EUTECH INSTRUMENTS | ALPHA COND 550 - REV 5 | User manual | Thermo Scientific Alpha Process Products

Add to my manuals
24 Pages

advertisement

EUTECH INSTRUMENTS | ALPHA COND 550 - REV 5 | User manual | Thermo Scientific Alpha Process Products | Manualzz

Thermo Scientific

Alpha Process Products

Thermo Scientific Alpha 2000 Series Controllers

pH

ORP

Conductivity

Dissolved Oxygen

One Source.

Total Solution

Recognised internationally for industry-leading quality and accuracy, companies all over the world choose

Thermo Scientific Process Products for reliable process monitoring and control across a broad range of water and wastewater applications:

• Wastewater Treatment

• Drinking Water

• Chemical Processing

• F&B Manufacturing

• Seawater Desalination

• Pharmaceutical

• Power

• Electroplating

• Semiconductor

Consistent monitoring and control of water quality are vital in many industries. Thermo Scientific products are built to stand up to the demands of on-line continuous use, even under the most severe conditions. With the Alpha 2000 series, Thermo Scientific brings electrochemical processes in water and wastewater applications to a new level of reliability and versatility, offering flexible process control at an excellent price point.

Whether it’s pH, ORP, Conductivity or

Dissolved Oxygen you are measuring, the Alpha 2000 series delivers, accurately and consistently. Because at Thermo Scientific, reliability and ease-of-use aren’t just features – they’re fundamentals.

Reliability and ease-of-use aren’t just features – they’re fundamentals

2 Thermo Scientific / Alpha Process Products / 2000 Series Controllers / www.thermo.com/process

Thermo Scientific Alpha 2000 Series

Controllers/Transmitters:

The Controllers/Transmitters:

Thermo Scientific’s Alpha 2000

Series controllers combine simple six-button operation with exceptional performance. Reliable and durable, each controller is built for ease-ofuse and versatility, so it is easy to customise your processor according to your application needs. Meters come in ½ DIN wall and panel-mount or ¼ DIN panel-mount versions to suit your installation requirements.

Alpha pH 2000D

employs differential measurement technology, giving you a reliable system that measures pH/ORP accurately in harsh environments

Alpha pH 2000

features symmetrical mode of operation option for clear, accurate pH/ORP readings in electrically noisy environments

• With 4-cell and 2-cell electrode options, the

Alpha COND 2000

allows you to measure a broad conductivity range, from ultra-pure water to highly conductive samples

Alpha DO 2000

supports either galvanic or amperometric measurements, allowing a broad dissolved oxygen range of 0 to 20 ppm or 0 to 200 % saturation

The Electrodes:

Thermo Scientific offers a wide selection of process electrodes, buffers, standards and accessories to complement your process requirements.

• High-quality, double-junction

pH and ORP electrodes

with Kynar ® or

Annular PTFE reference junctions operate in environment from

0 °C up to 110 °C. Each electrode comes with integral low-noise semi-conductor cables (unless otherwise stated)

• 2-cell and 4-cell

Conductivity electrodes

that incorporate 3-wire

Pt100 for automatic temperature compensation. Durable, low maintenance electrodes built with

Titanium or SS316 give consistent performances in high temperature of up to 200 °C

Dissolved Oxygen electrodes

designed for minimal maintenance and quick, stable readings within short response time. Rugged and long-lasting galvanic electrodes require no warm-up time; low maintenance amperometric electrodes capture DO readings as low as 0.01 ppm

Thermo Scientific / Alpha Process Products / 2000 Series Controllers / www.thermo.com/process 3

Main Features:

No Need for Recalibration

After Power-Failure

Built-in non-volatile memory retains calibration and setup information even when power is cut off

Push-Button Calibration

Quick, convenient and easy

Menu-Driven Controls

With intuitive prompts and Electrode offset adjustment from keypad connectors

Advanced Switch Mode

Power Supply (SMPS)

Auto-detection of power supply input from 80 V to

250 VAC/DC

Three Control Mode Options

Limit control mode for straightforward on/off operations, or proportional control mode and proportional integral control mode for processes that require finer controls

Front Panel

MIN Function

100% Yh

50%

Proportional Integral, Proportional or Limit Control, Operation of Relays

MAX Function

Relay tON

Pulse Length T tOFF

ON

Relay

Pulse Length T tON tOFF

ON

- Xw

0%

Xp

Prop.

Band SP 1 SP 2

Xp

Prop.

Band

Control Characteristic of Controller as Proportional Controller

+ Xw

OFF

Controller Signal of

Pulse Length Control

Time (S)

OFF Time (S)

Controller Signal of

Pulse Frequency Control

Three Layers of

Password Protection

Greater security against unauthorized or accidental tampering of data and critical settings

Minimal Maintenance

Required

Simply wipe occasionally with lint-free cloth to remove dust and dirt

Large UV Light-Protected

LCD Display with Backlight

For easy reading, even in dark places

IP66 (NEMA 4X) Rugged

Housing

Weatherproof and corrosion-resistant

Bi-Colour LEDs

Clearly indicates relay status and mode of operation. Visible even from a distance

Manual Override

Facilitates quick processing with direct control on closing pumps and cleaning system

4 Thermo Scientific / Alpha Process Products / 2000 Series Controllers / www.thermo.com/process

Liquid Ground

caters for symmetrical mode of operation to prevent measurement error in an electrically noisy environment

12 V DC Output

To power peripheral devices via controller

Two 4-20 mA or 0-20 mA

Current Outputs

Fully scaleable, galvanically-isolated outputs for parameter and temperature measurements. Out-ofrange current at 22 mA

Automatic Temperature

Compensation

Or manual temperature compensation without

ATC probe. 3-wire system compensates for cables line resistance errors. Independent settings for calibration and process temperatures for accurate temperature compensation

Installation and Wiring is

Easy

with pin terminals detachable blocks

Back Panel

Rated to IP66

(Nema 4X), the rugged weather-proof casing simplifies installation outdoors or indoors

Pipe Mount Wall Mount Panel Mount

Thermo Scientific / Alpha Process Products / 2000 Series Controllers / www.thermo.com/process 5

Four Fully Programmable Relays

Two Independent Relays

allow combination of high and low settings

Wash Relay

for periodical, automated cleansing of sensors – essential for accurate measurements

Single-Pole-Double-Throw

(SPDT) Alarm Relay

alerts when readings fall outside set points.

Comes with usercustomisable time-delay for minimal false alarms

Hold Function

to tie in a float or flow switch or other control device to lock out control functions if chemical feed levels are low

Prevents “Chattering”

Separately adjustable high and low hystereses prevent relay from switching between on and off during rapid fluctuations around set-points

Multiple Mounting Options

Wall-mount, pipe-mount or panel-mount: Flexible mounting options to suit your environment

Thermo Scientific Alpha pH 2000D

Differential pH/ORP Controller/Transmitter:

pH/ORP:

The Alpha pH 2000D Controller/

Transmitter and combination electrode system employs differential pH measurement technology.

What’s unique about this system is the differential electrode, which uses a pH glass bulb in place of the conventional reference system, making it more resistant to contamination or fouling in harsh wastewater applications. Replaceable salt bridge and reference solution add lifespan to the electrode when junction clogs.

The result: a reliable, long-lasting and cost-saving pH measurement system for demanding environments.

The pH Differential Electrode:

• Cost Savings, Reduced Downtime and Maintenance Costs –

soiled salt bridge and buffer reference solution can easily be replaced at a fraction of the cost of a new combination electrode

• Automatic Temperature Compensation

built-in temperature electrode

• Use Any Buffer Standard from 2 to 10 pH as a Reference –

very useful for applications that require adjustable isopotential points

• Built-In Electronic Amplifier

provides a low impedance output from the electrode, and allows analyzers to be located up to 3000 feet away from the electrode

• Ground Rod Minimizes Electrical

Interference

from motors, pumps, transformers and ground currents

• Wide pH range of -2.00 to 16.00 pH at two-decimal accuracy

• ORP mode measures in mV or % concentration, with independent calibration modes

• Control and measure pH and

ORP simultaneously with master-slave operation by placing two controllers side by side

• Wall mount, pipe mount or panel mount options. Compact design of electrode with universal 1.5”

NPT mounting threads enables optimal mounting positions

Easy replacement of salt bridge

Active (process) pH/ORP electrode

Ground rod shields active pH/ORP electrode from electrical noise

Reference electrode

Viton ® reference o-ring offers excellent resistance to aggressive chemicals

Universal 1.5” NPT threads for optimal mounting

Short 6” body with six wrench flats for easy assembly

Smooth surfaces reduce chances of fouling

Larger refillable internal reference chamber minimises effects of dilution and extends electrode life

Replaceable quad junction salt bridge

Better protection of sensing electrode

6 Thermo Scientific / Alpha Process Products / 2000 Series Controllers / www.thermo.com/process

About Differential pH Measurement:

The Conventional

Combination Electrode:

The conventional combination electrode consists of two parts: the glass electrode and the reference

(Figure 1)

.

The glass electrode contains a silver wire immersed in conductive solution. During measurement, the glass bulb and stem are the only parts which come in contact with the sample. Hence, contamination of the silver wire and conductive solution does not occur.

The reference electrode, on the other hand, is constructed simply with a porous junction and a silver wire. This porous junction allows the reference electrolyte to flow from its chamber to the sample, establishing a connection between the two. However, the porosity also allows contaminants to flow into the reference chamber, reacting with the electrolyte and/or silver wire.

The reference electrode is also susceptible to clogs. In the case of conventional combination electrodes, the entire electrode must be replaced due to a clogged reference electrode.

When contamination and clogging occurs, performance of the entire electrode is compromised.

The Differential Electrode:

The differential electrode replaces the exposed silver wire with a design built on the benefits of a pH glass electrode

(Figure 2)

.

Quadraple junction salt bridges and an electrolyte-filled chamber act as barriers to prevent contaminants from entering the inner chamber, while at the same time, allowing connection between the reference electrolyte and the sample.

In the event of clogs, the salt bridges and electrolyte are easily replaceable, reducing downtime and cost of maintenance.

An individual liquid ground is also built into the electrode, effectively eliminating noises from ground loops potential

(Figure 3)

:

(E1 – E3) – (E2 – E3) = E1 – E3

With these design improvements, the differential electrodes enjoy a longer lifespan in harsh wastewater environments as compared to conventional electrodes.

Measurement of pH in wastewater poses several challenges, one of which is fouling of the reference electrode system

Pre-Amplifier

pH

Glass

Electrode

(E1) pH

Reference

Electrode

(E2)

Figure 1: The conventional combination electrode is made up of a glass electrode and a reference electrode

Figure 2: Differential electrode replaces porous reference electrode with a glass electrode

Reference

Electrode

(E2)

Solution

Ground

(E3)

Process

Electrode

(E1)

Figure 3: Individual liquid ground loop eliminates noises from ground loop potentials

Thermo Scientific / Alpha Process Products / 2000 Series Controllers / www.thermo.com/process 7

Thermo Scientific Alpha pH 2000 pH/ORP Controller/Transmitter:

pH/ORP:

Built for the rugged industrial environment, the Alpha pH 2000 transmitter comes with an option for symmetrical mode of operation, making it ideal for electrically noisy environments. IP65 NEMA 4X casing is weatherproof and corrosionresistant, protecting meter against harsh elements.

The Alpha pH 2000 is user-friendly and intuitive: controller alerts when meter or electrode requires maintenance, and provides troubleshooting prompts in plain language.

• Wide pH range of -2.00 to 16.00 pH at two-decimal accuracy

• ORP mode measures in mV or % concentration, with independent calibration modes

• Quick, easy push-button calibration with auto-buffer recognition.

Meter displays electrode status after each calibration

• Accepts Antimony electrodes – useful in applications involving

Hydrofluoric (HF) Acid

• Control and measure pH and

ORP simultaneously with master-slave operation by placing two controllers side by side

Asymmetrical

PE N L S

3

0 / PA

2 1 22 27

Temp.

pH

BNC

Measuring Cable pH Combination

Electrode

Symmetrical

PE N L S

3 2 1 22

0 / PA

27

Temp.

Potential

Matching Pin (PM) pH

BNC

Measuring Cable pH Combination

Electrode

Asymmetrical mode works well in environments with where there is little or no electrical noise. When there is electrical noise, the noise acts as a common signal and is picked up by both the pH and the reference electrodes. However, since the reference electrode is grounded to the ground potential of the amplifier, electrical noise will be present only on the pH electrode.

This noise is amplified along with the pH signal, which causes reading fluctuations in an electrically noisy atmosphere. Electrical noise from a DC source (as in an electroplating tank) will typically result in stable but incorrect values.

Symmetrical mode avoids grounding the reference electrode by reconfiguring the input to a floating differential mode. The electrical noise appears equally on the pH and reference electrodes, and is therefore rejected by the operational amplifier.

The Alpha pH 2000 transmitter offers an option for Symmetrical Mode of Operation. To take advantage of

Symmetrical operation, you must have an electrode with a solution ground (potential matching) pin.

If your electrode does not have a solution ground, be sure to set the controller to Asymmetrical mode.

8 Thermo Scientific / Alpha Process Products / 2000 Series Controllers / www.thermo.com/process

Specification Information pH/ORP Controller/Transmitter

Order Code

Part No.

pH:

Range:

Resolution:

Accuracy:

ORP:

Range:

Resolution:

Accuracy:

Temperature:

Range:

Resolution:

Accuracy:

Sensor:

Compensation:

Set point & controller functions:

Set point 1 (SP1) / set point 2 (SP2):

Switching pH hysteresis:

Switching ORP hysteresis:

Function (switchable):

Adjustable period with pulse length controller:

Adjustable period with pulse frequency controller:

Integral action time (IAT):

Pickup/dropout delay:

Wash cycle:

Wash duration:

Contact outputs:

Switching voltage/current/power:

Alarm functions:

Function (switchable):

Pickup delay:

Switching voltage/current/power:

Electrical data & connections:

Transmitter function:

CU 22 function:

Isolated output voltage:

Hold function switch:

Load: pH/ORP input:

Connection terminal:

Display:

LCD:

Backlight:

Power supply:

Input:

Main fuse:

Pollution degree:

Transient overvoltage category:

EMC specifications:

Emitted interference:

Immunity to interference:

Environmental conditions:

Operating temperature range:

Max. relative humidity:

Mechanical specifications:

Dimensions (WxHxD):

Weight:

Ingress protection:

Alpha pH 2000D

(Wall mount)

TSPHCTP2000D

01X275375

Alpha pH 2000W

(Wall mount)

TSPHCTP2000W

01X275373

-2.00 to 16.00 pH or -1000 to 1000 mV or 0 to 100 %

0.1 to 1 pH

10 to 100 mV or 1 to 10.0 %

P/PI control (pulse length/pulse frequency/proportional integral); limit control; off

0.5 to 20 sec

60 to 120 pulse/min

0 to 999.9 min

0 to 2000 sec

0.1 to 200.0 hr

1 to 2000 sec

1 SPDT, 3 SPST relays

Max. 250 VAC / max. 3 A / max. 600 VA

Alpha pH 2000P

(Panel mount)

TSPHCTP2000P

01X275374

Two 0/4 to 20 mA scalable outputs for pH/ORP and temperature, galvanically isolated

22 mA current output

12 V ±0.5 V (max. 50 mA)

To freeze output current and deactivate control relays

2-pin terminal (Differential)

5-pin, 8-pin, 9-pin & 13-pin terminal, detachable blocks

315 mA time delay, 250 V,

Bussmann BK/GDC-315 mA

950 g (unit) / 1100 g (packed)

Max. 600 Ω

BNC (10 12 impedance); asymmetrical/symmetrical

3-pin, 8-pin, 9-pin & 13-pin terminal, detachable blocks

3-pin, 9-pin & 19-pin terminal, detachable blocks

80 to 250 VAC/DC ; 50/60 Hz ; 10 VA

144 x 144 x 110 mm

2

II

250 mA, anti-surge,

S504 Bussmann

745 g (unit) / 1100 g (packed)

IP66 (NEMA 4X)

-2.00 to 16.00 pH

0.01 pH

±0.01 pH

-1000 to 1000 mV / 0 to 100 %

1 mV / 0.1 %

±1 mV / ±0.2 %

-10.0 to 110 °C / 14.0 to 230.0 ºF

NTC 300; 2 wire

-10.0 to 125.0 °C / 14.0 to 257.0 ºF

0.1 ºC / 0.1 ºF

±0.5 ºC / ±1.0 ºF

Pt100 / Pt1000 (jumper selectable); 2 or 3 wire

Auto/manual (independent process/CAL temperature)

Steady or fleet (pulse)

0 to 2000 sec

Max. 250 VAC / max. 3 A / max. 600 VA

UV coat, backlit 14 segments display with symbols for status information

On/off selectable with four levels of brightness control

According to EN 61326

According to EN 61326

0 to 40 ºC

80 % up to 31 ºC decreasing linearly to 50 % at 40 ºC

96 x 96 x 175 mm

550 g (unit) / 950 g (packed)

IP54 (front panel)

* Includes side thread rods/brackets for mounting

Thermo Scientific / Alpha Process Products / 2000 Series Controllers / www.thermo.com/process 9

pH/ORP Electrodes

Order Code

Part No.

EC100GTSO20B EC100GTSO10B EC100GTSO05B ECARGTSO05B ECARHTTSO05B ECARTSOHF05B ECARTSO05B ECFP21A1A1

93X417005 93X417006 93X218865 93X218864 93X218860 93X218872 93X218859 93X218986 pH Electrodes

Classification pH range

Reference

Reference electrolyte

Operating temperature

Pressure tolerance

Temperature sensor

Potential matching pin/liquid ground

Material

Thread

Cable

Connector

Dimensions Length

Diameter

Weight

0 to 14 pH

Annular PTFE, double junction

0 to 14,

HF resistant

6 bars (87 psi)

Pt100

Saturated KCl, polymerized gel

0 to 110 ºC /

0 to 80 ºC / 32 to 176 ºF 32 to 230 ºF high temp.

9 bars (130 psi)

Platinum

PPS (Ryton ® )

¾” NPT

Integral 20 m

(65.6 ft) low-noise semi-conductor screened

Integral 10 m

(32.8 ft) low-noise semi-conductor screened

0 to 80 ºC / 32 to 176 ºF

6 bars (87 psi)

Integral 5 m (16.4 ft) low-noise semi-conductor screened

950 g 850 g 650 g

BNC

151 mm (excludes cable)

26 mm (external)

430 g

Differential pH

0 to 14

Kynar ® , differential pH electrode pH buffer solution

-5 to 95 ºC /

23 to 203 ºF

6.9 bars (100 psi)

NTC 300

SS316

CPVC

1½” NPT

Integral 3 m (9.8 ft) cables, tinned ends

152.7 mm

(excludes cable)

48.3 mm (external)

500 g

Order Code

Part No.

ECHTAUTSO05B

93X219128

ECHTPTTSO05B

93X219126

ECFR21A1A1

93X218987

ORP Electrodes

Classification

Sensor

ORP range

Reference

Reference electrolyte

Operating temperature

Pressure tolerance

Temperature sensor

Potential matching pin/liquid ground

Material

Thread

Cable

Connector

Weight

ORP

Gold

Annular PTFE, double junction

Saturated KCl, polymerized gel

± 1000 mV

0 to 80 ºC / 32 to 176 ºF

6 bars (87 psi)

Pt100

Platinum

PPS (Ryton ® )

¾” NPT

Integral 5 m (16.4 ft) low-noise semi-conductor screened

BNC

151 mm (excludes cable)

26 mm (external)

430 g

Differential ORP

Platinium

Kynar ® , differential pH electrode pH buffer solution

-5 to 95 ºC / 23 to 203 ºF

6.9 bars (100 psi)

NTC 300

SS316

CPVC

1½” NPT

Integral 3 m (9.8 ft) cables, tinned ends

152.7 mm (excludes cable)

48.3 mm (external)

500 g

10 Thermo Scientific / Alpha Process Products / 2000 Series Controllers / www.thermo.com/process

Electrode Selection Guide pH/ORP Electrodes

General pH measurement pH measurement with ATC pH measurement in noisy environment eg. electroplating pH measurement at high temperatures

(up to 110 °C/230 °F ; 9 bar/130 psi) pH measurement in the presence of Hydrofluoric Acid (HF)

Replaceable salt junctions & refillable reference electrolyte

General ORP/Redox measurement

ORP measurement in noisy environment

ORP measurement in Cyanide treatment

ORP measurements in oxidising applications (above 500 mV)

ORP measurements in reducing applications (below 500 mV)

Line Diagram (All dimensions are in mm unless specified otherwise)

pH/ORP Electrodes

EC100GTSO20B

EC100GTSO10B

EC100GTSO05B

ECARGTSO05B

ECARHTTSO05B

ECARTSOHF05B

ECARTSO05B

ECHTAUTSO05B

ECHTPTTSO05B

25.5

19 21

pH/ORP Differential

Electrodes

ECFP21A1A1

ECFR21A1A1

1.5” NPT

EC100GTSO20B 93X417005 EC100GTSO10B 93X417006 EC100GTSO05B 93X218865 ECARGTSO05B 93X218864 ECARHTTSO05B 93X218860 ECARTSOHF05B 93X218872 ECARTSO05B 93X218859 ECFP21A1A1 93X218986

• • • •

47

151

20.5

20.3

ECHTPTTSO05B 93X219126 ECFR21A1A1 93X218987

152.7

Accessories Ordering Information

Order Code Part Number Description

TSPHCTP2000D 01X275375 Alpha pH 2000 wall-mount pH/ORP differential controller/transmitter. Incl. terminal blocks, cable glands, gasket etc

TSPHCTP2000W 01X275373 Alpha pH 2000 wall-mount pH/ORP controller/transmitter (unit only). Incl. elbow BNC connector, terminal blocks, cable glands, gasket, etc

TSPHCTP2000P 01X275374

Alpha pH 2000 panel-mount pH/ORP controller/transmitter (unit only). Incl. elbow BNC connector, terminal blocks, gasket, threaded rods, catch, etc

EC100GTSO20B 93X417005 Ryton

EC100GTSO10B 93X417006 Ryton

EC100GTSO05B 93X218865 Ryton

®

®

®

ECARGTSO05B 93X218864 Ryton ®

ECARHTTSO05B 93X218860 Ryton ®

ECARTSOHF05B 93X218872 Ryton ®

-body pH combi electrode with Pt100 RTD (ATC) & 20 m cable with BNC & PMP

-body pH combi electrode with Pt100 RTD (ATC) & 10 m cable with BNC & PMP

-body pH combi electrode with Pt100 RTD (ATC) & 5 m cable with BNC & PMP

-body pH combi electrode with 5 m cable with BNC & connector for PMP (no ATC)

93X218859 Ryton ®

-body pH combi electrode with 5 m cable with BNC connector (no ATC); measures up to 110 ºC

-body pH combi electrode without ATC & 5 m cable with BNC connector. HF resistant glass

ECARTSO05B

ECFP21A1A1

-body pH combi electrode with 5 m cable with BNC connector (no ATC)

93X218986 PES-body differential pH combi electrode with NTC 300 (ATC) & 5 m tinned open-ended cable (Alpha pH 2000D only)

ECHTAUTSO05B 93X219128 Ryton ®

ECHTPTTSO05B 93X219126 Ryton ®

-body ORP gold electrode with 5 m cable with BNC & PMP (no ATC)

-body ORP platinum electrode with 5 m cable with BNC & PMP (no ATC)

ECFR21A1A1 93X218987 PES-body differential ORP platinum electrode with NTC 300 (ATC) & 20 ft tinned open-ended cable (Alpha pH 2000D only)

ECCBLO5SMK50 01X222801 Low-noise 50 m coaxial SMK cable for pH/ORP electrodes (without ATC), 5 mm, open-ended with no connectors

28X088001

ECCBL030510

28X088001 Male BNC connector for 5 mm extension cable; 1 unit (need BNC crimping tool to connect to extension cable)

01X222802

Low-noise 10 m coaxial cable for pH/ORP electrodes (without ATC; with PMP), 3 mm/5 mm, male-male BNC connectors

(for extending ECARGTSO05, ECHTAUTSO05B & ECHTPTTSO05B)

Low-noise 20 m coaxial cable for pH/ORP electrodes (without ATC; with PMP), 3 mm/5 mm, male-male BNC connectors

ECCBL030520 01X222803

(for extending ECARGTSO05, ECHTAUTSO05B & ECHTPTTSO05B)

ECCONBNCBNC 01X243102 BNC to BNC adapter (for extension of cable connection) – a pack of 10 units

ECAC021011

ECAK061014

81X220801

81X220802

CPVC electrode tee for pH/ORP electrodes with ¾” to 1” adapter

Kynar ® electrode tee for pH/ORP electrodes with ¾” to 1” adapter

ECPREAMP 01X228601 Pre-amplifier (for cable length exceeding 25 m) with female-female BNC connectors at each side of the junction box; batteries included

ECPHSIMULATOR 01X373301 Precision hi-low impedance & multiple buffers pH simulator (with BNC-BNC cable provided)

15X403101

01X211270

01X278701

15X403101 Salt bridge for ECFP21A1A1 & ECFR21A1A1 differential pH or ORP electrode

01X211270 Reference electrolyte for ECFP21A1A1 & ECFR21A1A1 differential pH or ORP electrode (500 ml)

01X278701 Panel mount kit for Alpha 2000 controllers/transmitters

Thermo Scientific / Alpha Process Products / 2000 Series Controllers / www.thermo.com/process 11

Thermo Scientific Alpha COND 2000

Conductivity Controller/Transmitter:

Conductivity:

Measure from ultra-pure water to highly conductive samples with the Alpha COND 2000 controller/ transmitter: this conductivity controller accepts 4-cell and

2-cell electrodes, and features programmable temperature coefficients to give you precise temperature compensation.

• Measures seven ranges, from

0 to 2 µS/cm to 0 to 1000 mS/cm

• Readings of up to three-decimal resolution at ±1 % accuracy

• Adjustable temperature coefficients for more precise temperature compensation:

Choose between programmable linear temperature compensation from 0.0 % to 10.0 %, and pure water temperature compensation to correct non-linearity of ultrapure water temperature correction curves

• Meter displays electrode status after each calibration

• User-adjustable cell constant values (k = 0.005 to k = 9.999)

• 4-cell electrode prevents polarizing effects and electrode fouling in samples with high conductivity

• Accepts 2- or 3-wire, Pt100 or

Pt1000 RTD sensors for automatic temperature compensation

• Allows user to input cell constant corresponding to connecting electrode independently during calibration

50

40

30

20

10

0

0

100

90

80

70

60

5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85

Temp ºC

Resistivity Of Pure Water

90 95 100

12 Thermo Scientific / Alpha Process Products / 2000 Series Controllers / www.thermo.com/process

Specification Information

Conductivity Controller/Transmitter

Order Code

Part No.

Conductivity:

Range:

Resolution:

Accuracy:

Temperature:

Range:

Resolution:

Accuracy:

Sensor:

Compensation:

Set-point & controller functions:

Set point 1 (SP1) / set point 2 (SP2):

Switching Conductivity hysteresis:

Function (switchable):

Adjustable period with pulse length controller:

Adjustable period with pulse frequency controller:

Integral action time (IAT):

Pickup/dropout delay:

Wash cycle:

Wash duration:

Contact outputs:

Switching voltage/current/power:

Alarm functions:

Function (switchable):

Pickup delay:

Switching voltage/current/power:

Electrical data & connections:

Transmitter function:

CU 22 function:

Isolated output voltage:

Hold function switch:

Load:

Conductivity input:

Connection terminal:

Display:

LCD:

Backlight:

Power supply:

Input:

Main fuse:

Pollution degree:

Transient overvoltage category:

EMC specifications:

Emitted interference:

Immunity to interference:

Environmental conditions:

Operating temperature range:

Max. relative humidity:

Mechanical specifications:

Dimensions (WxHxD):

Weight:

Ingress protection:

Alpha COND 2000W

(Wall mount)

TSCONCTP2000W

01X275376

Alpha COND 2000P

(Panel mount)

TSCONCTP2000P

01X275377

... to 2.000 µS/cm ; ... to 20.00 µS/cm ;

... to 200.0 µS/cm ; ... to 2000 µS/cm ;

... to 20.00 mS/cm ; ... to 200.0 mS/cm ;

... to 1000 mS/cm

0.001 µS/cm ; 0.01 µS/cm ;

0.1 µS/cm ; 1 µS/cm ;

0.01 mS/cm ; 0.1 mS/cm ;

1 mS/cm

±1 % of full scale reading (±2 % >500 mS/cm)

-10.0 to 125.0 °C (14.0 to 257.0 °F)

0.1 ºC / 0.1 ºF

±0.5 ºC / ±1.0 ºF

Pt100/Pt1000 (jumper selectable); 2 or 3 wire

Auto/manual (normalized at 25 °C)

... to 2.000 µS/cm or ... to 20.00 µS/cm or

... to 200.0 µS/cm or ... to 2000 µS/cm or

... to 20.00 mS/cm or ... to 200.0 mS/cm or

... to 1000 mS/cm

0 to 10 % of full scale

P/PI control (pulse length/pulse frequency/proportional integral); limit control; off

0.5 to 20 sec

60 to 120 pulse/min

0 to 999.9 min

0 to 2000 sec

0.1 to 200.0 hr

1 to 2000 sec

1 SPDT, 3 SPST relays

Max. 250 VAC / max. 3 A / max. 600 VA

Steady or fleet (pulse)

0 to 2000 sec

Max. 250 VAC / max. 3 A / max. 600 VA

Two 0/4 to 20 mA scalable outputs for pH/ORP and temperature, galvanically isolated

22 mA current output

12 V ±0.5 V (max. 50 mA)

To freeze output current and deactivate control relays

Max. 600 Ω

2-cell/4-cell terminal

5-pin, 8-pin, 9-pin & 13-pin terminal, detachable blocks

5-pin, 9-pin & 19-pin terminal, detachable blocks

UV coat, backlit 14 segments display with symbols for status information

On/off selectable with four level of brightness control

80 to 250 VAC/DC ; 50/60 Hz ; 10 VA

315 mA, anti-surge, 250 V

2

II

According to EN 61326

According to EN 61326

0 to 40 °C

80 % up to 31 °C decreasing linearly to 50 % at 40 °C

144 x 144 x 110 mm

745 g (unit) / 1100 g (packed)

IP66 (NEMA 4X)

96 x 96 x 175 mm

550 g (unit) / 950 g (packed)

IP54 (front panel)

Thermo Scientific / Alpha Process Products / 2000 Series Controllers / www.thermo.com/process 13

Conductivity Electrodes

Order Code

Part No.

EC91346S ECCS10-0-01T ECCS10-0-01TS ECCS10-0-01S ECCS10-0-01SS ECCS10-0-1S ECCS10-0-1SSP ECCS10-1-0S ECCS10-1-0SSP

93X219048 93X219019 93X219054 93X219018 93X219053 93X219020 93X219055 93X219021 93X219056

Conductivity

Electrodes

Conductivity range

Cell constant, K

Temperature sensor

Pressure rating at 25 ºC

Material

Fitting material

Thread

Cable

Dimensions

Weight

Length

Diameter

Up to 500 mS/cm

0.3, 4-cell

6 bar

(87 psi)

Ryton ®

SS316

,

¾” NPT

Integrated

7.6 m (24.9 ft),

8-wire doubleshielded, tinned ends

150.5 mm

(excludes cable)

22.2 mm

(external)

430 g

3.4 bar

(50 psi)

-5 to 50 ºC /

23 to 122 ºF

0.55 to 20 µS/cm

0.01, 2-cell

5.5 bar

(80 psi)

-5 to 80 ºC /

23 to 176 ºF

Titanium

Nylon plastic

Stainless steel

3.4 bar

(50 psi)

-5 to 50 ºC /

23 to 122 ºF

Pt100, 3-wire

5.5 bar

(80 psi)

-5 to 80 ºC /

23 to 176 ºF

Nylon plastic

Stainless steel

3.4 bar

(50 psi)

-5 to 50 ºC /

23 to 122 ºF

½” NPT

0.5 to 200 µS/cm

0.1, 2-cell

Nylon plastic

-5 to 200 ºC /

23 to 392 ºF

SS316

13.8 bar

(200 psi)

Stainless steel

600 g 680 g

Integrated 7.5 m (24.6 ft), 6-wire double-shielded, tinned ends

680 g

168 mm

(excludes cable)

660 g

12.8 mm

(external)

560 g 660 g

0.01 to 200 mS/cm

1.0, 2-cell

3.4 bar

(50 psi)

-5 to 50 ºC /

23 to 122 ºF

Nylon plastic

590 g

13.8 bar

(200 psi)

-5 to 120 ºC /

23 to 248 ºF

Stainless steel

660 g

Electrode Selection Guide

Conductivity Electrodes

General Conductivity measurements

Low Conductivity & Resistivity measurements

High Conductivity measurements

(4-cell electrodes)

Conductivity measurements with ATC

Conductivity measurements of ultrapure water

Conductivity measurements of pure water

Conductivity measurements of power plant

& condensate water

EC91346S 93X219048

• •

ECCS10-0-01T 93X219019

ECCS10-0-01TS 93X219054

ECCS10-0-01S 93X219018

ECCS10-0-01SS 93X219053

ECCS10-0-1S 93X219020

ECCS10-0-1SSP 93X219055

• •

ECCS10-1-0S 93X219021

ECCS10-1-0SSP 93X219056

14 Thermo Scientific / Alpha Process Products / 2000 Series Controllers / www.thermo.com/process

Line Diagram (All dimensions are in mm unless specified otherwise)

Conductivity Electrodes

EC91346S

Black Strain

Relief

150.5

Conductivity Electrodes

ECCS10-0-01T

ECCS10-0-01S

Cable Length: 30’

Black Strain

Relief

Ryton

Cap

3/4” NPT

41.5

3/4” NPT

22.9

½” MNPT

Plastic Nylon ø12.7

Titanium Body

110 ±2

Cable Length: 25’ Belden 9536

(6 Conductor Cable)

Conductivity Electrodes

ECCS10-0-1S

Black Strain

Relief

Ryton

Cap

½” MNPT

Plastic Nylon ø12.7

Stainless Steel

0.500”

110 ±2

Cable Length: 25’ Belden 9536

(6 Conductor Cable)

Conductivity Electrodes

ECCS10-1-0S

Black Strain

Relief

Ryton

Cap

½” MNPT

Plastic Nylon ø12.7

Stainless Steel

Roll Tip

316

Stainless Steel

Roll Over

PTFE

110 ±2

Cable Length: 25’ Belden 9536

(6 Conductor Cable)

Accessories Ordering Information

Order Code

TSCONCTP2000W 01X275376

TSCONCTP2000P

Part Number Description

01X275377

Alpha COND 2000 wall-mount Conductivity controller/transmitter. Incl. terminal blocks, cable glands, gasket, etc

Alpha COND 2000 panel-mount Conductivity controller/transmitter. Incl. terminal blocks, gasket, threaded

EC91346S

ECCS10-0-01T

ECCS10-0-01TS

ECCS10-0-01S

93X219048

93X219019

93X219054

93X219018 rods, catch, etc

4-cell Conductivity electrode with 3-wire Pt100, 30 ft tinned open-ended cable

Conductivity/Resistivity electrode with Pt100, cell constant K=0.01, titanium with 25 ft tinned open-ended cable

(with ½ inch nylon plastic cap threading)

Conductivity/Resistivity electrode with Pt100, cell constant K=0.01, titanium with 25 ft tinned open-ended cable

(with ½ inch stainless steel cap threading)

Conductivity/Resistivity electrode with Pt100, cell constant K=0.01, stainless steel with 25 ft tinned open-ended cable

ECCS10-0-01SS

ECCS10-0-1S

ECCS10-0-1SSP

ECCS10-1-0S

ECCS10-1-0SSP

ECAC021022

01X278701

93X219053

93X219020

93X219055

93X219021

93X219056

81X220803

01X278701

Conductivity/Resistivity electrode with Pt100, cell constant K=0.01, stainless steel with 25 ft tinned open-ended cable

Conductivity electrode with Pt100, cell constant K=0.1, stainless steel with 25 ft tinned open-ended cable

(with ½ inch nylon plastic cap threading)

Conductivity electrode with Pt100, cell constant K=0.1, stainless steel with 25 ft tinned open-ended cable

(with ½ inch stainless steel cap threading)

Conductivity electrode with Pt100, cell constant K=1.0, stainless steel with 25 ft tinned open-ended cable

(with ½ inch nylon plastic cap threading)

Conductivity electrode with Pt100, cell constant K=1.0, stainless steel with 25 ft tinned open-ended cable

(with ½ inch stainless steel cap threading)

CPVC electrode tee for Conductivity/Resistivity electrodes with ½” to 1” adapter

Panel mount kit for Alpha 2000 controllers/transmitters

Thermo Scientific / Alpha Process Products / 2000 Series Controllers / www.thermo.com/process 15

Thermo Scientific Alpha DO 2000

Dissolved Oxygen Controller/Transmitter:

Dissolved Oxygen:

With a wide measuring range of 0 to 20 ppm or 0 to 200 % saturation, Alpha DO 2000 controller/ transmitter is ideal for use in a variety of applications, ranging from water treatment and monitoring to pharmaceutical and food processing.

Meter comes in both Galvanic and

Amperometric versions.

• Readings values in mg/L or % air saturations, selectable from menu options

• Quick, easy calibration using atmospheric air as calibration media for 100 % air saturation

• Pressure, Salinity and Automatic

Temperature Compensation

The Galvanic Electrode

(ECDOGEN-S & ECDOTPII-S)

The Amperometric Electrode

(EC237150 & EC237450)

Rugged and long-lasting, the galvanic electrode requires no warm-up time and prevents zero drift by recreating H 2 O within the electrode. This prevents change of pH to the electrode and allows it to recycle within the probe. The result: galvanic sensors have long lifespans and require low maintenance costs.

The amperometric dissolved oxygen electrode contains special membranes that are designed to require almost zero maintenance. These unique membranes stay stable even under harsh ambient conditions and high pressure. This enables the sensors to capture accurate

DO readings within short response time at required flow as low as 0.02 m/s.

16 Thermo Scientific / Alpha Process Products / 2000 Series Controllers / www.thermo.com/process

Specification Information

Dissolved Oxygen

Controller/Transmitter

Order Code

Part No.

Dissolved Oxygen:

Range:

Resolution:

Accuracy:

Temperature:

Range:

Resolution:

Accuracy:

Sensor:

Compensation:

Temperature compensation:

Salinity compensation:

Pressure compensation:

Set point & controller functions:

Set point 1 (SP1) / set point 2 (SP2):

Switching DO hysteresis:

Function (switchable):

Adjustable period with pulse length controller:

Adjustable period with pulse frequency controller:

Integral action time (IAT):

Pickup/dropout delay:

Wash cycle:

Wash duration:

Contact outputs:

Switching voltage/current/power:

Alarm functions:

Function (switchable):

Pickup delay:

Switching voltage/current/power:

Electrical data & connections:

Transmitter function:

CU 22 function:

Isolated output voltage:

Hold function switch:

Load:

DO input:

Connection terminal:

Display:

LCD:

Backlight:

Power supply:

Input:

Main fuse:

Electromagnetic compliance:

Emitted interference:

Immunity to interference:

Environmental conditions:

Operating temperature range:

Max. relative humidity:

Mechanical specifications:

Dimensions (WxHxD):

Weight:

Ingress protection:

Alpha DO 2000W

(Galvanic)

(Wall mount)

TSDOCTP2000W

01X275378

0.00 to 25.00 mg/L or ppm;

0.0 to 300.0 %

±1.5 % of full scale reading

0.00 to 25.00 mg/L or ppm;

0.0 to 300.0 %

0 to 10.0 %

Alpha DO 2000WPG

(Amperometric)

(Wall mount)

TSDOCTP2000WPG

01X275379

0.00 to 20.00 mg/L or ppm;

0.1 to 1.0 mg/L or 1.0 to 10.0 %

P/PI control (pulse length/pulse frequency/proportional integral); limit control; off

0.5 to 20 sec

60 to 120 pulse/min

0.0 to 999.9 min

0 to 1999 sec

0.1 to 199.9 hr

1 to 1999 sec

0.0 to 200.0 %

1 SPDT, 3 SPST relays

Max. 250 VAC / max. 3 A / max. 600 VA

Alpha DO 2000PPG

(Amperometric)

Pt100/Pt1000 (jumper selectable);

2 or 3 wire

-10.0 to 125.0 °C / 14.0 to 257.0 °F

0.1 ºC / 0.1 ºF

±0.5 ºC / ±1.0 ºF

NTC 22 kΩ thermistor

(Panel mount)

TSDOCTP2000PPG

0.0 to 50.0 ppt (manual setting and automatic correction)

10.73 to 43.51 psi (manual setting and automatic correction)

01X275380

Two 0/4 to 20 mA outputs for DO values and temperature, galvanically isolated

22 mA current output –

12 V ±0.5 V (max. 50 mA)

To freeze output current and deactivate control relays

5-pin, 8-pin, 9-pin & 13-pin terminal, detachable blocks

Max. 600 Ω

2-pin terminal

2x 3-pin, 8-pin, 9-pin & 13-pin terminal blocks

3-pin, 9-pin & 19-pin terminal blocks

UV coat, backlit 14 segments display with symbols for status information

On/off selectable with four level of brightness control

315 mA, anti surge, 250 V

80 % up to 31 °C decreasing linearly to 50 % at 40 °C

144 x 144 x 111.5 mm

0.01 mg/L or ppm; 0.1 %

0.740 to 3.000 bar ; 555 to 2250 mmHg ;

Steady or fleet (pulse)

Max. 250 VAC / max. 3 A / max. 600 VA

80 to 250 VAC/DC ; 50/60 Hz ; 10 VA

250 mA, anti surge

According to EN 61326

According to EN 61326

IP66 (NEMA 4X)

0.00 to 20.00 mg/L or ppm;

±1 % of full scale reading

Auto/manual

0 to 1999 sec

0 °C to 50 °C

144 x 144 x 110 mm

745 g (unit) / 1100 g (packed)

0.0 to 200.0 %

175 x 96 x 96 mm

700 g (unit) / 850 g (packed)

IP54 (front panel)

Thermo Scientific / Alpha Process Products / 2000 Series Controllers / www.thermo.com/process 17

Dissolved Oxygen Electrodes

Order Code

Part No.

ECDOGEN-S

01X247507

Dissolved Oxygen

Electrodes

ECDOTPII-S

01X247508

EC237150

93X416401

EC237450

93X416402

Dissolved Oxygen range

Type

Flow rate

Response time

Temperature sensor

Pressure rating

Operating temperature

Material

Membrane

Cable

Weight

0.50 to 20 ppm 0.03 to 20 ppm

Galvanic

50 mm/sec (dependent on temperature and O

2

level)

40 to 50 sec to attain 95 % of actual reading

Pt100

6 bar (87 psi)

0 to 50 ºC / 32 to 122 ºF

Delrin housing

HDPE

Integral 3 m (9.8 ft) water-resistant, tinned ends

152.4 mm (excludes cable)

58.4 mm (external)

670 g

Electrode Selection Guide

0.04 ppm to saturation 0.01 ppm to saturation

Amperometric

20 mm/sec

22 kΩ NTC

4 bar (58 psi)

0 to 60 ºC / 32 to 140 ºF

Non-replaceable

0 to 130 ºC / 32 to 248 ºF

Stainless steel

FDA membrane

(sold separately)

12 mm shaft diameter (excludes cable)

178 (external)

670 g

Dissolved Oxygen Electrodes

DO measurements at low levels

Waterproof probes

Galvanic DO measurements system for general purposes, eg. wastewater

& aquaculture

Galvanic DO measurements system for low

DO level, eg. power plants, metal corrosion test facilities

Biocompatible electrode materials

Amperometric DO measurements system for biotechnology, steam sterilisation & autoclaving, fermentation applications

Amperometric DO measurements system for general purposes, eg. wastewater, fish farms, composting facilities

ECDOGEN-S 01X247507

ECDOTPII-S 01X247508

EC237150 93X416401

EC237450 93X416402

18 Thermo Scientific / Alpha Process Products / 2000 Series Controllers / www.thermo.com/process

Line Diagram (All dimensions are in mm unless specified otherwise)

Dissolved Oxygen Electrodes

ECDOGEN-S

ECDOTPII-S

152.4

26.7

58.4

Red Wire (+) Black Wire (-)

Black Cable

Black Nut =

DOGEN-S

Gray Nut =

DOTPII-S

Zinc Anode

DOGEN-S

Lead Anode

DOTPII-S

(Internal)

ø1.05”

(26.7 mm)

Silver

Cathode

(Internal) ø2.30”

(58.4 mm)

6.00”

(152.4 mm)

Galvanic Probe

(Construction)

Membrane

Dissolved Oxygen Electrode

EC237450

178

120

Accessories Ordering Information

Order Code

TSDOCTP2000W

Part Number Description

01X275378

Alpha DO 2000 galvanic wall-mount Dissolved Oxygen controller/transmitter. Incl. terminal blocks, cable

TSDOCTP2000WPG 01X275379

TSDOCTP2000PPG 01X275380

ECDOGEN-S

ECDOTPII-S

EC237150

EC237450

01X241605

01X241606

32X246702

15X241503

01X247507

01X247508

93X416401

93X416402

01X241605

01X241606

32X246702

15X241503

ECDOGENSOLNBT 01X211228

ECDOTPIISOLNBT 01X211229

EC237140

EC237118

EC237137

EC355089

EC355136

01X278701

01X416501

01X416601

15X416701

30X416801

30X416802

01X278701 glands, gasket, etc

Alpha DO 2000 polarographic wall-mount Dissolved Oxygen controller/transmitter. Incl. terminal blocks, cable glands, gasket, etc

Alpha DO 2000 polarographic panel-mount Dissolved Oxygen controller/transmitter. Incl. terminal blocks, gasket, threaded rods, catch, etc.

Delrin housing-body galvanic Dissolved Oxygen electrode with Pt100, 5 m tinned open-ended cable

Delrin housing-body, galvanic Dissolved Oxygen electrode with Pt100, 5 m tinned open-ended cable

Polarographic probe with 22 KOhm NTC, 5 m tinned open-ended cable

Fermentation polarographic probe with 22 KOhm NTC, with extra membrane, without cable

Set of 5 o-rings & membranes (DOGEN-S)

Set of 5 o-rings & membranes (DOTPII-S)

Large o-ring (DOGEN-S/DOTPII-S)

Tool for membrane housing (DOGEN-S/DOTPII-S)

DO refilling electrolyte for ECDOGEN-S (480 ml bottle)

DO refilling electrolyte for ECDOTPII-S (480 ml bottle)

3 FDA membrane bodies, electrolyte, pipette, spare o-ring, polishing strip (for EC237450)

Electrolyte for EC237450, 50 ml

Autoclavation connector cap for EC237450

5 m cable for EC237450

10 m cable for EC237450

Panel mount kit for Alpha 2000 controllers/transmitters

Thermo Scientific / Alpha Process Products / 2000 Series Controllers / www.thermo.com/process 19

Wall Mount – ½ DIN (in mm)

144

111.5

27.5

6

80

39

78

PG13.5 (3 pcs)

90

Holes for post mounting (4X)

Holes for wall mounting (2X)

Red plugs for wall mounting holes enclosure

(2X)

24

57.8

For PG13.5 cable glands

Panel Mount – ½ DIN (in mm)

Max. 80 84

1 - 22

29

Seal gasket

175

138 175

Threaded rod for panel mounting

Panel Mount – ¼ DIN (in mm)

138

Cut-out mounting plates for post mounting

Flat gasket

1

92+0.5

Panel cut-out

Max. 180

Max. 45

20 Thermo Scientific / Alpha Process Products / 2000 Series Controllers / www.thermo.com/process

Sensor Installation Tips

There are two main sensor installation configurations:

1. In-line Installation where electrodes are installed into the piping system

• In-line installation is best in a location which is always flooded.

It is recommended that the electrode be installed in the dead leg of a pipe tee fitting

(Figure 4)

This reduces the velocity of the

. fluid as it flows into and around the electrode before flowing upwards and away

• Take into consideration the amount of cleaning and calibrating your application imposes on the electrode. In-line mounting may mean the process has to be shut down, or that a portion of the pipe must be drained in order to remove the electrode

• Remove cables before unscrewing electrodes from the pipes to avoid twisting and damaging wires or shielding

2. Submersible Installation where electrodes are submerged in tanks

• Use a baffle in submersible installation to avoid air bubbles from being trapped inside or around the electrode, distorting readings

(Figure 5)

. Install your electrode away from mixers, sparges and inlets

• Mount electrodes near outlets of the tank, away from points where reagents are introduced.

This allows time for reagents to mix thoroughly and react with your effluent water. Some applications, such as neutralization processes that require pH control, are best executed with submersible configuration than with in-line configuration, where both pH and flow are variables

• Some sensors must be mounted at a ±30 ° angle

(Figure 6)

. This ensures that the electrolyte remains in contact with sensing tips, and prevents bubble entrapment within the electrode membrane. Some electrodes, however, like our DO Sensor,

EC237450, can be installed even upside-down – very useful for measurements in tanks that are nearly empty

• Keep connection between electrode and preamplifier dry and clean at all times. Moisture in this area may cause permanent damage

Whether you’re considering an in-line or submersible configuration, the principle considerations are to avoid air bubble entrapment, as well as to ensure maximum exposure of the sensor to your process water.

In processes when temperature and pressure of the process liquid exceeds the sensor ratings, a sidestream system may be used, where a portion of the process liquid is constantly directed to the sensor.

Sensor

Tee

Figure 4: In-line installation

Flow

Sensor

30 º Min

Weld Spud at Angle

Figure 5: Submersible installation Figure 6: Angle mounting installation

Thermo Scientific / Alpha Process Products / 2000 Series Controllers / www.thermo.com/process 21

Applications

pH Application

The pH neutralization system is crucial in the back-end process of treating effluent water in industries such as pharmaceutical, F&B manufacturing, electroplating and seawater desalination etc.

In a neutralization system, effluent water flows into a tank where a pH electrode picks up the pH level of the process water. When this pH level goes below or above set limits, the electrode sends a signal to the meter, which in turn, adjusts the pH level in the tank by operating two chemical pumps that release acid or alkali into the tank as needed.

Pump

Acid

Alpha pH 2000D

Recorder

Influent

Electrode

Mixer

Valve

Effluent

Base

Pump

Alpha pH 2000 Alpha pH 2000

pH & ORP Application

In applications where pH and ORP must be measured and controlled simultaneously, two Alpha pH/ORP 2000 meters can be placed side by side where the pH meter acts as the primary controller – the ORP controller will only operate its pumps when pH is within set limits.

Pump pH Adjustment

Influent

Electrode

Mixer

Valve

Pump

Effluent

ORP Adjustment

Alpha COND 2000

Conductivity Application

In the back-end process of the electroplating industry, conductivity monitoring systems monitor the conductivity of rinse water in metal finishing facilities, maintaining chemical concentrations at levels that provide adequate rinsing and at the same time, prevent excessive dragin to subsequent process tanks. When rinse water conductivity exceeds set limits, the meter activates a valve that introduces more water into the rinse tank, reducing conductivity of the rinse water.

Valve

Rinse

Tank Electrode

Mixer

Valve

Effluent

Recorders

Alpha DO 2000 /

Alpha DO 2000PG

pH & DO Application

In biological wastewater treatment, both dissolved oxygen level and pH level are vital in ensuring efficient breakdown of biological waste in the effluent water.

A dissolved oxygen sensor picks up the temperature and

DO level of the process water in the biological wastewater tank. It then relates this information to the DO meter, which controls an air pump to increase or reduce the amount of oxygen level that goes into the effluent water.

Pump

Acid

Alpha pH 2000

Influent

Electrodes

Mixer

Valve

To

Secondary

Treatment

Base

Pump

22 Thermo Scientific / Alpha Process Products / 2000 Series Controllers / www.thermo.com/process

Trademarks Used:

Viton ® is a registered trademark of DuPont

Performance Elastomers.

Kynar ® is a registered trademark of Arkema Inc.

Ryton ® is a registered trademark of Chevron

Philips Chemical Company LLC.

Warranty:

Thermo Fisher Scientific provides one year of warranty against manufacturing defects for meters, and six months for electrodes.

Disclaimers:

Specifications and terms are subject to change.

Not all products are available in all countries.

Please consult your local sales representative for details.

All drawings and diagrams are for illustration purposes only and are not drawn to scale.

About Thermo Fisher Scientific

Thermo Fisher Scientific (NYSE: TMO) is the world leader in serving science, enabling our customers to make the world healthier, cleaner and safer. With annual sales of more than $9 billion, we employ 30,000 people and serve over 350,000 customers within pharmaceutical and biotech companies, hospitals and clinical diagnostic labs, universities, research institutions, government agencies as well as environmental and industrial process control settings. Serving customers through two premier brands, Thermo

Scientific and Fisher Scientific, we help solve analytical challenges from routine testing to complex research and discovery. Thermo

Scientific offers customers a complete range of high-end analytical instruments as well as laboratory equipment, software, services, consumables and reagents to enable integrated laboratory workflow solutions. Fisher Scientific provides a complete portfolio of laboratory equipment, chemicals, supplies and services used in healthcare, scientific research, safety and education. Together, we offer the most convenient purchasing options to customers and continuously advance our technologies to accelerate the pace of scientific discovery, enhance value for customers and fuel growth for shareholders and employees alike.

Thermo Scientific / Alpha Process Products / 2000 Series Controllers / www.thermo.com/process 23

Environmental Instruments

Water Analysis Instruments

North America

166 Cummings Center

Beverly, MA 01915 USA

Toll Free: 1-800-225-1480

Tel: 1-978-232-6000

Dom. Fax: 1-978-232-6015

Int’l Fax: 978-232-6031 www.thermo.com/process

Europe

Denmark House, Angel Drove

Ely, Cambridgeshire

CB7 4ET, UK

Tel: 44-1353-666111

Fax: 44-1353-666001

Asia Pacific

Blk 55, Ayer Rajah Crescent

#04-16/24, Singapore 139949

Tel: 65-6778-6876

Fax: 65-6773-0836

© 2007 Thermo Fisher Scientific Inc.

All rights reserved.

www.thermo.com/process

CON 08X233408 05/08 Rev 0

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement