T Detection

T Detection
Detection
Applications
• Endpoint, kinetic, flash, area scanning assays
• DNA quantification, protein quantification
• Ion channel assays
• Enzyme activity
• Gene expression (reporter gene assays)
• Cell-based assays
• FRET assays
Optional Accessories
KC4TM or KCjuniorTM Data Analysis Software
Fluorescence Test Plate
Product Qualification Package
T
he FLx800TM provides researchers
outstanding value with a combination of
excellent performance and ease of use — at a
price much lower than traditional fluorescenceluminescence microplate readers. The design
features a low-noise detection system for
increased performance in both fluorescence
and luminescence modes. The FLx800 includes
several models with options that meet the
specific needs of research and clinical users,
including the ability to read multiple plate types
and PCR plates. The FLx800 includes onboard
data reduction software or can be interfaced
with BioTek’s easy-to-use PC software packages.
Features
• New low-noise detector operated in photon
integration mode enables high luminescence
sensitivity
• Selectable, automated top- or bottom-probe
reading for best sensitivity in fluorescence mode
• Flash luminescence, and fast kinetic
fluorescence reading with optional reagent
injector
• Incubation to 50°C for temperature-sensitive
assays
• PCR plate-reading option available
Models
FLx800: Multiple models including incubation, top/bottom reading, extended range PMT, PCR plate capability
and injector
See price list or Web site for complete model listings and
descriptions.
Specifications*
FLx800
Dimensions
15"W x 16"D x 9"H
(38.1 x 40.64 x 22.89 cm)
30 lbs (13.6 kg)
100 Watts max.
6-384
4ºC above ambient to 50ºC
Yes
Yes
Onboard
(optional PC software)
Weight
Power Consumption
Microplate Types
Temperature Control
Shaking
Top Optics Adjustment
Software
Fluorescence Intensity
Light Source
Wavelength Range
Wavelength Selection
Dynamic Range
Sensitivity Top
Tungsten Halogen
300 – 700 nm (850 nm option)
Deep blocking filters
5 decades
5 pM fluorescein typical
Luminescence
Wavelength Range
Dynamic Range
Sensitivity
300 – 700 nm
5 decades
100 amol ATP typical (flash)
Dispensers
Number of Injectors
Dispense Volume
Dead Volume
1
5 – 1000 μl in 1 μl increment
1 ml, 100 μl with backflush
* Specifications subject to change
BioTek Instruments, Inc.
Highland Park, P.O. Box 998
Winooski, Vermont 05404-0998, USA
Tel: 802-655-4040 • Toll-Free: 888-451-5171 • Outside the USA: 802-655-4740 • Fax: 802-655-7941
Rev - 3/14/2006