University of Wisconsin - Madison REQUEST FOR BID

University of Wisconsin - Madison REQUEST FOR BID
University of Wisconsin - Madison
REQUEST FOR BID
THIS IS NOT AN ORDER
AGENT: Carla Parker
Questions regarding this bid – see Section 3.1
For Submittal of Bid Instructions – see Section 1
Bid prices and terms shall be firm for sixty (60) days from
the date of bid opening, unless otherwise specified in this
Request for Bid by the UW-Madison Purchasing Services.
OFFICIAL SEALED
BID NUMBER:
16-0004
ISSUE DATE:
7/17/15
DUE DATE:
7/22/15
Unless otherwise noted, public openings will be on the
date and time specified. Award(s), if any, will be made as
soon thereafter as possible.
2:00 PM CDT
The bid abstract will be on file and is open to public
inspection only after award determination has been made.
If NO BID (check here)
and return.
16-0004 / AMENDMENT #1 / OXYGEN CONTROLLED ANIMAL CHAMBER
Amendment #1 issued to answer questions received from bidders.
If this amendment is not returned, it shall be assumed your original bid meets all conditions of the amendment.
All other terms and conditions remain the same. Please revise and submit your bid accordingly.
In signing this bid, we have read and fully understand and agree to all terms, conditions and specifications and acknowledge that the UW-Madison Purchasing Services
bid document on file shall be the controlling document for any resulting contract. We certify that we have not, either directly or indirectly, entered into any contract or
participated in any collusion or otherwise taken any action in restraint of free competition; that no attempt has been made to induce any other person or firm to submit or
not to submit a bid; that this bid has been independently arrived at without collusion with any other bidder, competitor or potential competitor; that this bid has not been
knowingly disclosed prior to the opening of bids to any other bidder or competitor; that the stated statement is accurate under penalty of perjury. I certify that the
information I have provided in this bid is true and I understand that any false, misleading or missing information may disqualify the bid.
By submitting a bid, the bidder certifies that no relationship exists between the bidder and the University that interferes with fair competition or is a Conflict of Interest,
and no relationship exists between such bidder and another person or firm that constitutes a Conflict of Interest. Further, bidder certifies that no employee of the
University whose duties relate to this request for bid assisted the bidder in preparing the bid in any way other than in his or her official capacity and scope of
employment.
The Bidder certifies by submission of the bid that neither it nor its principals is presently debarred, suspended, proposed for debarment, declared ineligible, or voluntarily
excluded from participation in this transaction by any federal department or agency.
COMPANY NAME:
COMPANY STREET ADDRESS:
COMPANY CITY, STATE & ZIP:
SIGNATURE:
DATE :
TYPE OR PRINT NAME:
TITLE:
TELEPHONE NUMBER: (
)
FAX NUMBER:
EMAIL ADDRESS:
FEIN NUMBER:
DUNS #:
(
)
AMENDMENT #1
QUESTIONS AND ANSWERS FOR 16-0004
(OXYGEN CONTROLLED ANIMAL CHAMBER)
Q1:
Will bids be considered to the attached if they control and monitor CO2 from 1-30% as
opposed to the bid criteria of 0-30%?
A1:
No - we need it to have the ability to go to 0% and to be accurate to 0.1% increments. Some
experiments require CO2 below 1%
Q2:
Can we get more clarification of number 14 in the specifications? Do you need savable
profiles, is this cycling between 4 different O2 and CO2 levels? Can you clarify timing for
each gas as well? This specification indicates that they are trying to cycle between multiple
oxygen levels and the timing is adjustable to the users needs but certain limitations may
also apply for the volume of the chamber and flow rates.
A2:
I don’t need a profile be savable — but I do need a savable log of what the instrument “did”. It
should be able to set the level and have a log of what the level was over time - say for 20 min it was
x % CO2 and x % O2 etc. and then 20min later it was z% and b%.
We need to be able to go back and be sure with the log that the levels where what we had
programmed in so we can be sure the data we get is meaningful.
For timing — it does depend on chamber size and we defined the chamber size in our bid. For
what we need - being able to change within an hour is fine.
Q3:
In regards to programmable 4 different oxygen levels. Do they need to cycle CO2 with this
or is CO2 a static level?
A3:
BOTH need to be able to be changed at the same time — as we said — oxygen AND carbon
dioxide