Thank You and Congratulations on your purchase of the ScanX Quantum Digital Imaging System, a rugged portable
imaging system manufactured by Air Techniques, Inc. for ALLPRO Imaging. Each system has been designed and manufactured
using state-of-the-art technology to produce the lightest, most rugged and dependable digital imaging system available for
the demanding conditions encountered in portable radiography applications.
This document provides abbreviated instructions to setup and operate the ScanX. Refer to the Operator’s Manual (P/N D5471)
included as a PDF file on the Drivers and Utilities Disk (P/N D5475) for complete instructions to ensure that your ScanX gives
the highest level of service.
Although each ScanX has been designed to minimize exposure of personnel to hazards, users must exercise every precaution to
ensure personnel safety, and be familiar with the warnings and cautions presented below and throughout the supplied Operator’s
Manual.
Use of Accessory Equipment.
The use of ACCESSORY equipment not complying with the equivalent safety requirements of this equipment may lead to a
reduced level of safety of the resulting system. Use of ACCESSORIES or cables other than those specified or provided by the
manufacturer may result in increased EMISSIONS or decreased IMMUNITY of the EQUIPMENT.
Electrical Safety Notes.
• The main power switch is the main power disconnect device.
• Use only the line cord provided with the unit.
• Use only grounded electrical connections.
• Check the device cables for possible damage before switching on. Damaged cables, plugs and sockets must be replaced
before use.
• Do not locate unit where it could be sprayed with water, or in a damp environment.
Do Not Attempt Internal Service.
The interior of each component of the ScanX is only accessible by removing hardware with tools and should only be opened
and serviced by an authorized dealer service technician. Contact your ALLPRO Imaging authorized dealer for service. Failure to
heed this directive may result in equipment damage and voiding the warranty.
Laser Class Warning.
The ScanX is a Class 1 Laser Product [Class 1 Laser Product (IEC 60825)]
The warning label shown on the right identifies the ScanX as such a product and describes the
potential danger to humans in the event the product is opened during service. There is no laser radiation
from this product when operated and maintained as instructed.
The Laser Product Accession Number is 0212282-00
CAUTION
CLASS 1 LASER PRODUCT
CLASS 3B LASER
RADIATION WHEN OPEN
AVOID EXPOSURE TO BEAM
DANGER
LASER RADIATION WHEN OPEN
AVOID DIRECT
EXPOSURE TO BEAM
COVER REMOVED MAKES
THIS DEVICE A
CLASS IIIb LASER PRODUCT
Customer-Supplied Computer and Imaging Software
To operate the ScanX, it must be connected to a compliant Computer System, and the computer must be loaded with an authorized
Imaging Software. See the Operator’s Manual for the minimum computer system requirements necessary to operate the ScanX.
Contact your dealer for available Computer System and Software options.
In addition to imaging software authorized for the ScanX, USB device drivers must also be installed on the computer in order
to operate the associated ScanX. The USB device drivers are provided on the Drivers and Utilities Disk (P/N D5475) included
with each unit. Also refer to the SYSTEM SETUP section of the Operator’s Manual included on the Drivers and Utilities Disk for
installation information for device drivers.
ScanX Quantum is a registered trademarks of ALLPRO Imaging.
© Copyright ALLPRO Imaging, 2009 • P/N D5471QB Rev. G - November 2013
BOX CONTENTS
Unpacking
As listed below, each ScanX System is shipped in a single carton
containing the ScanX main assembly and associated accessory
kit. Unpack each component of the ScanX and inspect for physical
damage such as scratched panels, damaged connectors, etc. If any
damage is noted, immediately notify your ALLPRO Imaging authorized
dealer immediately so corrective action can be taken. Save all cartons
and packing materials to protect the ScanX in the event of future transport
or shipping.
Included System Components
Each ScanX system consists of the indicated main assembly and
accessory kit as listed. Refer to the TECHNICAL DATA section of the
Operator’s Manual included on the Drivers and Utilities Disk for ratings and identification for specific ScanX models. Verify that all listed
items were received. If any item is missing, notify your dealer.
System Components - ScanX Quantum, P/N D5000-QB
Quick Start Instructions
Accessory Kit containing:
10-Foot Power Cord
6-Foot USB Cable
Dust Cover
Brush Kit
Fuse Kit
Size 0 Plate Guides, Pack of 4
Size 1 Plate Guides, Pack of 4
Size 2 Plate Guides, Pack of 4
Size 3 Plate Guides, Pack of 4
Size 4 Plate Guides, Pack of 4
Accessory Literature Kit:
Sample ScanX Cleaning Sheet Kit
CD Disk containing Drivers, Utilities and Operator's Manual
D5471QB
D5492B
61035
D5226
D5262
D5027
D5469
D5231
D5232
D5233
D5234
D5235
D5492-LIT
B2030
D5475
Important:Make sure to read and follow the Operator’s Manual (P/N D5471) included as a PDF file on the Drivers and Utilities
Disk (P/N D5475) to ensure that your ScanX gives the highest level of service.
SETUP & OPERATION
ScanX Drivers and Utilities Installation
Before connecting the ScanX to your computer or attempting to use it for the first time, run the Setup program on the ScanX
Drivers and Utilities Disk (P/N D5475) included with the ScanX. Normally, this program runs automatically when the CD is inserted into the drive for the first time. If not, run the Setup program located in the root directory of the CD (typically D:\AutoRun.exe).
Important: The fuse must be installed in order to operate the ScanX via the battery.
Battery Fuse Installation. Install the supplied 5A fuse as follows:
1. Insert the fuse cartridge into the fuse holder hole.
2. Secure the fuse by twisting the fuse cap 1/4 turn clockwise.
ScanX Connections. As shown by the illustration at right, connecting the
ScanX for operation consists of the following
1.Connect to the computer via supplied USB cable.
2.Connect to power source via supplied line cord.
Important: The built-in battery must always be charged prior to using for the first time.
Make sure the CHARGE LEVEL indicator illuminates green prior to use.
Battery Charging. Charge the battery as follows:
1.Place the Main Power switch to the ON (I) position and press the BATTERY
ON/OFF switch to the OFF position.
2.Allow the battery to charge for a full 3 hours before using for the first
time and on subsequent uses to reach the full operational power level.
3.Verify that the CHARGE LEVEL indicator illuminates green by removing
the AC power and depressing the BATTERY ON/OFF switch to the ON
position. If okay, turn the BATTERY ON/OFF switch to the OFF position.
Type B USB Connector
to ScanX
Type A USB
Connector to
Computer
Main Power Switch (I/0)
BATTERY CHARGE
IEC Connector
Note:
The Main Power Switch is the
AC Mains disconnect device.
Line Cord Plug
From AC Mains Outlet
IMPORTANT:
The fuse must be installed in order to operate the ScanX via the battery.
ScanX Operation. Operate the ScanX connected to a properly set up
computer as follows:
1.Perform one of the following steps depending on operating power desired.
a. When operating from the battery, depress the BATTERY switch and
verify the CHARGE LEVEL indicator shows an adequate power level.
b. When operating from AC Mains, place the Main Power switch to ON and
set the BATTERY ON/OFF switch to the ON position (flush with the
panel). Make sure the CHARGE LEVEL indicator illuminates green.
2.Press the READY switch to turn ON the ScanX.
3.Verify that the READY indicator illuminates.
4.Activate the scanner via the Imaging Software.
5.Verify that all track status indicators illuminate green.
6.Minimize exposure to ambient light and insert the PSP into the ScanX inlet
with the sensitive (front) side of PSP towards the ScanX.
7.Verify that all track status indicators illuminate yellow.
8.Observe that the scanned PSP exits through the scanner arch.
5A, 250VAC 3AG SLO-BLO Battery Fuse
BATTERY ON/OFF Switch
IMPORTANT:
Unless the ScanX is in use under battery power, the ON/OFF
(BATTERY) switch must always be in the OFF position to prevent
battery discharge.
CHARGE LEVEL Indicator
ERASER Switch
Eraser
Status
Indicator
Scanner
Track Status
Indicators
Corporate: 1295 Walt Whitman Road | Melville, New York 11747- 3062 | Phone: 800-247-8324 | Fax: 888-247-8481
West Coast: 291 Bonnie Lane, Suite 101 | Corona, CA 92880 - 2804 | Phone: 800-247-8324 | Fax: 951-898-7646
Website: a l l p ro i m a g i n g. c o m
READY
Switch
Ready/
Standby Status
Indicator
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertisement