compartment - Absorbance and Fluorescence
CUV-QPOD-2E-MPKIT
Temperature Controlled Sample Compartment -- Absorbance and
Fluorescence; USB/Bluetooth interface
CUV-QPOD
Temperature Controlled Sample Compartment - No Optics
CUV-QPOD-CL-UV
Collimating Lens for qpod Absorbance
CUV-QPOD-IL-UV
Imaging Lens for qpod Fluorescence
CUV-QPOD-MP
Mirror Plug for qpod Sample Compartment
CUV-QPOD-POL
Polarizer for qpod Sample Compartment
CUV-QPOD-FH
Holder for 12.5 mm Diameter Optical Filter for qpod
CUV-QPOD-SER
Serial Interface for qpod External Computer Control
CUV-QPOD-16.10-Q
Micro-volume (10 ul) transmission cuvette
CUV-QPOD-16.100F-Q
Micro-volume (100 ul) fluorescence cuvette
CUV-QPOD-16.10F-Q
Micro-volume (10 ul) fluorescence cuvette
CUV-QPOD-Q-02SH
Oxygen holder for the qpod sample compartment
CUV-QPOD-THERM
Thermistor for the qpod sample compartment
Technical Tip
Proper use of cuvettes
can help avoid measurement errors. For example,
cuvettes always should
be used in the same
orientation. Most cuvettes have index marks
as a guide. Also, it’s important not to touch the optical
surfaces of the cuvette. Oils from your skin, particles from
wiping tissues and other contaminants can affect the
readings.
For open-top square cuvettes, perhaps the most effective
approach is to use a slender transfer or Pasteur pipette to
add and remove fluids. The tiny tip allows for suction of
fluid from the corners, minimizing the carry-over volume.
The typical procedure is to rinse the cuvette with the next
sample to be analyzed at least three times. If the residual
fluid is less than 10% of the wash fluid (it’s more likely
to be 1% or less), the carry-over is reduced to 1/1000. It
is important that the pipettes also be washed with the
sample and not be allowed to touch or scratch the inside
optical surfaces.
www.oceanoptics.com Tel: +1 727-733-2447
111
Sampling Accessories
Sampling Accessories
Integrated Sampling Systems
These Integrated Sampling Systems are direct-attach cuvette holder and light source combinations created specifically for our USB2000+ and
USB4000 Spectrometers (page 14). Both systems receive power and control signals though a connector on the spectrometer.
USB-ISS-UV-VIS
Integrated Sampling System
The USB-ISS-UV-VIS Integrated Sampling System is a direct-attach sample holder and
deuterium tungsten halogen light source (200-1100 nm) combination for 1-cm square
cuvettes. The USB-ISS-UV-VIS allows you to adjust the intensity of the bulb via software. The
sampling system has an electronic shutter for taking dark measurements and comes with a
5-volt power supply.
USB-ISS-VIS
Integrated Sampling System
The USB-ISS-VIS Integrated Sampling System has a violet LED-boosted tungsten source
(390-900 nm) and a sample holder that bolts to the front of a USB4000 or USB2000+
Spectrometer. The spectrometer provides the power and control signals for the light sources.
The USB-ISS-VIS holds 1-cm cuvettes.
ISS-UV-VIS
Integrated Sampling System
The ISS-UV-VIS Integrated Sampling System is a combination RF deuterium source with a
tungsten bulb in a housing connected to a holder for 1-cm cuvettes. This sampling system
couples to an Ocean Optics spectrometer with optical fiber to create a small-footprint system
for relative absorbance. This sampling system is best used with Ocean Optics' 300 µm
solarization-resistant optical fiber.
ISS-2
Integrated Sampling System
The ISS-2 Integrated Sampling System is a fully integrated 1-cm cuvette holder and tungsten halogen light source for relative absorbance measurements. It couples to Ocean Optics
spectrometers with optical fiber to create a small-footprint system for VIS-NIR (~360-1100 nm)
applications.
Specifications
112
USB-ISS-UV-VIS
USB-ISS-VIS
ISS-UV-VIS
ISS-2
Dimensions (mm):
198 x 105.1 x 40.6
40.7 x 88.8 x 34.1
198 x 104.9 x 40.9
155 x 50 x 53.3
Weight:
200 g
130 g
400 g
240 g
Power consumption:
1.8 A @ 5 VDC
160 mA @ 5 VDC
420 mA @ 12 VDC
600 mA @ 12 VDC
Wavelength range (source):
~200-1100 nm (Typical)
390-900 nm (Typical)
~200-1100 nm (Typical)
~360-1100 nm (Typical)
Pathlength:
1 cm
1 cm
1 cm
1 cm
Cuvette shape:
Square
Square
Square
Square
Light source:
Deuterium tungsten
Tungsten and violet LED
Deuterium tungsten
Tungsten
Bulb life (hours):
800 (deut.); 2,000 (tung.)
2000 (tungsten); 45,000 (LED)
800 (deut.); 2,000 (tung.)
900
Time to stabilized output:
~30 minutes
~5 minutes
~30 minutes
~30 minutes
Filter slot:
None
None
None
6.35 mm
Recommended optical fibers:
None
None
QP400-025-SR
QP400-2-UV-VIS
Spectrometers:
USB2000+ and USB4000
USB2000+ and USB4000
All
All
“Z” dimension:
15 mm
15 mm
15 mm
15 mm
www.oceanoptics.com Tel: +1 727-733-2447
Plastic and Quartz Cuvettes and More
Disposable Cuvettes
CVD-VIS1M
CVD-UV1U
CVD-UV1S
Our CVD-Series Disposable Cuvettes are a low-cost, zero-maintenance
alternative to quartz cuvettes. They feature a 1-cm pathlength, 220-900 nm
or 350-900 nm wavelength range coverage and a variety of fill volumes.
Sampling Accessories
Sampling Accessories
Quick tip: You can use a single cuvette for your measurements, but take care
to rinse the cuvette thoroughly with the next sample being analyzed in
order to eliminate cross-over contamination. Oils from your skin, particles
from wiping tissues and other contaminants also can affect the readings.
In addition, proper positioning of cuvettes is important. The cuvettes are
marked with an arrow indicating the transmission path; the user consistently
must position the cuvette with the arrow facing the light source.
Item Code
Range
Material
Volume
Window (in mm)
Clear Sides*
Cover Needed
Quantity
CVD-UV1S
220-900 nm
Plastic
1.5 - 3.0 mL
4.5 x 23
4
Square
100 Pack
CVD-UV1S-SAM
220-900 nm
Plastic
1.5 - 3.0 mL
4.5 x 23
4
Square
8 Pack
CVD-UV1U
220-900 nm
Plastic
70 µL - 1.8 mL
2 x 3.5
2
Round
100 Pack
CVD-UV1U-SAM
220-900 nm
Plastic
70 µL - 1.8 mL
2 x 3.5
2
Round
8 Pack
CVD-VIS1S
350-900 nm
Polystyrene
1.5 - 3.0 mL
5 x 23
4
Square
100 Pack
CVD-VIS1M
350-900 nm
Polystyrene
2.5 – 4.0 mL
10 x 35
2
Square
100 Pack
*Cuvettes with 4 clear sides are suitable for fluorescence measurements.
Quartz Cuvette Cells
We offer several popular high-purity quartz cuvettes including macro, semi-micro, flow and
cylindrical cells. These Quartz Cuvette Cells are suitable for use from 200-2700 nm.
Item
Description
CV-Q-10
Standard
CVFL-Q-10
CVS-Q-10
Windows
Path
Lid
Exterior (mm)
Volume
2 clear
10 mm
Teflon cover
12.5 x 12.5 x 45
3.5 mL
Fluorescence
4 clear
10 mm
Teflon stopper
12.5 x 12.5 x 45
3.5 mL
Self-masking
2 clear
10 mm
Teflon stopper
12.5 x 12.5 x 48
1.4 mL
CVF-Q-10
Flow cell
2 clear
10 mm
M6 screws
12.5 x 12.5 x 35
0.42 mL
CV-Q-100
Cylindrical
2 clear
100 mm
Teflon stopper
22 OD x 102.5
28.2 mL
SpecVette Cuvettes
SpecVette™ Cuvettes are UV-transparent, short pathlength, disposable cuvettes especially designed for use with the
HR4000 and HR2000+ High-resolution Spectrometers. Economical and disposable, SpecVettes eliminate crosscontamination and reduce analyte consumption. SpecVettes are sold in packs of 25, with or without the universal
adapter. Sample packs are also available. SpecVettes are a product of ALine, Inc.
CSV-6-SP
6-piece sample pack
with adapter, 2 each:
250 µm pathlength,
500 µm pathlength and
1000 µm pathlength
CSV-250-25
25-pack, 250 µm
pathlength
CSV-250-25-A
25-pack with adapter,
250 µm pathlength
CSV-500-25
25-pack, 500 µm
pathlength
CSV-500-25-A
25-pack with
adapter, 500 µm
pathlength
CSV-1000-25
25-pack, 1000 µm
pathlength
CSV-1000-25-A
25-pack with
adapter, 1000 µm
pathlength
www.oceanoptics.com Tel: +1 727-733-2447
113
Sampling Accessories
Sampling Accessories
Flow Cells for Flow Injection Analysis
We offer a variety of optical flow cells with a Z configuration. For measuring the optical absorbance of fluids, couple Z cells directly to our
spectrometers to monitor chemical or biological processes and immunoassays.
In our FIA-Z-SMA Flow Cells, standard optical fibers (available separately) connect to the SMA 905 fittings to transmit and receive light through the
central axis of the Z. The FIA-Z-SMA cells use silica windows as wetting surfaces at each fiber optic junction and are available in PEEK polymer,
Plexiglas, Stainless Steel, Teflon and Ultem.
The FIA-Z-CELL Flow Cells are different from the FIA-Z-SMAs; instead of windows they use optical fibers in 1.58-mm ferrules, a design that allows you
to slide the ferrules in and out of the cell to adjust the optical pathlength from 0-10 mm.
Item Code: FIA-Z-CELL-####
Specifications
Cell materials:
PEEK, Plexiglas, Teflon, Stainless Steel, Ultem
Inner diameter:
1.5 mm
Window material:
UV-grade fused silica
Window thickness:
1 mm
Wavelength range:
200-2000 nm
FIA connectors:
1/4-28 fittings (included)
Fiber connectors:
SMA 905 for FIA-Z-SMA cells; 1.58 mm stainless
steel ferrules for FIA-Z-CELL cells
In this setup, a FIA-Z-SMA-PEEK
10-mm pathlength flow cell
is between a light source and
spectrometer. Tubing and connectors
are included.
Available Items
Note: Item Code Extensions indicate:
PEEK – PEEK
PLEX – Plexiglas
SS – Stainless Steel
10 mm Pathlength
FIA-Z-SMA-###
TEF – Teflon
ULT - Ultem
20 mm Pathlength
FIA-Z-SMA-20-###
Our standard fibers are designed for
the FIA-Z-SMA cells.
50 mm Pathlength
100 mm Pathlength
FIA-Z-SMA-50-###
FIA-Z-SMA-100-###
The FIA-P400-SR and FIA-P200-SR fiber
assemblies have ferrules for use with
the FIA-Z-CELL cells.
This FIA-Z-SMA-100-ULT is a
100 mm pathlength cell
made out of Ultem.
Fibers for Use with FIA Cells
A FIA-Z-SMA requires two 200 µm or 400 µm diameter fiber assemblies. Your application may require optical fibers that are optimized for a specific
wavelength range. The FIA-Z-CELL requires two fiber assemblies with ferrule terminations.
Item
114
Description
Use With
P400-2-UV-VIS
400 µm fiber assembly with SMA 905 connectors
FIA-Z-SMA
P200-2-UV-VIS
200 µm fiber assembly with SMA 905 connectors
FIA-Z-SMA
FIA-P400-SR
400 µm fiber assembly with ferrule terminations
FIA-Z-CELL
FIA-P200-SR
200 µm fiber assembly with ferrule terminations
FIA-Z-CELL
www.oceanoptics.com Tel: +1 727-733-2447
Fluid Analysis Systems
Our FIA-1000-Z Flow Cell Kit is a convenient, low-cost fluid
handling system that couples to Ocean Optics highsensitivity miniature spectrometers and light sources for
fast, quantitative analysis of solutions.
Sampling Accessories
Sampling Accessories
The FIA-1000-Z system consists of a two-channel
peristaltic pump, a 1-cm pathlength fiber optic flow cell,
a fittings kit and Windows-based operating software. A
pair of optical fibers or special bifurcated assembly (priced
separately) completes the package.
High-throughput Z-cell Option
The FIA-Z-SMA-PEEK-LENSED is a 10 mm Z-type flow cell
that includes two collimating lenses and special SMA
905 adapters that increase the UV-VIS throughput of the
flow cell by up to 10x compared with the standard Z-cell.
The high-throughput cell can be used as part of any
system that has an Ocean Optics spectrometer configured for optical absorbance measurements, but is
especially attractive as part of a multichannel Jaz system
configured for chromatographic applications.
The FIA-Z-SMA-PEEK-LENSED cell has good resistance to
a wide range of organic and inorganic materials. Its lens
material is UV-grade fused silica with response from
200-900 nm. By adding accessories such as the
FIA-PUMP-C and a pair of QP450-1-XSR 450 µm extreme
solarization resistant optical fibers, you can create a
system comparable to the FIA-1000-Z Flow Cell Kit.
Specifications
Criteria
Spectral range:
260-2000 nm
Cell materials:
PEEK, Plexiglas, Teflon, Stainless Steel or Ultem
Inner diameter:
1.5 mm
Window material:
UV-grade fused silica
Pathlength:
10 mm
Fiber connectors:
SMA 905
Tubing:
1/16" Teflon, ~3 m
Tubing connectors:
1/4-28
Pump dimensions:
105 mm x 105 mm x 185 mm
Pump weight:
2100 g
Flow rate:
0.5-2.0 mL/minute/channel (depending on RPM
and tubing diameter)
Pressure rating:
25 psi
Available Items
Item
Description
FIA-1000-Z
Z-SMA flow cell with peristaltic pump. Includes operating
software with code for computer control of pump. User
specifies cell material: Teflon, Plexiglas, Stainless Steel,
PEEK polymer or Ultem.
FIA-ZSMA
Replacement Z-SMA flow cell for FIA-1000-Z. User specifies cell material: Teflon, Plexiglas, Stainless Steel, PEEK
polymer or Ultem.
FIA-ZSMA-100
FIA-ZSMA-50
FIA-ZSMA-20
Z flow cell w/SMA 905, 100 mm pathlength
Z flow cell w/SMA 905, 50 mm pathlength
Z flow cell w/SMA 905, 20 mm pathlength
FIA-ZSMA-ML
Z flow cell with SMA 905 connector, microliter volume. User
specifies cell material: Teflon, Plexiglas, Stainless Steel,
PEEK polymer or Ultem.
FIA-ZSMA-ML-100
FIA-ZSMA-ML-50
FIA-ZSMA-ML-20
Z flow cell w/SMA 905, microliter volume 100 mm path
Z flow cell w/SMA 905, microliter volume 50 mm path
Z flow cell w/SMA 905, microliter volume 20 mm path
FIA-Z-CELL
Adjustable replacement cell for use with FIA-1000-Z. Connects to fiber optics w/ferrule termination. Standard cell
materials are Plexiglas, Stainless Steel, Teflon, Ultem or
PEEK.
FIA-P200-SR
Custom 200 µm ferrule-termination fiber optic for use with
FIA-Z-CELL; requires 2 fibers.
FIA-P400-SR
Custom 400 µm ferrule-termination fiber optic for use with
FIA-Z-CELL; requires 2 fibers.
FIA-PUMP-C
Replacement peristaltic pump (computer-controlled) for FIA1000 flow cells
www.oceanoptics.com Tel: +1 727-733-2447
115
Sampling Accessories
Sampling Accessories
Fluid Analysis
SpectroPipetter Microcell
The PIP-10-2 SpectroPipetter Microcell is a combination micropipetter and 10-mm pathlength microcell for
low-volume sampling. The SpectroPipetter conveniently couples to Ocean Optics high-sensitivity fiber optic
spectrometers and compact light sources for rapid measurement of microliter-level solutions. It requires only
2 µL of sample for a spectral measurement and is especially handy for applications such as protein analysis, DNA
experimentation and forensic toxicology.
Specifications
Sample cell pathlength:
Sample cell volume:
Wavelength range:
10 mm
2.0 µL
UV-VIS through Shortwave NIR (~240-1100 nm)
PIP-10-2
PIP-UCK
Micropipetter and 1-cm pathlength microcell sampling device
PIP-UCK-CS
Micropipetter cleaning kit
Cleaning solution
Longpass Flow Cells
LPC Longpass Flow Cells couple to Ocean Optics spectrometers and light sources for simple,
efficient measurement of low-volume or low-concentration aqueous samples. These liquid
waveguide capillary cells are available in pathlengths ranging from 50-500 cm, with internal
sampling volumes ranging from 125 µL for a 50 cm pathlength cell to 1,250 µL for a 500 cm
pathlength cell. Cells are responsive from ~230-800 nm, with higher starting wavelengths for
the longer pathlength versions.
The LPC Flow Cells were developed by our partner World Precision Instruments as fibercoupled alternatives to standard cuvettes. Units can be directly connected to a pump or fluid
injection analysis system or filled with a syringe. The LPCs are ideal for various kinds of flow
analysis and water monitoring.
Also available are very low-volume LPC flow cells ideal for micro-volume solution
measurements such as DNA and RNA quantification, protein determination and nutrient
analysis. These cells are available in 10 mm, 50 mm and 100 mm pathlengths, with sample
volumes of 2.4 µL, 12 µL and 24 µL, respectively.
LPC Specifications
LPC-050CM
LPC-100CM
LPC-250CM
LPC-500CM
Optical pathlength:
50 cm
100 cm
250 cm
500 cm
Internal volume:
125 μL
250 μL
625 μL
1250 μL
Fiber connection (SMA 905):
500 μm
500 μm
500 μm
500 μm
Noise [mAU]:
<0.1
<0.2
<0.5
<1.0
Maximum pressure:
100 PSI
100 PSI
100 PSI
100 PSI
Wetted material:
PEEK, Fused Silica, PTFE
Liquid input standard:
Micro Cell Specifications
LPC-10MM
LPC-50MM
LPC-100MM
Optical pathlength:
10 mm
50 mm
100 mm
Internal volume:
2.4 μL
12 μL
24 μL
Wavelength range (most efficient):
~230-800 nm
~230-800 nm
~230-800 nm
Fiber connection (SMA 905):
500 μm
500 μm
500 μm
Maximum pressure:
1000 PSI
1000 PSI
1000 PSI
Refractive index @ 280 nm:
<7 mAU
<15 mAU
Not reported
Wetted material:
116
10-32 Coned Port Fitting
PEEK, Fused Silica, PTFE
www.oceanoptics.com Tel: +1 727-733-2447
Longpass Flow Cell Options
Item Code
Description
LPC-050CM
Liquid Waveguide Capillary Cell,
50 cm pathlength
LPC-100CM
Liquid Waveguide Capillary Cell,
100 cm pathlength
LPC-250CM
Liquid Waveguide Capillary Cell,
250 cm pathlength
LPC-500CM
Liquid Waveguide Capillary Cell,
500 cm pathlength
LPC-10MM
Liquid Waveguide Capillary Cell,
10 mm pathlength
LPC-50MM
Liquid Waveguide Capillary Cell,
50 mm pathlength
LPC-100MM
Liquid Waveguide Capillary Cell,
100 mm pathlength
Fluid Analysis
CFV-Series FluoroVettes
Our FluoroVettes are ultra low-volume,
disposable cells for nano-molar range
fluorescence detection. These UVtransparent cuvettes require only 50 µl of
fluid and slide neatly into a 1-cm cuvette
adapter for use in setups with a
spectrometer and cuvette holder.
Sampling Accessories
Sampling Accessories
The design of our FluoroVettes makes them
quite suitable for a variety of applications
including real-time monitoring of chemical
or biological agents for competitive
displacement assays, high-sensitivity
quantification of double-stranded DNA and
ELISA assay development.
Here are some other possible applications:
-
Assay development with quantum dots
Protein conformation analysis
Cell marker identification
Enzyme inhibitors using fret assays
Specifications
Available Items
Volume
Item Code
Dead volume:
2 µL
Dead volume (tubing interface):
Depends on length of tubing at inlet side
of FluoroVette; typical ID is 0.030”, OD is
0.063” (1/16”); length of tubing is 1.5” on
each side
Contained volume:
50 µL (+/- 1 µL)
Dimensions
Description
CFV-PIP-SP
10 pack Pipettor UV-transparent
FluoroVettes with adapter
CFV-PUMP-SP
5 pack Pump UV-transparent
FluoroVettes with adapter
CFV-PIP
10 pack Pipettor UV-transparent
FluoroVettes
CFV-PUMP
5 pack Pump UV-transparent
FluoroVettes
Height x width:
50 mm x 9 mm
Thickness:
1 mm
CFV-ADP-2
Universal adapter for 1 cm x
1 cm fluorescence sample holder
with 2-way 90 degree viewing
Zeonor Film:
50 mm (0.002”) (UV-transparent to 220 nm)
CFV-ADP-4
Black Delrin:
0.75 mm (0.030”)
Bonding:
Medical-grade acrylic solvent for bonding
Delrin to Zeonor
Universal adapter for 1 cm x
1 cm sample holder with 4-way
viewing
CSV-6-SP
Compatibility:
Ethanol (up to 99.5%), Isopropyl alcohol
(up to 91% water), aqueous buffers with or
without detergents
Sample pack of 6 SpecVettes - 2
x 1 mm, 2 x 0.5 mm, and
2 x 0.25 mm + Adapter
Incompatibility:
Acetone or aliphatic solvents such as
hexane
Materials
Connectors
For CFV-PIP-SP:
Acrylic
For CFV-PUMP-SP:
Polysulfone
For tube connection:
Translucent FEP for tubing; glue is a
solvent-less UV-curing medical-grade
adhesive
Note: There are two types of FluoroVettes. The CFV-PIP-SP has
an inlet port for loading the sample into the FluoroVette with
a standard 20-200 µL pipetter and ordinary tips, making it a
snap to fill and perform measurements. The CFV-PUMP-SP has
tubing barbs at the inlet and outlet ports so the FluoroVette
can be used in continuous or flow injected measurements
using a syringe or peristaltic pump.
www.oceanoptics.com Tel: +1 727-733-2447
117
Sampling Accessories
Sampling Accessories
Cosine Correctors
CC-Series Cosine Correctors
Our Cosine Correctors couple to optical fibers and spectrometers
for relative and absolute spectral intensity measurements,
emission applications and evaluation of light sources.
When the CC-3 and CC-3-UV-S are screwed onto the end of an
optical fiber, the cosine corrector and optical fiber become an
irradiance probe. The probe couples to one of our spectrometers
to measure the intensity of light normal to the probe surface.
The CC-3-DA screws directly onto the SMA 905 Connector of our
spectrometers, creating a complete spectroradiometric system
and eliminating the need for an optical fiber.
Please note that the native response of Spectralon covers a wider
range than what we report here. Our specifications reflect its use
with our spectrometers as a light collector.
Available Items
CC-3
CC-3-UV-S
CC-3-DA
Diffusing material:
Opaline glass
Spectralon
PTFE
Spectralon
Typical range:
350-1000 nm
200-1700 nm
200-1100 nm
200-1100 nm
Dimensions:
6.35 mm OD
6.35 mm OD
6.35 mm OD
12.7 mm OD
Field of View:
180°
180°
180°
180°
Power Supply and Controller for LEDs
The LED-PS Power Supply works with our spectrometers and the FOIS-1
Fiber Optic Integrating Sphere for spectroradiometric and color measurements of LEDs. The LED-PS unit has easy-to-reach electrical connectors for
mounting LEDs that are 9.52-mm diameter or smaller with 2.77-mm lead
spacing.
The LED-PS Power Supply provides three useful functions. The LED-PS
holds the LED in place, powers the LED, and displays the LED’s drive
current. The drive current is adjustable, with a digital display to indicate
the current level. Use the adjustable drive current feature to increase or
decrease an LED’s current up to 50 mA. With the LED-PS-NIST, the current
meter is calibrated against a NIST-traceable standard.
Specifications
Dimensions:
56.8 mm x 56.8 mm x 56 mm
Weight:
170 g
Power consumption:
Up to 100 mA @ 12 VDC; depends on setting
LED drive current:
12-50 mA with 0.1 mA resolution
Drive current accuracy:
± 1.0%
LED mount:
2.77 mm lead spacing, PTFE base
Available Items
118
CC-3-UV-T
Item
Description
LED-PS
Power Supply and Controller for LEDs
LED-PS-NIST
NIST-traceable Power Supply and Controller for LEDs
LED-PS-RECAL
Recalibration services for your LED-PS-NIST
FOIS-1
Fiber Optic Integrating Sphere
www.oceanoptics.com Tel: +1 727-733-2447
Integrating Spheres
ISP-I Integrating Spheres
Our ISP-I Integrating Spheres are convenient sampling optics that couple to our spectrometers and optical fibers to
measure spectral output of LEDs, lasers and other light sources from 200-2500 nm.
Sampling Accessories
Sampling Accessories
Each fiber optic integrating sphere consists of a proprietary PTFE-based, sintered diffusing material in diameters of
30, 50 or 80 mm. This provides a Lambertian surface with >98% reflectivity in the visible range for irradiance measurements. Sample port sizes of 6 mm for the 30 mm sphere and 8 mm for the 50 mm and 80 mm spheres are available.
Available Items
Item Code
Description
Sample Port
ISP-30-6-I
Integrating sphere, 59 mm diameter, 58 mm high
6 mm
ISP-50-8-I
Integrating sphere, 80 mm diameter, 78 mm high
8 mm
ISP-80-8-I
Integrating sphere, 107 mm diameter, 117 mm high
8 mm
Specifications
ISP-LED-ADP Holds in place 3, 5 or 8 mm LED
for reproducibility; for use with
ISP-50-8-1
N/A
Dimensions:
ISP-50-I-USB
ISP-50-8-I with connector for directly attaching to a spectrometer
8 mm
Weight:
330 g (ISP-30); 730 g (ISP-50); 1,650 g (ISP-80)
Spectral range:
200-2500 nm
ISP-80-8-I with connector for directly attaching to a spectrometer
8 mm
Custom sample port machining of
8, 10, or 12 mm diameter
8, 10 or 12 mm
Custom sample port machining of
14,16 or 20 mm diameter
14, 16 or 20 mm
ISP-80-I-USB
ISP-PORT-1
ISP-PORT-2
ISP-30-6-I: 59 mm diameter, 58 mm height
ISP-50-8-I: 80 mm diameter, 78 mm height
ISP-80-8-I: 107 mm diameter, 117 mm height
Sphere diameter:
30 mm, 50 mm or 80 mm
Sample port diameter:
6 mm (ISP-30); 8 mm (ISP-50 and ISP-80)
Sphere coating:
Proprietary PTFE-based diffusing material
Reflectivity:
>98% (400-1500 nm); >95% (250-2500 nm)
LED adapter:
For 3-mm, 5-mm or 8-mm LEDs
FOIS-1 Fiber Optic Integrating Sphere
The FOIS-1 is a compact sampling optic that collects light from emission sources such as LEDs and lasers and can be used to measure light fields
with a 360° field of view.
The compact FOIS-1 measures just 56.8 mm x 62.4 mm x 38.1 mm and weighs only 240 g -- yet it is durable enough for use for many types of
applications. The interior of the FOIS-1 is made from Spectralon, a white diffusing material that provides a highly Lambertian reflecting surface.
The FOIS-1 is easy to operate. Simply connect an optical fiber (the read fiber) from the FOIS-1’s SMA 905-terminated output port to the SMA
termination of the spectrometer. The emission source is then inserted into the 0.375” input port of the FOIS-1. Or the setup can be configured so that
the light energy from the emission source can enter the input port.
Specifications
Effective spectral range:
250-2500 nm
Dimensions:
56.8 mm x 62.4 mm x 38.1 mm
Weight:
240 g
Sample port aperture:
9.5 mm
Sphere coating:
Spectralon
Top cap mounts:
(2) 8-32 threaded holes (hardware not included)
(1) 1/4”-20 threaded hole in center (screw/adapter included)
Side mounts:
SMA 905 connector for coupling optical fiber to
the spectrometer
8-32 threaded hole for post mounts
Connector:
SMA 905
www.oceanoptics.com Tel: +1 727-733-2447
119
Sampling Accessories
Sampling Accessories
Spheres for Reflectance
ISP-REF Integrating Sphere for Reflectance
The ISP-REF Integrating Sphere is designed for applications requiring even surface illumination for
reflectance measurements -- such as in determining the color of flat surfaces.
The ISP-REF is 1.5” in diameter and features a transfer optic assembly for restricting the fiber viewing
angle, a 0.4” aperture sample port and a built-in tungsten-halogen light source with 12 VDC adapter. The
sphere is coated with Spectralon®, a white diffusing material that provides a highly Lambertian reflecting
surface. A simple switch allows users to manipulate the sampling optic for the inclusion (I) or exclusion (E)
of specular reflectance.
The ISP-REF Integrating Sphere is small and compact -- measuring just 54 mm x 57 mm x 83 mm (LWH)
and weighing 865 g – yet, it is durable enough for use outside the laboratory.
Item Codes: ISP-REF Illuminated Integrating Sphere, ISP-REF-B Replacement Bulb for ISP-REF
Specifications
Spectral range (of illumination source):
360-1000 nm
Dimensions:
54 mm x 57 mm x 83 mm (LWH)
Sphere diameter:
38.1 mm
Weight:
865 g
Sample port aperture:
10.32 mm
Sphere coating:
Spectralon (doped with Barium Sulfate)
Reflectivity:
>98% (400-1500 nm)
>95% (250-2000 nm)
Reflectance measurements:
Specular included or excluded
Bulb life:
900 hours
Bulb color temperature:
3100 K
Connector:
SMA 905
ISP-R Integrating Spheres for Reflectance
The ISP-R Integrating Spheres are distinguished by their compact size and sturdy design. All ISP-R spheres
have two SMA 905 ports. The excitation input is angled at 8° and collimates the fiber input before
introduction into the sphere. The output port is angled at 90° (to connect to a spectrometer). A gloss-trap
version comes with two cylindrical inserts coated with either a black absorbing material (for excluding
the specular component of the reflection) or with the same material as the sphere (to include the
specular component). This insert fits into a hole angled at 8° at the top of the sphere.
Specifications
Dimensions:
120
ISP-REF
ISP-30-6-R
54 mm x 57 mm x 83 mm
59 mm dia., 58 mm high
ISP-50-8-R
80 mm dia., 78 mm high
ISP-80-8-R
ISP-50-8-R-GT
107 mm dia., 117 mm high
80 mm dia., 78 mm high
Weight:
864.7 g
330 g
730 g
1,650 g
743.3 g
Power consumption:
600 mA @ 12 VDC (lamp)
None
None
None
None
Spectral range:
360-2000 nm
200-2500 nm
200-2500 nm
200-2500 nm
200-2500 nm
Sphere diameter:
38.1 mm
30 mm
50 mm
80 mm
50 mm
Sample port diameter:
10.32 mm
6 mm
8 mm
8 mm
8 mm
Sphere coating:
Spectralon
PTFE material
PTFE material
PTFE material
PTFE material
Reflectance:
Diffuse or specular and
diffuse
Specular and diffuse
Specular and diffuse
Specular and diffuse
Diffuse or specular and
diffuse
Reflectivity:
>98% (400-1500 nm)
>95% (250-2000 nm)
>98% (400-1500 nm)
>95% (250-2000 nm)
>98% (400-1500 nm)
>95% (250-2000 nm)
>98% (400-1500 nm)
>95% (250-2000 nm)
>98% (400-1500 nm)
>95% (250-2000 nm)
Bulb:
900-hour bulb; 3100 K
color temp.
None
None
None
None
www.oceanoptics.com Tel: +1 727-733-2447
Spheres for Reflectance
Reflectance/Transmittance Spheres
The RT and RTC Spheres from Labsphere are basic measurement tools that can be used to measure the reflectance
or transmittance of light through a wide variety of sample media. These spheres feature 25.4 mm diameter ports to
accommodate sample and reference beams as well as port plugs required for a 9° single beam geometry.
Sampling Accessories
Sampling Accessories
RTC Spheres add further versatility with a center-mounted sample holder so users are able to measure reflectance
and transmittance versus incident angle of radiation. Five ports accommodate a sample and reference beam with
a center mount stage located at the top of the sphere, and a 12.7 mm detector port located at the bottom of the
sphere assembly.
Item Codes: LAB-RT-060SF, Reflectance/Transmittance Integrating Sphere; 6”; includes 0° sample holder and accessories
LAB-RTC-060SF, Reflectance/Transmittance Integrating Sphere; 6”; includes 0° and 8° sample holders, center mount and accessories
Specifications
RT-060-SF
RTC-060-SF
Sphere diameter:
152. 4 mm (6”)
152. 4 mm (6”)
Sphere coating:
Spectraflect®
Spectraflect®
Optimum spectral range:
250-2500 nm
250-2500 nm
Detector port diameter:
12.7 mm (0.5”)
12.7 mm (0.5”)
Sample and reference ports:
5
5
Sample and reference ports (diameters):
25.4 mm (1”) for all
25.4 mm (1”) for three and 31.75 mm (1.25”) for two
Center-mount sample holder:
NA
Jaw and clip styles
Sphere mount:
1/4-20 boss mounting post and base assembly
Adjustable H-frame assembly
www.oceanoptics.com Tel: +1 727-733-2447
121
Sampling Accessories
Sampling Accessories
Light Modification
MonoScan2000
The MonoScan2000 is a computer-controlled scanning monochromator with a 300-700 nm wavelength
range. The rapid system scans one nanometer in 15-20 milliseconds and covers its entire range within three
seconds. The MonoScan2000 is compatible with all Ocean Optics spectrometers, light sources,
accessories and optical fibers.
A light source directs light via optical fiber to a sample, interacts with the sample and sends light to the MonoScan2000. The monochromator captures the incoming light and transmits it via fiber to a singleelement detector (such as a photodiode) one wavelength at a time. The MonoScan2000 allows a high optical
throughput and provides an intense spectral signal. The MonoScan2000 has the ability to scan through a
wavelength range you select via software. Because the MonoScan2000 has no slit, the diameter size of the
optical fiber determines the optical resolution of the system. When using 200 µm fibers, for example, optical
resolution is <3 nm (FWHM).
Item Code: MONOSCAN2000
Specifications
Dimensions:
112 mm x 132 mm x 145 mm
Weight:
1.1 kg
Wavelength range:
300-700 nm
Optical resolution:
Depends on the diameter of the optical fiber; ~4 nm (FWHM) using a 400 µm diameter optical fiber
Holographic grating:
1250 lines/mm, blazed at 350 nm
Accuracy:
<0.5 nm
Repeatability:
0.2 nm
Transition speed:
A scan from 300 nm to 700 nm takes ~3 seconds
A one nanometer step takes ~15-20 milliseconds
Dispersion:
~10 nm per mm
Optical throughput:
>50% with a 1000 µm fiber at 350 nm
>30% with a 1000 µm fiber at 500 nm
Grating scan angle:
14.8° (300-700 nm)
Computer interface:
USB and RS-232
Power requirement:
12 VDC max. 1.2A (WT-12V-E)
Gearbox ratio:
1:261
Hydra Fiber-Coupled Light Mixer
Hydra from Ocean Optics is a fiber-coupled light mixer that features nine SMA 905 connectors. Each connector
can be used as input for light or an output.
Light coming into the Hydra’s SMA connectors are mixed homogeneously inside the Hydra’s chamber. The
intensities of the outgoing light are at the same level with a matching of better than 1%.
The physical properties of the Hydra are based on Spectralon® diffuse reflectance material and offer high
reflection efficiency (better than 90%) over a wavelength range from 200-2500 nm.
The Hydra is a compact 60 mm x 45 mm. The
unit’s integrating sphere is 20 mm in diameter.
Use Hydra to harness up to eight light sources
and accurately mix light intensities to fit your
application’s requirements.
Item Code: HYDRA
122
Specifications
Dimensions:
60 mm ht x 45 mm diameter
Integrating sphere:
20 mm diameter
Integrating sphere coating:
Spectralon
Integrating sphere reflectivity:
>90% from 200-2500 nm
Connectors:
(9) SMA 905
www.oceanoptics.com Tel: +1 727-733-2447
Reflection and Transmission Stages
Reflectance Stage
Our Single-Point Reflection stage is a probe holder that is perfect for reflection
measurements of optical layers and other substrates (up to 150 mm in diameter).
The probe holder accommodates fiber optic probes up to 6.35 mm in diameter
and slides up and down a stainless steel post for adjustment to heights as great as
~63.5 mm.
Sampling Accessories
Sampling Accessories
The Stage has an anodized base plate, scored in concentric circles of varying
diameters, that act as a guide when positioning round samples.
Item Code: STAGE
Reflection-Transmission Stage
The Stage-RTL-T is a unique and versatile sampling system for performing
transmission and reflection measurements in numerous configurations. When
combined with Ocean Optics spectrometers and light sources, the Stage-RTL-T
is ideal for applications ranging from materials analysis of coatings, plastics, glass
and semiconductor wafers to characterization of biological samples such as
plants, animal tissue and fruit.
The Stage-RTL-T includes a variable rail to which various additional sampling fixtures can be attached. In addition to the rail, the Stage-RTL-T comes with a sample
plate, adapter, a pair of 74-UV collimating lenses and a light trap.
Here are some of the ways you can utilize the Stage-RTL-T:
- A
ttach any Ocean Optics reflection probe with ¼” (6.35 mm) OD and position
the rail to measure an extensive range of sample shapes and sizes.
- W
ith the generic adapter, position the collimating lenses to measure transmission at a perpendicular angle or reflection at a 45-degree angle. The adapter
also will accommodate Ocean Optics reflection probes with ¼” (6.35 mm) OD.
dd an optional ISP-RTL-ADP adapter to accommodate our 30 mm and 50 mm
- A
ISP-series integrating spheres. This allows the user to take diffuse reflection and
transmission measurements.
The optical components in the Stage-RTL-T are responsive from 200-2500 nm,
making them feasible for the full range of Ocean Optics UV, Visible and NIR
spectrometers and light sources. The system’s anodized aluminum plates and rails
and its stainless steel posts and post holders are sturdy and dependable.
Item Code: STAGE-RTL-T
Specifications
STAGE
STAGE-RTL-T
Dimensions:
152.4 mm diameter (base)
101.6 mm diameter (sample area)
206.3 mm diameter (base)
152.4 mm diameter (sample area)
Weight:
620 g
4.5 kg
Height:
Adjustable 63.5 mm
Adjustable to 400 mm
Materials:
Anodized aluminum plate, stainless steel post and post holder
Anodized aluminum
www.oceanoptics.com Tel: +1 727-733-2447
123
Sampling Accessories
Sampling Accessories
Specular Reflectance Standards
We offer three Specular Reflectance Standards for use in measuring the reflectance of surfaces with high or low specular reflectivity. Each consists of
a 31.7-mm outer diameter optical reflectance material in a protective aluminum receptacle with screw-on top. The superior coatings on the substrates are environmentally stable and can withstand high temperature and mechanical stresses.
STAN-SSH
The STAN-SSH High-reflectivity Specular Reflectance Standard is a mirrored, fused-silica standard that can be used as a
reference when measuring surfaces with high specular reflectance values such as optical substrates, optical coatings,
machined metals and semiconductor materials. The STAN-SSH provides high reflectance across the range from 800-2500 nm.
STAN-SSH-NIST
The NIST-traceable STAN-SSH-NIST is calibrated to a NIST master (NIST part number NIST38060S, s/n 99G16) and is spectrally
flat to <5% from 250-2500 nm. With the shipment of the NIST-traceable version, the customer will receive a certificate of
calibration, a data sheet with the reflectance values as a function of wavelength and a diskette that contains calibration data
that can be transferred to Ocean Optics SpectraSuite Operating Software.
STAN-SSL
The STAN-SSL Low-reflectivity Specular Reflectance Standard is a black glass standard that can be used as a reference when
measuring surfaces with low specular reflectance values such as thin film coatings, anti-reflective coatings, blocking filters and
substrates. The coated surface of the STAN-SSL provides ~4.0% reflectance across the 200-2500 nm wavelength range.
STAN-HOLDER
This convenient accessory option keeps your standard securely in place during measurements and helps preserve its coating.
Specifications
STAN-SSH
STAN-SSH-NIST
STAN-SSL
Substrate dimensions:
31.75 mm outer diameter x 6.35 mm
height
31.75 mm outer diameter x 6.35 mm
height
31.75 mm outer diameter x 6.35 mm
height
Housing dimensions:
38 mm outer diameter x 19 mm height
38 mm outer diameter x 19 mm height
38 mm outer diameter x 19 mm height
Weight:
40 g
40 g
40 g
Reflectance material:
Front-surface protected aluminum mirror on fused silica substrate
Front-surface protected aluminum mirror on fused silica substrate
Schott ND9 glass
Reflectivity:
~87-93% (200-1000 nm)
~93-98% (1000-2050 nm)
~87-93% (200-1000 nm)
~93-98% (1000-2050 nm)
~5% (200-950 nm)
~4% (950-2500 nm)
Reflectance of STAN-SSL at 6°
Reflectance of STAN-SSH at 6°
10%
98%
9%
96%
7%
% REFLECTION
% REFLECTION
8%
6%
5%
4%
3%
2%
500
700
900
1100
1300
1500
WAVELENGTH (nm)
124
92%
90%
88%
1%
0
300
94%
www.oceanoptics.com Tel: +1 727-733-2447
1700
1900
86%
200
400
600
800
1000
1200
1400
WAVELENGTH (nm)
1600
1800
2000
Sampling Accessories
Sampling Accessories
Diffuse Reflectance Standards
WS-1 Diffuse Reflectance Standard
The WS-1 Diffuse Reflectance Standard is made of PTFE, a diffuse white plastic that
provides a Lambertian reference surface for reflectance experiments. The WS-1 comes
in an anodized aluminum housing, and is hydrophobic, chemically inert and very stable,
even in deep-ultraviolet applications. It is >98% reflective from 250-1500 nm and >95%
reflective from 250-2200 nm.
Item Code: WS-1
WS-1-SL White Reflectance Standard with Spectralon
The WS-1-SL is a diffuse reflectance standard from Labsphere and is made from their
patented diffuse reflectance material, Spectralon. Spectralon is hydrophobic and is thermally
stable to 350 °C. The durable material provides highly accurate, reproducible data.
Item Code: WS-1-SL
Specifications
WS-1
WS-1-SL
Dimensions:
38 mm diameter housing
32 mm OD, 10 mm thick tile
Weight:
30 g
30 g
Spectral range:
250-2000 nm
250-2500 nm
Housing:
Aluminum
Delrin holder, protective cover
Reflectivity:
>98% (250-1500 nm)
>95% (250-2200 nm)
99% (400-1500 nm)
>96% (250-2000 nm)
Sampling Accessories
Spectralon Standards from Labsphere
Spectralon® material is distinguished by its high diffuse reflectance values (>99% over a range from 400-1500 nm and >95% from 250-2500 nm) and
is available in components such as reflectance, color and fluorescence standards.
Diffuse reflectance standards are highly Lambertian, with reflectance values ranging from 2%-99%. Color standards come in various color sets and
are an excellent choice for developing consistent reproduction in manufacturing applications. Fluorescence standards use various combinations of
Spectralon and inorganic fluors and are photochemically stable compared with their organic counterparts.
Available Reflectance and Color Standards
Available Fluorescence Standards
LAB-RSS-040-010
Diffuse reflectance standard set, 1", 2%, 50%, 75%, 99%
reflectance values
LAB-USFS-500-010
Fluorescence standards set, includes five fluors,
uncalibrated
LAB-RSS-08-020
Diffuse reflectance standard set, 2", 2%, 5%, 10%, 20%,
40%, 60%, 80%, 99% reflectance values
LAB-USFS-200-010
Fluorescence standards, 1", Blue/White Fluor #200,
uncalibrated
LAB-SRS-02-010
Diffuse reflectance standard, 1", 2% reflectance value
LAB-USFS-205-010
LAB-SRS-10-010
Diffuse reflectance standard, 1", 10% reflectance value
Fluorescence standards, 1", Blue Fluor #205, uncalibrated
LAB-SRS-20-010
Diffuse reflectance standard, 1", 20% reflectance value
LAB-USFS-210-010
LAB-SRT-50-020
Diffuse reflectance target, UV-NIR, 2", 50% reflectance
value, single centerpoint calibration
Fluorescence standards, 1", Green Fluor #210,
uncalibrated
LAB-USFS-336-020
LAB-SRT-99-020
Diffuse reflectance target, UV-NIR, 2", 99% reflectance
value, single centerpoint calibration
Fluorescence standards, 1", Orange Fluor #336,
uncalibrated
LAB-USFS-461-010
LAB-SRT-MS-050
Diffuse reflectance target, multi-step, 250-2500 nm, 5"
x 5", 12%, 25%, 50%, 99% reflectance values, single
centerpoint calibration
Fluorescence standards, 1", Blue/White Fluor #461,
uncalibrated
LAB-CSS-04-010
Diffuse color standards set, 1”, Red-Green-Blue-Yellow
LAB-CSS-04A-010
Diffuse color standards set, 1”, Orange-Purple-VioletCyan
LAB-CSS-12-010
Diffuse color standards set, 1”, RGBY and OPVC, 2%,
20%, 50%, 99% reflectance values
www.oceanoptics.com Tel: +1 727-733-2447
125
Sampling Accessories
Sampling Accessories
Filter Holders
FHS-UV In-Line Filter Holder
The FHS-UV In-Line Filter Holder is a low-cost spectrophotometric accessory for fast,
convenient absorbance/transmission measurements of optical and other filters.
The FHS-UV In-Line Filter Holder features a pair of 74-UV 5-mm diameter f/2
collimating lenses that maximize light throughput, as well as a manual light-block wheel for
dark readings. The Filter Holder acts as a simple yet effective device for measuring filters. It
also provides a convenient place to accommodate filters for optical setups.
FHSA-Series Filter and Cuvette Holders
With our FHSA Filter and Cuvette Holders, you can configure the holders to sample either
cuvettes or filters. Use FHSAs to measure transmission of a 1-cm square cuvette or filter up
to 7 mm thick. Plus, FHSAs interface to your PC, allowing you to control many of their
functions via the included software.
With the FHSA-TTL, you have manual control of attenuation (adjustable 0-100%) and
manual or software control of a shutter.
Specifications
FHSA-TTL
FHS-UV
Dimensions:
150 mm x 50 mm 50 mm
50.6 mm x 140 mm x 43.1 mm
Weight:
490 g
240 g
Power consumption:
100 mA @ 12 VDC
None
Maximum filter size:
7 mm thick
25 mm diameter round; up to 6 mm thick
Wavelength range:
200-2000 nm
200-2000 nm
Cuvette dimensions:
10 mm x 10 mm
NA
Shutter frequency:
5 Hz/60 dB maximum
NA
Shutter response time:
7 µsec
NA
INLINE-FH Filter Holder for Optical Fibers
Our INLINE-FH is an in-line filter holder that accommodates filters up to 8 mm
in diameter and 3 mm in thickness. This filter holder includes two collimating
lenses and connects to two fibers for in-line filtering. Filter options are similar
to the 25.4 mm square filters listed on page 128. Please consult an Applications
Scientist for assistance.
The FH-SMA allows you to mount filters or diffusers at the end of the SMA
905-terminated optical fibers. The FH-SMA accepts 8 mm diameter filters in
thicknesses of 1-7 mm.
Specifications
126
FH-SMA
INLINE-FH
Dimensions:
16 mm diameter
15 mm diameter
Weight:
10 g
20 g
Filter size:
8 mm diameter, 1-7 mm thick
8 mm diameter, 3 mm thick
Material:
Anodized aluminum
Stainless steel (collimating lenses have anodized housings)
www.oceanoptics.com Tel: +1 727-733-2447
Linear Variable Filters
We’ve combined our patented high-pass and low-pass technology to create an exceptional Linear Variable Filter line that features an adjustable
center wavelength and bandpass. Each filter delivers an outstanding transmission band (~90%) and blocking band (99.8%).
Our LVF Filters have interference coatings applied to 57 mm x 10 mm quartz substrates and are particularly useful for spectrally shaping the
excitation energy from broadband sources used for fluorescence.
Sampling Accessories
Sampling Accessories
LVF-H High-pass Filter
Slide Carriers
100
These off-the-shelf filters are epoxied into slide
carriers that allow you to move the transmission or
blocking band throughout each filter’s wavelength
range.
%TRANSMISSION
80
60
40
20
0
250
Single High-pass and Single Low-pass Filter
The LVF-H High-pass Filter is a single filter that blocks light at 98.8% until a transition wavelength
that varies along its length. At that point, the LVF-H passes light better than 90%. The LVF-L
Low-pass Filter is a single filter that passes light at 88% until a transition wavelength that varies
along its length.
350
450
550
WAVELENGTH (nm)
650
750
The transition wavelength from blocking to transmission
band varies according to the filter’s position in front of the
collimating lens.
LVF-HL Variable Bandwidth Filter
Double High-pass and Double Low-pass Filters
The benefit of having double filters versus a single filter is that the optical density of the blocking
band increases to 99.96%. However, the transmission band is reduced to 80%.
High-pass and Low-pass Variable Bandpass Filters
By fastening together a high-pass filter and a low-pass filter, we created a variable bandpass
notch filter that allows you to adjust the center wavelength and the bandwidth. We preset the
transmission bandwidth at ~25 nm FWHM, but adjusting four screws allows you to slide the
filters against one another to create a wider or narrower transmission bandwidth.
Item
%TRANSMISSION
80
60
40
20
0
250
350
450
550
WAVELENGTH (nm)
650
These spectra were taken with an LVF-HL to show how a
transmission bandwidth can be set throughout the filter’s
range (300-750 nm).
Description
LVF-H
A single high-pass filter for 300-750 nm
LVF-L
A single low-pass filter for 300-750 nm
LVF-HH
Two LVF-H high-pass filters epoxied together for 300-750 nm
LVF-LL
Two LVF-L low-pass filters epoxied together for 300-750 nm
LVF-HL
An LVF-H high-pass filter and LVF-L low-pass filter fastened together to create an adjustable bandpass linear variable filter
LVF-UV-H
A single high-pass filter for 230-500 nm
LVF-UV-L
A single low-pass filter for 230-500 nm
LVF-UV-HH
Two LVF-UV-H high-pass filters epoxied together for 230-500 nm
LVF-UV-LL
Two LVF-UV-L low-pass filters epoxied together for 230-500 nm
LVF-UV-HL
An LVF-UV-H high-pass filter and LVF-UV-L low-pass filter fastened together to create an adjustable bandpass linear variable filter
LVF-CUV-ADP
Adapter for use with cuvette holders, to clamp filter slides into place
CVD-DIFFUSE
Teflon diffuser for cuvette holder-LVF setups, to redirect excitation energy to spectrometer
FHS-LVF
In-line system with two collimating lenses that accommodates slide carriers for straight-through (absorbance/transmission) measurements
LVF-KIT
Consists of the LVF-HL, LVF-CUV-ADP, FHS-LVF, and CVD-Diffuse
LVF-UV-KIT
Consists of the LVF-UV-HL, LVF-CUV-ADP, FHS-LVF and CVD-Diffuse
www.oceanoptics.com Tel: +1 727-733-2447
127
Sampling Accessories
Sampling Accessories
Accessories for Linear Variable Filters
FHS-LVF
The FHS-LVF is an in-line filter holder that is used in absorbance and transmission applications. This in-line LVF
holder features two collimating lenses with SMA 905 connectors. Its slot accommodates the LVF slide carrier.
Screws hold the FHS-LVF in place. For absorbance/transmission measurements only.
Item Code: FHS-LVF
LVF-CUV-ADP
The LVF-CUV-ADP is an adapter piece that fits onto our 1-cm cuvette holders and holds the LVF slide carrier. The
cuvette adapter slides over the top of the cuvette holder and includes screws to clamp the LVF’s slide carrier into
place. The LVF-CUV-ADP comes with a cover to block out ambient light and can accommodate both single-filter
and double-filter slide carriers
Item Code: LVF-CUV-ADP
CVD-DIFFUSE
The CVD-DIFFUSE is a 1-cm cuvette-shaped piece of Teflon with a 45° surface at the measurement height. Use it
in a fluorescence cuvette holder-with-filters configuration to redirect the excitation energy into the spectrometer.
This facilitates setting the filter position or selecting the wavelength passed by the filter.
Item Code: CVD-DIFFUSE
Packages
Select one of the following packages for a convenient, all-in-one filter solution.
Item Code: LVF-KIT includes: LVF-HL, LVF-CUV-ADP, FHS-LVF, CVD-DIFFUSE
Item Code: LVF-UV-KIT includes: LVF-UV-HL, LVF-CUV-ADP, FHS-LVF, CVD-DIFFUSE
Sampling Accessories
Absorbing Glass Filters
High-Pass Filters
We offer our OF2 filters for installing into the optical path of your spectrometer setup.
These high-pass filters are transmissive approximately 50% of the normal cutoff wavelength, >99% at wavelengths 50 nm higher than the cutoff and less than 0.1% at 50 nm
lower than the cutoff. High-pass filters are used to block second orders, test for stray
light and block excitation energy in fluorescence or Raman experiments.
Balancing Filters
Our Balancing Filters absorb energy in some regions while transmitting in others. The
BG34 filter, for example, reduces the intensity of light at 600 nm from a tungsten
halogen bulb while transmitting all of the light at the blue and red regions, where
detector sensitivity is lower.
OF2-WG305
pass >305 nm
square 25.4 x 25.4 x 3 mm
OF2-GG375
pass >375 nm
square 25.4 x 25.4 x 3 mm
OF2-GG395
pass >395 nm
square 25.4 x 25.4 x 3 mm
OF2-GG475
pass >475 nm
square 50.8 x 50.8 x 3 mm or
square 25.4 x 25.4 x 3 mm
OF2-OG515
pass >515 nm
square 25.4 x 25.4 x 3 mm
OF2-OG550
pass >550 nm
square 25.4 x 25.4 x 3 mm
Available Items
OF2-FG3
enhance blue and red
square 25.4 x 25.4 x
3 mm
Bandpass Filters
OF2-BG34R
enhance blue and red
round 12.7 mm OD
These filters transmit a particular wavelength range while rejecting energies higher
and lower than the selected range.
Available Items
Note: Some of 25.4 mm square filters may be available in dimensions suitable for use with the
INLINE-FH filter holder (page 126). Consult an Applications Scientist for details.
128
Available Items
www.oceanoptics.com Tel: +1 727-733-2447
OF2-KG3
>325 nm and
<700 nm
square 25.4 x 25.4 x
3 mm
OF2-U360
>340 nm and
<380 nm
square 25.4 x 25.4 x
3 mm
OF2RG780
>780 nm and 50%
transmission <2.7 µm
square 25.4 x 25.4 x
3 mm
Routing Control and More
Fiber Optic Dual Switch with TTL Line
The FOS-2X2-TTL Fiber Optic Dual Switch was designed to provide you with flexibility in routing, splitting
and controlling light. The FOS has two light channels. You can opt to have one light channel open at a time
or have both closed. The FOS is useful for monitoring the drift of the light source or for measuring two
samples with one spectrometer channel and one light source. The diagram below is an example of how
the FOS can be utilized. In this setup, the FOS eliminates the need for a second spectrometer channel.
Sampling Accessories
Sampling Accessories
Item Code: FOS-2X2-TTL
Sample 1
Light
Source
BIF400-UV-VIS
Sample 2
P400-025-SR
Spectrometer
P400-025-SR
BIF400-UV-VIS
FOS-2X2-TTL
In this setup, light enters a Bifurcated Optical Fiber Assembly and then splits into two arms, one for each sample. Light interacts with each sample
and travels through another fiber assembly, each into its own port in the FOS. Another Bifurcated Assembly collects the light from the FOS and
sends it to the spectrometer. Here you would switch the shutter on the FOS from one light channel to another in order to get clean data from each
sample. Without the FOS, you would need another spectrometer channel to monitor the two samples.
Electronic TTL Shutter
In spectrometer setups, the INLINE-TTL TTL-driven shutter allows you to block the light path
without disturbing the experiment -- for example, by turning the light source on and off.
The laser-cut shutter is installed between two collimating lenses, which attach to two optical fibers. The INLINE-TTL is driven by a small board with a TTL input. Included is a cable for
interfacing to a spectrometer.
Item Code: INLINE-TTL-S
Specifications
Dimensions:
140 mm x 50 mm x 50 mm
Weight:
~600 g
Shutter-Input:
TTL maximum 5 Hz
Power requirements:
12 VDC (power supply included)
Power consumption:
100 mA maximum
Maximum frequency:
5 Hz
Field of View Control
The Gershun Tube Kit (GER-KIT) controls the field of view of our SMA 905-terminated optical
fiber. It also directly attaches to a spectrometer with an SMA 905 Connector. User-interchangeable apertures provide many different fields of view from 1° to 28°. When the GER-KIT
is used with our optical fiber, the field of view cannot exceed the optical fiber’s 25° field of
view if you are measuring radiance.
Item Code: GER-KIT
Specifications
Material:
Black anodized aluminum
Interior:
Bead-blasted surface to reduce off-axis reflections
Connection:
Directly attaches to one of our spectrometers or couples to an SMA 905-terminated
optical fiber with included adapter barrel
Apertures:
1°, 3°, 8°, 10° and 14° apertures included, providing 1°, 2°, 3°, 6°, 8°, 10°, 14°, 16°,
20° and 28° fields of view
www.oceanoptics.com Tel: +1 727-733-2447
129
Sampling Accessories
Sampling Accessories
Filtering Light
MPM-2000 Optical Multiplexer
Our MPM-2000 Fiber Optic Multiplexers take light to your
spectrometer or from a light source (connected to one of the
input ports) and distribute it to either 8 or 16 outputs. The light
is distributed through the output ports in sequential order with
switching times between channels of less than 150 milliseconds.
These precision instruments are ideal for process environments
where multiple locations need to be measured with a single
spectrometer channel or light source.
High-Precision
All versions of our MPM-2000 include a DC motor that provides
excellent speed control without sacrificing power. An included
encoder converts movement into a digital pulsed output. Each
channel in the MPM-2000 has a collimating lens that connects to
an internal optical fiber system. Plus, the MPM-2000 provides accurate measurements with superior repeatability of 99%.
Software Controlled
Our MPM-2000 Multiplexers interface easily with your PC via an RS-232 port and come with software and drivers for complete PC control. The
MPM-2000’s software allows you full control of the switching order, switching delay time and system calibration.
Specifications
Available Items
Dimensions (600 µm
version):
400 mm x 170 mm x 130 mm
MPM-2000-UV-VIS400-1X16
1x input 16x output channels, RS-232 controlled,
24VDC. 400 µm fiber
Dimensions (400 µm
version):
200 mm x 170 mm x 130 mm
MPM-2000-UV-VIS400-2X8
2x input 8x output channels, RS-232 controlled,
24VDC. 400 µm fiber
Wavelength range:
250-800 nm – UV-VIS
350-2000 nm – VIS-NIR
MPM-2000-VIS400-1X16
1x input 16x output channels, RS-232 controlled,
24VDC. 400 µm fiber
Optical throughput:
>60% when using standard 400 µm fibers @
650 nm
MPM-2000-VIS400-2X8
2x input 8x output channels, RS-232 controlled,
24VDC. 400 µm fiber
Motor:
Direct-current
MPM-2000-VIS600-1X16
1x input 16x output channels, RS-232 controlled,
24VDC. 600 µm fiber
MPM-2000-VIS600-2X8
2x input 8x output channels, RS-232 controlled,
24VDC. 600 µm fiber
MPM-2000-UV-VIS600-1X16
1x input 16x output channels, RS-232 controlled,
24VDC. 600 µm fiber
MPM-200-UV-VIS600-2x8
2x input 8x output channels, RS-232 controlled,
24VDC. 600 µm fiber
Repeatability:
>99%
Switching time:
150 ms between adjacent positions
Interface:
RS-232 (optional USB)
Power requirement:
24 VDC, 1.2 A (includes WT-24V-E power supply)
Connectors:
SMA 905
Fiber Optic Variable Attenuator
Our FVA-UV Fiber Optic Variable Attenuator is an opto-mechanical device that helps you control
the amount of light transmitted between two fibers. Fibers screw into either side of the FVA-UV
via SMA 905 connectors with collimating lenses that project light across a metal disk in which a
slit has been cut. The width of the slit varies as a function of manually adjusted radial position.
Rotating the disc varies the attenuation from 0-100% uniformly across a 200-2000 nm wavelength range.
Item Code: FVA-UV Fiber Optic Variable Attenuator, FVA-ADP Adapter for FVA-UV
Specifications
130
Dimensions:
38.1 mm x 59.4 mm x 40 mm
Weight:
90 g
Assembly ports:
3/8-24 threads for collimating lenses
Wheel lock:
6-32 nylon thumbscrew
ADP adapter:
Directly attaches to a light source with a collimating lens
Connector:
SMA 905
www.oceanoptics.com Tel: +1 727-733-2447