86142B - Agilent / HP Optical Spectrum Analyzers

86142B - Agilent / HP Optical Spectrum Analyzers
Agilent 8614xB
Optical Spectrum Analyzer Family
Technical Specifications
August 2003
·
Filter Mode
Enables you to drop a single DWDM channel or
measure time resolved chirp (TRC) and calculate
dispersion penalty (DPC).
·
Excellent “Close-In” Dynamic Range
Accurately characterize 50 GHz WDM system
performance
·
High Throughput
Fast sweep speeds at high sensitivity to
maximize measurement throughput
·
Built-In Applications
Agilent’s powerful application concept makes
complex and repetitive measurements simple
·
Benchtop and Portable Platforms
Choose between a large screen or small
footprint package
Benchtop
Portable
Ideal for critical WDM system and component characterization
Agilent 86142B
Agilent 86145B
Ideal for a wide range of applications at value prices
Agilent 86140B
Agilent 86143B
Features filter mode, single mode monochromator output
Agilent 86146B
Agilent Technologies offers a wide variety of optical spectrum analyzers (OSA) to meet your test needs
whether it’s in R&D, manufacturing, installation, or maintenance and commissioning. Both benchtop and
portable models are available at different price and performance points so you can choose the most cost
effective solution to meet your test needs.
Specifications
Characteristics and Specifications
The distinction between specifications and characteristics is described as follows:
·
·
Specifications describe warranted performance.
Characteristics provide useful, but non-warranted information about the functions and performance of the
instrument.
The specifications apply to all functions autocoupled over the temperature range 0 to 55° C and relative
humidity <95% (unless otherwise noted).
All specifications apply after the instrument’s temperature has been stabilized after 1 hour continuous operation
and the auto-align routine has been run. Unless otherwise noted, specifications apply without USER CAL.
The 86146B specifications are for the 50 µm internal path only.
Wavelength
Range
Span Range
Accuracy
After calibration with internal calibration
source and with enhanced wavelength
calibration on for specified range. *
1480-1570 nm
1570-1620 nm
After calibration with external reference
source(s)
±10 nm of calibration reference point(s)*
After user calibration over full
wavelength range (600-1700 nm) *
Absolute Accuracy
Tuning Repeatability *
Reproducibility (£1 min) *
Span Linearity *
1525-1570 nm
for spans <40 nm
Resolution Bandwidth (RBW)
FWHM (3 dB Bandwidth) *
Noise Marker Bandwidth Accuracy
using noise markers 1525-1610 nm
³0.5 nm
0.2 nm
0.1 nm
0.06 nm
Agilent 8614xB
Notes
600 nm to 1700 nm
0.2 nm to full range and zero span
±0.01 nm
±0.025 nm
±0.01 nm
±0.2 nm
T(20-30°C)
±0.5 nm
±0.002 nm
±0.002 nm
Factory cal. 2 yr. cycle
Char ., T(20-30°C)
±0.01 nm,
±0.02 nm
Agilent 86140B, 86142B,
86143B, 86145B
0.06, 0.1, 0.2, 0.5, 1, 2, 5,
10 nm
±2%
±3%
±7%
±12%
Char. indicates the number is a characteristic.
T(#) indicates temperature dependence.
* With applied input fiber 9/125 µm.
Page 2 of 10
Agilent 86146B
0.06, 0.07, 0.1, 0.14,
0.2, 0.33, 0.5, 1, 2, 5,
10 nm
Notes
Resolution of 10 nm is
available for first order
grating response only.
Agilent 8614xB
Amplitude
Sensitivity
Sensitivity is defined as
signal value > 6 x RMS
noise value.
(0-30°C)
nd
T
, 2 Order
600-750 nm
750-900 nm
900-1250 nm
1250-1610 nm
1610-1700 nm
Maximum Measurement Power
-60dBm
-75 dBm
-75 dBm
-90 dBm
-80 dBm
1525-1700 nm
600-1000 nm
1000-1525 nm
Maximum Safe Power
Total safe power
Total power within any 10 nm
portion of the spectrum
Absolute Accuracy
At –20dBm, 1310 nm/1550 nm
Scale Fidelity
+15 dBm per channel, +30 dBm total
+15 dBm per channel, +30 dBm total
+12 dBm per channel, +30 dBm total
Autorange off
Autorange on
Display Scale (log scale)
Amplitude Stability
(1310 nm, 1550 nm)
1 minute
15 minutes
Flatness*
1290-1330 nm
1525-1570 nm
1525-1610 nm
1250-1610 nm
Polarization Dependence*
1310 nm
1530 nm, 1565 nm
1600 nm
1250-1650 nm
Notes
T
(0-30°C)
(20-30°C)
T
Resolution bandwith
setting < channel spacing.
Char.
+30 dBm
+23 dBm
± 0.5 dB
± 0.05 dB
± 0.07 dB
0.01-20 dB/DIV, -120 to +90 dBm
±0.01 dB
±0.02 dB
Agilent 86140B,
Agilent 86142B,
86143B
86145B, 86146B
±0.2 dB
±0.2 dB
±0.2 dB
¾
¾
±0.2 dB
±0.7 dB
±0.25 dB
±0.2 dB
±0.25 dB
±0.3 dB
The 86146B specifications are for the 50 mm internal path only.
Char. indicates the number is a characteristic.
T(#) indicates temperature dependence.
* With applied input fiber 9/125 mm.
Page 3 of 10
±0.12 dB
±0.05 dB
±0.08 dB
±0.25 dB
For resolution ³ 0.1 nm
Excluding amplitude errors
at low power levels due to
noise.
(20-30°C)
T
For signals within 8 dB of
top of screen
Char.
Absorption of light by
atmospheric moisture
affects flatness
at 1350-1420 nm
For resolution ³0.2 nm,
(room)
T
Dynamic Range
Agilent 86140B,
86143B
Agilent 86142B,
86145B, 86146B
In 0.1 nm Resolution Bandwidth *
1250-1610 nm (chop mode on)
±0.5 nm, ± 1nm, ±5 nm
1550 nm
At ±0.8 nm (±100 GHz at 1550 nm)
-70 dB
At ±0.5 nm (±62.5 GHz at 1550 nm)
At ±0.4 nm (±50 GHz at 1550 nm)
At ±0.2 nm (±25 GHz at 1550 nm)
-58 dB
-55 dB
Input Return Loss
Straight connector (9/125 mm)
Graphics export
Instrument Drivers
Notes
Depends on the qualitiy of the
attached connector.
Notes
Char.
<180 ms
<340 ms
<400 ms
<650 ms
Char.
<±0.5 ms
2 ms-6.5 ms
Char.
Agilent 86140B,
Agilent 86142B,
86143B
86145B, 86146B
<±0.2 dB (starting from dark)
<±0.2 dB
<±0.2 dB (30 dB
extinction)
Notes
Agilent 8614xB
Notes
Computer Interfacing
Floppy Disk
Data export
Char.
Agilent 8614xB
40 nm/56.3 ms
50 ms/trace point
Max. Sweep Rate
Max. Sampling Rate in Zero Span
Sweep Cycle Time
50 nm span, auto zero off
50 nm span, auto zero on
100 nm span
500 nm span
ADC Trigger Accuracy
Jitter (distributed uniformly)
Trigger delay range
Remote Control
Compatibility
Interfaces
-40 dB
>35 dB
Sweep
Turn On (³2ms after rising edge)
Turn Off (³10ms after falling edge)
Average of all states of
polarization
Char. (86140B, 86143B)
Agilent 8614xB
Monochromator Input
Pulse Mode Accuracy
Excluding multiple order grating
response
Char., Chop mode not available
on the 86146B model
-60 dB
-40 dB
Notes
Web enabled controls
IEEE-488.1, IEEE-488-2 (100%)
LAN, GPIB, Parallel Printer Port, External VGA
Monitor, Keyboard and Mouse (PS/2)
3,5” 1.44 MB, MS-DOS
Spreadsheet and Word Processor Compatible
(CSV ASCII)
CGM, PCL, GIF
Universal Instrument Drivers (PNP),
Compatible with VEE, Labview, Visual Basic
and C++
The 86146B specifications are for the 50 mm internal path only.
Char. Indicates the number is a characteristic.
T(#) indicates temperature dependence.
* With applied input fiber 9/125 mm
Page 4 of 10
Char.
Char. (86140B, 86143B, 86146B)
MS-DOS is a U.S. registered
trademark of Microsoft
Corporation
Labview is a U.S. registered
trademark of National
Instruments.
General Specifications
Dimensions
Weight
Environmental
Temperature
Humidity
EMI
Benchtop OSA Agilent
86140B, 86142B, 86146B
222 high x 425 wide x 427 mm long
16.5 Kg
Portable OSA
Agilent 86143B, 86145B
163 high x 325 wide x 427 mm long
14.5 Kg
Operating 0°C to 55°C, Storage –40°C to 70°C
Operating <95% RH, Storage: Noncondensing
Conducted and radiated interference is in compliance with CISPR pub 11,
IEC 801-3, IEC 801-4 and IEC 555-2
Power Requirements
Voltage and frequency
Maximum power consumption
90 Vac to 260 Vac, 44 to 444 Hz
230W
Additional Specifications: Agilent 86146B
Insertion Loss Stability** (For 0.1 nm filter bandwidth and greater)
1550 nm,15 minutes
0.5 dB
Insertion Loss***(For 0.1 nm filter bandwidth and greater)
1550 nm
10 dB max
Filter Bandwidth: (From 1530 nm - 1610 nm)
0.5 dB*
RBW Nominal Setting
0.04 nm
0.05 nm
0.07 nm
0.1 nm
0.2 nm
0.3 nm
0.5 nm
0.016
0.019
0.033
0.076
0.134
0.257
0.421
1.0 dB*
Actual Bandwidth
0.023
0.026
0.044
0.089
0.147
0.270
0.434
±20%
3.0 dB*
0.039
0.045
0.063
0.115
0.173
0.297
0.460
Filter Bandwidth: Adjacent Channel Rejection* (at 1550 nm)
0.04 nm
0.05 nm
0.07 nm
0.1 nm
0.2 nm
0.3 nm
0.5 nm
12.5 GHz
±0.1 nm
40 dB
40 dB
N/A
N/A
N/A
N/A
N/A
25 GHz
±0.2 nm
50 dB
50 dB
50 dB
40 dB
40 dB
N/A
N/A
50 GHz
±0.4 nm
55 dB
55 dB
55 dB
50 dB
45 dB
45 dB
45 dB
100 GHz
±0.8 nm
55 dB
55 dB
55 dB
55 dB
55 dB
55 dB
50 dB
Filter Bandwidth: Polarization Dependence (for 0.2 nm filter bandwidth and greater)
1550 nm***
±0.2 dB
* Characteristic value
** Immediately following enhanced single point auto align, at constant temperature
***At room temperature
All data applies across 0-55 degrees C operating range unless otherwise noted.
After warmup period of 2 hrs
Adjacent Channel Rejection limited to 60 dB below total integrated power
Page 5 of 10
Options and Accessories
Options (available on new
instruments only)
Benchtop OSA
Agilent 86140B, 86142B, 86146B
Portable OSA
Agilent 86143B, 86145B
8614xB-001
8614xB-002
8614xB-004
¾
¾
¾
8614xB-006
Included
Included
Included
Included
8614xB-006
Included
Included
Included
Included
Current Source *
White Light Source *
Built-in 1310 & 1550 nm EELED
Source
Wavelength Calibrator
DWDM Spectral Analysis Application
Passive Component Test Application
Amplifier Test Application
Source Test Application
Connector Interface
FC/PC: 81000PI
SC/PC: 81000KI
DIN: 81000SI
ST: 81000VI
Certificate of Calibration
Included
Included
OSA Fiber Sizes
Model
Number
86143B
86145B
86140B
86142B
86146B
Optical
Input
8614xB-002*
(White Light
Source)
8614xB-004*
(1310/1550
EELED)
8614xB-006
(Calibrator)
Photodiode
Input
Mono
Output 1
N/A
9 mm
62.5 mm
9 mm
* 8614xB-002 and 004 are exclusive.
Page 6 of 10
N/A
9 mm
50 mm
9 mm
Options and Accessories: Specifications
8614xB-001 Current Source
Range
Resolution (char)
Accuracy
Clamp Voltage (nominal)
Noise Density at 1 kHz (char)
Stability Within 30 Minutes (char)
Temperature Drift (char)
Pulse Mode
Pulse Range
Pulse Resolution
Duty Cycle Range
Benchtop OSA
Agilent 86140B, 86142B, 86146B
0 to ±200 mA (source or sink)
50 mA steps
2% ±50 mA
±2.7 V
<4 nA/ÖHz
<100 ppm ±500 nA
<(100 ppm ±500 nA)/°C
10 ms to 6.5 ms
100 ns
Pulse width/1 s to 100%
8614xB-002 White Light Source
Wavelength
Minimum Output Power Spectral
Density
(9/125 mm fiber)
900 to 1600 nm
900 to 1600 nm (typical)
1600 to 1700 nm
Minimum Output Power Spectral
Density (char)
50/125 mm fiber
62.5/125 mm fiber
Output Stability (characteristic)
Lamp Lifetime, Mean Time Between
Failures
(MTBF) (char)
900 nm to 1700 nm
-67 dBm/nm (0.2 nW/nm)
-64 dBm/nm (0.4 nW/nm)
-70 dBm/nm (0.1 nW/nm)
-50 dBm/nm (10 nW/nm)
-46 dBm/nm (25 nW/nm)
±0.02 dB over 10 minutes
>5000 hours
8614xB-004 EELED Sources
Minimum Spectral Power Density
1300 to 1320 nm, 1540 to 1560 nm
1250 to 1620 nm
Return Loss
With straight connector
Stability (ambient temp. <±1°C)
Over 15 minutes
Over 6 hours
>-40 dBm/nm (10nW/nm)
>-60 dBm/nm (1nW/nm)
>25dB
<±0.02 dB
<±0.05 dB
Page 7 of 10
8614xB-006 Wavelength Calibrator
The wavelength calibrator option provides an onboard wavelength reference that can be used to automatically
calibrate the optical spectrum analyzer. The calibrator is based on an EELED and an Acetylene gas absorption
cell, Figure 1. The Acetylene absorbs light at very specific wavelengths based on the molecular properties of gas.
The cell is illuminated by an EELED and the OSA uses the absorption pits to perform a wavelength calibration,
Figure 2. Since the absorption of the Acetylene gas is a physical constant it never needs calibrating.
The wavelength calibrator enhances the OSA to achieve better than ±10 pm wavelength accuracy and removes
the
need to use a tunable laser source and multi-wavelength meter as an external reference.
Additional Parts and
Accessories
Printer Paper (5 rolls / box)
Additional Connector Interfaces
9 µm Single Mode Connector Saver
Rack-mount Flange Kit
Transit Case
Soft Carrying Case
BenchLink Lightwave Software*
Benchtop OSA
Agilent 86140B, 86142B, 86146B
Portable OSA
Agilent 86143B, 86145B
9270-1370
See Agilent 81000 series
Standard
8614xB-AX4
9211-2657
N/A
Standard
9270-1370
See Agilent 81000 series
Standard
N/A
9211-5604
8614xB-042
Standard
* Agilent N1031A BenchLink Lightwave allows transfer of measurement results over a GPIB Interface to a PC for the purposes of archiving, printing and
further anlysis
Page 8 of 10
Definition of Terms
Wavelength
·
Absolute Accuracy (after user cal) refers to the wavelength accuracy after the user has performed the
internal wavelength calibration using a source of known wavelength.
·
Reproducibility refers to the amount of wavelength drift which can occur over the specified time while the
OSA is swept across a source of known wavelength.
·
Tuning Repeatability refers to the wavelength accuracy of returning to a wavelength after having tuned to a
different wavelength.
Resolution
·
FWHM refers to the Full-Width-Half-Maximum resolutions that are available. This indicates the width at half
powerlevel of the signal after passing through the resolution slits.
Amplitude
·
Scale Fidelity refers to the potential errors in amplitude readout at amplitudes other than at the calibration
point. This specification is sometimes called linearity.
·
Flatness defines a floating band which describes the error in signal amplitude over the indicated wavelength
range.
(This error may be removed at a given wavelength by performing the user amplitude calibration.)
·
Polarization Dependence refers to the amplitude change that can be seen by varying the polarization of the
light entering the OSA. This is not to be confused with amplitude variations caused by the varying
distribution
of energy between the different modes in fiber that are multimode at the wavelength of interest.
Sensitivity
·
Sensitivity is defined as the signal level that is equal to six times the RMS value of the noise. Displayed
sensitivity values are nominal. Slightly lower values may have to be entered to achieve specified sensitivity.
Dynamic Range
·
Dynamic Range is a measure of the ability to see low-level signals that are located very close (in
wavelength) to
a stronger signal. In electrical spectrum analyzers, this characteristic is generally called shape factor.
Sweep Time
·
Maximum Sweep Rate refers to the maximum rate that the instrument is able to acquire data and display it.
Thisrate may be limited by multiple internal processes when using default number of trace points.
·
Sweep Cycle Time refers to the time required to make a complete sweep and prepare for the next sweep. It
can be measured as the time from the start of one sweep to the start of the next sweep.
Page 9 of 10
Agilent Technologies’ Test and Measurement Support, Services, and Assistance
Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and
measurement capabilities you paid for and obtain the support you need. Our extensive support resources and services can help you choose the right
Agilent products for your applications and apply them successfully. Every instrument and system we sell has a global warranty. Support is available for at
least five years beyond the production life of the product. Two concepts underlie Agilent's overall support policy: "Our Promise" and "Your Advantage."
Our Promise
Our Promise means your Agilent test and measurement equipment will meet its advertised performance and functionality. When you are choosing new
equipment, we will help you with product information, including realistic performance specifications and practical recommendations from experienced test
engineers. When you use Agilent equipment, we can verify that it works properly, help with product operation, and provide basic measurement assistance
for the use of specified capabilities, at no extra cost upon request. Many self-help tools are available.
Your Advantage
Your Advantage means that Agilent offers a wide range of additional expert test and measurement services, which you can purchase according to your
unique technical and business needs. Solve problems efficiently and gain a competitive edge by contracting with us for calibration, extra-cost upgrades,
out-of-warranty repairs, and on-site education and training, as well as design, system integration, project management, and other professional engineering
services. Experienced Agilent engineers and technicians worldwide can help you maximize your productivity, optimize the return on investment of your
Agilent instruments and systems, and obtain dependable measurement accuracy for the life of those products.
By internet, phone, or fax, get assistance with all your test & measurement needs
Online assistance:
For related literature, please visit:
www.agilent.com/comms/lightwave
www.agilent.com/comms/osa
Phone or Fax
United States:
(tel) 1 800 452 4844
Canada:
(tel) 1 877 894 4414
(fax) (905) 282-4120
Europe:
(tel) (31 20) 547 2323
(fax) (31 20) 547 2390
Japan:
(tel) (81) 426 56 7832
(fax) (81) 426 56 7840
Latin America:
(tel) (305) 269 7500
(fax) (305) 269 7599
Australia:
(tel) 1 800 629 485
(fax) (61 3) 9210 5947
New Zealand:
(tel) 0 800 738 378
(fax) 64 4 495 8950
Asia Pacific:
(tel) (852) 3197 7777
(fax) (852) 2506 9284
Product specifications and descriptions in this document subject to change without notice.