Quickie Pulse® Series Power Wheelchair Service manual


Add to my manuals
72 Pages

advertisement

Quickie Pulse® Series Power Wheelchair Service manual | Manualzz

Quickie Pulse 6 Service Manual

©2011 Sunrise Medical Inc.

111425 Rev B

Quickie Pulse 6 Service Manual Contents

Introduction 0.1

Basic Setup 0.2

Motor - Drive Gear engaged

Motor - Free-wheel

0.2

0.2

Multimeter Tutorial

The Multimeter

The Probes

0.3

0.3

0.3

The Ports 0.3

Symbols 0.4

Health and Safety

Good Working Practices

Battery Safety

Battery Chargers

EMI Warnings

Electro Static Discharge 0.8

Batteries 0.9

Battery Diagnostics 0.10

Battery Types

VR2 Remote Controller

0.11

0.12

0.5

0.5

0.5

0.6

0.6

VR2 Plugs/Connectors

R-NET Remote Controller

R-NET Remote Controller w/Display

0.13

0.14

0.15

R-NET Plugs/Connectors

Main Wiring Diagram VR2 (PLS)

Main Wiring Diagrams Rnet (PLS)

Main Wiring Diagram VR2 and RNET

(PLS6A/PLS6B)

0.16

0.17

0.18

0.19

Main Wiring Diagrams Actuators

(PLS6A/PLS6B) 0.20

VR2 Dual Attendant System Connection

Basic Tool List

Battery Connection Test

Check Battery Wire Harness

Circuit Breaker Test

0.21

1.1

1.2

1.2

1.2

Main Harness

Section 2

VR2 Remote Controller Display

The Maximum Speed Indicator Ripples

The Maximum Speed Indicator Flashes

Battery Gauge is Steady

Battery Gauge Flashes Slowly

Battery Gauge Steps Up.

Battery Gauge Blinks Once Every 2.5 Seconds 2.1

Battery Gauge Flashes Rapidly 2.1

2.1

2.1

2.1

2.1

2.1

2.1

Section 3

VR2 Controller Diagnostic Codes

One Bar - Low Battery Voltage

Two Bars - Left Motor Disconnected

3.1

3.1

3.1

Three Bars - Left Motor Wiring Trip

Four Bars- Right Motor Disconnected

Five Bars - Right Motor Wiring Trip

Six Bars - Charger Connected

Seven Bars - Possible Joystick Trip

Seven Bars + Speed Profile Indicator error

Eight Bars - Possible Control System Trip

Nine Bars - Solenoid Brake Trip

Ten Bars - High Battery Voltage

Section 4

R-net Troubleshooting Procedures

Chair Will Not Power Up

R-net Troubleshooting Procedures (cont.)

Section 5

4.1

4.1

4.2

R-net Fault Codes 5.1

Power Chair Displays a Fault on the Hand Control or

Omni 5.1

Table 1, Error Codes

Example of R-10 Fault Isolation

5.1

5.4

3.4

3.4

3.4

3.5

3.5

3.2

3.2

3.3

3.4

Power Chair Will Not Drive Full Speed 5.4

Power Chair Will Not Drive in Creep Speed When

Tilted 5.6

Tilt Will Not Operate 5.7

Section 6

Motor/Gearbox Inspection 6.1

Removal Procedures

Section 7

Battery Removal

Battery Installation

Motor and Gearbox Removal

7.1

7.1

7.3

7.4

Motor Removal for chairs after

S/N PLS-101919 and s/n prefix PLS6A &PLS6B 7.6

Control Module Removal

Shock Removal

7.7

7.9

Front Caster Arm Removal

Rear Caster Arm Removal

Front or Rear Caster Fork Removal

Caster Removal

Tilt Actuator and Micro-Switch Removal

Section 8

Seat/Back Width adjustment

Legrest adjustment

Back-rest Angle Adjustment

Powered elevating or articulating legrest

7.10

7.12

7.13

7.14

7.15

8.1

8.4

8.5

8.6

PAGE 2 SEPT 2011 PULSE 6 SERVICE MANUAL SUNRISE MEDICAL

Introduction

Please read and follow instructions in this service manual before attempting to troubleshoot or repair this product for the first time. If there is anything in this Service Manual that is not clear, or if you require additional Technical assistance, contact Sunrise Medical at: (800) 333-4000 option 2, then option 1.

Safely troubleshooting and/or repair of this product depends on your diligence at following the instructions within this manual. Sunrise Medical is not responsible for injuries or damage resulting from a person’s failure to exercise good judgement and/or common sense.

This Service Manual is intended as a troubleshooting guide for the Quickie Pulse 6. Photographs and content may differ from the actual products in some cases due to changes in specifications and other factors.

This Service Manual is intended for use by persons with a basic working knowledge and the skills required in servicing and maintaining Power Wheelchairs. Persons without a general working knowledge and expertise in the servicing of this product should not carry out troubleshooting procedures. This can result in personal safety issues, problems with future servicing, and/or damage to the unit.

Parts, configuration, and/or specifications of Products included in this Service Manual are subject to change without prior notice.

There are warning symbols used in this document that are intended to focus attention on any hazard that could effect the safety of the individual troubleshooting the Power Wheelchair covered by this Service Manual.

Click !

For up-to-date parts, and the latest version of this Service Manual go to: www.sunmed.com

and click on the Parts Search button

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 0.1

Basic Setup

When setting up the components of the chair, complete the following checklist to ensure proper and safe operation of the equipment.

Check :

□ Are the batteries fully charged?

a. Test battery voltage with D.C. meter across the terminals of batteries. The measurement should be above 12 volts D.C. (Note: a fully charged battery is between 12.9 and 13.1 VDC) b. If not, fully charge the batteries.

□ Are all necessary power components installed and connected ?

a. Input device (normally Joystick) b. Cable from Joystick to the Bus Line c. Control Module; for the Pulse located at back of chair behind shroud d. Are batteries installed correctly? (refer to Section 1 of this manual

□ Are all necessary connections fastened and fully engaged?

a. Battery connectors to the batteries b. Cable between Joystick and the Control c. Both Motor Connectors to the Control Module.

□ Is the Drive Gear engaged? See (fi gures .2.1 and .2.2) a. With the power off, the chair should not move if pushed from behind. b. If the chair moves when pushed, the drive gear needs to be engaged on both motors. c. Does Display light up when Power On/Off switch is depressed?

If no - Follow checklist a second time, then refer to the section on Diagnostics.

If yes – The Power Wheelchair is ready to drive!

Motor - Drive Gear engaged Motor - Free-wheel

PAGE 0.2

SEPT 2011 fig .2.1

PULSE 6 SERVICE MANUAL fig. 4.3.4

SUNRISE MEDICAL

Multimeter Tutorial

The Multimeter

For Powerchairs, the multimeter is one of the most useful tools in the toolbox. It can be used to check wires, shorts, voltages, resistance, and all manner of electrical circuits. This tutorial is designed to help clarify the symbols and socket options.

MULTIMETER

The Probes

Probes are found on various multimeters. The

Probes connect the meter to the circuit. Simply touch them to the directed area with the Mulitmeter on the correct settings, and read the display.

Follow instructions in this manual carefully to avoid errant reads.

PROBES

The Ports

1.

2.

3.

4.

The Common Port.

Generally, the black probe plugs in here

(negative) and as the name suggests, it’s the common element to all of the testing circuits.

Think of it as the ground rail.

Voltage, Resistance and Continuity port.

This is a commonly used option. Connect the red (positive) probe to this port when fi nding voltage readings, resistance readings or when checking wire continuity. This is explained in more detail later in the tutorial.

Current up to 300mA.

This port is used for “counting electrons” in a circuit, and thus their rate of fl ow (current being the fl ow of electrons). You’ll notice that this side is “fused”, so that you don’t end up melting the meter’s circuits.

Current up to 10A.

Same as above, except it can take more current, as the name suggests.

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL

PORTS

SEPT 2011 PAGE 0.3

Multimeter Tutorial

Symbols

This section describes the basic symbols used in a typical multimeter.

AC

ALTERNATING CURRENT Use this when you want to test something that has AC current running through it. Typically you’d want to test the voltage of an inverter (for cold cathodes or neons) or a similar device.

DC

DIRECT CURRENT. This is the type of electrical power produced by a battery. With a battery connector, the black wire(s) should be connected to the negative(-) terminal of the battery and should be considered the common ground. The red wire(s) should be connected to the positive(+) terminal of the battery and is considered the “hot” lead.

Voltage

This means Voltage or Potential Difference. It will measure the potential difference between the two probes. To measure voltage, connect the positive probe to a port that is marked “V” or Voltage.

Note: “mV” means milli-volts = .001 Volt

Current

Since current is measured in Amps and the readout value is in amps, the symbol "A" is used. On this setting the unit measures current that is fl owing through the part of the circuit between the two probes (the meter itself). Typically, you need to plug the positive terminal into a port marked “A” or Current. You need to put the meter “In Series” in the circuit to use this feature correctly.

Resistance

This symbol represents Resistance and is measured in Ohms.

You can use this setting to measure the resistance between two points; for example across a piece of wire or a resistor (to check its value). If you don’t have a continuity check, then this can be used to check for shorts. Any value below 0.05 Ohms constitutes a short, meaning that whatever the probes are attached to is connected electrically.

Continuity

A commonly used function. By putting a current through the two terminals (the same as the Ohm-meter function) it can be determined if the resulting value is within the “contact” range, this is signifi ed with a beep. The feature found on some multimeters enables you to check for shorts without taking your eyes off your work. Other meters signify this with a small fl ashing light.

PAGE 0.4

SEPT 2011 PULSE 6 SERVICE MANUAL SUNRISE MEDICAL

Health and Safety

Good Working Practices

Health and Safety

While working on powered mobility products, it is essential to observe good working practices. Below are a series of safety guidelines and recommendations. Please note that these precautions are intended to serve only as a guide, not to supersede or replace any safety statute, NHS or other safety regulations.

General

• Always wear suitable protective clothing when handling batteries.

• Always wear suitable eye protection when drilling or inspecting.

• When safe to do so, wear protective gloves when handling the running gear or batteries, as these parts are exposed to paths, parks etc.

• If the drive wheels have to be raised off the fl oor, always use a pair of axle stands to secure the vehicle.

Battery Safety

Use extra caution when working with batteries.

Always make sure that the batteries are disconnected from the device before commencing work.

Always check that the battery charger is disconnected from the device

/batteries before commencing work.

Do not smoke while working on this device.

Keep batteries away from all sources of ignition.

Do not place objects on top of the batteries.

Always keep someone close to your work area so that they may come to your assistance if needed.

Always wear personal protection when handling batteries, including, eye/face protection and gloves.

Make sure there is easy access to soap and water in case of acid spills.

Avoid touching eyes or unprotected parts of the body while working on batteries.

Remember that non-sealed batteries can contaminate any packaging, housing, or boxes they may have been transported in so handle all packaging with care, especially during disposal.

If battery acid should come into contact with bare skin or clothing, be sure to wash contacted area immediately, using plenty of soap and water. If battery acid enters the eyes, fl ush with running cold water for as long as possible while medical help is being sought.

When the tops of batteries are exposed, take extra care when working on or around the terminals.

Do not allow metal tools to drop on to or touch the exposed terminals of the batteries or other exposed connections, as this could cause a short circuit, which may result in an explosion.

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 0.5

Health and Safety

Remove personal items of jewelry, such as rings, watches, chains etc. before working on batteries. Such items could cause short circuits resulting in serious burns.

Batteries are constructed of heavy materials. Therefore moving batteries requires appropriate lifting techniques. Safety footwear should also be worn. In addition, disposal of old batteries requires correct procedures. Contact your local authority for their recommendations.

Battery Chargers

Remember battery chargers are connected to household current.

Always observe all guidelines and laws relating to electrical equipment.

Never operate the battery charger in wet or damp conditions.

If you think that the charger has been exposed to water or excessive dampness, do not use it. Return the unit to the dealer/supplier for inspection/replacement.

If you think the battery charger is defective or is visibly damaged, return the unit to the dealer/supplier for inspection.

EMI Warnings

EMI means electromagnetic (EM) interference (I). EMI comes from radio wave sources, such as radio transmitters and transceivers. A

“transceiver” is a device that both sends and receives radio wave signals.)

There are a number of sources of intense EMI in our daily environment. Some of these are obvious and easy to avoid. Others are not, and we may not be able to avoid them.

Powered wheelchairs, although tested in accordance with EMC guidelines, may be susceptible to electromagnetic interference (EMI) emitted from sources such as, radio stations, TV stations, amateur radio (HAM) transmitters, two-way radios, and cellular phones.

EMI can also be produced by conducted sources or electro-static discharge (ESD).

What effect can EMI have?

1. EMI can, without warning, can cause a power chair to:

• Release its electronic brakes

• Move by itself

• Move in unintended directions.

If any of these occur, severe injury could result.

2. EMI can damage the control system of a power chair, resulting in a safety hazard and/or costly repairs.

PAGE 0.6

SEPT 2011 PULSE 6 SERVICE MANUAL SUNRISE MEDICAL

Health and Safety

Sources of EMI

1. Hand-Held Transceivers: Antenna is usually mounted directly on the unit.

These include:

Citizens band (CB) radios

“Walkie-talkies”

Security, fi re and police radios

Cellular phones

Lap top computers with phone or fax

Other personal communication devices

Note - These devices can transmit signals while they are on, even if not in use.

The wheelchair should be switched off when not in use.

2. Medium-Range Mobile Transceivers: include two-way radios used in police cars, fi re engines, ambulances and taxi cabs. The antenna is usually mounted on the outside of the vehicle.

3. Long-Range Transceivers: These include commercial radio and TV broadcast antenna towers, amateur (HAM) radios, and alarm systems.

NOTE- The following are Not likely to cause EMI problems: Lap-top computers

(without phone or fax), cordless phones, TV sets or AM/FM radios, CD or tape players.

EM energy rapidly becomes more intense as you get closer to the source. For this reason, EMI from handheld devices is of special concern. A person using one of these devices can bring high levels of EM energy very close to a power chair without the user’s knowledge.

Immunity level

The level of EM is measured in volts per metre (V/m). Every power wheelchair can resist EMI up to a certain level. This is called its “immunity level”. The higher the immunity level, the less the risk of EMI. It is believed that a 20 V/m immunity level will protect the power wheelchair user from the more common sources of radio waves.

For the Pulse 6, the confi guration tested and found to be immune to at least 20

V/m is: a right-handed mounted joystick system, 18” seat width, 18” seat depth, dual-post height-adjustable armrests, fi xed tapered legrests with one-piece solid footplate and Gp 24 gel cell batteries.

The following dealer installed speciality input devices have an unknown effect on the immunity level because they have not been tested with the Quickie control systems:

Breath Control (“Sip n Puff”)

Tri-Switch Head Array

Proximity Head Array

Proportional Mini-Joystick/Chin Control

Buddy Button

Wafer Board

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 0.7

Electro Static Discharge

To help prevent Electro Static Discharge (ESD) the following proper handling techniques should be followed:

ESD:

Do not place Printed Circuit Boards or their containers near sources of strong electrical fi elds (such as above a CRT).

To avoid the occurrence of static charge or discharge due to friction, keep the Printed Circuit Boards separate from one another and do not stack them directly on top of one another if not protected by antistatic bags.

Store each Printed Circuit Board in an antistatic bag with an external cushioning bubble-wrap layer until assembled to wheelchair. Antistatic bag must have metal content to protect the printed circuit board. Gray bag protects from ESD, pink bag or bubble wrap does not protect as well.

Always wear an ESD preventive wrist or ankle strap when handling electronic components. Connect one end of the strap to an ESD jack or an unpainted metal component on the system

(such as a captive installation screw).

Handle Printed Circuit Boards by the edges only; avoid touching the Printed Circuit Board and connector pins.

Place any removed Printed Circuit Board on an antistatic surface or in a static shielding bag.

Avoid contact between the Printed Circuit Boards and clothing.

The wrist strap only protects the card from ESD voltages on the body; ESD voltages on clothing can still cause damage.

Make sure that the Printed Circuit Board power is off by disconnecting the seating harness prior to attaching or removing printed circuit board.

Printed Circuit Board Flexing:

• The printed circuit board has surface-mount components that may break when the board is fl exed. To minimize the amount

• of board fl exing, observe the following precautions:

Hold the printed circuit board only by the edges.

Do not place the printed circuit board on a hard surface.

Tighten board mounting screws only hand tight (torque12.4 in lbs/1.4Nm) in a cross pattern to reduce stress on mounting holes and PCB board material.

PAGE 0.8

SEPT 2011 PULSE 6 SERVICE MANUAL SUNRISE MEDICAL

Batteries

Safety

If mishandled batteries can be dangerous and hazardous.

All mobility batteries, whether wet type or gel/sealed type, contain lead and sulfuric acid. Both of these materials are toxic and in the case of sulfuric acid, highly corrosive. Additionally, when batteries are charged, they produce hydrogen gas which is “highly” fl ammable and can cause explosion. This is why proper handling is mandatory at all times.

Battery explosion - This is frequently the result of too low an acid/ electrolyte level in the battery, which allows high concentrations of hydrogen to build up. This is possible with all batteries if improper charging or battery failure occurs, but not common in gel/sealed batteries.

< KEEP SPARKS AND FLAMES AWAY FROM BATTERIES >

Burns - dropping a wrench or screwdriver across battery terminals results in sparks, and intense heat. Improper assembly of battery boxes or battery box wiring may short the battery through the wiring and produce a possible electrical fi re.

Electronic damage - batteries that are improperly wired can short out electronic chair components resulting in expensive repairs.

Pollution - improper disposal of batteries could damage the environment. All batteries should be disposed of through a reliable battery recycler.

Battery Charge Cycle Illustration

Typical Flooded Battery

Discharge

POS = PbO2

NEG = Pb

ACID = H2SO2

+

+

+

+

+

H2SO2

-

-

-

-

-

+

+

+

+

+

-

-

-

-

-

+

+

+

+

+

-

-

-

-

-

+

+

+

+

+

H2O

-

-

-

-

-

POS = PbSO4

NEG = PbSO4

ACID = H2O

Recharge

As battery discharges, the sulfate from the electrolyte forms on the plates.

As battery recharges, the sulfate is driven back into the electrolyte

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 0.9

Batteries

Battery Diagnostics

Batteries should last an average of 1 to 1.5 years.

These are some Factors that affect battery performance:

Maintenance - Poor maintenance.

Charging - Improper charging shortens battery life.

Chair Components - Malfunctioning electronics, bad motors, electric brakes, and corroded wiring are also capable of degrading the battery performance.

Battery Servicing and Replacement

Automobile batteries, which are used for starting, are tested with a load tester to assure a high rate of energy production in a short burst.

Deep-cycle batteries produce energy more slowly and are designed to hold up to constant discharging and recharging. Testing a deep-cycle wheelchair or scooter battery requires different procedures than an automobile battery.

A routine for testing deep-cycle batteries should follow these guidelines:

Never replace just one battery at a time. This will create an imbalance when charging and ultimately damage both batteries.

Check batteries for a voltage difference. A voltage difference of more than .4 volts D.C. is a true indicator of a bad battery.

Voltage test - A dead battery cannot be effectively tested, yet many people mistakenly try to do just that.

Any battery that reads 11.0 volts or less is technically dead.

To perform any testing, especially a load test:

A. Batteries must be charged

B. The top charge must be taken from fully charged batteries if charge rate has just fi nished.

The voltmeters on load testers are not accurate enough to establish a state of charge.

Load Test - This test can only be done on fully charged batteries and can only diagnose one type of problem, an internal short.

Capacity/Discharge Test - This is the only accurate way to test a deep-cycle battery for adequate running time. The problem with this test is that it is time consuming.

Current / Voltage check

with a regular interval check - Another way of truly knowing how much time your battery will last, but it is also time consuming.

PAGE 0.10

SEPT 2011 PULSE 6 SERVICE MANUAL SUNRISE MEDICAL

Batteries

Battery Types

IT IS THE RESPONSIBILITY OF THE INSTALL-

ER TO KNOW WHAT KIND OF BATTERIES TO

INSTALL IN A CUSTOMER’S WHEELCHAIR!

Deep-cycle batteries are designed to be discharged and recharged on a regular basis.

Starting or automotive type batteries use a rapid burst of power to start an engine and are quickly recharged by an alternator or generator. They are rated by cold cranking amps, a measure that has no relevance to wheelchair application.

Marine and RV batteries frequently are not deep-cycle as they are often used for starting engines.

Only use Deep-Cycle sealed type batteries in a wheelchair.

Battery Size

• Batteries function as a power wheelchair’s fuel tank. The larger the group size, the farther the wheelchair will go.

• Use the size specifi ed by the wheelchair manufacturer. Never use undersized batteries.

22 NF

Battery Installation.

• •

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 0.11

VR2 Remote Controller

VR2 Controller Buttons

Battery Gauge

A series of ten LED’s, which indicate charge level, and is also used for determining fault codes.

On/Off Key- Press to power on or off the power chair or Controller.

Horn Key-

Activates a warning horn.

Speed/ Profile Decrease. Used to decrease the Speed/ Profile setting.

Speed/Profile indicator A series of five LED’s, whichdisplay speed and profile settings

Speed/ Profile Increase.

Used to

Increase the Speed/ Profile setting.

PAGE 0.12

SEPT 2011 PULSE 6 SERVICE MANUAL SUNRISE MEDICAL

VR2 Plugs/Connectors

Charger Port

Joystick 4pin connector

+

Motor Plug Port

+

_

_

VR2 Controller

2

3

1 1 = 24 Vdc (POS)

2 = 0 Vdc (NEG)

3 = Inhibit 1/

Programmer

Charger port

On-Board Charger - Used to power Switchcontrolled actuators and Tilt Inhibit.

1 = 24 Vdc

2 = 0 Vdc

3 = INHIBIT 1/

PROGRAMMER

M1 = LEFT SIDE MOTOR

M2 = RIGHT SIDE MOTOR

JSM = JOYSTICK MODULE

INH-2 = INHIBIT 2

A1 = ACTUATOR 1

A2 =ACTUATOR 2

OBC = ON BOARD

CHARGER

+ - =BATTERY

PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 0.13

SUNRISE MEDICAL

4 = Red (+)

3 = Yellow

2 = White

1 = Black (-)

R-NET Remote Controller

On/Off Key- Press to power on or off the power chair or Controller.

Speed Profile Decrease .

Used to decrease the

Speed/ Profile setting.

Horn Key- Activates a warning horn.

S

peed/ Profile Increase.

Used to Increase the

Speed/ Profile setting.

PAGE 0.14

SEPT 2011 PULSE 6 SERVICE MANUAL SUNRISE MEDICAL

R-NET Remote Controller w/Display

On/Off Key

- Press to power on or off the power chair or

Controller.

Speed Profile Decrease

.

Used to decrease the

Speed/ Profile setting.

Horn Key- Activates a warning horn.

S

peed/ Profile Increase.

Used to Increase the

Speed/ Profile setting.

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 0.15

R-NET Plugs/Connectors

Charger Port

1 2

3 4

4 = Black (-)

3 = Red (+)

2 = Blue

1 = White (-)

2

3

1

1 = 24 Vdc (POS)

2 = 0 Vdc

3 = Inhibit (NEG)

+

Motor Plug Port

+

_

_

Rnet Controller

PAGE 0.16

SEPT 2011

On-Board Charger - Used to power Switchcontrolled actuators and Tilt Inhibit.

1 = 24 Vdc

2 = 0 Vdc

3 = INHIBIT 1/

PROGRAMMER

PULSE 6 SERVICE MANUAL

M1 = LEFT SIDE MOTOR

M2 = RIGHT SIDE MOTOR

INH-2 = INHIBIT 2

A1 = ACTUATOR 1

A2 =ACTUATOR 2

OBC = ON BOARD

CHARGER

+ - =BATTERY

SUNRISE MEDICAL

Main Wiring Diagram VR2 (PLS)

4-Way

Tyco Bus

VR2

Attendant

Hand Control

VR2 BUS

SPLITTER

4-Way

Tyco Bus

VR2

Hand Control

Offboard

Charger

Drive

Left

4-Way

Tyco Bus

Power Module

VR2

2-Way PGD

Actuator

Inhibit2

4-Way

Intech

Motor

2-Way VR2

Intech Battery

Inhibit3

4-Way

Intech

Motor

2-way

SB50

Red

Fuse

Red Black

Battery

POWER HARNESS

2-way

SB50

Black

Circuit

Breaker

Fuse

SEE NOTE

"A"

Red Black

Battery

Drive

Right

VR2 DRIVE-THRU SINGLE ACTUATOR

4-Way Amp

Mate-N-Loc

ACTUATOR ADAPTOR

HARNESS

6-Way

Mini-Fit J r.

ACTUATOR

HARNESS

6-Way

Mini-Fit J r.

Switch

Seat Tilt

NOTE "A"

CIRCUIT BREAKER USED ON CHAIRS

PRIOR TO PLS-100808 ONLY

4-Way

Tyco Bus

VR2 BUS

SPLITTER

4-Way

Tyco Bus

VR2

Attendant

Hand Control

VR2

Hand Control

Offboard

Charger

Drive

Left

4-Way

Tyco Bus

2-Way PGD

Actuator

Power Module

VR2

Inhibit2

4-Way

Intech

Motor

2-Way VR2

Intech Battery

Inhibit3

4-Way

Intech

Motor

2-way

SB50

Red

Fuse

Red Black

Battery

POWER HARNESS

Circuit

Breaker

2-way

SB50

Black

Fuse

Red Black

Battery

ACTUATOR ADAPTOR

HARNESS

6-Way

Mini-Fit J r.

4-Way Amp

Mate-N-Loc

Drive

Right

SEE NOTE

"A"

SUNRISE MEDICAL

DUAL-TOGGLE

HARNESS

VR2 DUAL-TOGGLE SINGLE ACTUATOR

PULSE 6 SERVICE MANUAL

ACTUATOR

HARNESS

4-Way Amp

Mate-N-Loc

6-Way

Mini-Fit J r.

Switch

Seat Tilt

Dual-Toggle

Single Actuator

Driver

3.5 mm Stereo

Phone Jack

SEPT 2011

Dual-Toggle

Switch

PAGE 0.17

Main Wiring Diagrams Rnet (PLS)

Drive

Left

External 3rd Party

Devices

9-Way "D"

Connector

Output Mod le

4-Way

PGD Bus

4-Way

PGD Bus

External 3rd Party

Wireless Devices

Bluetooth Modu le

4-Way

PGD Bus

4-Way

PGD Bus

4-Way

PGD Bus

4-Way

PGD Bus

Connector Block

4 X CAN

4-Way

PGD Bus

4-Way

PGD Bus

Attendant Contr ol

4-Way PGD

Bus

4-Way PGD

Bus

Power Module

EL

2-Way PGD

Actuator

Inhibit2

4-Way

Intech

Motor

2-Way VR2

Intech Battery

Inhibit3

4-Way

Intech

Motor

2-way

SB50

Red

Fuse

Red Black

Battery

POWER HARNESS

2-way

SB50

Black

Circuit

Breaker

Fuse

SEE NOTE

"A"

Red Black

Battery

4-Way Amp

Mate-N-Loc

ACTUATOR ADAPTOR

HARNESS

6-Way

Mini-Fit J r.

Drive

Right

Ext. Switch

Switch/

Potentiometer

Option

Remote

Switch Option

Ext. Switch

Specialty Input

Device

9-Way "D"

Connector

9-Way "D"

Connector

Specialty Input

Device

Charger

ONMI

Universal Specialty

Control Interface

Infrared

Device(s)

4-Way

PGD Bus

Joystick

3-Way Neutrik

Charge Port

4-Way PGD

Bus

ACTUATOR

HARNESS

6-Way

Mini-Fit J r.

Switch

Seat Tilt

R-NET DRIVE-THRU SINGLE ACTUATOR

NOTE "A"

CIRCUIT BREAKER USED ON CHAIRS

PRIOR TO PLS-100808 ONLY

Drive

Left

External 3rd Party

Devices

9-Way "D"

Connector

Output Mod le

4-Way

PGD Bus

4-Way

PGD Bus

4-Way PGD

Bus

External 3rd Party

Wireless Devices

Bluetooth Modu le

4-Way

PGD Bus

4-Way

PGD Bus

2-Way PGD

Actuator

4-Way

PGD Bus

4-Way

PGD Bus

Connector Block

4 X CAN

4-Way

PGD Bus

4-Way

PGD Bus

Power Module

EL

Inhibit2

4-Way

Intech

Motor

2-Way VR2

Intech Battery

Inhibit3

4-Way

Intech

Motor

2-way

SB50

Red

Fuse

Red Black

Battery

POWER HARNESS

2-way

SB50

Black

Circuit

Breaker

SEE NOTE

"A"

Fuse

Red Black

Battery

104977-020

ACTUATOR ADAPTOR

HARNESS

6-Way

Mini-Fit J r.

4-Way Amp

Mate-N-Loc

Drive

Right

Attendant Contr ol

4-Way PGD

Bus

108277-100

DUAL-TOGGLE

HARNESS

Ext. Switch

Switch/

Potentiometer

Option

Remote

Switch Option

Ext. Switch

Specialty Input

Device

9-Way "D"

Connector

9-Way "D"

Connector

Specialty Input

Device

Charger

ONMI

Universal Specialty

Control Interface

Infrared

Device(s)

4-Way

PGD Bus

Joystick

3-Way Neutrik

Charge Port

4-Way PGD

Bus

020063-040

ACTUATOR

HARNESS

4-Way Amp

Mate-N-Loc

6-Way

Mini-Fit J r.

Dual-Toggle

Single Actuator

Driver

3.5 mm Stereo

Phone Jack

R-NET DUAL-TOGGLE SINGLE ACTUATOR

Switch

Seat Tilt

Dual-Toggle

Switch

NOTE: HARNESS 108277-100 IS INCLUDED IN RETRO ASSEMBLY 110842

PAGE 0.18

SEPT 2011 PULSE 6 SERVICE MANUAL SUNRISE MEDICAL

Main Wiring Diagram VR2 and RNET (PLS6A/PLS6B)

4-Way

Tyco Bus

102359

(Included with Attendant Control)

4-Way

Tyco Bus

4-Way

Tyco Bus

VR2 BUS

SPLITTER

101626

SEAT

102175

VR2

Attendant

Hand Control

Pulse BASE

VR2

Hand Control

105375 - 4 Button

105376 - 6 Button batc8ct

Offboard

Charger

SEAT

BASE

BASE

4-Way

Tyco Bus

2-Way PGD

Actuator

Power Module

VR2-90

Inhibit2

Inhibit3

VR2-90

4-Way

Intech

Motor

2-Way

Intech Battery

4-Way

Intech

Motor

113351-045

Drive

Left

104742

113212 (Pulse 6)

Fuse

Red Black

Battery

2-Way

SB50 Black

2-way

SB50

Red

104741

Fuse

Red Black

Battery

Drive

Right

113213 (Pulse 6)

118557-100

4-Way

R-net Bus

2-Way PGD

Actuator

Power Module

R

-Net EL-90

Inhibit2

Inhibit3

R-Net EL-90

4-Way

Intech

Motor

Drive

Left

113351-045

113212 (Pulse 6)

104742

Fuse

Red Black

Battery

2-Way

Intech Battery

2-way

SB50

Black

2-way

SB50

Red

4-Way

Intech

Motor

104741

Fuse

Red Black

Battery

Drive

Right

113213 (Pulse 6)

External 3rd Party

Wireless Devices

R-NET Input and Output

BUS Connections

External 3rd Party

Devices

9-Way "D"

Connector

Output Modle

108217

4-Way

PGD Bus

4-Way

PGD Bus

Bluetooth Module

108218

4-Way

PGD Bus

4-Way

PGD Bus

108316-120

108316-050

4-Way

PGD Bus

4-Way

PGD Bus

Connector Block

4 X CAN

108312

4-Way

PGD Bus

4-Way

PGD Bus

Intellegent

Seating

Module

(ISM)

108316-030

Attendant Control

108214

4-Way PGD

Bus

108316-120

108316-120

SEAT

Ext. Switch

Ext. Switch

Right or Left

Hand Mount

4-Way

PGD Bus

Switch/Pot

Option

Remote

Switch Option

Joystick

108207

108210

116525 batc8ct

Offboard

Charger

Specialty Input

Infrared

Device

Device(s)

9-Way "D"

Connector

9-Way "D"

Connector

OMNI

Universal Specialty

Control Interface

112731

Right or Left

Hand Mount

4-Way PGD

Bus

Specialty Input

Device

BASE

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 0.19

Main Wiring Diagrams Actuators (PLS6A/PLS6B)

3-AXIS ACTUATOR DRIVER - TILT

020063-100

ACTUATOR

HARNESS 6-Way

Mini-Fit Jr.

109230

Switch

Seat Tilt

2-Way PGD

Actuator

Inhibit2

Power Module

Inhibit3

ACTUATOR

SHORTING PLUG

(Use in place of tilt for legs only.)

119131

6-Way

Mini-Fit Jr.

116507-150

POWER/INHIBIT

HARNESS

3-Way Molex

Mini-Fit Jr.

020063-160

ACTUATOR

HARNESS 6-Way

Mini-Fit Jr.

DUAL ELRs:

LEFT: 111206

RIGHT: 111207

Right/

Extend

ELR

020063-160

ACTUATOR

HARNESS

6-Way

Mini-Fit Jr.

Left/Lift

ELR

6-Way

Mini-Fit Jr.

6-Way

Mini-Fit Jr.

6-Way

Mini-Fit Jr.

113277

Centermount

3-Axis Actuator Driver

116509

DB9 Switch Controls

139221

2-Way PGD

Actuator

Inhibit2

Power Module

Inhibit3

Drive Only

116467

SHORTING

PLUG

(7.5 KOhm)

2-Way PGD

Actuator

Inhibit2

Power Module

Inhibit3

Thru-Drive

PM Driven Tilt

111438-020

DRIVE

THRU

HARNESS

SEAT

4-Way Amp

Mate-N-Loc

BASE

6-Way

Mini-Fit Jr.

020063-070

104977-020 ACTUATOR

ADAPTER

HARNESS

Power Module AUX Connections

Single Actuator Driver

2-Way PGD

Actuator

Inhibit2

Power Module

Inhibit3

Tilt

SEAT

139920

Buddy Button

Switch

104977-020

ACTUATOR

ADAPTOR

HARNESS

6-Way

Mini-Fit Jr.

BASE

020063-070

4-Way Amp

Mate-N-Loc

3.5 mm Stereo

Phone Jack

Dual-Toggle

Single Actuator

Driver

TILT:

112563

PAGE 0.20

SEPT 2011 PULSE 6 SERVICE MANUAL SUNRISE MEDICAL

VR2 Dual Attendant System Connection

1.

A s shown below, the 4 or 6 button joystick module must be connected to the short branch of the attendant module intermediate cable and the attendant joystick must be connected to the long branch.

SHOR T

B R A NCH

L ONG

B R A NCH

2.

If these connections are reversed a “7-flash” communications fault will be displayed on the 4 or 6 button joystick module.

• 19mm socket wrench

• 18mm combination wrench

• 17mm Deep Socket wrench

• 17mm Open End wrench

• 13mm combination wrench

• 13mm" Socket wrench

• 10mm open end wrench

• 13mm Open End wrench

• 5mm socket wrench

• 3/8 combination wrench

Basic Tool List

Basic Tool List

This list of tools will be needed to accomplish

all of the tasks given in this Technical Manual

some are used often.

• 3mm Hex Key/THandle

• 4mm Hex Key/THandle

• 5mm Hex Key/THandle

• Phillips screwdriver #2

• Cutter for zip-tie

• Needle nose pliers

• Flat blade screwdriver

• 3mm or 1/8" Pin Punch

• Multi-Meter

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 0.21

Section 1

Troubleshooting: No Power

Battery Connection Test

1. Check that the female VR2 Bus plug on the chair has voltage. Set the meter to DC volts and measure pins 4 (using the red lead of the meter) and 1 (using the black lead of the meter) as shown in (fig 1.1.1)

2. If the voltage meter reads full voltage, then replace the joystick module

If the voltage meter reads zero voltage measure the corresponding pins on the VR2 controller as shown in (fig 1.1.2).

If the voltage meter reads full voltage, then replace the jumper cable. If the voltage meter reads zero, then measure the Battery Connector as shown in (fig 1.1.3) If the voltage meter reads full voltage, replace the controller, or proceed to the next step.

Fig 1.1.1

Fig 1.1.2

Fig 1.1.3

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 1.1

Section 1

Troubleshooting: No Power (cont.)

Check Battery Wire Harness

Check that the battery wire harness has the correct polarity. Set the meter to dc volts and measure the connector with the red lead on the

+ terminal and the black lead on the negative terminal as shown in (figure 1.2.1). If the voltage is absent proceed to battery fuse test. If the polarity is reversed correct battery wiring.

Battery Fuse

Check that the battery fuse is in good condition.

With the batteries disconnected, set the meter to ohms and measure the resistance across the fuse. see (figure 1.2.2). If the meter reads more than one ohm, change the wiring harness, or else proceed to the next step.

Fig 1.2.1

Fig 1.2.2

Circuit Breaker Test

For chairs prior to PLS-100808

To check the circuit breaker set the meter to ohms and measure the resistance across the circuit breaker as shown in (figure 1.2.3) if the meter reads more than 1 ohm, then change the circuit breaker, otherwise proceed to next step.

Fig 1.2.3

Main Harness

If the above steps did not correct the problem, change the main harness.

PAGE 1.2

SEPT 2011 PULSE 6 SERVICE MANUAL SUNRISE MEDICAL

Section 2

VR2 Remote Controller Display

The Maximum Speed Indicator Ripples

Indicates that the wheelchair is locked. To unlock the wheelchair, deflect the joystick forwards until the control system chirps. Then deflect the joystick in reverse until the control system chirps. Release the joystick, there will be a long beep. The wheelchair is now unlocked. To lock the wheelchair, while the control system is switched on, depress and hold the on/off button. After 1 second, the control system will chirp. Now release the on/off button, deflect the joystick forwards until the control system chirps, and deflect the joystick in reverse until the control system chirps. Release the joystick, there will be a long beep. The wheelchair is now locked.

The Maximum Speed Indicator Flashes

This indicates that the chair is charging . The chair will be ready to drive as soon as the charger is unplugged.

Battery Gauge is Steady

This indicates that all is well.

Battery Gauge Flashes Slowly

The control system is functioning correctly, but you should charge the battery as soon as possible. At

22 V, the red light starts to blink. Each bar represents a .5V value. The controller requires 18V to start and a minimum of 16V to work once started

Battery Gauge Steps Up.

Indicates the wheelchair batteries are being charged with the offboard charger. You will not be able to drive the wheelchair until the charger is disconnected and you have reset the control system by switching off the power and then powering up again.

Battery Gauge Blinks Once Every 2.5 Seconds

The control system has "gone to sleep" because the wheelchair has not been driven for a period of time. The time period depends on the programming of the system. To re-start, reset the system by switching off the power and then powering up again.

Battery Gauge Flashes Rapidly

Make sure the Joystick is completely released (Joystick should be centered and/or nothing is pushing the gimbale out of center). The control system safety circuits have been activated and the control system has been prevented from moving the wheelchair. This indicates a system trip, i.e. the VR2 has detected a problem somewhere in the wheelchair's electrical system. Please refer to Section 3

(VR2 Controller Diagnostics).

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 2.1

Section 3

VR2 Controller Diagnostic Codes

One Bar - Low Battery Voltage

This code could indicate discharged batteries, failed batteries, or poor battery connections.

Begin by recharging the batteries and then refer to Section 1 to check batteries and connections.

Two Bars - Left Motor Disconnected

Check that the batteries are fully charged and in good condition; and check all cables and connections. Check the connections to the left motor, look for a loose or damaged connector.

Use the meter to check the resistance across the two bottom contacts (thicker wires) on the 4-pin motor connector as shown in (figure 3.1.1). If the meter reads between 0 to 1.5 ohms, then replace the controller. If none of the above corrects the problem, replace the left motor.

Fig 3.1.1

Fig 3.1.1

Fig 3.1.2

Otherwise, check the brushes on the left motor

Using a Flat head screwdriver (figure 3.1.2) remove the brush cap to access the brush (figure

3.1.3) Ensure that they are not excessively worn, see curvature on brush (figure 3.1.4) Replace as required.

Fig 3.1.3

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011

Fig 3.1.4

PAGE 3.1

Section 3

VR2 Controller Diagnostics Codes (cont.)

Three Bars - Left Motor Wiring Trip

Check that the batteries are fully charged and in good condition; and check all cables and connections. Check the connections to the left motor, look for a loose or damaged connector.

Measure the resistance from the bottom contact of the red thick wire on the 4-pin left motor connector to each of the top contacts of the connector (figure 3.2.2). Measure the resistance from the bottom contact of the black thick wire on the 4-pin left motor connector to each of the top contacts of the connector. If all of the readings are open, then replace the controller. If any of the readings are short, then replace the left motor.

Test 1

Test 3

Fig 3.2.1

Test 2

Test 4

Four Bars- Right Motor Disconnected

Check that the batteries are fully charged and in good condition; and check all cables andconnections. Check the connections to the right motor, look for a loose or damaged connector.

Use the meter to check the resistance across the two bottom contacts of the thicker wires on the 4-pin motor connector as shown in (figure

3.3.1). If the meter reads between 0 to 1.5 ohms, then replace the controller. If none of the above corrects the problem, replace the right motor.

Fig 3.2.2

PAGE 3.2

SEPT 2011 PULSE 6 SERVICE MANUAL SUNRISE MEDICAL

Section 3

VR2 Controller Diagnostics Codes (cont.)

Otherwise, check the brushes on the Right motor

(figure 3.3.2). Using a Flat head screwdriver remove the brush cap to access the brush.

Ensure that they are not excessively worn, see curvature on brush (figure 3.2.1) Replace as required.

Fig 3.3.1

Fig 3.3.2

Five Bars - Right Motor Wiring Trip

Check that the batteries are fully charged and in good condition; and check all cables and connections. Check the connections to the right motor, look for a loose or damaged connector.

If the reading is short (resistance is less than

10 K ohms) on any of the readings, proceed to check the 4-pin motor connector. Measure the resistance from the bottom contact of the red thick wire on the 4-pin right motor connector to each of the top contacts of the connectors see

(figure 3.3.3). Measure the resistance from the bottom contact of the black thick wire on the 4-pin right motor connector to each the top contacts of the connector (below right). If all of the readings are open, then replace the controller. If any of the readings are short, then replace the right motor.

Test 1

Test 3

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL

Fig 3.3.3

Test 2

Test 4

SEPT 2011 PAGE 3.3

Section 3

VR2 Controller Diagnostics Codes (cont.)

Six Bars - Charger Connected

The Onboard Batteries are being charged with the off-board charger. You will not be able to drive the wheelchar until the charger is disconnected. You will have to reset the control system by switching off the power and then powering up again. The On-Board charger has no indication that the chair is charging, and the chair will not move until complete. If the condition still exists after the charger has been diconnected, and the chair has been switched off and then powered up again, the Joystick module may be defective.

Seven Bars - Possible Joystick Trip

A joystick trip is indicated. Make sure that the joystick is in the center position before switching on the control system. Check that the batteries are fully charged and in good condition, examine the joystick for damage. This fault can also be caused by a joystick that fails to center itself due to being dirty, bent or broken. If this is the case, replace the joystick module.

Note: If replacing the joystick does not resolve the issue, replace the cable connecting the joystick to the controller.

Seven Bars + Speed Profile Indicator error

Inspect the wiring between joystick module and controller. Replace the jumper or joystick module with damaged wiring. If the problem persists replace the controller.

Eight Bars - Possible Control System Trip

Controller Fault - A control system trip is indicated. Make sure that all connections are secure. Check that the batteries are fully charged and in good condition, and check all joystick connections and cables. If this does not correct the problem,disconnect the power to the controller for 2 minutes, replug in to reboot the module. If the condition still exits, then replace the controller.

PAGE 3.4

SEPT 2011 PULSE 6 SERVICE MANUAL SUNRISE MEDICAL

Section 3

VR2 Controller Diagnostics Codes (cont.)

Nine Bars - Solenoid Brake Trip

The parking brakes have a bad connection. Check the parking break and motor connections. Make sure the control system connections are secure.

Measure the two small contacts on the four-pin motor connector (figure 3.5.1). If both motor connectors read approximately 60 ohms, (for chairs prior to PLS-101619 only)

then replace the controller. Otherwise replace the motor that does not read approximately 60 ohms.

For chairs after PLS-101619 or with serial number prefix PLS6A, or PLS6B, it should read ≈13.3 ohms. If both motor connections read ≈13.3 ohms, then replace the controller.

Otherwise replace the motor that does not read

≈13.3 Ohms.

Figure 3.5.2 shows Motor in unlocked (freewheel) 9bar state.

Fig 3.5.1

fig. 3.5.2

Ten Bars - High Battery Voltage

An excessive voltage has been applied to the control system. This is usually caused by a poor battery connection. Check the battery connections.

Battery Fault

Check that the batteries are fully charged, the correct voltage and in good condition. Take a voltage reading from pin 1 and pin 2 of the charger port of the VR2 controller, see (figure 3.5.2) If the meter reads more than 30 volts, then check the charger. Otherwise, replace your controller.

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL

2 1

3

SEPT 2011

Fig .3.5.2

PAGE 3.5

Section 4

R-net Troubleshooting Procedures

Chair Will Not Power Up

1. Check for battery voltage at the hand control using a Multimeter. Connect the Multimeter between the two outside pins (figure 4.1.1), pin 1 is positive (+), pin 2 is negative (-). Note: Positive is on the right.

2

3

1

2. If battery voltage is present, replace the

following components in this order: a. Cables b. Hand control c. Control module. (refer to the Control Module

Replacement section of this manual.) d. Retest as necessary.

A

3. Verify that the buss cables are correctly mated between the hand control and the control module

(figure 4.1.2).

Note: This connector is mated incorrectly.

The connectors are designed to visually indicate when they are not mated correctly.

If yellow is showing between the halves (A), push them closer together until only black is visible. Retest as necessary.

Fig. 4.1.1

Fig. 4.1.2

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 4.1

Section 4

R-net Troubleshooting Procedures (cont.)

5. Disconnect the power connector from the control module, and use a Multimeter to check for battery voltage at the connector (figure 4.2.1).

If voltage is present, replace the control module.

Retest as necessary.

Note: The power connector is the larger 2-pin connector between the left and right motor cable.

6. Manually tilt the seat back and remove the shroud from the base. Refer to Battery Removal section of this manual.

7. For chairs prior to PLS-100808, verify that the circuit breaker located at the front of the power chair is not tripped (figure 4.2.2). Reset if it is tripped and retest. fig 4.2.1

fig. 4.2.2

8. Disconnect the batteries. (also see Subsection 7.1)

PAGE 4.2

SEPT 2011 PULSE 6 SERVICE MANUAL SUNRISE MEDICAL

Section 4

Rnet Troubleshooting Procedures (cont.)

9. Verify that battery voltage is present at each connector leading to the batteries (fi g. 4.6). fig. 4.3.1

10. If battery voltage is not present, use a

Multimeter and measure for continuity across the fusible links connected to positive (+) terminal of each battery (fi g. 4.7). Normal resistance is less than 1 ohm. If open, replace the defective battery

.

harness. Retest as necessary. fig. 4.3.2

Caution: To prevent damage to the Multimeter, ensure that the batteries are disconnected.

11. For chairs prior to PLS-100808, Check for continuity across the circuit breaker (fi gure 4.3.3).

If Normal resistance is less than 1 ohm. If open, replace the circuit breaker. Retest as necessary.

Note: Access to the circuit breaker terminals is easier if the front battery is removed..

12. If the above tests pass, replace the wiring harness leading from the batteries through the

.

circuit breaker and control module. Retest as necessary.

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL fig. 4.3.3

SEPT 2011 PAGE 4.3

Section 5

R-net Fault Codes

Power Chair Displays a Fault on the Hand Control or Omni

The R-Net control used on this power chair is constantly monitoring for conditions that can cause unsafe or erratic operation. When a fault is displayed, refer to the fault code table in this manual for a list of corrective actions.

Identifi ed Module

The following identifi es which module of the control system has registered the problem.

• PM-Power Module (Control Module)

• JSM-Joystick Module/Omni Module

• ISM-Intelligent Seating/lighting Module

Table 1, Error Codes

PM

Module Error

Trip Text

Trip Text

Joystick Error

Low Battery

High Battery

M1 Brake Error

M2 Brake Error

M1 Motor Error

M2 Motor Error

Inhibit Active

Jstick Cal Error

Latched Timeout

Brake Lamp Short

Left Lamp Short

Right Lamp Short

L Ind Lamp Short

R Ind Lamp Short

L Ind Lamp Failed

R Ind Lamp Failed

Trip Code

0506

Corrective Action

Ensure that the joystick is centered upon power up. If it is centered, replace the hand control.

Recharge the batteries. After charging, perform the Battery Testing section of this manual.

Verify that the battery charger is not defective. Leave the chair on for a few minutes to drain off the excess charge. Check the condition of the battery charger.

Verify that the left motor is connected to the control module. Complete the

Motor and Gearbox Inspection section of this manual.

Verify that the right motor is connected to the control module. Complete the Motor and Gearbox Inspection section of this manual.

Verify that the left motor is connected to the control module. Complete the

Motor and Gearbox Inspection section of this manual.

Verify that the right motor is connected to the control module. Complete the Motor and Gearbox Inspection section of this manual.

Cycle power. Check all cable connections. If this does not correct the fault, contact Sunrise Technical Service for assistance.

Calibrate the joystick. If the error is still present, the hand control may be defective.

A latch function has exceeded it preset time.

N/A

N/A

N/A

N/A

N/A

N/A

N/A

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 5.1

Section 5

R-net Fault Codes cont.

Table 1, Error Codes cont.

Over-current

Overtemp. (Acts)

Overtemp. (Lamps)

DIME Error

Memory Error

This fault occurs when the limits of an actuator circuit is exceeded.

Perform Tilt Will Not Operate troubleshooting section of this manual to test the end of travel limit switches.

This error indicates that the Intelligent Seating Module (ISM) has become excessively warm. Note: An ISM is only required when there are more than 2 actuators and may not be mounted on this power chair. Allow the unit to cool. If the error repeats, the actuator may be defective or over loaded.

Note: This fault can only occur if an Intelligent Seating Module is used on this power chair and connected to external lights.

This error indicated that one or more of the modules are not compatible.

Remove the last module installed and retest. Recycle the power. Contact

Sunrise Technical Service for assistance.

Recycle the power. Verify that the cables are correctly mated. If a module has been replaced recently, that module may be defective. Contact

Sunrise Technical Service for assistance.

PM Memory Error

Bad Cable

Bad Settings

Module Error

System Error

SID Detached

Recycle the power. Verify that the cables are correctly mated. If a module has been replaced recently, that module may be defective. Contact

Sunrise Technical Service for assistance.

Inspect and replace the defective cable(s).

Verify that the programming agrees with the installed equipment. If all settings are correct, the control module may be defective. Contact

Sunrise Technical Service for assistance.

Recycle the power. Verify that the cables are correctly mated. If a module has been replaced recently, that module may be defective. Contact

Sunrise Technical Service for assistance.

Recycle the power. Verify that the cables are correctly mated. If a module has been replaced recently, that module may be defective. Contact

Sunrise Technical Service for assistance.

The Omni has detected that a specialty control has become disconnected.

Recheck all cables. If the error is still present, replace the specialty control.

User Switch Detached Indicates that a user switch has become disconnected. Reconnect the switch.

Gone to Sleep The predetermined sleep time has been exceeded due to inactivity by the user.

Charging This indication is present when the battery charger is connected. There may also be an error in the control module. Contact Sunrise Technical

Service for assistance.

PAGE 5.2

SEPT 2011 PULSE 6 SERVICE MANUAL SUNRISE MEDICAL

Section 5

R-net Fault Codes cont.

Note: On hand controls that contain LED battery gauges, they will display the fault code by illuminating various LEDs . Refer to the Corrective Action column for a list of items to check for each fault.

Bar Indication Corrective Action

The battery needs charging or there is a bad connection to the battery. Check the connections to the battery. If the connections are good, recharge the battery.

Verify that the left motor is connected to the control module. Complete the Motor and Gearbox Inspection section of this manual.

The left motor has a short circuit to a battery connection.

Contact Sunrise Medical Technical Service for assistance.

The right motor has a bad connection. Complete the Motor and Gearbox Inspection section of this manual.

The right motor has a short circuit to a battery connection.

Contact Sunrise Medical Technical Service for assistance.

The wheelchair is being prevented from driving by an external signal. Verify that the battery charger is not connected. Contact Sunrise Medical Technical Service for assistance.

Ensure that the joystick is centered upon power up. If it is centered, replace the hand control, and or cable.

A control system fault is indicated. Make sure that all connections are secure. Contact Sunrise Medical

Technical Service for assistance.

The parking brakes have a bad connection. Complete the

Motor and Gearbox Inspection section of this manual.

Verify that the battery charger is not defective. Leave the chair on for a few minutes to drain off the excess charge.

Check the condition of the battery charger.

A communication fault is indicated. Make sure that joystick cable is securely connected and not damaged.

An Actuator trip is indicated. If more than one actuator is fi tted, check which actuator is working correctly. Check the actuator wiring.

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 5.3

Section 5

R-net Fault Codes cont.

Example of R-10 Fault Isolation

1. In this example, one of the right motor being disconnected (fi gure.5.4.1). The hand control displays the fault and the power chair will not operate.

2. To troubleshoot this problem, refer to Table 1,

Error Codes.

3. From this error we see that the problem is being recorded in the PM (Control Module).

4. The hand control tells us that motor 2 (M2) has the error and it is with the brake circuit. Motor 2 is the right motor.

5. Table 1 informs us to, “Verify that the right motor is connected to the control module. Complete the

Motor and Gearbox Inspection section of this manual.”

6. The next step is to complete the Motor and

Gearbox Inspection section of this manual and retest. fig. 5.4.1

Power Chair Will Not Drive Full Speed

An external micro-switch is mounted next to the actuator in the rear of the tilt system. This micro-switch is open when the tilt is lowered. The opening of this micro-switch blocks a signal from being sent to the control module through the 6-pin connector located on the back of the tilt. The absence of this signal informs the control module that the tilt is less than 20 degrees and that maximum speed should be used. If the tilt is more than 20 degrees, the switches closes and invokes “creep” speed. Creep speed is a predetermined speed programmed at time of manufacture. Creep is indicated by a “turtle” on either the hand control or Omni.

1. “Creep” speed as indicated by the turtle on the hand control ("A" fi gure 5.4.2).

A

2. Tilt the seating system to the rear.

PAGE 5.4

SEPT 2011 PULSE 6 SERVICE MANUAL SUNRISE MEDICAL fig. 5.4.2

Section 5

R-net Fault Codes cont.

3. Locate the creep micro-switch mounted at the rear of the actuator "A" (fi gure 5.5.1).

Note: The creep micro-switch may be seen from the top of the seating system on the right rear of the actuator. It may be necessary to remove the seat pan to access the switch.

4. Manually operate the creep micro-switch while observing the hand control or Omni for the turtle symbol to go on and go off with the micro-switch operation. If the symbol changes, the problem is in the programming. Contact Sunrise Medical

Technical Service for assistance. If the symbol does not change, proceed to step 5.

5. Lower the tilt system.

6. Locate the 6-pin connector "B" at the rear of the tilt unit (fi gure 5.5.2).

7. Disconnect the connector leading to the control module.

8. Test the micro-switch wiring by performing the following test. a. Ensure that the seat is in the full down position.

b. Use Multimeter and measure continuity between the indicated pins below on the 6-pin connector mounted on the tilt. When down the indication should be open (fi gure 5.5.3). c. If it is closed, replace the micro-switch assembly. Refer to Tilt Actuator and Micro-Switch

Removal section of this manual. d. Retest as necessary.

9. If the above test passes, replace in this order: a. Cable leading from tilt to control module.

b. Control module

10. Retest as necessary

SUNRISE MEDICAL

B

A

fig. 5.5.1

fig. 5.5.2

Open fig. 5.5.3

PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 5.5

Section 5

R-net Fault Codes cont.

Power Chair Will Not Drive in Creep Speed When Tilted

An external micro-switch is mounted next to the actuator in the rear of the tilt system. This micro-switch is open when the tilt is lowered. The opening of this micro-switch blocks a signal from being sent to the control module through the 6-pin connector located on the back of the tilt. The absence of this signal informs the control module that the tilt is less than 20 degrees and that maximum speed should be used. If the tilt is more than 20 degrees, the switches closes and invokes “creep” speed. Creep speed is a predetermined speed programmed at time of manufacture. Creep is indicated by a “turtle” on either the hand control or Omni.

1. Tilt the seat all the way back and observe the hand control or Omni to see if the turtle symbol

"A" is displayed.

2. If the turtle is displayed (fi gure 5.6.1), contact

Sunrise Technical Service for assistance. The programming for the power chair is incorrectly set.

3. Use the hand control and tilt the seating system beyond 20 degrees.

4. Locate the 6-pin connector "B" on the rear of the tilt (fi gure 5.6.2).

5. Disconnect the connector leading to the control module

B

A

fig. 5.6.1

6. Test micro-switch with the following test. a. Ensure that the tilt is beyond 20 degrees tilt. b. Use Multimeter and measure continuity between the indicated pins below on the 6-pin connector mounted on the tilt. When seat is tilted the indication should be closed (fi gure 5.6.3). c. If it is open, replace the micro-switch assembly.

Refer to Tilt Actuator and Micro-Switch Removal section 7 of this manual. d. Retest as necessary. fig. 5.6.2

Closed

PAGE 5.6

SEPT 2011 PULSE 6 SERVICE MANUAL fig. 5.6.3

SUNRISE MEDICAL

Section 5

R-net Fault Codes cont.

Tilt Will Not Operate

The tilt actuator used in the tilt system receives power through a 6-pin connector located at the rear of the tilt See "B" (Figure 5.6.2). The actuator also contains micro-switches that open at the end of their stroke to prevent stressing of the tilt system. Around these micro switches are diodes that allow reverse power to fl ow to the actuator when reversing direction.

1. Select tilt on the hand control (fi gure. 5.7.1).

Operate the joystick and listen closely to the control module for a click. This click indicates that the control module is closing the power relay inside and supplying power to the tilt actuator.

If no click is heard, replace the control module.

Refer to Control Module Replacement section this manual. Retest as necessary. If click is heard proceed to step 3.

2. Verify that tilt is selected on the hand control.

Operate the joystick and listen closely to the tilt actuator. If the actuator appears to be running but the tilt is not moving, replace the tilt actuator.

Refer to Tilt Actuator and Micro-Switch Removal section of this manual. Retest as necessary.

3. Locate the 6-pin connector on the rear of the tilt

See "B" (fi gure 5.6.2 previous page).

4. Disconnect the 6-pin cable leading to the control module.

5. Select tilt on the hand control and place a rubber band around the joystick to hold it displaced

(fi gure 5.7.2).

SUNRISE MEDICAL fig. 5.7.1

fig. 5.7.2

PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 5.7

Section 5

R-net Fault Codes cont.

Warning: Use caution in the next step. Do not short the leads of the Multimeter together or damage may occur to the control module.

6. At the 6-pin connector on the cable leading to the control module, verify that battery voltage is present between the indicated pins (fi gure 5.8.1).

Note: The polarity of the voltage is not important since it is reverses when the opposite direction is selected.

7. If voltage is not present, replace the cable leading to the control module. If this does not solve the problem, replace the control module.

Refer to Control Module Removal section of this manual. Retest as necessary. fig. 5.8.1

8. At the 6 pin connector "B" (fi gure 5.8.2) on the rear of the tilt, check continuity between the indicated pins (fi g. 5.8.3).

B

fig. 5.8.2

1 to 3 ohms the actuator fig. 5.8.3

PAGE 5.8

SEPT 2011 PULSE 6 SERVICE MANUAL SUNRISE MEDICAL

Section 5

R-net Fault Codes cont.

Note: Since the tilt actuator will not operate, it is only possible to check for one condition.

Pick the condition below that best describes the position of the tilt system.

9. Example 1, the tilt is in mid-stroke. The value recorded is the resistance through the actuator motor windings (Figure 5.9.1). If this reading is not correct, replace the tilt actuator and retest.

Note: The value recorded in one direction is the forward resistance through a diode and will vary with the type of Multimeter used.

This value is not important, as long as there is continuity in one direction only. This value may even be high. The value recorded in the other direction is across the open contact of the end of stroke micro-switch and should be open.

10. Example 2, the tilt is in the full down position

(Figure 5.9.2). If the readings are not correct, replace the tilt actuator. Refer to Tilt Actuator and Micro-Switch Removal section this manual.

Retest as necessary.

Note: The value recorded in one direction is the forward resistance through a diode and will vary with the type of Multimeter used.

This value is not important, as long as there is continuity in one direction only. This value may even be high. The value recorded in the other direction is across the open contact of the end of stroke micro-switch and should be open.

11. Example 3, the tilt is in the full up position

(fi gure 5.9.3). If the readings are not correct, replace the tilt actuator. Refer to Tilt Actuator and Micro-Switch Removal section this manual.

Retest as necessary.

1 to 3 ohms fig. 5.9.1

Positive probe here

Negative probe here

Continuity between pins

Negative probe here

Positive probe here

Open between pins

*SEE NOTE* fig. 5.9.2

Positive probe here

Negative probe here

Open between pins

Negative probe here

Positive probe here

Continuity between pins

*SEE NOTE* fig. 5.9.3

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 5.9

Section 5

R-net Fault Codes cont.

Battery Testing

The Pulse power chair uses two (2) group-22 batteries connected in series. The circuit is protected by a circuit breaker located under the shroud near the front of the power chair and a non-serviceable fusible link in each battery harness. Prior to performing any test, the battery terminals should be clean and tight. Refer to Battery Removal section of this manual.

1. Check for initial battery voltage at the joystick using the battery charger connector. Connect the

Multimeter to the two outside pins (fi gure 5.10.1).

Pin 1 is positive (+) and pin 2 is negative (-). Fully charged batteries are typically 25.6 VDC for the pair, or 12.8 VDC each. Discharged batteries are

24 VDC or less for the pair.

Note: Voltage greater than 26 to 28 VDC indicates that the batteries are either overcharged, or the batteries have just been disconnected from the battery charger and contain a surface charge. If the batteries are not freshly charged and the voltage is excessive, inspect the battery charger for correct operation.

2. Load test the batteries using either a commercial load tester or the following procedure. To use the power chair’s motors as a load, complete the following steps: fig. 5.10.1

Warning: Failure to disengage the gearbox in the next step can cause unexpected operation of the power chair. a. Disengage the motor/gearbox so the motors can run in freewheel (fi gure. 5.10.2).

b. Connect the Multimeter to the joystick and record the voltage (fi gure. 5.10.3). If the voltage indicates a surface charge, run the motors for a few minutes to dissipate the surface charge.

Note: in this example, a small surface charge is present.

c. After the voltage has stabilized, record the voltage. Typical value for a fully charged pair of batteries is 25.6 VDC (12.8 VDC each). fig. 5.10.2

fig. 5.10.3

PAGE 5.10

SEPT 2011 PULSE 6 SERVICE MANUAL SUNRISE MEDICAL

Section 5

R-net Fault Codes cont.

d. Operate the power chair at high speed by operating the joystick in a high-speed profi le. The motors put approximately a 5 to 10 amp load on the batteries. e. Continue to operate the motors for 60 seconds.

(fi gure 5.11.1) Record the readings and compare them to following: i. During the fi rst 5 seconds, normal voltage drop is approximately .5 to .6 VDC.

ii. During the remaining portion of the 60 seconds, the voltage should stabilize, dropping no more than an additional .1 to .2 VDC. iii. At the end of the 60 seconds, release the joystick. f. At the end of a 5 minute rest, note the voltage recorded on the Multimeter. Normal indication at the end of a 5 minute rest is typically the starting voltage, or no more than .1 VDC less. fig. 5.11.1

Notes:

• A rapid drop of approximately 1.5 or more during the fi rst 5 seconds indicates a shorted cell in one or both of the batteries. The battery terminals should be checked to ensure that they are tight and clean, and then the test should be repeated. If the terminals are tight, the batteries should be replaced. Refer to Battery Removal section of this manual.

• A slow drop of more than .8 volts during the fi rst few seconds indicates a weakened pair of batteries. Evaluate the age of the batteries and the use they have received, and consider replacing them.

• If the voltage does not return to .1 VDC of the starting voltage at the end of the 5 minute test, evaluate the age of the batteries and the use they have received, and consider replacing them.

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 5.11

Section 5

R-net Fault Codes cont.

3. The individual voltage of each battery may also be checked using a Multimeter. Connect the

Multimeter to the battery connector (fi gure 5.12.1).

Note: Refer to Battery Removal (Section 7).

Compare the values to the following. a. 12.8 VDC indicates a fully charged battery. b. The battery readings should be similar. Newer batteries typically read within .1 to .2 VDC. c. 13 VDC or more indicates an overcharged battery, or one that has a surface charge from being recently removed from the charger. Verify correct operation of the battery charger if the batteries have not recently been removed from charge. d. 11.9 to 12 VDC indicates a discharged battery.

Recharge as necessary. e. Less than 11.9 VDC indicates an excessively discharged battery. f. Values larger than 13 VDC indicate either an overcharged battery or one that has just been removed from charge, as in the example (fi gure

5.12.1) fig. 5.12.1

PAGE 5.12

SEPT 2011 PULSE 6 SERVICE MANUAL SUNRISE MEDICAL

Section 6

Motor/Gearbox Inspection

Note: This procedure may be done with or without removing the motor from the power chair. If the motor/gearbox is to be removed from the power chair, refer to Motor and Gearbox Removal,

(Section7) of this manual. Consider replacing any motor/gearbox assembly that does not pass the following test.

1. Inspect the output shafts for signs of grease leakage (fi gure. 6.1.1). fig. 6.1.1

2. Inspect the cotter pin connecting the gearbox release arm to the motor/gearbox assembly.

Ensure that it is tight and not damaged (fi gure.

6.1.2). Chairs after serial number PLS-101619 and with s/n prefi x PLS6A & PLS6B, did not utilize this feature.

3. Manually operate each gearbox release lever to ensure that they operate easily and fully engage and disengage

Note: When removing the brush assemblies, remove one brush assembly at a time. Note its orientation so that it can be inserted into the brush holder the same way it was removed.

Caution: Use care not to damage the brush cap with excessive force in the next step.

4. Remove the brush by using a common screwdriver to unscrew the cap (fi gure. 6.1.3).

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL fig. 6.1.2

SEPT 2011 fig. 6.1.3

PAGE 6.1

Section 6

Motor/gear Box Inspection (cont.)

Note: When replacing brush assemblies, always replace both motors. Run-in the brush assemblies prior to delivering the power chair to the customer by releasing the gearbox assembly and operating the motor for 20 to

30 minutes, alternating between forward and reverse.

5. Inspect the brush assembly (fi gure. 6.2.1).

Replace any brush assembly less than ¼ inch long. Ensure that the lead to the brush and spring has not over heated and that the ends of the brush are sharp and clean.

6. Inspect the motor’s commutator (fi gure 6.2.2).

Verify that it is clean, not overheated, and does not show signs of excessive arcing. Normal color is bright copper to dull brown copper. Contact

Sunrise Medical Technical Service if in doubt of its condition.

7. Inspect the motor connector at both the motor end and the control module end for signs of over heating (fi gure. 6.2.3).

Note: Disconnect the motor connector from the control module if the motor/gearbox assembly is mounted in the power chair. fig. 6.2.1

fig. 6.2.2

fig. 6.2.3

PAGE 6.2

SEPT 2011 PULSE 6 SERVICE MANUAL SUNRISE MEDICAL

Section 6

Motor/gear Box Inspection (cont.)

8. Measure the resistance of the park brake assembly by connecting a Multimeter to the small pins of the 4-pin connector (fi gure. 6.3.1). Normal resistance is approximately 12 to 15 ohms.

9. Measure the resistance through the motor brush assembly and motor by connecting a Multimeter to the large pins of the 4-pin connector (fi gure.

6.3.2). Normal resistance is approximately .5 to

.9 ohms.

Note: New brush readings may be different than those recorded above. The new brush assemblies should be run-in prior to measuring their resistance. Contact Sunrise Technical

Service if the readings are either signifi cantly higher or lower than those recorded above.

fig. 6.3.1

fig. 6.3.2

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 6.3

Section 7

Removal Procedures

Battery Removal

1. Remove the two thumb screws under the front of the seat (fi gure. 7.1.1)

2. Inspect the power chair base at the back and disconnect any connectors that may be damaged before tilting the seat back (fi gure 7.1.2).

3. Flip back the arms to prevent damage to the arms as the seat back is tilted (fi gure 7.1.3).

fig. 7.1.1

fig. 7.1.2

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011 fig. 7.1.3

PAGE 7.1

Section 7

Removal Procedures (cont)

Caution: While not required, it may be a good idea to place a box under the seat back prior to tilting the seat.

4. Tilt the seat back while ensuring that none of the cables attached to the electronics are damaged (fi gure 7.2.1).

5. Locate and remove the 6 plastic inserts that secure the shroud to the base (fi gure 7.2.2).

Effective on chairs with s/n prefi x PLS only .

fig. 7.2.1

fig. 7.2.2

6. Locate and remove the 4 thumb screws that secure the shroud to the base (fi gure 7.2.3) effective on chairs with s/n prefi x PLS6A and

PLS6B

PAGE 7.2

SEPT 2011 PULSE 6 SERVICE MANUAL fig. 7.2.3

SUNRISE MEDICAL

Section 7

Removal Procedures (cont.)

7. Locate the two batteries mounted in the base

(fi gure 7.3.1) Note the battery connectors. fig. 7.3.1

8. Disconnect the battery connectors

(fi gure 7.3.2).

8. Remove the batteries.

9. Replace the batteries by reversing the

previous procedures.

Battery Installation

Batteries installed incorrectly can blow the fuse for this system. Pay careful attention to install the battery harness across both batteries instead of to each of the batteries individually.

SUNRISE MEDICAL fig. 7.3.2

PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 7.3

Section 7

Removal Procedures (cont.)

Motor and Gearbox Removal

1. Remove the batteries. Refer to the Battery

Removal in this section.

2. Locate and disengage the connector leading from the motor to the control module (fi gure..

7.4.1).

3. Elevate the wheelchair by placing wood blocks under the battery box (fi g. 7.4.2). fig. 7.4.1

fig. 7.4.2

4. Remove the wheel’s hubcap by prying it off using a common screw drive (fi gure. 7.4.3) .

fig. 7.4.3

5. Remove the wheel by removing the 4 bolts using a 6 mm Allen wrench (fi gure. 7.4.4).

PAGE 7.4

SEPT 2011 PULSE 6 SERVICE MANUAL fig. 7.4.4

SUNRISE MEDICAL

Section 7

Removal Procedures (cont.)

6. Use 17 mm Rachet and 17 mm open end to remove bolt ,locknut and washer from frame

(fi gure 7.5.1).

fig. 7.5.1

7. View of 17mm nylock nut (fi gure 7.5.2).

Slide Motor Assembly down and out of frame.

fig. 7.5.2

8. Effective for chairs prior to PLS-101619 only.

With the motor removed from the chair, use a 3mm or 1/8" pin punch to drive out the spring pin

(fi gure 7.5.3).

9. When re-assembling be sure the holes for the spring pin are aligned and drive in the spring pin with a hammer. Reverse steps 1-7 to complete re-assembly.

SUNRISE MEDICAL fig. 7.5.3

PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 7.5

Section 7

Motor Removal for chairs after S/N PLS-101919 and s/n prefix PLS6A &PLS6B

1. Remove 4 bolts from wheel and elevate frame as shown. (Figure 7.6.1) fig. 7.6.1

2. Using a 17mm socket and open end wrench, remove front suspension bolt and nut combination shown in (figure 7.6.2) fig. 7.6.2

3. Remove motor from chair (figure 7.6.3) fig. 7.6.3

4. Remove the 5 bolts from the motor plate assembly with a 5mm Hex Key (figure 7.6.4)

Note: Upon re-assembly, make sure to re-tighten the motor plate back to 40 inch/pounds of torque.

5. Final motor removal step requires disconnecting Motor connection from control module and cutting the Ty-wrap (Figure 7.6.5).

Both parts should be re-attached when motor is re-installed..

PAGE 7.6

SEPT 2011 PULSE 6 SERVICE MANUAL fig. 7.6.4

fig. 7.6.5

SUNRISE MEDICAL

Section 7

Removal Procedures (cont.)

Control Module Removal

1. Remove the rear cover by lifting the two quick release pins from the cover (fi gure. 7.7.1).

Note: The control module is mounted under the rear cover of the power chair

(fi gure 7.7.2). fig. 7.7.1

fig. 7.7.2

2. Carefully disconnect the connectors leading to the control module (fi gure 7.7.3) fig. 7.7.3

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 7.7

Section 7

Removal Procedures (cont.)

3. Loosen the control module by loosening two 3 mm Allen screws (fi gure 7.8.1).

Note: It may be necessary to remove the batteries to gain access to the nuts on the other side of the control module.

4. Slide the control module up to remove it (fi gure.

7.8.2).

5. Reverse the above procedures to install the control module fig. 7.8.1

fig. 7.8.2

PAGE 7.8

SEPT 2011 PULSE 6 SERVICE MANUAL SUNRISE MEDICAL

Section 7

Removal Procedures (cont.)

Shock Removal

NOTE: The shocks used in this power base are pre-adjusted at the factory. Contact Sunrise

Medical Technical Service if adjustment is required.

1. Remove the batteries. Refer to the Battery

Removal section.

2. Remove the motor/gearbox assembly from the side on which you are removing the shock.

Refer to Motor/Gearbox Removal section of this manual.

3. Use a 17 mm wrench and remove the top bolt securing the shock to the base (fi gure 7.9.1).

4. Use two 19 mm wrenches and remove the bottom nut and bolt securing the shock to the base (fi gures . 7.9.2 & 7.9.3). fig. 7.9.1

fig. 7.9.2

5. Lift the shock from the base.

6. Reverse the previous procedures to install the shock.

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL fig. 7.9.3

SEPT 2011 PAGE 7.9

Section 7

Removal Procedures (cont.)

Front Caster Arm Removal

1. Remove the batteries. Refer to the Battery

Removal section of this manual.

2. Remove the motor/gearbox assembly. Refer to the Motor/Gearbox Removal section of this manual.

3. Remove the two braces connecting the front caster arm to the rear caster arm using 19 mm wrenches (fi gures 7.10.1 & 7.10.2). fig. 7.10.1

fig. 7.10.2

4. Remove the coupler that connects to the two braces (fi gure 7.10.3).

PAGE 7.10

SEPT 2011 PULSE 6 SERVICE MANUAL fig. 7.10.3

SUNRISE MEDICAL

Section 7

Removal Procedures (cont.)

5. Push down on the rear of the motor mount assembly to allow clearance for the front caster arm assembly to slide out (fi gure 7.11.1).

6. Remove the front caster arm assembly.

(fi gure 7.11.2)

7. Reverse the previous procedures to install the front caster arm assembly. fig. 7.11.1

fig. 7.11.2

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 7.11

Section 7

Removal Procedures (cont.)

Rear Caster Arm Removal

1. Remove the batteries. Refer to the Battery

Removal section of this manual.

2. Remove the two braces connecting the front caster arm to the rear caster arm using 19 mm wrenches. (fi gure 7.12.1 & 7.12.2) fig. 7.12.1

3. Remove the rear caster arm assembly (fi gure

7.12.3).

4. Reverse the previous procedures to install the rear caster arm assembly fig. 7.12.2

PAGE 7.12

SEPT 2011 PULSE 6 SERVICE MANUAL fig. 7.12.3

SUNRISE MEDICAL

Section 7

Removal Procedures (cont.)

Front or Rear Caster Fork Removal

1. Elevate the caster fork to be removed by placing wood blocks under the battery box of the power chair (fi gure 7.13.1).

2. Remove the plastic cap on the caster arm assembly using a common screwdriver (fi g.

7.13.2). fig. 7.13.1

fig. 7.13.2

3. Hold the caster steady with one hand while removing the caster fork nut with a 19 mm socket wrench (fi g. 7.13.3).

4. Remove the caster fork from the caster arm assembly (fi g. 7.13.4)

Note: If either of the bearings needs to be replaced, it may be necessary to use a bearing puller on the lower bearing to separate it from the fork stem, and a brass drift to drive out the upper or lower bearings from the caster arm assembly if they are stuck.

5. Reverse the previous procedures to install the caster fork.

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL fig. 7.13.3

SEPT 2011 fig. 7.13.4

PAGE 7.13

Section 7

Removal Procedures (cont.)

Caster Removal

1. Elevate the caster fork to be removed by placing wood blocks under the battery box of the power chair (fi gure 7.14.1).

2. Remove the caster using two 13 mm wrenches (fi gure 7.14.2).

3. Reverse the previous procedures to install the caster (fi gure 7.14.3). fig. 7.14.1

fig. 7.14.2

fig. 7.14.3

PAGE 7.14

SEPT 2011 PULSE 6 SERVICE MANUAL SUNRISE MEDICAL

Section 7

Removal Procedures (cont.)

Tilt Actuator and Micro-Switch Removal

The following procedures describe the removal of the tilt actuator and/or the micro-switch assembly used for the creep speed. Read the full set of directions prior to performing the task. Some steps may be eliminated if both items do not need to be replaced. Reverse the procedures to install the tilt actuator and/or micro-switch and retest as necessary.

A

1. Disconnect the 6-pin connector "A" located on the rear of the tilt system (fi g. 7.38).

2. Remove the cover over the 6 pin connector on the rear of the tilt system using a 3 mm Allen wrench (fi gure 7.15.2).

fig. 7.15.1

fig. 7.15.2

3. Locate the micro-switch assembly and remove using a 3 mm Allen wrench (fi g. 7.15.3).

Note: this step may be eliminated if the micro- switch is to be disconnected and left in place.

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011 fig. 7.15.3

PAGE 7.15

Section 7

Removal Procedures (cont.)

Note: Complete step 4 only if the micro-switch assembly is to be replaced separately.

4. Reconnect the 6-pin connector on the rear of the tilt and partially tilt the seating system back to expose the actuator. Disconnect the 6-pin connector

5. Locate the tilt actuator under the seat

(fi gure 7.16.1).

6. Locate the micro-switch assembly at the rear of the tilt (fi gure 7.16.2). Note the connections leading to the micro-switch. Separate and label the wires leading to the micro-switch. Reverse the procedures to replace it and retest as necessary. fig. 7.16.1

fig. 7.16.2

Warning: The tilt system is free to fall during the next step. Secure the seating system prior to removing the tilt actuator.

7. Use two 13 mm wrenches and remove the nuts at both ends of the actuator (fi gure 7.16.3).

Remove both pins from the tilt actuator.

PAGE 7.16

SEPT 2011 PULSE 6 SERVICE MANUAL fig. 7.16.3

SUNRISE MEDICAL

Section 7

Removal Procedures (cont.)

Caution: the actuator and micro-switch assembly wires are fragile. Use caution when removing the actuator and micro-switch assembly.

8. Remove the actuator and micro-switch assembly as one part if the entire assembly is to be replaced

(fi gure 7.17.1). Reverse the procedures to install and retest as necessary. fig. 7.17.1

Note: Perform this step if only the actuator is to be replaced.

9. Note the connections of the wires and separate them (fi gure 7.17.2). Remove the actuator.

10. Reverse the previous procedure to install the actuator/micro-switch assembly. Retest as necessary. fig. 7.17.2

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 7.17

Section 8

Seat/Back Width adjustment

ASAP II Seating 16” - 20” wide is shown.

(NOTE: The same procedure is used for 18”- 22” wide seating system)

The ASAP II seating system is fully adjustable to allow changes from 16” to 20” or 18” – 22”. This requires a frame weldment and seat pan plates adjust for seat growth. The pictures shown in

(figure 8.1.1) shows a 20” wide seat width . This detail shows how easy it is to change back to

16” wide.

A

To make seating width adjustments follow the following steps:

1. Have 3mm – 4mm – and 5mm Allen Hex keys available for the adjustments.

2. Loosen (do not remove) the front M8 Button

Head Cap bolt "A" (figure 8.1.1). with a 4mm

Hex key, or a T-handle Hex key so that the plates can slide for adjustment.

3. Using a 3mm Hex key, remove the socket head cap screws (figure 8.1.2) that attach the left and right wings to the seat frame.

fig. 8.1.1

fig. 8.1.2

4. Remove the 5mm socket head bolts with a

Hex key that are used to hold bottom seat in place (figure 8.1.3) – totally remove out of its position.

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011 fig. 8.1.3

PAGE 8.1

Section 8

Seat/Back Width adjustment

5. Remove the nut, and bolt that secures the rear towel bar using a 4mm Hex Key and a 13 mm open end wrench or ratchet as shown in

(figure 8.2.1).

fig. 8.2.1

6. Loosen Back bracket on the chair to allow for changes to the seat width. (figure 8.2.2).

NOTE: it is recommended to use the back width to accommodate the user, for example – so 16 with 16 wide – 18 with 18” wide.

If, however there is a ½ “ adjustment that is needed, the brackets shown in (figure

8.2.2) show that a ½” change equals a 1” accomadation to the user .

fig. 8.2.2

7. Using a 5mm Hex Key and either a 13mm open end wrench or ratchet., Remove the nuts and bolts from the back bracket attached to the back cane assembly.

PAGE 8.2

SEPT 2011 PULSE 6 SERVICE MANUAL fig. 8.2.3

SUNRISE MEDICAL

Section 8

Seat/Back width adjustment

8. After checking alignment and symmetry of the final adjustments, and measuring (figure 8.3.1) the final adjustment change, the nuts and bolts can be re-tightened in the reverse order of the steps listed. Make sure all bolts are tightened to appropriate Specifications.

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 8.3

Section 8

Legrest adjustment

1. To perform a Legrest adjustment (figure

8.4.1) the following tools are required: A 5mm

Hex Key, and a 13mm open end wrench and/or ratchet.

2 Loosen all 3 sets of bolts/nuts as shown in

(figure 8.4.2) and (figure 8.4.3) using the 5mm hex key and 13mm open-end wrench or ratchet.

fig. 8.4.1

fig. 8.4.2

3. There are 5 sets of holes as seen from the front bars and 6 sets of holes as seen on on

(figure 8.4.4) where bolts can go through to give adjustability of up to 6 inches from top to bottom.

To extend the front arms further, simply loosen

5mm size SHCS screws and move the bars further forward, then retighten.

4. Once desired height is achieved, Reset placement of bolts then tighten with 5mm Hex key and 13mm wrench. Tighten to 40 in/lbs of torque.

PAGE 8.4

SEPT 2011 PULSE 6 SERVICE MANUAL fig. 8.4.3

fig. 8.4.4

SUNRISE MEDICAL

Section 8

Back-rest Angle Adjustment

1 Make sure you have the right tools- 5mm

Hex Key, and 13mm open end wrench and/or ratchet.

2. Loosen the 2 BHCS bolts "A" (figure 8.5.1) with a 5mm Hex key and use 13mm open end

Wrench for the nuts "B".

B

A

Note:

Take care not to lose the saddles, or plastic washers between Bracket and seat rail while performing the next step.

3. To make a small angle change to one of four possible settings, (-4°, 0, 4°, 8°) pull bolt out and change the position of the front edge bracket.

(figure 8.5.2) Then re-attach the nut and tighten bolt into place. make sure plastic saddles and washers are in place behind the part.

fig. 8.5.1

fig. 8.5.2

-4°

SUNRISE MEDICAL PULSE 6 SERVICE MANUAL SEPT 2011 PAGE 8.5

Section 8

Powered elevating or articulating legrest

VR2/R-Net Control

The actuator buttons can operate any factory approved actuator. Operation is dependent on what options are fitted to your wheelchair. To operate the legrests, seat tilt, seat lift or back rest recline follow the following steps

1. Push one of the actuator buttons that has the seat icon on it to select actuator mode

2 Operate the joystick left or right to select the actuator required (actuator 1 or actuator 2).

Selection is indicated via the lighting of the red

LED adjacent to the desired actuator button.

3. Operate the joystick in the forward or rearward direction to move the leg rest/seat lift/ tilt or recline up or down.

(Note: R-Net Only)

4. Release the joystick when the desired angle is reached.

5. To return to drive mode press the actuator button again or Mode for R-Net.

The Multi Axis box

– (figure 8.6.1) has 3 settings available – with Toggling occurring in each of these settings. On this example below hitting the blue will cause combined legrest to extend – hitting yellow button will extend left mounted actuator and orange for right mounted actuator. Simply touch the switch for Right Leg and the leg will either extend outward or retract inward – once fully extended or retracted it is recommended to let go of switch.

(figure 8.6.2) shows all connectors to actuator are 6 pin molex connectors. The lead going back for power connects to the control moduole as shown below via OBCport (figure 8.6.3)

The test for power can be done quite simply by removing connector from actuator and hooking up Multi meter to pins shown. (figure 8.6.4) connection prior to depression of switch.

(figure 8.6.5) shows depression switch and what meter will read

PAGE 8.6

SEPT 2011 PULSE 6 SERVICE MANUAL fig. 8.6.1

fig. 8.6.2

fig. 8.6.3

fig. 8.6.4

fig. 8.6.5

SUNRISE MEDICAL

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Questions & Answers

2024-05-05

M M F

What causes JSM Joystick Error?
Possible joystick trip or issue with the joystick and its connections.
How to troubleshoot Joystick Error 0E00?
Check the joystick, connections, and power components. If issue persists, contact Sunrise Medical for technical assistance.
2024-05-05

M M F

What does the 'JSM Joystick Error' indicate?
The 'JSM Joystick Error' usually shows a possible issue with the joystick, causing an error code to appear on the controller display.
What are the possible solutions for the 'Joystick Error 0E00'?
For the 'Joystick Error 0E00,' perform diagnostic checks on the joystick and its connections, making sure that the joystick, cable, and control module are all properly installed and connected.