Agilent Technologies 7 400-DS Apparatus Operator’s Manual
Below you will find brief information for 400-DS Apparatus 7. The Agilent 400-DS Dissolution Apparatus is the first compendial small-volume dissolution apparatus for testing medical devices such as drug-eluting stents (DES). This integrated, small-footprint dissolution apparatus features several first-in-class innovations that greatly increase the convenience of release rate testing of combinatorial drug products. The 400-DS is ideal for dissolution testing of extended release products that either contain and / or release small amounts of API over time. All parameters are controlled using the 400-DS Dissolution Workstation software via the PC. Up to 4 units can be fully programmed to execute all dissolution method and sampling parameters from a single PC. Samples can be collected in pre-capped HPLC vials for direct transfer to an HPLC system. Septa are pierced using an innovative needle manifold, which lowers and raises at each sample point.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
400-DS Apparatus 7 Operator’s Manual Agilent Technologies Notices © Agilent Technologies, Inc. 2014 Warranty No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws. The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control. Manual Part Number 70-9064 Edition Rev C, March 2014 Agilent Technologies, Inc. 3501 Stevens Creek Blvd. Santa Clara, CA 95052 USA Technology Licenses The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license. Restricted Rights Legend U.S. Government Restricted Rights. Software and technical data rights granted to the federal government include only those rights customarily provided to end user customers. Agilent provides this customary commercial license in Software and technical data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Software) and, for the Department of Defense, DFARS 252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commercial Computer Software or Computer Software Documentation). Safety Notices CAUTION A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met. WA RNING A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met. 400-DS Operator’s Manual Contents Figures 1 7 Tables 9 Safety 11 Electrical Hazards 12 Warning 13 Caution 13 Note 13 Information Symbols 2 Introduction 14 15 Conventions Used in this Manual Serial Number Format 3 Setting Up the 400-DS 17 17 19 Initial Setup 20 Unpacking Procedure 20 Environmental Requirements for Installation 21 Clearance 21 Parts and Accessories 22 Leveling Your Apparatus 23 Electrical Connection 24 Cable Connections 25 Heater Jacket / Sample Holder Installation 26 Tubing Connections 27 400-DS Operator’s Manual 3 Contents PC Setup 28 Initial Power Up 29 Software Installation and Setup 30 400-DS Workstation Software Installation 30 Attaching the Database 31 Local Security Policy 32 E-mail Notification Configuration (Optional) 34 Starting 400-DS Workstation 36 Dissolution Server Selection 37 Configuring the Application - Adding Users 40 4 Operating the 400-DS 45 400-DS Workstation Logon 46 System Configuration 49 Initial Setup of 400-DS Configuration 49 Copying a System Configuration 53 Deleting a System Configuration 53 System Configuration Report 54 Editing an Existing System Configuration 54 Recovering a Deleted Configuration 55 Show Audit Trail 55 Verify Integrity 56 Import/Export XML File 56 400-DS System Calibration 58 Volume Calibration 58 Temperature Calibration 60 Manual Control/Diagnostics 64 Verifying Sample Tray Control Verifying Media Control 65 Verifying the Cell Temperature Verifying Reciprocation 67 4 64 66 400-DS Operator’s Manual Contents Evaporation Cover Control 67 Method Editor 68 Creating a Method 69 Copying Methods 76 Deleting Methods 76 Editing an Existing Method 76 Method Report 77 Recovering a Method 78 Method Audit Trail 78 Import/Export XML File 79 Active Channel Selection Valves Running a Method 82 Test Reports 87 Electronic Signatures Clean System 5 81 88 88 Troubleshooting and Maintenance 89 Preventive Maintenance 90 Calibration 90 Routine Maintenance (to be Performed Between Each Dissolution Method 90 Sample Holders 91 Troubleshooting 92 Obtaining Warranty and Other Services Index 400-DS Operator’s Manual 92 93 5 Contents This page was intentionally left blank, except for this message. 6 400-DS Operator’s Manual Figures Figure 1. Diagram of the electrical connection 24 Figure 2. Diagram of cable connections 25 Figure 3. License dongle 30 Figure 4. MSDE Manager screen 31 Figure 5. DissoService tools menu 34 Figure 6. DissoService Options screen 35 Figure 7. Logon screen 38 Figure 8. Add New Dissolution Server screen 38 Figure 9. Browse Dissolution Servers screen 39 Figure 10. Configuration Dialog screen 40 Figure 11. Local Users and Groups screen 41 Figure 12. Miscellaneous tab 42 Figure 13. Logon screen 46 Figure 14. 400-DS Workstation Screen 47 Figure 15. System Configuration screen 49 Figure 16. System Editor screen 50 Figure 17. 400-DS Workstation screen 51 Figure 18. System Calibration screen 58 Figure 19. Sample Tray Control box 64 Figure 20. Reciprocation box 67 Figure 21. Methods screen 69 Figure 22. Method Editor screen 70 Figure 23. Automated App7 Sequence tab 72 Figure 24. Automated App7 Options tab 74 Figure 25. System status screen 83 Figure 26. Method Start Options screen 84 Figure 27. Method Sample Information Screen 85 Figure 28. Test Report Selection screen 87 400-DS Operator’s Manual 7 Figures This page was intentionally left blank, except for this message. 8 400-DS Operator’s Manual Tables Table 1. Table 2. Table 3. Table 4. Table 5. Table 6. Table 7. Table 8. 400-DS Operator’s Manual 400-DS Workstation Screen Options 48 System Editor Screen 1 Options 50 System Editor Screen 2 Options 52 Media Control Options 65 Parameters Tab Options 71 Automated App7 Sequence Options 73 Automated App7 Options 75 Method Start Options 84 9 Tables This page was intentionally left blank, except for this message. 10 400-DS Operator’s Manual 400-DS Apparatus 7 Operator’s Manual 1 Safety Electrical Hazards 12 The 400-DS has been designed and tested so that when used properly you have an accurate, fast, flexible, and safe instrument. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. Operation of an Agilent 400-DS involves the use of aqueous liquids and various pharmaceutical dosage forms. Unskilled, improper, or careless use of this instrument can create shock hazards, fire hazards, or other hazards which can cause death, serious injury to personnel, or severe damage to equipment and property. Information on safety practices is provided with your instrument and operation manuals. Before using your instrument or accessories, you must thoroughly read these safety practices. Observe all relevant safety practices at all times. Agilent Technologies 11 1 Safety Electrical Hazards The 400-DS contains electrical circuits, devices, and components operating at dangerous voltages. Contact with these circuits, devices, and components can cause death, serious injury, or painful electric shock. Panels or covers that are retained by fasteners which require the use of a tool for removal may be opened only by Agilent-trained, Agilent-qualified, or Agilent-authorized service engineers. Consult the manuals or product labels supplied with the 400-DS to determine which parts are operator-accessible. Application of the wrong supply voltage, connection of the instrument to an incorrectly wired supply outlet, or lack of proper electrical grounding can create a fire hazard or a potentially serious shock hazard and could seriously damage the instrument and any attached ancillary equipment. Always use a three-wire outlet with ground connection which is adequately rated for the load. The installation must comply with local, state, and federal safety regulations. Do not connect the instrument to the main power supply until you have made sure that the operating voltage is correctly set for the main power supply in the specific outlet in your laboratory to which the equipment will be connected. 12 400-DS Operator’s Manual Safety 1 Warning WA RNING A ‘Warning’ message appears in the manual when failure to observe instructions or precautions could result in death or injury. Read all warnings and cautions carefully and observe them at all times. Caution CAUTION A ‘Caution’ message appears in the manual when failure to observe instructions could result in damage to equipment (Agilent supplied and / or other associated equipment). Note NOT E A ‘Note’ appears in the manual to give advice or information. 400-DS Operator’s Manual 13 1 Safety Information Symbols I 0 Switches main power on Switches main power off Indicates single-phase alternating current Indicates the product complies with the requirements of one or more European Union (EU) directives. Indicates specific equipment meets standards of safety. These products are safe for use in the workplace for North America. Indicates that this product must not be disposed of as unsorted municipal waste. All Agilent products that are subject to the WEEE directive shipped after August 13, 2005 are compliant with the WEEE marking requirements. Such products are marked with the “crossed out wheelie bin” WEEE symbol in accordance with European Standard EN 50419. For more information on collection, reuse, and recycling systems, please contact your local/regional waste administration, your local distributor, or Agilent. Indicates the product complies with regulatory compliance requirements of New Zealand and Australia. 14 400-DS Operator’s Manual 400-DS Apparatus 7 Operator’s Manual 2 Introduction Conventions Used in this Manual Serial Number Format 17 17 Agilent Technologies 15 2 Introduction The Agilent 400-DS Dissolution Apparatus is the first compendial small-volume dissolution apparatus for testing medical devices such as drug-eluting stents (DES). This integrated, small-footprint dissolution apparatus features several first-in-class innovations that greatly increase the convenience of release rate testing of combinatorial drug products. The 400-DS is ideal for dissolution testing of extended release products that either contain and / or release small amounts of API over time. All parameters are controlled using the 400-DS Dissolution Workstation software via the PC. Up to 4 units can be fully programmed to execute all dissolution method and sampling parameters from a single PC. Samples can be collected in pre-capped HPLC vials for direct transfer to an HPLC system. Septa are pierced using an innovative needle manifold, which lowers and raises at each sample point. Sample trays are available in 1.5 mL and 4 mL sizes. 16 400-DS Operator’s Manual Introduction 2 Conventions Used in this Manual • Items you are asked to click are in bold. For example, “Click OK.” Serial Number Format The serial number contains 10 characters and follows this syntax: CC1234xxxx Syntax Code Meaning Description CC Country of origin 2 alpha characters matching the required trade designation for the country of origin 12 Year of manufacture ‘09’ for 2009, ‘10’ for 2010, etc. 34 Week of manufacture ‘01’ for week 1 to ‘52’ for week 52 400-DS Operator’s Manual 17 2 Introduction This page was intentionally left blank, except for this message. 18 400-DS Operator’s Manual 400-DS Apparatus 7 Operator’s Manual 3 Setting Up the 400-DS Initial Setup 20 PC Setup 28 Initial Power Up 29 Software Installation and Setup 30 Agilent Technologies 19 3 Setting Up the 400-DS Initial Setup Complete the following sections to initially set up the Agilent 400-DS and all other system components. Unpacking Procedure The Agilent 400-DS is shipped in one crate. The crate contains the apparatus and accessories. 1 Open the crate and check the contents for damage, which may have occurred during shipping. Shipping damage rarely occurs, but if it does contact both the carrier who delivered the instruments and the Dissolution Systems Service Department. Though claims for damage should be filed with the carrier, we can help you file a claim. 2 Remove the top of the crate. 3 Remove one side of the crate and lift the accessories box out of the crate. 4 Disassemble the other three sides of the crate. 5 Remove any cushioning material and tape. If any item is missing, contact the Dissolution Systems Service Department or your local representative for replacements. 6 Carefully remove the apparatus from the base of the shipping crate. WA RNING 20 Because of its heavy weight, two people should lift the apparatus. Lift by holding underneath the base of the unit on each side. Support feet are located at each corner of the unit. Hands should be positioned between the support feet to ensure they are not trapped when lowering. 400-DS Operator’s Manual Setting Up the 400-DS 3 7 Place the apparatus on a clean, dry, and level section of the bench. NOT E The electrical connection at the back of the apparatus is the primary disconnect for the instrument. The apparatus should be positioned to allow accessibility to the power cords and circuit breaker switch for easy disconnection. Environmental Requirements for Installation • Humidity: max relative humidity 80% for temperatures up to 31 ºC decreasing linearly to 50% relative humidity at 40ºC • Indoor use only • Pollution Degree: 2 • Installation Category: II • Altitude: 2000 m • Temperature: 5 ºC to 40ºC • Power: 115/230 V ac, 50/60 Hz, 8 A Main supply voltage fluctuations are not to exceed ± 10% of the nominal supply voltage Clearance 1 Prepare the area where the equipment is to be located. 2 Ensure a minimum clearance of 40 cm above the unit and 10 cm at the rear and on both sides of the Agilent 400-DS. Approximately 65 cm x 65 cm total bench space is required per apparatus. NOT E Ensure space is also available for the PC, media containers, and waste. 3 Orient the Agilent 400-DS and all other system components appropriately. 400-DS Operator’s Manual 21 3 Setting Up the 400-DS Parts and Accessories Locate the following items, as applicable for your system configuration: Item 22 Quantity Agilent 400-DS Automated Apparatus 7 1 AC power cord 1 DB-9 cable 1 6-pin Mini-DIN cable 1 Sample cells 13 Cell heater jackets 13 Sample holders 13 Sample holder extraction tool 1 Roll of tubing (25 feet) 1 Sample tray with 12 rows 1 Additional sample tray for calibration 1 Magnetic plate 1 Evaporation cover with 2 guide rod stabilizer plates 1 2.5 mm Allen wrench located on rear of 400-DS 1 9/64” Allen key screwdriver 1 1.5 mm Allen key screwdriver 1 Modified temperature probe (calibration certificate included) 1 Sample vials, 12 x 32, Qty. 100 2 Sample vials, 15 x 35, Qty. 100 1 Level 1 400-DS Workstation software 1 License dongle 1 Technical Documentation CD 1 400-DS Operator’s Manual Setting Up the 400-DS 3 Leveling Your Apparatus 1 Place the level provided in the accessory kit on the vessel table. 2 Check the level of the instrument in the front and rear center of the unit as well as the left and right sides. 3 The apparatus has adjustable feet under each corner. These may be screwed in or out to raise or lower each leg. Adjust the feet so the bubble inside the level is within the circle at all four points. 400-DS Operator’s Manual 23 3 Setting Up the 400-DS Electrical Connection NOT E The electrical connection at the back of the apparatus is the primary disconnect for the instrument. The apparatus should be positioned to allow accessibility to the power cords and circuit breaker switch for easy disconnection. 1 Verify the Agilent 400-DS power switch is in the off position. 2 Connect the power cord between the receptacle on the drive unit rear panel and an outlet of the appropriate voltage. Figure 1 Diagram of the electrical connection Consult your monitor and printer manuals for details of their individual cabling requirements. 24 400-DS Operator’s Manual Setting Up the 400-DS 3 Cable Connections Figure 2 Diagram of cable connections Single System Setup 1 Connect a cable between COM1 on the PC and the left serial port (IN COM A) on the rear of the Agilent 400-DS Automated Apparatus 7. 2 Ensure the loop back switch (located to the right of the right serial port) is switched to the left (ON). Multiple System Setup 1 Connect a cable between COM1 on the PC and the left serial port (IN COM A) on the rear of the Agilent 400-DS Automated Apparatus 7. 2 Ensure the loop back switch (located to the right of the right serial port) is switched to the right (OFF). 3 Connect a 6-pin mini-DIN cable between the right serial port (OUT COM B) of the first apparatus and the left serial port of the second apparatus. 4 If this is the final Agilent 400-DS Automated Apparatus 7 in the series, ensure the loop back switch (located to the right of the right serial port) is switched to the left (ON). If this is not the final Agilent 400-DS Automated Apparatus 7 in the series, ensure the loop back switch (located to the right of the right serial port) is switched to the right (OFF). 5 Repeat steps 3 and 4 for each additional apparatus. 400-DS Operator’s Manual 25 3 Setting Up the 400-DS Heater Jacket / Sample Holder Installation 1 Locate the 1.5 mm Allen wrench to install the sample heater jackets. 2 Install the first cell heater by placing it over the thermistor at the desired cell position of the Agilent 400-DS. 3 While holding the jacket in place, use the Allen wrench to tighten the screw. Do not overtighten the screw. 4 Insert the 2-pin plug for the heater in the appropriate location on the sample heater jacket. 5 Repeat steps 2 - 4 to install the sample heater jackets for the remaining cell positions. 6 Install the magnetic plate on the reciprocation guide rod, which is the shorter rod in the middle of the 3 rods on the Agilent 400-DS. 7 Tighten the bolt on each end of the magnetic plate using the 9/64" Allen wrench. 8 Remove the guide rod stabilizers from the bottom of the evaporation cover. 9 Install the guide rod stabilizers on top of the magnetic plate using the two longer guide rods for alignment. 10 Tighten the two bolts on each stabilizer plate using the 9/64" Allen wrench. 11 Locate the sample cells and sample holders provided with your instrument. 12 Remove the sample cells and holders from the packing material to ensure no damage occurred during shipment. 13 To install a sample cell in the heater jacket, slide one end through the magnetic lift plate. Use the finger tabs on the heater jacket to spread it open. 14 Continue to lower the sample cell until it is secure and completely covers the bottom cell seal. CAUTION 26 Be careful when sliding the sample cells over the O-ring seals as damage can occur. Ensure the cell and seal are properly aligned before applying pressure. 400-DS Operator’s Manual Setting Up the 400-DS 3 15 With the magnet at the top, insert the sample holder into the sample cell. 16 Repeat steps 13 - 15 to install the remaining sample cells and holders. 17 Replace the evaporation cover and tighten the two thumb screws. Tubing Connections 1 If necessary, install the tubing connectors at each location of the sample valve, rinse, and waste lines. 2 Connect an adequate length of tubing to each location of the sample valve, rinse, and waste lines. 3 Place each media / rinse / waste line into an appropriate media / rinse / waste container. CAUTION If one of the provided media bottle caps is used as a waste container, remove the dip tubes before attaching the waste line. If an alternative waste container is used, ensure that the waste connection is not capable of siphoning back into the instrument. The inlet tubing must never fall below the level of waste in the container. The media bottles should be located at the same bench level as the 400-DS. 400-DS Operator’s Manual 27 3 Setting Up the 400-DS PC Setup The 400-DS Workstation software requires Microsoft Windows XP®, or Windows® 7 to be installed on the computer. For instructions on installing the operating system, refer to the documentation supplied with Windows. Minimum PC Configuration Windows XP or Windows 7 256 MB RAM 1024 x 768 pixel display resolution 10 GB free hard disk drive space 1x USB port If the PC is not equipped with a standard serial port, a USB-to-serial converter must be purchased. Refer to the manufacturer’s documentation for installation instructions. If Agilent is installing the 400-DS Workstation software for you, note that the installation of a Windows® operating system is not included as part of the standard instrument installation. The monitor, keyboard, mouse, and printer/plotter are connected to the PC via cables which plug into the back of the PC. 28 400-DS Operator’s Manual Setting Up the 400-DS 3 Initial Power Up 1 Ensure the circuit breaker on the rear of the instrument is on. 2 Turn on the power switch on the front panel and ensure the integrated red LED illuminates. 3 Repeat steps 1 and 2 for each Agilent 400-DS Automated Apparatus 7. 4 Turn on the PC, monitor, and printer. Verify the PC power light illuminates and the operating system initializes. 400-DS Operator’s Manual 29 3 Setting Up the 400-DS Software Installation and Setup NOT E To complete the installation of the 400-DS Workstation software, you must log on to the PC as an administrator. Attaching the Database and Local Security Policy are included as part of the software installation procedures. The CD contains all of the requirements for an installation. 400-DS Workstation Software Installation 1 Install the 400-DS Workstation CD and access the files. 2 Execute the file setup.exe and follow the on-screen prompts. NOT E If installing the software on a Windows 7 PC, right-click setup.exe and select Run as administrator. 3 When asked if you want to install the Microsoft .Net Framework V1.1, you must click Yes. 4 After the installation has successfully completed, install the license dongle by plugging it into an open USB port. The dongle LED illuminates indicating proper driver installation. Figure 3 30 License dongle 400-DS Operator’s Manual Setting Up the 400-DS 3 Attaching the Database Once the basic installation is complete, complete the following steps to attach the database files: 1 Click Start > All Programs > Agilent > Dissolution > MSDE Manager. The MSDE Manager screen displays. Figure 4 MSDE Manager screen 2 Select a machine name from the list of available servers. When connecting to a database installation on a local machine, the encrypted password is automatically entered in the password box. When connecting to a remote database, your system administrator may have to provide you with a password. 3 Click Connect to connect to the server or local PC. 400-DS Operator’s Manual 31 3 Setting Up the 400-DS 4 Select the Attach tab on the MSDE Manager screen and attach the database by clicking Attach. Leave all the values at their default settings. After successful attachment, the Database Attach screen displays. 5 Click OK. 6 Close the MSDE Manager screen. Local Security Policy For 21 CFR Part 11 compliance purposes, you must ensure that the following minimum requirements are met by your system’s security policy. NOT E It may be necessary to coordinate the following configurations with the domain administrator. 1 Click Start > Run to run the Local Security Settings Manager. Type secpol.msc and press Enter. The Local Security Settings manager displays. 2 Click Security Settings > Account Policies > Password Policy and set the following security policy configuration: 32 Policy Security Setting Enforce Password History 3 passwords remembered Maximum password age 30 days Minimum password length 6 characters Password Must Meet Complexity Requirements Enabled 400-DS Operator’s Manual Setting Up the 400-DS 3 3 Click Security Settings > Account Policies > Account Lockout Policy and set the following configuration: Policy Security Setting Account lockout duration 0 minutes (infinite) Account lockout threshold 3 invalid login attempts Reset account lockout counter 99999 minutes 4 Click Security Settings > Local Policies > Audit Policy and set the following configuration: Policy Security Setting Audit account logon events Success, Failure Audit account management Success, Failure Audit logon events Success, Failure Audit policy change Success, Failure 5 Close the Local Security Settings Management screen. 400-DS Operator’s Manual 33 3 Setting Up the 400-DS E-mail Notification Configuration (Optional) 1 On the computer which is hosting the DissoService, click Start > All Programs > Agilent > Dissolution > Service Manager. The DissoService Manager screen displays. 2 On the toolbar, click Tools and select Stop Service. The Options item on the Tools menu enables after the service stops (see Figure 5, “DissoService tools menu,” on page 34). Figure 5 34 DissoService tools menu 400-DS Operator’s Manual Setting Up the 400-DS 3 3 On the toolbar, click Tools > Options. The DissoService Options screen displays (see Figure 6, “DissoService Options screen,” on page 35). Figure 6 DissoService Options screen 4 Configure the SMTP server address. NOT E The e-mail alert software assumes the use of an SMTP server which does not require authentication for sending. Most e-mail servers only require authentication for receiving (POP3), but allow sending from any system on the same domain without the use of a password. If your corporate server does not require passwords, provide the address of the corporate SMTP server in order for Dissolution Server to send e-mail. Enter the SMTP server address as either a name (for example, server.domain.com) or as an IP address (for example, 192.168.1.1). If your corporate SMTP server requires passwords, consider installing Microsoft Internet Information Server (IIS) to a server on your intranet through which other Dissolution Servers could send e-mails. IIS includes a SMTP server which allows unrestricted e-mail sending. 400-DS Operator’s Manual 35 3 Setting Up the 400-DS 5 Once you have configured the SMTP server address, click OK to save the settings. 6 On the toolbar, click Tools and select Start Service. Starting 400-DS Workstation 1 Double-click the 400-DS Workstation icon on the Windows desktop to start the software. 2 If your system has Windows Firewall installed, the following displays: 3 Click Unblock to enable the program. 36 400-DS Operator’s Manual Setting Up the 400-DS 3 Dissolution Server Selection NOT E A dissolution server must be selected when logging into a server for the first time. The same user ID and password that was used to log on to Windows should be used to log on to the 400-DS Workstation software. 1 From the logon screen, enter a domain user identification and password. 2 Click the drop-down arrow corresponding to Dissolution server and select Add New. The Add New Dissolution Server screen displays 400-DS Operator’s Manual 37 3 Setting Up the 400-DS (see Figure 8, “Add New Dissolution Server screen,” on page 38). Figure 7 Logon screen Figure 8 Add New Dissolution Server screen 3 If you know the domain name of the computer on which the Dissolution Server is installed, enter the name in the Server Name box. Alternately, click the network browse button (…) to select a server from a list of computers on your network (see Figure 9, “Browse Dissolution Servers screen,” below). NOT E 38 If your network has a large number of systems connected, browsing the network may take several minutes. 400-DS Operator’s Manual Setting Up the 400-DS Figure 9 3 Browse Dissolution Servers screen 4 The default server port is 8787, but may be changed to any other port. Coordinate the selection of a port number with your I.T. department to satisfy any port restriction requirements they may have. 5 Click OK. 6 Click Logon on the Logon screen. 7 After successfully logging onto the software, click Tools > Options. The Configuration Dialog screen displays (see Figure 10, “Configuration Dialog screen,” on page 40). 400-DS Operator’s Manual 39 3 Setting Up the 400-DS Configuring the Application - Adding Users NOT E In order to configure the application, it is necessary to be logged on as an administrator. Figure 10 Configuration Dialog screen 1 To add a user to a group, select the Security tab on the Configuration Dialog screen. 40 400-DS Operator’s Manual Setting Up the 400-DS 3 2 Click User Administration at the bottom of the screen. The Local Users and Groups screen displays. Figure 11 Local Users and Groups screen 3 Double-click the Groups folder to expand the list of groups. 4 Click Add... to add users to the group. The Select Users, Computers, or Groups screen displays. 5 Enter the names of the users to add to the group and click OK. 6 Repeat steps 3-5 as appropriate for each group that requires new users. 7 Close the window. 400-DS Operator’s Manual 41 3 Setting Up the 400-DS 8 Enter an appropriate length of time in minutes the application can be idle before locking. Figure 12 Miscellaneous tab 9 Click OK. The application has been configured. 42 400-DS Operator’s Manual Setting Up the 400-DS 3 This page was intentionally left blank, except for this message. 400-DS Operator’s Manual 43 3 44 Setting Up the 400-DS 400-DS Operator’s Manual 400-DS Apparatus 7 Operator’s Manual 4 Operating the 400-DS 400-DS Workstation Logon 46 System Configuration 49 400-DS System Calibration 58 Manual Control/Diagnostics 64 Method Editor 68 Active Channel Selection Valves 81 Running a Method 82 Test Reports 87 Clean System 88 Agilent Technologies 45 4 Operating the 400-DS 400-DS Workstation Logon 1 Double-click the Agilent 400-DS Workstation icon on your desktop. The Logon screen displays. Figure 13 Logon screen 46 400-DS Operator’s Manual Operating the 400-DS 4 2 Enter your user identification and password. Verify the domain is correct and click Logon. The 400-DS Workstation screen displays. Figure 14 400-DS Workstation Screen NOT E 400-DS Operator’s Manual Options in the left pane may be grayed out depending on user permissions. 47 4 Operating the 400-DS Following is a description of the screen options listed on the navigation bar of the 400-DS Workstation screen: Table 1 400-DS Workstation Screen Options Option System Method Security Description Configuration Use this option to configure the system. See “System Configuration” on page 49. Diagnostics Use this option to check the diagnostics of a dissolution apparatus or accessory within the system. See “Manual Control/Diagnostics” on page 64. Editor Use this option to create a method, change or delete method parameters, run a report of the method setup, view the method audit trail, and verify the integrity of the method. See “Method Editor” on page 68. Run Method Use this option to run a method. See “Running a Method” on page 82. Test Reports Use this option to run a report of the completed method. See “Test Reports” on page 87. Change User To change the user, click Change User. The Logon screen displays. Enter the appropriate user identification and password and click Logon. Lock Application To lock the 400-DS Workstation screen, click Lock Application. The 400-DS Workstation Locked screen displays. Click the lock or the link to unlock the screen. The Logon screen displays. Enter the appropriate user identification and password as applicable and click Logon. Audit Trail Click Audit Trail. The Security Audit Trail screen displays. Click Show Report to display the report. Change the start date, end date, and/or user identification as desired and click Retrieve Records to change the parameters for the information displayed. Permissions Use this option to view permissions assigned to the current user and the descriptions of the corresponding privileges. Click Permissions. The User Group Membership screen displays. Status 48 Click one of the pie-shaped graphics under status to view the status of the corresponding dissolution system. 400-DS Operator’s Manual Operating the 400-DS 4 System Configuration It is necessary to set up the 400-DS systems that will be used. Systems can be added, modified, and removed from the database. All system configuration activity is recorded in the system audit log. The dissolution software allows the configuration of multiple systems. A maximum of four systems can be running methods at one time on a single PC. System configuration entails selecting the appropriate equipment and setting the communication and other physical properties of the system. Serial numbers are stored for each system to allow tracking of physical system changes. Initial Setup of 400-DS Configuration 1 From the navigation bar, click Configuration. The System Configuration screen displays. Figure 15 System Configuration screen 400-DS Operator’s Manual 49 4 Operating the 400-DS 2 Click Create. The System Editor screen displays (see Figure 16, “System Editor screen,” on page 50). Figure 16 System Editor screen Following is a description of the System Editor screen options: Table 2 50 System Editor Screen 1 Options Screen Section Parameter Description System System Name Enter a name for your system. Laboratory Enter a laboratory name. 400-DS Operator’s Manual Operating the 400-DS Table 2 4 System Editor Screen 1 Options Screen Section Parameter Restrict Execution To Specified Server Description Click to display the name of the workstation or enter the name of the workstation connected to the system. Click No Restriction to allow the system to be run from any workstation. Note: The system must be physically connected. Change Management If applicable, select the box under Change Management in order to restrict the system editing rights to the current user. Note: This option prevents system configuration modification by anyone other than the system owner or a user with VkModifyOthersSystems privilege. 3 Click Next. The next System Editor screen displays. Figure 17 400-DS Workstation screen 400-DS Operator’s Manual 51 4 Operating the 400-DS Following is a description of the System Editor screen options: Table 3 System Editor Screen 2 Options Screen Section Parameter Response Agilent 400-DS Automated Apparatus 7 Serial Number Click ... to display the serial numbers of all connected instruments. Highlight the appropriate serial number and click Select. Cell Size Click the drop-down arrow to indicate the appropriate cell size. Vial Size Click the drop-down arrow to indicate the appropriate vial size. Syringe Size Click the drop-down arrow to indicate the appropriate syringe size. Manufacturer Enter the manufacturer of the calibrated temperature probe. Model Name Enter the model name of the calibrated temperature probe. Serial Number Enter the serial number of the calibrated temperature probe. Set Point Temperature (ºC) Enter the set point for the cell temperature. Stop On Internal Moisture Detection Check this option to shut down the system when moisture is detected. Cell Calibration Temperatures Moisture Sensor Control Enter the temperature for each cell as determined from the diagnostic (see “Temperature Calibration” on page 60). Note: Enabling this option is recommended. Volume Calibration Coefficient Offset The current offset displays in this area. Slope The current slope displays in this area. 4 Click Next. The Serial Numbers screen displays. 5 Enter the serial number for the sample cells, sample holders, etc. and click Add. 6 Click Finish. 7 Repeat all the sections under “System Configuration” on page 49 for each additional system. 8 Close the System Configuration screen. 52 400-DS Operator’s Manual Operating the 400-DS 4 Copying a System Configuration To copy a system configuration, complete the following steps: 1 From the navigation bar, click Configuration. The System Configuration screen displays (see Figure 15, “System Configuration screen,” on page 49). 2 Select the desired system configuration. 3 Click Copy. 4 Click Paste. A new system configuration displays. The description of the new system configuration is Copy of.... 5 Close the System Configuration screen. 6 To edit the system configuration, see “Editing an Existing System Configuration” on page 54. Deleting a System Configuration To delete a system configuration, complete the following steps: 1 From the navigation bar, click Configuration. The System Configuration screen displays (see Figure 15, “System Configuration screen,” on page 49). 2 Select the desired system configuration. 3 Click Delete. 4 Click Yes. 5 Close the System Configuration screen. NOT E 400-DS Operator’s Manual The system configuration is never permanently deleted. It is marked as deleted in the database. 53 4 Operating the 400-DS System Configuration Report To display a report of the system configuration, complete the following steps: 1 From the navigation bar, click Configuration. The System Configuration screen displays (see Figure 15, “System Configuration screen,” on page 49). 2 Select the desired system configuration. 3 Click Report. By default, the most recent report version displays. NOT E Alternately, you can right-click the desired system configuration and select Report. The Version Selection screen displays as a result of either of these actions. 4 Use the up and down arrows to indicate the desired report version and click OK. 5 The system report displays. The report can be printed, exported, searched, or verified. NOT E PCL6 or PPD printer drivers are required for printing a report. Editing an Existing System Configuration To display and edit the properties of an existing system configuration, complete the following steps: 1 From the navigation bar, click Configuration. The System Configuration screen displays (see Figure 15, “System Configuration screen,” on page 49). 2 Select the desired system configuration. 54 400-DS Operator’s Manual Operating the 400-DS 4 3 Click Properties. The System Editor screen displays (see Figure 16, “System Editor screen,” on page 50). NOT E Alternately, you can double-click the desired system configuration or right-click the desired system configuration and select Properties. The System Editor screen displays as a result of any of these actions. 4 Select the appropriate tabs and change the relevant information in the same manner that the system was created. 5 Close the System Configuration screen. Recovering a Deleted Configuration To recover a deleted configuration, complete the following steps: 1 From the navigation bar, click Configuration. The System Configuration screen displays. 2 Click Recover. A list of the deleted systems is displayed. 3 Select the system to recover and click OK. 4 Verify the system is now in the current system list. Show Audit Trail To display the audit trail for a system configuration, complete the following steps: 1 From the navigation bar, click Configuration. The System Configuration screen displays (see Figure 15, “System Configuration screen,” on page 49). 2 Select the system configuration and click Show Audit Trail. The System Audit Trail screen displays. NOT E 400-DS Operator’s Manual Alternately, you can right-click the desired system configuration and select Show Audit Trail. The System Audit Trail screen displays as a result of either of these actions. 55 4 Operating the 400-DS 3 While holding the Ctrl key, select two or more versions and click Differences in the navigation bar. A change report displays. The report can be printed or exported. NOT E PCL6 or PPD printer drivers are required for printing a report. Verify Integrity To verify that the system configuration has not been changed outside of the 400-DS Workstation program, complete the following steps: 1 From the navigation bar, click Configuration. The System Configuration screen displays (see Figure 15, “System Configuration screen,” on page 49). 2 Select the system configuration and click Verify Integrity. Either the data is verified successfully or the user is directed to contact their system administrator. NOT E Alternately, you can right-click the desired system configuration and select Verify Integrity. The integrity of the system configuration is checked as a result of either of these actions. 3 Click OK to close the Data Verification screen. 4 Close the System Configuration screen. Import/Export XML File To use an existing system configuration from one 400-DS Workstation computer on a different 400-DS Workstation computer, you can export and import the system configuration as an XML file. 56 400-DS Operator’s Manual Operating the 400-DS 4 To export the system configuration, complete the following steps: 1 From the navigation bar, click Configuration. The System Configuration screen displays (see Figure 15, “System Configuration screen,” on page 49). 2 Select the desired system configuration. 3 Click Export. The Version Selection screen displays. NOT E Alternately, you can right-click the desired system configuration and select Export. The Version Selection screen displays as a result of either of these actions. 4 If applicable, indicate which version to export and click OK. The Export System to XML File screen displays. NOT E Each time a system configuration is saved, a new version is created. To export a version other than the most recently saved, indicate the appropriate version number on the Version Selection screen. 5 Indicate the directory and file name and click Save. The code is saved as an XML file. To import the XML file at another 400-DS Workstation, complete the following steps: 1 From the navigation bar, click Configuration. The System Configuration screen displays (see Figure 15, “System Configuration screen,” on page 49). 2 Click Import. The Import System from XML File screen displays. NOT E Alternately, you can right-click the desired system configuration and select Import. The Import System from XML File screen displays as a result of either of these actions. 3 Select the appropriate directory and file name and click Open. The system configuration displays on the System Configuration screen. 4 Close the System Configuration screen. 400-DS Operator’s Manual 57 4 Operating the 400-DS 400-DS System Calibration Volume Calibration 1 Place the tubing from the Media A valve into a suitable container with deionized water. 2 From the navigation bar, click Configuration. 3 Select desired configuration. 4 Click Calibration. The System Calibration screen displays. 5 Enter the type of media used during the calibration procedure. 6 Enter the density of the media used at ambient temperature. Figure 18 System Calibration screen 7 Label 13 empty 4 mL sample vials (Row 1 Vial 1 - Row 1 Vial 13). 58 400-DS Operator’s Manual Operating the 400-DS 4 8 Label 13 empty 1.5 mL sample vials (Row 2 Vial 1 - Row 2 Vial 13). 9 Weigh each of the vials and record the empty masses in the assigned cells in the table. 10 Place the sample vials in row 1 (4 mL) and row 2 (1.5 mL) of the sample tray. 11 Click Fill Calibration Volumes to fill the vials in the first row with 4 mL and the second row with 1 mL of deionized water. 12 After the fill process is complete, weigh each of the vials and record their filled masses in the table. The vial volume calculates and displays in the table automatically. 13 Click OK. The Reason for Change screen displays. NOT E The software performs a Least Squares Fit analysis. The volume deviation from the curve is 1.5% of the vial volume. An automated error message is generated if the calibration fails. 14 Enter a reason code and change description and click OK. 15 Highlight the appropriate system. 16 Click Properties. 17 Click Next and ensure the Volume Calibration Coefficients are visible. 18 Click Finish. 19 Close the System Configuration screen. 400-DS Operator’s Manual 59 4 Operating the 400-DS Temperature Calibration NOT E If a previous temperature calibration has been performed, it is necessary to reset the cell temperature values before executing this procedure: Step 1. Click Configuration. Step 2. Select the appropriate system from System Configuration. Step 3. Click Next (from System Editor). All channels should be set to the desired temperature (ºC). 1 Place the rinse media line into a suitable container holding deionized water. 2 Ensure sample holders are installed in each sample cell. 3 From the navigation bar, click Diagnostics. The Select Item screen displays. 60 400-DS Operator’s Manual Operating the 400-DS 4 4 Select the desired system and click OK. The System Diagnostics screen displays. 5 Ensure the media source option selected is Rinse Media. In the Cells box under Media Control, click Fill. Media moves into the cells. 6 In the Cell Heat box, set the cell temperature to the desired temperature. 7 Click (start heating). DO THE SAMPLE CELLS CONTAIN LIQUID? displays. 8 Click Yes. The temperatures display in the Cell Temperature box. 9 Set the reciprocation rate to at least 20 DPM and click . 10 Allow the media to equilibrate to the set temperature. 400-DS Operator’s Manual 61 4 Operating the 400-DS 11 Fill a beaker with a sufficient amount of deionized water and verify the temperature using a calibrated thermometer. Record the value. NOT E If using the standard QA Station probe, it will be necessary to first verify the temperature with the QA Station probe and then replace it with the Agilent 400-DS modified probe. 12 Using the modified temperature probe, confirm the temperature of the beaker of deionized water. Record the value. Temperatures should match ± 0.2 ºC. 13 In the Evaporation Cover box, click the up arrow to raise the evaporation cover. 14 Remove the two thumb screws from the evaporation cover and remove the evaporation cover. 15 Remove the sample holder for Cell 1 and replace it with the modified temperature probe. 16 Close the top cover of the Agilent 400-DS. Allow the temperature to stabilize and record the temperature. 17 Repeat steps 14 - 16 for each additional cell. 18 Reinstall the sample holders and the evaporation cover. 19 Close the Diagnostics screen. 20 From the navigation bar, click Configuration. The System Configuration screen displays. 21 Select the appropriate system and double-click to open. The System Editor screen displays. 22 Click Next. Enter the Setpoint temperature. 23 Enter the temperature values for each cell (as recorded in the table beneath Step 40) into the cell calibration field. 24 Click Finish. The Reason for Change screen displays. 25 Enter a reason code and change description and click OK. 26 Close the System Configuration screen. NOT E 62 To update the 400-DS Workstation Software with the new temperature values, initiate a new method. See “Creating a Method” on page 69 and “Running a Method” on page 82. 400-DS Operator’s Manual Operating the 400-DS 4 27 From the navigation bar, click Diagnostics. The Select Item screen displays. 28 Highlight the system being qualified and click OK. 29 In the Cell Heat box, set the cell temperature. 30 Click (start heating). DO THE SAMPLE CELLS CONTAIN LIQUID? displays. 31 Click Yes. The temperatures display in the Cell Temperature box. 32 If necessary, set the reciprocation rate and click . 33 Allow the media to equilibrate to the set temperature. 34 Record the temperatures displayed for each position. 35 Remove the two thumb screws from the evaporation cover and remove the cover. 36 Remove the sample holder for Cell 1 and replace it with the modified temperature probe. 37 Close the top cover of the Agilent 400-DS. Allow the temperature to stabilize. 38 After stabilization, record the temperature displayed on the QA Station. 39 Repeat steps 36 - 38 for each additional sample cell position. NOT E The temperatures displayed in the 400-DS Workstation Software should match the modified QA Station temperature probe ± 0.2 ºC to ensure accurate temperatures. 40 Replace the sample holders and the evaporation cover. 41 Click Purge to empty the sample cells. 42 Close the Diagnostics screen. 400-DS Operator’s Manual 63 4 Operating the 400-DS Manual Control/Diagnostics To check the diagnostics of a dissolution apparatus, complete the procedures on the following pages. These procedures are performed on one system at a time. Repeat the procedures as applicable for each additional system. 1 From the navigation bar, click Diagnostics. The Select Item screen displays. 2 Select the desired system and click OK. The System Diagnostics screen displays. Verifying Sample Tray Control 1 Click any vessel position corresponding to the desired row in the Sample Tray Control box. The tray moves and the indicated row raises. Figure 19 Sample Tray Control box 2 Click under the Sample Tray Control box. The drive unit returns to the home position. 3 Click 64 to stop the movement. 400-DS Operator’s Manual Operating the 400-DS 4 4 Click to eject or retract the sample tray. Select Ship to place the tray in ship mode. 5 Click to raise the sample tray row. 6 Click to lower the sample tray row. 7 Click to turn off any audible alarm. Verifying Media Control 1 Under the Media Control box, set the following parameters: Table 4 Media Control Options Option Description Aspirate Flow Rate (mL/s) Use the up and down arrows to indicate the rate at which the media is pulled into the syringe. Dispense Flow Rate (mL/s) Use the up and down arrows to indicate the rate at which the media is dispensed from the syringe. Aspirate Dwell Time (s) Use the up and down arrows to indicate the duration of time the plunger holds at the top of the stroke before dispensing. Note: The time specified is for full syringe volume. For smaller syringe volumes, it is calculated proportionately. Needle Clean Cycles Use the up and down arrows to indicate the number of times the unit performs the needle cleaning cycle. Prime Volume (mL) Use the up and down arrows to indicate the volume to be pulled before taking measurement. Media Source Use the drop-down arrow to indicate the source of the media. Media Volume Use the up and down arrows to indicate the desired volume of the media. 2 Under the Media Control box, click Prime. 3 Under the Cells box, click Fill. The media moves into the cells. CAUTION 400-DS Operator’s Manual Do not click Fill more than one time without purging in between. The media will overflow the cells. 65 4 Operating the 400-DS 4 Under the Sample Vials box, use the up and down arrows to indicate the desired row number to sample into and the desired volume. 5 Ensure the selected row contains vials in all positions. Under the Sample Vials box, click Fill Row. The media moves from the cell into the sample vial. 6 Under the Waste Pump box, click On to turn on the waste pump. 7 Under the Waste Pump box, click Off to turn off the waste pump. Verifying the Cell Temperature 1 In the Cell Heat box, use the up and down arrows to set the media cell temperature to an appropriate temperature (see the sample screen below). 2 Click CAUTION . DO THE SAMPLE CELLS CONTAIN LIQUID? displays. Do not begin heating the sample cells if they are empty as this can damage the heating jacket. Ensure the each cell is filled with media before activating the heater(s). 3 Click Yes. The temperatures display in the Cell Temperature box. 4 As applicable, click . 5 Under the Cells box, click Purge to purge the remaining media from the tube. 66 400-DS Operator’s Manual Operating the 400-DS 4 Verifying Reciprocation To set the dips per minute, complete the following steps: 1 In the Reciprocation box, use the up and down arrows in the box that corresponds to DPM, to set the dips per minute to a desired value and click . Dipping begins. Figure 20 Reciprocation box 2 Click . The dipping stops. Evaporation Cover Control To raise the evaporation cover, click To lower the evaporation cover, click box. 400-DS Operator’s Manual in the Evaporation Cover box. in the Evaporation Cover 67 4 Operating the 400-DS Method Editor 1 From the navigation bar, click Editor. The Methods screen displays. All test parameters are entered via the Method Editor screen. 68 Function Procedure Create a new method See “Creating a Method” on page 69. Copy a method See “Copying Methods” on page 76. Delete a method See “Deleting Methods” on page 76. Edit an existing method See “Editing an Existing Method” on page 76. Run a report of the method setup See “Method Report” on page 77. View the method audit trail See “Show Audit Trail” on page 78. Verify the integrity of a method See “Verify Integrity” on page 79. Import or export the method between two 400-DS Workstations See “Import/Export XML File” on page 79. 400-DS Operator’s Manual Operating the 400-DS 4 Creating a Method To create a new method, complete the following steps: 1 From the navigation bar, click Editor. The Methods screen displays. Figure 21 Methods screen 2 Click New Method. The Method Editor screen displays. 400-DS Operator’s Manual 69 4 Operating the 400-DS Figure 22 Method Editor screen 70 400-DS Operator’s Manual Operating the 400-DS 4 3 Select the Parameters tab. Following is a description of the Parameters tab options: Table 5 Parameters Tab Options Option Sample Information Description Product Name Enter the name of the product. Notes Enter any relevant notes regarding the sample. User Defined Labels The fields in this area have default values of LOT, BATCH, and GROUP. These fields are customizable; enter the information that best serves the needs of the method parameters. Label 1 Label 2 Label 3 Error Tolerance (±) Temperature Enter the desired tolerance. If the tube temperature goes over or under the set temperature (see “Verifying the Cell Temperature” on page 66) by the amount of this tolerance, an error is recorded as part of the results. Speed Enter the desired tolerance. If the RPM goes over or under the set speed (see “Verifying Reciprocation” on page 67) by the amount of this tolerance, an error is recorded as part of the results. Profile Interval Enter the timepoint in hh:mm format at which the temperature and speed settings are recorded. Note: Profile measurements are optional. Values are always recorded at sample timepoints independent of this setting. Change Management If applicable, select the box under Change Management in order to restrict the method editing rights to the current user or any user with VkModifyOthersMethod. 4 Select the Automated App7 Sequence tab (see Figure 23, “Automated App7 Sequence tab,” on page 72). 400-DS Operator’s Manual 71 4 Operating the 400-DS Figure 23 Automated App7 Sequence tab 72 400-DS Operator’s Manual Operating the 400-DS 4 Following is a description of the Automated App7 Sequence tab options: Table 6 Automated App7 Sequence Options Option Description Sample Row Use the up and down arrows to indicate the sample tray row number for which the parameters are being set. After clicking Add to List, use the up and down arrows to indicate the next row. Note: The sample row will increment automatically from the previous timepoint. Media Use the drop-down arrows to indicate Media A, Media B, Media C, Media D, Media E, or Rinse media. Sampled Volume (mL) Use the up and down arrows to indicate appropriate volume based on your vial size. Dissolution Volume (mL) Use the up and down arrows to indicate appropriate volume based on the system cell size. Cell Temperature Enter the desired tube temperature in degrees Celsius for the first timepoint. Hold Time (mm:ss) Enter the desired duration for the sample holder to remain stationary prior to dipping in mm:ss format. Dip Speed (DPM) Use the up and down arrows to indicate the desired dips per minute (DPM) for each applicable row of the 400-DS. Dip Interval (hhh:mm:ss) Enter the desired duration for dipping in hhh:mm:ss format. 5 Select the Automated App7 Options tab. 400-DS Operator’s Manual 73 4 Operating the 400-DS Figure 24 Automated App7 Options tab 74 400-DS Operator’s Manual Operating the 400-DS 4 Following is a description of the Automated App7 Options tab options: Table 7 Automated App7 Options Screen Section Parameter Description Media Description Media A Enter descriptions as appropriate for the different medias. Media B Media C Media D Media E Rinse Media Syringe Properties Aspirate Flow Rate (mL/s) Use the up and down arrows to indicate the rate at which the media is pulled into the syringe. Dispense Flow Rate (mL/s) Use the up and down arrows to indicate the rate at which the media is dispensed from the syringe. Aspirate Dwell Time (s) Use the up and down arrows to indicate the duration of time the plunger holds at the top of the stroke before dispensing. Note: The time specified is for full syringe volume. For smaller syringe volumes, it is calculated proportionately. Prime Volume (mL) Use the up and down arrows to indicate the volume to be pulled before taking a sample. Syringe Clean Cycles Use the up and down arrows to indicate the number of times to rinse the syringe between samples. Sample Cell Clean Cycles Use the up and down arrows to indicate the number of times to rinse the sample cell between samples. Needles Clean Cycles Use the up and down arrows to indicate the number of times to rinse the needles between samples 6 Select the Notifications tab. 7 Enter appropriate address(es) to receive notification of system operation (if applicable). 8 Indicate which options would require notification. 9 Click OK to close the Method Editor screen. 10 Close the Methods screen. 400-DS Operator’s Manual 75 4 Operating the 400-DS Copying Methods To copy a method, complete the following steps: 1 From the navigation bar, click Editor. The Methods screen displays (see Figure 21, “Methods screen,” on page 69). 2 Select the desired method. 3 Click Copy Method. 4 Click Paste. A new method displays. The description of the new method is Copy of.... 5 To change any of the parameters of the method, see “Editing an Existing Method” on page 76. Deleting Methods To delete a method, complete the following steps: 1 From the navigation bar, click Editor. The Methods screen displays (see Figure 21, “Methods screen,” on page 69). 2 Select the desired method. 3 Click Delete. 4 Click Yes. NOT E A deleted method can be recovered at any time. See “Recovering a Method” on page 78. Editing an Existing Method To edit a method already entered on the 400-DS Workstation, complete the following steps: 1 From the navigation bar, click Editor. The Methods screen displays (see Figure 21, “Methods screen,” on page 69). 2 Click the desired method. 76 400-DS Operator’s Manual Operating the 400-DS 4 3 Click Properties on the navigation bar. The Method Editor screen displays (see Figure 22, “Method Editor screen,” on page 70). NOT E Alternately, you can double-click the desired method or right-click the desired method and select Properties. The Method Editor screen displays as a result of any of these actions. 4 Select the appropriate tabs and change the relevant information in the same manner that the method was created. Method Report To display a report of the method parameters, complete the following steps: 1 From the navigation bar, click Editor. The Methods screen displays (see Figure 21, “Methods screen,” on page 69). 2 Select the desired system configuration. 3 Click Report. The Version Selection screen displays. 4 If applicable, indicate which version and click OK. NOT E Each time a method is changed, a new version is created. To create a report of a version other than the most recently saved, indicate the appropriate version number on the Version Selection screen. 5 The method report displays. The report can be printed, exported, searched, verified, and/or signed. 400-DS Operator’s Manual 77 4 Operating the 400-DS Recovering a Method To recover a method, complete the following steps: 1 From the navigation bar, click Editor. The Methods screen displays (see Figure 21, “Methods screen,” on page 69). 2 Click Recover. The Select Deleted Method(s) to Recover, screen displays. 3 Select the method to recover and click OK. Method Audit Trail Once a method has completed, the results are available for review, audited modification, and electronic signature. The software maintains complete history for all runs executed on the system. Results can be previewed and printed. Show Audit Trail To display the audit trail for a method, complete the following steps: 1 From the navigation bar, click Editor. The Methods screen displays (see Figure 21, “Methods screen,” on page 69). 2 Select the method and click Show Audit Trail. The Method Audit Trail screen displays. NOT E Alternately, you can right-click the desired method and select Show Audit Trail. The Method Audit Trail screen displays as a result of either of these actions. 3 While holding the Ctrl key, select two or more versions and click Differences in the navigation bar. A change report displays. The report can be printed or exported. 78 400-DS Operator’s Manual Operating the 400-DS 4 Verify Integrity To verify that the method has not been changed outside of the application, complete the following steps: 1 From the navigation bar, click Editor. The Methods screen displays (see Figure 21, “Methods screen,” on page 69). 2 Select the method and click Verify Integrity. Either the data is verified successfully or the user is directed to contact their system administrator. NOT E Alternately, you can right-click the desired method and select Verify Integrity. The integrity of the method is checked as a result of either of these actions. Import/Export XML File To use an existing method from one 400-DS Workstation on a different 400-DS Workstation, you can export and import the method as an XML file. To export the method, complete the following steps: 1 From the navigation bar, click Editor. The Methods screen displays (see Figure 21, “Methods screen,” on page 69). 2 Select the desired method. 3 Click Export. The Version Selection screen displays. NOT E Alternately, you can right-click the desired method and select Export. The Version Selection screen displays as a result of either of these actions. 4 If applicable, indicate which version to export and click OK. The Export Method to XML File screen displays. NOT E 400-DS Operator’s Manual Each time a method is saved, a new version is created. To export a version other than the most recently saved, indicate the appropriate version number on the Version Selection screen. 79 4 Operating the 400-DS 5 Indicate the directory and file name and click Save. The code is saved as an XML file which displays in Windows Notepad. 6 Close the file. To import the XML file at another 400-DS Workstation, complete the following steps: 1 From the navigation bar, click Editor. The Methods screen displays (see Figure 21, “Methods screen,” on page 69). 2 Click Import. The Import Method from XML File screen displays. NOT E Alternately, you can right-click the desired method and select Import. The Import Method from XML File screen displays as a result of either of these actions. 3 Select the appropriate directory and file name and click Open. The method displays on the Method screen. 80 400-DS Operator’s Manual Operating the 400-DS 4 Active Channel Selection Valves The Agilent 400-DS is equipped with channel selection valves which allow the you to isolate and shut off any unused sample cells. For example, if an experiment only required the use of eight sample vessels, Vessels 9 - 13 could be shut off by closing the red channel selection valves. 1 Identify the sample cells you want to disable and their corresponding valves. 2 Turn the red knob(s) clockwise one quarter turn. 3 To re-enable a valve, turn the applicable knob one quarter turn counterclockwise. 4 The number of active channels is programmed as part of the Method Start options just before the test begins. CAUTION 400-DS sample cells are numbered left to right. To prevent damage to the heater jackets, close the valves from right to left when less than 13 active channels are used. CAUTION Ensure sample vials are present in positions 1 and 13 of each sample tray row if less than 13 active channels are used. This provides proper sensing, alignment, and balance for each sample tray. 400-DS Operator’s Manual 81 4 Operating the 400-DS Running a Method 1 From the navigation bar, click Run Method. The Select System Step 1 of 2 screen displays. 2 Select the desired system to run the method and click Next. The Select Method Step 2 of 2 screen displays. 82 400-DS Operator’s Manual Operating the 400-DS 4 3 Select the desired method to run and click Finish. The system status screen displays. Figure 25 System status screen 4 If necessary, remove the evaporation cover. 5 Place the samples on or in the sample holder. 6 Insert the sample holders into the sample cells. 7 Reinstall the evaporation cover and tighten the two thumb screws. 8 Place the capped sample vials in the applicable sample rows. CAUTION 400-DS Operator’s Manual Ensure sample vials are present in positions 1 and 13 of each sample tray row if less than 13 active channels are used. This provides proper sensing, alignment, and balance for each sample tray. 83 4 Operating the 400-DS 9 Click Sampler > Load to load the tray. 10 Fill applicable media containers. 11 Click Start. The Method Start Options screen displays (see Figure 26, “Method Start Options screen,” on page 84). Figure 26 Method Start Options screen Following is a description of the Method Start Options screen options: Table 8 Method Start Options Screen Section Parameter Description Sample Information Product Name Enter the product name. Notes Enter any appropriate notation. Label 1 Label 2 Enter the appropriate information based on the user-defined labels (see “User Defined Labels” on page 71). Label 3 Active Channels Enter the number of active heating channels. Note: Heating channels are counted from left to right. See “Active Channel Selection Valves” on page 81 Injection Control 84 Enable Injections N/A 400-DS Operator’s Manual Operating the 400-DS Table 8 4 Method Start Options Screen Section Parameter Description Temperature Delayed Start Vessel Temperature Start N/A Bath Temperature Start N/A Time Delayed Start Use this option to program a delayed start. Enter the desired date and time to start the method. 12 Click OK. The Method Sample Information screen displays. Figure 27 Method Sample Information Screen 400-DS Operator’s Manual 85 4 Operating the 400-DS 13 Enter the sample information, if applicable. NOT E If the method requires more than 12 sample rows, complete the following steps: 1. Upon completion of filling the 12th sample row, click Sampler > Eject Tray. 2. Replace the sample rows with the required amount of fresh sample vials. 3. Load the new sample tray. 4. Click Sampler > Load New Tray. NOT E Vials may be removed or replaced at any time after a sample row has been completed, provided a sample is not occurring. 14 When the method is complete, click Test Report. The test report displays. The report can be printed, exported, searched, verified, and/or signed. 86 400-DS Operator’s Manual Operating the 400-DS 4 Test Reports To display a report of the completed method, complete the following steps: 1 From the navigation bar, click Test Reports. The Test Report Selection screen displays. Figure 28 Test Report Selection screen NOT E By default, the test report dialog only shows tests done in the previous week. To view tests completed earlier than the previous week, click the Start Date and/or End Date drop-down arrows to indicate date range for a test report selection and click Filter Date and System. Additionally, you can select a system name from the System combo box to restrict the selection to a particular system. To restrict the selection to a particular test identification, enter the value in the Test ID Filter box and click Filter Test ID. 2 Click a test in the list to select it. 400-DS Operator’s Manual 87 4 Operating the 400-DS 3 From the navigation bar in the Test Report Selection dialog, click Show Report. The report for the selected test displays. 4 Using the buttons on the report toolbar, you can do the following: first page previous page print the report skip to a specific page by number next page last page Electronic Signatures When the user is satisfied with the results, the results can be electronically signed. The software allows multiple electronic signings of a set of results. Each signing is accomplished using the signature dialog box shown. The user authenticity is determined by testing the user identification and password against the Windows security database. Electronic signatures are permanently linked to the results. The software always requires the signature to be executed using all the signature components. Any attempts to sign a set of results using an invalid user identification, password, or any combination thereof that is incorrect is automatically recorded to the system audit trail. Clean System From the system status screen, click . All instruments initialize. Media is pulled from the Rinse Media Container and cells are filled and expelled through the valves. The cells are aspirated and purged. NOT E 88 The 400-DS should be flushed / cleaned as soon as possible after a test. Multiple clean cycles may be necessary to properly rinse the system. 400-DS Operator’s Manual 400-DS Apparatus 7 Operator’s Manual 5 Troubleshooting and Maintenance Preventive Maintenance Troubleshooting 92 90 Agilent Technologies 89 5 Troubleshooting and Maintenance Preventive Maintenance WA RNING The 400-DS contains electrical circuits, devices, and components operating at dangerous voltages. Contact with these circuits, devices, and components can cause death, serious injury, or painful electric shock. Preventive maintenance needs may vary depending on frequency of instrument usage. Calibration Agilent recommends the 400-DS be calibrated at six-month intervals. However, based on your own SOPs, this can be extended for up to a year. Contact the Agilent Service Department to schedule preventive maintenance and qualification. The Agilent Service Department can be reached at 1.800.227.9770. Routine Maintenance (to be Performed Between Each Dissolution Method • Execute the Clean System cycle using the 400-DS Workstation software (“Clean System” on page 88). Ensure an appropriate cleaning solution is in place to thoroughly rinse the media lines and sample cells. 90 400-DS Operator’s Manual Troubleshooting and Maintenance 5 Sample Holders • Clean sample holders thoroughly after each use and place them in an ultrasonic bath for a few minutes, if necessary. • When using corrosive materials such as hydrochloric acid, surfactants, or medium containing salts, be sure to rinse the holders with deionized water immediately after use. Dry completely with a soft towel or cloth (preferably lint-free). • Do not clean the sample holders with abrasive cleansers or cloths. Use deionized water whenever possible. If you must use cleanser or solvent, be sure that it is as mild as possible, non-abrasive, and fully compatible with fluorocarbons and stainless steel before use. • It is suggested that the sample holders are cleaned by hand. Use of a laboratory dishwasher is NOT recommended. • Handle with care at all times. • Store the sample holders properly between each use. 400-DS Operator’s Manual 91 5 Troubleshooting and Maintenance Troubleshooting Obtaining Warranty and Other Services To place a service order (warranty or other services), please contact your local Customer Care Center. Contact information can be found at www.agilent.com under your country using the Contact Us link. Place your service request using the displayed phone number or E-mail address. 92 400-DS Operator’s Manual Index Numerics Creating a Method, 69 M 400-DS, accessories, 22 400-DS, parts, 22 D Database, Attaching, 31 Deleting a System Configuration, 53 Deleting Methods, 76 diagnostics, 64 Dissolution Server Selection, 37 DissoService Manager screen tools menu, 34 DissoService Options screen, 35 Maintenance Preventive, 90 Manual Control, 64 Method Editor, 68 Method Editor screen, 70 Method Report, 77 Method Start Options screen, 84 Methods screen, 69 Minimum PC Configuration, 28 MSDE Manager screen, 31 E P Editing an Existing Method, 76 Editing an Existing System Configuration, 54 Electronic Signatures, 88 E-mail Notification Configuration, 34 Error Tolerance, 71 Export XML File, 56, 79 Parts, 22 PC requirements, minimum, 28 Permissions, 48 Power Up, 29 Preventive, 90 Profile Interval, 71 I report toolbar, 88 Run Method, 82 A Accessories, 22 Add New Dissolution Server screen, 38 Application, configuring, 40 Audit Trail, 48, 78 Automated Apparatus 7 Workstation screen, 47 B Browse Dissolution Servers screen, 39 C Cable Connections Multiple System, 25 Single System, 25 Calibration, 90 Change Management, 71 Change User, 48 Clean System, 88 clearance, 21 Configuration Dialog screen Miscellaneous tab, 42 Security tab, 40 Configuring Automated Apparatus 7 Workstation, 49 Configuring the Application, 40 Conventions, 17 Copying a System Configuration, 53 Copying Methods, 76 400-DS Operator’s Manual Import XML File, 56, 79 Initial Power Up, 29 L Local Security Policy, 32 Local Users and Groups screen, 41 Lock Application, 48 Logon screen, 38, 46 R S Sample Information, 71 screen options Automated Apparatus 7 Workstation, 48 Method Start Options screen, 84 System Editor, 50 Security, 48 Selecting Dissolution Server, 37 Serial Numbers, 54 Show Audit Trail, 55, 78 93 Index Starting Dissolution Workstation, 36 System Configuration Report, 54 System Configuration screen, 49 system editing rights, 51 System Editor screen, 50 System status screen, 83 T Test Report Selection screen, 87 Test Reports, 87 U unpacking your equipment, 19 Unpacking Your System, 20 V Verify Integrity, 56, 79 W Windows Security Alert screen, 36 94 400-DS Operator’s Manual
advertisement
Key Features
- First compendial small-volume dissolution apparatus
- Ideal for dissolution testing
- Integrated, small-footprint
- Control using software via PC
- Samples can be collected in vials
- Innovative needle manifold