User’s and Service Guide Supplement

Agilent Technologies

E8247C and E8257C/D Option H30

User’s and Service Guide Supplement

Use this manual with the following documents:

PSG Family User’s Guides:

E8251-90253

E8251-90259

E8251-90353

E8251-90359

Manufacturing Part Number: E8251-90072

Printed in USA

November 2005

© Copyright 2003, 2005 Agilent Technologies, Inc. All rights reserved.

Warranty Statement

THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT

TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE

MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL

WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND

ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE

IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR

PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL

OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR

PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN.

SHOULD AGILENT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH

WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT

WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL

CONTROL.

DFARS/Restricted Rights Notice

If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR

2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of

Software is subject to Agilent Technologies’ standard commercial license terms and non-DOD Departments and Agencies of the U.S. Government will receive no greater than

Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or

DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.

ii

Safety Notes

The following safety notes are used throughout this document. Familiarize yourself with each of these notes and its meaning before performing any of the procedures in this document.

WARNING Warning denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning note until the indicated conditions are fully understood and met.

CAUTION Caution denotes a hazard. It calls attention to a procedure that, if not correctly performed or adhered to, could result in damage to or destruction of the instrument. Do not proceed beyond a caution sign until the indicated conditions are fully understood and met.

Definitions

Specifications describe the performance of parameters covered by the product warranty

(temperature –0 to 55

°C, unless otherwise noted.)

Typical describes additional product performance information that is not covered by the product warranty. It is performance beyond specification that 80% of the units exhibit with a 95% confidence level over the temperature range 20 to 30

°C. Typical performance does not include measurement uncertainty.

Nominal values indicate expected performance or describe product performance that is useful in the application of the product, but is not covered by the product warranty.

iii

iv

Contents

E8247C and E8257C\D Option H30

Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

Nominal Performance Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

Operation and Performance Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Documentation Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Option H30 Operation (520 or 540) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Performance Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

MID Board or Coaxial Switch. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

Upconverter Mixer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

Repair Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

Safety and Regulatory Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

Connector Care and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

Before Applying Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

Shipping Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

Compliance with Canadian EMC Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

Compliance with German Noise Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

Statement of Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

Cautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

Instrument Markings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

Contacting Agilent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

Contents-1

Contents

Contents-2

E8247C and E8257C\D Option H30

User’s and Service Guide Supplement 1

Description

The Option H30 adds a mixer and coax switch to enable the E8247C or the E8257C/D to upconvert a RF signal (input is provided on the rear panel) to a higher frequency which will be the sum and difference of the RF input signal and frequency set on the E8247C or

E8257C/D.

The coaxial switch allows the E8247C or E8257C/D with Option H30 to be switched from standard synthesizer mode to upconverter mode. This allows the E8247C to be used as a standard source instrument, as well as an upconverter. The synthesizer and upconverter modes can be selected on the front panel or by remote command. The upconverted output signal is the same as the front panel RF output. The input connector is a SMA type female that is located on the rear panel.

The Option H30 can be ordered with Option 520 or Option 540. For instruments that include Option 520, signals can be upconverted from 200 kHz to 26 GHz. For instruments that include Option 540, signals can be upconverted from 6 GHz to 46 GHz.

2 User’s and Service Guide Supplement

Nominal Performance Characteristics

Nominal values indicate expected performance or describe product performance that is useful in the application of the product, but is not covered by the product warranty.

Table 1 Option H30 with Option 520 (20 GHz)

Description

Rear panel frequency range

Conversion loss for upconverted signal to the front panel

Front panel RF upconverted frequency range

Usable IF Bandwidth (dependent on the

IF frequency used on rear panel)

Frequency

100 kHz to 6 GHz

–10 to –15 dBm

200 kHz to 26 GHz

< 2 dB over 1 GHz span ≤ 3 GHz IF Frequency

< 20 dB over 1 GHz span > 3 GHz IF Frequency

Table 2 Option H30 with Option 540 (40 GHz)

Description

Rear panel frequency range

Conversion loss for upconverted signal to the front panel

Front panel RF upconverted frequency range

Usable IF Bandwidth (depends on the

IF frequency used on rear panel)

1

Frequency

100 kHz to 6 GHz

–10 dB (6 GHz) to –28 dB (46 GHz)

6 to 46 GHz

< 2 dB over 1 GHz span ≤ 3 GHz IF Frequency

< 20 dB over 1 GHz span > 3 GHz IF Frequency

1. Conversion loss and IF bandwidth flatness will degrade contingent on the IF Input frequency and the upconverted frequency. Refer to Figure 1 and Figure 2 on page 4 .

NOTE When in upconverter mode, the displayed RF output power level is LO drive.

The displayed RF power in upconverter mode does not represent the power of the upconverted signal at the RF output.

The maximum RF power output of the E8247C and E825C Option H30 is reduced by

2 dBm due to the added insertion loss of the coax transfer switch required for the design.

For units with Option H30, subtract 2 dBm from the power output specifications listed in

Table 1 of the standard user’s guides E8251-90253 and E8251-90353.

User’s and Service Guide Supplement 3

Figure 1 Upconverted RF Flatness

20

15

10

5

0

0

0.

4

0.

8 1

1.

4

1.

6 2

2.

4

2.

6 3

3.

2

3.

6

3.

8 4

4.

4

4.

8 5

5.

2

5.

6

5.

8 6

IF Input frequency vs. flatness over

1 GHz (dB) for fundimental mixing

Figure 2 Upconverted RF Conversion Loss

30

25

20

15

10

5

50

45

40

35

3 6 9 12 15 18 21 24 27 30 33 36 39 41 44 46

Conversion Loss from IF input power level (dB) for fundimental mixing

4 User’s and Service Guide Supplement

Operation and Performance Verification

Documentation Conventions

The following key conventions are used throughout this document.

• [HARDKEYS] are labeled front panel keys.

• SOFTKEYS are unlabeled keys whose function is indicated on the instrument display.

Option H30 Operation (520 or 540)

To use the E8247C and E8257C/D Option H30 as a standard instrument, refer to the standard User’s Guide.

To use the E8247C and E8257C/D Option H30 as an RF upconverter, refer to the following procedure:

CAUTION Maximum power into the rear panel RF upconverter input must not exceed

+20 dBm.

CAUTION Damage to upconverter mixer can result if the reverse power into the RF output port on the front panel occurs while in upconverter mode.

NOTE Refer to “Nominal Performance Characteristics” on page 3 for performance considerations.

Step 1. Remove the protective connector from the RF up converter Input on the rear panel.

Step 2. Set the RF source (customer supplied) to within the recommended power and frequency ranges.

Step 3. Connect the RF source to the H30 rear panel RF upconverter input using a suitable RF cable and adapters as necessary. It is recommended that the user set the power level of the E8247C or E8257C/D Option H30 to +10 dBm for best performance.

To activate upconverter mode from the front panel, press:

[FREQUENCY] > MORE, MORE > UPC ON/OFF (to activate mode)

Use the same front panel key to return the instrument to normal synthesizer mode.

The command UPC ON can be used to remotely select the upconverter mode. Use the command UPC OFF to return the instrument to normal synthesizer mode.

User’s and Service Guide Supplement 5

Performance Verification

The E8247C and E8257C/D Option H30 functionality can be verified using the following equipment and methods:

The E8247C and E8257C/D Option H30 must first meet standard product specifications with the exception that the power output may be degraded by up to 2.0 dBm. (Subtract 2.0 dBm from the Option 1E1 output power specification).

Refer to the Table 3 on page 6 for a list of equipment needed in addition to the E8247C and

E8257C/D Option H30 being tested:

Table 3 Equipment Required

Agilent Instrument Agilent

Part Number

50 GHz Spectrum Analyzer

6 GHz RF Signal Source

RF Cable (SMA male 3 ft)

RF Cable (2.4 mm 3 ft)

8565EC

E4438C or equivalent

5061-6669

5063-9820

SMA Female to Type-N Adapter 1250-1250

2.4 mm to 2.4 mm Female Adapter 11900B (or comparable)

Step 1. Set the RF Source power level to –10 dBm and frequency to 6 GHz.

Step 2. Connect the RF Source to the E8247C and E8257C/D Option H30 rear panel

RF Upconverter Input, using an RF cable and adapters.

Press: [FREQUENCY] > MORE, MORE > UPC ON/OFF (to activate mode)

An audible click should be heard from the E8247C and E8257C/D Option H30 as the transfer switch switches the upconverter mixer LO to the coupler output and the mixer RF out to the 1E1 step attenuator’s input.

Step 3. For instruments with Option 540 set the instrument to upconverter mode, set power level to +8 dBm and CW frequency to 40 GHz. Verify that a 46 GHz signal appears on the spectrum analyzer when in upconverter mode and is not present when in standard synthesizer mode. The level of the upconverter IF signal

(6 GHz + 40 GHz = 46 GHz) is nominally –28 dBm.

or

For instruments with Option 520 set the instrument to the upconverter mode, set power level to +8 dBm and CW frequency to 20 GHz. Verify on the spectrum analyzer that a 26 GHz signal appears when in upconverter mode and is not present when in standard synthesizer mode. The level of the upconverter IF signal (6 GHz + 20 GHz = 26 GHz) is nominally –28 dBm. Since this is a functional test only, measurement uncertainty is not taken into account. This verifies the switch path and mixer are functioning.

6 User’s and Service Guide Supplement

Troubleshooting

If there is no signal a problem has occurred. First check the equipment setup, verifying outputs on the sources and connections. Refer to Figure 4 through Figure 7 .

Verify that the E8247C and E8257C/D Option H30 Upconverter switch is working:

MID Board or Coaxial Switch

The following procedure will aid in determining whether the MID board (E8251-60089) or the coax switch (87222E) is functional.

Step 1. Select upconverter mode on/off, then listen for the coax switch to click. If it doesn’t click, check the ribbon cable connection to the switch (under switch/mixer assembly). Remove the switch end of the cable from the switch and verify: pin 1 for +32 Vdc, pin 7 for +5 Vdc or 0 Vdc pin 8 for +5.2 Vdc pin 9 for Ground

If the voltages are incorrect, verify the ribbon cable connection to the MID board assembly. The ribbon cable should be connected to J14.

Step 2. Cycle the upconverter mode on/off repeatedly while looking at pin 7 voltage. Pin 7 should toggle from +5 Vdc (upconverter mode on) to 0 Vdc (upconverter mode off).

If pin 7 voltage does not toggle, either the MID board (E8251-60089) or a configuration error is at fault. Consult with Agilent Service for repair. Refer to

“Contacting Agilent” on page 17 .

Step 3. If the voltages and control signal are present at the ribbon cable as verified in step

1, reconnect the ribbon cable to the coax switch and select upconverter mode and toggle it ON/OFF.

To verify the switch is not ultra quiet, measure the ports on the coax switch while toggling the upconverter mode.

(Coupler out) PORT 1 PORT 2 (mixer LO input)

(Atten. input) PORT 4 PORT 3 (mixer RF output)

Port 4 should connect to port 3 in upconverter On mode and port 1 in upconverter

Off mode. If it doesn’t, replace the 87222E coax switch.

User’s and Service Guide Supplement 7

Upconverter Mixer

The following procedure will aid in determining if the upconverter mixer is functional:

CAUTION Possible damage may occur if RF upconverter input levels exceed 20 dBm or if reverse power is applied at the RF output connector. First verify all cable configurations are correct before applying a signal to the instrument.

The RF step attenuator cable to the coax switch is functional if the instrument operates as a standard synthesizer.

The RF cable from the coupler output is also functional if the instrument operates as a standard synthesizer.

Step 1. Verify that the RF cable is connected from the connector labeled RF upconverter on the rear panel to the mixer’s IF input port. Verify the power at the connection to the mixer with RF Source applied.

Step 2. Verify the power through the cables to LO Input of the mixer (bottom connector). The power should be close to the power level coming from the

Coupler Output. Replace the cables if power insertion loss is excessive and damage is suspected. These cables have a K-type connector and a 2.92 mm adapter should be used for mating to connectors.

Step 3. If all of the cables have been verified and are functional, verify the mixer orientation with the markings on the outside of the mixer LO down, RF

Connector up and IF Connector facing rear of instrument. Replace the mixer and verify its functionality by following the verification procedure.

8 User’s and Service Guide Supplement

Repair Information

Figure 4 Top View

W105

W102 W104 W101

Figure 5 Semi-rigid Cables (front view)

W102 MP101

A9 MID Board W103

A100 Mixer

W103

W105 SW1 W104

User’s and Service Guide Supplement

W101

9

Figure 6 Ribbon Cable

A100

MP101

W106

A9

Figure 7 Bracket (rear panel)

Washer Nut MP108

MP107

10 User’s and Service Guide Supplement

Replaceable Parts

Reference

Designator

W104

W105

W105

W106

A9

MP101

MP102

A100

A100

SW1

W101

W101

W102

W103

W104

MP103

MP104

MP105

MP107

MP108

MP109

MP104

MP105

Description

Mixer 6 GHz to 46 GHz (Opt H30 with Opt 540)

Mixer 6 GHz to 26 GHz (Opt H30 with Opt 520)

Coax transfer switch, 4 port, 50

Ω, 50 GHz, 87222E

RF Cable Coupler- SW (Opt H30 with Opt 540)

RF Cable Coupler- Switch (Opt H30 with Opt 520)

RF Cable Attenuator - SW

RF Cable Mixer IF - Rear Panel

RF Cable Mixer LO - SW (Opt H30 with Opt 540)

RF Cable Mixer LO - SW (Opt H30 with Opt 520)

RF Cable Mixer R - SW (Opt H30 with Opt 540)

RF Cable Mixer R - SW (Opt H30 with Opt 520)

Ribbon Cable Conn. to Coax Switch

MID board (J14 mixer-switch output)

Mixer switch bracket (mounts to RF attenuator)

Torx 3.0 mm with self lock for mounting switch to bracket

Torx 3.0 mm to mount bracket to RF step attenuator (1E1)

00-90 inch screw to mount mixer (540)

Hex nut for 00-90 inch screw (540)

Rear Pnl connector mounting plate

SMA Panel Connector (RF Upconverter Input)

Torx screw (3.0 mm x 6 mm to mount conn/bracket)

Metric Screw (2.5 mm x 8 mm) (520)

Metric 2.5 hex nut with lock washer (520)

User’s and Service Supplement

Chain Dust Cap (SMA female)

0515-1035

3030-0436

0608-0004

E8251-00093

1250-1753

0515-0372

0515-2007

0535-0123

E8251-90072

1250-2759

Agilent

Part Number

0955-1447

0955-0488

87222-60015

E8251-20082

E8251-20173

E8251-20083

E8251-20084

E8251-20085

E8251-20172

E8251-20086

E8251-20171

8121-6187

E8251-60089

E8251-00092

0515-0376

Qty

1

2

2

1

1

1

1

1

1

1

1

1

1

1

1

2

2

1

2

2

1

1

1

2

2

User’s and Service Guide Supplement 11

Safety and Regulatory Information

Introduction

Review this product and related documentation to familiarize yourself with safety markings and instructions before you operate the instrument. The documentation contains information and warnings that must be followed by the user to ensure safe operation and to maintain the product in a safe condition.

Connector Care and Cleaning

If alcohol is used to clean the connectors, the power cord to the instrument must be removed. All cleaning should take place in a well ventilated area. Allow adequate time for the fumes to disperse and moist alcohol to evaporate prior to energizing the instrument.

WARNING To prevent electrical shock, disconnect the Agilent Technologies model product from mains before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.

Before Applying Power

Verify that the product is configured to match the available main power source. If this product is to be powered by autotransformer, make sure the common terminal is connected to the neutral (grounded) side of the ac power supply.

Shipping Instructions

You must always call the Agilent Technologies Instrument Support Center to initiate service before retuning your instrument to a service office. See “Contacting Agilent” on page 17 . Always transport or ship the instrument using the original packaging if possible.

If not, comparable packaging must be used. Attach a complete description of the failure symptoms.

Compliance with Canadian EMC Requirements

This ISM device complies with Canadian ICES-001.

Cet appareil ISM est conforme a la norme NMB du Canada.

12 User’s and Service Guide Supplement

Compliance with German Noise Requirements

This is to declare that this instrument is in conformance with the German Regulation on

Noise Declaration for Machines (Laermangabe nach der Maschinenlaermrerordnung-3.

GSGV Deutschland).

Acoustic Noise Emission/Geraeuschemission

LpA<70 dB

Operator Position

Normal Operation per ISO 7779

Lpa<70 dB am Arbeitsplatz normaler Betrieb nach DIN 45635 t. 19

Declaration of Conformity

For a copy of the manufacturer’s Declaration of Conformity for this apparatus, contact your local Agilent Technologies office or sales representative on Page 17 .

Statement of Compliance

This instrument has been designed and tested in accordance with IEC Publication 1010,

Safety Requirements for Electronic Measuring Apparatus, and has been supplied in a safe condition. The instruction documentation contains information and warnings which must be followed by the user to ensure safe operation and to maintain the instrument in a safe condition.

User’s and Service Guide Supplement 13

WARNING

WARNING

WARNING

WARNING

WARNING

WARNING

WARNING

WARNING

WARNING

14

Warnings

The WARNING notice denotes a hazard. It calls attention to a procedure, practice, or the like, which if not correctly performed or adhered to, could result in personal injury. Do not proceed beyond a

WARNING notice until the indicated conditions are fully understood and met.

Warnings applicable to this instrument are:

If this instrument is not used as specified, the protection provided by the equipment could be impaired. This instrument must be used in a normal condition (in which all means for protection are intact) only.

For continued protection against fire hazard replace line fuse only with same type and rating:

• United States—F 3A/250V, Part Number 2110-0780

• Europe—F 3.15A/250V, Part Number 2110-0655

The use of other fuses or material is prohibited.

This is a Safety Class I product (provided with a protective earthing ground incorporated in the power cord). The mains plug shall be inserted only into a socket outlet provided with a protective earth contact. Any interruption of the protective conductor, inside or outside the instrument, is likely to make the instrument dangerous.

Intentional interruption is prohibited.

The power cord is connected to internal capacitors that may retain dangerous electrical charges for 5 seconds after disconnecting the plug from its power supply.

These servicing instructions are for use by qualified personnel only.

To avoid electrical shock, do not perform any servicing unless you are qualified to do so.

The opening of covers or removal of parts is likely to expose dangerous voltages. Disconnect the instrument from all voltage sources while it is being opened.

This product is designed for use in Installation Category II and

Pollution Degree 2 per IEC 1010 and 664 respectively.

No operator serviceable parts inside. Refer servicing to qualified personnel. To prevent electrical shock do not remove covers.

User’s and Service Guide Supplement

CAUTION

CAUTION

CAUTION

CAUTION

Cautions

The CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like, which if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.

Cautions applicable to this instrument are:

Always use the three-prong ac power cord supplied with this instrument. Failure to ensure adequate earth grounding (by not using this cord) can cause instrument damage.

This instrument has autoranging line voltage input; be sure the supply voltage is within the specified range.

Ventilation Requirements: When installing the instrument in a cabinet, the convection into and out of the instrument must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the instrument by 4

°C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, forced convection must be used.

User’s and Service Guide Supplement 15

Instrument Markings

!

When you see this symbol on your instrument, you should refer to the instrument’s instruction manual for important information.

This symbol indicates hazardous voltages.

The laser radiation symbol is marked on products that have a laser output.

This symbol indicates that the instrument requires alternating current (ac) input.

The CE mark is a registered trademark of the European

Community. If it is accompanied by a year, it indicates the year the design was proven.

The CSA mark is a registered trademark of the Canadian

Standards Association.

This symbol indicates the product meets the Australian

Standards.

This symbol indicates separate collection for electrical and electronic equipment, mandated under EU law as of August 13,

2005. All electric and electronic equipment are required to be separated from normal waste for disposal (Reference WEEE

Directive, 2002/96/EC).

This text indicates that the instrument is an Industrial Scientific and Medical Group 1 Class A product (CISPR 11, Clause 4).

This symbol indicates that the power line switch is ON.

This symbol indicates that the power line switch is OFF or in

STANDBY position.

Safety Earth Ground. This is a Safety Class I product (provided with a protective earthing terminal). An uninterruptible safety earth ground must be provided from the main power source to the product input wiring terminals, power cord, or supplied power cord set. Whenever it is likely that the protection has been impaired, the product must be made inoperative and secured against any unintended operation.

16 User’s and Service Guide Supplement

Contacting Agilent

By internet, phone, or fax, get assistance with all your test and measurement needs.

This information supersedes all prior contact information.

Online assistance: www.agilent.com/find/assist

Americas

Brazil

(tel) (+55) 11 3351 7012

(fax) (+55) 11 3351 7024

Canada

(tel) +1 877 894 4414

(alt) +1 303 662 3369

(fax) +1 800 746 4866

Mexico

(tel) 1800 254 2440 Ext 2703

(alt) from USA 18008374039

(fax) 1 800 254 422

United States

(tel) 800 829 4444

(alt) (+1) 303 662 3998

(fax) 800 829 4433

Australia

(tel) 1 800 225 574

(fax) 1 800 681 776

(fax) 1 800 225 539

Japan (Bench)

(tel) 0120 421 345

(alt) (+81) 426 56 7832

(fax) 0120 01 2144

Asia Pacific and Japan

China

(tel) 800 810 0189

(fax) 800 820 2816

Hong Kong

(tel) 800 933 229

(fax) 800 900 701

Japan (On-Site)

(tel) 0120 421 345

(alt) (+81) 426 56 7832

(fax) 0120 012 114

Malaysia

(tel) 1800 880 399

(fax) 1800 801 054

India

(tel) 1600 112 626

(alt) +65 6275 0800

(fax) 1600 113 040

New Zealand

(tel) +64 4 939 0635

(alt) 0800 738 378

(fax) +64 4 972 5364

Singapore

(tel) 1 800 275 0880

(fax) (+65) 6755 1214

South Korea

(tel) 080 778 0011

(fax) 080 778 0013

Taiwan

(tel) 0800 047 669

(fax) 0800 047 667

(fax) +886 3492 0779

Europe

Thailand

(tel) +66 2 267 5913

(tel) 1 800 2758 5822

(fax) 1 800 653 336

Austria

(tel) 0820 87 44 11*

(fax) 0820 87 44 22

France

(tel) 0825 010 700*

(fax) 0825 010 701*

Italy

(tel) (+39) (0)2 9260 8484

(fax) (+39) (0)2 9544 1175

Spain

(tel) (+34) 91 631 3300

(fax) (+34) 91 631 3301

Belgium

(tel) (+32) (0)2 404 9340

(fax) (+32) (0)2 404 9395

Germany

(tel) 01805 24 6333*

(fax) 01805 24 6336*

Luxemburg

(tel) (+32) (0)2 404 9340

(fax) (+32) (0)2 404 9395

Sweden

(tel) 0200 88 22 55*

(alt) (+46) (0)8 5064 8686

(fax) 020 120 2266*

Denmark

(tel) (+45) 7013 1515

(fax) (+45) 7013 1555

Ireland

(tel) (+353) 1 890 924 204

(fax) 1 890 924 024

Netherlands

(tel) (+31) (0)20 547 2111

(fax) (+31) (0)20 547 2190

Switzerland (French)

(tel) 0800 80 5353 opt. 2*

(fax) (0) 22 567 5313

Finland

(tel) (+358) (0) 10 855 2100

(fax) (+358) (0) 10 855 2923

Israel

(tel) (+972) 3 9288 504

(alt) (+972) 3 9288 544

(fax) (+972) 3 9288 520

Russia

(tel) (+7) 095 797 3963

(alt) (+7) 095 797 3900

(fax) (+7) 095 797 3901

Switzerland (German)

(tel) 0800 80 5353 opt. 1*

(fax) 0 44 272 7373

Switzerland (Italian)

(tel) 0800 80 5353 opt. 3*

(fax) (0) 22 567 5314

United Kingdom

(tel) (+44) (0)7004 666666

(fax) (+44) (0)7004 444555

(tel) = primary telephone number; (alt) = alternate telephone number; (fax) = FAX number; * = in country number 8/10/05

User’s and Service Guide Supplement 17

18 User’s and Service Guide Supplement

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Frequently Answers and Questions

What additional capability does the Option H30 offer?
The Option H30 adds upconversion functionality to the E8247C or E8257C/D, enabling it to convert an input RF signal to a higher frequency.
What is the frequency range of the RF input signal?
The rear-panel RF input frequency range is from 100 kHz to 6 GHz for both Option 520 (20 GHz) and Option 540 (40 GHz).
What is the frequency range of the upconverted RF output signal?
With Option 520, the upconverted RF output frequency range is from 200 kHz to 26 GHz. With Option 540, the range is from 6 GHz to 46 GHz.
What is the typical conversion loss for the upconverted signal?
For Option 520, the conversion loss is typically -10 to -15 dBm. For Option 540, it ranges from -10 dB (at 6 GHz) to -28 dB (at 46 GHz).
What is the maximum RF power output of the E8247C and E8257C with Option H30?
The maximum RF power output is reduced by 2 dBm due to the insertion loss of the coax transfer switch required for the design. Subtract 2 dBm from the power output specifications listed in Table 1 of the standard user's guides.
How does Option H30 affect the displayed RF output power level?
In upconverter mode, the displayed RF output power level represents the LO drive, not the power of the upconverted signal at the RF output.