BC Biomedical ESU-2050 Series Electrosurgical Unit Analyzer User Manual
The ESU-2050 Series Electrosurgical Unit Analyzers is a family of high-accuracy True RMS RF Voltmeters designed to be used in the routine performance verification of Electrosurgical Generators. They offer a higher degree of accuracy than previously attainable with conventional Electrosurgical Unit Analyzer designs. The ESU-2050 Series has been designed to be used in conjunction with an external RF Current Transformer (Pearson Electronics Model 411 and 4100 recommended) and external precision load resistors (Vishay Dale NH-250 Precision 1% tolerance resistors recommended) to measure various parameters relating to the routine service of Electrosurgical Generators. They are microprocessor based and utilize a combination of unique hardware and software to provide accurate and reliable test results, even from “noisy” ESU Generator waveforms such as “Spray”. The Patent Pending DFA® Technology utilized in the ESU-2050 Series allows the system to aggressively digitize the complex RF waveforms produced by Electrosurgical Generators, analyze each individual digital data point, and provide highly accurate measurement results.
Advertisement
Advertisement
ELECTROSURGICAL
UNIT ANALYZER
ESU-2050 SERIES
USER MANUAL
BC BIOMEDICAL
ESU-2050 SERIES
TABLE OF CONTENTS
WARNINGS, CAUTIONS, NOTICES ............................................................................ ii
DESCRIPTION ............................................................................................................. 1
TYPICAL TEST SETUP ............................................................................................... 4
OVERVIEW .................................................................................................................. 5
KEYS ............................................................................................................................ 21
POWER UP SETTINGS .............................................................................................. 23
GRAPH MODE ............................................................................................................. 25
ERROR MESSAGES ................................................................................................... 28
DFA
®
TECHNOLOGY .................................................................................................. 29
COMMUNICATION PROTOCOL ................................................................................. 30
COMMUNICATION COMMAND SUMMARY ............................................................... 34
MANUAL REVISIONS .................................................................................................. 36
LIMITED WARRANTY ................................................................................................. 36
SPECIFICATIONS ....................................................................................................... 37
NOTES ......................................................................................................................... 39 i
WARNING - USERS
The ESU-2050 Series Analyzers are for use by skilled technical personnel only.
WARNING - USE
The ESU-2050 Series Analyzers are intended for testing only and they should never be used in diagnostics, treatment or any other capacity where they would come in contact with a patient.
WARNING - MODIFICATIONS
The ESU-2050 Series Analyzers are intended for use within the published specifications. Any application beyond these specifications or any unauthorized user modifications may result in hazards or improper operation.
WARNING - CONNECTIONS
All connections to patients must be removed before connecting the DUT to the Analyzer. A serious hazard may occur if the patient is connected when testing with the Analyzer.
Do not connect any leads from the patient directly to the Analyzer or DUT.
WARNING - POWER ADAPTOR
Unplug the Power Adaptor before cleaning the surface of the Analyzer.
WARNING - LIQUIDS
Do not submerge or spill liquids on the Analyzer.
Do not operate the Analyzer if internal components not intended for use with fluids may have been exposed to fluid, as the internal leakage may have caused corrosion and be a potential hazard.
ii
CAUTION - SERVICE
The ESU-2050 Series Analyzers are intended to be serviced only by authorized service personnel.
Troubleshooting and service procedures should only be performed by qualified technical personnel.
CAUTION - ENVIRONMENT
The ESU-2050 Series Analyzers are intended to function between 15 and 30 °C.
Exposure to temperatures outside this range can adversely affect the performance of the Analyzer.
CAUTION - CLEANING
Do not immerse. The Analyzer should be cleaned by wiping gently with a damp, lint-free cloth. A mild detergent can be used if desired.
CAUTION - INSPECTION
The ESU-2050 Series Analyzers should be inspected before each use for wear and should be serviced if any parts are in question.
iii
NOTICE – CE
The ESU-2050 Series Analyzers bear the mark
Based on the following testing standards:
ELECTROMAGNETIC COMPATIBILITY DIRECTIVE
EMC – Directive 89/336/EEC and 2004/108/EC as amended by
92/31/EEC, 93/68/EEC and Directive 91/263/EEC [ TTE/SES ]
EN 61000-3-2
EN 61000-3-3
EN 61326-1:1997 + A1:1998 + A2:2001 + A3:2003
“Electrical equipment for measurement, control and laboratory use
– EMC requirements”
This equipment has been type tested by an independent, accredited testing laboratory and compliance was demonstrated to the above standard to the extent applicable.
EMISSIONS
Radiated and Line Conducted Emissions
Harmonic Current Emissions
Voltage Fluctuation and Flicker
IMMUNITY
– CLASS C
EN 61000-4-2
EN 61000-4-3
EN 61000-4-4
EN 61000-4-5
EN 61000-4-6
EN 61000-4-11
Electrostatic Discharge
Radiated Electric Field Immunity
Electrical Fast Transients / Bursts
Surge Voltage
Conducted Disturbance
Voltage Dips and Short Interrupts
LOW VOLTAGE DIRECTIVE
EC – Directive 73/23/EC
EN 61010-1:2001
“Safety requirements for electrical equipment for measurement, control, and laboratory use – General requirements”
This equipment has been type tested and compliance was demonstrated to the above standard to the extent applicable. iv
NOTICE – SYMBOLS
Symbol Description
Caution
(Consult Manual for Further Information)
RF Current Transformer
Per European Council Directive 2002/95/EC, do not
dispose of this product as unsorted municipal waste.
NOTICE – ABBREVIATIONS
A or Amps Amperes
C Celsius
CF
°
DFA
DUT
Hz k kg kHz lbs
Crest Factor degree
Digital Fast Acquisition
Device Under Test hertz kilo- (10
3
) kilogram(s) kilohertz pounds
L1, L2, L3 Location 1, 2, 3
M Mega- (10
6
)
MHz
µ m mA mm ms
Megahertz micro- (10
-6
) milli- (10
-3
) milliampere(s) millimeter(s) millisecond(s) mV
Ω
PC millivolt(s) ohm(s)
Personal Computer p or pk peak p-p or pk-pk
RF peak-to-peak
Radio Frequency
RMS
US
V
VDC
W
Root Mean Square
United States volt(s)
Direct Current Voltage watt(s) v
NOTICE – PERFORMING TESTS
REFER TO DUT MANUFACTURER’S SERVICE MANUAL FOR
TEST PROCEDURES AND MEASUREMENT LIMITS
NOTICE – DISCLAIMER
USER ASSUMES FULL RESPONISIBILITY FOR UNAUTHORIZED
EQUIPMENT MODIFICATIONS OR APPLICATION OF EQUIPMENT
OUTSIDE OF THE PUBLISHED INTENDED USE AND
SPECIFICATIONS. SUCH MODIFICATIONS OR APPLICATIONS
MAY RESULT IN EQUIPMENT DAMAGE OR PERSONAL INJURY.
NOTICE
BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO
MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT
ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE
DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST
POSSIBLE PRODUCT. THE INFORMATION IN THIS MANUAL HAS
BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE
ACCURATE. HOWEVER, NO RESPONSIBILITY IS ASSUMED FOR
INACCURACIES.
NOTICE
– DISCLAIMER
.
– CONTACT INFORMATION
BC BIOMEDICAL
BC GROUP INTERNATIONAL, INC.
3081 ELM POINT INDUSTRIAL DRIVE
ST. CHARLES, MO 63301
USA
1-800-242-8428
ESU-2050 Series User Manual www.bcgroupintl.com
03/12
1-314-638-3800 www.bcgroupintl.com
Copyright © 2012
Made in the USA
Rev 10 vi
BC BIOMEDICAL
ESU-2050 SERIES
ELECTROSURGICAL UNIT ANALYZER
The ESU-2050 Series Electrosurgical Unit Analyzers is a family of high-accuracy True
RMS RF Voltmeters designed to be used in the routine performance verification of
Electrosurgical Generators. They offer a higher degree of accuracy than previously attainable with conventional Electrosurgical Unit Analyzer designs. The ESU-2050 Series has been designed to be used in conjunction with an external RF Current Transformer
(Pearson Electronics Model 411 and 4100 recommended) and external precision load resistors (Vishay Dale NH-250 Precision 1% tolerance resistors recommended) to measure various parameters relating to the routine service of Electrosurgical Generators. They are microprocessor based and utilize a combination of unique hardware and software to provide accurate and reliable test results, even from “noisy” ESU Generator waveforms such as “Spray”. The Patent Pending DFA® Technology utilized in the ESU-2050 Series allows the system to aggressively digitize the complex RF waveforms produced by
Electrosurgical Generators, analyze each individual digital data point, and provide highly accurate measurement results.
The following are highlights of some of the main features:
TRUE RMS READINGS USING DFA® TECHNOLOGY
INDUSTRY STANDARD CURRENT SENSING TECHNOLOGY
MV, MV PEAK, MA, CREST FACTOR AND POWER (WATTAGE) RANGES
LARGE GRAPHICS DISPLAY WITH CURSOR SELECTION
OF OPTIONS AND SETUP OF PARAMETERS
1% OF READING MEASUREMENT ACCURACY
DIGITAL DATA OUTPUT VIA USB AND RS232
PC BASED INTERFACE AND DATA CAPTURE PROGRAM
DIGITAL CALIBRATION
– NO POTS TO TURN
SELECTABLE DISPLAY OPTIONS
DISPLAY CONTRAST IS SOFTWARE ADJUSTABLE
TACTILE KEYS WITH AUDIO FEEDBACK
GRAPHICAL ON-SCREEN REPRESENTATION OF MEASURED RF SIGNAL
STANDARD (1000 MV) AND LOW RANGE (100 MV) WITH AUTOSCALING
CAPABILITY
CAN BE USED WITH 0.1:1 OR 1:1 VOLTAGE:CURRENT CURRENT
TRANSFORMERS
1
INTERNALLY PROTECTED INPUT CIRCUITRY GUARDS AGAINST INPUT
OVERLOAD DAMAGE
INTERNAL DATA STORAGE FOR 3 FULL DATA SETS
ESU-2050P HAS ALL THE BASIC MODEL FEATURES PLUS:
PULSED RF WAVEFORM MEASUREMENT MODE FOR LOW DUTY CYCLE
PULSED OUTPUTS OFFERED BY SOME ELECTROSURGICAL
MANUFACTURERS
STANDARD ACCESSORIES:
BC20 – 00126
BC20 – 41352
BC20 – 41341
BC20
– 00232
BC20 – 30108
BC20 – 21104
BC20
– 205XX
ACCESSORY KIT (TEST LEADS)
COMMUNICATIONS CABLE (USB)
COMMUNICATIONS CABLE (RS232)
CT CABLE (BNC)
SOFT SIDED CARRYING CASE
UNIVERSAL POWER SUPPLY
STANDARD POWER ADAPTER
(International Options, see Page 23 for details)
OPTIONAL ACCESSORIES:
BC20 – 41353 CD, ESU2050 INTERFACE SOFTWARE
BC20 – 00231 PEARSON ELECTRONICS MODEL 411 0.1:1 RATIO
CURRENT TRANSFORMER
BC20
– 00230
PEARSON ELECTRONICS MODEL 4100 1:1 RATIO
CURRENT TRANSFORMER
VISHAY-DALE 1% TOLERANCE LOAD RESISTORS:
BC20 – 00200
BC20 – 00201
BC20 – 00202
BC20 – 00203
BC20 – 00204
BC20 – 00205
BC20 – 00206
BC20 – 00207
BC20 – 00208
BC20 – 00209
BC20 – 00210
BC20 – 00211
BC20 – 00212
BC20 – 00213
BC20
– 00214
BC20
– 00215
BC20
– 00216
BC20
– 00217
5 , 250-WATT, NH-250 SERIES RESISTOR
10 , 250-WATT, NH-250 SERIES RESISTOR
20 , 250-WATT, NH-250 SERIES RESISTOR
30 , 250-WATT, NH-250 SERIES RESISTOR
50 , 250-WATT, NH-250 SERIES RESISTOR
100 , 250-WATT, NH-250 SERIES RESISTOR
200 , 250-WATT, NH-250 SERIES RESISTOR
300 , 250-WATT, NH-250 SERIES RESISTOR
500 , 250-WATT, NH-250 SERIES RESISTOR
1000 , 250-WATT, NH-250 SERIES RESISTOR
2000 , 250-WATT, NH-250 SERIES RESISTOR
3000 , 250-WATT, NH-250 SERIES RESISTOR
4000 , 250-WATT, NH-250 SERIES RESISTOR
5000 , 250-WATT, NH-250 SERIES RESISTOR
1 , 50-WATT, NH-50 SERIES RESISTOR
125 , 250-WATT, NH-250 SERIES RESISTOR
150 , 250-WATT, NH-250 SERIES RESISTOR
400 , 250-WATT, NH-250 SERIES RESISTOR
2
BC20 – 00218
BC20 – 00219
BC20 – 00220
BC20 – 00240
800 , 250-WATT, NH-250 SERIES RESISTOR
1500 , 250-WATT, NH-250 SERIES RESISTOR
10 , 50-WATT, NH-50 SERIES RESISTOR
LOAD RESISTOR BANANA JACK ADAPTER SET (2)
3
TYPICAL TEST SETUP
Unlike all conventional ESU Analyzers with lesser degrees of accuracy, the ESU-2050
Series utilizes an external Current Transformer and external precision load resistors
(values to be determined by the manufacturer’s suggested test load for the generator you are testing or servicing) for typical Electrosurgical Generator testing. Many of the world’s leading Electrosurgical generator manufacturers utilize this exact same technique when they test, service and calibrate their generators.
This methodology has several distinct advantages over conventional ESU Analyzers:
Improved accuracy and resolution capabilities
100% manufacturer recommended test load compatibility
Smaller and lighter weight instrumentation
The current transformer senses the RF current flowing through the external test load and produces a proportional voltage as an input to the ESU-2050. This input is either a direct
1 : 1 input (for current transformers with a ratio of 1 : 1 volts : amps) or a 0.1 : 1 input (for current transformers with a ratio of 0.1 : 1 volts : amps). Combining the standard and low ranges of the ESU-2050 with the use of either a 0.1:1 or 1:1 ratio current transformer, the user has full control over the ability to get high accuracy and high resolution readings from all types of Electrosurgical Generators.
4
OVERVIEW
This section looks at the layout of the ESU-2050 and gives descriptions of the elements that are present.
10 Light Touch Keys for
Selecting Parameters and
Settings:
High Impact
Plastic Case
Backlit LCD
Graphical
Display
Swivel and
Locking
Handle
RF Current
Transformer
(Donut ) Input
Power
Kycon 3 pos locking connector
USB
RS232
DB9
5
Power Switch
Rocker Switch
MAIN SCREENS
– There are 7 main screens, 5 Display Screens which have 1, 2, 3, 4 and
5 display zones respectively, a Measurement List Screen which shows available measurements and the Quick Config Screen which displays the current hardware configuration. In the Display Screens, each Display Zone can be customized to show the desired parameter from the following options:
Parameter mV RMS mA RMS
Power in Watts mV Peak
Time Pulse - On
Time - Pulse Off
Time - Total Cycle
% Duty Cycle mV Pulse mA Pulse
Watts Pulse
Abbreviation mV mA
Watts mV Pk mV Peak - to - Peak mV P-P mV Peak /
Peak - to - Peak mV Positive Peak
Only
Crest Factor
Pk/P-P mV Pk+
CF
Ton
Toff
Tcyc
%Duty mV cyc mA cyc
Wcyc
Description
This is the mV measured directly from the RF donut.
This is the converted mA measurement based on the RF donut mV to mA attenuation ratio.
This is the computed power based on load setting and mA measured.
This is the maximum mV measured in the buffer.
NOTE: This is shown as absolute value.
This is the difference between maximum mV measured and min mV measured.
This is the ratio of Peak versus
Peak to Peak millivolts.
This is the maximum positive mV measured in the buffer. For asymmetric waveforms this can determine if the output polarity is reversed.
This is the ratio of peak to rms of the measured waveform.
ESU-2050P Only
This is the duration that the pulsed waveform is on.
(See Diagram 1)
ESU-2050P Only
This is the duration that the pulsed waveform is off.
(See Diagram 1)
ESU-2050P Only
This is the total cycle time of the pulsed waveform
(i.e. Ton + Toff). (See Diagram 1)
ESU-2050P Only
This is the ratio of the pulse on time (Ton) versus cycle time (Tcyc). (See Diagram 1)
ESU-2050P Only
This represents the RMS mV over one pulsed cycle.
(See Diagram 1)
ESU-2050P Only
This represents the RMS mA over one pulsed cycle.
(See Diagram 1)
ESU-2050P Only
This represents the RMS Watts over one pulsed cycle.
(See Diagram 1)
6
Ton Toff
Tcyc mV mA
Watts
CF
Diagram 1 (ESU-2050P Only) mVcyc mAcyc
Wcyc
The available screens can be toggled through using .
Display Screens
One Display Zone Screen with mV parameter selected
Two Display Zone Screen with mV and Watts parameters selected
Three Display Zone Screen with mV, Watts and mA parameters selected
Four Display Zone Screen with mV, Watts, mA and mV
Peak parameters selected
7
Measurement List Screen:
Measured Parameters
Parameter mV RMS mA RMS
Watts RMS mV Peak mV Peak - to - Peak mV Peak /
Peak - to - Peak mV Positive Peak
Crest Factor
Time Pulse – On *
Time - Pulse Off *
Time - Total Cycle *
% Duty Cycle * mV Pulse * mA Pulse *
Watts Pulse *
* ESU-2050P Only
Abbreviation mV mA
Watts mV Pk mV P-P
Pk/P-P mV Pk+
CF
Ton
Toff
Tcyc
%Duty mV cyc mA cyc
Wcyc
Five Display Zone Screen with mV, Watts, mA and mV
Peak and CF parameters selected
8
Quick Config Screen:
Input Range
100 mV
1000 mV
Auto
Input Mode
(ESU-2050P Only)
Continuous
Pulsed
Donut Atten
0.1 : 1
1 : 1
Load
Resistance
0 –
6,500.0 ohms
NOTE: Load Resistance can be
Adjustable or Selected from the table of available resistor combinations. (See Load Table for more information).
The Quick Config Screen allows the user to see the current configuration and provide a quick method of changing the RF Current Transformer (Donut) Attenuation, Load
Resistance , Input Range or Input mode parameters.
Use to highlight the parameter to change and then to sequence through the available options.
NOTE: On power up, the Quick Config Screen will display for a few seconds to indicate the current configuration and then the default Display Screen will be displayed. The Quick
Config Screen can be accessed using .
9
DISPLAY PARAMETERS
– There are five options of parameters that can be selected for each Display Zone on the Main Screens. This allows users to custom configure the displays to best suit their needs.
Use to highlight the Display Zone to change and then to sequence through the available parameters.
NOTE: To save a custom configuration, see Power Up Settings section.
SYSTEM CONFIGURATION SCREEN – The SYSTEM SETUP MODE allows the user to adjust the configuration of the unit. The Setup Screen can be entered using the
key. The parameters can be changed by using to highlight the line and to toggle the available options. The Setup Screen can be exited using the key.
Typical Setup Screen
10
The following is a breakdown of the parameters available in the configuration of the unit and their available options:
System Setup Configuration
Parameter
Donut Atten
Input Zero
Input Range
Input Mode
(ESU-2050P Only)
Load Resistance
Load Selection
Power up with
Num A/D Samples
Display Averaging
Description Range
Selects the RF Current Transformer Attenuation in
Volts : Amp for the RF Donut being used .
Default = 0.1 : 1
0.1 : 1
1 : 1
Volts : Amps
Zeros the input circuitry based on donut being used. Each donut can have a slightly different zero offset. This function will eliminate the offset from the readings. Press the SETUP key while this parameter is selected to perform the auto –zero function.
Sets the number of A/D converter readings used in each mV RMS computation. A higher setting requires more computation and is slower, but results in a more stable reading.
Default = 32,768
Sets which display averaging parameter is to be used. Three independent averaging modes can be configured for optimum system performance.
Press Setup
Determines the input range.
Default = Auto
100 mV
1000 mV
Auto
Determines whether the unit continuously monitors the Input Signal or looks for a Pulsed Input Signal.
Default = Continuous
Used for Power (Watts) calculation only. Can be adjustable or set by a table of fixed resistors. (See
Load Selection and Load Table for more information.)
Default = 500
Determines whether the Load Resistance
Parameter is adjustable by tenths of ohms or selected from the Load Table. The Load Table is created from the Load Resistance Values set in the
Factory Setup and the Load Sets. (See Load Sets,
Load Table and Factory Setup for more information.)
Default = Table
Continuous
Pulsed
0-6,500.0 ohms
Adjustable or Table
Determines the power up mode of the ESU-2050.
The default mode shows a single mV parameter display. Set this parameter to custom to display the saved startup mode. Set this parameter to
Save current as custom to save the settings for the next time power is cycled.
Default = Defaults
Defaults
Custom
Set Current as Custom
1024
2048
4096
8192
16384
32768
Fast
Medium
Slow
11
System Setup Configuration
Parameter
Slow Averaging
Medium Averaging
Fast Averaging
Averaging Window
Load Set 1
Load Set 2
Load Set 3
Load Set 4
Description
Sets the number of mV RMS readings that are averaged when the Display Averaging parameter is set to Slow. A higher number will cause the display to update slower, but will give a more stable reading
Default = 150
Sets the number of mV RMS readings that are averaged when the Display Averaging parameter is set to Medium. A higher number will cause the display to update slower, but will give a more stable reading
Default = 15
Sets the number of mV RMS readings that are averaged when the Display Averaging parameter is set to Fast. A higher number will cause the display to update slower, but will give a more stable reading
Default = 4
Sets the range of input readings that will be averaged. If a new mV reading deviates from the average by less than this amount, it will be averaged with the rest of the readings in the display averaging buffer. Otherwise, the input is considered a step change and the display averaging buffer is flushed.
Assigns a resistance value for Set 1 from a combination of the available loads as determined by the Load Resistance Values set in the Factory
Setup. (See Custom Load Sets for more information.)
Default = None Selected
Assigns a resistance value for Set 2 from a combination of the available loads as determined by the Load Resistance Values set in the Factory
Setup. (See Custom Load Sets for more information.)
Default = None Selected
Assigns a resistance value for Set 3 from a combination of the available loads as determined by the Load Resistance Values set in the Factory
Setup. (See Custom Load Sets for more information.)
Default = None Selected
Assigns a resistance value for Set 4 from a combination of the available loads as determined by the Load Resistance Values set in the Factory
Setup. (See Custom Load Sets for more information.)
Default = None Selected
Range
1-200 Readings
1-200 Readings
1-200 Readings
0.0 to 100.0 mV
0-6,500.0 ohms
0-6,500.0 ohms
0-6,500.0 ohms
0-6,500.0 ohms
12
Parameter
Load Set 5
Load Set 6
Load Set 7
Load Set 8
Load Set 9
Load Set 10
LCD Contrast
Access Code
Software
System Setup Configuration
Description
Assigns a resistance value for Set 5 from a combination of the available loads as determined by the Load Resistance Values set in the Factory
Setup. (See Custom Load Sets for more information.)
Default = None Selected
Assigns a resistance value for Set 6 from a combination of the available loads as determined by the Load Resistance Values set in the Factory
Setup. (See Custom Load Sets for more information.)
Default = None Selected
Assigns a resistance value for Set 7 from a combination of the available loads as determined by the Load Resistance Values set in the Factory
Setup. (See Custom Load Sets for more information.)
Default = None Selected
Assigns a resistance value for Set 8 from a combination of the available loads as determined by the Load Resistance Values set in the Factory
Setup. (See Custom Load Sets for more information.)
Default = None Selected
Assigns a resistance value for Set 9 from a combination of the available loads as determined by the Load Resistance Values set in the Factory
Setup. (See Custom Load Sets for more information.)
Default = None Selected
Assigns a resistance value for Set 10 from a combination of the available loads as determined by the Load Resistance Values set in the Factory
Setup. (See Custom Load Sets for more information.)
Default = None Selected
Sets the contrast of the display screen.
Deault – 10
In some cases it may be desirable to restrict access to the System Setup. This sets the number that must be matched in order to gain access to the System Setup. If set to 0, the Access Code feature is disabled. Default – 0
Displays current software program.
13
Range
0-6,500.0 ohms
0-6,500.0 ohms
0-6,500.0 ohms
0-6,500.0 ohms
0-6,500.0 ohms
0-6,500.0 ohms
0-20
0 to 9999
(Read Only)
INPUT ZERO
– The Input Zero offset can be slightly different between RF donuts. This parameter accesses an auto-zeroing function that eliminates this offset. Independent settings are saved for the 0.1:1 RF donut and 1:1 RF donut. The user can switch between the two donut types without having to re-zero the input. The input needs to be zeroed only when a new donut is introduced.
INPUT RANGE – The input range can be scaled to accommodate the signal that is being measured. The input can be set to fixed ranges of 100mV Peak, 1000mV peak or Auto-
Ranging. For Auto-Ranging mode, the low range will be used for readings from 0.00 to
30.00 mV RMS. The high range will be used for 20.0 to 700.0 mV RMS.
INPUT MODE (ESU-2050P Only) – There are two input modes to allow for measurement of continuous signals or pulsed signals. In Continuous Mode, the input is updated every
100 mS. This mode should be used for all Electrosurgical Generator waveform outputs in general use generators. A number of Electrosurgical generator manufacturers offer generators with pulsed outputs, where there is a long duty cycle (typically ½-second or more) and the actual RF output of the generator is active for a brief period of time within the duty cycle (typically 1/10 th
second or less). In Pulsed Mode, the input to the
ESU-2050P is only processed when a signal over 20 mV in amplitude is detected. The
ESU-2050P analyzes the pulsed input waveform and can provide the RMS readings for
either the overall input or the pulse only (see Diagram 1, Page 7).
NOTE: When set to Pulsed
Mode, a small graphic appears in the upper right corner of the screen to identify to the user that the ESU-2050P is looking for a pulsed RF input.
14
CUSTOM LOAD SETS
– With twelve available loads, there is a possibility of 4096 combinations of resistors to use as a combined load. To simplify the selection of commonly used load configurations, 10 custom resistor sets are available. Each resistor set can consist of any combination of the available calibrated loads. The number of loads and load calibration is performed in the Factory Setup Screen.
A custom load set can be configured in the Setup Screen by either using to highlight the line and to toggle the available custom set resistance, or by using the key to access the Custom Resistor Set Menu.
This menu shows the current values of the loads as set in the Factory Setup Screen. The value of a Load Set can be changed by using to highlight the load line and to include or remove the resistor from the custom set. The total series resistance of the selected resistors is shown to aid in customizing the custom resistor set.
NOTE: If individual selected load values are changed in the Factory Setup Screen, the resistance value of the sets will change accordingly.
The Setup Screen can be exited using the key.
15
LOAD TABLE – Up to twelve Load Resistance Values (each with a range from 0.0 to
6,500.0 ohms) may be set in the FACTORY SETUP Configuration. These values are used in combination with the Custom Load Sets to determine the Load Configuration Table.
These options are available if the Load Selection parameter is set to “Table”. The settings will be the individual calibrated loads followed by the Load Sets. Since these values can be set to the actual values of the real resistors, this allows for maximum accuracy in the wattage calculations. By default, the following values are loaded into this table when the
ESU-2050 instruments ships from the factory:
Load #1: 10
Load #2: 20
Load #3: 30
Load #4: 50
Load #5: 100
Load #6: 200
Load #7: 300
Load #8: 500
Load #9: 1000
Load #10: 2000
Load #11: 3000
Load #12: 4000
16
LOAD CALIBRATION SCREEN
– The LOAD SETUP MODE allows the user to adjust the calibration of the loads. The Load Setup Screen can be entered using the key while in the SYSTEM SETUP MODE. The parameters can be changed by using to highlight the line and to toggle the available options. The Load
Setup Screen can be exited using the key.
Typical Load Setup Screen
17
The following is a breakdown of the parameters available in the LOAD SETUP MODE and their available options:
Load Setup Configuration
Parameter
Number Of Loads
Load 1
Load 2
Load 3
Load 4
Load 5
Load 6
Load 7
Load 8
Load 9
Load 10
Load 11
Load 12
Description
Sets the number of load resistors present in the system. This determines the maximum combination of resistors available when the Load
Selection is set to Table.
Calibrates the Load 1 Resistance Value. This should be set to the actual resistance of the smallest resistor in the system.
Calibrates the Load 2 Resistance Value. This should be set to the actual resistance of the next higher resistor in the system.
Calibrates the Load 3 Resistance Value. This should be set to the actual resistance of the next higher resistor in the system.
Calibrates the Load 4 Resistance Value. This should be set to the actual resistance of the next higher resistor in the system.
Calibrates the Load 5 Resistance Value. This should be set to the actual resistance of the next higher resistor in the system.
Calibrates the Load 6 Resistance Value. This should be set to the actual resistance of the next higher resistor in the system.
Calibrates the Load 7 Resistance Value. This should be set to the actual resistance of the next higher resistor in the system.
Calibrates the Load 8 Resistance Value. This should be set to the actual resistance of the next higher resistor in the system.
Calibrates the Load 9 Resistance Value. This should be set to the actual resistance of the next higher resistor in the system.
Calibrates the Load 10 Resistance Value. This should be set to the actual resistance of the next higher resistor in the system.
Calibrates the Load 11 Resistance Value. This should be set to the actual resistance of the next higher resistor in the system.
Calibrates the Load 12 Resistance Value. This should be set to the actual resistance of the next higher resistor in the system.
Range
1-12
0.0-6500.0 ohms
0.0-6500.0 ohms
0.0-6500.0 ohms
0.0-6500.0 ohms
0.0-6500.0 ohms
0.0-6500.0 ohms
0.0-6500.0 ohms
0.0-6500.0 ohms
0.0-6500.0 ohms
0.0-6500.0 ohms
0.0-6500.0 ohms
0.0-6500.0 ohms
18
LINE POWER
– A Kycon 3 position locking receptacle is provided for the 6 VDC Universal
Power Supply input.
The Universal Power Supply takes a Standard Power Adapter Cable with Small
Standard Product Plug and Required International Connector (See Options Below).
19
SERIAL COMMUNICATION
– There is a serial port on the rear panel. The RS-232 Port is used for Firmware upgrades and to interface with a PC.
USB COMMUNICATION – There is a USB port on the rear panel. The USB Port is used to interface with a PC.
POWER SWITCH - The main power switch for the Analyzer is located on the left side on the rear panel.
20
KEYS
Ten tactile-touch keys are provided for system operation:
– In the Main Screen, these keys will scroll through the available display screens.
In the GRAPH MODE, these keys will scroll through the horizontal zoom level
for the graph.
– In the Main Screen, these keys will toggle through the available Display Zones.
In the SETUP MODE, these keys will scroll through the available parameters.
In the GRAPH MODE, these keys will select the waveform to be displayed.
– In the Main Screen, these keys will scroll through the available parameters.
In the SETUP MODE, these keys will scroll through the available settings
for the parameters.
In the GRAPH MODE, these keys will Scroll through the selected data set.
– This key is used to toggle between the entering the SETUP MODE and the
LOAD SETUP MODE, where the calibration can be viewed and adjusted.
21
– In the SETUP MODE, this key is used to exit and return to the previously viewed
Main Screen. This will also save any changes to the internal EEPROM memory
so they will be retained even with the power turned off.
In the GRAPH MODE, this key is used to exit and return to the previously
viewed Main Screen.
In the SAVE MODE, this key is used to exit without saving.
– In the Main Screen, this key is used to enter the GRAPH MODE.
In the GRAPH MODE, this key is used to enter the SAVE MODE.
In the SAVE MODE, this key is used to save the data set.
– This key is used to toggle the HOLD MODE on and off. The HOLD MODE will
lock the latest reading into the display and the current waveform in the register.
In HOLD MODE, a small “HOLD” message will be displayed in the upper right
corner of the screen.
22
POWER UP SETTINGS
The ESU-2050 Series allows the user to customize the settings that the unit will have on
Power Up. The “Power up with” parameter in the System Setup Menu allows for the selection of either Default or Custom selections.
Use to enter the SETUP MODE. Use to select the “Power up with” parameter.
Use to change the parameters to Default, Custom or Set Current as Custom.
The Setup screen can be exited using the key.
Default
If this option is selected, the unit will Power Up to the One Display Zone screen, showing the mV reading. The default parameters shown in the other Main Screens will be the same as shown in the Main Screens Section.
Custom
If this option is selected, the unit will Power Up using the unique sets of parameters that were last customized and saved by the user. Each Main Screen will use the parameters in the Display Zones that were last configured and saved by the user.
23
Set Current as Custom
This choice is provided to create the set of custom startup screen parameters. The user simply configures each of the five display screens to show the desired parameters in each
Display Zone, selects this option and presses RETURN. The current configuration is then saved as the Custom Power Up values and will be used when the “Power up with” parameter is set to Custom. This configuration will remain the Custom configuration until it is written over using the Set Current as Custom option in the “Power up with” parameter.
24
GRAPH MODE
The GRAPH MODE allows the user to view the measured waveform in the display. The horizontal axis can be zoomed in to display higher frequency waveform components. The vertical axis is auto-scaling and cannot be adjusted. Any of the stored waveforms can be graphed. Additionally, if the unit is placed in the HOLD MODE, the user can adjust which portion of the waveform is being displayed.
NOTE: Due to the limited number of pixels in the display, this should not be used as a calibrated reference, rather as a quick check of the waveform being measured.
Use to enter the GRAPH MODE.
Use to exit the GRAPH MODE.
Selecting a Waveform
Use to select the waveform to be graphed, Ram or Location 1-3.
Waveform
Location
RAM
L1
L2
L3
25
Location Indicator
Use to select a specific portion of the waveform buffer to be graphed.
The Location Indicator is a small square that moves along the bottom of the Graph Screen to indicate where the current viewing window portion of the waveform is from within the overall data set.
Zooming
Location
Indicator
Use to zoom the graph in and out.
The Zoom Indicator is a bar that moves along the left side of the Graph Screen to indicate the Zoom level shown in the current viewing window. It adjusts from Fully Out (-) to
Fully In (+).
Zoom
Indicator
Fully Zoomed Out Fully Zoomed In
26
Saving
To save the displayed waveform, use to enter the SAVE MODE.
Use to select the desired storage location, then to save the waveform, or can be pressed to cancel the save function. Once the save is complete, the newly saved waveform will be displayed.
27
ERROR MESSAGES
Several error messages are provided to indicate invalid operating conditions. Any values that are over range will be displayed as dashes.
Watts Calculation out of range
Peak Voltage Input Out of range
When the input voltage rises above the range that is measurable by the system, the
“WARNING Input Overload” message will be shown.
Input Voltage Overload
NOTE: Although the input is protected from damage at these voltages, the user should immediately remove any input voltage when this message is shown.
28
DFA
®
TECHNOLOGY
DFA
®
Digital Fast Acquisition Technology is a revolutionary new method of measuring
ESU generator output power. A high-speed analog to digital converter is used to digitize the high frequency, high power output of the ESU generator. An RF Current Transformer is used to convert the current signal to a voltage signal, which is read by the analog to digital converter. By digitizing the signal a more accurate, frequency independent measurement can be made.
29
COMMUNICATION PROTOCOL
The communication protocol provides a means to completely configure and use the
ESU-2050 from a PC. All of the functions available through the front panel can be performed through the communication ports. All of the measurements made by the
ESU-2050 are accessible as well. This provides for hands free or automated operation of the ESU-2050.
Communication Ports
The ESU-2050 has two communication ports. Both ports use the same command format. The Serial port is configured as 115,200 Baud Rate, 8 Data Bits, 1 Stop Bit, and No Parity. The USB port appears to a PC as a serial port and is configured for
748,800 Baud Rate, 8 Data Bits, 1 Stop Bit, and No Parity.
Command Syntax
The command description is broken into 3 columns; the KEYWORD, the PARAMETER
FORM and COMMENTS.
Keyword
The KEYWORD column provides the name of the command. The actual name of the command consists of one or more keywords since SCPI commands are based on a hierarchical structure, also known as a tree system.
In such a system, associated commands are grouped together under a common node in the hierarchy, analogous to the way leaves at a same level are connected at a common branch. This and similar branches are connected to fewer and thicker branches, until they meet at the root of the tree. The closer to the root, the higher a node is considered in the hierarchy. To activate a particular command, the full path to it must be specified.
This path is represented in the following tables by placing the highest node in the left-most position. Further nodes are indented one position to the right, below the parent node.
The highest level node of a command is called the Keyword, followed by the Node,
Subnode, and then the value.
30
Not all commands require the complexity of the full command path. For example, the
Status? command doesn’t have a Node or Subnode.
Some commands allow for reading and writing data and some commands are Read Only.
To indicate a read function, a question mark (?) is placed at the end of the command path.
For example, a write command to change the load resistance to 100.5 ohms would be
“CONFigure:LOAD:VALue 100.5<cr>”, where <cr> indicates a carriage-return. For example, a mArms read command would be “READ:MArms?<cr>”, which would return a value of “xxx.x<cr><lf>” where <cr> is a carriage-return and <lf> is a linefeed.
Lowercase letters indicate the long-form of the command (for example,
CONFigure:INPut:RANGe?) and can be omitted for simplification. Uppercase letters indicate the abbreviated, or short-form, of the commands and must be included (for example, CONF:INP:RANG?).
All commands sent to the unit are terminated with a Carriage Return.
NOTE: Commands can be entered in either upper or lowercase or a mixture of the two, uppercase and lowercase. Commands sent to the ESU
–2050 are not case sensitive.
Upper and lower cases are only used when documenting the commands.
Parameter Form
The PARAMETER FORM column indicates the number and order of parameters in a command and their legal values. Parameter formats are listed in angle brackets (<>) while string parameters are simply listed.
Square brackets ([]) are used to enclose one or more parameters that are optional.
The vertical bar (|) can be read as “or” and is used to separate alternative parameter options.
The query form of a command is generated by appending a question mark (?) to the last keyword. However, not all commands have a query form, and some commands exist only in the query form. The COMMENTS column is used to indicate this.
Comments
The COMMENTS column indicates any notes.
31
CONFigure Subsystem
This group allows the user to setup the display and operational settings for the unit.
KEYWORD
CONFigure
:DISPlay
PARAMETER FORM COMMENTS
:SxZy S<display_screen_number> Z<zone_number> nn display_screen_number = 1-7 nn = Parameter for selected Zone
1 = One Parameter
2 = Two Parameters
3 = Three Parameters
4 = Four Parameters
5 = Five Parameters
0 = mV RMS
1 = mA RMS
2 = Watts RMS
3 = mV Peak
4 = mV Pk-Pk
7 = Creast Factor
8 = Time-Pulse On
9 = Time-Pulse Off
10 = Time-Total Cycle
11 = % Duty Cycle
6 = Measurement List Display (Non-editable)
7 = Quick Configuration Screen (Non-editable)
:SCReen
:AVERaging
:HOLD ON | OFF
:LOAD
:MODE
:VALue
< numeric_value >
5 = mV Pk / mV Pk-Pk
6 = mV Pk+
FAST | SLOW | MEDium
TABle | ADJustable
< numeric_value >
:SETn
:NUMber
:Ln
< numeric_value >
< numeric_value >
< numeric_value >
12 = mV Pulse Cycle
13 = mA Pulse Cycle
14 = Watts Pulse Cycle
Range 1-7
1-5 = # display zones
6=Measurement List Display
7=Quick Config Screen
Table Mode: 1-12 for individual resistors, 13-22 for resistor sets 1-10
Adjustable Mode: 0-6500.0 n = Resistor set to configure
(1-10)
< numeric_value > = 16 bit binary value of resistors to include in set ‘n’ where bit
0 = Load 1, Bit 1 = load 2…
Bit 11 = Load 12 each bit selects whether the load is included in the set.
Bit = 1 includes the load
Bit = 0 excludes the load. set of valid numeric values n = Load to configure
< numeric_value > = Actual value of load ‘n’;
0-6500.0 ohms
INPut:
Atten:
RANGe:
NUMsamples:
MODE:
0.1 | 1
100 | 1000 | AUTo
1024 | 2048 | 4096 | 8192 | 16384 | 54768
CONTinous | PULs
32
SYSTem Subsystem
This group allows the user to setup the startup mode for the unit, as well as directly control the unit, as if pressing the keys on the front panel.
PARAMETER FORM COMMENTS KEYWORD
SYSTem:
POWer
CONtrast
KEY
VER?
DEFaults | CUStom | SETCurrent
< numeric_value >
DUP | DDN | SUP | SDN | VUP |
VDN | SETup | RETurn | GSAVe |
HOLD
Numbers 1-20
Read only
READ Subsystem
This group allows the user to get measurements from the unit.
KEYWORD
READ:
PARAMETER FORM
MVrms | MArms | Warms | MVPeak |
MVPP | MVP-PP | MVPK+ | CF |
TON | TOFF | TCYC | DCYC |
MVCyc | MACyc | WCyc
DATA
LOCn
COMMENTS
Read only
Read only
Read only, ‘n’ = stored waveform location to read (1-3)
STATus Sub-system
This subsystem provides status on the operating mode of the unit including messages that would normally be seen on the display.
KEYWORD PARAMETER FORM COMMENTS
STATus?
Bit Value
0
1
1
2
2
3
4
5
4
8
16
32
6 64
7 128
Definition
Hold Mode
Graph Mode
Calibration Mode
8 256
9 512
10 1024
11 2048
Error Present mV Out of Range mA Out of Range
Watts Out of Range
12 4096 mV Peak Out of Range
13 8192 Crest Factor Out of Range
14 16384
15 32768
Read Only
33
ESU-2050 Communication Command Summary
Keywords Nodes
CONFigure DISPlay
Subnodes Values
SxZy nn x is the Screen # (1-5) and y is the Zone # (1-5). nn=0 to 13:
0=mV RMS
1=mA RMS
2=Watts RMS
3=mV Peak
4=mV Pk-Pk
5=mV Pk / mV Pk-Pk
6=mV Pk+
7=Crest Factor
8=Time-Pulse On
9=Time-Pulse Off
10=Time-Total Cycle
11=% Duty Cycle
12=mV Pulse Cycle
13=mA Pulse Cycle
14=Watts Pulse Cycle
SCReen
Range: 1-7
1-5 = # display zones
6=Measurement List Display
7=Quick Config Screen
FAST, SLOW, MEDium
HOLD
LOAD
AVERaging
ON,OFF
MODE
VALue
TABle, ADJustable
Table Mode: 1-12 for individual Resistors, 13-22 for resistor sets 1-
10
Adjustable Mode: 0-6500.0 Ohms
SETn xxxx n = Resistor set to configure, 1-10
XXXX = 16 bit Binary value of resistors to include in set 'n'
Where Bit 0 = Load 1, Bit 1 =Load 2… Bit 11 = Load 12
Each bit selects whether the load is included in the set.
Bit = 1 includes the load, Bit = 0 excludes the load.
Example: a value of 9 would select loads 4 and 1 and exclude everything else
NUMber
1-12 (determines the number of loads present in system)
Ln xxxx n = Load to configure xxxx = Actual value of load 'n', 0-6500.0 ohms.
INPut ATTen 0.1, 1 (donut attenuation)
RANGe 100, 1000, AUTo
NUMsamples 1024, 2048, 4096, 8192, 16384, 32768
MODE CONTinuous, PULsed
SYSTem POWerup DEFaults, CUStom, SETCurrent
CONtrast 0-20
KEY DUP, DDN, SUP, SDN, VUP, VDN, SETup, RETurn, GSAVe, HOLD
VERsion? [read only]
34
ESU-2050 Communication Command Summary
Keywords Nodes
READ
Subnodes Values
MVrms? Returns: mV RMS [read only]
MArms? Returns: mA RMS [read only]
WArms? Returns: Watts RMS [read only]
MVPeak? Returns: mV Peak [read only]
MVPP? Returns: mV Peak to Peak [read only]
MVP-PP? Returns: mV Peak/Peak to Peak [read only]
MVPK+? Returns: mV Positive Peak [read only]
CF?
TON?
Returns: Crest Factor [read only]
Returns: Time - Pulse On [read only]
TOFF?
TCYC?
Returns: Time - Pulse Off [read only]
Returns: Time - Total Cycle [read only]
DCYC? Returns: % Duty Cycle [read only]
MVCyc? Returns: mV Pulse Cycle [read only]
MACyc? Returns: mA Pulse Cycle [read only]
WCyc? Returns: Watts Pulse Cycle [read only]
DATA?
Returns: Entire Data Buffer [read only] Length = NUMSamples * 2
Data Format is mV in signed double byte format, Decimal place is assumed based on Input
Range (Low Range = 2DP, High Range = 1DP)
LOCn?
Returns: Saved Data Buffer at Location 'n' , n= 1-3 [read only] Length = NUMSamples * 2
Data Format is mV in signed double byte format, Decimal place is assumed based on Input
Range (Low Range = 2DP, High Range = 1DP)
STATus?
Bit Value Definition
0 1 Hold Mode
1 2 Graph Mode
2 4 Calibration Mode
3 8
4 16
5 32
6 64
7 128
8 256 Error Present
9 512 mV Out of Range
10 1024 mA Out of Range
11 2048 Watts Out of Range
12 4096 mV Peak Out of Range
13 8192 Crest Factor Out of Range
14 16384
15 32768
35
MANUAL REVISIONS
Revision #
Rev 01
Rev 02
Rev 03
Rev 04
Rev 05
Rev 06
Rev 07
Rev 08
Rev 09
Rev 10
Program #
DT7376A
DT7376A
DT7376B
DT7376C
DT7376CD
DT7376CE
DT7376CE
DT7376CE
DT7376CE
DT7376CE
Revisions Made
Origination
Specifications Updates
Factory Setup, Graph/Save Functions, PC Interface and Format Updated
Input Range, Input Mode, Load Sets, Load Table and Specifications Updated
New Measurements Added and Specs Updated,
Load Setup Screen and Measurement List Screen
Implements 2050P model, Communication Package
100mV Accuracy Specification Updated,
Miscellaneous edits
Address Updated
Miscellaneous Edits
Specifications Updated, Format Updated, Photos
Updated
LIMITED WARRANTY
WARRANTY : BC GROUP INTERNATIONAL, INC. WARRANTS ITS NEW PRODUCTS TO BE FREE
FROM DEFECTS IN MATERIALS AND WORKMANSHIP UNDER THE SERVICE FOR WHICH THEY
ARE INTENDED. THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM THE DATE OF
SHIPMENT.
EXCLUSIONS: THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR
IMPLIED, INCLUDING, BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY
OR FITNESS FOR A PARTICULAR PURPOSE.
BC GROUP INTERNATIONAL, INC. IS NOT LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL
DAMAGES.
NO PERSON OTHER THAN AN OFFICER IS AUTHORIZED TO GIVE ANY OTHER WARRANTY OR
ASSUME ANY LIABILITY.
REMEDIES: THE PURCHASER'S SOLE AND EXCLUSIVE REMEDY SHALL BE: (1) THE REPAIR OR
REPLACEMENT OF DEFECTIVE PARTS OR PRODUCTS, WITHOUT CHARGE. (2) AT THE OPTION
OF BC GROUP INTERNATIONAL, INC., THE REFUND OF THE PURCHASE PRICE.
P:/Manuals/BCGroup/…/ESU-2000/ESU-2050_Series_UM_Rev10.doc
36
SPECIFICATIONS
INPUT
INPUT
COMPATIBILITY
VOLTAGE (RMS)
VOLTAGE (pk, pk-pk)
ACCURACY,
100 mV RANGE
RF MEASUREMENT
IMPEDANCE
CONNECTION
MAXIMUM
VOLTAGE
FREQUENCY
CURRENT
TRANSFORMER
50 Ω
BNC ( 50 Ω)
3.3 V pk-pk
Internally Protected
10 kHz – 10 MHz
Pearson Electronics (Typical)
ATTENUATION
RATIOS
100 mV RANGE
1000 mV RANGE
100 mV RANGE
1000 mV RANGE
Input
≤ 10.0 mV
10.0 mV < Input ≤
35.0 mV
Input
> 35.0 mV
0.1:1 (Pearson Model 411)
1:1 (Pearson Model 4100)
User Selectable
0.20 – 70.00 mV RMS
0.01 mV Resolution
2.0
– 700.0 mV RMS
0.1 mV Resolution f ≤ 1.0 MHz
0.20 – 100.0 mV
0.01 mV Resolution
2.0 – 1000.0 mV
0.1 mV Resolution
1.0 MHz < f ≤
2.5 MHz f > 2.5 MHz
± 1% Reading
Or ± 0.25 mV
± 1% Reading
Or ± 0.25 mV
± 1% Reading
Or ± 0.25 mV
± 1% Reading
Or ± 0.25 mV
± (1% Reading
+ 0.5% Range)
± (1% Reading
+ 0.5% Range)
± 1% Reading
Or ± 1.0 mV
± (6% Reading +
0.5% Range)
± (6% Reading +
5% Range)
ƒ ≤ 2.5 MHz ƒ > 2.5 MHz
ACCURACY,
1000 mV RANGE
Input ≤ 50 mV
50 mV < Input ≤
400.0 mV
Input > 400.0 mV
± 1% Reading
Or ± 1 mV
± 1% Reading
± (1% Reading +
0.25% Range)
± 1% Reading
Or ± 1 mV
± 4% Reading
± (4% Reading +
0.25% Range)
37
POWER
CALCULATED RANGES
CURRENT
(with 0.1:1 CT)
CURRENT
(with 1:1 CT)
VOLTAGE (mV pk, mV pk-pk)
POWER
CREST FACTOR
100 mV RANGE
1000 mV RANGE
100 mV RANGE
1000 mV RANGE
100 mV RANGE
1000 mV RANGE
2.0 - 700.0 mA RMS
0.1 mA Resolution
20 - 7000 mA RMS
1 mA Resolution
0.20 - 70.0 mA RMS
0.01 mA Resolution
0.2 - 700.0 mA RMS
0.1 mA Resolution
0.0 - 1.0 mV
0.1 mV Resolution
0.0 - 1.0 mV
0.1 mV Resolution
0 - 999.9 W
0.1 W Resolution
1.4 – 500
0.1 Resoluton
PULSE MODE TIMING MEASUREMENT (ESU-2050P ONLY)
RESOLUTION 0.1 ms
ACCURACY ± 0.2 ms
PHYSICAL, ENVIRONMENTAL, AND ELECTRICAL
DISPLAY
128 X 64 Pixels Graphical LCD,
White LED Backlight
SETUP EEPROM, All Parameters
MEMORY
OPERATING
RANGE
STORAGE RANGE
SIZE
CONSTRUCTION
WEIGHT
DATA STORAGE 3 Sets of 32768 Data Points
RETENTION 10 Years w/o Power
ENCLOSURE
FACE
15 - 30 ° C
20 - 80% RH, Non-Condensing
-40 - 60 ° C
3.4” x 9.1” x 8.0”
86.4 x 231.4 x 203.2 mm
ABS Plastic
Lexan, Back Printed
≤ 3 lbs. (1.36 kg)
CONNECTIONS
RS-232 COMM.
USB
DB-9 (Female) Receptacle
USB ‘B’ Receptacle, USB 2.0 Compliant
POWER Kycon 3-position Locking Receptacle
6 VDC, minimum 500 mA
38
NOTES
39
NOTES
40
NOTES
41
NOTES
42
BC GROUP INTERNATIONAL, INC.
3081 ELM POINT INDUSTRIAL DRIVE
ST. CHARLES, MO 63301
USA
1-800-242-8428
1-314-638-3800 www.bcgroupintl.com
ESU-2050 SERIES User Manual
03/12 – Rev 10
Copyright © 2012
Made in the USA

Download
Advertisement
Key features
TRUE RMS READINGS USING DFA® TECHNOLOGY
INDUSTRY STANDARD CURRENT SENSING TECHNOLOGY
LARGE GRAPHICS DISPLAY WITH CURSOR
1% OF READING MEASUREMENT ACCURACY
DIGITAL DATA OUTPUT VIA USB AND RS232
STANDARD (1000 MV) AND LOW RANGE (100 MV)
INTERNALLY PROTECTED INPUT CIRCUITRY
INTERNAL DATA STORAGE FOR 3 FULL DATA SETS
Frequently asked questions
The ESU-2050 Series offers a higher degree of accuracy than previously attainable with conventional Electrosurgical Unit Analyzer designs, with an accuracy of 1% of reading.
DFA® Digital Fast Acquisition Technology is a revolutionary new method of measuring ESU generator output power. A high-speed analog to digital converter is used to digitize the high frequency, high power output of the ESU generator.
The ESU-2050 Series has two communication ports: USB and RS232. Both ports use the same command format, allowing for complete configuration and operation of the device from a PC.
To save the displayed waveform, press the ‘GRAPH’ button to enter the SAVE MODE. Use the left and right arrow keys to select the desired storage location, then press the ‘ENTER’ button to save the waveform. The ‘ESC’ button can be pressed to cancel the save function.
The ESU-2050 Series provides several error messages to indicate invalid operating conditions. Any values that are over range will be displayed as dashes. If the input voltage rises above the range that is measurable by the system, the “WARNING Input Overload” message will be shown.