Life Science (UV/UF) ultrapure water system
Operation Manual
Series 1193
LT1193X1 • 6/22/00
Model No.
D11931
Voltage
100V - 240V
Table of Contents
Safety Information ..........................................................................................................................................................................................3
Alert Signals............................................................................................................................................................................................3
Warnings ................................................................................................................................................................................................3
Introduction ....................................................................................................................................................................................................5
General Usage........................................................................................................................................................................................5
General Specifications ....................................................................................................................................................................................6
Environmental Conditions ......................................................................................................................................................................7
Declaration of Conformity ......................................................................................................................................................................7
Unpacking ......................................................................................................................................................................................................8
Installation ......................................................................................................................................................................................................9
Choosing a Site ......................................................................................................................................................................................9
Ultrafilter Installation ............................................................................................................................................................................10
UV Lamp Installation ............................................................................................................................................................................11
Other Accessories ................................................................................................................................................................................11
Bench Mounting ....................................................................................................................................................................................12
Wall Mounting ......................................................................................................................................................................................12
Installing the Control Panel in a Remote Location ..............................................................................................................................13
Water Connections ..............................................................................................................................................................................14
Controls ........................................................................................................................................................................................................15
Main Power Switch ..............................................................................................................................................................................15
Control Panel ........................................................................................................................................................................................15
Switches................................................................................................................................................................................................16
Initial Operation ............................................................................................................................................................................................17
Cartridge Pack Installation ....................................................................................................................................................................17
Initial Rinse ..........................................................................................................................................................................................18
System Sanitization Procedure ............................................................................................................................................................19
Normal Operation..........................................................................................................................................................................................21
Dispensing Water..................................................................................................................................................................................21
Automatic Dispensing ..........................................................................................................................................................................21
User Settings ................................................................................................................................................................................................22
Adjusting Display Brightness ................................................................................................................................................................22
Setting the Displayed Language ..........................................................................................................................................................22
Setting the Date and Time ....................................................................................................................................................................22
Setting the Cell Constant ......................................................................................................................................................................23
Use of Standby Mode ..........................................................................................................................................................................25
Selecting the Set Point ........................................................................................................................................................................25
Resetting the Sanitization Timer ..........................................................................................................................................................26
Resetting the UV Timer ........................................................................................................................................................................26
Setting Volumetric Dispensing ..............................................................................................................................................................27
Setting Timed Dispensing ....................................................................................................................................................................28
Performing a System Flush ..................................................................................................................................................................28
Changing Purity Units ..........................................................................................................................................................................28
Temperature Compensation..................................................................................................................................................................29
Installing Float or Pressure Switch ..............................................................................................................................................................30
N.I.S.T. Calibration Module ..........................................................................................................................................................................31
Performing an Electronic Calibration Using the Optional N.I.S.T. Calibration Module ........................................................................31
Remote Dispenser ................................................................................................................................................................................32
Computer/Printer Setup ................................................................................................................................................................................33
Connecting NANOpure DIamond to Computer and Communicating Through the RS-232 Port Using Hyperterminal or Procomm ..33
Connecting and Starting the Printer ....................................................................................................................................................35
Printer Setup ........................................................................................................................................................................................36
Maintenance and Servicing ..........................................................................................................................................................................38
General Cleaning Instructions ..............................................................................................................................................................38
Cell Cleaning ........................................................................................................................................................................................38
System Sanitization ..............................................................................................................................................................................38
System Depressurization ......................................................................................................................................................................40
Cartridge Pack Replacement ................................................................................................................................................................40
0.2 Micron Filter Replacement..............................................................................................................................................................41
Ultrafilter Replacement ........................................................................................................................................................................42
UV Lamp Replacement ........................................................................................................................................................................42
Fuse Replacement................................................................................................................................................................................44
Shutdown ..............................................................................................................................................................................................45
Troubleshooting ............................................................................................................................................................................................46
Replacement Parts ......................................................................................................................................................................................47
Flow Chart Operational Overview)................................................................................................................................................................52
Wiring Diagram ............................................................................................................................................................................................53
Ordering Procedures ....................................................................................................................................................................................54
One Year Limited Warranty ..........................................................................................................................................................................55
2
Safety Information
Alert Signals
Warning
Warnings alert you to a possibility of personal injury.
Caution
Cautions alert you to a possibility of damage to the equipment.
Note
Notes alert you to pertinent facts and conditions.
Your Barnstead|Thermolyne NANOpure DIamond Life
Science (UV/UF) ultrapure water system has been designed with function, reliability, and safety in mind. It is your responsibility to install it in conformance with local electrical codes. This manual contains important safety information. You must carefully read and understand the contents of this manual prior to the use of this equipment.
For safe operation, please pay attention to the alert signals throughout the manual.
Water purification technology employs one or more of the following: chemicals, electrical devices, mercury vapor lamps, steam and heated vessels. Care should be taken when installing, operating or servicing Barnstead products. The specific safety notes pertinent to this
Barnstead product are listed below.
Note
The UV lamp contains mercury. If broken or no longer needed, do not dispose of the UV lamp in the trash.
Recycle or dispose of the UV lamp as hazardous waste.
Warnings
To avoid electrical shock, always:
1.
Use a properly grounded electrical outlet of correct voltage and current handling capacity.
2.
Do not locate the NANOpure DIamond Life
Science (UV/UF) directly over equipment that requires electrical service. Routine maintenance of this unit may involve water spillage and subsequent electrical shock hazard if improperly located.
3.
Replace fuses with those of the same type and rating.
4.
Disconnect from the power supply prior to maintenance and servicing.
To avoid personal injury:
1.
Do not use in the presence of flammable or combustible materials; fire or explosion may result. This device contains components which may ignite such materials.
2.
This device is to be used with water feeds only.
Sanitizing/cleaning agents must be used in compliance with instructions in this manual. Failure to comply with the above could result in explosion and personal injury.
3
S
AFETY
I
NFORMATION
3.
Avoid splashing disinfecting solutions on clothing or skin.
4.
Ensure all piping connections are tight to avoid chemical leakage.
5.
Ensure adequate ventilation.
6.
Carefully follow manufacturer’s safety instructions on labels of chemical containers and material safety data sheets.
7.
Depressurize system prior to removing the cartridge pack.
8.
This unit is equipped with an ultraviolet lamp.
Ultraviolet radiation is harmful to the eyes and skin. Do not attempt to observe the lamp directly.
9.
Refer servicing to qualified personnel.
10. A full cartridge pack may weigh about 20 lbs.
4
Introduction
Congratulations on your purchase of a Barnstead|
Thermolyne NANOpure DIamond Life Science
(UV/UF) ultrapure water system. This water purification system is designed to provide nuclease-free (RNase and
DNase), DNA-free, pyrogen-free, low TOC, high resistivity, reagent grade water that exceeds ASTM Type I, CAP and NCCLS Type I standards. It uses a four-stage deionization process combined with a UV lamp, an ultrafilter and a 0.2 micron filter to polish suitable feed water (distilled, deionized, or reverse osmosis) to produce pyrogenfree (<0.005 Eu/ml), low TOC (<5.0 ppb) water with a resistivity of up to 18.2 megohm-cm. Water resistivity is continuously monitored by a resistivity cell and displayed on a digital display.
The NANOpure DIamond Life Science (UV/UF) is designed to automatically clean the ultrafilter by flushing water through the filter for one minute every 12 hours.
The water is sent to drain.
The electronics can be verified and calibrated utilizing a
N.I.S.T. Traceable Calibration module. See accessory ordering information.
Please read the instructions carefully to ensure that you receive maximum benefit from the NANOpure DIamond
Life Science (UV/UF). Also, please fill out and return the enclosed warranty registration card as it will help us assure you of proper warranty coverage.
General Usage
Do not use this product for anything other than its intended usage.
5
General Specifications
Dimensions and Clearance Requirements
Dimensions
13.4” W x 19.5” H x 16.9” D (34.0 x 49.5 x 42.8 cm)
Clearances
Sides - 9” (22.9 cm) minimum for servicing.
Above - 3” (7.6 cm) minimum for removal of the top cover.
Front - 4.75” (12.1 cm) minimum for opening the front door.
Feed Water Requirements
Types RO, DI, distilled.
TOC
Turbidity
Less than 1.0 ppm.
1.0 N.T.U. maximum.
Pressure Range Gravity feed to 100 psig (7kg/cm2) maximum.
Temperature Range 4°C - 40°C (40-104°F)
TDS (CaCO3)
Silica
Silt
< 70 ppm
< 1 ppm
< 5% SDI
Product Water
Water Quality
Resistivity
TOC
Endotoxin
> 18.0 m
Ω
-cm
Less than 5.0 ppb
Less than 0.02 Eu/ml
Flow Rate
1.5 lpm maximum at minimum inlet feed water pressure 10 PSIG with a new final filter.
Volumetric Dispense
Accuracy: ± 5%
Repeatability: ± 3%
6
Electrical Requirements
The NANOpure DIamond Life Science (UV/UF) is equipped with 2 power cords to be plugged into an electrical outlet of the appropriate voltage.
Voltage and Frequency (Nominal)
Model D11931 100-240 VAC, 47-63 Hz, 1 phase
G
ENERAL
S
PECIFICATIONS
Environmental Conditions
Operating: 4°C - 49°C; 20% to 80% relative humidity, non-condensing. Installation
Storage:
Category II (over-voltage) in accordance with IEC 664. Pollution Degree 2 in accordance with IEC 664.
Altitude limit: 3,500 meters.
-25°C to 65°C; 10% to 85% relative humidity.
Declaration of Conformity
Barnstead|Thermolyne hereby declares under its sole responsibility that this product conforms with the technical requirements of the following standards:
EMC: EN 61000-3-2 Limits for Harmonic Current Emissions
EN 61000-3-3 Limits for Voltage Fluctuations and Flicker
EN 61326-1 Electrical Equipment for Measurement, Control, and Laboratory Use; Part I:
General Requirements
Safety: EN 61010-1 Safety Requirements for Electrical Equipment for Measurement, Control and
Laboratory Use; Part I: General Requirements per the provisions of the Low Voltage Directive 73/23/EEC, as amended by 93/68/EEC.
The authorized representative located within the European Community is:
Electrothermal Engineering, Ltd.
419 Sutton Road
Southend On Sea
Essex SS2 5PH
United Kingdom
Copies of the Declaration of Conformity are available upon request.
7
8
Unpacking
1.
Remove the unit from its shipping container.
Remove all contents carefully. Ensure that the
UV lamp, ultrafilter, feed and drain tubing, sanitization cartridge, wall bracket, accessory parts bag and power cords (see list below) are removed from the packaging materials
before discarding. Put the NANOpure
DIamond Life Science (UV/UF) on a bench.
UV Lamp
Sanitization Cartridge
Wall Bracket (Unit)
Wall Bracket (Remote Display)
Ultrafilter
LMX13
CMX25
BC1190X10
FL1192X1
Accessory Parts Bag Includes:
Feed Water Tube, 3/8” O.D.
Drain Tube, 14” O.D.
240 V Power Cord w/Fuses
120 V Power Cord w/Fuses
Display Cable
Blank Display
Plug Adapter
CRX70 (2, FZX54)
CRX72 (2, FZX47)
WHX20
CEX42
Also required for installation:
* Customer Supplied Fasteners to mount unit and remote display.
Installation
Caution
Wall composition, condition and construction as well as fastener type must be considered when mounting this unit. The mounting surface and fasteners selected must be capable of supporting a minimum of 150 lbs.(68 kg). Inadequate support and/or fasteners may result in damage to mounting surface and/or equipment. If you are unsure of mounting surface composition, condition and construction or correct fasteners, consult your building maintenance group or contractor.
Warning
Do not locate the NANOpure DIamond
Life Science (UV/UF) directly over equipment that requires electrical service. Routine maintenance of this unit may involve water spillage and subsequent electrical shock hazard if improperly located.
Do not use in the presence of flammable materials; fire or explosion may result. This device contains components which may ignite such materials.
NOTE ********************
The outlet of a gravity feed storage reservoir must be above or at the same level as the inlet of the
NANOpure DIamond Life Science
(UV/UF).
************************
Choosing a Site
The NANOpure DIamond Life Science (UV/UF) features a removable control panel display which allows the system to be mounted almost anywhere within the laboratory.
Use the wall bracket for wall mounted systems as a template to drill mounting holes. (The NANOpure DIamond
Life Science (UV/UF) does not include screws and fasteners for mounting.) Please refer to the “General
Specifications” section for clearance requirements.
9
I
NSTALLATION
Note
Do not install the ultrafilter until the
NANOpure DIamond Life Science
(UV/UF) is ready for use. If the unit sits for longer than 72 hours without use, it will become contaminated and may affect product water quality.
Note
The ultrafilter must be installed before the 0.2 micron filter. Please refer to
“0.2 Micron Filter Replacement” under
“Maintenance and Servicing” for instructions on how to replace the 0.2
micron filter.
Top side port with cap removed
Ultrafilter with fittings
Bottom side port with cap
Ultrafilter Installation
Remove the ultrafilter from its protective bag. Install the ultrafilter as follows:
1.
Open the front (right side) door and right side door.
2.
Remove restraints (twist tie) from the three tubes.
3.
Locate the metal bracket and separate the
Velcro ® strap.
4.
Remove the cap on the top side port of the ultrafilter (make sure the bottom cap on the bottom side port is secure). Insert side ports into the holes in the metal bracket.
5.
Align the ultrafilter with the metal bracket (“This
End Up” should be facing front) and push the bottom fitting tightly into the bottom connector.
6.
Push the top fitting of the ultrafilter into the top connector, making sure the connection is tight.
7.
Secure the ultrafilter with the velcro strap.
8.
Connect tubing with luer fitting to the top side port.
9.
Close and latch right side door and front (right side) door.
Top side port tubing with luer fitting
Top connector
Velcro strap
Metal bracket
Ultrafilter Installation
Bottom connector
® Velcro is a registered Trademark of Velcro Industries B.V.
10
Ultrafilter Connections (Right side view)
I
NSTALLATION
Note
The UV lamp contains mercury. If broken or no longer needed, do not dispose of the UV lamp in the trash.
Recycle or dispose of the UV lamp as hazardous waste.
Caution
DO NOT TOUCH THE GLASS POR-
TION OF THE LAMP! It is recommended that lint-free gloves be worn when handling the lamp. The glass portion must be free of fingerprints, perspiration, etc. Even a light coating of perspiration will reduce the effectiveness of the lamp. If the glass portion of the lamp is touched, clean it with a damp, lint-free cloth: use isopropyl alcohol as required.
UV Lamp Installation
Locate the UV chamber inside the left door of the
NANOpure DIamond Life Science (UV/UF). Install the UV
Lamp as follows:
1.
Disconnect the power cord from the unit.
2.
Access the left side of the unit by unlatching the screw securing the door with a flathead screwdriver. Open the door. The UV lamp will be installed vertically.
3.
Remove the UV lamp from its packaging. DO
NOT TOUCH THE GLASS PORTION OF THE
LAMP! It is recommended that lint-free gloves be worn when handling the lamp. The glass portion must be free of fingerprints, perspiration, etc.
Even a light coating of perspiration will reduce the effectiveness of the lamp. If the glass portion of the lamp is touched, clean it with a damp, lintfree cloth: use isopropyl alcohol as required.
4.
Remove black cap by sliding off UV housing.
5.
Insert the lamp halfway into the UV lamp chamber and plug it in, then fully insert the lamp into the chamber.
6.
Make sure the UV lamp cable is tucked behind the chamber before closing and relatching the door.
Other Accessories
Optional accessories for this unit include a float switch, low pressure switch, N.I.S.T. Traceable Calibration
Module, remote dispenser, and printer. Installation procedures for these accessories are explained in later sections of this manual.
UV Lamp Installation - Left side view
11
I
NSTALLATION
Bench Mounting
1.
Place the NANOpure DIamond Life Science
(UV/UF) on a bench top that is accessible to pretreated water, electricity and an atmospherically vented drain.
Caution
Wall composition, condition and construction, as well as fastener type, must be considered when mounting this unit. The mounting surface and fasteners selected must be capable of supporting a minimum of 150 lbs.(68 kg). Inadequate support and/or fasteners may result in damage to mounting surface and/or equipment. If you are unsure of mounting surface composition, condition and construction or correct fasteners, consult your building maintenance group or contractor.
Note
Please refer to the “General
Specifications” section for clearance requirements.
Warning
Do not locate the NANOpure DIamond
Life Science (UV/UF) directly over equipment that requires electrical service. Routine maintenance of this unit may involve water spillage and subsequent electrical shock hazard if improperly located.
Wall Mounting
Install the NANOpure DIamond Life Science (UV/UF) on a wall in a convenient location that is accessible to water, an atmospherically vented drain and electricity.
1.
Locate the wall bracket packed separately from the unit.
2.
Using the wall bracket as a template, locate and drill the mounting holes in the wall. A minimum of four (customer-supplied) fasteners will be required — two on the top and two on the bottom.
3.
Attach the wall bracket to the wall using the customer-supplied fasteners.
4.
Remove the locking screws on each side of the wall bracket.
5.
Pull the two locking slides on each side of the wall bracket out as far as they will go.
6.
Hang the unit on the wall bracket by sliding the mounting pins into the wall bracket slots.
7.
Push the locking slides on each side of the wall bracket in as far as they will go.
8.
Replace the locking screws.
NOTE ********************
The outlet of a gravity feed storage reservoir must be above or at the same level as the inlet of the
NANOpure DIamond Life Science
(UV/UF).
************************
12
I
NSTALLATION
Note
The removable control panel was not designed to be repeatedly removed from the unit, therefore, you may experience difficulty when attempting to remove it. It is therefore recommended that the control panel be permanently mounted in a remote location or remain intact with the unit.
Note
The blank protective cover on the rear side panel display connector should remain in place when the display is not remotely located.
Blank panel
Wall Mount
Bracket
Caution
Repeated removal and replacement of the control panel from the top cover may eventually cause it to become damaged.
Installing the Control Panel in a
Remote Location
For your convenience, the control panel can be removed from the unit and mounted at a convenient location within
10 ft. (3.1 m.) of the unit. To remove the control panel from the unit:
1.
Turn the unit OFF and disconnect it from the power supply.
2.
Slide the control panel upward about 1/4” (.64
cm) to 3/8” (.95 cm) - (just enough to clear connector on top cover), pull forward, and remove.
(The control panel fits tightly in the unit so it may be difficult to remove.)
3.
Replace the control panel with the blank panel provided with the unit to protect the electrical connector.
4.
Using the wall bracket as a template, locate and drill the mounting holes in the wall. A minimum of two (customer-supplied) fasteners will be required. Attach the wall bracket to the wall using the customer-supplied fasteners.
5.
Locate the 25-pin connector on the back top right of the unit. Remove the protective black cover and store it for future use if the control panel is returned to the top cover.
6.
Locate the 10 ft. (3.1 m.) 25 pin M-F cable included with the unit; attach the female end to the 25 pin connector on the back top right of the unit, and the other end to the connector on the bottom of the control panel.
7.
Tighten the cable screws on each end to secure the cable to the unit and control panel.
8.
The control panel may now be mounted on the wall.
13
I
NSTALLATION
Water Connections
Caution
Do not connect feed water until directed to do so during “Initial Sanitization.”
Pump
Interlock
Feed Water Connection
1.
Locate the length of 3/8” O.D. tubing provided with the unit with a quick disconnect insert on one end and a 3/8” (.95 cm) O.D. X 1/4” (.64
cm) NPT tubing adapter on the other.
2.
Install the tubing adapter onto your incoming water line. We recommend a customer supplied shut off valve be installed in your feed water line. Do not connect the feed water to your NANOpure
DIamond Life Science (UV/UF). You will connect the feedwater during the “Initial Sanitization” procedure explained later in this manual. Note:
The cartridge pack and ultrafilter must be installed before connecting water line.
3.
If a pressure switch is to be used, see the
“Installing a Float or Pressure Switch” section for more information.
Remote 2
Remote 1
Feed
Water Inlet
Atmospheric
Drain
Atmospheric Drain Connection
When the NANOpure DIamond Life Science (UV/UF) flushes its membrane, the water used is sent to drain through this connection. To install:
1.
Locate the drain water tubing. This is the 1/4”
(.64 cm) O.D. tubing that is approximately 6 ft.
(1.9 m.) long. The atmospheric drain fitting is located on the lower left side of the NANOpure
DIamond Life Science (UV/UF).
2.
Route the other end of the drain water tubing to an atmospherically vented drain and make a connection. Ensure there are no kinks in the tubing and that it proceeds in a downward plane. Proceed to “Initial Operation.”
Water Connections - Left side view
14
Controls
Note
The removable control panel was not designed to be repeatedly removed from the unit, therefore, you may experience difficulty when attempting to remove it. It is therefore recommended that the control panel be permanently mounted in a remote location or remain intact with the unit.
LED
NANOpure DIamond “
Main Power Switch
The main power switch on the NANOpure DIamond Life
Science (UV/UF) is located on the lower back right side of the unit (as you face the front of the unit), directly above the power cord receptacle.
Control Panel
The NANOpure DIamond Life Science (UV/UF) is controlled through a panel which incorporates switches to control its functions: START/STOP, STANDBY, DIS-
PENSE, DOWN, UP, ENTER and BACK. This control panel utilizes a digital display to show system information and the purity of the product water. The intensity (brightness) of the display may be adjusted according to user preference. See “Adjusting Display Brightness” in the
“User Settings” section of this manual.
A single LED located on the top of the control panel will illuminate to inform you of the system status as follows:
• LED is not illuminated: System is OFF or in
(Idle) mode.
• Solid Green LED: System is recirculating and operating normally.
• Blinking Green LED: System is in Standby mode.
• Red LED: System is alerting user to an operational error or maintenance issue.
The control panel can be removed from the unit and remotely mounted. Please refer to “Installing the Control
Panel in a Remote Location” in the “Installation” section of this manual for more information about how to remove the control panel from the unit.
STANDBY
DISPENSE
START
BACK
STOP
Barnstead
ENTER
NANOpure DIamond Life Science (UV/UF)
Control Panel
15
C
ONTROLS
Note
The unit MUST be in the normal recirculating mode (when the display is showing purity) in order to dispense water.
Switches
When the main power switch (on the lower back right side of the unit as you face the unit) is ON, the switches on the control panel function as follows:
START/STOP allows you to alternate the unit between the normal (recirculation) mode; when the display is showing purity, and the (Idle) mode.
STANDBY allows you to put the unit into standby, recirculating water for 10 minutes/hour. The display will read
“Standby” during periods of inactivity and “Recirculating” and the time remaining during the 10 minute recirculation.
DISPENSE allows you automatically deliver water from the unit. Please refer to”Automatic Dispensing” in the
“User Settings” section of this manual for information.
UP and DOWN arrows allow you to scroll between menu options/items and values.
BACK allows you to return to the previous menu option/item. Please note that use of the BACK switch from a selection screen will return you to the previous menu option/item while maintaining the selection’s value upon entry.
ENTER allows you to activate a selected menu option/item and also functions as “Yes” whenever an option with a question mark appears.
DISPENSE KNOB when the unit is in the recirculation mode, push to the right to deliver water continuously until pushed back to the middle. Push to the left to deliver water manually until button is released.
16
Initial Operation
Warning
Depressurize system prior to removing cartridge pack.
Caution
Do not allow the NANOpure DIamond
Life Science (UV/UF) to operate unless water is available to the unit.
Note
Do not install the 0.2 micron filter and bell assembly at this time.
Manifold
Cartridge packs will come bagged with four manifold connection caps.
Part Application
DIamond Kit Organic Free R/O & Distilled Feed
D50280 Ultra-Low Organics, Type 1 Water,
Reverse Osmosis or Distilled
Water Feed
DIamond Kit Organic Free Deionized Feed
D50281 Ultra-Low Organics, Type 1 Water,
Deionized Water Feed
DIamond Kit Type 1 R/O & Distilled Feed
D50282 Low Organics, Type 1 Water,
Reverse Osmosis or Distilled
Water Feed
DIamond Kit Type 1 Deionized Feed
D50283 Low Organics, Type 1 Water, DI
DI Water Feed
Each cartridge pack includes one 0.2 micron absolute final filter.
Cartridge Pack Installation
1.
Open front (left side) door. Turn the cartridge pack so the caps are facing you. Remove the caps.
2.
Verify that each of the four posts on the cartridge pack has an o-ring.
3.
Lift unit manifold upwards, insert cartridge pack and align the pack so that it mates with the unit manifold.
4.
Lower unit manifold until it is flush with the top of the cartridge pack.
5.
Hand tighten wing head screws securely.
6.
Close door.
17
I
NITIAL
O
PERATION
18
Warning
Use a properly grounded electrical outlet of correct voltage and current handling capacity.
Initial Rinse
1.
Attach the feed water line to the unit by snapping the quick disconnect coupling into the quick disconnect body in the lower left back of the unit.
2.
Attach an atmospheric drain line [1/4” (.64 cm) ] tubing by pushing into the quick connect fitting on the lower left back of unit. See layout connection under the “Water Connections” section.
3.
Check to ensure the dispense knob is in the
OFF (middle) position.
4.
Select the appropriate power cord, remove and install the two fuses into the fuse draw in the power entry module.
5.
Connect/attach to proper electrical outlet and the unit.
6.
Turn the unit on by turning the main power switch to the “I” (ON) position.
7.
The system greeting will display the type of unit you own. This greeting is “UV/UF.”
8.
The system electronics will initialize and check its calibration. (All units have been factory calibrated.) If calibration is OK the display will show
“Calibration (passed).” The display will next read
“Self Test (in progress) Self Test. Passed.”
Finally, the display will read “10 (±0.2) M
Ω
-cm.”
This is a reading of the calibration reference value.
9.
From the (Idle) mode, when the display reads,
“NANOpure Diamond (Idle) xx:xx:xx xx/xx/xx,” press ENTER to run an air purge.
10. Unit will display, “Air Purge?” Press ENTER.
11.
Upon completion of the air purge, complete the
“System Sanitization Procedure.”
I
NITIAL
O
PERATION
Warning
Avoid splashing disinfecting solution on clothing or skin.
Ensure all piping connections are tight to avoid leakage.
Ensure adequate ventilation.
Carefully follow manufacturer’s safety instructions and material safety data sheets.
This device to be used with water feeds only.
Sanitizing/cleaning agents must be used in compliance with instructions in this manual. Failure to comply with the above could result in explosion and personal injury.
Note
After sanitization is complete, the sanitization timer will automatically be reset for approximately six months.
System Sanitization Procedure
Your NANOpure DIamond Life Science (UV/UF) has been shipped with a cleaning syringe (Cat. No. CMX25). When the system is new or whenever a new cartridge pack is installed, the system should be sanitized (cleaned). For
UV/UF units, it is best to perform sanitizing procedures when the system or water is not required. The entire sanitization procedure will take approximately 3.5 hours to complete. Sanitize your NANOpure DIamond as follows:
1.
From the (Idle) display, press the UP or DOWN arrow until display reads, “Sanitize Unit?”
2.
Press ENTER.
3.
Unit display will ask if you have “Disconnected
Inlet Water (NO).”
4.
Disconnect feed water supply at the quick-disconnect inlet valve on left side of unit.
5.
Press the UP or DOWN ARROW to select (YES) and then press ENTER. Display will now read,
“Injected Sanitant (NO).”
6.
Open front (left side door) to expose cartridge pack. Remove solution syringe from wrapper.
7.
Remove luer cap on system injection port, located on the right side of the cartridge pack manifold by turning counterclockwise.
8.
Remove syringe luer cap and attach syringe to the system luer fitting.
9.
Slowly inject solution into system and remove syringe. Avoid injecting air into the system.
10. Replace luer cap on injection port and close door.
11.
Press the UP or DOWN arrow to select (YES) and then press ENTER. Display will now read,
“Reconnected Water (NO).”
12. Attach feed water supply removed in step 4.
13. Press the UP or DOWN arrow to select (YES)
19
I
NITIAL
O
PERATION
0.2 Micron
Filter
Note
For more demanding applications where low TOC water is required, a rinse of 15-20 liters through the cartridges and 0.2 micron filter may be necessary.
and press ENTER. Display will now read, “Unit
Sanitizing.”
14. Unit may now be left unattended until display reads, “Sanitization Complete: Press ENTER.”
15. After sanitization is complete the unit will return to the (Idle) mode.
16. Press the START/STOP switch to return the unit to the normal recirculation mode.
17. Remove the new 0.2 micron filter and bell assembly from the bag and insert it into the dispenser arm manifold. Gently turn it clockwise until it is fully seated.
18. Remove the protective cap from the filter bell.
Turn the dispense knob to the left or right and flush 1-2 liters of water through the 0.2 micron filter. The product water is now ready for use.
20
Note
On initial start-up, the purity meter may display “...M
Ω
-cm or uS/cm.” This is caused by air in the cell and should be replaced by a resistivity reading almost immediately. If the dots aren’t replaced by a value after one minute, refer to the “Troubleshooting” section of this manual.
Note
Do not turn off the NANOpure
DIamond Life Science UV/UF during non-work hours. Doing so will allow bacterial growth and other contamination of the water in the system. As a result, the system will require a rinseup period at the beginning of the work day to achieve high quality product water. We recommend using the
Standby Mode.
Note
From the normal recirculating mode when the display is showing purity, press ENTER to show the current temperature. The temperature will display for about 5 seconds before returning to the normal recirculating mode.
Normal Operation
1.
Turn the system power ON by depressing the main power switch to the “I” position.
2 The system greeting display indicates the type of unit you own. This greeting is “UV/UF.”
3.
The system electronics will initialize and check its calibration. (All units have been factory calibrated.) If calibration is OK the display will show
“Calibration (passed).” The display will next read
“Self Test (in progress) Self Test. Passed.”
Finally, the display will read “10 (±0.2) M
Ω
-cm.”
This is a reading of the calibration reference value.
4.
From an (Idle) display, press the START/STOP switch on the control panel to enter the normal recirculating mode. The unit’s pump will begin to run.
5.
The display will then begin reading resistivity of the product water. Initially it will read “.... M
Ω
cm” until the air is bled from the system.
6.
Allow the water’s resistivity to rise to the desired purity before drawing off water.
7.
The system should be left on or in standby during the work day. (See “Use of Standby Mode.”)
Dispensing Water
Use the dispense knob for manual dispensing or, use the
“Dispense” switch for volumetric or timed dispensing. See
“User Settings.”
Note
Press any switch except ENTER to stop dispensing. Display will read,
“Auto Dispense Interrupted by User.”
Automatic Dispensing
(Factory set to “OFF.”) This feature is accessible only from the normal recirculating mode when the display is showing purity.
1.
Place a container under the dispenser.
2.
Press DISPENSE. Water will start dispensing
immediately! Display will read, “xx.M
Ω
-cm xx
Liters Remaining” or “xx.M
Ω
-cm xx Seconds
Remaining,” or “AUTO DISPENSE <OFF>” depending on which dispensing method was last chosen in the auto dispense menu.
21
User Settings
22
Adjusting Display Brightness
1.
From the (Idle) mode, when the display reads,
“NANOpure Diamond (Idle) xx:xx:xx xx/xx/xx,” press the UP or DOWN arrow until display reads, “Advanced Menu?”
2.
Press ENTER.
3.
Press ENTER to proceed through the options until “Set Display Brightness” is displayed.
4.
Press the UP or DOWN arrow to choose “Yes.”
5.
Press ENTER.
6.
Press the UP or DOWN arrow until the desired value (100%, 75%, 50% or 25%) is displayed.
7.
Press ENTER.
8.
Press BACK until you return to the (Idle) display.
Note
Time values are displayed in 24 hr.
military time. For example; 6:00p.m. =
18:00:00 = hr./min./sec.)
Setting the Displayed Language
1.
From the (Idle) mode, when display reads,
“NANOpure DIamond (Idle) xx:xx:xx xx/xx/xx,” press the BACK switch until the display reads “LANGUAGE.”
2.
The currently selected language is displayed on line 2 of the display.
3.
Press the UP or DOWN arrow to select the language desired.
4.
Press ENTER.
Setting the Date and Time
1.
From the (Idle) mode, when the display reads,
“NANOpure DIamond (Idle) xx:xx:xx xx/xx/xx,” press the UP or DOWN arrow until display reads, “Advanced Menu?”
U
SER
S
ETTINGS
2.
Press ENTER.
3.
Press ENTER when the unit displays “Printer
Attached?”
4.
When prompted with “Date Format” press the
UP or DOWN arrow to show the date format desired. Press ENTER to select the desired format and continue.
5.
When prompted with “Set Date,” press the UP or DOWN arrow to choose “Yes,” then press
ENTER.
6.
Press the UP or DOWN arrow until you reach the desired month. Press ENTER.
7.
Press the UP or DOWN arrow until you reach the desired day. Press ENTER.
8.
Press the UP or DOWN arrow until you reach the desired year. Press ENTER.
9.
When prompted with “Set Time” press the UP or
DOWN arrow to choose “Yes,” then press
ENTER. This is to set the time of day.
10. Press the UP or DOWN arrow until you reach the desired hour value. Press ENTER.
11.
Press the UP or DOWN arrow until you reach the desired minute value. Press ENTER.
12. Press BACK twice to return to the (Idle) mode.
Setting the Cell Constant
The actual temperature and cell constants are attached to the cell cable internal to the NANOpure DIamond unit.
This data helps ensure that the purity displayed is as accurate as possible. The only time the user would need to enter this data is if the cell or main board is being replaced and/or updated.
1.
Shut power OFF to the unit and disconnect the power cord.
23
U
SER
S
ETTINGS
2.
Unlatch and open the right side door.
3.
Find and copy down the two cell constants attached to the gray cable, leaving the electronics board mounted on the inside of the right side door.
4.
Close and latch the right side door.
5.
Reconnect the power cord and turn unit ON.
6.
From the (Idle) mode, when the display reads,
“NANOpure DIamond (Idle) xx:xx:xx xx/xx/xx,” press the UP or DOWN arrow until the display reads, “Advanced Menu.”
7.
Press ENTER.
8.
Proceed through the options by pressing
ENTER until “Set Cell Constant” is displayed.
9.
Press the UP or DOWN arrow to select “Yes.”
10. Press ENTER.
11.
Note that the digits are entered from left to right.
12. Press the UP or DOWN arrow to change the
“blinking” digit. Once the digit is correct, press
ENTER to advance to the next digit.
13. After all digits have been entered, line 2 of the display will read, “(NO) Finished.”
14. If the cell constants displayed on line 1 of the display are correct, press the UP or DOWN arrow to select “Yes.”
15. Press ENTER.
16. Press BACK twice to return to the (Idle) mode.
24
U
SER
S
ETTINGS
Note
When in Standby mode, press
START/STOP to continue normal operation.
Use of Standby Mode
At the end of the work day, place the NANOpure
DIamond Life Science (UV/UF) in Standby mode for the night. To place the unit in Standby mode:
1.
From the (Idle) mode, when display reads,
“NANOpure DIamond (Idle) xx:xx:xx xx/xx/xx,” press STANDBY.
—OR—
2.
From the xx.x M
Ω
-cm or x.xx uS/cm operational display, Press STANDBY.
After STANDBY is pressed the display will read “Standby” and the green LED will slowly blink. For 10 minutes of every hour while the pump is energized, the display will read “Standby (Recirculating) xx minutes.” Every fourth time the pump goes on in the standby mode, the UV light will go on. The lamp is on 10 minutes out of every four hours. Press START/STOP to return to the recirculating mode.
Selecting the Set Point
The NANOpure DIamond Life Science (UV/UF) electronics include a user programmable set point which alerts you when water quality falls below the programmed set point. The set point is user selectable from 1-18 M
Ω
-cm
(factory set at 10 M
Ω
-cm.).
1.
From the (Idle) mode, when display reads,
“NANOpure DIamond (Idle) xx:xx:xx xx/xx/xx,” press the UP or DOWN arrow until display reads, “Advanced Menu?”
2.
Press ENTER.
3.
Proceed through the options by pressing
ENTER until “Adjust Set Point” is displayed.
4.
Press the UP or DOWN arrow to choose “Yes.”
5.
Press ENTER.
6.
Press the UP or DOWN arrow until the desired value is displayed.
25
U
SER
S
ETTINGS
26
7.
Press ENTER.
8.
Press BACK to return to the (Idle) mode.
The second line of the display will show a “Below Set point” message once every 8 to 10 seconds when the measurement is below the set point. (The set point indication is set at the factory at 10 M
Ω
-cm.)
Resetting the Sanitization Timer
1.
From the (Idle) mode, when the display reads,
“NANOpure DIamond (Idle) xx:xx:xx xx/xx/xx,” press the UP or DOWN arrow until the display reads, “Reset Timers?”
2.
Press ENTER. Display will read, “Reset UV
Timer (NO).”
3.
Press ENTER.
4.
Display will read, “Reset San Timer” and show on the right side of the second line when the user will start receiving “Sanitize Unit” reminders.
5.
Press the UP or DOWN arrow to select “Yes.”
6.
Press ENTER and the timer will be reset. This will reset the sanitization timer for approximately six months calendar time.
Resetting the UV Timer
When the NANOpure DIamond Life Science (UV/UF) is in an (Idle) position:
1.
From the (Idle) mode, when display reads,
“NANOpure DIamond (Idle) xx:xx:xx xx/xx/xx,” press the UP or DOWN arrow until the display reads, “Reset Timers?” Press ENTER.
2.
Press the UP or DOWN arrow until the display reads, “Reset UV Timer.”
U
SER
S
ETTINGS
3.
Press the UP or DOWN arrow to select “Yes.”
4.
Press ENTER and the timer will be reset for approximately five months of operational time.
The display will now display “Rest San Timer.” Press
BACK two times to return to the (Idle) mode.
Note
Volumetric dispensing is based upon total fluid volume entering the
NANOpure DIamond Life Science
(UV/UF). Point of drawoff accuracy will be compromised if you are dispensing water volumetrically while simultaneously drawing water manually through the remote dispenser accessory.
Note
Once Volumetric or Timed Dispensing values are set, they will remain in memory until changed by the user. To turn the dispense method OFF, select
“OFF” in the “Dispense Method” menu.
Setting Volumetric Dispensing
1.
From the normal recirculating mode, when the display is showing purity, press the UP or
DOWN arrow until the display reads, “Auto
Dispense Menu?”
2.
Press ENTER.
3.
Display will read, “Dispense Method.” Press the
UP or DOWN arrow to select VOLUMETRIC.
4.
Press ENTER. Display will read, “Volume
Setting: xx Liters.”
5.
Press the UP or DOWN arrow to set the desired volume value. Press ENTER.
UP arrow: +0.25 up to 1 L, then 5 L, then incremental by
5 L
DOWN arrow: Decrement by 1 L until 1 L, then by 0.25 L
Maximum: 60 L
Minimum: 0.25 L
27
U
SER
S
ETTINGS
28
Setting Timed Dispensing
1.
From the normal recirculating mode, when the display is showing purity, press the UP or DOWN arrow until the display reads, “Auto Dispense
Menu?”
2.
Press ENTER.
3.
Display will read, “Dispense Method.” Press the
UP or DOWN arrow to select TIME.
4.
Press ENTER. Display will read, “Time Setting: xx
Minutes.”
5.
Press the UP or DOWN arrows to set the desired time value. Press ENTER.
UP arrow: +1 up to 5 min., then increments by 5 minutes
DOWN arrow: Decrement by 1 minute
Maximum: 40 minutes
Minimum: 1 minute
Note
An automatic 1-minute flush will occur once every 12 hours. If a unit is (Idle) or in STANDBY for over 12 hours, an automatic system flush will occur immediately after the START/STOP switch is pressed.
Performing a System Flush
1.
From the normal recirculating mode, when the display is showing purity, press the UP or DOWN arrow until the display reads, “System Flush?”
Press ENTER.
2.
Press the UP or DOWN arrow to select “Short
Flush?” or “Extended Flush?” Press ENTER.
Note
Press the STOP/STOP switch to cancel the flush.
3.
Display will read, “System Flushing: (xxx) Seconds or (xxx) Minutes Remaining.”
Changing Purity Units
(Factory set to “m
Ω
-cm.”)
1.
From the normal recirculating mode, when the display is showing purity, press the Up or DOWN arrow until display will reads, “Change Units?”
Press ENTER.
U
SER
S
ETTINGS
2.
Press the UP or DOWN arrow to select “M
Ω
-cm” or “uS/cm” Press ENTER.
3.
Press BACK to return to the normal recirculating mode.
Note
Purity displayed will have an asterisk
(*) the the left of the reading (* xx.x
M
Ω
-cm or * x.xx uS/cm) to inform the user that uncompensated purity is being displayed.
Temperature Compensation
(Factory set to “ON.”)
1.
From the normal recirculating mode, when the display is showing purity, press the Up or
DOWN arrow until the display reads,
“Temperature Compensation?” Press ENTER.
2.
Press the UP or DOWN arrow to select “OFF” or “ON” Press ENTER.
3.
Press BACK to return to the normal recirculating mode.
29
30
Installing Float or Pressure
Switch
Accessories D8964 (float switch) and D2706 (pressure switch) are designed to protect the NANOpure DIamond
Life Science (UV/UF) pump by alerting the system of an inadequate feed water condition so that the pump can be shut down. If an inadequate feed water condition exists and the pump is shut down due to this condition, the display will read “Check Inlet.” Follow the proceeding instructions for installation.
1.
Disconnect the unit from the electrical power.
2.
If using the D8964 float switch, follow the installation instructions included with the float switch for installation into a tank.
3.
If using the D2706 low pressure switch, install the PVC tee (supplied with D2706) in the incoming water line. Screw the switch into the top of the tee, then connect the inlet tubing of the NANOpure DIamond Life Science (UV/UF) into the remaining opening of the PVC tee.
4.
Route the cable from the float or low pressure switch to the top left of the NANOpure DIamond
Life Science (UV/UF).
5.
Remove the jumper plug from the back, upper left portion of the unit and save for future use.
6.
Plug the cable into the jumper plug outlet.
7.
Reconnect the electrical power.
N.I.S.T Calibration Module
PC Board
J2
Connection
Note
The calibration (in progress) may take up to two minutes to complete.
Performing an Electronic
Calibration Using the Optional
N.I.S.T. Calibration Module
If you purchased the optional N.I.S.T. calibration module
(Catalog No. E896X5) you can perform a calibration of the NANOpure DIamond purity sensing electronics traceable to N.I.S.T. standards.
1.
Disconnect power cord from the unit.
2.
Unlatch and open the right door. Remove the
2 screws and remove the cover plate over the
PC boards.
3.
Locate the outlet cell connection on the circuit board and remove the outlet cell connection at P2.
4.
Connect the N.I.S.T. calibration module to the connection point (J2), “To Outlet Cell” on the calibration board (PC1190X2).
5.
Close and latch door.
6.
Reconnect the power cord and turn the unit on by depressing the main power switch.
7.
From the (Idle) mode, Press the UP or DOWN arrow until the display reads, “N.I.S.T.
Calibrate?” Press ENTER.
8.
Display will show “Calibration Module
Installed?” Press the UP or DOWN arrow to select (YES).
9.
Press ENTER.
10. Display will read “Calibration (In Progress).”
When the calibration is complete the display will read “Calibration (passed/failed) for 4 seconds and then return to the”N.I.S.T.
Calibrate?” display.
31
N.I.S.T. C
ALIBRATION
M
ODULE
11.
Press BACK to return to the (Idle) mode.
12. Turn the main power switch to the OFF “O” position. Unplug the unit.
13. Remove the calibration module and reconnect the cell.
14. Reattach the cover plate and close and latch the door.
15. Reconnect the power cord.
The calibration procedure is complete.
Remote Dispenser
Refer to the Operating Instructions for your remote dispenser.
32
Computer/Printer Setup
Note
Due to character set differences, hyperterminal may display the
Ω symbol as ê. Also, the ° symbol may be displayed as ø.
Connecting NANOpure
DIamond to Computer and
Communicating Though the
RS-232 Port Using
Hyperterminal or Procomm
Hyperterminal
RS-232 Capture Instructions
1.
Connect the 9 pin serial cable (part no. WHX18) from the NANOpure DIamond RS-232 port to
COM2: port or (COM1: port) on back of computer.
2.
From the (Idle) display, press the UP or DOWN arrow until display reads, “Advanced Menu.”
3.
Press ENTER.
4.
From the “Printer Attached” display, press the
Up or DOWN arrow for (NO).
5.
Press ENTER.
6.
Now, during the normal Recirculation mode, date, time, purity and temperature readings will be sent to the computer once per minute.
7.
Run your Microsoft Windows program. Open
Hyperterminal (located in the accessories directory), give your setting a name and choose a symbol.
8.
In the “Connect Using:” box, select “direct to”
COM2: or (COM1: ) as port (depending on your computer), and click OK.
9.
Using the selection boxes in the next screen, configure the options as 9600 baud, 8-bit, No parity, 1 stop bit, Flow Control: None.
10. To receive data to a file, select Transfer and
33
C
OMPUTER
/P
RINTER
S
ETUP capture text. Accepting default will put the file which you name in the c:/windows directory called capture.txt and the file will be an ASCII text file.
11.
You should see the data on the computer screen.
12. When you are finished storing data from the
NANOpure DIamond, save your file. You can exit Hyperterminal or set up a new experiment.
13. Hyperterminal data can be manipulated and graphed in Excel
®
.
Procomm
1.
Connect the 9 pin serial cable from the
NANOpure DIamond RS-232 port to COM2: port or (COM1: port) on back of computer.
2.
From the (Idle) display, press the UP or DOWN arrow until the display reads, “Advanced Menu.”
3.
Press ENTER.
4.
From the “Printer Attached” display. Press the
Up or DOWN arrow for (NO).
5.
Press ENTER.
6.
The unit will now, during the normal
Recirculation mode send a purity and temperature reading to the computer once per minute.
7.
Open DOS window and change directory to procomm directory.
8.
Open procomm executable.
9.
Press ALT-F10 for configuration screen.
10. Enter into Line Settings screen by typing ALT-P.
11.
Type in 11 <Enter>, sets parameters as 9600, 8,
N, 1.
34
C
OMPUTER
/P
RINTER
S
ETUP
12. Type in 21 <Enter>, sets to COM2: or (Type in
20 <Enter>, sets to COM1:), depending on your computer.
13. Type in 24 <Enter>, saves settings.
14. Type ESC to exit setup menu.
15. Procomm is now ready to accept input from the
NANOpure DIamond.
16. To begin downloading a file, Type ALT-F1, and give the file an appropriate name when prompted.
The optional printer (part no. AY1137X1) provides a paper tape record for future reference.
During normal Recirculation, the printer prints date, time, purity and temperature readings once every minute.
Connecting and Starting the
Printer
1.
Make sure the printer and NANOpure DIamond are turned OFF.
2.
Connect the NANOpure DIamond to the printer via the RS232 port. Use the tan printer cable included with the Epson printer. Plug the printer power cord into an electrical outlet.
3.
Turn the NANOpure DIamond ON.
4.
Turn the printer ON.
5.
From the (Idle) display, press the UP or DOWN arrow until display reads, “Advanced Menu.”
6.
Press ENTER.
35
C
OMPUTER
/P
RINTER
S
ETUP
7.
From the “Printer Attached” display, press the
Up or DOWN arrow for (YES).
8.
Press ENTER.
9.
Press BACK twice to return to the (Idle) mode.
10. During normal Recirculation mode, a purity and temperature reading will be printed once every
12 minutes.
ON/OFF Button
Figure A: Printer
Paper Feed Button Paper Out Light
Error Light
Power Light
Printer Setup
Power Switch
Located on the front of the printer (AY1137X1), this switch turns power to the printer ON and OFF. (See Figure A.)
The green POWER light will illuminate when the printer is
ON.
Paper Feed Button
Press this button once to advance the tape one line, or hold the button down to advance paper continuously.
When the paper roll nears the end, the red PAPER OUT light will illuminate.
If the red ERROR light illuminates, the printer is off line.
This could be a result of the print head being too hot, in which case the printer will resume printing once it cools down; or—the paper may be jammed in the printer. Turn the printer OFF before checking for jammed paper.
Remove the paper jam and turn the printer back ON. If the printer still will not print, unplug it and refer servicing to qualified personnel.
Figure B: Control Panel
36
C
OMPUTER
/P
RINTER
S
ETUP
Note
A colored stripe along the edge of the paper indicates that you are approaching the end of the tape. When the stripe appears, approximately 23 inches (58.4 cm) of paper remains, enough for 1-2 days. Replace the tape before it is completely exhausted to ensure a complete record of your analyses.
Installing the Paper Roll
When the red PAPER OUT light illuminates, it is time to replace the paper roll. Replace the paper roll (AY669X4) as follows:
1. Turn the printer ON and open the chamber cover.
2.
Make sure the edge of the paper roll is straight.
3.
Insert the paper roll into the printer. (See Figure
C.)
4.
Insert the paper straight into the paper slot (see
Figure D). The paper will feed automatically.
5.
Tear off a strip of paper and close the cover.
(The red PAPER OUT light will be OFF if paper roll installation is successful.)
Figure C: Inserting the Paper Roll
Figure D: Inserting the Paper Into the Paper Slot
37
Maintenance and Servicing
38
Warning
• Avoid splashing mild detergent or weak acid on clothing or skin
• Ensure all piping connections are tight to avoid chemical leakage
• Carefully follow manufacturer's safety instructions on labels of chemical containers and material safety data sheets.
Caution
Feed water must be disconnected and the system must be depressurized before performing the cell cleaning procedure.
General Cleaning Instructions
Disconnect electrical service to the unit. Wipe exterior surfaces with lightly dampened cloth containing mild soap solution.
Cell Cleaning
1.
Disconnect inlet water and depressurize. See
“System Depressurization” section.
2.
Turn unit OFF. Open right side door and remove pc board cover.
3. Disconnect cable from J2 and snip plastic cable tie located on the inside top of unit.
4.
Unscrew and remove the cell by turning the gray bushing counterclockwise. Carefully remove Oring to clean the cell.
5.
Wash the cell in a mild detergent solution or a
10% inorganic acid solution (follow acid manufacturers recommended handling procedure).
This may be done in an ultrasonic cleaner or with a soft brush.
6.
Thoroughly rinse the cell in deionized or distilled water following the detergent or acid cleaning.
7.
After cleaning, check the o-ring; replace if necessary.
8.
Reinstall the cell into the cell well and hand tighten.
9.
Reroute the cable and reconnect to J2 on circuit board.
10. Replace pc board cover and close door.
11.
Reconnect water and turn unit ON.
M
AINTENANCE AND
S
ERVICING
Note
The sanitization timer will display after six months, reminding you to sanitize the unit. Complete the system sanitization procedure and reset the sanitization timer according to the
“Resetting the Sanitization Timer” section in this manual.
Warning
Disconnect from the power supply prior to maintenance and servicing.
Warning
Refer servicing to qualified personnel.
Warning
Avoid splashing disinfecting solution on clothing or skin.
Ensure all piping connections are tight to avoid leakage.
Ensure adequate ventilation.
Carefully follow manufacturer’s safety instructions and material safety data sheets.
This device to be used with water feeds only.
Sanitizing/cleaning agents must be used in compliance with instructions in this manual. Failure to comply with the above could result in explosion and personal injury.
System Sanitization
For UV/UF units, it is best to perform cleaning procedures early in the afternoon. After sanitization is complete
(approximately 3.5 hours), place the unit into its normal
(recirculation) mode to allow the system to recirculate overnight after cleaning.
1.
From the (Idle) mode, press the UP or DOWN arrow until the display reads, “Sanitize Unit?”
2.
Press ENTER.
3.
Display will prompt, “Disconnected Inlet Water
(NO).”
4.
Disconnect feed water supply at the quick-disconnect inlet valve.
5.
Press the UP or DOWN ARROW to select (YES) and then press ENTER. Display will now prompt, “Injected Sanitant (NO)?”
6.
Remove solution syringe from wrapper.
7.
Open front (left side) door. Remove luer cap on system injection port, located on the right side of the cartridge pack manifold by turning counterclockwise.
8.
Remove the syringe luer cap and attach the syringe to the system luer fitting.
9.
Slowly inject the solution into the system and remove syringe. Avoid injecting air into system.
10. Replace luer cap on injection port. Close door.
11.
Press the UP or DOWN arrow to select (YES) and then press ENTER. Display will now read,
“Reconnected Water (NO).”
12. Attach feed water supply removed in step 4.
13. Press the UP or DOWN arrow to select (YES) and press ENTER. Display will now read, “Unit
Sanitizing.”
39
M
AINTENANCE AND
S
ERVICING
Warning
Depressurize system prior to removing cartridge pack.
14. Unit may now be left unattended until the display reads, “Sanitization Complete: Press
ENTER.”
15. After sanitization is complete and the user presses “ENTER” the unit will return to the (Idle) mode.
16. Press the START/STOP switch to return the unit to the normal recirculating mode.
System Depressurization
This should be done prior to removing a cartridge pack or an ultrafilter. The unit needs to be plugged in and operating to correctly perform this depressurization step.
1.
IMPORTANT: Disconnect inlet water from the left back of unit.
2.
From the (Idle) display, press ENTER.
3.
From “Air Purge?” press the UP arrow.
4.
From “Depressurize” press the UP arrow to select (YES).
5. Press ENTER. Display will show “System
Flushing” and the time remaining as the system depressurizes.
40
M
AINTENANCE AND
S
ERVICING
Warning
Depressurize system prior to removing cartridge pack.
Warning
A full cartridge pack may weigh about
20 lbs. (9 kg).
Cartridge Pack Replacement
The frequency with which you will need to clean your unit and replace your cartridge pack is dependent on your feed water’s characteristics, your purity requirements and your usage. Sanitize your NANOpure DIamond Life
Science (UV/UF) and replace the cartridge pack when the product water purity drops below acceptable levels of resistivity, when organic levels become too high, or if a new 0.2 micron filter clogs rapidly after installation even though the cartridge pack was thoroughly rinsed before the 0.2 micron filter was installed. The simple-to-use cleaning syringe is available from BarnsteadlThermolyne
(Catalog Number CMX25). This is used to effect a complete sanitization.
1.
Remove the feed water line by depressing the stainless steel thumb pad.
2.
Depressurize the system by opening the remote dispenser or, from an operational unit in the normal recirculating mode, push the dispense knob to dispense water until no more water flows.
3.
Shut power OFF to the unit and disconnect the electrical service to the unit.
4.
Open the front ( left side) door.
5.
Loosen the wing head screws on the cartridge manifold.
6.
Remove the exhausted cartridge pack by lifting the unit manifold and pulling the cartridge pack out.
7.
Install new cartridge pack, rinse and sanitize according to the instructions in the “Initial
Operation” section.
41
M
AINTENANCE AND
S
ERVICING
42
Caution
Do not overtighten the 0.2 micron filter or use excessive force in seating it.
The filter can be damaged by overtightening or excessive force.
Note
If a newly installed 0.2 micron filter clogs rapidly after installation, the
NANOpure DIamond Life Science
(UV/UF) may need to be sanitized to remove bacterial contaminants. See
“System Sanitization.”
0.2 Micron Filter Replacement
Replace the 0.2 micron final filter whenever any of the following conditions occur: every 30 days, the product water flow rate is reduced or bacteria break through. To replace the 0.2 micron filter assembly:
1.
Remove the old 0.2 micron filter assembly by turning it counterclockwise until it is free from the dispense arm manifold.
2.
Remove the new 0.2 micron filter assembly from the bag and insert it into the dispense arm manifold. Gently turn it clockwise until it is fully seated.
3.
Remove the protective cap from the filter bell.
4.
Rinse 1-2 liters of water through the filter to drain prior to using the product water.
0.2 Micron
Filter
Note
Depressurize the system before removing the ultrafilter.
Note
Some water will drain from the ultrafilter when it is disconnected. Make sure a sink, bucket or other water-proof container is available to place the ultrafilter in while it drains. Water will also drain from the tubing onto the bench or floor. To avoid slipping and falling, clean up all spills immediately.
Note
Do not install the ultrafilter until the
NANOpure DIamond Life Science
(UV/UF) is ready for use. If the unit sits for longer than 96 hours without use, it will become contaminated and may affect product water quality.
Ultrafilter Replacement
The length of your ultrafilter’s life will depend to some extent on conditions and use in your lab. When you are unable to completely sanitize your system (i.e. you can no longer get pyrogen-free water even after sanitizing the system), replace the ultrafilter as follows:
1.
Disconnect feed water and depressurize the system according to “System Depressurization under the “Maintenance and Servicing” section.
2.
Turn the NANOpure DIamond Life Science
(UV/UF) OFF and disconnect it from the power supply.
3.
Open the front (right side) door and right side door.
4.
Separate the velcro strap from the ultrafilter.
5. Simultaneously push and hold the collet in using the tool supplied with the ultrafilter while pulling the top fitting out of the connector. Refer to photo on next page.
6.
Unscrew tubing with luer fitting from the top side port of the ultrafilter.
M
AINTENANCE AND
S
ERVICING
7.
Pull the ultrafilter out of the unit and turn it to the left side. (This will make it easier to disconnect the bottom fitting.)
8.
Using the tool supplied with the ultrafilter, simultaneously push and hold the collet on the ultrafilter in while pulling the bottom fitting out of the connector. See photo.
9.
Install new ultrafilter according to “Ultrafilter
Installation” in the “Installation” section of this manual.
10. Sanitize the NANOpure DIamond Life Science
(UV/UF) according to the “System Sanitization” section of this manual.
Tool used for ultrafilter removal
Collet
Ultrafilter Installation
Tubing with luer fitting
Top connector
Collet
Velcro strap
Metal bracket
Collet
Bottom connector Ultrafilter Connections - Right side view
43
M
AINTENANCE AND
S
ERVICING
Warning
This unit is equipped with an ultraviolet lamp. Ultraviolet radiation is harmful to the eyes and skin. Do not attempt to observe the lamp directly.
Note
The NANOpure DIamond Life Science
(UV/UF) display will read “Check UV
Lamp” when the lamp is burned out or disconnected.
Note
After approximately 5 months of operation, the display will periodically show a reminder to change the lamp.
Note
The UV lamp contains mercury. If broken or no longer needed, do not dispose of the UV lamp in the trash.
Recycle or dispose of the UV lamp as hazardous waste.
UV Lamp Replacement
The ultraviolet lamp requires changing every 5-6 months.
Lamp Life will vary according to the number of times the
NANOpure DIamond Life Science (UV/UF) is turned on and off (the START/STOP mode). This estimated lamp life is based on the NANOpure DIamond being operated in the normal recirculating mode during regular working hours and then placed in the Standby mode during off hours. Every fourth time the pump goes on in the Standby mode, the UV light will go on. If the NANOpure DIamond
Life Science (UV/UF) is cycled between the normal and
Standby modes during the workday, this will result in a shorter lamp life. Therefore, it is recommended that the
NANOpure DIamond (UV/UF) be left in the normal recirculating mode during regular working hours.
To replace the UV lamp:
1.
Turn the NANOpure DIamond Life Science
(UV/UF) OFF and disconnect power cord from unit.
2.
Access the left side of the unit by unlatching the screw securing the door with a flathead screwdriver. Unlatch and open the door. The UV lamp will be installed vertically.
3.
Remove the black cap by sliding it off the UV chamber. Disconnect and remove old UV lamp and dispose of it in the proper manner (see
Note).
4.
Remove the UV lamp from its packaging. DO
NOT TOUCH THE GLASS PORTION OF THE
LAMP! It is recommended that lint-free gloves be worn when handling the lamp. The glass portion must be free of fingerprints, perspiration, etc. Even a light coating of perspiration will
UV Lamp Installation
44
M
AINTENANCE AND
S
ERVICING reduce the effectiveness of the lamp. If the glass portion of the lamp is touched, clean it with a damp, lint-free cloth: use isopropyl alcohol as required.
5.
Insert the lamp halfway into the UV lamp chamber and plug UV cable in, then fully insert the lamp into the chamber.
6.
Make sure the UV lamp cable is tucked behind the chamber before closing and relatching the door.
7.
Reconnect the power cord. Turn unit ON.
8.
Reset UV Timer according to “Resetting the UV
Timer.”
9.
Operate normally.
Warning
Replace fuses with those of the same type and rating.
Fuse Replacement
1.
Turn off the NANOpure DIamond Life Science
(UV/UF) and disconnect it from the power supply by removing the power cord directly below the main power switch on right side of unit.
Locate the fuse drawer above the power cord receptacle.
2.
Pull out the fuse drawer located in the power entry module.
3.
Remove old fuses and replace with fuses of the same type and rating. (See
“Replacement Parts.”)
4.
Replace fuse drawer.
5.
Reattach the power cord.
6.
Operate normally.
45
M
AINTENANCE AND
S
ERVICING
Shutdown
If the NANOpure DIamond Life Science (UV/UF) is to be shut down for an extended period of time, the unit should be completely drained and the cartridge pack removed to prevent the growth of bacteria. If the system has remained inactive and full of water for more than 96 hours, the unit should be drained, sanitized and new cartridge pack installed prior to use.
46
Troubleshooting
Problem Possible Causes Solutions
NANOpure DIamond completely No electrical power to NANOpure Ensure NANOpure DIamond power inactive (pump and display DIamond.
cord is connected to a live power not operating.) source and completely plugged into electrical outlet as well as power entry module on the unit. Make sure on/off switch is in the “I” (on) position.
Main fuse(s) blown.
Power supply fuse blown
Replace the main fuse(s) as indicated in “Fuse Replacement.”
Contact customer service concerning power supply fuse replacement.
Reduced or no product flow from the dispense arm or
0.2 micron filter clogged.
UF module clogged.
remote dispenser (if attached).
Feed water not connected.
Display reads, “Check Inlet” and pump does not run.
Pump protector (in reservoir), feed water line pressure switch
Replace final filter as indicated in “0.2
Micron Filter Replacement” section.
Reconnect feed water.
Connect the pump protector, jumper plug or pressure switch cord to the or jumper plug not connected to receptacle located on the upper left pump interlock.
hand section of the NANOpure
DIamond.
Display reads, “...M
Ω
-cm” for more than 10 seconds.
Recirculated water will not rinse up to desired purity level.
Short cartridge pack life.
Insufficient feed water pressure Open feed water line shut off, connect or volume to satisfy the NANOpure feed water line to NANOpure DIamond
DIamond.
and ensure sufficient pressure or sufficient water level in the feed water reservoir).
Air in cell.
If you just started your NANOpure
DIamond, wait 1 minute to see if it corrects itself.
Bad cell connection.
Exhausted cartridge pack.
Cartridge pack is beyond expiration date.
Check connection of the cell as it attaches to the circuit board.
Replace cartridge pack as indicated in
“Cartridge Pack Replacement.”
Check the expiration date. Cartridge packs begin to lose capacity after being stored for 2 years from the date of manufacture. Replace the cartridge pack with an unexpired one.
47
T
ROUBLESHOOTING
Display reads,
“Overpressure Error.”
Poor quality feed water.
0.2 micron filter clogged.
If a Barnstead ROpure is the feed water source, check that the membrane is functioning properly.
If a Barnstead Still is the feed water source, ensure that the distillate temperature to the NANOpure
DIamond does not exceed 40°C
(104°F).
The display periodically reads, UV lamp not connected properly.
Check to ensure lamp is connected
“Check UV Bulb.” properly.
UV lamp burned out.
Replace the UV lamp as described in
“UV Lamp Replacement.”
Purity display continuously
“OVERRANGE.”
If feed water is from a central water purification system, verify water quality and proper functioning of the system.
See “General Specifications.”
Air in system.
Bad cell connection.
Dispense more water through the
NANOpure DIamond to purge air from the unit.
Check the connection of the cell as it attaches to the meter circuit board mounted on the inside of the right door.
Incorrect N.I.S.T. calibration offset.
If you have a N.I.S.T. module, perform a N.I.S.T. calibration.
If you do not have a N.I.S.T. module, perform the calibration function
WITHOUT the calibration module (as if you had it; answer (YES) to prompt.
The display will read “Calibration
(failed)” but this will clear any incorrect offset to zero.
Replace final filter as indicated in “0.2
Micron Filter Replacement.”
48
T
ROUBLESHOOTING
Display reads, “Meter
Board Problem.”
PC board communication error.
System electronics failure.
Turn the NANOpure DIamond OFF
(I/O Switch) and restart the unit.
Replace the resistivity meter board mounted on the inside of the right door.
Tighten wing head screws on unit manifold.
Water is leaking between the Wing head screws are not tight enough.
manifold and the cartridge pack.
New cartridge pack: o-ring missing Disconnect inlet water and from post (each of the 4 posts depressurize system. Next, disconnect must contain an o-ring.) the manifold and verify/replace missing o-ring. Reinstall cartridge pack as indicated in “Cartridge Pack
Replacement.”
49
Replacement Parts
Recommended Spares
Consumables
Consumable parts are those required to support the day-to-day operation of this equipment.
Barnstead|Thermolyne establishes two types of consumables; those items that must periodically be replaced to maintain performance (filters, resin cartridges, etc.) and other items of limited life (indicator lights, fuses, etc.) that you can expect to replace on a more or less random basis. Where practical, Barnstead|Thermolyne recommends the frequency of replacement, or provides information on life expectancy from which you may calculate a replacement interval compatible with your usage pattern. The replacement of consumable parts is discussed in the Maintenance and Servicing section to assist you in accomplishing your own service.
Consumables may be ordered separately and in some cases, as an expendables kit. Check with your
Barnstead|Thermolyne representative for additional information on the expendables kit.
Description
DIamond Kit Organic Free R/O & Distilled Feed
DIamond Kit Organic Free Deionized Feed
Final Filter
Main Transformer Fuse, Type F Fast-Acting,
3.15 Amp, 250 Volt
Sanitization Solution
Ultrafilter
Catalog No.
D50280
D50281
D3750
Main Fuse, 120 Volt - Slow Blow, 1.6 Amp, 250V FZX47
Main Fuse, 230 Volt - Type T Time Lag, 0.63 Amp, 250V FZX54
Teflon® Tape, Roll
UV Lamp
06078
LMX13
AZ9027
CMX25
FL1192X1
Recommended
Quantity
1
1
2
1
2
2
2
1
2
1
General Maintenance Parts
General maintenance parts are defined as laboratory level repair parts which do not require great expertise or special tools for installation. Barnstead|Thermolyne recommends that you stock the general maintenance parts as an aid to ensuring the continued operation of this equipment.
Description
Quick Disconnect Inlet
Quick Disconnect Insert 3/8” tube
Cartridge Pack O-Rings
Catalog No.
CUX11
CUX9
GSX68
Recommended Quantity
1
1
4
® Teflon is a registered trademark of DuPont.
50
R
EPLACEMENT
P
ARTS
Safety Stock
For critical applications where performance with minimum downtime is required, Barnstead|Thermolyne recommends that you maintain a local stock of those parts listed in the General Maintenance Parts and Safety
Stock sections.
Description
Display
External Display cable (Remote Display Use)
Main PC Board
Recirculation Pump & Motor
Pressure Regulator
Resistivity Cell
Solid State Ballast UV
Quartz Sleeve, UV Lamp
Power Supply PC Board
Calibration PC Board
Resistivity Meter PC Board
Check Valve
Wire Harness, UV Bulb
Flow Sensor
Flush Solenoid
Unit Manifold
Dispenser Manifold
Dispense Switch
Catalog No.
SW1190X2
WHX20
PC1190X1
PU1190X1
E896X4
SC1191X1
TU733X1
TNX116
PCX48
PC896X2
PMX214
WH1191X1
ME1190X1
RY1190X3
SW1190X1
Recommended Quantity
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
Optional Accessory Ordering Information
Description
N.I.S.T. Calibration Module
Low-Pressure Switch
Float Switch
Remote Dispenser
Serial Cable for Communication with Computer
Printer (120-240 VAC system)
Printer Replacement Paper Roll
Catalog No.
E896X5
D2706
D8964
D11981
WHX18
AY1137X1
AY669X4
51
Flow Chart
power switch turned on
Barnstead /Thermolyne
Vx.xx Style English
Unit Initialization & Meter
Calibration Check
NANOpure DIamond (Idle) xx :xx :xx xx /xx /xx standby key pressed start_stop key pressed
Purity display back key pressed
(force print line) enter key pressed
(
°
C) dispense key pressed up or down keys pressed
Change Units?
Temperature
Compensation?
System flush?
Auto Dispense
Menu?
back key pressed
Language
Arrow Legend:
Keypress Options
Automatic Entry
Menu Wrap-around
52
Operational Overview up or down keys pressed
Sanitize Unit?
NIST calibrate?
Reset Timers?
Advanced menu?
Optional
Printer
Date Format
Set Date
Set Time
Set Cell
Constant
Set Display
Brightness
Adjust Set
Point
Diagnostics menu?
enter key pressed
Air Purge?
Wiring Diagram
L1 L2
CHASSIS
GND
GND
NEUTRAL
LINE
SO3
C1
POWER ENTRY MODULE SWITCH
& FUSES:
120V: (2) FSX47
230V: (2) FSX54
SO2 SO1
WM1
CHASSIS GND
WH3
J1
J2
WH2
A
S1
PS1
B
WH7
C
FS1
WH1
S2
REMOTE DISPLAY
CONNECTOR
BLU
BLU
(SPG)
J11
J12
J13
J14
J6
J10
J9
J4
J5
J18
J8
J23
J7
J13
J3
PC1
C
WH8
L1
MODEL #s D11911
& D11931 ONLY
PC4
WH6
A
CP1
WHT
WHT
BLK
RED
PU1
WH5
PC3
REF.
NO.
DESCRIPTION
D11901
PC2
PC3
PC4
PU1
PS1
S1
S2
SO1
SO2
SO3
C1
CP1
E1
FS1
L1
PC1
POWER ENTRY MODULE
PUMP CONNECTOR
CONDUCTIVITY CELL
FLOW SENSOR
UV LAMP
MAIN CONTROL BOARD
CALIBRATION BOARD
METER BOARD
BALLAST
PUMP
POWER SUPPLY
DISPENSE SWITCH
MEMBRANE SWITCH
DIVERTER SOLENOID
DRAWOFF SOLENOID
FLUSH SOLENOID
WH1 PRINTER HARNESS
WH2 DC POWER HARNESS
WH3 AC POWER HARNESS
WH4 CAL. BOARD HARNESS
WH5 PRESSURE SW. HARNESS
WH6 METER BOARD HARNESS
WH7 DISPLAY HARNESS
WH8 LAMP HARNESS
WH9 CAL./METER HARNESS
WM1 GROUND WIRE
DIAGRAM COMPONENT LIST
PU1190X1
TNX116
SW1190X1
SW1190X2
RY1190X2
RY1190X1
RY1190X3
WH1190X1
WH1190X2
WH1190X3
WH1190X4
WH1190X5
WH1190X6
WHX19
MODEL NO. AND OUR PART NO.(s)
D11901
CEX183
CE896X2
E896X4
ME1190X1
PC1190X1
PC1190X2
PC896X2
WM896X2
WM1190X2
D11911
CEX183
CE896X2
E896X4
ME1190X1
LMX13
PC1190X1
PC1190X2
PC896X2
SC1191X1
PU1190X1
TNX116
SW1190X1
SW1190X2
RY1190X2
RY1190X1
RY1190X3
WH1190X1
WH1190X2
WH1190X3
WH1190X4
WH1190X5
WH1190X6
WHX19
WH1191X1
WM896X2
WM1190X2
D11921
CEX183
CE896X2
E896X4
ME1190X1
PC1190X1
PC1190X2
PC896X2
PU1190X1
TNX116
SW1190X1
SW1190X2
RY1190X2
RY1190X1
RY1190X3
WH1190X1
WH1190X2
WH1190X3
WH1190X4
WH1190X5
WH1190X6
WHX19
WM896X2
WM1190X2
E1
WH9
J2
D11931
CEX183
CE896X2
E896X4
ME1190X1
LMX13
PC1190X1
PC1190X2
PC896X2
SC1191X1
PU1190X1
TNX116
SW1190X1
SW1190X2
RY1190X2
RY1190X1
RY1190X3
WH1190X1
WH1190X2
WH1190X3
WH1190X4
WH1190X5
WH1190X6
WHX19
WH1191X1
WM896X2
WM1190X2
J3
PC2
J1
BLK
RED B WH4
C
53
54
Ordering Procedures
Please refer to the Specification Plate located inside the left side door for the complete model number, serial number, and series number when requesting service, replacement parts or in any correspondence concerning this unit.
All parts listed herein may be ordered from the
Barnstead|Thermolyne dealer from whom you purchased this unit or can be obtained promptly from the factory.
When service or replacement parts are needed we ask that you check first with your dealer. If the dealer cannot handle your request, then contact our Customer Service
Department at 319-556-2241 or 800-553-0039.
Prior to returning any materials to Barnstead|Thermolyne
Corp., please contact our Customer Service Department for a “Return Goods Authorization” number (RGA).
Material Returned without an RGA number will be returned.
One Year Limited Warranty
Barnstead|Thermolyne Corporation warrants that if a product manufactured by Barnstead|Thermolyne and sold by it within the continental United States or Canada proves to be defective in material or construction, it will provide you, without charge, for a period of ninety (90) days, the labor, and a period of one (1) year, the parts, necessary to remedy any such defect. Outside the continental United States and Canada, the warranty provides, for one (1) year, the parts necessary to remedy any such defect. The warranty period shall commence either six (6) months following the date the product is sold by Barnstead|Thermolyne or on the date it is purchased by the original retail consumer, whichever date occurs first.
All warranty inspections and repairs must be performed by and parts obtained from an authorized
Barnstead|Thermolyne dealer or Barnstead|Thermolyne (at its own discretion). Heating elements, however, because of their susceptibility to overheating and contamination, must be returned to our factory, and if, upon inspection, it is concluded that failure is not due to excessive high temperature or contamination, warranty replacement will be provided by Barnstead|Thermolyne. The name of the authorized Barnstead|Thermolyne dealer nearest you may be obtained by calling 1-800-446-6060 (319-556-2241) or writing to:
Barnstead|Thermolyne
P.O. Box 797
2555 Kerper Boulevard
Dubuque, IA 52004-0797
USA
FAX: (319) 589-0516
E-MAIL ADDRESS: [email protected]
Barnstead|Thermolyne’s sole obligation with respect to its product shall be to repair or (at its own discretion) replace the product. Under no circumstances shall it be liable for incidental or consequential damage.
THE WARRANTY STATED HEREIN IS THE SOLE WARRANTY APPLICABLE TO Barnstead|Thermolyne
PRODUCTS. Barnstead|Thermolyne EXPRESSLY DISCLAIMS ANY AND ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR FITNESS FOR USE.
55
Barnstead|Thermolyne
2555 Kerper Blvd.
P.O. Box 797
Dubuque, IA 52004-2241
Phone: 319-556-2241 800-553-0039
Fax: 319-589-0516
E-Mail Address: [email protected]
ISO
9001
REGISTERED
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 3 Safety Information
- 3 Alert Signals
- 3 Warnings
- 5 Introduction
- 5 General Usage
- 6 General Specifications
- 7 Environmental Conditions
- 7 Declaration of Conformity
- 8 Unpacking
- 9 Installation
- 9 Choosing a Site
- 10 Ultrafilter Installation
- 11 UV Lamp Installation
- 11 Other Accessories
- 12 Bench Mounting
- 12 Wall Mounting
- 13 Installing the Control Panel in a Remote Location
- 14 Water Connections
- 15 Controls
- 15 Main Power Switch
- 15 Control Panel
- 16 Switches
- 17 Initial Operation
- 17 Cartridge Pack Installation
- 18 Initial Rinse
- 19 System Sanitization Procedure
- 21 Normal Operation
- 21 Dispensing Water
- 21 Automatic Dispensing
- 22 User Settings
- 22 Adjusting Display Brightness
- 22 Setting the Displayed Language
- 22 Setting the Date and Time
- 23 Setting the Cell Constant
- 25 Use of Standby Mode
- 25 Selecting the Set Point
- 26 Resetting the Sanitization Timer
- 26 Resetting the UV Timer
- 27 Setting Volumetric Dispensing
- 28 Setting Timed Dispensing
- 28 Performing a System Flush
- 28 Changing Purity Units
- 29 Temperature Compensation
- 30 Installing Float or Pressure Switch
- 31 N.I.S.T. Calibration Module
- 31 Performing an Electronic Calibration Using the Optional N.I.S.T. Calibration Module
- 32 Remote Dispenser
- 33 Computer/Printer Setup
- 35 Connecting and Starting the Printer
- 36 Printer Setup
- 38 Maintenance and Servicing
- 38 General Cleaning Instructions
- 38 Cell Cleaning
- 38 System Sanitization
- 40 System Depressurization
- 40 Cartridge Pack Replacement
- 41 0.2 Micron Filter Replacement
- 42 Ultrafilter Replacement
- 42 UV Lamp Replacement
- 44 Fuse Replacement
- 45 Shutdown
- 46 Troubleshooting
- 47 Replacement Parts
- 52 Flow Chart Operational Overview)
- 53 Wiring Diagram
- 54 Ordering Procedures
- 55 One Year Limited Warranty