ALLPRO PROVECTA V Veterinary X-Ray System Operator’s Manual
Below you will find brief information for Veterinary X-Ray System PROVECTA V. The PROVECTA V is a veterinary dental imaging equipment with remote control that can be used for film or digital radiology processing. It comes with 30 preset exposure times and allows manual control of exposure time in 44 steps. The system also allows you to customize the settings for the different tooth categories, patient sizes, and media types you work with. The PROVECTA V has a built-in tubehead simulator for training and testing the X-Ray Control operation.
PDF
Download
Document
Advertisement
Advertisement
PROVECTA V Veterinary X-Ray System with Remote Control Operator’s Manual TABLE OF CONTENTS Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 On-Line Warranty Registration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Key Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Safety Notice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . General Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Protection Against Radiation Warning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Knowledge of Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Environmental Risk and Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 6 6 6 7 7 7 Cleaning and Disinfecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Key Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Available Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . PROVECTA V System Parts Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 8 9 9 T e c h n i c a l S p e c i f i c a t i o n s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Technical Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Tubehead Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-11 X-Ray Tube Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Apparatus and Detachable Parts Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Selected Exposure Time and Corrected Actual Exposure Time . . . . . . . . . . . . . . . . . . 11 Exposure Time Correction Factors vs Line Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Technical Factors Measuring Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 X-Ray Tube Curves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 X - R a y T u b e h e a d C o m p o n e n t s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 X-Ray Tubehead Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Filament Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Filament Loading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . X-Ray Tubehead Cooling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cooling Curve of Tubehead . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 15 15 16 16 S y s t e m A c t i v a t i o n . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 R e m o t e X - R a y C o n t r o l P a n e l F u n c t i o n s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18-19 Keypad Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Increase Select Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Decrease Select Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Tooth Select Mode Select Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . SIZE Mode Select Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Digital/Film Mode Select Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Default Exposure Time Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 20 20 20 20 21 21 21 TABLE OF CONTENTS Keypad Operation: Custom Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Changing Default Settings to Customized Values . . . . . . . . . . . . . . . . . . . . . . . . . . . Changing Display to Pulses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Changing Display to Seconds . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Changing Default Tooth Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Changing Default Large and Small Multiplier Ratio . . . . . . . . . . . . . . . . . . . . . . . . . . Changing Film Multiplier Ratio Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Changing Digital Multiplier Ratio Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Factory Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 22 22 22 23 23 24 24 24 System Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Setting the Exposure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Automatic Exposure Time Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . PROVECTA V Available Preset Exposure Time Setting Combinations . . . . . . . . . . . . . . Manual Exposure Time Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Exposure Emission Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Reading the Exposure Counter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 25 25 25 26 27 27 X - R a y S y s t e m R e a c h a n d S t o r a g e D i m e n s i o n s . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 R a d i o g r a p h i c T r o u b l e s h o o t i n g . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29-30 S y s t e m M a i n t e n a n c e . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 S y s t e m E r r o r M e s s a g e s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31-34 N o t e s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 3 INTRODUCTION Congratulations, your new PROVECTA V Veterinary X-ray (hereafter referred to as PROVECTA V) has been engineered to provide you with years of reliable service. The PROVECTA V will create radiographs of excellent quality. It will perform equally well using digital or film-based imaging media. Review and follow the guidelines included in this Operator’s Manual to thoroughly become familiar with the operating and safety procedures. This will ensure that your PROVECTA V gives you the highest level of service. Your PROVECTA V Intraoral X-Ray model A2000RV contains the following components: ITEM Straight Arm (one only) PART NUMBER 16 Inches Long 24 Inches Long 35 Inches Long Scissor Arm 70kVp 8mA Tubehead 16-Inch on Center Wall Mount Plate Mounting Hardware 16 Inch on Center Wall Plate Template Single Post Mount Plate Template Remote X-ray Control Panel X-ray Arm Support Connection Enclosure Service Manual Users Manual Warranty Registration Information Card Form FDA 2579 Installation Accessory Kit A2140-3 A2140-2 A2140-1 A2120 A2250V A2570 A2020 A2691 A2077 A2290V A2500 A2501V A2507V 9922-190 A2012 A2372R WARRANTY The PROVECTA V Veterinary X-Ray is warranted to be free from defects in material and workmanship from the date of installation for a period of twenty-four (24) months. Any item returned to our factory through an ALLPRO Imaging Authorized Dealer, will be repaired or replaced at our option at no charge provided that our inspection shall indicate it to have been defective. Dealer labor, shipping and handling charges are not covered by this warranty. This warranty does not apply to damage due to shipping, misuse, careless handling or repairs by other than authorized service personnel. Warranty void if installed or serviced by other than Authorized ALLPRO Dealer service personnel. ALLPRO, Inc. is not liable for indirect or consequential damages or loss of any nature in connection with this equipment. This warranty is in lieu of all other warranties express or implied. No representative or person is authorized to assume for us any liability in connection with the sale of our equipment. ON-LINE WARRANTY REGISTRATION Quickly and easily register your new PROVECTA V on-line. Just have your product model and serial numbers available. Then go to the ALLPRO website, www.allproimaging.com, click the warranty link and complete the registration form. This on-line registration ensures a record for the warranty period and helps ALLPRO keep you informed of product updates and other valuable information. 4 KEY FEATURES 1. Microprocessor controlled. 2. Exposure compensation over input AC voltage range. 3. User selectable exposure time in seconds/pulses. 4. 30 preset exposure times. 5. 44 Manually selectable exposure times. 6. Programmable Size ratios. 7. Programmable Film setting. 8. Programmable Digital setting. 9. Two memorized exposure settings. 10. Stable positioning. 11. Remote Mounted X-ray Control Panel. 12. Optional Remote Exposure Switch that is also settable to use in a location where two switches are required to be activated simultaneously. 13. Single 4 x 4 post mount and 16” on center plate mounting. 14. Designed for ease of installation and use. Printed circuit boards are hinged to allow unit to be easily installed. 15. Built in tubehead simulator for training and testing the X-Ray Control operation. SAFETY NOTICE ATTENTION USERS: MANUFACTURING DATE CODE ON SERIAL NUMBER LABEL IS IN THE FORMAT Month YYYY. Warns users that uninsulated voltage within the unit may be of sufficient magnitude to cause electric shock. CL A S S IF I ED Alerts users to important Operating and Maintenance instructions. Read carefully to avoid any problems. MEDICAL ELECTRICAL EQUIPMENT WITH RESPECT TO ELECTRICAL SHOCK, FIRE, MECHANICAL AND OTHER SPECIFIED HAZARDS ONLY IN ACCORDANCE WITH UL-60601-1, CAN/CSA C22.2 NO.601.1 66CA Indicates type B equipment in accordance with IEC 601-1 DO NOT ATTEMPT INTERNAL SERVICE The interior of the Main Assembly is only accessible by removing hardware with tools and should only be opened and serviced by an authorized ALLPRO Dealer Service representative. Call your authorized ALLPRO dealer for service. Since the interior of the unit contains high voltage components, failure to heed this warning may result in equipment damage, personal injury and/or death. 5 SAFETY NOTICE GENERAL SAFETY PRECAUTIONS This equipment has been designed and manufactured in compliance with applicable safety requirements; moreover, this manual provides all the necessary information for correct and safe utilization. It provides warnings related to risks associated with the PROVECTA V and should be read very carefully. ALLPRO Imaging Inc. shall not be responsible for: any use of PROVECTA V equipment different from that for which it was been designed, any damage to the equipment, the operator or the patient caused either by incorrect installation or maintenance not compliant with the procedures contained in the relevant user’s and installation manuals provided with the equipment, or by incorrect operation techniques, any mechanical and/or electrical changes made during or after installation, different from those specified in the service manual. Only qualified service personnel, authorized by ALLPRO Imaging, Inc., or its dealers are allowed to perform repairs on the equipment. PROTECTION AGAINST RADIATION WARNING: Radiation protection is generally prescribed or regulated by local, State or Federal regulatory agencies. These regulations are beyond the scope of this manual. Operation and use of the device should be limited to trained personnel in accordance with local, state and federal regulations. a. The film, phosphor plate or sensor must be introduced in the patient’s mouth either manually or by means of appropriate holders; it must never be held by the operator. b. During radiation exposure, the operator must not be in contact with the tubehead or the collimator. c. During radiation exposure, the operator must keep at a certain distance from the X-ray source (at least 6-1/2 feet), in a direction opposite to that of X-ray emission. d. During radiation exposure, no one other than the operator and the patient must be present in the room. SYMBOLS USED The following symbols are used on the PROVECTA V and in this manual. AC Connection to Neutral Conductor Connection to Line Conductor Protection Grounding Functional Grounding OFF; Equipment not Connected to Power Line ON; Equipment Connected to Power Line X-ray Emission 6 SAFETY NOTICE KNOWLEDGE OF WARNINGS Users must exercise every precaution to ensure personnel safety, and be familiar with the warnings presented throughout this manual and summarized below. WARNINGS: Federal law restricts this device to sale by or on the order of a Veterinarian. This equipment complies with DHHS performance standard 2l CFR Subchapter J. This equipment complies with the requirements of IEC 60601-2-7, and UL 60601-1. The X-ray Tubehead Assembly P/N A2250 complies with IEC 60601-2-28 (1993). The equipment must be used in compliance with the procedures contained in this manual and shall not be used for purposes different from those described in this manual. PROVECTA V is classified as veterinary dental imaging equipment and must therefore be used only under the supervision of qualified veterinary staff, having the necessary knowledge in the field of radiation. The user bears legal responsibility related to the possession, installation and use of this device. Do not remove panels. Removing panels would create an electric shock hazard. Equipment is not suitable for use in the presence of flammable anesthetic mixture with air or with oxygen or nitrous oxide. For continued protection against risk of fire, replace only with the same type and rating of circuit breakers and fuses. Replace components only with authorized replacement parts identified in this manual. Before performing any maintenance, the equipment must be disconnect from the AC power source. Do not allow water or other liquids to penetrate inside the equipment to avoid short circuits and corrosion. Although the x-ray doses provided by modern equipment are reduced as low as reasonably possible, during exposure the operator must take all the necessary precautions and/or protection measures for the patient and themselves, in compliance with existing regulations. PROVECTA V must be switched off during the use of High Frequency surgical devices or similar instruments placed near the equipment. Do Not place ANY additional weight on the scissor arms. Note: ALLPRO Imaging, Inc., is not responsible for disposal of the apparatus or parts thereof and for the related expenses. ENVIRONMENTAL RISK AND DISPOSAL The equipment contains - in some of its parts - solid and liquid substances which must be disposed of at appropriate recycling centers conforming to all local, state and federal regulations. In particular, the equipment contains the following materials and/or components: Tubehead: non-biodegradable plastic materials, metals, glass, dielectric oil, lead, tungsten. Other parts of the equipment: non-biodegradable plastics, metals, printed circuits, and electronic components. 7 CLEANING & DISINFECTING CLEANING AND DISINFECTING PROCEDURES In order to ensure proper hygiene and cleaning of the equipment, the following procedures must be followed thoroughly: Warning: Disconnect the branch circuit breaker or AC power source before performing cleaning procedures. Active internal circuits remain connected to main voltage even when main power switch is turned off. Exterior Surfaces: Wipe the outside surfaces with a paper towel dampened with a disinfectant solution or household, non-abrasive cleaner. DO NOT SPRAY SOLVENTS OR LIQUID DIRECTLY ON THE XRAY UNIT. BE CAREFUL NOT TO ALLOW SOLVENTS TO RUN OR DRIP into the PROVECTA V. This could cause damage to the PROVECTA V. Allow to air dry before plugging in or turning back on. Parts in Contact with the Patient’s Skin: To ensure proper cleaning of these parts, periodic disinfection with an approved EPA Non Corrosive Surface Disinfectant such as Birex® se is recommended. PRODUCT DESCRIPTION KEY COMPONENTS The PROVECTA V is comprised of the following components: a. STRAIGHT ARM AND SCISSOR ARM Horizontal straight arms are available in 16, 24, or 35-inch lengths to meet all possible reach requirements. The scissor arm has a double joint, enabling linear and upward extension. The tubehead remains balanced in all positions. b. TUBEHEAD WITH BEAM LIMITING DEVICE 70 kVp voltage and 8 mA current reduce exposure times and the amount of radiation absorbed by the patient. The tubehead is equipped with a beam limiting device with a 20 cm (8 inches) source to skin distance and a 6 cm (2 3/8 inches) beam diameter at the output. The tubehead is connected to the arm by means of a rotating contact, allowing 360-degree horizontal rotation and 285-degree vertical rotation. c. X-RAY ARM SUPPORT CONNECTION ENCLOSURE The X-ray Arm Support Connection Enclosure mounts to the Wall Mount Plate providing support for the straight and scissor arms and tubehead. the enclosure also provides electrical connection points for the PROVECTA V. d. WALL MOUNT PLATE - 16-INCH ON CENTER The Wall Mount Plate is designed to span two 2 by 4 wall studs that are spaced 16 inches apart to provide a secure wall-mounted surface for the installation of the PROVECTA V. e. REMOTE X-RAY CONTROL PANEL The Remote X-ray Control Panel is a microprocessor-controlled digital timer allowing both manual selection and 5 automatic selections of exposure times. Automatic selection further allows the operator to choose among 30 pre-set times according to the type of patient (Small, Medium or Large), the type of tooth and digital or film type. 44 fixed times can be selected manually and ranges from 0.02 seconds to 3.20 seconds. The timer automatically compensates for input line voltage variations within a range between110 and 132V. 8 PRODUCT DESCRIPTION AVAILABLE OPTIONS: The following lists additional configurations options available for the PROVECTA V. A2000V-16 A2000VR-16 A2000V-24 A2000VR-24 A2000V-35 A2000VR-35 A2290V-1 A2090 A2210 PROVECTA V X-Ray System with 16" Straight Arm PROVECTA V X-Ray System with Remote Control Assembly and 16" Straight Arm PROVECTA V X-Ray System with 24" Straight Arm PROVECTA V X-Ray System with Remote Control Assembly and 24" Straight Arm PROVECTA V X-Ray System with 35" Straight Arm PROVECTA V X-Ray System with Remote Control Assembly and 35" Straight Arm Kit,X-Ray Remote Control with Wall Mount & Cover (for outside examination room) Kit,X-Ray Remote wall plate exposure switch (for outside examination room) Mount Plate Assembly 16" center to center exterior dimensions PROVECTA V System Parts Identification STRAIGHT ARM SCISSOR ARM X-RAY ARM SUPPORT CONNECTION ENCLOSURE WALL PLATE, 16-INCH ON CENTER REMOTE CONTROL PANEL TUBEHEAD OPTIONAL REMOTE X-RAY EXPOSURE BUTTON EXPOSURE BUTTON ON REMOTE CONTROL PANEL 9 TECHNICAL SPECIFICATIONS Technical Features Equipment Manufacturer IEC 60601-1 Classification Degree of protection against ingress of water Air Techniques Inc. Class I with type B applied parts Ordinary Equipment is not suitable for use in the presence of flammable anesthetic mixture with air or with oxygen or nitrous oxide. Mode of operation Continuous operation with intermittent loading FDA 21 CFR Classification Class II Line voltage 110V to 132V Line frequency 60Hz Rated current 8.2 Amps Power consumption 0.8 kVA @ 120V Line Voltage Regulation <3% Apparent Resistance of Supply Mains 0.43 ohms Circuit Breakers 10 Amp Circuit Board Fuses 10A 250V Ceramic Fast Acting (0.25 x 1.25 IN) and 630mA 250V Fast Acting (5 mm x 20 mm) Preset exposure times from 0.02 to 3.20 seconds in 44 steps Anatomic selection 30 pre-set times Exposure control Microprocessor controlled exposure times, with automatic compensation for line voltage variations. See Selected Exposure Time and Corrected Actual Exposure Time note on page 11. Exposure Time at 120 Volts Timer Accuracy on the selected Exposure Time +/- 5% or +/- 33ms of selected Exposure Time (The value of the operator selected timer setting.) (whichever is greater) when the AC input is 120 volts. Corrected Exposure Time at Voltages Other than 120 Volts Timer Accuracy on the Corrected Exposure Time +/- 5% or +/- 33ms of Corrected Exposure Time (The actual time of exposure, indicated on the (whichever is greater) when the AC input is other than Control Panel during emission and determined by the 120 volts. internal algorithm as a function of the line voltage). The range of adjustment is 73% to 149% as the AC line voltage changes within the rated voltage range. X-ray Arm Support Connection Enclosure dimensions X-ray Remote Control dimensions 12.5” H x 9.75” W x 5” D 15.62” H x 9.00” W x 2.25” D Tubehead Features Rated output voltage High voltage circuit type Tubehead current Tubehead Power Total filtration Transformer insulation Cooling Pre-heating time Interval between exposure / duty cycle Beam Limiting Device Tubehead Assembly dimensions (less yoke) 70 kVp +/- 15% over 110 to 132 VAC 110 to 132 VAC Single phase, self-rectifying 8 mA +/- 2 mA 0.8 kw 2mm Aluminum equivalent @70 kVp Oil bath Convection 100ms 32 times X-ray time / 1: 32 Incorporated in Tubehead Assembly (non-removable) 5” H x 7.5” W x 15” D 10 TECHNICAL SPECIFICATIONS Minimum source to skin distance Maximum symmetrical radiation field @ 20 cm (X-ray beam diameter) Radiation leakage at 1 m Technical factors for radiation leakage 2 3/8” (6 cm) <50 mR/h, duty cycle 1: 32 70 kVp, 8mA, 3 seconds X-ray Tube Features Manufacturer Type Focal spot Inherent filtration Anode thermal capacity CEI- Bologna- Italy OCX/70-G 0.8 mm (IEC 336) 0.5 mm Al equivalent 6kJ Environmental Conditions Operating temperature range Operating relative humidity range Temperature range for transport and storage Max. relative humidity for transport and storage Min. atmospheric pressure for storage and transport 20 cm (8”) +50°F to +104°F (+10°C to +40°C) 30% - 75% -4°F to +158°F (-20°C to +70°C) <95 % non condensing 0.6 atmospheres Apparatus and Detachable Parts Weight Gross weight including packaging Net apparatus weight in standard configuration 16” fixed arm 24” fixed arm 35” fixed arm Scissor arm X-ray Remote Control 100 lbs. 95 lbs. 5 lbs. 7 lbs. 9 lbs. 24 lbs. 7.6 lbs. 9.6 lbs. 18.5 lbs. 17.2 lbs. X-ray Arm Support Connection Enclosure Tubehead 16” on center plate Selected Exposure time and Corrected Actual Exposure Time The ALLPRO X-ray Control Timer incorporates a special feature which automatically corrects the selected exposure time when the line voltage varies from 120 V. A change in the AC line voltage affects the peak voltage applied to the X-Ray tube and the value of high voltage significantly affects the spectrum of the radiation that ultimately affects the optical density of the imaging media. The purpose of the X-ray Control timer correction is to obtain the same nominal optical density on the imaging media independent of the AC line voltage variations from 110-132 Volts. The automatic correction of the exposure time works as follows: The internal voltage measuring circuit in the X-ray Control Timer continuously monitors the AC line voltage. When the user selects the desired exposure time and presses the X-ray Exposure button, the exposure time will be automatically adjusted based upon the measured AC line voltage. The corrected actual exposure time will be displayed until the X-ray Exposure button is released. NOTE: Electromagnetic interference between the equipment and other devices can occur. Do not use the equipment in close conjunction with sensitive devices, or devices creating high electromagnetic disturbances. 11 TECHNICAL SPECIFICATIONS The corrected exposure is calculated applying a correction factor to the selected exposure time, based on an empirical law that correlates the dose with the high voltage peak and consequently with the line voltage. The qualitative relationship between the multiplying factor and the line voltage is shown in the following graph: 12 TECHNICAL SPECIFICATIONS TECHNICAL FACTORS MEASURING METHOD kVp The kVp is defined as the average of the peak high voltage after pre-heating time and settling. The kVp is assessed with a non-invasive instrument having a ± 1% accuracy at the nominal AC input line voltage. mA The output current is defined as the average value of the current after pre-heating time and settling. The output current can be measured by attaching a DC voltmeter (0-10 VDC) to the connections behind the circular cap on the side of the Tubehead yoke. The correlating relationship is 1 volt per mA of tube current. t The exposure time is defined as the time measured with a non-invasive meter. Accuracy is verified by using an Innovision 8000 mAx instrument. X-RAY TUBE CURVES OCX / 70-G Emission characteristics 13 X-RAY TUBEHEAD KEY COMPONENTS X-ray Tubehead Assembly X-Ray Beam Axis Guide Focal Point Guide Yoke Angular Alignment Gauge 14 X-RAY TUBEHEAD SPECIFICATIONS Filament Characteristics Loading 15 X-RAY TUBEHEAD SPECIFICATIONS Cooling Cooling Curve of Tubehead 16 SYSTEM ACTIVATION PROVECTA V TURN-ON PROCEDURE: Turn on the PROVECTA V by performing the following procedure: 1. Place the Main Power Switch located on the bottom of the Remote X-Ray Control Panel in the ON ( I ) position. 2. Observe that the system self check function initiates and the digital display displays following sequence : a. API b. Display Test 8.8.8. c. Firmware Version (e.g. A) d. Last setting e.g. (e.g. 0.25) Note: See System Error Messages,pages 46 through 49, for a listing and description of possible start-up error codes. 3. Once the self check function completes successfully, the system starts in the Lower Feline, and Lower Canine Incisors and Exotic, Medium and Film Processor modes and is ready for operation. Verify associated indicator illumination and exposure time display. Note: When power is turned off, the system retains the last selected manual and preset modes. Upon turn on, the system resumes operation as set prior to turning off. Remote X-Ray Control Panel MAIN POWER SWITCH 17 REMOTE X-RAY CONTROL PANEL FUNCTIONS PULSE Time Exposure Mode Indicator Lamp Digital Display SECONDS Time Exposure Mode Indicator Lamp TOOTH SELECT Mode Select Button Increase Select Button 8.88 Decrease Select Button Lower Feline, and Lower Canine Incisors and EXOTIC Tooth Select Indicator Lamp Upper Canine Tooth Select Indicator Lamp Upper Canine MOLAR, PRE-MOLAR and Lower Canine Tooth Select Indicator Lamp Upper Incisors (Canine) Tooth Select Indicator Lamp Upper Feline, and Lower Canine PRE-MOLAR and MOLAR Tooth Select Indicator Lamp MEDIUM Mode Select Indicator Lamp SMALL Mode Select Indicator Lamp LARGE Mode Select Indicator Lamp SIZE Mode Select Button FILM Processing Selection Indicator Lamp Digital Processing Selection Indicator Lamp ACTIVE X-RAY EMISSION CAUTION Indicator Lamp DIGITAL/FILM Mode Select Button Note: All automatic settings are confirmed by the corresponding LED illumination. 18 REMOTE X-RAY CONTROL PANEL FUNCTIONS X-Ray Control Panel - All functional operating control of the PROVECTA V X-Ray is provided by the X-ray Control Panel allowing selection of both automatic and manual of exposure time modes. The 5 automatic selection functions allow the operator to further choose among 30 pre-set times according to the patient type (Small, Medium or Large), tooth type and media type (digital or film). An audible signal sounds during any X-ray emission. In addition, this signal is heard to confirm keypad button selection and when certain errors occur. The location of each panel control and indicator is shown in the preceding illustration, while the function of each is summarized below and described on the following pages. DIGITAL DISPLAY Panel 8.88 A 3-character digital display that shows the selected time value (pulses or seconds) during operational phases as follows: 1. Exposure Preparation Phase - displays the selected exposure time in both automatic or manual modes. 2. Exposure Phase - displays the corrected exposure time adjusted to compensate for variations in the input AC line voltage. 3. Cooling Phase - displays remaining tubehead cooling time. INCREASE Select Button A membrane keypad switch that disables all automatic settings and allows the user to manually increase exposure time as necessary. DECREASE Select Button A membrane keypad switch that disables all automatic settings and allows the user to manually decrease exposure time as necessary. PULSE TIME EXPOSURE MODE Indicator Lamp SECONDS TIME EXPOSURE MODE Indicator Lamp TOOTH SELECT Mode Select Button PULSES An LED indicator lamp that illuminates green to confirm selection of the PULSE time mode for Digital Display operation. SECONDS An LED indicator lamp that illuminates green to confirm selection of the SECONDS time mode for Digital Display operation. TOOTH SELECT A membrane keypad switch that allows sequential mode selection of preset exposure time among five different teeth categories as follows: Lower Feline, Lower Canine Incisors and Exotic Tooth Select Indicator Lamp An LED indicator lamp that illuminates green to confirm selection of preset exposure time for the Lower Feline, Lower Canine Incisors and Exotic tooth categories. Upper Incisor (Canine) Tooth Select Indicator Lamp An LED indicator lamp that illuminates green to confirm selection of preset exposure time for the Canine Upper Incisor tooth category. Upper Feline, and Lower Canine Pre-Molar and Molar Tooth Select Indicator Lamp An LED indicator lamp that illuminates green to confirm selection of preset exposure time for the Upper Feline, Canine Lower Pre-Molar and Canine Lower Molar tooth categories. Upper Canine Molar, Pre-Molar and Lower Canine Tooth Select Indicator Lamp An LED indicator lamp that illuminates green to confirm selection of preset exposure time for the Canine Upper Molar, Canine Upper Pre-Molar and Lower Canine tooth categories. Upper Canine Tooth Select Indicator Lamp SIZE Mode Select Button SMALL MODE Indicator Lamp SIZE SMALL An LED indicator lamp that illuminates green to confirm selection of preset exposure time for the Upper Canine tooth category. A membrane keypad switch that toggles between selection of preset exposure time for SMALL, MEDIUM and LARGE exposure modes as follows: An LED indicator lamp that illuminates green to confirm selection of preset exposure time for the SMALL exposure mode. MEDIUM MODE Indicator Lamp MEDIUM An LED indicator lamp that illuminates green to confirm selection of preset exposure time for the MEDIUM exposure mode. LARGE MODE Indicator Lamp LARGE An LED indicator lamp that illuminates green to confirm selection of preset exposure time for the LARGE exposure mode. DIGITAL/FILM Mode Select Button DIGITAL FILM A membrane keypad switch that toggles between selection of preset exposure time for DIGITAL and FILM exposure modes as follows:. DIGITAL Processing Indicator Lamp An LED indicator lamp that illuminates green to confirm selection of preset exposure time for the DIGITAL processing mode. FILM Processing Indicator Lamp An LED indicator lamp that illuminates green to confirm selection of preset exposure time for the FILM processing mode (D-speed film). ACTIVE X-RAY EMISSION CAUTION Indicator Lamp An LED indicator lamp that illuminates yellow to alert the user that xray radiation is present during PROVECTA V current operation. 19 KEYPAD OPERATION INCREASE SELECT BUTTON DECREASE SELECT BUTTON The INCREASE and DECREASE Select Buttons are membrane keypad switches that allow the user to disable the automatic operating mode default settings (Time Units, Media Type, Patient Size, and Tooth Selection) and manually increase or decrease exposure time as necessary to any of the 44 values shown below in seconds or pulses. When a manual exposure time setting is desired, press the associated button (Increase or Decrease) once to display the last manually set exposure time. Manually change the time setting in one increment or decrement by releasing the button and then pressing once. Rapid advancement through the 44 time values is accomplished by keeping the button depressed. Release button to set desired exposure time. Return to the last automatic mode default time settings by pressing any Mode Selection Button (Digital/Film, Size, Tooth Select). TIME VALUES IN SECONDS 0.02 0.04 0.05 0.06 0.08 0.10 0.12 0.14 0.15 0.16 0.18 0.20 0.22 0.23 0.25 0.27 0.28 0.30 0.32 0.33 0.35 0.36 0.38 0.40 0.42 0.43 0.45 0.47 0.50 0.55 0.60 0.63 0.70 0.80 0.90 1.00 1.25 1.30 1.40 1.60 2.00 2.50 3.00 3.20 TIME VALUES IN PULSES 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 30 33 36 38 42 48 54 60 75 78 84 96 120 150 180 192 TOOTH SELECT MODE SELECT BUTTON A membrane keypad switch located in the center of the tooth symbols allows sequential mode selection of preset default exposure times for each of the five indicated teeth types. Each time the TOOTH/SELECT Mode Select Button, shown below, is pressed, the associated LED lamp illuminates and an audible signal sounds to indicate selection. Default preset times are based on D-speed film. The five selectable default preset exposure times correlating to tooth categories are as follows: TOOTH SELECT Mode Select Button Upper Canine Lower Feline, Lower Canine Incisors and Exotic 0.10 seconds Upper Incisor (Canine) 0.12 seconds Upper Feline, Lower Canine Pre-Molar and Canine Molar 0.15 seconds Upper Canine Molar, Canine Pre-Molar and Lower Canine 0.20 seconds Upper Canine 0.25 seconds LED Factory setting 0.25 sec. Lower Feline, Lower Canine Incisors and Exotic Factory setting 0.10 sec. Upper Incisor (Canine) LED Factory setting 0.12 sec. Upper Canine Molar, Canine Pre-Molar and Lower Canine Upper Feline, Lower Canine Pre-Molar and Canine Molar Factory setting 0.20 sec. Factory setting 0.15 sec. 20 KEYPAD OPERATION SIZE MODE SELECT BUTTON A membrane keypad switch that switches among the selection of SMALL, MEDIUM and LARGE preset exposure time exposure modes. Each time the SIZE Mode Select Button is pressed, the selection of Small, Medium and Large modes is made with the resulting change in exposure time. The associated LED lamp illuminates and an audible signal sounds to indicate selection. MEDIUM factory setting is 1.00 or 100% SIZE Mode Select Button The factory setting for the SMALL reduction is 0.85 or 85%. The factory setting for the LARGE increase is 1.15 (115%). DIGITAL/FILM MODE SELECT BUTTON A membrane keypad switch that toggles between automatic exposure settings for specific media types commonly used with the PROVECTA V. The user can select between D-speed film and digital radiology processing. When the FILM mode is selected the system is set to process film at 100% exposure time of D-speed film. When set in the DIGITAL mode, the exposure time is reduced to values suitable for the use of digital phosphor plate scanners such as A/T ScanX. DIGITAL/FILM Mode Select Button Film factory setting 1.00 or 100% for Dspeed film. Digital X-Ray factory-set reduction is 0.50 or 50%. An audible signal is sounded each time the DIGITAL/FILM Mode Select Button is depressed and the associated LED is illuminated to indicate selection. DEFAULT EXPOSURE TIME SETTINGS The following table lists the factory-set default setting combinations available by the operation of the TOOTH SELECT, SIZE and DIGITAL/FILM Mode Select Buttons. PROVECTA V Available Preset Exposure Time Setting Combinations SMALL MEDIUM LARGE TOOTH SELECT SECONDS PULSES SECONDS PULSES SECONDS PULSES SECONDS PULSES SECONDS PULSES SECONDS PULSES 0.04 3 0.09 5 0.05 3 0.10 6 0.06 3 0.12 7 0.05 3 0.10 6 0.06 4 0.12 7 0.07 4 0.14 8 0.06 4 0.13 8 0.08 5 0.15 9 0.09 5 0.17 10 0.09 5 0.17 10 0.10 6 0.20 12 0.12 7 0.23 14 0.11 6 0.21 13 0.13 8 0.25 15 0.14 9 0.29 17 21 KEYPAD OPERATION: CUSTOM PROGRAMMING Note: When power is turned off, the system retains the last selected manual and preset modes. Upon turn on, the system resumes operation as set prior to turning off. CHANGING DEFAULT SETTINGS TO CUSTOMIZE VALUES The PROVECTA V factory default settings are based on D-speed film exposure with the digital display set to show time values in seconds. These defaults allow automatic operation of the system by the selection of the TOOTH SELECT, SIZE and DIGITAL/FILM Mode Select Buttons. Although this automatic operation is adequate for most applications, the PROVECTA V may be customized to meet the needs of a specific application or individual preference. Customize one or all of the six default settings listed below as necessary by performing the procedures provided by the following paragraphs. When all changes have been accomplished, resume operation of the PROVECTA V per System Operation instructions provided on page 25. 1. Changing Display to Pulses 2. Changing Display to Seconds 3. Changing Default Tooth Setting 4. Changing Default Large and Small Multiplier Ratio 5. Changing Film Multiplier Ratio Settings 6. Changing Digital Multiplier Ratio Settings CHANGING DISPLAY TO PULSES The PROVECTA V is set at the factory to display exposure times in seconds. Perform the following procedure to change the Digital Display panel to show time values in pulses: 1. Turn the Power Switch OFF (0) and then to ON ( I ) again. 2. When API is shown in the Digital Display panel, press and hold the INCREASE () Select Button until PUL (pulses) appears in the display window. Verify that the PULSES indicator illuminates on the control panel. 3. Release the INCREASE () Select Button. CHANGING DISPLAY TO SECONDS When the PROVECTA V was changed to display in pulses and there is a need to change the time value back to seconds, perform the following procedure: 1. Turn the Power Switch OFF (0) and then to ON ( I ) again. 2. When API is shown on the Digital Display panel, press and hold the INCREASE () Select Button until SEC (seconds) appears in the display window. Verify that the SECONDS indicator illuminates on the control panel. 3. Release the INCREASE () Select Button. 22 KEYPAD OPERATION: CUSTOM PROGRAMMING CHANGING DEFAULT TOOTH SETTINGS (0.02 to 3.20 Seconds) The PROVECTA V provides 5 tooth settings and is factory set to start operation in the LOWER Feline, Canine L-INCISORS and EXOTIC tooth select mode. Perform the following procedure to change one or all of the default tooth settings: 1. Turn the Power Switch OFF (0) and then to ON ( I ) again. 2. Verify that upon device turn on, the firmware version (e.g. A) is displayed on the Digital Display panel as the second display item. TOOTH 3. When the device firmware version (e.g. A) is displayed, press the SELECT TOOTH SELECT Mode Select Button to enter the default tooth setting programming mode. 4. When SEL is shown on the Digital Display panel, press the INCREASE () Select Button. One of the five tooth select indicator lamps will flash and the Digital Display panel will show the current programmed default exposure setting for the selected tooth. 5. Within ten seconds, repeatedly press the INCREASE () Select Button until the desired tooth category is selected. TOOTH 6. Press the SELECT TOOTH SELECT Mode Select Button to confirm selection. 7. The Digital Display panel will flash together with the indicator lamp of the selected tooth to indicate the default setting can be changed. 8. Within ten seconds, press the INCREASE () or DECREASE () Select Button to the desired default setting. TOOTH 9. Within ten seconds, press the SELECT TOOTH SELECT Mode Select Button to exit the default tooth setting program. 10. Repeat steps 1- 9 until all desired tooth setting changes have been made. CHANGING DEFAULT LARGE (1.00 to 2.00) AND SMALL (0.05 to 1.00) MULTIPLIER RATIOS The PROVECTA V default Large setting is 115% (1.15 ) of the Medium setting while the Small setting is 85% (0.85 )of Medium. Perform the following procedure to change the default LARGE and SMALL mode multiplier ratio settings: 1. Turn the Power Switch OFF (0) and ON ( I ) again. 2. Verify that upon device turn on, the firmware version (e.g. A) is displayed on the Digital Display panel as the second display item. SIZE 3. When the firmware version (e.g. A) is displayed on the Digital Display panel, press the SIZE Mode Select Button to enter the LARGE multiplier ratio programming mode. TOOTH 4. When the L is shown on the Digital Display panel, press the SELECT TOOTH SELECT Mode Select Button to confirm the LARGE mode selection. (See step 8 for SMALL mode.) 5. Verify that the Digital Display panel shows the factory setting of 1.15 for the LARGE mode multiplier ratio setting. 6. Within 10 seconds press the INCREASE () or DECREASE () Select Button to set the desired multiplier ratio with respect to D-speed film. TOOTH 7. Within 10 seconds of selecting the desired multiplier ratio press the SELECT TOOTH SELECT Mode Select Button to select change. 8. With the L shown on the Digital Display panel, press the INCREASE () Select Button to confirm the SMALL mode selection (S is shown on the Digital Display panel). TOOTH 9. When the S is shown on the Digital Display panel, press the SELECT TOOTH SELECT Mode Select Button to confirm the SMALL mode selection and verify that the Digital Display panel shows the factory setting of 0.85 for the SMALL mode multiplier ratio setting. 10. Set the desired SMALL mode multiplier ratio with respect to D-speed film by performing steps 6 and 7 above. 23 KEYPAD OPERATION: CUSTOM PROGRAMMING Note: If using film other than D-Speed film, change the film multiplier ratio as recommended by the film manufacturer prior to use. CHANGING FILM MULTIPLIER RATIO SETTING (0.05 to 1.50) The PROVECTA V default FILM setting is 100% (1.00) of D-speed film exposure time. Perform the following procedure to change the default film multiplier ratio setting: 1. Turn the Power Switch OFF (0) and ON ( I ) again. 2. When the firmware version (e.g. A) is displayed, press the DIGITAL/FILM Mode Select Button to enter the DIGITAL/FILM multiplier ratio programming mode. TOOTH 3. When FIL is shown on the Digital Display panel, press the SELECT TOOTH SELECT Mode Select Button to confirm the film selection mode. 4. Verify that the Digital Display panel shows the factory setting of 1.00 (100% of D-speed film exposure time). 5. Within 10 seconds press the INCREASE () or DECREASE () Select Button to set the desired multiplier ration with respect to D-speed film exposure time. TOOTH 6. Within 10 seconds press the SELECT TOOTH SELECT Mode Select Button to select. CHANGING DIGITAL MULTIPLIER RATIO SETTING (0.05 to 1.50) The PROVECTA V default DIGITAL setting is 50% (0.50) of the FILM setting (D-speed film exposure time). Perform the following procedure to change the default digital media multiplier ratio settings: 1. Turn the Power Switch OFF (0) and ON ( I ) again. 2. When the firmware version (e.g. A) is displayed, press the enter the DIGITAL/FILM multiplier ratio programming mode. 3. When FIL is shown on the Digital Display panel, press the INCREASE () Select Button. 4. Verify that the Digital Display panel shows DIG to confirm the film selection mode. 5. Press the 6. Verify that the Digital Display panel shows the factory setting of 0.50. Within 10 seconds press the INCREASE () or DECREASE () Select Button to set the desired multiplier ratio with respect to D-speed film exposure time. TOOTH Within 10 seconds press the SELECT TOOTH SELECT Mode Select Button to select. 7. 8. TOOTH SELECT DIGITAL/FILM button to TOOTH SELECT Mode Select Button to confirm selection. FACTORY RESET Performing factory reset returns all operating values and settings of the PROVECTA V to the values set at the factory during manufacture. Refer to the PROVECTA V Available Preset Exposure Time Setting Combinations table on page 21 for a listing of the factory-set default exposure settings. Make sure to write down all manually customized settings prior to performing the factory reset sequence. Any settings changed during operation will be lost upon completion of reset. Reset all operating values and settings by performing the following procedure. 1. Turn the Power Switch OFF (0) and then to ON ( I ) again. 2. When the initials API appear in the Digital Display panel, press and hold the DECREASE () Select Button until FAC (Factory Reset) appears in the display. 3. Verify that the display shows 0.10 seconds and the SECONDS, LOWER Feline, Canine LINCISORS and EXOTIC, MEDIUM, and FILM LED indicator lamps are illuminated. The PROVECTA V is now reset to operate with the factory-set default settings. 24 SYSTEM OPERATION SETTING THE EXPOSURE The PROVECTA V is normally operated in the automatic operation mode where the selection combinations of the TOOTH SELECT, SIZE and DIGITAL/FILM Mode Select Buttons provide 30 separate default exposure settings for specific exposure applications. These preset exposure time values (seconds or pulses) associated with the mode select button setting combinations are displayed as shown in the table below. The suggested exposure times are based on D-speed film exposure. Note: Keypad button selection is confirmed by the sounding of an audible signal and illumination of the corresponding indicator lamp . Automatic Exposure Time Selection - Perform the following procedure to set the PROVECTA V to operate in the automatic mode using preset exposure time values. 1 TOOTH SIZE Press any Mode Selection Button SELECT to initiate the automatic exposure mode of operation. 2. SIZE Press the SIZE Mode Select Button to select the patient size parameter. Verify that the corresponding SMALL Mode, MEDIUM Mode or LARGE Mode indicator lamp illuminates and audible signal sounds for the desired preset exposure setting selection. 3. Press the DIGITAL/FILM Mode Select Button to select the type of media processing. Verify that the corresponding indicator lamp Digital Mode or Film Mode illuminates and audible signal sounds for the desired preset exposure setting selection. 4. Press the SELECT TOOTH SELECT Mode Select Button to sequentially advance through the 5 tooth categories until the corresponding indicator lamp illuminates for the desired tooth type selection. The resultant exposure setting is shown on the Digital Display in seconds or pulses as required. (Tooth Select, Digital/Film, Size) TOOTH PROVECTA V Available Preset Exposure Time Setting Combinations SMALL MEDIUM LARGE TOOTH SELECT SECONDS PULSES SECONDS PULSES SECONDS PULSES SECONDS PULSES SECONDS PULSES SECONDS PULSES 0.04 3 0.09 5 0.05 3 0.10 6 0.06 3 0.12 7 0.05 3 0.10 6 0.06 4 0.12 7 0.07 4 0.14 8 0.06 4 0.13 8 0.08 5 0.15 9 0.09 5 0.17 10 0.09 5 0.17 10 0.10 6 0.20 12 0.12 7 0.23 14 0.11 6 0.21 13 0.13 8 0.25 15 0.14 9 0.29 17 25 SYSTEM OPERATION NOTE: If using other than D-Speed film, change the film multiplier ratio as recommended by the film manufacturer prior to use (refer to the section on Custom Programming on page 22). The PROVECTA V also allows the user to manually set exposure settings to meet the needs of specific patient applications or individual user preference. Using the INCREASE () or DECREASE () Select Buttons, the manual exposure selection mode enables the user to set exposure time values (seconds or pulses) as required. These user-set exposure time values offer 44 specific value settings shown in the tables below. Manual Exposure Time Selection- Perform the following procedure to set the PROVECTA V to operate in the manual exposure selection mode using user-set exposure time values. 1. Press either the INCREASE () or DECREASE () Select Buttons to initiate the manual exposure selection mode of operation. The Digital Display panel shows the last manually set exposure time. Note: 2. 3. When set in the manual exposure selection mode, exposure time is increased by pressing the INCREASE () Select Button and decreased by pressing the DECREASE () Select Button. The resultant exposure setting is shown on the Digital Display panel in seconds or pulses. Press either the INCREASE () or DECREASE () Select Buttons once to manually change the time setting in one increment or decrement. Rapid advancement through the entire range of 44 time values is accomplished by keeping the button depressed. Release button to set desired exposure time. Verify that the desired exposure time value is shown on the Digital Display panel. TIME VALUES IN SECONDS 0.02 0.04 0.05 0.06 0.08 0.10 0.12 0.14 0.15 0.16 0.18 0.20 0.22 0.23 0.25 0.27 0.28 0.30 0.32 0.33 0.35 0.36 0.38 0.40 0.42 0.43 0.45 0.47 0.50 0.55 0.60 0.63 0.70 0.80 0.90 1.00 1.25 1.30 1.40 1.60 2.00 2.50 3.00 3.20 TIME VALUES IN PULSES 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 30 33 36 38 42 48 54 60 75 78 84 96 120 150 180 192 26 SYSTEM OPERATION EXPOSURE EMISSION PROCEDURE 1. Introduce the film, plate or sensor into the patient’s mouth according to the chosen technique (bisecting or parallel). 2. Move the Tubehead beam limiter near the patient and direct it exactly towards the tooth under examination. 3. Arrange the Tubehead with an angle suitable for the required exposure and positioning. 4. Select exposure time value via either the automatic or manual exposure time operating mode as follows: a. Set automatic exposure settings for specific patient size, media type and tooth type per procedure of page 25. b. Set manual exposure settings as necessary per procedure of page 26. 5. Move as far away as the hand control cable allows, in a direction opposite to the X-ray beam emission while maintaining visual contact with the X-ray Control panel and the patient. 6. Press the X-ray Exposure button and hold it depressed throughout the entire exposure. 7. The X-ray emission condition is indicated by the illumination of the ACTIVE X-RAY EMISSION CAUTION Indicator Lamp and an audible signal sounding. 8. At the end of the exposure, the system starts the Tubehead cooling cycle, which is 32 times the exposure time. The remaining cooling time is shown on the Digital Display panel. WARNINGS: 1. The X-ray Exposure button is a “Dead Man” control. The button must be kept depressed throughout the entire exposure. Releasing the button before the end of the exposure terminates the emission. When this happens the the exposure time display flashes, while the other panel indicators extinguish. The flashing display lasts until either the INCREASE () or DECREASE () button is depressed. At that time, system operation resumes in the tubehead cooling cycle and the display shows remaining cooling time. 2. If the combination of low AC line voltage and long exposure time results in a corrected exposure time of more than 4.00 seconds (240 pulses), the system will inhibit the exposure. In this case, the display shows the corrected exposure time. Reset the unit by pressing the INCREASE () or DECREASE () button and reduce the exposure setting so as not exceed 4.00 seconds (240 pulses). 3. If the AC line voltage reaches minimum/maximum operating limits, the display shows the following: Error Code “LLL” denotes that the voltage is too low (under 106 Volts). Error Code“HHH” denotes that the voltage is too high (over 134 Volts). Contact your ALLPRO authorized Dealer for service when either code is displayed. READING THE EXPOSURE COUNTER 1. Simultaneously press and hold the INCREASE () and DECREASE () Select Buttons for more than 3 seconds. 2. When P20 is shown on the Digital Display panel, release the INCREASE () and DECREASE () Select Buttons. 3. Press either the INCREASE () or DECREASE () Select Button to display the number of exposures in thousands (e.g. 004=4,000). 4. Press the INCREASE () or DECREASE () Select Button to display the number of exposures up to 999 (e.g. 354). 5. The total number of exposures is 4,000 + 354 = 4,354. TOOTH 6. Press the SELECT TOOTH SELECT Button to exit the exposure counter mode. 27 X-RAY SYSTEM REACH AND STORAGE DIMENSIONS Straight Arm Length Options A B C D E F G H J K L M 35.00 49.25 74.50 80.00 35.50 6.25 29.00 7.00 12.75 5.00 47.44 8.52 52.25 24.00 37.75 63.00 68.50 35.50 6.25 29.00 7.00 12.75 5.00 35.94 8.52 52.25 16.00 30.50 55.75 61.25 35.50 6.25 29.00 7.00 12.75 5.00 28.69 8.52 52.25 Storage Dimensions Width Height Depth 47.44 51.25 5.12 Remote X-Ray Control Panel Dimensions X-Ray Arm Dimension Options J M L K 28 RADIOGRAPH TROUBLESHOOTING CHECKS AND CORRECTION OF POSSIBLE FAULTS IN DENTAL RADIOGRAPHS TYPICAL DEFECTS OF INTRA-ORAL RADIOLOGY - Light radiographs: Possible causes: - Insufficient exposure to X-ray (short time) - Poor film processing technique - Insufficient development time - Deteriorated developer / contaminated developer - Developer temperature below recommended value - Incorrect developing fluid dilution - Expired film - Dark radiographs: Possible causes: - Excessive exposure to X-ray (long time) - Film exposed to light / secondary X-ray exposure - Poor film processing techniques - Contaminated developer - Excessive development time - Developer temperature above recommended value - Incorrect developing fluid dilution - Expired film - Blurred radiographs (details not visible): Possible causes: - The patient moved - The tubehead moved - Radiographs with herringbone marks: Some intra-oral films are provided with a thin lead foil having a herringbone type mark on its lower side. These films may be exposed to radiation on one side only. If the film is exposed on the wrong side, the lead foil will absorb a large amount of radiation during exposure. The result will be a lighter radiograph and the film will show herringbone marks. 29 RADIOGRAPH TROUBLESHOOTING PARTIALLY EXPOSED RADIOGRAPHS Possible causes: - X-ray Tubehead positioned incorrectly with respect to the imaging media - Low developer fluid level, with resulting partial film development - Two or more films placed against each other during development CLOUDED RADIOGRAPHS Possible causes: - Outdated film (check expiration date) - Film accidentally exposed to X-ray - Film accidentally exposed to other natural or artificial light sources - Improper development technique RADIOGRAPHS SHOWING A BLACK LINE This line appears when the film is excessively folded RADIOGRAPHS SHOWING SIGNS OF ELECTROSTATIC ELECTRICITY When a film is compressed too much and the air is dry, static electricity may occur, which displays black marks. RADIOGRAPHS WITH CHEMICAL SPOTS Developer or fixer fluid spattered on the film before development and fixing procedures produces spots on the radiograph; such spots are: - dark, when caused by developer fluid - light when caused by fixer fluid RADIOGRAPHS WITH EMULSION COMING OFF If the film is kept in a hot water bath too long (e.g. throughout the whole night), or in the developer solution too long, the emulsion may become softer and partially come off the film base. After development, the film will show scratches. TYPICAL DEFECTS CAUSED BY INCORRECT POSITIONING - Radiographs with elongated or shortened image: The main beam is not perpendicular to the bisecting angle formed by the tooth longitudinal axis and the film. - Radiographs with stretched out tip of tooth Probably caused by excessive film folding or bending inside patient’s mouth. 30 SYSTEM MAINTENANCE To ensure the safe and effective performance of the PROVECTA V follow the proper operation as described by the instructions in this manual. Periodic maintenance is determined by observations made by the operator and/or an ALLPRO Imaging Authorized Dealer Field Service Technician. Refer to the PROVECTA V Intraoral X-Ray System with Remote Control Installation & Service Manual for information pertaining to the installation and servicing of the PROVECTA V Intraoral X-Ray System with Remote Control. WARNING: PRE-OPERATION CHECKS MUST BE PERFORMED BEFORE THE PROVECTA V SYSTEM. IF ANY PROBLEMS OR ANOMALIES ARE NOTED, DO NOT OPERATE THE UNIT AND CONTACT YOUR ALLPRO IMAGING AUTHORIZED DEALER FOR SERVICE. The operator should observe before any operating session that: - the unit is solidly secure on its wall mounting support - the scissor arms are correctly balanced and stable - the Tubehead is free from oil residues - the Exposure switch cable is not damaged - there is no external damage on the X-Ray unit - the scissor arms are correctly balanced and stable - the labels and keypad surface are properly secured SYSTEM ERROR MESSAGES ERROR MESSAGES The PROVECTA V has been designed with safety features to protect the patient and operator in case of failure of an electrical component. The system automatically checks for errors and will report a malfunction by means of an error code on the Remote X-Ray Control Panel display. These error codes are shown in the table on the following pages. The table lists the code, identifies the potential source of the problem and recommends the corrective action required by the operator to clear the error. WARNING: WHEN A PROBLEM IS DETECTED FOLLOW THE IDENTIFIED "ACTION REQUIRED" IN THE ERROR CODE TABLE BELOW AND CALL YOUR ALLPRO IMAGING AUTHORIZED DEALER FOR SERVICE AS REQUIRED. FOR PROPER MAINTENANCE AND SAFETY, SERVICE MUST BE PERFORMED ONLY BY AN ALLPRO IMAGING AUTHORIZED DEALER SERVICE TECHNICIAN. WARNING: IMMEDIATELY REMOVE POWER IF ERROR CODE E02 APPEARS ON THE ERROR CODE E02 DISPLAY AND A PULSATING AUDIBLE TONE IS HEARD. X-RAY GENERATION IS CONTINUOUS AND THE POWER SWITCH MUST BE SET TO THE OFF (0) POSITION TO END EMISSIONS. CALL YOUR ALLPRO IMAGING AUTHORIZED DEALER FOR SERVICE. 31 SYSTEM ERROR MESSAGES ERROR CODE E01 AUDIBLE SIGNAL None SOURCE OF ANOMALY ACTION REQUIRED The X-ray relay is closed when power is turned on. Test with Tubehead Simulator. Reset power by turning the Main Power Switch OFF (0) and ON (I). If E01 error repeats, replace the X-ray Control Panel. X-RAY GENERATION IS INHIBITED. Voltage is present at the Tubehead when power is turned ON (I). E02 Continuously This is a result of multiple component Pulsating failures. X-RAY GENERATION E05 None IS CONTINUOUS. None X-RAY GENERATION IS INHIBITED. The INCREASE button () is depressed when power is turned ON (I). E08 None X-RAY GENERATION IS INHIBITED. The DECREASE () button is depressed when power is turned ON (I). E09 None X-RAY GENERATION E10 None Test with Tubehead Simulator. Reset power by turning the Main Power Switch OFF (0) and then ON (I). If E02 error repeats, replace the X-ray Control Panel. Check that the switches are not The main or Remote Exposure Switch is depressed. depressed when power is turned ON (I). Reset power by turning the Main Power switch OFF (0) and then ON (I). If E05 error repeats check the X-RAY GENERATION IS INHIBITED. switches and cables for defects or short circuits. Replace Switch assembly. The DIGITAL/FILM button is depressed when power is turned ON (I). E07 IMMEDIATELY REMOVE POWER. TURN THE POWER SWITCH TO OFF (0). IS INHIBITED. The SIZE button is depressed when power is turned ON (I). X-RAY GENERATION IS INHIBITED. 32 Check that the Control Panel switch is not depressed. Reset power by turning the Main Power switch OFF (0) and then ON (I). If E07 error repeats, replace Membrane Assembly. Check that the Control Panel switch () is not depressed. Reset power by turning the Main Power switch OFF (0) and then ON (I). If E08 error repeats, replace Membrane Assembly. Check that the Control Panel switch () is not depressed. Reset power by turning the Main Power switch OFF (0) and then ON (I). If E09 error repeats, replace Membrane Assembly. Check that the Control Panel switch is not depressed. Reset power by turning the Main Power switch OFF (0) and then ON (I). If E10 error repeats, replace Membrane Assembly. SYSTEM ERROR MESSAGES ERROR CODE E12 E20 AUDIBLE SIGNAL None None SOURCE OF ANOMALY Check that the Control Panel switch is The TOOTH SELECT button is not depressed. depressed when power is turned ON (I). Reset power by turning the Main Power switch OFF (0) and then ON (I). If E12 error repeats, replace X-RAY GENERATION IS INHIBITED. Membrane Assembly The X-ray relay is not closing within the required time after the Exposure Switch is pressed. X-RAY GENERATION E21 IS INHIBITED. The X-ray relay is not opening within the required time after the exposure is Continuously completed. Pulsating X-RAY GENERATION SAFETY TIMER. E22 None ACTION REQUIRED IS TERMINATED BY A The X-ray triac is not closing within the required time after the Exposure Switch is pressed. X-RAY GENERATION IS INHIBITED. Test with Tubehead Simulator. Reset power by turning the Main Power Switch OFF (0) and ON (I). If E20 error repeats, replace the X-ray Control Panel. IMMEDIATELY REMOVE POWER. TURN THE POWER SWITCH TO OFF (0). Test with Tubehead Simulator. Reset power by turning the Main Power switch OFF (0) and then ON (I). If E21 error repeats, replace the X-ray Control Panel. Test with Tubehead Simulator. Reset power by turning the Main Power Switch OFF (0) and ON (I). If E22 error repeats, replace the X-ray Control Panel. IMMEDIATELY REMOVE POWER. TURN THE POWER SWITCH TO OFF (0). E23 Test with Tubehead Simulator. Reset power by turning the Main Power switch OFF (0) and then ON (I). The X-ray triac is not opening within the If E23 error repeats, replace the X-ray required time after the exposure is Control Panel. completed or the Tubehead is not If E23 error does not repeat, turn the Continuously connected to the X-ray control. Main Power switch OFF (0) and verify Pulsating X-RAY GENERATION IS TERMINATED BY A that the resistance at the wires going to the Tubehead is nominally 1.5 ohms. SAFETY TIMER. If not, remove the Tubehead and verify the resistance at the Tubehead connector is nominally 1.5 ohms. If not, replace the Tubehead. If yes, check the wiring in the fixed and scissor arms 33 SYSTEM ERROR MESSAGES ERROR CODE E25 AUDIBLE SIGNAL None SOURCE OF ANOMALY ACTION REQUIRED The X-ray emission exceeds the Test with Tubehead Simulator. programmed time and is terminated by the hardware Reset power by turning the Main timer at nominally 5.5 seconds. Power Switch OFF (0) and ON (I). If E25 error repeats, replace the X-ray X-RAY GENERATION IS TERMINATED Control Panel. BY A SAFETY TIMER. LLL HHH Greater than 4.00 seconds or 240 pulses Flashing Exposure Time None None None None Press the INCREASE () or The AC input line voltage is less DECREASE () buttons to reset the than 106 Volts when the display. Exposure Switch is pressed. If LLL error repeats, check the AC line X-RAY GENERATION IS INHIBITED. voltage at the input terminals to the Xray Control Panel The AC input line voltage is greater than 134 Volts when the Exposure Switch is pressed. Press the INCREASE () or DECREASE () buttons to reset the display. X-RAY GENERATION If HHH error repeats, check the AC line voltage at the input terminals to the X-ray Control Panel IS INHIBITED. The combination of low AC input Press the INCREASE () or line voltage and long selected DECREASE () buttons to reset the exposure time when the display. Exposure Switch is pressed Reduce the selected exposure time X-RAY GENERATION IS INHIBITED. such that the corrected time is less than 4.00 seconds or 240 pulses. The Exposure Switch was released prior to the completion of the programmed X-ray emission time. X-RAY GENERATION WAS PREMATURELY TERMINATED. 34 Press the INCREASE () or DECREASE () buttons to reset the display. NOTES 35 Air Techniques and ALLPRO Imaging are leading manufacturers of fine dental, medical and veterinary equipment from air and vacuum systems and X-ray film processors, to an impressive line of new products incorporating the most recent technological advances. These new products, vital components of the innovative professional practice, include intraoral cameras, digital imaging systems, which utilizes phosphor plate technology and, most recently, an intraoral digital X-ray system using sensor technology. Air Techniques and ALLPRO Imaging have been manufacturing quality products for the dental, medical and veterinary professional since 1962. Air Techniques and ALLPRO Imaging products are distributed only through authorized dealers. Refer to www.airtechniques.com or www.allproimaging.com to find a dealer in your area. K K K K K K K K K K K K AccentJ Intraoral Digital X-ray Image System Acclaim® Intraoral Digital Video Camera System Acclaim® USB Only Intraoral Digital Video Camera System AirStar® A/T 2000® XR GuardianJ Amalgam Collector 1-800-AIR-TECH Peri-Pro® (1-800-247-8324) www.airtechniques.com Provecta 70J Rinsendo Root Canal Disinfection System ScanX® STSJ VacStarJ K K K K K K K K K K K 100 Plus 2010 Medscope Provecta V ScanX® 12 ScanX® DVM ScanX® NDT ScanX® 12 EV ScanX® 14 Portable ScanX® NDT Portable ScanX® 14 In-Counter ALLPRO Imaging is a Division of Air Techniques © Copyright 2006 P/N A2507V Rev. C 1-800-AIR-TECH (1-800-247-8324) www.allproimaging.com ">
![](http://s3.manualzz.com/store/data/018679135_1-c98e5e1c3d36e41b478374bab13c1495-210x147.png)
Advertisement
Key features
- Microprocessor controlled
- Exposure compensation over input AC voltage range
- User selectable exposure time in seconds/pulses
- 30 preset exposure times
- 44 Manually selectable exposure times
- Programmable Size ratios
- Programmable Film setting
- Programmable Digital setting
- Two memorized exposure settings
Frequently asked questions
The PROVECTA V offers 30 preset exposure times and 44 manually selectable exposure times ranging from 0.02 to 3.20 seconds.
You can customize the default exposure time settings for tooth categories, patient sizes, and media types by following the instructions in the manual.
The cooling curve for the tubehead is provided in the manual. It shows the time it takes for the tubehead to cool down after an exposure.
The dimensions for the PROVECTA V are provided in the manual. It includes the dimensions for the X-ray Arm Support Connection Enclosure, X-ray Remote Control, and the Tubehead.