SIGLENT SDS1000D, SDS1022D, SDS1042D, SDS1062D, SDS1102D, SDS1152D, SDS1202D digital storage oscilloscope Service manual
Below you will find brief information for SDS1000D SDS1000D, SDS1000D SDS1022D, SDS1000D SDS1042D, SDS1000D SDS1062D, SDS1000D SDS1102D, SDS1000D SDS1152D, SDS1000D SDS1202D. The SDS1000D series digital storage oscilloscopes are powerful tools for engineers and scientists. These oscilloscopes offer a wide range of features, including high bandwidth, high sampling rate, and a variety of measurement and analysis functions. They are ideal for a wide variety of applications, including circuit design, debuggin, and education.
Advertisement
Advertisement
Service Manual
SDS1000D Digital Storage Oscilloscope
June 2011
SIGLENT Technologies Co., Ltd.
Service Manual
Declaration
Copyright © SIGLENT Technologies Co.,Ltd. All rights reserved.
Contents in this Manual are not allowed to copy, extract and translate before being allowed by SIGLENT Company.
Service Manual
General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
1. To Avoid Fire or Personal Injury
● ● ● ●
Use Proper Power Cord. Use only the power cord specified for this product and certified for the country of use.
● ● ● ●
Connect and Disconnect Properly. Do not connect or disconnect probes or test leads while they are connected to a voltage source.
● ● ● ●
Ground the Product. This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded.
●
● ●
●
Connect the Probe Properly. The probe ground lead is at ground potential. Do not connect the ground lead to an elevated voltage.
●
● ●
●
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings and marking on the product. Consult the product manual for further ratings information before making connections to the product.
●
● ●
●
Do Not Operate Without Covers. Do not operate this product without covers or panels removed.
●
● ●
●
Use Proper Fuse. Use only the fuse type and rating specified for this product.
Service Manual
● ● ● ●
Avoid Exposed Circuitry. Do not touch exposed connections and components when power is present.
●
● ●
●
Do Not Operate With Suspected Failures. If you suspect there is damage to this product, have it inspected by qualified service personnel.
● ● ● ●
Do Not Operate in Wet/Damp Conditions.
● ● ● ●
Do Not Operate in an Explosive Atmosphere.
●
● ●
●
Keep Product Surfaces Clean and Dry.
Power Line Conducted Emission Limits (Class B)
Measuring standard: EN61326:1998+A1, 2002+A2, 2003
2. Safety Terms and Symbols
Terms used on the Product. These terms may appear on the product:
DANGER: Indicates an injury hazard immediately accessible as you read the marking.
WARNING: Indicates an injury hazard not immediately accessible as you read the marking.
CAUTION: Indicates a hazard to property including the product.
Symbols used on the Product. These symbols may appear on the product:
Service Manual
Document Summarize
Chapter 1 Specifications
List the specifications and general specifications of SDS1000D series digital oscilloscope.
Chapter 2 Prepare for Use
Introduce the preparatory work should be done before using the oscilloscope.
Chapter 3 Performance Test
Introduce how to execute the performance test to understand current performance status of the oscilloscope.
Chapter 4 Calibration
Introduce how to calibrate the oscilloscope.
Chapter 5 Disassembly and Assembly
Introduce how to disassemble and assemble the oscilloscope to understand its structure.
Chapter 6 Trouble
List the troubles may appear during measuring and the solutions.
Chapter 7 Replaceable Parts
List the replaceable parts for user’s repair or exchange.
Service Manual
Catalogue
Document Summarize.......................................................................................................................6
Chapter 1 Specifications ...................................................................................................................8
Specifications ............................................................................................................................8
Generic Specifications ............................................................................................................12
Chapter 2 Prepare for Use...............................................................................................................13
General Inspecting ..................................................................................................................14
Power-On Inspection...............................................................................................................15
Probe Compensation ...............................................................................................................16
Auto Setup ..............................................................................................................................17
Chapter 3 Performance Test ............................................................................................................19
Interface Test...........................................................................................................................20
USB Host Test................................................................................................................20
USB Device Test .............................................................................................................21
RS-232 Test.....................................................................................................................22
Pass/Fail Test...................................................................................................................24
Specification Test ....................................................................................................................25
Test DC Gain Accuracy...................................................................................................25
Test Bandwidth................................................................................................................26
Test Trigger Sensitivity ...................................................................................................27
Test Noise........................................................................................................................27
Chapter 4 Calibration ......................................................................................................................28
Self Calibration .......................................................................................................................28
Chapter 5 Disassembly and Assembly ............................................................................................30
Notice......................................................................................................................................31
Disassemble and Assemble On-off and Knobs .......................................................................32
Disassemble and Assemble Back Rind ...................................................................................33
Disassemble and Assemble Back Cover Board.......................................................................34
Disassemble and Assemble Top Cover Board.........................................................................35
Disassemble and Assemble Fan ..............................................................................................36
Disassemble and Assemble power board ................................................................................37
Disassemble and Assemble Metal Shelf .................................................................................38
Disassemble and Assemble Main board..................................................................................39
Disassemble and Assemble LCD ............................................................................................40
Disassemble and Assemble Keyboard ....................................................................................41
Chapter 6 Prompting Messages and Troubleshooting.....................................................................42
Prompting messages:...............................................................................................................42
Troubleshooting ......................................................................................................................44
Chapter 7 Updating Parts ................................................................................................................45
Service Manual
Chapter 1 Specifications
All specifications apply to 10X probe and SDS1000D Digital Storage Oscilloscopes. To verify that the oscilloscope meets specifications, the oscilloscope must first meet the following conditions:
●
The oscilloscope must have been operating continuously for thirty minutes within the specified operating temperature.
●
You must perform the Do Self Cal operation, accessible through the Utility menu, if the operating temperature changes by more than 5° C.
●
The oscilloscope must be within the factory calibration interval ① ①
① ① .
All specifications are guaranteed unless noted “typical”.
Specifications
Inputs
Input Coupling
Input Impedance
AC, DC, GND
1MΩ±2% || 16Pf±3Pf
Maximum input voltage 400V
(
DC+AC PK-PK
,
1MΩ input impedance
),
CAT
Probe attenuator
Probe attenuator Factors
Set
I,CAT II
1X
、
10X
1X、5X、10X、50X、100X、500X、1000X
Vertical System
Vertical Sensitivity 2mV/div - 10V/div(1-2-5 order)
Channel voltage
2mV-200mV: ±800mV offset range
206mV-10V: ±40V
Vertical Resolution 8 bit
Channels 2
Analog
Bandwidth
200MHz(SDS1202D)
150MHz(SDS1152D)
Service Manual
Single-shot
Bandwidth
100MHz(SDS1102D)
60MHz(SDS1062D)
40MHz(SDS1042D)
25MHz(SDS1022D)
200MHz(SDS1202D)
150MHz(SDS1152D)
100MHz(SDS1102D)
60MHz(SDS1062D)
40MHz(SDS1042D)
25MHz(SDS1022D)
Lower frequency limit (AC -3dB)
≤ 10Hz(at input BNC)
DC Gain Accuracy
<±
3.0%: 5mv/div to 10V/div in Fixed Gain Ranges
<±
4.0%:typical for 2mv/div and Variable Gain Ranges
±
[3%X
( reading+offset
)
1% of offset +0.2div+2mv] DC Measurement
Accuracy:
All Gain settings
≤
100mv/div
DC Measurement ±[3%X
( reading+offset
)
1% of offset +0.2div+100mv]
Accuracy:
All Gain settings
>
100mv/div
Rise time
<1.7ns (SDS1202D)
<2.3ns (SDS1152D)
<3.5ns (SDS1102D)
<5.8ns (SDS1062D)
<8.8ns (SDS1042D)
<14ns (SDS1022D)
Vertical coupling
Math operation
FFT input
Bandwidth limited
AC, DC, GND
+, -, *, /, FFT
Window mode: Hanning, Hamming, Blackman, Rectangular
Sampling points: 1024
20MHz (-3dB)(SDS1022D doesn’t have this function )
Horizontal System
Real
Sampling
Time Single Channel:500MSa/s,Double Channel: 250MSa/s
Rate
Equivalent
Sampling Rate
50GSa/s
Measure Display MAIN, WINDOW, WINDOW ZOOM, ROLL, X-Y
Service Manual
Modes
Time base
Accuracy
Horizontal Scan
Range
±
50ppm measured over 1ms interval
2.5nS/DIV - 50S/DIV
Roll: 100mS/DIV
~
50S/DIV (1-2.5-5 order)
Measure System
Auto
Measure
(32 Types)
Vpp, Vmax, Vmin, Vamp, Vtop, Vbase, Vavg, Mean,Crms, Vrms,
ROVShoot, FOVShoot, RPREShoot, FPREShoot, Rise time, Fall time,
Freq, Period,+ Wid,
-
Wid,
+
Dut,
FFR, FFF, LRR, LRF, LFR, LFF
-
Dut, BWid, Phase, FRR, FRF,
Cursor
Measure
Manual mode, Track mode and Auto mode
Trigger System
Trigger Types Edge, Pulse Width, Video, Slope, Alternative
Trigger Source CH1
、
CH2
、
EXT
、
EXT/5
、
AC Line
Trigger Modes Auto, Normal, Single
Trigger
Coupling
Trigger Level
Range
AC, DC, LF rej, HF rej
CH1,CH2: ±6divisions from center of screen
EXT: ±1.2V
EXT/5: ±6V
Trigger
Displacement
Pre-trigger
:(
Memory depth/
(
2*sampling
))
,
Delay Trigger: 260DIV
Holdoff range 100ns – 1.5s
Edge Trigger Edge type: Rising, Falling, Rising and Falling
Pulse Width
Trigger
Video Trigger
Trigger Modes: (
=
>
,
<
)Negative Pulse Width
,
=
)Positive Pulse Width, (
Pulse Width Range: 20ns – 10s
Support signal Formats: PAL/SECAM, NTSC
Trigger condition : odd field, even field, all lines, line Num
>
,
<
,
Slope Trigger
Alternative
Trigger
(
>
,
<
,
=
) Positive slope, (
>
,
<
,
=
) Negative slope
Time: 20ns-10s
CH1 trigger type: Edge, Pulse, Video, Slope
CH2 trigger type: Edge, Pulse, Video, Slope
X-Y Mode
X-pole Input /
Y-Pole Input
Phase Error
Sample
Frequency
Channel 1 (CH1) / Channel 2 (CH2)
±3 degrees
XY mode has a breakthrough that trade oscilloscopes restrict sampling rate at 1MSa/s and supports 5KSa/s
~
200MSa/s ,(1-2.5-5 order)
Service Manual
Hard Ware Frequency Counter
Reading resolution 6 Bytes
Range
Signal Types
DC Couple, 10Hz to MAX Bandwidth
Satisfying all Trigger signals(Except Pulse width trigger and
Video Trigger)
Control Panel Function
Auto Set Auto adjusting the Vertical, Horizontal system and Trigger
Position
Save/Recall Support 2 Group referenced Waveforms,20 Group setups,20
Group captured Waveforms internal Storage/Recall function and USB flash driver storage function.
Service Manual
Generic Specifications
Display System
Display Mode
Resolution
Display Color
Display
(
Typical state
)
Contrast
Backlight
(
Typical state
)
Intensity
Wave display range
Wave Display Mode
Persist
Menu Display
Screen-Saver
Skin waveform interpolation
Color model
Language
Color TFT 5.7in.(145mm)diagonal Liquid Crystal Display
320 horizontal by 234 vertical pixels
64K color
150:1
300nit
8 x 12 div
Point, Vector
Off, 1 sec, 2 sec, 5 sec, Infinite
2 sec, 5 sec, 10 sec, 20 sec, Infinite
Off,1min,2min,5min,10min,15min,30min,1hour,2hour,5hour
Classical, Modern, Tradition, Succinct
Sin(x)/x
Normal , Invert
Simplified Chinese, Traditional Chinese, English, Arabic, French,
German, Russian, Spanish, Portuguese
Japanese, Korean, Italian
Environments
Temperature
Cooling
Humidity
Height
Operating:10
℃
to + 40
℃
Not operating: -20
℃
to + 60
℃
The fan forces it cold.
Operating: 85%RH, 40
℃
, 24 hours
Not operating: 85%RH, 65
℃
, 24 hours
Operating: 3000m
Not operating: 15,266m
Power Supply
Input Voltage
Frequency Scope
Power
Mechanical
Dimension
100-240 VAC, CAT II, Auto selection
45Hz to 440Hz
50VA Max
Weight length
Width
Height
2.3 kg
305mm
133mm
154mm
Service Manual
Chapter 2 Prepare for Use
This chapter includes the following contents:
●
General Inspecting
●
Power-On Inspecting
●
Connect the Probe
●
Probe Compensation
●
Auto Setting
Service Manual
General Inspecting
After receiving a new SDS1000D series oscilloscope, please inspect the instrument as follows:
1
. . . .
Inspect the shipping container for damage.
Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has been checked mechanically and electrically.
2
. . . .
Inspect the whole instrument.
In case there is any mechanical damage or defect, or the instrument does not operate properly or fails performance tests, notify the SIGLENT Sales Representative.
If the shipping container is damaged, or the cushioning materials show signs of stress, notify the carrier as well as the SIGLENT Sales Department. Keep the shipping materials for carrier’s inspection.
3
.
. .
.
Check the accessories.
Accessories supplied with the instrument are listed below. If the contents are incomplete or damaged, notify the SIGLENT Sales Representative.
Standard Accessories:
● Two pieces 1:1,(10:1) Passive Probes
● An User Manual
● A Certification
● A Guaranty Card
● An CD(including EasyScope3.0 computer software system)
●
A Power Cord that fits the standard of destination country
●
An USB Cable
Service Manual
Power-On Inspection
Normal operating voltage for DS1000CA series digital oscilloscope is the range of 100-240VRMS with the frequency from 45Hz to 440Hz. Connect one terminal of the power cord to the socket in left side of the oscilloscope and the other to the power source.
Picture 2-1 Connect power line
Press the button on top of oscilloscope, some keys on the front panel will be lighted for about 6 seconds until the normal display appears. And then you can operate the oscilloscope.
After power-on, the oscilloscope performs the self test automatically. After the test, press the
“DEFAULT/SETUP” button, then recall the Factory settings.
Service Manual
Picture 2-2 Power-On Inspecting
Probe Compensation
You are suggested to compensate probe before first using so as to match it to the input channel.
Measurement error may generate without the compensation or because of deviation compensation. To compensate the probe, follow the steps below.
1
.
Set the probe option attenuation in the channel menu to 10X
. Set the switch to 10X on the probe.
Picture 2-3 Set the attenuation switch and probe scale
2
.
Connect the probe to channel 1 on the oscilloscope, and attach the probe tip to the PROBE
COMP
Picture 2-4 Connect PROBE COMP
3
. Press AUTO. Within a few seconds, a square wave will be displayed (1 kHz, approximately 3 V pk-pk). Check if the compensation is correct according to the shape of waveform.
Under compensated Compensated correctly Over compensated
Picture 2-5 Check the shape
4
.
If necessary, adjust your probe. Repeat as necessary.
Service Manual
Auto Setup
SDS1000D Series Digital Storage Oscilloscopes have auto setup function, which can identify the type of waveform and adjust controls to produce a usable display of the input signal.
AUTO is the auto setup button.
Table2-1 Auto setup function menu:
Option Introduction
(
Multi-cycle sine
)
Auto set the screen and display several cycle signal.
Set the screen and auto display single cycle signal.
(
Single-cycle sine
)
Auto set and show the rising time.
(
Rising edge
)
Auto set and show the falling time.
(
Falling edge
)
(
Undo
Setup
)
Causes the oscilloscope to recall the previous setup.
There are four measure parameter will be displayed in the screen. Auto set determines the trigger source based on the following conditions:
●
If multiple channels have signals, channel with the lowest frequency signal.
●
The input frequency must be higher than 10Hz, and the amplitude must be higher than
8MV
Table 2-2 Auto setup function item
Function
Function Function
Function Setting
Setting Setting
Setting
Acquire Mode
Display Format
Display Type
Adjusted to Sampling
Y-T
Set to Dots for a video signal, set to Vectors for an FFT spectrum; otherwise, unchanged
Vertical Coupling Adjusted to DC or AC according to the input signal
Service Manual
Bandwidth Limit Off(full)
V/div
VOLTS/DIV adjustability
Adjusted
Coarse
Signal inverted Off
Horizontal position Center
S/div
Trigger type
Trigger source
Adjusted
Edge
Auto detect the channel which has the input signal
Rising
Auto
Trigger slope
Trigger mode
Service Manual
Chapter 3 Performance Test
This chapter includes the following contents:
■
Interface Test
USB Host
USB Device
RS-232
Pass/Fail
■
Specification Test
Service Manual
Interface Test
USB Host Test
Purpose: Test if the USB Host interface works normally through U disc.
Tools
:
●
An Unit SDS1000D digital oscilloscope
●
An U disc
Steps
:
1
.
Insert a U disc into the interface of USB Host on the front panel of the oscilloscope.
Picture 3-1 Connect the U disc
2
.
Wait until a prompt message “USB Flash Drive Plug In!” appeared on the screen.
图
3-2 USB host connect
Successfully
3
.
Press Save/Recall Save to file, to name renewably or delete the information
in
U disc.
4
.
Remove the U disc and a prompt message “USB Flash Drive Plug In!” will appeared.
Service Manual
USB Device Test
Purpose: Test if the USB Device interface works normally through EasyScope for SDS1000D series oscilloscope.
Tools
:
●
An Unit SDS1000D digital oscilloscope
●
A computer with USB interface
●
A standard USB cable(Type AB)
●
EasyScope software for SDS1000D series
Steps
:
1
.
Install the EasyScope to computer.
2
.
Connect the oscilloscope with PC using an USB cable and install the driver program step by step following the instruction.
Picture 3-3 connecting the USB Device interface
3
.
Run EasyScope ,then click “Connect”, if successful, the two point lamps at the top right corner of the software will be changed from red to green.; if failed, a prompt message will be appeared as below.
Service Manual
Picture 3-4 USB Device connect failed
RS-232 Test
Purpose: Test if the RS-232 interface works normally through EasyScope for SDS1000D series oscilloscope.
Tools
:
●
An Unit SDS1000D digital oscilloscope
●
A computer with RS-232 interface
●
A standard RS-232 cable
●
EasyScope software for SDS1000D series.
Steps
:
1
.
Install the EasyScope to computer.
2
.
Connect the oscilloscope with PC using an USB cable
Picture 3-5 Connect the RS232 interface
3
.
Run EasyScope
, click “Setting” “Connect Setting” , set the IO as below.
Service Manual
Picture 3-6 EasyScope connect setting page3/4 IO Setting, select the Baud as 38400 4
.
Press “UTILITY”
5
.
Click “Connect”, if successful, the two point lamps at the top right corner of the software will be changed from red to green.; if failed, a prompt message will be appeared as below.
Picture 3-7 RS-232 connect failed
Service Manual
Pass/Fail Test
Purpose: Test if the Pass/Fail interface works normally through the signal from the back BNC of the oscilloscope.
Tools
:
●
An Unit SDS1000D digital oscilloscope , another digital oscilloscope
●
A BNC cable
●
A Probe
Steps
:
1. Power-On two oscilloscopes.
2. Connect the Pass/Fail interface of SDS1000D with the channel BNC of another oscilloscope using the BNC cable.
3. Connect the probe to channel 1 on SDS1000D, and press AUTO button.
4. Press “UTILITY” button page3/4 Pass/Fail, set the function item as below:
Table 3-1 Pass/Fail function item
Option Setting
Enable Test
Source
Msg Display
Output
On
CH1
On
Pass
Stop On Output Off
5. Click “Mask Setting” Create Mask, and back to Page1/2, then press “Operate” button.
6. Adjust the time base and volt of another scope, if there is a pulse wave as below appeared on the screen, which show the interface works normally, if no response or another wave on the screen, errors might happen to the interface.
Picture 3-8 Pass/Fail interface works normally
Service Manual
Specification Test
Purpose: Test if some specifications are in the specified range.
Table 3-2 require instruments
Minimum requirements
-6 mV to 30 V,
±0.1% accuracy
Description
DC Voltage Source
Standard sine wave generator
BNC 50
Ω
Feed-Through
Termination
10Hz
and 300MHz
,
±1% amplitude accuracy
DC to 1GHz
Examples
FLUKE 9500B
PL-50
Before beginning these procedures, two conditions must first be met:
●
The oscilloscope must have been operating continuously for thirty minutes within the operating temperature range specified.
●
You must do the Self Calibration procedure. If the ambient temperature changes by more than
5℃
, you must do the Self Calibration procedure again.
Picture 3-9 Test instrument connecting
Test DC Gain Accuracy
Steps
:
1
、
Set the DC voltage source output level to 30 V
2
、
Press the button “DEFAULT/SETUP” of oscilloscope
3
、
Set VOLT/DIV to 10V/div, set measure parameter as “CH1,Mean”
4
、
Connect the oscilloscope channel 1 to the DC voltage source, and record the value of
Mean appeared on the screen as V pos
.
5
、
Change the polarity of the DC voltage source
, and record the value of Mean appeared on the screen as V neg
.
6
、
Calculate V diff
=V pos
– V neg
, then compare V diff to accuracy limits in the table.
Service Manual
Volt/div setting
10v/div
5v/div
2v/div
DC voltage output level
+30V, -30V
+15V, -15V
+6V, -6V
+3V, -3V
+1.5V, -1.5V
+600mV, -600mV
+300mV, -300mV
+150mV, -150mV
+60mV, -60mV
+30mV, -30mV
Accuracy limits for V diff
58.2 V
~
61.8 V
29.1 V
~
30.9 V
11.64 V
~
12.36 V
5.82 V
~
6.18 V 1v/div
500mv/div
200mv/div
100mv/div
50mv/div
20mv/div
10mv/div
2.91 V
~
3.09 V
1.164 V
~
1.236 V
582 mV
291 mV
~
~
618 mV
309 mV
116.4 mV
~
123.6 mV
58.2 mV
~
61.8 mV
5mv/div
2mv/div
+15mV, -15mV
+6mV, -6mV
29.1 mV
~
30.9 mV
11.52 mV
~
12.48 mV
7
、
Change the VOLT/DIV to the next Volt/div in the table, and set the DC voltage source output level to the corresponding level in the table as above. Repeat the steps 4 through 6 until all VOLT/DIV of both channels have been test.
Test Bandwidth
Steps
: : : :
1
、
Press the button “DEFAULT/SETUP”, and set the VOLT/DIV to 500 mV/div, set
SEC/DIV to 50 mS/div.
2
、
Set the measure parameter as “CH1,Vpp”.
3
、
Connect the oscilloscope channel 1 to the standard sinewave generator
4
、
Set the standard sinewave generator frequency to 10Hz
5
、
Set the generator output level so the Vpp measurement is between 2.98 V and 3.02 V.
6、Set the generator frequency to:
25MHz if you are testing a SDS1022D
40MHz if you are testing a SDS1042D
60MHz if you are testing a SDS1062D
100MHz if you are testing a SDS1102D
150MHz if you are testing a SDS1152D
200MHz if you are testing a SDS1202D
7
、
Set the oscilloscope timebase to the minimum SEC/DIV
8
、
Check the Vpp measurement is
≥
2.12 V
9
、
Test channel 2 with the same method
Service Manual
Test Trigger Sensitivity
Steps
: : : :
1
、
Press the button “DEFAULT/SETUP”, and set the VOLT/DIV to 500mv/div, set
SEC/DIV to 25ns/div.
2
、
Connect the oscilloscope channel 1 to the standard sinewave generator
3
、
Set the standard sinewave generator frequency to 10 MHz
3
、
Set the generator output level so the Vpp measurement is about 500 mV.
4
、
Press the button “set to 50%”. Adjust LEVEL as necessary and then check that triggering is stable.
5
、Set the generator frequency to:
25MHz if you are testing a SDS1022D
40MHz if you are testing a SDS1042D
60MHz if you are testing a SDS1062D
100MHz if you are testing a SDS1102D
150MHz if you are testing a SDS1152D
200MHz if you are testing a SDS1202D
6
、
Set the generator output level so the Vpp measurement is about 750 mV.
7
、
Press the button “set to 50%”. Adjust LEVEL as necessary and then check that triggering is stable.
8
、
Test channel 2 with the same method
Test Noise
Steps
:
1
、
Set the both channels VOLT/DIV to 2mV/div
,
SEC/DIV to 2.5µS/div
2
、
Set two measure parameters as “CH1,Vpp” and “CH2, Vpp”
3
、
Check the Vpp measurement is
≤
1.2mV
( other VOLT/DIV is 0.6 DIV
)
4
、
Change the VOLT/DIV, and check the Vpp until all the VOLT/DIV of both channel have been tested.
Service Manual
Chapter 4 Calibration
Self Calibration
Self Calibration procedure can optimize the signal path at the most measurement precision.
You can run this procedure at any time. If the operating temperature changes by more than 5
℃ or the units run less than thirty minutes, you should do the self calibration.
Steps
:
1
.
Cut off all the probes and lead.
2
.
Press the button “UTILITY” to choose the Do Self Cal to show the self calibration menu, and execute self calibration program according to the prompting on the screen .
Picture 4-1 Calibration beginning interface
3
.
Press “SINGLE” to do self calibration.
Service Manual
Picture 4-2 Calibrating
4
.
After the calibration finished, Press “SINGLE” key to exit.
Picture 4-3 Calibration succeed interface
Generally speaking, the self calibration will take about 3-5 minutes, if the calibration does not
Pass after this time, or the progress bar stop at one of the calibration item, there may be a trouble in the instrument.
Service Manual
Chapter 5 Disassembly and Assembly
This chapter includes the following topics:
●
Notice
●
Disassemble and Assemble the on-off and knobs
●
Disassemble and Assemble back rind
●
Disassemble and Assemble back cover board
●
Disassemble and Assemble top cover board
●
Disassemble and Assemble fan
●
Disassemble and Assemble power board
●
Disassemble and Assemble metal shelf
●
Disassemble and Assemble main board
●
Disassemble and Assemble LCD
●
Disassemble and Assemble keyboard
Service Manual
Notice
Notices
: : : :
●
Don’t disassemble the products except the work need.
●
Disassemble only can be done by qualified person.
●
Cut the power before disassembling.
●
Take static glove when disassembling.
●
Use proper tools and follow the disassemble sequence.
●
Prevent metal parts from transfiguration and avoid being scratched when disassembling.
Disassemble tools:
●
screwdriver
●
forceps
●
wrench
Service Manual
Disassemble and Assemble On-off and Knobs
Picture 5-1 Disassemble and Assemble the on-off and knobs
Disassemble steps
:
1.
Nip on-off 1 button by forceps, and then pull up.
○
2.
Cock knobs
③、④
gently by forceps
Assemble steps
:
Please operate as reverse orders. The same is as below.
Service Manual
Disassemble and Assemble Back Rind
Picture 5-2 Disassemble and Assemble back rind 1
图
5-3 Disassemble and Assemble back rind 2
Disassemble steps
:
1.
Back out the two bolts
⑤
and two bolts
⑥
.
2.
Back out the two bolts 7 .
○
3.
Remove the back rind.
Service Manual
Disassemble and Assemble Back Cover Board
Disassemble steps
:
Picture 5-4 Disassemble and Assemble back cover board
1.
Remove the back cover board by the forceps.
Service Manual
Disassemble and Assemble Top Cover Board
Disassemble steps
:
Picture 5-5 Disassemble and Assemble top cover board
1
.
Back out the two bolts
○
9
.
2
.
Pull on the top cover board.
Service Manual
Disassemble and Assemble Fan
Picture 5-6 Disassemble and Assemble fan
Disassemble steps
:
1
.
Back out the four bolts 11 .
○
2
.
Remove the fan 12
○
Service Manual
Disassemble and Assemble power board
Picture 5-7 Disassemble and Assemble power board
Disassemble steps
:
1
.
Back out the four bolts 13 .
○
2
.
Pull out the power line from the main board and remove the power board.
Service Manual
Disassemble and Assemble Metal Shelf
Picture 5-8 Disassemble and Assemble metal shelf
Disassemble steps
:
1
.
Back out the four bolts 14 , and remove the metal shelf
○
15 .
○
2
.
Take off the silica gel key 16 .
○
Service Manual
Disassemble and Assemble Main board
Picture 5-9 Disassemble and Assemble main board
Disassemble steps
:
1
.
Back out the five bolts 17 .
○
2
.
Pull out the screen line and keyboard line from main board.
3
.
Back out the three big nuts from the BNC.
4
.
Pull out the main board 18 slowly.
○
Service Manual
Disassemble and Assemble LCD
Picture 5-10 Disassemble and Assemble LCD
Disassemble steps
:
1
.
Break off the LCD 20 with fingers and thumb.
○
Service Manual
Disassemble and Assemble Keyboard
Picture 5-11 Disassemble and Assemble keyboard
Disassemble steps
:
1
.
Back out the four bolts 21 and remove the keyboard..
○
Service Manual
Chapter 6 Prompting Messages and Troubleshooting
Prompting messages:
■ Trig level at limit! : Mention you that the trigger Level is at a limit when you turn the Trig level knob.
■ Horizon position at limit! : Mention you that the horizontal position is at a limit when you turn the horizon position knob.
■ Volts/Div at limit! : Mention you that the vertical voltage have already touched the Min
2MV/div or the Max 5V/div.
■ Volts position at limit! : The system would display this information when the vertical position is at a limit.
■ Sec/Div at limit! : Prompts that the Volts/Div is at full range while turning the vertical scale knob.
■ Holdoff time at limit! : Turn the “Universal” knob when holdoff time has been to max or min value, now the system will clew this information.
■ Functions isn’t useable! : Under several special modes, the some functions could not be running.
■ No signal! : The system would clew this information when the signal could not match the auto set condition. (Using in the auto set)
■ Adjust at limit! : You could adjust the pulse width by the universal knob till the pulse width has reached min20.0ns or max 10.0s.
■ Location Empty! : If you have no stored waveforms or setups on some location, the screen will display this information when you press the “Recall” button on this location.
■ USB Flash Drive Plug In! : This information will appear when you invert the USB Flash Drive to the USB Host port.
■ USB Flash Drive Pull Out! : This information will appear when you pull out the USB Flash
Service Manual
Drive.
■ Store Data Success! : Save setup data, waveform data or picture data to the internal of the oscilloscope or USB flash successful.
■ Ready Data Success! : Read setup data or waveform data from the internal of the oscilloscope or USB flash successful.
■ Please set back USB to printer! : Press the “S/div” knob will appear this information on the screen when the “Print Key” option is set to “Print Picture” and the “Back USB” option is set to
“Computer”.
■ USB Flash Drive isn’t connected! : When the “Save To” option is set to “File” or the “Print
Key” option is set to “Save Picture” in “Save/Recall” menu , Press the “Save” option button or the “S/div” knob before inverting the USB Flash Drive to the USB Host port will appear this information on the screen.
■ Record Wave Success!
:
: :
:
This message will appear when you finish recording waveforms.
Service Manual
Troubleshooting
1. After the oscilloscope is powered on
,
, ,
, the screen remains dark
,
, ,
, please do as following steps
: : : :
⑴
Check the power cable’s connection.
⑵
Ensure the power switch is turned on.
⑶
After the inspections above
, restart the oscilloscope.
⑷
If the oscilloscope is still not used after the checking, please connect with SIGLENT company
2. If there is no signal wave in the screen after gathering the signal, please do as following steps:
(1) Check whether the probe have connected with the signal cable or not.
(2) Check whether the signal cable have connected with the BNC connector or not.
(3) Check whether the probe has connected with the goods tested or not.
(4) Check whether tested goods have produced the signal or not.
(5) Gather the signal again.
3. The value of the tested voltage is 10 times higher/lower than the real one , please do as following steps:
Check the attenuation quotient whether match the probe attenuation proportion or not.
4. Display the wave, but not steady, please do as following steps:
(1) Check the signal source on the trigger interface whether or not matches the signal channel.
Check the trigger mode: normal signal should use the “edge” trigger mode. The video signal should use the “Video” Trigger mode. The signal would display steady, only using the matching trigger mode.
(2) Attempt to change the “coupling” into “HF Reject” or “LF Reject” display, so that the
High/low frequency noise disturb the trigger should be filtrated
5. Press “RUN/STOP” button, but no display.
Check the trigger mode on the trigger interface whether or not in the “normal” or “single”, and check the trigger level is whether or not over the wave range. If yes, please put the trigger level to the middle position or set the trigger mode to the “Auto” position. In another hand, you could choose the “Auto” button to set up automatically.
6. After the Acquisition is set to Averages or Display Persistence time is set too long, the waveform refreshes slowly.
It is normal in these settings
7. The signal is displayed as ladder like waveform
(1)This phenomenon is normal. The time base maybe is too slow .you should turn the horizontal
SCALE knob to increase horizontal resolution to improve the display.
(2) Maybe the display Type is set to “Vectors”, You could set it to Dots mode to improve the display.
Service Manual
Chapter 7 Updating Parts
13
14
15
16
17
18
19
7
8
9
10
11
12
4
5
6
No.
1
2
3
SIGLENT Provide some replaceable parts in order to maintain or update for users. Please see as the following table. Note that the numbers in the table.
Table 7-1 Updating Parts Item
Name
On-Off button
Back Rind
Big knobs
Small knobs
Bolts
Bolts
Specification
PM2.5*10
PA3*8
Bolts
Back cover board
Bolts
Top cover board
Bolts
Fan
Bolts
Bolts
Metal shelf
Silica gel key
Bolts
Main board
BNC nuts
PA3*10
KM3*8
KM3*30
BM3*8
PA3*8
BM3*8
20
21
22
23
LCD
Bolts
Keyboard
Front panel
All the parts are available by contacting your local SIGLENT office..
KM3*8
Service Manual

Public link updated
The public link to your chat has been updated.
Advertisement
Key features
- High Bandwidth
- High Sampling Rate
- Variety of Measurement and Analysis functions
- Auto setup function
- Save/Recall function
- Probe compensation function
- Self Calibration function
- Color TFT display
- USB Host and Device interface
- RS-232 interface