NON-INVASIVE

BLOOD PRESSURE

SIMULATORS

NIBP-1030 NIBP-1030-BE

NIBP-1020 NIBP-1010

USER MANUAL

BC BIOMEDICAL

NIBP-1000 SERIES

TABLE OF CONTENTS

WARNINGS, CAUTIONS, NOTICES ........................................................................... ii

DESCRIPTION ............................................................................................................. 1

OVERVIEW .................................................................................................................. 4

KEYS ............................................................................................................................ 7

SCREENS .................................................................................................................... 8

SETUP ......................................................................................................................... 13

CONNECTIONS ........................................................................................................... 16

THEORY OF OPERATIONS ........................................................................................ 19

RUNNING A TEST ....................................................................................................... 21

MANUAL REVISIONS .................................................................................................. 33

LIMITED WARRANTY ................................................................................................. 33

SPECIFICATIONS ....................................................................................................... 34

NOTES ......................................................................................................................... 36 i

WARNING - USERS

The NIBP-1000 Series is for use by skilled technical personnel only.

WARNING - USE

The NIBP-1000 Series is intended for testing only and should never be used in diagnostics, treatment or any other capacity where it would come in contact with a patient.

WARNING - CONNECTIONS

All connections to patients must be removed before connecting the DUT to the NIBP-1000

Series. A serious hazard may occur if the patient is connected when testing with the NIBP-1000

Series. Do not connect any leads from the patient directly to the NIBP-1000 Series or DUT.

CAUTION - MODIFICATIONS

The NIBP-1000 Series is intended for use within the published specifications. Any application beyond these specifications or any unauthorized user modifications may result in hazards or improper operation.

CAUTION - ADAPTER

Turn Power Off and unplug any Battery

Eliminator before cleaning the surface of the NIBP-1000 Series.

CAUTION - CLEANING

Do not immerse. The NIBP-1000 Series should be cleaned by wiping gently with a damp, lint-free cloth. A mild detergent can be used if desired.

ii

CAUTION - LIQUIDS

Do not submerge or spill liquids on the NIBP-

1000 Series. Do not operate the NIBP-1000

Series if internal components may have been exposed to fluid.

CAUTION - SERVICE

The NIBP-1000 Series is intended to be serviced only by authorized service personnel.

Troubleshooting and service procedures should only be performed by qualified technical personnel.

CAUTION - INSPECTION

The NIBP-1000 Series should be inspected before each use for wear and should be serviced if any parts are in question.

CAUTION - ENVIRONMENT

Exposure to environmental conditions outside the specifications can adversely affect the performance of the NIBP-1000 Series. Allow the

NIBP-1000 Series to acclimate to specified conditions for at least 30 minutes before attempting to operate it. iii

NOTICE – CE

The NIBP-1000 Series Simulators bear the mark

Based on the following testing standards:

ELECTROMAGNETIC COMPATIBILITY DIRECTIVE

EMC – Directive 89/336/EEC and 2004/108/EC as amended by

92/31/EEC, 93/68/EEC and Directive 91/263/EEC [ TTE/SES ]

EN 61326-1:1997 + A1:1998 + A2:2001 + A3:2003

“Electrical equipment for measurement, control and laboratory use – EMC requirements”

This equipment has been type tested by an independent, accredited testing laboratory and compliance was demonstrated to the above standard to the extent applicable.

EMISSIONS

Radiated and Line Conducted Emissions

EN 61000-3-2

EN 61000-3-3

Harmonic Current Emissions

Voltage Fluctuation and Flicker

IMMUNITY – CLASS C

EN 61000-4-2

EN 61000-4-3

EN 61000-4-4

EN 61000-4-5

EN 61000-4-6

EN 61000-4-11

Electrostatic Discharge

Radiated Electric Field Immunity

Electrical Fast Transients / Bursts

Surge Voltage

Conducted Disturbance

Voltage Dips and Short Interrupts

LOW VOLTAGE DIRECTIVE

EC – Directive 73/23/EC

EN 61010-1:2001

“Safety requirements for electrical equipment for measurement, control, and laboratory use – General requirements”

This equipment has been type tested and compliance was demonstrated to the above standard to the extent applicable. iv

NOTICE – SYMBOLS

Symbol Description

Caution

(Consult Manual for Further Information)

Center Negative

Per European Council Directive 2002/95/EC, do not dispose of this product as unsorted municipal waste.

NOTICE – ABBREVIATIONS

AC Alternating Current

BP

BPM brpm

Blood Pressure

Beats per minute breaths per minute

C

°

DUT

Dias

ECG

Celsius degree(s)

Device Under Test

Diastolic

Electrocardiogram

Euro

F

European

Fahrenheit

HR Heart Rate

Hz kg hertz kilogram

µV/V/mmHg microvolts(s) per Volt per mmHg mA mm mmHg mV min

Neo

NSR

Ω

Lbs sec

Syst

USA

VDC

VRMS milliampere(s) millimeter(s) millimeter(s) of mercury millivolt(s) minute(s)

Neonatal

Normal Sinus Rhythm ohm(s) pounds second(s)

Systolic

United States of America

Volts Direct Current

Volts Root Mean Square v

NOTICE – DISCLAIMER

BC GROUP INTERNATIONAL, INC. WILL NOT BE RESPONSIBLE

FOR ANY INJURIES SUSTAINED DUE TO UNAUTHORIZED

EQUIPMENT MODIFICATIONS OR APPLICATION OF EQUIPMENT

OUTSIDE OF THE PUBLISHED INTENDED USE AND

SPECIFICATIONS.

NOTICE

BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO

MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT

ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE

DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST

POSSIBLE PRODUCT. THE INFORMATION IN THIS MANUAL HAS

BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE

ACCURATE. HOWEVER, NO RESPONSIBILITY IS ASSUMED FOR

INACCURACIES.

NOTICE

– DISCLAIMER

– CONTACT INFORMATION

BC BIOMEDICAL

BC GROUP INTERNATIONAL, INC.

3081 ELM POINT INDUSTRIAL DRIVE

ST. CHARLES, MO 63301

USA

1-800-242-8428

1-314-638-3800 www.bcgroupintl.com

[email protected]

NIBP-1000 Series Users Manual Copyright © 2012 www.bcgroupintl.com

08/12

Made in the USA

Rev 15 vi

BC BIOMEDICAL

NIBP-1000 SERIES

NON-INVASIVE BLOOD PRESSURE SIMULATORS

The NIBP-1000 Series is a family of microprocessor-based, high-precision Non-Invasive

Blood Pressure (NIBP) Simulators. The units are small, easy to use and have multiple features to fit many different applications. The NIBP-1010 is the basic model, which includes built-in manometer function. The NIBP-1020 adds realistic, full QRS and respiration ECG waveforms. The NIBP-1030 additionally offers Invasive Blood Pressure,

Temperature, Arrhythmias and a Leak Rate test mode.

The graphical display provides multiple screens containing pressure in mmHg, a plot of the overall pressure, or a close-up of the BP waveform.

The following are highlights of the main features:

NIBP-1010 (BASIC FEATURES, NIBP ONLY):

LARGE BACKLIT GRAPHICAL LCD DISPLAY WITH CURSOR SELECTION

OF OPTIONS AND PARAMETERS

FULL RANGE MANOMETER

ADULT, NEONATAL, HYPERTENSIVE, AND HYPOTENSIVE MODES

± 500 mmHg PRESSURE RANGE

DIGITAL PRESSURE ENVELOPE OFFSET

OPTIONAL PEAK PRESSURE DETECT WITH SIMPLE RESET

SpO

2

READY – COMPATIBLE WITH MSP-2100 MODULE

DIGITAL CALIBRATION – NO POTS TO TURN

SELECTABLE DISPLAY OPTIONS AND DIGIT SIZES

SOFTWARE ADJUSTABLE CONTRAST

FLASH PROGRAMMABLE

BATTERY LIFE DISPLAY (0 TO 100%)

OPTIONAL RECHARGEABLE NiMH BATTERIES

BATTERY ELIMINATOR

RS-232 INTERFACE

NIBP-1020 ADDS:

ECG OUTPUT WITH FULL NSR WAVEFORM

SINUSOIDAL RESPIRATION SIMULATION

ECG TEST WAVEFORMS

PACE WAVEFORM

OPTIONAL PEAK PRESSURE DETECT WITH ECG ALARM TEST

ECG SYNCHRONIZED WITH BLOOD PRESSURE

1

NIBP-1030 ADDS:

SYNCHRONIZED INVASIVE BLOOD PRESSURE OUTPUT

SELECTABLE IBP SENSITIVITY 5 OR 40 μV/V/mmHg

STATIC IBP SIMULATION –10mmHg TO 400 mmHg

LEAK RATE TEST

ECG ARRHYTHMIA WAVEFORMS

ECG ARRHYTHMIA SEQUENCE

YSI 400 AND 700 SERIES TEMPERATURE SIMULATION

AVAILABLE MODELS:

NIBP-1010

NIBP-1010-P

NIBP-1010-BP

BASIC UNIT

+ PEAK

+ BATTERY & PEAK

NIBP-1010-KIT NIBP-1010-BP, 5 ADAPTERS & SOFT CARRYING CASE

NIBP-1020 BASIC UNIT WITH ECG

NIBP-1020-PA + PEAK & ALARM

NIBP-1020-BPA + BATTERY, PEAK & ALARM

NIBP-1020-KIT NIBP-1020-BPA, 5 ADAPTERS & SOFT CARRYING CASE

NIBP-1030

NIBP-1030-B

BASIC UNIT WITH ECG, IBP, TEMP, PEAK & ALARM + MANY MORE

+ BATTERY

NIBP-1030-BE + UNIVERSAL PATIENT LEAD POST ECG CONNECTORS

NIBP-1030-KIT NIBP-1030-B, 5 ADAPTERS & SOFT CARRYING CASE

NIBP-1030-EKIT NIBP-1030-BE, 5 ADAPTERS & SOFT CARRYING CASE

STANDARD ACCESSORIES:

AC POWER ADAPTER (EURO) BC20-40614

(OR)

BC20-40613

OPTIONAL ACCESSORIES:

AC POWER ADAPTER (USA)

BC20-40714

BC20-40715

BC20-40716

BC20-40717

BC20-40608

BC20-40602

BC20-40605

DINAMAP/CRITIKON ADAPTER

QUICK DISCONNECT ADAPTER

LUER ADAPTER

MARQUETTE ADAPTER

BULB ADAPTER

NIBP FITTING KIT (HP CUFF ADAPTER,

INFLATION BULB, CPC CONNECTOR, QUICK

COUPLER WITH 1/8” MNPT AND SILICONE

TUBING AND TEE)

NIBP ADAPTER KIT (11 CONNECTORS)

2

BC20-41333

BC20-41334

MSP-2100

FingerSim™ KIT

FingerSim™ SET

BC20-41337

UT-1 YSI-400 TEMPERATURE SIMULATION CABLE

UT-2 YSI-700 TEMPERATURE SIMULATION CABLE

PULSE OXIMETRY MODULE (ALLOWS NIBP TO

SIMULATE SpO

2

FingerSims ™)

OUTPUT WHEN USED WITH

COMPLETE STARTER KIT (INCLUDES 80%, 90% AND

97% FingerSims™, FingerSim™ HOLDER, AND

CARRYING CASE)

REPLACEMENT KIT (80%, 90% AND 97%

FingerSims™ ONLY)

RS-232 COMMUNICATIONS CABLE (MINI-DIN TO

DB-9F)

BC20-41339

BC20-30111

USB COMMUNICATIONS ADAPTER (DB-9M to USB-A)

FOR USE WITH BC20-41337

SOFT CARRYING CASE

INVASIVE BLOOD PRESSURE (IBP) CABLES:

FOR A COMPLETE LIST CONSULT THE BC GROUP WEBSITE AT

WWW.BCGROUPINTL.COM

3

OVERVIEW

This section looks at the layout of the NIBP-1030 and gives descriptions of the elements that are present.

NOTE: The NIBP-1030-BE model is different from the unit shown below in that it uses universal patient lead banana posts that allow for the use of snaps and/or banana-type patient leads.

LCD Graphical Display:

Shows Test Data and

Configuration Data

Male Quick

Connect for

Pressure Input

Patient Lead Connectors:

Left Side Right Side

RA/R V2/C2

LA/L V3/C3

RL/N

LL/F

V1/C1

V4/C4

V5/C5

V6/C6

6 Light Touch Keys for

Selecting Parameters and Settings:

POWER for Turning Unit

On and Off

UP and DOWN Arrows for Scrolling Through

Selected Options.

(When No Options are

Selected, for Changing

Display Screen.)

SELECT for Choosing

Setting or Parameter to

Change

SETUP/RETURN for

Entering and Exiting

Setup Menu

RESET for resetting peak pressure reading.

6 pin Mini-

Din connector for

IBP

Simulation

8-Pin Mini-DIN

Plug Connector for Temperature

Simulation

7 pin Mini-Din connector for

RS-232 or SpO

2

Simulation

Power

2.1 mm Jack,

4

This section looks at the layout of the NIBP-1020 and gives descriptions of the elements that are present.

LCD Graphical

Display:

Shows Test Data and

Configuration Data

Male Quick

Connect for

Pressure

Input

6 Light Touch Keys for

Selecting Parameters and Settings:

POWER for Turning Unit

On and Off

UP and DOWN Arrows for Scrolling Through

Selected Options.

(When No Options are

Selected, for Changing

Display Screen.)

SELECT for Choosing

Setting or Parameter to

Change

SETUP/RETURN for

Entering and Exiting

Setup Menu

RESET for resetting peak pressure reading.

Patient Lead Connectors:

Left Side Right Side

RA/R V2/C2

LA/L

RL/N

LL/F

V3/C3

V4/C4

V5/C5

V1/C1 V6/C6

7 pin Mini-Din connector for

RS-232 or SpO

2

Simulation

Power

2.1 mm Jack

5

This section looks at the layout of the NIBP-1010 and gives descriptions of the elements that are present.

LCD Graphical

Display:

Shows Test Data and

Configuration Data

Male Quick

Connect for

Pressure

Input

6 Light Touch Keys for

Selecting Parameters and Settings:

POWER for Turning Unit

On and Off

UP and DOWN Arrows for Scrolling Through

Selected Options.

(When No Options are

Selected, for Changing

Display Screen.)

SELECT for Choosing

Setting or Parameter to

Change

SETUP/RETURN for

Entering and Exiting

Setup Menu

RESET for resetting peak pressure reading.

7 pin Mini-Din connector for

RS-232 and

SpO

2

Simulation

Power

2.1 mm Jack

6

KEYS

Six tactile-touch keys are provided for system operation:

– This key turns the unit off and on. The unit will return to the screen that was active when it was turned off.

– In the DISPLAY MODE, these keys toggle the display through the available

main screens.

In the SELECT MODE, if a parameter has been highlighted, these keys will scroll through the available settings.

– On the Main screen, this key sequences through the available NIBP or ECG simulations. On the Setup screen, there are a number of parameters that may be selected and changed. This key sequences the cursor (Highlight) through those parameters.

– This key is used to RESET the peak pressure reading or to begin a Leak Test.

– This key toggles the unit into and out of the Setup Mode. Depressing this key will enter the Setup screen where the configuration can be viewed and adjusted.

Depressing the key again will exit the Setup Mode and return to the previously viewed main screen.

7

SCREENS

MAIN SCREENS

– There are five main screens: Pressure Only, Pressure with Output

Waveform, Pressure with Pressure Graph, ECG (optional) and Battery Indicator (optional).

The available screens can be toggled using .

PRESSURE ONLY SCREEN – This screen has a large pressure display, as shown below.

Also displayed on this screen is the peak pressure and selected output waveform

The display will resemble the following:

PRESSURE

OUTPUT SELECTION

Systolic/Diastolic (mean) Mean

Adult 120/80

Adult 120/80 w/Pace

Adult High 190/120

Adult Low 80/40

Neonatal 70/40

Alarm Test (Optional)

(Alternates every 30 sec)

Arrhythmia Seq

(NIBP-1030 Only)

Leak Test

(NIBP-1030 Only)

Manometer

Heart Rate

93 80

93 Pacemaker 80

150 80

53

55

-

80

94

47

160

-

-

-

80

-

-

Respiration

Rate (Optional)

35

35

35

35

47

00

80

35

-

-

PEAK PRESSURE

(Optional)

PRESSURE

ENVELOPE

OFFSET

This line is only displayed if a nonzero value is programmed into the Systolic &/or

Diastolic Shift. (See

Setup for more information).

8

OUTPUT WAVEFORM SCREEN – This screen shows the pressure, peak pressure, output selection, and output waveform.

NOTE

This waveform is not intended to be physiologically correct.

The display will resemble the following:

PRESSURE

OUTPUT SELECTION

Systolic/Diastolic (mean) Mean

Adult 120/80

Adult 120/80 w/Pace

High (Adult) 190/120

Low (Adult) 80/40

Neo (Neonatal) 70/40

Alarm Test (Optional)

(Alternates every 30 sec)

Arrhythmia Seq

(NIBP-1030 Only)

Leak Test (NIBP-1030 Only)

Manometer

Heart Rate

93 80

93 Pacemaker 80

150

53

55

80

80

94

-

47

160

-

-

-

80

-

-

Respiration

Rate (Optional)

35

35

35

35

47

00

80

35

-

-

PEAK PRESSURE

(Optional)

OUTPUT

WAVEFORM

(as generated by the

NIBP-1000 Series)

Dependant on pressure

9

PRESSURE GRAPH SCREEN – This screen provides a graph of the pressure, as well as, the pressure, peak pressure and output selection

PRESSURE

PEAK PRESSURE

(Optional)

PRESSURE

GRAPH

2mmHg per pixel

OUTPUT SELECTION

Systolic/Diastolic (mean) Mean

Adult 120/80

Adult 120/80 w/Pace

High (Adult) 190/120

Low (Adult) 80/40

Neo (Neonatal) 70/40

Alarm Test (Optional)

(Alternates every 30 sec)

Arrhythmia Seq

(NIBP-1030 Only)

Leak Test

(NIBP-1030 Only)

Manometer

Heart Rate

93 80

93 Pacemaker 80

150 80

53

55

-

80

94

47

160

-

-

-

80

-

-

Respiration

Rate (Optional)

35

35

35

35

47

00

80

35

-

-

10

ECG OUTPUT SCREEN – This screen shows the selected ECG output mode.

(NIBP-1020 & NIBP-1030 units only)

NOTE: While in this mode, the NIBP simulation does not run.

NOTE: NSR ECG output is active during NIBP testing at the rate stipulated for the selected test.

The display will resemble the following:

SELECTED

OUTPUT

LIST OF AVAILABLE

OUTPUTS

NSR

NSR

NSR

NSR

Sine Wave

Sine Wave

Sine Wave

30 Bpm

60 Bpm

120 Bpm

240 Bpm

10 Hz

60 Hz

100 Hz

Square Wave

Square Wave

Triangle Wave

PVC Arrhythmia*

Pair of PVC’s*

Run of 5 PVC’s*

0.125 Hz

Ventricular Tachycardia*

Ventricular Fibrillation*

2 Hz

2 Hz

Asystole*

*NIBP-1030 Only

11

BATTERY INDICATOR SCREEN – This screen shows the status of the battery voltage level with Built-In Battery Option only.

NOTE: It is only an estimate of the battery life remaining.

When the level reaches 10%, the BP simulation mode will be disabled; however, the

Manometer and ECG (optional) will continue to function. Once the battery level reaches

0%, the unit will automatically turn itself off to avoid damaging the batteries.

The display will resemble the following:

Batteries charge from provided AC power adapter. The charge time is about 4 hours from full discharge. While the batteries are charging, the display will flash “charging.”

12

SETUP

The Setup Mode allows the user to adjust the configuration of the simulator. The Setup screen can be entered using the key. The parameters can be changed by using

key to highlight the line and to toggle the available options.

The Setup screen can be exited using the key.

NIBP 1010 and NIBP 1020

NIBP 1030

13

The following is a breakdown of the parameters available in the configuration of the unit and their available options:

System Setup Configuration

Parameter

Systolic Shift

Diastolic Shift

Description

Adjusts the Systolic Output of the NIBP

Simulation. This is not a direct mmHg adjustment.

Adjusts the Diastolic Output of the NIBP

Simulation. This is not a direct mmHg adjustment.

Range

±50.0

±50.0

Static BP

(NIBP-1030 Only)

IBP Sens

(NIBP-1030 Only)

Adjusts the Static Blood Pressure Output.

Selects the Invasive Blood Pressure Output sensitivity.

-10 to 400 mmHg

5 μV/V/mmHg

40 μV/V/mmHg

Temp

(NIBP-1030 Only)

Selects the simulated Temperature output.

SpO

2

Output

Auto Off Timer

This parameter determines whether the SpO

2 output pulse is active. The output drives an MSP-

2100 FingerSim Module.

Determines the period of inactivity before the unit is turned OFF. A timer is started when the unit is turned ON and is reset each time a key is pressed.

When the timer reaches the value set in this parameter, the power is automatically turned OFF.

(NOTE: Setting this parameter to 0 disables the

Auto Off timer. When running from line power, the unit does not automatically shut off. Auto Off timer is inactive during a test.)

Contrast Adjust Sets the contrast of the display screen.

Backlight

Battery Life

Off – Always off

1-30 sec

– The elapsed time after which the backlight will automatically turn off.

ON – Always ON.

The Default setting is 30 seconds.

Available only with Battery Option installed.

Displays current life of the battery.

At 10%, a warning screen will appear.

At 0%, the unit will power down automatically.

Software Displays current software program.

0, 24, 30, 35, 37, 40,

42 C

Off/On

0-30 Minutes

0-20

Off, 1-30 sec, ON

0-100%

(Read Only)

(Read Only)

14

SYSTOLIC AND DIASTOLIC SHIFT – The NIBP-1000 Series is equipped with the option to shift test results to compensate for different methods of measuring Oscillometric NIBP by various manufacturers and models of devices under test.

CAUTION

These adjustments must be used with caution as they will allow the user to adjust the output results to invalid values.

These adjustments should only be used to aid in the simplification of testing and with documented controls.

There are no absolute standards for Oscillometric NIBP readings; therefore, for a number of reasons (including patents, technology, etc.), each manufacturer has established a different method for evaluating the oscillometric pulses. Due to these varying methods, precise ly the same waveforms will give different results on different manufacturer’s units.

The normal technique used is to run the monitor against a fixed source like the NIBP-1000, with the understanding that each manufacturer has a predictable error from this norm.

While this is generally the most direct method, users have asked for a method to correct for this difference, making the monitors read the same as the test unit. The Systolic and

Diastolic Shift settings allow for just such correction.

These adjustments are indicated in a line added to the main display to inform the user of any shift that has been programmed into the system. This is done so there is no misunderstanding of the meaning of the results.

15

CONNECTIONS

Connecting Pressure Cuffs

The NIBP-1010, NIBP-1020 and NIBP-1030 Blood Pressure Simulators are connected between the blood pressure cuff and the DUT (Patient Monitor). The blood pressure cuff should be disconnected from the DUT and a ‘T’ adapter inserted between the cuff and the

DUT. The NIBP simulator is then connected to the open side of the ‘T’ adapter as pictured below:

Blood

Pressure

Cuff

NIBP-

10xx

DUT

NOTE: Roll the cuff tightly on itself or around a mandrel.

Connecting Patient Leads

For models with ECG output, snaps or patient lead test posts (NIBP-1030-BE only) are provided along the sides and are identified by the markings on the overlay.

16

Connecting IBP Cables (NIBP-1030 ONLY)

For models with IBP output, connect the DUT-appropriate IBP test cable between the

NIBP-1030 IBP Connector (6-pin Mini-DIN as pictured below) and the DUT. The pin-out shown below for the Mini-DIN connector is the industry standard configuration used on most modern patient simulators. As such, there are many existing cables that may be used or you may obtain the proper cable from BC Group.

NOTE: Mini-DIN pinout viewed from NIBP Connector Side

Connecting Temperature Simulation Cables (NIBP-1030 ONLY)

For the models with Temperature output, connect the DUT-appropriate Temperature

Simulation test cable between the NIBP-1030 Temperature Connector (8-pin Mini-DIN as pictured below) and the DUT.

NOTE: Mini-DIN pinout viewed from NIBP Connector Side

17

Auxiliary Connector

The Auxiliary connector can be used for either standard RS-232 Serial communications using an optional accessory cable or for SpO

2 simulation using the optional accessory

MSP-2100 SpO

2

module available from BC Biomedical.

NOTE: Mini-DIN pinout viewed from NIBP Connector Side

18

THEORY OF OPERATIONS

NIBP (Non-Invasive Blood Pressure) Function

A stepping motor and piston pressure engine is used to generate the output waveforms. A precision pressure sensor is used to measure the pressure of the cuff. The pressure sensor is read by a 16-bit differential Analog to Digital converter.

The cuff pressure is monitored and the motor/piston assembly reacts to provide the correct response based on the selected mode.

IBP (Invasive Blood Pressure) Function

A Digital to Analog converter is used to generate the IBP waveforms. The output circuit is fully isolated and capable of switching between the two standard sensitivities (5

µV/V/mmHg and 40 µV/V/mmHg).

Temperature

There is an 8-pin mini-DIN plug connector on the right side of the unit for connection of a Temperature cable. Temperatures are simulated for both YSI 400 and YSI 700 probe types. There are seven different temperatures selectable for each.

19

ECG Function

A Digital to Analog converter is used to generate the ECG outputs based on stored waveform data for standard ECG and Arrhythmias.

Respiration Function

A Digital to Analog converter is used to generate the respiration waveform. The respiration output is present on the LA ECG lead only. Refer to Monitor manual to determine which lead the Monitor uses for respiration. It may be necessary to reverse the LA and LL leads for respiration to be detected.

20

RUNNING A TEST

Depending on the model, there are up to 9 Basic Test Modes and 3 Sub-Test Modes. This section will walk through each of the tests and their basic operation.

The Main Tests are accessible with a single key. The key will scroll through the following tests in a continuous loop:

Adult 120/80

Adult 120/80 w/Pace

Adult High 190/120

Adult Low 80/40

Neonatal 70/40

Alarm Test

Arrhythmia Seq

Leak Test

Manometer

BASIC TEST MODES

NIBP:

The first 5 test modes deal with various NIBP setups. To run an NIBP simulation, the cuff and monitor are connected to the pressure input. Then the measurement is initiated by the monitor and the NIBP Simulator will output the proper waveform based on the cuff pressure provided by the monitor and the selected simulation.

The NIBP output mode can be changed by pressing the key. Once the desired operating mode is selected, the output will automatically begin when the correct pressure is detected.

21

There are 5 selectable NIBP Basic Test Modes; Adult, Adult w/pace, Adult High, Adult Low and Neonatal. The displays will resemble the following examples:

NOTE: The screens for the first test will have the individual components labeled. The component labels for subsequent tests are the same.

Adult 120/80

(Typical Screens)

PEAK PRESSURE

(Optional)

PRESSURE

NIBP TEST

Type Adult 120/80

Heart Rate 80

Respiration Rate 35

(Optional)

(Pressure Only Screen)

PRESSURE ENVELOPE

OFFSET

This line is only displayed if a non-zero value is programmed into the

Systolic &/or Diastolic

Shift. (See Setup for more information).

PRESSURE

NIBP TEST

Type Adult 120/80

Heart Rate 80

Respiration Rate 35

(Optional)

PRESSURE

(Output Waveform Screen)

PEAK PRESSURE

(Optional)

OUTPUT

WAVEFORM

(as generated by the

NIBP-1000 Series)

Dependant on pressure

PEAK PRESSURE

(Optional)

NIBP TEST

Type Adult 120/80

Heart Rate 80

Respiration Rate 35

(Optional)

PRESSURE

GRAPH

2mmHg per pixel

(Pressure Graph Screen)

22

Adult 120/80 with Pace

(Typical Screens)

Adult High 120/80

(Typical Screens)

23

Adult Low 80/40

(Typical Screens)

Neonatal 70/40

(Typical Screens)

24

ALARM TEST:

Use the key to cycle through the Test Modes until Alarm Test appears. The purpose of this function is to help test the alarms of the monitor under test. The ECG output will alternate from a 47 BPM NSR with apnea respiration (0 brpm) to 160 BPM NSR with 80 brpm respiration. The time interval between alternations is 30 seconds. A count down timer is displayed to indicate the time still remaining for the current output.

NOTE: While in this mode, the NIBP simulation output is disabled. Pressure readings in the screens below are only manometer-type readings of static pressure applied.

The displays will resemble the following examples: (Typical Screens)

PRESSURE

(NIBP Simulation disabled, this is pressure of the static volume only)

PEAK PRESSURE

(NIBP Simulation disabled, this is peak pressure of the static volume only)

TIMER

ALARM TEST

Heart Rate 47/160

Respiration Rate 0/80

(Pressure Only Screen)

PRESSURE

ENVELOPE OFFSET

This line is only displayed if a non-zero value is programmed into the Systolic &/or

Diastolic Shift. (See

Setup for more information).

(Output Waveform Screen) (Pressure Graph Screen)

25

ARRHYTHMIA SEQUENCE: (NIBP-1030 Only)

Use the key to cycle through the Test Modes until Arrhythmia Sequence appears.

The purpose of this function is to provide a timed sequence of some of the more common

Arrhythmias. The test will continually cycle through the 6 arrhythmias in the sequence listed below. A count down timer is displayed to indicate the time still remaining in each step.

NOTE: While in this mode, the NIBP simulation output is disabled.

Step

1

2

Display

Single PVCs

Pair of PVCs

Total Time ECG Output

18 sec 11 NSR, 1 PVC, 11 NSR, 1 PVC

9 sec 10 NSR, Pair of PVCs

3 Single PVCs 18 sec 11 NSR, 1 PVC, 11 NSR, 1 PVC

4 Run of 5 PVCs 12 sec 11 NSR, Run of 5 PVCs

5

6

7

Vtach

Vfib

Asystole

30 sec

30 sec

30 sec

15 NSR, VTach

15 NSR, VFib

15 NSR, Asystole

The displays will resemble the following examples: (Typical Screens)

PRESSURE

(NIBP Simulation disabled, this is pressure of the static volume only)

PEAK PRESSURE

(NIBP Simulation disabled, this is peak pressure of the static volume only)

TIMER

ARRHYTHMIA

Single PVCs

Pair of PVCs

Single PVCs

Run of 5 PVCs

Vtach

Vfib

Asystole

(Pressure Only Screen)

PRESSURE ENVELOPE

OFFSET

This line is only displayed if a non-zero value is programmed into the

Systolic &/or Diastolic

Shift. (See Setup for more information).

(Output Waveform Screen) (Pressure Graph Screen)

26

LEAK TEST: (NIBP-1030 Only)

Use the key to cycle through the Test Modes until Leak Test appears. The purpose of this function is to provide a standard pressure leak test. The pressure input is connected to the system that is to be monitored. Pressure is then applied to the system.

The Leak Test is initiated by pressing the key. The unit traces and displays the pressure drop and the time since the test was initiated. The unit also calculates and displays the leak rate in mmHg per minute.

NOTE: While in this mode, the NIBP simulation output is disabled.

The displays will resemble the following examples: (Typical Screens)

PRESSURE

LEAK TEST

Drop in mmHg

Time in min:sec

Leak Rate in mmHg/minute

PEAK PRESSURE

(Pressure Only Screen)

(Output Waveform Screen) (Pressure Graph Screen)

27

MANOMETER:

Use the key to cycle through the Test Modes until Manometer appears. The purpose of this function is to allow the unit to be used strictly as a Manometer with a full scale range of ± 500 mmHg.

NOTE: The Graphic Screens are also available in this mode for maximum flexibility

The displays will resemble the following examples: (Typical Screens)

PRESSURE

PEAK PRESSURE

(Optional)

(Pressure Only Screen)

PRESSURE ENVELOPE

OFFSET

This line is only displayed if a non-zero value is programmed into the

Systolic &/or Diastolic

Shift. (See Setup for more information).

(Output Waveform Screen) (Pressure Graph Screen)

28

SUB TEST MODES

PEAK: (Optional)

The optional Peak test feature is available whenever the display is in a pressure mode. It will continuously monitor the pressure input and display and hold the maximum value. This value may be reset at any time by pressing the key.

The displays will resemble the following examples: (Typical Screens)

PEAK PRESSURE

(Optional)

PEAK PRESSURE

(Optional)

PEAK PRESSURE

(Optional)

29

ECG PERFORMANCE TEST: (NIBP-1020 & NIBP-1030)

The ECG output is selected by pressing until the ECG Waveforms menu is displayed. The key is used to choose the desired ECG output. There are 16 fixed waveforms available.

NOTE: While in this mode the NIBP simulation is disabled, but the IBP simulation is active

(NIBP-1030 only).

The display will resemble the following example:

SELECTED

OUTPUT

LIST OF AVAILABLE

OUTPUTS

NSR

NSR

NSR

NSR

30 Bpm

60 Bpm

120 Bpm

240 Bpm

Sine Wave

Sine Wave

Sine Wave

Square Wave

10 Hz

60 Hz

100 Hz

0.125 Hz

Square Wave

Triangle Wave

PVC Arrhythmia*

Pair of PVC’s*

Run of 5 PVC’s*

Ventricular Tachycardia*

2 Hz

2 Hz

Ventricular Fibrillation*

Asystole*

*NIBP-1030 Only

30

DYNAMIC IBP: (NIBP-1030 Only)

The IBP output matches the NIBP setting in the NIBP mode or the ECG setting in the ECG

Waveform mode.

NOTE: The IBP sensitivity (5 µV/V/mmHg or 40 µV/V/mmHg) must be correctly selected in the SETUP Mode before using IBP. (See Setup section for more information.)

STATIC IBP: (NIBP-1030 Only)

A static Blood Pressure output may be selected in the SETUP mode. This fixed value will remain on the IBP output until changed.

NOTE: The IBP sensitivity (5 µV/V/mmHg or 40 µV/V/mmHg) must be correctly selected in the SETUP Mode before using IBP. (See Setup section for more information.)

31

TEMPERATURE SIMULATION: (NIBP-1030 Only)

The temperature setting can be selected in the SETUP mode. This fixed value will remain on the Temperature output until changed. The output will simulate both YSI 400 and YSI

700 Temperature probes.

NOTE: Both outputs are available at the output connector simultaneously.

NOTE: On some Patient Monitors, if ECG simulation is active on the NIBP-1030 and you enable Temperature Simulation, the ECG trace may disappear from the Patient Monitor screen. This is a flaw of the Patient Monitor, not the NIBP-1030 Simulator.

32

MANUAL REVISIONS

Rev 14

Rev 15

Revision # Program #

Rev 01

Rev 02

Rev 03

Rev 04

Rev 05

Rev 06

Rev 07

Rev 08

Rev 09

Rev 10

Rev 11

Rev 12

Rev 13

DT7355CA

DT7355CA

DT7355CA

DT7355CF

DT7355CG

DT7355CI

DT7355CI

DT7359CA

Revisions Made

Origination

Specification Information Updated

Misc. Edits

Add Systolic/Diastolic adjustments

Output Selection edited

Misc. Edits

Warnings, Cautions, Notices Updated

NIBP-1030, IBP, ECG Arrhythmia Waveforms Added

DT7359CA

DT7359CA

Temperature Added

Temp Connector Updated, Static 120 Added

DT7355CO, Address Updated, Screen Captures Updated, Output

DT7359CH Selection Charts Updated

DT7355CO, Misc. Edits

DT7359CH

DT7355CO, Added NIBP-1030-BE model, revised accessories

DT7359CH listing, and other miscellaneous edits.

DT7355CO, Format Updated, Misc. Edits

DT7359CH

DT7355CO, Misc. Edits

DT7359CH

LIMITED WARRANTY

WARRANTY : BC GROUP INTERNATIONAL, INC. WARRANTS ITS NEW PRODUCTS TO BE FREE

FROM DEFECTS IN MATERIALS AND WORKMANSHIP UNDER THE SERVICE FOR WHICH THEY

ARE INTENDED. THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM THE DATE OF

SHIPMENT.

EXCLUSIONS: THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR

IMPLIED, INCLUDING, BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY

OR FITNESS FOR A PARTICULAR PURPOSE.

BC GROUP INTERNATIONAL, INC. IS NOT LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL

DAMAGES.

NO PERSON OTHER THAN AN OFFICER IS AUTHORIZED TO GIVE ANY OTHER WARRANTY OR

ASSUME ANY LIABILITY.

REMEDIES: THE PURCHASER'S SOLE AND EXCLUSIVE REMEDY SHALL BE: (1) THE REPAIR OR

REPLACEMENT OF DEFECTIVE PARTS OR PRODUCTS, WITHOUT CHARGE. (2) AT THE OPTION

OF BC GROUP INTERNATIONAL, INC., THE REFUND OF THE PURCHASE PRICE.

P:\Manuals\BCGroup\ …\NIBP1000_UM_Rev15.doc

33

SPECIFICATIONS

MANOMETER (PRESSURE MEASUREMENT)

RANGE ± 500 mmHg @ 20 °C

ACCURACY

RATE

± (1% of Reading + 0.5 mmHg @ 20 °C)

NON-INVASIVE BLOOD PRESSURE (NIBP) SIMULATION

80, 94 BPM

(NOTE: Synchronized to ECG on NIBP-1020/1030)

ACCURACY ± 1%

INVASIVE BLOOD PRESSURE (IBP) SIMULATION (NIBP-1030 ONLY)

STATIC PRESSURE

-10, -5, 0, 20, 40, 50, 60, 80, 100, 120, 150, 160, 200,

240, 250, 300, 320, 400 mmHg

ACCURACY

IMPEDANCE

± (1% Full Range + 1 mmHg) or

± (2% Setting + 2 mmHg) whichever is better

300 Ω, ± 10%

EXCITATION RANGE 2 to 16 VRMS

EXCITATION FREQUENCY DC to 5 kHz

SENSITIVITY 5 or 40 µV/V/mmHg

ECG SIMULATION (NIBP-1020 & NIBP-1030)

NSR RATE 30, 60, 120, 240 BPM

PERFORMANCE

WAVEFORMS

SINE RATE

SQUARE RATE

TRIANGLE RATE

10, 60, 100 Hz

0.125, 2 Hz

2 Hz

RATE ACCURACY ± 1%

AMPLITUDE 2.75 mV ± 2% @ Lead II

PACEMAKER SIMULATION WAVEFORMS (NIBP-1020 & NIBP-1030)

AMPLITUDE 3 mV ± 10%

WIDTH 6 ms ± 5%

34

RESPIRATION SIMULATION (NIBP-1020 & NIBP-1030)

RATE ACCURACY ± 1%

DELTA 3.0 Ω ± 10%

IMPEDANCE

BASELINE 1000 Ω ± 5%

TEMPERATURE SIMULATION (NIBP-1030 ONLY)

SENSOR TYPE

POINTS

ACCURACY

AC POWER ADAPTER

YSI Series 400 and 700

0, 24, 30, 35, 37, 40, 42 °C

(32, 75.2, 86, 95, 98.6, 104, 107.6 °F)

± 0.1 °C

ELECTRICAL

12 VDC, 500 mA

BC20-40613 (USA)

BC20-40614 (Euro)

BATTERY (OPTIONAL)

Rechargeable NiMH pack

(Not user serviceable)

BATTERY RUNTIME

DISPLAY

500 Test Cycles between charges

PHYSICAL & ENVIRONMENTAL

128 X 64 Pixels Graphical LCD,

White LED Backlight

SIZE

WEIGHT

NIBP-1000 through

NIBP-1030

NIBP-1030-BE

7.70 x 5.36 x 4.07 Inches

(195.6 x 136.1 x 103.4 mm)

7.70 x 6.15 x 4.07 Inches

(195.6 x 156.2 x 103.4 mm)

< 3.5 Lbs (1.6 kg)

CONSTRUCTION

OPERATING RANGE

STORAGE RANGE

ENCLOSURE

OVERLAY

ABS Plastic

Back-Printed Lexan

15 to 40 °C

(59 to 104 °F)

-20 to 65 °C

(-4 to 149 °F)

35

NOTES

36

NOTES

37

NOTES

38

BC GROUP INTERNATIONAL, INC.

3081 ELM POINT INDUSTRIAL DRIVE

ST. CHARLES, MO 63301

USA

1-800-242-8428

1-314-638-3800 www.bcgroupintl.com

[email protected]

NIBP Series User Manual

08/12

Made in the USA

– Rev 15