advertisement
Variants:
3741321009
3741321369
3757021009
3767021004
3767021009
3767021094
3767021099
3767021159
3767021249
Service Manual
for seca 374 / 375 / 376 CN
Service Manual Number:
17-05-01-329-b
Valid as of: 01.03.2008
Description: electronic babyscale multi load cell design, DMS low end module, LCD, for battery use, prepared for mains operation.
Content:
Operation mod. 374
Operation mod. 375/376
Operation mod. 376 Japan
Description DMS module
Description of faults
Function diagram
Calibration (lin.)
PC configuration program
Replacement
Spare parts
17-10-06-301 a
17-10-07-496 c
17-10-07-513 c
30-34-00-645 b
30-34-00-588 c
25-01-02-560 a
30-34-00-603
30-34-00-672 b
30-34-00-726
30-34-00-727 b
Manual number: 17-05-01-329-b
Service manual Description of the electronics
Brief description of the DMS module
Introduction
The DMS modules are used to determine weights using a load cell with strain gauges. The DMS low end module also accommodates the display and the operating unit. To allow the modules to be adapted to different scales, an EEPROM is integrated which is used to configure the module. A connection to the SeSAM bus (Seca’s Serial Autoconfiguring Multicontroller bus) is provided via which the scale can be calibrated and additional modules can be connected.
Display
Load cell
A/D Converter
Operation
Voltage
control
Temperature measurement
µC
EEPROM
Quartz
SeSAM Bus
Power On
Reset
Voltage measurement
Voltage control
Mains connection
Battery/rechargeable batteries
Weight measurement
The DMS modules use load cells with strain gauges in a double bridge circuit as sensors.
The entire analog part runs on its own supply voltage. This voltage can be switched by the µC to 3V,
5V or 10V. This voltage is also supplied to the load cells. At full load, they deliver a bridge output voltage of 2mV/V.
This voltage is measured by a dual slope a/d converter. The DMS ext module can evaluate up to four load cells which are multiplexed accordingly. Only one load cell can be connected to the other types of
DMS modules.
The conversion time is selected so that a new measured value is available every 0.1 s.
The resolution of the a/d converter is 0.05 µV/V. If several load cells are connected, keep in mind that the resolution relates to the mean output signal of all the load cells connected.
When these measured values are within the expected range, they are converted into a weight value using a third order polynomial. In order to increase accuracy, eight measured values are used to calculate the weight to be displayed.
The first value obtained after switch-on is used as the zero point. If the values obtained later on do not deviate strongly from this value, the zero point is corrected in order to compensate for possible drifts.
(Please also refer to temperature measurement.)
05.08.03 /Jensen / Rei 1 34-34-00-645 b
Service manual Description of the electronics
Display (only DMS low-end module and SLC highend)
Whenever eight new measured values are available, i.e. approx. every 0.8 s, the weight is displayed.
The display is an LCD which is controlled by duplex operation. A maximum of five digits as well as a few special characters can be displayed.
Explanation of the special characters:
•
The battery symbol is used to alert of the fact that the battery is going flat.
•
Hold indicates that the weight display has frozen.
•
Tare or Net indicate that a weight on the scale has been tared off.
•
The left arrow indicates if additional explanations are given next to the display.
•
The weight units on the right show the unit in which the measured weight is displayed.
Temperature measurement
In order to compensate for the load cell’s temperature drift, the temperature is measured regularly and the measured value is adjusted accordingly.
To perform this measurement, a capacitor is charged every 2 seconds alternately via a reference resistor and an NTC with a linearization resistor connected in parallel. The time until the capacitor has reached a specific voltage is measured. The temperature is calculated from the ratio of the two times.
To reduce measured value fluctuations, a mean value of 16 temperature values is always formed. The fluctuation then amounts to approx
±
0.2°C.
The difference between the current temperature and the temperature measured when the last zero point was obtained is used to determine a projected zero point. To do so, one assumes that the weight change per °C is constant. The weight value is then corrected in accordance with the zero point calculated and in accordance with the difference to the temperature measured when the scale was calibrated. A linear and a square term are used for this purpose.
Power supply
Power supply is provided by AA size batteries (batteries or rechargeable batteries) and a plug-in power supply unit. The batteries pass the voltage via reverse voltage protection diodes to a 3V in-phase regulator. A coil which suppresses interference and a buffer capacitor come next. This 3 V supply voltage is supplied to the digital part of the circuit.
Another in-phase regulator generates the voltage for the analog part. The µC can set this regulator to
3V, 5V or 10V.
In order to operate the module on rechargeable batteries a recharging circuit can be connected. This circuit charges the rechargeable batteries at a constant current when a power supply unit is connected.
When the supply voltage is applied, a power-on reset is initiated at the µC via an RC element.
To monitor the condition of the batteries, their voltage is measured every 2s. For this purpose, a 2.2 V reference voltage is integrated upwards and the time until the voltage has reached approx. 9% of the supply voltage is measured. Afterwards, a downwards integration is made from 2.2V to the threshold.
From these two results, a measured value for the voltage is calculated. If the voltage drops below a certain limit, the user is alerted by the battery symbol in the LCD. If the next threshold is undershot,
„bAtt“ is displayed and weighing is no longer possible.
Operation (only DMS low-end module and SLC highend, with the exception of vibration-sensitive switch and platform switch)
Four buttons are provided; one is used to switch the module on and off, the others can be programmed as required. These three buttons can be pressed just lightly or longer (longer than 1.5s), so that six events are available. The actions tare, hold and unit switch-over can be assigned to them.
If you press another button when switching on the scale, the scale will switch to the calibration counter and recalibration mode.
Instead of the normal start button a piezo diaphragm can be soldered to the PI connection. If the piezo diaphragm registers a vibration, a transistor switches and generates an interrupt signal which causes the µC to start.
05.08.03 /Jensen / Rei 2 34-34-00-645 b
Service manual Description of the electronics
If the start button is directly connected to this interrupt input, a button is obtained which is only active when the scale is switched off. It can be fitted under the platform, for instance, so that the scale is started by briefly stepping on it.
Programming
The DMS modules are equipped with an EEPROM which is controlled via the I²C bus protocol. On this
EEPROM all parameters are stored that can be used to influence the program sequence. For example the way the weight is displayed or the allocation of buttons. It also stores the coefficients for determining the weight and data for temperature compensation, e.g. the factors for the polynomial.
Furthermore, the scale can store values to be retrieved when the scale is switched on the next time, e.g. the weight unit used last.
In order to avoid errors, the EEPROM contents are divided into blocks which are monitored by cyclic redundancy checks. The hamming distance is 4 so that single errors can be corrected and double errors are certainly detected. In addition, values which are recalculated when the scale is in operation can be replaced by default values if an error is detected that cannot be remedied.
Serial bus
To allow the scale to be calibrated and additional modules, such as additional displays, to be connected, the DMS modules are equipped with an interface to the SeSAM bus.
On this bus the stabilized and non-stabilized supply voltages are available. Via a data and a clock pulse line all relevant data is transmitted to other modules. A special start line is provided via which the scale can be switched on or other modules can signal that they want to send data.
Both data and start line are open collectors so that they can be switched to low by all modules.
For further details on the bus, please refer to the relevant documentation.
Fault handling
If faults occur, the SeSAM bus initiates a fault message with the relevant fault code. Furthermore, each fault message includes the number of the module type on which the fault has occurred and a code which indicates how long the message is to stay on display and which displays are to be suppressed.
The DMS low end module displays its own fault messages on its LCD. Module type (0 for the low-end module) and fault number are displayed. The other DMS modules only transfer fault messages to the bus (module type = 1).
The following fault messages are implemented:
10 20 bit counter overflow
11 Counts out off range
12 Switch-on zero setting out off range
13 Zero tracking out off range
14 Error in the kg value calculation
15 Recalibration error bAtt Battery voltage too low
21 Error in battery voltage measurement
22 Error in temperature measurement tEMP Temperature range exceeded ( < 0°C or > 50°C)
30 General bus error
32 Command buffer full
33 More than 8 slaves detected
40 EEPROM error
41 Incorrect EEPROM access
50 Rom checksum error
51 Menu error
CArd Invalid Chipcard
OFF Chipcard has scale not released
6x Error in measurement of lenght
Notice : Chipcard is saving now
05.08.03 /Jensen / Rei 3 34-34-00-645 b
Service manual Description of the electronics
Time base
From the quartz’s 4.19MHz a time base is derived which controls all timed processes in the scale.
These are, for instance, various timeouts and the on-time.
After expiry of the on-time, the supply voltage for the analog part, for voltage measurement and for the
EEPROM is switched off. The LCD is switched off, all ports are set to zero-signal level and the quartz is stopped.
Recalibration
In order to compensate for linear measuring errors of the scale, which can, for instance, occur as a result of the different forces of gravity at different geographical points, the module has a recalibration function. To initiate the function, another button must be pressed when the scale is started. The calibration counter is now displayed which is incremented whenever the scale is recalibrated. It enables the user to check whether the scale was manipulated. If, within five seconds, the button is pressed longer than 1.5 s, the scale switches to recalibration mode. The scale now measures the zero point and waits to be loaded with at least ¼ of the maximum load. By briefly pressing the button the weight displayed can be decremented or incremented in steps of 10g until the weight of the load placed on the scale is reached. The direction can be changed by pressing the button longer. To stop recalibration the button must be pressed longer than 5s. The scale can be switched off any time using the on/off button or by not activating it for a while, thus aborting recalibration.
Optional functions (only DMS low-end module and SLC highend)
The functions “Tare“, “Hold“, “Autohold“ and “Replace weight in repeat measurements” can be implemented via a program on the EEPROM. The functions are explained below:
Tare:
The tare function can be assigned to a button by the EEPROM program. If the button is pressed, the weight is tared off as soon as the display has settled, if it is within the permissible range. Pressing the button again alternatively switches the tare function off or causes the weight to be tared off again.
Hold and Autohold:
If the hold function is assigned to a button by means of the EEPROM, the weight reading is held on display when the button is pressed as soon as the value has stabilized. Alternatively, the hold function can be automated so that the weight will always be frozen when it has exceeded a certain threshold and is stable.
The hold function is switched off after expiry of a predetermined time or by pressing the button again.
In addition the scale can be configured so that the hold function will be deactivated when the scale is relieved, when the scale is relieved and then loaded again and / or when the load on the scale changes significantly.
Replace weight in repeat measurements:
The scale can be configured so that the weight last displayed is displayed again in a measurement when the scale detects that it has been loaded with almost the same weight.
Calibration
To ensure that the module functions properly, data must be written to the EEPROM before the module is calibrated. Values that are only determined during the calibration process can be set to 0. The correct figures will then be set after calibration.
If the weight is to be calculated using a third order polynomial, the scale must be calibrated using at least four test loads. In order to increase accuracy, it is recommended to use even more weights.
If you wish to configure individual temperature compensation, the scale must be measured with a load and without a load at temperatures on different ends of the range. In order to determine the square term for slope compensation, the scale must also be measured at an average temperature.
To determine the required calibration values, please refer to the relevant documentation for EEPROM programming.
05.08.03 /Jensen / Rei 4 34-34-00-645 b
Service manual
Technical data
Supply voltage:
Supply current:
Load cell voltage + 0.5 V to 15 V approximate load cell current
Zero-signal current: typ. 2µA
Operating temperature: approx. 0°C to 50°C
Storage temperature: -10°C to 60°C
Description of the electronics
05.08.03 /Jensen / Rei 5 34-34-00-645 b
Service manual Description of faults
Description of the fault
Scales cannot be started
Possible cause
•
Battery is not installed or discharged
•
Voltage supply is defective
•
Bus connection cable is defective
•
Keypad is defective or not connected
•
Bus fault
•
Check the batteries
Remedy
•
Check the voltage at the appropriate connections of the electronics; eliminate the interruption
•
Check the plug-in connections; check the cables for interruption and replace them if necessary.
•
Check the cable connections; check the cables and the keypad for interruption and replace them if necessary
•
Disconnect power supply unit and battery from the scale, wait for
60 seconds, then reconnect.
•
Check the scales Scales show no weight or an incorrect weight
(repeatedly)
Scale shows only the segment test or the readings are varying to much.
Incomplete digits
•
Packaging / transport safeguard has not been fully removed
•
Scales are maladjusted
•
Sensor is damaged
•
Load lever system is damaged or transfer of force to the sensor is interrupted
•
Shortcircuit between sensor cable and housing
•
Display is defective
•
Display: Malfunction during the weighing procedure
Er:x:no
•
Er:x:10
•
Sensor is defective or not properly connected
•
Recalibrate the scales
•
Replace the sensor and recalibrate the scales
•
Check the base frame (if present, the tie rod) for damage or incorrect assembly
•
Search for damaged isolation of the cabel and repair it. ( If appropriated use multimeter.)
•
Replace the electronics
•
The electronics have detected an error.
!
X : module number
!
No: error number
•
Check and, if necessary, replace the sensor (if the sensor is replaced, a recalibration is required)
20.02.01/Winterberg 1 (2) 30-34-00-588 index c
Service manual Description of faults
Description of the fault
Possible cause
•
Er:x:11
•
Sensor is defective or supply lines are damaged
•
Er:x:12
•
Initial load is too high
Remedy
•
Check connection cables, check sensor and replace as required
(if parts a replaced, a recalibration is necessary)
• If several load cell are used, check the allocation (colour codes)
•
Er:X:14
•
Error during calculation of the kg value
•
Er:x:15
•
Error during recalibration
•
Er:x:21
•
Error while measuring the battery voltage
•
Er:x:22
•
Error while measuring the temperature
•
Check the initial load
•
Check the load lever (stuck or bent; check the sensor, replace it if necessary and carry out a recalibration
•
Incorrect coefficient in the read-only memory of the module; recalibrate the scales
•
Reference weight is outside the permissible range or operator error
•
Battery voltage is far too low (< 2.5 V) or electronics are defective
-> Replace the electronics
•
Replace the electronics
•
Er:x:32
•
Error: command buffer overflow
•
Er:x:40
•
No EEProm data
•
Er:x:41
•
Incorrect
EEProm access by the slave
•
Check the cabling of the bus, eliminate any EMC malfunction, briefly (60 sec.) de-energize the electronics
•
Recalibrate the scale or write the correct EEProm data by service program
•
Check the slave; disconnect or replace the slave
•
Er:x:50
•
Other errors
•
Error occurs directly after switching on -> wrong checksum in the
ROM
•
Er:x:51
•
Menu error
-> Replace the electronics
•
Incorrect use of a menu function (wrong data e.g.), try it again
20.02.01/Winterberg 1 (2) 30-34-00-588 index c
Vogel & Halke
Created on: 30.06.1998
Holger Panier / SE
File: 00603_e1.doc
Printed on: 07.07.1999
seca
Description of the calibration mode for scales with modular electronics
General:
Seca scales with modular electronics are fitted with a software-controlled calibration device that is operated using the existing controls. This device can be used to recalibrate the scales or to set the scales to different GAL values. During development of the calibration device, special attention was paid to the following requirements:
•
It must be possible to recalibrate the scales without any additional external equipment/tools.
•
It must be possible to calibrate the calibration device.
•
It must be possible to recalibrate the scales without calibrated test weights.
•
The calibration device shall be protected against inadvertent use.
The scales fitted with the new, modular electronics are equipped with a calibration counter allowing a software recalibration to be carried out according to the requirements for calibrated scales. To carry out such a recalibration, the calibration counter setting can be called up on the display of the scales and compared to the data given on the calibration label mark (applies only for calibrated products). If the values are identical, the scales are calibrated; whereas differing values match the situation of a broken calibration seal. Each completed recalibration procedure is registered by the calibration counter, i.e. the number is automatically incremented by 1.
Before the scales can be recalibrated, they must be switched to the calibration mode:
To do so, switch on the scales using the appropriate control. While doing so, simultaneously press another control key (for example, the kg/lbs key). This causes the contents of the calibration counter (i.e. the number of calibration procedures carried out so far) to be displayed for five seconds (flashing).
During the 5 seconds the contents is displayed, press the kg/lbs key again and keep it depressed (for more than 1.5 seconds) to switch the scales to the calibration mode.
If the kg/lbs key is not pressed within these 5 seconds, the scales are automatically switched back to normal weighing operation.
When switched to the calibration mode, the scales first are in a decrementing state. This state is indicated by a flashing "dec" that alternates with the currently measured weight.
Now place a test weight onto the scales. This test weight should have a weight of at least
25% of the maximum permissible load for the weight (we recommend a weight of approx.
66-75%). The weight shown on the display must match the test weight; otherwise, the scales are maladjusted and the recalibration procedure must carried out in full length. To do so, change the weight shown on the display using the kg/lbs key until it matches the test weight.
If the value shown on the display differs from the test weight (if not, recalibration is not required), estimate whether the required nominal weight is easier to reach by decrementing or by incrementing. To toggle from the decrementing mode to the incrementing mode or vice versa, press the kg/lbs key for more than 1.5 seconds (display indicates dec or inc).
Page -1-
00603_e1.doc
Vogel & Halke
Created on: 30.06.1998
Holger Panier / SE
File: 00603_e1.doc
Printed on: 07.07.1999
seca
Pressing the kg/lbs key for less than 1.5 seconds causes the weight shown on the display to be increased or decreased by 10 g (depending on the mode currently selected). As the resolution of the display is lower, the displayed value will not necessarily be increased or decreased each time the kg/lbs key is pressed. Repeat the above procedure until the displayed weight and the test weight match. Set the scales to the middle between two division values by additionally incrementing or decrementing the value by n=d/(2*10). The calibration process can now be completed by pressing the kg/lbs key for at least five seconds. The scales compute and store a new linearity coefficient and will then be switched off automatically. The scales are now calibrated. Remove the test weight.
Note
:
To prevent improper calibration by improper use, a timeout control has been implemented in the calibration mode, i.e. the time that passes until the next key is pressed must never exceed five seconds.
However, this timeout control is only active for the phase when switching to the calibration mode.
If, during this phase, 5 seconds have passed and no key has been pressed, the scales are automatically restored to the weighing mode so that the procedure must be repeated.
To reduce calibration errors to a minimum, recalibration is only possible using a test weight of more than 25% and less than 100% of the max. load; otherwise, “Er:0:15” will appear on the display.
If the scales need to be set to a different GAL value, the displayed value must be matched to the test weight value adapted to the GAL value correction.
Example of a recalibration procedure:
The selected test weight is 100 kg.
The current weight shown on the display is 99.7 kg (max. difference of 300 g).
In this example, you must switch the scales to the incrementing mode to carry out the recalibration procedure. After the kg/lbs key has been pressed (for more than 1.5 seconds), a flashing "inc" is shown.
Now increase the value by the first 10 g (by briefly pressing the kg/lbs key). After this has been done thirty times, the value 100.0 (rounded!) appears on the display although the internal measured value still is 99.95 kg (not rounded!).
To correct this rounding error, increase the value by another 10 g to get exactly 100.00 as the calculated value.
Terminate the recalibration procedure as described above.
Page -2-
00603_e1.doc
Vogel & Halke
Created on: 30.06.1998
Holger Panier / SE
File: 00603_e1.doc
Printed on: 07.07.1999
seca
Action carried out by the user
The scales are switched off.
Press the kg/lbs key and keep it depressed while simultaneously pressing the start key.
Release both keys.
Press the kg/lbs key for more than
1.5 seconds within the next 5 seconds.
5 seconds have passed and no key has been pressed.
Place the test weight onto the scales.
Result
No values or characters on the display.
The contents of the calibration counter is shown on the display for 5 seconds
(flashing).
The contents of the calibration counter is shown on the display for 5 seconds
(flashing).
The system switches to the calibration mode: first "dec" appears on the display (flashing) and then the current weight (0.0 - also flashing).
The scales are switched back to normal weighing operation.
The measured weight of the test weight appears on the display (flashing).
Decide whether to decrement or to increment:
Incrementing:
Press the kg/lbs key for more than 1.5
seconds.
The incrementing mode is activated; "inc" appears on the display.
Decrementing:
No action required as the scales are already in the dec mode.
Decrementing or incrementing:
Again briefly press the kg/lbs key (for less than 1.5 seconds).
Repeat this step until the measured weight matches the test weight (format: xx.0 kg).
To correct this rounding error, decrement or increment the value five times by another
10 g.
No output on the display.
The measured weight is decreased or increased by 10 g. Please remember that pressing the key does not always change the value on the display due to the low display resolution. Only the rounded value is given.
The display indicates the test weight (format: xx.0 kg).
The display output does not change the first four times.
After the fifth time the value has been decremented or incremented, the measured weight appears; it now matches the test weight.
Press the kg/lbs key for at least 5 seconds until the system is switched off.
The display indicates the exact test weight
(format: xx.00 kg).
The display does not show any characters or values any more.
Page -3-
00603_e1.doc
seca page 1 (2) PC configuration program
Instructions for PC-based configuration
PC-based configuration is used for diagnosis or recalibration of all seca scales with modular electronics.
The following equipment is required:
•
PC with serial interface (RS -232)
• seca software 'SERVA”
•
RS232 service set
10-02-01-004
68-60-00-002, consisting of:
•
RS-232-service module
•
Connecting cable PC/service module
•
Modular cable
68-60-00-003
08-06-16-121
•
Bus cable for DETF-LEM
•
Loading weights
08-06-16-115
08-06-16-125
•
For scales with DETF sensor at least 5 weight stages up to the MAX
•
For scales with DMS load cell at least 1 weight stage (=MAX)
Software Installation
If your 'SERVA'-software is in a ZIP-file, it has to be unpacked at first.
Execute the program's SETUP.EXE and follow the instructions displayed by the setup program.
Connecting and calibrating a scale
•
If the service module has not been connected to the PC yet, shut the PC down, connect the service module to the serial interface and reboot the PC.
Vorderseite / Frontside Rückseite / Backside
Waage \ Scale
Drucker
Printer
Anzeige / Display
PC
•
Switch off the scale!
•
For calibration, a connection between the service module and the measured-value acquisition module of the scale to be calibrated must be established. For this purpose, a modular cable must be connected to the scale’s SeSAM bus.
17.06.03- Wint/Rei 30-34-00-672 b
seca
Standard case
page 2 (2) PC configuration program
Connection to the existing display unit port.
DMS module with integrated display
08-06-18-082
The display can instead be connected to the second modular socket of the service module.
DETF module with integrated display
08-06-18-051
08-06-18-073
If there is no modular socket on the back of the PCB, retrofit one.
Pick off the signal directly from the PCB via the DETF LEM bus cable.
•
Once the cables have been correctly connected, start the scale.
•
Start the software 'SERVA' from the programs menu.
•
The 'Help'-menu of the 'SERVA'-program contains detailed instructions for available operations
17.06.03- Wint/Rei 30-34-00-672 b
Service Manual Replacement instructions
Mod. 374/375/376
Removing the power supply
Before starting any work, unplug the mains plug and remove the batteries!
ESD protection measures must be taken whenever work is performed on electronic components.
1. Opening the scale (page 2)
1.2 Open the battery compartment cover (B) and unscrew the screw (C). Then push the electronics compartment cover (D) slightly to the right and remove it.
1.3 When replacing load cells (WZ) or the mains connection (NB) also: turn the scale round and place it on a soft mat. Remove the 4 screws (G) and take off the frame
(H).
1.4 The adjusting feet (NF) can be unscrewed and replaced by new ones without disassembling the scale.
2.1 Electronic module (EM) (page 2)
After slackening the 7 screws (I), you can take out the PCB. The switching mats (J) and the viewing glass (K) are now accessible. To remove the viewing glass, carefully release the 4 catches along the circumference.
To replace the electronic module (EM), unsolder the cable connections before slackening the screws.
2.2 Load cells (WZ) (Pages 2, 3)
When replacing a load cell (WZ), unsolder the 4 associated litz wires (L) from the electronic module and detach the load cell from the scale by removing the 3 screws (F). It is easier to fit the new load cell if you use the connecting cable of the old load cell to pull in the connecting cable of the new load cell or an auxiliary cable.
2.3 Mains connection (pages 2, 3).
Unsolder the 2 associated litz wires (M) (page 3) from the electronic module (EM) and detach the mains socket (NB) (page 2) from the frame by unscrewing the screws (P).
2.4 Bubble level (page 3)
Warning: Only remove the bubble level if it is defective, as destruction of the bubble level cannot be avoided. A new bubble level can only be fitted by a skilled person working with great care!
Lever the bubble level (O) out. Scrape the holder free and clean it.
3. Closing the scale
3.1 Reassemble the scale in reverse order, following steps 2.3 to 1.1.
Important note !
When the electronic module, the load cell or the bubble level are replaced, the scale must be readjusted.
Follow instructions no. 30-34-00-603.
3.2 Apply for official calibration of the scale.
16.02.06/Reinhold 30-34-00-726
Dienstag, 21. Februar 2006 12:47:54
Service Manual Spare parts list
seca 374 / 375 / 376
Item. Articel no. Designation
01 02-01-01-267-009
Price stage
30
02 02-03-01-070-509 when order, please clearly indicate the print of puh buttons
40
03 01-22-13-319-009
04 glas 15
01-03-01-224-009 Support for load cell, mod. 375, 376
15
32
05
10
08 01-17-01-204-009 Bubble level 14 mm, mod. 375, 376
15
01-18-01-265-203 Button HOLD China
05
05
05
05
Button 05
Button
01-18-01-266-204 05
MB. Mod. 376, China
05
11 01-18-01-267-201 Button START
01-18-01-267-202
05
05
12 02-03-01-264-009 Top cover electronic
15 batt. compartment
15
10
16
17
42
14-05-01-971-009 Sticker for Display technical data Mod.374 01
14-05-01-971-601 Sticker for Display technical data Mod.374 Russia 01
14-05-01-971-602 Sticker for Display technical data Mod.375
14-05-01-971-603 Sticker for Display technical data Mod.376 Max. 15kg
14-05-01-971-604 Sticker for Display technical data Mod.376 Max. 20kg
01
01
01
08-06-16-081-119 Harness Battery connection 15
20
10 calibration count sticker
mod. 375/376
01
14-05-01-870-009 Foil for calibration count sticker
Mod. 375/376
01
05.02.08/Rei 30-34-00-727b
Dienstag, 5. Februar 2008 11:54:22
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project