PC Board Calibration Procedure
Equipment Alert
Anytime the PC board is removed you must perform the Calibration Procedure.
Failure to do so will cause the chair to malfunction.
If the chair has a retractable foot section, it must be in the stowed position before performing the Calibration Procedure.
Step 3: Calibrate PC board.
Wait five seconds, then press S2 button.
Note: During calibration, you will hear a steady pattern of “beeps” as
each function moves. When successful calibration is complete,
the chair will stop moving, and you will hear three “beeps”.
S2 Button
1
2
A) Move MODE SELECT switch 1 to ON (up).
B) Move MODE SELECT switch 2 to OFF (down).
N
O
Step 1: Put PC board into Calibration Mode.
2
J1
O N
1
2
MODE SELECT
Step 2: Reset PC board.
A) Unplug chair for a few seconds.
B) Plug chair back in.
Step 4: Return to normal operating mode.
A) Move MODE SELECT switches back to original position.
B) Unplug chair for a few seconds (to reset PC board).
C) Plug chair back in.
1
003-1490-00
Rev.B