Specifications apply under the following conditions:

90 minutes warm-up time at ambient temperature, specified environmental conditions met, calibration cycle adhered to, and all internal adjustments performed. Data designated „overrange“ and data without tolerances: typical values only. Unless stated otherwise, specifications apply to test ports and a nominal source power of –10 dBm.

CONTENTS

Measurement range ...................................................................................................................................................................................3

Measurement speed...................................................................................................................................................................................5

Sweep times of R&S

®

ZVB ......................................................................................................................................................................5

Measurement accuracy ..............................................................................................................................................................................6

Effective system data .................................................................................................................................................................................9

Test port output ..........................................................................................................................................................................................9

Test port input...........................................................................................................................................................................................12

Additional front panel connectors .............................................................................................................................................................12

Display......................................................................................................................................................................................................13

Rear panel connectors .............................................................................................................................................................................13

Options .....................................................................................................................................................................................................14

General data.............................................................................................................................................................................................14

Ordering information.................................................................................................................................................................................15

2 Vector Network Analyzer R&S

®

ZVB Version 06.00, August 2006

Measurement range

Impedance

Test port connector R&S

®

ZVB4 and R&S

®

ZVB8

R&S

®

ZVB20

Number of test ports

Frequency range

Static frequency accuracy

Frequency resolution

Number of measurement points

Measurement bandwidths

Dynamic range of R&S

®

ZVB4 two-port model and R&S

®

ZVB8 two-port model

(without optional step attenuators)

R&S

®

ZVB4

R&S

®

ZVB8

R&S

®

ZVB20 without optional oven quartz with optional oven quartz per trace

1/2/5 steps from PORT 1 to PORT 2

300 kHz to 5 MHz

5 MHz to 100 MHz

100 MHz to 4 GHz

4 GHz to 7 GHz (R&S

®

ZVB8 only)

7 GHz to 7.5 GHz (R&S

®

ZVB8 only)

7.5 GHz to 8 GHz (R&S

®

ZVB8 only) from PORT 1 to PORT 2 and from PORT 3 to PORT 4 (for four-port models only)

300 kHz to 5 MHz

5 MHz to 100 MHz

100 MHz to 500 MHz

500 MHz to 4 GHz

4 GHz to 7 GHz (R&S

®

ZVB8 only)

7 GHz to 7.5 GHz (R&S

®

ZVB8 only)

7.5 GHz to 8 GHz (R&S

®

ZVB8 only)

50 Ω type N, female

3.5 mm, male

2, 3, or 4

300 kHz to 4 GHz

300 kHz to 8 GHz

10 MHz to 20 GHz

8 × 10

–6

1 × 10

–7

1 Hz

2 to 20001

1 Hz to 500 kHz

>80 dB, typ. 100 dB

>100 dB, typ. 120 dB

>120 dB, typ. 130 dB

>117 dB, typ. 127 dB

>115 dB, typ. 120 dB

>108 dB, typ. 118 dB

Dynamic range of R&S

®

ZVB4 three-port and four-port models and R&S

®

ZVB8 three-port and four-port models (without optional step attenuators)

Dynamic range of R&S

®

ZVB20 from PORT 1 to PORT 2 and from PORT 3 to PORT 4 (for four-port model only)

10 MHz to 100 MHz

100 MHz to 700 MHz

>80 dB, typ. 100 dB

>100 dB, typ. 120 dB

>120 dB, typ. 130 dB

>123 dB, typ. 130 dB

>120 dB, typ. 130 dB

>115 dB, typ. 125 dB

>108 dB, typ. 125 dB

>80 dB, typ. 110 dB

>100 dB, typ. 130 dB

700 MHz to 8 GHz

8 GHz to 20 GHz

>120 dB, typ. 133 dB

>110 dB, typ. 122 dB

The dynamic range is defined as the difference between the maximum source power and the rms value of the data trace of the transmission magnitude produced by noise and crosstalk with test ports short-circuited. The specification is valid without system error correction and at 10 Hz measurement bandwidth. The dynamic range can be increased by using a measurement bandwidth of 1 Hz.

Version 06.00, August 2006 Vector Network Analyzer R&S

®

Dynamic range in dB versus frequency in GHz of R&S

®

ZVB4

Dynamic range in dB versus frequency in GHz range of R&S

®

ZVB8

Dynamic range in dB versus frequency in GHz range of R&S

®

ZVB20

4 Vector Network Analyzer R&S

®

ZVB Version 06.00, August 2006

Measurement speed

Measurement time

Measurement time per point for 201 measurements points, with span

100 MHz, 500 kHz measurement bandwidth, ALC and display switched off with center frequency 1.1 GHz with center frequency 5.1 GHz

CW mode, 500 kHz measurement bandwidth

Data transfer time for 201 measurements points via IEC/IEEE bus via VX11 over 100 Mbit/s LAN via RSIB over 100 Mbit/s LAN

Time for measurement and data transfer for 201 measurements points, with start frequency 1 GHz, stop frequency 1.1 GHz,

500 kHz measurement bandwidth, and display switched off

(No additional time for data transfer is needed, as it is done simultaneously during the measurement.)

Switching time between channels

Switching time between two preloaded instrument settings with not more than 2001 points with not more than 2001 points

<6 ms

<4.5 ms

<4.5 µs

<2.9 ms

<1.3 ms

<0.7 ms

<6 ms

<1 ms

<10 ms

Sweep times of R&S

®

ZVB

Number of measurement points 51 101 201 401 801

Start frequency 5 GHz, stop frequency 5.2 GHz, ALC off, and measurement bandwidth of 100 kHz with full one port calibration or with correction switched off

2.4 ms 3.9 ms 6.3 ms 11 ms 20.4 ms with TOSM calibration 4.7 ms 8.6 ms 16.4 ms 32.4 ms 65 ms

1601

40.2 ms

170 ms

Start frequency 6 GHz, stop frequency 8 GHz, ALC off, and measurement bandwidth of 100 kHz with full one port calibration or with correction switched off with TOSM calibration with full one port calibration or with correction switched off with TOSM calibration

3.4 ms

5.3 ms 9.8 ms 18 ms 33 ms 63 ms

Start frequency 10 MHz, stop frequency 4 GHz (R&S

®

ZVB4), 8 GHz (R&S

®

ZVB8), or 20 GHz (R&S

ALC off, and measurement bandwidth of 100 kHz

®

ZVB20),

8.4 ms 12.6 ms 19.5 ms 30.5 ms 53.2 ms

10.3 ms

6.2 ms

16.6 ms

11 ms

28 ms

17.3 ms

47 ms

28.2 ms

81 ms

49 ms

160 ms

88.2 ms

190 ms

Version 06.00, August 2006 Vector Network Analyzer R&S

®

Measurement accuracy

This data is valid between +18 °C and +28 °C, provided the temperature has not varied by more than 1 K after calibration. Validity of the data is conditional on the use of a suitable calibration kit by which the effective system data specified below is achieved.

Frequency points, measurement bandwidth, and sweep time have to be identical for measurement and calibration (no interpolation allowed).

Accuracy of transmission measurements

R&S

®

ZVB4 and R&S

®

ZVB8

300 kHz to 50 MHz +15 dB to –30 dB

Above 50 MHz

–30 dB to –45 dB

+15 dB to +5 dB

+5 dB to –55 dB

–55 dB to –70 dB

–70 dB to –85 dB

R&S

®

ZVB20

10 MHz to 50 MHz

50 MHz to 400 MHz

+15 dB to –30 dB

+15 dB to –30 dB

–30 dB to –45 dB

400 MHz to 700 MHz

700 MHz to 8 GHz

+15 dB to +5 dB

+5 dB to –35 dB

–35 dB to –50 dB

–50 dB to –65 dB

+15 dB to +5 dB

+5 dB to –55 dB

–55 dB to –70 dB

–70 dB to –85 dB

<0.2 dB or <2°

<1 dB or <6°

<0.2 dB or <2°

<0.1 dB or <1°

<0.2 dB or <2°

<1 dB or <6°

<1 dB or <6°

<0.2 dB or <2°

<1 dB or <6°

<0.2 dB or <2°

<0.1 dB or <1°

<0.2 dB or <2°

<1 dB or <6°

<0.2 dB or <2°

<0.1 dB or <1°

<0.2 dB or <2°

<1 dB or <6°

8 GHz to 20 GHz +15 dB to +5 dB

+5 dB to –35 dB

–35 dB to –50 dB

<0.2 dB or <2°

<0.1 dB or <1°

<0.2 dB or <2°

–50 dB to –65 dB <1 dB or <6°

Specifications are based on a matched DUT, a measurement bandwidth of 10 Hz, and a nominal source power of –10 dBm.

Magnitude

Phase

10

1

0.1

0

0.01

-20 -40 -60 -80

Transmission coefficient in dB

-100

100

10

1

0

0.1

-20 -40 -60 -80

Transmission coefficient in dB

-100

Typical accuracy of transmission magnitude and transmission phase measurements of R&S

®

ZVB4 and R&S

®

ZVB8 in the frequency range from 300 kHz to 4 GHz

6 Vector Network Analyzer R&S

®

ZVB Version 06.00, August 2006

Magnitude

10

1

0.1

0

0.01

-20 -40 -60 -80

Transmission coefficient in dB

-100

100

10

1

Phase

0

0.1

-20 -40 -60 -80

Transmission coefficient in dB

Typical accuracy of transmission magnitude and transmission phase measurements of R&S

®

ZVB8 in the frequency range from 4 GHz to 8 GHz

-100

Magnitude

Phase

10

100

1 10

0.1

1

0

0.01

-20 -40 -60 -80

Transmission coefficient in dB

-100

0

0.1

-20 -40 -60 -80

Transmission coefficient in dB

Typical accuracy of transmission magnitude and transmission phase measurements of R&S

®

ZVB20 in the frequency range from 10 MHz to 700 MHz

-100

10

1

0.1

0

0.01

-20 -40 -60 -80 -100

0

100

10

1

0.1

-20 -40 -60 -80 -100

Typical accuracy of transmission magnitude and transmission phase measurements of R&S

® in the frequency range from 700 MHz to 20 GHz

ZVB20

Version 06.00, August 2006 Vector Network Analyzer R&S

®

Accuracy of reflection measurements

R&S

®

ZVB4 and R&S

®

ZVB8

Above 300 kHz +10 dB to +3 dB

+3 dB to –15 dB

–15 dB to –25 dB

–25 dB to –35 dB

<0.6 dB or <4°

<0.4 dB or <3°

<1 dB or <6°

<3 dB or <20°

R&S

®

ZVB20

10 MHz to 50 MHz

50 MHz to 20 GHz

+3 dB to –15 dB

–15 dB to –25 dB

+10 dB to +3 dB

+3 dB to –15 dB

<1 dB or <6°

<3 dB or <20°

<0.6 dB or <4°

<0.4 dB or <3°

–15 dB to –25 dB

–25 dB to –35 dB

<1 dB or <6°

<3 dB or <20°

Specifications are based on an isolating DUT, a measurement bandwidth of 10 Hz, and a nominal source power of –10 dBm.

Magnitude

Phase

0.05

0.04

0.03

0.02

0.01

0.00

0 0.2

0.4

0.6

0.8

Reflection coefficient (linear)

1

10

8

6

4

2

0

0 0.2

0.4

0.6

Reflection coefficient (linear)

0.8

1

Trace stability

Trace noise of S11 (rms)

Typical accuracy of reflection magnitude and reflection phase measurements of R&S of R&S

®

® and of R&S

ZVB4 in the frequency range from 300 kHz to 4 GHz,

ZVB8 in the frequency range from 300 kHz to 8 GHz,

®

ZVB20 in the frequency range from 50 MHz to 20 GHz

Temperature dependence at 0 dBm source power and 0 dB reflection and 1 kHz measurement bandwidth

R&S

®

ZVB4 and R&S

®

ZVB8

Above 300 kHz <0.004 dB, typ. 0.001 dB

R&S

®

ZVB20

700 MHz to 8 GHz

8 GHz to 20 GHz at 0 dB transmission or reflection

<0.004 dB, typ. 0.001 dB

<0.015 dB, typ. 0.004 dB

<0.05 dB/K or <0.4°/K

8 Vector Network Analyzer R&S

®

ZVB Version 06.00, August 2006

Effective system data

This data is valid between +18 °C and +28 °C, provided the temperature has not varied by more than 1 K after calibration. The data is based on a measurement bandwidth of 10 Hz and system error calibration by means of a suitable calibration kit. Frequency points, measurement bandwidth, and sweep time have to be identical for measurement and calibration (no interpolation allowed).

R&S

®

ZVB4 and R&S

®

ZVB8

Directivity

Source match

Reflection tracking

Load match

Transmission tracking

300 kHz to 4 GHz

4 GHz to 8 GHz (R&S

®

ZVB8 only)

300 kHz to 4 GHz

4 GHz to 8 GHz (R&S

®

ZVB8 only)

300 kHz to 4 GHz

4 GHz to 8 GHz (R&S

®

ZVB8 only)

300 kHz to 4 GHz

4 GHz to 8 GHz (R&S

®

ZVB8 only)

300 kHz to 4 GHz

4 GHz to 8 GHz (R&S

®

ZVB8 only)

>46 dB, typ. 50 dB

>40 dB, typ. 50 dB

>40 dB, typ. 46 dB

>36 dB, typ. 40 dB

<0.04 dB, typ. 0.01 dB

<0.1 dB, typ. 0.01 dB

>46 dB, typ. 50 dB

>40 dB, typ. 46 dB

<0.06 dB, typ. 0.01 dB

<0.2 dB, typ. 0.05 dB

R&S

®

ZVB20

Directivity

Source match

Reflection tracking

Load match

Transmission tracking

10 MHz to 700 MHz

700 MHz to 20 GHz

10 MHz to 700 MHz

700 MHz to 20 GHz

10 MHz to 700 MHz

700 MHz to 20 GHz

10 MHz to 700 MHz

700 MHz to 20 GHz

10 MHz to 700 MHz

700 MHz to 20 GHz

>36 dB, typ. 40 dB

>40 dB, typ. 50 dB

>30 dB, typ. 48 dB

>30 dB, typ. 48 dB

<0.3 dB, typ. 0.05 dB

<0.3 dB, typ. 0.05 dB

>36 dB, typ. 40 dB

>40 dB, typ. 50 dB

<0.3 dB, typ. 0.1 dB

<0.3 dB, typ. 0.1 dB

Test port output

Power range

(without optional step attenuators)

Power accuracy

(with ALC on and

without power calibration)

Power linearity

Power resolution

R&S

®

ZVB4 and R&S

®

ZVB8

300 kHz to 50 MHz –40 dBm to +10 dBm, typ. –45 dBm to +14 dBm

50 MHz to 4 GHz –40 dBm to +13 dBm, typ. –45 dBm to +15 dBm

4 GHz to 7 GHz (R&S

®

ZVB8 only) –40 dBm to +10 dBm, typ. –45 dBm to +13 dBm

7 GHz to 8 GHz (R&S

®

ZVB8 only) –40 dBm to +8 dBm, typ. –45 dBm to +12 dBm

R&S

®

ZVB20

10 MHz to 13 GHz –30 dBm to +10 dBm, typ. –40 dBm to +15 dBm

13 GHz to 20 GHz –30 dBm to +5 dBm, typ. –40 dBm to +10 dBm

R&S

®

ZVB4 and R&S

®

ZVB8 at –10 dBm in temperature range +18 °C to +28 °C

<2 dB above 50 MHz <0.8 dB, typ. 0.3 dB

R&S

®

ZVB20 at –10 dBm in temperature range +18 °C to +28 °C

<3 dB

50 MHz to 20 GHz referenced to –10 dBm

<0.8 dB, typ. 0.3 dB

<2 dB in temperature range +18 °C to +28 °C above 50 MHz <0.8 dB, typ. 0.2 dB

0.01 dB

Version 06.00, August 2006 Vector Network Analyzer R&S

®

Harmonics

(Output power is referenced to maximum specified output power)

R&S

®

ZVB4 and R&S

®

ZVB8

300 kHz to 50 MHz at –3 dB typ. <–30 dBc

50 MHz to 4 GHz at –5 dB <–20 dBc, typ. <–30 dBc

4 GHz to 7 GHz at –2 dB (R&S

®

ZVB8) <–20 dBc, typ. <–30 dBc

7 GHz to 8 GHz at 0 dB (R&S

®

ZVB8)

R&S

®

ZVB20

<–20 dBc, typ. <–30 dBc

10 MHz to 50 MHz at –3 dB

50 MHz to 13 GHz at 0 dB

13 GHz to 20 GHz at 0 dB typ. <–30 dBc

<–20 dBc, typ. <–30 dBc

<–20 dBc, typ. <–30 dBc

Maximum output power in dBm versus frequency in GHz of R&S

®

ZVB4

Maximum output power in dBm frequency in GHz of R&S

®

ZVB8

Maximum output power in dBm versus frequency in GHz of R&S

®

ZVB20

10 Vector Network Analyzer R&S

®

ZVB Version 06.00, August 2006

Output power accuracy in dB versus frequency in GHz of R&S

®

ZVB4

Output power accuracy in dB versus frequency in GHz of R&S

®

ZVB8

Version 06.00, August 2006

Output power accuracy in dB versus frequency in GHz of R&S

®

ZVB20

Vector Network Analyzer R&S

®

Test port input

Match

Maximum nominal input level

Power measurement accuracy

Receiver linearity without system error correction

R&S

®

ZVB4

300 kHz to 8 GHz

R&S

®

ZVB8

300 kHz to 7 GHz

>16 dB

7 GHz to 8 GHz

R&S

®

ZVB20

10 MHz to 50 MHz

50 MHz to 2 GHz

2 GHz to 20 GHz

R&S

®

ZVB4 and R&S

®

R&S

®

ZVB20

10 MHz to 8 GHz

>16 dB

>14 dB

>10 dB

>16 dB

>8 dB

8 GHz to 20 GHz at –10 dBm without power calibration

+10 dBm

0 dBm in temperature range +18 °C to +28 °C

10 MHz to 8 GHz 1 dB

8 GHz to 20 GHz

(R&S

®

ZVB20 only) referenced to –10 dBm

2 dB in temperature range +18 °C to +28 °C

R&S

®

ZVB4 and R&S

®

ZVB8 for +20 dB to –60 dB

50 MHz to 4 GHz 0.1 dB

4 GHz to 6 GHz (R&S

®

ZVB8 only) 0.1 dB

6 GHz to 8 GHz (R&S

®

ZVB8 only) 0.2 dB

R&S

®

ZVB20 for +20 dB to –30 dB

Damage level

Damage DC voltage

Noise level

(without optional step attenuators)

50 MHz to 700 MHz for +20 dB to +10 dB

700 MHz to 8 GHz for +15 dB to +10 dB

0.1 dB

0.3 dB

8 GHz to 20 GHz for +10 dB to –45 dB

0.3 dB

700 MHz to 20 GHz 0.1 dB

+27 dBm

30 V at 10 Hz measurement bandwidth

R&S

®

ZVB4 and R&S

®

ZVB8

300 kHz to 100 MHz <–70 dBm

100 MHz to 4 GHz

4 GHz to 8 GHz (R&S

®

ZVB8 only)

<–110 dBm

<–105 dBm

R&S

®

ZVB20

100 MHz to 700 MHz

700 MHz to 8 GHz

<–70 dBm

<–105 dBm

8 GHz to 20 GHz

The noise level is defined as the rms value of the indicated noise floor.

<–100 dBm

Additional front panel connectors

USB (two) universal serial bus connectors for connecting USB devices (USB 1.1); two additional USB connectors at the rear panel

12 Vector Network Analyzer R&S

®

ZVB Version 06.00, August 2006

Display

Screen

Resolution

21 cm (8.4") diagonal color LCD

800 × 600 × 262144 (high color)

Rear panel connectors

IEC BUS

remote control in line with IEEE488, IEC60625; 24 pins first local area network connector, 8 pins, RJ-45

LAN 1

LAN 2

USB

second local area network connector, 8 pins, RJ-45

(two) Universal Serial Bus connectors for connecting USB devices (USB 1.1); two additional USB connectors at the front panel

10 MHz REF

Connector type

Input frequency

Maximum allowed deviation

Input power

Input impedance

Output frequency

Output frequency accuracy

Output power

DC MEAS 1 V

Connector type

Voltage range

Measurement accuracy

Input impedance

Damage voltage

DC MEAS 10 V

Connector type

Voltage range

Measurement accuracy

Input impedance

Damage voltage

PORT BIAS

Connector type

Maximum nominal input voltage

Maximum nominal input current

Damage voltage

Damage current

MONITOR

altenatively input or output for external frequency reference signal

BNC, female

10 MHz

1 kHz

0 dBm

± 3 dB

50

Ω

10 MHz

80 Hz

–3 dBm ± 8 dB at 50 Ω

DC measurement input

DC measurement input

DC bias input for PORT

4-pin Mini DIN, female

–1 V to +1 V

2.5 % of reading + 2.5 mV

>10 k

Ω

30 V

4-pin Mini DIN, female

–10 V to +10 V

2.5 % of reading + 25 mV

>10 k

Ω

30 V

BNC, female

30 V

200 mA

30 V

500 mA

IBM-PC-compatible VGA monitor connector, 15-pin D-Sub (for external monitor)

Version 06.00, August 2006 Vector Network Analyzer R&S

®

ZVB

USER CONTROL

several control and trigger signals, 25-pin D-Sub, 3.3 V TTL for controlling external generators, for limit checks, sweep signals, etc

FOOT SWITCH 1 and FOOT SWITCH 2 pin 24 and pin 25 (inputs)

DRIVE PORT 1 to DRIVE PORT 4 pin 16 to pin 19 (outputs) control inputs indicate driving port

CHANNEL BIT 0 to CHANNEL BIT 3

PASS 1 and PASS 2

BUSY

READY FOR TRIGGER

EXT GEN TRIGGER

EXT GEN BLANK

EXTERNAL TRIGGER pin 13 and pin 14 (outputs) pin 4 (output) pin 6 (output) pin 21 (output) pin 22 (input) pin 2 (input) pass/fail results of limit checks measurements running ready for trigger control signal for external generator handshake signal from external generator trigger input for analyzer

EXT TRIGGER

Connector type

TTL signal (edge triggerd)

Polarity (selectable)

Minimum pulse width

Input impedance trigger input for analyzer

BNC, female

3 V positive or negative

1 µs

>10 k Ω

Options

Generator Step Attenuators

Frequency range

Power range

Power accuracy

Power linearity

Dynamic range extend the lower limit of the output power range of R&S

®

ZVB8 by 70 dB

300 kHz to 8 GHz upper limit is reduced by 6 dB, lower limit is extended by 70 dB at –10 dBm without power calibration above –30 dBm identical to specifications without optional step attenuators identical to specifications without optional step attenuators

–30 dBm to –70 dBm

–70 dBm to –100 dBm

<2 dB

<3 dB is reduced by 6 dB

General data

Temperature loading operating temperature range storage temperature range

Damp heat

Mechanical resistance

Calibration interval

EMC

EMC

Safety vibration, sinusoidal vibration, random shock

RF emission other emissions and immunity

+5 °C to +40 °C

–40 °C to +70 °C in line with IEC 60068-2-1 and

IEC 60068-2-2

+40 °C at 95 % rel. humidity, in line with IEC 60068-2-30

5 Hz to 150 Hz, in line with IEC 60068-2-6

10 Hz to 300 Hz, in line with IEC 60068-2-64

40 g shock spectrum, in line with IEC 60068-2-27, MIL-STD 810

1 year in line with CISPR 11/EN 55011 Group 1

Class B (for a shielded test set-up) in line with IEC/EN 61326, emission class B (residential environment), immunity industrial environment (excluding operating frequency) in line with IEC 61010-1, EN61010-1, and

UL 3111-1

14 Vector Network Analyzer R&S

®

ZVB Version 06.00, August 2006

Power supply

Power consumption

Conformity mark

Dimensions ( W × H × D )

Weight

R&S

®

ZVB4 and R&S

®

ZVB8

R&S

®

ZVB20 with two or three ports

R&S

®

ZVB20 with four ports

R&S

®

ZVB4 and R&S

®

R&S

®

ZVB20 with two or three ports

R&S

®

ZVB20 with four ports

100 V to 240 V (AC) with tolerance ±10 %,

50 Hz to 60 Hz with tolerance ±5 %, safety class I to VDE 411

450 W, typ. 350 W (standby: typ. 10 W)

VDE, GS, CSA, CSA-NRTL/C, CE

465.1 mm × 241.8 mm × 417.0 mm

465.1 mm × 241.8 mm × 417.0 mm

465.1 mm × 286.2 mm × 417.0 mm

20 kg

25 kg

Ordering information

Designation Type

Vector Network Analyzer, 4 GHz, 2 ports R&S

®

ZVB4 1145.1010.04

Vector Network Analyzer, 4 GHz, 3 ports R&S

®

ZVB4 1145.1010.05

Vector Network Analyzer, 4 GHz, 4 ports R&S

®

ZVB4 1145.1010.06

Vector Network Analyzer, 8 GHz, 2 ports R&S

®

ZVB8 1145.1010.08

Vector Network Analyzer, 8 GHz, 3 ports R&S

®

ZVB8 1145.1010.09

Vector Network Analyzer, 8 GHz, 4 ports R&S

®

ZVB8 1145.1010.10

Vector Network Analyzer, 20 GHz, 2 ports R&S

®

ZVB20 1145.1010.20

Vector Network Analyzer, 20 GHz, 4 ports R&S

®

ZVB20 1145.1010.22

Options

Generator Step Attenuator Port 1 for R&S

®

ZVB8

(max. two step attenuators insertable)

Generator Step Attenuator Port 2 for R&S

®

ZVB8

(max. two step attenuators insertable)

Generator Step Attenuator Port 3 for R&S

®

ZVB8 with three ports or four ports

(max. two step attenuators insertable)

R&S

R&S

R&S

®

®

®

ZVB8-B21 1302.5480.02

ZVB8-B22 1302.5073.02

ZVB8-B23 1302.5496.02

Oven Quartz (OCXO)

Time Domain

Mixer and Harmonic Measurements

R&S

®

ZVAB-B4 1164.1757.02

R&S

®

ZVAB-K2 1164.1657.02

R&S

®

ZVB-K3 1164.1592.02

Version 06.00, August 2006 Vector Network Analyzer R&S

®

Certified Quality System

ISO

9001

DQS REG. NO 1954 QM

Certified Environmental System

ISO

14001

DQS REG. NO 1954 UM

For product brochure, see PD 0758.1529.12 and www.rohde-schwarz.com

(search term: ZVB)

www.rohde-schwarz.com

Europe: +49 1805 12 4242, [email protected]

USA and Canada: +1-888-837-8772, [email protected]

Asia: +65 65 130 488, [email protected]

Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project