- Industrial & lab equipment
- Measuring, testing & control
- Multimeters
- Agilent Technologies
- 54502A
- Service manual
advertisement
Errata
Title & Document Type: 16532A Digitizing Oscilloscope Service Manual
Manual Part Number: 16532-97002
Revision Date: July 1993
HP References in this Manual
This manual may contain references to HP or Hewlett-Packard. Please note that Hewlett-
Packard's former test and measurement, semiconductor products and chemical analysis businesses are now part of Agilent Technologies. We have made no changes to this manual copy. The HP XXXX referred to in this document is now the Agilent XXXX.
For example, model number HP8648A is now model number Agilent 8648A.
About this Manual
We’ve added this manual to the Agilent website in an effort to help you support your product. This manual provides the best information we could find. It may be incomplete or contain dated information, and the scan quality may not be ideal. If we find a better copy in the future, we will add it to the Agilent website.
Support for Your Product
Agilent no longer sells or supports this product. You will find any other available product information on the Agilent Test & Measurement website:
www.tm.agilent.com
Search for the model number of this product, and the resulting product page will guide you to any available information. Our service centers may be able to perform calibration if no repair parts are needed, but no other support from Agilent is available.
SERVICE MANUAL
HP 16532A
1 GSa/s Digitizing
Oscilloscope
COPYRIGHT HEWLETT-PACKARD COMPANY,
COLORADO SPRINGS DIVISION 1991-1993
1900 GARDEN OF THE GODS ROAD, COLORADO SPRINGS, COLORADO U.S.A.
ALL RIGHTS RESERVED
Manual Part No. 16532-97002 PRINTED: JULY 1993
SAFETY CONSIDERATIONS
GENERAL - This is a Safety Class I instrument (provided with terminal for protective earthing).
OPERATION - BEFORE APPLYING POWER verify that the power transformer primary is matched to the available line voltage, the correct fuse is installed, and Safety Precautions are taken (see the following warnings). In addition, note the instrument’s external markings which are described under
"Safety Symbols."
•
Servicing instructions are for use by service-trained personnel. To avoid dangerous electric shock, do not perform any servicing unless qualified to do so.
•
BEFORE SWITCHING ON THE INSTRUMENT, the protective earth terminal of the instrument must be connected to the protective conductor of the (mains) powercord. The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. The protective action must not be negated by the use of an extension cord
(power cable) without a protective conductor (grounding).
Grounding one conductor of a two-conductor outlet is not sufficient protection.
•
If this instrument is to be energized via an auto-transformer (for voltage reduction) make sure the common terminal is connected to the earth terminal of the power source.
•
Any interruption of the protective (grounding) conductor
(inside or outside the instrument) or disconnecting the protective earth terminal will cause a potential shock hazard that could result in personal injury.
•
Whenever it is likely that the protection has been impaired, the instrument must be made inoperative and be secured against any unintended operation.
•
Only fuses with the required rated current, voltage, and specified type (normal blow, time delay, etc.) should be used. Do not use repaired fuses or short circuited fuseholders. To do so could cause a shock or fire hazard.
•
Do not operate the instrument in the presence of flammable gasses or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard.
•
Do not install substitute parts or perform any unauthorized modification to the instrument.
•
Adjustments described in the manual are performed with power supplied to the instrument while protective covers are removed. Energy available at many points may, if contacted, result in personal injury.
•
Any adjustment, maintenance, and repair of the opened instrument under voltage should be avoided as much as possible, and when inevitable, should be carried out only by a skilled person who is aware of the hazard involved.
•
Capacitors inside the instrument may still be charged even if the instrument has been disconnected from its source of supply.
SAFETY SYMBOLS
Instruction manual symbol. The product will be marked with this symbol when it is necessary for the user to refer to the instruction manual in order to protect against damage to the product.
Indicates hazardous voltages
Earth terminal (sometimes used in manual to indicate circuit common connected to grounded chassis).
The WARNING sign denotes a hazard. It calls attention to a procedure, practice, or the like, which, if not correctly performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING sign until the indicated conditions are fully understood and met.
The CAUTION sign denotes a hazard. It calls attention to an operating procedure, practice, or the like, which, if not correctly performed or adhered to, could result in damage to or destruction of part or all of the product. Do not proceed beyond a CAUTION sign until the indicated conditions are fully understood or met.
SC1D984
HP 16532A - Table of Contents
CONTENTS
SECTION I
General Information
1-1. Introduction .............................................................................................................................. 1-1
1-2. Module Covered by This Manual ............................................................................................ 1-1
1-3. Safety Requirements ............................................................................................................... 1-1
1-4. Product Description ................................................................................................................. 1-2
1-5. Accessories Supplied .............................................................................................................. 1-2
1-6. Accessories Available............................................................................................................... 1-2
1-7. Specifications ........................................................................................................................... 1-2
1-8. Operating Characteristics ....................................................................................................... 1-2
1-9. Recommended Test Equipment ............................................................................................. 1-2
SECTION II
Installation
2-1. Introduction ............................................................................................................................. 2-1
2-2. Initial Inspection ...................................................................................................................... 2-1
2-3. Preparation for Use ................................................................................................................ 2-1
2-4. Power Requirements .............................................................................................................. 2-1
2-5. Safety Requirements .............................................................................................................. 2-1
2-6. Module Installation ................................................................................................................. 2-1
2-7. Operating Environment .......................................................................................................... 2-4
2-8. Storage .................................................................................................................................... 2-4
2-9. Packaging ............................................................................................................................... 2-4
2-10. Tagging for Service ................................................................................................................ 2-4
SECTION III
Performance Tests
3-1. Introduction .............................................................................................................................. 3-1
3-2. Recommended Test Equipment ............................................................................................. 3-1
3-3. Test Record .............................................................................................................................. 3-1
3-4. Performance Test Interval ....................................................................................................... 3-1
3-5. Performance Test Procedures ................................................................................................ 3-1
3-6. DC CAL OUTPUT Test ............................................................................................................. 3-2
3-7. AC CAL OUTPUT Test ............................................................................................................. 3-3
3-8. Input Resistance ...................................................................................................................... 3-4
3-9. Voltage Measurement Accuracy ............................................................................................. 3-6
3-10. Offset Accuracy ....................................................................................................................... 3-9
3-11. Bandwidth ................................................................................................................................ 3-11
3-12. Time Measurement Accuracy .................................................................................................. 3-13
3-13. Trigger Sensitivity .................................................................................................................... 3-15 v
HP 16532A - Table of Contents
SECTION IV
Calibration
4-1. Introduction............................................................................................................................... 4-1
4-2. Safety Requirements ................................................................................................................ 4-1
4-3. Calibration Interval.................................................................................................................... 4-1
4-4. Maintaining System Integrity.................................................................................................... 4-1
4-5. Calibration Protection Switch................................................................................................... 4-1
4-6. Recommended Test Equipment .............................................................................................. 4-2
4-7. Instrument Warmup.................................................................................................................. 4-2
4-8. Calibration................................................................................................................................. 4-2
4-9. Setting the PROTECT/UNPROTECT Switch to UNPROTECT................................................ 4-3
4-10. Equipment Required for Calibration ....................................................................................... 4-4
4-11. Loading the Default Calibration Factors................................................................................. 4-4
4-12. Self Cal Menu Calibrations ...................................................................................................... 4-5
4-13. Setting the PROTECT/UNPROTECT Switch to PROTECT .................................................... 4-8
4-14. Sealing the Mainframe............................................................................................................. 4-8
Section V
Replaceable Parts
5-1. Introduction .............................................................................................................................. 5-1
5-2. Abbreviations ........................................................................................................................... 5-1
5-3. Replaceable Parts List ............................................................................................................. 5-1
5-4. Ordering Information ............................................................................................................... 5-1
5-5. Exchange Assemblies ............................................................................................................. 5-1
5-6. Direct Mail Order System ........................................................................................................ 5-2
SECTION VI
Service
6-1. Introduction .............................................................................................................................. 6-1
6-2. Safety Requirements ............................................................................................................... 6-1
6-3. Recommended Test Equipment ............................................................................................. 6-1
6-4. Module Block Diagram and Theory of Operation ................................................................... 6-1
6-5. Attenuator Theory of Operation ............................................................................................... 6-1
6-6. Main Assembly Theory of Operation ....................................................................................... 6-1
6-7. Self-Tests .................................................................................................................................. 6-4
6-8. Troubleshooting ....................................................................................................................... 6-10
6-9. Oscilloscope Module Removal and Replacement ................................................................. 6-18
6-10. Trigger Cable Removal and Replacement ............................................................................. 6-20
6-11. Attenuator Removal and Replacement .................................................................................. 6-21 vi
HP 16532A - Table of Contents
LIST OF TABLES
TABLE TITLE PAGE
1-1. HP 16532A Specifications........................................................................................................ 1-3
1-2. HP 16532A Operating Characteristics..................................................................................... 1-4
1-3. Recommended Test Equipment .............................................................................................. 1-9
3-1. Performance Test Record ........................................................................................................ 3-17
4-1. Calibration Sequence............................................................................................................... 4-2
5-1. Reference Designator and Abbreviations................................................................................ 5-3
5-2. Replaceable Parts List .............................................................................................................. 5-5
LIST OF FIGURES
FIGURE TITLE PAGE
2-1. Endplate Overlap - Top Sequence........................................................................................... 2-2
2-2. Endplate Overlap - Bottom Sequence..................................................................................... 2-3
3-1. Voltage Measurement Accuracy Test........................................................................................ 3-7
4-1. Endplate Overlap - Top Sequence........................................................................................... 4-3
4-2. Location of the PROTECT/UNPROTECT Switch..................................................................... 4-3
4-3. Default Calibration Display ....................................................................................................... 4-4
4-4. Ext Trig Null Calibration Connections...................................................................................... 4-7
4-5. End of Calibration Display........................................................................................................ 4-8
4-6. Endplate Overlap - Top Sequence .......................................................................................... 4-8
5-1. Parts Identification.................................................................................................................... 5-4
6-1. HP 16532A Oscilloscope Block Diagram ................................................................................ 6-2
6-2. Startup Screen.......................................................................................................................... 6-4
6-3. Load Test System Screen ........................................................................................................ 6-5
6-4. Test System Screen.................................................................................................................. 6-5
6-5. Functional Tests Screen........................................................................................................... 6-6
6-6. Main Test Menu ........................................................................................................................ 6-6
6-7. A/D Test Run Screen ................................................................................................................ 6-7
6-8. Stop Field .................................................................................................................................. 6-7
6-9. Exit Test System Screen .......................................................................................................... 6-8
6-10. Troubleshooting Flowchart ...................................................................................................... 6-13
6-11. Endplate Overlap - Top Sequence........................................................................................... 6-18
6-12. Endplate Overlap - Bottom Sequence..................................................................................... 6-19
6-13. Oscilloscope Module Exploded View ...................................................................................... 6-20 vii
TABLE OF CONTENTS
General Information
1-1. Introduction...........................................................................................................................1-1
1-2. Module Covered by Manual .................................................................................................1-1
1-3. Safety Requirements.............................................................................................................1-1
1-4. Product Description..............................................................................................................1-2
1-5. Accessories Supplied ...........................................................................................................1-2
1-6. Accessories Available ...........................................................................................................1-2
1-7. Specifications........................................................................................................................1-2
1-8. Operating Characteristics.....................................................................................................1-2
1-9. Recommended Test Equipment ..........................................................................................1-2
HP 16532A - General Information
SECTION I
General Information
1-1. Introduction
The service policy for this instrument is replacement of defective assemblies. This service guidel contains information for finding a defective assembly by testing and servicing the HP 16532A
1 GSa/s Oscilloscope Module. Also included in the guide are installation procedures and a list of recommended test equipment. This guide is divided into six sections as follows:
I - General Information
II - Installation
III - Performance Tests
IV - Calibration
V- Replaceable Parts
VI - Service
The General Information Section includes safety requirements, a product description, and a list of accessories supplied and of accessories available.
Also included are tables listing specifications and operating characteristics, and a list of recommended test equipment.
To complete the service documentation for your
HP 16500 Logic Analysis System, place this service manual in the 3-ring binder with your other logic analysis system Service Guides.
1-2. Module Covered by This Manual
The information covered in this manual is for the
HP 16532A 1 GSa/s Oscilloscope Module.
1-3. Safety Requirements
Specific warnings, cautions, and instructions are placed wherever applicable throughout the manual. These must be observed during all phases of operation, service, and repair of the module. Failure to comply with them violates safety standards of design, manufacture, and intended use of this module.
Hewlett-Packard assumes no liability for the failure of the customer to comply with these safety requirements.
1-1
HP 16532A - General Information
1-4. Product Description
The HP 16532A Oscilloscope Module is a 1 GSa/s digitizing oscilloscope. Some of the main features are:
• 1 GSa/s digitizing for 250 MHz bandwidth single shot analysis.
• 8000 samples per channel.
• Automatic pulse parameters, displays time between markers, acquires until specified time between markers is captured, performs statistical analysis on time between markers.
• Lightweight miniprobes.
• 24-pin IC test clip (HP 10211A)
• BNC-to-BNC ac coupling capacitor
(HP 10240B)
10:1 Probe:
• 1-M
Ω
, 7.5-pf miniprobe, 1 m (HP 10435A)
1:1 Probes:
• 40-pf miniprobe, 1 m (HP 10438A)
• 64-pf miniprobe, 2 m (HP 10439A)
• 50-
Ω
miniprobe, 2 m (HP 10437A)
100:1 Probe:
• 10-M
Ω
, 2.5-pf miniprobe, 2 m (HP 10440A)
1-5. Accessories Supplied
The following accessories are supplied with the
HP 16532A Oscilloscope module. Quantity one unless shown otherwise.
• User’s Reference and Programming Reference set
• Service Guide
• HP 10441A 10:1 probes, Qty 2
• Right angle BNC adapters, Qty 2
• BNC miniprobe adapter
• Master-slave trigger cable
• Operating System Software
1-6. Accessories Available
• 10:1, 100:1, 50
Ω
, 10 pf resistive divider probe set, 1.2 m (HP 10020A)
• BNC-to-BNC cable, 1.2 m (HP 10503A)
1-7. Specifications
Module specifications are listed in table 1-1. These specifications are the performance standards against which the module is tested.
1-8. Operating Characteristics
Table 1-2 is a listing of the module operating characteristics. The operating characteristics are not specifications, but are the typical operating characteristics included as additional information for the user.
1-9. Recommended Test
Equipment
Equipment required to test and maintain the
HP 16532A Oscilloscope Module is listed in table
1-3. Other equipment may be substituted if it meets or exceeds the critical specifications listed in the table.
1-2
HP 16532A - General Information
Table 1-1. HP 16532A Specifications
HP 16532A SPECIFICATIONS
Type: 2-channel simultaneous acquisition
Bandwidth
(* ,1)
: dc to 250 MHz (real time, dc-coupled)
Maximum Sample Rate: 1 Gigasample per second
Number of Channels: 2, 4, 6, 8 simultaneous channels using the same time base setting(s)
OR
Up to 10 channels with independent time bases for each pair of channels.
Up to 18 channels with the HP 16501A Expansion Frame.
Rise Time
(2)
: 1.4 ns
ADC: 8-bit real time
Vertical Resolution: 8 bits over 4 vertical divisions (
±
0.4%)
Waveform Record Length: 8000 points
Time Interval Measurement Accuracy
(* ,3,6)
:
±
[(0.005% X
∆ t)+ (2 x 10
−
6
x delay setting)+ 150 ps]
Vertical (dc) Gain Accuracy
(4)
:
±
1.5% of full scale
DC Offset Accuracy
(* )
:
±
(1.0% of channel offset + 2.0% of full scale)
Voltage Measurement Accuracy
(* ,5)
:
±
[(1.5% of full scale + offset accuracy) + (0.008 x V/div)]
Trigger Sensitivity
(* )
: dc to 50 MHz: 0.063 x full scale, 50 to 250 MHz: 0.125 x full scale
Input Coupling: 1 M
Ω
: ac and dc, 50
Ω
: dc only
Input R (selectable)
(* )
: 1 M
Ω
:
±
1%, 50
Ω
:
±
1%
Input C: Approximately 7 pF
___________________________________________________________________________________________
NOTES:
* Specifications (valid within
±
10
°
C of auto-calibration temperature, excluding bandwidth
− see
note 1 for bandwidth specification.)
1. Upper bandwidth reduces by 2.5 MHz for every degree C above 35
°
C.
2. Rise time is calculated from the formula: t r
=
0.35
bandwidth
3. Specification applies to the maximum sampling rate. At lower rates, the specification is:
±
[(0.005% x
∆ t)+ (2 x 10
-6
x delay setting) + (0.15 x sample interval)] for bandwidth limited signals
1
(t r
= 1.4 x sample interval). Sample interval is defined as sample rate
4. Vertical gain accuracy decreases 0.08% per degree C from software calibration temperature.
5. Digitizing level = (# vertical divisions)(
1
2
)(
1
LSB
), where LSB= 2
# bits in ADC
6. The Time Interval Measurement Accuracy deteriorates across multiple modules connected as one
unit with each added module.
1-3
HP 16532A - General Information
Table 1-2. HP 16532A Operating Characteristics
HP 16532A OPERATING CHARACTERISTICS
VERTICAL (at BNC)
Vertical Sensitivity Range (1:1 Probe):
4 mV/div to 10 V/div in 1-2-4 increments
DC Offset Range (1:1 Probe):
Vertical Sensitivity
4 mV/div to 100 mV/div
> 100 mV/div to 500 mV/div
> 500 mV/div to 2.5 V/div
> 2.5 V/div to 10 V/div
Probe Factors:
Available Offset
±
2 V
±
10 V
±
50 V
±
250 V
Any integer ratio from 1:1 to 1000:1
Maximum Safe Input Voltage:
1 M
Ω
:
±
250 V [dc + peak ac (< 10 KHz)]
50
Ω
:
±
5 VRMS
Channel-to-Channel Isolation: dc to 50 MHz: 40 dB, 50 MHz to 250 MHz: 30 dB
TIMEBASE
Range:
1 ns/div to 5 s/div
Resolution:
20 ps
Delay Pre-trigger Range:
Time/div Setting
1
µ s to 5 s/div
1 ns to 500 ns/div
Delay Post-trigger Range:
Time/div Setting
100 ms to 5 s/div
1
µ s to 50 ms/div
1ns to 500 ns/div
Available Delay
-8 x (s/div)
-4
µ s
Available Delay
2.5 ks
33,500 x (s/div)
16.7 ms
1-4
HP 16532A - General Information
Table 1-2. HP 16532A Operating Characteristics (cont.)
TRIGGERING
Trigger Level Range:
Within display window (full scale and offset)
Trigger Modes:
Immediate:
Triggers immediately after arming condition is met.
Edge:
Triggers on rising or falling edge of any internal channel or external trigger, count adjustable from 1 to 32,000.
Pattern:
Triggers on entering, exiting, duration greater than, duration less than, and duration in range for a specified pattern of both internal channels and external trigger, count adjustable from 1 to 32,000. Duration time range is adjustable from 20 ns to 160 ms in 10 ns steps. Duration accuracy is approximately
±
3%
±
2 ns.
Auto-Trigger:
If enabled, the module will self-trigger if no trigger condition is found within approximately 50 ms after arming.
Events Delay:
The trigger can be set to occur on the nth edge or pattern, as specified by the user. The number of events (n) can be set from 1 to 32,000 events. Maximum count frequency is 70 MHz.
Intermodule:
Arms another measurement module or triggers the rear panel BNC.
External:
If enabled, the oscilloscope will trigger on an ECL level signal (-1.3 V).
DIGITIZER
Resolution: 8 bits (1 part in 256)
Digitizing Rate: Up to 1 Gigasample per second
Digitizing Technique: Real-time digitizing; each 8000 samples are acquired on a single acquisition
Acquisition Memory Size: 8000 samples per channel
1-5
HP 16532A - General Information
Table 1-2. HP 16532A Operating Characteristics (cont.)
WAVEFORM DISPLAY
Displayed Waveforms:
Eight waveform windows maximum, with scrolling across 96 waveforms.
Display Formats:
Waveforms can be displayed in an overlapping and/or non-overlapping format.
Display capability of A
−
B and A+ B is also provided.
Display Resolution:
500 points horizontal, 240 points vertical
DISPLAY MODES
Normal (Single):
New acquisition replaces old acquisition on screen.
Accumulate:
New acquisition is displayed in addition to previous acquisitions until screen is erased.
Average:
New acquisitions are averaged with old acquisitions, with the updated waveform displayed until erased. Maximum number of averages is 256.
Overlay:
Up to 8 acquired waveforms from separate channels can be overlayed in the same display area.
Connect-the-dots:
Provides a display of the sample points connected by straight lines.
Waveform Reconstruction:
When there is insufficient data to fill every horizontal location, a post acquisition reconstruction filter fills in missing data points for time base < 50 ns/div.
Waveform Math:
Display capability of A
−
B and A+ B functions is provided.
1-6
HP 16532A - General Information
Table 1-2. HP 16532A Operating Characteristics (cont.)
MEASUREMENT AIDS
Time Markers:
Two vertical markers, X and O, are provided for measurements of time and voltage. Capabilities are: measure voltage at point where X and O markers cross each analog waveform; measure time from trigger to X, trigger to O, and
X to O; automatic marker placement by specifying percentage of edge, edge number, and rising or falling edge type; run until X to O > than, < than, in range, and not in range provides selective event search; X to O statistics (mean, max, and min) provide analysis of time interval deviation.
Automatic Search:
Searches for a percentage or an absolute voltage level at a positive or negative edge, count adjustable from 1 to 100.
Auto Search Statistics:
Mean, maximum, and minimum values for elapsed time from X to O markers for multiple runs. Number of valid runs and total number of runs available.
Automatic Measurements:
The following pulse parameter measurements can be performed automatically:
Frequency
Period
V p-p
Rise time
Fall time
Preshoot
Overshoot
+ pulse width
− pulse width
V amplitude
*
(* gives difference between top and base
voltages; only available over bus)
Grid:
Graticules can be displayed in backround of waveform.
Voltage Markers:
Two horizontal markers, A and B, are provided for voltage measurements.
These markers can each be placed on any acquisition channel trace. A delta voltage display provides delta V between markers on the same channel or between two channels.
1-7
HP 16532A - General Information
Table 1-2. HP 16532A Operating Characteristics (cont.)
SETUP AIDS
Autoscale:
Autoscales the vertical and horizontal ranges, offset, and trigger level to display the input signals. Requires a frequency between 50 Hz and 250 MHz.
Presets:
Scales the vertical range, offset, and trigger level to predetermined values for displaying ECL or TTL waveforms.
Calibration:
Vertical, trigger, delay, and all defaults. Calibration factors stored in NV-RAM on the circuit board.
Probe Compensation Source:
External BNC supplies a square wave approximately 0.0 mV to -800 mV into the open circuit at approximately 1000 Hz.
OPERATING ENVIRONMENT
Temperature:
Instrument, 0° to 55° C (+ 32° to 131° F). Probes and cables, 0° to 65° C (+ 32° to 149° F).
Humidity:
Instrument, up to 95% relative humidity at + 40° C (+ 104° F). Recommended disk media, 8% to 80% relative humidity at + 40° C (+ 104° F).
Altitude:
To 4600 m (15,000 ft).
Vibration
Operation:
Random vibration 5-500 Hz, 10 minutes per axis, ~ 0.3 g (rms).
Non-operating:
Random vibration 5-500 Hz, 10 minutes per axis, ~ 2.41 g (rms); and swept sine resonant search, 5-500 Hz, 0.75 g (0- peak), 5 minute resonant dwell @ 4 resonances per axis.
1-8
HP 16532A - General Information
Table 1-3. Recommended Test Equipment
CRITICAL SPECIFICATIONS RECOMMENDED MODEL USE* EQUIPMENT
Signal Generator
Dc Power Supply
Frequency: 1 - 250 MHz at approx .
170 mV RMS resolution
Output Accuracy:
±
1 dB
1 MHz time base accuracy 0.25 ppm
HP 8656B Option 001
Range:
−
35.000 to + 35.000 Vdc,
±
1 mV HP 6114A
Digital Multimeter 0.1 mV resolution
Accuracy: better than 0.005%
Resistance measurement: better than
0.25% accuracy
Power Meter/Sensor 1 - 250 MHz
±
3% accuracy
HP 3458A
HP 436/8482A
HP 11667B Power Splitter Outputs differ by < 0.15 dB
Blocking Capacitor 50
Ω
BNC (m-to-f), 0.18
µ
F,
±
200 V
Adapter 50
Ω
BNC (m)(m)
Adapter
Adapter
Adapter
Tee Adapter
50
Ω
Type N(m)-to-BNC(m)
50
Ω
BNC(f)-to-Dual Banana Plug
50
Ω
Type N(m)-to-BNC(f)
50
Ω
BNC (m)(f)(f)
Cable
Cable
Cable
Cable
Type N(m-to-m) 24-inch
50
Ω
BNC (m-to-m) 48-inch
50
Ω
BNC (m-to-m) 9 inch
50
Ω
BNC-SMB
HP 10240B
HP 1250-1236
HP 1250-0082
HP 1251-2277
HP 1250-0780
HP 1250-0781
HP 11500B
HP 10503A
HP 10502A
HP 16532-61601
P
P
P
P
P
P
P,C
P
C
P
P
C
C
P
P
* P= Performance Tests C= Calibrations
1-9
TABLE OF CONTENTS
Installation
2-1. Introduction ......................................................................................................................... 2-1
2-2. Initial Inspection .................................................................................................................. 2-1
2-3. Preparation for Use ............................................................................................................. 2-1
2-4. Power Requirements .......................................................................................................... 2-1
2-5. Safety Requirements ........................................................................................................... 2-1
2-6. Module Installation .............................................................................................................. 2-1
2-7. Operating Environment ...................................................................................................... 2-4
2-8. Storage ................................................................................................................................ 2-4
2-9. Packaging ........................................................................................................................... 2-4
2-10. Tagging for Service ............................................................................................................. 2-4
HP 16532A - Installation
SECTION II
Installation
2-1. Introduction
This section explains how to initially inspect the
HP 16532A Oscilloscope Module, then how to prepare it for use, storage, and shipment. Also included are procedures for module installation.
2-2. Initial Inspection
Inspect the shipping container for damage. If the shipping container or cushioning material is damaged, it should be kept until the contents of the shipment have been checked for completeness and the module has been checked mechanically and electrically. The contents of the shipment should be as listed in the "Accessories
Supplied" paragraph located in Section I.
Procedures for checking electrical performance are in Section III. If the contents of the container are incomplete, there is mechanical damage or defect, or the instrument does not pass the performance tests, notify the nearest
Hewlett-Packard sales office.
If the shipping container is damaged, or the cushioning material shows signs of stress, notify the carrier as well as the Hewlett-Packard sales office. Keep the shipping material so the carrier can inspect it. The Hewlett-Packard office will arrange for repair or replacement at
Hewlett-Packard’s option without waiting for claim settlement.
2-3. Preparation for Use
2-4. Power Requirements
All power supplies required for operating the
HP 16532A Oscilloscope Module are supplied to the module through the backplane connector in the mainframe.
2-5. Safety Requirements
Specific warnings, cautions, and instructions are placed wherever applicable throughout the manual. These must be observed during all phases of operation, service, and repair of the module. Failure to comply with them violates safety standards of design, manufacture, and intended use of this module.
Hewlett-Packard assumes no liability for the failure of the customer to comply with these safety requirements.
2-6. Module Installation
Do not install, remove, or replace the module in the instrument unless the instrument power is turned off.
The HP 16532A Oscilloscope Module will take up one slot in the card cage. For every additional
HP 16532A Oscilloscope Module you install, you will need an additional slot. They may be installed in any slot and in any order. The installation procedure for the module is continued, step-by-step, on the next page.
Read the Safety Considerations in the front of this manual and in Section I before installing or operating this module.
2-1
HP 16532A - Installation
Module Installation (cont.)
The effects of ELECTROSTATIC DISCHARGE can damage electronic components. Use grounded wrist straps and mats when you are performing any kind of service to this module.
Installation Considerations
•
•
The HP 16532A Oscilloscope Modules can be installed in any available card slot.
For multiple modules connected as one unit, the master card must be placed in the slot above the expander cards. The master card and expander cards must be in adjoining slots.
• Cards or filler panels below the empty slots intended for module installation do not have to be removed to install a module above them.
Procedure:
1. Turn instrument power switch off, unplug power cord and disconnect any input connections.
2. Starting from the top, loosen the thumb screws on the filler panels and cards.
3. Starting from the top, begin pulling out the cards and filler panels halfway. See figure 2-1.
Figure 2-1. Endplate Overlap - Top Sequence
2-2
HP 16532A - Installation
4. Slide the oscilloscope module into the card cage approximately halfway.
5. If you have more oscilloscope cards to install, slide them into the card cage approximately halfway.
6. Firmly seat the bottom card into the backplane connector. Keep applying pressure to the center of the card endplate while tightening the thumb screws finger tight.
7. Repeat step 6 for all cards and filler panels in a bottom-to-top order. See figure 2-2.
8. Any filler panels that are not used should be kept for future use. Filler panels must be installed in all unused card slots for correct air circulation.
Figure 2-2. Endplate Overlap - Bottom Sequence
2-3
HP 16532A - Installation
2-7. Operating Environment
The operating environment is listed in table 1-2 of
Section I of this manual. Note should be made of the non-condensing humidity limitation.
Condensation within the instrument can cause poor operation or malfunction. Protection should be provided against internal condensation.
The HP 16532A Oscilloscope Module will operate at all specifications within the temperature and humidity range given in table 1-2. However, reliability is enhanced when operating the module within the following ranges.
• Temperature:
+
20 °C to + 35 °C (+ 68 °F to
+ 95 °F)
• Humidity: 20% to 80% non-condensing
2-8. Storage
The module may be stored or shipped in environments within the following limits:
• Temperature:
−
40 °C to + 75 °C
• Humidity: Up to 90% at 65 °C
• Altitude: Up to 15,300 meters (50,000 feet)
The module should also be protected from temperature extremes which cause condensation on the instrument.
2-9. Packaging
The following general instructions should be used for repacking the module with commercially available materials.
• Wrap the module in antistatic plastic.
• Use a strong shipping container. A doublewalled carton made of 350-lb test material is adequate.
• Use a layer of shock-absorbing material 70 to
100 mm (3- to 4-inch) thick around all sides of the module to provide firm cushioning and to prevent movement within the container.
• Seal the shipping container securely.
• Mark the shipping container FRAGILE to ensure proper handling.
• In any correspondence, refer to the module by its model number and serial number.
2-10. Tagging for Service
If the module is to be shipped to a
Hewlett-Packard sales office for service or repair, attach a tag showing owner (with address), complete serial number, and a description of the service required.
2-4
TABLE OF CONTENTS
Performance Tests
3-1. Introduction ......................................................................................................................... 3-1
3-2. Recommended Test Equipment ........................................................................................ 3-1
3-3. Test Record ......................................................................................................................... 3-1
3-4. Performance Test Interval .................................................................................................. 3-1
3-5. Performance Test Procedures ........................................................................................... 3-1
3-6. DC CAL OUTPUT Test ........................................................................................................ 3-2
3-7. AC CAL OUTPUT Test ........................................................................................................ 3-3
3-8. Input Resistance ................................................................................................................. 3-4
3-9. Voltage Measurement Accuracy ........................................................................................ 3-6
3-10. Offset Accuracy .................................................................................................................. 3-9
3-11. Bandwidth ........................................................................................................................... 3-11
3-12. Time Measurement Accuracy.............................................................................................. 3-13
3-13. Trigger Sensitivity ............................................................................................................... 3-15
HP 16532A - Performance Tests
SECTION III
Performance Tests
3-1. Introduction
The procedures in this section test the oscilloscope module’s electrical performance using the specifications listed in Section I as the performance standards. All tests can be performed without access to the interior of the instrument. At the end of this section is a form that can be used as a record of performance test results.
3-2. Recommended Test Equipment
Equipment recommended for performance tests is listed in table 1-3. Any equipment that satisfies the critical specifications given in the table may be substituted for the recommended models.
The results recorded on the test record may be used for comparison in periodic maintenance and troubleshooting or after repairs and adjustments have been made.
3-4. Performance Test Interval
Periodic performance verification of the
HP 16532A Oscilloscope Module is required at one-year intervals. The instrument’s performance should be verified after it has been serviced, or if improper operation is suspected. Calibration should be performed before any performance verification checks.
3-5. Performance Test Procedures
All performance tests should be performed at the environmental operating temperature of the instrument, after a 30-minute warmup period.
3-3. Test Record
Results of performance tests may be tabulated on the Performance Test Record (table 3-1) at the end of the procedures. The test record lists all of the tested specifications and their acceptable limits.
3-1
HP 16532A - Performance Tests
3-6. DC CAL OUTPUT Test
The DC CAL OUTPUT port on the rear panel is used for instrument calibration and probe calibration.
Though calibrator accuracy is not specified in the performance specifications, it must be within limits in order to provide accurate calibration.
Test Limits:
5.000 V
±
10 mV
Equipment Required:
The following equipment is required for the test.
Equipment Required
Digital Multimeter
Cable
Adapter
Critical Specification
0.1 mV resolution, better than 0.005% accuracy
50
Ω
BNC (m to m) 48-inch
50
Ω
BNC (f) to Dual Banana Plug
Recommended
Model/Part Number
HP 3458A
HP 10503A
HP 1251-2277
Procedure:
1. Use the BNC-to-banana adapter to connect the BNC cable between the multimeter and the oscilloscope module DC CAL OUTPUT connector.
2. On the oscilloscope module set the following parameters in the order given:
Calibration
Menu
Mode
Procedure
DC Volts
Selection Setting
Service Calibration
DC Cal BNC
0 V
3. The DVM should read close to 0.0000 V. Record the reading to four decimal places.
V1 = __________ .
4. In the Calibration menu set the DC Volts to 5 V.
5. The DVM should read close to 5.0000 V. Record the reading to four decimal places.
V2 = __________ .
6. In the Calibration menu set the DC Volts to 0 V.
7. Subtract V1 from V2. The difference should be between 4.990 and 5.010 V. Record the reading in the performance test record.
3-2
HP 16532A - Performance Tests
3-7. AC CAL OUTPUT Test
The AC CAL OUTPUT port on the rear panel is used for instrument calibration and probe compensation.
Though calibrator accuracy is not specified in the performance specifications, it must be within limits in order to provide accurate calibration.
Test Limits:
1000 Hz
±
10%, 0.8 Vp_p
±
10%
Equipment Required:
The following equipment is required for the test.
Equipment Required
Cable
Critical Specification
50
Ω
BNC (m to m) 48-inch
Recommended
Model/Part Number
HP 10503A
Procedure:
1. Use the BNC cable to connect the AC CAL OUTPUT to channel 1 input of the oscilloscope module.
2. On the oscilloscope module set the following parameters in the order given:
Calibration
Channel
Menu
Mode
Procedure
Signal
Coupling
Probe
Selection Setting
Service Calibration
Oscillator Output
Probe Compensation
1 M
Ω
/DC
1:1
3. In the Calibration menu select AUTOSCALE.
4. In the Auto-Measure menu verify that the waveform is approximately 0.8 Vp_p at approximately
1000 Hz. Record the reading in the performance test record.
3-3
HP 16532A - Performance Tests
3-8. INPUT RESISTANCE
This test verifies the input resistance of the oscilloscope module. A four-wire measurement is used for accuracy at 50
Ω
.
Specification:
1 M
Ω ±
1% and 50
Ω ±
1%
Equipment Required:
The following equipment is required for the test. Procedures are based on the model or part numbers recommended.
Equipment Required
Digital Multimeter
Cables (2)
Adapter
Adapters (2)
Critical Specification
Measure resistance (4-wire) better than 0.25% accuracy
50
Ω
BNC (m to m) 48-inch
50
Ω
BNC Tee (m)(f)(f)
50
Ω
BNC (f) to Dual Banana Plug
Recommended
Model/Part Number
HP 3458A
HP 10503A
HP 1250-0781
HP 1251-2277
Procedure:
1. Set up the multimeter to make a 4-wire resistance measurement.
2. Use the BNC-to-banana adapters to connect one end of each BNC cable to the 4-wire resistance connections on the multimeter, and connect the free ends of the cables to the BNC Tee.
3. Connect the male end of the BNC tee to the channel 1 input of the oscilloscope module. Set up the oscilloscope module according to the following parameters in the order given:
Menu
Channel
Trigger
Input
Probe
V/Div
Offset
Coupling
Mode
Selection
1
1:1
20 mV
0 V
50
Ω
/ DC
Immediate
Setting
3-4
HP 16532A - Performance Tests
4. Select RUN - SINGLE. The clicking of attenuator relays should be audible. Verify resistance readings on the digital multimeter of 50
Ω ±
0.5
Ω
(49.5 to 50.5
Ω
).
5. In the Channel menu change the Coupling field to 1 M
Ω
/DC.
6. Select RUN. The clicking of attenuator relays should be audible. Verify resistance readings on the digital multimeter of 1 M
Ω ±
10 k
Ω
(0.990 to 1.010 M
Ω
).
7. In the Channel menu change the Coupling field to 50
Ω
/DC and V/Div to 200 mV/Div. Repeat steps
4 through 6.
8. In the Channel menu change the Coupling field to 50
Ω
/DC and V/Div to 1 V/Div. Repeat steps 4 through 6.
9. In the Channel menu change the Coupling field to 50
Ω
/DC and V/Div to 4 V/Div. Repeat steps 4 through 6.
10. Connect the male end of the BNC tee to the channel 2 input of the oscilloscope module.
11. Repeat steps 3 through 9 for channel 2, replacing channel 1 with channel 2 where applicable.
Note
If a reading is not within limits, then the attenuator for the out-of-bounds channel should be replaced. See "Service" in Section VI.
3-5
HP 16532A - Performance Tests
3-9. VOLTAGE MEASUREMENT ACCURACY
This test verifies the DC voltage measurement accuracy of the instrument. This procedure uses a dual cursor measurement that nullifies offset error, resulting in the following measurement:
±
[(1.5% of full scale) + (0.016 x V/div)].
Specification:
±
[(1.5% of full scale + offset accuracy) + (0.008 x V/div)]
Equipment Required:
The following equipment is required for this test. Procedures are based on the model or part numbers recommended.
Equipment Required Critical Specification
DC Power Supply
Digital Multimeter
Cable
Adapter
(cable to power supply)
Adapter
Blocking Capacitor
-14 Vdc to + 14 Vdc, 0.1 mV resolution
Better than 0.1% accuracy
50
Ω
BNC (m to m) 48-inch
50
Ω
BNC (f) to Dual Banana Plug
50
Ω
BNC tee (m)(f)(f)
50
Ω
BNC (m to f) 0.18
µ
F, 200 V
Recommended
Model/Part Number
HP 6114A
HP 3458A
HP 10503A
HP 1251-2277
HP 1250-0781
HP 10240B
3-6
HP 16532A - Performance Tests
Procedure:
1. Using a BNC adapter, connect one end of the cable to the power supply. Connect the BNC tee, the blocking capacitor, and the shorting endcap to the other end of the cable. Refer to figure 3-1.
Figure 3-1. Voltage Measurement Accuracy Test
2. In the Waveform Selection menu delete channel 2. If channel 1 is not inserted, insert channel 1. Set the following parameters in the order given.
Menu
Channel
Trigger
Display
Input
Probe
Coupling s/Div
Mode
Mode
Average #
Grid
Markers
Selection
1
1:1
1 M
Ω
/ DC
500 ns
Immediate
Average
8
On
On
Setting
Set the Channel Coupling field to 1 M
Ω
/DC or damage to the equipment will result.
3-7
HP 16532A - Performance Tests
Use the following table for steps 3 through 9.
V/Div
4 V/Div
1 V/Div
400 mV/Div
40 mV/Div
40 mV/Div
400 mV/Div
1 V/Div
4 V/Div
Offset
-7.0 V
-1.75 V
-700.0 mV
-70.0 mV
70.0 mV
700.0 mV
1.75 V
7.0 V
Supply
-14.0 V
-3.50 V
-1.40 V
-140.0 mV
140.0 mV
1.40 V
3.50 V
14.0 V
Upper Limit
-13.7 V
-3.43 V
-1.37 V
-137.0 mV
143.0 mV
1.43 V
3.57 V
14.3 V
Lower Limit
-14.3 V
-3.57 V
-1.43 V
-143.0 mV
137.0 mV
1.37 V
3.43 V
13.7 V
3. With the power supply disconnected from the channel input, set the Channel menu to the V/Div range and offset values shown in the first line of the table.
4. On the oscilloscope, select RUN - REPETITIVE and wait for approximately five seconds, then select
STOP. Read the voltage displayed under the channel label on the left of the display. Enter this value on the performance test record in the "Zero Input" column.
5. Set the power supply to the first supply value listed in the table. Connect the power supply to the channel input as shown in figure 3-1.
6. On the oscilloscope, select RUN - REPETITIVE and wait for approximately five seconds, then select
STOP. Read the voltage displayed under the channel label on the left of the display. Enter this value on the performance test record in the "Measured" column.
7. To obtain the test results, subtract the Zero Input voltage from the Measured voltage listed in the performance test record. For example: (Measured) - (Zero Input) = (Actual). When calculating the results, observe the plus or minus value of the voltage measurements. Record the results in the performance test record in the "Actual" column.
8. Repeat steps 3 through 8 for the second line of the table, then for the rest of the lines of the table for channel 1.
9. In the Waveform Selection menu, select Input channel 2. Delete channel 1 and insert channel 2. Repeat steps 2 through 8 for channel 2, replacing channel 1 with channel 2 where applicable.
3-8
HP 16532A - Performance Tests
3-10. OFFSET ACCURACY
This test verifies the offset accuracy.
Specification:
±
(1.0% of channel offset + 2.0% of full scale)
Equipment Required:
The following equipment is required for the test. Procedures are based on the model or part numbers recommended.
Equipment Required
DC Power Supply
Digital Multimeter
Cable
Adapter
Critical Specification
−
35.000 to + 35.000 Vdc,
±
1 mV resolution
Better than 0.1% accuracy
50
Ω
BNC (m to m) 48-inch
50
Ω
BNC (f) to Dual Banana Plug
Recommended
Model/Part Number
HP 6114A
HP 3458A
HP 10503A
HP 1251-2277
Procedure:
Zero-Input Offset
1. Disconnect all cables and probes from the oscilloscope inputs. In the Waveform Selection menu delete channel 2. If channel 1 is not inserted, insert channel 1. Then set up the oscilloscope module according to the following parameters in the order given:
Menu
Channel
Trigger
Display
Input
Probe v/Div
Offset
Coupling s/Div
Mode
Mode
Average #
Grid
Markers
Selection
1
1:1
4 V
0 V
1 M
Ω
/ DC
500 ns
Immediate
Average
32
On
On
Setting
2. With the DC power supply disconnected from the channel input, make sure the markers are near the center of the display.
3-9
HP 16532A - Performance Tests
3. Select RUN - REPETITIVE. Wait approximately 15 seconds (averaging complete), then select
STOP. The display should read 0.00 V
±
320.0 mV as read from the markers voltage field. Record the reading in the performance test record.
4. In the Channel menu, change the vertical sensitivity to 1 V/Div. Select RUN - REPETITIVE and wait approximately 15 seconds (averaging complete), then select STOP. The display should read 0.00 V
±
80.0 mV. Record the reading in the performance test record.
5. In the Channel menu, change the vertical sensitivity to 100 mV/Div. Select RUN - REPETITIVE and wait approximately 15 seconds (averaging complete), then select STOP. The display should read
0.00 V
±
8.0 mV. Record the reading in the performance test record.
6. In the Channel menu, change the vertical sensitivity to 10 mV/Div. Select RUN - REPETITIVE and wait approximately 15 seconds (averaging complete), then select STOP. The display should read
0.00 V
±
800.0
µ
V. Record the reading in the performance test record.
DC Input Offset
Set the Channel Coupling field to 1M
Ω
/ DC or damage to the equipment will result.
7. Use the BNC-to-banana adapter to connect the BNC cable between the power supply and channel
1 input. Monitor the power supply using a voltmeter.
8. Use the following table for the next steps:
V/Div
1 V/Div
100 mV/Div
20 mV/Div
20 mV/Div
100 mV/Div
1 V/Div
Offset
−
35.00 V
−
10.00 V
−
2.00 V
+ 2.00 V
+ 10.00 V
+ 35.00 V
Supply
−
35.00 V
−
10.00 V
−
2.00 V
+ 2.00 V
+ 10.00 V
+ 35.00 V
−
35.4 V
−
10.1 V
−
2.02 V
+ 1.98 V
+ 9.90 V
+ 34.6 V
Min Max
−
34.6 V
−
9.90 V
−
1.98 V
+ 2.02 V
+ 10.1 V
+ 35.4 V
9. In the Channel menu set the V/Div range and offset per the first line of the table. Set the power supply per the first line of the table.
10. On the oscilloscope, select RUN - REPETITIVE and wait approximately 15 seconds (averaging complete), then STOP.
11. Read the voltage from the markers voltage field. It should be within the limits given in the table. Record the reading in the performance test record.
12. Repeat steps 7 through 10 for the remaining lines of the table using the V/Div range, offset, and supply voltages given in the table.
13. In the Channel menu select channel 2. In the Waveform Selection menu delete channel 1 and insert channel 2.
14. Repeat steps 1 through 12 for channel 2, replacing channel 1 with channel 2 where applicable.
3-10
HP 16532A - Performance Tests
3-11. BANDWIDTH
This test checks the bandwidth of the oscilloscope module.
Specification:
Bandwidth (dc coupled) dc to 250 MHz
Equipment Required:
The following equipment is required for this test. Procedures are based on the model or part number recommended.
Equipment Required
Signal Generator
Power Meter/Sensor
Power Splitter
Cable
Adapter
Critical Specification
1 - 250 MHz at approx 170 mVrms
1 - 250 MHz
±
3% accuracy
Outputs differ by < 0.15 dB
Type N (m to m) 24-inch
50
Ω
Type N (m) to BNC (m)
Recommended
Model/Part Number
HP 8656B
HP 436/8482A
HP 11667B
HP 11500B
HP 1250-0082
Procedure:
1. With the N cable, connect the signal generator to the power splitter input. Connect the power sensor to one output of the power splitter.
2. Using an N-to-BNC adapter, connect the other power splitter output to the Channel 1 input of the oscilloscope module.
3. In the Waveform Selection menu delete channel 2. If channel 1 is not inserted, insert channel 1.
Set the following parameters.
Menu
Channel
Trigger
Display
Input
Probe v/Div
Offset
Coupling s/Div
Mode
Level
Mode
Average #
Grid
Markers
Selection
1
1:1
80 mV
0 V
50
Ω
/ DC
200 ns
Edge
0 V
Average
32
On
Off
Setting
3-11
HP 16532A - Performance Tests
4. Set the signal generator for 1MHz at
−
2.4 dBm. On the oscilloscope, select RUN - REPETITIVE. The signal on the screen should be two cycles at three divisions amplitude.
5. After the measurement settles (averaging complete, about 15 seconds), select STOP. In the
Auto-Measure menu note the V p-p
reading. V
1MHz
= __________ mV.
6. Set the power meter Cal Factor % to the 1MHz value from the calibration chart on the power splitter.
Then press dB[REF] to set a 0 dB reference.
7. Change the signal generator to 250 MHz. Set power meter Cal Factor % to the 250 MHz value from the chart.
8. Adjust the signal generator amplitude for a power reading as close as possible to 0.0 dB[REL].
Reading = __________ dB.
9. Set the oscilloscope s/Div for 2 ns/Div and select RUN-REPETITIVE.
10. After the measurement settles (averaging complete), select STOP. In the Auto-Measure menu note the V p-p
reading. V
250MHz
= ________mV.
11. Calculate the response using the formula: response
( dB
) =
20 log
10
V
250 Mhz
V
1 MHz
=
20 log 10
(
_____
) =
______ dB
12. Correct the result from step 11 with any differences in the power meter from step 8. Observe signs.
For example:
Result from step 11 =
−
2.3 dB
Power meter reading =
−
0.2 dB[REL] then true response = (
−
2.3)
−
(
−
0.2)
−
2.1dB
(__________)
−
(__________) = ____________ dB
13. The result from step 12 should be
≤ −
3.0 dB. Record the result in the Performance Test Record.
14. In the Channel menu select channel 2. In the Waveform Selection menu delete channel 1 and insert channel 2.
15. Repeat steps 2 through 14 for channel 2, replacing channel 1 with channel 2 where applicable.
Note
Failure of the bandwidth test can be caused by a faulty attenuator or main assembly.
3-12
HP 16532A - Performance Tests
3-12. TIME MEASUREMENT ACCURACY
This test uses a precise frequency source to check the accuracy of time measurement functions.
Specification:
±
[(0.005% x
∆ t) + (2 x 10
−
6 x delay setting) + 150 ps]
Equipment Required:
The following equipment is required for this test. Procedures are based on the model or part number recommended.
Equipment Required
Signal Generator
Cable
Adapter
Critical Specification
200 MHz, timebase accuracy 0.25 ppm
50
Ω
BNC (m to m) 48-inch
50
Ω
Type N (m) to BNC (f)
Recommended
Model/Part Number
HP 8656B Opt. 001
HP 10503A
HP 1250-0780
Procedure:
1. Set the signal generator to 181.81818 MHz (5.5 ns period) and approximately 600 mV rms.
2. Use the N-to-BNC adapter to connect the cable to the signal generator. Connect the free end of the
BNC cable to the channel 1 input.
3. In the Waveform Selection menu delete channel 2. If channel 1 is not inserted, insert channel 1. Set the following parameters in the order given:
Menu
Channel
Trigger
Display
Input
Probe v/Div
Offset
Coupling s/Div
Mode
Source
Level
Slope
Mode
Grid
Selection
1
1:1
400 mV
0.0 V
50
Ω
/ DC
2.00 ns
Edge
Chan 1
0.0 V
Positive
Normal
On
Setting
3-13
HP 16532A - Performance Tests
4. Select the Markers Auto menu, and set the following parameters:
Menu Selection
Markers
Auto
X
O
→
→
Statistics
Setting
Set on chan 1 at Level 50%
Slope Positive
Occur 1
Set on chan 1 at Level 50%
Slope Positive
Occur 2
On
5. Select DONE. On the oscilloscope module select RUN - REPETITIVE. If the waveform is clipping, reduce the signal generator output voltage level until the waveform no longer clips. Allow the oscilloscope module to run approximately two minutes, then select STOP.
6. In the statistics field, check to see that the MEAN X-O field is approximately 5.500 ns. Check that both the MIN X-O and the MAX X-O are within 150 ps of the MEAN X-O. Record the results in the performance test record.
7. Select the Markers Auto menu, and set the following parameters:
Menu Selection
Markers
Auto
X
O
→
→
Statistics
Setting
Manual
Set on chan 1 at Level 50%
Slope Positive
Occur 1
On
8. Select DONE. In the Display menu, set MODE AVERAGE # 8 and DELAY 99.00 ns. Select RUN-RE-
PETITIVE and allow the oscilloscope module to run approximately two minutes. Select STOP.
9. In the statistics field, check to see that the MEAN X-O field is approximately 99.000 ns. Check that both the MIN X-O and the MAX X-O are within 155 ps of the MEAN X-O. Record the results in the performance test record.
Note
Failure of the time measurement accuracy test can be caused by a defective main assembly.
3-14
HP 16532A - Performance Tests
3-13. TRIGGER SENSITIVITY
This test checks channel triggers for sensitivity at rated bandwidth.
Specification: dc to 50 MHz : 0.063 x full scale (0.25 division)
50 to 250 MHz : 0.125 x full scale (0.5 division)
Equipment Required:
The following equipment is required for this test. Procedures are based on the model or part number recommended.
Equipment Required
Signal Generator
Cable
Adapter
Critical Specification
50 and 250 MHz, 30-80 mV rms
output
50
Ω
BNC 48-inch
50
Ω
Type N (m) to BNC (f)
Recommended
Model/Part Number
HP 8656B Opt. 001
HP 10503B
HP 1250-0780
Procedure:
1. In the Waveform Selection menu delete channel 2. If channel 1 is not inserted, insert channel 1. On the oscilloscope module set the following parameters in the order given:
Menu
Channel
Trigger
Display
Input
Probe v/Div
Offset
Coupling s/Div
Mode
Source
Level
Mode
Average #
Grid
Markers
Selection
1
1:1
400 mV
0 V
50
Ω
/ DC
20 ns
Edge
Chan 1
0 V
Average
8
On
Off
Setting
2. With an N-to-BNC adapter and BNC cable, connect the signal generator to channel 1 input.
3-15
HP 16532A - Performance Tests
3. Set the signal generator for a 50 MHz and 35 mV rms signal. On the oscilloscope module select
RUN - REPETITIVE.
4. In the Auto-Measure menu read V p_p
. Adjust the signal generator output level so that the signal is close to and not more than 100 mV p_p
.
5. In the Trigger menu adjust the trigger level for a stable display. The test passes if the trigger is stable. Record whether the trigger is stable or is not stable.
6. On the oscilloscope module select STOP. Set the signal generator frequency to 250 MHz and
70 mV rms signal level.
7. In the Channel menu change s/Div to 5 ns/Div and select RUN.
8. In the Auto-Measure menu read V p_p
. Adjust the signal generator output level so that the signal is close to and not more than 200 mV p_p
.
9. In the Trigger menu adjust the trigger level for a stable display. The test passes if the trigger is stable. Record whether the trigger is stable or is not stable in the performance test record.
10. On the oscilloscope module select STOP. In the Channel menu select channel 2. In the Trigger menu select Source channel 2. In the Waveform Selection menu delete channel 1 and insert channel 2.
11. Repeat steps 1 through 9 for channel 2, replacing channel 1 with channel 2 where applicable.
Note
Failure of the trigger sensitivity test can be caused by a defective main assembly.
3-16
HP 16532A - Performance Tests
Table 3-1. Performance Test Record
Hewlett-Packard
Model 16532A
Oscilloscope Module
Serial No.:
____________________
Recommended Performance
Verification Interval:
12 Months
Tested by ______________________________
Work Order No. _________________________
Date Tested ____________________________
Paragraph No.
3-6
3-7
3-8
3-9
Test
DC CAL OUTPUT
5.000Vdc
±
10mV
AC CAL OUTPUT
0.8Vp_p
±
10%
1000Hz
±
10%
Input Resistance
50
1M
Ω±
0.5
Ω
(49.5 to 50.5
Ω
)
Ω±
10K
Ω
(0.990 to
1.010M
Ω
)
Voltage Measurement Accuracy
Upper Limit Lower Limit
-13.7V
-3.43V
-1.37V
-137.0mV
143.0mV
1.43V
3.57V
14.3V
-14.3V
-3.57V
-1.43V
-143.0mV
137.0mV
1.37V
3.43V
13.7V
Results
Minimum Maximum Actual
4.990Vdc 5.010Vdc ____________Vdc
Minimum Maximum Actual
0.72Vp_p 0.88Vp_p ____________Vp_p
Minimum Maximum Actual
900Hz 1100Hz ____________Hz
Channel 1 Reading
50
Ω
@ 20mV/Div _____________
Ω
1M
Ω
@ 20mV/Div _____________
Ω
50
Ω
@ 200mV/Div _____________
Ω
1M
Ω
@ 200mV/Div _____________
Ω
50
Ω
@ 1V/Div _____________
Ω
1M
Ω
@ 1V/Div _____________
Ω
50
Ω
@ 4V/Div _____________
Ω
1M
Ω
@ 4V/Div _____________
Ω
Channel 2 Reading
50
Ω
@ 20mV/Div _____________
Ω
1M
Ω
@ 20mV/Div _____________
Ω
50
Ω
@ 200mV/Div _____________
Ω
1M
Ω
@ 200mV/Div _____________
Ω
50
Ω
@ 1V/Div _____________
Ω
1M
Ω
@ 1V/Div _____________
Ω
50
Ω
@ 4V/Div _____________
Ω
1M
Ω
@ 4V/Div _____________
Ω
Channel 1
Zero Input
__________
__________
__________
__________
__________
__________
__________
__________
Measured
__________
__________
__________
__________
__________
__________
__________
__________
Actual
__________V
__________V
__________V
__________mV
__________mV
__________V
__________V
__________V
3-17
HP 16532A - Performance Tests
Paragraph No.
3-9
3-10
3-11
Table 3-1. Performance Test Record (Continued)
Test
Voltage Measurement Accuracy
(continued)
Upper Limit Lower Limit
-13.7V
-3.43V
-1.37V
-137.0mV
143.0mV
1.43V
3.57V
14.3V
-14.3V
-3.57V
-1.43V
-143.0mV
137.0mV
1.37V
3.43V
13.7V
Channel 2
Zero Input
__________
__________
__________
__________
__________
__________
__________
__________
Results
Measured
__________
__________
__________
__________
__________
__________
__________
__________
Actual
__________V
__________V
__________V
__________mV
__________mV
__________V
__________V
__________V
Offset Accuracy
Zero-input Offset-Channel 1
V/Div Setting Reading Actual
4V/Div 0.00V
±
320.0mV ___________V
1V/Div 0.00V
±
80.0mV ___________V
100mV/Div 0.00V
±
8.0mV ___________V
10mV/Div 0.00V
±
800.0
µ
V ___________V
Offset Accuracy
DC Input Offset-Channel 1
Supply Min Max Actual
-35.00V -35.4V -34.6V -__________V
-10.00V -10.1V -9.90V -__________V
-2.00V -2.02V -1.98V -__________V
+2.00V +1.98V +2.02V +__________V
+10.00V +9.90V +10.1V +__________V
+35.00V +34.6V +35.4V +__________V
Offset Accuracy
Zero-input Offset-Channel 2
V/Div Setting Reading Actual
4V/Div 0.00V
±
320.0mV ___________V
1V/Div 0.00V
±
80.0mV ___________V
100mV/Div 0.00V
±
8.0mV ___________V
10mV/Div 0.00V
±
800.0
µ
V ___________V
Offset Accuracy
DC Input Offset-Channel 2
Supply Min Max Actual
-35.00V -35.4V -34.6V -__________V
-10.00V -10.1V -9.90V -__________V
-2.00V -2.02V -1.98V -__________V
+2.00V +1.98V +2.02V +__________V
+10.00V +9.90V +10.1V +__________V
+35.00V +34.6V +35.4V +__________V
Bandwidth
≤ −
3.0dB
Channel 1:
Actual Reading= ________ dB
Channel 2:
Actual Reading = ________ dB
3-18
HP 16532A - Performance Tests
Paragraph No.
3-12
3-13
Table 3-1. Performance Test Record (Continued)
Test
Time Measurement Accuracy
5.500ns
±
150ps
99.00ns
±
155ps
Trigger Sensitivity
Results
MEAN X-O __________ns
MIN X-O __________ns
MEAN X-O - MIN X-O = _________ps
MAX X-O __________ns
MAX X-O - MEAN X-O = _________ps
MEAN X-O __________ns
MIN X-O __________ns
MEAN X-O - MIN X-O = _________ps
MAX X-O __________ns
MAX X-O - MEAN X-O = _________ps
Channel 1:
Trigger Stable @ 100mVp_p Yes ____ No ____
Trigger Stable @ 200mVp_p Yes ____ No ____
Channel 2:
Trigger Stable @ 100mVp_p Yes ____ No ____
Trigger Stable @ 200mVp_p Yes ____ No ____
3-19
TABLE OF CONTENTS
Calibration
4-1. Introduction..........................................................................................................................4-1
4-2. Safety Requirements............................................................................................................4-1
4-3. Calibration Interval...............................................................................................................4-1
4-4. Maintaining System Integrity ...............................................................................................4-1
4-5. Calibration Protection Switch..............................................................................................4-1
4-6. Recommended Test Equipment .........................................................................................4-2
4-7. Instrument Warmup .............................................................................................................4-2
4-8. Calibration ............................................................................................................................4-2
4-9. Setting the PROTECT/UNPROTECT Switch to UNPROTECT ...........................................4-3
4-10. Equipment Required for Calibration...................................................................................4-4
4-11. Loading the Default Calibration Factors ............................................................................4-4
4-12. Self Cal Menu Calibrations .................................................................................................4-5
4-13. Setting the PROTECT/UNPROTECT Switch to PROTECT................................................4-8
4-14. Sealing the Mainframe ........................................................................................................4-8
HP 16532A - Calibration
SECTION IV
Calibration
4-1. Introduction
This section provides information on calibrating the HP 16532A 1 GSa/s Oscilloscope module.
Information is included about when to calibrate the module, the module warm up interval required prior to calibration, and how to do the actual calibration of the module. Also included are safety requirements to observe when working on the equipment, a list of recommended test equipment, and equipment setups for the different calibration procedures. Although the module has no hardware adjustments, it does contain a calibration PROTECT/UNPROTECT switch that you use before and after calibration. Actual calibration of the module is performed in software.
4-2. Safety Requirements
Read the safety summary at the front of this manual before any replacement, maintenance, or repair is performed.
Specific warnings, cautions, and instructions are placed wherever applicable throughout the manual. These must be observed during all phases of operation, service, and repair of the module. Failure to comply with them violates safety standards of design, manufacture, and intended use of this module. Hewlett-Packard assumes no liability for the failure of the customer to comply with these safety requirements.
• The oscilloscope software changes revision number and asterisks (* ) are on the calibration menu.
• Ambient temperature changes more than
10
°
C.
If you add or delete a card(s) from a calibrated multicard oscilloscope module, you must perform the Time Null calibration.
4-4. Maintaining System Integrity
Calibration factors depend on the mainframe, module, and module configuration all staying together. If oscilloscope modules are interchanged between mainframes, or the oscilloscope module slot configuration is changed, calibration must be repeated. In an environment in which the possibility of sharing oscilloscope modules exists, Hewlett-Packard recommends taking the following precautions in order to maintain system integrity:
With the system turned off, place the
PROTECT/UNPROTECT switch (all switches in a single or a chained, multiple oscilloscope module configuration) in the UNPROTECT position.
Boot the system and perform calibration for each module.
Turn off the system, place the
PROTECT/UNPROTECT switch (all switches in a single or a chained, multiple oscilloscope module configuration) in the PROTECT position, then seal the mainframe.
4-3. Calibration Interval
To maintain proper calibration, perform calibration at approximately six-month intervals, or every 1000 hours, when the instrument is being used under normal operating conditions. Immediate calibration is required if one of the following occurs:
• Oscilloscope module is installed, repaired, or replaced, or if the card slot configuration is changed.
4-5. Calibration Protection Switch
Each HP 16532A board stores the calibration factors for that board in non-volatile RAM on the board itself. The HP 16532A board contains a calibration protection (PROTECT/UNPROTECT) switch. With the switch in the PROTECT position, calibration values cannot be changed.
4-1
HP 16532A - Calibration
To run the scope calibration, the switch must be in the UNPROTECT position. When you have more than one HP 16532A board in a mainframe, this must be done for each one of the boards. After calibration is complete, all the switches must be set to the PROTECT position to save the calibration values. Also, all the modules or filler panels must be reinstalled in the mainframe, then a seal should be placed on the equipment such that the seal touches both the top-most module and the mainframe. This is done to assure that calibration integrity is maintained.
When ordered with a system, oscilloscope modules are calibrated at Hewlett-Packard as single units (not chained together). If you need to use the modules as one unit (chained together), perform the Time Null Calibration in section 4-12.
The calibration procedure sets the oscilloscope calibration factors and stores them in non-volatile
RAM on the board. The procedures use signals generated in the oscilloscope module itself to calibrate channel sensitivity, offset, and trigger parameters. The calibration factors set up the module each time the instrument is turned on.
4-6. Recommended Test Equipment
The only test equipment required for calibration are cables and adapters. These are listed in paragraph 4-10 and are called out in the applicable calibration procedure.
The calibration procedures in this section should be followed in their entirety and in the sequence listed in table 4-1. Though in this section the default calibrations are loaded before calibration, calibration can be done without loading them.
However, if they are loaded, they should be loaded before an actual calibration is done.
4-7. Instrument Warmup
Calibrate the oscilloscope module at its environmental ambient temperature, after a
30-minute warmup period of the
HP 16500A/16501A mainframe with the module installed.
4-8. Calibration
When calibrating the module, it must be installed in the mainframe, and the mainframe covers and filler panels must be in place. The module must be warmed up (refer to "Instrument Warmup" given earlier in this section).
This section provides software calibration procedures for the HP 16532A oscilloscope module. There are no hardware adjustments on the oscilloscope module. The calibration is performed without access to the interior of the instrument.
Sequence
Calibration procedures require the use of cables and adapters only. Non-service personnel can perform the calibration, however, to do the calibration, it is necessary to unprotect the current calibration values. This may not be allowed in some circumstances
−
check with your
Calibration Laboratory or Service Department before you attempt calibration. To perform calibration, follow the "Self Cal Menu Calibration" procedures given later in this section.
Table 4-1. Calibration Sequence
Calibration Paragraph
3
4
1
2
7
8
5
6
Set the PROTECT/UNPROTECT Switch to UNPROTECT
Default (Not required as part of the calibration procedures)
Self Cal Menu
Vertical Calibration (all channels)
Delay Calibration
−
Channel 1
Delay Calibration
−
Channel 2
Time Null Calibration (all channels)
Logic Trigger - Channel 1
External Trigger Null Calibration
Set the PROTECT/UNPROTECT Switch to PROTECT
4-9
4-11
4-12
4-12
4-12
4-12
4-12
4-12
4-13
4-2
4-9. Setting the PROTECT/UNPROTECT Switch to UNPROTECT
HP 16532A - Calibration
The effects of ELECTROSTATIC DISCHARGE can damage electronic components. Use grounded wrist straps and mats when you are performing any kind of service on this module.
The oscilloscope board contains a calibration PROTECT/UNPROTECT switch. This switch must be set to UNPROTECT before calibration can be done. Set the switch to UNPROTECT as follows:
1. Remove power from the instrument. Pull halfway out all of the modules located above the oscilloscope and remove the filler panels and cards located above the oscilloscope. See figure 4-1 for the removal sequence.
Figure 4-1. Endplate Overlap - Top Sequence
2. Remove the oscilloscope module and set the PROTECT/UNPROTECT switch to the UNPROTECT position. See Figure 4-2 to locate this switch.
Figure 4-2. Location of the PROTECT/UNPROTECT Switch
3. Repeat step 2 for every oscilloscope module if this is a multimodule calibration.
4. Reinstall the modules and filler panels in reverse order of steps 1 and 2.
4-3
HP 16532A - Calibration
4-10. Equipment Required for Calibration
The equipment listed below is required for calibration:
Cable (2)
Cable
Adapter
Adapter
50
Ω
BNC, 9-inch (equal length)
50
Ω
BNC - SMB
50
Ω
BNC tee (m)(f)(f)
50
Ω
BNC (m)(m)
4-11. Loading the Default Calibration Factors
HP 10502A
HP 16532-61601
HP 1250-0781
HP 1250-1236
Once the default calibration factors are loaded, all calibrations must be done.
This includes all of the calibrations in the Self Cal menu. The calibration must be performed in the exact sequence listed in table 4-1.
Note
The calibration PROTECT/UNPROTECT switch on the oscilloscope board must be set to UNPROTECT.
Load the default calibration factors as follows:
1. Touch the menu field (top row, second from the left), then select Calibration from the pop-up menu.
2. Touch the Mode field, then select Service Cal from the pop-up menu.
3. Touch the Procedure field, then select Default Values from the pop-up menu.
Note
After you touch the Start field, you can abort the calibration procedure either by touching the Cancel field, if it is available in the menu, or by touching either the
Mode or Procedure fields if the continue field is still displayed on the screen.
4. Touch the Start field and follow the instructions on the display. The display should now look similar to the display shown in figure 4-3.
4-4
Figure 4-3. Default Calibration Display
HP 16532A - Calibration
4-12. Self Cal Menu Calibrations
Note
After you touch the Start field, you can abort the calibration procedure either by touching the Cancel field, if it is available in the menu, or by touching either the
Mode or Procedure fields if the continue field is still displayed on the screen.
Messages will be displayed as each calibration routine is completed to indicate calibration has passed or failed. The resulting calibration factors are automatically stored to nonvolatile RAM at the conclusion of each calibration routine.
The Self Cal menu lets you calibrate vertical sensitivity (Vert Cal) for channels 1 and 2 individually or both channels on a board simultaneously if it is a stand alone scope. Also, the Self Cal menu lets you calibrate delay (Delay) for channel 1 and 2 separately, then Time Null for each channel separately (except channel 1 of the master board), and Ext Trig Null for the external trigger (ECL).
Calibrate Vert Cal of the Self Cal menu as follows:
1. Connect two BNC 50-
Ω
, 9-inch cables to the BNC tee adapter.
Note
When you touch Start, the instrument will prompt you to connect the cables to the appropriate locations on the rear panel of the module.
2. Touch the Mode field, then select Self Cal from the pop-up menu.
3. Touch the Procedure field, then select Vert Cal from the pop-up menu.
4. Touch the Channel field, then select a channel choice from the pop-up menu.
5. Touch the Start field and follow the instructions on the display.
6. After completion of vertical calibration remove the cables from the rear panel of the oscilloscope.
Calibrate Delay of the Self Cal menu as follows:
1. Obtain a BNC 50-
Ω
, 9-inch cable.
Note
When you touch Start, the instrument will prompt you to connect the cable to the appropriate location on the rear panel of the module.
2. Touch the Procedure field, then select Delay from the pop-up menu.
4-5
HP 16532A - Calibration
3. Touch the Channel field, then select S1 from the pop-up menu (where S is the slot letter where the oscilloscope module is installed in the mainframe).
4. Touch the Start field and follow the instructions on the display.
5. After completion of channel 1 delay calibration, remove the cable from the rear panel of the oscilloscope.
6. Repeat steps 3 through 5 for channel S2.
7. After completing all of the channel delay calibrations, remove the cable from the rear panel of the oscilloscope.
Calibrate the Time Null of the Self Cal menu as follows:
1. Connect two BNC 50-
Ω
, 9-inch cables to the BNC tee adapter.
Note
When you touch Start, the instrument will prompt you to connect the cables to the appropriate locations on the rear panel of the module.
2. Touch the Procedure field, then select Time Null from the pop-up menu.
3. Touch the Channel field, then select S2 from the pop-up menu (where S is the slot letter where the oscilloscope module is installed in the mainframe).
4. Touch the Start field and follow the instructions on the display.
5. Repeat steps 3 through 4 for each channel, replacing S2 with the appropriate channel number.
6. At the conclusion of the time null calibration, remove the cables from the rear of the module.
Calibrate the Logic Trigger of the Self Cal menu as follows:
1. Obtain one BNC 50
Ω
9-inch cable.
Note
When you touch Start, the instrument will prompt you to connect the cables to the appropriate locations on the rear panel of the module.
2. Touch the Procedure field, then select Logic Trigger from the pop-up menu.
3. Touch the Start field, then follow the instructions on the display.
4. At the conclusion of the Logic Trigger calibration, remove the cable from the rear of the module.
4-6
HP 16532A - Calibration
Calibrate the Ext Trig Null of the Self Cal menu as follows:
1. Connect the cables, BNC adapter, and BNC tee adapter together as shown in figure 4-4.
Note
When you touch Start, the instrument will prompt you to connect the cables to the appropriate locations on the rear panel of the module.
Figure 4-4. Ext Trig Null Calibration Connections
2. Touch the Procedure field, then select Ext Trig Null from the pop-up menu.
3. Touch the Start field and follow the instructions on the display.
4. At the conclusion of the ext trig null calibration, remove the cables from the rear of the module.
Messages will be displayed as each calibration routine is completed to indicate calibration has passed or failed. The resulting calibration factors are automatically stored to nonvolatile RAM at the conclusion of each calibration routine.
This is the conclusion of the calibration procedures. The display should now look similar figure 4-5.
4-7
HP 16532A - Calibration
Figure 4-5. End of Calibration Display
4-13. Setting the PROTECT/UNPROTECT Switch to PROTECT
The effects of ELECTROSTATIC DISCHARGE can damage electronic components. Use grounded wrist straps and mats when you are performing any kind of service on this module.
The PROTECT/UNPROTECT switch must be set to PROTECT after calibration is finished. Set the switch to PROTECT as follows:
1. Remove power from the instrument. Pull the modules located above the oscilloscope halfway out, then remove the filler panels and cards located above the oscilloscope. See figure 4-6 for the removal sequence.
Figure 4-6. Endplate Overlap - Top Sequence
2. Remove the oscilloscope module and set the PROTECT/UNPROTECT switch to the PROTECT position. See Figure 4-2 to locate this switch.
3. Repeat step 2 for each module calibrated.
4. Reinstall the modules and filler panels in reverse order of steps 1 and 2.
4-14. Sealing the Mainframe
After calibration has been completed, the mainframe should be sealed to protect the integrity of the system. Place a seal on the equipment such that the seal touches both the top-most module (or filler panel) and the mainframe.
4-8
TABLE OF CONTENTS
Replaceable Parts
5-1. Introduction .......................................................................................................................... 5-1
5-2. Abbreviations ....................................................................................................................... 5-1
5-3. Replaceable Parts List ......................................................................................................... 5-1
5-4. Ordering Information ........................................................................................................... 5-1
5-5. Exchange Assemblies ......................................................................................................... 5-1
5-6. Direct Mail Order System .................................................................................................... 5-2
HP 16532A - Replaceable Parts
SECTION V
Replaceable Parts
5-1. Introduction
The service policy for this instrument is replacement of defective assemblies. This service manual contains information for finding a defective assembly.
This section contains parts and ordering information for the HP 16532A Oscilloscope
Module. Table 5-1 lists the reference designator and abbreviations used throughout this manual.
Table 5-2 lists all replaceable parts by reference designator.
5-2. Abbreviations
Table 5-1 lists abbreviations used throughout the manual. In some cases two forms of the abbreviations are used, one in all capital letters, the other partially or not capitalized. This was done because the abbreviations in the parts list are always all capitals. However, in other parts of the manual other abbreviation forms are used with both lower and uppercase letters.
The information given for each part consists of the following:
• Complete reference designation.
• Hewlett-Packard part number.
• Total quantity.
• Description of part.
• Check digit.
The total quantity for each part is only given once at the first appearance of the part number in the list.
5-4. Ordering Information
To order a part listed in the replaceable parts table, quote the Hewlett-Packard part number, check digit, indicate the quantity required, and address the order to the nearest Hewlett-Packard office.
To order a part that is not listed in the replaceable parts table, include the instrument model number, instrument serial number, the description and function of the part, and number of parts required.
Address the order to the nearest Hewlett-Packard office.
5-3. Replaceable Parts List
Table 5-2 lists replaceable parts and is organized as follows:
• Exchange assemblies in alphanumerical order by reference designation.
• Electrical assemblies in alphanumerical order by reference designation.
• Chassis-mounted parts in alphanumerical order by reference designation.
5-5. Exchange Assemblies
Exchange assemblies are available when a repairable assembly is returned to
Hewlett-Packard. These assemblies have been set up on the "Exchange Assembly" program. This allows the customer to exchange the faulty assembly with one that has been repaired, calibrated, and performance verified by the factory. The cost is significantly less than that of a new assembly.
5-1
HP 16532A - Replaceable Parts
Exchange assemblies are listed in a separate section in the replaceable parts table. They have a part number in the form XXXXX-695XX (where the new parts would be XXXXX-665XX). Before ordering an "exchange assembly," check with your local parts or repair organization for procedures.
5-6. Direct Mail Order System
Within the USA, Hewlett-Packard can supply parts through direct mail order. The advantages are as follows:
• Direct ordering and shipment from
Hewlett-Packard Parts Center in Mountain
View, California.
• No maximum or minimum on any mail order
(there is a minimum order for parts ordered through local Hewlett-Packard offices when orders require billing and invoicing).
• Prepaid transportation (there is a small handling charge for each order).
• No invoices - to provide these advantages, check or money order must accompany each order.
Mail order forms and specific ordering information are available through your local Hewlett-Packard offices.
5-2
HP 16532A - Replaceable Parts
A
B
BT
C
CR
DL
DS
E
DWL
ECL
ELAS
EXT
F
FC
FD
FEM
FF
FL
FM
FR
FT
FW
FXD
GEN
GND
GP
GRAT
GRV
H
HD
HDND
HG
HGT
HLCL
HORIZ
HP
HP-IB
LG
LI
LK
LKWR
LS
LV
M
MACH
MAX
JKT
K
L
LB
LCH
LCL
LED
HR
HV
HZ
I/O
IC
ID
IN
INCL
INCAND
INP
INTEN
INTL
INV
JFET
A
A/D
AC
ADJ
AL
AMPL
ANLG
ANSI
CMR
CNDCT
CNTR
CON
CONT
CRT
CW
D
D/A
DAC
DARL
DAT
DBL
DBM
DC
DCDR
DEG
DEMUX
DET
DIA
DIP
DIV
DMA
DPDT
CAL
CC
CCW
CER
CFM
CH
CHAM
CHAN
CHAR
CM
CMOS
ASSY
ASTIG
ASYNCHRO
ATTEN
AWG
BAL
BCD
BD
BFR
BIN
BRDG
BSHG
BW
C
DRC
DRVR
= amperes
= analog-to-digital
= alternating current
= adjust(ment)
= aluminum
= amplifier
= analog
= American National
Standards Institute
= assembly
= astigmatism
= asynchronous
= attenuator
= American wire gauge
= balance
= binary-code decimal
= board
= buffer
= binary
= bridge
= bushing
= bandwidth
= ceramic;cermet
(resistor)
= calibrate;calibration
= carbon composition
= counterclockwise
= ceramic
= cubic feet/minute
= choke
= chamfered
= channel
= character
= centimeter
= complementary metal-
oxide-semiconductor
= common mode rejec-
tion
= conductor
= counter
= connector
= contact
= cathode-ray tube
= clockwise
= diameter
= digital-to-analog
= digital-to-analog
converter
= darlington
= data
= double
= decibel referenced
to 1mW
= direct current
= decoder
= degree
= demultiplexer
= detector
= diameter
= dual in-line package
= division
= direct memory access
= double-pole,
double-throw
= DAC refresh controller
= driver
= assembly
= fan;motor
= battery
= capacitor
= diode;diode thyristor;
varactor
= delay line
= annunciator;lamp;LED
= misc. electrical part
F
FL
H
J
L
MP
P
Table 5-1. Reference Designator and Abbreviations
REFERENCE DESIGNATOR
= fuse
= filter
= hardware
= electrical connector
(stationary portion);jack
= coil;inductor
= misc. mechanical part
= electrical connector
(moveable portion);plug
Q
R
RT
S
T
TB
TP
= transistor;SCR;
triode thyristor
= resistor
= thermistor
= switch;jumper
= transformer
= terminal board
= test point
U
V
VR
W
X
Y
= integrated circuit;
microcircuit
= electron tube; glow
lamp
= voltage regulator;
breakdown diode
= cable
= socket
= crystal unit(piezo-
electric or quartz)
ABBREVIATIONS
POLYE
POS
POT
POZI
PP
PPM
PRCN
PREAMP
PRGMBL
PRL
PROG
PSTN
PT
PW
PWR
R-S
RAM
RECT
RET
RF
RGLTR
RGTR
RK
RMS
PC
PCB
PD
PF
PI
PL
PLA
NUM
OBD
OCTL
OD
OP AMP
OSC
P
P/O
PLST
PNP
MFR
MICPROC
MINTR
MISC
MLD
MM
MO
MTG
MTLC
MUX
MW
N
NC
NMOS
NPN
NPRN
NRFR
NSR
= dowel
= emitter coupled logic
= elastomeric
= external
= farads;metal film
(resistor)
= carbon film/
composition
= feed
= female
= flip-flop
= flat
= foam;from
= front
= gain bandwidth
product
= full wave
= fixed
= generator
= ground(ed)
= general purpose
= graticule
= groove
= henries;high
= hardware
= hardened
= mercury
= height
= helical
= horizontal
= Hewlett-Packard
= Hewlett-Packard
Interface Bus
= hour(s)
= high voltage
= Hertz
= input/output
= integrated circuit
= inside diameter
= inch
= include(s)
= incandescent
= input
= intensity
= internal
= inverter
= junction field-
effect transistor
= jacket
= kilo(103)
= low
= pound
= latch
= local
= light-emitting
diode
= long
= lithium
= lock
= lockwasher
= low power Schottky
= low voltage
= mega(106);megohms;
meter(distance)
= machine
= maximum
= manufacturer
= microprocessor
= miniature
= miscellaneous
= molded
= millimeter
= metal oxide
= mounting
= metallic
= multiplexer
= milliwatt
= nano(10-9)
= no connection
= n-channel metal-
oxide-semiconductor
= negative-positive-
negative
= neoprene
= not recommended for
field replacement
= not separately
replaceable
= numeric
= order by description
= octal
= outside diameter
= operational amplifier
= oscillator
= plastic
= part of
= printed circuit
= printed circuit board
= power dissipation
= picofarads
= plug in
= plate(d)
= programmable logic
array
= plastic
= positive-negative-
positive
= polyester
= positive;position
= potentiometer
= pozidrive
= peak-to-peak
= parts per million
= precision
= preamplifier
= programmable
= parallel
= programmable
= position
= point
= potted wirewound
= power
= reset-set
= random-access
memory
= rectifier
= retainer
= radio frequency
= regulator
= register
= rack
= root-mean-square
RND
ROM
RPG
RX
S
SCR
SEC
SEG
SEL
SGL
SHF
SI
SIP
SKT
SL
SLDR
SLT
TPG
TPL
TRANS
TRIG
TRMR
TRN
TTL
TX
U
UL
TA
TBAX
TC
TD
THD
THK
THRU
TP
SOLD
SPCL
SQ
SREG
SRQ
STAT
STD
SYNCHRO
UNREG
VA
VAC
VAR
VCO
VDC
VERT
VF
VS
W
W/
W/O
WW
XSTR
ZNR oC oF oK
= Round
= read-only memory
= rotary pulse generator
= receiver
= Schottky-clamped;
seconds(time)
= screw;silicon
controlled rectifier
= second(time);secon-
dary
= segment
= selector
= single
= shift
= silicon
= single in-line
package
= skirt
= slide
= solder
= slot(ted)
= solenoid
= special
= square
= shift register
= service request
= static
= standard
= synchronous
= tantalum
= tubeaxial
= temperature coefficient
= time delay
= thread(ed)
= thick
= through
= test point
= tapping
= triple
= transformer
= trigger(ed)
= trimmer
= turn(s)
= transistor-transistor
= transmitter
= micro(10-6)
= Underwriters
Laboratory
= unregulated
= voltampere
= volt,ac
= variable
= voltage-controlled
oscillator
= volt,dc
= vertical
= voltage,filtered
= versus
= watts
= with
= without
= wirewound
= transistor
= zener
= degree Celsius
(Centigrade)
= degree Fahrenheit
= degree Kelvin
5-3
HP 16532A - Replaceable Parts
5-4
Figure 5-1. Parts Identification
HP 16532A - Replaceable Parts
A1
A2
H1
H2
H3
H4
H5
H6
H7
MP1
MP2
MP3
W1
W2
Reference HP Part C Qty
Designator Number D
Table 5-2. Replaceable Parts List
Description
16532-69501 2 1
16532-66501 6 1
16532-63401 9 2
16500-22401 5 2
0510-0684
1250-2075
9
5
2
4
_________
_________
0515-0430
0515-1246
2
2
3
1
3
4
16532-40501 2 1
16500-29101 6 1
16532-94301 3 1
16532-61601 7 1
16532-61602 8 1
Board Exchange Assembly
(does not include attenuator assemblies)
Board Assembly
(does not include attenuator assemblies)
Attenuator Assembly
Endplate Thumbscrew
Thumbscrew Retaining Ring
Nut, Hex, 1/2-inch, RF Connector
Washer, RF Connector (Included with W2)
Endplate Screw
Attenuator Retainer Screw
Rear Panel
Ground Spring
ID Label
Master-Slave Trigger Cable
External Trigger Cable
Mfr
Code
Mfr Part
Number
5-5
TABLE OF CONTENTS
Service
6-1. Introduction ......................................................................................................................... 6-1
6-2. Safety Requirements ........................................................................................................... 6-1
6-3. Recommended Test Equipment ........................................................................................ 6-1
6-4. Module Block Diagram and Theory of Operation............................................................... 6-1
6-5. Attenuator Theory of Operation ......................................................................................... 6-1
6-6. Main Assembly Theory of Operation .................................................................................. 6-1
6-7. Self-Tests ............................................................................................................................. 6-4
6-8. Troubleshooting .................................................................................................................. 6-10
6-9. Oscilloscope Module Removal and Replacement ............................................................ 6-18
6-10. Trigger Cable Removal and Replacement ......................................................................... 6-20
6-11. Attenuator Removal and Replacement .............................................................................. 6-21
HP 16532A - Service
SECTION VI
Service
6-1. Introduction
The service policy for this instrument is replacement of defective assemblies. This service manual contains information for finding a defective assembly. This section contains information for servicing the HP 16532A Oscilloscope Module.
Included are a block level theory and procedures for self diagnostics and troubleshooting. If the module or a cable is determined faulty, procedures are provided for module and cable replacement.
6-2. Safety Requirements
Specific warnings, cautions, and instructions are placed wherever applicable throughout the manual. These must be observed during all phases of operation, service, and repair of the module. Failure to comply with them violates safety standards of design, manufacture, and intended use of this module. Hewlett-Packard assumes no liability for the failure of the customer to comply with these safety requirements.
6-3. Recommended Test Equipment
Table 1-3 lists recommended test equipment. Any equipment that satisfies the critical specification given in the table may be substituted for the recommended models.
6-4. Module Block Diagram and Theory of Operation
The following paragraphs contain block level theory of operation. This theory is not intended for component level troubleshooting, rather it is to be used to help isolate a module failure to card level.
The HP 16532A Oscilloscope Module is contained on one board. It runs at a 1 GSa/s digitizing rate, with a 250 MHz single-shot (real-time) bandwidth,
8000 samples per measurement per channel, with
2-channel simultaneous acquisition which is expandable to up to 8 channels. See figure 6-1.
6-5. Attenuator Theory of Operation
The channel signals are conditioned by the attenuator/preamps, thick film hybrids containing passive attenuators, impedance converters, and a programmable amplifier. The channel sensitivity defaults to the standard 1-2-4 sequence (other sensitivities can be set also). However, the firmware uses passive attenuation of 1, 5, 25, and
125, with the programmable preamp, to cover the entire sensitivity range.
The input has a selectable 1 M
Ω
input impedance with ac or dc coupling or a 50
Ω
input impedance with dc coupling. Compensation for the passive attenuators is laser trimmed and is not adjustable.
After the passive attenuators, the signal is split into high-frequency and low-frequency components.
Low frequency components are amplified on the main assembly, where they are combined with the offset voltage. The ac coupling is implemented in the low frequency amplifier.
The high- and low-frequency components of the signal are recombined and applied to the input
FET of the preamp. The FET provides a high input impedance for the preamp. The programmable preamp adjusts the gain to suit the required sensitivity and provides the output signal to the main assembly. The output signal is then sent to both the trigger circuitry and ADC.
6-6. Main Assembly Theory of
Operation
Acquisition
The acquisition circuitry provides the sampling, digitizing, and storing of the signals from the channel attenuators. The channels are identical.
The external trigger (ECL) input cannot be displayed. Trigger signals from each channel and the external triggers synchronize acquisition through the time base circuitry. A 100 MHz oscillator and a time base provide system timing and sample clocking. A voltage-controlled oscillator (VCO), frequency divider, and digital
6-1
HP 16532A - Service
Figure 6-1. HP 16532A Oscilloscope Block Diagram phase detector provide the sample clock for higher sample rates. After conditioning and sampling, the signals are digitized, then stored in a hybrid IC containing a FISO (fast in, slow out) memory.
ADC. The eight-bit ADC digitizes the channel signal. Digitization is done by comparators in a flash converter. The sample clock latches the digitized value of the input to save it so that it can be sent to memory.
FISO Memory. 8000 samples of the FISO (fast in, slow out) memory are used per measurement per channel. Memory positions are not addressed directly. The configuration is a ring which loops continuously as it is clocked. Memory position is tracked by counting clocks. The clocking rate is the same as the ADC, however the clock frequency is half that of the ADC since the FISO clocks on both transitions of the clock period.
Data is buffered onto the CPU data bus for processing.
6-2
HP 16532A - Service
Triggering. There are two main trigger circuits that trigger four trigger sources. The two trigger circuits are the analog trigger and the logic trigger.
The analog trigger IC operates as a multichannel
Schmidt trigger/comparator. A trigger signal (a copy of the analog input signal) from each of the inputs is directed to the analog trigger IC inputs.
The trigger signal is continuously compared with the trigger reference level selected by the user.
Once the trigger condition is met, the trigger TRUE signal is fed to the logic trigger, which begins the acquisition and store functions by way of the time base.
The four trigger sources are Channel 1, Channel 2,
Intermodule Bus (IMB), and external BNC. The operation of the input channels was discussed previously. The IMB trigger signal is sent directly to the logic trigger. External triggering is provided by the BNC input of the 16500A Logic Analysis
System mainframe.
Time Base. The time base provides the sample clocks and timing necessary for data acquisition.
It consists of the 100 MHz reference oscillator and time base hybrid.
The 100 MHz reference oscillator provides the base sample frequency.
The time base hybrid has programmable dividers to provide the rest of the sample frequencies appropriate for the time range selected. The time base uses the time-stretched output of the fine interpolator to time-reference the sampling to the trigger point. The time base has counters to control how much data is taken before (pre-trigger data) and after (post-trigger data) the trigger event.
After the desired number of pre-trigger samples has occurred, the Time base hybrid sends a signal to the Logic Trigger (trigger arm) indicating it is ready for the trigger event. When the trigger condition is satisfied, the Logic Trigger sends a signal back to the time base hybrid. The time base hybrid then starts the post-trigger delay counter.
When the countdown reaches zero, the sample clocks are stopped and the CPU is signaled that the acquisition is complete. The Fine Interpolator is a dual-slope integrator that acts as a time-interval stretcher. When the logic trigger receives a signal that meets the programmed triggering requirements, it signals the time base.
The time base then sends a pulse to the fine interpolator. The pulse is equal in width to the time between the trigger and the next sample clock.
The fine interpolator stretches this time by a factor of approximately 500. Meanwhile, the time base hybrid runs a counter with a clock derived from the sample rate oscillator. When the interpolator indicates the stretch is complete, the counter is stopped. The count represents, with much higher accuracy, the time between the trigger and the first sample clock. The count is stored and used to place the recently acquired data in relationship with previous data.
AC Cal. The AC Cal is a multiplexer circuit that can provide several signals to the Probe
Compensation/AC Calibrator output on the rear panel. The signal provided depends on the mode of the instrument. It can be either a probe compensation signal, a pulse representing the trigger event, signals used for self-calibration, or the 100 MHz reference oscillator when sample period is 1 ns.
DC Cal. The DC Cal output, a rear panel signal, is used for self-calibration. It is one output from the
16-channel DAC.
Digital Interface. The Digital Interface provides control and interface between the system control and digital functions in the acquisition circuitry.
Analog Interface. The Analog Interface provides control of analog functions in the acquisition circuitry. It is primarily a 16-channel DAC with an accurate reference and filters on the outputs. It controls channel offsets and trigger levels, and provides the DC Cal output.
6-3
HP 16532A - Service
6-7. Self-Tests
The self-tests for the HP 16532A oscilloscope module will identify the improper operation of major functional areas in the module. They are not intended for component level diagnostics.
If there are multiple oscilloscope modules, the user must select the specific oscilloscope module to be tested at the main test system menu. Multiple HP 16532A modules that are connected as one unit are recognized as a single unit in the test menu. The external trigger cables connected to the ECL IN and ECL
OUT do not need to be removed in order to do the self-test for each board. All other cable connections must be removed.
All self-tests can be run without access to the interior of the instrument. If a self-test fails, refer to section 6-8, "Troubleshooting."
Self-Test Access Procedure:
1. Disconnect all oscilloscope inputs and turn on the power switch.
2. From the startup screen, shown in figure 6-2, touch the Configuration field. Then, touch Test in the pop-up menu.
Figure 6-2. Startup Screen
6-4
HP 16532A - Service
3. Install the Composite PV Test software disk in the inactive disk drive.
4. Touch the box labeled Touch box to Load Test System. See figure 6-3.
Note
When the self-test software is loaded, the user operating system is overwritten in the mainframe system RAM.
Figure 6-3. Load Test System Screen
5. From the test system screen in figure 6-4, touch Test System, then touch 1 GSa/s SCOPE D. (For the example shown in figure 6-4, the oscilloscope module is in slot D. If multiple oscilloscope modules are present, select the one to be tested.)
Figure 6-4. Test System Screen
6-5
HP 16532A - Service
Figure 6-5 shows the selection menu for choosing to perform functional tests. The status of the functional tests is given.
Figure 6-5. Functional Tests Screen
6. Touch the Functional Tests field.
Figure 6-6 is the main self-test menu. Self-tests can be either run individually by touching a specific test field, or run automatically through all tests one time by touching the All Tests field. When either All Tests or individual tests are run, the status will change to either PASSED or FAILED. If you do all tests by touching the All Tests field, skip to step 11. The procedures given in steps 7 through 12 use the A/D Test as an example of how individual tests are done.
Figure 6-6. Main Test Menu
6-6
HP 16532A - Service
7. Touch the A/D Test field.
The individual test run screens, see figure 6-7, give the test name, a brief description of the test, number of test runs, and the number of test failures.
Figure 6-7. A/D Test Run Screen
8. Touch Run, then drag your finger to Single or Repetitive.
During the time a Single run or a Repetitive run is executing, the Run field displays Stop.
9. To stop a Repetitive run, touch Stop. See figure 6-8.
Figure 6-8. Stop Field
6-7
HP 16532A - Service
10. Touch Done to exit the A/D test.
11. To exit the self tests, touch the following fields in the lettered sequence below: a. 1GSa/s Scope b. Test System c. Configuration d. Exit Test
12. Touch the box that reads "Touch box to Exit Test System". See figure 6-9.
Figure 6-9. Exit Test System Screen
6-8
HP 16532A - Service
Self-Test Descriptions
The following self-tests check the major components of the HP 16532A oscilloscope module as well as all associated circuitry. When the self-tests have all been completed with a "PASS" status, the major data and control pipelines in the HP 16532A oscilloscope module are functioning properly.
Data Memory Test
This test verifies the correct operation of the FISO (fast-in/slow-out) data memory on each board. Test patterns are written into the memory and then read and compared with known values.
Timebase Test
The pre-trigger and post-trigger delay modes are first tested by programming a predetermined time interval in the trigger counters. At the end of the time intervals, the arm, trigger, and run status bits are read and compared with known values. The coarse and fine interpolators are then checked by reading the values of the interpolator counters after a simulated acquisition. The counter values are then compared with a known value. Finally, the sample clock is checked by programming a sample clock frequency and then reading the status of the clock to detect when one clock period has elapsed. The clock period time interval is then compared with a known value.
A/D Test
This test verifies the correct operation of the A/D convertor on each board. A check of the trigger in
Trigger Immediate mode is first made. The A/D convertors are then exercised by setting the reference voltage and channel offset such that a simulated acquisition obtains data in the extremes and middle of the quantization range of the A/D convertor. After each simulated acquisition, the data is compared with known values.
D/A Test
This test verifies the correct operation of the D/A convertor on each board. Both the offset and trigger level D/A convertors for each channel are set to a reference level and then changed. The logic trigger IC is programmed to detect the changes. The detection of a correct trigger indicates that the D/A convertor is operating normally.
Trigger Test
This test verifies the correct operation of the trigger components on each board. First, the logic trigger memory is checked by writing and then reading known patterns. The logic qualifiers, logic trigger output, and trigger holdoff are checked.
IMB Test
This test verifies the correct operation of the oscilloscope card interface to the intermodule bus.
All Tests
This will automatically execute each test, one at a time, until all tests are done.
6-9
HP 16532A - Service
6-8. Troubleshooting
The troubleshooting flow chart given in figure 6-10 will isolate module failures to card level only. It is not intended for component level troubleshooting. If self-tests indicate a failure, begin at the Start of the troubleshooting flow chart. When a specific test fails, you will be instructed to replace a faulty card or you will be referred to other flow charts for the isolation of the faulty card or cable.
The effects of ELECTROSTATIC DISCHARGE can damage electronic components. Grounded wriststraps and mats should be used when you perform any kind of service to this instrument or the cards in it.
Troubleshooting Aids
The Calibration menu contains the Service Cal mode that can help you troubleshoot the oscilloscope module. When the problem is found and corrected, run all of the calibration procedures given in
Section IV, "Calibration."
If you suspect a fault in the system, you can load the system default values to see if the NVRAM parameters have possibly been corrupted. The procedure for loading the default values is given in Section IV under the heading "Loading the Default Calibration Factors."
If loading the default values does not fix the problem, you may want to verify that the system clock is working correctly. The procedure to verify the system clock is given below.
If none of these aids find the problem, go to the troubleshooting flow chart to find the problem. As stated above, when the problem is found and corrected, run all of the calibration procedures given in Section IV
"Calibration."
6-10
HP 16532A - Service
Verifying the System Clock
This procedure is provided as a troubleshooting aid to verify that the 100 MHz System Clock oscillator is functioning properly. The 100 MHz oscillator is the primary time reference used for all of the acquisition clock periods. The test limits provided are the test limits for the 100 MHz oscillator only. If the readings are outside of the test limits, the oscilloscope module will not operate properly.
Equipment Required
Frequency Counter
Oscilloscope
Cable
Critical Specification
> 100 MHz, 50
Ω
Bandwidth
≥
100 MHz, Real Time
50
Ω
BNC (m to m) 48-inch
Recommended
Model/Part Number
HP 5315A Option 003
HP 54502A
HP 10503A
Procedure:
1. Connect the BNC cable from the AC CAL OUTPUT connector on the rear panel of the HP 16532A oscilloscope module to the input of the frequency counter.
2. On the HP 16532A oscilloscope module, set the following parameters, in the order given.
Menu
Channel
Trigger
Calibration
Selection s/Div
Auto-Trig
Mode
Procedure
Signal
Setting
1 ns
Off
Service Cal
Osc Out
Sample Clk
3. Select RUN-REPETITIVE. Observe the frequency displayed on the frequency counter and verify that it is 100 MHz
±
0.005%.
4. Disconnect the BNC cable from the frequency counter and connect the cable to the input of the
HP 54502A Oscilloscope.
6-11
HP 16532A - Service
5. On the HP 54502A Oscilloscope, set the following parameters, in the order given.
Menu
Channel
Timebase
Trigger
Selection
Probe
V/Div
Offset
Coupling s/Div
Acquisition
Mode
Level
Slope
Setting
1.000:1
200 mV
0.0 V
DC
10 ns
Realtime
Edge
−
200 mV
Falling
6. In the HP 16532A Oscilloscope trigger menu, select Auto-Trig On. Verify that the 100 MHz System
Clock is gated.
7. Select STOP on both the HP 54502A and HP 16532A oscilloscopes.
8. Disconnect the cable from both oscilloscopes.
6-12
Troubleshooting Sheet 1
HP 16532A - Service
H
O
O
L
E
S
T
R
O
U
B
T
I
N
G
5
3
2
A
H
P
1
6
S
H
E
E
T
1
Figure 6-10. Troubleshooting Flowchart
6-13
H
O
O
L
E
S
T
I
N
G
T
R
O
U
B
S
H
E
E
T
5
3
2
A
H
P
1
6
2
HP 16532A - Service
Troubleshooting Sheet 2
Figure 6-10. Troubleshooting Flowchart
6-14
Troubleshooting Sheet 3
HP 16532A - Service
Figure 6-10. Troubleshooting Flowchart
S
H
E
E
T
3
O
O
T
E
S
H
I
N
G
U
B
L
T
R
O
H
P
1
6
5
3
2
A
6-15
E
S
H
U
B
L
T
R
O
I
N
G
O
O
T
H
P
1
6
5
3
2
A
S
H
E
E
T
4
HP 16532A - Service
Troubleshooting Sheet 4
Figure 6-10. Troubleshooting Flowchart
6-16
Troubleshooting Sheet 5
HP 16532A - Service
Figure 6-10. Troubleshooting Flowchart
H
O
O
L
E
S
T
R
O
U
B
T
I
N
G
5
3
2
A
H
P
1
6
S
H
E
E
T
5
6-17
HP 16532A - Service
6-9. Oscilloscope Module Removal and Replacement
The effects of ELECTROSTATIC DISCHARGE can damage electronic components. Use grounded wrist straps and mats when you are performing any kind of service on this module.
When you need to remove and replace a module, test the module first to determine the faulty part. If either of the trigger cables are faulty, remove the faulty cable and replace it with a new one. Refer to paragraph 6-10 for procedures to remove and replace a faulty cable. If either of the attenuators are faulty, remove the faulty attenuator and replace it with a new one. Refer to paragraph 6-11 for procedures to remove and replace a faulty attenuator.
Oscilloscope Module Removal Procedure:
Do not install, remove or replace the module in the instrument unless the instrument power is turned off.
1. Turn off the instrument power switch, unplug the power cord, and disconnect any input or output connections.
2. Starting from the top, loosen the thumb screws on the filler panels and the cards. See figure 6-11.
Figure 6-11. Endplate Overlap - Top Sequence
3. Starting from the top, pull out the cards and the filler panels halfway.
4. Pull out the oscilloscope module completely.
6-18
HP 16532A - Service
5. If a trigger cable was determined to be faulty, replace the faulty trigger cable on the oscilloscope module (refer to paragraph 6-10). Otherwise, skip to step 6.
6. If an attenuator was determined to be faulty, replace the faulty attenuator on the oscilloscope module (refer to paragraph 6-11). Otherwise, skip to step 7.
7. If the oscilloscope module was determined to be faulty, remove the attenuators from the oscilloscope module (refer to paragraph 6-11).
8. Acquire a replacement oscilloscope module and install the attenuators on the replacement oscilloscope module (refer to paragraph 6-11).
9. Reinstall the repaired (or replacement) oscilloscope module in the mainframe (refer to the next procedure "Oscilloscope Module Replacement Procedure").
Oscilloscope Module Replacement Procedure
1. If you have not already done so, pull the installed boards half way out of the card slots (refer to steps 1 through 3 of the preceding procedure "Oscilloscope Module Removal").
2. Determine which modules you want to install in the mainframe, and the slots where you want them installed.
3. Starting with the bottom slot, slide each module to be installed approximately halfway into the mainframe card slot.
4. Firmly seat the bottom card into the backplane connector. Keep applying pressure to the center of the card endplate while tightening the thumb screws finger tight.
5. Repeat step 4 for all cards and filler panels in a bottom-to-top order. See figure 6-12.
Figure 6-12. Endplate Overlap - Bottom Sequence
6. Any filler panels that are not used should be kept for future use. Filler panels must be installed in all unused card slots for correct air circulation.
6-19
HP 16532A - Service
6-10. Trigger Cable Removal and Replacement
The effects of ELECTROSTATIC DISCHARGE can damage electronic components. Use grounded wriststraps and mats when performing any kind of service to this instrument or the cards in it.
Trigger Cable Removal Procedure:
1. Follow the procedures given in paragraph 6-9 "Oscilloscope Module Removal and Replacement" to remove the HP 16532A Oscilloscope Module to be serviced.
2. Remove the hex nut (H4) and the washer (H5) from the SMB connector on the cable to be removed.
See figure 6-13.
6-20
Figure 6-13. Oscilloscope Module Exploded View
HP 16532A - Service
3. Pull the straight SMB connector on the cable through the hole in the rear panel.
4. Pull the right-angle SMB connector on the cable away from its mating connector on the board.
5. Install a new cable (refer to the next paragraph).
Trigger Cable Replacement Procedure:
1. Install the straight SMB connector on the cable through the hole in the rear panel. See figure 6-13.
2. Attach the washer (H5) and the hex nut (H4) to the straight SMB connector on the cable. Tighten the hex nut.
3. Connect the right-angle SMB connector on the cable to its mating connector on the board.
6-11. Attenuator Removal and Replacement
The effects of ELECTROSTATIC DISCHARGE can damage electronic components. Use grounded wriststraps and mats when performing any kind of service to this instrument or the cards in it.
Attenuator Removal Procedure:
1. Follow the procedures given in paragraph 6-9 "Oscilloscope Module Removal and Replacement" to remove the HP 16532A Oscilloscope Module to be serviced.
2. Remove the four hex nuts (H3) from the BNC connectors on the rear panel (MP1). See figure 6-13.
3. Remove the two hex nuts (H4) and the two washers (H5) from the SMB connectors on the rear panel.
4. Remove the three end plate screws (H6) holding the rear panel (MP1) and the ground spring (MP2) to the board assembly (A1).
5. Pull the rear panel (MP1) and the ground spring (MP2) from the board assembly (A1). You may need to loosen the attenuator retainer screws (H7) before removing the ground spring.
6. Remove the two attenuator retainer screws (H7) holding the attenuator assembly (A2) to the board assembly (A1).
7. Gently pull the attenuator assembly (A2) straight up from the board assembly (A1) being careful not to damage the connector and the components beneath the attenuator assembly.
8. Install a new attenuator assembly (refer to the next paragraph "Attenuator Replacement Procedure").
6-21
HP 16532A - Service
Attenuator Replacement Procedure: a. Gently push the attenuator assembly (A2) straight down on the board assembly (A1) being careful not to damage the connector and the components beneath the attenuator assembly. See figure
6-13.
b. Attach the attenuator assembly (A2) to the board assembly (A1) with the two attenuator retainer screws (H7).
c. Assemble the rear panel (MP1) and the ground spring (MP2) to the board assembly (A1) and attach them with the three end plate screws (H6). You may need to loosen the attenuator retainer screws
(H7) before assembling the ground spring to the board assembly, then tighten the attenuator screws (H7) when the assembly is finished.
d. Attach the SMB connectors to the rear panel (MP1) with two hex nuts (H4) and two washers (H5).
e. Attach the BNC connectors to the rear panel (MP1) with four hex nuts (H3).
Note
Tighten the hex nuts down so that they will not interfere with the installation of the board above the oscilloscope module. One of the flat surfaces on the outside edge of the nut should be parallel with the top edge of the rear panel.
6-22
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 20 1-1. Introduction
- 20 1-2. Module Covered by This Manual
- 20 1-3. Safety Requirements
- 20 1-4. Product Description
- 20 1-5. Accessories Supplied
- 20 1-6. Accessories Available
- 20 1-7. Specifications
- 20 1-8. Operating Characteristics
- 20 1-9. Recommended Test Equipment
- 21 2-1. Introduction
- 21 2-2. Initial Inspection
- 21 2-3. Preparation for Use
- 21 2-4. Power Requirements
- 21 2-5. Safety Requirements
- 21 2-6. Module Installation
- 21 2-7. Operating Environment
- 21 2-8. Storage
- 21 2-9. Packaging
- 21 2-10. Tagging for Service
- 22 3-1. Introduction
- 22 3-2. Recommended Test Equipment
- 22 3-3. Test Record
- 22 3-4. Performance Test Interval
- 22 3-5. Performance Test Procedures
- 22 3-6. DC CAL OUTPUT Test
- 22 3-7. AC CAL OUTPUT Test
- 22 3-8. Input Resistance
- 22 3-9. Voltage Measurement Accuracy
- 22 3-10. Offset Accuracy
- 22 3-11. Bandwidth
- 22 3-12. Time Measurement Accuracy
- 22 3-13. Trigger Sensitivity
- 42 4-1. Introduction
- 42 4-2. Safety Requirements
- 42 4-3. Calibration Interval
- 42 4-4. Maintaining System Integrity
- 42 4-5. Calibration Protection Switch
- 42 4-6. Recommended Test Equipment
- 42 4-7. Instrument Warmup
- 42 4-8. Calibration
- 42 4-9. Setting the PROTECT/UNPROTECT Switch to UNPROTECT
- 42 4-10. Equipment Required for Calibration
- 42 4-11. Loading the Default Calibration Factors
- 42 4-12. Self Cal Menu Calibrations
- 42 4-13. Setting the PROTECT/UNPROTECT Switch to PROTECT
- 42 4-14. Sealing the Mainframe
- 43 5-1. Introduction
- 43 5-2. Abbreviations
- 43 5-3. Replaceable Parts List
- 43 5-4. Ordering Information
- 43 5-5. Exchange Assemblies
- 43 5-6. Direct Mail Order System
- 44 6-1. Introduction
- 44 6-2. Safety Requirements
- 44 6-3. Recommended Test Equipment
- 44 6-4. Module Block Diagram and Theory of Operation
- 44 6-5. Attenuator Theory of Operation
- 44 6-6. Main Assembly Theory of Operation
- 44 6-7. Self-Tests
- 44 6-8. Troubleshooting
- 44 6-9. Oscilloscope Module Removal and Replacement
- 44 6-10. Trigger Cable Removal and Replacement
- 44 6-11. Attenuator Removal and Replacement