Moore & Wright MW Value Line Digitronic Micrometer 270 Series is a versatile tool designed for precise measurements. It features an LCD display for clear readings, friction and quick drives for easy operation, and ABS/INC and UNIT keys for absolute and relative measurements, as well as metric/inch conversion. With its pointed spindle and anvil, it can measure grooves and keyways. The 30-degree measuring point allows for accurate measurements in various positions.
Advertisement
Advertisement
DIGITRONIC MICROMETER INSTRUCTIONS
Value Line Digitronic Micrometer 270 Series
Please carefully review these instructions to ensure proper and accurate use of your new
Digitronic Micrometer.
7 & 8 6
3
1. Features
1. LCD display
2. Friction drive
3. Quick drive
4. ABS/INC…UNIT key
5. ON/OFF…SET key
6. Data output
7. Pointed spindle and anvil for measuring grooves and keyways
8. 30 degree measuring point
5
1
4
2
2. Keys
• ON/OFF…SET key: Power switch. Datum set
• ABS/INC…UNIT key: Absolute and relative measuring. Metric/inch conversion
3. LCD Display
in: Inch measuring mode
mm: Metric measuring mode
INC: Relative measuring mode
4. Operation
• Keys are pressed two ways to execute functions:
1. Press and immediately release
2. Press and hold for at least two seconds
4.1 On/Off - SET Key
• Press and release: Power on/off
• Press and hold (2 sec. or more): Datum setting for absolute measurement
• Origin of metric is 0, 25, 50, 75mm…275mm. Origin of inch is 0, 1", 2", 3"…11"
• Sets datum automatically after battery reset
For more information visit www.moore-and-wright.com Partners in Precision
115568_A5_Instructions_MW270.indd 1 05/12/2012 07:03
DIGITRONIC MICROMETER INSTRUCTIONS
4.2 ABS/INC - UNIT Key
• Press and release: Absolute and relative measuring mode conversion; ‘INC’ sign displayed on LCD in relative measuring mode. No sign displayed on LCD in absolute measuring mode
• Press and hold (2 sec. or more): metric/inch conversion; ‘in’ sign displayed on LCD for inch, otherwise mm
5. Power
• Use a CR2032 Lithium battery and replace the battery when the display begins to dim or the data on the LCD flashes
• If the micrometer is not used for five minutes the power will automatically shut off.
Power off the micrometer by pressing the ‘On/Off – SET‘ key to save the battery when not in use
6. General Specifications
• Measuring force: 5-10N
• Operating temperature: 0-40°C
• Power consumption: ≤20µA
• Storage temperature: -20-60°C
7. Precautions
• Do not subject the instrument to blows or shock. Do not drop it or apply excessive force
• Do not disassemble the instrument
• Do not press the keys with a pointed object
• Do not use or store the instrument under direct sunlight or in an excessively hot or cold area
• Avoid exposing the instrument to strong magnetic fields and high voltage
• Use a soft cloth to clean the instrument. Never use organic solvents such as acetone or benzene to clean. Alcohol may be used
• Clean measuring faces before use
• If the instrument is to be stored or left unused for extended periods, remove the battery
8. Troubleshooting
Problem(s)
Measuring data incorrect
No display on LCD
Display confused or remains blank
Cause(s)
Dirty measuring faces Preset data is not correct
Battery voltage under 2.7v
Battery is not properly set
Battery voltage under 2.7v
Solution(s)
Clean measuring faces Inspect preset data and reset it
Reset battery
(remove for 3 minutes)
Replace battery
For more information visit www.moore-and-wright.com Partners in Precision