- Industrial & lab equipment
- Measuring, testing & control
- Multimeters
- Agilent Technologies
- U1731C
- User's Guide
advertisement
Agilent U1731C, U1732C, and U1733C Handheld
LCR Meter
User’s Guide
Agilent Technologies
II
Notices
© Agilent Technologies, Inc. 2011–2012
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent
Technologies, Inc. as governed by United
States and international copyright laws.
Manual Part Number
U1731-90077
Edition
Seventh Edition, November 28, 2012
Agilent Technologies, Inc.
5301, Stevens Creek Blvd.
Santa Clara, CA 95051 USA
Warranty
The material contained in this document is provided “as is,” and is subject to change, without notice, in future editions. Further, to the maximum extent permitted by the applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose.
Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Software and technical data rights granted to the federal government include only those rights customarily provided to end user customers. Agilent provides this customary commercial license in Software and technical data pursuant to FAR 12.211 (Technical
Data) and 12.212 (Computer Software) and, for the Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial
Items) and DFARS 227.7202-3 (Rights in
Commercial Computer Software or Computer Software Documentation).
Safety Notices
C A U T I O N
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the likes of that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a
CAUTION notice until the indicated conditions are fully understood and met.
WA R N I N G
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the likes of that, if not correctly performed or adhered to, could result in personal injury or death.
Do not proceed beyond a WARN-
ING notice until the indicated conditions are fully understood and met.
U1731C/U1732C/U1733C User’s Guide
Safety Symbols
Direct current (DC)
Alternating current (AC)
Both direct and alternating current
Three-phase alternating current
Earth (ground) terminal
Protective conductor terminal
Frame or chassis terminal
Equipotentiality
The following symbols on the instrument and in the documentation indicate precautions which must be taken to maintain safe operation of the instrument.
Off (supply)
On (supply)
Caution, risk of electric shock
Caution, risk of danger (refer to this manual for specific Warning or Caution information)
Caution, hot surface
Out position of a bi-stable push control
In position of a bi-stable push control
Equipment protected throughout by double insulation or reinforced insulation
U1731C/U1732C/U1733C User’s Guide III
Safety Considerations
Read the information below before using this instrument.
The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards for design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements.
C A U T I O N
• Disconnect circuit power and discharge all high-voltage capacitors before testing.
• When measuring in-circuit components, first de-energize the circuits before connecting them to the test leads.
• This device is for indoor use at altitudes of up to 2000 m.
• Always use the specified battery type (listed in
Characteristics" on page 76). The power for the meter is supplied
with a single standard 9 V battery. Observe the correct polarity markings before you insert the battery to ensure proper insertion of the battery in the meter.
• Line operation is also possible using a 12 V AC to DC adapter. If a power adapter is selected, please be sure it meets the safety requirements of a relevant IEC standard.
IV U1731C/U1732C/U1733C User’s Guide
WA R N I N G
• Use this meter only as specified in this manual; otherwise, the protection provided by the meter may be impaired.
• Do not use the meter if it is damaged. Before you use the meter, inspect the case. Look for cracks or missing plastic. Pay particular attention to the insulation surrounding the connectors.
• Inspect the test leads for damaged insulation or exposed metal.
Check the test leads for continuity. Replace damaged test leads before you use the meter.
• Do not operate the meter around explosive gas, vapor, or wet environments.
• Never use the meter in wet conditions or when there is water on the surface. If the meter is wet, ensure that the meter is dried only by trained personnel.
• When servicing the meter, use only the specified replacement parts.
• When using the probes, keep your fingers behind the finger guards on the probes.
• Connect the common test lead before you connect the live test lead. When you disconnect the leads, disconnect the live test lead first.
• Remove the test leads from the meter before you open the battery cover.
• Do not operate the meter with the battery cover or portions of the cover removed or loosened.
• To avoid false readings, which may lead to possible electric shock or personal injury, replace the battery as soon as the low battery indicator appears and flashes.
U1731C/U1732C/U1733C User’s Guide V
Environmental Conditions
This instrument is designed for indoor use and in an area with low condensation. The table below shows the general environmental requirements for this instrument.
N O T E
Environmental conditions
Operating temperature
Operating humidity
Storage temperature
Storage humidity
Altitude
Pollution degree
Requirements
Full accuracy from –10 °C to 55 °C
Full accuracy up to 80% RH (relative humidity)
–20 °C to 70 °C
0% to 80% RH non-condensing
Up to 2000 meters
Pollution degree II
The U1731C/U1732C/U1733C Handheld LCR Meter complies with the following safety and EMC requirements:
• IEC61010-1:2001/EN61010-1:2001 (Second Edition)
• IEC 61326-1:2005/EN 61326-1:2006
• Canada: ICES/NMB-001:Issue 4, June 2006
• Australia/New Zealand: AS/NZS CISPR11:2004
VI U1731C/U1732C/U1733C User’s Guide
Regulatory Markings
The CE mark is a registered trademark of the European Community. This CE mark shows that the product complies with all the relevant European Legal
Directives.
ICES/NMB-001 indicates that this ISM device complies with the Canadian
ICES-001.
Cet appareil ISM est confomre a la norme NMB-001 du Canada.
This symbol indicates the time period during which no hazardous or toxic substance elements are expected to leak or deteriorate during normal use.
Forty years is the expected useful life of the product.
The C-tick mark is a registered trademark of the Spectrum
Management Agency of Australia. This signifies compliance with the Australia EMC Framework regulations under the terms of the
Radio Communication Act of 1992.
This instrument complies with the
WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.
U1731C/U1732C/U1733C User’s Guide VII
Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC
This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.
Product Category:
With reference to the equipment types in the WEEE directive Annex 1, this instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste.
To return this unwanted instrument, contact your nearest Agilent Service
Centre, or visit www.agilent.com/environment/product for more information.
VIII U1731C/U1732C/U1733C User’s Guide
Declaration of Conformity (DoC)
The Declaration of Conformity (DoC) for this instrument is available on the
Agilent website. You can search the DoC by its product model or description at the web address below.
http://regulations.corporate.agilent.com/DoC/search.htm
N O T E
If you are unable to search for the respective DoC, please contact your local Agilent representative.
U1731C/U1732C/U1733C User’s Guide IX
X
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
U1731C/U1732C/U1733C User’s Guide
Table of Contents
1 Introduction
2
3
3
6
7
8
9
10
11
13
2 Features and Functions
Auto Identification (Ai) function
26
29
31
35
U1731C/U1732C/U1733C User’s Guide XI
XII
Measuring dissipation factor/quality factor/phase angle
θ ) 37
Changing the test frequency 37
Selecting parallel/series circuit mode (P/S)
37
Setting the standard reference tolerance (Tol%)
38
39
39
Freezing the display (Hold) 40
Enabling the static recording mode (Rec)
40
Setting the high/low limit comparison (Limit) 42
Making relative measurements (Null)
45
46
3 Setup Options
56
Changing the initial power-on behavior 56
Changing the Ai function’s phase angle condition
63
Changing the power-on limit category and set
65
Changing the user high/low limit values 66
69
Changing the beep frequency 71
Changing the auto power-off and backlight time-outs 73
U1731C/U1732C/U1733C User’s Guide
4 Characteristics and Specifications
77
78
Impedance/Resistance/DCR specifications 78
80
Phase angle of impedance specifications
81
Dissipation/Quality factor specifications 82
Source impedance of impedance/resistance measurement 84
Source impedance of capacitance measurement
85
Source impedance of inductance measurement 86
87
U1731C/U1732C/U1733C User’s Guide XIII
XIV U1731C/U1732C/U1733C User’s Guide
List of Figures
Installing the batteries 4
Power-on button 5
Tilt-stand adjustment and IR cable connection 8
Agilent GUI Data Logger Software 9
Width dimensions 11
Height and depth dimensions 12
Front panel 13
Rear panel 14
Using the Ai function 26
Inductance measurement with Q factor 29
Measuring inductance 30
Capacitance measurement with D factor 31
Measuring capacitance 32
Resistance measurement 33
Measuring resistance 34
Impedance measurement with theta 35
Measuring impedance 36
Component above setting tolerance 38
ESR measurement with theta 39
DCR measurement 39
Using the Hold function 40
Using the Rec function 41
Using the Limit function 43
High and low limit values 44
nGo and Go indications 44
Using the Null function 45
Using the Cal function 47
Open calibration and short calibration prompts 48
Changing the power-on measurement type 57
Changing the power-on test frequency 58
Changing the power-on secondary parameter and measurement mode for inductance (L) measurements 59
Changing the power-on secondary parameter and measurement mode for capacitance (C)
U1731C/U1732C/U1733C User’s Guide XV
measurements 60
Changing the power-on secondary parameter and measurement mode for resistance (R) measurements 61
Changing the power-on open/short CAL 62
Changing the Ai function’s phase angle condition 64
Changing the power-on limit and category set 65
Changing the user high/low limit values 67
Changing the baud rate 68
Changing the parity check 69
Changing the data bits 70
Changing the beep frequency 71
Locking the push buttons 72
Changing the auto power-off and backlight time-outs 73
Resetting the Setup items 74
U1782B SMD tweezers 87
XVI U1731C/U1732C/U1733C User’s Guide
List of Tables
Battery level indicator 5
Power-on options 10
Front panel parts 13
Rear panel parts 14
Keypad functions 15
General annunciators 18
Measurement units display 21
Input terminal/socket connections 22
Auto identification phase angle rules 27
Auto identification series/parallel rules for resistance measurements 27
Auto identification series/parallel rules for capacitance measurements 28
Auto identification series/parallel rules for inductance measurements 28
Available test frequencies 37
Factory default high and low limit values 42
Setup menu key functions 52
Setup menu item descriptions 54
Auto identification phase angle rules 63
Default user high/low limit values 66
Impedance/Resistance/DCR specifications 78
Capacitance specifications 79
Inductance specifications 80
Phase angle of impedance specifications 81
Dissipation/Quality factor specifications 82
Test signal specifications 83
Source impedance of impedance/resistance measurement 84
Source impedance of capacitance measurement 85
Source impedance of inductance measurement 86
U1782B SMD tweezers electrical characteristics 88
U1731C/U1732C/U1733C User’s Guide XVII
XVIII U1731C/U1732C/U1733C User’s Guide
U1731C/U1732C/U1733C Handheld LCR Meter
User’s Guide
1
Introduction
This chapter teaches you how to set up your LCR meter for the first time. An introduction to all the features and capabilities of the LCR meter is also given.
Agilent Technologies
1
2
1 Introduction
About This Manual
About This Manual
The descriptions and instructions in this manual apply to the Agilent U1731C, U1732C, and U1733C handheld LCR meters (hereafter referred to as the LCR meter).
The model U1733C appears in all illustrations.
Documentation map
The following manuals and software are available for your
LCR meter. For the very latest version, please visit our website at: http://www.agilent.com/find/hhTechLib .
Check the manual revision on the first page of each manual.
• User’s Guide. This manual.
• Quick Start Guide. Printed copy for outdoor use, included with shipment.
• Agilent GUI Data Logger Software, Quick Start Guide,
and Help. Free download at the Agilent website.
Safety notes
Safety notes are used throughout this manual (see the
“Safety Notices” section for format examples). Familiarize
yourself with each of the notes and its meaning before operating your LCR meter.
More pertinent safety notes for using this product are located under the
section.
Do not proceed beyond a safety notice until the indicated conditions are fully understood and met.
U1731C/U1732C/U1733C User’s Guide
Introduction 1
Preparing Your LCR Meter
Preparing Your LCR Meter
Check the shipment
When you receive your LCR meter, check the shipment according to the following procedure.
1 Inspect the shipping container for damage. Signs of damage may include a dented or torn shipping container or cushioning material that indicates signs of unusual stress or compacting. Save the packaging material in case the LCR meter needs to be returned.
2 Carefully remove the contents from the shipping container, and verify that the standard accessories and your ordered options are included in the shipment according to the standard shipped items list found in the printed copy of the U1731C/U1732C/U1733C Quick Start
Guide.
3 For any question or problems, refer to the Agilent contact numbers on the back of this manual.
Install the battery
Your LCR meter is powered by a single 9 V alkaline battery
(included with the shipment). When you receive your LCR meter, the 9 V alkaline battery is not installed.
Use the following procedure to install the battery.
C A U T I O N
Before you proceed with the battery installation, remove all cable connections to the terminals and ensure that the LCR meter is turned
OFF. Use only the battery type specified in
“Product Characteristics” on page 76.
U1731C/U1732C/U1733C User’s Guide 3
1 Introduction
Preparing Your LCR Meter
1 Open the battery cover. Lift the tilt stand. Loosen the screw with a suitable Phillips screwdriver and remove the
battery cover as shown in Figure 1- 1 .
2 Insert the battery. Observe the proper battery polarity.
The terminal ends of the battery are indicated inside the battery compartment.
3 Close the battery cover. Place the battery cover back in its original position and tighten the screw.
4
WA R N I N G
Figure 1-1 Installing the batteries
The battery level indicator in the lower right- hand corner of the display indicates the relative condition of the battery.
describes the various battery levels the indicator represents.
To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the low battery indicator appears. Do not discharge the battery by shorting the battery or reverse the battery polarity.
U1731C/U1732C/U1733C User’s Guide
Introduction 1
Preparing Your LCR Meter
C A U T I O N
To avoid instruments being damage from battery leakage:
• Always remove dead batteries immediately.
• Always remove the battery and store it separately if the LCR meter is not going to be used for a long period.
Table 1-1 Battery level indicator
Indication Battery capacity
Full capacity
2/3 capacity
1/3 capacity
Almost empty (less than one day)
[1]
(Flashing periodically)
[1] Battery change advised. Always use the specified battery type listed in
Turn on your LCR meter
To power ON your LCR meter, press the power- on button once. The LCR meter powers up in the auto identification
(Ai) mode (see page 26 ) when turned on for the first time.
Power-on button
Figure 1-2 Power-on button
U1731C/U1732C/U1733C User’s Guide 5
6
1 Introduction
Preparing Your LCR Meter
N O T E
To power OFF your LCR meter, press the power- on button again.
You can change the power-on behavior of your LCR meter for subsequent power cycles. See
“Changing the initial power-on behavior” on page 56 for
more information on changing the LCR meter’s power-on setting.
Automatic Power-Off (APO)
Your LCR meter automatically turns off after 5 minutes
(default) if no keys are pressed. Pressing any key will turn the LCR meter back on after it is powered off automatically.
The annunciator is shown on the bottom left of the display when the APO function is enabled.
N O T E
• To change the time-out period or completely disable the APO function,
refer to “Changing the auto power-off and backlight time-outs” on page 73
.
• If an external power adapter is used, the APO function will be disabled.
Enabling the backlight
If viewing the display becomes difficult in low- light conditions, press Tol% for more than 1 second to activate the
LCD backlight.
To conserve battery life, a user- adjustable time- out controls how long the backlight stays on. The default time- out is 30 seconds.
N O T E
• To change the time-out period or completely disable the backlight, refer to
“Changing the auto power-off and backlight time-outs” on page 73
.
• If an external power adapter is used, the backlight time-out will be disabled.
U1731C/U1732C/U1733C User’s Guide
Introduction 1
Preparing Your LCR Meter
Selecting the range
Pressing Range
Auto
switches the LCR meter between manual and autoranging. It also cycles through the available LCR meter ranges when manual ranging is enabled.
Autoranging is convenient because the LCR meter automatically selects an appropriate range for sensing and displaying each measurement. However, manual ranging results in better performance, since the LCR meter does not have to determine which range to use for each measurement.
In autorange, the LCR meter selects the lowest range to display the highest available precision (resolution) for the input signal. If manual range is already enabled, press Range
Auto for more than 1 second to enter the autoranging mode.
If autoranging is enabled, press Range
Auto
to enter the manual range mode.
Each additional press of Range
Auto
sets the LCR meter to the next higher range, unless it is already in the highest range, at which point the range switches to the lowest range.
U1731C/U1732C/U1733C User’s Guide 7
8
1 Introduction
Preparing Your LCR Meter
Adjusting the tilt stand
To adjust the LCR meter to a 60° standing position, pull the tilt- stand outward to its maximum reach.
To PC (host)
IR-USB cable
Figure 1-3 Tilt-stand adjustment and IR cable connection
U1731C/U1732C/U1733C User’s Guide
Introduction 1
Preparing Your LCR Meter
Connecting the IR-USB cable
You can use the IR communication link (IR communication port, located at the rear panel) and the Agilent GUI Data
Logger software to control your LCR meter remotely. Hence, you can only perform data logging operations in the Agilent
GUI Data Logger application with the LCR meter connected via the USB- IR cable to the PC.
Ensure that the Agilent logo on the U5481A IR- USB cable
(purchased separately) connected to the LCR meter is facing up. Firmly push the IR head into the LCR meter’s IR communication port until it snaps into place (see
).
Refer to the Agilent GUI Data Logger Software Quick Start
Guide and Help for more information on the IR communication link and the Agilent GUI Data Logger software.
Figure 1-4 Agilent GUI Data Logger Software
The Agilent GUI Data Logger software and its supporting documents (Quick Start Guide and Help) are available for free download at http://www.agilent.com/find/hhTechLib .
You may purchase a U5481A IR- USB cable from an Agilent
Sales Office nearest to you.
U1731C/U1732C/U1733C User’s Guide 9
1 Introduction
Preparing Your LCR Meter
Power-on options
Some options can be selected only while you turn the LCR meter on. These power- on options are listed in the table below.
To select a power- on option, press and hold the specified
while turning the LCR meter ON ( ).
Table 1-2 Power-on options
Key
Hold
Rec
Range
Auto
Null
Cal
Description
Tests the LCD.
All annunciators are displayed in the LCD. Press any key to exit this mode.
Simulates the Auto Power-Off (APO) mode. Press any key to turn the LCR meter back on and resume normal operation.
Checks the firmware version.
The LCR meter’s firmware version will be shown on the primary display. Press any key to exit this mode.
Performs the Open/Short CAL on all frequencies and all ranges for the User mode ( ).
[1]
ZLCR
P S
Enters the Setup menu.
See
Chapter 3 , “Setup Options,” starting on page 51 for
more information. Press and hold ZLCR
P S
for more than 1 second to exit this mode.
[1] The Open/Short CAL requires approximately 1.5 minutes to complete.
10 U1731C/U1732C/U1733C User’s Guide
Your LCR Meter in Brief
Dimensions
Front view
87 mm
Introduction 1
Your LCR Meter in Brief
Figure 1-5 Width dimensions
U1731C/U1732C/U1733C User’s Guide 11
1 Introduction
Your LCR Meter in Brief
Rear and side view
41 mm
184 mm
12
Figure 1-6 Height and depth dimensions
U1731C/U1732C/U1733C User’s Guide
Introduction 1
Your LCR Meter in Brief
Overview
Front panel
The front panel parts of your LCR meter are described in this section. Click the respective “Learn more” pages in
for more information on each part.
1
2
Figure 1-7 Front panel
Table 1-3 Front panel parts
Legend
1
2
3
Description
Display screen
Keypad
Input terminals and sockets
U1731C/U1732C/U1733C User’s Guide
3
Learn more on:
13
1 Introduction
Your LCR Meter in Brief
Rear panel
The rear panel parts of your LCR meter are described in this section. Click the respective “Learn more” pages in
for more information on each part.
1
2
3
4
14
Figure 1-8 Rear panel
Table 1-4 Rear panel parts
Legend
1
2
3
4
Description
IR communication port
Tilt stand
Battery cover (lift the tilt stand for access)
External power adapter input jack
[1]
Learn more on:
-
[1] The external power adapter input jack requires an input voltage of +12 VDC.
U1731C/U1732C/U1733C User’s Guide
Introduction 1
Your LCR Meter in Brief
Keypad
The operation of each key is explained below. Pressing a key enables a function, displays a related annunciator, and emits a beep.
Each key operation of the U1731C/U1732C/U1733C keypad
(shown in
) is described in Table 1- 5 . Click the
respective “Learn more” pages in
information on each function.
Table 1-5 Keypad functions
Legend
Less than 1 second
Function when pressed for:
More than 1 second
Learn more on:
ESR
Hold
Rec
Save
Turns the LCR meter on or off.
-
Starts or stops the auto identification mode.
• Press again while the annunciator is shown to exit this mode.
Holds or releases the present reading on the display.
• Press Hold
Rec
again to update the reading automatically once it is stable.
• Press Hold
Rec
for more than 1 second to exit this mode.
Enables or disables the ESR (equivalent series resistance) mode.
• Press for more than 1 second to exit this mode. The LCR meter will return to capacitance measurement by default.
Starts or stops the static recording mode.
• Press Hold
Rec
again to cycle through the maximum (Max), minimum (Min), average
(Avg), and present (MaxMinAvg) readings.
• Press Hold
Rec
for more than 1 second to exit this mode.
U1731C/U1732C/U1733C User’s Guide 15
1 Introduction
Your LCR Meter in Brief
Table 1-5 Keypad functions (continued)
Legend
DQO
Limit
Freq.
DCR
Range
Auto
Less than 1 second
Function when pressed for:
More than 1 second
Switches between the dissipation factor (D), quality factor (Q), or phase angle ( measurement.
Selects a test frequency.
θ)
• Press Freq.
again to cycle through the various test frequencies (100 Hz to
100 kHz).
Enables or disables the limit comparison mode.
• While the annunciator is flashing,
DQO
Limit
Range
• press and
Auto
again to toggle between high (H) or low (L) limits, then
Freq.
• use the and keys to select a high/low limit set (1 to 16).
• Press Hold
Rec
to start the limit sorting (with the selected limit set), or
• If no activity is detected after 3 seconds, the limit comparison will begin.
• Press
DQO
Limit
for more than 1 second to exit this mode.
U1733C only: Enables or disables the DCR
(direct current resistance) mode.
Freq.
• Press for more than 1 second to exit this mode. The LCR meter will return to inductance measurement by default.
Disables autoranging and sets a manual range.
• Press Range
Auto
again to cycle through each available measurement range.
Enables autoranging.
Learn more on:
and
ZLCR
P S
and
Tol%
Switches between impedance (Z), inductance (L), capacitance (C), and resistance (R) measurement.
Toggles between parallel and series circuit mode.
Sets the tolerance mode.
• Connect/insert an appropriate component into the input terminals/sockets and press
Tol% to set the value shown on the secondary display as the standard reference value.
Tol% • Press again to cycle through the various tolerance values (1% to 20%).
Turns the LCD backlight on for 15 seconds
(default) or off.
• To change the backlight time-out refer to
“Changing the auto power-off and backlight time-outs” on page 73.
and
16 U1731C/U1732C/U1733C User’s Guide
Introduction 1
Your LCR Meter in Brief
Table 1-5 Keypad functions (continued)
Legend
Null
Cal
Less than 1 second
Function when pressed for:
More than 1 second
Sets the null/relative mode.
• The displayed value is saved as a reference to be subtracted from subsequent measurements.
• Press Null
Cal
again to cancel the null mode.
Enters the open/short calibration mode for the selected range and test frequency.
• Follow the prompts on the screen (open or short connector) and press Null
Cal
to begin the calibration process.
• The LCR meter will return to normal display when the calibration is complete.
Learn more on:
and
U1731C/U1732C/U1733C User’s Guide 17
1 Introduction
Your LCR Meter in Brief
Display screen
The function that each display annunciator of your LCR meter is associated to is described in this section. See also
“Measurement units” on page 21 for a list of available
measurement signs and notations.
General display annunciators
The general display annunciators of your LCR meter are described in the table below.
Each display annunciator of the U1731C/U1732C/U1733C
display screen (shown in Figure 1- 7
) is described in
. Click the respective “Learn more” pages in
for more information on each annunciator.
Table 1-6 General annunciators
Legend Description
Remote control via PC indicator
Equivalent series resistance indicator
Resistance measurement by direct current indicator
LCR meter using open/short CAL settings by factory
LCR meter using open/short CAL setting by user
Measuring frequency of test signal is 100 Hz
Measuring frequency of test signal is 120 Hz
Measuring frequency of test signal is 1 kHz
Measuring frequency of test signal is 10 kHz
Measuring frequency of test signal is 100 kHz
Learn more on:
18 U1731C/U1732C/U1733C User’s Guide
Introduction 1
Your LCR Meter in Brief
Table 1-6 General annunciators (continued)
Legend Description
Tolerance mode indicator for sorting L, C, or R
Learn more on:
Tolerance set to 1% for sorting capacitance
Tolerance set to 5% for sorting capacitance
Tolerance set to 10% for sorting capacitance
Tolerance set to 20% for sorting capacitance
Data hold mode indicator
Audible alert indicator for tolerance or limit mode
Dissipation factor indicator
Quality factor indicator
Phase angle of impedance indicator
Secondary display -
Measurement units for the secondary display
Impedance measurement indicator
Inductance measurement indicator
Capacitance measurement indicator
Resistance measurement indicator
U1731C/U1732C/U1733C User’s Guide 19
1 Introduction
Your LCR Meter in Brief
Table 1-6 General annunciators (continued)
Legend Description
Present reading shown on primary display
Maximum reading shown on primary display
Minimum reading shown on primary display
Learn more on:
Averaged reading shown on primary display
Relative (Null) indicator
Autoranging indicator
Limit mode indicator
Reading out of HI limit
Reading out of LO limit
Auto power-off indicator
Primary display -
Measurement units for the primary display
Parallel mode indicator
Series mode indicator
Battery capacity indicator
20 U1731C/U1732C/U1733C User’s Guide
Introduction 1
Your LCR Meter in Brief
Measurement units
The available signs and notations for each measurement function in your LCR meter are described in
. The units listed below are applicable to the primary display measurements of your LCR meter.
Table 1-7 Measurement units display k m
μ n
Sign/Notation
M p
°
%
μH, mH, H pF, nF, μF, mF
Ω, kΩ, MΩ kHz, Hz
Description
Mega 1E+06 (1000000) kilo milli
1E+03 (1000)
1E–03 (0.001) micro 1E–06 (0.000001) nano 1E–09 (0.000000001) pico 1E–12 (0.000000000001)
Degree, unit for phase angle measurement
Percentage, unit for tolerance measurement
Henry, units for inductance measurement
Farad, units for capacitance measurement
Ohm, units for resistance and impedance measurement
Hertz, units for frequency measurement
U1731C/U1732C/U1733C User’s Guide 21
1 Introduction
Your LCR Meter in Brief
Input terminals
The terminal and socket connections of your LCR meter are described in the table below.
WA R N I N G
To avoid damaging this instrument, do not exceed the input limit. Do not apply voltage to input terminals. Discharge the capacitor before testing.
Table 1-8 Input terminal/socket connections
Input terminal/socket
Description
+
Positive terminal/component socket
-
Negative terminal/component socket
GUARD
Guard terminal/component socket
22 U1731C/U1732C/U1733C User’s Guide
Introduction 1
Cleaning Your LCR Meter
Cleaning Your LCR Meter
WA R N I N G
To avoid electrical shock or damage to the LCR meter, ensure that the insides of the casing stay dry at all times.
Dirt or moisture in the terminals can distort readings.
Follow the steps below to clean your LCR meter.
1 Turn the LCR meter off and remove the test leads.
2 Turn the LCR meter over and shake out any dirt that may have accumulated in the terminals.
3 Wipe the case with a damp cloth and mild detergent — do not use abrasives or solvents.
4 Wipe the contacts in each terminal with a clean swab dipped in alcohol.
U1731C/U1732C/U1733C User’s Guide 23
1 Introduction
Cleaning Your LCR Meter
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
24 U1731C/U1732C/U1733C User’s Guide
U1731C/U1732C/U1733C Handheld LCR Meter
User’s Guide
2
Features and Functions
Auto Identification (Ai) function 26
Measuring dissipation factor/quality factor/phase angle (D/Q/
Changing the test frequency 37
Selecting parallel/series circuit mode (P/S) 37
Setting the standard reference tolerance (Tol%) 38
Freezing the display (Hold) 40
Enabling the static recording mode (Rec) 40
Setting the high/low limit comparison (Limit) 42
Making relative measurements (Null) 45
Performing the open/short CAL 46
This chapter provides detailed information on the features and functions that are available in your LCR meter.
Agilent Technologies
25
2 Features and Functions
Making Measurements
Making Measurements
Auto Identification ( Ai) function
Press to automatically identify the appropriate measurement required for the device- under- test (DUT).
26
N O T E
Figure 2-1 Using the Ai function
The annunciator will flash while the LCR meter identifies the DUT, and
• selects an appropriate measurement in the primary display (L, C, or R) and secondary display (D, Q, or
θ),
• selects an appropriate range, and
• selects an appropriate measuring mode (series or parallel).
The Ai function helps to identify L, C, and R measurements automatically according to the angle of impedance detected in the DUT. See
the phase angle rules.
The default phase angle condition is set to 10°. You can change this angle in the Setup menu from 5° to 45°. See
“Changing the Ai function’s phase angle condition” on page 63 for more information.
U1731C/U1732C/U1733C User’s Guide
Table 2-2 Auto identification series/parallel rules for resistance measurements
Resistance range
200 M Ω
20 M Ω
2000 k Ω
200 k Ω
20 k Ω
2000 Ω
200 Ω
20 Ω
2 Ω
Down range
Parallel
Parallel
Parallel
Parallel
Parallel
Parallel
Parallel
Series
Series
Features and Functions 2
Making Measurements
The measurement mode (series or parallel) will be automatically identified from the autoranging direction.
,
Table 2- 3 , and Table 2- 4 list down the
series/parallel rules used.
Table 2-1 Auto identification phase angle rules
Phase angle
[1]
–Set <
θ < +Set
θ ≥ +Set
θ ≤ –Set
[1] Where ±Set is the phase angle selected.
Primary display
R
L
C
Secondary display
θ
Q
D
Up range
Parallel
Parallel
Parallel
Parallel
Series
Series
Series
Series
Series
U1731C/U1732C/U1733C User’s Guide 27
2 Features and Functions
Making Measurements
20 mF
2000
μF
200
μF
20
μF
2000 nF
200 nF
20 nF
2000 pF
200 pF
20 pF
Table 2-3 Auto identification series/parallel rules for capacitance measurements
Range
100 Hz
Down Up
120 Hz
Down Up Down
1 kHz
Up
Series Series Series Series -
Series Series Series Series Series Series
10 kHz
Down Up
-
-
-
-
100 kHz
Down Up
-
-
-
-
Series Series Series Series Series Series Series Series -
Series Parallel Series Parallel Series Series Series Series Series Series
Series Parallel Series Parallel Series Parallel Series Series Series Series
Series Parallel Series Parallel Series Parallel Series Parallel Series Series
Parallel Parallel Parallel Parallel Series Parallel Series Parallel Series Parallel
Parallel Parallel Parallel Parallel Parallel Parallel Series Parallel Series Parallel
-
-
-
-
-
-
-
Parallel
-
Parallel
-
Parallel
Parallel
Parallel
Parallel
Series
Parallel
Parallel
Parallel
2000 H
200 H
20 H
2000 mH
200 mH
20 mH
2000
μH
200
μH
20
μH
Table 2-4 Auto identification series/parallel rules for inductance measurements
Range
100 Hz
Down Up
120 Hz
Down Up Down
1 kHz
Up
10 kHz
Down Up
Parallel Parallel Parallel Parallel Parallel Parallel -
Parallel Parallel Parallel Parallel Parallel Parallel Parallel Parallel
100 kHz
Down Up
-
-
-
-
Parallel Series Parallel Series Parallel Parallel Parallel Parallel Parallel Parallel
Parallel Series Parallel Series Parallel Series Parallel Parallel Parallel Parallel
Parallel Series Parallel Series Parallel Series Parallel Series Parallel Parallel
Series Series Series Series Parallel Series Parallel Series Parallel Series
Series Series Series Series Series Series Parallel Series Parallel Series
Series Series Series Series Parallel Series
Series Series Series Series
28 U1731C/U1732C/U1733C User’s Guide
Features and Functions 2
Making Measurements
Measuring inductance (L)
Set up your LCR meter to measure inductance as shown in
.
N O T E
It is recommended that you perform the Open/Short calibration (see
page 46 ) before testing to achieve optimum precision for all inductance,
capacitance, and resistance measurements at either the highest or lowest ranges.
1 Press to power on the LCR meter.
2 Press Freq.
to select a suitable test frequency, and i press to enable the auto identification function; or
ii alternatively press ZLCR
P S
to select inductance measurement.
Figure 2-2 Inductance measurement with Q factor
3 Insert an inductor into the component socket or connect the test clip to the component leads as required.
4 Press DQO
Limit
to change the secondary display measurement
(D, Q, or
θ).
5 Read the displays.
U1731C/U1732C/U1733C User’s Guide 29
2 Features and Functions
Making Measurements
3
1
Freq.
>
1
OR
Freq.
> ZLCR
P S o
2
Inductor
OR
Surface Mount
Device (SMD)
U1782B
SMD Tweezers
Figure 2-3 Measuring inductance
30 U1731C/U1732C/U1733C User’s Guide
Features and Functions 2
Making Measurements
Measuring capacitance (C)
Set up your LCR meter to measure capacitance as shown in
.
WA R N I N G
To avoid electrical hazards, discharge the capacitor to be tested before measuring.
1 Press to power on the LCR meter.
2 Press Freq.
to select a suitable test frequency, and i press to enable the auto identification function; or
ii alternatively press ZLCR
P S
to select capacitance measurement.
Figure 2-4 Capacitance measurement with D factor
3 Insert a capacitor into the component socket or connect the test clip to the component leads as required.
4 Press DQO
Limit
to change the secondary display measurement
(D, Q, or
θ).
5 Read the displays.
U1731C/U1732C/U1733C User’s Guide 31
2 Features and Functions
Making Measurements
3
1
Freq.
>
1
OR
Freq.
> ZLCR
P S
2
Capacitor
OR
Surface Mount
Device (SMD)
U1782B
SMD Tweezers
Figure 2-5 Measuring capacitance
32 U1731C/U1732C/U1733C User’s Guide
Features and Functions 2
Making Measurements
Measuring resistance (R)
Set up your LCR meter to measure resistance as shown in
.
C A U T I O N
To avoid possible damage to your LCR meter or to the equipment under test, disconnect the circuit power and discharge all high-voltage capacitors before measuring resistance.
1 Press to power on the LCR meter.
2 Press Freq.
to select a suitable test frequency, and i press to enable the auto identification function; or
ii alternatively press ZLCR
P S
to select resistance measurement.
Figure 2-6 Resistance measurement
3 Insert a resistor into the component socket or connect the test clip to the component leads as required.
4 Read the display.
U1731C/U1732C/U1733C User’s Guide 33
2 Features and Functions
Making Measurements
3
1
Freq.
>
1
OR
Freq.
> ZLCR
P S
2
Resistor
OR
Surface Mount
Device (SMD)
U1782B
SMD Tweezers
Figure 2-7 Measuring resistance
34 U1731C/U1732C/U1733C User’s Guide
N O T E
Features and Functions 2
Making Measurements
Measuring impedance (Z)
All circuit components, resistors, capacitors, and inductors have parasitic components. These include, for example, unwanted resistance in capacitors, unwanted capacitance in inductors, and unwanted inductance in resistors. Thus, simple components should be modeled as complex impedances.
Set up your LCR meter to measure impedance as shown in
.
To learn more about impedance measurement theories, refer to the
Impedance Measurement Handbook. This document can be downloaded from our website at http://www.agilent.com/find/lcrmeters .
1 Press to power on the LCR meter.
2 Press Freq.
to select a suitable test frequency, and press
ZLCR
P S
to select impedance measurement.
Figure 2-8 Impedance measurement with theta
3 Insert a component into the component socket or connect the test clip to the component leads as required.
U1731C/U1732C/U1733C User’s Guide 35
2 Features and Functions
Making Measurements
4 Press DQO
Limit
to change the secondary display measurement
(D, Q, or
θ).
5 Read the displays.
3
Surface Mount
Device (SMD)
2 Device-Under-Test
(DUT)
1
Freq.
> ZLCR
P S
OR
Figure 2-9 Measuring impedance
36
U1782B
SMD Tweezers
U1731C/U1732C/U1733C User’s Guide
Features and Functions 2
Making Measurements
Measuring dissipation factor/quality factor/phase angle
(D/Q/
θ)
The dissipation factor (D), quality factor (Q), and phase angle (
θ) values can be displayed interchangeably by pressing the DQO
Limit
key when the LCR meter is set to the inductance, capacitance, or impedance measurement mode.
This setting is not applicable for DCR measurement.
Changing the test frequency
The test frequency is set to 1 kHz by default. Press the Freq.
key to select a desired test frequency.
Table 2-5 Available test frequencies
Model
U1731C
U1732C
U1733C
100 Hz
✔
✔
✔
120 Hz
✔
✔
✔
1 kHz
✔
✔
✔
10 kHz
-
✔
✔
100 kHz
-
-
✔
Selecting parallel/series circuit mode (P/S)
The LCR meter can display parallel (
( ) mode data for all ranges.
) or series
Press the ZLCR
P S
key for more than 1 second to toggle the parallel and series mode.
Series mode is set as the default setting. You can, however, change this power- on behavior in the Setup menu. See
“Changing the initial power- on behavior” on page 56 for
more information on how to change the default measurement mode (parallel or series) for subsequent power cycles.
U1731C/U1732C/U1733C User’s Guide 37
2 Features and Functions
Making Measurements
Setting the standard reference tolerance (Tol%)
The tolerance ranges available are 1%, 5%, 10%, and 20%.
To enable the tolerance mode, insert an appropriate component as a standard value into the component socket or connect the test clip to the component leads, then press the
Tol% key to set this value as the standard reference tolerance.
Similarly, any value which appears on the display, such as
Hold or Max/Min/Avg (Rec), can be used as a standard value to sort components. Press Tol% again to cycle through 1%, 5%,
10%, and 20% tolerance as desired.
This function is designed for convenient component sorting.
The beeper will beep three times whenever the component under test exceeds the setting tolerance. Conversely, when the beeper beeps once, this indicates that the component is within the setting tolerance.
38
N O T E
Figure 2-10 Component above setting tolerance
• The tolerance mode cannot be activated if is shown on the display or when the tested capacitance value is below 50 counts.
• Tolerance mode is only available in manual ranging; therefore, activation while in autoranging will automatically set the LCR meter to manual ranging.
U1731C/U1732C/U1733C User’s Guide
Features and Functions 2
Making Measurements
Enabling ESR measurements
Press for more than 1 second to select the ESR measurement. Use the ESR measurement to measure the equivalent series resistance of the capacitor, independent of its capacitance.
Figure 2-11 ESR measurement with theta
Press for more than 1 second to exit this mode.
Enabling DCR measurements
Press Freq.
for more than 1 second to select the DCR measurement. The DCR measurement measures the resistance of an unknown component by 1 VDC.
Figure 2-12 DCR measurement
Press Freq.
for more than 1 second to exit this mode.
U1731C/U1732C/U1733C User’s Guide 39
2 Features and Functions
Additional Features
Additional Features
Freezing the display (Hold)
To freeze the display for any function, press the Hold
Rec
key.
The annunciator is shown on the display while the
Hold function is active.
40
Hold
Rec
Figure 2-13 Using the Hold function
Press Hold
Rec
again to update the reading automatically once it is stable. The annunciator flashes while waiting for the reading to be stable.
Press Hold
Rec
for more than 1 second to release the Hold function.
Enabling the static recording mode (Rec)
The static recording mode stores the maximum, minimum, and average input values during a series of measurements in the LCR meter’s memory.
U1731C/U1732C/U1733C User’s Guide
Features and Functions 2
Additional Features
When the input goes below the recorded minimum value or above the recorded maximum value, the LCR meter beeps and records the new value. The LCR meter also calculates an average of all readings taken since the static recording mode was activated.
From the LCR meter’s display, you can view the following statistical data for any set of readings:
• Max: highest reading since the static recording mode was enabled
• Min: lowest reading since the static recording mode was enabled
• Avg: average or mean of all readings since the static recording mode was enabled
• MaxMinAvg: present reading (actual input signal value)
Press the Hold
Rec
key for more than 1 second to enter the static recording mode.
> 1 s
Hold
Rec
Figure 2-14 Using the Rec function
Press Hold
Rec
again to cycle through the Max, Min, Avg, or
MaxMinAvg (present) input values.
To exit this mode, press and hold the
1 second.
Hold
Rec
key for more than
U1731C/U1732C/U1733C User’s Guide 41
2 Features and Functions
Additional Features
N O T E
• Static recording captures only stable values and updates the memory; it will not record any overload ( ) value for any of the LCR functions. In addition, the LCR meter will not record values below 50 counts in capacitance measurement.
• Static recording is only available in manual ranging; therefore, activation while in autoranging will automatically set the LCR meter to manual ranging.
42
Setting the high/low limit comparison (Limit)
The high and low limit comparison function helps you to sort components easily. There are 32 limit sets available (16 fixed factory sets, and 16 variable user sets).
The LCR meter will use the factory sets by default. You can set the LCR meter to use the user sets upon start- up from the Setup menu. See
“Changing the power- on limit category and set” on page 65 for more information.
shows the factory default limit values for each set.
Table 2-6 Factory default high and low limit values
F04
F05
F06
F07
Set
F01
F02
F03
F08
F09
High limit (H)
1000
1200
1500
1800
2200
2700
3300
3900
4700
Low limit (L)
900
1080
1350
1620
1980
2430
2970
3510
4230
U1731C/U1732C/U1733C User’s Guide
N O T E
Features and Functions 2
Additional Features
Table 2-6 Factory default high and low limit values (continued)
F12
F13
F14
F15
F16
Set
F10
F11
High limit (H)
5600
6800
8200
10000
12000
15000
18000
Low limit (L)
5040
6120
7380
9000
10800
13500
16200
The default values of the variable user sets are set to the same as the fixed user sets. Use the Setup menu to change the high and low limits for each
set. See “Changing the user high/low limit values” on page 66 for more
information.
Press the DQO
Limit
key for more than 1 second to activate the high/low limit mode. The last- known set number (H## or
L##) will be indicated in the secondary display.
> 1 s
DQO
Limit
Figure 2-15 Using the Limit function
U1731C/U1732C/U1733C User’s Guide 43
2 Features and Functions
Additional Features
While the annunciator is flashing, use the key to select an appropriate limit set.
or Freq.
You may press DQO
Limit
or Range
Auto
again to toggle between the high
(H) or low (L) values shown on the primary display.
44
Figure 2-16 High and low limit values
Press Hold
Rec
while the annunciator is flashing to start the comparison. (If no activity is detected after 3 seconds, the comparison will also begin.)
The LCR meter beeps three times and displays in the secondary display if the reading is greater ( ) than the high limit or lesser ( ) than the low limit.
If the reading is within the high and low limits, the meter beeps once and displays in the secondary display.
Figure 2-17 nGo and Go indications
The limit set used in the comparison is displayed after the
/ indication.
U1731C/U1732C/U1733C User’s Guide
Features and Functions 2
Additional Features
Press and hold DQO
Limit
for more than 1 second to exit this mode.
Making relative measurements (Null)
When making relative measurements, also called null, each reading is the difference between a stored (selected or measured) relative value and the input signal.
One possible application is to increase the accuracy of a resistance measurement by nulling the test lead resistance
(test leads shorted). Nulling the leads is also particularly important prior to making capacitance measurements (test leads open).
Press the Null
Cal
key to enter the relative mode and store the display reading as a reference value. The LCR meter will then display all subsequent readings relative to the reference value.
Null
Cal
Figure 2-18 Using the Null function
U1731C/U1732C/U1733C User’s Guide 45
2 Features and Functions
Additional Features
N O T E
The annunciator is shown on the display while the relative mode is active. Press Null
Cal
again to exit the relative mode.
• The relative mode cannot be activated if the display value is .
• Relative mode is only available in manual ranging; therefore, activation while in autoranging will automatically set the LCR meter to manual ranging.
• The relative mode cannot be activated if the LCR meter is set at auto-ranging with data hold activated.
Performing the open/short CAL
The CAL function offsets (corrects) the LCR meter's internal parameters as well as external connector residues for further measuring. Performing this action will help you to correct the influence for temporary uses.
There are three types of open/short CAL available:
• OS- Factory: Open/Short CAL is performed during factory calibration mode (security code protected). It covers all frequencies and all ranges.
• OS- User: Open/Short CAL is performed at every power- on option interval. It covers all frequencies and all ranges.
(see
for the OS- User setup).
• Open/Short CAL for single range and frequency by pressing and holding the Null
Cal
key for more than 1 second.
The corrections for the and are pre- stored in the LCR meter. They are both calibrated at the terminal ends.
You can set the LCR meter to start up using the or open/short CAL from the Setup menu (see
46 U1731C/U1732C/U1733C User’s Guide
N O T E
Features and Functions 2
Additional Features
The open/short CAL function is available for fixed measurement ranges.
It is highly recommended that open/short CAL be performed before making precision measurements.
> 1 s
Null
Cal
Figure 2-19 Using the Cal function
U1731C/U1732C/U1733C User’s Guide 47
2 Features and Functions
Additional Features
Figure 2-20 Open calibration and short calibration prompts
1 Press and hold the Null
Cal
key for more than 1 second to enter the open/short CAL mode for the selected frequency and range.
2 Open/Short CAL prompts will be shown on the display.
Follow the prompts for open connector ( OPn ) or short connector ( SHor) connection and press the Null
Cal
key.
The CAL annunciator on the upper right of the display will flash indicating that the correction is in process.
3 After the open/short CAL is completed, the LCR meter will be restored to the normal display and ready for normal usage.
48 U1731C/U1732C/U1733C User’s Guide
Features and Functions 2
Additional Features
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
U1731C/U1732C/U1733C User’s Guide 49
2 Features and Functions
Additional Features
50 U1731C/U1732C/U1733C User’s Guide
U1731C/U1732C/U1733C Handheld LCR Meter
User’s Guide
3
Setup Options
Changing the initial power-on behavior 56
Changing the Ai function’s phase angle condition 63
Changing the power-on limit category and set 65
Changing the user high/low limit values 66
Changing the beep frequency 71
Changing the auto power-off and backlight time-outs 73
The following chapter describes how to change the preset features of your LCR meter.
Agilent Technologies
51
3 Setup Options
Using the Setup Menu
Using the Setup Menu
The Setup menu allows you to change a number of nonvolatile preset features. Modifying these settings affects the general operation of your LCR meter across several functions. Select a setting to edit to perform one of the following:
• Switch between two values, such as on or off.
• Cycle through multiple values from a predefined list.
• Decrease or increase a numerical value within a fixed range.
The contents of the Setup menu are summarized in
Table 3-1 Setup menu key functions
Legend
ZLCR
P S
Description
Press and hold ZLCR
P S
while turning the LCR meter ON
( ) to access the Setup menu.
Press and hold ZLCR
P S
for more than 1 second to exit this mode.
DQO
Limit
Range
Auto
Press DQO
Limit
or Range
Auto
to step through the menu items.
Hold
Rec
Save
Freq.
ZLCR
P S
Press or
Freq.
at each menu item to change the preset settings. The menu item (in the secondary display) will flash to indicate that you can now change the menu item values.
Press or
Freq.
again to switch between two values, to cycle through multiple values from a list, or to decrease or increase a numerical value.
While the menu item is flashing, press
Hold
Rec
to save your changes.
While the menu item is flashing, press discard your changes.
ZLCR
P S
to
52 U1731C/U1732C/U1733C User’s Guide
Setup Options 3
Using the Setup Menu
Editing numerical values
When editing numerical values, use the DQO
Limit
and Range
Auto
to position the cursor on a numerical digit.
• Press DQO
Limit
to move the cursor to the left, and
• Press Range
Auto
to move the cursor to the right.
When the cursor is positioned over a digit, use the and
Freq.
keys to change the numerical digit.
• Press to increment the digit, and
• Press Freq.
to decrement the digit.
When you have completed your changes, save the new numerical value by pressing Hold
Rec
. (Or alternatively, if you wish to discard the changes you made press, ZLCR
P S
.)
U1731C/U1732C/U1733C User’s Guide 53
3 Setup Options
Setup Menu Summary
Setup Menu Summary
The Setup menu items are summarized in the table below.
Click the respective “Learn more” pages for more information on each menu item.
Table 3-2 Setup menu item descriptions
Legend Available settings Description
Ai, Z, L, C, R, ESR, or DCR
Set the measurement type that the LCR meter powers up in. Default is the auto identification (Ai) mode.
Learn more on:
100 Hz, 120 Hz, 1 kHz,
10 kHz, or 100 kHz
D, Q, or
D, Q, or
D, Q, or
θ and P or S
θ and P or S
θ and P or S
FACt or USEr
Set the test frequency that the LCR meter powers up in.
Default is 1 kHz.
Set the inductance (L) secondary parameter and measurement mode that the LCR meter powers up in.
Default is quality factor (Q) and series (S).
Set the capacitance (C) secondary parameter and measurement mode that the LCR meter powers up in.
Default is dissipation factor (D) and series (S).
Set the resistance (R) secondary parameter and measurement mode that the LCR meter powers up in.
Default is phase angle ( θ) and series (S).
Set the open/short CAL mode that the LCR meter powers up in. Default is factory (FACt).
05° to 45 °
Set the phase angle condition for the auto identification
(Ai) mode. Default is 10 °.
Ft01 to Ft16 or Ur01 to Ur16
Set the limit category (factory or user) and set (01 to 16) that the LCR meter powers up in. Default is Ft01.
54 U1731C/U1732C/U1733C User’s Guide
Setup Options 3
Setup Menu Summary
Table 3-2 Setup menu item descriptions (continued)
Legend Available settings Description
H01 to H16 or L01 to L16
0 to 19999
Set the high and low limits for each variable user set.
for the user default values.
Learn more on:
9600 or 19200
En, nonE, or odd
7bit or 8bit
Set the baud rate for remote communication with a PC
(9600 or 19200). Default is 9600.
Set the parity bit for remote communication with a PC
(even, none, or odd). Default is none.
Set the data bit length for remote communication with a
PC (7-bit or 8-bit). Default is 8-bit.
2000 Hz, 3000 Hz, 4000 Hz, or oFF
Set the LCR meter’s beep frequency (2000 Hz, 3000 Hz,
4000 Hz, or off). Default is 4000 Hz.
oFF or on Lock the LCR meter’s push buttons. Default is off.
01 to 99 mins or oFF
01 to 99 s or oFF
Set the auto power-off time-out period from 1 to 99 minutes (1 hour, 39 minutes) or off. Default is 5 minutes.
Set the LCD backlight timeout period from 1 to 99 seconds
(1 minute, 39 seconds) or off. Default is 30 seconds.
dEFA Reset the LCR meter to its factory default settings.
U1731C/U1732C/U1733C User’s Guide 55
3 Setup Options
Setup Menu Items
Setup Menu Items
Changing the initial power-on behavior
You can change the power- on behavior of your LCR meter for subsequent power cycles.
Parameter
Pon-tYPE
Pon-FrEq
Pon-AUto (L)
Pon-AUto (C)
Pon-AUto (R)
Pon-oSC
Range
Ai, Z, L, C, R, ESR, or DCR
100 Hz, 120 Hz, 1 kHz, 10 kHz, or 100 kHz
• D, Q, or ° θ
• Parallel or Series
• D, Q, or °
θ
• Parallel or Series
• D, Q, or ° θ
• Parallel or Series
FACt or USEr
Default setting
Ai
1 kHz
• Q
• Series
• D
• Series
• ° θ
• Series
FACt
Changing the power-on measurement type
Use this Setup item to change the LCR meter’s initial measurement type. You can set the LCR meter to start up in the
• auto identification mode (Ai),
• impedance measurement (Z),
• inductance measurement (L),
• capacitance measurement (C),
• resistance measurement (R),
• equivalent series resistance mode (ESR), or
• direct current resistance mode (DCR) for U1733C only
56 U1731C/U1732C/U1733C User’s Guide
Setup Options 3
Setup Menu Items
The LCR meter will start up in the selected measurement type for subsequent power cycles.
Press Press
Press Press
Press Press
Figure 3-1 Changing the power-on measurement type
U1731C/U1732C/U1733C User’s Guide 57
3 Setup Options
Setup Menu Items
Changing the power-on test frequency
Use this Setup item to change the LCR meter’s initial test frequency. You can set the LCR meter to start up using a test frequency from 100 Hz to 100 kHz.
The LCR meter will start up using the selected test frequency for subsequent power cycles.
Press Press
Press
Press
Figure 3-2 Changing the power-on test frequency
58 U1731C/U1732C/U1733C User’s Guide
Setup Options 3
Setup Menu Items
Changing the power-on secondary parameter and measurement mode for inductance (L) measurements
Use this Setup item to change the inductance ( L) measurement’s initial secondary parameter — dissipation factor ( D), quality factor (Q), or phase angle (
θ) — and measurement mode — parallel or series.
The inductance ( L) measurement will start up using the selected secondary parameter and measurement mode for subsequent power cycles.
Press
Press
Press Press Press Press
Press Press
Figure 3-3 Changing the power-on secondary parameter and measurement mode for inductance (L) measurements
U1731C/U1732C/U1733C User’s Guide 59
3 Setup Options
Setup Menu Items
Changing the power-on secondary parameter and measurement mode for capacitance (C) measurements
Use this Setup item to change the capacitance ( C) measurement’s initial secondary parameter — dissipation factor ( D), quality factor (Q), or phase angle (
θ) — and measurement mode — parallel or series.
The capacitance ( C) measurement will start up using the selected secondary parameter and measurement mode for subsequent power cycles.
Press
Press
Press Press Press Press
Press Press
Figure 3-4 Changing the power-on secondary parameter and measurement mode for capacitance (C) measurements
60 U1731C/U1732C/U1733C User’s Guide
Setup Options 3
Setup Menu Items
Changing the power-on secondary parameter and measurement mode for resistance (R) measurements
Use this Setup item to change the resistance ( R) measurement’s initial secondary parameter — dissipation factor ( D), quality factor (Q), or phase angle (
θ) — and measurement mode — parallel or series.
The resistance ( R) measurement will start up using the selected secondary parameter and measurement mode for subsequent power cycles.
Press
Press
Press Press Press Press
Press Press
Figure 3-5 Changing the power-on secondary parameter and measurement mode for resistance (R) measurements
U1731C/U1732C/U1733C User’s Guide 61
3 Setup Options
Setup Menu Items
Changing the power-on open/short CAL
Use this Setup item to change the LCR meter’s initial open/short CAL to either the factory open/short CAL ( FACt), or user open/short CAL ( USEr).
The LCR meter will start up using the selected open/short
CAL for subsequent power cycles.
Press Press
Figure 3-6 Changing the power-on open/short CAL
62 U1731C/U1732C/U1733C User’s Guide
Setup Options 3
Setup Menu Items
Changing the Ai function’s phase angle condition
This setting is used with the Ai function (see page 26 ). The
Ai function helps to identify L, C, and R measurements automatically according to the angle of impedance detected in the DUT.
Use this Setup item to change the default phase angle for the Ai function between 5° and 45°.
Parameter
Ai
Range
(5 to 45)°
Default setting
10°
shows the correlation between the phase angle detected and the L, C, and R measurements selected.
Table 3-3 Auto identification phase angle rules
Phase angle
[1]
–Set <
θ < +Set
θ ≥ +Set
θ ≤ –Set
Primary display
R
L
C
[1] Where ±Set is the phase angle selected.
Secondary display
θ
Q
D
U1731C/U1732C/U1733C User’s Guide 63
3 Setup Options
Setup Menu Items
Press Press
Press
Press
Press Press
Figure 3-7 Changing the Ai function’s phase angle condition
64 U1731C/U1732C/U1733C User’s Guide
Setup Options 3
Setup Menu Items
Changing the power-on limit category and set
This setting is used with the Limit comparison function
). There are 32 limit sets available (16 fixed factory sets, and 16 variable user sets).
Use this Setup item to change the default category (factory or user) and set (1 to 16) for subsequent power cycles.
Parameter
Pon
Range
• Factory (Ft01 to Ft16) or
• User (Ur01 to Ur16)
Default setting
Ft01
Press Press
Press
Press
Press Press
Press Press
Figure 3-8 Changing the power-on limit and category set
U1731C/U1732C/U1733C User’s Guide 65
3 Setup Options
Setup Menu Items
Changing the user high/low limit values
This setting is used with the Limit comparison function
). There are 16 variable user sets available.
Use this Setup item to change the high and low limits of each variable user set.
N O T E
The low limit can be set from 0 to less than or equal to the high limit, and the high limit can be set from more than or equal to the low limit to less than or equal to the maximum display count (19999).
66
Parameter
• H(01 to 16) or
• L(01 to 16)
Range
0 to 19999
Default setting
See
shows the user default limit values for each set.
Table 3-4 Default user high/low limit values
U04
U05
U06
U07
Set
U01
U02
U03
U08
U09
U10
High limit (H)
1000
1200
1500
1800
2200
2700
3300
3900
4700
5600
Low limit (L)
900
1080
1350
1620
1980
2430
2970
3510
4230
5040
U1731C/U1732C/U1733C User’s Guide
Press
Setup Options 3
Setup Menu Items
Table 3-4 Default user high/low limit values (continued)
Set
U11
U12
U13
U14
U15
U16
High limit (H)
6800
8200
10000
12000
15000
18000
Low limit (L)
6120
7380
9000
10800
13500
16200
Press
Press
Press
Press Press
Press
Press
Press Press
Press Press Press Press
Press Press Press Press
Figure 3-9 Changing the user high/low limit values
U1731C/U1732C/U1733C User’s Guide 67
3 Setup Options
Setup Menu Items
Changing the baud rate
This setting is used with the IR communication link and the
Agilent GUI Data Logger software to control your LCR meter
Use this Setup item to change the baud rate for remote communications with a PC.
Parameter bPS
Range
(9600 or 19200) bits/second
Default setting
9600 bits/second
Press Press
Figure 3-10 Changing the baud rate
68 U1731C/U1732C/U1733C User’s Guide
Setup Options 3
Setup Menu Items
Changing the parity check
This setting is used with the IR communication link and the
Agilent GUI Data Logger software to control your LCR meter
Use this Setup item to change the parity check for remote communications with a PC.
Parameter
PAr
Range nonE, En, or odd
Default setting nonE
Press Press
Press Press
Figure 3-11 Changing the parity check
U1731C/U1732C/U1733C User’s Guide 69
3 Setup Options
Setup Menu Items
Changing the data bits
This setting is used with the IR communication link and the
Agilent GUI Data Logger software to control your LCR meter
Use this Setup item to change the number of data bits (data width) for remote communications with a PC. The number of the stop bit is always 1, and this cannot be changed.
Parameter dAt
Range
7-bit or 8-bit
Default setting
8-bit
Press Press
Figure 3-12 Changing the data bits
70 U1731C/U1732C/U1733C User’s Guide
Setup Options 3
Setup Menu Items
Changing the beep frequency
The LCR meter’s beeper alerts users to the presence of newly sensed values for static recordings, sensed values that are out of tolerance or limits set, as well as invalid key operations.
Use this Setup item to change the driving frequency of the beeper.
Parameter bEP
Range
(2000, 3000, 4000) Hz or oFF
Default setting
4000 Hz
Press Press
Press Press
Press Press
Figure 3-13 Changing the beep frequency
U1731C/U1732C/U1733C User’s Guide 71
3 Setup Options
Setup Menu Items
Locking the push buttons
Use this Setup item to lock the push buttons (keys) of your
LCR meter. If enabled, all keys will be locked (rendered unoperational) when you exit the Setup menu.
Unlock the push buttons again by entering the Setup menu
through the power- on options ( page 10
).
Parameter
LPb
Range on or oFF
Default setting oFF
Press Press
Figure 3-14 Locking the push buttons
72 U1731C/U1732C/U1733C User’s Guide
Setup Options 3
Setup Menu Items
Changing the auto power-off and backlight time-outs
The LCR meter’s automatic power- off (see page 6 ) and backlight (see page page 6
) features use timers to determine when to turn off the backlight and when to automatically turn the LCR meter off.
Parameter
APo bLt
Range
(01 to 99) minutes or oFF
(01 to 99) seconds or oFF
Default setting
05 minutes
30 seconds
Press Press Press Press
Press
Press
Press
Press
Press Press Press Press
Figure 3-15 Changing the auto power-off and backlight time-outs
U1731C/U1732C/U1733C User’s Guide 73
3 Setup Options
Setup Menu Items
Resetting the Setup items
The Setup items can be reset to their default values through this Setup item.
Press Hold
Rec
to perform the reset. The LCR meter will beep once, exit the Setup menu, and return to normal operation.
Parameter rSt
Range dEFA
Default setting dEFA
Figure 3-16 Resetting the Setup items
74 U1731C/U1732C/U1733C User’s Guide
U1731C/U1732C/U1733C Handheld LCR Meter
User’s Guide
4
Characteristics and Specifications
Impedance/Resistance/DCR specifications 78
Phase angle of impedance specifications 81
Dissipation/Quality factor specifications 82
Source impedance of impedance/resistance measurement 84
Source impedance of capacitance measurement 85
Source impedance of inductance measurement 86
SMD Tweezers Specifications 87
This chapter lists the characteristics, assumptions, and specifications of the U1731C, U1732C, and U1733C handheld
LCR meters.
Agilent Technologies
75
4 Characteristics and Specifications
Product Characteristics
Product Characteristics
N O T E
Product characteristics specified in the table below are applicable for all
U1731C, U1732C, and U1733C models unless stated otherwise.
POWER SUPPLY
Battery type:
• 1 × 9 V Alkaline battery (ANSI/NEDA 1604A or IEC 6LR61), or
• 1 × 9 V Zinc Chloride battery (ANSI/NEDA 1604D or IEC 6F22)
Battery life:
• 16 hours typical (based on new Alkaline batteries without backlight enabled)
• Low battery indicator will flash when the battery voltage drops below 7.2 V
(approximately)
External DC adapter
• DC 12 V ± 10% or 10.8 V
MIN
to 13.2 V
MAX
POWER CONSUMPTION
225 mVA maximum (without backlight enabled)
DISPLAY
Dual display liquid crystal display (LCD)
• Primary display is 4 1/2 digits with maximum of 19999 counts
• Secondary display is 3 digits with maximum of 999 counts
MEASUREMENT RATE
• 1 time/second, nominal
OPERATING ENVIRONMENT
• Operating temperature from –10 °C to 55 °C, 0% to 80% RH
• Full accuracy up to 80% RH for temperatures up to 30 °C, decreasing linearly to 50% RH at 55 °C
• Altitude up to 2000 m
• Pollution degree II
STORAGE COMPLIANCE
–20 °C to 70 °C, 0% to 80% RH
76 U1731C/U1732C/U1733C User’s Guide
Characteristics and Specifications 4
Specification Assumptions
SAFETY AND ELECTROMAGNETIC COMPATIBILITY (EMC) COMPLIANCE
• IEC61010-1:2001/EN61010-1:2001 (Second Edition)
• IEC 61326-1:2005/EN 61326-1:2006
• Canada: ICES/NMB-001:Issue 4, June 2006
• Australia/New Zealand: AS/NZS CISPR11:2004
TEMPERATURE COEFFICIENT
0.1 × (specified accuracy) / °C (from –10 °C to 18 °C, or 28 °C to 55 °C)
INPUT PROTECTION
Reset-able over-current protection.
DIMENSIONS (W × H × D)
87 × 184 × 41 mm
WEIGHT
337 grams (with battery)
WARRANTY
Please refer to http://www.agilent.com/go/warranty_terms
• Three years for product
• Three months for product’s standard accessories, unless otherwise specified
• Please take note that for the product, the warranty does not cover:
• Damage from contamination
• Normal wear and tear of mechanical components
• Manuals and standard disposable batteries
CALIBRATION CYCLE
One year
Specification Assumptions
• Accuracy is given as ±(% of reading + counts of least significant digit) at 23 °C ± 5 °C, with relative humidity less than 80% RH.
• The measurement performed at the component test socket and necessary open and short corrections must be done prior to verifying the instrument’s accuracy.
• The accuracy is verified by design and specified type tests.
U1731C/U1732C/U1733C User’s Guide 77
4 Characteristics and Specifications
Electrical Specifications
Electrical Specifications
Specification assumptions are given on page 77 .
N O T E
Impedance/Resistance/DCR specifications
Table 4-1 Impedance/Resistance/DCR specifications
DCR 100 Hz
Accuracy = A
Z
+ Offset
120 Hz 1 kHz 10 kHz 100 kHz
Range Resolution
U1733C only All models All models All models
U1733C and
U1732C only
U1733C only
2
20
200 Ω
2000 Ω
20 k Ω
200 k Ω
2000 k Ω
20 M
200 M
0.0001
0.001
0.01
0.1
0.01 k
0.1 k
Ω
Ω
0.001 k
Ω
0.001 M
0.01 M
Ω
Ω
Ω
Ω
Ω
Ω
0.7% + 50
0.7% + 8
0.2% + 3
0.2% + 3
0.2% + 3
0.5% + 5
0.5% + 5
2.0% + 8
6.0% + 80
0.7% + 50
0.7% + 8
0.2% + 3
0.2% + 3
0.2% + 3
0.5% + 5
0.5% + 5
2.0% + 8
6.0% + 80
0.7% + 50
0.7% + 8
0.2% + 3
0.2% + 3
0.2% + 3
0.5% + 5
0.5% + 5
2.0% + 8
6.0% + 80
0.7% + 50
0.7% + 8
0.2% + 3
0.2% + 3
0.2% + 3
0.5% + 5
0.5% + 5
2.0% + 8
6.0% + 80
0.7% + 50
0.7% + 8
0.2% + 3
0.2% + 3
0.2% + 3
0.5% + 5
0.7% + 5
5.0% + 8
-
1.0% + 50
0.7% + 8
0.5% + 5
0.5% + 5
0.5% + 5
0.7% + 8
-
-
-
Notes:
1 The accuracy for the 2 Ω to 200 Ω range is specified after the Null function is used to subtract the resistance of test leads and the contact resistance.
2 For the 20 M Ω and 200 MΩ range, the RH is specified for <60%.
3 Resistance measurement is specified to Q <10 and D >0.1; otherwise accuracy is specified as ( A
Z
+ Offset ) ×
1 + Q
2
.
4 ESR (Equivalent Series Resistance) measurement is specified according to the impedance measurement and range. The maximum display is up to 199.99 k Ω and the accuracy is specified as ( A
Z
+ Offset ) × 1 + Q
2
.
78 U1731C/U1732C/U1733C User’s Guide
Characteristics and Specifications 4
Electrical Specifications
Capacitance specifications
Table 4-2 Capacitance specifications
Range
20 mF
2000 μF
200 μF
20 μF
2000 nF
200 nF
20 nF
2000 pF
200 pF
Notes:
Resolution
0.001 mF
0.1 μF
0.01 μF
0.001 μF
0.1 nF
0.01 nF
0.001 nF
0.1 pF
0.01 pF
0.001 pF
100 Hz
All models
0.5% + 8
0.5% + 5
0.3% + 3
0.2% + 3
0.2% + 3
0.2% + 3
0.5% + 5
0.5% + 10
-
-
120 Hz
Accuracy = A
C
+ Offset
1 kHz 10 kHz
All models All models
U1733C and
U1732C only
0.5% + 8
0.5% + 5
0.3% + 3
0.2% + 3
0.2% + 3
0.2% + 3
0.5% + 5
0.5% + 10
-
-
-
0.5% + 8
0.5% + 5
0.2% + 3
0.2% + 3
0.2% + 3
0.2% + 3
0.5% + 5
0.5% + 10
-
-
-
0.5% + 8
0.5% + 5
0.2% + 3
0.5% + 3
0.5% + 3
0.5% + 3
0.8% + 10
1.0% + 20
100 kHz
U1733C only
-
-
-
5.0% + 10
0.7% + 10
0.7% + 10
0.7% + 10
2.0% + 10
2.0% + 10
2.5% + 10
1 The accuracy for the 20 pF to 2000 pF range is specified after the Null function is used to subtract the stray capacitance of the test leads.
2 The accuracy for the ceramic capacitor will be influenced depending on the dielectric constant (K) of the material used to make the ceramic capacitor. For related influence factors, please refer to the Component dependency factors section in the Impedance Measurement Handbook, downloadable for free at http://www.agilent.com/find/lcrmeters .
U1731C/U1732C/U1733C User’s Guide 79
20 μH
200 μH
2000 μH
20 mH
200 mH
2000 mH
20 H
200 H
2000 H
4 Characteristics and Specifications
Electrical Specifications
Inductance specifications
Table 4-3 Inductance specifications
Range
0.001 μH
0.01 μH
0.1 μH
0.001 mH
0.01 mH
0.1 mH
0.001 H
0.01 H
0.1 H
Resolution
100 Hz
All models
-
-
0.7% + 10
0.5% + 3
0.5% + 3
0.2% + 3
0.2% + 3
0.7% + 5
1.0% + 5
120 Hz
Accuracy = A
L
+ Offset
1 kHz 10 kHz
All models All models
U1733C and
U1732C only
-
-
0.7% + 10
0.5% + 3
-
1.0% + 5
0.5% + 3
0.2% + 3
1.0% + 5
0.7% + 3
0.5% + 3
0.3% + 3
0.5% + 3
0.2% + 3
0.2% + 3
0.7% + 5
1.0% + 5
0.2% + 3
0.2% + 3
0.5% + 5
1.0% + 5
2.0% + 8
0.2% + 3
0.5% + 5
1.0% + 5
2.0% + 8
-
100 kHz
U1733C only
2.5% + 20
2.5% + 20
0.8% + 20
0.8% + 10
1.0% + 10
1.0% + 10
2.0% + 10
-
-
80 U1731C/U1732C/U1733C User’s Guide
Characteristics and Specifications 4
Electrical Specifications
Phase angle of impedance specifications
Table 4-4 Phase angle of impedance specifications
Range
–180° to 180°
Resolution
0.1°/1°
Accuracy =
θ e
⎛
⎝
AZ +
Z x
⎠
×
π
Condition
D < 1 or Q > 1
Notes:
1 The A
Z
2 The
π variable is rounded up to 3.14159.
3 The table below shows an example of the calculations for the phase angle of impedance:
Impedance
1999.9
Ω
199.9
Ω
19.9
Ω
1.9
Ω
Z
X
19999
1999
199
19
A
Z
0.2%
0.2%
0.2%
0.2%
Offset
3
3
3
3
θ e
±0.12°
±0.20°
±0.98°
±9.16°
U1731C/U1732C/U1733C User’s Guide 81
4 Characteristics and Specifications
Electrical Specifications
Dissipation/Quality factor specifications
Table 4-5 Dissipation/Quality factor specifications
Range Resolution
Z 0.001 to 999
Accuracy = D e
AZ +
Zx
×
Condition
D < 1 or Q > 1
L 0.001 to 999
AL +
Lx
×
D < 1 or Q > 1
C 0.001 to 999
AC +
Cx
×
D < 1 or Q > 1
Notes:
1 The A
Z
, A
L
, A
C
, and Offset variables are the accuracy specified at
2 The Z x
, L x
, and C x
variables are the display count of the reading. For example, the C x
value is 8888 if the capacitance is
88.88
μF for the range of 200 μF.
3 The quality factor is the reciprocal of the dissipation factor.
4 The calculation example below is for the 120 Hz frequency:
Capacitance
88.88 μF
C
X
8888
A
C
0.3%
Offset
3
D e
0.334% + 3
82 U1731C/U1732C/U1733C User’s Guide
Test signal specifications
Table 4-6 Test signal specifications
100 Hz
120 Hz
1 kHz
10 kHz
100 kHz
DCR
Selection
All models
All models
All models
U1733C and
U1732C only
U1733C only
U1733C only
Level
Test signal level
Accuracy
0.74 Vrms
0.74 Vrms
0.74 Vrms
0.05 Vrms
0.05 Vrms
0.05 Vrms
0.70 Vrms
0.70 Vrms
1.235 V
0.05 Vrms
0.05 Vrms
0.05 V
Characteristics and Specifications 4
Electrical Specifications
Test frequency
Frequency Accuracy
100 Hz
120.481 Hz
1 kHz
0.01%
0.01%
0.01%
10 kHz
100 kHz
-
0.01%
0.01%
-
U1731C/U1732C/U1733C User’s Guide 83
4 Characteristics and Specifications
Electrical Specifications
Source impedance of impedance/resistance measurement
2 Ω
20 Ω
200 Ω
2000 Ω
20 k Ω
200 k Ω
2000 k Ω
20 M Ω
200 M Ω
Table 4-7 Source impedance of impedance/resistance measurement
Typical source impedance
120 Hz 1 kHz
Range
DCR 100 Hz
U1733C only
100 Ω
100 Ω
100 Ω
1 k Ω
10 k Ω
100 k Ω
100 k Ω
100 k Ω
100 k Ω
All models
100 Ω
100 Ω
100 Ω
1 k Ω
10 k Ω
100 k Ω
100 k Ω
100 k Ω
100 k Ω
All models
100 Ω
100 Ω
100 Ω
1 k Ω
10 k Ω
100 k Ω
100 k Ω
100 k Ω
100 k Ω
All models
100 Ω
100 Ω
100 Ω
1 k Ω
10 k Ω
100 k Ω
100 k Ω
100 k Ω
100 k Ω
10 kHz
U1733C and
U1732C only
100 Ω
100 Ω
100 Ω
1 k Ω
10 k Ω
10 k Ω
10 k Ω
100 k Ω
-
100 kHz
U1733C only
100 Ω
100 Ω
100 Ω
1 k Ω
1 k Ω
1 k Ω
-
-
-
84 U1731C/U1732C/U1733C User’s Guide
Characteristics and Specifications 4
Electrical Specifications
Source impedance of capacitance measurement
Table 4-8 Source impedance of capacitance measurement
Range
100 Hz 120 Hz
Typical source impedance
1 kHz 10 kHz 100 kHz
All models All models All models U1733C and
U1732C only
U1733C only
20 mF
2000 μF
200 μF
20 μF
2000 nF
200 nF
20 nF
2000 pF
200 pF
20 pF
100 Ω
100 Ω
100 Ω
100 Ω
1 k Ω
10 k Ω
100 k Ω
100 k Ω
-
-
100 Ω
100 Ω
100 Ω
100 Ω
1 k Ω
10 k Ω
100 k Ω
100 k Ω
-
-
-
100 Ω
100 Ω
100 Ω
100 Ω
1 k Ω
10 k Ω
100 k Ω
100 k Ω
-
-
-
100 Ω
100 Ω
100 Ω
100 Ω
1 k Ω
10 k Ω
10 k Ω
10 k Ω [1]
-
-
-
100 Ω
100 Ω
100 Ω
100 Ω
1 k Ω
1 k Ω
1 k Ω
[1] This specification value is applicable for firmware versions 00.21 and above only. For firmware versions below 00.21, the impedance value is set at 100 k Ω.
U1731C/U1732C/U1733C User’s Guide 85
4 Characteristics and Specifications
Electrical Specifications
Source impedance of inductance measurement
Table 4-9 Source impedance of inductance measurement
Range
100 Hz 120 Hz
Typical source impedance
1 kHz 10 kHz 100 kHz
20 μH
200 μH
2000 μH
20 mH
200 mH
2000 mH
20 H
200 H
2000 H
All models
-
-
100 Ω
100 Ω
100 Ω
100 Ω
1 k Ω
10 k Ω
100 k Ω
All models
-
-
100 Ω
100 Ω
100 Ω
100 Ω
1 k Ω
10 k Ω
100 k Ω
All models
-
100 Ω
100 Ω
100 Ω
100 Ω
1 k Ω
10 k Ω
100 k Ω
100 k Ω
U1733C and
U1732C only
100 Ω
100 Ω
100 Ω
100 Ω
1 k Ω
10 k Ω
10 k Ω
10 k Ω [1]
-
U1733C only
100 Ω
100 Ω
100 Ω
100 Ω
1 k Ω
1 k Ω
1 k Ω
-
-
[1] This specification value is applicable for firmware versions 00.21 and above only. For firmware versions below 00.21, the impedance value is set at 100 k Ω.
86 U1731C/U1732C/U1733C User’s Guide
Characteristics and Specifications 4
SMD Tweezers Specifications
SMD Tweezers Specifications
The Agilent U1782B is a tweezers to be used with the U1700
Series Handheld LCR Meters. This tweezers is useful when measuring SMD- type components. Plug in the base of the tweezers to the LCR meter's + (HI- SENSE), – (LO- SENSE) and GUARD ends. Ensure that the orientation of the base matches the polarity of the LCR meter.
It is recommended to measure the SMD components length as well as the maximum opening of the tweezers. The length of the tweezers is approximately 815 mm (32.08 inches) (see
).
815 ± 20 mm
150 mm
8.4 mm
Figure 4-1 U1782B SMD tweezers
SMD tweezers base
U1731C/U1732C/U1733C User’s Guide 87
4 Characteristics and Specifications
SMD Tweezers Specifications
Electrical characteristics
Table 5 U1782B SMD tweezers electrical characteristics
Parameters
Cp
Parallel capacitance
Rs
Series resistance
Ls
Series inductance
Notes:
Test condition
Tweezers open
Tweezers short
Tweezers short
100 Hz
<0.7 pF
<0.5 Ω
<1.2 µH
120 Hz
<0.7 pF
<0.5 Ω
<1.2 µH
1 kHz
<0.7 pF
<0.5 Ω
<1.2 µH
10 kHz
<0.7 pF
<0.5 Ω
<1.2 µH
1 The accuracy is specified at 23 °C ± 5 °C and <75% R.H.
2 You are recommended to perform an open/short calibration on the LCR meter before using the tweezers.
100 kHz
<0.7 pF
<0.5 Ω
<1.2 µH
88 U1731C/U1732C/U1733C User’s Guide
www.agilent.com
Contact us
To obtain service, warranty, or technical assistance, contact us at the following phone or fax numbers:
United States:
(tel) 800 829 4444 (fax) 800 829 4433
Canada:
(tel) 877 894 4414 (fax) 800 746 4866
China:
(tel) 800 810 0189 (fax) 800 820 2816
Europe:
(tel) 31 20 547 2111
Japan:
(tel) (81) 426 56 7832 (fax) (81) 426 56 7840
Korea:
(tel) (080) 769 0800 (fax) (080) 769 0900
Latin America:
(tel) (305) 269 7500
Taiwan:
(tel) 0800 047 866 (fax) 0800 286 331
Other Asia Pacific Countries:
(tel) (65) 6375 8100 (fax) (65) 6755 0042
Or visit Agilent World Wide Web at: www.agilent.com/find/assist
Product specifications and descriptions in this document are subject to change without notice. Always refer to the Agilent website for the latest revision.
© Agilent Technologies, Inc., 2011–2012
Seventh Edition, November 28, 2012
U1731-90077
Agilent Technologies
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project