Agilent Technologies U1732C User`s guide


Add to my manuals
107 Pages

advertisement

Agilent Technologies U1732C User`s guide | Manualzz

Agilent U1731C, U1732C, and U1733C Handheld

LCR Meter

User’s Guide

Agilent Technologies

II

Notices

© Agilent Technologies, Inc. 2011–2012

No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent

Technologies, Inc. as governed by United

States and international copyright laws.

Manual Part Number

U1731-90077

Edition

Seventh Edition, November 28, 2012

Agilent Technologies, Inc.

5301, Stevens Creek Blvd.

Santa Clara, CA 95051 USA

Warranty

The material contained in this document is provided “as is,” and is subject to change, without notice, in future editions. Further, to the maximum extent permitted by the applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose.

Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.

Technology Licenses

The hardware and or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.

Restricted Rights Legend

U.S. Government Restricted Rights. Software and technical data rights granted to the federal government include only those rights customarily provided to end user customers. Agilent provides this customary commercial license in Software and technical data pursuant to FAR 12.211 (Technical

Data) and 12.212 (Computer Software) and, for the Department of Defense, DFARS

252.227-7015 (Technical Data - Commercial

Items) and DFARS 227.7202-3 (Rights in

Commercial Computer Software or Computer Software Documentation).

Safety Notices

C A U T I O N

A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the likes of that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a

CAUTION notice until the indicated conditions are fully understood and met.

WA R N I N G

A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the likes of that, if not correctly performed or adhered to, could result in personal injury or death.

Do not proceed beyond a WARN-

ING notice until the indicated conditions are fully understood and met.

U1731C/U1732C/U1733C User’s Guide

Safety Symbols

Direct current (DC)

Alternating current (AC)

Both direct and alternating current

Three-phase alternating current

Earth (ground) terminal

Protective conductor terminal

Frame or chassis terminal

Equipotentiality

The following symbols on the instrument and in the documentation indicate precautions which must be taken to maintain safe operation of the instrument.

Off (supply)

On (supply)

Caution, risk of electric shock

Caution, risk of danger (refer to this manual for specific Warning or Caution information)

Caution, hot surface

Out position of a bi-stable push control

In position of a bi-stable push control

Equipment protected throughout by double insulation or reinforced insulation

U1731C/U1732C/U1733C User’s Guide III

Safety Considerations

Read the information below before using this instrument.

The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards for design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements.

C A U T I O N

Disconnect circuit power and discharge all high-voltage capacitors before testing.

When measuring in-circuit components, first de-energize the circuits before connecting them to the test leads.

This device is for indoor use at altitudes of up to 2000 m.

Always use the specified battery type (listed in

“Product

Characteristics" on page 76). The power for the meter is supplied

with a single standard 9 V battery. Observe the correct polarity markings before you insert the battery to ensure proper insertion of the battery in the meter.

Line operation is also possible using a 12 V AC to DC adapter. If a power adapter is selected, please be sure it meets the safety requirements of a relevant IEC standard.

IV U1731C/U1732C/U1733C User’s Guide

WA R N I N G

• Use this meter only as specified in this manual; otherwise, the protection provided by the meter may be impaired.

• Do not use the meter if it is damaged. Before you use the meter, inspect the case. Look for cracks or missing plastic. Pay particular attention to the insulation surrounding the connectors.

• Inspect the test leads for damaged insulation or exposed metal.

Check the test leads for continuity. Replace damaged test leads before you use the meter.

• Do not operate the meter around explosive gas, vapor, or wet environments.

• Never use the meter in wet conditions or when there is water on the surface. If the meter is wet, ensure that the meter is dried only by trained personnel.

• When servicing the meter, use only the specified replacement parts.

• When using the probes, keep your fingers behind the finger guards on the probes.

• Connect the common test lead before you connect the live test lead. When you disconnect the leads, disconnect the live test lead first.

• Remove the test leads from the meter before you open the battery cover.

• Do not operate the meter with the battery cover or portions of the cover removed or loosened.

• To avoid false readings, which may lead to possible electric shock or personal injury, replace the battery as soon as the low battery indicator appears and flashes.

U1731C/U1732C/U1733C User’s Guide V

Environmental Conditions

This instrument is designed for indoor use and in an area with low condensation. The table below shows the general environmental requirements for this instrument.

N O T E

Environmental conditions

Operating temperature

Operating humidity

Storage temperature

Storage humidity

Altitude

Pollution degree

Requirements

Full accuracy from –10 °C to 55 °C

Full accuracy up to 80% RH (relative humidity)

–20 °C to 70 °C

0% to 80% RH non-condensing

Up to 2000 meters

Pollution degree II

The U1731C/U1732C/U1733C Handheld LCR Meter complies with the following safety and EMC requirements:

IEC61010-1:2001/EN61010-1:2001 (Second Edition)

IEC 61326-1:2005/EN 61326-1:2006

Canada: ICES/NMB-001:Issue 4, June 2006

Australia/New Zealand: AS/NZS CISPR11:2004

VI U1731C/U1732C/U1733C User’s Guide

Regulatory Markings

The CE mark is a registered trademark of the European Community. This CE mark shows that the product complies with all the relevant European Legal

Directives.

ICES/NMB-001 indicates that this ISM device complies with the Canadian

ICES-001.

Cet appareil ISM est confomre a la norme NMB-001 du Canada.

This symbol indicates the time period during which no hazardous or toxic substance elements are expected to leak or deteriorate during normal use.

Forty years is the expected useful life of the product.

The C-tick mark is a registered trademark of the Spectrum

Management Agency of Australia. This signifies compliance with the Australia EMC Framework regulations under the terms of the

Radio Communication Act of 1992.

This instrument complies with the

WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.

U1731C/U1732C/U1733C User’s Guide VII

Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC

This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.

Product Category:

With reference to the equipment types in the WEEE directive Annex 1, this instrument is classified as a “Monitoring and Control Instrument” product.

The affixed product label is as shown below.

Do not dispose in domestic household waste.

To return this unwanted instrument, contact your nearest Agilent Service

Centre, or visit www.agilent.com/environment/product for more information.

VIII U1731C/U1732C/U1733C User’s Guide

Declaration of Conformity (DoC)

The Declaration of Conformity (DoC) for this instrument is available on the

Agilent website. You can search the DoC by its product model or description at the web address below.

http://regulations.corporate.agilent.com/DoC/search.htm

N O T E

If you are unable to search for the respective DoC, please contact your local Agilent representative.

U1731C/U1732C/U1733C User’s Guide IX

X

THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

U1731C/U1732C/U1733C User’s Guide

Table of Contents

1 Introduction

About This Manual

2

Documentation map 2

Safety notes 2

Preparing Your LCR Meter 3

Check the shipment

3

Install the battery

3

Turn on your LCR meter 5

Automatic Power-Off (APO)

6

Enabling the backlight 6

Selecting the range

7

Adjusting the tilt stand

8

Connecting the IR-USB cable

9

Power-on options

10

Your LCR Meter in Brief

11

Dimensions 11

Overview

13

Keypad 15

Display screen 18

Input terminals 22

Cleaning Your LCR Meter 23

2 Features and Functions

Making Measurements 26

Auto Identification (Ai) function

26

Measuring inductance (L)

29

Measuring capacitance (C)

31

Measuring resistance (R) 33

Measuring impedance (Z)

35

U1731C/U1732C/U1733C User’s Guide XI

XII

Measuring dissipation factor/quality factor/phase angle

(D/Q/

θ ) 37

Changing the test frequency 37

Selecting parallel/series circuit mode (P/S)

37

Setting the standard reference tolerance (Tol%)

38

Enabling ESR measurements

39

Enabling DCR measurements

39

Additional Features 40

Freezing the display (Hold) 40

Enabling the static recording mode (Rec)

40

Setting the high/low limit comparison (Limit) 42

Making relative measurements (Null)

45

Performing the open/short CAL

46

3 Setup Options

Using the Setup Menu 52

Editing numerical values 53

Setup Menu Summary 54

Setup Menu Items

56

Changing the initial power-on behavior 56

Changing the Ai function’s phase angle condition

63

Changing the power-on limit category and set

65

Changing the user high/low limit values 66

Changing the baud rate 68

Changing the parity check

69

Changing the data bits 70

Changing the beep frequency 71

Locking the push buttons 72

Changing the auto power-off and backlight time-outs 73

Resetting the Setup items 74

U1731C/U1732C/U1733C User’s Guide

4 Characteristics and Specifications

Product Characteristics 76

Specification Assumptions

77

Electrical Specifications

78

Impedance/Resistance/DCR specifications 78

Capacitance specifications 79

Inductance specifications

80

Phase angle of impedance specifications

81

Dissipation/Quality factor specifications 82

Test signal specifications 83

Source impedance of impedance/resistance measurement 84

Source impedance of capacitance measurement

85

Source impedance of inductance measurement 86

SMD Tweezers Specifications

87

Electrical characteristics 88

U1731C/U1732C/U1733C User’s Guide XIII

XIV U1731C/U1732C/U1733C User’s Guide

List of Figures

Figure 1-1

Installing the batteries 4

Figure 1-2

Power-on button 5

Figure 1-3

Tilt-stand adjustment and IR cable connection 8

Figure 1-4

Agilent GUI Data Logger Software 9

Figure 1-5

Width dimensions 11

Figure 1-6

Height and depth dimensions 12

Figure 1-7

Front panel 13

Figure 1-8

Rear panel 14

Figure 2-1

Using the Ai function 26

Figure 2-2

Inductance measurement with Q factor 29

Figure 2-3

Measuring inductance 30

Figure 2-4

Capacitance measurement with D factor 31

Figure 2-5

Measuring capacitance 32

Figure 2-6

Resistance measurement 33

Figure 2-7

Measuring resistance 34

Figure 2-8

Impedance measurement with theta 35

Figure 2-9

Measuring impedance 36

Figure 2-10

Component above setting tolerance 38

Figure 2-11

ESR measurement with theta 39

Figure 2-12

DCR measurement 39

Figure 2-13

Using the Hold function 40

Figure 2-14

Using the Rec function 41

Figure 2-15

Using the Limit function 43

Figure 2-16

High and low limit values 44

Figure 2-17

nGo and Go indications 44

Figure 2-18

Using the Null function 45

Figure 2-19

Using the Cal function 47

Figure 2-20

Open calibration and short calibration prompts 48

Figure 3-1

Changing the power-on measurement type 57

Figure 3-2

Changing the power-on test frequency 58

Figure 3-3

Changing the power-on secondary parameter and measurement mode for inductance (L) measurements 59

Figure 3-4

Changing the power-on secondary parameter and measurement mode for capacitance (C)

U1731C/U1732C/U1733C User’s Guide XV

measurements 60

Figure 3-5

Changing the power-on secondary parameter and measurement mode for resistance (R) measurements 61

Figure 3-6

Changing the power-on open/short CAL 62

Figure 3-7

Changing the Ai function’s phase angle condition 64

Figure 3-8

Changing the power-on limit and category set 65

Figure 3-9

Changing the user high/low limit values 67

Figure 3-10

Changing the baud rate 68

Figure 3-11

Changing the parity check 69

Figure 3-12

Changing the data bits 70

Figure 3-13

Changing the beep frequency 71

Figure 3-14

Locking the push buttons 72

Figure 3-15

Changing the auto power-off and backlight time-outs 73

Figure 3-16

Resetting the Setup items 74

Figure 4-1

U1782B SMD tweezers 87

XVI U1731C/U1732C/U1733C User’s Guide

List of Tables

Table 1-1

Battery level indicator 5

Table 1-2

Power-on options 10

Table 1-3

Front panel parts 13

Table 1-4

Rear panel parts 14

Table 1-5

Keypad functions 15

Table 1-6

General annunciators 18

Table 1-7

Measurement units display 21

Table 1-8

Input terminal/socket connections 22

Table 2-1

Auto identification phase angle rules 27

Table 2-2

Auto identification series/parallel rules for resistance measurements 27

Table 2-3

Auto identification series/parallel rules for capacitance measurements 28

Table 2-4

Auto identification series/parallel rules for inductance measurements 28

Table 2-5

Available test frequencies 37

Table 2-6

Factory default high and low limit values 42

Table 3-1

Setup menu key functions 52

Table 3-2

Setup menu item descriptions 54

Table 3-3

Auto identification phase angle rules 63

Table 3-4

Default user high/low limit values 66

Table 4-1

Impedance/Resistance/DCR specifications 78

Table 4-2

Capacitance specifications 79

Table 4-3

Inductance specifications 80

Table 4-4

Phase angle of impedance specifications 81

Table 4-5

Dissipation/Quality factor specifications 82

Table 4-6

Test signal specifications 83

Table 4-7

Source impedance of impedance/resistance measurement 84

Table 4-8

Source impedance of capacitance measurement 85

Table 4-9

Source impedance of inductance measurement 86

Table 5

U1782B SMD tweezers electrical characteristics 88

U1731C/U1732C/U1733C User’s Guide XVII

XVIII U1731C/U1732C/U1733C User’s Guide

U1731C/U1732C/U1733C Handheld LCR Meter

User’s Guide

1

Introduction

About This Manual 2

Documentation map 2

Safety notes 2

Preparing Your LCR Meter 3

Check the shipment 3

Install the battery 3

Turn on your LCR meter 5

Automatic Power-Off (APO) 6

Enabling the backlight 6

Selecting the range 7

Adjusting the tilt stand 8

Connecting the IR-USB cable 9

Power-on options 10

Your LCR Meter in Brief 11

Dimensions 11

Overview 13

Keypad 15

Display screen 18

Input terminals 22

Cleaning Your LCR Meter 23

This chapter teaches you how to set up your LCR meter for the first time. An introduction to all the features and capabilities of the LCR meter is also given.

Agilent Technologies

1

2

1 Introduction

About This Manual

About This Manual

The descriptions and instructions in this manual apply to the Agilent U1731C, U1732C, and U1733C handheld LCR meters (hereafter referred to as the LCR meter).

The model U1733C appears in all illustrations.

Documentation map

The following manuals and software are available for your

LCR meter. For the very latest version, please visit our website at: http://www.agilent.com/find/hhTechLib .

Check the manual revision on the first page of each manual.

• User’s Guide. This manual.

• Quick Start Guide. Printed copy for outdoor use, included with shipment.

• Agilent GUI Data Logger Software, Quick Start Guide,

and Help. Free download at the Agilent website.

Safety notes

Safety notes are used throughout this manual (see the

“Safety Notices” section for format examples). Familiarize

yourself with each of the notes and its meaning before operating your LCR meter.

More pertinent safety notes for using this product are located under the

“Safety Considerations”

section.

Do not proceed beyond a safety notice until the indicated conditions are fully understood and met.

U1731C/U1732C/U1733C User’s Guide

Introduction 1

Preparing Your LCR Meter

Preparing Your LCR Meter

Check the shipment

When you receive your LCR meter, check the shipment according to the following procedure.

1 Inspect the shipping container for damage. Signs of damage may include a dented or torn shipping container or cushioning material that indicates signs of unusual stress or compacting. Save the packaging material in case the LCR meter needs to be returned.

2 Carefully remove the contents from the shipping container, and verify that the standard accessories and your ordered options are included in the shipment according to the standard shipped items list found in the printed copy of the U1731C/U1732C/U1733C Quick Start

Guide.

3 For any question or problems, refer to the Agilent contact numbers on the back of this manual.

Install the battery

Your LCR meter is powered by a single 9 V alkaline battery

(included with the shipment). When you receive your LCR meter, the 9 V alkaline battery is not installed.

Use the following procedure to install the battery.

C A U T I O N

Before you proceed with the battery installation, remove all cable connections to the terminals and ensure that the LCR meter is turned

OFF. Use only the battery type specified in

“Product Characteristics” on page 76.

U1731C/U1732C/U1733C User’s Guide 3

1 Introduction

Preparing Your LCR Meter

1 Open the battery cover. Lift the tilt stand. Loosen the screw with a suitable Phillips screwdriver and remove the

battery cover as shown in Figure 1- 1 .

2 Insert the battery. Observe the proper battery polarity.

The terminal ends of the battery are indicated inside the battery compartment.

3 Close the battery cover. Place the battery cover back in its original position and tighten the screw.

4

WA R N I N G

Figure 1-1 Installing the batteries

The battery level indicator in the lower right- hand corner of the display indicates the relative condition of the battery.

Table 1- 1

describes the various battery levels the indicator represents.

To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the low battery indicator appears. Do not discharge the battery by shorting the battery or reverse the battery polarity.

U1731C/U1732C/U1733C User’s Guide

Introduction 1

Preparing Your LCR Meter

C A U T I O N

To avoid instruments being damage from battery leakage:

Always remove dead batteries immediately.

Always remove the battery and store it separately if the LCR meter is not going to be used for a long period.

Table 1-1 Battery level indicator

Indication Battery capacity

Full capacity

2/3 capacity

1/3 capacity

Almost empty (less than one day)

[1]

(Flashing periodically)

[1] Battery change advised. Always use the specified battery type listed in

page 76 .

Turn on your LCR meter

To power ON your LCR meter, press the power- on button once. The LCR meter powers up in the auto identification

(Ai) mode (see page 26 ) when turned on for the first time.

Power-on button

Figure 1-2 Power-on button

U1731C/U1732C/U1733C User’s Guide 5

6

1 Introduction

Preparing Your LCR Meter

N O T E

To power OFF your LCR meter, press the power- on button again.

You can change the power-on behavior of your LCR meter for subsequent power cycles. See

“Changing the initial power-on behavior” on page 56 for

more information on changing the LCR meter’s power-on setting.

Automatic Power-Off (APO)

Your LCR meter automatically turns off after 5 minutes

(default) if no keys are pressed. Pressing any key will turn the LCR meter back on after it is powered off automatically.

The annunciator is shown on the bottom left of the display when the APO function is enabled.

N O T E

To change the time-out period or completely disable the APO function,

refer to “Changing the auto power-off and backlight time-outs” on page 73

.

If an external power adapter is used, the APO function will be disabled.

Enabling the backlight

If viewing the display becomes difficult in low- light conditions, press Tol% for more than 1 second to activate the

LCD backlight.

To conserve battery life, a user- adjustable time- out controls how long the backlight stays on. The default time- out is 30 seconds.

N O T E

To change the time-out period or completely disable the backlight, refer to

“Changing the auto power-off and backlight time-outs” on page 73

.

If an external power adapter is used, the backlight time-out will be disabled.

U1731C/U1732C/U1733C User’s Guide

Introduction 1

Preparing Your LCR Meter

Selecting the range

Pressing Range

Auto

switches the LCR meter between manual and autoranging. It also cycles through the available LCR meter ranges when manual ranging is enabled.

Autoranging is convenient because the LCR meter automatically selects an appropriate range for sensing and displaying each measurement. However, manual ranging results in better performance, since the LCR meter does not have to determine which range to use for each measurement.

In autorange, the LCR meter selects the lowest range to display the highest available precision (resolution) for the input signal. If manual range is already enabled, press Range

Auto for more than 1 second to enter the autoranging mode.

If autoranging is enabled, press Range

Auto

to enter the manual range mode.

Each additional press of Range

Auto

sets the LCR meter to the next higher range, unless it is already in the highest range, at which point the range switches to the lowest range.

U1731C/U1732C/U1733C User’s Guide 7

8

1 Introduction

Preparing Your LCR Meter

Adjusting the tilt stand

To adjust the LCR meter to a 60° standing position, pull the tilt- stand outward to its maximum reach.

To PC (host)

IR-USB cable

Figure 1-3 Tilt-stand adjustment and IR cable connection

U1731C/U1732C/U1733C User’s Guide

Introduction 1

Preparing Your LCR Meter

Connecting the IR-USB cable

You can use the IR communication link (IR communication port, located at the rear panel) and the Agilent GUI Data

Logger software to control your LCR meter remotely. Hence, you can only perform data logging operations in the Agilent

GUI Data Logger application with the LCR meter connected via the USB- IR cable to the PC.

Ensure that the Agilent logo on the U5481A IR- USB cable

(purchased separately) connected to the LCR meter is facing up. Firmly push the IR head into the LCR meter’s IR communication port until it snaps into place (see

Figure 1- 3

).

Refer to the Agilent GUI Data Logger Software Quick Start

Guide and Help for more information on the IR communication link and the Agilent GUI Data Logger software.

Figure 1-4 Agilent GUI Data Logger Software

The Agilent GUI Data Logger software and its supporting documents (Quick Start Guide and Help) are available for free download at http://www.agilent.com/find/hhTechLib .

You may purchase a U5481A IR- USB cable from an Agilent

Sales Office nearest to you.

U1731C/U1732C/U1733C User’s Guide 9

1 Introduction

Preparing Your LCR Meter

Power-on options

Some options can be selected only while you turn the LCR meter on. These power- on options are listed in the table below.

To select a power- on option, press and hold the specified

key in Table 1- 2

while turning the LCR meter ON ( ).

Table 1-2 Power-on options

Key

Hold

Rec

Range

Auto

Null

Cal

Description

Tests the LCD.

All annunciators are displayed in the LCD. Press any key to exit this mode.

Simulates the Auto Power-Off (APO) mode. Press any key to turn the LCR meter back on and resume normal operation.

Checks the firmware version.

The LCR meter’s firmware version will be shown on the primary display. Press any key to exit this mode.

Performs the Open/Short CAL on all frequencies and all ranges for the User mode ( ).

[1]

ZLCR

P S

Enters the Setup menu.

See

Chapter 3 , “Setup Options,” starting on page 51 for

more information. Press and hold ZLCR

P S

for more than 1 second to exit this mode.

[1] The Open/Short CAL requires approximately 1.5 minutes to complete.

10 U1731C/U1732C/U1733C User’s Guide

Your LCR Meter in Brief

Dimensions

Front view

87 mm

Introduction 1

Your LCR Meter in Brief

Figure 1-5 Width dimensions

U1731C/U1732C/U1733C User’s Guide 11

1 Introduction

Your LCR Meter in Brief

Rear and side view

41 mm

184 mm

12

Figure 1-6 Height and depth dimensions

U1731C/U1732C/U1733C User’s Guide

Introduction 1

Your LCR Meter in Brief

Overview

Front panel

The front panel parts of your LCR meter are described in this section. Click the respective “Learn more” pages in

Table 1- 3

for more information on each part.

1

2

Figure 1-7 Front panel

Table 1-3 Front panel parts

Legend

1

2

3

Description

Display screen

Keypad

Input terminals and sockets

U1731C/U1732C/U1733C User’s Guide

3

Learn more on:

page 18

page 15

page 22

13

1 Introduction

Your LCR Meter in Brief

Rear panel

The rear panel parts of your LCR meter are described in this section. Click the respective “Learn more” pages in

Table 1- 4

for more information on each part.

1

2

3

4

14

Figure 1-8 Rear panel

Table 1-4 Rear panel parts

Legend

1

2

3

4

Description

IR communication port

Tilt stand

Battery cover (lift the tilt stand for access)

External power adapter input jack

[1]

Learn more on:

page 9

page 8

page 3

-

[1] The external power adapter input jack requires an input voltage of +12 VDC.

U1731C/U1732C/U1733C User’s Guide

Introduction 1

Your LCR Meter in Brief

Keypad

The operation of each key is explained below. Pressing a key enables a function, displays a related annunciator, and emits a beep.

Each key operation of the U1731C/U1732C/U1733C keypad

(shown in

Figure 1- 7

) is described in Table 1- 5 . Click the

respective “Learn more” pages in

Table 1- 5 for more

information on each function.

Table 1-5 Keypad functions

Legend

Less than 1 second

Function when pressed for:

More than 1 second

Learn more on:

ESR

Hold

Rec

Save

Turns the LCR meter on or off.

-

page 5

Starts or stops the auto identification mode.

Press again while the annunciator is shown to exit this mode.

Holds or releases the present reading on the display.

Press Hold

Rec

again to update the reading automatically once it is stable.

Press Hold

Rec

for more than 1 second to exit this mode.

Enables or disables the ESR (equivalent series resistance) mode.

Press for more than 1 second to exit this mode. The LCR meter will return to capacitance measurement by default.

Starts or stops the static recording mode.

Press Hold

Rec

again to cycle through the maximum (Max), minimum (Min), average

(Avg), and present (MaxMinAvg) readings.

Press Hold

Rec

for more than 1 second to exit this mode.

page 26

page 40

U1731C/U1732C/U1733C User’s Guide 15

1 Introduction

Your LCR Meter in Brief

Table 1-5 Keypad functions (continued)

Legend

DQO

Limit

Freq.

DCR

Range

Auto

Less than 1 second

Function when pressed for:

More than 1 second

Switches between the dissipation factor (D), quality factor (Q), or phase angle ( measurement.

Selects a test frequency.

θ)

Press Freq.

again to cycle through the various test frequencies (100 Hz to

100 kHz).

Enables or disables the limit comparison mode.

While the annunciator is flashing,

DQO

Limit

Range

• press and

Auto

again to toggle between high (H) or low (L) limits, then

Freq.

• use the and keys to select a high/low limit set (1 to 16).

Press Hold

Rec

to start the limit sorting (with the selected limit set), or

If no activity is detected after 3 seconds, the limit comparison will begin.

Press

DQO

Limit

for more than 1 second to exit this mode.

U1733C only: Enables or disables the DCR

(direct current resistance) mode.

Freq.

Press for more than 1 second to exit this mode. The LCR meter will return to inductance measurement by default.

Disables autoranging and sets a manual range.

Press Range

Auto

again to cycle through each available measurement range.

Enables autoranging.

Learn more on:

page 37

and

page 42

page 37

page 7

ZLCR

P S

page 27 to

page 35

and

page 37

Tol%

Switches between impedance (Z), inductance (L), capacitance (C), and resistance (R) measurement.

Toggles between parallel and series circuit mode.

Sets the tolerance mode.

Connect/insert an appropriate component into the input terminals/sockets and press

Tol% to set the value shown on the secondary display as the standard reference value.

Tol% Press again to cycle through the various tolerance values (1% to 20%).

Turns the LCD backlight on for 15 seconds

(default) or off.

To change the backlight time-out refer to

“Changing the auto power-off and backlight time-outs” on page 73.

page 38

and

page 6

16 U1731C/U1732C/U1733C User’s Guide

Introduction 1

Your LCR Meter in Brief

Table 1-5 Keypad functions (continued)

Legend

Null

Cal

Less than 1 second

Function when pressed for:

More than 1 second

Sets the null/relative mode.

The displayed value is saved as a reference to be subtracted from subsequent measurements.

Press Null

Cal

again to cancel the null mode.

Enters the open/short calibration mode for the selected range and test frequency.

Follow the prompts on the screen (open or short connector) and press Null

Cal

to begin the calibration process.

The LCR meter will return to normal display when the calibration is complete.

Learn more on:

page 45

and

page 46

U1731C/U1732C/U1733C User’s Guide 17

1 Introduction

Your LCR Meter in Brief

Display screen

The function that each display annunciator of your LCR meter is associated to is described in this section. See also

“Measurement units” on page 21 for a list of available

measurement signs and notations.

General display annunciators

The general display annunciators of your LCR meter are described in the table below.

Each display annunciator of the U1731C/U1732C/U1733C

display screen (shown in Figure 1- 7

) is described in

Table 1- 6

. Click the respective “Learn more” pages in

Table 1- 6

for more information on each annunciator.

Table 1-6 General annunciators

Legend Description

Remote control via PC indicator

Equivalent series resistance indicator

Resistance measurement by direct current indicator

LCR meter using open/short CAL settings by factory

LCR meter using open/short CAL setting by user

Measuring frequency of test signal is 100 Hz

Measuring frequency of test signal is 120 Hz

Measuring frequency of test signal is 1 kHz

Measuring frequency of test signal is 10 kHz

Measuring frequency of test signal is 100 kHz

Learn more on:

page 9

page 46

page 37

18 U1731C/U1732C/U1733C User’s Guide

Introduction 1

Your LCR Meter in Brief

Table 1-6 General annunciators (continued)

Legend Description

Tolerance mode indicator for sorting L, C, or R

Learn more on:

Tolerance set to 1% for sorting capacitance

Tolerance set to 5% for sorting capacitance

Tolerance set to 10% for sorting capacitance

Tolerance set to 20% for sorting capacitance

Data hold mode indicator

Audible alert indicator for tolerance or limit mode

Dissipation factor indicator

page 38

page 40

page 71

Quality factor indicator

page 37

Phase angle of impedance indicator

Secondary display -

Measurement units for the secondary display

Impedance measurement indicator

Inductance measurement indicator

Capacitance measurement indicator

Resistance measurement indicator

page 21

page 35

page 29

page 31

page 33

U1731C/U1732C/U1733C User’s Guide 19

1 Introduction

Your LCR Meter in Brief

Table 1-6 General annunciators (continued)

Legend Description

Present reading shown on primary display

Maximum reading shown on primary display

Minimum reading shown on primary display

Learn more on:

page 40

Averaged reading shown on primary display

Relative (Null) indicator

Autoranging indicator

Limit mode indicator

Reading out of HI limit

Reading out of LO limit

Auto power-off indicator

page 45

page 7

page 42

page 6

Primary display -

page 21

Measurement units for the primary display

Parallel mode indicator

Series mode indicator

Battery capacity indicator

page 37

page 5

20 U1731C/U1732C/U1733C User’s Guide

Introduction 1

Your LCR Meter in Brief

Measurement units

The available signs and notations for each measurement function in your LCR meter are described in

Table 1- 7

. The units listed below are applicable to the primary display measurements of your LCR meter.

Table 1-7 Measurement units display k m

μ n

Sign/Notation

M p

°

%

μH, mH, H pF, nF, μF, mF

Ω, kΩ, MΩ kHz, Hz

Description

Mega 1E+06 (1000000) kilo milli

1E+03 (1000)

1E–03 (0.001) micro 1E–06 (0.000001) nano 1E–09 (0.000000001) pico 1E–12 (0.000000000001)

Degree, unit for phase angle measurement

Percentage, unit for tolerance measurement

Henry, units for inductance measurement

Farad, units for capacitance measurement

Ohm, units for resistance and impedance measurement

Hertz, units for frequency measurement

U1731C/U1732C/U1733C User’s Guide 21

1 Introduction

Your LCR Meter in Brief

Input terminals

The terminal and socket connections of your LCR meter are described in the table below.

WA R N I N G

To avoid damaging this instrument, do not exceed the input limit. Do not apply voltage to input terminals. Discharge the capacitor before testing.

Table 1-8 Input terminal/socket connections

Input terminal/socket

Description

+

Positive terminal/component socket

-

Negative terminal/component socket

GUARD

Guard terminal/component socket

22 U1731C/U1732C/U1733C User’s Guide

Introduction 1

Cleaning Your LCR Meter

Cleaning Your LCR Meter

WA R N I N G

To avoid electrical shock or damage to the LCR meter, ensure that the insides of the casing stay dry at all times.

Dirt or moisture in the terminals can distort readings.

Follow the steps below to clean your LCR meter.

1 Turn the LCR meter off and remove the test leads.

2 Turn the LCR meter over and shake out any dirt that may have accumulated in the terminals.

3 Wipe the case with a damp cloth and mild detergent — do not use abrasives or solvents.

4 Wipe the contacts in each terminal with a clean swab dipped in alcohol.

U1731C/U1732C/U1733C User’s Guide 23

1 Introduction

Cleaning Your LCR Meter

THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

24 U1731C/U1732C/U1733C User’s Guide

U1731C/U1732C/U1733C Handheld LCR Meter

User’s Guide

2

Features and Functions

Making Measurements 26

Auto Identification (Ai) function 26

Measuring inductance (L) 29

Measuring capacitance (C) 31

Measuring resistance (R) 33

Measuring impedance (Z) 35

Measuring dissipation factor/quality factor/phase angle (D/Q/

θ)

37

Changing the test frequency 37

Selecting parallel/series circuit mode (P/S) 37

Setting the standard reference tolerance (Tol%) 38

Enabling ESR measurements 39

Enabling DCR measurements 39

Additional Features 40

Freezing the display (Hold) 40

Enabling the static recording mode (Rec) 40

Setting the high/low limit comparison (Limit) 42

Making relative measurements (Null) 45

Performing the open/short CAL 46

This chapter provides detailed information on the features and functions that are available in your LCR meter.

Agilent Technologies

25

2 Features and Functions

Making Measurements

Making Measurements

Auto Identification ( Ai) function

Press to automatically identify the appropriate measurement required for the device- under- test (DUT).

26

N O T E

Figure 2-1 Using the Ai function

The annunciator will flash while the LCR meter identifies the DUT, and

selects an appropriate measurement in the primary display (L, C, or R) and secondary display (D, Q, or

θ),

selects an appropriate range, and

selects an appropriate measuring mode (series or parallel).

The Ai function helps to identify L, C, and R measurements automatically according to the angle of impedance detected in the DUT. See

Table 2-1 for

the phase angle rules.

The default phase angle condition is set to 10°. You can change this angle in the Setup menu from 5° to 45°. See

“Changing the Ai function’s phase angle condition” on page 63 for more information.

U1731C/U1732C/U1733C User’s Guide

Table 2-2 Auto identification series/parallel rules for resistance measurements

Resistance range

200 M Ω

20 M Ω

2000 k Ω

200 k Ω

20 k Ω

2000 Ω

200 Ω

20 Ω

2 Ω

Down range

Parallel

Parallel

Parallel

Parallel

Parallel

Parallel

Parallel

Series

Series

Features and Functions 2

Making Measurements

The measurement mode (series or parallel) will be automatically identified from the autoranging direction.

Table 2- 2

,

Table 2- 3 , and Table 2- 4 list down the

series/parallel rules used.

Table 2-1 Auto identification phase angle rules

Phase angle

[1]

Set <

θ < +Set

θ ≥ +Set

θ ≤ –Set

[1] Where ±Set is the phase angle selected.

Primary display

R

L

C

Secondary display

θ

Q

D

Up range

Parallel

Parallel

Parallel

Parallel

Series

Series

Series

Series

Series

U1731C/U1732C/U1733C User’s Guide 27

2 Features and Functions

Making Measurements

20 mF

2000

μF

200

μF

20

μF

2000 nF

200 nF

20 nF

2000 pF

200 pF

20 pF

Table 2-3 Auto identification series/parallel rules for capacitance measurements

Range

100 Hz

Down Up

120 Hz

Down Up Down

1 kHz

Up

Series Series Series Series -

Series Series Series Series Series Series

10 kHz

Down Up

-

-

-

-

100 kHz

Down Up

-

-

-

-

Series Series Series Series Series Series Series Series -

Series Parallel Series Parallel Series Series Series Series Series Series

Series Parallel Series Parallel Series Parallel Series Series Series Series

Series Parallel Series Parallel Series Parallel Series Parallel Series Series

Parallel Parallel Parallel Parallel Series Parallel Series Parallel Series Parallel

Parallel Parallel Parallel Parallel Parallel Parallel Series Parallel Series Parallel

-

-

-

-

-

-

-

Parallel

-

Parallel

-

Parallel

Parallel

Parallel

Parallel

Series

Parallel

Parallel

Parallel

2000 H

200 H

20 H

2000 mH

200 mH

20 mH

2000

μH

200

μH

20

μH

Table 2-4 Auto identification series/parallel rules for inductance measurements

Range

100 Hz

Down Up

120 Hz

Down Up Down

1 kHz

Up

10 kHz

Down Up

Parallel Parallel Parallel Parallel Parallel Parallel -

Parallel Parallel Parallel Parallel Parallel Parallel Parallel Parallel

100 kHz

Down Up

-

-

-

-

Parallel Series Parallel Series Parallel Parallel Parallel Parallel Parallel Parallel

Parallel Series Parallel Series Parallel Series Parallel Parallel Parallel Parallel

Parallel Series Parallel Series Parallel Series Parallel Series Parallel Parallel

Series Series Series Series Parallel Series Parallel Series Parallel Series

Series Series Series Series Series Series Parallel Series Parallel Series

Series Series Series Series Parallel Series

Series Series Series Series

28 U1731C/U1732C/U1733C User’s Guide

Features and Functions 2

Making Measurements

Measuring inductance (L)

Set up your LCR meter to measure inductance as shown in

Figure 2- 3

.

N O T E

It is recommended that you perform the Open/Short calibration (see

page 46 ) before testing to achieve optimum precision for all inductance,

capacitance, and resistance measurements at either the highest or lowest ranges.

1 Press to power on the LCR meter.

2 Press Freq.

to select a suitable test frequency, and i press to enable the auto identification function; or

ii alternatively press ZLCR

P S

to select inductance measurement.

Figure 2-2 Inductance measurement with Q factor

3 Insert an inductor into the component socket or connect the test clip to the component leads as required.

4 Press DQO

Limit

to change the secondary display measurement

(D, Q, or

θ).

5 Read the displays.

U1731C/U1732C/U1733C User’s Guide 29

2 Features and Functions

Making Measurements

3

1

Freq.

>

1

OR

Freq.

> ZLCR

P S o

2

Inductor

OR

Surface Mount

Device (SMD)

U1782B

SMD Tweezers

Figure 2-3 Measuring inductance

30 U1731C/U1732C/U1733C User’s Guide

Features and Functions 2

Making Measurements

Measuring capacitance (C)

Set up your LCR meter to measure capacitance as shown in

Figure 2- 5

.

WA R N I N G

To avoid electrical hazards, discharge the capacitor to be tested before measuring.

1 Press to power on the LCR meter.

2 Press Freq.

to select a suitable test frequency, and i press to enable the auto identification function; or

ii alternatively press ZLCR

P S

to select capacitance measurement.

Figure 2-4 Capacitance measurement with D factor

3 Insert a capacitor into the component socket or connect the test clip to the component leads as required.

4 Press DQO

Limit

to change the secondary display measurement

(D, Q, or

θ).

5 Read the displays.

U1731C/U1732C/U1733C User’s Guide 31

2 Features and Functions

Making Measurements

3

1

Freq.

>

1

OR

Freq.

> ZLCR

P S

2

Capacitor

OR

Surface Mount

Device (SMD)

U1782B

SMD Tweezers

Figure 2-5 Measuring capacitance

32 U1731C/U1732C/U1733C User’s Guide

Features and Functions 2

Making Measurements

Measuring resistance (R)

Set up your LCR meter to measure resistance as shown in

Figure 2- 7

.

C A U T I O N

To avoid possible damage to your LCR meter or to the equipment under test, disconnect the circuit power and discharge all high-voltage capacitors before measuring resistance.

1 Press to power on the LCR meter.

2 Press Freq.

to select a suitable test frequency, and i press to enable the auto identification function; or

ii alternatively press ZLCR

P S

to select resistance measurement.

Figure 2-6 Resistance measurement

3 Insert a resistor into the component socket or connect the test clip to the component leads as required.

4 Read the display.

U1731C/U1732C/U1733C User’s Guide 33

2 Features and Functions

Making Measurements

3

1

Freq.

>

1

OR

Freq.

> ZLCR

P S

2

Resistor

OR

Surface Mount

Device (SMD)

U1782B

SMD Tweezers

Figure 2-7 Measuring resistance

34 U1731C/U1732C/U1733C User’s Guide

N O T E

Features and Functions 2

Making Measurements

Measuring impedance (Z)

All circuit components, resistors, capacitors, and inductors have parasitic components. These include, for example, unwanted resistance in capacitors, unwanted capacitance in inductors, and unwanted inductance in resistors. Thus, simple components should be modeled as complex impedances.

Set up your LCR meter to measure impedance as shown in

Figure 2- 9

.

To learn more about impedance measurement theories, refer to the

Impedance Measurement Handbook. This document can be downloaded from our website at http://www.agilent.com/find/lcrmeters .

1 Press to power on the LCR meter.

2 Press Freq.

to select a suitable test frequency, and press

ZLCR

P S

to select impedance measurement.

Figure 2-8 Impedance measurement with theta

3 Insert a component into the component socket or connect the test clip to the component leads as required.

U1731C/U1732C/U1733C User’s Guide 35

2 Features and Functions

Making Measurements

4 Press DQO

Limit

to change the secondary display measurement

(D, Q, or

θ).

5 Read the displays.

3

Surface Mount

Device (SMD)

2 Device-Under-Test

(DUT)

1

Freq.

> ZLCR

P S

OR

Figure 2-9 Measuring impedance

36

U1782B

SMD Tweezers

U1731C/U1732C/U1733C User’s Guide

Features and Functions 2

Making Measurements

Measuring dissipation factor/quality factor/phase angle

(D/Q/

θ)

The dissipation factor (D), quality factor (Q), and phase angle (

θ) values can be displayed interchangeably by pressing the DQO

Limit

key when the LCR meter is set to the inductance, capacitance, or impedance measurement mode.

This setting is not applicable for DCR measurement.

Changing the test frequency

The test frequency is set to 1 kHz by default. Press the Freq.

key to select a desired test frequency.

Table 2-5 Available test frequencies

Model

U1731C

U1732C

U1733C

100 Hz

120 Hz

1 kHz

10 kHz

-

100 kHz

-

-

Selecting parallel/series circuit mode (P/S)

The LCR meter can display parallel (

( ) mode data for all ranges.

) or series

Press the ZLCR

P S

key for more than 1 second to toggle the parallel and series mode.

Series mode is set as the default setting. You can, however, change this power- on behavior in the Setup menu. See

“Changing the initial power- on behavior” on page 56 for

more information on how to change the default measurement mode (parallel or series) for subsequent power cycles.

U1731C/U1732C/U1733C User’s Guide 37

2 Features and Functions

Making Measurements

Setting the standard reference tolerance (Tol%)

The tolerance ranges available are 1%, 5%, 10%, and 20%.

To enable the tolerance mode, insert an appropriate component as a standard value into the component socket or connect the test clip to the component leads, then press the

Tol% key to set this value as the standard reference tolerance.

Similarly, any value which appears on the display, such as

Hold or Max/Min/Avg (Rec), can be used as a standard value to sort components. Press Tol% again to cycle through 1%, 5%,

10%, and 20% tolerance as desired.

This function is designed for convenient component sorting.

The beeper will beep three times whenever the component under test exceeds the setting tolerance. Conversely, when the beeper beeps once, this indicates that the component is within the setting tolerance.

38

N O T E

Figure 2-10 Component above setting tolerance

The tolerance mode cannot be activated if is shown on the display or when the tested capacitance value is below 50 counts.

Tolerance mode is only available in manual ranging; therefore, activation while in autoranging will automatically set the LCR meter to manual ranging.

U1731C/U1732C/U1733C User’s Guide

Features and Functions 2

Making Measurements

Enabling ESR measurements

Press for more than 1 second to select the ESR measurement. Use the ESR measurement to measure the equivalent series resistance of the capacitor, independent of its capacitance.

Figure 2-11 ESR measurement with theta

Press for more than 1 second to exit this mode.

Enabling DCR measurements

Press Freq.

for more than 1 second to select the DCR measurement. The DCR measurement measures the resistance of an unknown component by 1 VDC.

Figure 2-12 DCR measurement

Press Freq.

for more than 1 second to exit this mode.

U1731C/U1732C/U1733C User’s Guide 39

2 Features and Functions

Additional Features

Additional Features

Freezing the display (Hold)

To freeze the display for any function, press the Hold

Rec

key.

The annunciator is shown on the display while the

Hold function is active.

40

Hold

Rec

Figure 2-13 Using the Hold function

Press Hold

Rec

again to update the reading automatically once it is stable. The annunciator flashes while waiting for the reading to be stable.

Press Hold

Rec

for more than 1 second to release the Hold function.

Enabling the static recording mode (Rec)

The static recording mode stores the maximum, minimum, and average input values during a series of measurements in the LCR meter’s memory.

U1731C/U1732C/U1733C User’s Guide

Features and Functions 2

Additional Features

When the input goes below the recorded minimum value or above the recorded maximum value, the LCR meter beeps and records the new value. The LCR meter also calculates an average of all readings taken since the static recording mode was activated.

From the LCR meter’s display, you can view the following statistical data for any set of readings:

Max: highest reading since the static recording mode was enabled

Min: lowest reading since the static recording mode was enabled

Avg: average or mean of all readings since the static recording mode was enabled

MaxMinAvg: present reading (actual input signal value)

Press the Hold

Rec

key for more than 1 second to enter the static recording mode.

> 1 s

Hold

Rec

Figure 2-14 Using the Rec function

Press Hold

Rec

again to cycle through the Max, Min, Avg, or

MaxMinAvg (present) input values.

To exit this mode, press and hold the

1 second.

Hold

Rec

key for more than

U1731C/U1732C/U1733C User’s Guide 41

2 Features and Functions

Additional Features

N O T E

Static recording captures only stable values and updates the memory; it will not record any overload ( ) value for any of the LCR functions. In addition, the LCR meter will not record values below 50 counts in capacitance measurement.

Static recording is only available in manual ranging; therefore, activation while in autoranging will automatically set the LCR meter to manual ranging.

42

Setting the high/low limit comparison (Limit)

The high and low limit comparison function helps you to sort components easily. There are 32 limit sets available (16 fixed factory sets, and 16 variable user sets).

The LCR meter will use the factory sets by default. You can set the LCR meter to use the user sets upon start- up from the Setup menu. See

“Changing the power- on limit category and set” on page 65 for more information.

Table 2- 6

shows the factory default limit values for each set.

Table 2-6 Factory default high and low limit values

F04

F05

F06

F07

Set

F01

F02

F03

F08

F09

High limit (H)

1000

1200

1500

1800

2200

2700

3300

3900

4700

Low limit (L)

900

1080

1350

1620

1980

2430

2970

3510

4230

U1731C/U1732C/U1733C User’s Guide

N O T E

Features and Functions 2

Additional Features

Table 2-6 Factory default high and low limit values (continued)

F12

F13

F14

F15

F16

Set

F10

F11

High limit (H)

5600

6800

8200

10000

12000

15000

18000

Low limit (L)

5040

6120

7380

9000

10800

13500

16200

The default values of the variable user sets are set to the same as the fixed user sets. Use the Setup menu to change the high and low limits for each

set. See “Changing the user high/low limit values” on page 66 for more

information.

Press the DQO

Limit

key for more than 1 second to activate the high/low limit mode. The last- known set number (H## or

L##) will be indicated in the secondary display.

> 1 s

DQO

Limit

Figure 2-15 Using the Limit function

U1731C/U1732C/U1733C User’s Guide 43

2 Features and Functions

Additional Features

While the annunciator is flashing, use the key to select an appropriate limit set.

or Freq.

You may press DQO

Limit

or Range

Auto

again to toggle between the high

(H) or low (L) values shown on the primary display.

44

Figure 2-16 High and low limit values

Press Hold

Rec

while the annunciator is flashing to start the comparison. (If no activity is detected after 3 seconds, the comparison will also begin.)

The LCR meter beeps three times and displays in the secondary display if the reading is greater ( ) than the high limit or lesser ( ) than the low limit.

If the reading is within the high and low limits, the meter beeps once and displays in the secondary display.

Figure 2-17 nGo and Go indications

The limit set used in the comparison is displayed after the

/ indication.

U1731C/U1732C/U1733C User’s Guide

Features and Functions 2

Additional Features

Press and hold DQO

Limit

for more than 1 second to exit this mode.

Making relative measurements (Null)

When making relative measurements, also called null, each reading is the difference between a stored (selected or measured) relative value and the input signal.

One possible application is to increase the accuracy of a resistance measurement by nulling the test lead resistance

(test leads shorted). Nulling the leads is also particularly important prior to making capacitance measurements (test leads open).

Press the Null

Cal

key to enter the relative mode and store the display reading as a reference value. The LCR meter will then display all subsequent readings relative to the reference value.

Null

Cal

Figure 2-18 Using the Null function

U1731C/U1732C/U1733C User’s Guide 45

2 Features and Functions

Additional Features

N O T E

The annunciator is shown on the display while the relative mode is active. Press Null

Cal

again to exit the relative mode.

The relative mode cannot be activated if the display value is .

Relative mode is only available in manual ranging; therefore, activation while in autoranging will automatically set the LCR meter to manual ranging.

The relative mode cannot be activated if the LCR meter is set at auto-ranging with data hold activated.

Performing the open/short CAL

The CAL function offsets (corrects) the LCR meter's internal parameters as well as external connector residues for further measuring. Performing this action will help you to correct the influence for temporary uses.

There are three types of open/short CAL available:

OS- Factory: Open/Short CAL is performed during factory calibration mode (security code protected). It covers all frequencies and all ranges.

OS- User: Open/Short CAL is performed at every power- on option interval. It covers all frequencies and all ranges.

(see

page 10

for the OS- User setup).

Open/Short CAL for single range and frequency by pressing and holding the Null

Cal

key for more than 1 second.

The corrections for the and are pre- stored in the LCR meter. They are both calibrated at the terminal ends.

You can set the LCR meter to start up using the or open/short CAL from the Setup menu (see

page 62 ).

46 U1731C/U1732C/U1733C User’s Guide

N O T E

Features and Functions 2

Additional Features

The open/short CAL function is available for fixed measurement ranges.

It is highly recommended that open/short CAL be performed before making precision measurements.

> 1 s

Null

Cal

Figure 2-19 Using the Cal function

U1731C/U1732C/U1733C User’s Guide 47

2 Features and Functions

Additional Features

Figure 2-20 Open calibration and short calibration prompts

1 Press and hold the Null

Cal

key for more than 1 second to enter the open/short CAL mode for the selected frequency and range.

2 Open/Short CAL prompts will be shown on the display.

Follow the prompts for open connector ( OPn ) or short connector ( SHor) connection and press the Null

Cal

key.

The CAL annunciator on the upper right of the display will flash indicating that the correction is in process.

3 After the open/short CAL is completed, the LCR meter will be restored to the normal display and ready for normal usage.

48 U1731C/U1732C/U1733C User’s Guide

Features and Functions 2

Additional Features

THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

U1731C/U1732C/U1733C User’s Guide 49

2 Features and Functions

Additional Features

50 U1731C/U1732C/U1733C User’s Guide

U1731C/U1732C/U1733C Handheld LCR Meter

User’s Guide

3

Setup Options

Using the Setup Menu 52

Editing numerical values 53

Setup Menu Summary 54

Setup Menu Items 56

Changing the initial power-on behavior 56

Changing the Ai function’s phase angle condition 63

Changing the power-on limit category and set 65

Changing the user high/low limit values 66

Changing the baud rate 68

Changing the parity check 69

Changing the data bits 70

Changing the beep frequency 71

Locking the push buttons 72

Changing the auto power-off and backlight time-outs 73

Resetting the Setup items 74

The following chapter describes how to change the preset features of your LCR meter.

Agilent Technologies

51

3 Setup Options

Using the Setup Menu

Using the Setup Menu

The Setup menu allows you to change a number of nonvolatile preset features. Modifying these settings affects the general operation of your LCR meter across several functions. Select a setting to edit to perform one of the following:

Switch between two values, such as on or off.

Cycle through multiple values from a predefined list.

Decrease or increase a numerical value within a fixed range.

The contents of the Setup menu are summarized in

Table 3- 2 on page 54.

Table 3-1 Setup menu key functions

Legend

ZLCR

P S

Description

Press and hold ZLCR

P S

while turning the LCR meter ON

( ) to access the Setup menu.

Press and hold ZLCR

P S

for more than 1 second to exit this mode.

DQO

Limit

Range

Auto

Press DQO

Limit

or Range

Auto

to step through the menu items.

Hold

Rec

Save

Freq.

ZLCR

P S

Press or

Freq.

at each menu item to change the preset settings. The menu item (in the secondary display) will flash to indicate that you can now change the menu item values.

Press or

Freq.

again to switch between two values, to cycle through multiple values from a list, or to decrease or increase a numerical value.

While the menu item is flashing, press

Hold

Rec

to save your changes.

While the menu item is flashing, press discard your changes.

ZLCR

P S

to

52 U1731C/U1732C/U1733C User’s Guide

Setup Options 3

Using the Setup Menu

Editing numerical values

When editing numerical values, use the DQO

Limit

and Range

Auto

to position the cursor on a numerical digit.

Press DQO

Limit

to move the cursor to the left, and

Press Range

Auto

to move the cursor to the right.

When the cursor is positioned over a digit, use the and

Freq.

keys to change the numerical digit.

Press to increment the digit, and

Press Freq.

to decrement the digit.

When you have completed your changes, save the new numerical value by pressing Hold

Rec

. (Or alternatively, if you wish to discard the changes you made press, ZLCR

P S

.)

U1731C/U1732C/U1733C User’s Guide 53

3 Setup Options

Setup Menu Summary

Setup Menu Summary

The Setup menu items are summarized in the table below.

Click the respective “Learn more” pages for more information on each menu item.

Table 3-2 Setup menu item descriptions

Legend Available settings Description

Ai, Z, L, C, R, ESR, or DCR

Set the measurement type that the LCR meter powers up in. Default is the auto identification (Ai) mode.

Learn more on:

page 56

100 Hz, 120 Hz, 1 kHz,

10 kHz, or 100 kHz

D, Q, or

D, Q, or

D, Q, or

θ and P or S

θ and P or S

θ and P or S

FACt or USEr

Set the test frequency that the LCR meter powers up in.

Default is 1 kHz.

Set the inductance (L) secondary parameter and measurement mode that the LCR meter powers up in.

Default is quality factor (Q) and series (S).

Set the capacitance (C) secondary parameter and measurement mode that the LCR meter powers up in.

Default is dissipation factor (D) and series (S).

Set the resistance (R) secondary parameter and measurement mode that the LCR meter powers up in.

Default is phase angle ( θ) and series (S).

Set the open/short CAL mode that the LCR meter powers up in. Default is factory (FACt).

05° to 45 °

Set the phase angle condition for the auto identification

(Ai) mode. Default is 10 °.

Ft01 to Ft16 or Ur01 to Ur16

Set the limit category (factory or user) and set (01 to 16) that the LCR meter powers up in. Default is Ft01.

page 58

page 59

page 60

page 61

page 62

page 63

page 65

54 U1731C/U1732C/U1733C User’s Guide

Setup Options 3

Setup Menu Summary

Table 3-2 Setup menu item descriptions (continued)

Legend Available settings Description

H01 to H16 or L01 to L16

0 to 19999

Set the high and low limits for each variable user set.

See Table 3-4 on page 66

for the user default values.

Learn more on:

page 66

9600 or 19200

En, nonE, or odd

7bit or 8bit

Set the baud rate for remote communication with a PC

(9600 or 19200). Default is 9600.

Set the parity bit for remote communication with a PC

(even, none, or odd). Default is none.

Set the data bit length for remote communication with a

PC (7-bit or 8-bit). Default is 8-bit.

2000 Hz, 3000 Hz, 4000 Hz, or oFF

Set the LCR meter’s beep frequency (2000 Hz, 3000 Hz,

4000 Hz, or off). Default is 4000 Hz.

page 68

page 69

page 70

page 71

oFF or on Lock the LCR meter’s push buttons. Default is off.

page 72

01 to 99 mins or oFF

01 to 99 s or oFF

Set the auto power-off time-out period from 1 to 99 minutes (1 hour, 39 minutes) or off. Default is 5 minutes.

Set the LCD backlight timeout period from 1 to 99 seconds

(1 minute, 39 seconds) or off. Default is 30 seconds.

page 73

dEFA Reset the LCR meter to its factory default settings.

page 74

U1731C/U1732C/U1733C User’s Guide 55

3 Setup Options

Setup Menu Items

Setup Menu Items

Changing the initial power-on behavior

You can change the power- on behavior of your LCR meter for subsequent power cycles.

Parameter

Pon-tYPE

Pon-FrEq

Pon-AUto (L)

Pon-AUto (C)

Pon-AUto (R)

Pon-oSC

Range

Ai, Z, L, C, R, ESR, or DCR

100 Hz, 120 Hz, 1 kHz, 10 kHz, or 100 kHz

D, Q, or ° θ

Parallel or Series

D, Q, or °

θ

Parallel or Series

D, Q, or ° θ

Parallel or Series

FACt or USEr

Default setting

Ai

1 kHz

Q

Series

D

Series

° θ

Series

FACt

Changing the power-on measurement type

Use this Setup item to change the LCR meter’s initial measurement type. You can set the LCR meter to start up in the

auto identification mode (Ai),

impedance measurement (Z),

inductance measurement (L),

capacitance measurement (C),

resistance measurement (R),

equivalent series resistance mode (ESR), or

direct current resistance mode (DCR) for U1733C only

56 U1731C/U1732C/U1733C User’s Guide

Setup Options 3

Setup Menu Items

The LCR meter will start up in the selected measurement type for subsequent power cycles.

Press Press

Press Press

Press Press

Figure 3-1 Changing the power-on measurement type

U1731C/U1732C/U1733C User’s Guide 57

3 Setup Options

Setup Menu Items

Changing the power-on test frequency

Use this Setup item to change the LCR meter’s initial test frequency. You can set the LCR meter to start up using a test frequency from 100 Hz to 100 kHz.

The LCR meter will start up using the selected test frequency for subsequent power cycles.

Press Press

Press

Press

Figure 3-2 Changing the power-on test frequency

58 U1731C/U1732C/U1733C User’s Guide

Setup Options 3

Setup Menu Items

Changing the power-on secondary parameter and measurement mode for inductance (L) measurements

Use this Setup item to change the inductance ( L) measurement’s initial secondary parameter — dissipation factor ( D), quality factor (Q), or phase angle (

θ) — and measurement mode — parallel or series.

The inductance ( L) measurement will start up using the selected secondary parameter and measurement mode for subsequent power cycles.

Press

Press

Press Press Press Press

Press Press

Figure 3-3 Changing the power-on secondary parameter and measurement mode for inductance (L) measurements

U1731C/U1732C/U1733C User’s Guide 59

3 Setup Options

Setup Menu Items

Changing the power-on secondary parameter and measurement mode for capacitance (C) measurements

Use this Setup item to change the capacitance ( C) measurement’s initial secondary parameter — dissipation factor ( D), quality factor (Q), or phase angle (

θ) — and measurement mode — parallel or series.

The capacitance ( C) measurement will start up using the selected secondary parameter and measurement mode for subsequent power cycles.

Press

Press

Press Press Press Press

Press Press

Figure 3-4 Changing the power-on secondary parameter and measurement mode for capacitance (C) measurements

60 U1731C/U1732C/U1733C User’s Guide

Setup Options 3

Setup Menu Items

Changing the power-on secondary parameter and measurement mode for resistance (R) measurements

Use this Setup item to change the resistance ( R) measurement’s initial secondary parameter — dissipation factor ( D), quality factor (Q), or phase angle (

θ) — and measurement mode — parallel or series.

The resistance ( R) measurement will start up using the selected secondary parameter and measurement mode for subsequent power cycles.

Press

Press

Press Press Press Press

Press Press

Figure 3-5 Changing the power-on secondary parameter and measurement mode for resistance (R) measurements

U1731C/U1732C/U1733C User’s Guide 61

3 Setup Options

Setup Menu Items

Changing the power-on open/short CAL

Use this Setup item to change the LCR meter’s initial open/short CAL to either the factory open/short CAL ( FACt), or user open/short CAL ( USEr).

The LCR meter will start up using the selected open/short

CAL for subsequent power cycles.

Press Press

Figure 3-6 Changing the power-on open/short CAL

62 U1731C/U1732C/U1733C User’s Guide

Setup Options 3

Setup Menu Items

Changing the Ai function’s phase angle condition

This setting is used with the Ai function (see page 26 ). The

Ai function helps to identify L, C, and R measurements automatically according to the angle of impedance detected in the DUT.

Use this Setup item to change the default phase angle for the Ai function between 5° and 45°.

Parameter

Ai

Range

(5 to 45)°

Default setting

10°

Table 3- 3

shows the correlation between the phase angle detected and the L, C, and R measurements selected.

Table 3-3 Auto identification phase angle rules

Phase angle

[1]

Set <

θ < +Set

θ ≥ +Set

θ ≤ –Set

Primary display

R

L

C

[1] Where ±Set is the phase angle selected.

Secondary display

θ

Q

D

U1731C/U1732C/U1733C User’s Guide 63

3 Setup Options

Setup Menu Items

Press Press

Press

Press

Press Press

Figure 3-7 Changing the Ai function’s phase angle condition

64 U1731C/U1732C/U1733C User’s Guide

Setup Options 3

Setup Menu Items

Changing the power-on limit category and set

This setting is used with the Limit comparison function

( page 42

). There are 32 limit sets available (16 fixed factory sets, and 16 variable user sets).

Use this Setup item to change the default category (factory or user) and set (1 to 16) for subsequent power cycles.

Parameter

Pon

Range

Factory (Ft01 to Ft16) or

User (Ur01 to Ur16)

Default setting

Ft01

Press Press

Press

Press

Press Press

Press Press

Figure 3-8 Changing the power-on limit and category set

U1731C/U1732C/U1733C User’s Guide 65

3 Setup Options

Setup Menu Items

Changing the user high/low limit values

This setting is used with the Limit comparison function

( page 42

). There are 16 variable user sets available.

Use this Setup item to change the high and low limits of each variable user set.

N O T E

The low limit can be set from 0 to less than or equal to the high limit, and the high limit can be set from more than or equal to the low limit to less than or equal to the maximum display count (19999).

66

Parameter

H(01 to 16) or

L(01 to 16)

Range

0 to 19999

Default setting

See

Table 3-4

Table 3- 4

shows the user default limit values for each set.

Table 3-4 Default user high/low limit values

U04

U05

U06

U07

Set

U01

U02

U03

U08

U09

U10

High limit (H)

1000

1200

1500

1800

2200

2700

3300

3900

4700

5600

Low limit (L)

900

1080

1350

1620

1980

2430

2970

3510

4230

5040

U1731C/U1732C/U1733C User’s Guide

Press

Setup Options 3

Setup Menu Items

Table 3-4 Default user high/low limit values (continued)

Set

U11

U12

U13

U14

U15

U16

High limit (H)

6800

8200

10000

12000

15000

18000

Low limit (L)

6120

7380

9000

10800

13500

16200

Press

Press

Press

Press Press

Press

Press

Press Press

Press Press Press Press

Press Press Press Press

Figure 3-9 Changing the user high/low limit values

U1731C/U1732C/U1733C User’s Guide 67

3 Setup Options

Setup Menu Items

Changing the baud rate

This setting is used with the IR communication link and the

Agilent GUI Data Logger software to control your LCR meter

remotely ( page 9 ).

Use this Setup item to change the baud rate for remote communications with a PC.

Parameter bPS

Range

(9600 or 19200) bits/second

Default setting

9600 bits/second

Press Press

Figure 3-10 Changing the baud rate

68 U1731C/U1732C/U1733C User’s Guide

Setup Options 3

Setup Menu Items

Changing the parity check

This setting is used with the IR communication link and the

Agilent GUI Data Logger software to control your LCR meter

remotely ( page 9 ).

Use this Setup item to change the parity check for remote communications with a PC.

Parameter

PAr

Range nonE, En, or odd

Default setting nonE

Press Press

Press Press

Figure 3-11 Changing the parity check

U1731C/U1732C/U1733C User’s Guide 69

3 Setup Options

Setup Menu Items

Changing the data bits

This setting is used with the IR communication link and the

Agilent GUI Data Logger software to control your LCR meter

remotely ( page 9 ).

Use this Setup item to change the number of data bits (data width) for remote communications with a PC. The number of the stop bit is always 1, and this cannot be changed.

Parameter dAt

Range

7-bit or 8-bit

Default setting

8-bit

Press Press

Figure 3-12 Changing the data bits

70 U1731C/U1732C/U1733C User’s Guide

Setup Options 3

Setup Menu Items

Changing the beep frequency

The LCR meter’s beeper alerts users to the presence of newly sensed values for static recordings, sensed values that are out of tolerance or limits set, as well as invalid key operations.

Use this Setup item to change the driving frequency of the beeper.

Parameter bEP

Range

(2000, 3000, 4000) Hz or oFF

Default setting

4000 Hz

Press Press

Press Press

Press Press

Figure 3-13 Changing the beep frequency

U1731C/U1732C/U1733C User’s Guide 71

3 Setup Options

Setup Menu Items

Locking the push buttons

Use this Setup item to lock the push buttons (keys) of your

LCR meter. If enabled, all keys will be locked (rendered unoperational) when you exit the Setup menu.

Unlock the push buttons again by entering the Setup menu

through the power- on options ( page 10

).

Parameter

LPb

Range on or oFF

Default setting oFF

Press Press

Figure 3-14 Locking the push buttons

72 U1731C/U1732C/U1733C User’s Guide

Setup Options 3

Setup Menu Items

Changing the auto power-off and backlight time-outs

The LCR meter’s automatic power- off (see page 6 ) and backlight (see page page 6

) features use timers to determine when to turn off the backlight and when to automatically turn the LCR meter off.

Parameter

APo bLt

Range

(01 to 99) minutes or oFF

(01 to 99) seconds or oFF

Default setting

05 minutes

30 seconds

Press Press Press Press

Press

Press

Press

Press

Press Press Press Press

Figure 3-15 Changing the auto power-off and backlight time-outs

U1731C/U1732C/U1733C User’s Guide 73

3 Setup Options

Setup Menu Items

Resetting the Setup items

The Setup items can be reset to their default values through this Setup item.

Press Hold

Rec

to perform the reset. The LCR meter will beep once, exit the Setup menu, and return to normal operation.

Parameter rSt

Range dEFA

Default setting dEFA

Figure 3-16 Resetting the Setup items

74 U1731C/U1732C/U1733C User’s Guide

U1731C/U1732C/U1733C Handheld LCR Meter

User’s Guide

4

Characteristics and Specifications

Product Characteristics 76

Specification Assumptions 77

Electrical Specifications 78

Impedance/Resistance/DCR specifications 78

Capacitance specifications 79

Inductance specifications 80

Phase angle of impedance specifications 81

Dissipation/Quality factor specifications 82

Test signal specifications 83

Source impedance of impedance/resistance measurement 84

Source impedance of capacitance measurement 85

Source impedance of inductance measurement 86

SMD Tweezers Specifications 87

Electrical characteristics 88

This chapter lists the characteristics, assumptions, and specifications of the U1731C, U1732C, and U1733C handheld

LCR meters.

Agilent Technologies

75

4 Characteristics and Specifications

Product Characteristics

Product Characteristics

N O T E

Product characteristics specified in the table below are applicable for all

U1731C, U1732C, and U1733C models unless stated otherwise.

POWER SUPPLY

Battery type:

1 × 9 V Alkaline battery (ANSI/NEDA 1604A or IEC 6LR61), or

1 × 9 V Zinc Chloride battery (ANSI/NEDA 1604D or IEC 6F22)

Battery life:

16 hours typical (based on new Alkaline batteries without backlight enabled)

Low battery indicator will flash when the battery voltage drops below 7.2 V

(approximately)

External DC adapter

DC 12 V ± 10% or 10.8 V

MIN

to 13.2 V

MAX

POWER CONSUMPTION

225 mVA maximum (without backlight enabled)

DISPLAY

Dual display liquid crystal display (LCD)

Primary display is 4 1/2 digits with maximum of 19999 counts

Secondary display is 3 digits with maximum of 999 counts

MEASUREMENT RATE

1 time/second, nominal

OPERATING ENVIRONMENT

Operating temperature from –10 °C to 55 °C, 0% to 80% RH

Full accuracy up to 80% RH for temperatures up to 30 °C, decreasing linearly to 50% RH at 55 °C

Altitude up to 2000 m

Pollution degree II

STORAGE COMPLIANCE

–20 °C to 70 °C, 0% to 80% RH

76 U1731C/U1732C/U1733C User’s Guide

Characteristics and Specifications 4

Specification Assumptions

SAFETY AND ELECTROMAGNETIC COMPATIBILITY (EMC) COMPLIANCE

IEC61010-1:2001/EN61010-1:2001 (Second Edition)

IEC 61326-1:2005/EN 61326-1:2006

Canada: ICES/NMB-001:Issue 4, June 2006

Australia/New Zealand: AS/NZS CISPR11:2004

TEMPERATURE COEFFICIENT

0.1 × (specified accuracy) / °C (from –10 °C to 18 °C, or 28 °C to 55 °C)

INPUT PROTECTION

Reset-able over-current protection.

DIMENSIONS (W × H × D)

87 × 184 × 41 mm

WEIGHT

337 grams (with battery)

WARRANTY

Please refer to http://www.agilent.com/go/warranty_terms

Three years for product

Three months for product’s standard accessories, unless otherwise specified

Please take note that for the product, the warranty does not cover:

• Damage from contamination

• Normal wear and tear of mechanical components

• Manuals and standard disposable batteries

CALIBRATION CYCLE

One year

Specification Assumptions

Accuracy is given as ±(% of reading + counts of least significant digit) at 23 °C ± 5 °C, with relative humidity less than 80% RH.

The measurement performed at the component test socket and necessary open and short corrections must be done prior to verifying the instrument’s accuracy.

The accuracy is verified by design and specified type tests.

U1731C/U1732C/U1733C User’s Guide 77

4 Characteristics and Specifications

Electrical Specifications

Electrical Specifications

Specification assumptions are given on page 77 .

N O T E

Impedance/Resistance/DCR specifications

Table 4-1 Impedance/Resistance/DCR specifications

DCR 100 Hz

Accuracy = A

Z

+ Offset

120 Hz 1 kHz 10 kHz 100 kHz

Range Resolution

U1733C only All models All models All models

U1733C and

U1732C only

U1733C only

2

Ω [1]

20

Ω [1]

200 Ω

[1]

2000 Ω

20 k Ω

200 k Ω

2000 k Ω

20 M

Ω [2]

200 M

Ω [2]

0.0001

0.001

0.01

0.1

0.01 k

0.1 k

Ω

Ω

0.001 k

Ω

0.001 M

0.01 M

Ω

Ω

Ω

Ω

Ω

Ω

0.7% + 50

0.7% + 8

0.2% + 3

0.2% + 3

0.2% + 3

0.5% + 5

0.5% + 5

2.0% + 8

6.0% + 80

0.7% + 50

0.7% + 8

0.2% + 3

0.2% + 3

0.2% + 3

0.5% + 5

0.5% + 5

2.0% + 8

6.0% + 80

0.7% + 50

0.7% + 8

0.2% + 3

0.2% + 3

0.2% + 3

0.5% + 5

0.5% + 5

2.0% + 8

6.0% + 80

0.7% + 50

0.7% + 8

0.2% + 3

0.2% + 3

0.2% + 3

0.5% + 5

0.5% + 5

2.0% + 8

6.0% + 80

0.7% + 50

0.7% + 8

0.2% + 3

0.2% + 3

0.2% + 3

0.5% + 5

0.7% + 5

5.0% + 8

-

1.0% + 50

0.7% + 8

0.5% + 5

0.5% + 5

0.5% + 5

0.7% + 8

-

-

-

Notes:

1 The accuracy for the 2 Ω to 200 Ω range is specified after the Null function is used to subtract the resistance of test leads and the contact resistance.

2 For the 20 M Ω and 200 MΩ range, the RH is specified for <60%.

3 Resistance measurement is specified to Q <10 and D >0.1; otherwise accuracy is specified as ( A

Z

+ Offset ) ×

1 + Q

2

.

4 ESR (Equivalent Series Resistance) measurement is specified according to the impedance measurement and range. The maximum display is up to 199.99 k Ω and the accuracy is specified as ( A

Z

+ Offset ) × 1 + Q

2

.

78 U1731C/U1732C/U1733C User’s Guide

Characteristics and Specifications 4

Electrical Specifications

Capacitance specifications

Table 4-2 Capacitance specifications

Range

20 mF

2000 μF

200 μF

20 μF

2000 nF

200 nF

20 nF

2000 pF

[1]

200 pF

[1]

20 pF

[1]

Notes:

Resolution

0.001 mF

0.1 μF

0.01 μF

0.001 μF

0.1 nF

0.01 nF

0.001 nF

0.1 pF

0.01 pF

0.001 pF

100 Hz

All models

0.5% + 8

0.5% + 5

0.3% + 3

0.2% + 3

0.2% + 3

0.2% + 3

0.5% + 5

0.5% + 10

-

-

120 Hz

Accuracy = A

C

+ Offset

1 kHz 10 kHz

All models All models

U1733C and

U1732C only

0.5% + 8

0.5% + 5

0.3% + 3

0.2% + 3

0.2% + 3

0.2% + 3

0.5% + 5

0.5% + 10

-

-

-

0.5% + 8

0.5% + 5

0.2% + 3

0.2% + 3

0.2% + 3

0.2% + 3

0.5% + 5

0.5% + 10

-

-

-

0.5% + 8

0.5% + 5

0.2% + 3

0.5% + 3

0.5% + 3

0.5% + 3

0.8% + 10

1.0% + 20

100 kHz

U1733C only

-

-

-

5.0% + 10

0.7% + 10

0.7% + 10

0.7% + 10

2.0% + 10

2.0% + 10

2.5% + 10

1 The accuracy for the 20 pF to 2000 pF range is specified after the Null function is used to subtract the stray capacitance of the test leads.

2 The accuracy for the ceramic capacitor will be influenced depending on the dielectric constant (K) of the material used to make the ceramic capacitor. For related influence factors, please refer to the Component dependency factors section in the Impedance Measurement Handbook, downloadable for free at http://www.agilent.com/find/lcrmeters .

U1731C/U1732C/U1733C User’s Guide 79

20 μH

200 μH

2000 μH

20 mH

200 mH

2000 mH

20 H

200 H

2000 H

4 Characteristics and Specifications

Electrical Specifications

Inductance specifications

Table 4-3 Inductance specifications

Range

0.001 μH

0.01 μH

0.1 μH

0.001 mH

0.01 mH

0.1 mH

0.001 H

0.01 H

0.1 H

Resolution

100 Hz

All models

-

-

0.7% + 10

0.5% + 3

0.5% + 3

0.2% + 3

0.2% + 3

0.7% + 5

1.0% + 5

120 Hz

Accuracy = A

L

+ Offset

1 kHz 10 kHz

All models All models

U1733C and

U1732C only

-

-

0.7% + 10

0.5% + 3

-

1.0% + 5

0.5% + 3

0.2% + 3

1.0% + 5

0.7% + 3

0.5% + 3

0.3% + 3

0.5% + 3

0.2% + 3

0.2% + 3

0.7% + 5

1.0% + 5

0.2% + 3

0.2% + 3

0.5% + 5

1.0% + 5

2.0% + 8

0.2% + 3

0.5% + 5

1.0% + 5

2.0% + 8

-

100 kHz

U1733C only

2.5% + 20

2.5% + 20

0.8% + 20

0.8% + 10

1.0% + 10

1.0% + 10

2.0% + 10

-

-

80 U1731C/U1732C/U1733C User’s Guide

Characteristics and Specifications 4

Electrical Specifications

Phase angle of impedance specifications

Table 4-4 Phase angle of impedance specifications

Range

–180° to 180°

Resolution

0.1°/1°

Accuracy =

θ e

AZ +

Z x

×

π

Condition

D < 1 or Q > 1

Notes:

1 The A

Z

and Offset variables are the accuracy specified at Table 4-1 , “Impedance/Resistance/DCR specifications,” on page 78.

2 The

π variable is rounded up to 3.14159.

3 The table below shows an example of the calculations for the phase angle of impedance:

Impedance

1999.9

Ω

199.9

Ω

19.9

Ω

1.9

Ω

Z

X

19999

1999

199

19

A

Z

0.2%

0.2%

0.2%

0.2%

Offset

3

3

3

3

θ e

±0.12°

±0.20°

±0.98°

±9.16°

U1731C/U1732C/U1733C User’s Guide 81

4 Characteristics and Specifications

Electrical Specifications

Dissipation/Quality factor specifications

Table 4-5 Dissipation/Quality factor specifications

Range Resolution

Z 0.001 to 999

Accuracy = D e

AZ +

Zx

×

Condition

D < 1 or Q > 1

L 0.001 to 999

AL +

Lx

×

D < 1 or Q > 1

C 0.001 to 999

AC +

Cx

×

D < 1 or Q > 1

Notes:

1 The A

Z

, A

L

, A

C

, and Offset variables are the accuracy specified at

Table 4-1

, Table 4-2

, and Table 4-3 respectively.

2 The Z x

, L x

, and C x

variables are the display count of the reading. For example, the C x

value is 8888 if the capacitance is

88.88

μF for the range of 200 μF.

3 The quality factor is the reciprocal of the dissipation factor.

4 The calculation example below is for the 120 Hz frequency:

Capacitance

88.88 μF

C

X

8888

A

C

0.3%

Offset

3

D e

0.334% + 3

82 U1731C/U1732C/U1733C User’s Guide

Test signal specifications

Table 4-6 Test signal specifications

100 Hz

120 Hz

1 kHz

10 kHz

100 kHz

DCR

Selection

All models

All models

All models

U1733C and

U1732C only

U1733C only

U1733C only

Level

Test signal level

Accuracy

0.74 Vrms

0.74 Vrms

0.74 Vrms

0.05 Vrms

0.05 Vrms

0.05 Vrms

0.70 Vrms

0.70 Vrms

1.235 V

0.05 Vrms

0.05 Vrms

0.05 V

Characteristics and Specifications 4

Electrical Specifications

Test frequency

Frequency Accuracy

100 Hz

120.481 Hz

1 kHz

0.01%

0.01%

0.01%

10 kHz

100 kHz

-

0.01%

0.01%

-

U1731C/U1732C/U1733C User’s Guide 83

4 Characteristics and Specifications

Electrical Specifications

Source impedance of impedance/resistance measurement

2 Ω

20 Ω

200 Ω

2000 Ω

20 k Ω

200 k Ω

2000 k Ω

20 M Ω

200 M Ω

Table 4-7 Source impedance of impedance/resistance measurement

Typical source impedance

120 Hz 1 kHz

Range

DCR 100 Hz

U1733C only

100 Ω

100 Ω

100 Ω

1 k Ω

10 k Ω

100 k Ω

100 k Ω

100 k Ω

100 k Ω

All models

100 Ω

100 Ω

100 Ω

1 k Ω

10 k Ω

100 k Ω

100 k Ω

100 k Ω

100 k Ω

All models

100 Ω

100 Ω

100 Ω

1 k Ω

10 k Ω

100 k Ω

100 k Ω

100 k Ω

100 k Ω

All models

100 Ω

100 Ω

100 Ω

1 k Ω

10 k Ω

100 k Ω

100 k Ω

100 k Ω

100 k Ω

10 kHz

U1733C and

U1732C only

100 Ω

100 Ω

100 Ω

1 k Ω

10 k Ω

10 k Ω

10 k Ω

100 k Ω

-

100 kHz

U1733C only

100 Ω

100 Ω

100 Ω

1 k Ω

1 k Ω

1 k Ω

-

-

-

84 U1731C/U1732C/U1733C User’s Guide

Characteristics and Specifications 4

Electrical Specifications

Source impedance of capacitance measurement

Table 4-8 Source impedance of capacitance measurement

Range

100 Hz 120 Hz

Typical source impedance

1 kHz 10 kHz 100 kHz

All models All models All models U1733C and

U1732C only

U1733C only

20 mF

2000 μF

200 μF

20 μF

2000 nF

200 nF

20 nF

2000 pF

200 pF

20 pF

100 Ω

100 Ω

100 Ω

100 Ω

1 k Ω

10 k Ω

100 k Ω

100 k Ω

-

-

100 Ω

100 Ω

100 Ω

100 Ω

1 k Ω

10 k Ω

100 k Ω

100 k Ω

-

-

-

100 Ω

100 Ω

100 Ω

100 Ω

1 k Ω

10 k Ω

100 k Ω

100 k Ω

-

-

-

100 Ω

100 Ω

100 Ω

100 Ω

1 k Ω

10 k Ω

10 k Ω

10 k Ω [1]

-

-

-

100 Ω

100 Ω

100 Ω

100 Ω

1 k Ω

1 k Ω

1 k Ω

[1] This specification value is applicable for firmware versions 00.21 and above only. For firmware versions below 00.21, the impedance value is set at 100 k Ω.

U1731C/U1732C/U1733C User’s Guide 85

4 Characteristics and Specifications

Electrical Specifications

Source impedance of inductance measurement

Table 4-9 Source impedance of inductance measurement

Range

100 Hz 120 Hz

Typical source impedance

1 kHz 10 kHz 100 kHz

20 μH

200 μH

2000 μH

20 mH

200 mH

2000 mH

20 H

200 H

2000 H

All models

-

-

100 Ω

100 Ω

100 Ω

100 Ω

1 k Ω

10 k Ω

100 k Ω

All models

-

-

100 Ω

100 Ω

100 Ω

100 Ω

1 k Ω

10 k Ω

100 k Ω

All models

-

100 Ω

100 Ω

100 Ω

100 Ω

1 k Ω

10 k Ω

100 k Ω

100 k Ω

U1733C and

U1732C only

100 Ω

100 Ω

100 Ω

100 Ω

1 k Ω

10 k Ω

10 k Ω

10 k Ω [1]

-

U1733C only

100 Ω

100 Ω

100 Ω

100 Ω

1 k Ω

1 k Ω

1 k Ω

-

-

[1] This specification value is applicable for firmware versions 00.21 and above only. For firmware versions below 00.21, the impedance value is set at 100 k Ω.

86 U1731C/U1732C/U1733C User’s Guide

Characteristics and Specifications 4

SMD Tweezers Specifications

SMD Tweezers Specifications

The Agilent U1782B is a tweezers to be used with the U1700

Series Handheld LCR Meters. This tweezers is useful when measuring SMD- type components. Plug in the base of the tweezers to the LCR meter's + (HI- SENSE), (LO- SENSE) and GUARD ends. Ensure that the orientation of the base matches the polarity of the LCR meter.

It is recommended to measure the SMD components length as well as the maximum opening of the tweezers. The length of the tweezers is approximately 815 mm (32.08 inches) (see

Figure 4- 1

).

815 ± 20 mm

150 mm

8.4 mm

Figure 4-1 U1782B SMD tweezers

SMD tweezers base

U1731C/U1732C/U1733C User’s Guide 87

4 Characteristics and Specifications

SMD Tweezers Specifications

Electrical characteristics

Table 5 U1782B SMD tweezers electrical characteristics

Parameters

Cp

Parallel capacitance

Rs

Series resistance

Ls

Series inductance

Notes:

Test condition

Tweezers open

Tweezers short

Tweezers short

100 Hz

<0.7 pF

<0.5 Ω

<1.2 µH

120 Hz

<0.7 pF

<0.5 Ω

<1.2 µH

1 kHz

<0.7 pF

<0.5 Ω

<1.2 µH

10 kHz

<0.7 pF

<0.5 Ω

<1.2 µH

1 The accuracy is specified at 23 °C ± 5 °C and <75% R.H.

2 You are recommended to perform an open/short calibration on the LCR meter before using the tweezers.

100 kHz

<0.7 pF

<0.5 Ω

<1.2 µH

88 U1731C/U1732C/U1733C User’s Guide

www.agilent.com

Contact us

To obtain service, warranty, or technical assistance, contact us at the following phone or fax numbers:

United States:

(tel) 800 829 4444 (fax) 800 829 4433

Canada:

(tel) 877 894 4414 (fax) 800 746 4866

China:

(tel) 800 810 0189 (fax) 800 820 2816

Europe:

(tel) 31 20 547 2111

Japan:

(tel) (81) 426 56 7832 (fax) (81) 426 56 7840

Korea:

(tel) (080) 769 0800 (fax) (080) 769 0900

Latin America:

(tel) (305) 269 7500

Taiwan:

(tel) 0800 047 866 (fax) 0800 286 331

Other Asia Pacific Countries:

(tel) (65) 6375 8100 (fax) (65) 6755 0042

Or visit Agilent World Wide Web at: www.agilent.com/find/assist

Product specifications and descriptions in this document are subject to change without notice. Always refer to the Agilent website for the latest revision.

© Agilent Technologies, Inc., 2011–2012

Seventh Edition, November 28, 2012

U1731-90077

Agilent Technologies

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement