Assembly instructions induSENSOR Serie LVP

Assembly instructions induSENSOR Serie LVP
Sensor rod
2x M4, length
of thread (min.)
16 mm (.63 “)
Plunger
Notes on CE Identification
The following applies for the induSENSOR, LVP series:
EU directive 2004/108/EC
EU directive 2011/65/EG, “RoHS“ category 9
The induSENSOR, LVP series, satisfies the requirements of the standards
-- EN 61326-1: 2006-10
-- DIN 61326-2-3: 2007-05
Sensor mounting with mounting kit MBS 12/8
53 (2.09)
38.5
(1.52)
Mounting plate
7-pin
socket
radial
Dimensional drawing housing ZA, dimensions in mm (inches), not to scale
Mounting of the housing GA, dimensions in mm (inches), not to scale
The sensor is screwed to the mounting plate using the M18 thread, see
figure above.
Measuring range
A
50 (1.97)
77 (3.03)
100 (3.94)
138 (5.43)
200 (5.91)
261 (10.3)
B
M2
M3
M3
C
10 (.39)
12 (.47)
12 (.47)
ø 30.5
(1.20 dia.)
SW 32
12 (.47)
38.5
(1.52)
ø 12
(.47 dia.)
B
ø8
(.31 dia.)
12
(.47)
Dimensional drawing housing GA, dimensions in mm (inches), not to scale
min. 30 3
(1.18)
2x M4,
length of thread
(min.) 12 mm (.47 “)
C
D
50 (1.97)
100 (3.94)
200 (7.87)
1) Plunger at start of measuring range, I OUT = 4 mA
2) Plunger at end of measuring range, I OUT = 20 mA
3) Bending radius sensor cable: > 30 mm (once) > 90 mm (repeated)
min. 30 3
(1.18)
38.5 (1.52)
ø 12
(.47)
Do not move the plunger to the back-stop of the sensor rod.
>> Damage to or destruction of the sensor
2
ø 30.5 (1.20 dia.)
Avoid bending the coil rod or the measuring tube.
>> Damage to or destruction of the sensor
Sensor
housing
38.5
(1.52)
Mounting Kit MBS 12/8
Avoid shocking and knocking the sensor.
>> Damage to or destruction of the sensor
D
(.98±.04)
53 (2.09)
>> Damage to or destruction of the sensor
25±11
A
Mounting of the Housing ZA
The sensor is mounted with peripheral clamping on the sensor rod, see
figure below.
MICRO-EPSILON recommends to use the mounting set MBS 12/8, which is
available as an accessory. At installation locations where there are no forces and vibrations the sensor can also be mounted by the sensor rod using
radial point clamping with set screws. Plastic set screws must be used so
that the sensor rod is not damaged or deformed.
48 (1.89)
Power supply must be connected in accordance with the safety regulations
for electrical equipment.
>> Danger of injury
7-pin
socket
axial
SW 32
ø8
(.31 dia.)
>> Damage to or destruction of the sensor
Installation and Mounting
Centering and Mounting the Plunger
Screw the plunger to the target using the thread. The screw joint must
either be secured with a screw locking compound (e.g. Loctite) or counter-screwed with the locknut supplied. When mounting, please ensure that
the plunger remains freely movable in the sensor and that tilting is avoid.
Observe the measuring tube position at the zero point (= 4 mA output).
48 (1.89)
13.8
(.54)
M 18x1.5
Warnings
The power supply may not exceed the specified limits.
>> Danger of injury
Mounting of the Housing GA
IP 67
-40 to +85 °C (-40 to +185 °F)
RL ≤ 500 Ohm)
-40 to +100 °C (-40 to +212 °F)
5 - 95 % (non condensing)
Atmospheric pressure
EN 61326-1: 2006-10 and
DIN 61326-2-3: 2007-05
M18x1.5
Assembly Instructions
induSENSOR, LVP series
Proper Environment
-- Protection class for sensor:
-- Operating temperature: -- Storage temperature: -- Humidity: -- Ambient pressure:
-- EMC acc. to: 12 (.47)
Mounting of housing ZA
Mounting block
MBS12/8
4 mounting holes for
screws M4 DIN 912
Precautionary Measure
The plunger must not contact the sensor rod during operation.
>> Damage to or destruction of the sensor through abrasion.
Power Supply and Display / Output Device
Power supply and signal output are effected through the 7-pin. connector
at the sensor’s electronic housing. The pin assignment is shown in drawing
and table stated below.
Avoid bending or shortening the plunger.
>> Loss of specified technical data
The minimum bending radius is 30 mm (1.2 “) once, repeated: 90 mm
(3.5 “) for the sensor cable C 703x.
>> Damage to or destruction of the sensor cable
i
The specified technical data are valid only, if the plunger supplied
from MICRO-EPSILON is used!
Output signal (mA)
Measuring Principle
54
2
6
7
1
View: Solder
pin side,
female cable
connector
20
Pin Assignment
Color C703
1
Supply + (18 ... 30 VDC)
white
2
0 V Ground
I OUT 4 ... 20 mA
(U OUT 1 ... 5 V) 1
brown
3
4
green
Signal ground
yellow
5
SCL (sensor calibration)
gray
6
SDA (sensor calibration)
pink
7
not connected (n.c.)
blue
With a preset load resistor the maximum permitted operating temperature
is calculated as:
0.04 °C * R L
3.3 °C * U B
;
Note: T max ≤ 85 °C
+
T max = 150 °C Ohm
V
1
1/2
0
Measuring range
Plunger
i
Sensor rod
Ue
Ia
Pin 2 is connected with pin 4 on the sensor electronics. The screen of the
C703 sensor cable is connected with the connector housing.
Connect the screen of the C703 sensor cable with the protective earth
conductor on power side.
The sensors are connected according to the pin assignment, see opposite
table and Figures “Signal monitoring“. Notice the different criteria:
The maximum load resistor RL is limited by the operating voltage UB.
(U B - 10 V)
R L max =
20 mA
18 ... 30 VDC
2
3
+
UB
_
2
3
+
RL
U OUT
V
Signal monitoring with load resistor and voltmeter
+
UB
_
Make sure, that the output noise of the power supply units does not
exceed 5 mVss, if the sensors are supplied through switched-mode
power supply units.
18 ... 30 VDC
LVP
4
LVP
1) With sensor cable C703-5/U
1
1
_
R L = Load resistor
U B = Operating voltage
T max = Maximum operating temperature
Pin and color assignment of 7-pin. connector and sensor cable C703-5
respectively C703-5/U
12
4
3
A small load resistor increases the thermal load of the sensor electronics.
With a maximum operating temperature of 85 °C (+185 °F), the minimum
permitted load resistor RL is calculated as:
82.5 Ohm * U B
- 1625 Ohm (If the result is negative: R L = 0 )
R L min =
V
RL
+
I OUT
A
R L = Load resistor
U B = Operating voltage
T max = Maximum operating temperature
More information about the sensor can be read in the instruction manual.
You will find these online at:
www.micro-epsilon.de/download/manuals/man--induSENSOR-Serie-LVP-de-en.pdf
_
4
Signal monitoring with amperemeter
i
R L can be inserted as an option to adapt the power loss to high
ambient temperatures.
If the signal is monitored with a voltmeter, the load resistor R L is dimensioned in accordance with the desired output voltage U OUT.
Formula: U OUT = R L * I Signal
MICRO-EPSILON MESSTECHNIK GmbH
& Co. KG
Königbacher Str. 15 · 94496 Ortenburg
www.micro-epsilon.com
X9771074-A011065HDR
*X9771074-A01*