PC 700 - REV 3 | User manual | POL-EKO-APARATURA TOP+ Laboratory Thermostatic Chamber Catalogue 90 Pages
PC 700 - REV 3 | User manual | POL-EKO-APARATURA TOP+ Laboratory Thermostatic Chamber Catalogue
The POL-EKO-APARATURA TOP+ is a professional range of ovens, incubators, cooled incubators and climatic chambers. The TOP+ version features a full colour 5.7’’ touch screen and an Ethernet network connection that enables the user to program the equipment virtually from any place in the world. The unit is equipped with a full colour, clear 5.7'' LCD touch screen, easy-to-use intuitive software, USB port for data transfer and a 7-day programming mode. It also features temperature (and humidity for the KK range) calibration at 3 points straight from the control panel.
advertisement
TALOGUE 20
2
POL-EKO-APARATURA sp.j.
POL-EKO-APARATURA has been present in the scientific market since 1990. In addition to manufacturing and trading activities, the company provides professional advice on selection, operation and maintenance of measuring, analysis and control equipment. We organise seminars on laboratory analysis, waste management and disposal, laboratory management systems and validation of analytical methods.
Since 2006 we have exported our equipment worldwide. Nowadays, our export sale makes up to 40% of all manufactured products that have been delivered to many countries of Europe, Asia, Africa, North and South America, and Australia. Independently of developing the manufacturing activity, we have been an authorized distributor in Poland of measuring and control equipment provided by the following companies: ARCTIKO,
EUTECH, HAMILTON, KNICK, THERMO SCIENTIFIC, WTW.
2
In 2005 we moved our headquarters to Wodzis³aw Œl¹ski. New, modern office buildings and production halls of a total area of 4 000 m provide safe and friendly work environment.
To deliver complex services and quality products to our customers, we implemented an Integrated Management System in compliance to the PN-EN
ISO 9001 norm in 2004. We also established a Measurement Laboratory that provides calibration of laboratory equipment and has been accredited since 2008 by the Polish Centre for Accreditation (PCA) a member of ILAC.
Table of contents
Table of contents
About us
LABORATORY THERMOSTATIC CHAMBERS
Thermostatic cabinets
Laboratory refrigerators
Thermostatic boxes
Laboratory freezers
Laboratory incubators
Cooled incubators
Drying ovens
Pass-through drying ovens
Laboratory sterilizers
Equipment with photoperiodic and phytotron system
Climatic chambers
Climatic chambers with phytotron system
OPTIONS AND ACCESSORIES
Parameters
Options and accessories
Temperature protection
STEEL LABORATORY FURNITURE
Fume hoods
Steel laboratory furniture
OTHER LABORATORY EQUIPMENT
Colony counter
Laboratory shakers
Stationary samplers
POL-EKO LABORATORIUM POMIAROWE COMPANY
POL-EKO Measurement Laboratory
35
38
40
43
20
29
32
9
14
19
3
4
7
67
69
71
46
46
51
52
57
84
86
87
88
73
81
83
3
POL-EKO-APARATURA founded
Start of production
POL-EKO-APARATURA
New product launched: waste water receipt station
New product launched: stationary sampler
Starting cooperation with
WTW Germany;
POL-EKO-APARATURA as general distributor in Poland
New products launched: thermostatic cabinet and laboratory refrigerator
New product launched: colony counter
Starting cooperation with foreign laboratory and measuring equipment producers, such as
AQUALYTIC, EUTECH
New products launched: stainless steel drying oven, incubator and cooled incubator
Moving the company to new headquarters in Wodzis³aw Œl¹ski
Development of export sale
New product launched: laboratory sterilizer
Eurolab Trade Fairs
Award - CyberScan 6500
New products launched: instruments in TOP version with Ethernet connection
Measurement
Laboratory founded
New products launched: climatic chambers and climatic chambers with phytotron system
4
POL-EKO-APARATURA chosen the 2008
”Local Leader of Enterprise”
Starting building another production hall
Eurolab Trade Fairs
Award - Cooled incubator
ILW 115 TOP+
New products launched: laboratory furniture and fume hoods
POL-EKO-APARATURA exhibits at ACHEMA in Frankfurt am Main
Eurolab Trade Fairs
Award - Monitoring station PP 2002M
New product launched:
Pass-through drying oven
POL-EKO Laboratorium pomiarowe company founded
development history
New products launched: pH and dissolved oxygen on-line meters
Eurolab Trade Fairs
Award - CyberScan 5500
Eurolab Trade Fairs
Award - BOD OxiDirect
New products launched: stainless steel drying oven, incubator and cooled incubator
New product launched: homogenizer
Eurolab Trade Fairs
Award - EcaFlow 150 GLP
Eurolab Trade Fairs
Award - stationary sampler PP 2002
Implementing ISO 9001 and 18001 systems
Measurement Laboratory receives accreditation from the Polish Centre for Accreditation
POL-EKO-APARATURA exhibits at IFAT expo
in Munich
New production hall opened
Eurolab Trade Fairs
Award - Orion STAR PLUS
City Mayor of Wodzis³aw
Œl¹ski awards
POL-EKO-APARATURA for involvement in cooperation between Polish and Czech enterprises and municipalities
New product launched: laboratory shaker
POL-EKO-APARATURA co-exhibits at ARABLAB
Fairs in Dubai
POL-EKO-APARATURA receives an award from
Local Chamber of
Commerce: “Friendly to the Environment”
POL-EKO-APARATURA co-exhibits at PITTCON in the USA and ANALITIKA in Moscow
Graphite revolution
New design of
POL-EKO-APARATURA equipment
5
Contact Us
Contact Us
POL-EKO-APARATURA: tel. +48 32 453 91 70 e-mail: [email protected]
Export Managers: tel. +48 32 453 91 88; +48 32 453 91 93
Head of Marketing Department: tel. +48 32 453 91 75
Shipping and invoicing: tel. +48 32 453 91 90
Fairs and exhibitions
Some of the lab exhibitions we have participated in: EuroLab - Poland, IFAT - Germany, Achema - Germany, ArabLab - United Arab Emirates,
WodKan - Poland, POLEKO - Poland, Medicina and Technology - Croatia, LabComplEX - Ukraine, Rich-Mac - Italy, Pittcon - United States of
America, Analitica Latin America - Brazil, Lab Asia - Malaysia, A-TESTex/Analitika - Russia.
EuroLAB
6
ArabLAB
IFAT
ACHEMA
Laboratory thermostatic chambers
I
8
Laboratory thermostatic chambers
Thermostatic cabinets
single chamber
double chamber
Laboratory refrigerators single chamber
double chamber
Thermostatic boxes
Laboratory freezers
Incubators, cooled incubators, drying ovens, sterilizers, climatic chambers
STD - standard version
TOP+ - professional version
TOP+ Control and EasyLab-T PLUS software
Laboratory Incubators
Cooled incubators
Drying ovens
Pass-through drying ovens
Laboratory sterilizers
Equipment with photoperiodic and phytotron system
Climatic chambers
Climatic chambers with phytotron system
Options and accessories icon description
OPTIONS AND ACCESSORIES
Parameters list
Parameters description
Options and accessories
Temperature protection
29
32
35
38
24
25
26
27
18
19
20
9
12
13
14
17
51
52
54
57
40
43
46
46
50
67
Application
BOD determination microbiological research plant growing and microorganisms breeding at specified temperature storage of liquids and samples for physicochemical analysis
Thermostatic cabinets ST
Thermostatic cabinets are equipped with both heating and cooling systems and can provide a stable temperature in a range of +3...+70°C regardless of ambient conditions.
LCD graphic display acces port fan internal socket
(option) external glass door (option) door lock
ST 2/2+
7
9
ST Thermostatic cabinets
10
Advantages available in two versions - basic and TOP+ single or double chamber versions (possibility of combination
ST/CHL 2+, 3+, 4+ 5+, ZLN 85+) capacity from 68 to 1460 liters aluminum or stainless steel interior (depending on model) housing: powder coated sheet, stainless steel linen finish
(INOX/G) or polished (INOX/S) (depending on model) thermal insulation - polyurethane foam solid, glass (ST*/A option) or double (ST*/C option) door type forced air convection (optional fan speed control) heating and cooling systems to hold set temperature regardless of ambient conditions defrosting function (see page 56) wheels in ST 1200 and 1450 model two temperature ranges available
+3…+40°C
+3…+70°C (ST/70 option)
LCD graphic display illuminated touch buttons microprocessor temperature controller temperature sensor fail alarm (see page 56) power failure control system (see page 56) real time clock sound alarm (see page 56) self-check function (auto control) automatic fan shut down after completing the program test results memory
Program possibilities in basic version six segment temperature-time profile cycle run of the program
3 user programs memory adjustable start delay feature (from 1 min to 99:59 h) adjustable hold at set temperature point from 1 min to
31 days / 1 min to 99:59 h or continuous operating recording min, average and max temperature value separately for each segment overview of set parameters during operating time over/under temperature sound alarm (see page 56) temperature program priority; temperature or time program priority with FOT option (see page 56)
Program possibilities in TOP+ version, see page 26.
Temp. [°C]
70°C
T2 55°C off 100%
T1 37°C
T0 20°C
D
0,5 h
100% 100% 100% 100% off
Maximal temperature
TIME 2
TIME 1
37 °C 55°C
3 h 1,5 h temperature
Time [h]
- parameter set by User:
D - adjustable start delay (1 min...99,59 h)
- fan speed control
T1/T2 - set temperature
TIME 1/2 - set time
All POL-EKO-APARATURA products can be supplied with a calibration certificate issued by an accredited
Measurement Laboratory. Further information is available on pages 87-89 and our website www.polekolab.com
Thermostatic cabinets ST
Thermostatic cabinets in TOP+ version
The ST thermostatic cabinets are available in the TOP+ professional version too. The TOP+ version features a full colour 5,7’’ touch screen. These are the only thermostatic cabinets which can be controlled from a PC, as they can be connected to an Ethernet network that enables the user to program the equipment virtually from any place in the world. Details on page 26.
Standard equipment solid door temperature range +3…+40°C (+3…+70°C in TOP+ version)
RS 232 interface and USB port ; RS 232 or USB cable must be purchased additionally (RSK or USBK) wire shelves (with slides set) in case of aluminum interior and stainless steel wire shelves in case of stainless steel interior manfacturer’s test certificate at +37°C operation manual in English and English menu (other languages on request) over temperature protection 1.0 class according to DIN 12880 door lock open door alarm access port for external sensor (diameter: 30 mm)
TOP+ standard equipment see page 26
Options and accesories
(icon description see page 50)
INOX
USBK
50...100%
KAFKA
90%
30%
9
for basic version only
5
DIN
3.3
+70°C
KG
SMS
IQ
OQ
PQ
for basic version and TOP+
RS 422 RS 485
RS 232 RS 422 RS 485
for TOP+ only
11
ST Single chamber cabinets
Available options
:
ST/70 - extended temperature range up to +70°C
ST*/C - internal glass door
ST*/A - external glass door
PLUS - automatic defrosting during operating
FOT - photoperiodic system
Thermostatic chambers are available with photoperiod option, see page 43.
ST 2+ with ST/FOT 6 option
Model
ST 1+ ST 2+ ST 3+ ST 4+ ST 5+ ST 500 ST 700 ST 1200 ST 1450
12
Parameter air convection chamber capacity
1
[l]
[cu ft] door type temperature range
[°C]
[°F] controller interior housing overall dims [mm]**
3 internal dims [mm] width height
depth max shelf workload [kg]*
-
PW version max unit workload [kg]
W version nominal power [W] weight [kg] temperature resolution [°C]
4 temperature fluctuation at +37°C [°C] over/under temperature protection voltage 50/60 Hz [V] shelves fitted / max
6 warranty manufacturer width height depth
68
2,4
20
160
30
±0,2
2/2
* - at even surface workload
** - ST 1-5 in TOP+ version are 60 mm higher
dims of power cable (50 mm) not included
1 - working capacity of chamber can be smaller
2 - additional internal glass door
570
600
670
470
430
300
10
150
5,2 forced
200
7,0
250
8,8
300
10,5
493
17,4
2 solid / glass or double (option)
625
22,0
+3...+40 / up to +70 (option) / +3...+70 in TOP+ version
+37...+104 / up to +158 (option) / +37...+158 in TOP+ version microprocessor with external LCD graphic display
1365
48,2 powder coated sheet / stainless steel linen finish INOX/G or polished INOX/S
620
860
640
520
660
420
10
30
620
1060
640
520
860
420
10 on request
40
620
1260
640
520
1060
420
10
50
620
1460
640
520
1260
420
10
60 on request
645
2025
820
510
1510
640
20
100
100
735
2025
870
600
1510
690
30
100
150
1440
2045
860
1310
1510
690
30
100
300
170
48
170
52 aluminum / stainless steel
330
61
330
66 every 0,1
400
105
400
115
550
195
±0,2 ±0,3 ±0,3 ±0,3 ±0,2 ±0,2 ±0,3 class 1.0 according to DIN 12880 / class 3.3 (option) / class 3.3 in TOP+ version
3/4 3/4 4/6
115/230
4/7 3/11 3/11 2 x 3/11
5
24 months
POL-EKO-APARATURA
3 - dims of units with double door can be smaller
4 - fluctuation measured in centre of the chamber
5 - two columns with 3 shelves each
6 - "fitted” number of shelves fitted, included in price
6 "max" number of shelves which can be fitted
1460
51,6
550
200
±0,3
2 x 3/11
5
1450
1970
950
1340
1460
750
30
100
300
Double chamber cabinets ST
Available options for each chamber
ST/70 - extended temperature range up to +70°C
ST*/C - internal glass door
ST*/A - external glass door
PLUS - automatic defrosting during operating
FOT - photoperiodic system
There is a possibility to combine different sizes of ST+ and CHL+ chambers between models ST/CHL+ 2-5 (e.g ST 2/CHL 3+).
ST 2/2+ with additional options
Example of configuration
To configure a double-chamber thermostatic cabinet, e.g.:
- upper chamber equipped with solid door and temperature up to +70°C
- lower chamber equipped with double door and temperature up to +40°C, you select: for ST 1/1+: ST 1/1+ and ST/70 option for upper chamber + ST 1+/C option for lower chamber
Model ST 1/1+ ST 1/1/1+ ST 2/2+ ST 2/3+ ST 2/4+ ST 3/3+ ST 350/350
Parameter air convection chamber capacity
1 door type temperature range controller interior housing overall dims [mm]**
[l]
[cu ft]
[°C]
[°F]
3 internal dims [mm] max shelf workload [kg]* max unit workload [kg] nominal power [W] weight [kg] temperature resolution [°C] over/under temperature protection voltage 50/60 Hz [V] shelves fitted / max
5 warranty manufacturer
* - at even surface workload
** - dims of power cable (50 mm) not included
1 - working capacity of chamber can be smaller
2 - additional internal glass door width height depth width height
depth
-
PW version
W version
570
1170
670
470
430
300
10
68/68
2,4/2,4
68/68/68 150/150 forced
150/200 150/250 200/200
2,4/2,4/2,4 5,2/5,2 5,2/7,0 5,2/8,5
2 solid / glass or double (option)
7,0/7,0
+3...+40 / up to +70 (option) / +3...+70 in TOP+ version
+37...+104 / up to +158 (option) / +37...+158 in TOP+ version microprocessor with external LCD graphic display aluminum / stainless steel powder coated sheet / stainless steel linen finish INOX/G or polished INOX/S
570
1740
670
470
430
300
10
620
1680
640
520
660
420
10
620
1875
640
520
660/860
420
10 on request
620
2080
640
520
660/1060
420
10
620
2080
640
520
860
420
10
294/294
10,3/10,3
720
2045
860
600
700
700
30
20/20
320
65
±0,2
20/20/20
480
100
30/30
350
100
30/40 on request
350
105 every 0,1
30/50
350
110
40/40
350
115
±0,2 ±0,3 ±0,3 ±0,3 ±0,3 class 1.0 according to DIN 12880 / class 3.3 (option) / class 3.3 in TOP+ version
75/75
800
162
±0,5
2/2 / 2/2 / / 3/4 / 3/4
115/230
3/4 / 3/4
24 months
POL-EKO-APARATURA
3 - dims of units with double door can be smaller
4 - fluctuation measured in centre of the chamber
5 - "fitted” number of shelves fitted, included in price
6 "max" number of shelves which can be fitted
3/4 / 4/6 2/6 / 2/6
13
CHL
Laboratory refrigerators
Application storage of water and sewage samples, piezometer leachate storage of ASA, GC or HPLC calibration standards storage of reagents chemical storage general storage
Laboratory refrigerators are equipped with a cooling system and can provide a stable temperature lower than ambient.
RS 232 and
USB port
LCD graphic display adjustable shelf with slide set illuminated touch buttons
CHL 2+ fan
14
Laboratory refrigerators
CHL
Advantages available in two versions - basic and TOP+ single or double chamber versions (possibility of combination
ST/CHL 2+, 3+, 4+ 5+, ZLN 85+) capacity from 68 to 1460 liters aluminum or stainless steel interior (depending on model) housing: powder coated sheet, stainless steel linen finish
(INOX/G) or polished (INOX/S) (depending on model) thermal insulation - polyurethane foam solid, glass (CHL*/A option) or double (CHL*/C option) door type forced air convection (optional fan speed control) cooling systems to hold set temperatureregardless of ambient conditions defrosting function (see page 56) wheels in CHL 1200 and 1450 model
Program possibilities in basic version single segment temperature-time profile
1 user program memory temperature program priority (see page 56) adjustable start delay feature (from 1 min to 99:59 h) adjustable hold at set temperature point from 1 min to
31 days / 1 min to 99:59 h or continuous operating two temperature ranges available
0…+10°C
-10…+10°C (CHL*/T option) for CHL 500-1450
LCD graphic display illuminated touch buttons microprocessor temperature controller temperature sensor fail alarm (see page 56) power failure control system (see page 56) real time clock sound alarm (see page 56) self-check function (auto control) automatic fan shut down after completing the program test results memory recording min, average and max temperature value overview of set parameters during operating time over/under temperature sound alarm (see page 56)
Temp. [°C] off
D
Program possibilities in TOP+ version, see page 26.
TIME
1
- parameter set by User:
D - adjustable start delay (1 min...99:59 h)
T1 - temperature
TIME 1 - time
0,5h 5h Time [h]
All POL-EKO-APARATURA products can be supplied with a calibration certificate issued by an accredited
Measurement Laboratory. Further information is available on pages 87-89 and our website www.polekolab.com.
15
CHL
Laboratory refrigerators
Laboratory refrigerators CHL
All refrigerators, thermostatic cabinets and cooled incubators can optionally be equipped with an automatic defrosting function
(PLUS option). It features a countinuous operating without a necessity to defrost, keeping a stable temperature in the chamber.
Details on page 65.
Standard equipment solid door temperature range 0…+10°C
RS 232 interface and USB port ; RS 232 or USB cable must be purchased additionally (RSK or USBK) wire shelves (with slides set) in case of aluminum interior and stainless steel wire shelves in case of stainless steel interior manfacturer’s test certificate at +4°C operation manual in English and English menu (other languages on request) over temperature protection 1.0 class according to DIN 12880 door lock open door alarm access port for external sensor (diameter: 30 mm)
TOP+ standard equipment see page 26
16
Options and accesories
(icon description see page 50)
INOX
DIN
3.2
SMS RS 422 RS 485
RS 232 RS 422 RS 485 USBK
IQ
OQ
PQ
for basic version only for basic version and TOP+
KAFKA
50...100%
9 5
for TOP+ only
Single chamber refrigerators
CHL
Available options
CHL*/T - extended temperature range down to -10
°C
CHL*/A - external glass door
PLUS - automatic defrosting function during operating
A mini refrigerator
TB 35 INOX with an additional battery power supply guarantees continuous operating in case of power failure.
Details on page 19.
TB 35 INOX
Model
CHL 1+ CHL 2+ CHL 3+ CHL4+ CHL 5+ CHL 500 CHL 700 CHL 1200 CHL 1450
Parameter air convection chamber capacity
1
[l]
[cu ft] door type temperature range
[°C]
[°F] controller interior housing overall dims [mm]** width height depth
3 internal dims [mm] width height
depth max shelf workload [kg]*
-
PW version max unit workload [kg]
W version nominal power [W] weight [kg] temperature resolution [°C]
4 temperature fluctuation at +4°C [°C] over/under temperature protection voltage 50/60 Hz [V] shelves fitted / max
6 warranty manufacturer
* - at even surface workload
** - CHL 1-5 in TOP+ version are 60 mm higher
dims of power cable (50 mm) not included
1 - working capacity of chamber can be smaller
2 - additional internal glass door
68
2,4
570
600
670
470
430
300
10
20
160
30
±0,5
2/2
620
860
640
520
660
420
10
150
5,2
200 250 forced
300 493 625 1365 1460
7,0 8,8 10,5 17,4
2 solid / glass or double (option)
22,0 48,2
+0...+10 +0...+10 / -10...+10 (option)
+32...+50 +32...+50 / +14...+50 (option) microprocessor with external LCD graphic display
51,6 aluminum / stainless steel powder coated sheet / stainless steel linen finish INOX/G or polished INOX/S
30
620
1060
640
520
860
420
10 on request
40
620
1260
640
520
1060
420
10
50
620
1460
640
520
1260
420
10
60 on request
645
2025
820
510
1510
640
20
100
100
735
2025
870
600
1510
690
30
100
150
1440
2025
860
1310
1510
690
30
100
300
1450
1970
950
1340
1460
750
30
100
300
170
48
170
52
330
61
330
66 every 0,1
400
105
400
115
550
195
±0,5 ±0,5 ±0,5 ±0,5 ±0,5 ±0,5 ±0,5 class 1.0 according to DIN 12880 / class 3.2 (option) / class 3.2 in TOP+ version
115/230
3/4 3/4 4/6 4/7
24 months
3/11 3/11
2 x 3/11
5
POL-EKO-APARATURA
550
200
±0,5
2 x 3/11
5
3 - dims of units with double door can be smaller
4 - fluctuation measured in centre of the chamber
5 - two columns with 3 shelves each
6 - "fitted” number of shelves fitted, included in price
6 "max" number of shelves which can be fitted
17
CHL
Double chamber refrigerators
Available options for each chamber
CHL*/T - extended temperature range to -10
°C
CHL*/A - external glass door
PLUS - automatic defrosting during operating
There is a possibility to combine different sizes of ST+ and CHL+ chambers between models ST/CHL+ 2-5 (e.g ST 2/CHL 3+).
Example of configuration
To configure a double chamber laboratory refrigerator, e.g.:
- upper chamber with solid door and temperature range 0…+10°C
- lower chamber with solid door and temperature range -10…+10°C, you select: for CHL 350/350 model: CHL 350/350 and CHL 350/T option for lower chamber
CHL 2 / ZLN 85+
Model CHL 1/1+ CHL 1/1/1+ CHL 2/2+ CHL 2/3+ CHL 2/4+ CHL 3/3+ CHL 350/350
18
Parameter air convection chamber capacity
1
[l]
[cu ft] door type temperature range
[°C]
[°F] controller interior housing overall dims [mm]**
3 internal dims [mm] width height depth width height
depth max shelf workload [kg]*
-
PW version max unit workload [kg]
W version nominal power [W] weight [kg] temperature resolution [°C]
4 temperature fluctuation at +4°C [°C] over/under temperature protection voltage 50/60 Hz [V] shelves fitted / max
5 warranty manufacturer
* - at even surface workload
** - dims of power cable (50 mm) not included
1 - working capacity of chamber can be smaller
2 - additional internal glass door
570
1170
670
470
430
300
10
68/68
2,4/2,4
68/68/68 150/150 forced
150/200 150/250 200/200 294/294
2,4/2,4/2,4 5,2/5,2 5,2/7,0 5,2/8,5
2 solid / glass or double (option)
0...+10 / -10...+10 (option only for CHL 350/350)
7,0/7,0
+32...+50 / +14...+50 (option only for CHL 350/350) microprocessor with external LCD graphic display aluminum / stainless steel powder coated sheet / stainless steel linen finish INOX/G or polished INOX/S
10,3/10,3
570
1740
670
470
430
300
10
620
1680
640
520
660
420
10
620
1875
640
520
660/860
420
10 on request
620
2080
640
520
660/1060
420
10
620
2080
640
520
860
420
10
720
2045
860
600
700
700
30
20/20
320
65
±0,5
20/20/20
480
100
30/30
350
100
30/40 on request
350
105 every 0,1
30/50
350
110
40/40
350
115
±0,5 ±0,5 ±0,5 ±0,5 ±0,5 class 1.0 according to DIN 12880 / class 3.2 (option) / class 3.2 in TOP+ version
75/75
800
162
±0,5
2/2 / 2/2 2/2
/
2/2
/
2/2 3/4 / 3/4
115/230
3/4 / 3/4 3/4 / 4/6
24 months
POL-EKO-APARATURA
3 - dims of units with double door can be smaller
4 - fluctuation measured in centre of the chamber
5 - "fitted” number of shelves fitted, included in price
6 "max" number of shelves which can be fitted
3/4 / 3/4 2/6 / 2/6
Thermostatic boxes
TB
Advantages chamber capacity from 25 to 106 litres transport in stable temperature, e.g. +4°C adjustable temperature from
-18 to +10°C
-10 to +10°C for TB 35 INOX stable temperature regardless of ambient conditions temperature display battery, car socket (not for TB 80A and TB 105A) or standard mains power supply, wireless temperature data loggers available (option)
Application transport of:
water and sewage samples
biological substances, tissues etc.
blood and haematoid material
according to certain directives
Thermostatic boxes, unlike normal car freezers, guarantee a stable temperature during transport.
TB 35 INOX is a solution not only for customers who carry samples. With an additional battery power supply, it features also the functions of a mini refrigerator which is able to hold set temperature in case of power failure.
Parameter chamber capacity temperature range
[l]
[cu ft] o
[ C] o
[ F] interior housing overall dims [mm] width height depth internal dims [mm] nominal power [W] weight [kg] o temperature resolution [ C] o o temperature fluctuation at +4 C [ C] voltage [V] warranty manufacturer width height depth chamber with bottles for samples
Model TB 32 A TB 50 A
25
0,8
265
300
60
20
580
385
360
315
TB 80 A wide range of
wireless temperature data loggers is available
Options and accesories
(only for TB 35 INOX)
9
TB 105 A TB 35 INOX
630
480
360
315
410
265
45
25
49
1,7
80
2,8
-18...+10
0...+50 powder coated aluminium
plastic
106
3,7
790
455
500
500
365
400
65
45 every 1,0
± 1,0
100-240 50/60 Hz or 12 DC
12 months
790
555
500
500
465
400
65
50
POL-EKO-APARATURA
35
1,2
-10...+10
+14...+50 stainless steel
powder coated sheet
600
425
440
340
310
340
90
40
19
ZL Laboratory freezers
Application long-term storage of samples and biological material for research storage of easily decomposing material (e.g. solid state) freeze resistance tests (e.g. of building materials: concrete, wood etc.) pre-freezing
Laboratory freezers can freeze and store frozen samples.
acces port
LCD graphic display
RS 232 and USB port
(at the rear) perforated shelf door handle with door lock
ZLN 300
20
Laboratory freezers ZL
Advantages single chamber models, possibility to combine ZLN 85+ with models: CHL 2+ to CHL 5+ and ST 2+ to ST 5+ capacity from 85 to 299l plastic or stainless steel interior
(depending on model) housing: powder coated sheet, stainless steel linen finish
(
INOX/G) or polished (INOX/S) (depending on model) two temperature ranges depending on model:
-25...0°C
-40...0°C thermal insulation - polyurethane foam cooling system
Program possibilities single segment temperature-time profile
1 user program memory temperature program priority (see page 56) adjustable start delay feature (from 1 min to 99:59 h) adjustable hold at set temperature point from 1 min to
99:59 h or continuous operating
Temp. [°C]
20°C off
D
11
T1 -10°C TIME 1
-10°C
5h 0,5h natural air convection
LCD graphic display illuminated touch buttons microprocessor temperature controller temperature sensor fail alarm (see page 56) power failure control system (see page 56) real time clock sound alarm (see page 56) self-check function (auto control) test results memory recording min, average and max temperature value
(except ZLN 75, 145, 180) overview of set parameters during operating time over/under temperature sound alarm (see page 56)
Time [h]
- parameter set by User:
D - adjustable start delay (1 min...99,59 h)
T1 - set temperature
TIME 1 - set time
All POL-EKO-APARATURA products can be supplied with a calibration certificate issued by an accredited
Measurement Laboratory. Further information is available on pages 87-89 and our website www.polekolab.com.
21
ZL Laboratory freezers
ZLN 85 + freezer
ZLN 85+ laboratory freezer can be combined with laboratory refrigerators CHL 2+, 3+, 4+, 5+ or thermostatic cabinets ST 2+, 3+, 4+, 5+ and delivered as a double chamber model.
Standard equipment solid door
RS 232 interface and USB port ; RS 232 or USB cable must be purchased additionally (RSK or USBK) glass shelves in models ZLN 75, 145, 180; wire shelves (with slides set) in model ZLN 85+ and perforated shelves in models ZLN-T 125, 200, 300 manufacturer’s test certificate at -10°C operation manual in English and English menu (other languages on request) door lock in models ZLN 85+ and ZLN-T 125, 200, 300 open door alarm acces port for external sensor (diameter: 30 mm) in models ZLN 85+ and ZLN-T 125, 200, 300
22
Options and accesories
(icon description see page 50)
INOX
KG
KAFKA 9
RS 485
RS 232 RS 422 RS 485
for ZLN 75, 145, 180 models
USBK
5
for ZLN 85+ and ZLN-T
IQ
OQ
PQ
SMS RS 422
Laboratory freezers ZL
Available options
*/W - reinforced version
*/PW - reinforced shelf
QLK/* - wheels
ZLN 85+
Model ZLN 75 ZLN 180 ZLN 85+ ZLN-T 125 ZLN-T 200 ZLN-T 300
Parameter air convection chamber capacity door type temperature range controller interior housing
1
[l]
[cu ft]
[°C]
[°F]
85
3,0 natural
196
6,9
86
3,0
125
4,4
200
7,0 solid
-25...0
-40...0
-13...+32 -40...+32 microprocessor with external LCD graphic display powder coated sheet stainless steel powder coated sheet
/ stainless steel (INOX/G or INOX/S) overall dims [mm]** width height depth internal dims [mm] width height
depth max shelf workload [kg]*
-
PW version max unit workload [kg]
W version nominal power [W] weight [kg] temperature accuracy [°C]
2 temperature fluctuation at -10°C [°C]
50 Hz [V] voltage
60 Hz [V]
560
850
610
410
630
390
10
-
30
-
120
34
±0,4
230
-
600
1455
650
450
1160
420
10
-
50
-
100
55
±0,7
620
885
640
420
590
395
10
-
30
-
200
60
±0,4 every 0,1
675
1170
800
396
600
524
10
50
50
100
450
90
±0,7
2/2 5/5 2/4 2/3 warranty 24 months manufacturer POL-EKO-APARATURA
* - at even surface workload
** - dims of power cable (50 mm) not included
1 working capacity of chamber can be smaller
2 - fluctuation measured in centre of the chamber
3 - "fitted” number of shelves fitted, included in price, "max" number of shelves which can be fitted
230
775
1385
800
496
770
524
10
50
65
130
450
120
±0,7
2/4
299
10,5
775
1735
800
496
1120
524
10
50
80
160
450
185
±1,0
3/6
23
24
Incubators, cooled incubators, drying ovens , sterilizers, climatic chambers
Laboratory incubators (CL), drying ovens (SL), cooled incubators (IL) are available in 2 versions: STD and TOP+, sterilizers (SR) in STD version only and climatic chambers (KK) in TOP+ version only.
ST, CHL, ZLN STD TOP+
displays fan stainless steel interior automatic air-flap control
(at the rear)
CLW 53 STD
STD - standard version STD
Equipment available in STD version
Incubators (CL)
Cooled incubators (IL)
Drying ovens (SL)
Sterilizers (SR)
STD-standard version extended programming possibilities wide range of optional equipment affordable price
Program possibilities six segment temperature-time profile temperature program priority (see page 56)
3 user programs memory cycle run of the program adjustable heating up/cooling down system adjustable start delay feature (from 1 min to 99:59 h or date/time) adjustable hold at set point (from 1 min to 999:59 h), or continuous operating (temperature; light with FOT and FIT options) overview of set parameters during operating time recording min, average and max temperature value for each segment over/under temperature sound alarm (see page 56) fan speed control (for instruments with forced air convection) automatic air-flap control (adjustable)
Temp. [°C] off 100% 50% 100% 40% 100% 50% 100% off
T3 70°C
T1 50°C TIME 1
TIME
0/1
50°C
T0 25°C
T2 20°C
D
TIME 1/2
1,5h
TIME 2
20°C
3h 1h 2h 5h
Standard equipment stainless steel interior housing: powder coated sheet, stainless steel linen finish
(INOX/G) or polished (INOX/S) (depending on model) solid door forced or natural air convection (depending on model) automatic fan shut down after completing the program
(for instruments with forced air convection)
LCD graphic display illuminated touch buttons microprocessor temperature controller temperature sensor fail alarm (see page 56) open door alarm access port for external sensor (diameter: 30 mm)
1,5h
TIME
2/3
TIME 3
70°C
3h
TIME 3/4 temperature
1,5h
Time [h]
- parameter set by User:
D - adjustable start delay
(1 min...99:59 h)
- fan speed control
T1/T2/T3 - set temperature
TIME 1/2/3 - set time
TIME 0/1...
- time to achieve set
temperature point power failure control system (see page 56) real time clock sound alarm (see page 56)
RS 232 interface and USB port self-check function (auto control) test results memory manufacturer’s test certificate operation manual in English and English menu
(other languages on request) stainless steel wire shelves (1-6 pieces, depending on model) with slides set over temperature protection according to DIN 12880: 2.0 class
25
TOP+ TOP+ - professional version
Equipment available in TOP+ version
Incubators (CL 53-1000)
Cooled incubators (IL)
Drying ovens (SL 53-1000)
Climatic chambers (KK)
Thermostatic cabinets (ST)
Laboratory refrigerators (CHL)
TOP+ - professional version
World's first full colour touch screen thermostatic equipment with Ethernet network connection!
To meet the expectations of the most demanding customers, POL-EKO-APARATURA has introduced a new professional range of ovens, incubators, cooled incubators and climatic chambers called “TOP+”. First in the market to feature Ethernet network connection which allows to connect several units and program, monitor or record their parameters on any computer inside a laboratory or remotely via internet.
Equipped with a full colour, clear 5.7'' LCD touch screen, easy-to-use intuitive software, USB port for data transfer make the operating comfortable. The large size of the display enables the user to view the unit's current parameters and program stage with only one touch of the screen. Additionally, the controller has been equipped with a 7-day programming mode and temperature (and humidity for the KK range) calibration feature at 3 points straight from the control panel.
Temp. °C
26
T3 70°C off
0%
100%
100%
50%
50%
100%
40%
40%
50%
100%
40%
T1 50°C TIME 1
TIM
E
0/1
50°C
T0 25°C
T2 20°C
D
TIME 1/2
1,5h
TIME 2
20°C
3h 1h 2h 5h
Standard equipment stainless steel interior housing: powder coated sheet, stainless steel linen finish
(INOX/G) or polished (INOX/S) (depending on model) solid door forced or natural air convection (depending on model) automatic fan shut down after completing the program
(for instruments with forced air convection) automatic air-flap control (adjustable) full colour, transparent 5.7” LCD touch screen microprocessor temperature controller temperature (and humidity in KK) sensor fail alarm
(see page 56) power failure control system (see page 56) real time clock sound alarm (see page 56) access port for external sensor (diameter: 30 mmm) e-mail info (see page 56)
1,5h
TIME
2/3
50%
100%
TIME 3
70°C
3h
100%
100% off
0%
TIME 3/4 temperature
1,5h
Time [h]
- parameter set by User:
D - adjustable start delay
(1 min...99:59 h)
- fan speed
- air flap control
T1/T2/T3 - set temperature
TIME 1/2/3 - set time
TIME 0/1...
- time to reach set
temperature point
USB port
RS 232 interface self-check function (auto control) automatic defrosting in KK range manufacturer’s test certificate operation manual in English and English menu
(other languages on request) stainless steel wire shelves
(2-6 pieces, depending on model) with slides set over/under temperature protection according to DIN 12880: 3.1 class in CL/SL; 3.2 class in CHL;
3.3 class in IL/KK/ST
Ethernet connection and remote control via Internet
TOP+ Control software, temperature (and humidity in KK, light with FIT and FOT options) control and monitoring via PC open door alarm fan speed control (for instruments with forced air convection)
TOP+ - professional version TOP+
Advanced program possibilities of the “TOP+” range: administrator function that allows to manage user accounts access control via login
10 000 segment temperature-time profile or 7 days programming temperature (and humidity in KK) calibration at 3 points automatic air-flap control (each 10% regulation) cycle run of the program (1 - 99 h or continuous) adjustable heating up / cooling down system adjustable start delay feature (from 1 min to 99:59 h or date/time) adjustable hold at set point (from 1 min to 999:59 h), or continuous operating (temperature; humidity in KK; light with FOT and FIT options) overview of set parameters during operating time recording min, average and max temperature (and humidity in KK) value for each segment over/under temperature (and humidity in KK) sound alarm (see page 56) fan speed control (for instruments with forced air convection) time or temperature priority program
5.7” LCD full colour touch screen panel
GLP supporting functions: password protected settings
20 user programs memory test results memory with possibility of displaying the values in tabular or graphic form measurement data saving on external memory through the USB port
Main screen User’s menu Temperature and humidity chart
Program settings Administrator panel settings Administrator panel settings
Administrator panel settings 7 days programming Measurement recording
27
EasyLab Temperature and humidity monitoring
TOP+ Control
To facilitate the configuration of complex programs, a TOP+ Control software has been introduced. Moreover, the user is able to program and control the “TOP+” equipment with ease from any corner of the world by accessing the unit via Internet!
Device manager Temperature chart
EasyLab-T PLUS
EasyLab-T PLUS software features temperature and humidity monitoring in all thermostatic equipment manufactured by POL-EKO-
APARATURA.
The user may record constant or temporary values, accumulate them and convert into charts. RS 232 or USB port allows the recording process (it is necessary to purchase a connection cable along with the software). If the unit is equipped with an additional Pt 100 temperature sensor, the EasyLab-T PLUS software enables simultaneous recording. Additionally EasyLab-T PLUS software enables programming of devices in TOP+ version, thanks to integrated TOP+ Control application.
EasyLab-T PLUS features high quality tools for chart making and approximation.
Standard features of EasyLab-T PLUS software:
report creator
data import from external memories
multilanguage (Polish, English, German, Spanish, Hungarian, Portuguese, Romanian)
28
Tabular report
Temperature and humidity chart
Laboratory incubators CL
Application incubation of samples for microbiological examinations analysis of thermal resistance of samples subjected to higher temperatures (e.g. pigments and lacquers) antibody testing bacterial research chemical storage crystallization studies cultivating thermophilic bacteria denaturalizing tests for food industry microbiological determinations pharmaceutical stability testing
Laboratory incubators can store samples in a higher temperature of up to 100°C.
acces port air flap
(at the rear) illuminated touch buttons stainless steel housing linen finish
(INOX/G)
CLW 115 STD INOX/G
29
CL Laboratory incubators
30
Advantages available in STD and TOP+ versions capacity from 15 to 1005 litres stainless steel interior housing: powder coated sheet or stainless steel linen finish (INOX/G) (depending on model) forced or natural air convection (depending on model) fan speed regulation from 0 to 100% for CLW 15, 32,
53, 115 and 240; from 10 to 100% for CLW 400, 750 and 1000 heating system (temperature in the chamber cannot be lower than ambient temperature)
CL 750 and 1000 models equipped with wheels
LCD graphic display in STD version
LCD touch screen in TOP+ version illuminated touch buttons in STD version microprocessor temperature controller temperature sensor fail alarm (see page 56) power failure control system (see page 56) real time clock sound alarm (see page 56) self-check function (auto control) test results memory
Program possibilities
STD version (see page 25)
TOP+ version (see page 26-27)
Standard equipment solid door
RS 232 interface and USB port ; RS 232 or USB cable must be purchased additionally (RSK or USBK) stainless steel wire shelves (1-6 pieces, depending on model) with slides set manufacturer’s test certificate at +37°C open door alarm air-flap diameter 37 mm for CL 15, 32, 53 and 115;
60 mm for CL 240, 400, 750 and 1000 operation manual in English and English menu
(other languages on request) over temperature protection according to DIN 12880:
2.0 class in STD version, 3.1 class in TOP+ version access port for external sensor (diameter: 30 mm)
Options and accesories
(icon description see page 50)
INOX
HEPA
IQ
OQ
PQ
for STD only
DIN
3.1
KG
SMS RS 422 RS 485
RS 232 RS 422 RS 485 USB
for STD and TOP+
KAFKA 9 5
All POL-EKO-APARATURA products can be supplied with a calibration certificate issued by an accredited
Measurement Laboratory. Further information is available on pages 87-89 and our website: www.polekolab.com.
Laboratory incubators CL
Available options
CL*/A - door with viewing window
CL*/C - internal glass door
CL*/W - reinforced version
PW - reinforced shelf
Model CL 15 CL 32 CL 53 CL 115 CL 240 CL 400 CL 750 CL 1000
Parameter air convection chamber capacity door type warranty
1 temperature range controller interior housing overall dims* [mm] internal dims [mm] max shelf workload [kg] max unit workload [kg] nominal power [W] weight [kg] temperature resolution [°C] temperature fluctuation at +37°C [°C] voltage 50 /60 Hz [V] shelves fitted / max
4
2 over temperature protection
[l]
[cu ft]
[°C]
[°F] width height depth width height depth
-
PW version
-
W version
CLN
CLW
15
0,5 solid
515
550
440
320
240
180
10
20
-
-
300
27
±0,5
±0,5
1/2
32 natural (CLN) / forced (CLW)
1,1 solid or double
3
(option)
56
2
112
3,9
245
8,6
424
15 forced (CLW)
3
749
26,4 solid / door with viewing window or double (option) o
5 above ambient temperature...+100 o o
9 above ambient temperature...+212 o microprocessor with external LCD graphic display
595
630
495
400
320
240
10
-
30 acid-proof stainless steel according to DIN 1.4301
powder coated sheet / stainless steel linen finish (INOX/G)
595
700
600
395
395
360
10
50
40
655
845
700
460
540
450
10
50
60
815
1130
760
600
800
510
10
100
90
1015
1375
760
800
1040
510
10
100
120
-
300
35
80
400
50
120
400
65
300
800
126
300
1200
174
300
1800
260
2/5 every 0,1
±0,5
±0,2
±0,2
±0,2
±0,2
±0,2
±0,2
±0,2
-
±0,2
-
±0,2 class 2.0 according to DIN 12880 in STD / class 3.1 (option) / class 3.1 in TOP+ version
1/3 2/7
115/230
3/10 3/14 5/16
24 months
POL-EKO-APARATURA
1260
1590
865
1040
1200
600
-
100
140 manufacturer
* - dims of power cable (50 mm) not included
1 working capacity of chamber can be smaller
2 - fluctuation measured in centre of the chamber
3 - additional internal glass door
4 - "fitted" - number of shelves fitted, included in price, "max" - number of shelves which can be fitted
1005
35,5
1260
2000
865
1040
1610
600
-
100
-
300
1800
330
-
±0,2
6/22
31
IL Cooled incubators
Application microbiological research plant growing, microorganisms and insects breeding at specified temperature samples incubation
Cooled incubators can store samples in a stable environment regardless of ambient conditions, as the temperature range is -10...+100°C.
stainless steel interior stainless steel wire shelf interior lighting
(option) door with viewing window
(option)
ILW 115 STD INOX/G
32
Cooled incubators IL
Advantages available in STD and TOP+ versions chamber capacity from 56 to 749 litres stainless steel interior housing: powder coated sheet or stainless steel linen finish (INOX/G) (depending on model) forced air convection fan speed control from 10 to 100% heating and cooling systems to hold set temperature regardless of ambient conditions thermal insulation - polyurethane foam
ILW 750 models equipped with wheels
LCD graphic display in STD version
LCD touch screen in TOP+ version illuminated touch buttons in STD version microprocessor temperature controller temperature sensor fail alarm (see page 56) power failure control system (see page 56) real time clock sound alarm (see page 56) self-check function (auto control) test results memory
Program possibilities
STD version (see page 25)
TOP+ version (see page 26-27)
Standard equipment solid door
RS 232 interface and USB port ; RS 232 or USB cable must be purchased additionally (RSK or USBK) stainless steel wire shelves (2-6 pieces, depending on model) with slides set manufacturer’s test certificate at +37°C operation manual in English and English menu
(other languages on request) over temperature protection according to DIN 12880:
2.0 class in STD version, 3.3 class in TOP+ version open door alarm access port for external sensor (diameter: 30 mm)
Options and accesories
(icon description see page 50)
INOX
90%
30%
KAFKA 9 5 IQ
OQ
PQ
DIN
3.3
KG
SMS RS 422 RS 485
RS 232 RS 422 RS 485 USB
for STD only for STD and TOP+ for TOP+ only
All POL-EKO-APARATURA products can be supplied with a calibration certificate issued by an accredited
Measurement Laboratory. Further information is available on pages 87-89 and our website: www.polekolab.com.
33
IL Cooled incubators
Available options
IL*/A - door with viewing window
IL*/C - internal glass door
IL*/W - reinforced version
PW - reinforced shelf
IL*/T - extended temperature range to -10°C
FOT - photoperiodic system
FIT - phytotron system
34
Model ILW 53 ILW 115 ILW 240 ILW 400 ILW 750
Parameter air convection chamber capacity
1
[l]
[cu ft] door type temperature range
[°C]
[°F] controller interior housing overall dims* [mm] internal dims [mm] width height depth
max shelf workload [kg]
PW version
max unit workload [kg]
W version nominal power [W] weight [kg] temperature resolution [°C]
2 temperature fluctuation at +37°C [°C] over temperature protection voltage 50 /60 Hz [V] shelves fitted / max
4 warranty manufacturer width height depth
56
2
590
960
620
395
395
360
10
50
40
80
400
69
±0,2
2/5
112
3,9 forced
245
8,6
424
15
749
26,4
0...+70 (+100 in TOP+ version) / -10...+70 (+100 in TOP+ version) (option)
+32...+158 (+212 in TOP+ version) / +14...+158 (+212 in TOP+ version) (option) microprocessor with external LCD graphic display acid-proof stainless steel according to DIN 1.4301
powder coated sheet / stainless steel linen finish (INOX/G)
650
1100
710
460
540
450
10
50
60
120
815
1430
770
600
800
510
10
100
90
300
1015
1680
770
800
1040
510
10
100
120
300
400
90
±0,2 every 0,1
±0,2 ±0,2 class 2.0 in STD / class 3.3 (option) / class 3.3 in TOP+ version
2/7
800
140
115/230
3/10
1200
185
3/14
24 months
POL-EKO-APARATURA
1260
1890
870
1040
1200
600
10
100
140
300
1800
275
±0,2
5/16
* - dims of power cable (50 mm) not included
1 working capacity of chamber can be smaller
2 - fluctuation measured in centre of the chamber
3 - additional internal glass door
4 - "fitted" - number of shelves fitted, included in price, "max" - number of shelves which can be fitted
Drying ovens SL
Application thermal resistance analysis of building materials, electronic and electrotechnical components etc.
monitoring properties of products subjected to high temperatures drying of wires of papermaking machines drying of laboratory glass general aging and curing general heating ordnance material testing preheating vulcanization studies agricultural genetic conditioning digestion of proteins drying of plant tissues drug metabolism life and exposure studies paper drying serum protein analysis suspended solids evaluation
Drying ovens can provide temperatures higher than ambient up to 300°C.
RS 232 and
USB port
(at the rear) visual alarm stainless steel housing linen finish
(INOX/G)
SLW 240 STD INOX/G
35
SL Drying ovens
36
Advantages
available in STD and TOP+ versions chamber capacity from 15 to 1005 litres stainless steel interior housing: powder coated sheet or stainless steel linen finish (INOX/G) (depending on model) forced or natural air convection (depending on model) fan speed regulation from 0 to 100% for SLW 15, 32, 53, 115 and 240 from 10 to 100% for SLW 400, 750 and 1000 heating system (the temperature in chamber cannot be lower than the ambient temperature)
SLW 750 and SLW 1000 models equipped with wheels
LCD graphic display in STD version
LCD touch screen in TOP+ version illuminated touch buttons in STD version microprocessor temperature controller temperature sensor fail alarm (see page 56) power failure control system (see page 56) real time clock sound alarm (see page 56) self-check function (auto control) test results memory
Program possibilities
STD version (see page 25)
TOP+ version (see page 26-27)
Standard equipment solid door
RS 232 interface and USB port ; RS 232 or USB cable must be purchased additionally (RSK or USBK) stainless steel wire shelves
(1-6 pieces, depending on model) with slides set manufacturer’s test certificate at +105°C open door alarm air-flap diameter 37 mm for CL 15, 32, 53 and 115;
60 mm for CL 240, 400, 750 and 1000 operation manual in English and English menu
(other languages on request) over temperature protection according to DIN 12880:
2.0 class in STD version, 3.1 class in TOP+ version access port for external sensoe (diameter: 30 mm)
Options and accesories
(icon description see page 50)
INOX DIN
3.1
KG
KAFKA 9
SMS RS 422 RS 485
RS 232 RS 422 RS 485 USB
5 IQ
OQ
PQ
HEPA
for STD only for STD and TOP+
All POL-EKO-APARATURA products can be supplied with a calibration certificate issued by an accredited
Measurement Laboratory. Further information is available on pages 87-89 and our website: www.polekolab.com.
Drying ovens SL
Available options
SL*/A - door with viewing window o
(temperature limited to +250 C)
SL*/W - reinforced version
PW - reinforced shelf
Model SL 15 SL 32 SL 53 SL 115 SL 240 SL 400 SL 750 SL 1000
Parameter air convection chamber capacity door type warranty
1 temperature range controller interior housing overall dims* [mm] internal dims [mm] max shelf workload [kg] max unit workload [kg] nominal power [W] weight [kg] temperature resolution [°C] temperature fluctuation at +105°C [°C] voltage 50 /60 Hz [V] shelves fitted / max
3
2 over temperature protection
[l]
[cu ft]
[°C]
[°F] width height depth width height depth
-
PW version
-
W version
SLN
SLW
15
0,5
515
550
440
320
240
180
10
-
20
-
600
27
±0,7
±0,7
1/2 solid
595
630
495
400
320
240
32
1,1 natural (SLN) / forced (SLW) forced (SLW)
56
2
112
3,9
245
8,6
424
15 solid / door with viewing window (option) o
5 above ambient temperature...+300 o o
9 above ambient temperature...+572 o microprocessor with external LCD graphic display
749
26,4 acid-proof stainless steel according to DIN 1.4301
powder coated sheet / stainless steel linen finish (INOX/G)
595
700
600
395
395
360
655
845
700
460
540
450
815
1130
760
600
800
510
1015
1375
760
800
1040
510
1260
1590
865
1040
1200
600
10
-
30
-
1100
35
10
50
40
80
1600
50
10
50
60
120
10
100
90
300
2400
65 every 0,1
3000
126
10
100
120
300
3900
174
-
100
140
300
5400
260
±0,4
±0,4
±0,3
±0,3
±0,5
±0,5
±0,6
±0,5
-
±0,5
-
±0,5 class 2.0 according to DIN 12880 in STD / class 3.1 (option) / class 3.1 in TOP+ version
1/3
115/230
2/5 2/7 3/10 3/14
230/400
5/16
24 months
POL-EKO-APARATURA manufacturer
* - dims of power cable (50 mm) not included
1 working capacity of chamber can be smaller
2 - fluctuation measured in centre of the chamber
3 - "fitted" - number of shelves fitted, included in price, "max" - number of shelves which can be fitted
1005
35,5
-
100
-
300
5400
330
1260
2000
865
1040
1610
600
-
±0,5
6/22
37
SLWP Pass-through drying ovens
Application drying of painted and lacquered components drying/sterilizing of components between clean and dirty zones drying of components on production line
RS 232 and
USB port function indications
(at the rear) air-flap
(on the top) reinforced shelves
(option)
SLWP 115 STD
38
Pass-through drying ovens SLWP
Our main goal is to provide the customer with the highest satisfaction if it comes to flexibility of our offer and individual solutions. Passthrough drying ovens are based on standard models of drying ovens and are available in 2 versions: 112, 245 liters. Each request for such equipment is given professional advice and is discussed in details, therefore we kindly ask you to contact us on this matter.
Model SLWP 115 SLWP 240
Parameter air convection chamber capacity door type
1
[l]
[cu ft] temperature range
[°C]
[°F] controller interior housing overall dims* [mm] internal dims [mm] width height depth
max shelf workload [kg]
PW version
max unit workload [kg]
W version nominal power [W] weight [kg] temperature resolution [°C]
2 temperature fluctuation at +105°C [°C] over temperature protection voltage 50 /60 Hz [V] shelves fitted / max
3 warranty manufacturer width height depth
112
3,9 solid / door with viewing window (option) o
5 above ambient temperature...+300 o o
9 above ambient temperature...+572 o microprocessor with external LCD graphic display acid-proof stainless steel according to DIN 1.4301
powder coated sheet / stainless steel linen finish (INOX/G)
655
885
815
1180
815 700
460
540
480
600
800
540
10 10
50
60
120
100
90
300
2400
65
3000
126
±0,5 forced every 0,1
24 months
POL-EKO-APARATURA
* - dims of power cable not included
1 working capacity of chamber can be smaller
2 - fluctuation measured in centre of the chamber
3 - "fitted" - number of shelves fitted, included in price, "max" - number of shelves which can be fitted
245
8,6
±0,5 class 2.0 according to DIN 12880 in STD / class 3.1 (option)
115/230
2/7 3/10
39
SR Laboratory sterilizers
Application drying of laboratory glass hot-air sterilization required by pharmacies, dentists and prosthetics
Sterilizers have been equipped with a couple of additional functions that protect samples. They can sterilize at temperatures up to 250°C.
Pt 100 connection
(option)
(at the rear) additional
Pt 100 sensor
(option) fan stainless steel wire shelves and interior
SRW 240 STD INOX/G
40
Laboratory sterilizers SR
Advantages available in STD version chamber capacity from 56 to 1005 litres stainless steel interior housing: powder coated sheet or stainless steel linen finish (INOX/G) (depending on model) forced or natural air convection fan speed regulation from 0 to 100% for SLW 53, 115 and 240 from 10 to 100% for SLW 400, 750 and 1000 factory configured sterilizing programs automatic door locking system during operating time air-flap closed automatically after launching the program
SRW 750 and SRW 1000 models equipped with wheels
LCD graphic display illuminated touch buttons microprocessor temperature controller temperature sensor fail alarm (see page 56) power failure control system (see page 56) real time clock sound alarm (see page 56) self-check function (auto control) test results memory
Program possibilities
STD version (see page 25)
Standard equipment solid door
RS 232 interface and USB port ; RS 232 or USB cable must be purchased additionally (RSK or USBK) stainless steel wire shelves (2-6 pieces, depending on model) with slides set open door alarm access port for external sensor (diameter: 30 mm) air-flap diameter 37 mm for SR 53 and 115;
60 mm for SR 240, 400, 750 and 1000 manufacturer’s test certificate at +170°C operation manual in English and English menu (other languages on request) over temperature protection according to
DIN 12880: class 2.0
Options and accesories
(icon description see page 50)
INOX
DIN
3.1
KG
KAFKA
USB
SMS RS 422 RS 485
RS 232 RS 422 RS 485
9 5 IQ
OQ
PQ
HEPA
All POL-EKO-APARATURA products can be supplied with a calibration certificate issued by an accredited
Measurement Laboratory. Further information is available on pages 87-89 and our website: www.polekolab.com.
41
SR Laboratory sterilizers
Available options
SR*/A - door with viewing window
SR*/W - reinforced version
PW - reinforced shelf
42
Model SR 53 SR 115 SR 240 SR 400 SR 750 SR 1000
Parameter air convection chamber capacity door temperature range controller interior housing at +170°C [°C] manufacturer
1 overall dims* [mm] internal dims [mm] max shelf workload [kg] max unit workload [kg] nominal power [W] weight [kg] temperature resolution [°C] temperature fluctuation
2 over temperature protection voltage 50 /60 Hz [V] shelves fitted / max
3 warranty
[l]
[cu ft]
[°C]
[°F] width height depth width height
depth
PW version
SRN
SRW
10
50
40
1600
50
595
700
600
395
395
360
56
2 natural (SRN) / forced (SRW)
112
3,9
245 424 forced (SRW)
749
8,6 15 solid / door with viewing window (option)
26,4 o
5 above ambient temperature...+250 o o
9 above ambient temperature...+482 o microprocessor with external LCD graphic display acid-proof stainless steel according to DIN 1.4301
powder coated sheet / stainless steel linen finish (INOX/G)
655
845
700
460
540
450
815
1135
760
600
800
510
1015
1375
760
800
1040
510
1260
1590
865
1040
1200
600
±0,4
±0,4
2/5
10
50
60
2400
65
115/230
2/7
10
100
90
3000
126
10
100
120
3900
174
-
100
140
5400
260 every 0,1
±0,6
±0,5
±0,7
±0,5
-
±0,6
-
±0,6 class 2.0 according to DIN 12880 in STD / class 3.1 (option)
3/10
24 months
3/14
POL-EKO-APARATURA
230/400
5/16
* - dims of power cable (50 mm) not included
1 working capacity of chamber can be smaller
2 - fluctuation measured in centre of the chamber
3 - "fitted" - number of shelves fitted, included in price, "max" - number of shelves which can be fitted
1005
35,5
6/22
1260
2000
865
1040
1610
600
-
100
-
5400
330
-
±0,6
Application plant and fungus growing microorganisms and insects breeding
Photoperiodic system FOT
Photoperiodic system is an additional option available for thermostatic cabinets (ST) and cooled incubators (IL). Some applications require day and night simulation.
Photoperiodic system consists of illumination system placed in the door, on the side walls or in the ceiling of the unit (depending on model).
For each segment of temperature-time profile the user sets either light ON or OFF. Standard light is daylight. Other colours on request.
photoperiodic system
(in the door) photoperiodic system
(in the ceiling)
ILW 53 STD INOX/G + IL/FOT
43
FOT Photoperiodic system
Advantages day and night simulation software able to control illumination and temperature temperature range for “night” simulation: -10 ... +60°C for IL range and +3 ... +60°C for ST range
(depending on the selected temperature option) temperature range for "day" simulation: +5 ... +50°C for IL range and +10 ... +50°C for ST range selection of light colour, intensity and place of installation of the lamps (door, wall or shelves) - depending on model aluminum or stainless steel interior (depending on model)
"photoperiodic system" option does not feature illumination intensity temperature program priority in STD version temperature or time program priority in TOP+ version
Program possibilities
Temp. [°C] off
ON 100% (day)
100% 100%
T3 23°C
100%
OFF 0% (night)
100% 100%
ON 100% (day)
100%
TIME 3
T1 20°C TIME 1
TIME
0/1
T0 15°C
T2 10°C
D 20°C
2h 4h
- parameter set by User:
6h
TIME 1/2 TIME 2
10°C
9h 3h
D - adjustable start delay (1 min...99:59 h)
TIME 0/1...
- time to achieve set temperature point
(only for IL and TOP+ version)
- day/night simulation
4h
TIM
E2
/3
OFF 0% (night)
100% off
23°C
TIME 3/4 temperature
Time [h] 8h 3h
- fan speed control
TIME 1/2/3 - set time
T1/T2/T3 - set temperature
All POL-EKO-APARATURA products can be supplied with a calibration certificate issued by an accredited
Measurement Laboratory. Further information is available on pages 87-89 and our website: www.polekolab.com.
44
Photoperiodic system FOT
Photoperiodic system (*/FOT option) for single and double chamber thermostatic cabinets (ST)**
Option available for models
ST/FOT2
ST 1+
ST 1/1+
ST/FOT4
ST 2+
ST 2/2+
ST 350/350
ST/FOT6
ST 2+; ST 3+
ST 2/2+
ST 3/3+
ST 350/350
ST/FOT8
ST 5+
ST/FOT10
ST 500
ST 700
ST/FOT15
ST 1200 temperature range with photoperiod ON [°C] +10 ... +50 numbers of lamps on walls 2 4 6 8 10 no
15
( 3 colums with
5 pieces ) no adjustable illumination intensivity no no no
** for ST models with */FOT option, inner dims can be narrower by 4 cm on each side, “FOT" option is factory preinstalled.
There is no possibility to order it separately.
no
Photoperiodic system (*/FOT option) for cooled incubators (IL)**
Option IL/FOT2S IL/FOT3S IL/FOT5D IL/FOT6D ST/FOT10D available for models ILW 53 ILW 115 ILW 53
ILW 115
ILW 240
ST/FOT8D
ILW 240
ILW 400
ILW 750
ILW 750 temperature range with photoperiod ON [°C] numbers of lamps in door numbers of lamps in ceiling
-
2
-
3 adjustable illumination intensivity no
** “FOT" option is factory preinstalled. There is no possibility to order it separately.
no
5
no
+5 ... +50°C
6
no
8
no
10
no
Phytotron system (*/FIT option) features illumination intensity in the day/night simulation. Lighting elements are located in over-shelf panels, side walls or door
(depending on the selected model). Option available for thermostatic cabinets (ST 500-1450 in the TOP+ version and INOX interior), cooled incubators
(ILW 115-750 in TOP+ version) and climatic chambers.
Phytotron system (*/FIT option) for thermostatic cabinets (ST in TOP+ version with INOX interior) and climatic chambers (KK)**
ST/KK 1200/FIT P Option*** temperature range with photoperiod ON [°C] numbers of shelves with over-shelf lighting std/max lamps in door lamps in walls adjustable illumination intensivity
ST/KK 350/500/700/FIT DS ST/KK 500/700/FIT P ST/KK 500/700/FIT S
+10 ... +50°C
yes yes yes
+10 ... +50°C
1/3 no no yes
+10 ... +50°C
1/3 no yes yes
*** FIT DS - lamps in door and walls; FIT S - lamps in walls; FIT P - over-shelf panels
+10 ... +50°C
1/3 no no yes
Phytotron system (*/FIT option) for cooled incubators (IL in TOP+ version) and climatic chambers (KK)**
Îption
IL/KK 115/FIT P IL/KK 240/FIT P IL/KK 400/FIT P IL/KK 750/FIT P temperature range with photoperiod ON [°C]
+10 ... +50°C +10 ... +50°C +10 ... +50°C +10 ... +50°C numbers of shelves with over-shelf lighting std/max adjustable illumination intensivity
1/1 yes
1/3 yes
** “FOT" and "FIT" options are factory preinstalled. There is no possibility to order them separately.
1/3 yes
1/3 yes
45
KK
Climatic chambers
Climatic chambers with phytotron system
Application plant and fungus growing seeds germination microorganisms and insects breeding photostability tests food preservation experiments all research and tests that require stable temperature and humidity conditions (optionally light) tests of building materials
Climatic chambers can control humidity and temperature to provide a precise environment.
container for deionized water
(at the rear) powder painted housing light bulbs wheels
KK 350 TOP+ with KK 350/FIT
46
Climatic chambers
Climatic chambers with phytotron system
KK
Advantages available TOP+ version chamber capacity from 112 to 1460 litres stainless steel interior housing: powder coated sheet, stainless steel linen finish (INOX/G) (depending on model) forced air convection fan speed control from 10 to 100% heating and cooling systems to hold the set temperature regardless of ambient conditions thermal insulation - polyurethane foam automatic defrosting function (see page 65)
Program possibilities
TOP+ version (see page 26-27)
LCD touch screen in TOP+ version microprocessor temperature and humidity controller
(illumination intensity controller optionally – */FIT option) temperature and humidity sensor fail alarm (see page 56) power failure control system (see page 56) real time clock sound alarm (see page 56) self check function (auto control) test results memory
Temp. [°C] off off
70%
100%
90% (day)
90%
100%
40%
100%
0% (night)
50%
100%
70%
100%
80% (day)
90%
100%
TIME 3
30%
100% off off
T3 23°C
T1 20°C
TIME
0/1
TIME 1
TIM
E2
/3
T0 15°C D
20°C
T2 10°C
2h 4h
- parameter set by User:
6h
Climatic chambers time, temperature and humidity control temperature range: -10 to +60°C
TIME 1/2
3h
- set humidity
TIME 2
10°C
9h 4h
T1/T2/T3 - set temperature
- fan speed control TIME 1/2/3 - set time
D - adjustable start delay (1 min...99:59 h)
23°C
TIME 3/4 temperature
8h 3h Time [h]
- day/night simulation with illumination intensivity control
TIME 0/1...
- time to achieve set temperature point
Climatic chambers with phytotron system (*/FIT option) time, temperature, humidity and illumination intensity control illumination system: day and night simulation system with intensity control (every 10%):
- placed in the door and in the side walls for KK 350, 500 and 700 FIT DS
- placed in the side walls for KK 500 and 700 FIT S
- fixed as over-shelf panel with intensity control for KK 115, 240, 500, 400, 700, 750, 1200, 1450 FIT P temperature range - lighting off: -10 to +60°C temperature range - lighting on: +10 to +50°C light colour selection illumination power of up to 15 000 LUX per shelf (measured in geometrical center of the shelf placed 25 cm under illumination panel) program priority: temperature or time
47
KK
Climatic chambers
Climatic chambers with phytotron system
48
KK 350 TOP+ interior
Temperature and humidity sensors in the interior of a climatic chamber.
Standard equipment double door (solid outside, glass inside) wheels with brakes (except KK 115 and 240)
RS 232 interface and USB port ; RS 232 or USB cable must be purchased additionally (RSK or USBK) stainless steel wire shelves (2-6 pieces, depending on model) with slides set manufacturer’s test certificate at 25°C and 60% humidity operation manual in English and English menu (other languages on request) over temperature protection according to DIN 12880: class 3.3
container for deionised water container for waste water open door alarm access port for external sensor (diameter: 30 mm)
Options and accesories
(icon description see page 50)
INOX
90%
30%
5
SMS
IQ
OQ
PQ
H 2 O ////////
(((((
////////
RS 422 RS 485
RS 232 RS 422 RS 485 USB
KAFKA 9
POL-EKO Measurement Laboratory provides
Qualification procedures are required by pharmaceutical (also by veterinary pharmacy), cosmetic and biotechnology industries.
A complete qualification procedure consists of the following parts: installation qualification (IQ), operational qualification (OQ), and performance qualification (PQ).
All POL-EKO-APARATURA products can be supplied with a calibration certificate issued by an accredited Measurement Laboratory.
Further information is available on pages 87-89 and our website: www.polekolam.com.
Climatic chambers
Climatic chambers with phytotron system
KK
Chambers of this kind can be also produced according to customer’s requirements.
Available options
KK*/A - external glass door
FIT - phytotron system
Model KK 115 KK 240 KK 350 KK 400 KK 500 KK 700 KK 750 KK 1200 KK 1450
Parameter air convection chamber capacity
1
[l]
[cu ft] door type temperature range
[°C]
[°F] humidity range [%] phytotron system**
(available models)
FIT DS option
FIT P option
FIT S option controller interior housing overall dims [mm]
2 width height depth width internal dims [mm] max shelf workload [kg]*
height depth
-
PW version max unit workload [kg] nominal power [W]
-
W version
FIT version weight [kg] temperature resolution [°C] over temperature protection voltage 50/60 Hz [V] wire shelves fitted/max
3 over-shelf illumination panels fitted/max
3
FIT P option
115
3,9
-
YES
-
650
1200
710
460
540
450
10
50
60
120
1000
1200
90
2/7
1/1 max illumination power [lux] warranty
FIT version manufacturer
* - at even surface workload
** - FIT DS - lamps in door and walls
FIT P - over-shelf panel with lamps
FIT S - lamps in walls
FIT options can not be combined!
5000
-
YES
-
815
1490
770
600
800
510
10
100
90
300
1500
1800
140
245
8,6
335 424 forced
493 625 749
11,8 15 17,4 22 26,4 double (external solid, internal glass) / external glass (option)
-10...+60 (with phytotron ON +10...+50)
+14...+140 (with phytotron ON +50...+122 )
30...90 (with temperature range +10...+60)
645
2025
980
500
1340
500
10
100
YES
-
-
YES YES
YES
-
YES
YES
YES
YES
microprocessor with external display acid-proof stainless steel according to DIN 1.4301
-
YES
powder coated sheet / stainless steel (INOX/G)
1015 645 735 1260
1890 1740
770
800
1040
510
10
100
120
2025
1025
510
1525
635
20
100
2025
1075
600
1510
690
30
100
100 on request
1400
2600
125
300
2000
2300
185
100 on request
1600
2000
130 every 0,1
150 on request
1600
2000
170
870
1040
1200
600
10
100
140
300
2500
2900
275
3/11 3/14 class 3.3
115/230
3/11 3/11 5/16
-/1/2 1/3 1/3 1/2
1365
48,2
-
YES
-
1470
2000
1075
1310
1510
690
30
100
300 on request
2200
3000
220
1450
1970
1175
1340
1460
750
30
100
300 on request
2200
3000
230
2 x 3/11
1/3
1460
51,6
-
YES
-
2 x 3/11
1/3
3/10
1/2
10000 15000 15000 15000 15000
24 months
POL-EKO-APARATURA
1 - working capacity of chamber can be smaller
2 - overall dims for units without FIT DS/FIT D option, dims of power cable (50 mm) and shelf for water container
(250 mm) not included
3 - "fitted” number of shelves/panels fitted, included in price
6 "max" number of shelves/panels which can be fitted
15000 15000 15000
49
50
Options and accessories
Internal glass door
Order number: */C
External glass door
Order number: */A
Door with viewing window
Order number: */A
Wire shelf
Order number: */P
INOX
Stainless steel wire shelf
Order number: */P INOX
Perforated shelf
Order number: */PP
Glass shelf
Order number: */PS
Reinforced shelf
Order number: */PW
Stainless steel drawers
Order number: */SW
50...100%
Fan speed control
Order number: ST/CHL WENT
Photoperiodic system
Order number: */FOT
Phytotron system
Order number: */FIT
90%
30%
FIT panels independent control
Order number: FIT/R3
Door openings counter
Order number: LOD
Additional temperature sensor
Order number: PT 100
Internal socket
Order number: GNZ
HEPA
HEPA-fresh air filter
Order number: HEPA
Humidity measurement
Order number: PHR
Door lock
Order number: ZKL
Stainless steel cuvettes
Order number: KUW GN */*
Wheels
Order number: QLK*
SMS
H 2 O
SMS Info
Order number: SMS INFO
Container for deionized water
Order number: KK/Z
DIN
3.1
DIN
3.2
DIN
3.3
Table with wheels
Order number: */S or */S INOX
Over temperature protection
3.1 class according to DIN 12880
Order number: */3.1
Under temperature protection
3.2 class according to DIN 12880
Order number: */3.2
Over/under temperature protection
3.3 class according to DIN 12880
Order number: */3.3
Automatic defrosting function
Order number: * PLUS
KG
Low temperature version
Order number: */T
+70°C
Extended temperature range to +70°C
Order number: ST/70
Reinforced version
Order number: */W
Interior lighting
Order number: OWW
////////
Container for waste water
Order number: KK/K
(((((
////////
Low water level sensor
Order number: KK/CP
RS 422 interface (instead of RS 232)
Order number: RS422
RS 422
RS 485 interface (instead of RS 232)
Order number: RS485
RS 485
RS 232
RS 232 cable
Order number: RSK
RS 422
RS 422 cable
Order number: RSK/422
RS 485
RS 485 cable
Order number: RSK/485
USB
USB cable
Order number: USBK
Dot printer
Order number: EPSON
KAFKA
„Kafka” thermal printer
Order number: KAFKA
EasyLab-T PLUS software
Order number: EasyLab-T PLUS
5 Calibration at 5 points on 1 shelf
Order number: BRT/1P/L
9 Calibration at 9 points in the chamber
Order number: BRT/9/L
IQ
OQ
PQ
IQ, OQ, PQ qualification
Order number: IQ/OQ/PQ
Options and accessories
II
Parameters list
52
Features interior housing air convection plastic (ZLN 75, 145, 180) aluminum stainless steel (ST/CHL INOX; ZLN 85+; ZLN-T) powder coated sheet stainless steel (ST/CHL INOX; ZLN 85+, ZLN-T INOX) natural forced
0...100% (CL/SL 15...240, SR 53...240)
10...100% (CL/SL/SR 400-1000) fan speed control automatic fan shut down after completing the program automatic air-flap control (adjustable) system heating cooling defrosting function automatic defrosting function external display microprocessor temperature (humidity in KK; light with FIT and FOT options) controller graphic LCD
5.7” LCD touch screen temperature-time profile (humidity in KK ; light with FIT and FOT options )
1 segment
6 segment
multi-segment (max 10 000) temperature time (ST and IL in STD with FOT/FIT option) program priority cycle run of the program user programs memory start delay adjustable heating up time adjustable hold at set point (temperature; humidity in KK ; light with FIT and FOT options ) overview of set parameters during operating time recording min, average and max temperature (and humidity in KK) value for each segment (unavailable for ZLN 75, 145, 180) over/under temperature and humidity sound alarm (and humidity in KK) over/under temperature protection class according to DIN 12880 temperature sensor fail alarm (and humidity sensor fail alarm in KK) power failure control system real time clock sound alarm test results memory access port for external sensor (diameter: 30 mm) (option for ZLN 75, 145, 180) door lock (option for ZLN 75, 145, 180) open door alarm interface / port
Ethernet connection and remote control via Internet self-check function (auto control)
Administrator function access control via login e-mail info manufacturer’s test certificate (for: ZL at -10°C; CHL at +4°C; KK at +25°C and 60% rH; ST, CL, IL at +37°C; SL at +105°C; SR at +170°C)
24 months warranty
CE mark manufacturer’s certificate PN-EN ISO 9001
1
3
20
1 min...99:59 h date/time
1 min...99:59 h
1 min...99:59 h
1 min...999:59 h
1 min...31 days
RS 232
USB
40.
41.
42.
43.
44.
45.
46.
34.
35.
36.
37.
38.
39.
47.
48.
49.
50.
51.
52.
53.
54.
55.
56.
17.
18.
19.
20.
21.
22.
23.
24.
25.
26.
27.
28.
29.
30.
31.
32.
33.
13.
14.
15.
16.
8.
9.
10.
11.
12.
5.
6.
7.
1.
2.
3.
4.
Parameters list
ST
P* TOP+
CHL
P* TOP+
ZL
22.
23.
24.
25.
26.
27.
28.
29.
30.
14.
15.
16.
17.
18.
19.
20.
21.
31.
32.
33.
10.
11.
12.
13.
7.
8.
9.
3.
4.
5.
6.
1.
2.
42.
43.
44.
45.
46.
47.
48.
49.
50.
51.
52.
34.
35.
36.
37.
38.
39.
40.
41.
53.
54.
55.
56.
o o
1.0
o
3.3
o o
1.0
o
3.2
STD
CL
TOP+
IL SL
STD TOP+ STD TOP+
SR
STD
KK
TOP+
2.0
o
3.1
o o
2.0
o o
3.3
o
2.0
o
3.1
o
2.0
o
3.3
o
53
Parameter description
Interior aluminum
Aluminum interior is standard feature of ST/CHL ranges
(not in INOX version).
Housing stainless steel
Stainless steel interior is standard feature of CL/IL/SL/SR/KK ranges; ST/CHL INOX version and ZLN 85+, ZLN-T models.
54 powder coated sheet (structural painting)
Powder coated (structural painting) housing is standard feature of ST/CHL/ZL/CL/IL/SL/SR/KK (not in INOX version).
The standard housing is coated with a light grey colour (RAL 7035). The front panel is graphite (RAL 7016).
INDICATIONS
The type of interior and housing has been stated in product codes of the equipment.
Interior
ALU - aluminum
INOX - stainless steel stainless steel - INOX/G
Stainless steel linen finish (INOX/G) housing is standard feature of ST/CHL/ZL/ CL/IL/SL/SR/KK INOX/G models.
Housing:
PAINT - powder coated sheet
INOX/S - stainless steel polished
INOX/G - s tainless steel linen finish
In a standard version (for ST/CHL – aluminum interior, powder coated housing; for CL/IL/SL/SR/KK – stainless steel interior, powder coated housing), the indication interior/housing in the product code has been omitted.
stainless steel - INOX/S
Stainless steel polished (INOX/S) housing is standard feature of ST/CHL/ZL INOX/S models.
EXAMPLES:
ST 1+ thermostatic cabinet, aluminum interior, powder coated housing: ST 1+
ST 1+ thermostatic cabinet, stainless steel interior, polished stainless steel housing: ST 1+ INOX INOX/G
CLW 53 STD incubator, powder coated housing: CLW 53 STD
CLW 53 STD incubator, stainless steel linen finish housing: CLW 53 STD INOX/G
(for CL/IL/SL/SR/KK the interior is always made of stainless steel)
Parameter description
Air-flap control
BACK FRONT air-flap in STD version
The CL/SL/SR ranges feature an automatic air-flap control (adjustable).
Diameter of the air-flap:
- for 15, 32, 53, 115 models - 37 mm
- for 240, 400, 750, 1000 models - 60 mm
External display
LCD graphic display
The ST/CHL/ZL range feature an LCD graphic display with illuminated touch buttons.
5.7” LCD touch screen
The ST/CL/IL/SL/KK ranges in TOP+ version feature a 5,7” LCD touch screen.
LCD graphic display
The CL/IL/SL/SR ranges in STD version feature an LCD graphic display with illuminated touch buttons.
55
Parameters description
56
Defrosting function
This is standard feature of ST/CHL ranges.
Defrosting is performed automatically but is has to be chosen manually by the user at the most suitable time, e.g. when there are no samples in the chamber. It consists in temperature increase in the chamber.
Over/under temperature (and humidity in KK) sound alarm
It is possible to set temperature (and humidity in KK) offset in the program menu. If the temperature or humidity (humidity in KK only) exceeds or falls beyond the set point, the alarm will go off and the
“ALARM” message will appear on the display.
Temperature (and humidity in KK) sensor fail alarm
If the sensors are not working correctly, an error message appears on the display.
Sound alarm
This function sounds the alarm at a time specified by the user.
E-MAIL info
This is standard feature of any product in
TOP+ version.
It consists of e-mail notifications about set temperature (and humidity in KK range) values exceedings (NOT about power shortages), on max 2 e-mail addresses.
Ethernet connection and remote control via internet
This is standard feature of ST/CHL/CL/IL/SL/KK ranges in TOP+ version. The equipment can be controlled and monitored via internet. It is also possible to connect several units in-line and control them from one PC.
Test results memory
All POL-EKO-APARATURA products are equipped with this feature. This feature is test results memory module that allows to safe data in the internal memory of the unit and brows it any time. To enable logging for units which are not equipped with internal memory, they need to be connected to a PC computer at all times.
All data can be sent to a PC though the EasyLab-T PLUS software with an RSK or USBK cable.
Acces port for external sensor
Access port for external sensor
This is standard feature of all product ranges except
ZLN 75, 145, 180 models. Access port diameter: 30 mm, placed in upper part of the left chamber side.
Temperature or time program priority
For each program the user is able to choose either temperature or time priority. Temperature priority consists in: the unit achieves set temperature first and then starts time countdown.
Time priority consists in: the unit starts the process of achieving set temperature simultaneously with time being countdown.
Over/under temperature protection system
All POL-EKO-APARATURA products (except ZL range) are equipped with over temperature protection 1.0 class according to DIN 12880.
Depending on the product range and version, various temperature protections are available. See details in the products parameters tables and on pages 52-53.
Power failure control system
A temporary power failure while running the power is unnoticeable due to autoresume function but the time of power shortage is displayed on the screen.
Administrator function
This is standard feature of ST/CHL/CL/IL/SL/KK ranges in TOP+ version. It allows to manage user accounts as support to GLP.
7 days programming
This is standard feature of ST/CHL/CL/IL/SL/KK ranges in TOP+ version.
It allows user to set program for each day of the week (e.g. Monday, 9.00-15.00, at 37°C)
Open door alarm
All POL-EKO-APARATURA products are equipped with this feature. The user sets the time (1 to 5 min) after which the alarm goes off and a message appears on the display when the door is open.
RS 232 interface / USB port
All POL-EKO-APARATURA products are equipped with RS 232 and USB ports. This feature enables
PC connection, data archiving and transmition
(e.g. temperature or humidity values). For this, it is necessary to purchase a connection cable
(RSK or USBK) and EasyLab-T PLUS software.
Only TOP+ version units feature direct USB flash drive data saving facility. Basic and STD version units allow a USB-PC transmission for data archiving.
Temperature (and humidity in KK) calibration at 3 points
This is standard feature of ST/CHL/CL/IL/SL/KK ranges in TOP+ version. Thanks to this feature the user has got a possibility to calibrate the chamber at 3 points of temperature and additionally at 3 points of humidity in KK range (only TOP+ version).
Options and accessories
Options and accessories
P
6
ST
TOP+ P
6
CHL
TOP+
ZL internal glass door
1, 5 external glass door 1, 8 door with viewing window
4, 7 internal socket
1, 5 interior lighting
1, 4, 5 door lock
1, 12 wire shelf
1 stainless steel wire shelf
1, 9 perforated shelf
1 glass shelf
3 reinforced shelf extended temperature range to +70°C
1 reinforced version 12 low temperature version 1, 10 photoperiodic system - FOT
1 phytotron system - FIT
11 automatic defrosting function over temperature protection system
1 according to DIN 12880 stainless steel cuvettes stainless steel drawers humidity measurement
5 door openings counter 1 fan speed control additional Pt 100 temperature sensor
HEPA - fresh air filter
RS 422 interface (instead of RS 232) 1
RS 485 interface (instead of RS 232)
1 wheels table with wheels
2
RS 232 cable
1
RS 422 cable
1
RS 485 cable
1
USB cable 1
EasyLab-T PLUS software dot printer
“Kafka” thermal printer calibration at 9 points
1 chamber calibration at 5 points on 1 shelf
1
IQ, OQ, PQ qualification
1
SMS Info 1, 12 container for deionised water container for waste water water level sensor
FIT panels independent control
11 s
3.3
s s
(3.3) s s
3.2
s s
(3.2) s
* - model (e.g. ST 1+, IL 53, ZL 75)
** - over temperature protection system (e.g. 3.1)
1 - for double chamber units, the option available for both chambers separately
2 - unavailable for 400, 500, 700, 750, 1200, 1000 and 350/350 models, ST/CHL 4+ and 5+
3 - only for ZL 75, 145, 180
4 - in case of SL range, maximum temperature is reduced to +250
ºC
option unavailable for CL/SL 15/32
5 - in case of CL/IL in TOP+ version, maximum temperature is reduced to +70
ºC s
CL
STD TOP+ STD
IL
TOP+ STD
SL
TOP+
SR
STD
KK
TOP+ s
3.1
s s
(3.1) s
3.3
s s
(3.3) s
3.1
s s
(3.1) s
3.1
s s s
(3.3) s
Order no.
*/C
*/A
*/A
GNZ
OWW
ZKL
*/P
*/P INOX
*/PP
*/PS
*/PW
ST/70
*/W
*/T
*/FOT
*/FIT
*PLUS
*/**
KUW GN */*
*/SW
PHR
LOD
ST/CHL WENT
PT100
HEPA
RS422
RS485
QLK*
*/S, */INOX
RSK
RSK/422
RSK/485
USBK
EasyLab-T PLUS
EPSON
KAFKA
BRT/9/L
BRT/5/L
IQ/OQ/PQ
SMS INFO
KK/Z
KK/K
KK/CP
FIT/R3
6 - basic version of thermostatic and laboratory refrigerators
7 - for KK 115, 240, 400, 750
8 - for KK 500, 700, 1200
9 - for ZLN 85+
10 - for CHL 500, 700, 1200, 1450, 350/350
11 - thermostatic cabinets ST 500, 700, 1200, 1450
12 - except ZL 75, 145, 180 s - standard equipment
- available option
57
58
Options and accessories
Internal glass door
This is standard feature of KK models.
This is an additional option available for ST/CHL/CL/IL ranges.
Order number: */C.
In case of CL/IL in TOP+ version, maximum temperature is reduced to +70ºC.
External glass door
This is an additional option available for ST/CHL ranges and for KK 500,700, 1200, 1450 models.
Order number: */A (option factory preinstalled).
Door with viewing window
This is an additional option available for CL/IL/SL/SR ranges, except CL/SL 15, 32 and for KK 115, 240, 400, 750 models.
Order number: */A (option factory preinstalled).
In case of SL range, maximum temperature is reduced to +250°C.
Internal socket
This is an additional option available for ST/CHL/CL/IL ranges. In case of CL/IL in STD and TOP+ version, maximum temperature is reduced to +70°C with internal socket.
Order number: GNZ (option factory preinstalled).
Internal socket allows to plug in additional equipment inside the chamber, e.g. laboratory shaker.
Max socket peak load 200W.
Interior lighting
This is an additional option available for all product ranges.
Order number: OWW (option factory preinstalled).
Interior lighting features 1 light point. The user switches it on with enter button located in the front panel.
This option does not allow day/night simulation (see FIT and
FOT options). If a unit is equipped with the open door sensor the light can switch on when the door opens. Max working temperature of the unit is reduced to +70°C, for SL/SR ranges to +250°C.
Options and accessories
Door lock
This is standard feature of ST/CHL and ZL 85+ and ZLN-T.
This is an additional option available for all product ranges.
Order number: ZKL.
Door lock allows to lock the door with a key.
Wire shelf
This is standard equipment – wire shelf in
ST/CHL ranges with slides set.
This is an additional option available for
ST/CHL ranges.
Order number: */P.
Wire shelf is provided with slides set.
Perforated shelf
This is standard equipment in ZLN-T models.
This is an additional option available for
ST/CHL/CL/IL/SL/SR/KK ranges and ZLN 85+ model.
Order number: */PP.
Perforated shelf is provided with slides set.
Glass shelf
This is standard equipment in ZLN 75, 145, 180 models.
This is an additional option available for ZLN 75, 145, 180 models.
Order number: */PS.
Stainless steel wire shelf
This is standard equipment in CL/IL/SL/SR/KK ranges, ZLN 85+ and in ST/CHL ranges with stainless steel interior. This is an additional option available for ST/CHL/CL/IL/SL/SR/KK ranges and ZLN 85+.
Order number: */P INOX.
INOX wire shelf is stainless steel wire shelf with slides set.
INOX
59
60
KG
Options and accessories
Reinforced shelf
This is standard equipment for the CL/IL/SL 750 and 1000 models and all CL/IL/SL models in the reinforced version
(order number: */W). This is an additional option available for CL/IL/SL/SR ranges, ST/CHL/KK 500, 700, 1200 models, and ZLN-T models.
Order number: */PW.
Reinforced shelf is provided with slides set. Reinforced shelf allows to store heavy loads on it. For maximum shelf workloads and maximum unit workloads look in the table with parameters for certain ranges of equipment (reinforced shelf - PW; reinforced version of the unit - W).
Reinforced version
This is standard equipment for the CL/SL 1000 models.
This is an additional option available for CL/IL/SL ranges and ZLN-T 200, 300 models.
Order number: */W (option factory preinstalled).
Reinforced version of equipment allows to store heavy loads in the chamber. It consists of reinforced construction of the chamber and reinforced shelves. In this way we prevent damage to the unit caused by heavy loads. For maximum shelf workloads and maximum unit workloads look in the table with parameters for certain ranges of equipment (reinforced shelf - PW; reinforced version of the unit - W).
When a unit in reinforced version is purchased, the reinforced shelves are supplied instead of wire shelves.
Stainless steel drawers
This is an additional option available for ST/CHL/CL/IL/SL/SR/KK ranges and ZLN 85+, ZLN-T models.
Order number: */SW.
Stainless steel drawer is provided with slides set and can be ordered instead of wire shelf. Drawer is perforated and its depth must be specified in the order.
Different kinds of perforation and depths are available on request.
Stainless steel cuvettes
This is an additional option available for any chamber.
Order number: KUW.
Stainless steel cuvettes can be placed on the shelves.
Different sizes available.
Options and accessories
Photoperiodic system
This is an additional option for the ST/IL ranges.
Order number: */FOT (option factory preinstalled).
The illumination system can simulate day and night.
Depending on model, it is placed either in the door, or on side walls, or in the ceiling of the chamber.
For the IL equipment, the temperature range when the illumination is switched off can be -10…+60°C and
+5…+50°C when it is switched on. For the ST equipment, the temperature range when the illumination is switched off can be +3…+60°C (depending on selected temperature option) and +10…+50°C when it is switched on. Standard light colour: daylight, other colours available on request.
Phytotron system
This is an additional option for the KK range, IL in TOP+ version and ST 500-1450 in TOP+ INOX version.
Order number: */FIT (option factory preinstalled).
The illumination system can simulate day and night and control illumination smoothly by every 10%.
FIT P and FIT PANEL options (over-shelf panels with illumination) are available for: KK/IL 115, 240, 400,
750 and ST 500-1450. FIT DS option (illumination system placed in the side walls and in the door of the chamber) is available for: KK 350 and ST/KK 500-700. FIT S option
(illumination system placed in the side walls of the chamber) is available for: ST/KK 500-700. For KK/IL models, maximum temperature range when the illumination is switched off is -10…+60°C and +10...+50°C when it is switched on.
For ST models, maximum temperature when the illumination is switched off is +3…+60°C and +3…+50°C when it is switched on. Illumination can be 5 000-15 000 LUX depending on model (see page 49). Standard light colour: daylight; other colours available on request.
Additional Pt 100 temperature sensor
This is an additional option available for CL/IL/SL/SR/KK ranges and ST/CHL in TOP+ version.
Order number: PT 100 (option factory preinstalled).
This option consists of additional temperature sensor, internalsocket, access port and value browse possibility in the display. The sensor may be supplied with a calibration certificate. It enables more detailed temperature monitoring during incubation. Additional Pt 100 sensor also allows to choose in the unit's menu which sensor should be used as primary and which as secondary. This way unit can work according to the sample temperature in which additional
Pt 100 sensor is placed.
Wheels
This is standard equipment provided for ST/CHL 1200, 1450; and CL/IL/SL/SR 750, 1000; IL 400; and KK models (except
KK 115 and 240). This is an additional option available for all product ranges.
Order number: QLK*(option factory preinstalled).
For units of bigger capacities wheels are indispensable equipment.
However some users may require wheels in smaller units.
This is why we are able to equip each unit with the wheels.
61
62
KAFKA
HEPA
Options and accessories
EasyLab-T PLUS software
This is an additional option available for all product ranges.
Order number: EasyLab-T PLUS.
All thermostatic products manufactured by POL-EKO-APARATURA are equipped with an RS232 interface and USB port and can be connected to a PC. The EasyLab-T PLUS software allows to record temperature and/or humidity values. The data (day, time, temperature/humidity values) can be transferred to a PC computer in a tabular form that can be used to generate charts. It can be further stored or exported to the .xls format. It is necessary to purchase an RS232 or USBK interface cable along with the software.
Dot printer
This is an additional option available for all product ranges.
Order number: EPSON.
DOT printer enables current temperature, time and date printing.
Printing interval is to be set in the unit’s menu. There must be RSK
(PC and printer connection cable) purchased along with DOT printer.
“Kafka” thermal printer
This is an additional option available for all product ranges.
Order number: KAFKA.
“Kafka” thermal printer enables current temperature, time and date printing. Printing interval is to be set in the unit’s menu. There must be RSK (PC and printer connection cable) purchased along with “Kafka” printer.
HEPA-fresh air filter
This is an additional option available for CL/SL/SR ranges.
Order number: HEPA (option factory preinstalled).
HEPA filter is installed at the air inlet to the chamber.
Options and accessories
Table with wheels
This is an additional option available for ST/CHL 1+, 2+ ,3+ ;
ZLN 75, 85+, SL/CL 15, 32; CL/IL/SL/SR 53, 115, 240 models.
Order number: */S (powder painted) or */S INOX (stainless steel).
Table with wheels provides you with the highest comfort of using our products. We offer a wide range of tables equipped with wheels.
Different sizes of the tables are available on request.
The user can choose the most suitable height.
RS 232 cable
This is an additional option available for all product ranges.
Order number: RSK.
RS 232 cable is a connection cable for PC or printer.
Standard cable length: 5m.
RS 232
RS 422
RS 422 cable
This is an additional option available for all product ranges.
Order number: RSK/422.
RS 422 is a connection cable for PC.
Standard cable length: 5m.
RS 485
RS 485 cable
This is an additional option available for all product ranges.
Order number: RSK/485.
RS 485 is a connection cable for PC.
Standard cable length: 5m.
USB cable
This is an additional option available for all product ranges except TOP+ version.
Order number: USBK.
The cable is useful to connect the unit to a PC computer.
Standard cable length: 5m.
USBK
63
64
H 2 O
(((((
////////
RS 422
RS 485
////////
Options and accessories
RS 422 interface / RS 485 interface
These are additional options available for all product ranges.
Order number: RS422 or RS485 (option factory preinstalled).
This option consists of a converter from RS 232
(standard built in the device) to RS 422 or RS 485.
It allows to plug a few pieces of equipment in-line.
Container for waste water
This is standard equipment for KK range.
This is an additional option available for KK range.
Order number: KK/K.
This is a plastic container for waste water.
The container is recommended when it is not possible to connect the unit to a drain system.
Container for deionised water
This is standard equipment for KK range.
This is an additional option available for KK range.
Order number: KK/Z.
Plastic container for deionised water is standard equipment for all climatic chambers (KK), however also an additional container can be ordered. The container is recommended when it is not possible to connect the unit to a deionizer.
Low water level sensor
This is an additional option available for the KK range.
Order number: KK/CP (option factory preinstalled).
An alarm goes off when the water level is low.
Options and accessories
90%
30%
FIT panels independent control
This is an additional option available for ST 500-1450
KK/IL ranges in TOP+ version with FIT option (in case of phytotron system consisting of 3 over-shelf illumination panels).
Order number: FIT/R3 (option factory preinstalled).
This option consists in possibility to control each of 3 over-shelf illumination panels independently
(e.g. one panel intensity can be set at 100%, the other one at 50%).
Automatic defrosting function
This is standard feature of KK range.
This is an additional option available for ST/CHL/IL ranges.
Order number: * PLUS (option factory preinstalled).
Defrosting in this case is programmable (the user sets periodicity and duration). This function is carried out while the unit is working. This advanced technology holds the temperature stable, allowing only a minor increase in the chamber (considerably higher temperature rise is caused by opening the door).
+70°C
Extended temperature range ST/70
This is standard feature of ST in TOP+ version.
This is an additional option available for ST range.
Order number: ST/70 (option factory preinstalled).
It consists of extended temperature range up to +70°C (standard temperature range in ST ranges is +3°C…+40°C).
Low temperature version
This is an additional option available for
CHL 500, 700, 1200 and CHL 350/350 models and IL range.
Order number: */T (option factory preinstalled).
It extends temperature range down to -10°C
(standard temperature range starts from 0°C).
SMS
SMS info
This is an additional option available for all product ranges (except ZLN 75, 145 180 models).
Order number: SMS INFO (option factory preinstalled).
It consists in SMS notifications about set temperature
(and humidity in KK range) values exceedings as well as about power shortages, on max 2 mobile phone numbers.
DIN
3.1
Over temperature protection
3.1 class according to DIN 12880
This is a standard function for the CL/SL equipment in the TOP+ version, and optional for the CL/SL/SR ranges in the STD version.
Order number: */3.1 (option factory preinstalled).
It features a sample protection function: the user can set the protection temperature and once it has been exceeded, the program will cut off the heaters.
When the temperature falls down below the set limit, the unit will resume operating automatically.
DIN
3.2
Under temperature protection
3.2 class according to DIN 12880
This is standard function for CHL range in TOP+ version and optional for CHL in basic version.
Order number: */3.2 (option factory preinstalled).
It features a sample protection function: the user can set the protection temperature and once it has been exceeded, the program will cut off the compressor.
When the temperature goes above the set limit, the unit will resume operating automatically.
DIN
3.3
Over/under temperature protection
3.3 class according to DIN 12880
This is a standard function for the KK, ST and IL ranges in the TOP+ version, and optional for ST in basic version and IL in STD version.
Order number: */3.3 (option factory preinstalled).
It features a sample protection function: the user can set the over/under protection temperature and once it has been exceeded, the program will cut off the heaters or compressor. When the temperature goes back to the permitted range, the unit will resume operating automatically.
65
66
Options and accessories
Humidity measurement
This is an additional option available for CL/IL ranges and ST/CHL in TOP+ version.
Order number: PHR (option factory preinstalled).
This option is not humidity control but humidity measurement. The user can browse humidity values in the unit’s display.
50...100%
Fan speed control
This is an option for the ST/CHL range in basic version which allows to control the fan speed from 50% up to 100%.
Order number: ST/CHL WENT
It can be set for each program segment separately.
Door openings counter
This is an additional option available for all product ranges.
Order number: LOD (option factory preinstalled).
This function counts door openings (how many timesthe door was opened during the program run).
9
Calibration at 9 points in the chamber
This is an additional option available for all product ranges.
Order number: BRT/9/L (option factory preinstalled).
This option consist of measurements performed at 9 points of the chamber (corners + geometric center) at temperature according to customer's specification;
ACCREDITATED TEMPERATURE RANGE IS -20°C...+200°C.
5
Calibration at 5 points on 1 shelf
This is an additional option available for all product ranges.
Order number: BRT/1P/L (option factory preinstalled).
This option consists of measurements performed at 5 points on the unit's shelf (corners + geometric center) at temperature according to customer's specification; ACCREDITATED TEMPERATURE
RANGE IS -20°C...+200°C.
IQ
OQ
PQ
IQ, OQ, PQ qualification
This is an additional option available for all product ranges.
Order number: IQ/OQ/PQ (option factory preinstalled).
It consists of: operation, qualification and performance qualification tests performed in our factory; empty installation qualification documentation; enclosed qualification book.
For more information see page 89.
Temperature protection
Over temperature protection class 1.0 and class 2.0 according to DIN 12880
TB action
Temp. [°C]
100°C
70°C
TWB action
Over temperature protection class 1.0 (TB) the max temperature.
Over temperature protection class 2.0 (TWB)
- programmable set temperature
Unit's malfunction*
Unit's malfunction*
* triac, temperature sensor,
A/D converter, program failure, unintentional change of set parameters
20°C
Time [s]
Over temperature protection class. 1.0 to DIN 12880 is a standard function for the ST/CHL/CL/IL/SL/SR/KK equipment.
It is factory set at approx. 10°C above the max temperature.
Over temperature protection class 2.0 to DIN 12880 is a standard function for the CL/IL/SL/SR equipment in the STD version.
It features a sample protection function: the user can set the protection temperature and once it has been exceeded, the program will cut off the heaters. To resume operating, the user has to switch the unit off and turn it on again.
Over temperature protection class 3.1 according to DIN 12880
TB action
Temp. [°C]
100°C
TWW action
Over temperature protection class 1.0 (TB) factory set at approx 10 ° C above the max temperature.
Over temperature protection class 2.0 (TWB)
- programmable set temperature 70°C
Unit's malfunction*
Unit's malfunction*
* triac, temperature sensor,
A/D converter, program failure, unintentional change of set parameters
20°C
Time [s]
Over temperature protection class 3.1 to DIN 12880 is a standard function for the CL/SL equipment in the TOP+ version, and optional for the CL/SL/SR ranges in the STD version. Order number: */3.1 (option factory preinstalled).
It features a sample protection function: the user can set the protection temperature and once it has been exceeded, the program will cut off the heaters. When the temperature falls down below the set limit, the unit will resume operating automatically.
67
Temperature protection
68
Under temperature protection class 3.2 according to DIN 12880
Temp. [°C]
* triac, temperature sensor,
A/D converter, program failure, unintentional change of set parameters
Unit's malfunction*
TWW action
20°C
15°C set temperature
Under temperature protection class 3.2 (TWW)
- programmable
Time [s]
Under temperature protection class 3.2 to DIN 12880 is a standard function for CHL in TOP+ version and optional for CHL range in basic version. Order number: */3.2 (option factory preinstalled). It features a sample protection function: the user can set the protection temperature and once it has been exceeded, the program will cut off the compressor. When the temperature goes above the set limit, the unit will resume operating automatically.
Over/under temperature protection class 3.3 according to DIN 12880
TB action
Temp. [°C]
100°C
TWW action
Over temperature protection class 1.0 (TB) factory set at approx 10 ° C above the max temperature.
Over temperature protection class 3.3 (TWW)
- programmable set temperature 70°C
Unit's malfunction*
Unit's malfunction*
TWW action
* triac, temperature sensor,
A/D converter, program failure, unintentional change of set parameters
20°C
15°C set temperature
Under temperature protection class 3.3 (TWW)
- programmable
Time [s]
Over/under temperature protection class 3.3 to DIN 12880 is a standard function for the KK, ST and IL ranges in the TOP+ version, and optional for ST in basic version and IL in the STD version. Order number: */3.3 (option factory preinstalled). It features a sample protection function: the user can set the over/under protection temperature and once it has been exceeded, the program will cut off the heaters or compressor. When the temperature goes back to the permitted range, the unit will resume operating automatically.
Steel laboratory furniture
III
70
Steel laboratory furniture
Fume hoods
Steel laboratory furniture
underbench cabinet for C-type frame of 750 mm height
underbench cabinets with castors; height 480 mm
underbench cabinet for C-type frame of 900 mm height
underbench cabinets with service supply for C-type frame of 900 mm height
underbench cabinets with castors; height 630 mm
octagonal underbench cabinet, for C-type frame of 900 mm height
wall mounted shelves with the door
wall mounted open shelves
C-type frame of 750 mm and 900 mm height
accessories: columns, shelves
innovative designing
77
77
78
79
80
75
76
76
77
71
73
74
74
Fume hoods
Advantages steel construction work safety conformity with PN EN 14175
AUTO and MANUAL PROTECT functions for window closing – saving the air you breathe wide range of working chamber materials for use with non-flammable materials
FH
Auto-protect function (option) features a motion sensor, window level sensor and electrical opening/closing system. If the sensor does not detect any motion in front of the fume hood, the system will close the window automatically. Control buttons are also a part of the system, as well as manual sliding window detector and light barrier that stops the motion of the window if an obstacle has been spotted.
Q-DYNAMIC 1200
Q-Flow-Air Flow Sensor (standard equipment):
controlling functions with alarm status indicated by visual and acoustic signals
alarm indicating that the sash window exceeds the maximum opening height digital LED display showing current airflow in m³/h
controlled and indicated alarm status
power failure alarm
power suspension system
control of fume cupboard illumination
71
FH Fume hoods
Parameter overall dims [mm] working space dims [mm] recommended airflow [m³/h] maximum power consumption [kW] voltage 50 Hz [V] exit air stub pipe dimensions [mm] water supply sewage connection [mm] standard airflow sensor of the Q-Flow type
(or Schneider Elektronik of the FM 100 type) width height depth width height depth airflow sensor with non-gradual airflow regulation, depending on the extent of sash window opening
(Schneider Elektronik of the iCM type) optional motion detector with automatic closing device the window closes automatically when the fume cupboard is not in use (within defined time) optional motion detector with manual window closing upon time limit acoustic and visual signals indicate that the window should be closed electrical insulation class permissble work board load [kPa] warranty
Laboratory fittings covered with the chemically resistant polyamid 11 coating:
average coating thickness 250-300 microns
melting point 184-186°C
inflammability self-extinguishing
shore hardness D up to 20°C
colour grey RAL 7035
The sash window:
metal
in an aluminum frame
in an MDF frame painted with chemically-resistant epoxy paint
with the system of obstacle detection
safety at every move
Utilities:
gas
compressed air
vacuum
demineralised water
nitrogen
others at client`s request
the possibility of revision so as to supply individual valves
Model
Q-Dynamic 1200 Q-Dynamic 1500 Q-Dynamic 1800
1280
2325...2572
942
1070
1100
750
450...650
Ø 160
1580
2325...2572
942
1370
1100
750
650...850
3,5
230
Ø 200
G ½”
Ø 50
1880
2325...2572
942
1670
1100
750
850...1350
Ø 250 yes, confomity with standards PN-EN 14 175-2,3 optional PN-EN 14 175-2,3 optional auto protect function
2 kPa optional manual protect function class I
4 kPa
24 months
5 kPa
72
Steel laboratory furniture MM
Steel laboratory furniture
The furniture that we offer is remarkable for its mechanical resistance. It has been constructed from highest quality materials. There is a wide selection of standard frames, cupboards, panels and worktops, but customized solutions are available too.
Advantages
steel construction featuring easy leveling, based on C-type frames
2 available versions: 750mm frame height – to assist sitting and 900mm to assist standing work large selection of panels, sinks and fittings made of various materials wide range of worktops (e.g. laminate, trespa, durcon) free configuration of frames to form segments work safety according to PN-EN 13150 norm professional advice, designing and visualization are a part of the offer
Standard equipment
made of galvanized steel, epoxy powder coated - chemical resistant, grey colour RAL7035 underbench cabinets for C-type frames
BLUM slides and hinges
Indications
SZ – underbench cabinet
SZK – underbench cabinet with castors
SZM – underbench cabinet with service supply
SZO – octagonal underbench cabinet
SZW – wall mounted shelves with the door
SZWR – wall mounted open shelves
S – frame
450-900 – cupboard width – frame height
2D – double wing door
1DP/1DL – single left/right door
1DP1S/1DL1S - single left/right door, one drawer
2D1S – double wing door, one drawer
2D2S – double wing door, two drawers
3S – three drawers
4S – four drawers
73
MM Steel laboratory furniture
underbench cabinets for C-type frame of 750 mm height
underbench cabinets for C-type frame of 750 mm height, cabinet height 480 mm, cabinet depth 500 mm
Cabinet 450-750 1D Cabinet 600-750 1D Cabinet 900-750 2D Cabinet 1200-750 2D model width door drawers order number
450 mm
1 Right (P)/Left (L)
-
MP_SZ 450-750 1DP
MP_SZ 450-750 1DL
Cabinet 450-750 3S model
74 width door drawers order number
450 mm
-
3
MP_SZK 450-750 3S
600 mm
1 Right (P)/Left (L)
-
MP_SZ 600-750 1DP
MP_SZ 600-750 1DL
Cabinet 600-750 3S
900 mm
2
-
MP_SZ 900-750 2D
Cabinet 900-750 3S width door drawers order number
450 mm
-
3
MP_SZ 450-750 3S
600 mm
-
3
MP_SZ 600-750 3S
900 mm
-
3
MP_SZ 900-750 3S
underbench cabinets with castors; height 480 mm
underbench cabinets with castors, height 480 mm, depth 500 mm
Cabinet 450-750 1D with castors Cabinet 600-750 1D with castors model width door drawers order number
450 mm
1 Right (P)/Left (L)
-
MP_SZK 450-750 1DP
MP_SZK 450-750 1DL
Cabinet 450-750 3S with castors
600 mm
1 Right (P)/Left (L)
-
MP_SZK 600-750 1DP
MP_SZK 600-750 1DL
Cabinet 600-750 3S with castors model
600 mm
-
3
MP_SZK 600-750 3S
1200 mm
2
-
MP_SZ 1200-750 2D
Underbench cabinet 600-900 4S
Steel laboratory furniture
underbench cabinets for C-type frame of 900 mm height
Underbench cabinets with service supply for C-type frame of 900 mm height, cabinet height 630 mm, cabinet depth 500 mm
Cabinet 450-900 1D Cabinet 600-900 1D Cabinet 900-900 2D Cabinet 1200-900 2D model
MM width door drawers order number
450 mm
1 Right (P)/Left (L)
-
MP_SZ 450-900 1DP
MP_SZ 450-900 1DL
Cabinet 450-900 4S model
600 mm
1 Right (P)/Left (L)
-
MP_SZ 600-900 1DP
MP_SZ 600-900 1DL
Cabinet 600-900 4S
900 mm
2
-
MP_SZ 900-900 2D
Cabinet 900-900 4S
1200 mm
2
-
MP_SZ 1200-900 2D width door drawers order number
450 mm
-
4
MP_SZ 450-900 4S
Cabinet 450-900 1D1S model
600 mm
-
4
MP_SZ 600-900 4S
Cabinet 600-900 1D1S
900 mm
-
4
MP_SZ 900-900 4S
Cabinet 900-900 2D1S Cabinet 1200-900 2D1S width door drawers order number
450 mm
1 Right (P)/Left (L)
1
MP_SZ 450-900 1DP1S
MP_SZ 450-900 1DL1S
600 mm
1 Right (P)/Left (L)
1
MP_SZ 600-900 1DP1S
MP_SZ 600-900 1DL1S
900 mm
1
MP_SZ 900-900 2D1S
Cabinet 900-900 2D2S
1200 mm
2
1
MP_SZ 1200-900 2D1S
Cabinet 1200-900 2D2S model width door drawers order number
900 mm
2
2
MP_SZ 900-900 2D2S
1200 mm
2
2
MP_SZ 1200-900 2D2S
75
MM Steel laboratory furniture
underbench cabinets with service supply for C-type frame of 900 mm height
underbench cabinets with service supply for C-type frame of 900 mm, cabinet height 630 mm, cabinet depth 500 mm
Cabinet 600-900 1D Cabinet 900-900 1D Cabinet 1200-900 2D model width door drawers order number
600 mm
1 Right (P)/Left (L)
-
MP_SZM 600-900 1DP
MP_SZM 600-900 1DL
900 mm
2
-
MP_SZM 900-900 2D
1200 mm
2
-
MP_SZM 1200-900 2D
underbench cabinets with castors; height 630 mm
underbench cabinets with castors; height 630 mm, depth 500 mm
Cabinet 450-900 1D with castors Cabinet 600-900 1D with castors model width door drawers order number
450 mm
1 Right (P)/Left (L)
-
MP_SZK 450-900 1DP
MP_SZK 450-900 1DL
Cabinet 450-900 4S with castors
600 mm
1 Right (P)/Left (L)
-
MP_SZK 600-900 1DP
MP_SZK 600-900 1DL
Cabinet 600-900 4S with castors model width door drawers order number
450 mm
-
4
MP_SZK 450-900 4S
600 mm
-
4
MP_SZK 600-900 4S
Cabinet 450-900 1D1S with castors Cabinet 600-900 1D1S with castors model
76 width door drawers order number
450 mm
1 Right (P)/Left (L)
1
MP_SZK 450-900 1DP1S
MP_SZK 450-900 1DL1S
600 mm
1 Right (P)/Left (L)
1
MP_SZK 600-900 1DP1S
MP_SZK 600-900 1DL1S
Steel laboratory furniture
octagonal underbench cabinet, for C-type frame of 900 mm height
octagonal underbench cabinet, cabinet height 630 mm
Octagonal underbench cabinet 1DP Octagonal underbench cabinet 1DL model
MM width door drawers order number
1000 mm
1 Right (P)
-
MP_SZO 1000-900 1DP
1000 mm
1 Left (L)
-
MP_SZO 1000-900 1DL
wall mounted cabinet with door
wall mounted shelves with the door, depth 365 mm
Cabinet 450 1D Cabinet 600 1D model
Cabinet 900 2D width height door order number model
450 mm
480 / 630 / 780 mm
1 Right (P)/Left (L)
MP_SZW 450-480 1DP
MP_SZW 450-480 1DL
MP_SZW 450-630 1DP
MP_SZW 450-630 1DL
MP_SZW 450-780 1DP
MP_SZW 450-780 1DL width height order number
450 mm
480 / 630 / 780 mm
MP_SZWR 450-480
MP_SZWR 450-630
MP_SZWR 450-780
600 mm
480 / 630 / 780 mm
1 Right (P)/Left (L)
MP_SZW 600-480 1DP
MP_SZW 600-480 1DL
MP_SZW 600-630 1DP
MP_SZW 600-630 1DL
MP_SZW 600-780 1DP
MP_SZW 600-780 1DL
wall mounted open shelves
wall mounted open shelves, depth 320 mm
Open shelf 450 Open shelf 600
600 mm
480 / 630 / 780 mm
MP_SZWR 600-480
MP_SZWR 600-630
MP_SZWR 600-780
900 mm
480 / 630 / 780 mm
2
MP_SZW 900-480 2D
MP_SZW 900-780 2D
MP_SZW 900-630 2D
Open shelf 900
900 mm
480 / 630 / 780 mm
MP_SZWR 900-480
MP_SZWR 900-630
MP_SZWR 900-780
Cabinet 1200 2D
1200 mm
480 / 630 / 780 mm
2
MP_SZW 1200-480 2D
MP_SZW 1200-780 2D
MP_SZW 1200-630 2D
Open shelf 1200
1200 mm
480 / 630 / 780 mm
MP_SZWR 1200-480
MP_SZWR 1200-630
MP_SZWR 1200-780
77
MM Steel laboratory furniture
C-type frame of 750 mm height
C-type frame of 750 mm height basic frame model add-on frame width order number
450 / 600 / 900 / 1200 / 1500 mm
MP_S 450-750 P
MP_S 600-750 P
MP_S 900-750 P
MP_S 1200-750 P
MP_S 1500-750 P
C-type frame of 900 mm height
C-type frame of 900 mm height basic frame add-on frame model
450 / 600 / 900 / 1200 / 1500 mm
MP_S 450-750 R
MP_S 600-750 R
MP_S 900-750 R
MP_S 1200-750 R
MP_S 1500-750 R octagonal add-on frame width order number
450 / 600 / 900 / 1200 / 1500 mm
MP_S 450-900 P
MP_S 600-900 P
MP_S 900-900 P
MP_S 1200-900 P
MP_S 1500-900 P
example of installation
450 / 600 / 900 / 1200 / 1500 mm
MP_S 450-750 R
MP_S 600-750 R
MP_S 900-750 R
MP_S 1200-750 R
MP_S 1500-750 R
-
MP_SO 900 R
78
Example of installation before mounting the worktop
Steel laboratory furniture
Accessories: columns, shelves
columns column column with socket column with socket column with socket column with socket column for media model
MM description order number fully covered cassette
MP_K P
1 AC power socket
MP_K 1G shelf for a column
2 AC power sockets
MP_K 2G
3 AC power sockets
MP_K 3G
4 AC power sockets
MP_K 4G cabinet door lock
1 fitting for media
MP_K M model width order number
825/1000/1070/1075/1100/1125/1200/1225/1230/1275/
1325/1350/1400/1430/1500/1675/1750/1800/2100 mm
MP_PN 825 ... 2100
-
MP_SZ_ZKL
Columns with shelves
A mini column with AC power sockets
Adjustable shelves every 5 cm
79
MM Steel laboratory furniture
Innovative designing
CATIA
The furniture has been designed with a professional software that supports the whole engineering process. All single parts of the furniture are extremely precise and perfectly dimensioned.
3 DVIA
Projects prepared in the CATIA program can be further exported to the 3DVIA program to prepare screens and 3D animations.
Project
Every offer for furniture is complete with a customized 3DVIA visualization to overview a target configuration in a laboratory.
Additional simulations, such as openings of drawers and doors, can be very helpful to guarantee an ideal, trouble free final configuration.
The program allows to prepare professional visualizations and fully dimensioned projections.
80
Other laboratory equipment
IV
82
Other laboratory equipment
Colony counter
Laboratory shakers
Stationary samplers
83
84
86
Colony counter
Advantages automatic weight compensation of Petri plates anti-shock counting technology ringlight technology enables even illumination of the counting field possibility of working with bright or dark background mean value calculation function standard marker included optional marker ZM 2002 for external counting
Standard equipment colony counter magnifying glass standard marker bright and dark background
Petri plates adapters
Wolfhuegel scale plate
Petri plates adapters (diameter < 120mm) removable Wolfhuegel scale plate adjustable push force sound and visual counting control adjustable position of the magnifying glass affordable price
Accessories marker ZM 2002 for external counting
Colony counter is invaluable help in every microbiological laboratory since the most time consuming activity is counting the colonies on Petri plates.
An easy-to-use device featuring quick and precise counting.
LKB 2002
LKB
ZM 2002
Model
Parameter counting field diamater [mm] display magnifying glass illumination dims [mm] weight [kg] nominal power [W] voltage 50/60 Hz [V] warranty manufacturer width x depth x height
LKB 2002
120
LED (0...999)
2,5 X
20W ringlight
300 x 325 x 90
4,9
22
115/230
24 months
POL-EKO-APARATURA
83
LS Laboratory shakers
Advantages orbital movement microprocessor control of rotation and time orbital diameter: 10...25 mm max shaking weight: 10 kg variable speed control: 30...300 rpm shaking mode: from 1 min to 99 h, or continuous
LCD digital display anti-skid mat (option) various shaking tables can be located inside cooled incubators
Accessories universal shaking table separating funnel attachment fixing clip support dish attachment test tube support
Erlenmeyer flasks (25...2000 ml) attachment
LS 280
84
Parameter movement controller display speed range [rpm] accuracy [rpm] amplitude [mm]
1 max load capacity [kg] shaking mode dimensions without / with shaking table [mm] width height depth fits to cooled incubator nominal power [W] weight with shaking table [kg] ambient temperature
[°C] humidity [%] voltage 50/60 Hz [V] warranty manufacturer
1 - requested amplitude must be specified in the order
ILW 53 with LS 280
Model LS 280
5
320
120 / 220
330
ILW 53
60
10
Laboratory shakers have been designed to fit inside cooled incubators (IL range)
LS 350 LS 500 orbital microprocessor
LCD display
30 ... 300
10
5 or 10
10
1 min...99 h or continuous
390
120 / 220
400
ILW 115
60
15
+10...+40 up to 70
100/240
24 months
POL-EKO-APARATURA
550
120 / 220
440
ILW 240
60
22
LS 700
700
120 / 220
420
ILW 400
60
25
universal platform anti-skid mat
Accessories for Laboratory shakers
LS platform for fixing flasks handles platform for Petri plates shaking platform for separatory funnels
Universal platform
Universal platform for various kinds of vessels, with 4 roller clamps (without anti-skid mat).
Platform for fixing flasks handles
Platform for fixing flasks handles, suitable for flasks of the following capacities: 25 ml, 50 ml,
100 ml, 250 ml, 500 ml, the handles shall be ordered separately.
Platform for Petri plates shaking
Platform for shaking Petri plates, bacteria culture flasks and other vessels of low centre of gravity.
Platform for separatory funnels
Platform for separatory funnels with 3 roller clamps for shaking, salting, extraction and concentration.
Anti-skid mat
Anti-skid mat for LS laboratory shakers.
85
PP Stationary samplers
Advantages representative sample taking according to PN-ISO 5667 sampling system:
vacuum
peristaltic pump sampling mode:
time proportional
flow proportional
event (e.g.: pH value exceeding)
combined intuitive menu up to 5 configurable sampling programs bottle filling overview suitable for continuous outdoor use can be implemented into a monitoring system refrigerated chamber
SD card recording system: pH, conductivity, redox, dissolved oxygen, flow, chamber temperature etc.
Representative sample taking according to PN-ISO 5667 directive.
Model
PP 2002+
86
Parameter sampling system sample storing menu language medium hose blowing sampling mode sampling height [m] sample volume [ml] hose length [m] hose diameter [mm] distributor number of bottles x capacity [l] overall dims [mm] weight [kg] housing ambient temperature [°C] nominal power [W] controller programming data logging input signals output signals communication installation site voltage 50 Hz [V] warranty manufacturer width height depth
PP 2002+/24
PP 2002E PP 2002M vacuum system peristaltic pump peristaltic pump
/vacuum system stable temperature +4°C regardless of ambient conditions
English, French, Polish, Czech, Romanian, Lithuanian, Italian liquid media of min conductivity 20 µS/cm and max temp. 60°C before and after sample taking automatic time proportional, flow proportional, event or manual max. 8 / up to 30 (option) regulated
30...250/500 regulated
10...9990
regulated
30...250/500 or 10...9990
8 standard
12 round
24 x 1; 12 x 2,9; 4 x 10; 1 x 25
630
1070
660
630
1325
660
100 90 acid-proof stainless steel with 40 mm insulation
-20…+45
350 450 microprocessor, graphic display bottle filling overview
5 programs, 8 tasks each
SD card (option)
8 analogue, 4 binary
4 binary
RS 232 or RS 485 / PROFIBUS, GSM modem (option)
SD card indoor or outdoor
230
24 months
POL-EKO-APARATURA
POL-EKO Laboratorium
Pomiarowe company
V
LP POL-EKO Measurement Laboratory
88
POL-EKO Measurement Laboratory is Accredited by the
Polish Centre for Accreditation (a member of ILAC) and provides accredited services
calibration of chambers for steam sterilization (such as autoclaves, Koch steam sterilizers etc.)
method temperature range: +60…+130°C
measurement capability: 0,5
°C
place of calibration: customer’s site calibration of lab furnaces method temperature range: +100…+1000 °C
calibration and measurement capability: 3,0 °C place of calibration: Measurement Laboratory or at customer’s site calibration of thermostatic and climatic chambers (ovens, incubators, sterilizers, cooled incubators, thermostatic cabinets, refrigerators, freezers, climatic chambers etc.) method temperature range: -25…+200°C
calibration and measurement capability 0,15 °C place of calibration: Measurement Laboratory or at customer’s site calibration of water baths and thermoreactors method temperature range: -20…+200 °C
calibration and measurement capability: 0,15 °C place of calibration: Measurement Laboratory or at customer’s site
The calibration certificate features:
average value in each point uncertainty of the measurement
temperature homogenity in chamber or on the surface
load effect (optionally)
POL-EKO Measurement Laboratory LP calibration of electric and electronic thermometers in Measurement Laboratory method temperature range: -25…+1000 °C
calibration and measurement: 0,05 °C in range -20 ... +80 °C;
0,06 in range +80 ... +200 °C; 2,5 in range +200 ... +1000 °C
calibration of electric and electronic thermometers at customer’s site
method temperature range: +30…+200°C calibration and measurement capability: 0,2
°C calibration of data loggers for air temperature measurement method temperature range: 0…+130 °C
calibration and measurement capability: 0,2 °C in range 0 ... +60°C; 0,3°C in range +60 ... +130°C
place of calibration: Measurement Laboratory calibration of thermohygrometers method temperature range: +10…+60 °C relative humidity range: 30…90% place of calibration: Measurement Laboratory
The calibration certificate features:
average value of temperature (humidity)
temperature (humidity) correction measurement uncertainty
Services beyond accreditation
calibration of climatic chambers relative humidity range: 30…90% place of calibration: Measurement Laboratory or customer’s site
IQ, OQ, PQ qualification procedures place of calibration: Measurement Laboratory or customer's site mapping of room temperatures
Qualification procedures are required by pharmaceutical (also by veterinary pharmacy), cosmetic and biotechnology industries. A complete qualification procedure consists of the following parts: installation qualification (IQ), operational qualification (OQ), and performance qualification (PQ)
The qualification procedure after conducting measurements and completing documentation is followed by issuing the “Qualification book” (full list of conducted measurements and tests is available upon request).
POL-EKO-APARATURA also provides many services concerning various laboratory equipment
pH-meters
conductivity meters
multiparameter metres
photometers and spectrophotometers
EcaFlow 150 GLP analyser of trace concentrations of metals
monometric BOD equipment, e.g. OxiTop, OxiDirect etc.
checking and cleaning of: pH and redox electrodes; oxygen sensors;
conductivity cells
implementing new photometric methods into photometers and spectrophotometers
The Measurement Laboratory issues a ”Verification Certificate” after the control.
89
POL-EKO-APARATURA
manufacturer of laboratory equipment for laboratory analysis and technological processes, authorised distributor in Poland of:
ARCTIKO, EUTECH,
HAMILTON, KNICK,
THERMO SCIENTIFIC, WTW
POL-EKO-APARATURA sp.j.
ul. Kokoszycka 172C
44 - 300 Wodzis³aw Œl¹ski
Tel: +48 32 453 91 70
Fax: +48 32 453 91 85 e-mail: [email protected]
www.polekolab.com
Catalogue “Products of POL-EKO-APARATURA” version 9.0/2013.
In spite of taking necessary steps to avoid mistakes, we cannot guarantee that the technical data in this catalogue is completly correct
POL-EKO-APARATURA company reserves the right to change some of the technical parameters. All dimensions are given exact to ±5 %.
advertisement
Key Features
- Full colour 5.7” touch screen
- Ethernet network connection
- 7-day programming mode
- Temperature (and humidity in KK) calibration at 3 points
- USB port for data transfer
- Remote control via Internet