OPMI
®
VISU 200 BrightFlex™
Surgical Microscope
Instructions for use
G-30-1439-en
Issue 3.0
Printed on 21. 11. 2002
Contents
G-30-1439-en
– VISU 200 BrightFlex™ surgical microscope
– Notes on installation and use
– When using a fundus imaging system (e.g. BIOM II)
– Phototoxic retinal injury in eye surgery
VISU 200 BrightFlex™ Surgical Microscope
– Controls, displays, connections
– Tube and eyepieces for main microscope
– Tube and eyepieces for assistant's microscope
– Mounting the binocular tubes, eyepieces and the objective lens 46
– Setting the microscope tilt to angles greater than 15°
– Adjusting the surgical microscope
– When using a fundus imaging system (e.g. BIOM II)
OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Maintenance / Further information
– Magnifications / Fields of view
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Functions at a glance
Functions at a glance
VISU 200 BrightFlex™ surgical microscope
5
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
6
Functions at a glance
3
4
1
2
5
VISU 200 BrightFlex™ surgical microscope
1 Resetting the X-Y coupling, focus and zoom to their initial positions Page 30
2 Adjusting the tilt of the surgical microscope Page 50
3 Setting the interpupillary distance
4 Adjusting the eyecups
Page 40
Page 40
5 Setting your prescription Page 40
6 Display of the magnification factor of the zoom system Page 32
7 Setting the red reflex illumination (+2° and -2°) Page 28
8 Releasing the magnetic brakes of the suspension system Page 32
9 Changing the magnification on assistant's microscope Page 34
10 Focusing of assistant's microscope Page 34
11 Setting the 6° illumination
12 Arrows indicating the focusing range
Page 36
Pages
30, 56
Page 34 13 Locking screw for assistant's microscope (vertical)
14 Selecting light stops
15 Connecting the light guide
16 Locking screw for assistant's microscope (horizontal)
Page 36
Page 36
Page 36
6 7 8 9 10 5 4 11 12 8 10 9 13 14 15 16
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Functions at a glance
7
Key to symbols
Different symbols used in this user's manual draw your attention to safety aspects and useful tips. The symbols are explained in the following.
Warning!
The
warning triangle
indicates potential sources of danger which may constitute a risk of injury for the user or a health hazard.
Caution:
The
square
indicates situations which may lead to malfunction, defects, collision or damage of the instrument.
Note:
The
hand
indicates hints on the use of the instrument or other tips for the user.
OPMI
®
OPMI
® is a registered trademark of Carl Zeiss.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
8
Functions at a glance
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Safety
Safety
When using a fundus imaging system (e.g. BIOM II)
Phototoxic retinal injury in eye surgery
9
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
10
Safety
The instrument described in this manual has been developed and tested in accordance with Carl Zeiss safety standards and with national and international regulations. A high degree of instrument safety is thus ensured.
We would like to inform you on the safety aspects involved in operating the instrument. This chapter contains a summary of the most important precautions to be observed.
Further safety notes are also contained in other parts of this user's manual; they are marked with a warning triangle containing an exclamation mark as shown here. Please pay special attention to these safety notes.
Safety is only ensured when this instrument is operated properly. Please read through this manual carefully before turning the instrument on. Also read through the user's manuals of the other equipment used with this instrument. You may obtain further information from our service organization or authorized representatives.
Directives and standards
The instrument described in this manual has been designed in compliance with the following standards:
– EN
– IEC
– UL
– CSA
In accordance with Directive 93/42/EEC, the complete quality management system of the company Carl Zeiss has been certified by the DQS
Deutsche Gesellschaft zur Zertifizierung von Managementsystemen mbH, a notified body, under registration number 250758 MR2.
• The instrument must be connected to a special emergency backup line supply in accordance with the regulations or directives which apply in your country.
• This is a class I instrument as defined by Directive 93/42 /EEC.
• Please observe all applicable accident prevention regulations.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Safety
11
Notes on installation and use
Safe working order
• Do not operate the equipment contained in the delivery package in
– explosion-risk areas,
– the presence of inflammable anesthetics or volatile solvents such as alcohol, benzine or similar chemicals.
• Do not station or use the instrument in damp rooms. Do not expose the instrument to water splashes, dripping water or sprayed water.
• Immediately unplug any equipment that gives off smoke, sparks or strange noises. Do not use the instrument until our service representative has repaired it.
• Do not place any fluid-filled containers on top of the instrument. Make sure that no fluids can seep into the instrument.
• Do not force cable connections. If the male and female parts do not readily connect, make sure that they are appropriate for one another.
If any of the connectors are damaged, have our service representative repair them.
• Potential equalization: The instrument can be incorporated into potential equalization measures. For this purpose, contact our service department.
• Do not use a mobile phone in the vicinity of the equipment because the radio interference can cause the equipment to malfunction. The effects of radio interference on medical equipment depend on a number of various factors and are therefore entirely unforeseeable.
• Modifications and repairs on these instruments or instruments used with them may only be performed by our service representative or by other authorized persons.
• The manufacturer will not accept any liability for damage caused by unauthorized persons tampering with the instrument; this will also forfeit any rights to claim under warranty.
• Use this instrument only for the applications described.
• Only use the instrument with the accessories supplied. Should you wish to use other accessory equipment, make sure that Carl Zeiss or the equipment manufacturer has certified that its use will not impair the safety of instrument.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
12
Safety
G-30-1439-en
• Only personnel who have undergone training and instruction are allowed to use this instrument. It is the responsibility of the customer or institution operating the equipment to train and instruct all staff using the equipment.
• Keep the user's manuals where they are easily accessible at all times for the persons operating the instrument.
• Never look at the sun through the binocular tube, the objective lens or an eyepiece.
• Do not pull at the light guide cable, at the power cord or at other cable connections.
• This instrument is a high-grade technological product. To ensure optimum performance and safe working order of the instrument, its safety must be checked once every 12 months. We recommend having this check performed by our service representative as part of regular maintenance work.
If a failure occurs which you cannot correct using the trouble-shooting table, attach a sign to the instrument stating it is out of order and contact our service representative.
Requirements for operation
Our service representative or a specialist authorized by us will install the instrument. Please make sure that the following requirements for operation remain fulfilled in the future:
– All mechanical connections (details in the user's manual) which are relevant to safety are properly connected and screw connections tightened.
– All cables and plugs are in good working condition.
– The voltage setting on the instrument conforms to the rated voltage of the line supply on site.
– The instrument is plugged into a power outlet which has a properly connected protective earth contact.
– The power cord being used is the one designed for use with this instrument.
Before every use and after re-equipping the instrument
• Make sure that all ”Requirements for operation” are fulfilled.
• Go through the checklist.
• Re-attach or close any covers, panels or caps which have been removed or opened.
OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Safety
13
• Pay special attention to warning symbols on the instrument (triangular warning signs with exclamation marks), labels and any parts such as screws or surfaces painted red.
For every use of the instrument
• Avoid looking directly into the light source, e.g. into the microscope objective lens or a light guide.
• Any kind of radiation has a detrimental effect on biological tissue.This
also applies to the light illuminating the surgical field. Please therefore reduce the brightness and duration of illumination on the surgical field to the absolute minimum required.
• When operating on the eye, always use a GG 475 protection filter to ensure that the patient's retina is not exposed to unnecessary (blue) radiation (retinal injury).
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
14
Safety
When using a fundus imaging system (e.g. BIOM II)
Risk of collision!
1
2
When using a fundus imaging system (e.g. BIOM II from the company Oculus) which is usually installed between the surgical microscope and the patient, make sure that the patient is neither put at risk nor injured by the motorized focusing system or the movement of the stand arm.
Only use accessories expressly certified by the manufacturer for combination with the surgical microscope described in this manual.
Caution!
• With the fundus imaging system swung out of position, always position the microscope body in such a way that index dot (1) of the microscope's focus is in the middle of triangle (2) of the marking.
• Select a medium magnification (e.g. 1.0).
• Lower the surgical microscope towards the surgical field until you see the patient's cornea sharply defined.
• Turn the screw for limiting the downward movement clockwise as far as it will go.
• It is vital that you read the user’s manual on the fundus imaging system (e.g. BIOM II from the company Oculus).
Phototoxic retinal injury in eye surgery
General
Several papers have been published dealing with the problems of phototoxicity during eye surgery. A comprehensive review of these publications reveals five aspects of particular concern:
– Illumination characteristics (spectral composition)
– Intensity of illumination
– Angle of illumination
– Focus of the light source
– Exposure time to light
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
G-30-1439-en
Safety
15
In the following, comments on these aspects are given and a description of how Carl Zeiss, as a manufacturer, makes allowance for them in its instruments.
Illumination characteristics (spectral composition)
Studies on exposure of the eye to light of varying spectral composition date back to the early 1950s. These studies suggest that the potential hazard of phototoxic injury to the patient's retina can be reduced by blocking out the blue and ultraviolet light below a wavelength of 475 nm.
Carl Zeiss provides a GG 475 retina protection filter for surgical microscopes recommended for use in ophthalmic surgery. This reduces not only the light exposure of the patient's eye, but also that of the surgeon's.
It should be noted in this context that the use of filters inevitably leads to a change in the color of the light. The surgeon may therefore have to get used to the changed appearance of anatomical structures.
Intensity of illumination
The majority of researchers suggest that the surgeon should use the lowest light intensity necessary to guarantee good viewing during surgery.
Carl Zeiss has addressed this concern by providing a device for continuously varying the brightness of the light source. This permits the surgeon to optimally adapt the light intensity at the patient's eye to the conditions existing in each case. Carl Zeiss strongly discourages the use of xenon or other high-intensity light sources in ophthalmology.
Angle of illumination
A number of publications suggest that the microscope should be tilted to reduce the exposure of the macula to direct illumination.
Carl Zeiss ophthalmic surgical microscopes are therefore equipped with the following:
– Tilting mechanism for the microscope body
– Oblique illumination with brightness control
Focus of the light source
Studies show that injuries are likely to occur if the filament of the light source is imaged on the patient's retina. The peak intensity of a filament is considerably higher than the peak intensity of an even and extended light source such as a fiber guide.
OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
16
Safety
This is the reason why fiber optic illumination is used in surgical microscopes from Carl Zeiss.
Exposure time to light
According to some publications, the phakic or aphakic eye should not be exposed to the light source longer than a few minutes. In every operation the exposure of the retina to light is dependent on the type and duration of surgery and on any complications which occur. It is therefore recommended in ophthalmic surgery to keep the light intensity as low as possible, or to use a device which prevents the light from entering through the patient's pupil. Also, the surrounding light sources should not cause additional strain to the patient's eye.
Carl Zeiss has provided an answer to this problem in the form of a swingin retinal protection device for insertion into the beam path of the surgical microscope. This device ensures total eclipsing of the pupil, preventing light from entering into the patient's eye. It can be swung out when a red reflex is required.
Intensity scale
The intensity scale of our suspension system is calibrated in units of the
”Spectrally weighted radiance for the photochemical hazard of the phakic eye (L
B
)”
L
B is the spectral radiance integrated over the spectral range from 380 nm to 700 nm and weighted with B(
λ
):
700
L
B
=
∑
L(
λ
) B(
λ
)
∆λ
380 in which B(
λ
) is the spectral weighting function for the photochemical hazard of the retina of the phakic eye.
The value L
B
= 500mW/cm² sr is the reference value and is defined as
1.0 on the intensity scale of the suspension system as recommended in the ISO 10936-2 standard draft for surgical microscopes used in ophthalmic applications. At this reference value, photoretinitis might be expected to occur from the surgical microscope light source in a total retinal exposure time of 10 minutes. This applies to the exposure of a specific point on the retina with an uninterrupted illumination beam. In a cataract procedure, instruments such as the phacoemulsification handpiece, the
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Safety
17 use of fluids in the eye, manipulation and movement of the eye, among other factors, result in the interruption of the illumination from the surgical microscope light source. Such factors would be expected to significantly extend the time at which photoretinitis might occur.
In conclusion
Carl Zeiss recommends:
– Use of the GG 475 eye protection filter.
– Reduction of the illumination of the surgical area to the extent required for the patient's safety and for microscopic visualization.
– Tilting of the microscope body as required.
– Insertion of the retinal protection device.
– Maximum reduction of the exposure of the patient's eye to light from surrounding light sources.
These measures should help the surgeon to reduce the likelihood of phototoxic retinal injury.
Note:
The VISU 150 microscope always contains a UV blocking filter.
The use of this filter ensures that the illumination intensity lies below 50
µW/cm
2 in the range between 305 nm and 400 nm.
These measures help the surgeon to reduce the risk of phototoxic retinal injury.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
18
Safety
Warning labels and notes
Caution:
Observe all warning labels and notes!
If any label is missing on your instrument or has become illegible, please contact us or one of our authorized representatives. We will supply the missing labels.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Description
Description
VISU 200 BrightFlex™ Surgical Microscope
Controls, displays, connections
Tube and eyepieces for main microscope
Tube and eyepieces for assistant's microscope
19
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
20
Description
VISU 200 BrightFlex™ Surgical Microscope
Intended use
The VISU 200 BrightFlex™ surgical microscope has been designed for surgical procedures in the field of ophthalmology, i.e. the microscope meets the special requirements of this discipline.
Description of the modules
The VISU surgical microscope is comprised of the following modules:
1
X-Y coupling
The X-Y coupling allows fine, motorized positioning of the surgical microscope in a horizontal plane. The range of travel is 40 mm x 40 mm.
The speed of travel can be set on the display field of the suspension system.
The X-Y coupling is provided with a recentering mechanism. When you press the actuator button,
– the X-Y coupling adopts its center position,
– the focusing system of the surgical microscope is reset to its initial position and
– the zoom system will set a preselected magnification factor if the
XYZ-RES function has been activated (possible with S8 suspension systems only).
You can also trigger the recentering movement using the foot control panel..
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Description
21
1
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
22
Description
2
Support arm for the surgical microscope
The support arm incorporates a tilt device. This allows the viewing direction of the surgical microscope to be adapted to the requirements of the surgical field. Using the knob for fine tilt, you can position the surgical microscope in a range from +180° to -180° (+ in the direction of the surgeon and - in the opposite direction). The +90° setting is ideal for surgery on patients in a seated position or lying on their side.
Caution:
Do not tilt the main microscope beyond + / - 180°, as this could damage the microscope cable or the light guide.
3
Main microscope
The apochromatic optics of the main microscope provide superb optical quality. The microscope image displays optimum contrast and excellent detail recognition along with a large depth of field. The bright microscope image is a particular benefit in vitreoretinal surgery. A 1:6 ratio zoom system allows the magnification of the overall system to be set as required by the surgical procedure.
Two apochromatic objective lenses with focal lengths of 175 mm and
200 mm are available for different working distances.
A 180° tiltable tube is used as a viewing device for the surgeon. The large tilt range allows work with minimum fatigue.
The standard equipment includes eyepieces with a magnification factor of 12.5x (option: 10x).
The illumination system has been designed for use in ophthalmology.
A light guide directs the light from the light source in the suspension system to the surgical microscope.
6° illumination can be continuously faded in. This allows the illumination intensity in the surgical field to be adjusted continuously, without changing the color of the light. The illumination of the surgical field at an angle of 6° produces an image with an outstanding impression of depth.
An additional illumination system produces an intensive red reflex even when the eye is decentered. The angle of illumination can be switched from +2° to -2°.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Description
23
2
3
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
24
Description
To protect the patient's eye against photo-retinitis, a retinal protection device is provided. This device can be swung into the illumination beam path, if no red reflex is required.
At the light source integrated in the stand, a GG 475 eye protection filter can be swung into the illumination beam path. This filter markedly reduces the exposure of the patient's and surgeon's eyes to radiation.
A video camera from our MediLive video camera line can be optionally connected to the VISU 200 surgical microscope. The light loss for the surgeon is only 20% and cannot be perceived subjectively.
Caution!
– Avoid looking directly into the light source, e.g. into the microscope objective lens or into the light guide!
– When selecting the brightness level for the patient's eye, always take care to keep the strain on the patient's eye to a minimum.
– If the red reflex is not necessary, move the retinal protection device into the beam path. Only use the retro-illumination contrast stop, if the surgical procedure requires a red reflex.
– When operating on the eye, always use a GG 475 eye protection filter to ensure that the patient's retina is not exposed to unnecessary (blue) radiation (retinal injury)!
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Description
25
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
26
Description
4
0° assistant's microscope
The assistant's microscope is an integral part of the VISU 200 Bright-
Flex™ surgical microscope, i.e. it cannot be separated from the main microscope. The assistant sees the same image as the main surgeon.
The system's excellent image quality is also available to the assistant.
The assistant's microscope has two working positions. They are located on the right and left of the main surgeon at an angle of 90° to the main surgeon's viewing direction. No locking mechanism has been provided, allowing the assistant to move the microscope by a certain amount out of the 90° position, if necessary.
Warning!
To prevent the assistant's microscope from moving downward of its own accord when the main microscope is being tilted, the assistant's microscope must be adjusted and locked in position using screw (5) before surgery.
The assistant's microscope is equipped with a focusing system and a
5-step magnification changer. This enables the assistant to adjust his microscope image independently of the main surgeon.
The binocular tube can be turned by ± 12° about the optical axis of the assistant's microscope. In addition, the assistant's microscope can be tilted by 15°. If the assistant finds the viewing angle too steep, an optical wedge (option) can be installed between the microscope body and the binocular tube to permit horizontal viewing.
The standard equipment includes eyepieces with a magnification factor of 10x, providing a low initial magnification. This provides the benefit of a wide field of view and an improved overview of the surgical field. The assistant sees the red reflex in both eyepieces.
5
Locking screw for assistant's microscope
After adjusting the assistant's microscope as required, secure it in position using this screw.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Description
27
4
5
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
28
Description
Illumination system
6° illumination
You can continuously fade out the 6° illumination. This results in markedly reduced reflection from the cornea, while providing nevertheless a highcontrast, informative microscope image.
For cataract surgery, we recommend opening the 6° illumination system approx. 25%.
2° illumination
You can switch the illumination from +2° to -2°. This ensures good visibility of the red reflex even when the patient's eye is decentered.
1
6° illumination and 2° illumination
2
2° illumination with centered patient's eye
Here, the red reflex is clearly visible in both positions (+2° and -2°).
3
2° illumination with decentered patient's eye
The red reflex is no longer visible at +2°. After switching to -2°, the red reflex is again clearly visible.
+ 6°
G-30-1439-en
- 2° + 2°
OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
1
Description
29
2
3
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
30
Description
Controls, displays, connections
1
Instrument connector
2
Microscope shaft
3
X-Y coupling
4
Actuator button
– recenters the X-Y coupling.
– resets the focus to its initial position in the focusing range, and
– the zoom system will set a preselected magnification factor if the
XYZ-RES function has been activated (possible with S8 suspension systems only).
Note:
Press this button to start the recentering movement. To stop this movement, press the button again.
You can also stop the recentering movement by briefly tipping on one of the direction keys on the foot control panel.
5
Cable and light guide clip
6
Support arm with tilt device
7
Knob for setting the tilt angle of the surgical microscope;
+180° in the direction of the surgeon,
-180° in the opposite direction.
8
Arrows indicating the focusing range
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Description
3
4
5
6
1
2
7
8
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
31
32
Description
9
Dust cover
10
Release buttons for the magnetic brakes of the stand
Only in connection with suspension systems with magnetric brakes.
A second button with the same function is located on the opposite side in the knob for the additional red reflex illumination.
– Button pressed
– Button released magnetic brakes are released, the unit can be moved as required.
magnetic brakes are locked, the unit cannot be moved.
11
Maneuvering grip including release button. See (10).
12
Display of the magnification factor of the zoom system
13
Securing screw
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Description
33
9
10
G-30-1439-en
11 12 13
OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
34
Description
14
Manual adjustment of zoom system
15
Knob for adjusting the red reflex illumination
The angle of illumination can be switched between + 2° and - 2°.
16
Release button for magnetic brakes
(A second button with the same function is located in maneuvering grip (10).
– Button pressed
– Button released magnetic brakes are released, the unit can be moved as required.
magnetic brakes are locked, the unit cannot be moved.
17
Securing screw for locking the assistant's tube within its 12° range of movement.
18
Binocular tube of the assistant's microscope
19
Focusing knob for adjusting the assistant's microscope independently of the main surgeon.
20
Five-step manual magnification changer
21
Screw for locking the assistant's microscope within its 15° tilt range.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Description
35
14
15
16
17
18
19
G-30-1439-en
20 21
OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
36
Description
22
Light stop knob
Retro-illumination contrast stop.
This stop reduces the straylight reflected from the sclera.
Diameter approx. 16 mm (with objective lens f = 200 mm): Free passage.
Diameter over approx. 16 mm: Partially transmitting periphery.
Free passage. The field of view is fully illuminated.
Retinal protection device. This device does not snap in, i.e. it can be moved continuously in the image.
23
Adjusting lever for the 6° illumination for gradual fading in/out the coaxial illumination. Fading out the 6° illumination improves the visualization of structures in retro-illumination.
24
Connector socket for S light guide
25
Locking screw for assistant's microscope for locking the assistant's microscope in both working positions on the right and left of the main microscope.
Warning!
To prevent the assistant's microscope from moving downward of its own accord when the main microscope is being tilted, the assistant's microscope must be adjusted and locked in position using screw (25) before surgery.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Description
37
G-30-1439-en
22 23 24 25
OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
38
Description
Hand grips (option)
In addition to the maneuvering grips provided, larger hand grips can be mounted. The mounting procedure is described in the instructions enclosed.
This does not affect the functioning of the surgical microscope. Only the operation of the surgical microscope is slightly changed.
1
Maneuvering grip with integrated release button
By turning one of the two maneuvering grips you will activate the button for magnetic brakes of the suspension system.
(Only in combination with suspension systems with magnetic brakes
(e.g. S8 Floor Stand)).
– Button pressed
– Button released
Magnetic brakes are released, the unit can be moved as required.
Magnetic brakes are locked, the unit cannot be moved.
2
Knob for adjusting the red reflex illumination
By sliding the knob, you can switch the angle of illumination between
+ 2° and - 2°.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Description
39
1
2
1
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
40
Description
Tube and eyepieces for main microscope
1
180° tiltable tube
2
PD adjustment knob
The correct interpupillary distance has been set when both eyepiece images merge into one.
3
Push-in widefield eyepiece
Note:
When you remove the eyepiece from the tube, please note that this eyepiece is fitted with a magnetic coupling. When mounted, the eyepieces display a very weak magnetic field. For this reason, the usual rules for the handling of magnets must only be observed with eyepieces which have not been mounted on the microscope:
• Do not place the eyepieces close to instruments where there is any risk of magnetization.
• Do not place eyepieces on sensitive electronic instruments such as infusion pumps, cardiac pace-makers, measuring instruments or magnetic data carriers such as disks, audiotapes and videotapes or credit cards.
• Always store the eyepieces in their original packaging when not in use.
4
Eyecup
Always adjust the eyecups in such a way that you can see the full field of view.
– Viewing with eyeglasses:
– Viewing without eyeglasses:
Screw in eyecups all the way.
Screw eyecups outward until you can see the full field of view.
5
Diopter setting ring
The eyepieces provide ametropia compensation from -8 D to +5 D. If you wear your eyeglasses during work, adjust the ring to 0 D. Turn the ring until you have found the optimum setting. A built-in catch lever locks the ring in the position set.
6
Diopter scale
For reading the prescription set.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
3
4
1
2
5
6
Description
41
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
42
Description
Tube and eyepieces for assistant's microscope
7
Push-in widefield eyepiece
Note:
When you remove the eyepiece from the tube, please note that this eyepiece is fitted with a magnetic coupling. When mounted, the eyepieces display a very weak magnetic field. For this reason, the usual rules for the handling of magnets must only be observed with eyepieces which have not been mounted on the microscope:
• Do not place the eyepieces close to instruments where there is any risk of magnetization.
• Do not place eyepieces on sensitive electronic instruments such as infusion pumps, cardiac pace-makers, measuring instruments or magnetic data carriers such as disks, audiotapes and videotapes or credit cards.
• Always store the eyepieces in their original packaging when not in use.
8
Eyecup
Always adjust the eyecups in such a way that you can see the full field of view.
– Viewing with eyeglasses:
Screw in eyecups all the way.
– Viewing w/out eyeglasses:
Screw eyecups outward until you see the full field of view.
9
Diopter scale
For reading the prescription set.
10
Diopter setting ring
The eyepieces provide ametropia compensation from -8 D to +5 D.If
you wear your eyeglasses during work, adjust the ring to 0 D. Turn the ring until you have found the optimum setting. A built-in catch lever locks the ring in the position set.
11
Binocular assistant's microscope
Note:
Always use the assistant's microscope with the appropriate (shown) binocular assistant's tube. If you use a different tube, you will obtain an inverted image.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Description
7
8
9
10
11
43
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
44
Description
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Preparations for use
45
Preparations for use
Mounting the binocular tubes, eyepieces and the objective lens 46
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
46
Preparations for use
Mounting the binocular tubes, eyepieces and the objective lens
Note:
Always use the assistant's microscope with the appropriate binocular assistant's tube (shown). If you use a different tube, you will obtain an inverted image.
• Loosen securing screw (5) by a few turns.
• Remove cover (1) and store it in a safe place.
• Place binocular tube (2) on the surgical microscope and tighten securing screw (5) firmly.
• You can install other units between the binocular tube and the microscope body. Lock these units in position in the same way using securing screw (5).
• Place binocular tube (10) on the assistant's microscope and tighten securing screw (7)firmly.
• You can install other units between the binocular tube and the microscope body (e.g. an optical wedge, 30°). Lock these units in position in the same way using securing screw (7).
• Insert
12.5x
widefield eyepieces (4) into the mounts intended for them
(3) as far as they will go. The magnetic coupling reliably secures them in position.
• Insert
10x
widefield eyepieces (9) into the mounts intended for them
(8) as far as they will go. The magnetic coupling reliably secures them in position.
Note:
If a documentation device is used, one of the eyepieces can be delivered with a reticle to aid focusing. Retrofitting the reticle can only be performed at the factory or by our service representative. Always use the eyepiece with reticle on the same side of the binocular tube as the documentation device is installed.
• Screw objective lens (6) into the microscope body and tighten itfirmly.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Preparations for use
47
Caution:
Before using and after re-equipping the unit, make sure that the two binocular tubes (2 and 10) are securely locked in position. Make sure that
– securing screws (5 and 7) and
– objective lens (6) have been tightened firmly.
1
2
3
4
5
6
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
10
9
8
7
48
Preparations for use
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Operation
49
Operation
Setting the microscope tilt to angles greater than 15°
Adjusting the surgical microscope
When using a fundus imaging system (e.g. BIOM II)
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
50
Operation
Preparations for use
Adjusting the tilt angle
Using knob (1), you can position the surgical microscope in a range from
+180° to -180° (+ in the direction of the surgeon and - in the opposite direction). The +90° setting is ideal for surgery on patients in a seated position or lying on their side.
Warning!
– To prevent the assistant's microscope from moving downward of its own accord when the main microscope is being tilted, the assistant's microscope must be adjusted and locked in position using screw (2) before surgery.
– Check that the assistant's microscope is firmly seated.
– Do not tilt the main microscope beyond + / - 180°, as this could damage the microscope cable or the light guide.
• Using locking screw (2), secure the assistant's microscope in the working position set (on the left or right of the main microscope).
• Turn knob (1) until the surgical microscope is in the viewing position required.
After the viewing angle has been set, the surgical microscope remains in this position. The gear drive is self-locking.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Operation
51
1
G-30-1439-en
2
OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
52
Operation
Setting the microscope tilt to angles greater than 15°
• Using locking screw (2), secure the assistant's microscope in the working position set (on the left or right of the main microscope).
• Check that the assistant's microscope is firmly seated.
• Remove the binocular tube and all accessories mounted on the assistant's microscope.
• Remove all accessories mounted on the objective lens (e.g. 0° or 8° assistant's microscope).
• Turn knob (1) until the surgical microscope is in the viewing position required.
After the viewing angle has been set, the surgical microscope remains in this position. The gear drive is self-locking.
1
2
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Operation
53
Adjusting the surgical microscope
Bring the surgical microscope into starting position within the focusing range. Set the minimum magnification on the surgical microscope. Bring the surgical microscope into working position. Adjust the interpupillary distance on the binocular tube. Set the diopter scales on the eyepieces. Please note that instrument myopia may occur.
Eyepieces without reticles:
Eyepieces with reticles:
Emmetropes set the diopter scale to 0 diopter (D).
Eyeglass wearers
(who perform surgery wearing their glasses) set the diopter scale to 0 diopter (D).
Ametropes
(who know their refractive powers and perform surgery without wearing their glasses) set the diopter scale ring to their refractive powers.
Ametropes
(who do not know their refractive powers and perform surgery without wearing their glasses):
Set both eyepieces on the main microscope to +5 D.
Set both eyepieces on the assistant’s microscope to +5 D (if applicable).
Remove the binocular tube and eyepieces from the microscope body and point them at a distant object
*)
, i.e. use them like a pair of binoculars.The image of the object is still unsharp.
Set both eyepieces on the main microscope to +5 D.
Set both eyepieces on the assistant’s microscope to +5 D (if applicable).
Turn the diopter scale ring on the eyepiece slowly clockwise until the reticle is imaged sharply. Stop turning the ring.
Turn the diopter scale ring of one eyepiece slowly clockwise until the image of the object is sharp. Stop turning the ring. If necessary, repeat this procedure three times and take the average of the readings found.
Adjust the second eyepiece in the same manner.
Replace the tube and eyepieces on the microscope body and firmly tighten the securing screw.
Focus the surgical microscope on the object. The reticle and the object must be imaged sharply at the same time.
Adjust the second eyepiece in the same way until the object is seen sharply.
Adjust the eyecups in such a way that the full field of view can be seen. Set the maximum magnification on the microscope and focus on the object. Set the working magnification required. When the magnification is changed, the focal plane is retained, but the depth of field changes.
Note: If several surgeons use the instrument, it is advisable to draw up a table showing the individual refractive powers of each surgeon and to keep it in a handy location near the instrument.
*) CAUTION:
Never point the eyepieces at the sun!
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
54
Operation
Checklist
G-30-1439-en
Always check the following points before surgery (without the patient!):
VISU 200 BrightFlex™ surgical microscope
– Check that the correct rated voltage has been set for the suspension system.
– Check that all cables have been connected.
– Check that the light guide has been connected.
• Turn on the instrument at the power switch of the suspension system.
Note:
After the system has been switched on, the S8 suspension system(ophthalmology) automatically performs a self-test. The self-test is completed after approx. five seconds.
Zoom
– Check this function using the appropriate key on the foot control panel.
Focus
– Check this function using the appropriate key on the foot control panel.
Friction adjustment of the surgical microscope
– Check that the friction of the surgical microscope's rotation has been adjusted as required using the friction adjustment knob on the suspension system.
Locking the assistant's microscope in position
– Check that the assistant's microscope is locked in the working position required (on the right or left of the main microscope).
Speeds of the surgical microscope's functions
(only in connection with S8 suspension system (ophthalmology))
– Check that the speeds of the microscope's functions have been set as required on the S8 suspension system (ophthalmology).
Eyepieces / Tiltable binocular tubes
– Check that the tiltable tube of the main microscope and the assistant's microscope are in positions convenient for you.
– Check that the correct interpupillary distance has been set.
– Check that the eyecups have been adjusted in such a way that you can see the full field of view.
– Check that the correct prescription has been set on the diopter scale.
OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Operation
55
– Check that image quality is the same throughout the entire magnification range.
Accessories
– Using the relevant user's manuals, check that the other equipment (illumination system, video system, etc.) is working properly.
Warning!
If a function fails, you must not use this instrument for safety reasons. Correct the fault (see the "Troubleshooting table“) or contact our service dept.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
56
Operation
When using a fundus imaging system (e.g. BIOM II)
Risk of collision!
1
2
When using a fundus imaging system (e.g. BIOM II from the company Oculus) which is usually installed between the surgical microscope and the patient, make sure that the patient is neither put at risk nor injured by the motorized focusing system or the movement of the stand arm.
Only use accessories expressly certified by the manufacturer for combination with the surgical microscope described in this manual.
Caution!
• With the fundus imaging system swung out of position, always position the microscope body in such a way that index dot (1) of the microscope's focus is in the middle of triangle (2) of the marking.
• Select a medium magnification (e.g. 1.0).
• Lower the surgical microscope towards the surgical field until you see the patient's cornea sharply defined.
• Turn the screw for limiting the downward movement clockwise as far as it will go.
• It is vital that you read the user’s manual on the fundus imaging system (e.g. BIOM II from the company Oculus).
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Operation
57
Procedure
• Turn on the power switch of the suspension system.
• Adjust the brightness of the surgical field illumination as required on the stand.
• Check the unit using the checklist.
• Move the surgical microscope over the surgical field and bring it into a convenient position within the working distance.
• Press the activator button.
– The X-Y coupling is recentered.
– The focus is reset to its initial position in the focusing range.
• Select the lowest magnification (zoom function on the foot control panel).
• For coarse focusing, look through the eyepieces and lower the surgical microscope using the suspension arm until the surgical field comes into focus.
• Select the highest magnification (zoom function on the foot control panel).
• Look through the eyepieces and activate the focusing function on the foot control panel until the microscope is sharply focused on the surgical field.
• Select the magnification required (zoom). Look through the eyepieces of the binocular tube. Adjust the eyepieces (including those of the assistant's microscope) in such a way that you can see both the edge of the field of view and the microscope image sharply. Also see “Adjusting the surgical microscope“.
• Switch the unit off when it is not in use.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
58
Operation
Caution!
– Avoid looking directly into the light source, e.g. into the microscope objective lens or into a fiber light guide!
– When selecting the brightness level for the patient's eye, always take care to keep the strain on the patient's eye to a minimum.
– If the red reflex is not necessary, move the retinal protection device into the beam path. Only use the retro-illumination contrast stop, if the surgical procedure requires a red reflex.
– When operating on the eye, always use a GG 475 protection filter to ensure that the patient's eye is not exposed to unnecessary (blue) radiation (retinal injury).
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Operation
59
What to do in an emergency
Failure of zoom system
• Manually adjust the magnification using zoom control (1) on the microscope (if necessary, use a tool, e.g. screwdriver, coin, etc.).
G-30-1439-en
1
OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
60
Operation
Failure of X-Y coupling
• If the X-Y coupling fails, you can manually position the surgical microscope utilizing the possibilities of the suspension system you are using.
Failure of focusing system
• Disconnect the cable between the surgical microscope and the suspension system (illumination continues to be on.)
• Focus the surgical microscope by moving the suspension arm or the lift arm of the suspension system you are using.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Maintenance / Further information
Maintenance / Further information
Magnifications / Fields of view
61
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
62
Maintenance / Further information
Trouble-shooting table
• This instrument is a high-grade technological product. To ensure optimum performance and safe working order of the instrument, its safety must be checked once every 12 months. We recommend having this check performed by our service representative as part of regular maintenance work.
If a failure occurs which you cannot correct using the trouble-shooting table, attach a sign to the instrument stating it is out of order and contact our service representative.
VISU 200 BrightFlex™ Surgical Microscope
Problem
No function at all.
Possible cause See
Power cord of suspension system not plugged in.
Plug in power cord.
Power switch of suspension system not on.
Press power switch. Green light in power switch must come on.
-
-
Circuit breaker in suspension system power switch activated.
Press power switch again.
-
Line power failure.
Contact in-house electrician.
-
Lamp module has no contact.
Insert lamp module as far as it will go.
-
Surgical field illumination at microscope inoperative.
Illumination system of suspension switched on.
system not
Switch on illumination system of suspension system.
-
S light guide not properly inserted in lamp or microscope.
Insert S light guide as far as it will go.
-
Failure of suspension system electronics.
Illuminate surgical field using an OR illuminator.
Contact service dept.
-
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Maintenance / Further information
Problem Possible cause See
Insufficient surgical field illumination.
Brightness level set too low. Adjust brightness on suspension system or using foot control panel.
-
Defective S light guide (no homogeneous illumination).
Contact service dept.
The light guide must probably be changed.
-
Focusing function inoperative.
Connectors on suspension system not correctly inserted.
Plug in connectors correctly. -
Focusing system always moves into upper or lower end position.
Check suspension system.
Move the suspension arm to adjust the focus.
Contact service dept.
-
Zoom function inoperative.
Connectors on suspension system not correctly inserted.
Plug in connectors correctly. -
Zoom system always moves into upper or lower end position.
Check suspension system.
Manually adjust magnification using the zoom knob.
Contact service dept.
page 34
X-Y coupling inoperative.
Connectors on suspension system not correctly inserted.
Plug in connectors correctly. –
Defective motor.
Position microscope manually.
Contact service dept.
-
X-Y coupling always moves in opposite direction.
Function not correctly adjusted on suspension system.
Check suspension system.
-
Zoom and focusing functions inoperative.
Functions not correctly adjusted on suspension system.
Check suspension system.
-
Microscope motion too stiff.
Friction adjustment knob on suspension system tightened too firmly.
Loosen friction adjustment.
-
63
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
64
Maintenance / Further information
Magnifications / Fields of view
Using the magnification factor
γ of the zoom system, you can calculate the total magnification of the surgical microscope according to the following formula:
M
T
= f
----------f obj
⋅ ⋅
M eye where: f tube is the focal length of the binocular tube f obj is the focal length of the main objective lens
γ is the magnification factor set on the zoom system
M eye is the magnification of the eyepiece
Example: f tube
= 170 mm, f obj
= 200 mm,
γ
= 1.6 and M eye
= 12.5 x.
The resulting total magnification is:
M
T
=
170 mm
200 mm
⋅ ⋅
=
17.0
If the total magnification M
T of-view diameter FoV
D of the surgical microscope is known, the fieldcan be calculated using the formula:
FoV
D
=
FoVN M
M
T eye
The field-of-view diameter FoV
D is the diameter of the circular area of the surgical field which can be seen through the eyepieces.
FoV
N in the above formula stands for the field-of-view number of the eyepiece. This number is marked on our widefield eyepieces.
Using M
T
= 17.0 from the example above, the field-of-view diameter obtained with a 12.5x eyepiece with a field-of-view number FoV
N is calculated as follows: of 18 mm
FoV
D
=
18 mm 12.5
13.2 mm
17.0
=
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Maintenance / Further information
65
Care of the unit
Cleaning optical surfaces
The multilayer T* (T-star) coating of our optical components (e.g. eyepieces, objective lenses) results in optimum image quality.
Image quality is impaired by even slight contamination of the optics or by a fingerprint. In order to protect the internal optics from dust, the instrument should never be left without the objective lens, binocular tube and eyepieces installed when it is not in use. Always store objective lenses, eyepieces and accessories which are not being used in dust-free cases.
The external surfaces of optical components (eyepieces, objective lenses) should only be cleaned when required:
• Do not use any chemical cleaning agents.
• Blow off dust on the optical surfaces using a squeeze blower or remove the dust using a clean and grease-free brush.
Fogging of optical surfaces
To protect the eyepiece optics from fogging, we recommend using an anti-fogging agent.
Note:
Anti-fogging agents provided by eyecare professionals for use with eyeglass lenses are also suitable for Zeiss eyepieces.
• Please observe the instructions for use supplied with each anti-fogging agent.
Anti-fogging agents do not only ensure fog-free optics. They also clean the eyepiece optics and protect them against dirt, grease, dust, fluff and fingerprints.
Cleaning mechanical surfaces
All mechanical surfaces of the equipment can be cleaned by wiping with a moist cloth. Do not use any aggressive or abrasive cleaning agents.
Wipe off any residue with a mixture of 50% ethyl alcohol and 50% distilled water plus a dash of household dish-washing liquid.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
66
Sterilization
Maintenance / Further information
The asepsis sets available from Carl Zeiss contain rubber caps, sleeves and handgrips which can be sterilized in autoclaves. We recommend the following program for sterilization:
Sterilization temperature: 134° C
Sterilization time: 10 minutes
Sterile single-use drapes are available to cover the instrument.
Note:
When draping the system, make sure there is enough slack in the drapes to allow for movement of the microscope carrier and surgical microscope.
It is especially important that the drapes are completely loose around the handgrips. The surgeon must be able to operate the keys through the drape.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Maintenance / Further information
67
Ordering data
VISU 200 BrightFlex™ Surgical Microscope
Description
VISU 200 BrightFlex™ surgical microscope including
X-Y coupling and assistant's microscope
180° tiltable tube
Cat.No.
1179-101
30 37 91-0000
10x eyepiece for assistant's microscope (2x)
12.5x/18 B eyepiece for main microscope (2x)
Objective lens, f = 175 mm
Dust cover
30 55 42-0000
30 55 43-0000
30 26 51-9902
1055-278
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
68
Spare parts
Maintenance / Further information
VISU 200 BrightFlex™ Surgical Microscope
Description
S light guide (not shown) 2.0 m
Dust cover (not shown)
Asepsis set (not shown)
Cat. No.
30 34 81-9020
1055-278
1009.100
Accessories
Please observe the following:
Only operate the instrument with the accessories included in the delivery package. If you want to use other accessories, make sure that Carl Zeiss or the manufacturer of the accessories has proved and confirmed that these accessories meet the respective technical safety standards and can be used without risk.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Maintenance / Further information
VISU 200 BrightFlex™ Surgical Microscope
Description
1 Objective lens, f = 200 mm
2 Optical wedge, 30°
3 Sleeve (for insertion of eyepiece (4))
4 10x eyepiece with magnetic coupling
5 OPMI VISU hand grips
Asepsis for OPMI VISU hand grips
(1x included in delivery package 1049- 171)
Cat. No.
30 26 52-9903
1006-145
30 55 42-0107
30 55 42-0000
1049- 171
1054- 074
69
1
G-30-1439-en
2 3 4 5
OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
70
Adapter cables
Maintenance / Further information for the electrical connection of the surgical microscope to different types of suspension system.
1 EMD II Electromechanical Ceiling Mount
1 S3 Ceiling Mount
2 S4 Ceiling Mount
1 S23 Ceiling Mount
3
3
1
S6 Ceiling Mount
S61 Ceiling Mount
S22 Floor Stand
2 S4 Floor Stand
1 S3 Floor Stand
Cat. No.
30 26 02- 9053
30 26 02- 9053
30 26 02- 9054
30 26 02- 9053
30 26 02- 9055
30 26 02- 9055
30 26 02- 9053
30 26 02- 9054
30 26 02- 9053
1 2 3
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Technical data
Technical data
71
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
72
Technical data
Technical data
VISU 200 BrightFlex™ Surgical Microscope
Optical data
Magnification Main microscope:
4.9x - 29x
Focusing
Motorized zoom system with apochromatic optics, 1:6 zoom ratio, magnification factor
γ
= 0.4x - 2.4x.
Motorized, focusing range 50 mm
At the press of a button, the focusing drive returns to its initial position.
Objective lens focal length f = 175 mm (option: f = 200 mm)
Tubes / Eyepieces Main microscope:
180° tiltable binocular tube, f=170 mm. 12.5x
screw-type, widefield eyepieces (option: 10x ).
Illumination
Assistant's microscope:
5-step magnification changer, separate fine focusing, adjustable tilt.
Binocular tube, can be turned by ±12° about its optical axis,
10x widefield eyepieces with magnetic coupling.
Fiber illumination with two lamp modules, each with two 12 V 100 W halogen reflector lamps, fully automatic lamp change in the event of lamp failure, filter against UV exposure, protection against IR exposure,
6° illumination (continuous fading), patented 2° illumination for red reflex, usable in
+2° and -2° positions
Pat. No. DE40 28 60 5 C2
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Technical data
73
Swing-in stops Patented retinal protection device,
Pat. No. DE 33 39 17 2 C2
Patented field stop for reduced glare,
Pat. No. G 91 03 43 3.7
Mechanical data
Front-to-back tilt of microscope
With self-locking gear drive, manually adjustable using a knob.
Tilt angle ± 180°
Swivel range of assistant's microscope
X-Y coupling
Swivel angle ± 90°, manually lockable
Adjustment range: max. 40 mm x 40 mm
Weight
Automatic centering at the press of a button approx. 13 kg
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
74
Technical data
Ambient requirements
For operation
For transportation and storage
Temperature
Rel. humidity
Air pressure
Temperature
Rel. humidity
(without condensation)
Air pressure
+10 °C...+40 °C
30%...75%
700 hPa...1,060 hPa
- 40 °C...+70 °C
10%...100%
500 hPa...1,060 hPa
The unit meets the essential requirements stipulated in
Annex I to the 93/42/EEC Directive governing medical devices. The unit is marked with:
Subject to change.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Index
Index
G-30-1439-en
Symbols
+90° position
Numerics
2° illumination
A
Accessories
Accident prevention regulations
Actuator button
Adjusting lever for 6° illumination
Adjusting the microscope tilt to angles greater than 15°
Adjusting the surgical microscope
Angle of illumination
Anti-fogging agent
Arrows indicating the focusing range
Assistant's microscope 0°
B
Binocular tubes, eyepieces and objective lens, mounting of
Buttons for the release of the magnetic brakes
C
Care of the unit
Connector socket for S light guide
Controls, displays, connections
D
Description
Directives
Diopter scale
Directive 93/42/EEC
E
Emergency backup line supply
Eyecup
OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
75
76
Index
Eyepiece optics, fog-free
F
Failure of a function
Failure of zoom system
Fields of view
Focus of the light source
Focusing knob
Focusing system failure
Fogging of optical surfaces
Fundus imaging system
G
GG 475 eye protection filter
GG 475 protection filter
H
Hand grips (option)
I
Illumination characteristics
Installation, notes
Instrument class
Instrument connector
Intensity of illumination
Intensity scale
ISO 10936-2 draft for surgical microscopes used in ophthalmic applications
K
Key to symbols
Knob for adjusting the red reflex illumination
L
LB
Light source, focus
Light stop knob
Locking screw for assistant's microscope
M
Magnification changer
Main microscope
Main microscope, tilted
Maintenance
Manual adjustment of zoom system
Mechanical surfaces, cleaning
Microscope shaft
Mobile phones
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
G-30-1439-en
Index
N
Notes on safety
O
Operation
Operation, requirements
Optical surfaces, fogging
Ordering data
P
PD adjustment knob
Phototoxic retinal injury in eye surgery
Potential equalization
Preparations for use
R
Red reflex illumination
Re-equipping the instrument
Release button for magnetic brakes
Requirements for operation
Risk of collision!
S
Safety
Safety check
Safety standards
Spectral composition
Standards
Support arm for surgical microscope
Support arm with tilt device
T
Time of exposure to light
Tube and eyepieces for assistant's microscope
Tube and eyepieces for main microscope
U
Use of the instrument, for every
Use, notes
W
OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
77
78
Index
X
X-Y coupling
X-Y coupling failure
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
79
G-30-1439-en
Carl Zeiss
Surgical Products Division
73446 Oberkochen
Germany
Fax: +49 (0) 7364 20-4823
E-mail: [email protected]
www.zeiss.com/surgical
Printed in Germany MS-TD
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 5 Functions at a glance
- 6 VISU 200 BrightFlex™ surgical microscope
- 7 Key to symbols
- 9 Safety
- 10 Directives and standards
- 11 Notes on installation and use
- 14 When using a fundus imaging system (e.g. BIOM II)
- 14 Phototoxic retinal injury in eye surgery
- 18 Warning labels and notes
- 19 Description
- 20 VISU 200 BrightFlex™ Surgical Microscope
- 20 Intended use
- 20 Description of the modules
- 28 Illumination system
- 30 Controls, displays, connections
- 38 Hand grips (option)
- 40 Tube and eyepieces for main microscope
- 42 Tube and eyepieces for assistant's microscope
- 45 Preparations for use
- 46 Mounting the binocular tubes, eyepieces and the objective lens
- 49 Operation
- 50 Preparations for use
- 50 Adjusting the tilt angle
- 52 Setting the microscope tilt to angles greater than 15˚
- 53 Adjusting the surgical microscope
- 54 Checklist
- 56 When using a fundus imaging system (e.g. BIOM II)
- 57 Procedure
- 59 What to do in an emergency
- 59 Failure of zoom system
- 60 Failure of X-Y coupling
- 60 Failure of focusing system
- 61 Maintenance / Further information
- 62 Trouble-shooting table
- 64 Magnifications / Fields of view
- 65 Care of the unit
- 66 Sterilization
- 67 Ordering data
- 68 Spare parts
- 68 Accessories
- 70 Adapter cables
- 71 Technical data
- 72 Technical data
- 74 Ambient requirements
- 75 Index