Illumina Infinium Assay Best Practices Guide
Below you will find brief information for Infinium Assay. This guide provides best practices for optimizing the performance of Infinium Assays, including tips on reagent handling, hybridization, and data analysis. It also covers common issues that users may encounter and provides solutions to these problems.
Advertisement
Advertisement
Best Practices for
Infinium
®
Assays
ILLUMINA PROPRIETARY
Pub. No. 370-2009-010
Current as of 11 June 2009
Topics
2 Evaluate Data in GenomeStudio First
Optimizing Intensity Around the Loading Port
4 Check Z-Height on Robot Deck
Place Lid on Hyb Chamber Base After Adding PB2
Cool Sample Plate for 30 Minutes After Denaturing
Do Not Precondition Hyb Chambers
Check Loading Port for Bolus of Liquid
Do Not Modify the Hybridization Environment
Optimizing Intensity in Other Areas of the BeadChip
10 Cool Hyb Chamber for 25 Minutes After Incubation
10 Remove BeadChip IntelliHyb Seal Slowly
11 Keep Hyb Chamber Lids and Bases Together
12 Do Not Replace PB2 with RA1
12 Place BeadChip in PB1 Immediately After Removing Seal
13 Remove Seal Residue under PB1
13 Disassemble Flow-Through Chambers Immediately After XStain
2
Introduction
This document contains the latest Illumina-supported improvements and best practices for the Infinium Assay. It is a supplemental guide to the commercially released Infinium Assay protocols, designed to optimize sample processing in the varying lab conditions of Illumina’s customers worldwide. Revisions to this document will be made in the event of additional recommendations.
Evaluate Data in
GenomeStudio First
When evaluating assay performance, it is important to analyze data in
Illumina’s GenomeStudio data analysis software before inferring the impact of any observation on overall data quality. For example, sample evaporation around the loading port on a BeadChip is associated with decreased intensity. However, observed sample evaporation during BeadChip processing may not ultimately have a significant impact on data quality. In addition, it is important to note that decreased intensity on areas of the array away from the sample loading port does not correlate with sample evaporation. It is essential to conduct an investigation in GenomeStudio before drawing any conclusions about assay performance.
Reference Material
Please refer to the following documentation for complete instructions on assay protocols. Electronic copies of the latest documentation are available at http://www.illumina.com/documentation. An iCom login is required.
Infinium HD Ultra Assay
`
Infinium HD Ultra Assay Guide (PN 11328087)
`
Infinium HD Ultra Assay Experienced User Card for Automated
Processing (PN 11328108)
`
Infinium HD Ultra Assay Experienced User Card for Manual Processing
(PN 11328095)
Infinium HD Super Assay
`
Infinium HD Super Assay Protocol Guide (PN 11322427)
`
Infinium HD Super Assay Experienced User Card for Automated
Processing (PN 11322291)
`
Infinium HD Super Assay Experienced User Card for Manual Processing
(PN 11322312)
Infinium HD Gemini Assay
`
Infinium HD Gemini Assay Guide (PN 11311007)
`
Infinium HD Gemini Assay Experienced User Card for Automated
Processing (PN 11311023)
`
Infinium HD Gemini Assay Experienced User Card for Manual Processing
(PN 11311015)
Pub. No. 370-2009-010
Current as of 11 June 2009
Technical
Assistance
We are always interested in hearing how other members of the Illumina
Community are optimizing their research. If you would like to share your methods for getting the most out of the Infinium Assay, please contact
Illumina Technical Support. If you have any questions about this document or about how to implement the best practices it describes, contact Technical
Support or your local Field Applications Scientist.
Table 1 Illumina Customer Support Contacts
Contact
Toll-free Customer Hotline
International Customer Hotline
Illumina Website
Number
1-800-809-ILMN (1-800-809-4566)
1-858-202-ILMN (1-858-202-4566) http://www.illumina.com
3
Infinium HD Assay Best Practices Guide
4
Optimizing Intensity Around the Loading Port
The practices described in this section have been found to reduce the incidence of decreased intensities on Infinium BeadChips around the loading port.
Check Z-Height on
Robot Deck
Step in Assay
`
All steps that involve the Tecan robot, and especially the Hybridize
BeadChip step
Best Practice
Review the plate’s sample volumes throughout the protocol to ensure that the Tecan robot is dispensing and aspirating correctly to all wells.
If you encounter any problems with sample evaporation, check the Z-height of the tips on the Tecan robot and verify the amount of sample remaining in the amplification plate. It may be that the Tecan robot is underloading samples. If you suspect that the Tecan tip height is out of specification, contact an Illumina Field Service Engineer.
Background Information
Correct tip height is an important component of good assay performance. If the robot tips are too high and far away from their optimized position within the wells of the sample plate, then the robot cannot aspirate the required amount for sample hybridization. If tip height is too high during sample loading, for example, then less sample will be dispensed onto the BeadChip.
If significant enough, underloading can lead to sample evaporation during hybridization that can affect data quality.
If the tip height is too low and the tips bottom out in the sample well, this can also impair proper aspiration of the plate contents, by impeding the tip’s ability to draw up the correct sample volume.
Use Fresh RA1 for
Each Step
Step in Assay
`
Resuspend the MSA/AMP Plate
`
Single-Base Extension and Stain HD BeadChip (XStain)
Best Practice
The RA1 reagent is used in the Resuspension and XStain steps of the Infinium
Assay. RA1 is shipped frozen and must be gradually warmed to room temperature, then mixed gently to dissolve any crystals. RA1 that has been stored properly and has not been dispensed for use in either the XStain or
Resuspension step is considered fresh RA1. It is important to use fresh RA1 for each protocol step in the assay where it is required. This is especially critical for the resuspension step.
After RA1 has been poured out into a reservoir and exposed to room temperature air for extended periods of time, it is no longer fresh.
Pub. No. 370-2009-010
Current as of 11 June 2009
To make best use of RA1, only pour out the amount needed for the current step. If you plan to perform additional assay steps requiring RA1 that same day, then leave the remaining thawed reagent in the original, closed bottle at room temperature until it is needed. Otherwise, follow the standard RA1 storage procedures described in the assay guide for next-day processing and prolonged storage conditions.
Background Information
Figure 1 shows results from a controlled experiment done to measure the
changes in the osmolality of RA1 over prolonged exposure to specific environmental conditions. The experiment took place over an 8-hour period where a half-reservoir of the reagent at various fill volumes was exposed to a
12% relative humidity environment. The data points are extrapolated from
osmometer measurements for varying volumes of RA1. The graph in Figure 1
shows a line for each reservoir fill volume: 10 ml, 15 ml, 20 ml, 25 ml, and
30 ml.
5
Figure 1 Time Series Model of RA1 Evaporation at RT
The X-axis represents elapsed time in hours, and the Y-axis represents the theoretical formamide+salt concentration as a total percent. The graph clearly indicates that prolonged exposure of RA1 over the course of a day increases the overall concentration of formamide+salt in the reagent solution as water is lost through evaporation. The effect is more pronounced for lower starting volumes in the reagent reservoir.
Place Lid on Hyb
Chamber Base
After Adding PB2
Step in Assay
`
Hybridize Multi Bead Chips | Assemble the Hyb Chambers
Best Practice
1. After you fill the Hyb Chamber reservoirs with PB2, place the lid on the
Hyb Chamber right away to prevent evaporation. The lid does not need to be locked down.
2. After the Hyb Chamber has been loaded with PB2, load BeadChips into the Hyb Chamber within one hour.
Infinium HD Assay Best Practices Guide
6
Figure 2 Hyb Chamber with Lid
Background Information
Changes to the osmolality of PB2 shift the vapor pressure balance between
RA1 (sample hyb buffer) and PB2 away from its optimized point. An imbalance in the vapor pressure between RA1 and PB2 can lead to increased
sample evaporation during the hybridization process. As the data in Figure 3
show, the osmolality of PB2 changes at a much faster rate if the chamber lid is left open.
Figure 3 compares the evaporation rate of PB2 in closed Hyb Chambers
versus open Hyb Chambers at room temperature with 37% relative humidity.
The PB2 was serially loaded into Hyb Chambers at 2 hours, 1.5 hours, 1 hour, and 0.5 hours before sample loading. One chamber was left open throughout, and the other one was closed after each addition.
Figure 3 PB2 Evaporation Rate for Closed and Open Hyb Chambers
The Y-axis shows the increasing range of osmolality (measure of osmoles of solute per kilogram of solvent), as measured by an osmometer when the PB2 was left exposed to the lab environment.
Pub. No. 370-2009-010
Current as of 11 June 2009
Cool Sample Plate for 30 Minutes
After Denaturing
Step in Assay
`
Hybridize Multi Bead Chips | Hybridize Sample to Multi BeadChip
Best Practice
1. After the 20-minute incubation at 95º C, remove the amplification plate from the heat block and place it on the lab bench.
2. Leaving the seal intact, cool the plate at room temperature for
30 minutes.
3. Continue with the instructions in the assay guide, beginning by pulse centrifuging the plate to 280 xg.
Cool plate for 30 minutes
Figure 4 Denaturation Step in Infinium Assay
Background Information
Figure 5 compares the water loss between two amplification plates. After the
20 minute incubation at 95º C, the seal was removed from one plate, and then both plates were left to cool at room temperature for 30 minutes. The plate weight was measured every five minutes. Cumulative water loss per sample was calculated based on the weight loss of the entire plate over that period. The unsealed plate lost approximately 9 μl water per sample, while the sealed plate lost less than 1 μl.
7
Figure 5 Water Loss From Sealed and Unsealed Plates
These data show that sample evaporation can be caused by exposing the heated samples to the lab environment immediately after denaturation. The loss of sample volume from the plate wells can lead to lower sample hybridization volumes on the BeadChip. Underloading the BeadChips makes them more susceptible to lower intensity areas caused by evaporation.
Infinium HD Assay Best Practices Guide
8
It is important to note that the majority of the sample volume loss occurs within the first 10 minutes after removing the plate seal. Letting the sample plate cool for 30 minutes on the lab bench allows the seal to capture the condensate volume that would have evaporated had the plate seal been immediately removed. Pulse centrifuging the plate after cooling it down draws the condensate back into the sample well.
Do Not
Precondition Hyb
Chambers
Step in Assay
`
Hybridize Multi Bead Chips | Load BeadChip
Best Practice
Hyb Chambers should be at room temperature when you load the
BeadChips. They should not be preconditioned in the Illumina Hybridization
Oven prior to loading the BeadChips. Heating the PB2 and then opening the
Hyb Chamber to add BeadChips causes some of the PB2 to evaporate, leading to a change in the osmolality of PB2 and an imbalance in the vapor pressure between PB2 and RA1 (sample hyb buffer).
Check Loading Port for Bolus of Liquid
Step in Assay
`
Hybridize Multi Bead Chips | Load BeadChip
Best Practice
This best practice applies only to manual loading. Any loading problems that occur during robotic loading should be referred to an Illumina Field Service
Engineer.
1. After loading all DNA onto the BeadChip, wait for the sample to disperse over the entire surface.
Figure 6 Loading BeadChip with Sample
2. Inspect the loading port to see if a large bolus of liquid remains. Excess sample volume in the BeadChip loading port helps prevent low-intensity areas resulting from evaporation.
Pub. No. 370-2009-010
Current as of 11 June 2009
Figure 7 Bolus of Liquid at Loading Port
3. If no excess liquid is visible, it is acceptable to add additional sample from the leftover volume in the amplification plate until there is a large bolus around the loading port.
CAUTION
Do not top off with RA1 (sample hyb buffer), as this will dilute the sample.
4. Record the top-off activity on the lab tracking form.
Do Not Modify the
Hybridization
Environment
Step in Assay
`
Hybridize Multi Bead Chips | Set up Multi BeadChip for Hyb
Best Practice
After loading the BeadChips into the Hyb Chambers, place the Hyb
Chambers into the Illumina Hybridization Oven immediately. Do not modify the hybridization environment by adding additional fixtures or humidifying elements. Leave the Hyb Chambers in the oven at the correct orientation and temperature until hybridization is complete. Changes to the hybridization environment can have unexpected effects on data quality.
9
Infinium HD Assay Best Practices Guide
Figure 8 Hyb Chambers in Hyb Oven
10
Optimizing Intensity in Other Areas of the BeadChip
The practices described in this section have been found to reduce the incidence of decreased intensities in areas that are not around the sample loading ports of the BeadChips.
Cool Hyb Chamber for 25 Minutes
After Incubation
Step in Assay
`
Wash BeadChip
Best Practice
1. After removing the incubated Hyb Chamber from the Illumina
Hybridization Oven, leave the chamber with the lid closed on the benchtop at room temperature for 25 minutes.
For efficiency, use the 25 minutes to prepare wash dishes, spacers, and glass back plates for washing.
2. After this cooling period, follow the instructions for removing the seals and washing the BeadChips.
Cool plate for
25 minutes
Figure 9 Step in Assay
Background Information
Internal Illumina studies have shown that allowing BeadChips to cool down before removing the seals reduces the occurrence of dim areas on the side of the BeadChip where the IntelliHyb seal was first pulled off.
Remove BeadChip
IntelliHyb Seal
Slowly
Step in Assay
`
Wash BeadChip
Best Practice
When removing the seal from the BeadChip, slowly creep the seal diagonally along the surface of the BeadChip until the first corner of the sample area on the BeadChip is exposed. Continue peeling off the seal in a slow and steady motion until it is completely removed. Immediately place the peeled
BeadChip into wash buffer.
It is especially important at the beginning of the seal removal process that the seal adhesive is broken away slowly from the BeadChip surface. This technique is most effective when preceded by the 25 minute cooldown period after incubation.
Pub. No. 370-2009-010
Current as of 11 June 2009
Figure 10 Remove IntelliHyb Seal Slowly and Steadily
Background Information
If the IntelliHyb seal is removed too quickly, the corners of the stripes at the bottom of the sample section where you first removed the seal may display decreased intensity. In severe cases, there may be negative effects on call rates and logRdev. Slow, deliberate removal also helps ensure that the glue does not remain on the surface of the BeadChip. Residual glue can interfere with the effectiveness of the XStain process to a greater or lesser degree, depending on the location and severity of the residue.
General Best Practices
The following best practices have not been formally tested for correlation with intensity, but have been found to ensure better results in the assay.
Mix Reagents
Thoroughly
Step in Assay
`
All Steps
Best Practice
It is important that all samples are completely thawed and well mixed before being used in the assay. After thawing reagents to room temperature, always invert them 10 times and pulse centrifuge them to 280 xg for 1 minute.
Keep Hyb Chamber
Lids and Bases
Together
Step in Assay
`
Hybridize Multi Bead Chips | Assemble the Hyb Chambers
Best Practice
To ensure optimal results from Hyb Chambers:
`
Adopt a labeling convention that keeps each Hyb Chamber base paired with its original lid.
`
Check Hyb Chamber lid-base pairs regularly to ensure that the fit remains secure.
Infinium HD Assay Best Practices Guide
11
12
`
Check hinges regularly for any signs of abnormal wear or loose fittings. It is important that the hinges provide adequate clamping strength to ensure an airtight seal between the lid and the base.
`
Record the Hyb Chamber that was used for each BeadChip, so that Hyb
Chambers can be investigated and evaluated in the event of sample evaporation or other lab processing anomalies.
Do Not Replace
PB2 with RA1
Step in Assay
`
Hybridize Multi Bead Chips | Assemble the Hyb Chambers
Best Practice
Do not replace PB2 in the Hyb Chamber with RA1. This will decrease the stringency and may negatively affect sample call rates and logRdev. PB2 is formulated to produce the appropriate amount of humidity within the Hyb
Chamber environment to prevent sample from evaporating during hybridization.
Place BeadChip in
PB1 Immediately
After Removing
Seal
Step in Assay
`
Wash BeadChip
Best Practice ti
1. After removing the IntelliHyb seal from the hybridized BeadChip, immediately place the BeadChip into PB1.
CAUTION
Do not attempt to remove any glue residue from the
BeadChip surface at this time.
Figure 11 Placing BeadChip Immediately into Wash Rack with PB1
2. Repeat this for each BeadChip until they are all placed in the PB1.
3. Ensure that all BeadChips have been in the PB1 buffer for at least one minute before proceeding.
This neutralizes the BeadChip surface in order to maintain high data quality and assay performance.
Pub. No. 370-2009-010
Current as of 11 June 2009
Remove Seal
Residue under PB1
Step in Assay
`
Wash BeadChip
Best Practice
1. After all the BeadChips have been in PB1 for at least one minute, place them into the Alignment Fixture as directed.
2. Before placing the spacer on top of the BeadChips, inspect the surface of each BeadChip for residue left by the seal.
3. If you see any, carefully use a pipette tip to remove any residue under buffer. Be very careful not to scratch the bead area.
4. When the BeadChips are free of residue, proceed to place the spacers and assemble the Flow-Through Chambers as instructed.
Disassemble Flow-
Through Chambers
Immediately After
XStain
Step in Assay
`
Stain BeadChip
Best Practice
It is important that BeadChips are not allowed to dry after staining.
1. Remove the Flow-Through Chambers from the Chamber Rack immediately after staining and place them horizontally on a lab bench at room temperature.
2. Disassemble the Flow-Through Chambers and submerge the BeadChips in the wash dish as soon as possible.
13
Infinium HD Assay Best Practices Guide
Figure 12 Stain BeadChip Step in Assay
14
Pub. No. 370-2009-010
Current as of 11 June 2009
Illumina, Inc.
9865 Towne Centre Drive
San Diego, CA 92121-1975
+1.800.809.ILMN (4566)
+1.858.202.4566 (outside North America) [email protected]
www.illumina.com
Advertisement
Key features
- Optimizing intensity around the loading port
- Optimizing intensity in other areas of the BeadChip
- General best practices
- Detailed instructions on reagent management
- Tips for troubleshooting common issues
- Data quality monitoring and analysis